Procedures
General anesthesia (100%FiO2)
Right femoral vein (RFV) and right femoral artery (RFA) were cannulated and
engaged with 6F and 5F sheath respectively
5F pigtail diagnostic catheter was introduced into RFA Desc. Ao guiding
Terumo wire Arcus PDA graphy measure PDA diameter
PDAgraphy : diameter PDA 2,2 mm
ampulla PDA 8 mm
Saturation of aorta was measured and intravenous 1000 IU heparin was given
6F MP2 diagnostic catheter was introduced into RFV IVC RA RV
MPA LPA measure PA pressure
6F MP2 diagnostic catheter was introduced to PDA and Desc. Ao guiding by
terumo wire pull out terumo wire change terumo wire with amplatzer wire
pull out 6F MPA2 and changed with 6F delivery sheath Lifetech introduced
to IVC RA RV PA PDA Ao Desc. Guiding Amplatzer Wire
Pull Out amplatzer wire and dilator delivery sheath
6/8 Lifetech Duct Ocluder was deployed using 6F delivery Lifetech
Post procedure angiography