3D Pan and Ceph X-Ray Calibration and Alignment


5/24/2019
Summary

These instructions serve as supplemental assembly guidance of the


Panoura. They are not meant to replace the installation manual,
but rather serve to complement it.

If there are any questions, please do not hesitate to contact us

914-592-6100

ImageWorks Proprietary – Confidential Information


2
Contents

5/24/2019
1. Computer setup
2. Tube head aging
3. 2D exposure field Adjustment
4. 2D Sensor Calibration Setting
5. 3D exposure field adjustment
6. 3D Digital calibration
7. Cephalometric exposure field adjustment and swing offset
adjustment
8. Cephalometric sensor calibration settings
9. Ceph 2nd slit adjustment
10. Ear rod adjustment
11. Save Adjustment results to computer
12. Final QA Images

ImageWorks Proprietary – Confidential Information


3
5/24/2019
1. Computer Setup
This procedure assumes PC was either provided or staged
by ImageWorks

Start PC
Launch eBUSPlayer64.exe (located in the Hamamtsu
Program Group).
Click Select/Connect Button

ImageWorks Proprietary – Confidential Information


4
1. Computer Setup

5/24/2019
On the "Device Selection" window,
select the "NTx-Mini PT01 -" item
displayed with the no entry symbol.
After making the selection, select the
"Show unreachable Network Devices"
check box, and click the "Set IP
Address..." button.

Under "GigE Vision Device


Configuration", enter 5 for the
location in the "IP Address" item
indicated below, and confirm that
the "IP Address" item is
192.168.16.5. After you confirm
this, click the "OK" button to close
the window.

ImageWorks Proprietary – Confidential Information


5
5/24/2019
1. Computer Setup
On the "Device Selection" window,
confirm that the no entry symbol
has been removed from the "NTx-
Mini PT01 -" item that had
displayed it, and then click the
"OK" button to close the window.

Click the "Device control" button.

Dental Fix
ImageWorks Proprietary – Confidential Information
6
5/24/2019
1. Computer Setup
The Device Control window will be
displayed. For the Visibility item,
select "Beginner". Next, configure
the items under the
"TransportLayerControl" menu as
shown in the following screenshot.
After you configure these settings,
click the "×" button at the top right
to close the window.

Select "Tools" > "Buffer


Options..." to open the Buffer
Options window.

Dental Fix
ImageWorks Proprietary – Confidential Information
7
1. Computer Setup

5/24/2019
On the Buffer Options window,
enter 128 for the "Buffers used for
streaming", and then click the
"OK“ button.

Select "Tool" again, and select


"Save Preferences" on the
displayed items. After making
the selection, if the
eBUSPlayer window is
displayed, click the "×" button

Dental Fix
to close it.

ImageWorks Proprietary – Confidential Information


8
2. Tube Head Aging

5/24/2019
Open the X-ERA
equipment adjustment
tool. enter password
p18p18 and press OK.

Select "Aging" in the X-ERA


Adjustment Tool.

ImageWorks Proprietary – Confidential Information


9
2. Tube Head Aging

5/24/2019
The aging window will be
launched. Click the "Start
aging" button to start aging.

The following window will be


displayed. Press the hand
switch for the equipment to
emit X-rays.

Continue until all 20 shots are exposed. Note X-ray cools


for 60 seconds between shots. Close the Aging window.

ImageWorks Proprietary – Confidential Information


10
3. 2D exposure field Adjustment

5/24/2019
Set the chin rest to the highest position, and remove the 3D-
compatible sliding chin rest block.

Launch the X-ERA Adjustment Tool and then click the


"Exposure field adjustment (2D)" button on the tool window to
launch the Exposure field adjustment (2D) window.

ImageWorks Proprietary – Confidential Information


11
3. 2D exposure field adjustment

5/24/2019
Click the "Select" button, select “Standard panoramic
exposure field “from the list, and then click "OK”.

ImageWorks Proprietary – Confidential Information


12
3. 2D exposure field Adjustment

5/24/2019
The adjustment window for the item
selected in (2) will be displayed. Check if
the correction values are displayed in the
collimator correction value input items. *
This section explains the procedure
assuming that "Standard panoramic
exposure field" was selected.

On the Exposure field adjustment (2D)


window, click the "Start AJAT Dental
Imaging Software" button to launch
AJAT, and then select Tools >
Maintenance > Mechanical Alignment
to open the Mechanical Alignment
window. (After AJAT Dental Imaging
Software has launched, the Exposure
field adjustment (2D) window will move
to the bottom right of the desktop.)

ImageWorks Proprietary – Confidential Information


13
3. 2D exposure field Adjustment

5/24/2019
Click the "Detect" button of the 2D
exposure field adjustment tool
window, and then detect the
collimator values by the procedure
as follows. (* It detects the rough
position of the exposure field.)

ImageWorks Proprietary – Confidential Information


14
3. 2D exposure field Adjustment

5/24/2019
Click the "Initial position" button of
the displayed "Detection" window,
and the collimator exposure port will
move to the initial position.

Click the "Exposure


(Horizontal)" button and wait
until the display of the clicked
button is changed to
"Preparation is completed".

ImageWorks Proprietary – Confidential Information


15
3. 2D exposure field Adjustment

5/24/2019
Perform X-ray exposure. During the
exposure, the X-ray exposure shutters of
the collimator moves to the left toward the
window. Release the exposure switch at
the time when the edge of the exposure
field has started to appear in the
Mechanical Alignment window.
(*) In case of selecting the Aluminum
block calibration, the X-ray exposure port
moves to the right toward the window.

Click the "Exposure (Vertical)" button


and wait until the display of the clicked
button is changed to "Preparation is
completed".

ImageWorks Proprietary – Confidential Information


16
3. 2D exposure field Adjustment

5/24/2019
Perform X-ray exposure. During the
exposure, the X-ray exposure port of the
collimator moves to upward. Release
the exposure switch at the time when
the edge of the exposure field has
started to appear in the Mechanical
Alignment window.

Click the close (×) button to return to


the exposure field adjustment
window. The dialog window as below
is displayed. Click the "OK“ button,
and the correction values of the
collimator that were generated
previously in this procedure are
displayed on the exposure field
adjustment window.

Click the "Apply" button.

ImageWorks Proprietary – Confidential Information


17
3. 2D exposure field Adjustment

5/24/2019
Click the "Take an exposure" button on the Exposure field
adjustment (2D) tool window to set the equipment to the
capture standby state, and then press the hand switch to
emit X-rays.

The area hit by the X-rays is displayed in white


on the green grid (thin lines) of the Mechanical
Alignment window. (The grid spacing is 10 mm
vertical and 1 mm horizontal. The size of the
green grid displayed on the window is 151.0 mm
× 6.4 mm.) Check if this white area fits in the
defined area in the below capture modes. If it
does not fit in the defined area, adjust the
collimator correction values with the Exposure
field adjustment (2D) tool so that the white area
fits in the defined area.

ImageWorks Proprietary – Confidential Information


18
3. 2D exposure field Adjustment

5/24/2019
Change each of the collimator correction values (top, bottom, left, and right) so that the white
area fits in the defined area, write the values to the scanner, and then Click “Take an
exposure” and initiate the scan with the Exposure Switch and check the exposure field.

ImageWorks Proprietary – Confidential Information


19
3. 2D exposure field Adjustment

5/24/2019
To expand the exposure field:
• Top: Enter a value larger than the current
correction value (+ direction).
• Bottom: Enter a value smaller than the
current correction value (- direction).
• Right: Enter a value smaller than the current
correction value (- direction).
• Left: Enter a value larger than the current
correction value (+ direction).

To shrink the exposure field:


• Top: Enter a value smaller than the current
correction value (- direction).
• Bottom: Enter a value larger than the current
correction value (+ direction).
• Right: Enter a value larger than the current
correction value (+ direction).
• Left: Enter a value smaller than the current
correction value (- direction).

ImageWorks Proprietary – Confidential Information


20
3. 2D Exposure field adjustment

5/24/2019
In the Exposure Field Adjustment Tool Select Bitewing panoramic exposure field click the
"OK" button.

Repeat alignment steps until Bitewing exposure field is OK as pictured.

ImageWorks Proprietary – Confidential Information


21
3. 2D Exposure field adjustment

5/24/2019
In the Exposure Field Adjustment Tool Select
Panoramic sensor aluminum block calibration click the
"OK" button.

Repeat alignment steps until


Aluminum filter exposure field is OK
as pictured.

ImageWorks Proprietary – Confidential Information


22
4. 2D Sensor Calibration Setting

5/24/2019
Launch the "Sensor calibration settings" tool.

Click the "Display sensor calibration image" button to launch AJAT Dental Imaging Software
and display the calibration window.

ImageWorks Proprietary – Confidential Information


23
4. 2D Sensor Calibration Setting

5/24/2019
In AJAT Dental Imaging Software, select Tools > Maintenance > Calibration to open the
calibration window, and then click the "Calibrate" button.

ImageWorks Proprietary – Confidential Information


24
4. 2D Sensor Calibration Setting

5/24/2019
Click the "Take an exposure" button, confirm that the window display has changed to "Please
take an exposure.", and then press the X-ray exposure switch to emit X-rays. X-ray exposure
is repeated a total of four times.

After the 4 exposures confirm that the signal light displayed to the left of the window has
changed to green

ImageWorks Proprietary – Confidential Information


25
5. 3D exposure field adjustment

5/24/2019
Launch the Exposure field
adjustment (3D) tool from X-ERA
Adjustment Tool.

Click the "Select" button to


select “C10900D CT scan
entire field” from the list.

ImageWorks Proprietary – Confidential Information


26
5. 3D exposure field adjustment

5/24/2019
Click the "Detect" button
of the 3D exposure field
adjustment tool window,
and then detect the
collimator values by the
procedure as follows

Click the "Take an


exposure" button and wait
until the display of the
clicked button is changed
to "Preparation is
completed

ImageWorks Proprietary – Confidential Information


27
5. 3D exposure field adjustment

5/24/2019
Click the close (×) button of the "Detection" window to return to the
exposure field adjustment window.

The dialog window as below is displayed. Click the "OK" button, and
the correction values of the collimator that were gotten at the
procedure 1) to 2) will be displayed on the exposure field adjustment
window.

Click the "Apply" button

ImageWorks Proprietary – Confidential Information


28
5. 3D exposure field adjustment

5/24/2019
Click the "Take an exposure" button, confirm that the window
display has changed from "Taking an exposure" to "Please take an
exposure.", and then press the hand switch to emit X-rays..

Use the Exposure field adjustment (3D) tool and roughly adjust the
exposure field. Check the captured image displayed in the center of
the tool window. The portions of the image hit by X-rays
are displayed in white, and the portions of the image not hit by X-rays
are displayed in black.

When yellow index lines are displayed:


Adjust the collimator positions (correction values) so that the
white section extends slightly out from the yellow index
lines.

When yellow index lines are not displayed:


Adjust the collimator positions (correction values) so that the
entire image is displayed in white.

ImageWorks Proprietary – Confidential Information


29
5. 3D exposure field adjustment

5/24/2019
ImageWorks Proprietary – Confidential Information
30
5. 3D exposure field adjustment

5/24/2019
Check the four corners of the exposure field and the pixel
value by the exposure field adjustment window

ImageWorks Proprietary – Confidential Information


31
5. 3D exposure field adjustment

5/24/2019
[How to adjust the exposure field]
If the pass criterion is not satisfied or one or more part of the four
corners are displayed in black, adjust the exposure field. The
adjustment method is to change the collimator correction values with
the Exposure field adjustment (3D) tool to expand or shrink the
left/right/top/bottom edges with the arrows on the adjustment tool.

ImageWorks Proprietary – Confidential Information


32
5. 3D exposure field adjustment

5/24/2019
Repeat field alignment for the “C10900D Partial Medium FOV 110
offset scan” field and the ”C10900D CT scan partial” field

After the corner adjustments are Green for all three FOV
close the Exposure field adjustment window.

ImageWorks Proprietary – Confidential Information


33
5/24/2019
6. 3D Digital Calibration

From the X-ERA Adjustment


Tool window, select "Digital
calibration" to display the
Digital calibration window

ImageWorks Proprietary – Confidential Information


34
5/24/2019
6. 3D Digital Calibration
Assembly the 3D Calibration Phantom

Install on the chin rest and set as shown

ImageWorks Proprietary – Confidential Information


35
5/24/2019
6. 3D Digital Calibration
Select the capture mode 3D medium FOV/High quality (center)

Click the "Start calibration" button to perform preparations for


imaging.

ImageWorks Proprietary – Confidential Information


36
5/24/2019
6. 3D Digital Calibration
When the window display has changed from "Preparing for exposure." to "Please take an
exposure.", press the X-ray exposure switch to capture the image. During imaging, the
window display will change to "Taking an exposure.".

After imaging, wait while "Saving RAW data" is displayed.

ImageWorks Proprietary – Confidential Information


37
5/24/2019
6. 3D Digital Calibration
After "Calibration complete" is
displayed, the calibration results
will be displayed on the right side
of the window.

The Center channel U (pixel) must be between 300 and 316.

The Center channel V (pixel) must be less than 608.

If the results are within tolerance then click the "Copy the result" button to apply
the center imaging calibration results to the equipment.

ImageWorks Proprietary – Confidential Information


38
5/24/2019
6. 3D Digital Calibration
Perform calibration for the offset image.

Change the capture mode to "Offset".

Assembly the 3D QA
phantom to the center
position of the phantom
base, and attach it to
the chin rest.

ImageWorks Proprietary – Confidential Information


39
5/24/2019
6. 3D Digital Calibration
Click the "Start calibration" button to
perform preparations for imaging.

When the window display has


changed from "Preparing for
exposure." to "Please take an
exposure.", press the X-ray exposure
switch to capture the image.
During imaging, the window display
will change to "Taking an exposure.".
After imaging has ended, "Image to
confirm center channel is being
constructed." will be displayed.

ImageWorks Proprietary – Confidential Information


40
5/24/2019
6. 3D Digital Calibration
17 images will be created automatically. It
will take several minutes to generate the
images. Open and view the images using
Photo Viewer.

Check the FOV center section of the


aluminum disc displayed in the
confirmation image, find the
image with no vertical offset (split), and
confirm the optimum center channel value
in the file name.

Enter the value of the


best image in the
Center Channel U box
and click “Copy the
result.

ImageWorks Proprietary – Confidential Information


41
5/24/2019
7. Cephalometric exposure field
adjustment and swing offset adjustment

• Repeat alignment for ceph.


• See Installation Manual Section 2-8 for details

Dental Fix
ImageWorks Proprietary – Confidential Information
42
5/24/2019
8. Cephalometric sensor calibration
settings

• Repeat Calibration for ceph.


• See Installation Manual Section 2-9 for details

Dental Fix
ImageWorks Proprietary – Confidential Information
43
9. Ceph 2nd Slit Adjustment

5/24/2019
Attach the second slit to the cephalometric unit.
Push on the side that has already been fastened
with set screws, and fasten it with 4 hexagon nuts
from the rear (M3×L6).

After the adjustments have been finished, gently


push the opposite side with the included set
screws, and protect the set screws that come out
with hexagon cap nuts

Launch the X-ERA Adjustment Tool, and then click


the "2nd slit adjustment" button on the tool window.

In the 2nd slit adjustment tool, click the "Take an


exposure" button, and capture an image of
"Center".

Dental Fix
ImageWorks Proprietary – Confidential Information
44
9. Ceph 2nd Slit Adjustment

5/24/2019
Check the center position exposure field, and
confirm that the exposure field fits in the
frame.

You can make large adjustments in the


horizontal direction with the belt/pulley.
Loosen the 2 hexagon socket head bolts,
and then adjust by shifting the
engagement of the belt and pulley

Dental Fix
You can make fine adjustments with the hollow set
screws attached to the left and right of the second slit.

ImageWorks Proprietary – Confidential Information


45
5/24/2019
9. Ceph 2nd Slit Adjustment

In the Adjust 2nd Slit tool, click the


"Take an exposure" button, and
capture an image of "Finish position",
"Center", and "Start position".
Check the exposure field from the start
position to the end position, and
confirm that the exposure field
fits in the frame.

If the exposure field at the finish


position is offset as shown in the
following screenshot, correct the
swing offset correction value.

Dental Fix
ImageWorks Proprietary – Confidential Information
46
5/24/2019
9. Ceph 2nd Slit Adjustment

Launch the Exposure field adjustment


tool from X-ERA Adjustment Tool
again, and select "Exposure field
adjustment (Ceph)".

Set the correction value so that the


exposure fields for "Center" and
"Finish position" become uniform
on the 2nd slit adjustment window.

Dental Fix
ImageWorks Proprietary – Confidential Information
47
5/24/2019
9. Ceph 2nd slit adjustment

Make fine adjustments so that the exposure fields become roughly uniform as
shown in Figure A. Figure B shows the exposure fields when the value is too
small.

Dental Fix
ImageWorks Proprietary – Confidential Information
48
5/24/2019
10. Ceph Ear rod adjustment
Turn the ear rods and nose
support position to the
cephalometric LA position,
and open up the ear rods
to their widest point.

Launch the X-ERA


Adjustment Tool, and click
the "Ear rod adjustment"
button.

Dental Fix
The Ear Rod Adjustment
window will open.

ImageWorks Proprietary – Confidential Information


49
5/24/2019
10. Ceph Ear rod adjustment
Click the "Take an exposure" button to open the
"Standing by for exposure" window.
Press the X-ray exposure switch on the main unit
to start X-ray emissions and image capturing
while the unit operates.

The captured image is


displayed on the left side of the
window.

Use mouse to zoom/pan.

Dental Fix
ImageWorks Proprietary – Confidential Information
50
5/24/2019
10. Ceph Ear rod adjustment
Confirm the center position of the
ear markers on the captured image.

The X-ray tube head side has the


larger ear marker.

See Installation manual for details regarding adjusting ear rods.

Dental Fix
ImageWorks Proprietary – Confidential Information
51
11. Save Adjustment results to computer

5/24/2019
After all adjustments have been finished, save the adjustment
results to the computer.

Launch the Equipment configuration setting


tool from the Adjustment Tool.

On the Equipment configuration


settings tool, click the "Read the value
to change from file" button to
read the adjustment results from the
equipment.

On the Equipment configuration settings tool, click the "Backup the current equipment value"
button to save the adjustment results from the equipment.

Dental Fix
* The adjustment results are saved in C:\YCINC\XP81\MaintenanceTools\ComponentSetting.
The adjustment results are saved to a file named ComponentSettingXXXXXXXX.Xml in the
above folder. (XXXXXXXX is the date.)

ImageWorks Proprietary – Confidential Information


52
12. Final Test Exposures

5/24/2019
Final QA Pictures
2D Pan Phantom set offset to 0 mm. Set 70kV, 2.0 mA and 12.0
seconds.

Requirement: All ball round and symmetrically positioned.

3D QA Phantom.
Large FOV (low dose) use 70 kV, 2.0 mA, 12.0 s
Small FOV (HQ) use 70 kV, 2.0 mA, 12.0 s

Requirements:
Distance between steel balls 29.6 mm to 30.4 mm
Disk Thickness Axial < 1.0 mm
Short Axis Pin 4.6 mm to 5.3 mm
Standard Deviation < 150

ImageWorks Proprietary – Confidential Information


53
5/24/2019
Summary

These instructions serve as supplemental assembly guidance of the


Panoura. They are not meant to replace the installation manual,
but rather serve to complement it.

If there are any questions, please do not hesitate to contact us

914-592-6100