B. I. M. S
BRINELL IMPRESSION
MEASUREMENT SYSTEM
1
CONTENTS
1. General
1.1. Preface . .03
1.2. Description . .03
2. Installation.
2.1. Installation of Video Capture. .04
2.2. B.I.M.S. Software Installation. .04
2.3. System Requirements. .04
3. Getting Started.
3.1. How to Start.? .05
3.2. For the First Time .06
3.3. Main Window. .09
3.4. New Main Window. .10
4. Toolbar and Menus.
4.1. Toolbar. .11
4.1.1. New Batch. .12
4.1.2. Open Batch. .13
4.1.3. View Indentation. .15
4.1.4. Manual. .16
4.1.5. Semi Auto. .18
4.1.6. Auto. .19
4.1.7. Remote Auto .19
4.1.8. Calibration. .22
4.1.9. Report. .23
4.1.10. Save Image .25
4.1.11. Load image .25
4.2. Menu bar.
4.2.1 File -> Video-Grab Size 26
4.2.2 Settings -> Printer Page .27
4.2.3 Settings -> Password Change. .28
4.2.4 Settings -> Video Settings .29
4.2.5 Settings -> COM Port .30
2
1. General
1.1 PREFACE :
The Hardness Tester software’ is user friendly windows based
software. Apart from its simplicity in operating it covers all technical
requirements for measurements of Brinell hardness at just one click.
1.2 DESCRIPTION :
The CCD camera and the software gives high precision for the
automated image analysis. So here the hardness report presents the
high standard of precision.
3
1. INSTALLATION
2.1 Installation of Video Capture :
• Start PC. If you have any antivirus software enabled, please disable it during
installation of the software. Before you first connect EASYCAP to your
computer, please insert the software CD-ROM into CD drive. Open folder
‘Driver WIN8_7_XP’ and run ‘setup.exe’ follow onscreen instructions to finish
installation. This completed Easycap USB video capture installation. When
prompted for restart, restart computer. Plug in USB cable of Video capture to
one of the USB ports of personal Computer. Windows will
detect and automatically install driver.
4
3 GETTING STARTED
3.1 How To Start
Click Start → Programs → B.I.M.S.
The following splash screen appears.
The above screen appears for few seconds and the main window of the
software is shown.
5
3.2 For the First Time
At first start a dialog is shown for Party/ Company Name , Address and
password for software. Put relevant data and click ‘DONE’. For next
start ‘Enter Password’ screen appears. User has to enter same
password. This is the software entry password and can be modified by
customer.
6
Next is Video source selection as shown:
Pull down ‘Select video source’ box and select ‘USB 2861 Device’.
Default Brightness and contract values are applied and shown
7
The property windows for ‘USB 2861 Device’ are shown below:
8
3.3 Main Window
After correct password the main window with live video display,
Menu and toolbar will appear. Use view to see the test area on the
component.
9
3.4 New Main Window
Now the main window shows the view port area with indentation.
10
4 Toolbar and Menus
4.1 Toolbar
The software includes user-friendly toolbar controls as follows: -
→ To view Report.
11
4.1.1 New Batch
This is for new batch entry. Enter batch details with hardness
conversion parameters(if required). One can select Name and Address
from Old data or can add new. To open a new batch click on “New
Batch” in File menu or New in toolbar.
12
4.1.2 Open Batch
This is for opening previous entered batches for viewing or
change of data. One needs valid “Data Entry Password” to change the
previous entered batch contents.
Click “Open Batch ’in File menu or click Open in toolbar. Select
appropriate batch by either company name or certificate No. Above
picture shows selection by certificate no.
13
Following shows selection by Company name.
14
4.1.3 View Indentation
Click `Indentation’ in View menu or View in toolbar to
view live view of image. The view appears as follows
15
4.1.4 Manual
This mode of measurement is to provide control to the operator and
user can mark the diagonals by clicking on the edges of the impression.
First click `manual’ in toolbar or Manual in Measure menu to select
manual mode of measurement. The image is freeze. Move mouse
pointer to outer edge of the impression and click left mouse button to
mark first point of the diameter. Move pointer to appropriately mark
second point of the diagonal and left click to mark the diagonal. Thus
one diagonal is marked as shown in figure
16
Click “Next Diameter” to mark second diameter or click `Done’ to
calculate the Brinell hardness number.
17
4.1.5 Semi Auto
Here we have to click in the black area of indentation desired for
measurement. Click Semi Auto in toolbar or in Measure menu. After
image is freeze click on the black portion inside the indentation. The
image finding analysis and is done automatically by the software and
the result window is shown as follows
18
4.1.6 Auto
Click `Auto’ in toolbar or in Measure menu to auto find indention
and give the result.
19
4.1.8 Calibration
This enables recalibration and calibration check of the system.
Select “Calibration of Objective” in Settings menu or calibration in
toolbar. First password window appears and after valid password entry
is done calibration window is as shown: -
20
OK to accept. Here red, green and blue markers can be moved as in
manual mode in measure.
Now one can check the calibration by clicking `Check’ and marking the
distance. Subsequently it shall be displayed in digits in mm as shown
on the next page.
21
Click CLOSE to close the window and return to main window.
22
4.1.8 Report
23
Select `Readings’ in select page to view table of all readings in batch (
shown above).
24
4.1.9 Save Image
We can save image from ‘VIEW’ in bitmap format. Give folder and file
name to save to a file. This saved image can be recalled and measured
in Semi Auto and Manual mode of measurement.
25
4.2. Menu bar
The software includes user friendly menu bar and includes
following additional menus
Small-Large indicates small video window and large freeze window for
measurement. The image size is adjusted as per screen size. Use this
menu when live video is breaking due to lower speed of PC.
Large-Large indicates large video window and large freeze window for
measurement. The video and image size is adjusted as per screen
size. Use this menu when live video is not breaking due to lower speed
of PC.
26
4.2.2 Settings -> Printer Page
Use this to set page setting for Left and Top margin in inches.
User can put either plus or minus figures up to one decimal place as
0.2 for increasing margin by 0.2 inches or -0.5 to reduce it by 0.5 inch.
Firstly window to input Left margin is shown as below
27
4.2.2 Settings ->Password change
This facility is provided to allow the user to change any password
as per requirement. The three options are provided as “Program”,
“Objective Calibration” and “Change Data”. Click “Program” in
Password Change of Settings menu to change password for the
software program.
Enter old and New passwords and click Done to accept the
change. If old password is correct then new password is implemented
and
Message box “Password is successfully changed” occurs.
Similarly user can change passwords for calibration by clicking
“Objectives Calibration” and data change in Open Batch window using
“Change data”.
28
4.2.3 Settings -> Video Setting
Click Video in settings menu. Here we can adjust Brightness and
contrast as per requirements. To change the setting move respective
slider and corresponding value is displayed in text box near to slider.
29
4.2.4 COM Port
This is to select RS232 port for remote auto measurement by unit.
The window is as shown.
User can either select or put any other no for COM port. Remember to
select a port that is free, otherwise “COM Port not available” is
displayed. Click ‘Select’ to select it and “COM Port selected” message
appears.
The selected port no is saved till further change.