10 3290@j Jad A42929 PDF

You might also like

You are on page 1of 9

Do Different Application Modes Improve the Bonding

Performance of Self-etching Ceramic Primer to Lithium


Disilicate and Feldspathic Ceramics?
Andres Felipe Millan Cardenasa / Angela Sisley Quintero-Calderonb / Fabiana Suelen Figuerêdo de
Siqueirac / Veridiana Silva Camposd / Michel Wendlingere / Camilo Andres Pulido-Moraf /
María José Masson-Palaciosg / Martha Lucia Sarmiento-Delgadoh / Alessandro D. Loguercioi

Purpose: To evaluate the effect of different application modes of a recently introduced self-etching ceramic primer
on the microshear bond strength (μSBS) and ceramic surface-etching pattern of two glass-ceramic surfaces.
Materials and Methods: Twenty-two CAD/CAM blocks of lithium disilicate (LD) and feldspathic glass ceramic (VTR)
were each cut into 4 rectangular sections (n = 88 for ceramic surface). The LD and VTR specimens were divided
into one control group (hydrofluoric acid + silane coupling agent [HF+SII]), and 10 experimental groups using Mono-
bond Etch and Prime (MEP) applied for a combination of scrubbing times (5, 10, 20, 40, and 60 s) and reaction
times (20 or 40 s). After each treatment, Tygon matrices (n = 8) were filled with a resin cement and light cured for
each ceramic specimen. The specimens were stored in water at 37°C for 24 h and subjected to the μSBS test. The
failure pattern and μSBS were statistically evaluated (α = 0.05). In addition, the ceramic surface etching pattern
was analyzed using scanning electron microscopy.
Results: For the LD groups, the application of MEP 60/40 resulted in a higher mean μSBS compared to HF+SI
(p < 0.05). Groups 5/40, 10/40, and 20/40 resulted in mean μSBS similar to that obtained by HF+SI (p > 0.05).
For VTR, no significant difference was observed among the groups (p = 0.32). Compared with MEP, HF better pro-
moted the dissolution of glass matrix for both ceramics. However, prolonged MEP scrubbing or reaction caused sig-
nificant dissolution of the glass matrix for both evaluated ceramics.
Conclusion: Active and prolonged application of MEP may be a viable alternative to HF+SI for increasing the bond
strength to LD.
Keywords: hydrofluoric acid, lithium disilicate, scanning electron microscopy, self-etching ceramic primer.

J Adhes Dent 2019; 21: 319–327. Submitted for publication: 23.08.18; accepted for publication: 14.04.19
doi: 10.3290/j.jad.a42929

M inimally invasive dentistry has been promoted and en-


abled by the use of indirect restorations constructed
with materials that have better mechanical properties.
Nowadays, different materials with esthetic properties and
clinical indications are used as indirect restorative materi-
als.16,21

a Professor, Postgraduate Program in Dentistry, CEUMA University, São Luis, f Professor, Department of Dentistry, National University of Colombia, Bogota,
MA, Brazil. Research idea, designed testing assembly, performed bond Colombia. Designed testing assembly and contributed to the discussion.
strength experiments, co-wrote paper. g Professor, Department of Dentistry, Equinoccial Technological University,
b MSc Student, Department of Dentistry, National University of Colombia, Bo- Quito, Ecuador. Contributed substantially to discussion.
gota, Colombia. Performed microshearbond strength experiments, co-wrote
h Professor, Department of Dentistry, National University of Colombia, Bogota,
paper.
Colombia. Provided consulting for statistical analysis, co-wrote paper.
c Professor, Postgraduate Program in Dentistry, CEUMA University, São Luis,
i Professor, Department of Restorative Dentistry, State University of Ponta
MA, Brazil. Research idea, designed testing assembly, performed bond
strength experiments, co-wrote paper. Grossa, Ponta Grossa, PR, Brazil. Provided consulting for statistical analysis,
contributed substantially to the discussion, proofread the manuscript.
d PhD Student, Department of Restorative Dentistry, State University of Ponta
Grossa, Ponta Grossa, PR, Brazil. Research idea, designed testing assembly,
Correspondence: Prof. Dr. Andres Felipe Millan Cardenas, Department of Post-
co-wrote paper.
graduate Program in Dentistry, Universidade CEUMA, Rua José Montello 1,
e Undergraduate Student, Department of Dentistry, CEUMA University, São Luis, Jardim Renascença, São Luís, Maranhão, Brazil. 65075120. Tel: +55-98-3214-
MA, Brazil. Performed scanning electron microscopy, co-wrote paper. 4164; e-mail: andresfelipemillancardenas@hotmail.com

Vol 21, No 4, 2019 319


Cardenas et al

Control group:
Lithium disilicate HF + Silane (HF + SI)
glass ceramic
(EMX, IPS e.max
CAD)
MEP 5/20

MEP 10/20

MEP 20/20 Microshear bond strength


(μSBS)
MEP 40/20

MEP 60/20
Monobond Etch & Ceramic surface etching
Prime (MEP) MEP 5/40 pattern
Feldspathic glass
ceramic MEP 10/40
(VTR, Vitablocs
RealLife) MEP 20/40

MEP 40/40

MEP 60/40

Fig 1 Flow chart of the experimental design.

Among these materials, polycrystalline ceramics are con- the glass-matrix ceramic restoration, was recently intro-
sidered a “glass-free matrix” due to the absence of a glass duced on the market. As per manufacturer’s recommenda-
phase; they provide strength and fracture toughness but tion, MEP should be scrubbed on the ceramic surface for
with limited translucency.3,5 Due to the nature of polycrys- 20 s and then left to react for 40 s. Regarding this proto-
talline ceramics, etching with hydrofluoric acid (HF) is diffi- col, several studies recently concluded that the bond
cult, and consequently, creating reliable, durable bonding is strength to lithium disilicate when using MEP was similar to
challenging.33 For this reason, different surface treatments that achieved by the gold standard protocol (HF +
have been suggested to increase the bond strength of resin SI);1,13,35,39 however, the etching pattern produced by MEP
luting cements to polycrystalline ceramics.33 was inferior to that produced by HF.1,13,29,35,39 Neverthe-
On the other hand, the esthetics and optical properties of less, to the best of our knowledge, different application pro-
glass-matrix ceramics such as lithium disilicate and feldspathic tocols of this new self-etching ceramic primer have not yet
glass ceramics are superior to those of glass-free matrices.16 been evaluated, and it is known that different application
However, to achieve successful bonding during cementation of modes influence the bond strength to lithium disilicate.15,20
glass-matrix ceramics, prior conditioning of the intaglio surface In addition, according to the manufacturer, MEP is used for
with HF and a silane coupling agent (SI) is a standard proce- conditioning glass-matrix ceramic restorations; however, to
dure for adhesive cementation,5 and are essential steps that our best knowledge, only a few studies have evaluated the
will influence the success of the restoration.9,44 conditioning of MEP on different ceramics.13,29,35,42
Etching with HF can selectively dissolve the glass-matrix Thus, the aim of the present study was to evaluate the mi-
ceramic and generate multiple roughness patterns on the croshear bond strength (μSBS) and etching pattern of MEP
ceramic surface, which promotes microretention.8,18,22 After when applied using different application modes to two different
HF, a silane is applied. Silane is a bifunctional molecule that glass-matrix ceramics, and compare them with the perfor-
induces a chemical interaction between the silica in the mance of HF + SI. The null hypotheses were as follows: the
glass phase of the ceramic and the methacrylate groups of different MEP application protocols have no influence on 1) the
the resin monomers through siloxane bonds,14,30,41 thereby microshear bond strength and 2) the etching pattern for the
increasing the bonding strength.10,14,41 However, these ap- two glass ceramics, as compared to those yielded by HF + SI.
plication steps increase the chairside time, and conse-
quently, the risk of mistakes during the bonding procedures.
To simplify this conditioning procedure, a self-etching ce-
ramic primer, Monobond Etch & Prime (MEP),19 which si-
multaneously etches and silanizes the intaglio surface of

320 The Journal of Adhesive Dentistry


Cardenas et al

Table 1 Materials used, composition, application mode, and batch number

Material (manufacturer) Composition Application mode* Batch number

Condac porcelana 5%, (FGM Hydrofluoric acid, water, thickener, Apply for 20 s, rinse with water for 30 s, 110615
Prod Odonto; Joinville, SC, surfactant, coloring clean ultrasonically with distilled water
Brazil) for 180 s

Vita Ceramics 5% Etch (Vita Sulfuric acid, hydrofluoric acid, ethanol Apply for 60 s, rinse with water for 30 s, 42530
Zahnfabrik; Bad Säckingen, clean ultrasonically with distilled water
Germany) for 180 s

Monobond Plus (Ivoclar Ethanol, 3-trimethoxysilylpropyl Apply with a brush and allow to react for 1401001210
Vivadent; Schaan, methacrylate, methacrylated phosphoric 60 s, disperse excess with a strong
Liechtenstein) acid ester (10-MDP), disulfide acrylate stream of air to ensure solvent
evaporation*

Monobond Etch & Prime (Ivoclar Butanol, tetrabutylammonium Apply onto the luting surface with a U10661
Vivadent) dihydrogen trifluoride, methacrylated microbrush and rub for 20 s, leave on
phosphoric acid ester, trimethoxypropyl on the surface for a further 40 s, rinse
methacrylate monomer thoroughly, dry for 10 s

Variolink II (Ivoclar Vivadent) Urethane dimethacrylate, inorganic Mix at a 1:1 ratio on a mixing pad, pack U50748
fillers, ytterbium trifluoride, initiators, resin cement carefully inside each tube,
stabilizers, pigments light cure for 20 s at 1200 mW/cm2

*According to manufacturers’ recommendations.

MATERIAL AND METHODS with water for 30 s, ultrasonic cleaning in water for 180 s,
and silane application for 60 s, and Monobond Etch and
Specimen Preparation Prime (MEP; Ivoclar Vivadent) was applied with scrubbing
Two CAD/CAM materials were selected: 1) lithium disilicate for 5, 10, 20, 40, and 60 s and left to react for 20 or 40 s
(EMX, IPS e.max CAD, Ivoclar Vivadent; Schaan, Liechten- (Fig 1). The composition, application mode, batch number,
stein) and 2) feldspathic glass ceramic (VTR, Vitablocs Re- and experimental groups are described in Tables 1 and 2.
alLife, VITA Zahnfabrik; Bad Säckingen, Germany).
Twenty-two CAD/CAM blocks of each material were used. Microshear Bond Strength Test
For each material, the blocks (12 x 12 x 6 mm) were cut The specimens (n = 66) were mounted on a polyvinyl chloride
into 4 rectangular sections (6 x 6 x 6 mm) (n = 88 per ce- (PVC) ring filled with acrylic resin (AutoClear, DentBras; Piras-
ramic surface) using a diamond disk at slow speed (Isomet, sununga, SP, Brazil), where the specimen surface was ex-
Buehler; Lake Bluff, IL, USA) under water cooling. After posed on the top of the cylinder with a height of 3 mm.
cleaning ultrasonically with distilled water for 15 min, the The specimens were etched according to the different
EMX specimens were fired in a furnace (Programat P300, experimental protocols (Fig 1) and thoroughly rinsed with
Ivoclar Vivadent) following the crystallization program at water spray. Monobond P or MEP (Ivoclar Vivadent) was ap-
840–850ºC for 20–31 min. plied according to the respective manufacturer’s instruc-
tions and experimental groups (Table 1). A single operator
Experimental Design performed all the bonding procedures. Then, eight transpar-
For each indirect material, the specimens were randomly ent polyethylene Tygon tubes (Tygon Medical Tubing Formu-
assigned (http://www.sealedenvelope.com) into 11 groups lations 54-HL, Saint Gobain Performance Plastics; Akron,
(n = 8), with 66 specimens used for the μSBS test and 22 OH, USA) with an internal diameter of 0.8 mm and a height
specimens used to evaluate the ceramic surface-etching of 0.5 mm were positioned over each ceramic specimen.
pattern according to the following independent variables. In The resin cement Variolink II (Ivoclar Vivadent) (Table 1)
the control group (HF+SI) 5% HF (FGM; Joinville, SC, Brazil) was carefully packed inside each tube, and a clear Mylar
+ silane coupling agent (Monobond P, Ivoclar Vivadent) matrix strip was placed over the filled Tygon tube and
were applied according to the manufacturer’s recommenda- pressed gently into place. The luting composite specimens
tions for each ceramic surface. For EMX, HF 5% was applied were simultaneously light cured for 20 s using an LED light-
for 20 s, followed by rinsing with water for 30 s, ultrasonic curing unit set at 1200 mW/cm2 (Radii-cal, SDI; Bayswater,
cleaning in water for 180 s and silane application for 60 s. Victoria, Australia) in close contact with the Mylar strip. A
For VTR, HF 5% was applied for 60 s, followed by rinsing radiometer (Demetron LED Radiometer, Kerr Sybron Dental

Vol 21, No 4, 2019 321


Cardenas et al

Table 2 Surface ceramic and application mode of experimental groups

Ceramic Etching Experimental Application mode Resin


surface agent groups cement
Lithium Hydrofluoric 1. Control group: 1. 5% hydrofluoric acid etching for 20 s Variolink II*
disilicate acid (HF) HF + silane (HF + 2. Rinse with water for 30 s
glass ceramic SI) 3. Clean ultrasonically with distilled water for 180 s
(EMX, IPS 4. Apply silane solution and Monobond Plus* with a brush and allow to react for 60 s.
e.max CAD)* Subsequently, disperse the excess with a strong stream of air to ensure solvent
evaporation*

Monobond 1. MEP scrubbing for 5 s, let react on the surface for 20 s Rinse thoroughly, dry Variolink II*
Etch & Prime 2. Group 5/20 2. MEP scrubbing for 10 s, let react on the surface for 20 s for 10 s
(MEP) 3. Group 10/20 3. MEP scrubbing for 20 s, let react on the surface for 20 s
4. Group 20/20 4. MEP scrubbing for 40 s, let react on the surface for 20 s
5. Group 40/20 5. MEP scrubbing for 60 s, let react on the surface for 20 s
6. Group 60/20 6. MEP scrubbing for 5 s, let react on the surface for 40 s
7. Group 5/40 7. MEP scrubbing for 10 s, let react on the surface for 40 s
8. Group 10/40 8. MEP scrubbing for 20 s, let react on the surface for 40 s*
9. Group 20/40 9. MEP scrubbing for 40 s, let react on the surface for 40 s
10. Group 40/40 10. MEP scrubbing for 60 s and left to reacted on the
11. Group 60/40 surface for 40 s

Feldspathic Hydrofluoric 1.Control group: 1. 5% hydrofluoric acid etching for 60 s VariolinkII*


glass ceramic acid (HF) HF + Silane (HF + 2. Rinse with water for 30 s
(VTR, SI) 3. Clean ultrasonically with distilled water for 180 s
Vitablocs 4. Apply silane solution and Monobond Plus* with a brush and allow to react for 60 s.
RealLife)** Subsequently, disperse the excess with a strong stream of air to ensure solvent
evaporation*

Monobond 1. MEP scrubbing for 5 s, let react on the surface for 20 s Rinse thoroughly, Variolink® II*
Etch & Prime 2. Group 5/20 2. MEP scrubbing for 10 s, let react on the surface for 20 s dry for 10 s
(MEP)*** 3. Group 10/20 3. MEP scrubbing for 20 s, let react on the surface for 20 s
4. Group 20/20 4. MEP scrubbing for 40 s, let react on the surface for 20 s
5. Group 40/20 5. MEP scrubbing for 60 s, let react on the surface for 20 s
6. Group 60/20 6. MEP scrubbing for 5 s, let react on the surface for 40 s
7. Group 5/40 7. MEP scrubbing for 10 s, let react on the surface for 40 s
8. Group 10/40 8. MEP scrubbing for 20 s, let react on the surface for 40 s*
9. Group 20/40 9. MEP scrubbing for 40 s, let react on the surface for 40 s
10. Group 40/40 10. MEP scrubbing for 60 s and left to reacted on the
11. Group 60/40 surface for 40 s

* According to the manufacturer’s (Ivoclar Vivadent) recommendations; ** Vita Zahnfabrik; *** the manufacturer of MEP did not have a specific application
protocol for Vita Blocs.

Specialties; Middleton, WI, USA) was used to check the The μSBS (in MPa) was calculated by dividing the load at
light intensity every 8 specimens. These procedures were failure by the surface area (mm2) of each specimen. After
carried out under magnifying loupes.27 testing, the specimens were examined under an optical mi-
After storing the specimens in distilled water for 24 h at croscope (SZH-131, Olympus; Tokyo, Japan) at 100X mag-
37°C, the Tygon tubes were carefully removed with a blade, nification to identify the location of bond failure. The failure
exposing the cement cylinders. Each specimen was exam- mode was classified as cohesive in resin cement (CR), that
ined under a stereomicroscope at 10X magnification. The is, failure exclusively within the resin cement; cohesive in
cement cylinder was discarded if pores or gaps were ob- ceramic (CC), ie, failure exclusively within the ceramic; ad-
served at the interface. The specimens were attached to a hesive (A), ie, failure at the resin cement-ceramic interface;
shear-testing fixture (Odeme Biotechnology; Joaçaba, SC, and mixed (M), that is, failure included cohesive failure of
Brazil) and tested in a universal testing machine (Kratos the neighboring substrates.
IKCL 3-USB, Kratos Equipamentos Industriais; Cotia, SP,
Brazil). Each specimen was positioned in the universal test- Ceramic Surface-etching Pattern
ing machine and a thin orthodontic wire (0.2 mm diameter) To etch surfaces, the specimens (n = 22) of each ceramic
was looped around the base of each composite cylinder. were surface conditioned with HF or MEP according to the
The orthodontic wire contacted the composite cement cylin- experimental conditions previously described (Table 1).
der on half of its circumference. The setup was kept aligned Then, the specimens were rinsed and dried. Furthermore,
(cement-ceramic interface, the wire loop, and the center of two additional surfaces of each ceramic (6 x 6 x 6 mm) were
the load cell) to ensure correct orientation of the shear added to evaluate the non-conditioned ceramic surface. All
forces.37 The crosshead speed was set at 1 mm/min until the specimens were dried and dehydrated in a desiccator
failure. for 12 h, and the conditioned ceramic surfaces were sputter

322 The Journal of Adhesive Dentistry


Cardenas et al

Table 3 Number of specimens (%) according to fracture mode for all experimental groups

EMX VTR

Experimental groups A M CR CC A M CR CC
Hydrofluoric acid + silane 43 (90) 5 (10) 0 (0) 0 (0) 31 (68) 7 (15) 2 (4) 6 (13)
MEP 5/20 45 (94) 1 (2) 2 (4) 0 (0) 40 (84) 2 (4) 0 (0) 6 (12)
10/20 46 (98) 1 (2) 0 (0) 0 (0) 43 (90) 2 (4) 0 (0) 3 (6)
20/20 45 (94) 3 (6) 0 (0) 0 (0) 44 (94) 1 (2) 0 (0) 2 (4)
40/20 41 (85) 7 (5) 0 (0) 0 (0) 29 (60) 3 (7) 2 (4) 14 (29)
60/20 41 (85) 6 (11) 2 (4) 0 (0) 31 (65) 5 (10) 0 (0) 12 (25)
5/40 48 (100) 0 (0) 0 (0) 0 (0) 41 (84) 3 (6) 2 (4) 3 (6)
10/40 48 (100) 0 (0) 0 (0) 0 (0) 39 (81) 2 (2) 1 (2) 6 (13)
20/40 40 (84) 5 (10) 3 (6) 0 (0) 28 (59) 6 (13) 6 (13) 7 (15)
40/40 44 (92) 4 (8) 0 (0) 0 (0) 34 (71) 6 (12) 0 (0) 8 (17)
60/40 41 (85) 7 (15) 0 (0) 0 (0) 27 (57) 7 (15) 1 (2) 12 (26)
EMX: lithium disilicate; VTR: feldspathic glass ceramic. Fracture modes: A: adhesive; m: mixed; CR: cohesive in resin cement; CC: cohesive in ceramic.

Table 4 Mean (± SD) microshear bond strengths in MPa for all experimental groups

Experimental groups EMX VTR


Hydrofluoric acid + silane 31.17 ± 1.0 B 27.62 ± 1.5 a
MEP 5/20 14.43 ± 0.3 D 27.61 ± 1.3 a
(scrubbing time/reaction time)
10/20 20.00 ± 1.5 C 27.31 ± 2.5 a
20/20 25.07 ± 0.8 C 28.15 ± 1.9 a
40/20 32.40 ± 1.8 AB 27.70 ± 1.6 a
60/20 33.57 ± 1.1 AB 29.10 ± 1.2 a
5/40 32.03 ± 1.1 B 27.90 ± 1.0 a
10/40 31.50 ± 0.6 B 27.09 ± 1.4 a
20/40 33.27 ± 1.2 AB 27.78 ± 1.4 a
40/40 31.90 ± 1.7 AB 28.05 ± 1.3 a
60/40 35.30 ± 1.6 A 28.49 ± 0.3 a
Different capital (EMX) or lowercase letters (VTR) in each column indicate statistically significant differences for each material (one-way ANOVA, Tukey’s test,
p < 0.05)

coated with gold-palladium in a vacuum evaporator (SCD tion, after which Barlett’s test for equality of variances was
050, Balzers; Schaan, Liechtenstein). The entire treated sur- performed to determine if the assumption of equal vari-
face was examined under a scanning electron microscope ances was valid.26 After confirming the normality of data
(SEM, MIRA3 LM, Tescan Orsay Holding; Warrendale, PA, distribution and the equality of the variances, the μSBS (in
USA), and three photomicrographs of representative surface MPa) data were subjected to appropriate statistical analy-
areas were taken at different magnifications. sis. The mean μSBS of all the specimens from the same
ceramic surface were averaged for statistical purposes.
Statistical Analysis One-way ANOVA was used to analyze the μSBS data for
The data were first analyzed using the Kolmogorov-Smirnov each glass ceramic material. Tukey’s post-hoc test was ap-
test to assess whether the data followed a normal distribu- plied with α = 0.05.

Vol 21, No 4, 2019 323


Cardenas et al

a b c d e f

g h i j k l

Fig 2 Representative SEM images of EMX surface (25,000X): the untreated EMX surface does not show vitreous matrix dissoution (a).
EMX etched with 5% hydrofluoric acid for 20 s shows significant vitreous matrix dissolution and exposed crystals (arrow). Different etching
patterns observed for different MEP application modes (g). Poor vitreous matrix dissolution and exposed crystals can be observed in (b)
MEP-5/20, (c) MEP-10/20, (d) MEP-20/20, (h) MEP-5/40, and (i) 10/40. The increase in MEP application time leads to slightly greater
vitreous matrix dissolution with an increase in surface roughness, porosity (white hands), and exposed crystals (e, f, k, and l) when compared
to the manufacturer’s protocol (j).

RESULTS characterized by an increase in porosity and roughness and


exposure of vitreous matrix dissolution for both ceramics.
Microshear Bond Strength MEP caused partial dissolution of the vitreous matrix,
For each ceramic surface, the majority of specimens showed showing fewer pores and less pronounced roughness
adhesive failures, as shown in Table 3. For the lithium disili- (Figs 2 and 3). However, a progressive effect of the differ-
cate ceramic, there are significant differences among the ent MEP application protocols was manifest as a greater
groups (ANOVA test; p = 0.001; Table 4). The application of dissolution of the glass matrix of the ceramics with the for-
MEP 60/40 resulted in a higher mean μSBS than did the mation of larger, deeper craters and pits with increasing
application of HF + SI (Tukey’s test; p < 0.05; Table 4). The scrubbing and reaction times (Figs 2 and 3), up to a total
mean μSBS of groups 5/40, 10/40, and 20/40 were simi- application time of 60 s.
lar to that of the HF + SI control group (Tukey’s test; p >
0.05; Table 4). On the other hand, groups 5/20, 10/20,
and 20/20 showed lower mean μSBS than that obtained by DISCUSSION
the HF + SI control group (Tukey’s test; p < 0.05; Table 4).
For the feldspathic ceramic, there were no significant dif- The results of the present study showed that the conven-
ferences among the groups (ANOVA; p = 0.32; Table 4). tional application of 5% HF followed by silane coupling
That is, no significant differences were observed in the agent achieved a well-defined conditioning pattern and a
mean μSBS when MEP was applied regardless of protocol, high μSBS to both of the evaluated glass-matrix ceramics.
as compared to the HF + SI control group (ANOVA; p = 0.32; This could be attributed to two factors: 1. the application of
Table 4). HF resulted in some major pores with severe irregularities
on the intaglio ceramic surface,8,18,22 as observed in the
Ceramic Surface-etching Pattern present study (Figs 2 and 3); 2. the use of a silane coupling
The representative SEM images of the ceramic surface- agent created a chemical bond between the ceramic and
etching patterns for different experimental groups are the resin monomers,14,30,41 which increased the wettability
shown in Figs 2 and 3. The untreated lithium disilicate and of the resin cement.5
feldspathic ceramics showed a smooth surface without any However, due to the toxicity and volatility of HF, its use
craters or pits. HF promoted the dissolution of glass matrix, has been questioned, especially given the potential health

324 The Journal of Adhesive Dentistry


Cardenas et al

a b c d e f

g h i j k l

Fig 3 Representative SEM images of VTR surface (3000X): untreated VTR shows no pores or surface roughness (a). Dissolution of the
vitreous matrix with an aggressive etched pattern and deeper pores (white hands) can be observed on the VTR surface after etching
with hydrofluoric acid for 60 s (g). Progressive vitreous matrix dissolution can be observed for different MEP application protocols (b, c, d, h,
and i). A more aggressively etched pattern with major pores (white hands) and matrix dissolution can be observed with increasing MEP
application time (e, f, j, k, and l).

hazard upon contacting unprotected skin.4,32 This is a seri- ester (10-MDP), which is responsible for increasing the po-
ous problem in intraoral ceramic repair.31 One alternative tential for chemical interaction.11,19 In addition, MEP con-
evaluated in the present study is provided by the new self- tains trimethoxypropyl methacrylate (methacrylate silane),
etching ceramic primer. According to the manufacturer, MEP which leaves a thin, chemically bonded layer of silane on
is less toxic and poses less health risk.19 the surface even after washing with water and drying.19 It is
Unfortunately, although MEP caused a partial dissolution worth mentioning that silane coupling agents play an impor-
of the glass-ceramic matrix,19 it yielded a less-pronounced tant role in promoting bifunctional adhesion on a ceramic
etching pattern when to both the ceramic surfaces com- surface.24,25 The presence of both molecules could be re-
pared to HF (Figs 2 and 3). This could be attributed to the sponsible for the lower contact angle with the glass ceramic
milder acidic nature of MEP, which has been previously re- surface when MEP was applied.35,42
ported1,13,29,35,39 and could be responsible for the lower Further, it is noteworthy that some questions have been
bond strength (groups 5/20, 10/20, and 20/20) to lithium raised regarding the stability of silane molecules in acidic
disilicate ceramic at shorter application times. environments.20,48 Continuous hydrolyzation of silanol
However, the results of the present study showed that groups occurs during long-term storage in a low-pH solution.
active, prolonged application of MEP might compensate for Therefore, although the manufacturer of MEP claims that
the milder etching potential of ammonium polyfluoride pres- the silane coupling agent is stable during storage in the
ent in MEP, which leads to the formation of several irregu- presence of ammonium polyfluoride, this hypothesis needs
larities and pores due to the dissolution of the glass matrix to be evaluated in the near future, since the condensation
(Figs 2 and 3). Moreover, although a longer scrubbing time reaction of the silanol groups is suppressed and the forma-
along with a longer reaction time of MEP did not yield the tion of a highly reactive monomeric silanol is continually
same conditioning pattern as that of HF, but bond strengths promoted at a steady rate during storage.19 Based on these
similar to that of HF + SI were obtained when the MEP ap- considerations, the bonding mechanism of MEP is expected
plication time was at least 60 s (scrubbing + reaction time) to be related to the interaction between the functional phos-
on the lithium disilicate ceramic surface. phoric monomers and ceramic ions rather than the methac-
The exact action mechanism of MEP is not entirely rylate silane bonding to the glass ceramic.
clear.13,35,42 According to the manufacturer, MEP contains Therefore, when MEP was applied for 60 s (reaction
functional monomers such as methacrylate phosphoric acid time) instead of 20 s, the interaction of the acidic resinous

Vol 21, No 4, 2019 325


Cardenas et al

monomers (methacrylated phosphoric acid ester) with the It is worth mentioning that the occurrence of cohesive
ceramic surface improved, similar to that on enamel, failure could not be attributed to the load application de-
thereby increasing not only the potential of etching, but also vice. Orthodontic wire was used in this study instead of a
monomer impregnation into the underlying surface.34,43 chisel, because the former distributes stress more evenly
Furthermore, 60 s of scrubbing application can not only around the adhesive interface area6,12,45 as it has a large
eliminate water, alcohol, and other products of the silane contact area compared to a chisel.6,28 In contrast, a chisel
reaction, but also help complete the condensation reaction promotes puncture loading, resulting in a smaller contact
between silane and silica, promoting the formation of silox- area and causing severe stress concentration at the load
ane.36 Thus, evaporation of alcohol can increase the den- application area, thus inducing a larger number of cohesive
sity of local connections available for the solution of silane failures.6,28 This is the reason that several recently pub-
to react with the ceramic.36 Therefore, we hypothesized lished papers used orthodontic wire instead of a chisel for
that the prolonged, active application increased the poten- evaluating the μSBS of adhesive interfaces.7,22,23,28,35,40
tial of etching associated with solvent evaporation and Thus, the results of the present study lead to partial re-
helped promote the formation of siloxane groups, thus im- jection of the first null hypothesis and complete rejection of
proving the μSBS for the lithium disilicate ceramic surface. the second null hypothesis. However, further studies are
Figure 2 clearly shows the positive effect of active and required to clarify the effect of different MEP application
prolonged MEP application on the surface topography. How- protocols on the pretreatment of other glass-matrix ceram-
ever, the μSBS to the lithium disilicate surface did not show ics, mainly focusing on the chemical interaction with the
an increasing trend, mainly after at least 20 s of scrubbing surface. Finally, although MEP showed good results in the
(minimum 40 s of reaction; 20/40 and 40/40) or at least present study as in previous studies, future randomized
40 s of scrubbing associated with a minimum 20 s of reac- clinical studies are required to evaluate MEP as a condi-
tion (40/20 and 60/20). It is known that the higher surface tioner of glass ceramics.
roughness is not directly related to increased bond
strength.47,49 However, the presence of methacrylate phos-
phoric acid ester and methacrylate silane could be respon- CONCLUSION
sible for the higher potential for chemical interaction11,19
and lower contact angle with the glass ceramic surface,35,42 Although the active and prolonged application of MEP re-
which would make the surface changes less necessary. sulted in progressive glass matrix dissolution for both the
This can also explain the absence of significant differ- evaluated glass-matrix ceramics, none of the cases showed
ences in the μSBS with even a slight increase in the etching an etching pattern similar to that obtained with traditional
pattern upon MEP application to the feldspathic ceramic. HF treatment. MEP provided results similar to those ob-
The results are in agreement with those of Venturi et al,46 tained with HF+SI, and active, prolonged application of MEP
who reported that increasing the etching pattern using a improved the bond strength for lithium disilicate compared
higher concentration of HF did not result in any difference in to HF+SI. The scrubbing and application times of MEP were
the bond strength for a feldspathic ceramic. It is known that material dependent.
feldspathic ceramics have a higher amount of amorphous
glass matrix,17,38 and due to the micromechanical nature of
the adhesion between the ceramic surface and resin ce- REFERENCES
ment, any microtopographical change plays an important 1. Alrahlah A, Awad MM, Vohra F, Al-Mudahi A, Al jeaidi ZA, Elsharawy M. Ef-
role in its chemical adhesion with silane. Thus, we hypoth- fect of self etching ceramic primer and universal adhesive on bond
strength of lithium disilicate ceramic. J Adhes Sci Technol 2017;31:
esized that irrespective of the active and prolonged applica- 2611–2619.
tion, the microtopographical alterations on the feldspathic 2. Apel E, Deubener J, Bernard A, Holand M, Muller R, Kappert H, Rhein-
ceramic surface created by MEP were sufficient to produce berger V, Holand W. Phenomena and mechanisms of crack propagation
in glass-ceramics. J Mech Behav Biomed Mater 2008;1:313–325.
bond strength simliar to that induced by HF + SI.
3. Bachhav VC, Aras MA. Zirconia-based fixed partial dentures: a clinical re-
Additionally, the results revealed a strong association view. Quintessence Int 2011;42:173–182.
between the failure mode, surface treatment, and type of 4. Bertolini JC. Hydrofluoric acid: a review of toxicity. J Emerg Med 1992;10:
material. The highest percentage of adhesive failure was 163–168.
5. Blatz MB, Sadan A, Kern M. Resin-ceramic bonding: a review of the litera-
observed for shorter application times (groups 5/20, ture. J Prosthet Dent 2003;89:268–274.
10/20, 20/20, 5/40, and 10/40) for the lithium disilicate 6. Braga RR, Meira JB, Boaro LC, Xavier TA. Adhesion to tooth structure: a
ceramic surface, while a high percentage of cohesive failure critical review of “macro” test methods. Dent Mater 2010;26:e38–49.
was observed in feldsphatic glass ceramics independent of 7. Cardenas AM, Siqueira F, Hass V, Malaquias P, Gutierrez MF, Reis A,
Perdigao J, Loguercio A. Effect of MDP-containing silane and adhesive
surface treatment. These findings could be attributed to the used alone or in combination on the long-term bond strength and chemi-
difference in the microstructure and the lower intrinsic cal interaction with lithium disilicate ceramics. J Adhes Dent 2017;19:
203–212.
strength (brittleness) of the feldspathic ceramic. The addi-
8. Chen JH, Matsumura H, Atsuta M. Effect of different etching periods on
tion of long lithium disilicate crystals results in greater flex- the bond strength of a composite resin to a machinable porcelain. J Dent
ural strength and fracture resistance than feldspathic ce- 1998;26:53–58.
ramic, due to the spatial configuration of their crystalline 9. Conrad HJ, Seong WJ, Pesun IJ. Current ceramic materials and systems
with clinical recommendations: a systematic review. J Prosthet Dent
grid and resistance to crack propagation through crystals.2 2007;98:389–404.

326 The Journal of Adhesive Dentistry


Cardenas et al

10. de Carvalho RF, Martins ME, de Queiroz JR, Leite FP, Ozcan M. Influence of 32. Özcan M, Allahbeickaraghi A, Dundar M. Possible hazardous effects of
silane heat treatment on bond strength of resin cement to a feldspathic hydrofluoric acid and recommendations for treatment approach: a review.
ceramic. Dent Mater J 2011;30:392–397. Clin Oral Investig 2012;16:15–23.
11. de Souza G, Hennig D, Aggarwal A, Tam LE. The use of MDP-based mate- 33. Özcan M, Bernasconi M. Adhesion to zirconia used for dental restorations:
rials for bonding to zirconia. J Prosthet Dent 2014;112:895–902. a systematic review and meta-analysis. J Adhes Dent 2015;17:7–26.
12. DeHoff PH, Anusavice KJ, Wang Z. Three-dimensional finite element anal- 34. Perdigao J, Gomes G, Lopes MM. Influence of conditioning time on
ysis of the shear bond test. Dent Mater 1995;11:126–131. enamel adhesion. Quintessence Int 2006;37:35–41.
13. El-Damanhoury HM, Gaintantzopoulou MD. Self-etching ceramic primer 35. Prado M, Prochnow C, Marchionatti AME, Baldissara P, Valandro LF,
versus hydrofluoric acid etching: Etching efficacy and bonding perfor- Wandscher VF. Ceramic surface treatment with a single-component
mance. J Prosthodont Res 2018;62:75–83. primer: resin adhesion to glass ceramics. J Adhes Dent 2018;20:99–105.
14. Filho AM, Vieira LC, Araujo E, Monteiro Junior S. Effect of different ceramic 36. Shen C, Oh WS, Williams JR. Effect of post-silanization drying on the bond
surface treatments on resin microtensile bond strength. J Prosthodont strength of composite to ceramic. J Prosthet Dent 2004;91:453–458.
2004;13:28–35. 37. Shimada Y, Yamaguchi S, Tagami J. Micro-shear bond strength of dual-
15. Giraldo TC, Villada VR, Castillo MP, Gomes OM, Bittencourt BF, Dominguez cured resin cement to glass ceramics. Dent Mater 2002;18:380–388.
JA. Active and passive application of the phosphoric acid on the bond 38. Sinthuprasirt P, van Noort R, Moorehead R, Pollington S. Evaluation of a
strength of lithium disilicate. Braz Dent J 2016;27:90–94. novel multiple phase veneering ceramic. Dent Mater 2015;31:443–452.
16. Gracis S, Thompson VP, Ferencz JL, Silva NR, Bonfante EA. A new classi- 39. Siqueira FS, Alessi RS, Cardenas AF, Kose C, Souza Pinto SC, Bandeca
fication system for all-ceramic and ceramic-like restorative materials. Int J MC, Loguercio AD, Gomes JC. New single-bottle ceramic primer: 6-month
Prosthodont 2015;28:227–235. case report and laboratory performance. J Contemp Dent Pract 2016;17:
17. Griggs JA. Recent advances in materials for all-ceramic restorations. Dent 1033–1039.
Clin North Am 2007;51:713–727, viii. 40. Siqueira FSF, Cardenas AM, Ocampo JB, Hass V, Bandeca MC, Gomes
18. Guarda GB, Correr AB, Goncalves LS, Costa AR, Borges GA, Sinhoreti JC, Reis A, Loguercio AD. Bonding performance of universal adhesives to
MA, Correr-Sobrinho L. Effects of surface treatments, thermocycling, and eroded dentin. J Adhes Dent 2018;20:121–132.
cyclic loading on the bond strength of a resin cement bonded to a lithium 41. Söderholm KJ, Shang SW. Molecular orientation of silane at the surface
disilicate glass ceramic. Oper Dent 2013;38:208–217. of colloidal silica. J Dent Res 1993;72:1050–1054.
19. Ivoclar Vivadent. Monobond Etch & Prime. 2015, 2017. 42. Tribst J, Anami LC, Ozcan M, Bottino MA, Melo RM, Saavedra G. Self-
20. Kalavacharla VK, Lawson NC, Ramp LC, Burgess JO. Influence of etching etching primers vs acid conditioning: Impact on bond strength between
protocol and silane treatment with a universal adhesive on lithium disili- ceramics and resin cement. Oper Dent 2018;43:372–379.
cate bond strength. Oper Dent 2015;40:372–378. 43. Tsuchiya H, Tsubota K, Iwasa M, Ando S, Miyazaki M, Platt JA. Influence
21. Kelly JR, Benetti P. Ceramic materials in dentistry: Historical evolution of adhesive application time on enamel bond strength of single-step self-
and current practice. Aust Dent J 2011;56 Suppl 1:84–96. etch adhesive systems. Oper Dent 2010;35:77–83.
22. Lise DP, Perdigao J, Van Ende A, Zidan O, Lopes GC. Microshear bond 44. van den Breemer CR, Gresnigt MM, Cune MS. Cementation of glass-ce-
strength of resin cements to lithium disilicate substrates as a function of ramic posterior restorations: A systematic review. Biomed Res Int
surface preparation. Oper Dent 2015;40:524–532. 2015;2015:148954.
23. Lopes GC, Perdigao J, Baptista D, Ballarin A. Does a self-etching ceramic 45. Van Meerbeek B, Peumans M, Poitevin A, Mine A, Van Ende A, Neves A,
primer improve bonding to lithium disilicate ceramics? Bond strengths De Munck J. Relationship between bond-strength tests and clinical out-
and FESEM analyses. Oper Dent 2018. comes. Dent Mater 2010;26:e100–121.
24. Lung CY, Matinlinna JP. Aspects of silane coupling agents and surface 46. Venturini AB, Prochnow C, Rambo D, Gundel A, Valandro LF. Effect of hy-
conditioning in dentistry: an overview. Dent Mater 2012;28:467–477. drofluoric acid concentration on resin adhesion to a feldspathic ceramic.
25. Matinlinna JP, Vallittu PK. Bonding of resin composites to etchable ce- J Adhes Dent 2015;17:313–320.
ramic surfaces - an insight review of the chemical aspects on surface 47. Xiaoping L, Dongfeng R, Silikas N. Effect of etching time and resin bond
conditioning. J Oral Rehabil 2007;34:622–630. on the flexural strength of IPS e.max Press glass ceramic. Dent Mater
26. Montgomery DC. Design and analysis of experiments, ed 5. New York: 2014;30:e330–336.
John Wiley & Sons, 1991. 48. Yoshihara K, Nagaoka N, Sonoda A, Maruo Y, Makita Y, Okihara T, Irie M,
27. Munoz MA, Baggio R, Emilio Mendes YB, Gomes GM, Luque-Martinez I, Yoshida Y, Van Meerbeek B. Effectiveness and stability of silane coupling
Loguercio AD, Reis A. The effect of the loading method and cross-head agent incorporated in ‘universal’ adhesives. Dent Mater 2016;32:
speed on resin-dentin microshear bond strength. Int J Adhes Adhes 1218–1225.
2014;50:136–141. 49. Zogheib LV, Bona AD, Kimpara ET, McCabe JF. Effect of hydrofluoric acid
28. Muñoz MA, Baggio R, Emilio Mendes YB, Gomes GM, Luque-Martinez I, etching duration on the roughness and flexural strength of a lithium disili-
Loguercio AD, Reis A. The effect of the loading method and cross-head cate-based glass ceramic. Braz Dent J 2011;22:45–50.
speed on resin–dentin microshear bond strength. Int J Adhes Adhes
2014;50:136–141.
29. Murillo-Gomez F, Palma-Dibb RG, De Goes MF. Effect of acid etching on
tridimensional microstructure of etchable CAD/CAM materials. Dent
Mater 2018;34:944–955. Clinical relevance: MEP provided results similar to
30. Naves LZ, Soares CJ, Moraes RR, Goncalves LS, Sinhoreti MA, Correr-So-
brinho L. Surface/interface morphology and bond strength to glass ce-
those of HF+SI. Active, prolonged application of MEP
ramic etched for different periods. Oper Dent 2010;35:420–427. may be a viable alternative for increasing the bond
31. Özcan M. Evaluation of alternative intra-oral repair techniques for fractured strength only for lithium disilicate.
ceramic-fused-to-metal restorations. J Oral Rehabil 2003;30:194–203.

Vol 21, No 4, 2019 327

You might also like