Professional Documents
Culture Documents
PSI 300
Particle Size Analyzer
Operating Manual
Operating Manual
Code 10000004257e Revision A Page 2 (37)
Contents
6. DEFAULT PARAMETERS...........................................................................................................................35
Assays..........................................................................................................................................35
Display settings ............................................................................................................................35
Analyzer system...........................................................................................................................36
Sample measurement ..................................................................................................................36
Sampler control ............................................................................................................................37
Optional devices...........................................................................................................................37
DCS Parameters ..........................................................................................................................37
Operating Manual
Code 10000004257e Revision A Page 3 (37)
They are,
PSI 300 Standby switch; when switched to STANDBY the analyzer stops
the measurement operation, flushes the flow tank and goes to a safe
standby state, and when switched to NORMAL the analyzer starts the
measurement sequence.
Calibration sample button, to cut a calibration sample out of the next
sample line to be analyzed
Zero Check button to request a zero check. Zero check maintains
accuracy of the PSI 300 particle size measurement.
Alarm status lamp to indicate that there is an active alarm in the system.
In addition the analyzer system can have a remote Maintenance switch in the
pneumatic control valve boxes of the primary samplers.
Local display
The local touch screen display functions as a user interface for various tasks
within the analyzer system.
Operating Manual
Code 10000004257e Revision A Page 4 (37)
For most parameter inputs pressing the component opens a keypad dialog
window that will work in the same way as the keyboard.
a When starting the analyzer, make sure that there are no other personnel
working on any part of the analyzer.
If PSI 300 includes density meter, refer to density meter operating manual.
Density meter requires field configuration on first startup after
installation.
If you are starting the analyzer for the first time or need to change the
parameters, turn the Standby switch to STANDBY. See chapter 1.9
Options displays how to change analyzer parameters.
8. The PSI 300 software is loaded, this will take for a while.
9. The main operating window is shown on the local display of PSI 300.
10. The analyzer starts with the Initialization sequence. Check that pneumatic
valves are working as show on the Status display, see detailed information
later on this section. For detailed information of display messages or
troubleshooting, see section 8 in this manual, or refer to the PSI 300
Maintenance Manual.
11. The next phase is the Zero Check sequence. If the zero check is not
acceptable an alarm is set and the reason of the alarm needs to be
corrected.
12. Finally the Auto Measurement is started.
The start up procedure is now complete and the analyzer has started measuring
the sample.
Now the analyzer has been shut down properly and it is easy to restart.
Initialization
Zero Check
Auto Measurement
Zero Check
Standby or
Flushing
The analyzer starts through the Initialization state. Next follows the Zero
check with additional primary sampler (when in automatic control) flushes. In
zero check, 40-120 measurements are performed during water flush. When the
standard deviation of the measurements falls below the limit (default 5 um), the
calculated mean value of this 40 measurements is set as a new zero level. If the
mean value exceeds the alarm limits, zero check alarm is activated and the
Operating Manual
Code 10000004257e Revision A Page 8 (37)
value is not accepted. Fraction calculation is then normalized with the new
granted zero level to eliminate the wear of the measurement elements.
After this system passes to the Auto Measurement. The analyzer starts to take
secondary sample cuts and making particle size assays according to the
defined sample sequence. Sample measurements consists of 120 individual
measurements which are collected into the measuring file. After the
measurement assays are calculated based on these measurements updated on
the local display and outputs. Also the calibration sampling (if was active) is
completed. Next the next line in sequence will be selected for measurement.
An alarm condition, the standby switch or a remote standby command from the
automation system brings the analyzer to the Standby state. Flow tank and
measurement probe are drained and flushed and the primary samples (when in
automatic control) are flushed and stopped. The Tank’s drain valve is left open.
When all these operations are completed the Going Standby message turns to
Standby message. When the system is on Standby the user can start Zero
Check by pressing Zero Check button or flush the tank by pressing Flush
button on the Status display.
After the standby switch is set to NORMAL or all current alarms are
acknowledged, the analyzer continues from Standby to Initialization state.
At any point of time, switching the standby switch on the Probe Control Set
(PCS) to STANDBY causes the analyzer to go to the Standby state. The flow
tank’s drain valve is left open and all water valves are closed. Any automatically
controlled primary samplers are flushed and closed. Now the analyzer is safe
for maintenance operations.
The flow tank and measurement probe can by flushed in Standby state by
pressing Flush button on the Status display, to make an additional manual
flush.
a When using the manual valve commands, make sure that there are no
other personnel working on these parts of the analyzer.
When the Standby switch for a primary sampler is switched to STANDBY all
pneumatic controls stop operation immediately. The water and sample line
Operating Manual
Code 10000004257e Revision A Page 9 (37)
valves close and NLA’s cleaning knife is drawn into the cutter. The primary
sampler is safe for maintenance operations. Note that the pipelines may contain
pressurized water and process sample.
The analyzer keeps running normally, except that the primary sample line is set
to BYPASS.
Zero Check can also be activated by pressing Zero Check button when the
system is in Standby state. In Standby state, it is also possible to activate a
manual tank flushing by pressing a Flush button on the Status display.
There are several sub-screens and pop-ups under the four main screens. The
structure is as follows,
Main
– Last assay values
Trend
– Three 2 hours trend displays
– Three 8 hours trend displays
– Current particle size distribution histogram graph and filtered particle
size distribution histogram graph (for each sample line) .
Status
– Alarms
– Sample line status
– Zero level history and 70 weeks zero level history.
– Calibration results
Options
– Assay options
– Display settings options
– Analyzer system options
– Sample measurement options
– Primary sampler cleaning options
– Advanced options
– Instruments
– DCS parameters
– System test
Navigation bar clock time is displayed in 24 hour format and day format
depends on panel regional settings (default: US).
Last assay values display shows current values of the maximum 40 assays.
Operating Manual
Code 10000004257e Revision A Page 11 (37)
For more information of the status, alarm, and warning messages, see section
5.
Last assay values display shows current values of the maximum 40 assays.
Operating Manual
Code 10000004257e Revision A Page 12 (37)
arrow buttons to rescale the y-axis or reset the scale back to original
settings.
link button to the Distribution display..
The distribution displays show the current (last measured) particle size
distribution. (There is one display for each sample line). The filtered distribution
display show the current filtered particle size distribution.
When a sample line is switched to bypass it will not be measured, that is, it is
skipped in the sample sequence.
If the sample line is in AUTOMATIC mode switching the line to bypass will
request a primary sampler cleaning and the sample flow is stopped. When the
bypassed sample line is switched back on, the primary sampler is flushed and
the flow is started again.
In an alarm situation, the system brings the analyzer to Going Standby –state
and drains and flushes the flow tank. The alarm lamp flashes on the front panel
of the Probe Control Set, and the alarm messages are displayed on the Main,
Status and Alarms displays. The measurement does not continue until the
reason for the alarm is removed.
If the instrument air / water pressure is low the analyzer does not continue
functioning until sufficient air / water pressure is reached.
value is removed from the left border of the chain so that together 70 values are
shown.
There is also an average weekly zero check history display. The display
includes history from last 70 weeks. This display can be accessed from Zero
level history display.
To enter to the parameter displays press the Options button. This opens first the
Password pop-up window. Enter the three digits password number and press
ok. The password is not requested if it is disabled on the Advanced Options
display.
Operating Manual
Code 10000004257e Revision A Page 18 (37)
Note: Make sure that you remember the password after changing it on the
parameters display.
Each options display has three buttons on the right hand side,
– saves the parameter values so that they are kept after a
shutdown.
The save and load operations concern all parameters at the same time. After
any parameter change a save is required for the changes to remain in effect in
an analyzer restart.
Display settings
The Display Settings display contains the setup of the numerical assay values
shown on the Main display and the variable selection and settings for the Trend
display.
Calibration Sampling
Cuts – no. of cuts for a single calibration sample
Cut Duration – cut duration in calibration sample cutting
Operating Manual
Code 10000004257e Revision A Page 21 (37)
Composite Sampling
Cuts – no. of cuts for a single composite sample
Interval – time interval between composite sample cuts (zero means that
no composite sample is performed)
Zero Check
Interval – time interval between zero checks (zero means that automatic
zero check is not performed)
Zero Level High Limit – high limit for mean value of measurements during
zero check (default 1100 um).
Zero Level Low Limit – low limit for mean value of measurements during
zero check (default 100 um).
St. Dev. Limit in Zero Check – high limit for standard deviation of
measurements during zero check (default 5 um).
Tank Cleaning
Slurry Draining Time – slurry draining time of the flow stabilization tank
before flushing
Extra Flushing Time Before ZC – extra water flushing time of the flow
stabilization tank before zero check measurement starts
Flushing Time Between Samples – water flushing time of the flow
stabilization tank between measurements of sample lines
Water Draining Time – water draining time of the flow stabilization tank
after flushing
Sample Prepare Time – delay ensures a stable and representative process
sample flow for fraction measurements
Flushing Interval (Single Line Mode) – when only one inlet is activated, flow
stabilization tank is flushed at regular intervals, not after every fraction
measurement
Based on the level switch signal, the slurry is dosed to the flow tank. The slurry
feeding adapts automatically to the availability of the sample flow from the
primary sampler. Sample dosing is based on short pulses in order to keep the
level at the level switch. System increases or decreases the rate of short pulses
automatically based on the primary sample flow rate. After measurement used
pulse duration is saved on the Pulse Duration parameter value.
The level switch can be simulated with simulation parameter. When the
simulation is active, PSI 300 ignores level switch operation and continues
measurement. This is useful if PSI 300 is needed to run without slurry flow.
Startup delay – time that tank is filled after level switch is already on
(overfilling in the first pulse)
Pulse duration – filling pulse duration, default 200 ms, changes
automatically according to the primary sample flow rate
Time Btw Pulses SP – set point for the time between pulses
Pulse Delay – minimum time between pulses, waits the level switch
Max Tank Fill Time – maximum tank filling time
Simulation fill time – time for tank filling before going to simulation mode
(must be less than max tank fill time).
Simulation active – When active, PSI 300 goes to simulation mode after
simulation fill time.
Probe Flow – sample flow through the measurement probe
The setup display contains also the time parameters of the automatic
sequences. For the sampler control function, see section 2.
Exit - stop the controller software application (functions only when the
system is on maintenance)
Instruments display
Instruments display can be used on PSI 300 installation or maintenance. It
includes raw input signal from measurement probe. the display includes two
kinds of values: values that are valid when probe motor is on and values that
are valid when the maintenance measurement is active.
Following values are shown only when the maintenance measurement is on,
Last value – measurement value without zero level.
Last value (tot.) – Measurement value with zero level.
When the analyzer goes to Standby all automatically controlled sample lines
setup to AUTOMATIC CONTROL are stopped, and when the system completes
Initialization they are started again. The same applies to any individual sample
line that is switched to BYPASS or ON by the user or through the plant
automation system.
On stopping any of the sample lines, the automatic sampler control makes the
following operations:
upstream flush
sample valve close
upstream flush
downstream flush
A periodic primary sampler flush can be set up for each sample line. The flush
sequence is a combination of the stop and start sequences.
Operating Manual
Code 10000004257e Revision A Page 27 (37)
Calibration samples are cut in a similar way to the measured samples, using the
cutters on the opposite side of the secondary sampler. The number of the cuts
can be adjusted by the parameters at Analyzer System display to get a suitable
sample size. Other parameters related to calibration sampling on the Analyzer
System display are cut duration and cylinder delay which adjust movement and
speed of the cylinder.
When collecting the calibration samples, it is essential that the samples are not
mixed with each other. It is also important that the samples are not
contaminated in any way. The whole sample collection and handling procedure
and the sample marking method should be clear to all users to avoid mistakes.
Sample buckets
Figure 23. Placing the sample bucket under the daily /calibration sampler.
When collecting the calibration samples, make sure that the process conditions
are suitable, that there are no alarms or warnings indicated and the sampling
equipment functions normally. Check also that the samples flow normally, the
sample bucket is clean and marked with the sample line name, and that you
have a pen available for writing the sample ID tags on the buckets.
It is important that both the analyzer and the related sampling equipment
functions normally during the sample collection.
9. See the sample identification stamp on the Main or Status display. Write
the ID stamp on the side of the bucket.
10. Flush the calibration sampler and sample cutters with water.
11. See calibration results from calibration results display in status display.
Calibration results
When the calibration sampling finishes, PSI will update calibration results in
calibration results display:
The panel includes external compact flash storage. A calibration log is saved to
flash storage. The calibration log includes maximum 199 calibration samples.
After 199 calibration sample, the log is emptied and Calibration ID starts from 1.
The calibration log can be copied to external USB removable disk from buttons
in calibration results display and status display.
The calibration log can also be accessed from web server if the PSI 300 is
connected to plant network.
Operating Manual
Code 10000004257e Revision A Page 30 (37)
The web server can be accessed through Internet explorer by typing the PSI
300 IP address to browser address bar. This shows a web interface for PSI 300.
Calibration log can be found from “Custom documents -> Calibration”.
When PSI 300 calibration has already been done, these assay values can be
used for verification of the PSI 300 analysis results against the laboratory
analysis.
Calibration parameters
Calibration coefficients (A0, A1, A2 and A3) are obtained from the regression
analysis done with the calibration data. They are set separately for each particle
size by four calibration parameter buttons on the Assay Configuration display.
4. DAILY MAINTENANCE
The PSI 300 analyzer is designed to operate independently and requires little
daily maintenance. However, you should conduct general check-ups regularly.
Table 1. Maintenance of the PSI 300 analyzer.
Operation Interval
Operation of the secondary Daily
sampler and the primary sample
flow
Condition of the calibration and Every time the sample buckets are
composite sample cutter collected for laboratory analysis
Check Local Display’s cleaning Weekly
requirement
If the analyzer does not control the primary sampler automatically, make
sure that the primary sampler is checked and cleaned regularly to
maintain the best possible performance.
Operating Manual
Code 10000004257e Revision A Page 32 (37)
Zero Check Green Zero Check measurement maintains accuracy of the PSI 300.
It follows 1) after turning power on, 2) after turning standby switch OFF, 3) after
Zero Check button is pressed, 4) after certain time interval that is set in
Analyzer System display.
Auto Measurement Green Measures the process samples in the sequence order. Updates the assay
variables after every measurement.
Going Standby Green The analyzer is going to Standby. Flow tank is drained and flushed before
analyzer is in Standby state.
Also when an alarm occurs, tank is drained and flushed and is going to
Standby. Analyzer tries restart after 10 minutes time interval or when Reset
alarms button at Status display is pressed.
Standby Green Before Standby tank is drained and flushed at Going Standby state.
When system is in Standby state, the drain valve is open and other valves are
closed.
Standby – Zero Check Green Manual Zero Check in Standby state. Zero Check is activated by pressing Zero
Check button.
Single Line Green PSI 300 operates in single sample line mode (two other sample lines are
bypassed)
Flushing *) Green Tank is flushed, water valves and quick drain valve are open.
Tank Draining *) Green Tank is drained, slurry draining before flushing or water draining after flushing.
Quick drain valve is open.
Zero Check Request Yellow Zero Check button is pressed. The analyzer is waiting for the end of
measurement cycle to start Zero Check.
New Zero Check Measurement *) Green Zero Check measurement failed. After three trials an alarm is activated.
Calibration Sample Request for Line Yellow The calibration sample button is pressed. The analyzer is waiting for next
# measurement cycle to start the sampling.
Taking Calibration Sample for Line # Green The analyzer is taking calibration sample cuts.
Standby Request Yellow The Standby switch is turned to ON. The analyzer is going to Standby.
Mech. Cleaning Primary Sample Line Yellow / Primary sampler’s mechanical cleaning (knife cuts). Mechanical cleaning is
# blinking done before measurement for the sample line that that will be measured next.
Full Cleaning Primary Sample Line # Yellow / The primary sampler full cleaning is running. Full cleaning is done with Zero
blinking Check. Can also to be activated time based.
Primary Sampler in Maintenance # Yellow The primary sampler is safe for maintenance operations. The analyzer
continues running and is waiting for the primary sample.
Alarm Blockage Flow nozzle at tank bottom is blocked or Check flow nozzle and quick drain
Quick drain duct is blocked or Check tank level switch
Tank level switch is broken Clean flow nozzle and quick drain duct
Operating Manual
Code 10000004257e Revision A Page 35 (37)
6. DEFAULT PARAMETERS
These are factory default parameters for PSI 300 analyzer. Parameters are
classified according to Options display in PSI 300 user interface.
Assays
Parameter Value Unit Other
Line 1 label Line 1 -
Filtering const. 1 -
Display settings
Parameter Value Unit Other
Main display: visible No - Display objects 1-6
Analyzer system
Parameter Value Unit Other
Calibration cuts 15 - For lines 1-3
Sample measurement
Parameter Value Unit Other
Startup delay 500 ms For lines 1-3
Sampler control
Parameter Value Unit Other
Primary sampler 0 min For lines 1-3
cleaning interval
Primary sampler control Auto - For lines 1-3
mode
Flush upstream 2.0 sec For lines 1-3
duration
Flush upstream 2.0 sec For lines 1-3
duration
Flush downstream 2.0 sec For lines 1-3
duration
Knife cuts 3 repeats For lines 1-3
Optional devices
Parameter Value Unit Other
Ph measurement off - -
DCS Parameters
READER COMMENTS
The design team of Outotec is committed to ensure the quality and usefulness of this manual
and the PSI 300 Particle Size analyzer system. Your comments, recommendations and
suggestions will help make our continuous effort more effective.
Please take few moments to write your comments either directly on the form provided or on a
photocopy of it.
Outotec, Analyzers
Fax to: FAX: INT + 358 205 292998
From:
Comments /
recommendations: