You are on page 1of 10

Polymerization Efficiency of LED Curing Lights

DANIEL L. LEONARD, DDS*


DAVID G. CHARLTON, DDS, MSDt
HOWARD W. ROBERTS, DDSt
MARK E. COHEN, P H D ~

ABSTRACT
Purpose: The purpose of this study was to compare the curing efficiency of three commercially
available light-emitting diode (LED)-based curing lights with that of a quartz tungsten halogen
(QTH)curing light by means of hardness testing. In addition, the power density (intensity) and
spectral emission of each LED light was compared with the QTH curing light in both the 380-
to 520-nm and the 450- to 500-nm spectral ranges.
Materials and Methods: A polytetrafluoroethylene mold 2 mm high and 8 mm in diameter was
used to prepare five depth-of-cure test specimens for each combination of exposure duration, com-
posite type (Silux Plus [microfill], 2-100 [hybrid]), and curing light (ZAP Dual CuringTMLight,
LumaCureTM;VersaLuxm, Optilux 401TM).After 24 hours, Knoop hardness measurements were
made for each side of the specimen, means were calculated, and a bottodtop Knoop hardness
(B/T KH) percentage was determined. A value of at least 80% was used to indicate satisfactory
polymerization. A linear regression of BTT KJd percentage versus exposure duration was performed,
and the resulting equation was used to predict the exposure duration required to produce a BK KH
percentage of 80% for the test conditions. The power densities ( p o w e r h i t area) of the LED curing
lights and the QTH curing light (Optilux 401TM)were measured 1 mm from the target using a labo-
ratory-grade, laser power meter in both the full visible light spectrum range (380-780 nm) and the
spectral range (between 450 and 500 nm), using a combination of long- and short-wave edge filters.
Results: The emission spectra of the LED lights more closely mirrored the absorption spectrum of
the commonly used photoinitiator camphorquinone. Specifically, 95 % of the emission spectrum of
the VersaLux, 87% of the LumaCure, 84% of the ZAP LED, and 78% of the ZAP combination
LED and QTH fell between 450 and 500 nm. In contrast, only 56% of the emission spectrum of
the Optilux 401TMhalogen lamp fell within this range. However, the power density between 450
and 500 nm was at least four times greater for the halogen lamp than for the purely LED lights. As
a result, the LED-based curing lights required from 39 to 61 seconds to cure a 2-mm thick hybrid
resin composite and between 83 and 131 seconds to adequately cure a microfill resin composite. By
comparison, the QTH light required only 21 and 42 seconds to cure the hybrid and microfill resin
composites, respectively.
CLINICAL SIGNIFICANCE
The first-generation LED-based curing lights in this study required considerably longer exposure
durations than the QTH curing light to adequately polymerize a hybrid and a microfill resin
composite.
(JEsthet Restor Dent 14:286-295,2002)

'Colonel, USAF Dental Corps, Chiefi USAF Dental Investigation Service, Great Lakes, Illinois
tColonel, USAF Dental Corps, Director, Dental Materials Evaluation, USAF Dental Investigation Service,
Great Lakes, Illinois
$Lieutenant Colonel, USAF Dental Corps, Director, Technical Evaluations, USAF Dental Investigation
Service, Great Lakes, Illinois
SStatistician, Naval Dental Research Institute, Great Lakes, Illinois
286 JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY
LEONARD ET AL

dental profession extremely well


G rowing demand for esthetic
dentistry and the use of light-
cured materials have made an effi-
despite some inherent limitations.
For example, the halogen lamp
that employ camphorquinone as a
photoactivator. Other advantages
follow from the first: LED units
cient, dependable curing light indis- used in these units generates light require less power to operate, since
pensable. Visible light-curing units with wavelengths outside the effec- they produce light only within a
are commonly used to polymerize tive spectrum for activation of the narrow spectral range. This lower
light-sensitive restorative materials, photoinitiator and, therefore, power requirement may eliminate
such as resin composites, resin- contains an internal filter to restrict the need for a cooling fan since heat-
modified glass ionomers, polyacid- the light to the appropriate wave- producing infrared wavelengths are
modified resin composites, and pit lengths. It has been shown that eliminated. As a result, these lights
and fissure sealants. In addition, light delivered by halogen curing have the potential to be powered
visible light-curing units are required units diminishes in intensity with with rechargeable batteries. This
for most bonding systems, an use mainly owing to degradation of feature makes it possible for the
increasing number of bases and the halogen bulb and its reflector, units to be cordless, portable, and
liners, various luting agents, and blistering and cracking of the inter- relatively lightweight. Another
some provisional restorative nal filter, and damage to the fiber- advantage is the extended life of
materials. Adequate polymerization optic tips used to focus the light on LEDs. At most, halogen bulbs last
of these materials depends on the the restorative material.6 All of 100 hours,16 whereas LEDs can last
light source intensity (irradiance these problems can be corrected thousands of hours." Also, unlike
or power density), wavelength, and easily and inexpensively. Even so, halogen bulbs and their filters and
exposure duration. Unless all three studies have shown that the major- reflectors, LEDs do not degrade over
parameters are adequate, the mate- ity of curing lights in use are not time or need reflectors and filters, so
rials incompletely polymerize and properly maintained, and a sub- the light they produce is constant in
exhibit poor physical properties stantial number deliver less than intensity. Finally, LED lights produce
that may lead to early failure.'I2 the minimum power density less heat, so there is less potential for
required to adequately polymerize gingival and pulpal irritation.
Most visible light-activated resins light-cured rest~ratives.~-~
use a diketone photoinitiator, such Previous studies have shown LED
as camphorquinone, to create free Light-emitting diode (LED) technol- technology to have promise. How-
radicals that initiate the polymeriza- ogy has been introduced as an alter- ever, these studies used prototype
tion p r o c e s ~The
. ~ ~effective
~ wave- native energy source for polymeriz- LED units rather than commercial
length range to activate camphor- ing dental restorative materials.1° products, and in some studies, the
quinone has been reported to be Several recent articles have described intensity of the halogen light with
between 410 nm and 500 nm, with and evaluated the purported advan- which the LED unit was compared
a peak wavelength of 470 nm.294 tages of LED technology compared was reduced to coincide with the
More recent research suggests a with those of halogen lights.'l-l5 reduced power density output of
further restriction of the adequate First, unlike halogen lights, LED the LED ~ n i t . l ' - * ~
wavelengths to between 450 nm units produce light within a narrow
and 490 nm.5 spectral range. The diodes use gal- The purpose of this study was to
lium nitride as a semiconductor and compare the curing efficiency of
Halogen dental curing lights, intro- produce light with wavelengths three commercially available LED-
duced in the late 1970s, have been between 450 and 490 nm with a based curing lights with that of a
the mainstay for over 20 years. peak at 460 nm. This energy range is quartz tungsten halogen (QTH)
These units continue to serve the nearly ideal for activating materials curing light by means of hardness

V O L U M E 14, N U M B E R 5 , 2002 287


POLYMERIZATION EFFICIENCY OF L E D CURING LIGHTS

testing. In addition, the power den- was used to prepare five depth-of- ized the distance from the light
sity (intensity)and spectral emissioncure test specimens. To prepare source to the resin composite and
of each LED light were compared each specimen, the mold was placed provided a smooth, non-air-inhib-
with those of the QTH curing light on a clear glass slide (1mm thick) ited surface for subsequent hard-
in both the 380- to 520-nm and the and the resin composite was placed ness testing. One side of the speci-
450- to 500-nm spectral ranges. in the mold. The resin composite was men was then exposed to the curing
covered with a second glass slide to light. One group of five specimens
MATERIALS A N D METHODS ensure that the exposed surface of was made in this way, and addi-
Four light-curing units were evalu- the composite was flat and parallel tional groups of five specimens were
ated in this study: one using a QTH to the surface of the mold. A glass fabricated at increasing exposure
source, one with a combination of slide was used rather than a Mylar durations, depending on the partic-
LED and QTH sources, and two strip because pilot studies demon- ular curing light, until adequate
using LEDs only (Figure 1; Table 1). strated minimal power loss (i.e., depth of cure was obtained. A
less than 5 % ) through the glass hybrid resin composite (2-100 shade
Testing of Curing Efficiency slide compared with a 10% loss A-3,3M ESPE Dental Products,
A polytetrafluoroethylene mold with the Mylar strip. In addition, St. Paul, Minnesota) and a microfill
2 mm high and 8 mm in diameter the 1-mm thick glass slide standard- resin composite (Silux Plus, univer-

Pigwe 1. Curing lights evaluated: A, ZAP Dual CuringR*;


~ ;Optilux 401 ”.
B, LumaCure” C, V e r ~ a L u xD,

288 J O U R N A L OF E S T H E T I C A N D R E S T O R A T I V E D E N T I S T R Y
LEONARD ET AL

sal shade, 3 M ESPE Dental Prod- TABLE 1. CURING LIGHTS EVALUATED.


ucts) were used to address reported unit Mmkctur rrp.
differences in degree of cure between ZAP Dual Curing CMS-Dental, Copenhagen, Denmark LED and halogen
hybrid and microfill resins.18-21 LumaCure LumaLite, hc.,Spring Valley, California LED
Following fabrication, specimen VCrsaLW Centrix, Inc., Shelton, Connecticut LED
thickness was measured with an optilux401 Kur/Duneaon, Danbury, conncccian Hdogco
electronic digital caliper (Fowler
& NSK, Newton, Massachusetts)
to ensure consistent thickness
(2.0 * 0.1 mm). Specimens were Power Density Measurement The power density between 450
stored dry for 24 hours in a light- The power density ( p o w e r h i t and 500 nm was calculated 1 mm
proof container prior to Knoop area) of the curing lights was from the target in the following
hardness testing (M-400-G2, LECO, measured in milliwatts per square manner: the power was measured
St. Joseph, Michigan), using a 100-g centimeter (mW/cm2)with the 1 mm from the target without the
load and a 10-s dwell time. For light tip placed 1 mm from the edge filters, using a Power Max
each specimen, three hardness mea- target, using a laboratory-grade 5OOD laser power meter with
surements were made on the top laser power meter (Molectron PMlO probe. At 1 mm, all the light
surface and three on the bottom Power Max 5OOD with PMlO from the lamp fell on the detector.
surface. Mean hardness values were probe, Molectron Detector, Inc., The power was then measured 20
calculated for each surface. These Portland, Oregon). Power density mm from the target without the
values were then averaged for the at 1 mm is frequently reported in filter. At 20 mm, all of the light did
five specimens to obtain mean top the literature since this is the aver- not fall on the detector, and, there-
and bottom surface values. The age distance of the light tip from fore, a ratio of the two measure-
bottom value was divided by the the composite being cured in clini- ments gave the percentage of power
top value and multiplied by 100 to cal practice. It is generally reported lost. This ratio was multiplied by
obtain percentage depth of cure. that a power density of 300 to the power measurement at 20 mm
If that mean value exceeded 80%, 400 mW/cm2 as measured between (the closest distance the light tip
the specimen was considered to be 380 and 520 nm is necessary to could be positioned from the detec-
adequately p ~ l y m e r i z e d . ~ ~ - ~ ~ adequately cure a 2-mm thick tor when the filters were in place),
composite ~pecimen.’?~’ However, using the edge filters to obtain the
For each resin composite and cur- the spectral range of 450 to 490 calculated power at 1 mm for the
ing light, the data were plotted, and nm has been shown to be the most 450- to 500-nm wavelengths. The
a simple linear regression was per- effective in initiating the poly- power density between 450 and
formed with b o t t o d t o p Knoop merization process and is a more 500 nm was calculated using the
hardness (B/T KH) percentage as accurate indicator of a light’s effi- following formula:
the dependent variable and expo- ciency.’ Therefore, the power den-
power density = measured
sure duration as the independent sity was also determined for the
power(mW)/area of light tip/
variable. The resulting equation spectral range between 450 and
%transmittance of filters
was used to predict the exposure 500 nm ( 2 10 nm), using a combi-
duration required under the test nation of long- and short-wave The only assumption made is
conditions to produce a specimen edge filters (Thermo Corion, that the light diverges the same
with a B/T KH percentage of 80 %. Franklin, Massachusetts). amount with the filter as without

V O L U M E 14, N U M B E R 5 , 2 0 0 2 289
POLYMERIZATION EFFICIENCY OF LED CURING LIGHTS

TABLE 2. EXPOSURE DURATION REQUIRED TO OBTAIN 80% BOTTOM/TOP KNOOP the filter. This assumption is not
HARDNESS PERCENTAGE FOR A 2-MM THICK LAYER OF RESIN COMPOSITE. strictly true but is considered to be
Exporum Duration (4
sufficient if the majority of light
( M n g Lght 2-100 suux Plus falls on the detector.
ZAP Dual Curing 39 83

LumaCure 53 103 Spectral Emission Measurements


131 The spectral emission plots of the
VersaLux 61
VersaLux, LumaCure, ZAP Dual
Optilw 401 21 42
Curing Light, and Optilux 401
(Demetron Research Corp.,

100 100

80 80 -

- -E
I 60 I 60 -
Y Y
Q a

5 40-

20 20 -
P = 0.91 r2 = 0.76
I KH% = 21.72 + 1.48 x exposure duration (s)
B
T B R KH% = 15.38 + 0.78 x exposure duration (8)
0 0- I

0 20 40 60 80 100 120 140 180


A Exposure Duration (s)

20 - 20 - i
P = 0.88 P=0.87
B/TKH%= 32.19 + 0.91 x exposure duration (s) BTT KH% = 24.54 t 0.54x exposure duration (s)

Figure 2. Correlationand linear regression between the BottomlTop Knoop hardness percentages of A, Z-100 and exposure
duration using the Zap Dual Curing Light; B, Silux Plus and exposure duration using the Zap Dual Curing Light; C, Z-100 and
exposure duration using the LumaCure Curing Light; D,Silux Plus and exposure duration using the LumaCure Curing Light.

290 J O U R N A L OF ESTHETIC A&ID RESTORATIVE D E N T I S T R Y


L E O N A R D E T AL

Danbury, Connecticut) were bination LED-QTH light, and Data for B/T KH percentage using
determined with a spectral the halogen light required to pro- the four curing lights at the differ-
telephotometerkolorimeter (PR-650 duce 2-100 and Silux Plus speci- ent exposure durations are pre-
SpectraColorimeter, Photo Research, mens with a B/T KH percentage of sented in Figure 2. Included in the
Inc., Chatsworth, California). 80% are reported in Table 2. All graphs is the regression equation
the LED-based lights required and r2. The r2 value represents the
RESULTS longer exposure durations to poly- proportion of the total variability
Curing Efficiency merize the two resin composites in B/T KH percentage that can be
The predicted exposure durations than did the halogen light. accounted for by exposure dura-
for the two LED lights, the com- tion alone. As seen in Figure 2, A,

100 ,
80

-
E
I 60
Y

40

m
4
2o 11 rz = 0.87
B i l KH% = 35.00 + 0.74 x exposure duration (5) I
20

0
f
9 = 0.82
BTT KH% = 27.67 + 0.40 x exposure duration (s)
0 1 I
0 20 40 60 80 100 120 140 160 0 20 40 60 80 loo 120 140 160

E Exposure Duration (9) F Exposure Duration (s)

100

a
loo 0
-
E
5
Q.
80

60
/
5
P
40

20 7.0-

13 = 0.81 13 = 0.91
B/T KH% = 62.71+ 0.82 x exposum duration (s) BIT KH% = 27.50 + 1.25 x exposurs duration (s)
0
0 20 40 80 80 100 120 140 1M
G Exposure Duration (s)

Figure 2 continued. E, Z-100 and exposure duration using the VersaLux Curing Light; F, Silux Plus and exposure duration
using the VersaLux Curing Light; G,Z-100 and exposure duration using the Optilux 401 Curing Light; and H,Silux Plus and
exposure duration using the Optilux 401 Curing Light.

VOLUME 14, N U M B E R 5 , 2002 291


POLYMERIZATION EFFICIENCY O F LED CUHINC LIGHTS

the first mean B/T KH percentage Using the LumaCure Light to poly- With the Optilux 401 light and
value above 80% for the Zap Cur- merize Silux Plus revealed that the 2-100 (see Figure 2, G), the first
ing Light using the hybrid resin exposure duration of 108 seconds mean BTT KH percentage value above
composite 2-100 was recorded was necessary to produce the first 80% was recorded when the expo-
when the exposure duration was 34 specimen with a mean B/T KH sure duration was 20 seconds. The
seconds. Analysis showed that there percentage value above 80% (see exposure duration correlated signifi-
was a statistically significant corre- Figure 2, D). Exposure duration cantly with the B/T KH percentages
lation between exposure duration was found to correlate significantly ( p c .0001, r2 = 0.81). Based on the
and B/T KH percentage of the spec- with the B/T KH percentages of the regression equation, the predicted
imens ( p c .0001, r2 = 0.91). From specimens ( p < .0001, r2 = 0.87). exposure duration required to pro-
the regression equation, the pre- The regression equation revealed duce a specimen with a B/T KH
dicted exposure time required to that the predicted exposure duration percentage of 80% is 21 seconds.
produce a specimen with B/T KH required to produce a specimen with
percentage of 80% is 39 seconds. B/T KH percentage of 80% under The first mean B/T KH percentage
these test conditions is 103 seconds. value above 80% measured for
For the microfill resin composite Silux Plus using the Optilux 401
Silux Plus, the Zap Curing Light Figure 2, E, shows that the first mean was recorded when the exposure
produced the first mean BIT KH B/T KH percentage value above 80% duration was 40 seconds (see
percentage value above 80% with when polymerizing 2-100 with the Figure 2, H). Exposure duration
an exposure duration of 85 seconds VersaLux Curing Light was recorded was found to correlate significantly
(see Figure 2, B). Exposure duration with an exposure duration of 60 sec- with the B/T KH percentages
was found to correlate significantly onds. Exposure duration correlated ( p c .0001, r2 = 0.91). The regres-
with the B/T KH percentages significantly with the B/T KH per- sion equation predicted that an
( p c .0001, r2 = 0.76). The regres- centages of the 2-100 specimens exposure of 42 seconds is required
sion equation revealed that the pre- ( p < .0001, r2 = 0.87). As calculated with Silux Plus to produce a B/T
dicted exposure duration required using the regression equation, the KH percentage of 80%.
to produce a specimen with B/T KH exposure duration required to pro-
percentage of 80% under these test duce a specimen with a B/T KH Power Density and Spectral Emission
conditions is 83 seconds. percentage of 80% is 61 seconds. The power density measured for
each light in the emission spectrum
The first mean B/T KH percentage The first mean B/T KH percentage of 380 to 520 nm, and in the more
value above 80% measured for the value above 80% produced with effective 450- to 500-nm range, is
LumaCure Light using 2-100 the VersaLux Light using Silux Plus presented in Figure 3. Spectral
(see Figure 2, C) was recorded (see Figure 2, F) was observed when emission plots are presented in Fig-
when the exposure was 48 seconds. the exposure duration was 120 sec- ure 4. These figures demonstrate
Again, the exposure duration corre- onds. Statistical testing found that that the LED lights more closely
lated significantly with the B/T KH exposure duration correlated signif- mirror the absorption spectrum of
percentages ( p c .0001, r2 = 0.86). icantly with the B/T KH percentages camphorquinone. Specifically, 95 %
Using the regression equation, the ( p < .0001, r2 = 0.82). From the of the emission spectrum of the
predicted exposure duration regression equation, an exposure Versalux, 87% of the LumaCure,
required to produce a 2-100 speci- time of 131 seconds is necessary 84% of the ZAP LED, and 78% of
men with B/T KH percentage of to obtain a B/T KH percentage of the ZAP combination LED and
80% is 53 seconds. 80% with this resin composite. QTH fell between 450 and 500 nm.

292 J O U R N A L O F ESTHETIC A N D RESTORATIVE DENTISTRY


LEONARD E T A L

In contrast, 56% of the Optilux


401 halogen lamp's emission spec-
1000
trum fell within this range. How-
f
ever, as seen in Figure 3 , the power 5
750
density between 450 and 500 nm
was at least four times greater for
s
the halogen lamp than for the
3 500

purely LED-based lights. il


n.
250

DISCUSSION
0
Because of the increasing number LurmCun V-Lux ZAPFED) LI\p OPT401
(LED+H.logm)
of light-activated dental materials,
Figure 3 . Power density (mWlcm2)for the LumaCure, VersaLux, ZAP,
the visible light-curing unit is a and Optilux 401 curing lights in both 450- to SOO-nm and 380- to
common piece of equipment in den- S20-nm spectral ranges.
tal offices. The proper performance
of these units (i.e., their ability to
provide adequate power density) is light when both were adjusted to Organization for Standardization
crucial to optimizing the physical provide an output of 100 mW/crn2." (ISO) 4049 depth of cure testing.13
properties of light-activated materi- Similarly, Mills and co-workers The compressive strengths of sam-
als. Inadequate polymerization has demonstrated a significantly greater ples produced by both lights were
been associated with inferior physi- depth of cure with a 21-LED array statistically equivalent.
cal properties, higher solubility, light than with a halogen curing
retention failures, and adverse pul- light; however, in their study the All of these studies used either
pal responses caused by residual halogen light's output was adjusted experimental LED devices with a
down
unpolymerized r n ~ n o m e r s . ~ ~ ~ ~ ~ ~ ~ to 300 mW/cm2 to coincide relatively large number of LEDs
The performance of halogen lights with the output of the LED light.12 and concentrating light guide tips
is known to diminish over time if Stahl and colleagues compared or reduced the output of the halo-
the bulb, reflector, fiber-optic tips, the flexural strength and flexural gen light to coincide with the lower
and filter are not properly main- modulus of resin composites poly- output of the LED light or both."-14
tained.6 It is also known that prac- merized with a halogen curing The commercially available LED
titioners are often unaware of this light (755 mW/cm2) and a 27-LED lights used in this study only had
degradation and continue to use array light with a concentrating 7- or 12-LED arrays. Therefore,
poorly performing lights, which leads light guide tip (350 mW/cm2).14 In the findings of this study are con-
to inadequate p~lymerization.~-~ most cases, no significant differ- sistent with those of Kurachi and
ences in the physical properties c o - ~ o r k e r swho
, ~ ~ compared a
Recent literature suggests that were found. In a parallel study 6-LED array light having a power
LEDs, specifically gallium nitride comparing the same light sources, density of 79 mW/cm2 with a
blue LEDs, offer an effective means Jandt and colleague^'^ reported halogen light having an output of
of polymerizing light-activated that the halogen light produced a 475 m W / ~ m ~In. ' that ~ study,
materials.11-13 Fujibayashi and significantly greater depth of cure hybrid resin composite samples
colleagues" found that a 61-LED than the LED light but that both (0.35, 1.25, and 1.8 mm thick)
array produced a greater depth of lights met the minimum require- were exposed for the typical 40 sec-
cure than did a halogen curing ments for the International onds with the halogen light and for

VOLUME 14, N U M B E R 5, 2 0 0 2 293


POLY\lERI% \TIOX EFFIGIES(:\ OF L E D C:l-RlSG LIGHTS

1.s approximately twice the exposure


time needed by the hybrid resin
composite (2-100) to produce the
-Ea same depth of cure.
C

a 0.5 The results indicate that the com-


ii
0
mercially available LED-based
n curing lights used in this study can
0
380 396 412 428 444 480 476 492 508 524 be expected to effectively polymer-
ize a hybrid resin composite in
Wavelength (nm) an acceptable amount of time
Figure 4. Spectral emission plots of the LumaCure, VersaLux, ZAP, and (39-61 s), but not a microfill resin
Optilux 401 curing lights. The plots have each been normalized such that
the maximum power density of each light source is defined as one arbitrary composite (83-131 s). However,
unit (a.u.). LED technology for polymerizing
light-activated dental materials
holds great promise. Future curing
20,40, 60,120, and 180 seconds Degree of polymerization is based units with a greater number of
with the LED light. Under those on several factors, including the LEDs and power-concentrating
conditions, samples exposed to the output and wavelength of the cur- light guides should provide the
LED light for 40 seconds had lower ing light, exposure duration, and power density necessary to ade-
hardness values than those poly- the particular light-activated mater- quately polymerize materials in a
merized with the halogen light for ial. In this study, microhardness timely manner. The many advan-
40 seconds. All samples, regardless testing was used as an indicator of tages of LED units compared with
of thickness, required 120 seconds adequate polymerization. Ideally, those using halogen bulbs make
of exposure with the LED light to the B/T KH percentage would be further development of this technol-
produce hardness values compara- 100%; however, previous studies ogy exciting and promising.
ble to those produced by an expo- have used 80% as the minimum
sure duration of 40 seconds with value required to indicate adequate CONCLUSION

the halogen light. p ~ l y m e r i z a t i o n . ~Microfill


~ - ~ ~ resin The polymerization efficiency of
composites have been shown to three commercially available LED-
The LED curing lights in this study require more power than hybrid based curing lights was compared
have a higher percentage of their resin composites for adequate poly- with that of a QTH light by means
output in the absorption spectrum merization.18-21?28 It is believed that of hardness testing. In addition, the
of camphorquinone compared with microfills exhibit a reduced depth power density (intensity) of each
the Optilux halogen curing light of cure because their small filler light in both the normally reported
(see Figures 3 and 4). Nevertheless, particles cause light scattering, 380- to 520-nm range and the more
even if the output is limited to which decreases the effectiveness effective 450- to 500-nm spectral
between 450 and 500 nm, as sug- of the curing light. This concept is range was measured and compared.
gested by Nomoto: the halogen reinforced by the results of this Although a smaller percentage of
light still provides greater power study because the microfill resin the QTH unit’s power density fell
density than LED lights. composite (Silux Plus) required within the important 450- to 500-nm

294 J O U R N A L OF E S T H E T I C A N D R E S T O R A T I V E D E N T I S T R Y
LEONARD E T A L

absorption range of the camphor- 7. Barghi N, Berry T, Hatton C. Evaluating 19. Atmadja G, Bryant RW. Some factors
intensity output of curing lights in private influencing the depth of cure of visible
quinone photoactivator, it was still dental offices. J Am Dent Assoc 1994; light-activated composite resins. Aust Dent
at least four times more powerful 125:992-996. J 1990; 35:213-218.
than the LED-based units in this 8. Martin FE. A survey of the efficiency of 20. Rueggeberg FA, Caughman WF, Curtis
range. As a result, the LED-based visible light curing units. J Dent 1998; JW, Davis HC. A predictive model for the
26:239-243. polymerization of photo-activated resin
units required considerably longer composites. Int J Prosthodont 1994;
9. Miyazaki M, Hattori T, Ichiishi Y, Kondo 7:159-166.
exposure times to adequately poly- M, Onose H, Moore BK. Evaluation of
merize a hybrid and a microfill curing units used in private dental offices. 21. Leonard DL, Charlton DG, Roberts HW,
Oper Dent 1998; 2350-54. Hilton TJ, Zionic A. Determination of the
resin composite. minimum irradiance required for adequate
10. Mills RW. Blue light emitting diodes: polymerization of a hybrid and a microfill
Another method of light curing? Br Dent composite. Oper Dent 2001; 26:176-180.
DISCLOSURE AND J 1995; 178:169.
ACKNOWLEGMENTS 22. Watts DC, Amer 0, Combe EC. Charac-
11. Fujibayashi K, Ishimaru K, Takhashi N, teristics of visible-light-acdvated compos-
The views expressed in this article Kohno A. Newly developed curing unit ite systems. Br Dent J 1984; 156:209-215.
are those of the authors and do not using blue light-emitting diodes. Dent Jpn
1998; 34:49-53. 23. Breeding LC, Dixon DL, Caughman WF.
reflect the official policy of the The curing potential of light-activated
12. Mills RW, Jandt KD, Ashworth SH. Den- composite resin luting agents. J Prosthet
Department of Defense or other Dent 1991; 65512-518.
tal composite depth of cure with halogen
departments of the United States and blue light emitting diode technology.
Br Dent J 1999; 186:388-391. 24. Manga RK, Charlton DG, Wakefield CW.
Government. In vitro evaluation of a curing radiometer
13. Jandt KD, Mills RW, Blackwell GB, as a predictor of polymerization depth.
Ashworth SH. Depth of cure and com- Gen Dent 1995; 43:241-243.
The authors thank 3M ESPE for pressive strength of dental composites
supplying the resin composites used cured with blue light emitting diodes 25. Bayne SC, Taylor DF. Dental materials. In:
(LEDs). Dent Mater 2000; 16:41-47. Sturdevant CM, Ro berson TM, Heymann
in this study. HO, Sturdevant JR, eds. The art and sci-
14. Stahl F, Ashworth SH, Jandt KD, Mills ence of operative dentistry. 3rd Ed. St.
RW. Light-emitting diode (LED) poly- Louis: Mosby, 1995:260.
REFERENCES merisation of dental composites: flexural
1. Rueggeberg FA, Caughman WF, Curtis properties and polymerisation potential. 26. Blankenau RJ, Kelsey WP,Powell GL,
JW. Effect of light intensity and exposure Biomaterials 2000; 21:1379-1385. Shearer GO, Barkmeier WW, Cave1 WT.
duration on cure of resin composite. Oper Degree of composite resin polymerization
Dent 1994; 19:26-32. 15. Kurachi C, Tuboy A, Magalhaes D, with visible light and argon laser. Am J
Bagnato V. Hardness evaluation of a Dent 1991; 4:40-42.
2. Cook W. Spectral distributions of dental dental composite polymerized with experi-
photopolymerization sources. J Dent Res mental LED-based devices. Dent Mater 27. Ferracane JL, Mitchem JC, Condon JR,
1982; 61:1436-1438. 2001; 17:309-315. Todd R. Wear and marginal breakdown
of composites with various degrees of
3. Council on Dental Materials. Visible light- 16. Rueggeberg FA, Twiggs SW, Caughman cure. J Dent Res 1997; 76:1508-1516.
cured and activating units. J Am Dent WF, Khajotia S. Lifetime intensity profiles
ASSOC1985; 11O:lOO-103. of 11 light-curing units. J Dent Res 1996; 28. Ruyter IE,Oysaed H. Conversion in differ-
75:380. (Abstr) ent depths of ultraviolet and visible light
4. McCabe JF, Carnick TE. Output from activated composite materials. Acta Odon-
visible-light activation units and depth of 17. Haitz RH, Craford MG, Weissman RH. to1 Scand 1982; 40:179-192.
cure of light activated composites. J Dent Light emitting diodes. In: Bass M, ed.
Res 1989; 68:1534-1539. Handbook of optics. 2nd Ed. New York:
McGraw Hill, 1995:12.1-12.39.
5. Nomoto R. Effect of light wavelength on Reprint requests: Col. David G . Churlton,
polymerization of light-cured resins. Dent 18. Ferracane JL, Aday P, Matsumoto H, Det I, USAFSAM, 31 OC B St., Bldg IH,
Mater J 1997; 16:60-73. Marker VA. Relationship between shade Great Lakes, 1L 60088; e-mail:
and depth of cure for light-activated dental dauid.charlton&dri.med.navy.mil
6. Friedman J. Variability of lamp character- composite resins. Dent Mater 1986; Q2002 BC Decker lnc.
istics in dental curing lights. J Esthet Dent 2:80-84.
1989; 1:189-190.

VOLUME 1 4 , NUMBER 5, 2 0 0 2 295

You might also like