33K1-4-1177-1
TECHNICAL MANUAL
CALIBRATION PROCEDURE
FOR
3403172-1
(LOCKHEED CORP)
Distribution Statement C - Distribution authorized to U. S. Government agencies and their contractors for
official use or for administrative or operational purposes only, 30 June 1988. Requests for this document
shall be referred to 562 CBSG/GBHA, 813 Irving-Wick Dr W, Heath, OH 43056-1199.
Destruction Notice - For unclassified, limited documents, destroy by any method that will prevent
disclosure of the contents or reconstruction of the document.
Published under Authority of the Secretary of the Air Force
30 JUNE 1998
CHANGE 4 - 30 MARCH 2008
T.O. 33K1-4-1177-1
INSERT LATEST CHANGED PAGES. DESTROY SUPERSEDED PAGES.
LIST OF EFFECTIVE PAGES NOTE: The portion of the text and illustrations affected by the changes is indicated by a vertical
line in the outer margins of the page.
Title .......................................4
A. ...........................................4
1 .............................................1
2 .............................................4
3 .............................................3
4 .............................................0
5 .............................................2
6 - 16......................................0
A Change 4 USAF
T.O. 33K1-4-1177-1
3403172-1
(LOCKHEED CORP)
1 CALIBRATION DESCRIPTION:
Table 1.
Torque Indicator
Change 1 1
T.O. 33K1-4-1177-1
2 EQUIPMENT REQUIREMENTS:
2 Change 4
T.O. 33K1-4-1177-1
3 PRELIMINARY OPERATIONS:
3.1 Review and become familiar with entire procedure before beginning Calibration Process.
WARNING
WARNING
For PMEL use, earth ground is not provided thru J1 power connector. An
external earth ground is required to comply with electrical safety. Failure to do
so may result in hazardous voltage on the TI case.
3.3 Connect external earth ground to TI case and connect TI J1 POWER as follows:
A 28 VDC HI
B 115 VAC/400 Hz HI
3.4 Connect Test Equipment to 115 VAC/60 Hz, POWER switches to on, and allow 15 minutes warm-up.
4 CALIBRATION PROCESS:
NOTE
Unless otherwise specified, verify the results of each test and take corrective
action whenever the test requirement is not met, before proceeding.
Change 3 3
T.O. 33K1-4-1177-1
4.1.3 Connect the DC Voltage Standard output to the TI Volt Ohmmeter input observing polarity.
4.1.4 Set the DC Voltage Standard output for the first applied value listed in Table 2, and STANDBY/ OPERATE
switch to OPERATE.
4.1.5 The TI Volt Ohmmeter must indicate within the corresponding limits listed in the Limits column of Table 2.
4.1.7 Repeat steps 4.1.4 through 4.1.6 for each remaining applied value listed in Table 2.
Table 2.
4.1.9 Set the TI Volt Ohmmeter FUNCTION to OHMS and RANGE to 100.
4
T.O. 33K1-4-1177-1
4.2.4 Adjust TI TACH RPM control fully CCW and verify the PERCENT RPM indicator displays .F.
4.2.5 Slowly adjust TI TACH RPM control fully CW. The PERCENT RPM indicator will begin to display at about
16.7 percent. Adjust TI TACH RPM control fully CW and verify display indicates 110.0 percent or greater.
4.2.7 Set Oscilloscope controls to measure a sine wave of 20 V p-p at about 70 Hz.
4.2.8 Disconnect Phasemeter REF and SIG inputs and connect to Oscilloscope CH A and B respectively.
4.2.9 The Oscilloscope must display a sinusoidal waveform with an amplitude of 18 to 22 V p-p on both
channels.
4.2.10 Set the Frequency Counter controls to measure a sine wave of 20 V p-p at about 70 Hz. Ensure input
Impedance switch is set to 1 MΩ and ATTEN switch is set to X10 or X20.
4.2.11 Disconnect the Oscilloscope. Connect TI TACH A and TORQUE DC (Black) to Frequency Counter HI and
LO respectively.
4.2.12 Adjust TI TACH RPM control for 100% as monitored on TI PERCENT RPM indicator.
Change 2 5
T.O. 33K1-4-1177-1
4.2.14 The Frequency Counter must indicate 69.8 to 70.2 Hz (14245 to 14326 μs).
4.2.15 Adjust TI TACH RPM control for 30% as monitored on TI PERCENT RPM indicator.
4.2.17 The Frequency Counter must indicate 20.8 to 21.2 Hz (47169 to 48076 μs).
4.2.18 Set TI TACH RPM control fully CCW. Set TI 115 VAC POWER to OFF, disconnect setup and test cables
from TI.
C S1
D S3
POWER ON
LOCAL/REMOTE LOCAL
SYNCHRO/RESOLVER SYNCHRO
REFERENCE 26 VAC
4.3.4 Set TI PRESSURE SIMULATOR control to 0°, 90°, 180° and 270°. The TI PRESSURE SIMULATOR
indicator must indicate within ±0.1° of the API.
6
T.O. 33K1-4-1177-1
CH A HI TORQUE (red)
CH A LO TORQUE (black)
CH B HI REFERENCE (red)
CH B LO REFERENCE (black)
4.4.4 Verify both signals displayed on the Oscilloscope are triangular with an amplitude of 2.8 to 3.8 V p-p.
4.4.5 Disconnect Oscilloscope and connect each input sequentially to the Frequency Counter inputs. Verify the
Frequency Counter reads 6.4 to 8.4 kHz.
4.4.7 Set TI TORQUE PHASE switch to each position listed in the Applied column of Table 3. The Phasemeter
must indicate within the corresponding limits listed in Table 3.
7
T.O. 33K1-4-1177-1
Table 3.
36 32 to 40
4.4.9 Connect the Digital Multimeter, set to measure DC Volts, to the TI TORQUE DC test jacks (red HI, black
LO).
4.4.10 Set the TI TORQUE PHASE switch to the first position listed in the Torque Phase switch column of
Table 4.
4.4.11 The Digital Multimeter must indicate within the corresponding limits listed in the Limits column of
Table 4.
4.4.12 Repeat steps 4.4.10 and 4.4.11 for each remaining value listed in Table 4.
Table 4.
36 0.30 to 0.42
4.4.13 Set the TI 115 VAC power OFF, and disconnect the Digital Multimeter and cables.
28 VDC POWER ON
4.5.2 Connect the Oscilloscope GND to the TI FUEL FLOW TRANSMITTER A test jack.
8
T.O. 33K1-4-1177-1
4.5.3 Connect the Oscilloscope CH A input to the first TI FUEL FLOW SENSOR connector pin listed in Table 5.
4.5.4 Set Oscilloscope controls as necessary to view at least one complete cycle of the waveform.
4.5.5 Verify the signal on the Oscilloscope is a distorted square wave, and is within the corresponding limits listed
in Table 5.
4.5.6 Repeat steps 4.5.3 through 4.5.5 for each remaining pin connection listed in Table 5.
Table 5.
Fuel Flow
Connector Pin Limits (Period ms) Limits (V p-p)
4.5.8 Connect the Oscilloscope CH A input, through a X10 probe, to the first TI FUEL FLOW SENSOR connector
pin listed in Table 6.
4.5.9 Set Oscilloscope controls as necessary to view at least one complete cycle of the waveform.
4.5.10 Verify the signal on the Oscilloscope is a distorted sine wave, and amplitude and period is within limits
listed in Table 6.
4.5.11 Repeat steps 4.5.8 through 4.5.10 for each remaining pin connection listed in Table 6.
Table 6.
Fuel Flow
Connector Pin Limits (Period ms) Limits (V p-p)
9
T.O. 33K1-4-1177-1
4.5.15 Remove TI from case and disconnect plug from TI Fuel Flow Transmitter inside TI.
SPECTRUM ANALYZER
AC SIGNAL SOURCE
H L G
TI
AC VOLTAGE
DIVIDER
A D
B C
Figure 1.
FREQUENCY 800 Hz
RESOLUTION BANDWIDTH 30 Hz
FREQ SPAN/DIV 10 Hz
4.5.20 Set TI FUEL FLOW TRANSMITTER dial to the first value listed in the Transmitter column of Table 7. Set
AC Voltage Divider to the corresponding value listed in the AC Voltage Divider column of Table 7.
10
T.O. 33K1-4-1177-1
4.5.21 Set the Spectrum Analyzer controls as necessary to view the amplitude of the second harmonic (800 Hz)
signal. Adjust the TI FUEL FLOW TRANSMITTER dial for minimum amplitude (null) of the second harmonic
signal displayed on the Spectrum Analyzer.
4.5.22 The TI FUEL FLOW TRANSMITTER indication must be within the values listed in the Limits column of
Table 7.
4.5.24 Repeat steps 4.5.20 through 4.5.23 for each remaining Transmitter Setting listed in Table 7.
Table 7.
SPECTRUM ANALYZER
AC SIGNAL SOURCE
H L G
TI
AC VOLTAGE
DIVIDER
A D
B C
B0 B10
Figure 2.
11
T.O. 33K1-4-1177-1
Table 8.
Table 9.
4.6.2 Connect the Digital Multimeter, set to measure ohms, to the OIL QUANTITY INDICATOR jack pins A
and B.
4.6.4 Disconnect the Digital Multimeter from pin A, and connect to pin C.
12
T.O. 33K1-4-1177-1
4.7.1 Use K Thermocouple and connect the Digital Thermometer to TI Temperature Indicator connector pins
D (Chromel/yellow) and A (Alumel/red).
4.7.3 Set TI TEMP switch to the first value listed in the applied column of Table 10.
4.7.4 The Digital Thermometer must indicate within the corresponding limits listed in Table 10.
4.7.5 Repeat steps 4.7.3 and 4.7.4 for each remaining Applied value listed in Table 10.
Table 10.
4.7.8 Connect the Digital Multimeter, set to measure VDC, to TI TEMPERATURE DC and GND test points.
4.7.11 The Digital Multimeter must indicate in mV 10 X ambient temperature ±60 mV.
13
T.O. 33K1-4-1177-1
90 87.3 to 92.7
30 20.8 to 21.2
0 359.9 to 0.1
90 89.9 to 90.1
36 32 to 40
14
T.O. 33K1-4-1177-1
36 0.30 to 0.42
15
T.O. 33K1-4-1177-1
A to B 209 to 215
B to C 215 to 220
16