Professional Documents
Culture Documents
Overview PDF
Overview PDF
Ion-selective electrodes (ISEs) are electrochemical ion sensors that convert the ac-
tivity of a target ion into an electrical potential as the measurable signal.
Related terms:
Biosensor, Fluorine Atom, Sodium Atom, Chloride, Fluoride, Ion, Ionophore, Be-
havior as Electrode, Reference Electrode, pH Value
(10)
Here, I denotes the ion one aims to measure (also called primary ion) while j is any
other ion of the same charge that may interfere in the measurement. This equation
is a direct extension of the Nernst equation and incorporates an additional weighted
activity term that describes the selectivity. The selectivity coefficient is written in
the form KI, jpot, where the superscript “pot” indicates a selectivity coefficient for
a potentiometric probe and the subscript indicates in the first position the primary
ion and in the second position the interfering ion. For a potassium-selective elec-
trode with sodium as potential interference, one writes for example the selectivity
coefficient as KK, Napot. Smaller selectivity coefficients indicate better selectivity.
As their values can be very small in some cases, it is customary to write them as
logarithmic values, such as logKI, jpot. Large negative values indicate small selectivity
coefficients and excellent selectivity.
(11)
In the absence of primary ions, Eq. (11) predicts correctly the appropriate Nernst
equation for the interfering ion, but it unfortunately fails if both ions substantially
contribute to the potential. Thermodynamically sound descriptions have been put
forward in the literature, but one may alternatively use the following general equa-
tion that is reasonably applicable for ions of any valency.13 For any two such ions,
(12)
The anticipated error of the measurement, pI,j(%), from the interfering ion is de-
scribed quite readily if one assumes less than about a 10% error12:
(13)
From this relationship, the required selectivity coefficient for a given tolerated error
up to about 10% is given as:
(14)
(15)
For pH electrodes particularly, one varies the concentration of the primary ion in a
wide range to observe deviation from Nernst behavior at low ion concentrations.
This comprises the fixed interference method. As shown in Fig. 2, the selectivity
coefficient is obtained from extrapolating the Nernstian response range to the
potential of the background electrolyte, which gives the detection limit discussed
below, ai(LDL). Together with the knowledge of the background ion activity one gets:
(16)
Note that one should confirm a near-Nernstian response slope for each ion to avoid
experimental bias.14 For very selective membranes one strongly benefits from opti-
mized protocols to achieve this, for example using chelating agents or precipitation
reactions to reduce the primary ion concentration, or, especially with polymeric
membrane electrodes, reconditioning steps or the extrapolation of time dependent
selectivity coefficients.14
(17)
This method is particularly justified for potentiometric probes that function in un-
usual ways that are difficult to describe theoretically, as with kinetically responding
probes, some gas sensors or enzyme based potentiometric biosensors, or other
probes that rely on the combination of different recognition principles.
Ion-selective electrodes are now well understood in terms of the underlying theory,
and this has made it possible for new sensing principles to emerge that make
use of the thousands of chemical receptors originally developed for ion-selective
electrodes. One is the field of optical sensors, which has not been discussed here
because it is outside the focus of this chapter. Such so-called bulk optodes do
not require electrical connectivity between the sensing and detection unit and are
therefore more easily brought into various shapes and sizes, including particle
formats, which suit the need of modern chemical analysis.
ANALYTICAL APPLICATION OF
MEMBRANES
K. Scott, in Handbook of Industrial Membranes (Second Edition), 1995
Ion-Selective Electrodes
Ion-selective electrodes have many applications in water analysis and environmental
analysis. Electrodes include those for the determination of pH, F−, CN−, NH3 and
total hardness (Ca2+ + Mg2+). The principle of operation of the electrode is shown
in Fig 10. The measurement of the potential between the ion-selective electrode
and the reference electrode allows the determination of the ion Mn+ between the
analysed solution and an internal standard solution. In most cases the ion-selective
electrode and reference electrode are mounted in a single (combination) probe (see
Fig 10).
FIGURE 10. Basic electrode principle for determination of a ion Mn+ with an ion-se-
lective electrode.
As, in principle, all factors associated with the use of the electrode, other than the
ion concentration to be measured, remain constant (eg Cs), the measurement of the
cell potential is directly proportional to the logarithm of the ion concentration.
A variety of different solid state membranes are used for different ion detection.
Many of these use a combination of Ag2S + Ag X (X = Cl−, Br− or SCN−) in the form
of a pressed disc and the electrode responds to X”. For the determination of cations
(M+) the membrane material is a mixture of Ag2S + Mn/2 S.
Heterogeneous membranes use similar active components as solid state devices but
instead the active material is deposited into the pores of an inert support, eg silicone
rubber.
Ion-Selective Electrodes
Ion-selective electrodes (ISEs) are very similar in use to pH electrodes. They are used
for chloride, potassium, calcium, carbon dioxide/carbonate, oxygen, and a variety
of other ions. These methods are particularly suited for field analysis and online
measurements.
Lenz, B.L. and J.R. Mold, Ion—selective electrode method compared to
standard methods for sodium determination in mill liquors, Tappi J. 54(12):
2051–55(1971).Sodium ion can be quickly measured directly with ion-specific
electrodes, with about 1% error, over a wide range of concentrations. A small
sample is diluted with a small amount of ionic strength adjuster (ISA, an
ammonium buffer solution) and the solution measured as if for pH, but in
this case for p(Na+). This method would probably be useful for measuring
sodium loss in pulp by incubating the pulp with ISA until ammonium ion
has exchanged sodium ion. It could also measure sodium concentrations at
various stages of the recovery process.
Cooper, Jr., H.B.H., Continuous measurement of sodium sulfide in black
liquor, Tappi 58(6):59–62(1975).This is a general article on the subject.
Schwartz, J.L. and T.S. Light, Analysis of alkaline pulping liquor with sulfide
ion—selective electrode, Tappi J. 53(1):90–95(1970).New electrodes often use
a salicylate buffer with 1:3 sample:buffer dilution. This article has much exper-
imental detail.
Surfactants☆
María C. Prieto-Blanco, ... Dario Prada-Rodríguez, in Encyclopedia of Analytical
Science (Third Edition), 2019
Electroanalytical Techniques
New ion-selective electrodes were applied for the end-point detection in poten-
tiometric titration of ionic surfactants in raw technical materials, formulations and
even wastewater of industrial plants. Different polyvinyl chloride (PVC) membrane
electrodes with different analytical properties were designed by modifying their
electroactive material. For the analysis of anionic surfactants, the electroactive ma-
terial can be formed by a cationic surfactant and tetraphenylborate, although ionic
liquids or cationic complexes of metals with organic reagents were also proposed.5
To meet the demand for miniaturized systems for an in situ surfactant analysis,
screen-printed carbon paste electrodes measured in small volumes can be carried
for field titration in portable devices. Major anionic surfactants found in cleansing
products, alkyl sulfates and alkylbenzenesulfonates, and CPC and BAK in phar-
maceutical formulations are the most common analytes titrated using surfactant
selective electrodes. Similarly to ionic surfactants, nonionic surfactants were also
analyzed by PVC membrane electrodes containing a complex of nonionic surfactants
with metal cations as electroactive material. Interferences between surfactants as
that detected in titration of SLS in presence of CAPB had to be controlled.
7.1 INTRODUCTION
Potentiometric ion-selective electrodes (ISEs) are one of the most important groups
of chemical sensors. The application of ISEs has evolved to a well-established routine
analytical technique in many fields, including clinical and environmental analysis,
physiology, and process control. The essential part of ISEs is the ion-selective
membrane that is commonly placed between two aqueous phases, i.e., the sample
and inner solutions that contain an analyte ion. The membrane may be a glass, a
crystalline solid, or a liquid (1). The potential difference across the membrane is
measured with two reference electrodes positioned in the respective aqueous phases
Under equilibrium conditions, the measured potential (or emf of the cell), E, can be
expressed as
(7.1.1)
where zI is the charge of the analyte ion, I, aIw is its activity in the sample solution,
and the constant term, E10, is unique for the analyte and also includes the sum
of the potential differences at all the interfaces other than the membrane/sample
solution interface. This well-known “Nernst” equation for ISEs represents their
unique response properties, i.e., Nernstian responses, where the sensor signal,E ,
is proportional to logarithm of the analyte activity rather than the activity itself. The
slope in an E versus lnaIw (or more commonly logaIw) plot is used for identification
of the analyte based on the charge. A wide range of the analyte activity can be
determined because of the logarithmic dependence of the potential on the analyte
activity. Moreover, a very low analyte activity can be determined only by measuring
the potential difference rather than detecting a very small signal. A well-known
example is a glass membrane pH electrode, which has a detection limit of down
to 10–12 M for H+.
How can we create such a membrane for a wider range of analytes? The most
successful approach is to use ion-selective liquid membranes (2, 3). The liquid mem-
branes are hydrophobic and immiscible with water, and most commonly made of
plasticized poly(vinyl chloride). The selectivity is achieved by doping the membranes
with a hydrophobic ion (ionic site) and a hydrophobic ligand (ionophore or carrier)
that selectively and reversibly forms complexes with the analyte (Figure 7.1). Whereas
the technique has been well established experimentally since the 1960s, it is only re-
cently that the response mechanisms are fully understood. In this chapter, principles
of liquid membrane ISEs will be introduced using simple concepts of ion-transfer
equilibrium at water/liquid membrane interfaces. Non-equilibrium effects on the
selectivity and detection limits will also be discussed. This information will enable
practitioners of ISEs to better optimize experimental conditions and also to interpret
data. Additionally, examples of ISEs based on commercially available ionophores
are listed. More comprehensive lists of ionophore-based ISEs developed so far are
available in recent IUPAC reports (4–6).
Figure 7.1. Schematic view of the equilibrium between sample, ion-selective mem-
brane, and inner filling solution (cell 1). The cation-selective membranes are based
on (A) cation exchanger (R−), (B) electrically neutral ionophore (L) and anionic sites
(R−), and (C) charged ionophore (L−) and cationic sites (R+). The aqueous solutions
contain an analyte cation (I+) and its counter anion (X−). Adapted from reference (2).
Magnesium Silicate
Iyad Rashid, ... Adnan A. Badwan, in Profiles of Drug Substances, Excipients and
Related Methodology, 2011
7 Analytical Performance
The properties of an ion-selective electrode are characterized by parameters such as
linearity, limit of detection (LOD), limit of quantification (LOQ), sensitivity, selectiv-
ity, reproducibility, accuracy, robustness, response time, and lifetime.
7.2 Sensitivity
Providing that some conditions are gotten together, the theoretical slope (sensitivity)
of the calibration curve of an ISE toward a given singly charged ion, at 25 °C, should
be 59.2/z mV/decade, as formulated by the Nernst equation. The ISE membranes
must have a sufficient perm-selectivity. For the following two situations, a
non-Nernstian may be seen:
Surfactants and proteins, and other species that may be present together with the
target ions in pharmaceuticals and biological fluids, can also interfere with the
normal ion-extraction process resulting in nontheoretical (non-Nerstian) behaviors.
TABLE 11.4. Composition of the MIP-Based Hydroxyzine CPEs and Their Poten-
tiometic Response Characteristics, Measured in Phosphate Buffer Solutions (pH 2.0)
7.3 Selectivity
The potentiometric selectivity coefficients can be determined by different proce-
dures, namely, the so-called fixed interference method (FIM), the separate solution
method (SSM), and the matched potential method (MPM). The coefficients
describe the preference of the suggested electrode for an interfering ion, X, with
reference to the analyte ion. The IUPAC recommended the use of SSM and FIM,
two different procedures to determine the Nicolskii coefficients of ISEs [54]. The
SSM involves the measurement of two separate solutions, each containing a salt
of the determined ion only. The Nicolskii coefficient is then calculated from the
two observed emf values. In the FIM, an entire calibration curve is measured for
the analyte ion in a constant interfering ion background, aj(BG). The linear response
curve of the electrode as a function of the analyte ion activity is extrapolated until,
at the lower detection limit aI(DL), it intersects with the observed potential for the
background alone. The FIM coefficient is calculated from the resulting the lower
detection limit to the interfering ion activity ratio (Eq. 11.2):
(11.2)
The MPM was introduced in the mid-1980s by Gadzekpo and Christian to provide
a selectivity formalism that would give empirically more meaningful results [55].
According to the MPM method, the specified activity (concentration) of the primary
ions is added to a reference solution, and the potential is measured. In another
experiment, the interfering ions (X) are successively added to an identical reference
solution, until the measured potential matches that obtained before the addition of
the primary ions. The MPM selectivity coefficient, , is then given by the ratio of
resulting primary ion activity (concentration) to the interfering ion activity increases
in the two experiments (Eq. 11.3):
(11.3)
The MPM and FIM selectivity coefficients for the hydroxyzine ion-selective elec-
trode at the constant pH value of 2.0 are listed in Table 11.5 [12]. As is obvious from
this table, when the MIP sensor is applied to determine hydroxyzine, all the other
substances (except for cetirizine) hardly interfere with the determination. In most
cases, the selectivity coefficients are on the order of 5.0 × 10−4 and lower. The MPM
selectivity sequence of the employed MIP for different drugs and organic materials
(Fig. 11.9) approximately obeys the order: cetirizine > piperazine > pipyridine >
tri-ethyl ammonium chloride > promethazine > salbutamol ≈ metochlorpramide
> pyrrole ≈ aniline. As a result, the MIP molecular recognition is based both on
the template molecular structure (shape) and on the interactions between the print
molecule and the imprinted polymer. Cetirizine is of a similar chemical structure,
in which an acetic acid (-CH2-COOH) moiety presents instead of the ethanol (-CH2-
–CH2-OH) moiety in the aliphatic chain. This behavior confirmed that this molecule
could interfere in the hydroxyzine detection.
FIGURE 11.9. Structure of the investigated drugs and organic materials as interfer-
ences in MIP-based hydroxyzine sensor.
7.5 Lifetime
The lifetime of the MIP-based electrode can be studied by periodically recalibrating
the potentiometric response to the target drug in the standard solutions. After the
conditioning step, the electrode was repeatedly calibrated three times every month,
and any change in the electrode performance was recorded.
The average lifetime for most of the reported PVC-plasticized membrane electrodes
is in the range of 1–2 months [39,40,42–46]. After this time, a slow deterioration
of selectivity and response will be observed. It is well established that the loss of
plasticizer or ionic site from the polymeric film due to leaching into the sample
is a primary reason for the limited lifetimes of ISEs [33]. From this point, the
chemically modified electrodes were found to have significant advantages over their
corresponding plasticized PVC membranes. Since these electrodes did not contain
any special plasticizer or membrane solvent, they were more durable and less toxic
than the plasticized PVC membranes to be applied to ion sensors.
TABLE 11.6. Recovery Results for the Hydroxyzine Potentiometric Sensor in Urine
and Serum Samples
For the evaluation of reproducibility, the analytical method must be repeated several
times. The precision of the procedure can be reported in terms of relative standard
deviation. The analytical performance of the hydroxizine potentiometric sensor was
evaluated with five repeated potentiometric measurements of the 1.0 × 10−5 M hy-
droxizine solution in different laboratories. The precision of the described procedure
in terms of relative standard deviation was 6.5% [12].
7.7 Robustness/Ruggedness
Robustness is defined as the persistence of a system’s characteristic behavior under
perturbations or conditions of uncertainty. In other hands, robustness is a measure
of its capacity to remain unaffected by small but deliberate variations in the analytical
procedure parameters. A robust chemosensor capable of specifically sensing the
presence of chloropropanols in complex sample matrices with adequate sensitivity
was reported by K.P.L. Mitch and co-workers [57]. This MIP-based sensor showed
good selectivity for 3-chloro-1,2-propanediol (3-MCPD) since both 3-MCPD and
1,3-DCP often coexist as contaminants in epoxy resins used in the paper industry
and in food products.
The correct interpretation of the relation between the activity and concentration
requires a proper choice of experimental conditions and a careful interpretation
of the results. The most common procedure involves the use of an ionic strength
adjustment solution (buffer). It consists of an inert salt at constant concentration,
that which does not interact with the analyte (e.g., KNO3), providing a solution of
constant ionic strength, not influenced by small changes of the composition of
the sample solution. Usually the other components have additional functions. For
example, in total fluoride determination, such as Al3 + or Fe3 +, which bind fluoride
ions, should be complexed with cyclohexane-1,2-diamine-N,N.N ,N -tetraacetic
acid to liberate fluoride ions. In determination of total calcium content, controlled
complexation of the analyte by addition of iminodiacetic acid eliminates the ef-
fects caused by the presence of other weaker complexants of various strength and
concentration. Usually such an ionic strength adjustment solution also contains a
pH-buffer, which provides optimal pH value for the ISE used. Under conditions
of constant ionic strength there exists an exact correlation between the potential
reading and the logarithm of concentration of the analyte ion according to the
Nernst equation.
In principle the ion-selective electrodes are sensitive only to free (hydrated) ions of
the analyte and do not respond to various complexed species of the analyte. This
property could be exploited in the speciation analysis, distinguishing the free and
bound analyte species. However, it must be remembered that most procedures of
sample pretreatment: dilution, extraction, addition of reagents, change the initial
speciation, corresponding to the nontreated sample. Therefore, only when the
species are kinetically inert, and the sample pretreatment is carried out with special
caution is there a possibility of analyte speciation. For example, calcium in milk is
present as a free (hydrated) ion or in combination with proteins, lipids or larger
molecular mass species. The evaluation of each of the species is the aim of speciation
analysis. The task is relatively simple when the species investigated are inert, do not
dissociate, or otherwise undergo decompose during sample pretreatment. A similar
problem is common in analysis of other biological samples, e.g., in blood analysis,
where even 20-fold dilution results in nearly complete dissociation of the calcium
complexes with various ligands. Similarly the measurements with iodide selective
electrode does not respond to iodine bound to organic matrix.
Potentiometric (as well as other electrochemical) measurements have the important
advantage over optical measurements in that the color or opacity of the sample
solution do not obscure the proper determination of the analyte. It was found that
in many instances the ISE measurements give superior precision than spectropho-
tometric determinations. On the other hand, functioning of the electrodes in the
presence of colloids and suspensions may also produce erroneous results. This is
described as a suspension effect that is due to blocking the electrode surface or the
porous plug connecting the reference electrode. This requires frequent checking the
proper functioning of the electrode system, or even using special electrode systems
to avoid troubles. Measurements using flow systems often diminish such effects.
The main limitation of the use of ISEs in practical analysis is in the range of analyte
determination and the selectivity of the electrodes. Most commercially available
electrodes allow precise measurements of the analyte down to 10− 5–10− 6 mol l− 1
concentration. For some crystalline electrodes this may be shifted down by some
orders of magnitude, but only in the case when the species measured remain in
an labile equilibrium in an ion (e.g., metal ion) buffer. Obviously measurements in
the lowest concentration range suffer all the difficulties typical for trace analysis
(e.g., contamination), which may affect the determination. The concentration range
of analyte that may be measured using ISEs is usually between 10− 2 and 10− 4 mol l− 1,
and the sample size and subsequent dilution should be adjusted to those conditions.