You are on page 1of 8

DIABETES TECHNOLOGY & THERAPEUTICS

Volume 20, Number 1, 2018


ª Mary Ann Liebert, Inc.
DOI: 10.1089/dia.2017.0287

ORIGINAL ARTICLE

Flash Glucose Measurements in Children


with Type 1 Diabetes in Real-Life Settings:
To Trust or Not to Trust?
Agnieszka Szadkowska, MD, PhD,1 Andrzej Gawrecki, MD, PhD,2
Arkadiusz Michalak,1 Dorota Zozulińska-Zió1kiewicz, MD, PhD,2
Downloaded by Gothenburg University Library from online.liebertpub.com at 12/14/17. For personal use only.

Wojciech Fendler, MD, PhD,1,3 and Wojciech M1ynarski, MD, PhD1

Abstract
Background and Aims: To evaluate the clinical accuracy of a flash glucose monitoring device FreeStyle Libre
(FSL) among children with type 1 diabetes in real-world settings during a summer camp.
Materials and Methods: During a summer camp, children with type 1 diabetes (n = 79, aged 8–18 years) were
provided with FSLs for 12 days. On days 3, 7, and 11 of the study, they underwent supervised glucose testing at
8 timepoints. Glycemia was estimated by using FSL and measured with a personal glucometer within a period
of 2 min. The glucose trend arrows were recorded.
Results: The study was completed by 78 children (median: age 12.8 years, diabetes duration 5.8 years, HbA1c
58.5 mmol/mol). Mean absolute relative difference (MARD) between the FSL and the glucometer was
13.5% – 12.9%. FSL was the most accurate in stable glycemic conditions: MARD 11.4% – 10.4%, less accu-
rate when glycemia was falling >2 mg/(dL$min) [0.111 mmol/(L$min)—MARD 22.6% – 18.6%; P < 0.001 vs.
stable conditions] and when the device could not determine the glucose trends (16.5% – 16.3%, P = 0.01 vs.
stable conditions). The FSL demonstrated lower accuracy during the day than the night [MARD 14.9% – 14%
vs. 11.2% – 10.6%, P < 0.0001]. Out of 1655 data pairs of glucometer and FSL, using the Surveillance Error
Grid methodology we determined that 80.36% of FSL readings were associated with no clinical risk, 18.73%
with slight risk and only one high-risk measurement was detected.
Conclusion: FSL is accurate in children, but its accuracy depends on the glucose trend. Results flagged by the
rapid fall flag and ‘‘trend undetermined’’ should be verified by blood glucose measurements.

Keywords: Type 1 diabetes mellitus, Blood glucose self-monitoring, Children, Flash glucose monitoring,
FreeStyle Libre, Accuracy.

Introduction of diabetic ketoacidosis.5 The guidelines of the American


Diabetes Association have recently increased the recommended

T he key aim of therapy in type 1 diabetes mellitus


(T1DM) is to safeguard patients from acute and chronic
complications and to improve health-related quality of life.
number of BG measurements, especially in patients with a
short duration of diabetes or long life expectancy6 who should
perform SMBG tests 6 to 12 times daily. Such a number is,
The DCCT and EDIC trials proved that lowering HbA1c however, difficult to achieve, especially for children and young
protects from vascular complications and cardiovascular people, for whom the invasiveness of SMBG is a significant
disease.1–3 barrier for self-monitoring.7
Maintaining a high frequency of self-monitoring of blood To improve this situation, other systems have been intro-
glucose (SMBG) is the cornerstone for meeting HbA1c goals duced to replace traditional capillary BG measurement. Con-
and preventing serious hypoglycemia.4 An increased num- tinuous glucose monitoring (CGM) offers efficient glycemic
ber of tests is also associated with significantly lower rates control with a reduced need for manual tests and associated
Departments of 1Pediatrics, Oncology, Hematology and Diabetology, and 3Biostatistics and Translational Medicine, Medical University
of Lodz, Lodz, Poland.
2
Department of Internal Medicine and Diabetology, Poznan University of Medical Science, Poznan, Poland.

1
2 SZADKOWSKA ET AL.

inconveniences. Recent studies have demonstrated that the The children had basic training in using FSL, limited to
use of CGM systems reduces glycemic variability and time device handling and explanation of trends and alerts. On the
spent in hypoglycemia of subjects with diabetes, and HbA1c, seventh day, the authors collected feedback from children
in well-controlled individuals and those using insulin and created initial clinical guidelines for the interpretation of
pumps.8–12 A health-economic analysis by Roze et al. found glucose trend arrows, which were subsequently taught to the
that sensor-augmented pump therapy represents good value children.
for money in terms of improving quality-adjusted life ex- On the 3rd, 7th, and 11th days of FSL use, the children
pectancy and delaying diabetes complications13; however, underwent supervised glucose testing at eight predefined time
the high cost of CGM still hinders its widespread use. points each day: before and 2 h after breakfast, before and 2 h
Recently, a new flash glucose monitoring (FGM) system after their midday meal, before and 2 h after dinner, at mid-
(Abbott Diabetes Care, Alameda, CA) was introduced into night, and at 3 a.m. At each timepoint, glycemia was mea-
the market along with a dedicated device: FreeStyle Libre sured twice within 2 min: once in the interstitial fluid by
(FSL). It offers a factory-calibrated sensor that can be worn for scanning the FGM sensor, and again in the capillary blood by
14 days without manual calibration by using blood testing, a using a Contour Plus One glucometer (Ascensia Diabetes Care
reader combined with a standard glucometer, and dedicated Poland, Warszawa, Poland), compliant with ISO 15197:2013
software. The user receives a current glucose measurement accuracy criteria.15 The glucose trend arrows of the FGM were
along with historic results from the preceding 8 h by scanning recorded. The sensors were removed on the 12th (last) day of
Downloaded by Gothenburg University Library from online.liebertpub.com at 12/14/17. For personal use only.

the sensor with the FreeStyle reader. Recordings are taken in the camp, and the skin was carefully examined by physicians.
15-min intervals. The glucose trends are displayed as arrows After sensor removal, data from the readers were downloaded
on the reader along with alerts. As the device requires active and analyzed. Each reader was also reviewed manually.
scanning by the user, it does not report alerts in real time. Medical history was collected, and physical examination
FSL was proposed as an alternative for those who cannot was performed by pediatricians. At the end of the camp, body
afford CGM, find CGM too complicated, or complain of alarm height and weight were measured and capillary blood sam-
fatigue. Children and adolescents could clearly benefit from ples were taken for HbA1c assessment (D-10 Hemoglobin
using the FGM system.14 However, initial observations sug- A1c Program [Bio-Rad Laboratories, Hercules, CA, Bio-
gested that its users may forego SMBG measurements and rely Rad, Marnes-la-Coquette, France]). Body mass index (BMI)
solely on FGM. This raises the question—Is FGM accurate was recalculated into standard deviation Z-scores based on
enough to replace standard glucometers, or should it be just a local BMI charts.16
complementary addition to SMBG? So far, only one study On the last day, a custom-designed questionnaire was ad-
tested the accuracy of FSL in a pediatric population, and the ministered to assess the degree of satisfaction with FGM use.
data obtained by FGM were hidden from participants.14 The children rated their experience and satisfaction with FSL
This study examines the introduction of FSL to children on a five-point scale from strongly agree to strongly disagree.
with T1D in a real-world setting during a summer camp. During The questionnaire included 13 closed questions regarding
the study, the children could observe current glucose levels, trend ease of using the device, pain when applying the sensor to the
arrows, and the 8-h glucose concentration history on a reader arm, the comfort of wearing the sensor, possible inconve-
screen. The main aim of this study was to establish the accuracy nience, whether the sensor got in the way of daily activities,
of FSL FGM system in reference to the glucometer and pinpoint and the occurrence of pain or itching at the site of insertion.
the factors that may affect said accuracy. The secondary objec- The detailed questionnaire is included in the Supplemen-
tive was to assess the children’s attitude toward the device and its tary Data (Supplementary Data are available online at http://
possible association with measurement accuracy. online.liebertpub.com/doi/suppl/10.1089/dia.2017.0287). The
users were also asked about their subjective opinion regard-
ing the decrease in the number of hypo- and hyperglycemia
Materials and Methods
episodes during the use of FGM. In the final open question,
This was a prospective single-arm study conducted during an the children were asked to describe the advantages and dis-
annual summer camp for children and adolescents with T1D. advantages of the device.
The study protocol was approved by the Ethical Committee of
the Medical University of Lodz (No. RNN/223/16/KE) and
Statistical analysis
registered in the German Clinical Trials Register (DRKS-ID:
DRKS00011751). Informed written consent was obtained from FSL readings were compared with glucometer readings as
parents, and so was informed assent from the children. the reference method. FSL readings beyond the sensor value
Children from 12 diabetes care centers from Poland took part range (<40 mg/dL reported as LO (low) and >500 mg/dL
in the camp, which took place in the period from July 24 to reported as HI (high)) were included in the analysis as 40 and
August 5, 2016. They were provided with full medical care 500 mg/dL, respectively. These extreme readings, however,
and asked whether they would like to use FSL. Out of 80 ap- were excluded from the analysis of glycemia trends, as such
proached children, 79 accepted the opportunity to participate in recordings had no defined trend mark by default. The accu-
the study. One child did not complete the study. FSL readers racy was assessed by calculating the mean absolute relative
and sensors were provided by Abbott Laboratories Poland LLC. difference ( MARD) for each patient by using the fol-
On the second day of the camp, the children were introduced lowing equation: MARD = meanj[(glucometer-FreeStyle
to the FGM system and the sensors were attached to the back Libre)/glucometer) · 100%]j.
of their upper arms, according to the manufacturer’s instruc- The Pearson’s test was used to determine the correlation of
tions. In the case of the sensor dislodging or any technical the two methods, and the Bland–Altman regression was used
issues, the sensors were replaced in each child up to two times. to assess bias (mean difference was compared with 0 by using
ACCURACY OF FREESTYLE LIBRE IN CHILDREN 3

the t-test). The clinical accuracy of the readings was tested by


constructing a Clarke Error Grid, a Consensus Error Grid, and
a Surveillance Error Grid for paired FGM and glucometer
measurements.17–19 To determine the impact of clinical vari-
ables, t-tests were used for two-group comparisons, one-way
analysis of variance was used for multiple groups (with the
post hoc Tukey’s test performed for significant results), and the
Spearman’s rank correlation was used for continuous variables
that were not distributed normally.
The questionnaire assessing satisfaction with FSL was
assessed by contingency tables. The patients were divided
into subgroups depending on the answers given in the ques-
tionnaire, and their FGM accuracies were compared between
the groups by using the Kruskal–Wallis test.

Results
Downloaded by Gothenburg University Library from online.liebertpub.com at 12/14/17. For personal use only.

A total of 79 children agreed to use FSL for the duration of FIG. 1. FreeStyle Libre measurements compared to BG
the summer camp; however, one girl returned home before with Bland–Altman regression plot. BG, blood glucose.
study completion and was excluded from the analysis. The
analyzed group, therefore, numbered 78 children (43% boys,
88.6% treated with CSII) with a median age of 12.8 years differences related to glucometer reference measurements pro-
(interquartile range [IQR]: 11.6–14.7 years) and a median duced a mean relative difference of -4.42% with a high standard
T1D duration of 5.8 years (IQR: 3.8–8.5 years). Their me- deviation (17.87%). The MARD between FSL and the gluc-
dian HbA1c at inclusion was 7.5% (IQR: 7.0%–8.0%) or ometer was 13.5% – 12.9%. The MARD for each patient ranged
58.5 mmol/mol (53–63.9 mmol/mol). The median standard- from 10.4% to 24.7%. The distribution of MARDs by indi-
ized BMI in the group was 0.29 (IQR: -0.3 to 0.89), four viduals is presented in Figure 2. The accuracy of FGM did not
children were overweight (BMI between 90th and 95th per- differ between days of sensor use (P = 0.11) and was not
centile standardized for gender and age), and two children associated with BMI Z-score (R = -0.11, P = 0.31) or HbA1c
were obese (>95th percentile). (R = 0.07, P = 0.55). However, a weak positive correlation
During the summer camp, children wore the sensors for was found between age and MARD (R = 0.25, P = 0.024), and
median time of 10.5 days (IQR 10 to complete 11 days) and the sensors worn by boys displayed a lower accuracy than
10 sensors were replaced in eight children due to detachment. those worn by girls—median MARD 14.8% (IQR: 12.4%–
The analysis included 1655 paired sensor/glucometer 17.3%) versus 12% (9.9%–13.6%), P = 0.0001.
readings (median 23 pairs/patient, IQR: 21–24). Due to the In addition, the accuracy of FSL strongly depended on
need for sensor replacement, it was not possible to collect all the current trend of glucose concentration (P < 0.0001)
measurements scheduled for the 3rd, 7th, and 11th day of (Fig. 3). FSL measurements were the least accurate (MARD
FGM. In total, 668 readings were collected from the first 22.6% – 18.6%, mean relative difference -17.23% – 23.73%)
timepoint, 539 from the second, and 448 from the third. The when glycemia was falling at a rate exceeding 2 mg/(dL$min)
glycemic variability in the participants over the camp dura- [0.111 mmol/(L$min)] compared with other trends (P < 0.0001).
tion is presented in Table 1. The greatest accuracy (MARD 11.4% – 10.4%, mean relative
FSL readings showed a strong positive correlation with glu- difference -2.32% – 15.3%) was achieved in stable condi-
cose meter measurements (r = 0.95, P < 0.0001). However, the tions, when glycemia changes did not exceed 1 mg/(dL$min)
Bland–Altman regression (Fig. 1) showed that the mean dif- [0.056 mmol/(L$min)] in either direction. Moreover, when the
ference (7.35 mg/dL, 0.4 mmol/L) was significantly different reading device could not determine the glucose trends and
from the expected value of 0 (P < 0.0001), suggesting that FSL displayed no glucose trend arrow, the accuracy of FSL (MARD
was biased toward an overestimation of glycemia. In line, raw 16.5% – 16.3%, mean relative difference -6.72% – 22.28%)
was significantly inferior to measurements performed in stable
Table 1. Parameters of Glycemic Variability glycemic conditions (P = 0.001).
in the Studied Children During Flash Glucose We detected a single extreme point in our data (absolute
Monitoring Use relative difference of 125%, no glucose trend arrow) that may
Mean glycemia 167 mg/dL, 9.27 mmol/L have caused overestimation of the disparity between ‘‘unable
[mg/dL, mmol/L] to determine tendency’’ and ‘‘stable glycemic conditions.’’
% of time in range 37.7 Therefore, we removed this outlier and repeated the ANOVA
(70–140 mg/dL, analysis, which yielded significant results similar to the
3.9–7.8 mmol/L) original ones (data not shown).
Time above range [%] 55.4 Further, daytime measurements taken by FSL were found
Time in hypoglycemia [%] 8.2 to be less accurate than those taken at nighttime (i.e., mid-
Episodes of hypoglycemia 1.35 – 0.93 night, 3 a.m. and after waking up) [MARD 14.9% – 14% vs.
per day of use [N] 11.2% – 10.6%, P < 0.0001].
Average duration of 77 – 27 min
hypoglycemia episode Out of 1655 data pairs, 98.43% satisfied the clinical ac-
curacy criteria (class A or B) of the Clarke Error Grid (Fig. 4a
4 SZADKOWSKA ET AL.
Downloaded by Gothenburg University Library from online.liebertpub.com at 12/14/17. For personal use only.

FIG. 2. A bar graph of MARD values of individual patients. MARD, mean absolute relative difference.

and Table 2) and 99.1% satisfied that of the Consensus Error Consensus Error Grid was similar among the patients (Sup-
Grid (Fig. 4b and Table 2). The Surveillance Error Grid plementary Figure S1).
(Fig. 4c and Table 3) labeled 1330 measurements (80.36%) as No serious adverse effects were found to be associated
associated with no risk and 310 (18.73%) as associated with with FSL sensor use during the study. In two boys, edema and
slight risk. The distribution of records in different classes of the rash were observed after sensor removal. During the camp,

FIG. 3. Accuracy of FreeStyle Libre in different glycemic conditions. The glycemic trends were determined by in-built
software in the FreeStyle Libre reading device.
ACCURACY OF FREESTYLE LIBRE IN CHILDREN 5
Downloaded by Gothenburg University Library from online.liebertpub.com at 12/14/17. For personal use only.

FIG. 4. Error Grid Analysis of FreeStyle Libre results plotted against BG concentrations with: (a) Clarke Error Grid.
(b) Consensus Error Grid. (c) Surveillance Error Grid (graph prepared through the courtesy of Prof. Boris P. Kovatchev,
Prof. Marc D. Breton, Dr. David C. Klonoff, and Mr. Christian Wakeman). The codes and definitions for each class are
summed up in Tables 2 and 3.

none of the children reported feeling pain or itching. How- open question regarding the advantages and disadvantages of
ever, in the retrospective questionnaire, 4 reported pain on the device, most of the participants regarded the discretion
sensor insertion, 13 reported itching, and 2 reported moderate and painlessness of measurements as key advantages, and its
pain during sensor use. inaccuracy, as perceived by the children, as a disadvantage.
Sixty-eight (87.2%) children expressed a willingness to Interestingly, sensor accuracy was not clearly associated
continue FGM use (see Supplementary Figure S2). In the with the accuracy of FSL, as perceived by children (P = 0.72),

Table 2. Summary of Clarke Error Grid and Consensus Error Grid Analysis
Clarke Error Grid Consensus Error Grid
Coding Definition N (%) Definition N (%)
Class A B Within 20% of reference value 1329 (80.3) Clinically accurate, no effect 1460 (88.22)
on clinical action
Class B  Outside of 20% range of 300 (18.13) Possibly causing an altered 180 (10.88)
reference value but not leading clinical action but with little
to inappropriate treatment or no effect on clinical outcome
Class C > Leading to unnecessary treatment 9 (0.54) Likely to affect the clinical outcome 14 (0.84)
Class D A Indicating a potentially dangerous 16 (0.97) Associated with significant risk 0 (0)
failure to detect hypoglycemia
or hyperglycemia
Class E
* Would confuse treatment of
hypoglycemia for hyperglycemia
1 (0.06) Having possible dangerous
consequences
1 (0.06)
6 SZADKOWSKA ET AL.

Table 3. Summary of Surveillance Error however, as no objective measure of physical activity was
Grid Analysis included in our study group, this is impossible to confirm.
FSL accuracy was also found to be associated with the
Degree of risk Color coding N (%) current glycemia trend: When glucose was falling rapidly, the
None Deep green 1330 (80.36) FSL reading differed by up to 20% from BG measurements.
Slight, underestimated Light green 265 (13.01) This corresponds with results of Bonora et al., who report a
Slight, overestimated Yellow 45 (2.72) significant increase in the MARD between FGM and CGM
Moderate, underestimated Light orange 14 (0.95) during hypoglycemic conditions.22 Moreover, when glycemia
Moderate, overestimated Dark orange 0 (0) was falling by more than 2 mg/(dL$min) [0.111 mmol/(L$min)],
Great, underestimated Light red 1 (0.06) the mean difference reached -20 mg/dL (-1.1 mmol/L), which
Great, overestimated Dark red 0 (0) caused the FGM unit to overestimate the true current glycemia
Extreme Brown 0 (0) value. This discrepancy may have been caused by the lag time
known to exist between glucose concentrations in the inter-
stitial fluid reaching those in the blood.20 This phenomenon
their trust in the sensor readings (P = 0.33), or their willing- was first described by Aussedat et al.23 in rats and has become
ness to continue using FSL (P = 0.52). one of the focal points in theorycrafting of artificial pancreas.
It complicates the interpretation of glucose concentrations
Downloaded by Gothenburg University Library from online.liebertpub.com at 12/14/17. For personal use only.

measured in the interstitial fluid as the delay must be accounted


Discussion
for in closed-loop algorithms and by users of FGM and CGM
The study examines the utility of FSL in children and ad- devices who want to make therapeutic decisions.
olescents with T1D during a summer camp. FSL readings were Importantly, the lag is unfortunately no constant difference
found to correlate closely with BG measurements. So far, only but its magnitude, direction, and dynamics may change, de-
three studies have analyzed the accuracy of FSL in real-life pending on glucose uptake, utilization, and elimination in
conditions. In adults, Bailey et al. observed MARDs as high as peripheral tissues.24 In humans, physiological lag between
12% in reference to a glucometer and 12.1% in reference to venous and interstitial glycemia was measured to be around 6
venous glycemia measured with YSI.20 Similarly, Ji et al. re- to 10 min in adults without25 and with type 1 diabetes.26 So
ported MARDs of 10% in reference to BG measurements and far, no study has investigated the physiology of the glycemic
of 10.7% in reference to venous measurements.21 Edge et al. lag in children. The tendency of FGM to overestimate falling
reported a slightly higher MARD in children (13.9%), which glycemia fits well with the ‘‘push and pull’’ hypothesis
corresponds well with our findings.14 formulated by Aussedat et al.23 and their observation that
We were also the first to evaluate FGM readings system- during insulin-driven hypoglycemia the interstitial glucose
atically with the Bland–Altman analysis, in which FSL concentrations fall slower than intravenous ones.
demonstrated a fixed tendency to overestimate glucometer- Moreover, FSL demonstrated significantly lower accuracy
measured glycemia. This bias was small (mean 7.35 mg/dL, (compared with stable conditions), when the reading device
0.4 mmol/L) and negligible in most clinical situations, but on could not determine the glycemic trend. This and the rapid
very rare occasions it may lead to the wrong clinical deci- decline scenario are, therefore, indicators for the patient to
sions. For this reason, all collected data pairs were assessed verify the result with capillary glucose measurement by using
against published error grid scores17,18: the Clark, Consensus, a glucometer. These results are in contrast to those reported
and Surveillance Error grids. The most recently developed by Edge et al.,14 who report a steady relative difference of
grid is the Surveillance Error Grid, which was developed in 6%–7% between FSL readings and BG under extreme glu-
2014. It was constructed by using feedback from medical cose trends (rapidly falling and rapidly rising) but negligible
professionals who represent a modern approach to diabetes differences between the two methods (around 1%) for other
care, based on DCCT study results and incorporation of trends. However, their study examined children using FSL in
technological advances such as insulin pumps into therapy. home-based conditions, whereas this study evaluates children
Our study demonstrated that most FSL readings were lo- during a summer camp: Increased physical activity, irregular
calized in zone A or B of the error grids for more than 98% of eating patterns, and other factors may have contributed to the
cases, which means that they would not impact clinical de- differences between the reported results and the ones shown
cisions in a severe manner. Other studies have reported in our study. These conditions might make both studies in-
similar values regarding grid-assessed clinical accuracy for comparable in this regard.
FGM versus glucometer in adult20,21 and pediatric popula- In the studies regarding CGM, the main barriers to con-
tions.14 A single incidence of significant disparity was found tinue using them were unrealistic expectations of users and
between an FSL reading (<40 mg/dL, <2.2 mmol/L, reported lack of proper education.27 In our study, the children were
as LO) and BG measurement (215 mg/dL, 11.9 mmol/L), asked to assess the usefulness of FSL. In general, children
which could be associated with potential clinically important graded the 2-week-long experience with FSL positively.
consequences (class E by both Clark and Consensus Error Most of the children expressed willingness to continue using
Grid). This record was associated with sensor failure. this device. It is possible that the good reception of FGM may
FGM accuracy did not correlate with the patients’ BMI or result from its minimal invasiveness, when compared with
HbA1c, which was in line with results reported by Bailey and standard BG measurements, and added value in the form of
by Edge.20 However, in our study, MARD values were easily interpretable glucose trend arrows.
influenced by sex and age: The accuracy was a little worse in As noted in other studies, no serious adverse effects were
the boys. This may be associated with the higher physical observed while using FSL.14,20 Skin allergic reactions were
activity demonstrated by the boys during the summer camp; only found in fewer than 3% of patients.
ACCURACY OF FREESTYLE LIBRE IN CHILDREN 7

The limitation of our study is its short (12 days) duration. camp organizers (charity foundation ‘‘Diabeciaki’’), but Ab-
Studies with longer observation times are needed to evaluate bott Laboratories Poland had no involvement in the study
whether FSL could provide lasting improvement in glycemic design and course. A.S. reports being a speaker for Abbott
outcomes. Moreover, our results are limited by the high rate Laboratories Poland and Bayer. A.G. reports being a speaker
of sensor removal. More effective means are needed to for Abbott Laboratories Poland and Bayer. A.M., D.Z.-Z.,
safely fix the sensor in place for studies in children. W.F., and W.M. report no conflict of interest.
A minor limitation was also comparing FSL readings only
with glucometer measurements. However, use of the labora- References
tory reference method during the camp would demand draw-
1. Writing Group for the DCCT/EDIC Research Group TJ,
ing venous blood in a nonsterile environment and would create
Orchard TJ, Nathan DM, et al.: Association between 7
logistic problems with storage and transport of samples. years of intensive treatment of type 1 diabetes and long-
Similarly, using the Hemocue system for reference was not term mortality. JAMA 2015;313:45–53.
possible in this study for a few reasons. First, although the 2. de Boer IH, Rue TC, Cleary PA, et al.: Long-term renal
Hemocue microcuvettes can be stored in room temperature, outcomes of patients with type 1 diabetes mellitus and
the measurements were often taken in the outdoor conditions microalbuminuria: an analysis of the Diabetes Control and
in temperatures exceeding 25C; the microcuvettes are highly Complications Trial/Epidemiology of Diabetes Interventions
moisture sensitive and could be easily damaged. Further, and Complications cohort. Arch Intern Med 2011;171:
Downloaded by Gothenburg University Library from online.liebertpub.com at 12/14/17. For personal use only.

the sampling size needed for Hemocue is around 4 lL of blood. 412–420.


We used the glucometer that required a considerably 3. Nathan DM, Cleary PA, Backlund J-YC, et al.: Intensive
smaller volume of only 0.6 lL to prevent children’s pain diabetes treatment and cardiovascular disease in patients
and discomfort and to ensure good compliance. Finally, the with type 1 diabetes. N Engl J Med 2005;353:2643–2653.
number of participants in our study would demand many 4. Miller KM, Beck RW, Bergenstal RM, et al.: Evidence of a
Hemocue analyzers to be used at once to carry out the testing. strong association between frequency of self-monitoring of
Finally, the comparison of FGM with reference laboratory blood glucose and hemoglobin A1c levels in T1D exchange
methods has been already done by the producer and other clinic registry participants. Diabetes Care 2013;36:2009–
researchers14,20,21—the added value of our study is that it 2014.
provides an opportunity to analyze the accuracy from a user’s 5. Ziegler R, Heidtmann B, Hilgard D, et al.: Frequency of
point of view. The reference-related error was minimized by SMBG correlates with HbA1c and acute complications in
using the Contour Plus One glucometer (Bayer HealthCare children and adolescents with type 1 diabetes. Pediatr Dia-
betes 2011;12:11–17.
LLC, Diabetes Care, Whippany, NJ) as a comparator: It has
6. American Diabetes Association: Standards of Medical Care
high accuracy with an MARD of below 10% in comparison
in Diabetes—2014. Diabetes Care 2014;37(suppl 1):S14–
with reference methods.15 On the other hand, such a protocol S80.
provided great insight into FGM accuracy from the users’ 7. Wagner J, Malchoff C, Abbott G: Invasiveness as a barrier
perspective, as most clinical decisions are based on SMBG to self-monitoring of blood glucose in diabetes. Diabetes
performed with a glucometer. Technol Ther 2005;7:612–619.
8. Juvenile Diabetes Research Foundation Continuous Glu-
Conclusions cose Monitoring Study Group, Beck RW, Hirsch IB, et al.:
The effect of continuous glucose monitoring in well-controlled
Results showed that the FSL FGM device offers good type 1 diabetes. Diabetes Care 2009;32:1378–1383.
accuracy in comparison to capillary BG in children and ad- 9. Danne T, de Valk HW, Kracht T, et al.: Reducing gly-
olescents with T1DM. However, the performance of FGM is caemic variability in type 1 diabetes self-management with
strongly affected by the glycemia change trend at the time of a continuous glucose monitoring system based on wired
measurement. Therapeutic decisions should not be solely enzyme technology. Diabetologia 2009;52:1496–1503.
based on measurements flagged by a rapid change flag on the 10. Hirsch IB, Abelseth J, Bode BW, et al.: Sensor-augmented
FGM: Such measurements should first be verified by using a insulin pump therapy: results of the first randomized treat-
BG measurement. to-target study. Diabetes Technol Ther 2008;10:377–383.
11. Battelino T, Phillip M, Bratina N, et al.: Effect of contin-
Acknowledgments uous glucose monitoring on hypoglycemia in type 1 dia-
betes. Diabetes Care 2011;34:795–800.
The authors would like to thank Prof. Boris P. Kovatchev, 12. Bergenstal RM, Tamborlane WV, Ahmann A, et al.: Ef-
Prof. Marc D. Breton, Dr. David C. Klonoff, and Mr. Christian fectiveness of sensor-augmented insulin-pump therapy in
Wakeman, who provided the tool for calculating the Sur- type 1 diabetes. N Engl J Med 2010;363:311–320.
veillance Grid Score risks, and the medical staff for col- 13. Roze S, Saunders R, Brandt A-S, et al.: Health-economic
lecting data during the summer camp. The authors would also analysis of real-time continuous glucose monitoring in
like to thank the organizers of the camp, charity foundation people with Type 1 diabetes. Diabet Med 2015;32:618–626.
‘‘Diabeciaki’’ for the opportunity to carry out the study and 14. Edge J, Acerini C, Campbell F, et al.: An alternative
for providing FSL sensors for children. sensor-based method for glucose monitoring in children
and young people with diabetes. Arch Dis Child 2017;102:
Author Disclosure Statement 543–549.
15. Bailey T, Wallace JF, Greene C, et al.: Accuracy and user
All authors took part in the study. None of the authors have performance evaluation of the Contour Next Link 2.4 blood
any competing financial interests with the content of the glucose monitoring system. Clin Chim Acta 2015;448:
study. The FSL sensors were provided by the producer to the 139–145.
8 SZADKOWSKA ET AL.

16. Nawarycz T, Ostrowska-Nawarycz L: [Body mass index in tinuous subcutaneous glucose monitoring. Am J Physiol
the school age children and youth from the city of Lodz]. Endocrinol Metab 2000;278:E716–E728.
Pol Merkur Lekarski 2007;23:264–270. 24. Kulcu E, Tamada JA, Reach G, et al.: Physiological differ-
17. Klonoff DC, Lias C, Vigersky R, et al.: The surveillance ences between interstitial glucose and blood glucose mea-
error grid. J Diabetes Sci Technol 2014;8:658–672. sured in human subjects. Diabetes Care 2003;26:2405–2409.
18. Parkes JL, Slatin SL, Pardo S, Ginsberg BH: A new con- 25. Basu A, Dube S, Slama M, et al.: Time lag of glucose
sensus error grid to evaluate the clinical significance of from intravascular to interstitial compartment in humans.
inaccuracies in the measurement of blood glucose. Diabetes Diabetes 2013;62:4083–4087.
Care 2000;23:1143–1148. 26. Basu A, Dube S, Veettil S, et al.: Time lag of glucose from
19. Clarke WL, Cox D, Gonder-Frederick LA, et al.: Evaluat- intravascular to interstitial compartment in type 1 diabetes.
ing clinical accuracy of systems for self-monitoring of blood J Diabetes Sci Technol 2015;9:63–68.
glucose. Diabetes Care 1987;10:622–628. 27. Kubiak T, Mann CG, Barnard KC, Heinemann L: Psy-
20. Bailey T, Bode BW, Christiansen MP, et al.: The per- chosocial aspects of continuous glucose monitoring: con-
formance and usability of a factory-calibrated flash glu- necting to the Patients’ Experience. J Diabetes Sci Technol
cose monitoring system. Diabetes Technol Ther 2015;17: 2016;10:859–863.
787–794.
21. Ji L, Guo X, Guo L, et al.: A multicenter evaluation of Address correspondence to:
the performance and usability of a novel glucose monitor- Agnieszka Szadkowska, MD, PhD
Downloaded by Gothenburg University Library from online.liebertpub.com at 12/14/17. For personal use only.

ing system in Chinese adults with diabetes. J Diabetes Sci Department of Pediatrics, Oncology,
Technol 2017;11:290–295. Hematology and Diabetology
22. Bonora B, Maran A, Ciciliot S, et al.: Head-to-head com- Medical University of Lodz
parison between flash and continuous glucose monitoring Sporna 36/50
systems in outpatients with type 1 diabetes. J Endocrinol Lodz 91-738
Invest 2016;39:1391–1399. Poland
23. Aussedat B, Dupire-Angel M, Gifford R, et al.: Interstitial
glucose concentration and glycemia: implications for con- E-mail: agnieszka.szadkowska@wp.pl

You might also like