You are on page 1of 7

Nutrition, Metabolism & Cardiovascular Diseases (2018) 28, 180e186

Available online at www.sciencedirect.com

Nutrition, Metabolism & Cardiovascular Diseases


journal homepage: www.elsevier.com/locate/nmcd

FreeStyle Libre and Dexcom G4 Platinum sensors: Accuracy


comparisons during two weeks of home use and use during
experimentally induced glucose excursions
F. Boscari a, S. Galasso a, A. Facchinetti b, M.C. Marescotti a, V. Vallone a, A.M.L. Amato a,
A. Avogaro a, D. Bruttomesso a,*
a
Division of Metabolic Diseases, Department of Medicine, University of Padova, Padova, Italy
b
Department of Informatics Engineering, University of Padova, Padova, Italy

Received 17 July 2017; received in revised form 12 October 2017; accepted 29 October 2017
Handling Editor: A. Giaccari
Available online 11 November 2017

KEYWORDS Abstract Background and aims: This study compared the accuracy of the FreeStyle Libre (Ab-
Type 1 diabetes; bott, Alameda, CA) and Dexcom G4 Platinum (DG4P, Dexcom, San Diego, CA) CGM sensors.
Continuous glucose Methods and results: Twenty-two adults with type 1 diabetes wore the two sensors simulta-
monitoring (CGM); neously for 2 weeks. Libre was used according to manufacturer-specified lifetime (MSL); DG4P
Flash glucose was used 7 days beyond MSL. At a clinical research center (CRC), subjects were randomized to
monitoring; receive the same breakfast with standard insulin bolus (standard) or a delayed and increased (de-
Accuracy; layed & increased) bolus to induce large glucose swings during weeks 1 and 2; venous glucose was
Hypoglycaemia; checked every 5e15 min for 6 h. Subjects performed 4 reference fingersticks/day at home. Ac-
Rate of change curacy was assessed by differences in mean absolute relative difference (%MARD) in glucose levels
compared with fingerstick test (home use) and YSI reference (CRC). During home-stay the Libre
MARD was 13.7  3.6% and the DG4P MARD 12.9  2.5% (difference not significant [NS]). With
both systems MARD increased during hypoglycaemia and decreased during hyperglycaemia,
without significant difference between sensors. In the euglycaemic range MARD was smaller with
DG4P [12.0  2.4% vs 14.0  3.6%, p Z 0.026]. MARD increased in both sensors following delayed
& increased vs. standard bolus (Libre: 14.9  5.5% vs. 10.9  4.1%, p Z 0.008; DG4P: 18.1  8.1% vs.
13.1  4.6%, p Z 0.026); between-sensor differences were not significant (p Z 0.062). Libre was
more accurate during moderate and rapid glucose changes.
Conclusions: DG4P and Libre performed similarly up to 7 days beyond DG4P MSL. Both sensors
performed less well during hypoglycaemia but Libre was more accurate during glucose swings.
Trial registration: The study was registered in ClinicalTrials.gov (NCT02734745) April 12, 2016.
ª 2017 Published by Elsevier B.V. on behalf of The Italian Society of Diabetology, the Italian Society
for the Study of Atherosclerosis, the Italian Society of Human Nutrition, and the Department of
Clinical Medicine and Surgery, Federico II University.

Introduction hypoglycaemia in individuals with type 1 diabetes (T1D)


[1e3]. Continuous glucose monitoring (CGM), offers real-
Good glycaemic control reduces the risk of micro- and time blood glucose values, alerts about impending hypo-
macro-vascular complications but increases the risk of glycaemia and hyperglycaemia, reduces HbA1c and has

* Corresponding author. Department of Medicine, University Hospital of Padova, Via Giustiniani 2, 35128 Padova, Italy.
E-mail address: daniela.bruttomesso@unipd.it (D. Bruttomesso).

https://doi.org/10.1016/j.numecd.2017.10.023
0939-4753/ª 2017 Published by Elsevier B.V. on behalf of The Italian Society of Diabetology, the Italian Society for the Study of Atherosclerosis, the Italian Society of Human
Nutrition, and the Department of Clinical Medicine and Surgery, Federico II University.
FreeStyle Libre and Dexcom G4 Platinum sensors 181

psychological benefits in T1D subjects treated with interstitial glucose levels every minute. By scanning the
continuous subcutaneous insulin infusion (CSII) and mul- reader over the sensor one obtains a real-time glycaemic
tiple daily insulin injection (MDI) [4e12]. Most glucose value, the glycaemic trend over the preceding 8 h, the
sensors, however, require calibration, have short life time, direction towards which glucose levels are moving and
measure glucose in the interstitial fluid and cannot always glucose rate of change. The sensor lasts 14 days.
replace blood glucose self-monitoring (SMBG). The Dexcom G4P consists of a small sensor inserted
The FreeStyle Libre Flash glucose monitoring system under the skin which measures subcutaneous glucose
(FGM, Abbott Diabetes Care, Alameda, CA) is factory- concentration every 5 min and sends readings to a
calibrated and has 14 days of wear lifetime but lacks receiver. Twice-daily calibration with fingerstick blood
real-time alarms when limit glucose values are exceeded. glucose is required. The manufacturer specified lifetime
It is intended as a replacement for the SMBG, except dur- (MSL) for the sensor is 7 days, however it is possible to
ing times of rapid glucose changes, when hypoglycaemia extend sensor use for a further week by reactivating it [15].
requires confirmation or when displayed glucose values
are not in accordance with symptoms. Procedures
A study evaluating the accuracy of Libre system at home
for 10e14 days [13] and another [14], comparing the ac- Patients were trained on the use of the two devices at the
curacy of Libre to that of two commercially available sen- CRC. Sensors were simultaneously placed on the back of
sors, found that Libre had comparable accuracy. To date, no one upper arm (Libre) and on the abdominal region (DG4P).
study has compared Libre’s performance with that of
comparators during rapid glucose swings or beyond sensor Home phase
certified wear lifetime. During at-home CGM use, subjects were instructed to
The most commonly used CGM sensor in Italy is Dex- perform SMBG at 4 times per day (before meals and at
com G4 Platinum (DG4P, Dexcom Inc., San Diego, CA), bed-time) using the blood glucose (BG) meter built into
which is provided by the health care system as adjunctive the hand-held Libre reader. Patients were instructed to
to SMBG. The Libre system is purchased by patients or scan the sensor immediately after each BG test and to
provided by the health care system without adjunctive based their insulin dosage adjustments on the BG reading.
SMBG for clinical decision-making. Patients calibrated the DG4P according to manufactures’
The aim of this study was to compare the accuracy of specification against the BG measurements. After 7 days,
Libre vs. DG4P in adults with T1D as assessed by mean subjects began a second 7-day session with the original
absolute difference (% MARD) during glycaemic excursions DG4P sensor to assess accuracy beyond the manufacturer
induced in at a clinical research center and during home specified lifetime. In case of sensor failure, loss of signal,
use. skin issues or accidental dislodgment, the sensor was
substituted. After 14 days, data from FreeStyle Libre and
Methods DG4P were downloaded.

Study design and participants CRC studies


After calibration of the DG4P system, blood sampling
This was a prospective, single-centre, randomised, cross- through a venous line started at 8:00 a.m. and glucose
over study performed at the Clinical Research Center (CRC) concentration was measured on 25-mL-samples using YSI
of Padova University between April and November 2300 STAT PLUS glucose and lactate analyser (YSI Inc.,
2016. The study was registered in ClinicalTrials.gov Yellow Springs, OH).
(NCT02734745), approved by the institutional ethics review At 8:15 a.m. subjects received a regular breakfast and
board and done according the Declaration of Helsinki. were randomized 1:1 to either their usual pre-meal insulin
Written informed consent was obtained before study start. dose or a delayed and increased bolus to induce significant
Inclusion criteria were: 18 years, T1D for 1 year and glycaemic excursions. At one visit, breakfast insulin dose
treatment with CSII or MDI. was calculated using carbohydrate-to-insulin ratio, adding
Exclusion criteria were: pregnancy, lactation, medica- a correction bolus if fasting glucose was >5.6 mmol/L
tions impacting on glucose metabolism and inability to (>100 mg/dL). Blood was sampled every 15 min between
comply with study procedures. Participants avoided drugs 8:00 to 11:00 a.m. and every 30 min during the next 3 h. At
impairing sensor-related measurement of glucose (i.e. the other visit, the insulin dose was doubled and admin-
acetaminophen) during the study. istered 30 min after breakfast to induce an early post-meal
hyperglycaemia followed by a decrease in blood glucose.
Study devices Blood was sampled every 15 min between 8:00 a.m. to
9:00 a.m., every 10 min between 9:00 a.m. to 12:00 p.m.,
The FreeStyle Libre System consists of a sensor and a every 15 min between 12:00 p.m. to 1 p.m., and every
receiver/reader. The sensor is provided with a sterile 30 min between 1:00 p.m. to 2:00 p.m. Sensor readings
catheter (0.4 mm calliper), which is inserted 5 mm under were obtained simultaneously to venous YSI samples and
the skin and connected to a round disk (35 mm  5 mm) capillary BG, which was tested hourly. If hypoglycaemia
applied on the back of the upper arm. The sensor measures was detected, subjects received rescue carbohydrates
182 F. Boscari et al.

when glucose was <3.0 mmol/L (<54 mg/dL) or at physi- used to compare normally and not normally distributed
cian’s discretion and sampling was every 5 min in case of variables. The means of more than two group were
hypoglycaemia. Patients left the CRC at 2:00 p.m. compared with one-way analysis of variance (ANOVA). The
level of significance accepted was 0.05 using two-tailed
Sensor accuracy assessments tests. Statistical analysis was performed using MATLAB,
Statistics Toolbox (Release 2016a, The MathWorks, Inc.,
The systems were tested over 14 days of continuous use at- Natick, Massachusetts, United States).
home and during two six-hour CRC admissions scheduled
at day 3e5 and 9e11 (Fig. S1). Accuracy was expressed as Results
mean absolute relative difference (%MARD). Plasma
glucose values were used as reference during the CRC Subject disposition
visits and SMBG values were used as reference during the
home phase. Sensor data were matched with venous (CRC Twenty-four adults with type 1 diabetes were enrolled.
assessments) or capillary (home use assessment) BG One subject was excluded from analysis as home data
measurements. When CGM and reference values could not could not be uploaded; one subject left the study for poor
be obtained at the same time, CGM values were linearly device acceptance. Twenty-two subjects completed the
interpolated to match reference values, provided CGM study: 13 female/9 male (10 CSII/12 MDI), mean age
values were obtained within 5 min of reference values. 36.3  12.9 years (mean  SD), diabetes duration
MARD analyses were performed separately for break- 18.9  11.1 years, BMI 23.5  2.7 kg/m2, HbA1c 7.3  0.75%
fast with normal bolus vs. breakfast with increased and (56.7  8.19 mmol/mol).
delayed bolus, and for week 1 vs. week 2 of sensor use. Ten subjects had breakfast with delayed and increased
MARD was calculated over the entire glycaemic range and bolus during the first week of sensor use, 12 during the
for glucose values: <3.9 mmol/L (70 mg/dL), and 3.9e10 second week. The Libre device was scanned 14,4  6 time/
mmol/L (70e180 mg/dL) and >10 mmol/L (>180 mg/dL). day.
Sensor accuracy was also assessed by calculating both the
percentage of data points in zones A and A þ B of the Home phase
Clarke Error Grid (CEG) [16] and the proportion of values
that fulfilled ISO 15197:2013 criteria (percentage of sensor As shown in Fig. 1, there was no significant difference in
data within 15% of reference value for glucose concen- overall accuracy between Libre and DG4P during at-home
trations 5.6 mmol/L (100 mg/dL), and within 0.8 mmol/ use. Both systems demonstrated similar performances in
L (15 mg/dL)) of reference value for glucose concentrations the hypoglycaemic and hyperglycaemic ranges; however,
<5.6 mmol/l (100 mg/dL). DG4P showed better accuracy in the euglycaemic range.
During CRC sessions, MARD was also calculated by Both systems showed worse accuracy in the hypo-
consolidating BG references into five groups based on the glycaemic range.
rate of change (ROC), calculated as the first-order differ- Comparison of week 1 to week 2 showed similar ac-
ence between the current and the previous sample divided curacy with both sensors during home use (Fig. 1). The
by the time distance between the two. The five ROC ranges MARD for both systems worsened in week 2, but not
were: increasing glucose, >0.08 mmol/L/min (>1.5 mg/dL/ significantly (Libre: 13.0  3.7% week 1 vs. 14.6  4.9%
min) and >0.03 to 0.08 mmol/L/min (>0.5 to 1.5 mg/dL/ week 2, p Z 0.080; DG4P: 12.5  4.1% week 1 vs.
min); stable glucose, 0.03 to 0.03 mmol/L/min (0.5 to 13.4  3.2% week 2, p Z 0.092) (Table S1).
0.5 mg/dL/min); and decreasing glucose, >0.03 to As shown in Table S1, Libre’s accuracy worsened
0.08 mmol/L/min (>0.5 to 1.5 mg/dL/min), and >0.08 significantly during days 11e14 compared with days 1e10
mmol/L/min (>1.5 mg/dL/min). Sensor accuracy based on (12.6  6.0% on days 1e10 vs 15.0  8.2% on days 11e14,
ROC values was measured considering all CRC sessions p Z 0.006). DG4P had the worst accuracy on days 1 and 8
(with and without induced glycaemic excursions), and (14.5  7.1% vs. 11.7  7.4%, p Z 0.015), which was ex-
separating first from second week visits. pected; CGM performance on day 1 of sensor wear is
Because sensor accuracy was expected to be lower on known to be worse than performance on subsequent days.
day 1 of sensor insertion [15,17] and during the second
week of monitoring (due to possible sensor performance CRC studies
degradation), MARD analyses were also performed
comparing: day 1 vs. all the other days; days 1 and 8 vs. all Induced glycaemic excursions
other days; week 1 vs. week 2; and combined days 1e10 MARD increased significantly in both the Libre and DG4P
vs. combined days 11e14. Data are presented as systems following delayed and increased bolus at break-
mean  standard deviation. fast; the betweenesensor difference was not significant
(Table 1). Libre showed a lower MARD vs. DG4P during
Statistical analysis hyperglycaemia, whether induced by a breakfast with
delayed and increased bolus or following a breakfast with
Univariate analyses were performed for descriptive sta- correctly administered insulin bolus (Table 1). Overall
tistics. The t-test and the Wilcoxon signed-rank test were MARD values (with/without induced glycaemic
FreeStyle Libre and Dexcom G4 Platinum sensors 183

Figure 1 Top: Mean absolute relative difference (MARD) per day  95% confidence interval for Libre or DG4P. Bottom: Mean absolute relative
difference (MARD) between Libre or DG4P glucose readings and capillary glucose reference concentration at home.

Table 1 Mean absolute relative difference (%MARD) between Libre and DG4 P glucose readings and reference glucose concentration in venous
blood in subjects with type 1 diabetes receiving a breakfast with standard or delayed and increased insulin bolus.

Glucose profile Breakfast w/standard insulin bolus Breakfast w/delayed and increased insulin bolus
Libre DG4P Data pairs P value Libre DG4P Data pairs P value
Overall 10.9  4.1 13.1  4.6 424 0.055 14.9  5.5 18.1  8.1 710 0.062
Hypoglycaemia 10.8  6.9 12.9  11.9 9 0.824 21.7  14.4 27.3  22.9 87 0.148
<3.9 mmol/L (<70 mg/dL)
Euglycaemia 13.3  4.9 13.7  6.7 245 0.894 17.3  6.9 18.5  7.2 362 0.421
3.9e10 mmol/L (70e180 mg/dL)
Hyperglycaemia 7.8  4.5 11.2  5.1 170 0.010 10.2  4.9 14.5  6.1 261 0.031
>10 mmol/L (>180 mg/dL)
184 F. Boscari et al.

excursions) were similar during week 1; however, Libre and 1 sensor was interrupted at day 12 for acetaminophen
MARD was significantly lower during week 2 (Table S2). assumption by the subject.
Libre MARD values were also significantly lower in the
hyperglycaemic range during weeks 2 (Table S2). Safety and adverse events
No sensor failure requiring removal or infection at the
Clark error grid analysis and fulfilment of ISO insertion site were reported.
15197:2013 criteria
There was no difference in the systems’ clinical perfor- Discussion
mance, most values fell within the clinically acceptable
error zones (A þ B) (Fig. 2, Table S3). This study compared the Libre and DG4P continuous
A notable proportion of glucose values from Libre and glucose monitoring systems during at-home use and dur-
DG4P fulfilled ISO 15197:2013 accuracy criteria: 70.2% and ing induced glycaemic excursions. It is important to note
73.5%, respectively, during home phase; 74.1% (Libre) and that the systems compared are not strictly equivalent,
65.0% (DG4P) during CRC assessment with normal insulin since Libre data can be used in place of SMBG for insulin
bolus and by 69.5% (Libre) and 57.0% (DG4P) during CRC adjustments, while DG4P is considered adjunctive to
assessment with increased and delayed insulin bolus SMBG. Furthermore, MSL is 14 days for Libre and only 7
(Table S3). days for DG4P. Despite these differences, we felt it was
important to compare the two systems because they are
Rates of glucose change widely used in Italy, and because many patients wear
Analysis of accuracy during glucose swings was possible DG4P beyond MSL.
only where YSI values were available. No significant dif- In designing the study protocol, we relied on evidence
ference in sensor accuracy was observed when glucose showing that DG4P accuracy remains unchanged 7 days
was stable 0.03 to 0.03 mmol/L (0.5 to 0.5 mg/dL/min), beyond MSL [15]. As shown in our study, we found that the
while Libre was significantly more accurate during periods Libre and Dexcom G4 sensors functioned with similar ac-
of moderate and rapid glucose changes (Table 2). Accuracy curacy during home use for 2 weeks. Surprisingly, DG4P
among both sensors within the DG4P MSL (7 days), was accuracy was even better in the euglycaemic range, which
similar except during glucose changes at >0.08 mmol/L/ is likely due to the 2 daily calibrations recommended by
min (>1.5 mg/dL/min) (Table 2). the manufacturer.
We also found that Libre’s accuracy decreased between
Sensors lifetime days 11 and 14, which is an important consideration for
Mean sensor lifetime was 13.45 days (1.3) for DG4P and patients who base insulin dosing on sensor readings only.
13.5 (1.2) for Libre. Eighteen (86.4%) Libre sensors func- The increase in MARD and variability of DG4P on day 8 was
tioned for the full 14 days; 2 sensors were removed at day due to sensor recalibration, which is intrinsic to the Dex-
10 and 1 at day 13 for adhesive issues and 1 sensor was com algorithm [15].
replaced at day 3 due to accidental dislodgment. Sixteen With both systems, no difference in accuracy was found
(63.3%) DG4P functioned for the full 2 weeks, 1 sensor was comparing week 1 vs. week 2 (in CRC and at home). As
removed at day 9 and 4 at day 13 for adhesive problems, expected, the accuracy of both systems worsened

Figure 2 Clarke Error Grid Analysis. Left: Libre (grey dots) and DG4P (black dots) vs. capillary measurements during 14 days at home. Right: Libre
(grey dots) and DG4P (black dots) vs. venous measurements after delayed and increased insulin administration.
FreeStyle Libre and Dexcom G4 Platinum sensors 185

significantly during rapid glucose excursions due to the

P value

<0.001

<0.001
time delay between plasma and interstitial glucose [18].

0.019

0.039

0.248
Additionally, both sensors deviated approximately 20%
from reference value in the hypoglycaemic range during
Table 2 Mean absolute relative difference (%MARD) between Libre or DG4P readings and reference venous blood glucose according to different rates of change of glucose concentration. home use, indicating that low glucose readings should

Data pairs
routinely be confirmed by SMBG.
In studies done at the CRC during the first week Libre

144

157
71
55

66
and DG4P had similar accuracy across all glucose values
but Libre had a lower MARD when glucose decreased by

25.5  12.2
14.6  12.1

14.6  10.9

14.9  12.7
13.9  18.7
>0.08 mmol/L/min (>1.5 mg/dL/min). During the second
MARD (%)-YSI sessions

week Libre was more accurate than DG4P in the hyper-


DG4P

glycaemic range and when blood glucose levels increased


>0.03e0.08 mmol/L/min (>0.5 to 1.5 mg/dl/min). These
results are similar to a recent study of Aberer and col-
17.8  14.0

12.4  13.6
8e14 days

10.6  9.0

12.6  8.5

10.8  9.0
leagues, comparing Libre with DG4P and Medtronic Enlite
over 12 h in an inpatient setting under real-life conditions
Libre

(e.g., meals, exercise, hypoglycaemia and hyperglycaemia)


[14]. However, the effect of changes in glucose levels on
accuracy was not considered in the study.
P value

0.362

0.702

0.684
0.049

We found that Libre and DG4P had similar accuracy if


0.07

used within MSL and with glucose rates of change


<0.08 mmol/L/min (<1.5 mg/dL/min). For greater rates of
pairs
Data

change Libre was more accurate. It can be concluded that


177

180
106
99
72

in patients with good glycaemic control, using Libre or


DG4P would have little effect on dosing accuracy. How-
20.7  14.8
15.8  13.1

13.0  15.3

13.2  14.5
18.3  19.4

ever, Libre may be preferable for patients with high


MARD (%)-YSI sessions

glucose variability. We found no differences in sensor


DG4P

lifetime, nor in the need to replace sensors due to loss of


adhesion, accidental displacement or skin irritation.
The strength of this study is that it was performed in-
16.8  12.0
13.8  12.0

13.5  12.6

12.7  10.6
13.7  13.6

dependent from sensor manufacturers. Limitations may


1e7 days

include the fact that home glucose values were measured


Libre

by subjects themselves and that the DG4P has not been


approved by FDA for clinical-decision-making. Another
possible limitation is that home glucose values were
P value

<0.001

<0.001
<0.001
0.010

0.083

collected using the BG meter embedded in the Libre sys-


tem. Although Libre is factory calibrated and does not use
capillary glucose values, the factory calibration could have
pairs
Data

been designed and/or optimized to better match the blood


170
127

321

337
172

glucose values collected with the Libre meter. However,


the results of the Libre’s accuracy at home vs CRC, where
23.0  14.1
15.2  12.7

14.0  13.7

14.3  13.6
16.9  18.9
MARD (%)-All YSI sessions

the BG references have been collected with an indepen-


dent system (YSI), do not evidence any advantage for the
DG4P

Libre. It should be noted that the DG4P sensor used in the


study utilized an older algorithm, which has since been
updated (505 software). Dexcom has also launched a new
17.0  14.7
12.4  11.0

12.9  11.0

13.2  13.3
1e14 days

11.8  9.9

sensor (Dexcom G5 Mobile), which is the first CGM device


approved by the U.S. Food and Drug Administration (FDA)
Libre

for non-adjunctive use. It would be interesting to compare


Libre with the Dexcom G5 Mobile system.
>0.03 to 0.08 (>0.5 to 1.5)

0.03 to 0.03 (0.5 to 0.5)

>0.03 to 0.08 (>0.5 to 1.5)


mmol/L/min (mg/dL/min)

Conclusion

We demonstrated that DG4P is as accurate as Libre over 2


Rate of change,

weeks irrespective of MSL. Libre and DG4P have similar


>0.08 (>1.5)

>0.08 (>1.5)
Decreasing
Increasing

accuracy during rapid glucose swings if used within MSL


and when glucose changes by <0.08 mmol/L/min
Stable

(<1.5 mg/dL/min). Above this rate of change, Libre is more


accurate.
186 F. Boscari et al.

Funding [6] Hommel E, Olsen B, Battelino T, Conget I, Schütz-Fuhrmann I,


Hoogma R, et al. Impact of continuous glucose monitoring on
quality of life, treatment satisfaction and use of medical care re-
This research did not receive any specific grant from sources: analyses from the SWITCH study. Acta Diabetol 2014;51:
funding agencies in the public, commercial, or not-for- 845e51.
profit sectors. [7] Pickup JC, Freeman SC, Sutton AJ. Glycaemic control in type 1
diabetes during real time continuous glucose monitoring
compared with self monitoring of blood glucose: meta-analysis of
Author contributions randomised controlled trials using individual patient data. BMJ
2011;343. d3805.
[8] Heinemann L, DeVries JH. Evidence for continuous glucose
D.B. and F.B conceived the study, had full data access, monitoring: sufficient for reimbursement? Diabet Med 2014;31:
drafted the manuscript and are responsible of the integrity 122e5.
[9] Lind M, Polonsky W, Hirsch IB, Heise T, Bolinder J, Dahlqvist S,
of the data. S.G., A.F., and A.A. critically reviewed the
et al. Continuous glucose monitoring vs conventional therapy for
manuscript. A.F. analyzed the data. M.C.M. analyzed blood glycemic control in adults with type 1 diabetes treated with
sample. V.V. and A.M.L.A. collected the data. multiple daily insulin injections: the GOLD Randomized Clinical
Trial. JAMA 2017;317:379e87.
[10] Battelino T, Conget I, Olsen B, Schütz-Fuhrmann I, Hommel E,
Conflicts of interest Hoogma R, et al., The SWITCH Study Group. The use and efficacy of
continuous glucose monitoring in type 1 diabetes treated with
insulin pump therapy: a randomised controlled trial. Diabetologia
Authors declare no competing interests. 2012;55:3155e62.
[11] Beck RW, Riddlesworth T, Ruedy K, Ahmann A, Bergenstal R,
Haller S, et al. Effect of continuous glucose monitoring on glycemic
Acknowledgements control in adults with type 1 diabetes using insulin injections: the
DIAMOND Randomized Clinical Trial. JAMA 2017;317:371e8.
The authors thank subjects who participated to the study. [12] The Juvenile Diabetes Research Foundation Continuous Glucose
Monitoring Study Group. Continuous glucose monitoring and
intensive treatment of type 1 diabetes. N Engl J Med 2008;359:
Appendix A. Supplementary data 1464e76.
[13] Ólafsdóttir AF, Attvall S, Sandgren U, Dahlqvist S, Pivodic A,
Skrtic S, et al. A clinical trial of the accuracy and treatment
Supplementary data related to this article can be found at experience of the Flash glucose monitor FreeStyle Libre in adults
https://doi.org/10.1016/j.numecd.2017.10.023. with type 1 diabetes. Diabetes Technol Ther 2017;19(3):164e72.
[14] Aberer F, Hajnsek M, Rumpler M, Zenz S, Baumann PM, Elsayed H,
et al. Evaluation of subcutaneous glucose monitoring systems
References under routine environmental conditions in patients with type 1
diabetes. Diabetes Obes Metab 2017 Feb 16. https:
//doi.org/10.1111/dom.12907 [Epub ahead of print] available from:
[1] Diabetes Control and Complications Trial Study Group. The effect
https://www.ncbi.nlm.nih.gov/pubmed/28205324.
of intensive treatment of diabetes on the development and pro-
[15] DeSalvo DJ, Ly TT, Wadwa RP, Messer L, Westfall E, Gopisetty D,
gression of long-term complications in insulin-dependent dia-
et al. Continuous Glucose Sensor survival and accuracy over 14
betes mellitus. N Engl J Med 1993;329:977e86.
consecutive days. Diabetes Care 2016;39(8):e112e3.
[2] DCCT Research Group. Epidemiology of severe hypoglycemia in
[16] Wentholt IM, Hart AA, Hoekstra JB, Devries JH. How to assess and
the diabetes control and complications trial. Am J Med 1991;90:
compare the accuracy of continuous glucose monitors? Diabetes
450e9.
Technol Ther 2008;10(2):57e68.
[3] Davis S, Alonso MD. Hypoglycemia as a barrier to glycemic control.
[17] Christiansen M, Bailey T, Watkins E, Liljenquist D, Price D,
J Diabetes Complicat 2004;18(1):60e8.
Nakamura K, et al. A new-generation continuous glucose moni-
[4] Deiss D, Bolinder J, Riveline JP, Battelino T, Bosi E, Tubiana-Rufi N,
toring system: improved accuracy and reliability compared with a
et al. Improved glycemic control in poorly controlled patients with
previous-generation system. Diabetes Technol Ther 2013;15(10):
type 1 diabetes using real-time continuous glucose monitoring.
881e8.
Diabetes Care 2006;29(12):2730e2.
[18] Cobelli C, Schiavon M, Dalla Man C, Basu A, Basu R. Interstitial fluid
[5] Hermanides J, Nørgaard K, Bruttomesso D, Mathieu C, Frid A,
glucose is not just a shifted-in-time but a distorted mirror of blood
Dayan CM, et al. Sensor augmented pump therapy lowers HbA1c
glucose: insight from an in silico study. Diabetes Technol Ther
in suboptimally controlled Type 1 diabetes; a randomized
2016;18(8):505e11.
controlled trial. Diabet Med 2011;28:1158e67.

You might also like