Direct digital phase noise measurement


Back to projects

DUT

REF

This apparatus measures phase noise down to wideband noise floor levels below -170 dBc/Hz. Historically, such measurements were either difficult or
expensive to make. Based on the work [1] of Grove et al., the method described here is cheap, sensitive, accurate and requires no calibration. It is a
differential measurement between a device under test (DUT) and a reference oscillator, which are connected to the SMA ports at the extreme left. Ideally,
the reference oscillator should be an order of magnitude (or more) quieter, so DUT noise predominates.

The largest board is a Xilinx SP605 FPGA evaluation kit, attached to which (by the FMC connector) is a Linear Technology DC1525A-A quad ADC evaluation
board, minus its top-right corner. The two identical boards to which the DUT and reference are connected are custom-made power splitters, feeding four
ADC inputs through coaxial cables. The part-populated yellow board, spare from my GPS project, was re-purposed as a clean power supply for a Crystek
CPRO33-77.760 crystal oscillator, which can just be seen mounted vertically on the ADC clock input SMA.

ADC resolution is 14-bits. Samples are received by the FPGA on eight LVDS pairs, each of which can run at up to 1Gbps. In the FPGA, the four ADC channels
are digitally down-converted to IQ complex baseband, low pass filtered, decimated, and phase-demodulated using a CORDIC arc-tangent block. The
resultant phase data is streamed to a Windows laptop via UDP over Ethernet. Phase data is converted to power spectral density by Fast Fourier Transform
(FFT) and displayed in real-time. The quality of the measurement is refined by correlation / averaging the longer it runs.

The inputs are down-converted by mixing with a quadrature NCO. NCO rate is set close to input frequency, but cannot remain exactly equal, so the
baseband signal is close to but rarely at zero frequency. The CORDIC output is phase noise superimposed on a gentle linear ramp, the gradient of which
depends on the small unavoidable difference between input frequency and NCO rate.

Cancel & correlate


It's possible to measure phase noise of a DUT, relative to the sampling clock, using a single ADC channel. In fact, I did so as a first step. So why the two
inputs and four ADC channels? Because using a single 14-bit ADC, even assuming zero sampling clock jitter, the simulation-predicted noise floor is worse
than -150 dBc/Hz. Typically, more like -140 dBc/Hz. Grove et al. did two things to get the noise floor down below -170 dBc/Hz.

The first neat trick is to make a differential measurement between DUT and reference, by subtracting their phases. The phase noise of the sampling clock is
cancelled-out. The quality of the measurement depends on the reference, not the ADC clock, which can be slightly inferior. Although phases are subtracted,
uncorrelated noise powers add. The result is the sum of DUT and reference noise. The reference can be an order of magnitude quieter than the DUT, or, if
two identical oscillators are compared, half the power (-3dB) can be attributed to each.

Cancelling clock noise still leaves ADC quantisation noise and thermal noise, which raise the noise floor above -150 dBc/Hz, as already mentioned. The
second neat trick is to duplicate the measurement. This is the reason for the power splitters. Everything in the FPGA (and downstream in software) is

www.aholme.co.uk/PhaseNoise/Main.htm 1/5
2/3/2020 Direct digital phase noise measurement
replicated. We simultaneously make two completely independent measurements. Both contain the same (correlated) DUT+Reference noise; but different
(uncorrelated) ADC noises. The latter are then greatly attenuated using the cross spectrum experimental method described in [2] by E. Rubiola and F.
Vernotte.

DUT and reference do not have to be the same frequency. If the sampling points of all four ADC channels are displaced by a small time offset Δt due to jitter
on the common sampling clock, it affects the measured phases in proportion to their frequencies: Δθ = ω.Δt. Reference phase must be scaled by the
DUT/REF frequency ratio to cancel clock jitter. ADC noises remain uncorrelated, scaled or not. It can be beneficial to use a higher reference frequency,
because its phase noise contribution, ΦREF, will then be scaled down. All we want from it is stability.

Phase wraps
The 32-bit fixed-point binary output of the CORDIC is encoded in semi-circles with 31 bits after the binary point. The most significant bit represents ±1
semi-circle or ± π radians. The angle can be interpreted as signed or unsigned, without ambiguity. Due to unavoidable frequency differences, the measured
phases will always be slowly but steadily ramping, either up or down. Phase 'wraps' occur every 2π radians. Real measurements have phase noise
superimposed on the ramp and can "chatter" back and forth several times as they go through the wrapping point:

These discontinuities must be removed before the FFT. One way is to unwrap them. Integer precision can be extended before the binary point and 1.0 circles
(2.0 semi-circles) are added or subtracted every time the raw data wraps. Another way - the method I am using - is to convert phase to instantaneous
frequency, ω, by differentiating. Each phase wrap generates a carry or a borrow from the differentiator; but there are no jumps or glitches in its 32-bit
output. The FFT then calculates power spectral density of angular frequency fluctuation, which is converted to power spectral density of phase by dividing by
ωm² afterwards.

Differentiation converts sin(ωm.t) to ωm.cos(ωm.t) so has gain j.ωm. The 90 degree phase shift doesn't affect power. Modulating frequency f = ωm/2π is
the offset from the carrier.

Frequency domain
The final output, ℒ(f), is the "one-sided" power spectral density of phase, relative to the carrier, in dBc/Hz. This is plotted on a log/log scale, typically over
the frequency range 0.1 Hz to 100 kHz. Samples are transformed into the frequency domain by Fast Fourier Transform. Making two independent
measurements doubles the amount of processing required. In order to execute this in real-time, with the same number of data points in each decade, the
processing is broken up into a succession of identical stages, decimating by 10 at each step:

www.aholme.co.uk/PhaseNoise/Main.htm 2/5
2/3/2020 Direct digital phase noise measurement

The above diagram includes cross-correlation. Complex FFT outputs bins from one measurement are multiplied by the complex conjugate of the
corresponding bin from the other. The complex product is averaged for minutes, hours or days. Uncorrelated noises are attenuated by 5log10(N) where N is
the number averaged. After a while, the average real part tends to contain only correlated noise. The average imaginary part is an indicator of system noise
floor. 5dB steps can been seen in the imaginary part between decades because of decimation. Each decade averages 10 times more data than the next lower
in frequency.

ADC sampling rate is 77.76 Msps, decimated to 607.5 ksps in the FPGA. FFT length is typically 1000 points, and bins 10 to 99 are plotted from each decade.
Length can be increased to get finer detail. Bins LEN/100 to LEN/10 are always plotted in the middle decades; but more are required (up to LEN *
100/607.5) in the first stage to reach 100 kHz. Fewer are needed in the last stage because the graph starts from 0.1 Hz. Low pass filters are 6th order
Butterworth IIR with a normalised cut-off frequency of 20/1000.

Results must be scaled to compensate for CIC pass-band droop; differentiator and LPF gain; FFT growth; FFT bin size (converting to dB/Hz) and equivalent
noise bandwidth (leakage) of a raised cosine window function, which is applied before the FFT. For efficiency, combined adjustments are pre-computed in dB
and applied as offsets to the final outputs.

Spurs
Does this sounds too good to be true? Are you wondering what the snag is? The biggest problem is instrument-generated spurs, due to ADC quantisation.
These appear when harmonics of the input signal fall close to harmonics of the sampling clock. I discovered them the hard way. The DC1525A ADC board
works up to 125 Msps and I was making measurements of 5 MHz sources using a 124.998 MHz sampling clock! None of the papers I read beforehand led me
to expect this, and I was puzzled for a while, until I reproduced the spurs using a very simple simulation. Moving to 77.76 Msps improved matters.

The same signal can be fed to both the DUT and reference inputs with a third power splitter. I used this configuration to estimate system noise floor, before I
knew about the imaginary part of the averaged cross-product. When I started making differential measurements, I discovered another problem: low-
frequency spurs, due to crosstalk, at the DUT - Reference difference frequency. Fortunately, although they are quite noticeable in the imaginary part of the
cross-product, these spurs barely push through into the real part.

Results
Each of the plots below was averaged over several hours. DUT and reference inputs were 5 MHz. Blue trace is ℒ(f), the one-sided power spectral density of
phase, as estimated by the real part of the average cross-product. Green trace is system noise floor in the imaginary part. The Wenzel ULN is an ultra-low-
noise reference oscillator:

www.aholme.co.uk/PhaseNoise/Main.htm 3/5
2/3/2020 Direct digital phase noise measurement

The Wenzel 500-03220 is about 1 Hz off-frequency, judging by the low-frequency crosstalk spurs around 1 and 2 Hz, which are evident in three plots, most
strongly in the imaginary part. All plots have spurs at the 50 Hz power line frequency. Harmonics of 50 Hz and other "real" spurs are visible on the
synthesizer plots. The -160 dBc spur around 21 kHz in (a) is 77759000*58 - 5000001*902. The two Dana 7020 "Digiphase" synthesizers are several years
apart in age; but very similar in performance, except #2 has a probem at 410 Hz. Their closed loop responses match figure 26-8 in Garry Gillette's essay on
page 290 of "Analog Circuit Design" edited by Jim Williams. Digiphase close-in performance (-90dBc/Hz @ 1Hz) beats the Marconi 2019A by 20dB. The
2019A has a lower wide band noise floor and much narrower PLL loop bandwidth. Wenzel may be as responsible as Dana for the measurements at 0.1 Hz.
The Dana synthesizers were set to 5.000000, so fractional compensation was not operative.

Software - Updated October 2019


Windows 10 can be used with a workaround; but Windows 7 or XP is recommended. The software has been built successfully with Visual Studio 6, Visual
Studio 2013 and Visual Studio 2019. Source files and installation instructions can be found in a zip linked below. The UI is minimal. The following commands
are added to the system menu:

Load FPGA Send FPGA configuration bitstream


Settings Edit measurement frequencies, network settings e.t.c.
Levels Check ADC input amplitudes
Start Start measurement
Stop Stop and output results (to csv file)

The IP address of the PC must be specified on the Settings dialog Ethernet tab. Only direct cable connections have been tested. It might be possible to work
through a switch or router; but packet loss and re-ordering cannot be tolerated. Heavy network or CPU-intensive activities are best avoided whilst
performing measurements.

Data is crudely plotted in real-time, then dumped to a file, from which the below-linked Python script produces proper graphs like the above. Hold down the
shift key whilst resizing the window to change the number of horizontal and/or vertical decades visible in the real-time view. Resize normally afterwards to
re-centre the view. The dBc/Hz power spectral density plot is normalised to 1 Hz bandwidth. Click near a spur to display its frequency and true amplitude in
dBc. Left-click for PN spurs; right-click for AN.

Verilog - Updated October 2019


8-lane LVDS source-synchronous deserializer
Interpolating quadrature NCO
Parameterised CIC filter
CORDIC arc-tangent function
UDP Ethernet sender
32x32 and 14x34 multipliers

The software zip contains a pre-generated bit stream in XSVF format for the "Load FPGA" menu option. The FPGA has been built successfully using Xilinx ISE
13.3 and 14.7. Project files can be found in a zip linked below. The build procedure is: open QuadChannel.xise project; migrate to 14.7 if necessary; locate
and select fifo - fifo_generator_v8_3 in the hierarchy list; run "Regenerate Core" in the process list; scroll back up and click on top-level module
QuadChannel.v in the hierarchy; run "Generate Programming File" in the process list.

Acknowledgments
I wish to thank Joël Imbaud of Institut FEMTO-ST who replicated the SP605/DC1525A setup and verified its accuracy against a commercial Symmetricom
5120A. Instead of custom boards, Joël used Mini-Circuits attenuators; ZMSC-2-1+ connectorized power splitters; SPB10.7+ and SLP5+ filters; and AMP-77
amplifiers. He removed the SFP cage to avoid sawing his ADC board. I checked the Gerbers before cutting mine.

I wish to thank Adrian Rus who obtained good agreement between his SP605/DC1525A and a Rohde and Schwarz FSWP for one Wenzel ULNO; and against
the manufacturer's specification of another.

Adrian tested several ways of feeding the ADC before settling on a preferred configuration:

www.aholme.co.uk/PhaseNoise/Main.htm 4/5
2/3/2020 Direct digital phase noise measurement

Adrian uses two high-quality reference oscillators with low output impedances connected directly to the ADC inputs. Their noises are removed by cross-
correlation and the measurement is DUT noise only. Sometimes he inserts LNOM multipliers after the 3dB split to measure very low noise floors. The clock
shaper and attenuator are mounted directly on the clock input SMA without cabling.

Power Splitters
My power splitter schematic is linked below. The design is not ideal for measuring ultra-low noise floors. It was a first attempt based on guess work.
Primarily, I wanted anti-alias filters, around which I placed resistive attenuators for wideband impedance matching. MMIC amplifiers restore the level and
provide reverse isolation. Unfortunately, the amplifiers and attenuators compromise overall noise figure. Input configuration is best optimized by trial and
error using SMA connectorized modules.

Prequel
This project began with an attempt to make very precise average frequency measurements using a digital PLL. Sampling at 1 Msps, using the 14-bit ADC on
the Spartan 3E Starter Kit, it was possible to measure average frequency of a 100 kHz source to µHz precision over a 1 second averaging period:

In order to measure exactly 100000.000000 Hz, the ADC was clocked using a standard frequency output from the same generator that produced the 100
kHz. Despite this, a small frequency offset was observed as the equipment warmed-up. This was due to a time-varying delay (phase shift) in the signal path,
equivalent to a frequency shift ω = dθ/dt, which settled down once thermal equilibrium was reached.

The most significant bits of the detected product are zero when the loop is locked, but less significant bits are noisy, and were routed out via a DAC to view
the residual noise on an oscilloscope. It was the PLL method of phase noise measurement, implemented digitally. I googled "digital phase noise
measurement" and found the Grove et al. paper.

Source code
QuadChannel Spartan6 FPGA Project (Xilinx ISE)
LivePlot Windows C++ Application (Microsoft Visual Studio)
Python Matplotlib graphs
Power splitter - schematic (PDF)

Links
5330A Programmable Cross Spectrum Analyzer Operation and Service Manual by John Miles
A 1 MHz to 50 GHz Direct Down-Conversion Phase Noise Analyzer with Cross-Correlation Rohde & Schwarz

References
1. Grove, J. et al., "Direct-digital phase-noise measurement," Proc. of 2004 IEEE International Frequency Control Symposium, Montreal, Canada, pp. 287-
291, August 2004.
2. E. Rubiola and F. Vernotte, "The cross-spectrum experimental method," Mar. 2010, arXiv:1003.0113v1 [physics.ins-det].