Infant radiant warmer Service manual
Contents
1. Introduction ……………………………………………………………….….1
4. Functions instruction……………………………………………..…….…....4
1
Infant radiant warmer Service manual
1 Introduction
Infant radiant warmer of model BN-100 adopts anti-explosion quartz infrared radiation tube for its
radiation source. Infant radiant warmer of model BN-100 has three operation modes: pre-heat, auto
and manual, the indicator shows under which mode the equipment is running, the testing temperature
and setting temp is shown independently in the screen, The equipment will alarm with sound and light
if failure and exceptional situation (such as power failure, sensor failure, over temp, temp deviation)
comes out when using the infant radiant warmer.
This instruction consists of instruction of usage and techniques, and is suitable for the
installation, usage, cleanout, maintenance and analysis and solution of common failure of infant
radiant warmer of model BN-100.
Some parts of the products used may not be the same as the ones in the introduction list and
circuit diagram, which is because our company put improved parts and circuits into use before
putting them into printed introduction, but it isn’t affect use this equipment.
2
Infant radiant warmer Service manual
Shake
brilliantly
T h e w o r k in g
S ta rt
s t a t e in it ia liz in g
O s c i ll a t o r Ee p ro m
I n it ia liz e S a m p le d a t a
A /D S h o w t h e d r iv e
5 4 1 0 AD
Y
S e t u p th e s ta te T r o u b le ?
T re a tm e n t F a u lt a la r m t o in it ia liz e
N
Exp o rt o f
t h e c o n t r o llin g
S e n s o r ja c k Y S e t u p th e ke y ?
te m p C o n tro l a n d e x p o rt
N
N
O v e r t im e ? Show
The key F a u lt a la r m
is d e a lt w it h
2.4. Parameter revision method (user don’t revise by yourself without experimental condition).
2.4.1. Temperature revision.
When the warmer in operating under “auto” mode, press “set” button all along till the temp display
screen shows “---”, first character is twinkling, press “increase” button once, the screen shows “1--”. Press
“decrease” button once, second character is twinkling, press “increase” twice; the screen shows “12-”.
Press “decrease” button, third character is twinkling, press “increase” three times, the screen shows “123”,
then press “set” button, the system into main menu: the temp set display screen shows “PA.-”.
Press “increase” button once, temp set display screen shows “PA.1”. Then press “set” button, the
screen shows “0”(“0”: to give up; “1”: to determine), press “increase” button, it shows “1”. Press “set”
button, now system go into “temp revision” submenu.
Temp display screen shows P××( the initial value is P2.0), under this state user can presses the
“increase” or “decrease” button to revise the deviation of the temp displayed by the skin temp and the temp
in the standard thermometer(the revise range:±2℃). Then press “set” button, now the temp display screen
shows U××(the initial value is U1.0), under this state revision of temp value for over-temp alarm can be
made(the revise range:±1℃).
After finishing operation above press “set” button twice, the equipment goes back to normal working
state.
Be careful:
The skin temp sensor’s head and the standard thermometer must be put under the same temp(36℃-
37℃), then find out the deviation and revise after it is steady.
3
Infant radiant warmer Service manual
4
Infant radiant warmer Service manual
4 Functions instruction
4.1 LCD display
5
Infant radiant warmer Service manual
Fig.7-2
6
Infant radiant warmer Service manual
Notice: when inspecting this function, the alarm time should not be too long for fear of consuming
much battery energy
9.1.2 Lock the brake wheel of the warmer tight so as to prevent it from moving while in use.
9.1.3 Insert the skin temp sensor into the skin temp sensor jack.
9.1.4 Insert one end of the power wire into the power input jack behind the warmer, insert the
other end into AC 220V, 50Hz power socket.
Be careful: the grounding of the power socket must be tight and reliable
APGAR timer have timer reset and APGAR timer start / hold two key, can control APGAR
timer reset, start or hold.
Main screen
have one
lock key to
lead into main menu, press ,main screen upper right have a choice
7
Infant radiant warmer Service manual
8
Infant radiant warmer Service manual
9
Infant radiant warmer Service manual
Press ,
During
auto
mode,
and in
whole state(100%
heating) skin temp
sensor induction
temperature not
change the system
will take protect limit
temp action(30%
10
Infant radiant warmer Service manual
Press key,
reduce set
value, press key
increase set value,
adjustment range is 0-
10 , stepping is 2 ,
corresponding heating
power is 0-50%. Press
11
Infant radiant warmer Service manual
12
Infant radiant warmer Service manual
Aux temp set value can make one auxiliary temperature alarm point as one auxiliary function of over
temperature detection, temperature monitoring. If aux real temperature over the set value will have
over temperature alarm, audio and video indication same time cut off power stop heating.
13
Infant radiant warmer Service manual
W
hen amending the aux temp displayed value, press
increase, press reduce, default is 2.0, amend range
is 0-4.0 , corresponding temperature is ±2.0℃
When amending skin temperature display value,
press increase, press reduce,amend range is 0-
4.0, corresponding temperature is ±2.0℃
W
hen
amending over temp alarm, press increase, press
reduce, amend range is 0-2.0, corresponding alarm
temperature is 37.0℃-39.0℃
Press or key can downward or upward choice 4 type situation.
Press key can exit main menu or auto exit
main menu.
14
Infant radiant warmer Service manual
When the power supply is cut off abruptly or has troubles, it will have alarm both audio and visual, you
should turn off the switch immediately (long time alarm will consume the energy of the battery, it may
affect the alarm next time), and check it. When the trouble is eliminated, turn on the device again.
Warning: (power supply breaks off or reset) after turn on the device again, you must check the control
mode again and set the temperature.
15
Infant radiant warmer Service manual
! automatically at least 6 hours, insuring the sufficiency of power to alarm for power
failure.
The battery could be changed by authorized person only!
In the process of examining the alarm of power failure, if the alarm can not sound, or the sound is too
small, charge the battery up or change a new one in time in case that the battery liquid flow out and
damage the parts of the equipment. If the battery liquid flowed out, must wash it cleanly with the cleanser,
so as not to corrode the equipment. While changing the battery, pull out the plug of supply power firstly,
and unload the screws of the operation panel. Demount the rechargeable battery which lies in the PCB
board with electric iron and solder a new one (model:D80H5A2H )onto the circuit board. Collect them
together and dispose them to avoid environment polluting. Don’t abandon the changed batteries at will.
16
Infant radiant warmer Service manual
17
Infant radiant warmer Service manual
a. The skin temp sensor has not been Insert the skin temp sensor.
inserted. Check the plug and the
Sensor alarm b. The plug of the skin temp sensor falls off connection.
or bad tangency. Replace the skin temp
c. Skin temp sensor failure. sensor.
Panel button
Panel button failure. Replace the panel..
failure
a. Spotlight lamp socket connect not well
a. deal with the socket
b. Power cable connector not well
Spotlight lamp b. fixed the power cable
c. Transformer or treatment lamp
not bright c. change transformer
transformer is damage
d. change lamp
d. LED lamp is damage
18
Infant radiant warmer Service manual
7. APPENDIXⅠ
7.1. Use
This product is used for light treatment to reduce concentration of bilirubin in the body of infants.
7.2. Features
7.2.1 With its small bulk and light weight, the product has a high efficiency of radiation.
7.2.2.With electron adaptor, power have high adaptive capacity
7.2.3 LCD display the total time of therapy.
7.2.4 With low,middle,high three level to adjust phototherapy radiant.
7.2.5 Auto save 20000 hours accumulated use time.
Uniformity >0.4
19
Infant radiant warmer Service manual
7.4. Installation
Instruction: This type phototherapy unit is specially designed to be installed on the infant radiant warmer. So
before installing this phototherapy unit, please install the infant radiant warmer first, then install this phototherapy
unit.
20
Infant radiant warmer Service manual
packing of phototherapy unit, insert phototherapy fixed seat into the slideway of side of warmer. Then adjust
phototherapy unit to suitable height(advice 450-500mm). Tighten M8*20 screw thread handle.
slideway
7.4.4 When use,adjust rotating arm to the clinic needed position. Reset it after use.
Arm can be adjusted
21
Infant radiant warmer Service manual
7.4.5 Make sure the radiant warmer main body connect well (including grounding connect well), insert output
end of special phototherapy connect cable into power socket of rear of warmer.
Front panel
1.prot
ect
22
Infant radiant warmer Service manual
label
2.Irradiance key .Adjust the irradiance level by press key. sign means high level sign means middle level sign
means low level sign means pause
3. Clear key: Clear time of this phototherapy.
4.Increasing key. a) check total cumulative time b) increasing value when set count-down time.
5.Decreasing key. a)press it few seconds, enter setting count down time. b) decreasing value when set count down
time.
6.LCD display screen
Back panel
OFF Switch ON
7.6.
Instruction
for use
Before treating one patient, phototherapy unit should rinse thoroughly and disinfect, avoiding affect the
phototherapy effect.especially the lens surface.
Put one end of power cable into bear of input connecter.the other end insert auxiliary mains socket-outlet of a
infant warmer.
After use blinder and perineum cover for patient, move patient to suitable position
Switch on phototherapy source and turn on the lamp(default is high level irradiance), timer and treatment is
starting.
The relationship of distance between phototherapy equipment and effective surface,effective surface,irradiance
level and irradiance. Fig7-1
Attention: usually, distance between the phototherapy equipment and the effective surface area is about 50cm,and the
centre of effective surface area is coincide with baby mattress centre, Ebi max(maximum total irradiance for bilirubin) is the
maximumm value of the effective surface; Ebi avg is the average value of total irradiance for bilirubin between Ebi max and
Ebi min.
Warning: when using phototherapy equipment,should choice the suitable level according
to the direct of doctor and the needed of patient. Review the therapy effect after 4 hours, to
decide whether change the irradiance level or not.
23
Infant radiant warmer Service manual
This unit have high,middle,low three level.can adjust the irradiance by press key,and it will display on the
screen.the order is :high – middle – low – pause.
As the fig 7-2. means high level; means middle
level, means low level. All irradiance sign not lighten means phototherapy now is paused.
Irradiance level
Fig.7-2
Attention : when switch off phototherapy unit or cut off power,the unit can remember the
level of irradiance.
It is necessary to measure patient’s body temperature and give patient some water
according to needed.
2) Phototherapy unit have the function to auto turn off the lamp at the scheduled time and accumulation timing.if not
set this function,default treating time is 100 hours,then stop auto. Accumulation time is the total value from initial
value to now. and is display as “xxxxx.x”. The max.time is 19999.9 hours.
3) Reset timer: press and don’t move. A few seconds later,timer will reset to zero.
4) Set the time for phototherapy (turn off the lamp on time). As fig.8-3
a) press key some seconds,enter count down mode,default value is 1 hour. Then can press or to increase or
reduce count down time.the range is : 30 minutes *n (n=1-96),when n=97 and display 0, then the range of irradiate
time is 00:30:00-48:00:00’
b) if set the irradiate time when the lamp is lighten,then the irradiance level not change; if set the irradiate time when
the lamp paused,then the irradiance level is high level. And the timer is starting count down till display 0. then
irradiate stopped.(display 03:30:00 means it still have three hours and half till end of irradiate.)
24
Infant radiant warmer Service manual
Fig.8-3
5)check the time :press some seconds,timer display cumulative time,then auto change former value.(attention ,the
unit is hour,display with decimalism)
If want to end the phototherapy,switch off phototherapy unit and cut off power.
Attention:
1).normally is increasing timer for display phototherapy time, the max value is 100 hours;count down timer is
decreasing timer for display left phototherapy time,the max value is 48 hours.
2).if turn off the phototherapy,the timer display 0,but it will be remembered as accumulative time unless use CLEAR
key.
.
Plug in, some LED 1. LED lamps are broken 1.Change the LED lamp
lamps are not lightening 2. Element protection 2. Inspect the circuit.
Calculagraph are not 1. Line panel get loose 1.Weld the lines well
lightening Circuit board break down. 2.Change for a new circuit board.
25
Infant radiant warmer Service manual
8. APPENDIXⅡ:
Low-pressure suction apparatus (JE-DY-1)
Operating Manual
II. OPERATION
AND
MAINTENANCE
a. Operation
• Inspect the
apparatus
according to the procedure of assembly and adjustment. Ensure its performances are satisfactory.
Then connect the disinfected suction hose and catheter used in clinic and set vacuum level by
adjusting the vacuum regulator, then ready for operation.
26
Infant radiant warmer Service manual
• The storage bottle should be emptied out immediately when the liquid reaches near to 1/2 capacity
of it, which can prevent the pump from damaging.
Note: The aspirator should be operated by professional doctor according to the directions of medical science and
the instruction in the operating manual strictly.
If there is any doubt about operation, please get in touch with supplier or manufacturer.
b. Replacement for fuses
When the unit fuse fixed behind the unit case need to be replaced, cut off the power supply first,
align to the cross-groove of the fuse seating with a screw-driver and turn around in counter-
clockwise to withdraw the seating cover, then inspect or replace the fuses.
c. Maintenance
• Before stop working, sucking a little clear water into the suction tubing will be helpful to make it
clean.
• After stop working, empty storage bottle, eliminate dirt inside storage bottle with soft brush or
cloth and then rinse those including the suction tubing throughout.
• It is recommended that immerse the storage bottle and caps as well as suction tubing into
kangweida disinfect solution (refer to its instructions) for one hour.
Note: Avoid the storage bottles made of glass being knocked against sharp body or falling down.
• The outer casing of the unit may be wiped with disinfected cloth which should be twisted dryer as
to prevent liquid from seeping into the chinks of case.
• The unit should be kept in a dry and clean place, and run it periodically ( once a week in normal
use) to ensure the electronic components not to damage.
III. TROUBLESHOOTING
Note : It is recommended that the pump should be unassembled or repaired by professionals. If necessary, please get in touch with
27
Infant radiant warmer Service manual
the manufacturer.
9. APPENDIX III:
Infant resuscitator (BR-100)
Operating Manual
9.1 Features
a) PIP (Peak Inspiratory Pressure) and PEEP (Positive End Expiratory Pressure) are pre-set by the user,
and displayed on the manometer.
b) Accurately regulate and maintain a constant and uniform PIP (Peak Inspiratory Pressure).
c) Accurately regulate and maintain a constant and uniform PEEP (Positive End Expiratory Pressure).
d) Respiratory rate and IE (inspiration and expiration ratio) are controlled by the time of pressing and
releasing air outlet.
e) Tidal volume is adjusted by adjusting the gas flow and the speed of pressing.
f) The maximum airway pressure can be set.
g) If really necessary, oxygen with content of 100% is available.
h) Simple to operate and the operator can complete with less fatigue.
i) Easily portable.
28
Infant radiant warmer Service manual
The Max Pressure Relief that may be delivered to the patient. (factory set at 40
cmH2O)
29
Infant radiant warmer Service manual
Performance Specifications
The relation between the adjustment range of the PIP and the gas supply
Input Gas Flow Peak Inspiratory Pressure(PIP)
(L/min) cmH2O (mbar)
*5 2~70
*8 3~74
*10 4~76
*15 10~77
The relation between the adjustment range of the PEEP and the gas supply
Input Gas Flow Positive End Expiratory Pressure(PEEP)
(L/min) cmH2O (mbar)
*5 1~6
*8 1.5~10
*10 3~15
*15 5~25
Input Gas Flow 5L/min~15L/min
Note:Figures with* listed above are representative only.
PEEP and PIP values stated are typical ones. Higher PEEP values can be achieved if higher
PIP values are set.
NOTE:Factory setting of pressure limit valve is at 40 cmH 2O (mbar). Pressure limit valve is to
limit the pressure in the circuit. In particular cases, adjust the upper limit value of pressure to,
or the pressure in the circuit will not exceed 40 cmH2O (mbar).
Technical Specification
Item Parameters
Height 270 mm
Width 200 mm
Depth 110 mm
Weight 1.8 kg
Manometer Range -10 cm~+80 cmH2O
Manometer accuracy ±2%
Storage Temperature: -40℃~+60℃;
environmental condition Humidity: ≤95% RH
Operating Temperature: -18~+50℃;
environmental condition Humidity: 5%~95%RH
Dead Space 6 mL
Recommended Patient Body Weight ≤10 Kg
Delivered Oxygen Concentration Over 85%
30
Infant radiant warmer Service manual
5.
31
Resuscitation
Mask
32
Infant radiant warmer Service manual
33
Infant radiant warmer Service manual
34
Infant radiant warmer Service manual
35
Infant radiant warmer Service manual
36
Infant radiant warmer Service manual
9.6. To resuscitate
① Set the clinical specifications according to the Chapter9.5.4 and 9.5.5
② Fit the resuscitation mask to the PEEP valve output port and place over the baby’s mouth and nose.
37
Infant radiant warmer Service manual
③ Resuscitate by pressing and removing over the PEEP outlet port alternating with forefinger to allow
patient for inspiration and expiration. (Refer to below figure)
38
Infant radiant warmer Service manual
cleaning procedures.
c) Dust all surfaces with a clean damp soft cloth.
d) After the clean of the air supply tube, put it in the specific clean place for using next time.
e) In normal case, clean and maintenance is performed rarely.
9.7.2 Sterilization
9.7.2.1 Test Lung
Test lung can be disinfected by autoclaving. Recommended way: It can be disinfected by autoclaving at
121℃, 15psi for 30 minutes. Since there are different sizes and types of autoclave, the
pressure, temperature, humidity and time are different. Please refer to the manufacturer's
instructions or the standard of the sterilization. The way of disinfection with liquid is not
recommend since some agentia may result in the product discoloration.
9.7.2.2 Resuscitation masks
Disinfect before using, as in the above 10.2.1.
9.7.2.3 Resuscitation breathing tube
Resuscitation breathing tube is a single-use part. It can be used if the package is complete. (It is
suggested to disinfect it with ethylene oxide before using)
9.7.2.4 Gas Supply tube
It can be reusable. Clean and disinfect it regularly as requirement with ethylene oxide.
9.7.3 Notice: Do not autoclave or gas-sterilize any part of the main body of the infant
resuscitator.
39
Infant radiant warmer Service manual
40
Infant radiant warmer Service manual
41
Infant radiant warmer Service manual
42
Infant radiant warmer Service manual
43
Infant radiant warmer Service manual
3 M6 screw 4 J030058
5 M8 handle 1 J040019
6 M4 screw 4 J040086
44