You are on page 1of 97

Transmitters for Hazardous

Areas
Series HMT360

USER'S GUIDE
M010056EN-E
August 2005
PUBLISHED BY

Vaisala Oyj Phone (int.): +358 9 8949 1


P.O. Box 26 Fax: +358 9 8949 2227
FIN-00421 Helsinki
Finland

Visit our Internet pages at http://www.vaisala.com/

© Vaisala 2005

No part of this manual may be reproduced in any form or by any means,


electronic or mechanical (including photocopying), nor may its contents be
communicated to a third party without prior written permission of the copyright
holder.

The contents are subject to change without prior notice.

Please observe that this manual does not create any legally binding obligations for
Vaisala towards the customer or end user. All legally binding commitments and
agreements are included exclusively in the applicable supply contract or
Conditions of Sale.
________________________________________________________________________________

Table of Contents

CHAPTER 1
GENERAL INFORMATION............................................................................ 7
About This Manual ................................................................... 7
Version Information ............................................................... 7
Related Manuals ................................................................... 7
General Safety Considerations ............................................. 8
Feedback............................................................................... 8
Product Related Safety Precautions ...................................... 9
ESD Protection ......................................................................... 9
Recycling ................................................................................ 10
Regulatory Compliances ....................................................... 10
European requirements....................................................... 10
US requirements (FM)......................................................... 10
Canadian requirements (CSA) ............................................ 11
Australian requirements (TestSafe) .................................... 11
Trademarks ............................................................................. 11
License Agreement ................................................................ 11
Warranty.................................................................................. 11

CHAPTER 2
PRODUCT OVERVIEW................................................................................ 13
Output Quantities ................................................................ 14
Probe Options ..................................................................... 14
Maintenance........................................................................ 15

CHAPTER 3
INSTALLATION............................................................................................ 17
General Instructions .............................................................. 17
Mounting Probe Cable in Gas Group IIC Spaces ............... 17
Selecting Location............................................................... 18
Temperature Differences .................................................... 18
Checking Temperature Reading ......................................... 18
Mounting of Transmitter ...................................................... 19
Mounting the Probes.............................................................. 20
HMP362 - Small Pressure-Tight Probe............................... 21
HMP363 Probe for Confined Spaces.................................. 23
HMP365 Probe for High Temperatures .............................. 24
HMP364 for Pressurized Spaces........................................ 24
HMP368 Probe For Measuring Humidity in Pressurized
Pipelines or Moisture in Liquids .......................................... 27
Tightening Clasp Nut ..................................................... 29
Installing Sensor Head Through Ball Valve Assembly........ 30

VAISALA ________________________________________________________________________ 1
User's Guide ______________________________________________________________________

CHAPTER 4
ELECTRICAL CONNECTIONS....................................................................33
Installation in Hazardous Locations.....................................34
US and Canadian Requirements.........................................34
European Requirements .....................................................34
CATEGORY 1 (Zone 0)..................................................34
CATEGORY 2 or 3 (Zone 1 or 2) ...................................34
Maximum Cable Resistance Calculation for the Barrier
(Vaisala Code: 210664)..................................................35
HMT360 Connected to Galvanic Isolator ............................36
HMT360 Connected to Zener Barrier ..................................37
Examples of Connections....................................................38
Grounding............................................................................39

CHAPTER 5
OPERATION.................................................................................................41
HMT360 with Display..............................................................41
HMT360 for Natural Gas .....................................................41
HMT360 without Display ........................................................42
DIP Switch Functions.............................................................42

CHAPTER 6
OPERATION ERRORS ................................................................................45
Cleaning Sensor from Glycol in Natural Gas Application..45

CHAPTER 7
DISPLAY/KEYPAD COMMANDS................................................................47
Setting Pressure for Calculations.........................................47
Selecting Output Quantities ..................................................48
HMT360 for Natural Gas .....................................................48
Upper Half of Display ..........................................................48
Lower Half of Display ..........................................................48
Selecting Analog Outputs......................................................49
Scaling Analog Outputs.........................................................50

CHAPTER 8
SERIAL INTERFACE ...................................................................................53
Serial Communication Settings ............................................54
ASEL Selecting Analog Outputs ...........................................54
S Scaling Analog Outputs .....................................................55
Adjustment Commands .........................................................56
CRH Relative Humidity Adjustment.....................................56
CT Temperature Adjustment ...............................................56
ACAL Analog Output Calibration.........................................57
Output Commands .................................................................57
ITEST Testing Analog Outputs............................................57
SEND Outputting Measurement Values..............................58
R Activating Continuous Output ..........................................58
S Stopping Continuous Output............................................58

2 ___________________________________________________________________ M010056EN-E
________________________________________________________________________________

INTV Setting Output Interval ............................................... 58


PRES Setting Ambient Pressure for Calculations .............. 59

CHAPTER 9
MEASURING AT OVERPRESSURE ........................................................... 61
Pressure Regulator Recommended ..................................... 61

CHAPTER 10
CALIBRATION AND ADJUSTMENT........................................................... 63
Calibration Interval................................................................. 63
Factory Calibration and Adjustment .................................... 63
User Calibration and Adjustment ......................................... 63
Removing Electronics Unit for Calibration .......................... 64
Connections ........................................................................ 65
Calculating Correspondence of Current Values and Output
Quantities................................................................................ 66
Relative Humidity Calibration ............................................... 67
Automatic Two-Point Calibration (Only with HMT360 with
Display) ............................................................................... 67
Manual Calibration .............................................................. 69
Low End Adjustment ...................................................... 69
High End Adjustment (Two-Point Calibration) ............... 70
Temperature Calibration in One Point ................................ 71

CHAPTER 11
SPECIFICATIONS........................................................................................ 73
Performance ........................................................................... 73
Relative Humidity ................................................................ 73
Moisture in Natural Gas (HMT362N/364N/368N) ............... 74
Dewpoint Measurement ................................................. 74
Calculated Variables Available (Typical Ranges) ............... 74
Water Activity in Jet Fuel Applications ................................ 75
Temperature........................................................................ 75
Calculated Variables (Typical Ranges)............................... 76
With HMP361 Probe ...................................................... 76
With HMP363, HMP364, HMP365 & HMP368 Probes.. 76
Outputs.................................................................................... 76
Classification with Current Outputs..................................... 77
General .................................................................................... 78
Probes ..................................................................................... 79
Options and Accessories ...................................................... 80
Accuracy of Calculated Variables (Not Valid with HMT360
for Natural Gas) ...................................................................... 81
Accuracy of Dewpoint Temperature °C ......................... 81
Accuracy of Mixing Ratio g/kg........................................ 81
Accuracy of Wet Bulb Temperature °C.......................... 82
Accuracy of Absolute Humidity g/m³ .............................. 82

VAISALA ________________________________________________________________________ 3
User's Guide ______________________________________________________________________

APPENDIX A
DIMENSIONS ...............................................................................................83
HMP361....................................................................................83
HMP362....................................................................................84
HMP363....................................................................................85
HMP364....................................................................................85
HMP365....................................................................................85
HMP368....................................................................................86
Mounting Plate ........................................................................86
Rain Shield ..............................................................................87
Protection Cover.....................................................................88

APPENDIX B
WIRING FOR INTRINSICALLY SAFE OPERATION, FM ...........................89

APPENDIX C
WIRING FOR INTRINSICALLY SAFE OPERATION, CSA.........................91

APPENDIX D
CERTIFICATES ............................................................................................93

List of Figures
Figure 1 Transmitter for Hazardous Areas Series HMT360 ...................13
Figure 2 HMT360 Probe Options............................................................14
Figure 3 Mounting the Transmitter and Detaching the Probe ................19
Figure 4 Different Parts of the Transmitter .............................................20
Figure 5 HMP362 Installation (without Sampling Cell) ...........................21
Figure 6 HMP302SC (Optional Sampling Cell) ......................................22
Figure 7 Installation of Sensor Head HMP363 in Channel with Flange
and Supporting Bar...................................................................23
Figure 8 Mounting HMP365 Sensor Head in Duct or Channel...............24
Figure 9 Mounting the HMP364 Probe ...................................................25
Figure 10 Marking Nut and Fitting Screw .................................................26
Figure 11 Cleaning Tightening Cone ........................................................26
Figure 12 HMP368 Probe Dimension: Standard 178 mm and Optional 400
mm............................................................................................28
Figure 13 Tightening Clasp Nut ................................................................29
Figure 14 Installing Sensor Head Through DMP248BVS Ball Valve
Assembly ..................................................................................30
Figure 15 HMT360 Connected to Galvanic Isolator .................................36
Figure 16 HMT360 Connected to Zener Barrier .......................................37
Figure 17 STAHL 9160/13-11-11 (Galvanic Isolator) ...............................38
Figure 18 STAHL 9001/51-280-091-141 (Zener Barrier)..........................38
Figure 19 Grounding .................................................................................39
Figure 20 DIP Switch Functions ...............................................................42
Figure 21 Cleaning Sensor .......................................................................46
Figure 22 Detaching Electronics Unit with Probe for Calibration..............64
Figure 23 Connecting Power Supply and Multimeter for Calibration........65

4 ___________________________________________________________________ M010056EN-E
________________________________________________________________________________

Figure 24 Display Chart of the Automatic Calibration Procedure............. 68


Figure 25 Effect of Process Temperature on Dewpoint Measurement
Accuracy................................................................................... 74

List of Tables
Table 1 Manual Revisions ....................................................................... 7
Table 2 Related Manuals ........................................................................ 7
Table 3 Available Quantities.................................................................. 49
Table 4 Serial Communications Settings .............................................. 54
Table 5 Output Quantities and Their Abbreviations .............................. 55
Table 6 Pressure Conversion Chart...................................................... 60
Table 7 Relative Humidity Specifications .............................................. 73
Table 8 Dewpoint Measurement Specifications .................................... 74
Table 9 Calculated Variables Specifications ......................................... 74
Table 10 Water Activity in Jet Fuel Applications Specifications.............. 75
Table 11 Temperature Specifications ..................................................... 75
Table 12 HMP361 Calculated Variables Specifications .......................... 76
Table 13 HMP363, HMP364, HMP365 & HMP368 Probes Calculated
Variables Specifications ........................................................... 76
Table 14 Output Specifications ............................................................... 76
Table 15 General Specifications ............................................................. 78
Table 16 Probe Specifications................................................................. 79
Table 17 Options and Accessories Table ............................................... 80
Table 18 Accuracy of Dewpoint Temperature °C.................................... 81
Table 19 Accuracy of Mixing Rratio g/kg................................................. 81
Table 20 Accuracy of Wet Bulb Temperature °C .................................... 82
Table 21 Accuracy of Absolute Humidity g/m³ ........................................ 82

VAISALA ________________________________________________________________________ 5
User's Guide ______________________________________________________________________

This page intentionally left blank.

6 ___________________________________________________________________ M010056EN-E
Chapter 1 ________________________________________________________ General Information

CHAPTER 1
GENERAL INFORMATION

This chapter provides general notes for the product.

About This Manual


This manual provides information for installing, operating, and
maintaining Transmitters for Hazardous Areas Series HMT360.

Version Information
Table 1 Manual Revisions
Manual Code Description
U336en-1.1 1st version, March 1999
M010056en-A 2nd version, January 2001
M010056en-B 3rd version, March 2002
M010056en-C 4th version, November 2002
M010056EN-D 5th version, May 2004
M010056EN-E 6th version, July 2005

Related Manuals
Table 2 Related Manuals
Manual Code Manual Name
M210185en-B Humidity Calibrator HMK15 User's Guide

VAISALA ________________________________________________________________________ 7
User's Guide ______________________________________________________________________

General Safety Considerations


Throughout the manual, important safety considerations are
highlighted as follows:

WARNING Warning alerts you to a serious hazard. If you do not read and follow
instructions very carefully at this point, there is a risk of injury or
even death.

CAUTION Caution warns you of a potential hazard. If you do not read and
follow instructions carefully at this point, the product could be
damaged or important data could be lost.

NOTE Note highlights important information on using the product.

Feedback
Vaisala Customer Documentation Team welcomes your comments
and suggestions on the quality and usefulness of this publication. If
you find errors or have other suggestions for improvement, please
indicate the chapter, section, and page number. You can send
comments to us by e-mail: manuals@vaisala.com

8 ___________________________________________________________________ M010056EN-E
Chapter 1 ________________________________________________________ General Information

Product Related Safety Precautions


The Transmitter for Hazardous Areas Series HMT360 delivered to
you has been tested for safety and approved as shipped from the
factory. Note the following precautions:

WARNING Ground the product, and verify outdoor installation grounding


periodically to minimize shock hazard.

CAUTION Do not modify the unit. Improper modification can damage the
product or lead to malfunction.

ESD Protection
Electrostatic Discharge (ESD) can cause immediate or latent damage
to electronic circuits. Vaisala products are adequately protected
against ESD for their intended use. However, it is possible to damage
the product by delivering electrostatic discharges when touching,
removing, or inserting any objects inside the equipment housing.

To make sure you are not delivering high static voltages yourself:

- Handle ESD sensitive components on a properly grounded and


protected ESD workbench. When this is not possible, ground
yourself to the equipment chassis before touching the boards.
Ground yourself with a wrist strap and a resistive connection cord.
When neither of the above is possible, touch a conductive part of
the equipment chassis with your other hand before touching the
boards.
- Always hold the boards by the edges and avoid touching the
component contacts.

VAISALA ________________________________________________________________________ 9
User's Guide ______________________________________________________________________

Recycling
Recycle all applicable material.

Dispose of batteries and the unit according to statutory regulations.


Do not dispose of with regular household refuse.

Regulatory Compliances

European requirements
EU directive 94/9/EC (ATEX 100a).

Standards EN 50014 and EN 50020.

Standard EN 50284 for apparatus of equipment group II, category 1 G


(intended for hazardous locations).

Marking EEx ia IIC T4: at temperature range -20 ... +60 °C and
pressure range 0.8 ... 1.1 bar.

US requirements (FM)
FM approved for Classes I, II & III, Division 1, Groups A, B, C, D, E,
F & G and Division 2, Groups A, B, C, D, F and G. Intrinsically safe
when connected as shown in Appendix B.

10 __________________________________________________________________ M010056EN-E
Chapter 1 ________________________________________________________ General Information

Canadian requirements (CSA)


CSA approved for Class I, Division 1 and 2, Groups A, B, C, D; Class
II Division 1 and 2, Groups G and Coal Dust; Class III. Intrinsically
safe when connected as shown in Appendix C.

Australian requirements (TestSafe)


TestSafe approved for Ex ia IIC T5 IP65 (Zone 0).

Trademarks
HUMICAP® is a registered trademark of Vaisala.

License Agreement
All rights to any software are held by Vaisala or third parties. The
customer is allowed to use the software only to the extent that is
provided by the applicable supply contract or Software License
Agreement.

Warranty
For certain products Vaisala normally gives a limited one-year
warranty. Please observe that any such warranty may not be valid in
case of damage due to normal wear and tear, exceptional operating
conditions, negligent handling or installation, or unauthorized
modifications. Please see the applicable supply contract or Conditions
of Sale for details of the warranty for each product.

VAISALA _______________________________________________________________________ 11
User's Guide ______________________________________________________________________

This page intentionally left blank.

12 __________________________________________________________________ M010056EN-E
Chapter 2 __________________________________________________________ Product Overview

CHAPTER 2
PRODUCT OVERVIEW

The series HMT360 transmitters are microprocessor based two-wire


transmitters for measuring relative humidity and temperature in
hazardous areas.

NOTE Connect the transmitter always via galvanic isolators or Zener


barriers in hazardous environments.

Figure 1 Transmitter for Hazardous Areas Series HMT360

VAISALA _______________________________________________________________________ 13
User's Guide ______________________________________________________________________

Output Quantities
The transmitter has a local display and two output current channels.
The available quantities are described in Table 3 on page 49.

Probe Options
The HMT360 transmitter series has various options for transmitter
units, sensor heads and cable lengths (2, 5 and 10 m). The available
probe types are presented in Figure 2.

0505-265

Figure 2 HMT360 Probe Options

The following numbers refer to Figure 2 above:


1 = HMP361 probe for wall mounting
2 = HMP362 probe for pressurized spaces up to 167 bars
3 = HMP363 probe head (Ø 13.5 mm) for tight spaces
4 = HMP364 probe for pressurized spaces up to 100 bars
5 = HMP365 probe for high temperatures up to 180°C
6 = HMP368 probe for installations in pressurized pipelines up to
40 bars; pressure-tight sliding clasp nut

14 __________________________________________________________________ M010056EN-E
Chapter 2 __________________________________________________________ Product Overview

Maintenance
The electronics unit of the transmitter can be removed and replaced in
the field, as well as the probe; see section Removing Electronics Unit
for Calibration on page 64 for detailed information.

All other maintenance is performed at the factory. If a transmitter is


damaged, contact Vaisala or Vaisala distributor and send the
instrument to the addresses below:

Please, include a short description of the fault, and the date and place
of purchase.

VAISALA _______________________________________________________________________ 15
User's Guide ______________________________________________________________________

This page intentionally left blank.

16 __________________________________________________________________ M010056EN-E
Chapter 3 _______________________________________________________________ Installation

CHAPTER 3
INSTALLATION

General Instructions
NOTE Mounting Probe Cable in Gas
Group IIC Spaces
The following applies only to installation within the EU!

The following instructions shall be followed to fulfil the


specifications of the EN 50284 for non conductive layer of the probe
cable:
- Never mount or handle the probe cable when hazardous gases are
present.
- For achieving a conductive shield, cover the probe cable with
conductive material like metal or conductive tape or mount the
probe cable in a metal conduit.
- Assure that the conductive shield fulfils requirements of the
standard EN50284 (resistance less than 1 GΩ) and make sure that it
can not come loose in any operation situation.

NOTE During the installation work of the sensor heads in gas group IIC
areas (requiring category I devices), it has to be guaranteed that even
in fault cases sparks generated by impacts or friction on the surface of
the housing can never occur.

VAISALA _______________________________________________________________________ 17
User's Guide ______________________________________________________________________

Selecting Location
Select a place with stable conditions for mounting the transmitter. Do
not expose the transmitter to direct sunlight or rain. A rain shield for
HMT360 transmitter is available. When mounting the sensor head,
select a place representing the process conditions.

Temperature Differences
If there is a difference between the process and ambient temperatures,
it is recommended to mount the sensor head and at least one meter of
cable inside the process. This is to avoid incorrect relative humidity
readings due to temperature difference. If this installation is not
possible and the sensor head must be installed through the wall (for
example, with a flange), the sensor head and the cable must be
insulated carefully.

Checking Temperature Reading


The actual temperature of the process can be measured with a
reference instrument to be compared with the transmitter reading. The
heat transfer is less evident if you remove the protective filter of the
sensor for a short-term test. However, never use the sensor long
periods without the filter as this may cause a faster contamination of
the sensor. The transmitter fulfils the specified EMC regulations with
the protective filter on the sensor head.

18 __________________________________________________________________ M010056EN-E
Chapter 3 _______________________________________________________________ Installation

Mounting of Transmitter

0505-266

Figure 3 Mounting the Transmitter and Detaching the Probe

1. Attach the mounting plate to the wall with 4 screws.


2. Press down the transmitter so that it slides along the rails of the
mounting plate.
3. Fasten the transmitter to the mounting plate with the Allen
screw1).
4. The probe can be detached and replaced when needed by simply
unfastening the two Allen screws.
1) 3 mm Allen key provided

VAISALA _______________________________________________________________________ 19
User's Guide ______________________________________________________________________

0505-267

Figure 4 Different Parts of the Transmitter

The following numbers refer to Figure 4 above:


1 = Electronics unit
2 = Transmitter base
3 = Probe; including a part of measurement electronics (for
example, calibration memory and the sensor head)
4 = Protective covers
5 = Flat cable
6 = Sensor head
7 = Cable glands
8 = Grounding terminals
9 = RS232 connector

Mounting the Probes


NOTE - Do not unsolder and then again resolder the sensor head cable from
and to the printed board during installation.
- Do not shorten or lengthen the probe cable

These procedures may alter the humidity calibration of the


transmitter.

20 __________________________________________________________________ M010056EN-E
Chapter 3 _______________________________________________________________ Installation

HMP362 - Small Pressure-Tight


Probe
The HMP362 probe is a small pressure-tight probe equipped with
installation flange. It is suitable for dewpoint measurements in natural
gas. When sampling in pressurized processes, the sampling cell
HMP302SC is available as an optional accessory.

0505-268

Figure 5 HMP362 Installation (without Sampling Cell)

VAISALA _______________________________________________________________________ 21
User's Guide ______________________________________________________________________

4
1

5
2
6

0505-269

Figure 6 HMP302SC (Optional Sampling Cell)

The following numbers refer to Figure 6 above:


1 = Gas in
2 = Probe
3 = Gas out
4 = Probe
5 = Sampling cell
6 = Clamp (not needed if sampling cell is suspended on the
piping)

CAUTION In pressurized processes it is essential to tighten the supporting nuts


and screws very carefully to prevent loosening of the probe by the
action of pressure.

NOTE If the HMP362 is installed in a process with a pressure differing from


normal atmospheric pressure, please enter the pressure value of the
process (in hPa or mbar) into the transmitter memory via the serial
line (see page 59) or via the display/keypad (see page 47).

22 __________________________________________________________________ M010056EN-E
Chapter 3 _______________________________________________________________ Installation

HMP363 Probe for Confined Spaces


The HMP363 is a small size general-purpose probe suitable for
installation in ducts and channels with the installation kit available.
The installation kit includes a flange, a supporting bar for the sensor
head cable and screws for attaching the flange to the duct wall. The
distance between the sensor and the channel wall can be adjusted. The
adjustment range is 100 - 320 mm, measured from the tip of the sensor
head to the flange.

1
2

70 3
Ø13.5
4

6 5

0505-270

Figure 7 Installation of Sensor Head HMP363 in Channel


with Flange and Supporting Bar

The following numbers refer to Figure 7 above:


1 = Duct wall
2 = Plugged hole for reference measurements is recommended
3 = Flange
4 = Seal
5 = Supporting bar
6 = Sensor head

VAISALA _______________________________________________________________________ 23
User's Guide ______________________________________________________________________

HMP365 Probe for High


Temperatures
When installing the HMP365 sensor head in a duct or channel, there
should not be significant difference between temperatures inside and
outside of the duct. This can cause incorrect humidity readings.

0505-271

Figure 8 Mounting HMP365 Sensor Head in Duct or Channel

The following numbers refer to Figure 8 above:


1 = A plugged hole for reference measurements
2 = Sensor head
3 = Mounting flange

HMP364 for Pressurized Spaces


The HMP364 pressure-tight probe is designed for the humidity
measurements in pressurized rooms and industrial processes. This
probe is also available for dewpoint measurements in natural gas. The
probe is provided with a nut, a fitting screw and a sealing washer to
make the installation easier.

CAUTION In pressurized processed it is essential to tighten the supporting nuts


and screws very carefully to prevent loosening of the probe by the
action of pressure.

24 __________________________________________________________________ M010056EN-E
Chapter 3 _______________________________________________________________ Installation

1. Unscrew the fitting screw from the nut and the sensor head.
2. Fasten the fitting screw to the chamber wall with a sealing
washer. Tighten the fitting screw into the threaded sleeve with a
torque spanner. The tightening torque is 150 ± 10 Nm (110 ± 7
ft-lbs)
3. Insert the body of the sensor head into the fitting screw and
screw the nut manually to the fitting screw until the connection
feels tight.
4. Mark both the fitting screw and the nut hex (see Figure 10).
5. Tighten the nut a further 30° (1/12 turn) or if you have a torque
spanner tighten it with a torque of 80 ± 10 Nm (60 ± 7 ft-lbs).

NOTE After detachment the nut must be tightened without increased effort.

6. Clean and grease the tightening cone of the fitting screw after
every tenth detachment. Change the sealing washer every time
the fitting screw is detached. Use high-vacuum grease (for
example; Down Corning, Europe) or a similar grease.

0505-272

Figure 9 Mounting the HMP364 Probe

The following numbers refer to Figure 9 above:


1 = Tightening cone
2 = Nut
3 = Fitting screw
4 = Sealing washer
5 = Sensor head; Ø 13.5 mm

VAISALA _______________________________________________________________________ 25
User's Guide ______________________________________________________________________

0505-273

Figure 10 Marking Nut and Fitting Screw

0505-274

Figure 11 Cleaning Tightening Cone

The following numbers refer to Figure 11 above:


1 = Fitting screw
2 = Sealing washer
3 = Tightening cone
4 = Clean cotton stick

NOTE If the HMP364 is installed in a process with a pressure differing from


normal atmospheric pressure, please enter the pressure value of the
process (in hPa or mbar) into the transmitter memory via the serial
line (see page 59) or via the display/keypad (see page 47) .

26 __________________________________________________________________ M010056EN-E
Chapter 3 _______________________________________________________________ Installation

HMP368 Probe For Measuring


Humidity in Pressurized Pipelines
or Moisture in Liquids
The HMP368 probe is available with two lengths, standard 178 mm
and optional 400 mm. The probe is provided with a sliding fit which
enables different installation depths and makes the installation and
removal easier in pressurized processes. The probe is specially
designed for the measurements in pipelines. This probe is also
available for dewpoint measurements in natural gas.

When the probe is used in a pressurized pipeline or in liquids (for


example, feeder routes) the sensor head should preferably be installed
directly in the process through a ball valve assembly, see page 30.

CAUTION In pressurized processed it is essential to tighten the supporting nuts


and screws very carefully to prevent loosening of the probe by the
action of pressure.

- When the HMP368 is installed in a process with a pressure


NOTE differing from normal atmospheric pressure, please enter the
pressure value of the process (in hPa or mbar) into the transmitter
memory via the serial line (see page 59) or via the display/keypad
(see page 47).
- Make sure that the temperature at the measurement point is equal
to that of the process, otherwise the humidity reading may be
incorrect.

VAISALA _______________________________________________________________________ 27
User's Guide ______________________________________________________________________

0505-275

Figure 12 HMP368 Probe Dimension: Standard 178 mm and


Optional 400 mm

The following numbers refer to Figure 12 above:


1 = Clasp nut, 24 mm hex head
2 = Fitting screw, 27 mm hex head
A = Probe 178 mm; adjustment range 120 mm
Probe 400 mm; adjustment range 340 mm

CAUTION Take care not to damage the probe body. A damaged body makes the
probe head less tight and may prevent it from going through the clasp
nut.

28 __________________________________________________________________ M010056EN-E
Chapter 3 _______________________________________________________________ Installation

Tightening Clasp Nut


1. Adjust the probe to a suitable depth according to the type of
installation.
2. Tighten the clasp nut first manually.
3. Mark the fitting screw and the clasp nut.
4. Tighten the nut a further 50 - 60° (ca.1/6" turn) with a fork
spanner. If you have a suitable torque spanner, tighten the nut to
45 ± 5 Nm (33 ± 4 ft-lbs).

0505-276

Figure 13 Tightening Clasp Nut

The following numbers refer to Figure 13 above:


1 = Probe
2 = Clasp nut
3 = Pen
4 = Fitting screw

NOTE Take care not to tighten the clasp nut more than 60° to avoid
difficulties when opening it.

VAISALA _______________________________________________________________________ 29
User's Guide ______________________________________________________________________

Installing Sensor Head Through


Ball Valve Assembly
The best way for installing the sensor head to pressurized process or
pipeline is to install it through the ball valve assembly. This
installation makes it unnecessary to empty or shut down the process
for installing or removing the sensor head. Use the Vaisala
DMP248BVS ball valve set or a 1/2" ball valve assembly with a ball
hole of Ø14 mm or more. If you install the sensor head (Ø 13.5 mm)
in a process pipe, please note that the nominal size of the pipe must be
at least 1 inch (2.54 cm). The manual press tool can be used to press
the sensor head into the pressurized (< 10 bar) process or pipeline.

0505-277

Figure 14 Installing Sensor Head Through DMP248BVS Ball


Valve Assembly

The following numbers refer to Figure 14 above:


1 = Manual press tool
2 = Handle of the ball valve
3 = Sensor head
4 = Process chamber/pipeline
5 = Groove on the probe indicating the upper adjustment limit
6 = Filter
7 = Ball of the ball valve

30 __________________________________________________________________ M010056EN-E
Chapter 3 _______________________________________________________________ Installation

Install the sensor head transversely against the direction of the process
flow.

NOTE Installation or removing the probe through a ball valve assembly is


possible without shutting the process when the pressure is less than
10 bar. With 10 ... 40 bar pressures you must shut down the process
before installing or removing the probe. The maximum operation
pressure for the probe HMP368 is 40 bar.

VAISALA _______________________________________________________________________ 31
User's Guide ______________________________________________________________________

This page intentionally left blank.

32 __________________________________________________________________ M010056EN-E
Chapter 4 ______________________________________________________ Electrical Connections

CHAPTER 4
ELECTRICAL CONNECTIONS

Refer to local requirements regarding cabling, grounding and galvanic


isolator or barrier connections.

NOTE Connect the transmitter always via galvanic isolators or Zener


barriers in hazardous environments.

WARNING Be sure that the main power switch of the transmitter is set off before
making any electrical installations in hazardous areas.

1. Open the transmitter cover and remove the protective cover of


the transmitter base.
2. Thread the power supply wires through the cable gland (Figure
4, item 7).
3. Connect the unpowered power supply wires to the connectors:
CH1 (humidity) and CH2 (temperature). Both channels require
an own power supply.

NOTE As CH1 is a main output, the transmitter does not operate if only CH2
is connected (CH2 is optoisolated from transmitter electronics).

4. Replace the protective cover. Turn the transmitter on ON with


the ON/OFF switch (Figure 22, item 2).
5. Close the cover. The transmitter is ready for use.
When using the transmitter in hazardous places, the use of galvanic
isolators or barriers is essential. The following barrier & isolator are

VAISALA _______________________________________________________________________ 33
User's Guide ______________________________________________________________________

available in Vaisala: barrier No. 210664 (STAHL 9001/51-280-091-


141) and galvanic isolator No. 212483 (STAHL 9160/13-11-11).
Examples of connections are presented on page 36 and more
information on installation in hazardous locations is given below.

Installation in Hazardous Locations

US and Canadian Requirements


USA (FM): Wiring for intrinsically safe operation is shown in
Appendix B.

Canada (CSA): Wiring for intrinsically safe operation is shown in


Appendix C.

European Requirements

CATEGORY 1 (Zone 0)
The HMT360 has to be connected to Exia-certified associated
apparatus with galvanic isolation, gas group IIB or IIC.

NOTE If both analog outputs are in use, the channel 1 (-) and channel 2 (-)
must be short circuited (see figure 4-3).

CATEGORY 2 or 3 (Zone 1 or 2)
The HMT360 has to be connected either to a Zener barrier or galvanic
isolator

NOTE If both analog outputs are in use with a galvanic isolator, the channel
1 (-) and channel 2 (-) must be short circuited (see Figure 17).

Figure 15 and Figure 16 present examples of galvanic isolators and


Zener barrier connections (only CH1 connected).

34 __________________________________________________________________ M010056EN-E
Chapter 4 ______________________________________________________ Electrical Connections

Maximum Cable Resistance Calculation for the


Barrier (Vaisala Code: 210664)
General specifications of the HMT360:

Supply voltage Uin = 24 V (12 ... 35 V)


Maximum current Iout = 20 mA
(100 %RH)
Minimum operating voltage for HMT360 Umin = 12 V (15 V with
serial port)

VAISALA _______________________________________________________________________ 35
User's Guide ______________________________________________________________________

HMT360 Connected to Galvanic


Isolator
HAZARDOUS AREA SAFE AREA

a) current signal controller


HAZARDOUS AREA SAFE AREA

b) voltage signal controller


0505-278, 0505-279

Figure 15 HMT360 Connected to Galvanic Isolator

36 __________________________________________________________________ M010056EN-E
Chapter 4 ______________________________________________________ Electrical Connections

HMT360 Connected to Zener


Barrier
HAZARDOUS AREA SAFE AREA

a) current signal controller


HAZARDOUS AREA SAFE AREA

b) voltage signal controller


0505-280, 0505-281

Figure 16 HMT360 Connected to Zener Barrier

VAISALA _______________________________________________________________________ 37
User's Guide ______________________________________________________________________

Examples of Connections

Make the connection


between the (-) terminals
always when using a
galvanic isolator !

0505-282

Figure 17 STAHL 9160/13-11-11 (Galvanic Isolator)

0505-283

Figure 18 STAHL 9001/51-280-091-141 (Zener Barrier)

38 __________________________________________________________________ M010056EN-E
Chapter 4 ______________________________________________________ Electrical Connections

Grounding
When grounding the transmitter, follow the local requirements. Use at
least 4 mm2 grounding cable when grounding the transmitter or
barrier. Note that the allowed resistance between barrier and system
ground must be less than 1 ohm. Use ground terminal located inside or
outside of transmitter. See Figure 19 below:

HAZARDOUS AREA SAFE AREA

0505-284

Figure 19 Grounding

VAISALA _______________________________________________________________________ 39
User's Guide ______________________________________________________________________

This page intentionally left blank.

40 __________________________________________________________________ M010056EN-E
Chapter 5 ________________________________________________________________ Operation

CHAPTER 5
OPERATION

Turn the transmitter on ON with the ON/OFF switch (Figure 22, item
2).

HMT360 with Display


Measurement readings appear on the display after switching power
ON. The analog output signals can be read from the system or load
resistor.

HMT360 for Natural Gas


When switching on the HMT360 for natural gas, the existing pressure
setting appears on the display; the default setting is 1000 hPa (=1 bar
or 14.5 psi).

To modify the pressure setting, turn the internal CALIBRATION


ENABLED/DISABLED DIP switch of the transmitter to position ENABLED
(up). Adjust the pressure reading with buttons ▲ and ▼ on the
display cover; the adjustment step is 100 hPa). Acknowledge the value
with button E. To complete the pressure setting, turn the DIP switch
back to position DISABLED (down). If the pressure setting is not
modified, the measurement readings appear automatically on the
display after 60 seconds.

NOTE Avoid static discharge. Always use a wet cloth for wiping the display.

VAISALA _______________________________________________________________________ 41
User's Guide ______________________________________________________________________

HMT360 without Display


If the transmitter does not have a display, the red LED indicator on the
cover indicates various phases of calibration and possible errors. In
normal operation it is not lit. If the LED is lit and all calibration or test
DIP switches are disabled, it is an indication of an internal error.

Otherwise, the LED indicates the following:

- LED blinking = calibration of the dry end (offset).


- LED lit = calibration of the wet end (gain).

NOTE If calibration is concluded but DIP switch settings are not restored,
the LED indicator keeps blinking.

DIP Switch Functions


The table below is also printed on the protection board:

0505-285

Figure 20 DIP Switch Functions

n: ANALOG OUTPUT TEST ON/OFF:

If you turn the switch to ON position (up), you can force the outputs to
states 4 mA, 12 mA and 20 mA by pressing the buttons ▲ and ▼on
the cover.

42 __________________________________________________________________ M010056EN-E
Chapter 5 ________________________________________________________________ Operation

o: CALIBRATION DISABLED/ENABLED:

The EEPROMs are write protected. If this switch is in the DISABLED


position (down), it does not allow any calibrations or scalings.

NOTE Keep this switch always in the disabled position during normal use of
the transmitter.

p and q CALIBRATION RH, T, ANALOG:

With these combinations you can perform relative humidity,


temperature or analog output calibrations with a multimeter or with
the transmitter display unit. Turn the DIP switches to the desired
position according to the table printed on the protective cover.

r OUTPUT QUANTITIES:

Determines whether the output units are metric (down) on non-metric.

s, t and u SELECT OUTPUT QUANTITIES:

With the three DIP switches on the right, you can select the output
quantities according to the table printed on the right side of the
protective cover. Note that only the ordered quantities can be selected.
The SPECIAL option gives you the choice of setting any ordered
quantity to each channel.

NOTE Always restore the DIP switch settings after having tested the analog
outputs or performed the calibration.

VAISALA _______________________________________________________________________ 43
User's Guide ______________________________________________________________________

This page intentionally left blank.

44 __________________________________________________________________ M010056EN-E
Chapter 6 ___________________________________________________________ Operation Errors

CHAPTER 6
OPERATION ERRORS

The following indicates an operation error of the HMT360:

- Analog output(s) current drops below 4 mA.


- Display shows ERROR.
- LED on the cover is lit and no calibration is going on (transmitters
without a display).
In the case of error:

- Check that the sensor is connected properly.


- Check if there is condensed water in the probe, if yes, let the probe
dry.

Cleaning Sensor from Glycol in Natural Gas


Application
If a glycol layer on the Vaisala HUMICAP® sensor is disturbing the
moisture measurement, the sensor can be cleaned on site soaking the
sensor to de-ionized water or to IPA (isopropanol, propan-2-ol).

VAISALA _______________________________________________________________________ 45
User's Guide ______________________________________________________________________

0505-286

Figure 21 Cleaning Sensor

1. Carefully remove the sintered filter.


2. Dip the sensor in de-ionized water or IPA (max 1 minute).
3. Gently blow the sensor dry with dry gas (nitrogen) if available.
4. When the sensor and probe are dry, set the filter back on the
probe.

NOTE Do not touch the sensor surface, especially if removing the Vaisala
HUMICAP® sensor from the probe.

Only dipping in de-ionized water or IPA is allowed, other agents like


ethanol cannot be used.

Do not keep the sensor for more than one minute in the water or IPA.

The water or IPA temperature must be below 30 °C (86 °F).

Do not use a cotton stick or any other mechanical means for cleaning
and/or drying the sensor.

In case of constant error, please contact Vaisala Service (see Chapter


2, section Maintenance on page 15.

46 __________________________________________________________________ M010056EN-E
Chapter 7 __________________________________________________ Display/Keypad Commands

CHAPTER 7
DISPLAY/KEYPAD COMMANDS

With these commands you can scale the outputs and select special
output quantities if the transmitter configuration allows that.

NOTE Chapter 10 Calibration and Adjustment describes separately


display/keypad commands for the calibration.

Setting Pressure for Calculations


In humidity transmitters, ambient pressure is used for the mixing ratio
and wet bulb calculations. In dewpoint transmitters (natural gas
applications), the process pressure in the measurement point is
required to achieve the specified accuracy.

To modify the pressure settings, turn the internal CALIBRATION


ENABLED/DISABLED DIP switch of the transmitter to position ENABLED
(up). Press button C on the display cover: text "SCAL" appears on the
display. Then press button E until the following display appears (the
numeric value is always the existing setting, in this example 1013.25):

0505-287

Adjust the pressure reading with buttons ▲ and ▼. Acknowledge the


value with button E. To complete the pressure setting, turn the dip
switch back to position DISABLED (down).

See the pressure conversion table on page 60.

VAISALA _______________________________________________________________________ 47
User's Guide ______________________________________________________________________

Selecting Output Quantities

HMT360 for Natural Gas


Two quantities are shown on the LCD. The upper half of the display
shows the moisture, whereas the lower half is reserved for the
temperature reading.

Upper Half of Display


To modify the displayed quantity, turn the CALIBRATION
ENABLED/DISABLED DIP switch of the transmitter to position ENABLED
(up). Select the required quantity with button ▲ on the display cover,
and acknowledge the value with button E. To complete the selections,
turn the DIP switch back to position DISABLED (down).

When the OUTPUT QUANTITIES METRIC/NON METRIC DIP switch is in


the METRIC (down) position, the available quantities are dew/frost
point temperature (Td °C), parts per million (ppm), and absolute
humidity in mg/m3 (W). When the DIP switch is in the NON METRIC
(up) position, the available quantities are dew/frost point temperature
(Td °F), parts per million (ppm), and absolute humidity in lb/mmscf
(W).

Lower Half of Display


The OUTPUT QUANTITIES METRIC/NON METRIC DIP switch can be used
to select between °C and °F.

It is possible to check the pressure setting of the transmitter by


pressing button ▼. To return to the temperature reading, press button
▼ again.

48 __________________________________________________________________ M010056EN-E
Chapter 7 __________________________________________________ Display/Keypad Commands

Selecting Analog Outputs


You can select the output quantities for channels 1 and 2 by turning
the CALIBRATION ENABLED/DISABLED DIP switch of the transmitter to
position ENABLED (up) and the three OUTPUT SELECTION DIP switches
to position SPECIAL (all up).

Press button C on the display cover: text "SCAL" appears on the


display. Then press button E until the following display appears:

0505-288

The characters on the second line of the display in this menu


correspond to the quantities according to the following table:

Table 3 Available Quantities


Character Quantity Abbreviation Availability
0 relative humidity RH A, D
1 temperature T A, D, F, H, N
2 dewpoint Td D, N
temperature
3 absolute humidity a D
4 mixing ratio x D
5 wet bulb Tw D
temperature
6 volume ppm H, N
concentration
7 water activity aw H, F
8 relative humidity RS H
of saturation
9 saturation Ts H
temperature
A water content w N
A: available for HMT360A
D: available for HMT360D
F: available for HMT360F
H: available for HMT360H
N: available for HMT360N

Select the quantity for channel 1 with buttons ▲ and ▼ and


acknowledge the selection with button E.

If the transmitter is equipped with two analog channels, select the


quantity for channel 2 in the same way, for example:

VAISALA _______________________________________________________________________ 49
User's Guide ______________________________________________________________________

0505-289

Press button C to exit the display command mode or continue by


setting the pressure.

NOTE Remember to restore the DIP switch settings.

Scaling Analog Outputs


Turn the internal CALIBRATION ENABLED/ DISABLED DIP switch of the
transmitter to upward position (ON). Press button C on the display
cover and a text similar to the following appears:

0505-290

Numbers on the second line indicate the low end scaling of channel 1
currently stored in the transmitter memory. The text SET LO on the
lower left corner indicates that you can now change the low end
scaling with buttons ▲ and ▼. Acknowledge the value with button E
and a text similar to the following appears:

0505-291

Numbers on the second line indicate the high end scaling of channel 1.
The text SET HI on the lower left-hand corner indicates that you can
now change the high end scaling with buttons ▲ and ▼.
Acknowledge with button E.

If there is another channel available, the display shifts to the scaling


menu of channel 2. You can now scale the analog outputs for the
channel 2 in the same way as described above.

50 __________________________________________________________________ M010056EN-E
Chapter 7 __________________________________________________ Display/Keypad Commands

Press button C to exit the display command mode or continue by


selecting the output quantities. This menu starts automatically after the
scaling menu only if the OUTPUT SELECTION DIP switches are on the
position SPECIAL (all up) from the beginning.

NOTE Remember to restore the DIP switch settings.

VAISALA _______________________________________________________________________ 51
User's Guide ______________________________________________________________________

This page intentionally left blank.

52 __________________________________________________________________ M010056EN-E
Chapter 8 ____________________________________________________________ Serial Interface

CHAPTER 8
SERIAL INTERFACE

WARNING The serial interface MUST NOT be used in hazardous areas.

Use the serial interface for calibration and testing purposes in safe
areas only. Always use the serial interface cable (optional accessory,
order code: 25905ZZ). Connect one end of the cable to the serial port
of your computer and the other to the connector marked "RS 232C" on
the electronics unit of the transmitter (see Figure 4, item 9).

NOTE With serial communication the current consumption increases


approximately to 7 mA and the transmitter is not able to operate with
4 mA. Therefore, it is recommended to use serial communication
only temporarily for changing settings or for calibrating the
transmitter in a safe area. With serial port, the minimum supply
voltage is 15 VDC.

NOTE The transmitter incorporates a serial interface detector. However, not


all terminals or PC serial ports (for example, optoisolated or ports not
meeting RS232C standards) recognize this standard. If the
communication is not possible via a serial interface, use a forced
activation by pressing simultaneously buttons ▲ and E on the
transmitter cover. To deactivate the forced activation, press these
buttons again or reset the transmitter. If the transmitter does not
receive any commands for half an hour, it automatically closes the
serial communication.

VAISALA _______________________________________________________________________ 53
User's Guide ______________________________________________________________________

Serial Communication Settings


Table 4 Serial Communications Settings
Parameter Value
bauds 2400
parity none
data bits 8
stop bits 1

To start giving commands, make sure that the HMT360 is connected


to a serial port of your computer and that the terminal session is open.
Issue commands by typing them on your computer according to the
following instructions. In these commands, <cr> stands for pressing
Enter (on your computer keyboard).

ASEL Selecting Analog Outputs


Syntax: ASEL xxx yyy<cr>

where
xxx = Quantity of channel 1
yyy = Quantity of channel 2

Turn the internal DIP switch CALIBRATION ENABLED/DISABLED to ON-


position before the selecting the analog outputs and return it to OFF-
position after selecting.

54 __________________________________________________________________ M010056EN-E
Chapter 8 ____________________________________________________________ Serial Interface

Table 5 Output Quantities and Their Abbreviations


Output Quantity Abbreviation
Relative humidity RH
Temperature T
Dewpoint temperature Td
Absolute humidity a
Mixing ratio x
Wet bulb temperature 1) Tw
Mass proportion 1) ppm
Water activity 1) aw
Relative humidity of saturation 1) RS
Saturation temperature1 Ts
Water content 2) w

1) Available only for HMT368H for moisture measurements in liquid.


2) Available only for HMT360N for natural gas.

Example:
>asel rh t<cr>
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo : -40.00 'C
Ch2 T hi : 60.00 'C
>

S Scaling Analog Outputs


Syntax: Szz aa.a bb.b<cr>

where
zz = Quantity (RH, T, Td, x, a, Tw, ppm, aw, RS, Ts)
aa.a = Lower limit of the quantity
bb.b = Upper limit of the quantity

Turn the internal DIP switch CALIBRATION ENABLED/DISABLED to ON-


position before the scaling the analog outputs and return it to OFF-
position after scaling.

Example:
>SRH 0 100<cr>
RH lo 0.00 %RH
RH hi 100.00 %RH
>
Exception:

When you have natural gas transmitter and you scale the quantity of
water content w (mg/m³ or lb/mmscf), issue command SNG1.

VAISALA _______________________________________________________________________ 55
User's Guide ______________________________________________________________________

Adjustment Commands
Turn the internal DIP switch CALIBRATION ENABLED/DISABLED to ON-
position before the calibration and return it to OFF-position after the
calibration.

Note that HMT360 for natural gas can be checked but not adjusted by
the user.

CRH Relative Humidity Adjustment


NOTE Read also the calibrator, for example HMK15 manual and refer to
page 56 of this manual for more detailed instructions on salt bath
calibration.

Syntax: CRH<cr>

The transmitter asks and measures relative humidity and calculates the
calibration coefficients.

Example:
>CRH<cr>
RH : 1.82 1. ref ? 0 <cr>
Press any key when ready... <cr>
RH : 74.22 2. ref ? 75 <cr>
OK
>
The OK indicates that the calibration has succeeded.

CT Temperature Adjustment
Syntax: CT<cr>

The transmitter asks and measures temperature readings and calculates


the calibration coefficients.

Example:
>CT<cr>
T : 0.81 1. ref ? 0.5 <cr>
Press any key when ready... <cr>
T : 99.12 2. ref ? 99.5 <cr>
OK

56 __________________________________________________________________ M010056EN-E
Chapter 8 ____________________________________________________________ Serial Interface

>
The OK indicates that the calibration has succeeded. When
performing one-point calibration press only Enter for 2. reference.

ACAL Analog Output Calibration


Syntax: ACAL<cr>

Connect the HMT360 to a multimeter. Issue the ACAL command:

1. Disconnect the serial cable from the transmitter while reading


the multimeter value for the CH1 (I1).
2. Reconnect the serial cable. Type the multimeter reading and
press Enter.
3. Type the higher current multimeter reading and press Enter.

Example:
>ACAL<cr>
Ch1 I1 (mA) ? 4.846
Ch1 I2 (mA) ? 19.987
>

Output Commands

ITEST Testing Analog Outputs


Syntax: ITEST aa.aaa bb.bbb<cr>

where
aa.aaa = Current value to be set for channel 1 (mA)
bb.bbb = Current value to be set for channel 2 (mA)

This command outputs the current value of each channel and the
corresponding control signal of the digital-to-analog converter.

Example:
>ITEST 8 12 <cr>
8.000 3F8 12.000 70O
>

VAISALA _______________________________________________________________________ 57
User's Guide ______________________________________________________________________

>ITEST<cr>
6.349 25A 19.001 E93
>

The set current values remain valid until you issue the command
ITEST without readings. With this command the desired outputs of
the transmitter are shown.

NOTE When outputting low currents from channel 1, remember to remove


RS232C-cable while reading the current output, because of the
increased current consumption for using RS port.

SEND Outputting Measurement


Values
Syntax: SEND<cr>

This command outputs the measured values in one point.

R Activating Continuous Output


Syntax: R<cr>

With the command R the transmitter outputs measured values


continuously.

S Stopping Continuous Output


Syntax: S<cr>

The continuous outputting is stopped with the command S. If


outputting is active, this command is not echoed.

INTV Setting Output Interval


Syntax: INTV n xxx<cr>

where
n = 1 ... 255
xxx = S, MIN, or H

58 __________________________________________________________________ M010056EN-E
Chapter 8 ____________________________________________________________ Serial Interface

Sets the output interval when the transmitter outputs measurement


readings. The time interval is used when the continuous output is
active.

Example, the output interval is set to 10 minutes:


>INTV 10 min<cr>
Output intrv. : 10 min
>

PRES Setting Ambient Pressure for


Calculations
Syntax: PRES aaaa.a<cr>

where
aaaa.a = Pressure (hPa)

Turn the internal DIP switch CALIBRATION ENABLED/DISABLED to ON-


position before setting the pressure and return it to OFF-position after
making the setting.

Example:
>PRES 2400<cr>
Pressure : 2400
>

VAISALA _______________________________________________________________________ 59
User's Guide ______________________________________________________________________

Table 6 Pressure Conversion Chart


FROM mmHg/Torr inHg atm bar psi
hPa/mb 1 1.333224 33.86388 1013.25 1000 68.94757
ar
mmHg/ 0.7500617 1 25.40000 760 750.0617 51.71493
TO

Torr
inHg 0.02952999 0.03937008 1 29.921 29.52999 2.036021
atm 0.00098692 0.00131597 0.033422 1 0.98692 0.068046
bar 0.001 0.001333224 0.03386388 1.01325 1 0.06894757
psi 0.01450377 0.01933678 0.4911541 14.6962 14.50377 1

Example:

29.9213 inHg = 29.9213 x 33.86388 = 1013.25 hPa / mbar

NOTE Conversions from mmHg and inHg are defined at 0 °C.

60 __________________________________________________________________ M010056EN-E
Chapter 9 __________________________________________________ Measuring at Overpressure

CHAPTER 9
MEASURING AT OVERPRESSURE

The probes HMP362, HMP364 and HMP368 are designed for


humidity measurement at overpressure. The maximum measurement
pressures depend on the probe as follows:

HMP362: 0 ... 167 bar (16,7 MPa), for natural gas, sample cell
HMP302SC available

HMP364: 0 ... 100 bar (10 MPa), for pressurized rooms and
processes, probe is provided with a nut, fitting screw and
sealing washer

HMP368: 0 ... 40 bar (4 MPa), for pressurized pipelines, ball valve


set available

The accuracy of the dewpoint measurement is affected by pressure in


the measurement chamber. The actual pressure in the sampling cell
shall be set to the transmitter by using the serial line command PRES
(PRES Setting Ambient Pressure for Calculations, page 59 or by using
the keypad commands, page 47).

Pressure Regulator Recommended


When sampling pressurized processes exceeding the maximum
measurement pressure of the probe, the pressure in the measurement
chamber must be regulated to acceptable level or below. It is
recommended to use pressure regulator before the measurement
chamber to prevent remarkable pressure variations.

VAISALA _______________________________________________________________________ 61
User's Guide ______________________________________________________________________

This page intentionally left blank.

62 __________________________________________________________________ M010056EN-E
Chapter 10 __________________________________________________Calibration and Adjustment

CHAPTER 10
CALIBRATION AND ADJUSTMENT

Calibration Interval
HMT360 is calibrated as shipped from the factory. Typical humidity
calibration interval is one year. Depending on the application it may
be good to make the first calibration check earlier.

Factory Calibration and Adjustment


The device (or only the electronics unit) can be sent to Vaisala Service
Centers for calibration and adjustment, see contact information on
page 15.

NOTE HMT360 for natural gas must be adjusted always in Vaisala Service
Center.

User Calibration and Adjustment


Note that HMT360 for natural gas can be checked but not adjusted by
a user.

Calibration is carried out by using the keypad or by using serial


commands. The following additional equipment is needed in
calibration of the HMT360:

VAISALA _______________________________________________________________________ 63
User's Guide ______________________________________________________________________

- power supply for the channel(s) to be calibrated (12 ... 24 VDC),


- a multimeter for the HMT360 without a display, and
- the calibrated references.
First, the electronics unit is removed from the transmitter base to
perform a calibration in safe area. Second, the power supply and a
multimeter is connected to the electronics unit according to the
following instructions.

Removing Electronics Unit for


Calibration

0505-292

Figure 22 Detaching Electronics Unit with Probe for


Calibration

1. Unfasten the screws and open the cover.


2. Switch the transmitter off with the ON/OFF switch.
3. Disconnect the flat cable by lifting it carefully, for example with
a screw-driver.
4. Turn the electronics unit slightly upwards to release it from the
hinges. Leave the transmitter base with the cable connections on
place. When putting the electronics unit back to place, attach the
upper hinge first.

64 __________________________________________________________________ M010056EN-E
Chapter 10 __________________________________________________Calibration and Adjustment

Connections

B1 C1

B2 C2
A1
A2

0505-293

Figure 23 Connecting Power Supply and Multimeter for


Calibration

1. Connect with banana plugs a power supply (12 ... 24 VDC, with
a serial port the minimum supply voltage is 15 VDC ) to the
terminals B1 (-) and C1 (+) (channel 1), see Figure 23.
2. Connect the multimeter in series with the supply, B1 (-) and
A1 (+).
3. Follow the same procedure with channel 2 using the terminals
B2 (-) and C2 (+). When calibrating both channels at the same
time, use two galvanically separated power supplies.
You can now calibrate or check the humidity and temperature or the
analog outputs according to the instructions given in calibration
chapters.

If the current measurement is needed in a hazardous area, the


multimeter is connected to the terminals A1/A2 (+) and B1/B2 (-).
Use only an approved multimeter.

VAISALA _______________________________________________________________________ 65
User's Guide ______________________________________________________________________

WARNING The power supply connectors (C1 and C2) MUST NOT be used in
hazardous areas. For calibration in a hazardous area use only an
approved multimeter, which fulfils the safety factors printed on the
protective cover.

Calculating Correspondence of Current Values


and Output Quantities
When using HMT360 without a display, calibration is carried out by
using a multimeter. Use the following equations to calculate the
current values corresponding the reference output quantities.

(Qref - Qmin )
I = 4 mA + 16 mA ⋅ (1)
(Qmax - Qmin )

where
Qref = Reference value of the calibrated quantity
Qmin = Value corresponding to 4 mA
Qmax = Value corresponding to 20 mA

Example 1:

Relative humidity scaling 0 - 100 %RH, reference 11.3 %RH:

(11.3 %RH - 0 %RH)


I = 4 mA + 16 mA ⋅ = 5.808 mA (2)
(100 %RH - 0 %RH)

Example 2:

Temperature scaling -40 - +120 °C, reference 22.3 °C:

(22.3 º C - 40 º C)
I = 4 mA + 16 mA ⋅ = 10.230 mA (3)
(120 º C - 40 º C)

66 __________________________________________________________________ M010056EN-E
Chapter 10 __________________________________________________Calibration and Adjustment

Relative Humidity Calibration

Automatic Two-Point Calibration


(Only with HMT360 with Display)
The automatic calibration procedure is a user friendly way to calibrate
with salt solutions; user do not have to feed the reference values to the
HMT360. The transmitter displays the accurate value based on the
measured temperature and the Greenspan table stored into the
transmitter memory. The display chart of the calibration procedure is
presented in Figure 24 Display Chart of the Automatic Calibration
Procedure.

- Set the calibration DIP switch to ENABLED position and select RH


calibration by using switches 3 and 4 (see Chapter 5 DIP Switch
Functions on page 42).
- Remove the filter and insert the probe head into a measurement
hole of the LiCl salt chamber in the humidity calibrator.
Acknowledge by pressing the button E. Alternatively select the 0.1
(humidity in nitrogen) with buttons ▲ and ▼. Acknowledge with
the button E.
- The transmitter remains to wait for the process stabilization (10-15
min) and then stores the correction.
- Insert the probe head into a measurement hole of the NaCl salt
chamber in the humidity calibrator. Acknowledge by pressing the
button E. Alternatively you can select K2SO4 value with buttons
▲ and ▼.
- The transmitter remains to wait for the stabilization and then stores
the correction. The text CAL PASS is displayed after the calibration
is performed.

NOTE Remember to restore the DIP switch settings.

VAISALA _______________________________________________________________________ 67
User's Guide ______________________________________________________________________

RH RH

Set Lo Set Lo

RH RH

Set Lo Set Lo

RH RH
RH

Set Lo Set Lo
Set Lo

RH

Set Lo

0505-294

Figure 24 Display Chart of the Automatic Calibration


Procedure

68 __________________________________________________________________ M010056EN-E
Chapter 10 __________________________________________________Calibration and Adjustment

Manual Calibration

Low End Adjustment


- Remove the filter and insert the probe head into a measurement
hole of the dry end reference chamber (for example LiCl: 11 %RH)
in the humidity calibrator.

NOTE If you use serial commands, please refer to page 56.

- Set the calibration DIP switch 2 to ENABLED (up) position and


select RH calibration by using switches 3 and 4 (see Chapter 5 DIP
Switch Functions on page 42).
- Press button C to ignore the automatic calibration procedure. The
humidity reading measured by the transmitter appears on the
display and on the lower left corner you can see the text SET LO. If
the transmitter has no display, the LED indicator is blinking.
- According to the calibrator used, wait at least 10-15 minutes for the
sensor to stabilize; use a multimeter or the transmitter display to
monitor the stabilization.
- Press the buttons ▲ and ▼ to adjust the display reading or
multimeter reading to correspond the reference value (when using a
multimeter calculate the current value corresponding to the
reference humidity by using the equations presented in page 66).
Press the button E to conclude the low end adjustment.
- If the low end adjustment is sufficient, press the button E again to
conclude the calibration and restore the DIP switch settings; set the
calibration DIP switch to position DISABLED (down). To continue
with the high end adjustment follow the instructions in next
chapter.

VAISALA _______________________________________________________________________ 69
User's Guide ______________________________________________________________________

High End Adjustment (Two-Point Calibration)


- After having made the low end adjustment, insert the probe head
into a measurement hole of the NaCl (75 %RH) salt chamber in the
humidity calibrator. Please, note that the difference between the
two humidity references must be at least 30 %RH.

NOTE If you use serial commands, please refer to page 56.

- The readings measured by the transmitter appear on the display, as


well as the text, SET HI on the lower left hand corner. If the
transmitter has no display, the LED indicator lights up.
- Wait at least 10 - 15 minutes for the sensor to stabilize; use a
multimeter or the transmitter display to monitor the stabilization.
- Press the buttons ▲ and ▼ to adjust the display or multimeter
reading (when using a multimeter calculate the current value
corresponding to the reference humidity by using the equations
presented in page 66) and conclude the calibration with the button
E.
- Restore the DIP switch settings. Set the calibration DIP switch to
position DISABLED (down).

70 __________________________________________________________________ M010056EN-E
Chapter 10 __________________________________________________Calibration and Adjustment

Temperature Calibration in One


Point
Always use a high quality standard for calibrating the temperature.

NOTE If you use serial commands, please refer to page 56.

- Set the calibration DIP switch to position ENABLED (up).


- Set the calibration DIP switches to position TEMPERATURE
CALIBRATION according to the table printed on the protective cover.
The temperature value measured by the transmitter appears on the
display as well as the text SET LO on the lower left corner. If the
transmitter has no display, the LED indicator starts to blink.
- Let the sensor stabilize; use a multimeter or the transmitter display
to monitor the stabilization.
- Press the buttons ▲ and ▼ to adjust the display or multimeter
reading to correspond the reference value (when using a multimeter
calculate the current value corresponding to the reference humidity
by using the equations presented in page 66). Conclude the one
point calibration by pressing the button E twice.

VAISALA _______________________________________________________________________ 71
User's Guide ______________________________________________________________________

This page intentionally left blank.

72 __________________________________________________________________ M010056EN-E
Chapter 11 ____________________________________________________________ Specifications

CHAPTER 11
SPECIFICATIONS

Performance

Relative Humidity
Table 7 Relative Humidity Specifications
Property Description / Value
Measurement range 0 ... 100 %RH
Accuracy (including non-
linearity and repeatability);
when calibrated against:
high quality, certified humidity ± 1 %RH (0 ... 90 %RH)
standards ± 2 %RH (90 ... 100 %RH)

salt solutions ± 2 %RH (0 ... 90 %RH)


(ASTM E104-85) ± 3 %RH (90 ... 100 %RH)
Response time (90 %) at +20°C 15 seconds
in still air (with sintered filter)
Sensors:
HUMICAP®180 for typical applications
HUMICAP®L for applications with a demanding
chemical environment (max. +40 °C in
high humidities)

VAISALA _______________________________________________________________________ 73
User's Guide ______________________________________________________________________

Moisture in Natural Gas


(HMT362N/364N/368N)

Dewpoint Measurement
Table 8 Dewpoint Measurement Specifications
Property Description / Value
Measurement range -50 ... +50 °C (-58 ... +122 °F) Td
Accuracy ± 2 °C (± 3.6 °F) (see graph in Figure 25
below)
Response time 63 % [90 %]
-20 → +10 ° C (-4 → +50 ° F) 6 s [32 s]
+10 → -20 ° C (+50 → -4 ° F) 120 s [370 s]
Sensor Vaisala HUMICAP®180M

0505-295

Figure 25 Effect of Process Temperature on Dewpoint


Measurement Accuracy

Calculated Variables Available


(Typical Ranges)
Table 9 Calculated Variables Specifications
Property Description / Value
Water content 0 ... 20 lb/mmscf, 0 ... 326 mg/m³
Parts per million by volume 0 ... 500 ppmv

74 __________________________________________________________________ M010056EN-E
Chapter 11 ____________________________________________________________ Specifications

Water Activity in Jet Fuel


Applications
Table 10 Water Activity in Jet Fuel Applications
Specifications
Property Description / Value
Measurement range 0 ... 1 aw at -40 ... +180 °C
(-40 ... +356 °F)
Accuracy
when calibrated against:
high quality, certified humidity ± 0.01 (0 ... 0.9)
standards ± 0.02 (0.9 ... 1.0)

salt solutions ± 0.02 (0 ... 0.9)


(ASTM E104-85) ± 0.03 (0.9 ... 1.0
Response time (90 %) at +20 ° 10 min
C in still oil (stainless steel filter)
Humidity sensor Vaisala HUMICAP®
Calculated variables available water content
relative humidity for saturation
temperature
Accuracy of the water content better than ± 15 % of the reading

Temperature
Table 11 Temperature Specifications
Property Description / Value
Measurement range -40 ... +180 °C
Typical accuracy of electronics ± 0.1 °C
at +20 °C
Typical temperature 0.005 °C/°C
dependence of electronics
Sensor Pt 1000 RTD ⅓ Class B IEC 751

VAISALA _______________________________________________________________________ 75
User's Guide ______________________________________________________________________

Calculated Variables (Typical


Ranges)

With HMP361 Probe


Table 12 HMP361 Calculated Variables Specifications
Property Description / Value
Dewpoint temperature -40 ... +60 °C
Mixing ratio 0 ... 160 g/kg d.a.
Absolute humidity 0 ... 160 g/m³
Wet bulb temperature 0 ... +60 °C

With HMP363, HMP364, HMP365 & HMP368 Probes


Table 13 HMP363, HMP364, HMP365 & HMP368 Probes
Calculated Variables Specifications
Property Description / Value
Dewpoint temperature -40 ... +100 °C
Mixing ratio 0 ... 500 g/kg d.a.
Absolute humidity 0 ... 600 g/m³
Wet bulb temperature 0 ... +100 °C

Outputs
Table 14 Output Specifications
Property Description / Value
Two analog outputs
(one standard, one optional) two wire 4 ... 20 mA
Typical accuracy of analog
outputs at +20 °C ± 0.05 %/°C full scale
Typical temperature
dependence of analog outputs ± 0.005 %/°C full scale
RS232C serial output for connector type RJ45
service use only in safe area
Output connections via safety barriers

76 __________________________________________________________________ M010056EN-E
Chapter 11 ____________________________________________________________ Specifications

Classification with Current Outputs


EU (94/9/EC, ATEX100a) II 1 G EEx ia IIC T4
PTB 00 ATEX 2112 X
Safety factors: Ui=28 V, Ii=100 mA, Pi =0.7 W, Ci=1 nF,
Li=0 H
Environmental specifications: Tamb = -20 ... +60 °C
Pamb = 0.8 ... 1.1 bar
Dust classification (with II 1 D (IP65 T= 70 °C)
protection cover): VTT 04 ATEX 023X
U.S.A. (FM) Classes I, II, III
Division 1: Groups A-G
Division 2: Groups A-D, F and G
FM Project ID:3010615
Safety factors: Vmax = 28VDC, Imax = 100mA, Ci = 1nF,
Li = 0, Pi = 0.7 W, Tamb = 60 °C, T5
Canada (CSA) Class 1, Divisions 1 and 2,
Groups A, B, C, D
Class II, Divisions 1 and 2, Groups G and
Coal Dust
Class III
CSA File No: 213862 0 000;
CSA report 1300863
Tamb= 60 °C, T4
Japan (TIIS) Ex ia IIC T4
Code number: C15354
Safety factors: Ui = 28 VDC, Ii = 100 mA, Ci = 1nF,
Li = 0, Pi = 0.7 W Tamb = 60 °C
Australia (TestSafe) Certificate No: Ex AUS Ex 3738X
Marking code: Ex ia IIC T5 IP65
AUS Ex 3738X
Safety factors: Vmax = 28 VDC, Ii = 100 mA, Pi = 0.7 W,
Ci = 1nF, Li = 0 mH

VAISALA _______________________________________________________________________ 77
User's Guide ______________________________________________________________________

General
Table 15 General Specifications
Property Description / Value
Operating voltage 12 ... 28 V
with serial port (service 15 ... 28 V
mode)
Connections screw terminals, 0.33 ... 2.0 mm² wires
(AWG 14-22)
Cable bushing Pg11 (5 ... 12 mm)
Conduit fitting Pg11/NPT 1/2"-14
Operating temperature range
for electronics -40 ... +60 °C
with display and/or with
storage temperature range -40 ... +70 °C
Housing material G-AlSi 10 Mg (DIN 1725)
Housing classification IP 65 (NEMA 4)
Housing dimensions 164 × 115 × 62 mm
Housing weight 950 g
Fully electromagnetically EN 61326-1:1997 +Am :1998, Electrical
compatible according to equipment for measurement, control and
standards laboratory use - EMC requirements;
Industrial environment

NOTE IEC 1000-4-5 complies only when using external EXi approved surge
arrester on safe area.

78 __________________________________________________________________ M010056EN-E
Chapter 11 ____________________________________________________________ Specifications

Probes
Table 16 Probe Specifications
Property Description / Value
Probe cable length 2, 5 or 10 m
Probe cable diameter 5.5 mm
Sensor protection options stainless steel sintered filter
stainless steel grid
PPS grid with steel netting
filter
PPS grid
HMP361 Probe - wall mount
Temperature range -40 ... +60 °C
HMP362 Probe - small pressure-tight
Temperature range -40 ... +110 °C
Pressure range 0 ... 16.7 MPa (0 ... 167 bar)
HMP363 Probe - confined spaces
Temperature range -40 ... +120 °C
HMP364 Probe - high pressures
Temperature range -40 ... +180 °C
Pressure range 0 ... 10 MPa (0 ... 100 bar)
HMP365 Probe - high temperatures
Temperature range -40 ... +180 °C
HMP368 Probe - pressurized pipelines
Temperature range -40 ... +180 °C
Pressure range 0 ... 4 MPa (0 ... 40 bar)

VAISALA _______________________________________________________________________ 79
User's Guide ______________________________________________________________________

Options and Accessories


Table 17 Options and Accessories Table
Description Value / Order Code
Display two-line LCD
character size (1st line/2nd 12 mm/10 mm
line)
Calculated output variables dewpoint temperature
mixing ratio
absolute humidity
wet bulb temperature
Additional analog output 4 ... 20 mA
Duct mounting installation kit order code: HMP233FAH
(for HMP363 sensor head)
Installation flange (for HMP365
sensor head)
aluminium HMP235FA
stainless HMP235FS
Ball valve set (for HMP368 DMP248BVS
sensor head)
pressure range at 20 °C 0 ... 40 bar
Sampling cell HMP302SC
Serial interface cable for PC 25905ZZ
connectors RJ45 - D9 female
Rain shield HMT360SAR
Protection cover (for use in the 214101
presence of combustible dust)
HMK15 adapter fitting for 12 211011
mm probes
Galvanic isolator 212483
Barrier 210664

80 __________________________________________________________________ M010056EN-E
Chapter 11 ____________________________________________________________ Specifications

Accuracy of Calculated Variables (Not Valid


with HMT360 for Natural Gas)
Accuracy of the calculated variables depend on the calibration
accuracy of the humidity and temperature sensors; here the accuracy
are given for ±2 %RH and ±0.2 °C.

Accuracy of Dewpoint Temperature °C


Table 18 Accuracy of Dewpoint Temperature °C
Relative humidity
Temp. 10 20 30 40 50 60 70 80 90 100
-40 1.86 1.03 0.76 0.63 0.55 0.5 0.46 0.43 — —
-20 2.18 1.19 0.88 0.72 0.62 0.56 0.51 0.48 — —
0 2.51 1.37 1 0.81 0.7 0.63 0.57 0.53 0.5 0.48
20 2.87 1.56 1.13 0.92 0.79 0.7 0.64 0.59 0.55 0.53
40 3.24 1.76 1.27 1.03 0.88 0.78 0.71 0.65 0.61 0.58
60 3.6 1.96 1.42 1.14 0.97 0.86 0.78 0.72 0.67 0.64
80 4.01 2.18 1.58 1.27 1.08 0.95 0.86 0.79 0.74 0.7
100 4.42 2.41 1.74 1.4 1.19 1.05 0.95 0.87 0.81 0.76
120 4.86 2.66 1.92 1.54 1.31 1.16 1.04 0.96 0.89 0.84
140 5.31 2.91 2.1 1.69 1.44 1.26 1.14 1.05 0.97 0.91
160 5.8 3.18 2.3 1.85 1.57 1.38 1.24 1.14 1.06 0.99

Accuracy of Mixing Ratio g/kg


(Ambient pressure 1013 mbar)

Table 19 Accuracy of Mixing Rratio g/kg


Relative humidity
Temp. 10 20 30 40 50 60 70 80 90 100
-40 0.003 0.003 0.003 0.003 0.003 0.004 0.004 0.004 — —
-20 0.017 0.018 0.019 0.021 0.022 0.023 0.025 0.026 — —
0 0.08 0.09 0.09 0.1 0.1 0.11 0.11 0.12 0.13 0.13
20 0.31 0.33 0.35 0.37 0.39 0.41 0.43 0.45 0.47 0.49
40 0.97 1.03 1.1 1.17 1.24 1.31 1.38 1.46 1.54 1.62
60 2.68 2.91 3.16 3.43 3.72 4.04 4.38 4.75 5.15 5.58
80 6.73 7.73 8.92 10.34 12.05 14.14 16.71 19.92 24.01 29.29
100 16.26 21.34 28.89 40.75 60.86 98.85 183.66 438.56 — —
120 40.83 74.66 172.36 — — — — — — —

VAISALA _______________________________________________________________________ 81
User's Guide ______________________________________________________________________

Accuracy of Wet Bulb Temperature °C


Table 20 Accuracy of Wet Bulb Temperature °C
Relative humidity
Temp. 10 20 30 40 50 60 70 80 90 100
-40 0.2 0.2 0.2 0.2 0.2 0.2 0.2 0.2 — —
-20 0.21 0.21 0.22 0.22 0.22 0.22 0.23 0.23 — —
0 0.27 0.28 0.28 0.29 0.29 0.29 0.3 0.3 0.31 0.31
20 0.45 0.45 0.45 0.44 0.44 0.44 0.43 0.43 0.42 0.42
40 0.84 0.77 0.72 0.67 0.64 0.61 0.58 0.56 0.54 0.52
60 1.45 1.2 1.03 0.91 0.83 0.76 0.71 0.67 0.63 0.6
80 2.23 1.64 1.32 1.13 0.99 0.89 0.82 0.76 0.72 0.68
100 3.06 2.04 1.58 1.31 1.14 1.01 0.92 0.85 0.8 0.75
120 3.85 2.4 1.81 1.48 1.28 1.13 1.03 0.95 0.88 0.83
140 4.57 2.73 2.03 1.65 1.41 1.25 1.13 1.04 0.97 0.91
160 5.25 3.06 2.25 1.82 1.55 1.37 1.24 1.13 1.05 0.99

Accuracy of Absolute Humidity g/m³


Table 21 Accuracy of Absolute Humidity g/m³
Relative humidity
Temp. 10 20 30 40 50 60 70 80 90 100
-40 0.004 0.004 0.005 0.005 0.005 0.006 0.006 0.006 — —
-20 0.023 0.025 0.027 0.029 0.031 0.032 0.034 0.036 — —
0 0.1 0.11 0.12 0.13 0.13 0.14 0.15 0.15 0.16 0.17
20 0.37 0.39 0.41 0.43 0.45 0.47 0.49 0.51 0.53 0.55
40 1.08 1.13 1.18 1.24 1.29 1.34 1.39 1.44 1.49 1.54
60 2.73 2.84 2.95 3.07 3.18 3.29 3.4 3.52 3.63 3.74
80 6.08 6.3 6.51 6.73 6.95 7.17 7.39 7.61 7.83 8.05
100 12.2 12.6 13 13.4 13.8 14.2 14.6 15 15.3 15.7
120 22.6 23.3 23.9 24.6 25.2 25.8 26.5 27.1 27.8 28.4
140 39.1 40 41 42 43 44 45 45.9 46.9 47.9
160 63.5 64.9 66.4 67.8 69.2 70.7 72.1 73.5 74.9 76.4

82 __________________________________________________________________ M010056EN-E
Appendix A ______________________________________________________________Dimensions

APPENDIX A
DIMENSIONS

HMP361

VAISALA _______________________________________________________________________ 83
User's Guide ______________________________________________________________________

0505-296

HMP362
92.5 (3.64)
Flange intersection

ø12 (0.47)
41(1.61)
45°
3)
.6

M6
(0
16

9(0.35)
ø36 (1.42)
71.5 (2.81) 7 (0.28)

81 (3.19)
Sample cell HMP302SC
58.7 (2.31)

45° 1/8"-27 NPT


45°

ø24
.5
ø12.5 (0.49)
1/8"-27 NPT

M
6

ø32 (1.26) 71 (2.80)

0505-297

84 __________________________________________________________________ M010056EN-E
Appendix A ______________________________________________________________Dimensions

HMP363

0505-298

HMP364

0505-299

HMP365

VAISALA _______________________________________________________________________ 85
User's Guide ______________________________________________________________________

0505-300

HMP368

0505-301

Mounting Plate

86 __________________________________________________________________ M010056EN-E
Appendix A ______________________________________________________________Dimensions

0505-302

Rain Shield

VAISALA _______________________________________________________________________ 87
User's Guide ______________________________________________________________________

0505-303

Protection Cover
160
8.2 96
Ø

82
224

184
208

189

0505-304

88 __________________________________________________________________ M010056EN-E
Appendix B __________________________ WIRING FOR INTRINSICALLY SAFE OPERATION, FM

APPENDIX B
WIRING FOR INTRINSICALLY SAFE
OPERATION, FM

HAZARDOUS AREA SAFE AREA


Use FM approved associated apparatus;
HMT360 connection board zener barriers or galvanic separators
with entity concept parameters:
-Ch1+ -Ch2+
Voc < 28 V
Isc < 100mA
Ca > Ci + Ccable
La > Li + Lcable

Vmax=28VDC

Load Ch1

Vmax=28VDC

Load Ch2

HMT360-series transmitters are approved


for use in Classes I, II and III,
Division 1, Groups A - G and
Division 2, Groups A - D, F and G.

Safety factors for HMT360-series transmitters


are: Vmax=28V, Imax=100mA, Ci=1n F, Li=0, Pi=0.7W

NOTE:
1. Barrier installation must be completed
in accordance with ANSI/IS A RP 12.6
and the National Electrical Code.
2. Intrinsically safe barrier ground must
be less than 1 ohm.
3. Maximum safe area voltage is 250 V.

Material Weight Finish General tolerance

Part list

Creator Status
01-09-04 ARH
Checked Title
Vaisala Oyj
Accepted CONTROL DRAWNING Vanha Nurmijärventie 2 1
Handler Vantaa
Rev
Replaces Scale Finland Code

Replaced
by
1:1 DRW211603 A

VAISALA _______________________________________________________________________ 89
User's Guide ______________________________________________________________________

This page intentionally left blank.


0505-305

90 __________________________________________________________________ M010056EN-E
Appendix C _________________________ WIRING FOR INTRINSICALLY SAFE OPERATION, CSA

APPENDIX C
WIRING FOR INTRINSICALLY SAFE
OPERATION, CSA

VAISALA _______________________________________________________________________ 91
User's Guide ______________________________________________________________________

This page intentionally left blank.

92 __________________________________________________________________ M010056EN-E
Appendix D ______________________________________________________________ Certificates

APPENDIX D
CERTIFICATES

VAISALA _______________________________________________________________________ 93
User's Guide ______________________________________________________________________

94 __________________________________________________________________ M010056EN-E
Appendix D ______________________________________________________________ Certificates

VAISALA _______________________________________________________________________ 95

You might also like