Specifications:
Measuring range:
0.01 ppb to 20ppm O2 (at 25°C) or
0 – 200% saturation
Optional:
Big LC display for the reading of measuring
value, sample temperature, sample flow, oper- Instrument certificate
ating status and battery charge condition.
Easy user menus in English, German, French
and Spanish. Simple programming of all pa-
rameters by keypad.
Electronic record of major process events and
calibration data.
Data logger for 1’500 data records stored at a
selectable interval.
One current output (0/4 - 20 mA) for measured
signal.
Accuracy / Repeatability:
Accuracy ± 1.5 % of reading or ± 0.2 ppb
Repeatability:± 1 % of read. or ± 0.15 ppb
Response time
t90 < 30 sec. (rising concentration)
Transmitter Specifications and Storage of event log, alarm log and cali- Control functions
Functionality bration history. Relays or current outputs programmable
Storage of the last 1’500 data records in for 1 or 2 pulse dosing pumps, solenoid
Electronics case: Cast aluminum logger with selectable time interval. valves or for one motor valve.
Protection degree: IP 66 / NEMA 4X Programmable P, PI, PID or PD control
Display: LCD, 75 x 45 mm Safety features parameters.
Electrical connectors: screw clamps No data loss after power failure, all data
Dimensions: 180 x 140 x 70 mm is saved in non-volatile memory. 1 Communication interface
Weight: 1.5 kg Overvoltage protection of in- and outputs. USB Stick for logger data.
Ambient temperature: -10 to +50°C Galvanic separation of measuring inputs
Humidity: 10 - 90% rel., non condensing and signal outputs. Monitor Data
Power supply - Battery Transmitter temperature monitoring Sample conditions
Use original power adapter only. with programmable high/low alarm limits.Flow rate: 8 to 25 l/h
Voltage: 85 - 265 VAC, 50/60 Hz Temperature: up to 45 °C
Power consumption: max. 20 VA 1 Alarm relay Inlet pressure (25 °C): 0.2 to 1 bar
Charging time: ~ 6h One potential free contact for summary Outlet pressure: pressure free
Battery type: Li-Ion alarm indication for programmable alarm pH: not lower than pH 4
During charging protect from heat impact values and instrument errors. Suspended solids: less than 10 ppm
and keep splash-proof (not IP66). Maximum load: 1A / 250 VAC
Flow cell and connections
Operating time 1 Input Flow cell made of acrylic glass with built-
Stand-alone (Battery): > 24h One input for potential-free contact. in flow adjustment valve and digital sam-
Connected adapter: continuous Programmable hold or remote off func- ple flow meter.
Controlled shut-down when battery is tion.
empty, remaining time is displayed. Inlet: ¼” Swagelok tube adapter
2 Relay outputs Outlet: flexible tube 8 x 6 mm
Operation Two potential-free contacts programma-
Easy operation based on separate men- ble as limit switches for measuring val- Panel
us for “Messages”, “Diagnostics”, ues, controllers or timer for system clean- Dimensions: 275 x 320 x 240 mm
“Maintenance”, “Operation” and “Installa- ing with automatic hold function. Material: anodized aluminum
tion”. User menus in English, German, Rated load: 100 mA / 50 V Total weight: 4.5 kg
French and Spanish.
Separate menu specific password protec- 1 Signal output
tion. One programmable signal output for
Display of process value, sample flow, measured value (freely scalable, linear or
alarm status, time and battery charge bilinear) or as continuous control outputs
condition. (control parameters programmable).
Current loop: 0/4 - 20 mA
Maximum burden: 510