4
Richard B. Price
Table 4.1 Radiometric terminology used to describe the output from a light source [3]
Term Units Symbol Notes
Radiant power or Watt W Radiant energy delivered per unit time (J/s)
radiant flux
Radiant energy Joule J This describes the energy from the source (W/s)
Radiant exposure Joule per square J/cm2 This describes the energy received per unit area.
centimeter Sometimes this is incorrectly described as “energy
density”
Radiant energy density Joule per cubic J/cm3 This is the energy in a defined volume (cm3)
centimeter
Radiant exitance or Watt per square W/cm2 Radiant power (flux) that is emitted from a surface,
Radiant emittance centimeter e.g., the tip of a curing light
Irradiance (incident Watt per square W/cm2 Radiant power (flux) incident on a known surface
irradiance) centimeter area. This is an averaged value over the surface area
Spectral radiant power Watt per nanometer W/nm Radiant power at each wavelength of the
electromagnetic spectrum in nanometers (nm)
Spectral irradiance Watt per square W/cm2/nm Irradiance at each wavelength of the
centimeter per electromagnetic spectrum in nanometers (nm)
nanometer
it changes the area of the light tip from 78.5 mm2 4.1.3 Light Beam Uniformity
(10-mm diameter) to 38.5 mm2 (7-mm diame-
ter). This doubles the radiant exitance (irradi- A laser beam profiler records the spatial intensity
ance) at the tip. Thus, just by changing the tip profile of a light beam, and can be used to measure
diameter, two curing lights can deliver the same the beam uniformity from dental curing lights [14–
irradiance, but very different radiant power out- 19]. A digital camera records an image of the light
puts [14]. Consequently, in the era of bulk filling output on a specified diffusing screen. The light
and bulk curing of large restorations, the radiant received by each camera pixel is then displayed as a
power (in mWatts), the active tip diameter (mm), color-coded two- or three-dimensional image. Such
and the irradiance (mW/cm2) should all be images of curing lights show that many curing
reported. Otherwise, the clinician may not rec- lights have “hot spots” of high irradiance and “cold
ognize that they are actually using a low-pow- spots” where the irradiance is lower [14–19].
ered curing light that uses a small diameter light Examples of inhomogeneous distributions of irradi-
tip to boost the irradiance. ance across the emitting tips of different dental light
curing units are seen in Fig. 4.2.
The beam profile images show that the irradi-
4.1.2 Spectral Radiant Power ance homogeneity depends on the design of the
curing light and that using a single irradiance
To the human eye, all dental curing lights appear value does not describe the irradiance across the
to deliver the same blue light, but this is not the entire light tip. Instead this is an average of the
case. Instead, the emission spectrum from dental high and low irradiance values that are present
curing lights can be very different (Fig. 4.1 A-D), across the light tip. In addition, the use of multiple
and thus the lights will have very different effects LED pads, each producing a different color, in
on the photoinitiators found within the resins order to produce a broad spectral range in multi-
used in dentistry. Consequently, it is important peak or polywave® curing lights can cause prob-
that the spectral radiant power from the curing lems of spectral inhomogeneity as well as
light be reported so that the user can judge if the irradiance inhomogeneity across the light tip [16].
light they are using is ideal for the resin they are The beam profile images also show that unless
trying to photocure (Fig. 4.1). the operator is careful, the irradiance and radiant
4 The Dental Curing Light 45
Fig. 4.1 Emission
spectra from four
contemporary curing
lights. Note the
differences between the
lights
exposure received at the gingival margins of the duced a broad spectrum of electromagnetic energy,
proximal boxes from some curing lights may be from near ultra violet (UV) and into the infrared
insufficient to adequately polymerize the adhesive region. To deliver just the blue light required to pho-
or the resin restoration at the bottom of the proximal tocure the resin, the light from the QTH bulb was
boxes [14, 18, 19]. Not only are these regions the aggressively filtered to remove the UV and infrared
most difficult to reach with the curing light, they are wavelengths of light.
also the furthest away from the light source [20, 21]. Inside the QTH bulb is a halogen gas and a
Consequently, resin at these locations will receive tungsten filament [1, 25]. When electrical current
the least amount of light and may well be under- passes through the filament, the tungsten wire
polymerized, unless the operator pays close atten- becomes incandescent and atoms are vaporized
tion to ensure that they position the light tip properly from its surface. This releases a large amount of
and use the light for long enough so that the resin electromagnetic energy, most of which is emitted
receives an adequate amount of light. in the infrared region. When the current is turned
off, the filament cools, and the halogen gas rede-
posits the vaporized tungsten atoms back onto the
4.2 evelopment of Dental
D filament surface [25]. This process is called “the
Curing Lights halogen cycle.” A fan inside the unit cools the
reflector, bulb, and filters, but the surfaces still
Four types of blue light sources have been used in become very hot. Vapors from solvents, cleaning
dentistry: quartz-tungsten-halogen (QTH), plasma agents, or moisture in the operatory air can all be
arc (PAC), argon-ion laser, and light- emitting deposited onto the reflector surface. This dulls
diodes (LED). Examples of PAC, QTH, and LED the reflector surface and gradually reduces the
curing lights are shown below (Fig. 4.3), General amount of light delivered by the curing light. All
guidelines for using curing lights have been pro- this can occur without any outward sign of a
posed [1–3, 22–24]. decrease in light output, thus it is essential to
monitor the light output on a regular basis. Under
ideal conditions, the QTH bulb should last about
4.2.1 Quartz-Tungsten-Halogen 50 h before it burns out [1, 2], but the lifespan of
(QTH) Lights the bulb will be shortened if the user turns the
power supply off immediately after light curing
Blue light-emitting curing lights initially used a in order to stop the noise from the cooling fan.
quartz-tungsten-halogen (QTH) light bulb that pro- This occurs because the filament inside the bulb
4.2.4 L
ight-Emitting Diode (LED)
4.2.2 Plasma Arc (PAC) Lights Technology
In an attempt to reduce light exposure times, high- The next innovation in dental photocuring was
power PAC lights were introduced. These lights introduced in the early 1990s when blue LED
claimed to deliver a very high irradiance and only curing lights became available [27, 28]. Using
required a short 3–5-s exposure time to light cure a LEDs as the light source has many advantages
2-mm-thick increment of resin composite. In con- because the units are solid state, battery operated,
trast to an incandescent filament, PAC lights pro- and lightweight, and the light they produce does
duce light from two tungsten electrodes that are not need to be filtered. The LED emitters are very
surrounded by xenon gas. When a high voltage is efficient, and they deliver at least two to three
applied, a spark is formed that ionizes the xenon times the luminous output per watt compared to a
gas to produce an electrically conductive gaseous typical QTH bulb [2]. Also, when used correctly,
medium (a plasma) that maintains the spark [1, the LED emitters can provide a long working life
25]. Intense optical filtering is required within PAC of many thousands of hours [1, 2] instead of just
lights to prevent the emission of unwanted ionizing 50 or so hours for a QTH bulb [1, 2].
radiation and infrared light that would produce an LEDs are semiconductor light sources that
unacceptable temperature increase. have been doped with impurities to create
Although they are excellent curing lights, specific electron-excessive (n) and electron-
PAC lights are expensive, they have a noisy fan, depleted (p) surfaces in two different semi-
they are large, and they cannot be battery oper- conductor materials [2]. The semiconductor
ated. Consequently, they have become less pop- substrate on the n side, or cathode, has an excess
ular in recent years. of electrons and will rise in potential when suf-
48 R.B. Price
as Lucirin TPO®; however, most will activate are only weakly activated by second-generation
the recently introduced germanium-based pho- LED curing lights that emit little light below
toinitiator, Ivocerin™, that can be activated by 420 nm (Fig. 4.5). To solve the problem caused
wavelengths of light up to 460 nm [29, 30]. by the narrow emission spectrum from the blue-
only LED units, additional LED emitters were
added to the blue LED array. Each additional
4.2.7 T
hird-Generation LED Lights: LED pad produces a different narrow range of
Multi-wave, Multi-peak, wavelengths, and by incorporating several differ-
and Polywave® ent LED pads into the curing light, a broad spec-
tral range can now be delivered. These curing
The camphorquinone (CQ) photoinitiator used in lights have been called “third-generation” LED
the majority of dental resins has a bright yellow curing lights, meaning that they emit light of
color, and this yellow color is problematic when more than one wavelength range, usually violet
trying to produce very light or translucent shades and blue (Fig. 4.6). Curing lights of this genera-
of restorative resin. As a result, some manufac- tion have also been described as “multi-wave” or
turers have incorporated co-initiators to boost the “multi-peak” dental curing lights, with one man-
effectiveness of CQ so that they could reduce the ufacturer trademarking the name “Polywave®” to
overall concentration of CQ, while others started signify this concept within their product line.
using more efficient “alternative” photoinitiators The emission spectrum from broad-spectrum
to reduce the concentration of CQ [1]. These LED lights should be able to activate all the pho-
“alternative” initiators, such as Lucirin TPO®, toinitiators currently used in dental resins and
impart less of a yellow color than those that con- bonding agents. Since manufacturers rarely dis-
tain CQ. However, these alternative photoinitia- close all of the proprietary constituents used in
tors are activated by shorter wavelengths of light their resins, it is best to assume that if a resin
closer to violet light (at or below 410 nm), and manufacturer produces a broad spectrum LED
although they can be activated by the broader curing light, then their resins will likely benefit
emission spectrum from QTH curing lights, they from the use of a broad spectrum LED curing
Fig. 4.6 Emission
spectra from three
broad-spectrum, LED
curing lights. Note the
emission spectra are all
different among these
lights and from the
second generation
lights in Fig. 4.5
light, e.g., Ivoclar Vivadent or Ultradent Products. Unfortunately, the irradiance delivered to a
Conversely, if the manufacturer, e.g., 3M, makes resin from these turbo light guides declines rap-
a single peak wavelength curing light only, then idly as the distance from the tip increases [21]
their resins are unlikely to require the use of a and at 6 mm away from the tip; these light guides
broad-spectrum polywave® LED curing light. usually deliver less irradiance than a conven-
tional light guide [21]. Some curing lights also
include a special boosted output mode, where the
4.2.8 “Turbo” Light Guides voltage to the light source is raised for a short
time. However, overdriving the light source to
Conventional, glass-fibered light guides come in boost the light output shortens its life expectancy,
a variety of diameters, but usually the proximal and it also generates more heat.
entrance and the distal exit apertures of the light
guide are the same physical size. In an attempt to
increase the irradiance, the proximal entrance to 4.2.9 Exposure Reciprocity
the “turbo” light guide is larger than the exit
diameter (Fig. 4.7). Both the entrance and the exit Photopolymerization reaction rates are affected
of these light guides contain the same number of by many factors such as the irradiance received,
fibers, but the fibers at the exit end are smaller the concentration of the photoinitiators, the
than at the entrance. Thus, the same number of quantum yield for radical generation, the effects
photons that enter into the wide entrance to the of radical recombination/termination, the impact
light guide are emitted over a smaller exit area. of oxygen inhibition, the resin viscosity, the
Since the same radiant power (number of pho- amount of heat generated, the material thickness,
tons) is now emitted over a smaller area, the radi- the matching of the resin and filler refractive
ant exitance, “irradiance” (mW/cm2), from these indices, the filler particle size, and the shade. As
turbo light guides is increased [31]. a general recommendation, it has been proposed
4 The Dental Curing Light 51
that a QTH curing light delivering an irradiance of cure within the resin increases by the loga-
of 400 mW/cm2 should be used for 60 s to ade- rithm of the radiant exposure, i.e., there is an
quately polymerize a 1.5–2-mm-thick increment exponential relationship between depth of cure
of resin composite [32]. If this irradiance is mul- and the amount of energy delivered, -doubling
tiplied by the exposure time, this means that a the radiant exposure received will not double the
radiant exposure of 24–36 J/cm2 should ade- depth of cure [42]. Thus, it is inadvisable to
quately polymerize most resin composites, rec- exceed the resin manufacturer’s recommended
ognizing that brand, shade, and opacity of the increment thickness and hope that by using a
resin will all affect the exact amount of energy longer exposure time, the resin will be fully
required [33, 34]. Some believe that using a cured.
high-intensity curing light is a way to provide
timesavings and increase profitability [35].
Behind this is the belief that if the curing light 4.2.10 High-Output Curing Lights
delivers double the irradiance, then you can and Stress Development
halve the time spent light curing. However, this During Polymerization
is not the case. True reciprocity between the
duration of light exposure and the irradiance to Fast polymerization of dental resin-based com-
achieve the same level of resin polymerization posites is thought to adversely affect both the
does not exist [36–40]. Instead, it has been cal- mechanical properties of the polymer network
culated that the reaction rate is at best related to [43, 44] and the physical properties of the resin
the square root of the irradiance [41]. Thus, even [45]. With some contemporary dental resins that
under ideal circumstances, reducing the light use CQ photoinitiator systems, rapid photopoly-
exposure time from 20 to 5 s will require 16x the merization produces shorter chain lengths
irradiance to achieve the same level of polymer- because there is insufficient time for long poly-
ization within the resin. Thus, delivering mer chains to form before the solid state is
4000 mW/cm2 for 5 s will not achieve the same reached and no further movement is possible
resin polymerization as 500 mW/cm2 delivered [43]. Also, as long as the system is a liquid, it
for 40 s, despite the fact that in both cases the can physically deform, and little stress is devel-
resin would receive the same 20 J/cm2 of radiant oped. However, beyond the “gel point,” the resin
exposure. Additionally, it is known that the depth becomes a solid, and the polymerization shrink-
52 R.B. Price
age deforms the resin thus creating stresses both 4.2.12 Batteries
within the network and at the bonded interfaces.
If this gel point is reached very rapidly, as would Originally, LED curing lights were powered by
occur when using a high irradiance, this may nickel-cadmium (NiCad) batteries. These bat-
result in increased stress [46], increased bond teries can be damaged if they are deeply dis-
failure, and more gaps between the tooth and charged, or if they are overcharged. Smaller,
restorative material [47, 48]. high-capacity, nickel-metal hydride batteries
soon replaced these NiCad batteries [1], but
most currently available curing lights now use
4.2.11 Soft-Start Exposures either lithium-ion batteries or “ultracapacitors.”
Ultracapacitors can deliver high bursts of power
In an attempt to decrease the damaging effects for a short time and are used in at least one LED
of rapid photopolymerization, a range of differ- curing light (Demi Ultra, Kerr Corp., Orange,
ent light curing cycles have been proposed. An CA) that can deliver approximately 25, 10-s light
example is the “soft-start” curing mode, where a exposures before the unit must be recharged.
low irradiance value is applied to the resin sur- Although ultracapacitors cannot store the same
face for a short time and then, either immedi- amount of electrical energy as lithium-ion bat-
ately or over a short period of time, the light teries, they can be fully recharged in as little as
output increases to its full operating level for the 40 s and should provide a sufficient number of
remainder of the exposure [47, 49, 50]. This light exposures to treat one patient.
technique is intended to allow the partially pho-
topolymerized resin to deform and thus relieve
the stresses that are developing within the resin 4.2.13 Curing Light Output
that is cured and rigid. Another method, the Monitoring
pulse-delay, delivers an initial low-level, short-
duration exposure, and then the light is turned Although it has been reported that only 46.8% of
off. After waiting some 3–5 min, the final expo- dentists use a radiometer regularly to monitor the
sure is provided at the full light output [47]. It irradiance [6], it is important that the light output
was hoped that the use of such an exposure from the curing light be monitored on a regular
mode would reduce the rate at which the photo- basis because the light output will decrease for
polymerized material reacted, thus allowing many reasons [25] such as aging and degradation
movement within the unreacted or partially of the light source, the buildup of scale on the
reacted resin that would then reduce the overall fiber-optic light probe after autoclaving [58],
amount of polymerization contraction stress breakage or fracture at the light tip [59, 60], or
[46, 47, 51–54]. However, it has been reported the presence of cured resin on the light tip [59,
that using the pulse-delay method results in a 61] (Fig. 4.8). In addition, disinfectant sprays can
lower cross-link density within the photocured erode the O-rings that stabilize the light guide,
resin. This produced polymers that were more and liquids can become baked onto the lens or
susceptible to softening in ethanol [55] and that reflectors inside the housing, thus reducing the
also stained more readily [56]. Clinically, none light output [62].
of these attempts to control the polymerization Since, dental curing lights have different
shrinkage rate and stress development have emission spectra and different light tip diame-
been found to provide significantly better per- ters, low-cost dental radiometers are often
formance when compared to a continuous light unable to accurately report the irradiance value
exposure [1, 57], and in one study restorations from a range of curing lights [63–67]. This can
photopolymerized with the pulse-delay tech- lead clinicians to believe that they are delivering
nique had the highest failure rate after 7 years of sufficient energy to their restorations, when this
service in the mouth [57]. may not be the case [64]. A new type of dental
4 The Dental Curing Light 53
Fig. 4.9 Examples of four commercial dental radiometers. The Bluephase Meter II on the left has a large entrance aperture
and currently is the only dental radiometer that is able to measure all the power from a 13-mm diameter light tip
close to soft tissue, or where the remaining tooth ated retinal aging and accelerate age-related macu-
structure is thin, there is a risk of thermal damage lar degeneration (ARMD) [84, 85]. Most countries
when the LCU delivers greater than 1200 mW/ follow international guidelines, such as those from
cm2 [75]. It is recommended that when the LCU the International Commission on Non-Ionizing
is used for extended curing times, then the tooth Radiation Protection (ICNIRP) and American
be air-cooled, or the operator should wait at least Conference of Governmental Industrial Hygienists
2–3 s between every 10 s of light curing [76]. (ACGIH), that limit exposure to all types of haz-
ardous optical radiation [83, 84]. If the operator
does not use appropriate eye protection, these limits
4.3.2 The “Blue Light Hazard” may be easily reached during a normal workday by
dental personnel using high-power curing lights. It
The dentist has a duty to protect both the patient and has been shown that, after looking at the light for
their employees from harm [77]. Recently, there has even just the first second of the curing cycle before
been concern that blue light from high-power curing averting the eyes, it may take as little as seven light
lights may cause ocular damage [77–81]. This haz- exposures to exceed the maximum daily cumulative
ard can be prevented by using appropriate eye pro- exposure [79]. It should be noted that the maximum
tection that is provided by the orange-colored shields recommended exposure times have been calculated
or glasses (blue light blockers) that are supplied for individuals who have normal photosensitivity.
with the curing lights, but these items are not univer- Children, persons who have had cataract surgery, or
sally used in the dental office [5, 6, 82]. those who are taking photosensitizing medications
The “blue light hazard” is greatest at 440 nm [83, have a greater susceptibility for retinal damage.
84], and this is within the output range from dental In these circumstances, ocular injury may occur
LCUs [1, 3] (Fig. 4.10). Blue light is transmitted after shorter exposure times [83, 84]. If appropri-
through the ocular media and absorbed by the retina ate eye protection is used when light curing, instead
at the back of the eye. While high levels of blue of looking away from the bright blue light on the
light can cause immediate and irreversible retinal tooth, the operator can safely watch what they are
burning, of greater concern is the chronic exposure doing and so can increase the amount of light they
to low levels of blue light that may cause acceler- deliver to the restoration [86–90] (Fig. 4.11).
Fig. 4.10 Graphical
depiction of the blue
light hazard related to
the wavelength of light.
Superimposition of the
emission spectrum from
a second-generation
LED dental curing light
shows that all the blue
light delivered from this
LED curing light is in
the hazardous range
4 The Dental Curing Light 55
a b
Fig. 4.12 A large diameter light tip (a) can cover an entire molar tooth, whereas smaller diameter light tip (b) will
require multiple exposures to cover an entire molar mesial-occlusal-distal restoration
4.3.6 E
rgonomics and Clinical
Access
a b
Fig. 4.14 Effect of distance from the light tip on the irradiance. The distance between the light tip and the resin can
easily be 6 to 8-mm (a). Depending on their design, curing lights can deliver a similar irradiance at the light tip (C and
D), but very different at 6 to 10-mm distances
a b
Fig. 4.15 Example of the ability of different dental curing lights to access the mandibular second molar tooth. Light b
can only access the second molar at an angle whereas light (a) has direct access to the tooth
Some commercial barriers can reduce the minimal effect on light output, but it can appear
radiant exitance by up to 40% [15, 98–100]. messy [60, 98, 100].
Latex-based barriers have been reported to cause When using cold sterilizing techniques, only
the greatest reduction in irradiance. This can pro- cleaning solutions approved by the light manu-
duce significantly lower resin conversion values facturer should be used because some disinfec-
[100] unless the exposure time is extended. tant sprays can damage the plastic casing, remove
Consequently, the output from curing lights the paint, or damage the rubber O-rings used to
should be tested with the barrier over the light tip. stabilize light guides. The disinfectant spray can
Clear, plastic food wrap is a highly effective and then leak into the unit and be baked onto the light
inexpensive infection control barrier that has source or reflectors inside the curing light, thus
58 R.B. Price
4.3.9 Choosing a Curing Light depth of cure scrape tests using different
shades of resin that have been light-cured at
When light curing, the clinician should deliver clinically representative distances and angles.
sufficient radiant exposure at the correct wave- 2. Learn how to use the curing light to maximize
lengths of light required by the photoinitiator(s) the amount of energy delivered to the resin.
in the resin they are using. Thus, when making a 3. Maximize the output from the curing light by
decision about which curing light to purchase, routinely examining the light tip for damage
the clinician should ask: and remove remnants of previously cured
resin. Clean or replace the tip as necessary.
1. What is the radiant power output (watts) 4. Protect the eyes of everyone in the operatory
from the light? who could be exposed to the bright light
2. What is the emission spectrum from the cur- from the curing light by using appropriate
ing light? Does the spectral emission from the orange (blue light blocking) safety glasses or
light match the sensitivity of the resin I use? shields.
3. What is the active tip size, and how much of 5. When light curing, remember that light travels
a typical restoration will receive useful light? in a straight line. Place the central axis of the
4. Is the light beam homogeneous? tip of the curing light directly over and normal
5. What is the effect of distance from the light to the resin surface; the emitting end should
tip on the irradiance received by my restora- be parallel to the resin surface being exposed.
tions? Does the irradiance drop dramatically Where undercuts that cause shadows are pres-
as the distance from the tip increases? ent, move the light tip around and increase the
6. Does the unit have an ergonomic shape that exposure time. Use supplementary buccolin-
is comfortable to use, and does it allow full gual curing (but beware of overheating).
access to all regions of the mouth? 6. Protect the oral mucosa from the light with
7. Could the light produce an unacceptable gauze and air-cool, or wait several seconds
temperature rise in the pulp or oral mucosa? between each light curing cycle when using a
8. How do I disinfect the light? powerful curing light that may produce a
9. What barriers are available and how do they damaging increase in temperature.
affect the light output?
10. If it is battery operated, how frequently will Acknowledgments The author wishes to acknowledge
it require replacing, and at what cost? the valuable contributions of Professor F. Rueggeberg
when preparing this text and his contributions to dental
11. Is the curing light approved for use in my education.
country?
12. Who do I contact if I have a problem with the
light, or if the patient has a complaint? References
1. Rueggeberg FA. State-of-the-art: dental photocuring –
a review. Dent Mater. 2011;27:39–52.
4.3.10 Clinical Recommendations 2. Jandt KD, Mills RW. A brief history of LED photopo-
When Using a Curing Light lymerization. Dent Mater. 2013;29:605–17.
3. Price RB, Ferracane JL, Shortall AC. Light-curing
Having chosen a curing light, the following clini- units: a review of what we need to know. J Dent Res.
2015;94:1179–86.
cal recommendations should help improve how 4. Santini A, Turner S. General dental practitioners’
the curing light is used [22]. knowledge of polymerisation of resin-based compos-
ite restorations and light curing unit technology. Br
1. Monitor the performance of curing light, and Dent J. 2011;211:E13.
5. McCusker N, Bailey C, Robinson S, Patel N, Sandy
keep a logbook of the output from the light JR, Ireland AJ. Dental light curing and its effects
from the date of purchase. To guide any on color perception. Am J Orthod Dentofac Orthop.
adjustments in exposure duration, carry out 2012;142:355–63.
60 R.B. Price
6. Kopperud SE, Rukke HV, Kopperud HM, Bruzell 22. Price RB, Shortall AC, Palin WM. Contemporary
EM. Light curing procedures - performance, knowl- issues in light curing. Oper Dent. 2014;39:4–14.
edge level and safety awareness among dentists. J 23.
Shortall AC, Price RB, MacKenzie L, Burke
Dent. 2017;58:67–73. FJ. Guidelines for the selection, use, and mainte-
7. El-Mowafy O, El-Badrawy W, Lewis DW, Shokati nance of LED light-curing units - part II. Br Dent J.
B, Soliman O, Kermalli J, et al. Efficacy of halogen 2016;221:551–4.
photopolymerization units in private dental offices in 24.
Shortall AC, Price RB, MacKenzie L, Burke
Toronto. J Can Dent Assoc. 2005;71:587. FJ. Guidelines for the selection, use, and mainte-
8. Ernst CP, Busemann I, Kern T, Willershausen nance of LED light-curing units - Part 1. Br Dent J.
B. Feldtest zur Lichtemissionsleistung von 2016;221:453–60.
Polymerisationsgeräten in zahnärztlichen Praxen. 25. Rueggeberg F. Contemporary issues in photocuring.
Dtsch Zahnarztl Z. 2006;61:466–71. Compend Contin Educ Dent Suppl. 1999:S4–15.
9. Barghi N, Fischer DE, Pham T. Revisiting the inten- 26. Yaman BC, Efes BG, Dorter C, Gomec Y, Erdilek D,
sity output of curing lights in private dental offices. Buyukgokcesu S. The effects of halogen and light-
Compend Contin Educ Dent. 2007;28:380–4. quiz emitting diode light curing on the depth of cure and
5-6 surface microhardness of composite resins. J Conserv
10. Hegde V, Jadhav S, Aher GB. A clinical survey
Dent. 2011;14:136–9.
of the output intensity of 200 light curing units in 27. Fujibayashi K, Shimaru K, Takahashi N, Kohno
dental offices across Maharashtra. J Conserv Dent. A. Newly developed curing unit using blue light-
2009;12:105–8. emitting diodes. Dent Jpn. 1998;34:49–53.
11. Al Shaafi M, Maawadh A, Al QM. Evaluation of light 28. Mills RW, Jandt KD, Ashworth SH. Dental compos-
intensity output of QTH and LED curing devices in ite depth of cure with halogen and blue light emitting
various governmental health institutions. Oper Dent. diode technology. Br Dent J. 1999;186:388–91.
2011;36:356–61. 29. Moszner N, Fischer UK, Ganster B, Liska R,
12. Hao X, Luo M, Wu J, Zhu S. A survey of power Rheinberger V. Benzoyl germanium derivatives as
density of light-curing units used in private dental novel visible light photoinitiators for dental materials.
offices in Changchun City, China. Lasers Med Sci. Dent Mater. 2008;24:901–7.
2015;30:493–7. 30. Burtscher P. Ivocerin in comparison to camphorqui-
13. Maghaireh GA, Alzraikat H, Taha NA. Assessing
none. Ivoclar Vivadent Report, No. 19. July 2013.
the irradiance delivered from light-curing units in 31. Curtis JW Jr, Rueggeberg FA, Lee AJ. Curing
private dental offices in Jordan. J Am Dent Assoc. efficiency of the turbo tip. Gen Dent. 1995;43:
2013;144:922–7. 428–33.
14. AlShaafi MM, Harlow JE, Price HL, Rueggeberg
32. Rueggeberg FA, Caughman WF, Curtis JW Jr. Effect
FA, Labrie D, AlQahtani MQ, et al. Emission of light intensity and exposure duration on cure of
characteristics and effect of battery drain in "budget" resin composite. Oper Dent. 1994;19:26–32.
curing lights. Oper Dent. 2016;41:397–408. 33. Fan PL, Schumacher RM, Azzolin K, Geary R,
15. Vandewalle KS, Roberts HW, Rueggeberg FA. Power Eichmiller FC. Curing-light intensity and depth
distribution across the face of different light guides of cure of resin-based composites tested accord-
and its effect on composite surface microhardness. ing to international standards. J Am Dent Assoc.
J Esthet Restor Dent. 2008;20:108–17. discussion 18 2002;133:429–34.
16.
Price RB, Labrie D, Rueggeberg FA, Felix 34. Calheiros FC, Daronch M, Rueggeberg FA, Braga
CM. Irradiance differences in the violet (405 nm) and RR. Degree of conversion and mechanical properties
blue (460 nm) spectral ranges among dental light- of a BisGMA:TEGDMA composite as a function of
curing units. J Esthet Restor Dent. 2010;22:363–77. the applied radiant exposure. J Biomed Mater Res B
17. Michaud PL, Price RB, Labrie D, Rueggeberg
Appl Biomater. 2008;84:503–9.
FA, Sullivan B. Localised irradiance distribution 35.
Christensen GJ. Ask Dr. Christensen. Dental
found in dental light curing units. J Dent. 2014;42: Economics. 2009;99. http://www.dentaleconomics.
129–39. com/articles/print/volume-99/issue-9/departments/
18. Price RB, Labrie D, Rueggeberg FA, Sullivan B,
ask-dr-christensen/ask-dr-christensen.html.
Kostylev I, Fahey J. Correlation between the beam 36. Musanje L, Darvell BW. Polymerization of resin
profile from a curing light and the microhardness of composite restorative materials: exposure reciprocity.
four resins. Dent Mater. 2014;30:1345–57. Dent Mater. 2003;19:531–41.
19. Shimokawa CA, Turbino ML, Harlow JE, Price HL, 37. Leprince JG, Hadis M, Shortall AC, Ferracane
Price RB. Light output from six battery operated den- JL, Devaux J, Leloup G, et al. Photoinitiator type
tal curing lights. Mater Sci Eng C Mater Biol Appl. and applicability of exposure reciprocity law in
2016;69:1036–42. filled and unfilled photoactive resins. Dent Mater.
20. Yearn JA. Factors affecting cure of visible light acti- 2011;27:157–64.
vated composites. Int Dent J. 1985;35:218–25. 38. Hadis M, Leprince JG, Shortall AC, Devaux J, Leloup
21. Price RB, Dérand T, Sedarous M, Andreou P, Loney G, Palin WM. High irradiance curing and anomalies
RW. Effect of distance on the power density from two of exposure reciprocity law in resin-based materials. J
light guides. J Esthet Dent. 2000;12:320–7. Dent. 2011;39:549–57.
4 The Dental Curing Light 61
n anocomposites and conventional dental composites. 86. Price RB, McLeod ME, Felix CM. Quantifying light
Clin Oral Investig. 2009;13:309–16. energy delivered to a class I restoration. J Can Dent
71. Leprince J, Devaux J, Mullier T, Vreven J, Leloup Assoc. 2010;76:a23.
G. Pulpal-temperature rise and polymeriza- 87. Seth S, Lee CJ, Ayer CD. Effect of instruction on
tion efficiency of LED curing lights. Oper Dent. dental students’ ability to light-cure a simulated res-
2010;35:220–30. toration. J Can Dent Assoc. 2012;78:c123.
72. Matalon S, Slutzky H, Wassersprung N, Goldberg- 88. Federlin M, Price R. Improving light-curing
Slutzky I, Ben-Amar A. Temperature rises beneath instruction in dental school. J Dent Educ. 2013;77:
resin composite restorations during curing. Am J 764–72.
Dent. 2010;23:223–6. 89. Mutluay MM, Rueggeberg FA, Price RB. Effect
73. Shortall A, El-Mahy W, Stewardson D, Addison O, of using proper light-curing techniques on energy
Palin W. Initial fracture resistance and curing tem- delivered to a class 1 restoration. Quintessence Int.
perature rise of ten contemporary resin-based com- 2014;45:549–56.
posites with increasing radiant exposure. J Dent. 90. Price RB, Strassler HE, Price HL, Seth S, Lee
2013;41:455–63. CJ. The effectiveness of using a patient simula-
74. Gomes M, DeVito-Moraes A, Francci C, Moraes
tor to teach light-curing skills. J Am Dent Assoc.
R, Pereira T, Froes-Salgado N, et al. Temperature 2014;145:32–43.
increase at the light guide tip of 15 contemporary 91. Nitta K. Effect of light guide tip diameter of LED-
LED units and thermal variation at the pulpal floor light curing unit on polymerization of light-cured
of cavities: an infrared thermographic analysis. Oper composites. Dent Mater. 2005;21:217–23.
Dent. 2013;38:324–33. 92. Ash MM, Nelson SJ, Ash MM. Dental anatomy,
75. Mouhat M, Mercer J, Stangvaltaite L, Örtengren
physiology, and occlusion. 8th ed. Philadelphia:
U. Light-curing units used in dentistry: factors asso- W.B. Saunders; 2003.
ciated with heat development—potential risk for 93. Corciolani G, Vichi A, Davidson CL, Ferrari M. The
patients. Clin Oral Investig. 2017;21:1687–96. influence of tip geometry and distance on light-
76. Onisor I, Asmussen E, Krejci I. Temperature rise dur- curing efficacy. Oper Dent. 2008;33:325–31.
ing photo-polymerization for onlay luting. Am J Dent. 94. Vandewalle KS, Roberts HW, Andrus JL, Dunn
2011;24:250–6. WJ. Effect of light dispersion of LED curing lights
77. Stamatacos C, Harrison JL. The possible ocular haz- on resin composite polymerization. J Esthet Restor
ards of LED dental illumination applications. J Tenn Dent. 2005;17:244–54. discussion 54-5
Dent Assoc. 2013;93:25–9. quiz 30-1 95. Xu X, Sandras DA, Burgess JO. Shear bond strength
78. Bruzell Roll EM, Jacobsen N, Hensten-Pettersen
with increasing light-guide distance from dentin. J
A. Health hazards associated with curing light in the Esthet Restor Dent. 2006;18:19–27. discussion 8
dental clinic. Clin Oral Investig. 2004;8:113–7. 96. Price RB, Labrie D, Whalen JM, Felix CM. Effect
79. Labrie D, Moe J, Price RB, Young ME, Felix
of distance on irradiance and beam homogeneity
CM. Evaluation of ocular hazards from 4 types of cur- from 4 light-emitting diode curing units. J Can Dent
ing lights. J Can Dent Assoc. 2011;77:b116. Assoc. 2011;77:b9.
80. McCusker N, Lee SM, Robinson S, Patel N, Sandy 97. Konerding KL, Heyder M, Kranz S, Guellmar A,
JR, Ireland AJ. Light curing in orthodontics; should Voelpel A, Watts DC, et al. Study of energy trans-
we be concerned? Dent Mater. 2013;29:e85–90. fer by different light curing units into a class III
81. Price RB, Labrie D, Bruzell EM, Sliney DH, Strassler restoration as a function of tilt angle and distance,
HE. The dental curing light: a potential health risk. J using a MARC patient simulator (PS). Dent Mater.
Occup Environ Hyg. 2016;13:639–46. 2016;32:676–86.
82. Hill EE. Eye safety practices in U.S. dental
98. Scott BA, Felix CA, Price RB. Effect of dispos-
school restorative clinics, 2006. J Dent Educ. able infection control barriers on light output from
2006;70:1294–7. dental curing lights. J Can Dent Assoc. 2004;70:
83. American Conference of Governmental Industrial
105–10.
Hygienists (ACGIH): TLVs and BEIs Based on 99. Coutinho M, Trevizam NC, Takayassu RN, Leme
the Documentation for Threshold Limit Values AA, Soares GP. Distance and protective barrier
for Chemical Substances and Physical Agents and effects on the composite resin degree of conversion.
Biological Exposure Indices. 2015; Cincinnati, OH. Contemp Clin Dent. 2013;4:152–5.
84. International Commission on Non-Ionizing Radiation 100. Sword RJ, Do UN, Chang JH, Rueggeberg
Protection. Guidelines on limits of exposure to broad- FA. Effect of curing light barriers and light types on
band incoherent optical radiation (0.38 to 3 μm). radiant exposure and composite conversion. J Esthet
Health Phys. 1997;73:539–54. Restor Dent. 2016;28:29–42.
85. Ham WT Jr, Ruffolo JJ Jr, Mueller HA, Clarke AM, 101. Strassler HE. Successful light curing- not as easy as
Moon ME. Histologic analysis of photochemical it looks. Oral Health. 2013;103:18–26.
lesions produced in rhesus retina by short- wave- 102. Price RB, Felix CM, Whalen JM. Factors affecting
length light. Investig Ophthalmol Vis Sci. 1978;17: the energy delivered to simulated class I and class V
1029–35. preparations. J Can Dent Assoc. 2010;76:a94.