You are on page 1of 9

ORIGINAL ARTICLE

Comparison of adult and child radiation


equivalent doses from 2 dental cone-beam
computed tomography units
Anas Al Najjar,a Dan Colosi,b Lawrence T. Dauer,c Robert Prins,d Gayle Patchell,e Iryna Branets,f
Arthur D. Goren,g and Richard D. Faberh
Stony Brook and New York, NY, and Ft. Belvoir, Va

Introduction: With the advent of cone-beam computed tomography (CBCT) scans, there has been a transition
toward these scans’ replacing traditional radiographs for orthodontic diagnosis and treatment planning. Children
represent a significant proportion of orthodontic patients. Similar CBCT exposure settings are predicted to result
in higher equivalent doses to the head and neck organs in children than in adults. The purpose of this study was
to measure the difference in equivalent organ doses from different scanners under similar settings in children
compared with adults. Methods: Two phantom heads were used, representing a 33-year-old woman and
a 5-year-old boy. Optically stimulated dosimeters were placed at 8 key head and neck organs, and equivalent
doses to these organs were calculated after scanning. The manufacturers’ predefined exposure settings were
used. Results: One scanner had a pediatric preset option; the other did not. Scanning the child’s phantom
head with the adult settings resulted in significantly higher equivalent radiation doses to children compared
with adults, ranging from a 117% average ratio of equivalent dose to 341%. Readings at the cervical spine level
were decreased significantly, down to 30% of the adult equivalent dose. When the pediatric preset was used for
the scans, there was a decrease in the ratio of equivalent dose to the child mandible and thyroid. Conclusions:
CBCT scans with adult settings on both phantom heads resulted in higher radiation doses to the head and neck
organs in the child compared with the adult. In practice, this might result in excessive radiation to children
scanned with default adult settings. Collimation should be used when possible to reduce the radiation dose to
the patient. While CBCT scans offer a valuable tool, use of CBCT scans should be justified on a specific
case-by-case basis. (Am J Orthod Dentofacial Orthop 2013;143:784-92)

T
a
Resident, Department of Orthodontics and Pediatric Dentistry, School of Dental he development of cone-beam computed
Medicine, State University of New York at Stony Brook, Stony Brook, NY. tomography (CBCT) has introduced a transition
b
Clinical assistant professor and director, Division of Diagnostic Imaging,
Department of Prosthodontics and Digital Technology, State University of New toward these scans' replacing traditional
York at Stony Brook, Stony Brook, NY. radiographs taken for orthodontic diagnosis and
c
Associate attending physicist, Departments of Medical Physics and Radiology, Me- treatment planning. Since children are a significant
morial Sloan-Kettering Cancer Center, New York, NY.
d
Chief, Consequence Management Assistance Program, Defense Threat Reduction proportion of orthodontic patients, the use of the
Agency, Ft. Belvoir, Va. same CBCT exposure settings for children and adults is
e
Radiologic technologist and instructional support, School of Dental Medicine, predicted to result in higher equivalent doses to the
State University of New York at Stony Brook, Stony Brook, NY.
f
Clinical instructor, Department of Cariology and Comprehensive Care, College of head and neck organs in children than in adults.1,2
Dentistry, New York University, New York, NY. The CBCT radiation exposure to the surface tissues
g
Clinical associate professor, Department of Prosthodontics and Digital Technology, of adults has been previously measured with adult
School of Dental Medicine, State University of New York at Stony Brook, Stony
Brook, NY; clinical associate professor, Department of Cariology and Comprehen- male phantom heads using thermo-luminescent
sive care, College of Dentistry, New York University, New York, NY. dosimeters (TLDs).3-5 However, to our knowledge,
h
Program director and clinical associate professor, Department of Orthodontics and a comparative characterization of the radiation
Pediatric Dentistry, School of Dental Medicine, State University of New York at
Stony Brook, Stony Brook, NY. exposure from different CBCT scanners in children and
The authors report no commercial, proprietary, or financial interest in the adults is not available.
products or companies described in this article. Once cancer risk from x-ray radiation was confirmed
Reprint requests to: Anas Al Najjar, Department of Orthodontics and Pediatric
Dentistry, SUNY at Stony Brook, Stony Brook University School of Dental Med- early in the 20th century, the International Commission
icine, Sullivan Hall, Stony Brook, NY 11794; e-mail, anas_82@hotmail.com. on Radiological Protection (ICRP) has worked hard to
Submitted, June 2012; revised and accepted, January 2013. develop guidelines for x-ray exposure and safety proto-
0889-5406/$36.00
Copyright Ó 2013 by the American Association of Orthodontists. col.6 In the 1990s, the ICRP moved toward using
http://dx.doi.org/10.1016/j.ajodo.2013.01.013 the equivalent and effective dose (measured in
784
Al Najjar et al 785

Fig 1. Anthropomorphic phantom head.

Table I. Dosimeters per site


Adult (n) Child (n)
Brain 8 3
Cranium 4 3
Eye 2 2
Mandible 2 2 Fig 2. Optically stimulated luminescent dosimeter slots in
Cervical spin 1 2 the phantom head.
Thyroid 2 4
Parotid 2 1
Eye lens 2 2
to be less than that from multidetector CT used in med-
icine, it is still greater relative to conventional dental ra-
microsieverts) as a measure of organ dose and biologic diographs.4
risk, instead of the more traditional absorbed dose.5,7 Radiographic examination should be performed only
The equivalent dose takes into account tissue when indicated by the patient’s dental history, the den-
sensitivity to radiation and provides a more accurate tist’s physical examination, or laboratory findings.9 Jus-
picture of the detrimental effects of radiation to the tification of radiographic examination in children is
patient. Although simple periapical x-rays expose especially important since children’s tissues are different
patients to the equivalent of a few days’ worth of in size and position and are under constant growth,
background radiation and potentially increase cancer replication, and differentiation, making them signifi-
risk by less than 1 in 1,000,000, medical computed cantly more susceptible to radiation damage compared
tomography (CT) scans expose patients to the with adult tissues. Radiation risk highly depends on
equivalent of 2 to 3 years of background radiation and age, with risk inversely proportional to age; for
potentially increase cancer risk by 1 in 1000 to 1 in the same radiation doses, children less than 10 years of
10,000.3 CBCT scans offer an alternative to medical CT age have about a 3-times higher risk than a 30-year-
imaging, with comparable quality of bony images but old adult.7 In addition, the risk for females is typically
much less radiation, and they are significantly less ex- higher than that for males.2
pensive. Previous studies evaluating radiation doses to
In recent years, more orthodontists have moved phantom heads from CBCT scans have used TLDs on adult
toward using CBCT scans for diagnosis, as they offer male phantom heads.3-5 In this study, we used optically
a “1 scan covers all” type of convenience to stimulated luminescent (OSL) dosimeters (NanoDot;
practitioners.8 The rapid adoption of CBCT into practice Landauer, Glenwood, Ill). This is a newer technology of
has outstripped the time needed to conduct well- dosimeters, which we used to characterize CBCT
controlled studies on x-ray radiation from CBCT and radiation exposure to a woman’s phantom head and
to implement or suggest guidelines on CBCT use. a child’s phantom head. The advantages of OSL
Whereas radiation exposure from CBCT has been shown dosimeters over TLDs include faster readouts and more

American Journal of Orthodontics and Dentofacial Orthopedics June 2013  Vol 143  Issue 6
786 Al Najjar et al

Reference Imaging System, Norfolk, Va) corresponding


Table II. Exposure settings
to an average 33-year-old woman who is 160 cm in
Resolution Exposure Kilovoltage Current height and 55 kg in weight (Fig 1) and a child male
Type of scan (mm) time (s) (kV[p]) (mA) anthropomorphic phantom (model 705; Computerized
i-CAT Reference Imaging System) corresponding to an aver-
Full head 0.3 8.9 120 5
Maxilla 0.2 26.9 120 5
age 5-year-old boy who is 110 cm in height and 19
Maxilla 0.4 8.9 120 5 kg in weight. OSL dosimeters were placed at 8 key
Mandible 0.2 26.9 120 5 head and neck organs in the 2 phantom heads to as-
Mandible 0.4 8.9 120 5 sess the radiation doses (Table I; Fig 2). Equivalent
Iluma doses to the head and neck organs were calculated
Full head 0.3 7.8 120 3.8
Maxilla-mandible 0.2 40 120 3.8
based on the 2007 recommendations of the ICRP.3
Maxilla-mandible 0.4 20 120 3.8 The i-CAT Platinum CBCT scanner (Imaging Sciences
Maxilla-mandible 0.2 7.8 120 3.8 International, Hatfield, Pa) with 5 manufacturer-
predefined exposure settings and the Iluma CBCT
scanner (IMTEC, Ardmore, Okla) with 4 predefined set-
Table III. i-CAT ratio of child/adult equivalent dose tings were used in this study. Preset pediatric settings,
only available in the Iluma scanner, were used with the
Full head Maxilla Maxilla Mandible Mandible child phantom head. For the Iluma, 2 scanning
0.3 mm 0.2 mm 0.4 mm 0.2 mm 0.4 mm
(%) (%) (%) (%) (%) techniques were used: the full-head scan with no
Brain 152 200 214 233 180 collimation and the collimated scan (collimated to
Cranium 160 295 279 286 257 the maxillary-mandibular area, excluding the eyes).
Mandible 127 325 341 94 99 Both scanning techniques had the same machine pa-
Cervical spine 30 29 34 34 32 rameters of 120 kV(p) and 3.8 mA. For the i-CAT,
Thyroid 140 130 143 118 133
Parotid 90 92 101 115 119
full-head, maxilla-only, and mandible-only scans
Average 117 170 173 145 137 were taken, all using the same machine parameters
child/adult of 120 kV(p) and 5 mA. Table II shows the parameters
for each scan. The scans were repeated by the same in-
vestigators (L.T.D., G.P., R.P., A.D.G.) 3 times (scout
accurate readings, as the OSL dosimeters are optically and volumetric acquisitions), and radiation exposure
stimulated with green light, which is more efficient than to the OSL dosimeters was read with a MicroStar
the thermally stimulated TLDs. OSL dosimeters also allow reader (Landauer, Glenwood, Ill). Specific labeled do-
for multiple exposures and readouts from 1 dosimeter, simeters were used for each site. The dosimeters
have lower depletion over time, and are more sensitive were measured before exposure for a background
than TLDs.10,11 reading, exposed, and then read again. All dosimeters
Few studies have evaluated the differences in were read 3 times, and background radiation readings
radiation dose from CBCT scans to adults compared were subtracted from the average readings to give the
with children. Data characterizations of the equivalent radiation dose to the dosimeter. Individual readings of
organ doses from different CBCT scanners under absorbed dose for each scan (scout plus volumetric
similar controlled settings are limited in the litera- acquisitions) were obtained by dividing the average
ture.5,12 The purpose of this study was to measure readings minus the background by 3. Dosimetrical
the differences in equivalent organ dose from statistical evaluations were calculated using Excel (Mi-
different scanners under similar settings in children crosoft, Redmond, Wash). Averaged radiation readings
compared with adults. We investigated the equivalent were then converted to equivalent doses using the
doses to the lens of the eye, the thyroid, and several tissue-weighting factors of the 2007 ICRP’s recom-
head and neck organs of an anthropomorphic mendations.3 The calculated child equivalent doses
adult female phantom head and an anthropomorphic were divided by the adult equivalent dose, giving a per-
child male phantom head. The data were obtained centage ratio of child over adult comparison of the
from 2 commercially available dental CBCT scanners equivalent dose. This study will help to develop a base-
using several specific exposure settings. line for other CBCT studies to follow.
MATERIAL AND METHODS RESULTS
Two phantom heads were used: an adult female Readings from the i-CAT, which did not have
anthropomorphic phantom (model 702; Computerized a pediatric preset option, showed significantly higher

June 2013  Vol 143  Issue 6 American Journal of Orthodontics and Dentofacial Orthopedics
Al Najjar et al 787

Table IV. Iluma ratio of child/adult equivalent dose


Full scan Collimated maxilla-mandible Collimated maxilla-mandible Collimated maxilla-mandible
7.8 s (%) 40 s (%) 20 s (%) 7.8 s (%)
Brain 212 329 346 413
Cranium 479 373 378 367
Mandible 85 101 102 86
Cervical spine 68 61 54 63
Thyroid 93 75 77 75
Parotid 117 97 103 132
Average child/adult 176 173 177 189

Fig 3. Ratio of child/adult average equivalent doses by organ for all scans. C Spine, Cervical spine.

Table V. i-CAT full-head 0.3-mm scan


Adult Child
Ratio of
Organ dose WT Equivalent Organ WT Equivalent equivalent dose
(cGy) SD (ICRP 103) dose (mSv) dose (cGy) SD (ICRP 103) dose (mSv) child/adult (%)
Brain 0.066 0.008 0.01 6.6 0.1 0.01 0.01 10.0 152
Cranium 0.05 0.01 0.01 5.0 0.08 0.01 0.01 8.0 160
Mandible 0.079 0.008 0.01 7.9 0.1 0.02 0.01 10.0 127
Cervical spine 0.109 0.002 0.01 10.9 0.033 0.0004 0.01 3.3 30
Thyroid 0.02142 0.00009 0.04 8.6 0.03 0.01 0.04 12.0 140
Parotid 0.11 0.02 0.01 11.0 0.0986 0.0001 0.01 9.9 90
Average child/adult 116

WT, Tissue weighting factor; cGy, centigray (absorbed dose).

equivalent doses to children compared with adults, collimation. Readings at the cervical-spine level, how-
ranging from a 117% average ratio of equivalent ever, were decreased significantly to 30% of the adult
dose increase for a 3-mm full-head scan, to a 173% equivalent dose. The mandible and the parotid also
average increase for the 0.4-mm maxilla scan showed decreases in equivalent doses to children, al-
(Table III). The child’s mandible received an equivalent though not as dramatic as those at the cervical-spine
radiation dose 341% higher than the adult dose for the level (Table III).
0.4-mm maxilla scan, the highest increase in relative On the other hand, the Iluma had a pediatric preset
effective dose. Brain and cranial tissues received equiv- that was used for the scans. Increases for average
alent doses of 152% to 295% higher than adults, equivalent dose ranged from 158% for the 7.8-second
depending on the resolution of the scan and full-head scan to 266% for the 20-second collimated

American Journal of Orthodontics and Dentofacial Orthopedics June 2013  Vol 143  Issue 6
788 Al Najjar et al

Table VI. i-CAT maxilla 0.2-mm scan


Adult Child
Ratio of
Organ dose WT Equivalent Organ WT Equivalent equivalent dose
(cGy) SD (ICRP 103) dose (mSv) dose (cGy) SD (ICRP 103) dose (mSv) child/adult (%)
Brain 0.03 0.02 0.01 3.0 0.06 0.03 0.01 6.0 200
Cranium 0.0088 0.0009 0.01 0.9 0.026 0.001 0.01 2.6 295
Mandible 0.083 0.001 0.01 8.3 0.27 0.02 0.01 27.0 325
Cervical spine 0.1074 0.0006 0.01 10.7 0.0315 0.0001 0.01 3.2 29
Thyroid 0.023 0.002 0.04 9.2 0.03 0.01 0.04 12.0 130
Parotid 0.3 0.02 0.01 30.0 0.277 0.004 0.01 27.7 92
Average child/adult 179

WT, Tissue weighting factor; cGy, centigray (absorbed dose).

Table VII. i-CAT maxilla 0.4-mm scan


Adult Child
Ratio of
Organ dose WT Equivalent Organ WT Equivalent equivalent dose
(cGy) SD (ICRP 103) dose (mSv) dose (cGy) SD (ICRP 103) dose (mSv) child/adult (%)
Brain 0.014 0.009 0.01 1.4 0.03 0.01 0.01 3.0 214
Cranium 0.0043 0.0009 0.01 0.4 0.012 0.001 0.01 1.2 279
Mandible 0.0411 0.0005 0.01 4.1 0.14 0.01 0.01 14.0 341
Cervical spine 0.0468 0.0008 0.01 4.7 0.016 0.003 0.01 1.6 34
Thyroid 0.0112 0.0006 0.04 4.5 0.016 0.007 0.04 6.4 143
Parotid 0.146 0.009 0.01 14.6 0.1469 0.0002 0.01 14.7 101
Average child/adult 185

WT, Tissue weighting factor; cGy, centigray (absorbed dose).

Table VIII. i-CAT mandible 0.2-mm scan


Adult Child
Ratio of
Organ WT Equivalent Organ WT Equivalent equivalent dose
dose (cGy) SD (ICRP103) dose (mSv) dose (cGy) SD (ICRP103) dose (mSv) child/adult (%)
Brain 0.009 0.006 0.01 0.9 0.021 0.009 0.01 2.1 233
Cranium 0.0028 0.0006 0.01 0.3 0.008 0.001 0.01 0.8 286
Mandible 0.28 0.02 0.01 28.0 0.262 0.006 0.01 26.2 94
Cervical spine 0.249 0.005 0.01 24.9 0.0856 0.0002 0.01 8.6 34
Thyroid 0.0593 0.0002 0.04 23.7 0.07 0.04 0.04 28.0 118
Parotid 0.25 0.02 0.01 25.0 0.288 0.004 0.01 28.8 115
Average child/adult 147

WT, Tissue weighting factor; cGy, centigray (absorbed dose).

maxilla-mandible scan (Table IV). Brain and cranial in equivalent doses relative to the adult readings, to as
tissues received an equivalent radiation dose increase low as 75% (Table IV).
of up to 479% in the 7.8-second full-head scan. Comparing the equivalent dose percentages between
Preliminary data on the eye dosimeters measured the the i-CAT and the Iluma showed that the Iluma’s
highest increase at 671% in equivalent dose for the pediatric settings decreased the radiation to the
20-second collimated maxilla-mandible scan. Further mandible and thyroid, with similar results at the parotid.
findings on the eye and lens dosimeters will be published However, the i-CAT had significantly lower radiation
later. Similar to the i-CAT measurements, the cervical delivery to the brain, cranium, and cervical spine
spine showed significant decreases in equivalent radia- (Tables III and IV; Fig 3). Tables V through XIII show
tion, down to 54% to 68% of the adult equivalent the equivalent radiation dose readings for adults and
dose. The mandible and thyroid also showed decreases children for each organ from the various scans.

June 2013  Vol 143  Issue 6 American Journal of Orthodontics and Dentofacial Orthopedics
Al Najjar et al 789

Table IX. i-CAT mandible 0.4-mm scan


Adult Child
Ratio of
Organ WT Equivalent Organ WT Equivalent equivalent dose
dose (cGy) SD (ICRP 103) dose (mSv) dose (cGy) SD (ICRP 103) dose (mSv) child/adult (%)
Brain 0.005 0.003 0.01 0.5 0.009 0.004 0.01 0.9 180
Cranium 0.0014 0.0003 0.01 0.1 0.0036 0.0001 0.01 0.4 257
Mandible 0.136 0.003 0.01 13.6 0.135 0.008 0.01 13.5 99
Cervical spine 0.132 0.001 0.01 13.2 0.0425 0.0001 0.01 4.3 32
Thyroid 0.03 0.001 0.04 12.0 0.04 0.02 0.04 16.0 133
Parotid 0.123 0.003 0.01 12.3 0.146 0.004 0.01 14.6 119
Average child/adult 137

WT, Tissue weighting factor; cGy, centigray (absorbed dose).

Table X. Iluma full-head 7.8-second scan


Adult Child
Ratio of
Organ WT Equivalent Organ WT Equivalent equivalent dose
dose (cGy) SD (ICRP 103) dose (mSv) dose (cGy) SD (ICRP 103) dose (mSv) child/adult (%)
Brain 0.042 0.001 0.01 4.2 0.089 0.004 0.01 8.9 212
Cranium 0.0169 0.0004 0.01 1.7 0.081 0.002 0.01 8.1 479
Mandible 0.101 0.002 0.01 10.1 0.086 0.004 0.01 8.6 85
Cervical spine 0.066 0.001 0.01 6.6 0.045 0.003 0.01 4.5 68
Thyroid 0.0204 0.0005 0.04 8.2 0.019 0.02 0.04 7.6 93
Parotid 0.097 0.002 0.01 9.7 0.1135 0.0002 0.01 11.4 117
Average child/adult 176

WT, Tissue weighting factor; cGy, centigray (absorbed dose).

Table XI. Collimated maxilla-mandible 40-second scan (Iluma)


Adult Child
Ratio of
Organ WT Equivalent Organ WT Equivalent equivalent dose
dose (cGy) SD (ICRP 103) dose (mSv) dose (cGy) SD (ICRP 103) dose (mSv) child/adult (%)
Brain 0.045 0.002 0.01 4.5 0.148 0.004 0.01 14.8 329
Cranium 0.015 0.001 0.01 1.5 0.056 0.002 0.01 5.6 373
Mandible 0.35 0.01 0.01 35.0 0.36 0.01 0.01 35.5 101
Cervical spine 0.347 0.009 0.01 34.7 0.212 0.006 0.01 21.2 61
Thyroid 0.106 0.004 0.04 42.5 0.08 0.003 0.04 32.0 75
Parotid 0.395 0.008 0.01 39.5 0.385 0.008 0.01 38.5 97
Average child/adult 173

WT, Tissue weighting factor; cGy, centigray (absorbed dose).

DISCUSSION the lens of the eye, the thyroid, and other key head
Despite the spike in the use of CBCT scans for rou- and neck organs. Because no male adult phantom
tine examinations, characterization of the radiation ex- head was available that accommodated OSL
posure from different CBCT scanners to adults and dosimeters, we were limited to using a female adult
children is still incomplete. Furthermore, previous char- phantom head. The data were obtained from 2
acterizations were done with TLDs, which may be less commercially available dental CBCT scanners using
accurate than OSL dosimeters in the lower radiation several specific exposure settings. The equivalent
dose range.10,11 Here, we aimed to assess the doses for the child phantom head for a specific scan
differences in organ doses from several scanners was divided by the equivalent dose measured for the
under similar settings and analyzed the differences in adult phantom head with the same scan, and this
equivalent doses to children compared with adults to number was multiplied by 100 to give a percentage
ratio of the child-to-adult radiation comparison.

American Journal of Orthodontics and Dentofacial Orthopedics June 2013  Vol 143  Issue 6
790 Al Najjar et al

Table XII. Collimated maxilla-mandible 20-second scan (Iluma)


Adult Child
Ratio of
Organ WT Equivalent Organ WT Equivalent equivalent dose
dose (cGy) SD (ICRP 103) dose (mSv) dose (cGy) SD (ICRP 103) dose (mSv) child/adult (%)
Brain 0.024 0.002 0.01 2.4 0.083 0.003 0.01 8.3 346
Cranium 0.008 0.001 0.01 0.8 0.031 0.002 0.01 3.1 378
Mandible 0.19 0.01 0.01 19.1 0.194 0.003 0.01 19.4 102
Cervical spine 0.201 0.006 0.01 20.1 0.108 0.006 0.01 10.8 54
Thyroid 0.054 0.004 0.04 21.5 0.042 0.002 0.04 16.6 77
Parotid 0.199 0.004 0.01 19.9 0.204 0.005 0.01 20.5 103
Average child/adult 177

WT, Tissue weighting factor; cGy, centigray (absorbed dose).

Table XIII. Collimated maxilla-mandible 7.8-second scan (Iluma)


Adult Child
Ratio of
Organ WT Equivalent Organ WT Equivalent equivalent dose
dose (cGy) SD (ICRP 103) dose (mSv) dose (cGy) SD (ICRP 103) dose (mSv) child/adult (%)
Brain 0.0081 0.0008 0.01 0.8 0.033 0.001 0.01 3.3 413
Cranium 0.0029 0.0007 0.01 0.3 0.011 0.001 0.01 1.1 367
Mandible 0.1 0.01 0.01 9.7 0.083 0.004 0.01 8.3 86
Cervical spine 0.065 0.003 0.01 6.5 0.041 0.004 0.01 4.1 63
Thyroid 0.023 0.003 0.04 9.3 0.017 0.001 0.04 7.0 75
Parotid 0.087 0.006 0.01 8.7 0.115 0.005 0.01 11.5 132
Average child/adult 189

WT, Tissue weighting factor; cGy, centigray (absorbed dose).

It is evident from the i-CAT radiation readings that (Tables V-IX). Thus, collimation when possible is an
higher equivalent doses were found in children than in important modality for reducing radiation exposure to
adults with the same exposure settings (no pediatric patients. Similarly, reducing the resolution of
setting was available on this model of i-CAT) (Table the scan reduced the absorbed dose significantly; the
III). This increase in equivalent dose was generalized absorbed dose for the maxilla 0.4-mm scan or
over most of the child’s tissues and was significant, the mandible 0.4-mm scan was about half that of the
with the exception of the cervical spine and the parotid maxilla 0.2-mm scan or the mandible 0.2-mm scan
tissues. These differences can be explained by a number (Tables VI-IX).
of factors, including the difference in the size of the The Iluma CBCT scanner had a pediatric setting,
child’s organs compared with the adult, the difference which was used for the child scans. A similar pattern of
in the position of these organs, and the difference in percentage increases of child to adult was observed,
tissue penetration and thickness, allowing more with the brain and cranium receiving significant
radiation to penetrate through the child’s tissues and increases in equivalent radiation (Table IV). The cervical
perhaps also leading to increased internal scatter spine in children received a lower equivalent dose
radiation. relative to adults because of its size and its position in-
The position of the cervical spine is lower relative to ferior to the border of the mandible. Pediatric exposure
the mandible in children than in adults and thus is out of settings lowered the child dose to the inferior head
the collimation range in children. This explains the sig- and neck organs compared with the adult dose when
nificantly lower radiation received by those tissues in using adult exposure settings (cervical spine, thyroid,
children. and mandible).
The full-head scan resulted in the lowest increase in In all cases, the mandibular scan resulted in the
the ratio of equivalent dose to children relative to adults; highest equivalent dose to the thyroid compared with
however, the absolute absorbed doses of the collimated the maxilla and full-head scans (Tables III-XIII). The
scans were significantly less than a full-head scan variability in measured doses when comparable

June 2013  Vol 143  Issue 6 American Journal of Orthodontics and Dentofacial Orthopedics
Al Najjar et al 791

settings were used with the different CBCT scanners overall radiation dose to the patient and thus is an im-
shows that different CBCT scanners produce different portant part of achieving the ALARA principle.
amounts of radiation, and that one size does not fit CBCT scans can be beneficial, but they should not be
all (Tables V-XIII). Therefore, each CBCT scanner used for routine diagnosis, screening, and
requires its own calibration readings and dosage examination and should be reserved for complex cases
measurements, and a universal statement for all CBCT on a specific case-by-case basis, such as those involv-
scanners would be inherently inaccurate. Future ing dento-osseous pathology, impacted canines, dental
directions include comparisons of adult and pediatric anomalies, cleft lip and palate patients before bone
settings of the radiation dose to children’s organ sites grafting, and patients with dentofacial deformities.
in different scanners, as well as an accurate Facilities must develop a comprehensive radiation
characterization of the difference in readouts between protection policy with regard to CBCT use in accor-
TLDs and OSL dosimeters in the range of low radiation dance with national and international recommended
doses encountered in CBCT applications. Furthermore, radiation protection principles and guidelines.
image quality assessment at various settings, to ensure
clinically diagnostic images at lower settings, would be REFERENCES
a valuable assessment and would allow clinicians to 1. Prins R, Dauer LT, Colosi DC, Quinn B, Kleiman NJ, Bohle GC,
maintain the “as low as reasonably achievable” (ALARA) et al. Significant reduction in dental cone beam computed to-
principle with CBCT scans. Evaluating the effects of mography (CBCT) eye dose through the use of leaded glasses.
body mass index, ethnicity, age, and sex would also Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2011;112:
502-7.
add an interesting dimension to the necessary knowl-
2. Committee to Assess Health Risks from Exposure to Low Levels
edge on the effects of radiation. of Ionizing Radiation; Board of Radiation Effects, Research
Although radiographic examinations are an impor- Division on Earth and Life Studies,National Research Council
tant component of dental care, they should be taken of the National Academies. Health risks from exposure to low
only after a thorough clinical examination and levels of ionizing radiation: BEIR VII, Phase 2. National
Academy of Sciences. Washington, DC: National Academies
consideration of the patient’s dental history, preferably
Press; 2006.
including a study of any previous dental radiographs.13 3. Ludlow J, Davies-Ludlow L, White S. Patient risk related to
In addition, adequate education and practical training common dental radiographic examinations. The impact of 2007
in radiologic protection should be provided for all who International Commission on Radiological Protection recommen-
work with radiation in dentistry. This is especially dations regarding dose calculation. J Am Dent Assoc 2008;139:
1237-43.
important in CBCT dental radiography. The basic
4. Ludlow J, Ivanovic M. Comparative dosimetry of dental CBCT
principles recommended by the National Council on devices and 64-slice CT for oral and maxillofacial radiology.
Radiation Protection and Measurements for radiation Oral Surg Med Oral Pathol Oral Radiol Endod 2008;106:
protection in dentistry, as well as the European CBCT 106-14.
radiation protection guidelines, are important starting 5. Roberts JA, Drage NA, Davies J, Thomas DW. Effective dose from
cone beam CT examinations in dentistry. Br J Radiol 2009;82:
points for developing a comprehensive radiation
35-40.
protection policy for CBCT use.9,12-28 6. Committee on the Biological Effects of Ionizing Radiations. Board
of Radiation Effects, Research Division on Earth and Life Studies,
CONCLUSIONS National Research Council of the National Academies. Health
Equivalent doses for children’s organs were effects of exposure to low levels of ionizing radiation: BEIR V.
National Academy of Sciences. Washington, DC: National
generally higher than for adults when similar exposure
Academies Press; 1990.
settings were used. Pediatric settings (Iluma) resulted in 7. 1990 recommendations of the International Commission on
lower equivalent doses to some child organs compared Radiological Protection. Ann ICRP 1991;21:1-201.
with those calculated for the adult phantom head 8. National Committee on Radiation Protection. Radiation protection
scanned using standard settings (i-CAT). Certain organs in dentistry. Report no. 145. Bethesda, Md: National Council on
Radiation Protection and Measurements; 2004.
received more radiation in children than adults, most
9. Brenner DJ. Effective dose: a flawed concept that could and should
likely due to the difference in their size and location. be replaced. Br J Radiol 2008;81:521-3.
Different CBCT scanners expose patients to different 10. Lian OB. Luminescence dating: optically stimulated luminescence.
amounts of radiation. Equivalent doses to most organs Encyclopedia of quaternary science. University of London, UK:
varied with the center, size, and height of the field of Royal Holloway; 2007. p. 1491-505.
11. Zacharias N, Stuhec M, Knezevic Z, Fountoukidis E, Michael CT,
view. When possible, collimation should be used to re-
Bassiakos Y. Low dose environmental dosimetry using thermo- and
duce the field of view as much as possible. Despite the optically stimulated luminescence. Nuclear instruments
reduction in the field of view, collimation increases the and methods in physics research section A. Accelerators, Spectrome-
resolution of the image obtained while reducing the ters. Detectors and Associated Equipment 2007;580:698-701.

American Journal of Orthodontics and Dentofacial Orthopedics June 2013  Vol 143  Issue 6
792 Al Najjar et al

12. Radiation protection: cone beam CT for dental and maxillofacial 21. Scarfe WC, Farman AG. What is cone-beam CT and how does it
radiology, evidence based guidelines. SEDENTEXCT Project Con- work? Dent Clin North Am 2008;62:707-30.
sortium. Version 2.0; 2011. 22. Tetradis S, Ansley P, Graff-Radford S. Cone beam computed
13. Hechler SL. Cone-beam CT: applications in orthodontics. Dent Clin tomography in the diagnosis of dental disease. Tex Dent J 2011;
North Am 2008;52:809-23. 128:620-8.
14. Nakajima A, Sameshima GT, Arai Y, Homme Y, Dougherty H Sr. Two- 23. Kaeppler G. Applications of cone beam computed tomography
and three-dimensional orthodontic imaging using limited cone in dental and oral medicine. Int J Comput Dent 2010;13:
beam-computed tomography. Angle Orthod 2005;75:895-903. 203-19.
15. Holberg C, Steinhauser S, Geis P, Rudzki-Janson I. Cone-beam 24. Dawood A, Patel S, Brown J. Cone beam CT in dental practice. Br
computed tomography in orthodontics: benefits and limitations. Dent J 2009;207:23-8.
J Orofac Orthop 2005;66:434-44. 25. Vos W, Casselman J, Swennen GRJ. Cone-beam computerized
16. Mah JK, Huang JC, Choo H. Practical applications of cone-beam tomography (CBCT) imaging of the oral and maxillofacial region:
computed tomography in orthodontics. J Am Dent Assoc 2010; a systematic review of the literature. Int J Oral Maxillofac Surg
141(Supp 3):7S-13S. 2009;38:609-25.
17. Kapila S, Conley RS, Harrell WE Jr. The current status of cone beam 26. Quereshy FA, Barnum G, Demko C, Horan M, Palomo JM, Baur DA,
computed tomography imaging in orthodontics. Dentomaxillofac et al. Use of cone beam computed tomography to volumetrically
Radiol 2011;40:24-34. assess alveolar cleft defects—preliminary results. J Oral Maxillofac
18. Merrett SJ, Drage NA, Durning P. Cone beam computed tomogra- Surg 2012;70:188-91.
phy: a useful tool in orthodontic diagnosis and treatment 27. Radiological protection of the worker in medicine and dentistry.
planning. J Orthod 2009;36:202-10. Ann ICRP 1989;20:1-83.
19. Cattaneo PM, Melsen B. The use of cone-beam computed 28. Use of cone-beam computed tomography in endodontics. Joint
tomography in an orthodontic department in between research position statement of the American Association of Endodontists
and daily clinic. World J Orthod 2008;9:269-82. and the American Academy of Oral and Maxillofacial Radiology.
20. Fanning B. CBCT—the justification process, audit and review of the Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2011;111:
recent literature. J Ir Dent Assoc 2011;57:356-61. 234-7.

June 2013  Vol 143  Issue 6 American Journal of Orthodontics and Dentofacial Orthopedics

You might also like