Professional Documents
Culture Documents
Introduction: With the advent of cone-beam computed tomography (CBCT) scans, there has been a transition
toward these scans’ replacing traditional radiographs for orthodontic diagnosis and treatment planning. Children
represent a significant proportion of orthodontic patients. Similar CBCT exposure settings are predicted to result
in higher equivalent doses to the head and neck organs in children than in adults. The purpose of this study was
to measure the difference in equivalent organ doses from different scanners under similar settings in children
compared with adults. Methods: Two phantom heads were used, representing a 33-year-old woman and
a 5-year-old boy. Optically stimulated dosimeters were placed at 8 key head and neck organs, and equivalent
doses to these organs were calculated after scanning. The manufacturers’ predefined exposure settings were
used. Results: One scanner had a pediatric preset option; the other did not. Scanning the child’s phantom
head with the adult settings resulted in significantly higher equivalent radiation doses to children compared
with adults, ranging from a 117% average ratio of equivalent dose to 341%. Readings at the cervical spine level
were decreased significantly, down to 30% of the adult equivalent dose. When the pediatric preset was used for
the scans, there was a decrease in the ratio of equivalent dose to the child mandible and thyroid. Conclusions:
CBCT scans with adult settings on both phantom heads resulted in higher radiation doses to the head and neck
organs in the child compared with the adult. In practice, this might result in excessive radiation to children
scanned with default adult settings. Collimation should be used when possible to reduce the radiation dose to
the patient. While CBCT scans offer a valuable tool, use of CBCT scans should be justified on a specific
case-by-case basis. (Am J Orthod Dentofacial Orthop 2013;143:784-92)
T
a
Resident, Department of Orthodontics and Pediatric Dentistry, School of Dental he development of cone-beam computed
Medicine, State University of New York at Stony Brook, Stony Brook, NY. tomography (CBCT) has introduced a transition
b
Clinical assistant professor and director, Division of Diagnostic Imaging,
Department of Prosthodontics and Digital Technology, State University of New toward these scans' replacing traditional
York at Stony Brook, Stony Brook, NY. radiographs taken for orthodontic diagnosis and
c
Associate attending physicist, Departments of Medical Physics and Radiology, Me- treatment planning. Since children are a significant
morial Sloan-Kettering Cancer Center, New York, NY.
d
Chief, Consequence Management Assistance Program, Defense Threat Reduction proportion of orthodontic patients, the use of the
Agency, Ft. Belvoir, Va. same CBCT exposure settings for children and adults is
e
Radiologic technologist and instructional support, School of Dental Medicine, predicted to result in higher equivalent doses to the
State University of New York at Stony Brook, Stony Brook, NY.
f
Clinical instructor, Department of Cariology and Comprehensive Care, College of head and neck organs in children than in adults.1,2
Dentistry, New York University, New York, NY. The CBCT radiation exposure to the surface tissues
g
Clinical associate professor, Department of Prosthodontics and Digital Technology, of adults has been previously measured with adult
School of Dental Medicine, State University of New York at Stony Brook, Stony
Brook, NY; clinical associate professor, Department of Cariology and Comprehen- male phantom heads using thermo-luminescent
sive care, College of Dentistry, New York University, New York, NY. dosimeters (TLDs).3-5 However, to our knowledge,
h
Program director and clinical associate professor, Department of Orthodontics and a comparative characterization of the radiation
Pediatric Dentistry, School of Dental Medicine, State University of New York at
Stony Brook, Stony Brook, NY. exposure from different CBCT scanners in children and
The authors report no commercial, proprietary, or financial interest in the adults is not available.
products or companies described in this article. Once cancer risk from x-ray radiation was confirmed
Reprint requests to: Anas Al Najjar, Department of Orthodontics and Pediatric
Dentistry, SUNY at Stony Brook, Stony Brook University School of Dental Med- early in the 20th century, the International Commission
icine, Sullivan Hall, Stony Brook, NY 11794; e-mail, anas_82@hotmail.com. on Radiological Protection (ICRP) has worked hard to
Submitted, June 2012; revised and accepted, January 2013. develop guidelines for x-ray exposure and safety proto-
0889-5406/$36.00
Copyright Ó 2013 by the American Association of Orthodontists. col.6 In the 1990s, the ICRP moved toward using
http://dx.doi.org/10.1016/j.ajodo.2013.01.013 the equivalent and effective dose (measured in
784
Al Najjar et al 785
American Journal of Orthodontics and Dentofacial Orthopedics June 2013 Vol 143 Issue 6
786 Al Najjar et al
June 2013 Vol 143 Issue 6 American Journal of Orthodontics and Dentofacial Orthopedics
Al Najjar et al 787
Fig 3. Ratio of child/adult average equivalent doses by organ for all scans. C Spine, Cervical spine.
equivalent doses to children compared with adults, collimation. Readings at the cervical-spine level, how-
ranging from a 117% average ratio of equivalent ever, were decreased significantly to 30% of the adult
dose increase for a 3-mm full-head scan, to a 173% equivalent dose. The mandible and the parotid also
average increase for the 0.4-mm maxilla scan showed decreases in equivalent doses to children, al-
(Table III). The child’s mandible received an equivalent though not as dramatic as those at the cervical-spine
radiation dose 341% higher than the adult dose for the level (Table III).
0.4-mm maxilla scan, the highest increase in relative On the other hand, the Iluma had a pediatric preset
effective dose. Brain and cranial tissues received equiv- that was used for the scans. Increases for average
alent doses of 152% to 295% higher than adults, equivalent dose ranged from 158% for the 7.8-second
depending on the resolution of the scan and full-head scan to 266% for the 20-second collimated
American Journal of Orthodontics and Dentofacial Orthopedics June 2013 Vol 143 Issue 6
788 Al Najjar et al
maxilla-mandible scan (Table IV). Brain and cranial in equivalent doses relative to the adult readings, to as
tissues received an equivalent radiation dose increase low as 75% (Table IV).
of up to 479% in the 7.8-second full-head scan. Comparing the equivalent dose percentages between
Preliminary data on the eye dosimeters measured the the i-CAT and the Iluma showed that the Iluma’s
highest increase at 671% in equivalent dose for the pediatric settings decreased the radiation to the
20-second collimated maxilla-mandible scan. Further mandible and thyroid, with similar results at the parotid.
findings on the eye and lens dosimeters will be published However, the i-CAT had significantly lower radiation
later. Similar to the i-CAT measurements, the cervical delivery to the brain, cranium, and cervical spine
spine showed significant decreases in equivalent radia- (Tables III and IV; Fig 3). Tables V through XIII show
tion, down to 54% to 68% of the adult equivalent the equivalent radiation dose readings for adults and
dose. The mandible and thyroid also showed decreases children for each organ from the various scans.
June 2013 Vol 143 Issue 6 American Journal of Orthodontics and Dentofacial Orthopedics
Al Najjar et al 789
DISCUSSION the lens of the eye, the thyroid, and other key head
Despite the spike in the use of CBCT scans for rou- and neck organs. Because no male adult phantom
tine examinations, characterization of the radiation ex- head was available that accommodated OSL
posure from different CBCT scanners to adults and dosimeters, we were limited to using a female adult
children is still incomplete. Furthermore, previous char- phantom head. The data were obtained from 2
acterizations were done with TLDs, which may be less commercially available dental CBCT scanners using
accurate than OSL dosimeters in the lower radiation several specific exposure settings. The equivalent
dose range.10,11 Here, we aimed to assess the doses for the child phantom head for a specific scan
differences in organ doses from several scanners was divided by the equivalent dose measured for the
under similar settings and analyzed the differences in adult phantom head with the same scan, and this
equivalent doses to children compared with adults to number was multiplied by 100 to give a percentage
ratio of the child-to-adult radiation comparison.
American Journal of Orthodontics and Dentofacial Orthopedics June 2013 Vol 143 Issue 6
790 Al Najjar et al
It is evident from the i-CAT radiation readings that (Tables V-IX). Thus, collimation when possible is an
higher equivalent doses were found in children than in important modality for reducing radiation exposure to
adults with the same exposure settings (no pediatric patients. Similarly, reducing the resolution of
setting was available on this model of i-CAT) (Table the scan reduced the absorbed dose significantly; the
III). This increase in equivalent dose was generalized absorbed dose for the maxilla 0.4-mm scan or
over most of the child’s tissues and was significant, the mandible 0.4-mm scan was about half that of the
with the exception of the cervical spine and the parotid maxilla 0.2-mm scan or the mandible 0.2-mm scan
tissues. These differences can be explained by a number (Tables VI-IX).
of factors, including the difference in the size of the The Iluma CBCT scanner had a pediatric setting,
child’s organs compared with the adult, the difference which was used for the child scans. A similar pattern of
in the position of these organs, and the difference in percentage increases of child to adult was observed,
tissue penetration and thickness, allowing more with the brain and cranium receiving significant
radiation to penetrate through the child’s tissues and increases in equivalent radiation (Table IV). The cervical
perhaps also leading to increased internal scatter spine in children received a lower equivalent dose
radiation. relative to adults because of its size and its position in-
The position of the cervical spine is lower relative to ferior to the border of the mandible. Pediatric exposure
the mandible in children than in adults and thus is out of settings lowered the child dose to the inferior head
the collimation range in children. This explains the sig- and neck organs compared with the adult dose when
nificantly lower radiation received by those tissues in using adult exposure settings (cervical spine, thyroid,
children. and mandible).
The full-head scan resulted in the lowest increase in In all cases, the mandibular scan resulted in the
the ratio of equivalent dose to children relative to adults; highest equivalent dose to the thyroid compared with
however, the absolute absorbed doses of the collimated the maxilla and full-head scans (Tables III-XIII). The
scans were significantly less than a full-head scan variability in measured doses when comparable
June 2013 Vol 143 Issue 6 American Journal of Orthodontics and Dentofacial Orthopedics
Al Najjar et al 791
settings were used with the different CBCT scanners overall radiation dose to the patient and thus is an im-
shows that different CBCT scanners produce different portant part of achieving the ALARA principle.
amounts of radiation, and that one size does not fit CBCT scans can be beneficial, but they should not be
all (Tables V-XIII). Therefore, each CBCT scanner used for routine diagnosis, screening, and
requires its own calibration readings and dosage examination and should be reserved for complex cases
measurements, and a universal statement for all CBCT on a specific case-by-case basis, such as those involv-
scanners would be inherently inaccurate. Future ing dento-osseous pathology, impacted canines, dental
directions include comparisons of adult and pediatric anomalies, cleft lip and palate patients before bone
settings of the radiation dose to children’s organ sites grafting, and patients with dentofacial deformities.
in different scanners, as well as an accurate Facilities must develop a comprehensive radiation
characterization of the difference in readouts between protection policy with regard to CBCT use in accor-
TLDs and OSL dosimeters in the range of low radiation dance with national and international recommended
doses encountered in CBCT applications. Furthermore, radiation protection principles and guidelines.
image quality assessment at various settings, to ensure
clinically diagnostic images at lower settings, would be REFERENCES
a valuable assessment and would allow clinicians to 1. Prins R, Dauer LT, Colosi DC, Quinn B, Kleiman NJ, Bohle GC,
maintain the “as low as reasonably achievable” (ALARA) et al. Significant reduction in dental cone beam computed to-
principle with CBCT scans. Evaluating the effects of mography (CBCT) eye dose through the use of leaded glasses.
body mass index, ethnicity, age, and sex would also Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2011;112:
502-7.
add an interesting dimension to the necessary knowl-
2. Committee to Assess Health Risks from Exposure to Low Levels
edge on the effects of radiation. of Ionizing Radiation; Board of Radiation Effects, Research
Although radiographic examinations are an impor- Division on Earth and Life Studies,National Research Council
tant component of dental care, they should be taken of the National Academies. Health risks from exposure to low
only after a thorough clinical examination and levels of ionizing radiation: BEIR VII, Phase 2. National
Academy of Sciences. Washington, DC: National Academies
consideration of the patient’s dental history, preferably
Press; 2006.
including a study of any previous dental radiographs.13 3. Ludlow J, Davies-Ludlow L, White S. Patient risk related to
In addition, adequate education and practical training common dental radiographic examinations. The impact of 2007
in radiologic protection should be provided for all who International Commission on Radiological Protection recommen-
work with radiation in dentistry. This is especially dations regarding dose calculation. J Am Dent Assoc 2008;139:
1237-43.
important in CBCT dental radiography. The basic
4. Ludlow J, Ivanovic M. Comparative dosimetry of dental CBCT
principles recommended by the National Council on devices and 64-slice CT for oral and maxillofacial radiology.
Radiation Protection and Measurements for radiation Oral Surg Med Oral Pathol Oral Radiol Endod 2008;106:
protection in dentistry, as well as the European CBCT 106-14.
radiation protection guidelines, are important starting 5. Roberts JA, Drage NA, Davies J, Thomas DW. Effective dose from
cone beam CT examinations in dentistry. Br J Radiol 2009;82:
points for developing a comprehensive radiation
35-40.
protection policy for CBCT use.9,12-28 6. Committee on the Biological Effects of Ionizing Radiations. Board
of Radiation Effects, Research Division on Earth and Life Studies,
CONCLUSIONS National Research Council of the National Academies. Health
Equivalent doses for children’s organs were effects of exposure to low levels of ionizing radiation: BEIR V.
National Academy of Sciences. Washington, DC: National
generally higher than for adults when similar exposure
Academies Press; 1990.
settings were used. Pediatric settings (Iluma) resulted in 7. 1990 recommendations of the International Commission on
lower equivalent doses to some child organs compared Radiological Protection. Ann ICRP 1991;21:1-201.
with those calculated for the adult phantom head 8. National Committee on Radiation Protection. Radiation protection
scanned using standard settings (i-CAT). Certain organs in dentistry. Report no. 145. Bethesda, Md: National Council on
Radiation Protection and Measurements; 2004.
received more radiation in children than adults, most
9. Brenner DJ. Effective dose: a flawed concept that could and should
likely due to the difference in their size and location. be replaced. Br J Radiol 2008;81:521-3.
Different CBCT scanners expose patients to different 10. Lian OB. Luminescence dating: optically stimulated luminescence.
amounts of radiation. Equivalent doses to most organs Encyclopedia of quaternary science. University of London, UK:
varied with the center, size, and height of the field of Royal Holloway; 2007. p. 1491-505.
11. Zacharias N, Stuhec M, Knezevic Z, Fountoukidis E, Michael CT,
view. When possible, collimation should be used to re-
Bassiakos Y. Low dose environmental dosimetry using thermo- and
duce the field of view as much as possible. Despite the optically stimulated luminescence. Nuclear instruments
reduction in the field of view, collimation increases the and methods in physics research section A. Accelerators, Spectrome-
resolution of the image obtained while reducing the ters. Detectors and Associated Equipment 2007;580:698-701.
American Journal of Orthodontics and Dentofacial Orthopedics June 2013 Vol 143 Issue 6
792 Al Najjar et al
12. Radiation protection: cone beam CT for dental and maxillofacial 21. Scarfe WC, Farman AG. What is cone-beam CT and how does it
radiology, evidence based guidelines. SEDENTEXCT Project Con- work? Dent Clin North Am 2008;62:707-30.
sortium. Version 2.0; 2011. 22. Tetradis S, Ansley P, Graff-Radford S. Cone beam computed
13. Hechler SL. Cone-beam CT: applications in orthodontics. Dent Clin tomography in the diagnosis of dental disease. Tex Dent J 2011;
North Am 2008;52:809-23. 128:620-8.
14. Nakajima A, Sameshima GT, Arai Y, Homme Y, Dougherty H Sr. Two- 23. Kaeppler G. Applications of cone beam computed tomography
and three-dimensional orthodontic imaging using limited cone in dental and oral medicine. Int J Comput Dent 2010;13:
beam-computed tomography. Angle Orthod 2005;75:895-903. 203-19.
15. Holberg C, Steinhauser S, Geis P, Rudzki-Janson I. Cone-beam 24. Dawood A, Patel S, Brown J. Cone beam CT in dental practice. Br
computed tomography in orthodontics: benefits and limitations. Dent J 2009;207:23-8.
J Orofac Orthop 2005;66:434-44. 25. Vos W, Casselman J, Swennen GRJ. Cone-beam computerized
16. Mah JK, Huang JC, Choo H. Practical applications of cone-beam tomography (CBCT) imaging of the oral and maxillofacial region:
computed tomography in orthodontics. J Am Dent Assoc 2010; a systematic review of the literature. Int J Oral Maxillofac Surg
141(Supp 3):7S-13S. 2009;38:609-25.
17. Kapila S, Conley RS, Harrell WE Jr. The current status of cone beam 26. Quereshy FA, Barnum G, Demko C, Horan M, Palomo JM, Baur DA,
computed tomography imaging in orthodontics. Dentomaxillofac et al. Use of cone beam computed tomography to volumetrically
Radiol 2011;40:24-34. assess alveolar cleft defects—preliminary results. J Oral Maxillofac
18. Merrett SJ, Drage NA, Durning P. Cone beam computed tomogra- Surg 2012;70:188-91.
phy: a useful tool in orthodontic diagnosis and treatment 27. Radiological protection of the worker in medicine and dentistry.
planning. J Orthod 2009;36:202-10. Ann ICRP 1989;20:1-83.
19. Cattaneo PM, Melsen B. The use of cone-beam computed 28. Use of cone-beam computed tomography in endodontics. Joint
tomography in an orthodontic department in between research position statement of the American Association of Endodontists
and daily clinic. World J Orthod 2008;9:269-82. and the American Academy of Oral and Maxillofacial Radiology.
20. Fanning B. CBCT—the justification process, audit and review of the Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2011;111:
recent literature. J Ir Dent Assoc 2011;57:356-61. 234-7.
June 2013 Vol 143 Issue 6 American Journal of Orthodontics and Dentofacial Orthopedics