You are on page 1of 5

Radiography 20 (2014) 178e182

Contents lists available at ScienceDirect

Radiography
journal homepage: www.elsevier.com/locate/radi

Gonad shielding in paediatric pelvic radiography: Effectiveness


and practice
Thomas Warlow a, Peter Walker-Birch b, Philip Cosson b, *
a
Radiology Department, King’s Mill Hospital, Mansfield Road, Sutton in Ashfield, Nottinghamshire NG17 4JL, UK
b
Medical Imaging, Teesside University, Middlesbrough TS1 3BA, UK

a r t i c l e i n f o a b s t r a c t

Article history: The use of Gonad Shields (GS) has been advocated during pelvic radiography since the 1950’s, particu-
Received 30 September 2013 larly in children where the risks from radiation are higher. Previous literature reports that GS are often
Received in revised form omitted and rarely used correctly.
6 January 2014
Objectives: Presentation of findings concerning use of GS in the context of previous data in the literature,
Accepted 16 January 2014
Available online 8 February 2014
and recommend any appropriate actions.
Method: A retrospective analysis of images from an existing DICOM Digital Teaching Library (DTL) was
conducted. Images of the pelvis from paediatric patients were reviewed and scored on whether a GS was
Keywords:
Pelvis
present and (if present) whether the shield was considered to adequately protect the gonads.
Ovary Results: 130 images were reviewed. 70 male and 60 female. The gonads were deemed to be protected by
Gonads a shield in 22 images (17%), inadequately protected when a shield was used in 44 images (34%) with the
Protective devices remaining 64 images (49%) having no shield at all. A lack of adequate protection for the gonads was
Radiation protection found, with females more likely to be inadequately protected than males (c2 ¼ 19.009, df ¼ 1, p < 0.001).
Radiography These findings become more clinically significant when reports of ovaries lying outside of the pelvic
basin (in paediatric patients) are considered.
Conclusions: The current practice of gonad shielding is neither effective nor beneficial for female pae-
diatric patients, incorrect shield placement can often require repeat exposures. This finding is
commensurate with previous literature. Therefore, gonad shielding is no longer an appropriate opti-
mization tool for female paediatric patients during conventional radiography of the pelvis, and should be
abandoned.
Ó 2014 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.

Introduction risk for children aged 0e9 is twice that compared with an adult in
their thirties.4
Projection Radiography of the pelvis is a currently indicated Radiation protection during radiography of the pelvis has always
investigation in assessing hip pathologies in paediatrics.1,2 The been portrayed as challenging due to the location of reproductive
treatment of childhood diseases such as developmental hip organs (the gonads). Irradiating the germ cells within the gonads is
dysplasia, slipped upper femoral epiphysis (SUFE) and Perthes’ reported to cause genetic mutation which has the potential to
disease can often require long term investigation and numerous develop into hereditary disease passed down and expressed in
radiographs. future generations, making the risks during pelvic radiography
The European Directive on health protection identifies children therefore two-fold, with the potential for both somatic and genetic
as a special practice that merit particular attention from the radi- mutations.5,6,2
ation protection point of view.3 Radiation induced cancer risks to The International Commission on Radiological Protection (ICRP)
children are higher than the average for the whole population, the has traditionally seen the gonads as a radiosensitive organ with the
highest tissue weighting factor of any organ.7 However, this was
recently revised down from a tissue weighting factor of 0.2 to 0.088
(below that given to Bone-marrow, Colon, Lung, Stomach, and
Breast; all of which have a tissue weighting of 0.12). This revision
* Corresponding author. Tel.: þ44 01642 384175.
E-mail addresses: Thomas.Warlow@SFH-TR.NHS.UK (T. Warlow), P.Walker- was based largely on the lack of direct evidence supporting the link
Birch@tees.ac.uk (P. Walker-Birch), p.cosson@tees.ac.uk, philip.cosson@ between gonad irradiation and heritable disease in future
shaderware.com (P. Cosson).

1078-8174/$ e see front matter Ó 2014 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
http://dx.doi.org/10.1016/j.radi.2014.01.002
T. Warlow et al. / Radiography 20 (2014) 178e182 179

generations.5,8 The high tissue weighting of the colon is certainly and females was investigated; it was predicted that GS use would
notable in this scenario as lower sections of the colon are irradiated be both more frequent and more accurate in males than females,
during a pelvis radiograph. allowing the following directional hypothesises to be postulated:
The professional responsibility of the radiographer as operator H1 e Application frequency of gonad shielding is significantly
or practitioner should ensure that they optimise the examination higher in males than females.
for each individual patient. Existing optimisation guidelines for H0 e There is no significant difference in the frequency of
radiological investigations where the gonads lie within or close to application of gonad shielding between males and females and.
the primary beam (i.e. during pelvic radiography) from the ICRP H2 e Accuracy of gonad shield placement is significantly higher
and National Radiological Protection Board (NRPB) state that good in males than females.
radiographic technique includes standard use of a gonad shield H0 e There is no significant difference in the placement accuracy
(GS) without obscuring essential diagnostic information.5,9 Many of gonad shields between males and females.
radiographic technique books also advocate this practice. However, The objective was to present findings in the context of previous
optimisation has the aim of reducing the patient risk to as low as data in the literature, and recommend any appropriate actions.
reasonably practicable (ALARP). The radiographer must consider
the practicality of accurately positioning gonad shielding against Methods
the potential reduction in risk to the patient.
A systematic literature search10 was conducted to identify pre- The study adopted a non-experimental design, and was con-
vious work in which GS placement was assessed. The results are ducted as a retrospective survey on radiographic images that were
presented in Table 1. All the studies conducted a retrospective stored in an existing DICOM digital teaching library (DTL). The DTL
analysis of radiographic images of the pelvis archived, usually in comprises 886,000 anonymised radiographic DICOM images ob-
Picture Archiving and Communication System (PACS). No data tained as a result of exposures conducted over a period of six
pertaining to images that had been rejected or repeated by the months. These were then collected from a number of UK NHS
radiographer were included. All the studies (excluding the repeat Health Care providers.17
audits of McCarty et al.11) reviewed the images after they were The project collected quantitative nominal data, allowing the
taken, protecting them from the Audience-effect. use of the Chi2 test to determine statistically significant differences
While the studies vary in size and credibility, they all demon- between groups and testing hypotheses. It was determined that 87
strate relatively low rates of adherence to gonad protection advice. images need to be analysed for statistical power in a Chi2 test with 1
Moreover, a United Kingdom (UK) wide audit on GS use has been degree of freedom. A standard alpha level of 0.05 is assumed, and
suggested.10 Kenny and Hill12 were first to recognise that adequate the population effect size is medium.
placement of a GS was less common in females than in males. They The target population and inclusion criterion for images used
identified this as statistically noteworthy finding (choosing p  0.05 was all storage phosphor radiography projections of the pelvis in
as the level of significance, a). Their finding has been replicated by paediatric patients (aged under 16 years). Images from theatre
two other studies10,2 Additionally, the complete omission of a GS cases were excluded. 130 images were available from the 259 ac-
has been found to be more common in females.14,10 cessions meeting the inclusion criteria, 70 from male patients and
Compounding the low rates of adherence is the problem of 60 from female patients. All 130 images were included in the study.
inadequately placed shields obstructing bony landmarks.2,15,16 Each of the images were reviewed using a pro-forma adapted
These obstructions can lead to repeat radiographs. The extra radi- from existing published literature.2,6,10,14,15,18 The patient’s gender
ation dose from these repeats negates any initial benefit to the was noted, the presence or absence of a GS, and whether this was
patient and is more hazardous than a single unshielded exposure. on the first or second image in a series. If present, the adequacy of
Several studies describe the practice of omitting a GS in the very the shield placement in protecting the reproductive organs was
first instance to limit this problem,2,10,14,15 with the use of a GS in all reviewed. The age of each patient was not recorded.
subsequent imaging being advised. The shield should not obscure any bony anatomy. If it did, this
was recorded. Shielding of gonads was considered adequate if the
Aims and objective area covered by the shield protects the traditionally expected
location of the reproductive organs (gonad area). In males, this was
This study aimed to contribute to the evidence on whether GS defined as covering the soft tissue margins of the scrotum inferiorly
use can be a practical optimisation tool in paediatric patients to the symphysis pubis. In females, the pelvic basin was used as
referred for pelvic radiography. Difference between GS use in males popular gonad shield designs presume the ovaries to lie within the

Table 1
Summary of audit findings from literature search.

Number Adequately Shield absent Shield present Comments


of images protected but inadequate
reviewed

Kenny and Hill (1992)14 346 29% 40% 31%


Wainwright (2000)2 200 36% 42% 22%
Liakos et al. (2001)1 61 2% * * Only provided data on % of ovaries covered
McCarty et al. (2001)13 218 27% 17% 56%
Sikand et al. (2003)6 267 23% 67% 10%
Doolan et al. (2004)15 198 0% 98% 2%
Gul et al. (2005)18 948 49% 29% 22%
Masud et al. (2008)22 100 22% * * No data to differentiate inadequacy
through omission or placement
Fawcett and Barter (2009)19 2405 18% 52% 30%
Frantzen et al. (2011)16 500 9% Female * * Results were defined into specific
34% Male groups and no summative findings
were presented
180 T. Warlow et al. / Radiography 20 (2014) 178e182

pelvic basin (true pelvis).19 Shielding 100% of the pelvic basin


without obstructing any bony anatomy would require unrealisti-
cally tight precision of shield placement. Consequently, a subjective
estimate was made and coverage was deemed adequate if 90% of
the pelvic basin was protected. Inadequate protection may be a
result of poor positioning or inadequate choice of shield size; this
information was recorded.
Descriptive statistics in the form of percentages were produced
to discuss the distribution of the three possible groups: (1) no
shield present; (2) shield present with inadequate protection of the
gonad area; and (3) shield present with adequate protection of the
gonad area.

Ethical considerations

In accordance with Section 33 of the UK Data Protection Act


(1998)20 written consent was not required from participants in this
investigation as data was obtained from the pre-existing DTL. The
DTL is a historical collection of radiology data entirely anonymized
with all staff and patient identifiable data removed. This anonym-
ization process was carried out with guidance from the National Figure 2. Pie chart showing percentage distribution and difference in shield adequacy
Information Governance Board for Health and Social Care inde- in male patients.

pendently of this study.17


The anonymous DTL ensures complete participant confidenti-
ality, to the point that not even the image assessor gathering data
had access to participant, staff or hospital details. Ethical approval
was granted to conduct the study, including the use of the database
and its comprising images in the manner described.

Results

According to this study’s criteria, the gonads were likely pro-


tected in 22 (17%) images, two female and 20 male. Shields were
present but inadequately protected the gonad area in 44 images
(34%), while the remaining 64 images (49%) had no shield at all,
these results are displayed in Fig. 1, while the difference between
male and female patients are displayed in Fig. 2 and Fig. 3.
No significant difference in frequency of GS application between
gender was found (c2 ¼ 0.293, df ¼ 1, p ¼ 0.588). The adequacy of
GS placement in protecting the gonad area was found to be higher
in males than females (c2 ¼ 19.009, df ¼ 1, p < 0.001). For all

Figure 3. Pie chart showing percentage distribution and difference in shield adequacy
in female patients.

patients the frequency of GS application was higher in the second


image of an accession than the 1st (c2 ¼ 18.692, df ¼ 1, p < 0.001).
Gonad shields (GSs) were present in a total of 51% of images
assessed. Due to the difference between frequency of GS applica-
tion in the first and second images in an accession, it is likely
protocols from the originating departments stated that gonad
shielding should be omitted in the patient’s first image.
Where a gonad shield was placed, three distinct errors were
identified: (1) the shield was too small to offer protection; (2) it was
badly positioned or; (3) it was obscuring bony anatomy (Table 2).

Table 2
Errors in shield use identified.

Male n (%) Female n (%)

Bony obstruction 26 (77) 28 (88)


Incorrect positioning 11 (32) 24 (75)
Figure 1. Pie chart showing percentage distribution of shield adequacy in both male
Too small 8 (24) 26 (81)
and female patients.
T. Warlow et al. / Radiography 20 (2014) 178e182 181

majority of patients in this sample, thus shielding poses the risk of


obscuring pathological information. Winfeld et al.23 reviewed
gonad shielding on abdominal images taken in neonatal intensive
care, and concluded that their results did not support the concern
that a shield may obscure relevant radiographic findings in this
context. However, the clinical indications for pelvic imaging are
different and similarly placed shielding may generate different
conclusions in this scenario.
Adequacy of protection has been defined in this study as
coverage of the gonad area. However, the variable position of the
ovaries has been demonstrated by Counsell et al.24 and Feather-
stone et al.25 (Fig. 5). While the majority of adult ovaries lay within
the pelvic basin,19 some were found outside the pelvic basin, such
as over the sacro-iliac joints, and these could not be protected
without obscuring bony anatomy. Indeed, Fawcett et al.19 found
approximately 13% of the ovaries in adult patients lay outside the
true pelvis.
Patient age is a particular factor affecting ovary location. In a
newborn, the ovaries are located above the true pelvis, descending
down by puberty.19,26,27 Fawcett et al.19 found in 51.4% of patients
Figure 4. Analysis of Gonad Shield errors in Females. under 7 years old this migration was yet to complete, and they did
not have both ovaries within the true pelvis. Thus, the variable
location of the ovaries raises real concerns about assuming gonad
Fig. 4 is an area proportional Venn diagram21 that demonstrates shielding even when assessed as adequate.
there was often more than one error in gonad shield use, with 19 Even when a GS is correctly placed, it is not likely to be effective
(63%) of images from female data exhibiting all three errors. The in protecting paediatric female patients. Frantzen et al.16 conducted
male data showed that all the images had more than one error, but a thorough audit and dose risk evaluation of the efficacy of gonad
with only 8 (24%) exhibiting all three. shielding. They concluded that even with appropriate shielding, the
reduction in hereditary disease was very small. Risk calculation
Discussion should reference all radiosensitive organs. During a pelvic exami-
nation sections of the colon are irradiated. Practitioners rigidly
In contrast to Fawcett and Barter10 and Kenny and Hill,14 no abiding by the recently revised ICRP tissue weighting factors may
significant difference between the application frequency of shield be unsure how to adapt their practice. Female gonad shielding may
use in male and female patients was found in this instance. become an anachronism.
Therefore the findings of this study do not support the theory that
radiographers intentionally omit shielding more often in female Limitations of the study
patients for fear of obscuring bony anatomy.
For the 32 images where a GS was applied for female patients, Employing a retrospective analysis for assessment is a limita-
only two images (6%) were likely to provide adequate protection to tion,2,15 as it does not account for images that may have been
the gonad area. In comparison, 20 out of 34 images (59%) showing rejected and repeated by the radiographer due to mal-positioning
application of GS for male patients provided adequate protection to of the GS (and obscuring bony anatomy) in the initial image.
the gonad area. This means that inadequate GS placement in this Reviewing in this fashion only allows review of images sent to PACS
study was found to be 94% in female patients and 41% in male and which were then subsequently imported into the DTL.
patients. A comparison of this finding with previous work can be
seen in Table 3. (Sikand et al.,6 Gul et al.18 and Masud et al.22 were
omitted, as they did not present enough data to determine the
individual inadequacy rates for each gender.) The inadequate
shielding rate for females is always higher than that of the
males.6,18 This finding points to a cause other than radiographer
ability.
Of the 20 images of male patients with adequate gonad pro-
tection, all had some level of bony anatomy obscured. One of the
two adequately protected female images had bony anatomy
obscured by the shield. While this is less important with patients
who have a first image unshielded, this was not the case with the

Table 3
Inadequate GS placement (when a GS is used, % of those GSs used inadequately) in
this and previous studies.

Male Female

Kenny and Hill (1992)14 44% 57%


Wainwright (2000)2 38% 57%
McCarty et al. (2001)13 63% 72%
Doolan et al. (2000)15 100% 100%
Figure 5. Schematic diagram of pelvis with positions of 128 ovaries plotted, located
This study 41% 94%
using ultrasound; reproduced with permission from Featherstone et al. (1999).24
182 T. Warlow et al. / Radiography 20 (2014) 178e182

Including images rejected and repeated by the radiographer would 5. International Commission on Radiological Protection. ICRP publication 105:
radiological protection in medicine. Ann ICRP 2007;37(6).
provide more depth to the investigation.
6. Sikand M, Stinchcombe S, Livesley PJ. Study on the use of gonadal protection
Current guidelines on GS placement, and the pro forma used in shields during paediatric pelvic X-rays. Ann Royal Coll Surg Engl 2003;85:
the image assessment and data collection for this study was based 422e5.
upon studies that located the ovaries in adult females.24,25 There- 7. International Commission on Radiological Protection. Radiological protection
and safety in medicine. ICRP publication 73. Ann ICRP 1996;26(2).
fore, it is likely that the figures for adequate protection are over- 8. Recommendations of the International Commission on Radiological Protection.
estimated given the increased variability in the paediatric patient. ICRP publication 103. Ann ICRP 2007;37(2e4).
Future research would be enhanced by analysing the outcome with 9. International Commission on Radiological Protection. ICRP publication 121:
radiological protection in paediatric diagnostic and interventional radiology.
respect to patient age; lack of this data makes it more difficult to Ann ICRP 2013;42(2).
confirm the generalizability of these findings. However, the nature 10. Fawcett SL, Barter SJ. The use of gonad shielding in paediatric hip and pelvis
of the database ensures a representative sample of UK practice. radiographs. Br J Radiol 2009;82(1):363e70.
11. National Radiological Protection Board. Documents of the NRPB: patient dose
reduction in diagnostic radiology, 1(3); 1990.
Conclusion 12. Richardson W, Wilson MC, Nishikawa J, Hayward RSA. The well-built clinical
question: a key to evidence-based decisions. ACP J Club 1995;123(A):12.
13. McCarty M, Waugh R, McCallum H, Montgomery RJ, Aszkenasy OM. Paediatric
While radiography has progressed significantly in the last few pelvic imaging: improvement in gonad shield placement by multidisciplinary
decades, the method for gonad protection has remained almost audit. Pediatr Radiol 2001;31(1):646e9.
entirely unchanged. Technical advances in imaging and protocol 14. Kenny N, Hill J. Gonad protection in young orthopaedic patients. Br Med J
1992;30:1411e3.
optimisation have resulted in potentially significant dose re-
15. Doolan A, Brennan PC, Rainford LA, Healy J. Gonad protection for the antero-
ductions for a pelvic examination while the risk burden has been posterior projection of the pelvis in diagnostic radiography in Dublin hospi-
diminished per Sievert by the lowered tissue weighting factor for tals. Radiography 2004;10:15e21.
gonads. This study and review concludes that female gonad pro- 16. Frantzen MJ, Robben S, Postma AA, Zoetelief J, Wildberger JE, Kemerink GJ.
Gonad shielding in paediatric pelvic radiography: disadvantages prevail over
tection is not a practical tool for optimization and should be benefit. Insights Imag 2011;2.
abandoned as it has the potential to generate repeat exposures. For 17. Cosson P, Willis N. Digital teaching library (DTL) development for radiography
males, the significant dose reduction and ease of locating and education. Radiography 2011;18(2):112e6.
18. Gul A, Zafar M, Maffulli N. Gonadal shields in pelvic radiographs in pediatric
protecting the testes still justifies the use of a gonad shield. Slovis patients. Bull Hosp Jt Dis 2005;63(1):13e4.
and Strauss28 believe that a greater reduction in radiation dose can 19. Fawcett SL, Gomez AC, Barter SJ, Ditchfield M, Set P. More harm than good? the
be achieved through less focus on teaching (successfully or un- anatomy of misguided shielding of the ovaries. Br J Radiol 2012;85(1016).
20. Great Britain. Data Protection Act 1998. Chapter 29. [Act of Parliament online].
successfully) the positioning of a gonad shield, and more on good London: HMSO; 1998 [cited 2013 Sept 30]. Availaible from: URL, http://www.
radiographic practice. Female gonad shielding has been a distrac- legislation.gov.uk/ukpga/1998/29.
tion. Radiographers should take up the challenge of focussing on 21. Kestler HA, Müller A, Kraus JM, Buchholz M, Gress TM, Liu H, et al. VennMaster:
area-proportional euler diagrams for functional GO analysis of microarrays.
reducing overall dose by employing tighter collimation, appro- BMC Bioinforma 2008;9:67.
priate radiographic technique to prevent repeat images, and pro- 22. Masud S, Mehra A, Clothier JC. Study on the use of gonadal shields in paediatric
tocol optimisation. pelvic X-rays. J Bone Jt Surg 2008;90(Suppl. II):338.
23. Winfeld M, Strubel N, Pinkney L, Lala S, Milla S, Babb J, et al. Relative distri-
bution of pertinent findings on portable neonatal abdominal radiographs: can
References we shield the gonads? Pediatr Radiol 2013;43:1295e302.
24. Counsell R, Bain G, Williams MV, Dixon AK. Artificial radiation menopause:
1. Liakos P, Schoenecker PL, Lyons D, Gordon JE. Evaluation of the efficacy of where are the ovaries? Clin Oncol 1996;8(1):250e3.
pelvic shielding in preadolescent girls. J Pediatr Orthop 2001;21:433e5. 25. Featherstone C, Harnett AN, Brunt AM. Ultrasound localization of the ovaries
2. Wainwright AM. Shielding reproductive organs of orthopaedic patients during for radiation-induced ovarian ablation. Clin Oncol 1999;11(1):393e7.
pelvic radiography. Ann Royal Coll Surg Engl 2000;82:318e21. 26. Occhipinti KA, Frankel SD, Hricak H. The ovary: computed tomography and
3. Teunen D. The European directive on health protection of individuals against magnetic resonance imaging. Endocr Radiol 1993;31(5):1115e32.
the dangers of ionising radiation in relation to medical exposures (97/43/ 27. Bardo DM, Black M, Schenk K, Zaritzky MF. Location of the ovaries in girls from
EURATOM). J Radiol Prot 1998;18(2):133e7. newborn to 18 years of age: reconsidering ovarian shielding. Pediatr Radiol
4. Wall BF, Haylock R, Jansen JTM, Hillier MC, Hart D, Shrimpton PC. Radiation 2009;39:253e9.
risks from medical X-ray examinations as a function of the age and the sex of the 28. Slovis TL, Strauss KJ. Commentary: gonadal shielding for neonates. Pediatr
patient. Oxford, UK: Heath Protection Agency; 2011. Radiol 2013;43(10):1265e6.

You might also like