You are on page 1of 6

Radiography 24 (2018) 328e333

Contents lists available at ScienceDirect

Radiography
journal homepage: www.elsevier.com/locate/radi

A questionnaire study of radiography educator opinions about patient


lead shielding during digital projection radiography
C. Shanley*, K. Matthews
Radiography and Diagnostic Imaging, School of Medicine, University College Dublin, Ireland

a r t i c l e i n f o a b s t r a c t

Article history: Background: In projection radiography, lead rubber shielding has long been used to protect the gonads
Received 5 September 2017 both within and outside the collimated field. However, the relative radio-sensitivity of the gonads is
Received in revised form considered lower than previously, and doses from digital projection radiography are reported as being
26 March 2018
lower than in previous eras. These factors, along with technical difficulties encountered in placing lead
Accepted 5 April 2018
shielding effectively, lead to varied opinions on the efficacy of such shielding in peer reviewed literature.
Available online 30 April 2018
This current study has investigated what is currently being taught as good practice concerning the use of
lead shielding during projection radiography.
Keywords:
Patient protection
Method: An online questionnaire was distributed to a purposive sample of 44 radiography educators
Organ shielding across 15 countries, with the aim of establishing radiography educators' opinions about patient lead
Optimisation shielding and its teaching.
Radiation protection Results: From the 27 responding educators, 57% (n ¼ 15) teach students to apply gonadal shielding across
Good practice a range of radiographic examinations; only 22% (n ¼ 6) do the same for the breast, despite respondents
Survey being aware that the breast has higher relative radio-sensitivity than the gonads. Radiation protection
was the primary reason given for using shielding. Students are generally expected to apply patient lead
shielding during assessments, although a small number of respondents report that students must justify
whether or not to apply lead shielding. Educators generally held the opinion that no matter what they
are taught, students are influenced by what they see radiographers do in clinical practice.
Conclusions: The current study has not found consensus in literature or in radiography educators'
opinions concerning the use of patient lead shielding. Findings suggest that a large scale empirical study
to establish a specific evidence base for the appropriate use of lead shielding across a range of projection
radiography examinations would be useful.
© 2018 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.

Introduction of the gonads when used within the primary beam2e4 and to
provide patient reassurance.2 However, several authors advise
In projection radiography, lead rubber aprons and shielding (all against shielding the ovaries during pelvis radiography because of
termed lead shielding in this paper) may be used both within and the risk of retake due to inaccurate positioning, anatomical variance
outside the collimated field for the purpose of patient radiation and the potential negative impact on AEC function.1,3,5e7
protection, and traditionally have most commonly been applied to Shielding of both male and female gonads outside of the colli-
protect the gonads.1 However, peer reviewed, English language mated beam is reported by several authors to convey dose savings
literature published between 1998 and 2017 reveals mixed opin- during spine and chest radiography.8e12 Dose reductions to the
ions on the use of such lead shielding. breast and ovaries are reported, particularly in scoliosis imaging of
During pelvic radiography, lead shielding of the female2 and the spine in female patients, with dose reductions of 80% to breast
male3 gonads has been reported to both reduce radiation exposure tissue reported in some cases.8,11,12 However, an empirical study
suggests that the gonadal dose levels are so low anyway that lead
shielding is not warranted.13
Several papers have a paediatric focus.1,3,4,6,14 For example,
Now practising in Our Lady's Children's Hospital Crumlin, Dublin.
neonatal gonadal shielding is advocated during portable abdomen
* Corresponding author.
E-mail address: caoimhe.shanley@ucdconnect.ie (C. Shanley). examinations because these may be serial examinations.4 In direct

https://doi.org/10.1016/j.radi.2018.04.001
1078-8174/© 2018 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
C. Shanley, K. Matthews / Radiography 24 (2018) 328e333 329

response to this paper, other author's question whether the very explanation or additional opinions to be given. The questionnaire
low doses justify using such lead shielding.14,15 Similarly in pelvis was reviewed by an academic radiography colleague before being
radiography, shielding is not advocated to be used due to risk of tested in a small pilot study.
retake of projections.1,6 A pilot study was conducted to gather feedback on any ambi-
An Australian paper3 combines the findings of a literature re- guity or lack of clarity which could be addressed before the main
view with empirical consideration of dose levels, and proposes that study.19 The pilot study also established how long the questionnaire
for a range of single examinations, gonad shielding both within and took to complete, the potential response rate, and approximately
outside collimation is ineffective at reducing the risk of genetic how long it took for the questionnaire to be returned. The ques-
effects because the doses in question are so small anyway. tionnaire was sent to two academic teaching centres with one ac-
This low consensus on the efficacy of gonad shielding exists in ademic and one clinical lecturer invited to participate from each:
an era when the relative radio-sensitivity of the gonads is consid- this constituted approximately 10% of the proposed sample.20 Pilot
ered lower than previously. Relative radio-sensitivity is a concept study participants were requested to complete the questionnaire
implicit in tissue weighting factor (Wt): any factor greater than within two weeks. A response rate of 25% (one participant) was
0 indicates higher than average radio-sensitivity, and the weighting received. Reasons for the remaining three participants who did not
factors of all the body tissues add up to 1.16 In 2007, the Interna- respond were not given. The same approach for selection of par-
tional Commission on Radiological Protection (ICRP) reduced the ticipants for the pilot study questionnaire was adopted for the main
gonadal Wt from 0.2 to 0.08,17 implying that the relative radio- study. Minor grammatical amendments were made, and a decision
sensitivity of the gonads is considered less than it was previously. was taken to follow up with reminders in the main study to
The main reason for decrease in gonadal Wt is the “reduced sig- enhance the response rate.
nificance” attached to genetic effects.17 In the same guidelines, the
ICRP state that the radio-sensitivity of the breast is higher than Survey sample
previously specified, with a Wt increasing from 0.05 to 0.12. The
main reason for the increase in Wt for the breast is the “focus on The research was directed at establishing opinions from radi-
cancer incidence in detriment calculations”.17 ography educators in centres with a reasonably comparable stan-
The Wt of different tissues permits consideration of their dard of radiography education. An overall population was not
contribution to effective dose, which is a single value figure established. Purposive sampling was followed to draw a sample of
determined by the risks of cancer, hereditary diseases or genetic educators from third level institutions that have at least a three year
mutations due to ionising radiation.17 It is relevant to these changed programme encompassing some clinical placement, and where the
ICRP weighting factors that two recent papers advocate application educators speak sufficient English to be able to complete the
of lead shielding over the female breast during lumbar spine ex- questionnaire: these were the inclusion criteria. The indicative
amination11 and during spine imaging for scoliosis.12 sample was found through a combination of internet browsing and
This context of mixed opinion and changed advice on tissue institutional associations, and the research supervisor reviewed
radio-sensitivity engendered consideration of what actually is good that the level of practice of radiography was similar though not
practice in the application of lead shielding during projection identical across all. Arising from the search and supervisor guid-
radiography. Consideration of the ALARA principle may persuade ance, 23 third level institutions in 11 different countries were
some radiographers that lead shielding should be applied no identified: nine in the United Kingdom and Ireland; seven in other
matter how small the dose saving, because any dose reduction European countries; three in Australia/New Zealand and four in the
enhances optimisation. Conversely, other radiographers may have United States/Canada. A lecturer in each institution with an
the opinion that the various reported risks of misplacement along accessible e-mail address was contacted and asked to forward the
with the low doses in projection radiography are sufficiently participant invitation to one academic lecturer and one lecturer
compelling to dispense with gonad shielding. Also, current tissue teaching in clinical practice. The potential sample of 44 educators is
weighting factors may be a stimulus for radiographers to recon- presented as indicative only, and power sampling was not applied.
sider which organs should be shielded. The initial basis for good Clearly there are limitations to the sampling process, however the
practice and application of the ALARA principle has to be what findings represent an initial attempt to establish opinions on this
radiographers learn in their pre-registration education.18 The cur- topic across an international cohort of educators.
rent research therefore set out to establish what is being taught as
good practice in the placement of patient lead shielding.
Survey distribution
Methods
For the main study an active link to the online questionnaire was
The proposed research was approved by the institutional Ethics e-mailed to participants in early December 2016. In order to pro-
Committee. This research did not receive any specific grant from mote response rate, a polite reminder was e-mailed two weeks
funding agencies in the public, commercial or not-for-profit sectors. later. The questionnaire did not deactivate for a further two weeks
after this to allow for as many respondents to respond as possible.
Questionnaire design
Analysis
The authors conducted a literature review to establish published
evidence concerning the use of lead shielding. This evidence, along Descriptive statistics were applied to establish the frequency of
with knowledge of local clinical practice, was considered in specific opinions about the use of lead shielding. Likert scale re-
developing an online questionnaire using commercial software sponses were evaluated by establishing median Likert scores. Open
(Survey Monkey ®) to investigate opinions about lead shielding, text responses were subject to simple thematic analysis to establish
what is taught about lead shielding, and what influences curricu- more common opinions or practices, although these must be
lum content. The questions required either categorical or Likert appreciated in the context of the relatively small number of
type responses, with open text options to permit further responses.
330 C. Shanley, K. Matthews / Radiography 24 (2018) 328e333

Results Reasons for lead shielding

The response rate was 61% (27 of 44 invited participants), with Radiation protection was the reason most frequently reported
24 participants providing geographic information as presented in for teaching the use of lead shielding, as shown in Fig. 2.
Table 1.
All respondents had worked as a clinical radiographer during Assessment of student use of lead shielding
their career, most (74%, n ¼ 20) had been in a teaching position for
more than five years. Most respondents (65%, n ¼ 18) were Educators reported that student use of lead shielding is assessed
teaching as an academic lecturer, with 35% (n ¼ 9) holding a by a variety of assessments and coursework including written,
clinical teaching position. As can be anticipated not all questions practical and clinical tests. Respondents explained in open text how
were answered by all respondents and therefore reported per- an assessment grade would be affected if what was taught about
centages are calculated from the number of responses to each lead shielding was not applied. Only four educators reported that
question. grades would not be affected at all. In 28 of 38 comments (re-
spondents offered multiple comments), the implication of not us-
Opinions about lead shielding ing gonadal lead shielding would be at least a grade reduction, and
eight of these comments indicated the possibility of an automatic
Radiography educators reported their agreement with a series fail of clinical assessment. Six further comments suggested a case
of opinion statements on a balanced five point Likert scale from by case consideration, with students being judged on their ability to
strongly agree (scoring 5) through neutral (scoring 3) to strongly justify whether to apply lead shielding or not.
disagree (scoring 1). Median Likert Scores (MLS) across all re-
spondents are presented in Fig. 1. Opinions compared with what is taught
In open text responses, 12 educators reported their opinion on
the use of lead shielding had changed over the course of their What educators teach about the use of lead shielding (Table 2)
career. The questionnaire design was flawed in that educators were was compared with the opinions they reported about lead shield-
not required to specify are they more or less likely to advise use of ing (Fig. 1) and also with their length of experience using Pearson's
lead shielding currently. However, four respondents specifically Chi-square analysis. No statistically significant associations were
proposed that improvements in collimation and digital detectors found, with p values for the various analyses ranging from 0.22 to
reduce the validity of using lead shielding. 0.81.

Placement of lead shielding Discussion

For five broad categories of examination, educators reported Arising from mixed opinions in peer-reviewed literature con-
whether they would teach students to shield the gonads or breast cerning the application of lead shielding during projection radi-
for reasons of any of protection, image quality or patient reassur- ography, this small survey was undertaken to establish what is
ance. Not all respondents answered all questions, the responses being taught in this regard in undergraduate programmes. Results
given are summarised in Table 2. are reported from 27 staff in 23 institutions across 11 countries. The
Generally the educators that did report teaching the place- mixed opinions on the efficacy of lead shielding that are reported in
ment of lead shielding did so across all patient types. However, literature are apparent in similar mixed opinions and approaches
for each group of examinations a small number of respondents among educators. Similarly as seen in other literature, organs
(always three or fewer) reported they would teach the use of radiographers state need to be protected might not be performed in
gonadal lead shielding mainly for female, paediatric and/or practice.21 The authors make no claim that the current findings are
pregnant patients. Depending on the examination type, between necessarily representative of all opinions and approaches, rather
12 and 35% of respondents would not protect the gonads at all. that they may be sufficiently indicative to raise questions for
When considering the breast, the proportion not teaching lead investigation in a larger scale study.
shielding at all rose to between 59 and 71% depending on ex- Reference to Fig. 1 indicates educators strongly agree (MLS ¼ 5)
amination type. However, 40% of respondents would teach breast that lead shielding should be used over organs with higher radio-
shielding for female patients during examinations of the sensitivity. However, educators opinions are neutral (MLS ¼ 3) on
abdomen, spine and pelvis. whether lead shielding is more important for the gonads
(Wt ¼ 0.08) than the female breast (Wt ¼ 0.12), and in disagree-
ment that lead shielding is more important for the colon
Table 1 (Wt ¼ 0.12) than the ovaries (Wt ¼ 0.08).
Geographical data concerning participants. These comparisons are relevant if the aim of using lead shielding
Country Number of educators Number of educators is to offer protection, which in the current study is indeed the
currently working here who trained here primary reason that use of shielding is taught (Fig. 2). This being the
United Kingdom 8 10
case, it is logical that such shielding should be applied to organs
Canada 3 3 that are more radio-sensitive. However in the current study, edu-
Finland 2 2 cators reported teaching the use of gonadal shielding over and
Netherlands 2 2 above shielding of other tissues with higher tissue weighting factor.
Slovenia 2 2
For example, during upper extremity examinations, 63% (n ¼ 17) of
Australia 1 1
Belgium 1 1 educators advocate for gonadal shielding whereas 56% (n ¼ 15) of
Denmark 1 1 educators teach that breast shielding is unnecessary. The authors
Ireland 1 0 acknowledge that in all examinations, elements such as the direc-
Italy 1 1 tion of the beam and position of the patient must be considered.
New Zealand 1 0
However, all other elements optimised, the point is simplistically
C. Shanley, K. Matthews / Radiography 24 (2018) 328e333 331

Figure 1. Median Likert Scores as indicators of educator opinions on lead shielding during projection radiography.

Table 2
Sites and reasons for applying lead shielding in different examination categories.

Examination Response frequency Reason

Over the gonads Over the breast

Upper Extremity 15 8 Radiation Protection


Lower Extremity 17 8
Skull/Cervical spine/Shoulder 13 9
Chest/Thoracic spine 17 5
Abdomen/Pelvis/Lumbar spine 19 11
Upper Extremity 0 0 Image Quality
Lower Extremity 0 0
Skull/Cervical spine/Shoulder 0 0
Chest/Thoracic spine 0 0
Abdomen/Pelvis/Lumbar spine 1 0
Upper Extremity 5 1 Patient reassurance
Lower Extremity 4 0
Skull/Cervical spine/Shoulder 2 0
Chest/Thoracic spine 1 0
Abdomen/Pelvis/Lumbar spine 2 0
Upper Extremity 4 10 Not necessary
Lower Extremity 3 9
Skull/Cervical spine/Shoulder 7 8
Chest/Thoracic spine 5 10
Abdomen/Pelvis/Lumbar spine 1 6
332 C. Shanley, K. Matthews / Radiography 24 (2018) 328e333

Figure 2. Reasons for placing lead shielding across all examination types.

made that if lead shielding is going to be used, it should be used Conclusions


over organs that have higher than average radio-sensitivity.
This brings the consideration to whether lead shielding outside The current study is not definitive, but has raised several points
of the beam offers any meaningful dose reduction to the shielded for consideration.
organ, an issue which has been questioned by some authors.13e15 There is mixed literature evidence on whether to adopt the use
Educators had mixed opinions on the dose saving when lead of patient lead shielding both within and outside the primary beam
shielding is outside of the primary beam, with the wide spread of during projection radiography.
response resulting in a neutral MLS ¼ 3. This ambivalence reflects There are some apparent contradictions between radiography
the mixed opinions that are apparent in literature, and in educators' opinions regarding organ radio-sensitivity and their
conjunction with the reported teaching about shielding, suggests teaching, and gonadal lead shielding for all patients across a range
that educators are advising students to protect the gonads when of examinations seems to be the predominant practice. The breast
sometimes no substantial dose reductions are perceived. No-one is acknowledged as a radiosensitive organ, but is not so widely
could dispute that to err on the side of caution is probably good protected with lead shielding as are the gonads. Students are
practice in the context of optimisation, yet not one of the re- generally expected to apply gonadal shielding in clinical and writ-
spondents disagreed that their teaching about lead shielding is ten assessments, although a small number of educators expected
based on literature evidence. Given the uncertainty in literature their students to carefully justify whether or not to use lead
about the protective efficacy of lead shielding, a large scale shielding.
empirical study to establish unequivocal evidence for the use or Findings suggest the need for a large scale empirical study to
otherwise of lead shielding seems to be needed. establish a well-founded and specific evidence base for the
The authors were interested to understand how educator's appropriate use of lead shielding across a range of projection
opinions on lead shielding related to student assessment. In open radiography examinations. Such evidence would greatly assist both
text responses, 28 of 38 separate statements suggested that lead academic and clinical radiographers in establishing and promul-
shielding should always be applied during assessment: several of gating good practice in the use of patient lead shielding.
these comments were received from educators who agreed that
minimal dose reduction would accrue. Contradictions between
taught theory and applied assessment criteria are not conducive to Conflict of interest
either deep learning or eventual good clinical practice. Again in
light of the low literature consensus, it seems a better approach to There is no conflict of interest.
assessment would be to assess whether students can justify the use
or non-use of lead shielding on an individual patient basis. This
References
approach was reported in only six of 38 comments received.
The current study focussed on what is being taught about lead 1. Doolan A, Brennan P, Rainford L, Healy J. Gonad protection for the antero-
shielding because what is being taught will inevitably promulgate posterior projection of the pelvis in diagnostic radiography in Dublin hospi-
tals. Radiography 2004;10(1):15e21.
into practice.22 It is noteworthy then, that educators strongly
2. Goodman RT, Amurao M. Medical imaging radiation safety for the female pa-
agreed (MLS ¼ 5) that students will be influenced by what radi- tient: rationale and implementation. Radiographics 2012;32:1829e37.
ographers do in practice irrespective of what they are taught. This 3. Karami V, Zabihzadeh M, Shams N, Malehi SA. Gonad shielding during pelvic
opinion highlights the importance of all staff, academic and clinical, radiography: a systematic review and meta-analysis. Arch Iran Med 2017;20(2):
112e23.
taking on the promulgation of good practice, and adds weight to the 4. Winfeld M, Strubel N, Pinkney L, Lala S, Milla S, Babb J, et al. Relative distri-
need for a study on lead shielding identifying what good practice bution of pertinent findings on portable neonatal abdominal radiographs: can
should be. we shield the gonads? Paediatr Radiol 2013;43(10):1295e302.
C. Shanley, K. Matthews / Radiography 24 (2018) 328e333 333

5. Wainwright MA. Shielding reproductive organs of orthopaedic patients during 14. Slovis LT, Strauss JK. Gonadal shielding for neonates. Pediatr Radiol 2013;43:
pelvic radiography. Roy Coll Surg Engl 2000;82:318e21. 1265e6.
6. Frantzen JM, Robben S, Postma AA, Zoetelief J, Wildberger EJ, et al. Gonad 15. Bowen AS, Clarke R, Dickson SS, McLennan MC, Stonehouse KM, Wilcher FT,
shielding in paediatric pelvic radiography: disadvantages prevail over benefit. et al. The use of gonadal shielding in singular common diagnostic radiographic
Insight Into Imag 2012;3(1):23e32. procedures. Radiographer 2007;54(2):13e7.
7. Warlow T, Birch-Walker P, Cosson P. Gonad shielding in paediatric pelvic 16. International Commission of Radiological Protection. ‘The 2007 recommenda-
radiography: effectiveness and practice. Radiography 2014;20:178e82. tions of the international commission on radiological protection’, ICRP Publi-
8. Palmer HS, Starritt CH, Patterson M. Radiation protection of the ovaries in cation 103. Ann ICRP 2007;37(2e4).
young scoliosis patients. Eur Spine J 1998;7(4):278e81. 17. Wrixon DA. ‘New ICRP recommendations’. J Radiol Prot 2008;28:161e8.
9. Jackson G, Brennan PC. Radio-protective aprons during radiological examina- 18. CORU. Radiographers registration board code of professional conduct and
tions of the thorax: an optimum strategy. Radiat Protect Dosim 2006;121(4): Ethics. Regul Health Soc Prof Care 2016:1e14.
391e4. 19. Bryman A. Social research methods. 2nd ed. Oxford: Oxford University Press;
10. Clancy LC, Reilly OG, Brennan CP, McEntee FM. The effect of patient shield 2004.
position on gonad dose during lumbar spine radiography. Radiography 20. Hertzog MA. Considerations in determining sample size for pilot studies. Res
2010;16:131e5. Nurs Health 2008;31(2):180e91.
11. Mekis N, Zontar D, Skrk D. The effect of breast shielding during lumbar spine 21. Hayre CM, Blackman S, Carlton K, Eyden A. Attitudes and perceptions of radi-
radiography. Radiol Oncol 2013;47(1):26e31. ographers applying lead (Pb) protection in general radiography: an ethnographic
12. Presciutti MS, Karukanda K, Lee M. Management decisions for adolescent study. 2017.
idiopathic scoliosis significantly affect patient radiation exposure. Spine J 22. Doyle J. An investigation into knowledge of radiation dose as an influential factor
2014;14(9):1984e90. in collimation practice. Unpublished undergraduate thesis: BSc Radiography.
13. Hashimoto M, Kato H, Fujibuchi T, Ochi S, Morita F. Gonad protective effect of University College Dublin; 2016.
radiation protective apron in chest radiography. Nihon Hoshasen Gakkai Zasshi
2004;60(12):1704e12.

You might also like