10th Edition
Yokogawa Electric Corporation
CONTENTS
Contents
i IM 01E20D01-01E
CONTENTS
6. OUTLINE................................................................................................................. 6-1
ii IM 01E20D01-01E
1. INTRODUCTION
1. INTRODUCTION
This instrument has been adjusted at the factory before • If the customer or any third party is harmed by the
shipment. use of this product, Yokogawa assumes no responsi-
bility for any such harm owing to any defects in the
To ensure correct use of the instrument, please read
product which were not predictable, or for any
this manual thoroughly and fully understand how to
indirect damages.
operate the instrument before operating it.
NOTE
NOTE
This manual describes the hardware configura- For details of the AXFA11G magnetic flowmeter
tion of integral flowmeter and remote flowtube of converter, see the IM01E20C01-01E instruction
the AXF magnetic flowmeters. manual. For details on the AXFA14G/C magnetic
flowmeter converter, see the IM01E20C02-01E
For details of the “basic operating procedures”, instruction manual.
“parameter description”, “operation via BRAIN
terminal (BT200)”, “operation via HART commu-
Safety and Modification Precautions
nicator”, and “actual operation” for the AXF
integral flowmeter, see the user’s manual of the • The following general safety precautions must be
AXF Integral Flowmeter [Software Edition] (IM observed during all phases of operation, service, and
01E20C02-01E). repair of this instrument. Failure to comply with
For FOUNDATION Fieldbus protocol (Converter these precautions or with specific WARNINGS
Output Signal and Communication suffix code; - given elsewhere in this manual violates safety
F), please refer to IM 01E20F02-01E. standards of design, manufacture, and intended use
For PROFIBUS PA protocol (Converter Output of the instrument. Yokogawa assumes no liability for
Signal and Communication suffix code; -G), the customer's failure to comply with these require-
please refer to IM 01E20F12-01E. ments. If this instrument is used in a manner not
specified in this manual, the protection provided by
Regarding This User’s Manual this instrument may be impaired.
• This manual should be provided to the end user. • Yokogawa will not be liable for malfunctions or
• Before use, read this manual thoroughly to compre- damage resulting from any modification made to this
hend its contents. instrument by the customer.
• The contents of this manual may be changed • The following safety symbol marks are used in this
without prior notice. user's manual and instrument.
• All rights are reserved. No part of this manual may
be reproduced in any form without Yokogawa's WARNING
written permission.
• Yokogawa makes no warranty of any kind with A WARNING sign denotes a hazard. It calls
regard to this material, including, but not limited to, attention to procedure, practice, condition or the
implied warranties of merchantability and suitability like, which, if not correctly performed or adhered
for a particular purpose. to, could result in injury or death of personnel.
• All reasonable effort has been made to ensure the
accuracy of the contents of this manual. However,
if any errors or omissions are found, please inform
Yokogawa. CAUTION
• Yokogawa assumes no responsibilities for this
A CAUTION sign denotes a hazard. It calls
product except as stated in the warranty.
attention to procedure, practice, condition or the
• Please note that this user's manual may not be
like, which, if not correctly performed or adhered
revised for any specification changes, construction
to, could result in damage to or destruction of
changes or operating part changes that are not
part or all of the product.
considered to affect function or performance.
1-1 IM 01E20D01-01E
1. INTRODUCTION
(2) Wiring
IMPORTANT
• The wiring of the magnetic flowmeter must be
An IMPORTANT sign denotes that attention is performed by expert engineer or skilled person-
required to avoid damage to the instrument or nel. No operator shall be permitted to perform
system failure. procedures relating to wiring.
• When connecting the wiring, check that the
supply voltage is within the range of the voltage
specified for this instrument before connecting
NOTE the power cable. In addition, check that no
voltage is applied to the power cable before
A NOTE sign denotes information necessary for connecting the wiring.
essential understanding of operation and fea- • The protective grounding must be connected
tures. securely at the terminal with the mark to
avoid danger to personnel.
Protective grounding terminal (3) Operation
Functional grounding terminal • When opening the cover, wait for more than 10
(This terminal should not be used as a protective minutes after turning off the power. Only
grounding terminal.) expert engineer or skilled personnel are permit-
ted to open the cover.
Alternating current • Be sure to set parameters as “Protect” on the
Direct current write protect function after finish of parameter
setting work.
Under extremely rare case, the infra-red
1.1 Using the Magnetic switches may respond unexpectedly in such
conditions as sticking ball of water or extrane-
Flowmeter Safely ous substances on the surface of display panel
glass according to the principle of infra-red
WARNING switch operation.
Its probability rises in such cases as sticking
(1) Installation
rain water by storm or other similar situation
• Installation of the magnetic flowmeter must be
and washing up work near flowmeter installa-
performed by expert engineer or skilled person-
tion place.
nel. No operator shall be permitted to perform
Either to illuminate or stop illuminating the infra-
procedures relating to installation.
red switches by the flashlight may cause the
• The magnetic flowmeter is a heavy instrument.
mis-reaction.
Be careful that no damage is caused to person-
Refer to Chapter 6 “Menu P: Parameter Protec-
nel through accidentally dropping it, or by
tion Items” and section “10.2.2” in the manual
exerting excessive force on the magnetic
IM 01E20C02-01E how to use the write protect
flowmeter. When moving the magnetic flowme-
function in detail.
ter, always use a trolley and have at least two
(4) Maintenance
people carry it.
• Maintenance of the magnetic flowmeter should
• When the magnetic flowmeter is processing hot
be performed by the trained personnel having
fluids, the instrument itself may become ex-
knowledge of safety standard. No operator
tremely hot. Take sufficient care not to get
shall be permitted to perform any operations
burnt.
relating to maintenance.
• Where the fluid being processed is a toxic
• When opening the cover, wait for more than 10
substance, avoid contact with the fluid and
minutes after turning off the power.
avoid inhaling any residual gas, even after the
• Always conform to maintenance procedures
instrument has been taken off the piping line for
outlined in this manual. If necessary, contact
maintenance and so forth.
Yokogawa.
• Do not apply excessive weight, for example, a
person stepping on the magnetic flowmeter.
• All procedures relating to installation must
comply with the electrical code of the country
where it is used.
1-2 IM 01E20D01-01E
1. INTRODUCTION
• Care should be taken to prevent the build up of The guarantee will not apply in the following
dirt, dust or other substances on the display cases:
panel glass or data plate. If these surfaces do • Damage due to negligence or insufficient mainte-
get dirty, wipe them clean with a soft dry cloth. nance on the part of the customer.
(5) Explosion Protected Type Instrument • Problems or damage resulting from handling,
• Magnetic flowmeters with the model name operation or storage that violates the intended use
AXFC are products which have been and specifications.
certified as explosion proof type instruments. • Problems that result from using or performing
Strict limitations are applied to the structures, maintenance on the instrument in a location that
installation locations, external wiring work, does not comply with the installation location
maintenance and repairs, etc. of these instru- specified by Yokogawa.
ments. Sufficient care must be taken, as any • Problems or damage resulting from repairs or
violation of the limitations may cause danger- modifications not performed by Yokogawa or
ous situations. someone authorized by Yokogawa.
Be sure to read Chapter 8 “EXPLOSION • Problems or damage resulting from inappropriate
PROTECTED TYPE INSTRUMENT” before reinstallation after delivery.
handling the instruments. The description in • Problems or damage resulting from disasters such as
Chapter 8 is prior to the other description in this fires, earthquakes, storms, floods, or lightning strikes
user’s manual. and external causes.
For TIIS flameproof type instruments, be sure Trademarks:
to read “INSTALLATION AND OPERATING ADMAG, AXF and ADMAG AXF are registered
PRECAUTIONS FOR TIIS FLAMEPROOF trademarks of Yokogawa Electric Corporation.
EQUIPMENT” at the end of this manual. Company names and product names used in this
• Only trained persons use this instrument in the material are registered trademarks or trademarks of
industrial location. their respective owners.
• The protective grounding must be connected
to a suitable IS grounding system.
• Take care not to generate mechanical spark
1.3 Combination Remote
when access to the instrument and peripheral Converters
devices in hazardous locations.
(6) European Pressure Equipment Directive
(PED) IMPORTANT
• When using the instrument in compliance with
• The AXF remote flowtube (size 2.5 (0.1 in.) to
PED, be sure to read Chapter 7 before use.
400 mm (16 in.)) should be used in combina-
tion with one of the following converters:
- AXFA11 remote converter
1.2 Warranty - AXFA14 remote converter
Contact Yokogawa before using it in combina-
• The terms of this instrument that are guaranteed are tion with converters other than those listed
described in the quotation. We will make any repairs above.
that may become necessary during the guaranteed • The AXF remote flowtube (size 500 (20 in.) to
term free of charge. 2600 mm (104 in.)) should be used in combina-
• Please contact our sales office if this instrument tion with AXFA11 remote converter.
requires repair. Contact Yokogawa before using it in combina-
• If the instrument is faulty, contact us with concrete tion with converters other than AXFA11.
details about the problem and the length of time it • In case of TIIS Flameproof type, a remote
has been faulty, and state the model and serial flowtube is available for combined use with the
number. We would appreciate the inclusion of AXFA14 remote converter only.
drawings or additional information. • If the converter combined with the AXF mag-
• The results of our examination will determine netic flowmeter’s remote flowtube is changed
whether the meter will be repaired free of charge or from the AXFA11 to AXFA14 or vice versa, the
on an at-cost basis. meter factor of the remote flowtube must be
readjusted according to its flow calibration.
1-3 IM 01E20D01-01E
1. INTRODUCTION
1-4 IM 01E20D01-01E
1. INTRODUCTION
SK PL
CZ
SLO
LT H
BG
LV
EST RO
1-5 IM 01E20D01-01E
2. HANDLING PRECAUTIONS
2. HANDLING PRECAUTIONS
This instrument has been inspected carefully at the *1)
0038
factory before shipment. When the instrument is
delivered, visually check that no damage has occurred
during transportation.
Read this section carefully as it contains important
information on handling this instrument. Refer to the
relevant sections for information not contained in this
section. If you have any problems or questions, please
contact Yokogawa sales office.
2.2 Accessories
Check that the parts shown below are included in the
package:
Remote Flowtube(size 2.5 to 1000 mm(0.1 to 40 in.)):
Centering device (wafer type only): 1 pc.
Hexagonal wrenche: 2 pcs. (one each of 1.5 mm
and 3 mm nominal sizes)
Integral Flowmeter:
Centering device (wafer type only): 1 pc.
Hexagonal wrenche: 2 pcs. (one each of 1.5 mm
and 3 mm nominal sizes)
F0201.EPS
2-1 IM 01E20D01-01E
2. HANDLING PRECAUTIONS
2-2 IM 01E20D01-01E
3. INSTALLATION
3. INSTALLATION
3.1 Piping Design Pre- Based on JIS B7554 "Electromagnetic Flowmeters"
and our piping condition test data, we recommend the
cautions piping conditions as shown in the following figures.
When installing two or more magnetic flowmeters on a
WARNING single pipe, provide a run of at least 5D between them.
Installation of the magnetic flowmeter must be D: Flowtube Size
Gate valve Reducer Expander
performed by expert engineer or skilled person- fully open pipe pipe
nel. No operator shall be permitted to perform
procedures relating to installation. 2D
5D or more or more 0 is allowable. 0 is allowable. 10D or more 2D or more
Tee 90-degree bent
Various valves
IMPORTANT
2D
5D or more 0 is allowable. 5D or more 0 is allowable.10D or more or more
Design piping correctly, referring to the following F08.EPS
to prevent damage to flowtubes and to assure Figure 3.1.1 Required Lengths of Straight Runs
accurate measuring.
*1: Do not install anything in the vicinity that may
interfere with the magnetic field, induced
NOTE signal voltages, or flow velocity distributions of
the flowmeter.
This chapter describes the remote flowtube as
an example. The same attention must be paid to *2: A straight run may not be required on the
the integral flowmeter. downstream side of the flowmeter. However, if
a downstream valve or other fitting causes
(1) Location irregularity or deviation in flows, provide a
straight run of 2D to 3D on the downstream
IMPORTANT side.
Install the flowmeter in a location where it is not *3: The valves shall be mounted on the down-
exposed to direct sunlight. The minimum ambi- stream side so that deviated flows do not
occur in the flowtube and to avoid startup from
ent temperature is limited by the minimum fluid
temperature of the flowtube (the lining). For an empty condition.
more information, refer to Chapter 6 “OUTLINE”.
The flowmeter may be used in an ambient (4) Maintaining Stable Fluid Conductivity
humidity where the relative humidity ranges from
0 to 100%. However, avoid long-term continuous
IMPORTANT
operation at relative humidity above 95%.
Do not install the flowmeter where fluid conduc-
(2) Noise Avoidance tivity tends to become uneven. If chemicals are
fed near the upstream side of a magnetic
IMPORTANT flowmeter, they may affect the flow rate’s indica-
tions. To avoid this situation, it is recommended
The flowmeter should be installed away from that the chemical feed ports be located on the
electrical motors, transformers, and other power downstream side of the flowmeter. If it is un-
sources in order to avoid interference with avoidable that chemicals must be fed on the
measurement. upstream side, provide a sufficient length of
straight run (approximately 50D) to ensure the
(3) Required Lengths of Straight Runs proper mixture of fluids.
To maintain accurate measurement, see JIS B7554
“Electro Magnetic Flowmeters” which explains the
requirements for upstream piping conditions of
magnetic flowmeters.
3-1 IM 01E20D01-01E
3. INSTALLATION
IMPORTANT
F0302.EPS
It is essential that pipes remain fully filled at all
Figure 3.1.2 Chemical Injection times, otherwise flow rate indications may be
affected and measurement errors may be
(5) Precautions for Use of Liquid Sealing caused.
Compounds
Block valve In cases where fluids contain air bubbles, piping must
F0303.EPS
be designed to prevent them from accumulating in the
Figure 3.1.3 Bypass Line measurement pipe of a flowtube.
(8) Supporting the Flowmeter If a valve exists near the flowmeter, try to mount the
flowmeter on the valve’s upstream side in order to
CAUTION
prevent a possible reduction of pressure inside the pipe,
thereby avoiding the possibility of air bubbles.
Do not secure the flowmeter separately to
prevent the vibrations, shocks, and expansion (Correct) (Incorrect)
(Correct)
and contraction forces of the piping from affect-
(Incorrect)
ing it. Fix the pipes first, then support the flow- Valve
meter with the pipes. With extra small-sized F10.EPS
flowmeters (2.5-10 mm), in particular, fix the Figure 3.1.5 Avoiding Air Bubbles
flowmeter in parallel with the piping on a mount-
ing base.
3-2 IM 01E20D01-01E
3. INSTALLATION
• Mounting orientation
CAUTION
Correct
Incorrect Incorrect
Air bubble
Water can
Electrode Precipitate seep into
Electrode
the terminal
box.
F0306.EPS
WARNING CAUTION
The magnetic flowmeter is a heavy instrument. Care should be taken not to drop the flowmeter
Be careful that no damage is caused to person- or expose it to excessive shock. In particular, be
nel through accidentally dropping it, or by careful not to subject the flange surface to
exerting excessive force on the magnetic flow- shock. This may lead to liner damage which will
meter. When moving the magnetic flowmeter, result in inaccurate readings.
always use a trolley and have at least two
people carry it.
(3) Flange Protection Covers
IMPORTANT
NOTE
Keep the protective covering (i.e. the corrugated
This chapter describes the remote flowtube as
cardboard or other cushioning material) in place
an example. The same attention must be paid to
over the flange except when mounting the
the integral flowmeter.
flowmeter to the pipe.
(4) Terminal Box Cover (2) Inside a newly installed pipeline, there may be
some foreign substances such as residue from
IMPORTANT welding or wood chips. Remove them by flushing
the piping before mounting the flowmeter. This
As it is possible that the insulation will deterio-
prevents the lining from being damaged, as well as
rate, do not open the terminal box cover until it
the occurrence of erroneous measured signals
is time to wire it.
resulting from foreign substances passing through
the flowtube during measurement.
(5) Long-term Non-use
3-4 IM 01E20D01-01E
3. INSTALLATION
0
A B C -0.1 IMPORTANT
CAUTION
+0.3 0
0 -0.1
A B C D Ceramics pipes may be damaged if the nuts are
tightened when the center axes are not properly
aligned.
4 10(0.39)
(0.16) 35(1.38)
(1) Mounting Direction Tighten the union joint nuts according to the
Mount the flowmeter so that the flow direction of the tightening torque values in Table 3.3.2. For
fluid to be measured is in line with the direction of the permeable fluid (such as nitric acid, hydrofluoric
arrow mark on the flowmeter. acid or sodium hydrate at high temperature),
tighten the nuts according to the torque values in
Table 3.3.3.
IMPORTANT
As the gasket material is fluorocarbon PTFE, it is
If it is impossible to match the direction of the possible that the nuts may loosen as time
arrow mark, the direction of the electrical con- passes. Retighten the nuts if this is the case. Be
nection can be changed. Refer to Section 5.1 to sure to use the gasket (thickness is 1.5 mm)
do this properly. which comes with the flowmeter.
In case the fluid being measured flows against
the arrow direction, refer to the parameter J20:
Flow Direction in the user’s manual of the
AXFA11 Magnetic Flowmeter Remote Converter
(IM 01E20C01-01E) or the AXFA14 Magnetic
Flowmeter Remote Converter/AXF Integral
Flowmeter [Software Edition] (IM 01E20C02-
01E).
3-5 IM 01E20D01-01E
3. INSTALLATION
Connecting fitting
Union joint nut
*Piping 3.3.2 Nominal Diameter 2.5 mm (0.1
in.) to 40 mm (1.5 in.), Wafer
Type
IMPORTANT
3-6 IM 01E20D01-01E
3. INSTALLATION
(3) Positioning the Flowmeter For permeable fluids (such as nitric acid, hydrofluoric
Pass two through-bolts through the adjacent holes of acid, or sodium hydrate at high temperatures), tighten
both flanges and position the flowmeter so that the the nuts according to the torque values in Table 3.3.6.
Mini-flanges and the centering devices come in close
contact with each other. Pass the other through-bolts
through the other holes (refer to Figures 3.3.2 and CAUTION
3.3.3). In case stud-type through-bolts are used,
position them in such a way that the centering devices For a flowmeter with fluorocarbon PFA lining, it
come in contact with the bolt threads. is possible that the nuts may loosen as time
passes, so tighten them regularly. Be sure to
tighten the nuts according to the prescribed
(4) Tightening Nuts torque values. Tighten them diagonally with the
Tighten the nuts according to the torque values for same torque values, step by step up to the
metal piping in Table 3.3.4. For PVC piping, select an prescribed torque value.
option code of /GA, /GC, or /GD, use rubber gaskets
and tighten the nuts to the torque values for PVC
piping in Table 3.3.5.
Mini-flange
Piping-side flange
F0310a.EPS
Horizontal mounting
F0310b.EPS Vertical mounting
F0310c.EPS
Figure 3.3.2 Mounting Procedure for Wafer Type (size: 2.5 (0.1) to 15 mm (0.5 in.))
3-7 IM 01E20D01-01E
3. INSTALLATION
Mini-flange
Vertical mounting
F0311c.EPS
Figure 3.3.3 Mounting Procedure for Wafer Type (size: 25 (1.0), 32 (1.25), and 40 mm (1.5 in.))
Table 3.3.4 Wafer Type Tightening Torque Values for Metal Piping
Tightening torque values for PFA/Polyurethane Rubber lining type (N-m / {kgf-cm} / [in-lbf])
Gasket types
No gasket (standard)
within flowtube
Gasket types
Non-asbestos fiber gasket, PTFE-sheathed non-asbestos gasket (optional codes BCF and BSF), or the equivalent in hardness
for user’s flange
Flange ratings
Size JIS 10K, ANSI Class 150, and DIN PN10 JIS 20K, ANSI Class 300, and DIN PN16 DIN PN40
mm (inch)
2.5 (0.1) 7.2 to 8.4 / {73.42 to 85.66} / [63.72 to 74.35] 7.3 to 8.4 / {74.44 to 85.66} / [64.61 to 74.35] 7.6 to 8.4 / {77.5 to 85.66} / [67.26 to 74.35]
5 (0.2) 7.2 to 8.4 / {73.42 to 85.66} / [63.72 to 74.35] 7.3 to 8.4 / {74.44 to 85.66} / [64.61 to 74.35] 7.6 to 8.4 / {77.5 to 85.66} / [67.26 to 74.35]
10 (0.4) 7.2 to 8.4 / {73.42 to 85.66} / [63.72 to 74.35] 7.3 to 8.4 / {74.44 to 85.66} / [64.61 to 74.35] 7.6 to 8.4 / {77.5 to 85.66} / [67.26 to 74.35]
15 (0.5) 7.2 to 8.4 / {73.42 to 85.66} / [63.72 to 74.35] 7.3 to 8.4 / {74.44 to 85.66} / [64.61 to 74.35] 7.6 to 8.4 / {77.5 to 85.66} / [67.26 to 74.35]
25 (1.0) 23.5 to 27.3 / {239.6 to 278.4} / [208 to 241.6] 23.7 to 27.3 / {241.7 to 278.4} / [209.8 to 241.6] 22.3 to 27.3 / {227.4 to 278.4} / [197.4 to 241.6]
32 (1.25) 26.2 to 30.5 / {267.2 to 311} / [231.9 to 269.9] 26.6 to 30.5 / {271.2 to 311} / [235.4 to 269.9] 28.0 to 30.5 / {285.5 to 311} / [247.8 to 269.9]
40 (1.5) 36.2 to 42.4 / {369.1 to 432.4} / [320.4 to 375.3] 36.9 to 42.4 / {376.3 to 432.4} / [326.6 to 375.3] 39.1 to 42.4 / {398.7 to 432.4} / [346.1 to 375.3]
Tightening torque values for Ceramics lining type (N-m / {kgf-cm} / [in-lbf])
Gasket types
Fluororesin with ceramic fillers (Valqua #7020) (standard) gasket, or fluororesin with carbon gasket (optional code GF)
within flowtube
Gasket types
Non-asbestos gasket, PTFE-sheathed non-asbestos gasket (optional codes BCF and BSF), or the equivalent in hardness
for user’s flange
Flange ratings
Size JIS 10K, ANSI Class 150, and DIN PN10 JIS 20K, ANSI Class 300, and DIN PN16 DIN PN40
mm (inch)
15 (0.5) 2.9 to 4.8 / {29.57 to 48.95} / [25.67 to 42.48] 2.9 to 4.8 / {29.57 to 48.95} / [25.67 to 42.48] 3.0 to 5.0 / {30.59 to 50.99} / [26.55 to 44.25]
25 (1.0) 8.2 to 13.6 / {83.62 to 138.7} / [72.57 to 120.4] 8.2 to 13.7 / {83.62 to 139.7} / [72.57 to 121.3] 7.9 to 13.1 / {80.56 to 133.6} / [69.92 to 115.9]
40 (1.5) 14.1 to 23.6 / {143.8 to 240.7} / [124.8 to 208.9] 14.4 to 24.1 / {146.8 to 245.8} / [127.4 to 213.3] 15.5 to 25.8 / {158.1 to 263.1} / [137.2 to 228.3]
T0305.EPS
3-8 IM 01E20D01-01E
3. INSTALLATION
Table 3.3.5 Wafer Type Tightening Torque Values for PVC Piping
Tightening torque values for PFA lining type (N-m / {kgf-cm} / [in-lbf])
Gasket types
Fluororubber gasket (optional codes GA, GC, and GD)
within flowtube
Gasket types
Fluororubber gasket, chloroprene rubber gasket (optional codes BSC and BCC), or the equivalent in hardness
for user’s flange
Flange ratings
Size JIS 10K, ANSI Class 150, and DIN PN10 JIS 20K, ANSI Class 300, and DIN PN16 DIN PN40
mm (inch)
2.5 (0.1) 1.5 to 2.5 / {15.3 to 25.49} / [13.28 to 22.13] 1.5 to 2.5 / {15.3 to 25.49} / [13.28 to 22.13] 1.5 to 2.4 / {15.3 to 24.47} / [13.28 to 21.24]
5 (0.2) 1.5 to 2.5 / {15.3 to 25.49} / [13.28 to 22.13] 1.5 to 2.5 / {15.3 to 25.49} / [13.28 to 22.13] 1.5 to 2.4 / {15.3 to 24.47} / [13.28 to 21.24]
10 (0.4) 1.5 to 2.5 / {15.3 to 25.49} / [13.28 to 22.13] 1.5 to 2.5 / {15.3 to 25.49} / [13.28 to 22.13] 1.5 to 2.4 / {15.3 to 24.47} / [13.28 to 21.24]
15 (0.5) 1.5 to 2.5 / {15.3 to 25.49} / [13.28 to 22.13] 1.5 to 2.5 / {15.3 to 25.49} / [13.28 to 22.13] 1.5 to 2.4 / {15.3 to 24.47} / [13.28 to 21.24]
25 (1.0) 4.9 to 8.1 / {49.97 to 82.6} / [43.37 to 71.69] 5.0 to 8.3 / {50.99 to84.64 } / [44.25 to 73.46] 4.3 to 7.2 / {43.85 to 73.42} / [38.06 to 63.72]
32 (1.25) 5.5 to 9.2 / {56.08 to 93.81} / [48.68 to 81.43] 5.7 to 9.5 / {58.12 to 96.87} / [50.45 to 84.08] 5.4 to 8.9 / {55.06 to 90.75} / [47.79 to 78.77]
40 (1.5) 7.7 to 12.9 / {78.52 to 131.5} / [68.15 to 114.2] 8.1 to 13.4 / {82.6 to 136.6} / [71.69 to 118.6] 7.5 to 12.5 / {76.48 to 127.5} / [66.38 to 110.6]
Tightening torque values for Ceramics lining type (N-m / {kgf-cm} / [in-lbf])
Gasket types
within flowtube Fluororubber gasket (optional codes GA, GC, and GD)
Gasket types
Fluororubber gasket, chloroprene rubber gasket (optional codes BSC and BCC), or the equivalent in hardness
for user’s flange
Flange ratings
Size JIS 10K, ANSI Class 150, and DIN PN10 JIS 20K, ANSI Class 300, and DIN PN16 DIN PN40
mm (inch)
15 (0.5) 0.6 to 1.0 / {6.118 to 10.2} / [5.31 to 8.85] 0.6 to 1.0 / {6.118 to 10.2} / [5.31 to 8.85] 0.6 to 1.0 / {6.118 to 10.2} / [5.31 to 8.85]
25 (1.0) 1.7 to 2.8 / {17.34 to 28.55} / [15.05 to 24.78] 1.7 to 2.8 / {17.34 to 28.55} / [15.05 to 24.78] 1.5 to 2.5 / {15.3 to 25.49} / [13.28 to 22.13]
40 (1.5) 3.0 to 5.0 / {30.59 to 50.99} / [26.55 to 44.25] 3.1 to 5.2 / {31.61 to 53.03} / [27.44 to 46.02] 2.9 to 4.8 / {29.57 to 48.95} / [25.67 to 42.48]
T0306.EPS
Table 3.3.6 Wafer Type Tightening Torque Values for Metal Piping and Permeable Fluids
Tightening torque values for PFA lining type (N-m / {kgf-cm} / [in-lbf])
Gasket types
No gasket (standard)
within flowtube
Gasket types
PTFE-sheathed non-asbestos gasket (optional codes BCF and BSF), or the equivalent in hardness
for user’s flange
Flange ratings
Size JIS 10K, ANSI Class 150, and DIN PN10 JIS 20K, ANSI Class 300, and DIN PN16 DIN PN40
mm (inch)
2.5 (0.1) 10.8 to 12.4 / {110.1 to 126.4} / [95.59 to 109.7] 10.8 to 12.4 / {110.1 to 126.4} / [95.59 to 109.7] 11.1 to 12.4 / {113.2 to 126.4} / [98.24 to 109.7]
5 (0.2) 10.8 to 12.4 / {110.1 to 126.4} / [95.59 to 109.7] 10.8 to 12.4 / {110.1 to 126.4} / [95.59 to 109.7] 11.1 to 12.4 / {113.2 to 126.4} / [98.24 to 109.7]
10 (0.4) 10.8 to 12.4 / {110.1 to 126.4} / [95.59 to 109.7] 10.8 to 12.4 / {110.1 to 126.4} / [95.59 to 109.7] 11.1 to 12.4 / {113.2 to 126.4} / [98.24 to 109.7]
15 (0.5) 10.8 to 12.4 / {110.1 to 126.4} / [95.59 to 109.7] 10.8 to 12.4 / {110.1 to 126.4} / [95.59 to 109.7] 11.1 to 12.4 / {113.2 to 126.4} / [98.24 to 109.7]
25 (1.0) 34.9 to 40.1 / {355.9 to 408.9} / [308.9 to 354.9] 35.2 to 40.1 / {358.9 to 408.9} / [311.5 to 354.9] 32.3 to 37.1 / {329.4 to 378.3} / [285.9 to 328.4]
32 (1.25) 38.8 to 44.6 / {395.6 to 454.8} / [343.4 to 394.7] 39.2 to 44.6 / {399.7 to 454.8} / [346.9 to 394.7] 40.6 to 46.7 / {414.0 to 476.2} / [359.3 to 413.3]
40 (1.5) 53.5 to 61.5 / {545.5 to 627.1} / [473.5 to 544.3] 54.2 to 61.5 / {552.7 to 627.1} / [479.7 to 544.3] 56.4 to 61.5 / {575.1 to 627.1} / [499.2 to 544.3]
Tightening torque values for Ceramics lining type (N-m / {kgf-cm} / [in-lbf])
Gasket types
Fluororesin with ceramic fillers (Valqua #7020) gasket (standard), or fluororesin with carbon gasket (optional code GF)
within flowtube
Gasket types
PTFE-sheathed non-asbestos gasket (optional codes BCF and BSF), or the equivalent in hardness
for user’s flange
Flange ratings
Size JIS 10K, ANSI Class 150, and DIN PN10 JIS 20K, ANSI Class 300, and DIN PN16 DIN PN40
mm (inch)
15 (0.5) 4.2 to 7.1 / {42.83 to 72.4} / [37.17 to 62.84] 4.3 to 7.1 / {43.85 to 72.4} / [38.06 to 62.84] 4.4 to 7.3 / {44.87 to 74.44} / [38.94 to 64.61]
25 (1.0) 12.1 to 20.2 / {123.4 to 206.0} / [107.1 to 178.8] 12.2 to 20.3 / {124.4 to 207.0} / [108.0 to 179.7] 11.3 to 18.9 / {115.2 to 192.7} / [100.0 to 167.3]
40 (1.5) 20.8 to 34.7 / {212.1 to 353.8} / [184.1 to 307.1] 21.1 to 35.2 / {215.2 to 358.9} / [186.7 to 311.5] 22.2 to 37.0 / {226.4 to 377.3} / [196.5 to 327.5]
T0307.EPS
3-9 IM 01E20D01-01E
3. INSTALLATION
3-10 IM 01E20D01-01E
3. INSTALLATION
*Nut
*Through-bolt Piping-side flange
*Gasket *: These items can be ordered optionally.
(two units) If they are provided by the user, choose nuts and
bolts in compliance with the flange ratings.
Housing
Mini-
flange
Horizontal mounting
F0312c.EPS
Centering device
(four units)
F0312a.EPS
Vertical mounting
F0312b.EPS
Figure 3.3.4 Mounting Procedure for Wafer Type (size: 50 mm (2 in.) to 300 mm (12 in.))
Table 3.3.7 Wafer Type Tightening Torque Values for Metal Piping
N-m
Tightening torque values for PFA/Polyurethane Rubber/Natural Soft Rubber/EPDM Rubber lining type Unit: {kgf-cm}
[in-lbf]
Gasket types
No gasket (standard)
within flowtube
Gasket types Non-asbestos fiber gasket, PTFE-sheathed non-asbestos gasket
for user’s flange (optional codes BCF and BSF), or the equivalent in hardness
Flange ratings JIS F12
Size JIS 10K ANSI Class 150 DIN PN10 JIS20K ANSI Class 300 DIN PN16 DIN PN40
mm (inch) (JIS 75M)
45.0 to 56.8 45.0 to 56.8 22.5 to 25.9 22.5 to 25.9 50.0 to 57.5
50 (2.0) {458.9 to 579.2} {458.9 to 579.2} — {229.4 to 264.1} {229.4 to 264.1} — {509.9 to 586.3} —
[398.3 to 502.7] [398.3 to 502.7] [199.1 to 229.2] [199.1 to 229.2] [442.5 to 508.9}
61.3 to 70.5 61.3 to 70.5 30.8 to 35.4 30.8 to 35.4 56.1 to 70.8
65 (2.5) {625.1 to 718.9} {625.1 to 718.9} — {314.1 to 361.0} {314.1 to 361.0} {572.1 to 722.0} — —
[542.5 to 624.0] [542.5 to 624.0] [272.6 to 313.3] [272.6 to 313.3] [496.5 to 626.6]
35.0 to 40.3 76.0 to 80.9 39.9 to 45.9 39.9 to 45.9 39.9 to 45.9 68.4 to 78.7
80 (3.0) {356.9 to 410.9} {775.0 to 825.0} — {406.9 to 468.1} {406.9 to 468.1} {406.9 to 468.1} — {697.5 to 802.5}
[309.8 to 356.7] [672.6 to 716.0] [353.1 to 406.2] [353.1 to 406.2] [353.1 to 406.2] [605.4 to 696.5]
46.1 to 53 46.1 to 53 52.9 to 60.8 52.9 to 60.8 52.9 to 60.8 88.6 to 101.9
100 (4.0) {470.1 to 540.5} {470.1 to 540.5} — {539.4 to 620.0} {539.4 to 620.0} {539.4 to 620.0} — {903.5 to 1039}
[408.0 to 469.1] [408.0 to 469.1] [468.2 to 538.1] [468.2 to 538.1] [468.2 to 538.1] [784.1 to 901.9]
73.7 to 84.8 73.7 to 84.8 80.5 to 92.6 80.5 to 92.6 80.5 to 92.6 75.1 to 86.4
125 (5.0) {751.5 to 864.7} {751.5 to 864.7} — {820.9 to 944.3} {820.9 to 944.3} {820.9 to 944.3} — {765.8 to 881.0}
[652.3 to 750.5] [652.3 to 750.5] [712.5 to 819.5] [712.5 to 819.5] [712.5 to 819.5] [664.7 to 764.7]
85.4 to 98.2 85.4 to 98.2 61.0 to 70.2 61.0 to 70.2 91.2 to 96.3 86.3 to 99.2
150 (6.0) {870.8 to 1001} {870.8 to 1001} — {622.0 to 715.8} {622.0 to 715.8} {930.0 to 982.0} — {880.0 to 1012}
[755.8 to 869.1] [755.8 to 869.1] [539.9 to 621.3] [539.9 to 621.3] [807.2 to 852.3] [763.8 to 878.0]
78.8 to 90.6 113.6 to 135.8 113.6 to 135.8 87.5 to 100.6 87.5 to 100.6 87.5 to 100.6 88.6 to 101.9
200 (8.0) {803.5 to 923.9} {1158 to 1385} {1158 to 1385} {892.3 to 1026} {892.3 to 1026} {892.3 to 1026} — {903.5 to 1039}
[697.4 to 801.8] [1005 to 1202] {1005 to 1202] [774.4 to 890.3] [774.4 to 890.3] [774.4 to 890.3] [784.1 to 901.9]
119.4 to 137.3 119.4 to 137.3 119.4 to 137.3 158.1 to 181.8
250 (10) {1218 to 1400} {1218 to 1400} {1218 to 1400} — — — — {1612 to 1854}
[1057 to 1215] [1057 to 1215] {1057 to 1215] [1399 to 1609]
83.0 to 99.2 105.2 to 121.0 105.2 to 121.0 146.6 to 168.6
300 (12) {846.4 to 1012} {1073 to 1234} {1073 to 1234} — — — — {1495 to 1719}
[734.6 to 878] [931.1 to 1071] {931.1 to 1071] [1297 to 1492]
N-m
Tightening torque values for Ceramics lining type Unit: {kgf-cm}
[in-lbf]
Gasket types Fluororesin with ceramic fillers (Valqua #7020) gasket (standard), or fluororesin with carbon gasket (optional code GF)
within flowtube
Gasket types Non-asbestos gasket, PTFE-sheathed non-asbestos gasket (optional codes BCF and BSF), or the equivalent in hardness
for user’s flange
Flange ratings JIS F12
Size JIS 10K ANSI Class 150 DIN PN10 JIS20K ANSI Class 300 DIN PN16 DIN PN40
mm (inch) (JIS 75M)
29.9 to 49.8 29.9 to 49.8 30.5 to 50.9 30.5 to 50.9 32.7 to 54.5
50 (2.0) {304.9 to 507.8} {304.9 to 507.8} — {311.0 to 519.0} {311.0 to 519.0} — {333.4 to 555.7} —
[264.6 to 440.8] [264.6 to 440.8] [269.9 to 450.5] [269.9 to 450.5] [289.4 to 482.4]
37.1 to 61.8 37.1 to 61.8 37.6 to 62.7 37.6 to 62.7 37.6 to 62.7 56.3 to 93.8
80 (3.0) {378.3 to 630.2} {378.3 to 630.2} — {383.4 to 639.4} {383.4 to 639.4} {383.4 to 639.4} — {574.1 to 956.5}
[328.4 to 547.0] [328.4 to 547.0] [332.8 to 554.9] [332.8 to 554.9] [332.8 to 554.9] [498.3 to 830.2]
48.9 to 81.5 48.9 to 81.5 49.9 to 83.1 49.9 to 83.1 49.9 to 83.1 74.2 to 123.7
100 (4.0) {498.6 to 831.1} {498.6 to 831.1} — {508.8 to 847.4} {508.8 to 847.4} {508.8 to 847.4} — {756.6 to 1261}
[432.8 to 721.3] [432.8 to 721.3] [441.6 to 735.5] [441.6 to 735.5] [441.6 to 735.5] [656.7 to 1095]
101.4 to 169.0 101.4 to 169.0 104.4 to 174.0 104.4 to 174.0 104.4 to 174.0 82.2 to 137.0
150 (6.0) {1034 to 1723} {1034 to 1723} — {1065 to 1774} {1065 to 1774} {1065 to 1774} — {838.2 to 1397}
[897.5 to 1496] [897.5 to 1496] [924.0 to 1540] [924.0 to 1540] [924.0 to 1540] [727.5 to 1213]
142.3 to 237.2 142.3 to 237.2 142.3 to 237.2 98.5 to 164.2 98.5 to 164.2 98.5 to 164.2 86.7 to 144.6
200 (8.0) {1451 to 2419} {1451 to 2419} {1451 to 2419} {1004 to 1674} {1004 to 1674} {1004 to 1674} — {884.1 to 1475}
[1259 to2099] [1259 to2099] [1259 to2099] [871.8 to 1453] [871.8 to 1453] [871.8 to 1453] [767.3 to 1280]
T0308.EPS
3-11 IM 01E20D01-01E
3. INSTALLATION
Table 3.3.8 Wafer Type Tightening Torque Values for PVC Piping
N-m
Tightening torque values for PFA lining type Unit: {kgf-cm}
[in-lbf]
Gasket types Fluororubber gasket (optional codes GA, GC, and GD)
within flowtube
Gasket types Fluororubber gasket, chloroprene rubber gasket (optional codes BSC and BCC), or the equivalent in hardness
for user’s flange
Flange ratings JIS F12
Size JIS 10K ANSI Class 150 DIN PN10 JIS20K ANSI Class 300 DIN PN16 DIN PN40
mm (inch) (JIS 75M)
9.9 to 16.5 9.9 to 16.5 10.6 to 17.6 10.6 to 17.6 9.5 to 15.9
50 (2.0) {101.0 to 168.3} {101.0 to 168.3} — {108.1 to 179.5} {108.1 to 179.5} — {96.9 to 162.1} —
[87.6 to 146.0] [87.6 to 146.0] [93.8 to 155.8] [93.8 to 155.8] [84.1 to 140.7]
14.2 to 23.7 14.2 to 23.7 15.5 to 25.9 15.5 to 25.9 28.2 to 51.8
65 (2.5) {144.8 to 241.7} {144.8 to 241.7} — {158.1 to 264.1} {158.1 to 264.1} {287.6 to 528.2} — —
[125.7 to 209.8] [125.7 to 209.8] [137.2 to 229.2] [137.2 to 229.2] [249.6 to 458.4]
8.0 to 13.3 17.4 to 26.7 9.7 to 16.1 9.7 to 16.1 9.7 to 16.1 15.4 to 25.6
80 (3.0) {81.6 to 135.6} {177.4 to 272.3} — {98.9 to 164.2} {98.9 to 164.2} {98.9 to 164.2} — {157.0 to 261.0}
[70.8 to 117.7] [154.0 to 236.3] [85.8 to 142.5] [85.8 to 142.5] [85.8 to 142.5] [136.3 to 226.6]
11.3 to 18.8 11.3 to 18.8 14.2 to 23.6 14.2 to 23.6 14.2 to 23.6 21.1 to 35.1
100 (4.0) {115.2 to 191.7} {115.2 to 191.7} — {144.8 to 240.7} {144.8 to 240.7} {144.8 to 240.7} — {215.2 to 357.9}
[100.0 to 166.4] [100.0 to 166.4] [125.7 to 208.9] [125.7 to 208.9] [125.7 to 208.9] [186.7 to 310.6]
18.8 to 31.3 18.8 to 31.3 22.3 to 37.2 22.3 to 37.2 22.3 to 37.2 18.5 to 30.8
125 (5.0) {191.7 to 319.2} {191.7 to 319.2} — {227.4 to 379.3} {227.4 to 379.3} {227.4 to 379.3} — {188.6 to 314.1}
[166.4 to 277.0] [166.4 to 277.0] [197.4 to 329.2] [197.4 to 329.2] [197.4 to 329.2] [163.7 to 272.6]
22.5 to 37.6 22.5 to 37.6 27.2 to 45.3 27.2 to 45.3 40.7 to 62.1 21.8 to 36.3
150 (6.0) {229.4 to 383.4} {229.4 to 383.4} — {277.4 to 461.9} {277.4 to 461.9} {415.0 to 633.2} — {222.3 to 370.2}
[199.1 to 332.8] [199.1 to 332.8] [240.7 to 400.9] [240.7 to 400.9] [360.2 to 549.6] [192.9 to 321.3]
22.1 to 36.9 31.9 to 55.3 31.9 to 55.3 27.3 to 45.3 27.3 to 45.3 27.3 to 45.3 23.8 to 39.6
200 (8.0) {225.4 to 376.3} {325.3 to 563.9} {325.3 to 563.9} {278.4 to 461.9} {278.4 to 461.9} {278.4 to 461.9} — {242.7 to 403.8}
[195.6 to 326.6] [282.3 to 489.4] [282.3 to 489.4] [241.6 to 400.9] [241.6 to 400.9] [241.6 to 400.9] [210.6 to 350.5]
N-m
Tightening torque values for Ceramics lining type Unit: {kgf-cm}
[in-lbf]
Gasket types Fluororubber gasket (optional codes GA, GC, and GD)
within flowtube
Gasket types Fluororubber gasket, chloroprene rubber gasket (optional codes BSC and BCC), or the equivalent in hardness
for user’s flange
Flange ratings JIS F12
Size JIS 10K ANSI Class 150 DIN PN10 JIS20K ANSI Class 300 DIN PN16 DIN PN40
mm (inch) (JIS 75M)
4.5 to 7.4 4.5 to 7.4 4.8 to 7.9 4.8 to 7.9 4.3 to 7.1
50 (2.0) {45.89 to 75.46} {45.89 to 75.46} — {48.95 to 80.56} {48.95 to 80.56} — {43.85 to 72.4} —
[39.83 to 65.49] [39.83 to 65.49] [42.48 to 69.92] [42.48 to 69.92] [38.06 to 62.84]
4.4 to 7.3 4.4 to 7.3 4.8 to 7.9 4.8 to 7.9 4.8 to 7.9 8.6 to 14.4
80 (3.0) {44.87 to 74.44} {44.87 to 74.44} — {48.95 to 80.56} {48.95 to 80.56} {48.95 to 80.56} — {87.7 to 146.8}
[38.94 to 64.61] [38.94 to 64.61] [42.48 to 69.92] [42.48 to 69.92] [42.48 to 69.92] [76.12 to 127.4]
6.4 to 10.7 6.4 to 10.7 7.2 to 11.9 7.2 to 11.9 7.2 to 11.9 12.2 to 20.3
100 (4.0) {65.26 to 109.1} {65.26 to 109.1} — {73.42 to 121.3} {73.42 to 121.3} {73.42 to 121.3} — {124.4 to 207.0}
[56.64 to 94.7] [56.64 to 94.7] [63.72 to 105.3] [63.72 to 105.3] [63.72 to 105.3] [108.0 to 179.7]
15.1 to 25.2 15.1 to 25.2 17.6 to 29.3 17.6 to 29.3 17.6 to 29.3 14.9 to 24.8
150 (6.0) {154.0 to 257.0} {154.0 to 257.0} — {179.5 to 298.8} {179.5 to 298.8} {179.5 to 298.8} — {151.9 to 252.9}
[133.6 to 223.0] [133.6 to 223.0] [155.8 to 259.3] [155.8 to 259.3] [155.8 to 259.3] [131.9 to 219.5]
23.4 to 39.0 23.4 to 39.0 23.4 to 39.0 18.6 to 31.6 18.6 to 31.6 18.6 to 31.6 17.1 to 28.6
200 (8.0) {238.6 to 397.7} {238.6 to 397.7} {238.6 to 397.7} {189.7 to 322.2} {189.7 to 322.2} {189.7 to 322.2} — {174.4 to 291.6}
[207.1 to 345.2] [207.1 to 345.2] [207.1 to 345.2] [164.6 to 279.7] [164.6 to 279.7] [164.6 to 279.7] [151.3 to 253.1]
T0309.EPS
3-12 IM 01E20D01-01E
3. INSTALLATION
Table 3.3.9 Wafer Type Tightening Torque Values for Metal Piping and Permeable Fluids
N-m
Tightening torque values for PFA lining type Unit: {kgf-cm}
[in-lbf]
Gasket types
No gasket (standard)
within flowtube
Gasket types
PTFE-sheathed non-asbestos gasket (optional codes BCF and BSF), or the equivalent in hardness
for user’s flange
Flange ratings JIS F12
Size JIS 10K ANSI Class 150 DIN PN10 JIS20K ANSI Class 300 DIN PN16 DIN PN40
mm (inch) (JIS 75M)
66.2 to 76.1 66.2 to 76.1 33.1 to 38.0 33.1 to 38.0 71.2 to 118.6
50 (2.0) {675.1 to 776.0} {675.1 to 776.0} — {337.5 to 387.5} {337.5 to 387.5} — {726.0 to 1209} —
[585.9 to 673.5] [585.9 to 673.5] [292.9 to 336.3] [292.9 to 336.3] [630.1 to 1050]
89.5 to 102.9 89.5 to 102.9 44.9 to 51.6 44.9 to 51.6 81.8 to 103.2
65 (2.5) {912.6 to 1049} {912.6 to 1049} — {457.9 to 526.2} {457.9 to 526.2} {834.1 to 1052} — —
[792.1 to 910.7] [792.1 to 910.7] [397.4 to 456.7] [397.4 to 456.7] {724.0 to 913.4]
51.3 to 59.0 111.3 to 118.4 58.1 to 66.8 58.1 to 66.8 58.1 to 66.8 100.8 to 115.9
80 (3.0) {523.1 to 601.6} {1135 to 1207} — {592.5 to 681.2} {592.5 to 681.2} {592.5 to 681.2} — {1028 to 1182}
[454.0 to 522.2] [985.0 to 1048] [514.2 to 591.2] [514.2 to 591.2] {514.2 to 591.2] [892.1 to 1026]
66.7 to 76.7 66.7 to 76.7 76.1 to 87.5 76.1 to 87.5 76.1 to 87.5 129.8 to 149.3
100 (4.0) {680.2 to 782.1} {680.2 to 782.1} — {776.0 to 892.3} {776.0 to 892.3} {776.0 to 892.3} — {1324 to 1522}
[590.3 to 678.8] [590.3 to 678.8] [673.5 to 774.4] [673.5 to 774.4] {673.5 to 774.4] [1149 to 1321]
106.1 to 122.0 106.1 to 122.0 114.5 to 131.7 114.5 to 131.7 114.5 to 131.7 109.6 to 126.0
125 (5.0) {1082 to 1244} {1082 to 1244} — {1168 to 1343} {1168 to 1343} {1168 to 1343} — {1118 to 1285}
[939.0 to 1080] [939.0 to 1080] [1013 to 1166] [1013 to 1166] {1013 to 1166] [970.0 to 1115]
122.2 to 140.5 122.2 to 140.5 86.8 to 99.8 86.8 to 99.8 129.8 to 136.9 125.6 to 144.4
150 (6.0) {1246 to 1433} {1246 to 1433} — {885.1 to 1018} {885.1 to 1018} {1324 to 1396} — {1281 to 1472}
[1082 to 1243] [1082 to 1243] [768.2 to 883.3] [768.2 to 883.3] {1149 to 1212] [1112 to 1278]
111.6 to 128.3 161.0 to 192.3 161.0 to 192.3 122.0 to 140.3 122.0 to 140.3 122.0 to 140.3 128.0 to 147.2
200 (8.0) {1138 to 1308} {1642 to 1961} {1642 to 1961} {1244 to 1431} {1244 to 1431} {1244 to 1431} — {1305 to 1501}
[987.7 to 1136] [1425 to 1702] [1425 to 1702] [1080 to 1242] [1080 to 1242] {1080 to 1242] [1133 to 1303]
167.7 to 192.9 167.7 to 192.9 167.7 to 192.9 227.6 to 261.7
250 (10) {1710 to 1967} {1710 to 1967} {1710 to 1967} — — — — {2321 to 2669}
[1484 to 1707] [1484 to 1707] [1484 to 1707] [2014 to 2316]
115.2 to 137.6 146.0 to 167.9 146.0 to 167.9 209.1 to 240.5
300 (12) {1175 to 1403} {1489 to 1712} {1489 to 1712} — — — — {2132 to 2452}
[1020 to 1218] [1292 to 1486] [1292 to 1486] [1851 to 2129]
N-m
Tightening torque values for Ceramics lining type Unit: {kgf-cm}
[in-lbf]
Gasket types Fluororesin with ceramic fillers (Valqua #7020) gasket (standard), or fluororesin with carbon gasket (optional code GF)
within flowtube
Gasket types PTFE-sheathed non-asbestos gasket (optional codes BCF and BSF), or the equivalent in hardness
for user’s flange
Flange ratings JIS F12
Size JIS 10K ANSI Class 150 DIN PN10 JIS20K ANSI Class 300 DIN PN16 DIN PN40
mm (inch) (JIS 75M)
29.9 to 49.8 29.9 to 49.8 30.5 to 50.9 30.5 to 50.9 32.7 to 54.5
50 (2.0) {304.9 to 507.8} {304.9 to 507.8} — {311.0 to 519.0} {311.0 to 519.0} — {333.4 to 555.7} —
[264.6 to 440.8] [264.6 to 440.8] [269.9 to 450.5] [269.9 to 450.5] [289.4 to 482.4]
48.5 to 80.8 48.5 to 80.8 46.6 to 77.6 46.6 to 77.6 46.6 to 77.6 56.3 to 93.8
80 (3.0) {494.6 to 823.9} {494.6 to 823.9} — {475.2 to 791.3} {475.2 to 791.3} {475.2 to 791.3} — {574.1 to 956.5}
[429.3 to 715.1] [429.3 to 715.1] [412.4 to 686.8] [412.4 to 686.8] [412.4 to 686.8] [498.3 to 830.2]
60.5 to 100.9 60.5 to 100.9 61.5 to 102.5 61.5 to 102.5 61.5 to 102.5 74.2 to 123.7
100 (4.0) {616.9 to 1029} {616.9 to 1029} — {627.1 to 1045} {627.1 to 1045} {627.1 to 1045} — {756.6 to 1261}
[535.5 to 893.0] [535.5 to 893.0] [544.3 to 907.2] [544.3 to 907.2] [544.3 to 907.2] [656.7 to 1095]
125.0 to 208.3 125.0 to 208.3 128.0 to 213.3 128.0 to 213.3 128.0 to 213.3 82.2 to 137.0
150 (6.0) {1275 to 2124} {1275 to 2124} — {1305 to 2175} {1305 to 2175} {1305 to 2175} — {838.2 to 1397}
[1106 to 1844] [1106 to 1844] [1133 to 1888] [1133 to 1888] [1133 to 1888] [727.5 to 1213]
174.7 to 291.2 174.7 to 291.2 174.7 to 291.2 120.1 to 200.2 120.1 to 200.2 120.1 to 200.2 86.7 to 144.6
200 (8.0) {1781 to 2969} {1781 to 2969} {1781 to 2969} {1225 to 2041} {1225 to 2041} {1225 to 2041} — {884.1 to 1475}
[1546 to2577] [1546 to2577] [1546 to2577] [1063 to 1772] [1063 to 1772] [1063 to 1772] [767.3 to 1280]
T0310.EPS
Table 3.3.10 Centering Device Identification (AXF Standard Models, PFA/Polyurethane Rubber/Natural Soft Rubber/EPDM
Rubber lining)
Flange JIS ANSI DIN
Size ratings
mm (inch) 10K 20K F12 (75M) 150 300 PN10 PN16 PN40
50 (2.0) B B — B F — — F
65 (2.5) B B — B G — F —
80 (3.0) B F H F C — G —
100 (4.0) B F H C H — F —
125 (5.0) B C C G D — F —
150 (6.0) C D D C E — C —
200 (8.0) C D D D E C C —
250 (10) C — D N — C C —
300 (12) C — D P — C C —
*: Each centering device is engraved with a character as identification. T0311.EPS
Table 3.3.11 Centering Device Identification (AXF Standard Models, Ceramics lining)
Flange JIS ANSI DIN
Size ratings
mm (inch) 10K 20K F12(75M) 150 300 PN10 PN16 PN40
50 (2.0) B B — B F — — F
80 (3.0) B F H F C — G —
100 (4.0) B F H C H — F —
150 (6.0) B C G B D — B —
200 (8.0) B C C G J B B —
*: Each centering device is engraved with a character as identification. T0312-1.EPS
3-13 IM 01E20D01-01E
3. INSTALLATION
Table 3.3.12 Centering Device Identification (Replacement Models, PFA/Polyurethane Rubber lining)
Flange JIS ANSI DIN
Size ratings
mm (inch) 10K 20K F12(75M) 150 300 PN10 PN16 PN40
50 (2.0) B B — B F — — F
80 (3.0) B F H F C — G —
100 (4.0) B F H C H — F —
150 (6.0) C D D C E — C —
200 (8.0) C D D D E C C —
T0312-2.EPS
*: Each centering device is engraved with a character as identification.
Table 3.3.13 Inner Diameter of Grounding Ring, Outer Diameter for Effective Sealing
Be sure to choose a gasket with an inner and outer diameter that does not protrude inside the piping.
If the inner diameter of the gasket is too large, or outer diameter of the gasket is too small, fluid leakage may result.
AXF Standard:
Size 2.5 to 400 mm Unit : mm (in.)
Lining PFA/Polyurethane Rubber/
Natural Soft Rubber/ Ceramics
EPDM Rubber
Inner diameter of Grounding ring Outer diameter for effective sealing Inner diameter of Outer diameter
[øA] [øB] Grounding ring for effective
Size Water Flange Water Flange [øA] sealing [øB]
2.5 (0.1) 15 (0.59) [12 (0.47)] *1 38 (1.50) 30 (1.18) - -
5 (0.2) 15 (0.59) [12 (0.47)] *1 38 (1.50) 30 (1.18) - -
10 (0.4) 15 (0.59) [12 (0.47)] *1 38 (1.50) 30 (1.18) - -
15 (0.5) 15 (0.59) 38 (1.50) 34 (1.34) 15 (0.59) 33 (1.30)
25 (1.0) 28 (1.10) 53 (2.09) 53 (2.09) 27 (1.06) 50 (1.97)
32 (1.25) 34 (1.34) 58 (2.28) 58 (2.28) - -
40 (1.5) 41 (1.61) 71 (2.80) 71 (2.80) 40 (1.57) 68 (2.68)
50 (2.0) 53 (2.09) 84 (3.31) 84 (3.31) 52 (2.05) 82 (3.23)
65 (2.5) 66 (2.60) 103 (4.06) 103 (4.06) - -
80 (3.0) 77 (3.03) 114 (4.49) 114 (4.49) 81 (3.19) 112 (4.41)
100 (4.0) 102 (4.02) 140 (5.51) 140 (5.51) 98 (3.86) 134 (5.28)
125 (5.0) 128 (5.04) 165 (6.50) 165 (6.50) - -
150 (6.0) 146.1 (5.75) 190 (7.48) 190 (7.48) 144 (5.67) 188 (7.40)
200 (8.0) 193.6 (7.62) 240 (9.45) 240 (9.45) 192 (7.56) 240 (9.45)
250 (10) 243.7 (9.60) 243 (9.57) 300 (11.81) 315 (12.40) - -
300 (12) 294.7 (11.60) 291.3 (11.47) 348 (13.70) 360 (14.17) - -
350 (14) - 323.4 (12.73) - 405 (15.94) - -
400 (16) - 373.5 (14.70) - 465 (18.31) - -
T0313-01.EPS
Size 500 to 2600 mm Unit : mm (in.) Replacement Model for earlier ADMAG or ADMAG AE:
Lining Polyurethane Rubber Unit : mm (in.)
Inner diameter of Outer diameter for Lining PFA/Polyurethane Rubber
Size Grounding ring [øA] effective sealing [øB] Outer diameter for
Inner diameter of effective sealing [øB]
500 (20) 468 (18.42) [485 (19.09)] *2 570 (22.44) Grounding ring [øA]
Size Wafer Flange
600 (24) 563 (22.16) [589 (23.18)] *2 675 (26.57)
700 (28) 665 (26.18) [689 (27.12)] *2 785 (30.91) 2.5 (0.1) 15 (0.59) 38 (1.50) -
800 (32) 765 (30.11) [788 (31.02)] *2 890 (35.04) 5 (0.2) 15 (0.59) 38 (1.50) -
900 (36) 855 (33.66) [888 (34.96)] *2 990 (38.98) 10 (0.4) 15 (0.59) 38 (1.50) -
1000 (40) 942 (37.08) [990 (38.97)] *2 1095 (43.11) 15 (0.5) 15 (0.59) 38 (1.50) -
1100 (44) 1085 (42.71) 1230 (48.43) 25 (1.0) 27 (1.06) 56 (2.20) -
1200 (48) 1185 (46.65) 1334 (52.52) 40 (1.5) 40 (1.57) 71 (2.80) -
1350 (54) 1335 (52.55) 1490 (58.66) 50 (2.0) 52 (2.05) 85 (3.35) -
1500 (60) 1485 (58.46) 1646 (64.80) 80 (3.0) 81 (3.19) 115 (4.53) -
1600 (64) 1585 (62.40) 1756 (69.13) 100 (4.0) 98 (3.86) 144 (5.57) -
1800 (72) 1785 (70.27) 1948 (76.69) 150 (6.0) 140.7 (5.6) 190 (7.48) 205 (8.07)
2000 (80) 1985 (78.14) 2160 (85.04) 200 (8.0) 188.9 (7.5) 240 (9.45) 255 (10.04)
2200 (88) 2185 (86.02) 2370 (93.31) 250 (10) 243 (9.57) - 315 (12.40)
T0313-03.EPS
2400 (96) 2385 (93.89) 2580 (101.57)
2600 (104) 2585 (101.77) 2770 (109.06)
T0313-02.EPS
*1: The I.D. of the process connection code: DD4, DJ1, DJ2 is values in brackets [ ].
*2: The I.D. of the process connection code: CG1 is values in brackets [ ].
3-14 IM 01E20D01-01E
3. INSTALLATION
CAUTION
IMPORTANT
3-15 IM 01E20D01-01E
3. INSTALLATION
*Piping-side flange
*Gasket (two units)
*: These items must be provided by the user.
Choose nuts and bolts in compliance with
the flange ratings.
Flowmeter-side flange
*Bolt
*Nut
F0313.EPS
Figure 3.3.5 Mounting Procedure for Flange Type (size: 2.5 mm (0.1 in.) to 400 mm (16 in.))
Table 3.3.14 Flange Type Tightening Torque Values for Metal Piping
N-m
Tightening torque values for PFA/Polyurethane Rubber/Natural Soft Rubber/EPDM Rubber lining type Unit: {kgf-cm}
[in-lbf]
Gasket types No gasket (standard)
within flowtube
Gasket types
Non-asbestos gasket, PTFE-sheathed non-asbestos gasket, or the equivalent in hardness
for user’s flange
Flange ratings JIS F12
Size JIS 10K ANSI Class 150 DIN PN10 JIS20K ANSI Class 300 DIN PN16 DIN PN40
mm (inch) (JIS 75M)
2.5 (0.1) 3.8 to 6.3 3.8 to 6.3 3.8 to 6.3 3.8 to 6.3 3.8 to 6.4
{38.75 to 64.24} {38.75 to 64.24} — {38.75 to 64.24} {38.75 to 64.24} — {38.75 to 65.26} —
(with 10-mm flanges) [33.63 to 55.76] [33.63 to 55.76] [33.63 to 55.76] [33.63 to 55.76] [33.63 to 56.64]
2.5 (0.1) 4.6 to 7.7 4.6 to 7.7 4.7 to 7.8 4.7 to 7.8 4.7 to 7.9
{46.91 to 78.52} {46.91 to 78.52} — {47.93 to 79.54} {47.93 to 79.54} — {47.93 to 80.56} —
(with 15-mm flanges) [40.71 to 68.15] [40.71 to 68.15] [41.6 to 69.03] [41.6 to 69.03] [41.6 to 69.92]
5 (0.2) 3.8 to 6.3 3.8 to 6.3 3.8 to 6.3 3.8 to 6.3 3.8 to 6.4
{38.75 to 64.24} {38.75 to 64.24} — {38.75 to 64.24} {38.75 to 64.24} — {38.75 to 65.26} —
(with 10-mm flanges) [33.63 to 55.76] [33.63 to 55.76] [33.63 to 55.76] [33.63 to 55.76] [33.63 to 56.64]
5 (0.2) 4.6 to 7.7 4.6 to 7.7 4.7 to 7.8 4.7 to 7.8 4.7 to 7.9
{46.91 to 78.52} {46.91 to 78.52} — {47.93 to 79.54} {47.93 to 79.54} — {47.93 to 80.56} —
(with 15-mm flanges) [40.71 to 68.15] [40.71 to 68.15] [41.6 to 69.03] [41.6 to 69.03] [41.6 to 69.92]
10 (0.4) 3.8 to 6.3 3.8 to 6.3 3.8 to 6.3 3.8 to 6.3 3.8 to 6.4
{38.75 to 64.24} {38.75 to 64.24} — {38.75 to 64.24} {38.75 to 64.24} — {38.75 to 65.26} —
(with 10-mm flanges) [33.63 to 55.76] [33.63 to 55.76] [33.63 to 55.76] [33.63 to 55.76] [33.63 to 56.64]
10 (0.4) 4.6 to 7.7 4.6 to 7.7 4.7 to 7.8 4.7 to 7.8 4.7 to 7.9
{46.91 to 78.52} {46.91 to 78.52} — {47.93 to 79.54} {47.93 to 79.54} — {47.93 to 80.56} —
(with 15-mm flanges) [40.71 to 68.15] [40.71 to 68.15] [41.6 to 69.03] [41.6 to 69.03] [41.6 to 69.92]
4.6 to 7.7 4.6 to 7.7 4.7 to 7.8 4.7 to 7.8 4.7 to 7.9
15 (0.5) {46.91 to 78.52} {46.91 to 78.52} — {47.93 to 79.54} {47.93 to 79.54} — {47.93 to 80.56} —
[40.71 to 68.15] [40.71 to 68.15] [41.6 to 69.03] [41.6 to 69.03] [41.6 to 69.92]
13.1 to 15.1 13.1 to 15.1 13.2 to 15.2 13.2 to 15.2 11.9 to 13.7
25 (1.0) {133.6 to 154.0} {133.6 to 154.0} — {134.6 to 155.0} {134.6 to 155.0} — {121.3 to 139.7} —
[115.9 to 133.6] [115.9 to 133.6] [116.8 to 134.5] [116.8 to 134.5] [105.3 to 121.3]
14.4 to 15.6 14.4 to 15.6 14.6 to 16.8 14.6 to 16.8 15.0 to 17.3
32 (1.25) {146.8 to 159.1} {146.8 to 159.1} — {148.9 to 171.3} {148.9 to 171.3} — {153.0 to 176.4} —
[127.4 to138.1] [127.4 to138.1] [129.2 to 148.7] [129.2 to 148.7] [132.8 to 153.1]
21.9 to 25.2 21.9 to 25.2 22.2 to 25.5 22.2 to 25.5 23.2 to 26.7
40 (1.5) {223.3 to 257.0} {223.3 to 257.0} — {226.4 to 260.0} {226.4 to 260.0} — {236.6 to 272.3} —
[193.8 to 223.0] [193.8 to 223.0] [196.5 to 225.7] [196.5 to 225.7] [205.3 to 236.3]
28.0 to 32.2 28.0 to 32.2 27.8 to 32.0 27.8 to 32.0 28.8 to 33.1
50 (2.0) {285.5 to 328.3} {285.5 to 328.3} — {283.5 to 326.3} {283.5 to 326.3} — {293.7 to 337.5} —
[247.8 to 285.0] [247.8 to 285.0] [246.0 to 283.2] [246.0 to 283.2] [254.9 to 293.0]
41.6 to 47.8 41.6 to 47.8 19.5 to 28.5 19.5 to 28.5 41.4 to 47.6
65 (2.5) {424.2 to 487.4} {424.2 to 487.4} — {198.8 to 290.6} {198.8 to 290.6} {422.2 to 485.4} — —
[368.2 to 423.0] [368.2 to 423.0] [172.6 to 252.2] [172.6 to 252.2] [366.4 to 421.3]
23.2 to 26.7 52.7 to 53.6 26.1 to 30.0 26.1 to 30.0 26.1 to 30.0 46.0 to 52.9
80 (3.0) {236.6 to 272.3} {536.9 to 546.2} — {266.1 to 305.9} {266.1 to 305.9} {266.1 to 305.9} — {469.1 to 539.4}
[205.3 to 236.3] [466.0 to 474.1] [231.0 to 265.5] [231.0 to 265.5] [231.0 to 265.5] [407.1 to 468.2]
30.9 to 35.5 30.9 to 35.5 34.8 to 40.0 34.8 to 40.0 34.8 to 40.0 60.9 to 70.0
100 (4.0) {315.1 to 362.0} {315.1 to 362.0} — {354.9 to 407.9} {354.9 to 407.9} {354.9 to 407.9} — {621.0 to 713.8}
[273.5 to 314.2] [273.5 to 314.2] [308.0 to 354.0] [308.0 to 354.0] [308.0 to 354.0] [539.0 to 619.5]
45.6 to 52.4 45.6 to 52.4 48.6 to 55.9 48.6 to 55.9 48.6 to 55.9 48.1 to 55.3
125 (5.0) {465.0 to 534.3} {465.0 to 534.3} — {495.6 to 570.0} {495.6 to 570.0} {495.6 to 570.0} — {490.5 to 563.9}
[403.6 to 463.8] [403.6 to 463.8] [430.1 to 494.7] [430.1 to 494.7] [430.1 to 494.7] [425.7 to 489.4]
64.5 to 74.2 64.5 to 74.2 44.8 to 51.5 44.8 to 51.5 66.7 to 80.8 67.5 to 77.6
150 (6.0) {657.7 to 756.6} {657.7 to 756.6} — {456.8 to 525.2} {456.8 to 525.2} {680.2 to 823.9} — {688.3 to 791.3}
[570.8 to 656.7] [570.8 to 656.7] [396.5 to 455.8] [396.5 to 455.8] [590.3 to 715.1] [597.4 to 686.8]
59.2 to 68.1 100.3 to 102.1 100.3 to 102.1 63.6 to 73.1 63.6 to 73.1 63.6 to 73.1 69.9 to 80.4
200 (8.0) {603.7 to 694.4} {1023 to 1041} {1023 to 1041} {648.5 to 745.4} {648.5 to 745.4} {648.5 to 745.4} — {712.8 to 819.9}
[523.9 to 602.7] [887.8 to 903.3] [887.8 to 903.3] [562.9 to 647.0] [562.9 to 647.0] [562.9 to 647.0] [618.6 to 711.6]
144.0 to 165.6 144.0 to 165.6 144.0 to 165.6 156.3 to 179.7 127.2 to 148.8 156.3 to 179.7 190.5 to 219.1
250 (10) {1468 to 1689} {1468 to 1689} {1468 to 1689} {1594 to 1832} {1297 to 1517} {1594 to 1832} — {1943 to 2234}
[1274 to 1466] [1274 to 1466] [1274 to 1466] [1383 to 1590] [1126 to 1317] [1383 to 1590] [1686 to 1939]
119.3 to 137.2 163.1 to 197.4 163.1 to 197.4 123.1 to 136.3 152.3 to 169.4 173.3 to 199.3 168.0 to 193.2
300 (12) {1217 to 1399} {1663 to 2013} {1663 to 2013} {1255 to 1390} {1553 to 1727} {1767 to 2032} — {1713 to 1970}
[1056 to 1214] [1443 to 1747] [1443 to 1747] [1089 to 1206] [1348 to 1499] [1534 to 1764] [1487 to 1710]
164.7 to 189.4 256.4 to 293.3 164.7 to 189.4 238.6 to 274.4
350 (14) {1679 to 1931} {2615 to 2991} {1679 to 1931} — — — — {2433 to 2798}
[1458 to 1676] [2269 to 2596] [1458 to 1676] [2112 to 2429]
225.2 to 258.9 225.2 to 258.9 225.2 to 258.9 305.8 to 351.7
400 (16) {2296 to 2640} {2296 to 2640} {2296 to 2640} — — — — {3118 to 3586}
[1993 to 2291] [1993 to 2291] [1993 to 2291] [2706 to 3113]
T0314.EPS
3-16 IM 01E20D01-01E
3. INSTALLATION
Table 3.3.15 Flange Type Tightening Torque Values for PVC Piping
N-m
Tightening torque values for PFA lining type Unit: {kgf-cm}
[in-lbf]
Gasket types Fluororubber gasket (optional codes GA, GC, and GD)
within flowtube
Gasket types Fluororubber gasket, chloroprene rubber gasket, or the equivalent in hardness
for user’s flange
Flange ratings JIS F12
Size JIS 10K ANSI Class 150 DIN PN10 JIS20K ANSI Class 300 DIN PN16 DIN PN40
mm (inch) (JIS 75M)
2.5 (0.1) 0.8 to 1.3 0.8 to 1.3 0.8 to 1.3 0.8 to 1.3 0.8 to 1.3
{8.158 to 13.26} {8.158 to 13.26} — {8.158 to 13.26} {8.158 to 13.26} — {8.158 to 13.26} —
(with 10-mm flanges) [7.08 to 11.51] [7.08 to 11.51] [7.08 to 11.51] [7.08 to 11.51] [7.08 to 11.51]
2.5 (0.1) 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6
{9.177 to 16.32} {9.177 to 16.32} — {9.177 to 16.32} {9.177 to 16.32} — {9.177 to 16.32} —
(with 15-mm flanges) [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16]
5 (0.2) 0.8 to 1.3 0.8 to 1.3 0.8 to 1.3 0.8 to 1.3 0.8 to 1.3
{8.158 to 13.26} {8.158 to 13.26} — {8.158 to 13.26} {8.158 to 13.26} — {8.158 to 13.26} —
(with 10-mm flanges) [7.08 to 11.51] [7.08 to 11.51] [7.08 to 11.51] [7.08 to 11.51] [7.08 to 11.51]
5 (0.2) 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6
{9.177 to 16.32} {9.177 to 16.32} — {9.177 to 16.32} {9.177 to 16.32} — {9.177 to 16.32} —
(with 15-mm flanges) [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16]
10 (0.4) 0.8 to 1.3 0.8 to 1.3 0.8 to 1.3 0.8 to 1.3 0.8 to 1.3
{8.158 to 13.26} {8.158 to 13.26} — {8.158 to 13.26} {8.158 to 13.26} — {8.158 to 13.26} —
(with 10-mm flanges) [7.08 to 11.51] [7.08 to 11.51] [7.08 to 11.51] [7.08 to 11.51] [7.08 to 11.51]
10 (0.4) 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6
{9.177 to 16.32} {9.177 to 16.32} — {9.177 to 16.32} {9.177 to 16.32} — {9.177 to 16.32} —
(with 15-mm flanges) [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16]
0.9 to 1.6 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6 0.9 to 1.6
15 (0.5) {9.177 to 16.32} {9.177 to 16.32} — {9.177 to 16.32} {9.177 to 16.32} — {9.177 to 16.32} —
[7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16] [7.966 to 14.16]
2.7 to 4.5 2.7 to 4.5 2.7 to 4.5 2.7 to 4.5 2.3 to 3.9
25 (1.0) {27.53 to 45.89} {27.53 to 45.89} — {27.53 to 45.89} {27.53 to 45.89} — {23.45 to 39.77} —
[23.9 to 39.83] [23.9 to 39.83] [23.9 to 39.83] [23.9 to 39.83] [20.36 to 34.52]
3.0 to 4.9 3.0 to 4.9 3.0 to 5.0 3.0 to 5.0 2.9 to 4.9
32 (1.25) {30.59 to 49.97} {30.59 to 49.97} — {30.59 to 50.99} {30.59 to 50.99} — {29.57 to 49.97} —
[26.55 to 43.37] [26.55 to 43.37] [26.55 to 44.25] [26.55 to 44.25] [25.67 to 43.37]
4.5 to 7.6 4.5 to 7.6 4.7 to 7.8 4.7 to 7.8 4.4 to 7.4
40 (1.5) {45.89 to 77.5} {45.89 to 77.5} — {47.93 to 79.54} {47.93 to 79.54} — {44.87 to 75.46} —
[39.83 to 67.26] [39.83 to 67.26] [41.6 to 69.03] [41.6 to 69.03] [38.94 to 65.49]
5.9 to 9.8 5.9 to 9.8 2.9 to 4.8 2.9 to 4.8 5.5 to 9.2
50 (2.0) {60.16 to 99.93} {60.16 to 99.93} — {29.57 to 48.95 } {29.57 to 48.95 } — {56.08 to 93.81 } —
[52.22 to 86.74] [52.22 to 86.74] [25.67 to 42.48] [25.67 to 42.48] [48.68 to 81.43]
9.0 to 15.0 9.0 to 15.0 2.1 to 4.4 2.1 to 4.4 4.4 to 7.3
65 (2.5) {91.8 to 153.0} {91.8 to 153.0} — {21.4 to 44.9} {21.4 to 44.9} {44.9 to 74.4} — —
[79.7 to 132.8] [79.7 to 132.8] [18.6 to 38.9] [18.6 to 38.9] [38.9 to 64.6]
4.9 to 8.1 11.1 to 16.3 5.5 to 9.1 5.5 to 9.1 5.5 to 9.1 9.7 to 12.2
80 (3.0) {50.0 to 82.6} {113.2 to 166.2} — {56.1 to 92.8} {56.1 to 92.8} {56.1 to 92.8} — {98.9 to 124.4}
[43.4 to 71.7] [98.2 to 144.3] [48.7 to 80.5] [48.7 to 80.5] [48.7 to 80.5] [85.8 to 108.0]
6.7 to 11.2 6.7 to 11.2 7.5 to 12.6 7.5 to 12.6 7.5 to 12.6 13.3 to 22.2
100 (4.0) {68.3 to 114.2} {68.3 to 114.2} — {76.5 to 128.5} {76.5 to 128.5} {76.5 to 128.5} — {135.6 to 226.4}
[59.3 to 99.1] [59.3 to 99.1] [66.4 to 111.5] [66.4 to 111.5] [66.4 to 111.5] [117.7 to 196.5]
9.9 to 16.5 9.9 to 16.5 10.7 to 17.8 10.7 to 17.8 10.7 to 17.8 10.5 to 17.6
125 (5.0) {101.0 to 168.3} {101.0 to 168.3} — {109.1 to 181.5} {109.1 to 181.5} {109.1 to 181.5} — {107.1 to 179.5}
[87.6 to 146.0] [87.6 to 146.0] [94.7 to 157.5] [94.7 to 157.5] [94.7 to 157.5] [92.9 to 155.8]
14.4 to 24.0 14.4 to 24.0 9.8 to 16.3 9.8 to 16.3 14.6 to 25.6 15.2 to 25.3
150 (6.0) {146.8 to 244.7} {146.8 to 244.7} — {99.9 to 166.2} {99.9 to 166.2} {148.9 to 261.0} — {155.0 to 258.0}
[127.4 to 212.4] [127.4 to 212.4] [86.7 to 144.3] [86.7 to 144.3] [129.2 to 226.6] [134.5 to 223.9]
13.4 to 22.3 22.7 to 33.4 22.7 to 33.4 14.6 to 24.3 14.6 to 24.3 14.6 to 24.3 16.1 to 26.9
200 (8.0) {136.6 to 227.4} {231.5 to 340.6} {231.5 to 340.6} {148.9 to 247.8} {148.9 to 247.8} {148.9 to 247.8} — {164.2 to 274.3}
[118.6 to 197.4] [200.9 to 295.6] [200.9 to 295.6] [129.2 to 215.1] [129.2 to 215.1] [129.2 to 215.1] [142.5 to 238.1]
T0315.EPS
3-17 IM 01E20D01-01E
3. INSTALLATION
Table 3.3.16 Flange Type Tightening Torque Values for Metal Piping and Permeable Fluids
N-m
Tightening torque values for PFA lining type Unit: {kgf-cm}
[in-lbf]
Gasket types No gasket (standard)
within flowtube
Gasket types
Non-asbestos gasket, PTFE-sheathed non-asbestos gasket, or the equivalent in hardness
for user’s flange
Flange ratings JIS F12
Size JIS 10K ANSI Class 150 DIN PN10 JIS20K ANSI Class 300 DIN PN16 DIN PN40
mm (inch) (JIS 75M)
2.5 (0.1) 5.7 to 6.6 5.7 to 6.6 5.7 to 6.6 5.7 to 6.6 5.7 to 6.6
{58.12 to 67.3} {58.12 to 67.3} — {58.12 to 67.3} {58.12 to 67.3} — {58.12 to 67.3} —
(with 10-mm flanges) [50.45 to 58.41] [50.45 to 58.41] [50.45 to 58.41] [50.45 to 58.41] [50.45 to 58.41]
2.5 (0.1) 6.9 to 7.9 6.9 to 7.9 7.0 to 8.1 7.0 to 8.1 7.0 to 8.1
{70.36 to 80.56} {70.36 to 80.56} — {71.38 to 82.6} {71.38 to 82.6} — {71.38 to 82.6} —
(with 15-mm flanges) [61.07 to 69.92] [61.07 to 69.92] [61.95 to 71.69] [61.95 to 71.69] [61.95 to 71.69]
5 (0.2) 5.7 to 6.6 5.7 to 6.6 5.7 to 6.6 5.7 to 6.6 5.7 to 6.6
{58.12 to 67.3} {58.12 to 67.3} — {58.12 to 67.3} {58.12 to 67.3} — {58.12 to 67.3} —
(with 10-mm flanges) [50.45 to 58.41] [50.45 to 58.41] [50.45 to 58.41] [50.45 to 58.41] [50.45 to 58.41]
5 (0.2) 6.9 to 7.9 6.9 to 7.9 7.0 to 8.1 7.0 to 8.1 7.0 to 8.1
{70.36 to 80.56} {70.36 to 80.56} — {71.38 to 82.6} {71.38 to 82.6} — {71.38 to 82.6} —
(with 15-mm flanges) [61.07 to 69.92] [61.07 to 69.92] [61.95 to 71.69] [61.95 to 71.69] [61.95 to 71.69]
10 (0.4) 5.7 to 6.6 5.7 to 6.6 5.7 to 6.6 5.7 to 6.6 5.7 to 6.6
{58.12 to 67.3} {58.12 to 67.3} — {58.12 to 67.3} {58.12 to 67.3} — {58.12 to 67.3} —
(with 10-mm flanges) [50.45 to 58.41] [50.45 to 58.41] [50.45 to 58.41] [50.45 to 58.41] [50.45 to 58.41]
10 (0.4) 6.9 to 7.9 6.9 to 7.9 7.0 to 8.1 7.0 to 8.1 7.0 to 8.1
{70.36 to 80.56} {70.36 to 80.56} — {71.38 to 82.6} {71.38 to 82.6} — {71.38 to 82.6} —
(with 15-mm flanges) [61.07 to 69.92] [61.07 to 69.92] [61.95 to 71.69] [61.95 to 71.69] [61.95 to 71.69]
6.9 to 7.9 6.9 to 7.9 7.0 to 8.1 7.0 to 8.1 7.0 to 8.1
15 (0.5) {70.36 to 80.56} {70.36 to 80.56} — {71.38 to 82.6} {71.38 to 82.6} — {71.38 to 82.6} —
[61.07 to 69.92] [61.07 to 69.92] [61.95 to 71.69] [61.95 to 71.69] [61.95 to 71.69]
19.6 to 22.5 19.6 to 22.5 19.7 to 22.7 19.7 to 22.7 17.5 to 20.1
25 (1.0) {199.9 to 229.4} {199.9 to 229.4} — {200.9 to 231.5} {200.9 to 231.5} — {178.5 to 205.0} —
[173.5 to 199.1] [173.5 to 199.1] [174.4 to 200.9] [174.4 to 200.9] [154.9 to 177.9]
21.5 to 24.7 21.5 to 24.7 21.6 to 24.8 21.6 to 24.8 22.1 to 25.4
32 (1.25) {219.2 to 251.9} {219.2 to 251.9} — {220.3 to 252.9} {220.3 to 252.9} — {225.4 to 259.0} —
[190.3 to 218.6] [190.3 to 218.6] [191.2 to 219.5] [191.2 to 219.5] [195.6 to 224.8]
32.5 to 37.4 32.5 to 37.4 32.8 to 37.7 32.8 to 37.7 33.8 to 38.9
40 (1.5) {331.4 to 381.4} {331.4 to 381.4} — {334.5 to 384.4} {334.5 to 384.4} — {344.7 to 396.7} —
[287.6 to 331.0] [287.6 to 331.0] [290.3 to 333.7] [290.3 to 333.7] [229.2 to 344.3]
41.3 to 47.5 41.3 to 47.5 20.6 to 23.7 20.6 to 23.7 42.2 to 48.5
50 (2.0) {421.1 to 484.4} {421.1 to 484.4} — {210.1 to 241.7} {210.1 to 241.7} — {430.3 to 494.6} —
[365.5 to 420.4] [365.5 to 420.4] [182.3 to 209.8] [182.3 to 209.8] [373.5 to 429.3]
61.2 to 70.4 61.2 to 70.4 14.3 to 21.0 14.3 to 21.0 30.5 to 35.1
65 (2.5) {624.1 to 717.9} {624.1 to 717.9} — {145.8 to 214.1} {145.8 to 214.1} {311.0 to 357.9} — —
[541.6 to 623.1] [541.6 to 623.1] [126.6 to 185.9] [126.6 to 185.9] [269.9 to 310.6]
34.2 to 39.3 77.6 to 78.8 38.5 to 44.3 38.5 to 44.3 38.5 to 44.3 68.1 to 78.3
80 (3.0) {348.7 to 400.7} {791.3 to 803.5} — {392.6 to 451.7} {392.6 to 451.7} {392.6 to 451.7} — {694.4 to 798.4}
[302.7 to 347.8] [686.8 to 697.4] [340.7 to 392.1] [340.7 to 392.1] [340.7 to 392.1] [602.7 to 693.0]
45.2 to 52.0 45.2 to 52.0 51.0 to 58.7 51.0 to 58.7 51.0 to 58.7 89.6 to 103.0
100 (4.0) {460.9 to 530.3} {460.9 to 530.3} — {520.1 to 598.6} {520.1 to 598.6} {520.1 to 598.6} — {913.7 to 1050}
[400.0 to 460.2] [400.0 to 460.2] [451.4 to 519.5] [451.4 to 519.5] [451.4 to 519.5] [793.0 to 911.6]
66.8 to 76.8 66.8 to 76.8 70.8 to 81.4 70.8 to 81.4 70.8 to 81.4 70.7 to 81.3
125 (5.0) {681.2 to 783.1} {681.2 to 783.1} — {722.0 to 830.1} {722.0 to 830.1} {722.0 to 830.1} — {720.9 to 829.0}
[591.2 to 679.7] [591.2 to 679.7] [626.6 to 720.4] [626.6 to 720.4] [626.6 to 720.4] [625.7 to 719.5]
93.9 to 108.8 93.9 to 108.8 65.4 to 75.2 65.4 to 75.2 97.3 to 118.0 98.8 to 113.6
150 (6.0) {957.5 to 1109} {957.5 to 1109} — {666.9 to 766.8} {666.9 to 766.8} {992.2 to 1203} — {1007 to 1158}
[831.1 to 962.9] [831.1 to 962.9] [578.8 to 665.5] [578.8 to 665.5] [861.1 to 1044] [874.4 to 1005]
85.8 to 98.7 145.4 to 147.9 145.4 to 147.9 91.5 to 105.2 91.5 to 105.2 91.5 to 105.2 101.8 to 117.1
200 (8.0) {874.9 to 1006} {1483 to 1508} {1483 to 1508} {933.0 to 1073} {933.0 to 1073} {933.0 to 1073} — {1038 to 1194}
[759.4 to 873.5] [1287 to 1309] {1287 to 1309] [809.8 to 931.1] [809.8 to 931.1] [809.8 to 931.1] [901.0 to 1036]
207.8 to 239.0 207.8 to 239.0 207.8 to 239.0 222.9 to 256.3 181.4 to 212.2 222.9 to 256.3 277.9 to 319.6
250 (10) {2119 to 2437} {2119 to 2437} {2119 to 2437} {2273 to 2614} {1850 to 2164} {2273 to 2614} — {2834 to 3259}
[1839 to 2115] [1839 to 2115] {1839 to 2115] [1973 to 2268] [1605 to 1878] [1973 to 2268] [2460 to 2829]
171.0 to 196.7 233.7 to 283.0 233.7 to 283.0 130.7 to 144.8 161.8 to 180.0 184.1 to 211.7 243.8 to 280.4
300 (12) {1744 to 2006} {2383 to 2886} {2383 to 2886} {1333 to 1477} {1650 to 1835} {1877 to 2159} — {2486 to 2859}
[1513 to 1741] [2068 to 2505] {2068 to 2505] [1157 to 1282] [1432 to 1593] [1629 to 1874] [2158 to 2482]
234.7 to 269.9 365.4 to 418.0 234.7 to 269.9 350.6 to 403.2
350 (14) {2393 to 2752} {3726 to 4262} {2393 to 2752} — — — — {3575 to 4112}
[2077 to 2389] [3234 to 3699] {2077 to 2389] [3103 to 3568]
320.0 to 368.0 320.0 to 368.0 320.0 to 368.0 448.3 to 515.5
400 (16) {3263 to 3753} {3263 to 3753} {3263 to 3753} — — — — {4571 to 5257}
[2832 to 3257] [2832 to 3257] {2832 to 3257] [3968 to 4562]
T0316.EPS
3-18 IM 01E20D01-01E
3. INSTALLATION
A 4-Anchor bolt
D D
Fig3.2.18.eps
Tab.3.2.10a.eps
Figure 3.3.6 Lifting Flowtube
3-19 IM 01E20D01-01E
3. INSTALLATION
A
16-R hole
E
F
Tab.3.2.11a.eps
3-20 IM 01E20D01-01E
3. INSTALLATION
Table 3.3.19 Flange Type Tightening Torque Values for Material Piping
Tightening torque values for Polyurethane Rubber lining (N-m / {kgf-cm} / [in-lbf])
Gasket types
Rubber gasket, or the equivalent in hardness
for user's
Flange ratings
JIS F12
Size JIS 10K ANSI Class150 DIN PIN10
(JIS 75M)
mm (inch)
500 (20) 43 to 72 / {438.5 to 734.2}/[380.6 to 637.2] 51 to 85 / {520.1 to 866.8}/[451.4 to 752.3] 43 to 71 / {438.5 to 724.0}/[380.6 to 628.4] 61 to 102 / {622.0 to 1040}/[539.9 to 902.7]
600 (24) 52 to 87 / {530.3 to 887.2}/[460.2 to 770.0] 82 to 137 / {836.2 to 1397}/[725.7 to 1213] 64 to 107 / {652.6 to 1091}/[566.4 to 947.0] 65 to 108 / {662.8 to 1101}/[575.3 to 955.8]
700 (28) 69 to 116 / {703.6 to 1183}/[610.7 to 1027] — 71 to 118 / {724.0 to 1203}/[628.4 to 1044] 84 to 141 / {856.6 to 1438}/[743.4 to 1248]
800 (32) 71 to 119 / {724.0 to 1213}/[628.4 to 1053] — 94 to 157 / {958.5 to 1601}/[831.9 to 1390] 81 to 136 / {826.0 to 1387}/[716.9 to 1204]
900 (36) 93 to 155 / {948.3 to 1581}/[823.1 to 1372] — 101 to 168 / {1030 to 1713}/[893.9 to 1487] 98 to 163 / {999.3 to 1662}/[867.3 to 1443]
1000 (40) 150 to 249 / {1530 to 2539}/[1328 to 2204] — 147 to 245 / {1499 to 2498}/[1301 to 2168] 117 to 195 / {1193 to 1988}/[1035 to 1076]
3-21 IM 01E20D01-01E
3. INSTALLATION
The sanitary type can be mounted to the piping using a (a) Clamp type
clamp, a union, or a welded joint. (process connection codes: HAB, HDB, and HKB)
1) Welding ferrule
Weld a ferrule to the piping.
NOTE
Gasket
Ferrule
F0314.EPS
Figure 3.3.8 Mounting Procedure for Clamp Connection
Type
3-22 IM 01E20D01-01E
3. INSTALLATION
Gasket
IMPORTANT
Sleeve
• Be sure to weld the sleeve after passing the
Nut
piping through the nut.
• When welding the sleeve, pay attention to the
Adapter mounting screw
edge preparation, level differences between the
sleeve and the piping, and the welding current
to avoid deforming the piping or causing
stagnation of some of the fluid. F0315.EPS
5) Confirmation of adapter mounting screw • Be sure to weld the ferrule, the sleeve, or the
After installation of the magnetic flowmeter, be alternative after removing the butt weld adapter
sure to retighten the adapter mounting screw from the flowmeter itself.
according to Table 3.3.21 or Table 3.3.22. • When welding the ferrule, the sleeve, or the
Be sure to confirm that leakage from adapter alternative, pay attention to the edge prepara-
connection point does not occur by filling the tion, level differences between them and the
pipe of the flowtube with fluid. piping, and the welding current to avoid deform-
ing the piping or causing stagnation portion of
the fluid.
NOTE
3-23 IM 01E20D01-01E
3. INSTALLATION
CAUTION
3-24 IM 01E20D01-01E
4. WIRING
4. WIRING
4.1 Wiring the Integral • When waterproof glands or union equipped
waterproof glands are used, avoid tightening
Flowmeter the glands with an excessive torque.
• In case of 24 V power supply version, it comes
This section describes the wiring of the integral flowmeter.
with a plug. Use this plug to cover the unused
WARNING wiring port when wiring the instrument with only
one, four-core cable.
The wiring of the magnetic flowmeter must be • Be sure to turn the power off before opening
performed by expert engineer or skilled person- the terminal box cover.
nel. No operator shall be permitted to perform • Before turning the power on, tighten the
procedures relating to wiring. terminal box cover securely.
• The terminal box cover is locked by the special
screw. In case of opening the terminal box
cover, use the hexagonal wrench attached. For
CAUTION handling the locking screw, refer to Figure
4.1.5.
Once all wiring is complete, check the connec- • Be sure to lock the cover by the special screw
tions before applying power to the instrument. using the hexagonal wrench attached after
Improper arrangements or wiring may cause a installing the cover. For handling the locking
unit malfunction or damage. screw, refer to Figure 4.1.15.
• Explosion protected types must be wired in
accordance with specific requirement (and, in
certain countries, legal regulations) in order to
4.1.1 Wiring Precautions preserve the effectiveness of their explosion
protected features.
Be sure to observe the following precautions when
wiring:
4.1.2 Power Cable/Output Cable
CAUTION JIS C3401 control cable equivalent
JIS C3312 power cable equivalent
• In cases where the ambient temperature 14 AWG Belden 8720 equivalent
exceeds 50°C (122°F), use external heat-
resistant wiring with a maximum allowable Outer Diameter:
temperature of 70°C (158°F) or above. With no gland option:
• Do not connect cables outdoors in wet weather 6.5 to 12 mm (0.26 to 0.47 in.)
in order to prevent damage from condensation With gland options EG, EU and EW:
and to protect the insulation, e.g. inside the 10.5 or 11.5 mm (0.41 to 0.45 in.)
terminal box of the flowmeter. With gland options EP:
• All the cable ends must be provided with round 6 to 12 mm (0.24 to 0.47 in.)
crimp-on terminals and be securely wired. Nominal Cross Section:
• The signal cables must be routed in separate Single wire; 0.5 to 2.5 mm2
steel conduit tubes 16 (JIS C 8305) or flexible Stranded wire; 0.5 to 1.5 mm2
conduit tubes 15 (JIS C 8309).
• Always route the power and output signal In case of power cable, Green/Yellow covered conduc-
cables in separate steel conduit tubes, except tor shall be used only for connection to PROTECTIVE
when the power supply voltage is 24 V and CONDUCTOR TERMINALS. Conform to IEC227,
four-core cables are used for wiring. Keep IEC245 or equivalent national authorization.
conduits or flexible tubes watertight using
sealing tape.
4-1 IM 01E20D01-01E
4. WIRING
NOTE
IMPORTANT
PF1/2
(1) When waterproof property is unnecessary
Washer
(When there are no particular optional
specifications) Tightening gland
IMPORTANT
4-2 IM 01E20D01-01E
4. WIRING
+ L/+ N/ -
SUPPLY
POWE
I
+-
D
+-
DIO
-
Drain valve
F0406.EPS
WARNING
F0405.EPS
Figure 4.1.5 Removing the Terminal Box Cover • Ensure that the power supply is OFF in order to
prevent electric shocks.
• Ensure the protective grounding terminal is
grounded before turning the power on.
• Use insulating sleeve crimp terminals (for 4-mm
screws) for the power supply wiring and protec-
tive grounding wiring.
4-3 IM 01E20D01-01E
4. WIRING
446 (1460)
+ L/+ N/ -
SUPPLY
POWE
200 ( 660)
I
0
20.4 22 24 26 28.8
+
Power supply
D
F0408.EPS
F0407.EPS
3) Setting Power Supply Frequency
Figure 4.1.7 Electric Cable Wiring
IMPORTANT
4-4 IM 01E20D01-01E
4. WIRING
For explosion proof type except TIIS, follow the 600 V vinyl-insulated cable
(2 mm2 or larger)
domestic electrical requirements as regulated in
each country.
IMPORTANT
Protective grounding
terminals In case grounding rings In case grounding rings are
are used. not used.
(Available only for metal piping)
F0410.EPS
F0409.EPS
(6) Connecting to External Instruments
Figure 4.1.8 Protective Grounding Terminal Location
WARNING
IMPORTANT
Before wiring with external instruments, be sure
Improper grounding can have an adverse effect to turn off the magnetic flowmeter and any
on the flow measurement. Ensure that the external instruments.
instrument is properly grounded.
Connect the AXF integral flowmeter terminal to • To avoid communication (BRAIN/ HART)
external instruments, giving attention to the following failure, it is recommended to use the shield
points. cable.
For FOUNDATION Fieldbus protocol, please refer to IM
01E20F02-01E.
For PROFIBUS PA protocol, please refer to IM NOTE
01E20F12-01E.
For pulse output from the DO terminals, param-
4 to 20 mA DC Current Output eters must be set. Refer to “Parameter Descrip-
tion” in the user’s manual of the AXF Integral
Flowmeter [Software Edition] (IM 01E20C02-
AXF integral flowmeter 01E).
l+
Receiver
l- Instrument
Protective diode
Resistive load max. 750 Ω. AXF integral flowmeter DO+
(When using BRAIN/ HART communication,
more than 250 Ω, less than 600 Ω). Mechanical Counter
PULSE OUT
DO-
F0411.EPS
F0412.EPS
shield cable or sufficiently reduce the input Figure 4.1.12 Active Pulse Output Connection
impedance of the electronic counter within the (Optional code EM)
electromagnetic flowmeter pulse output specifi-
cation range.
• The active pulse output (Optional code EM)
cannot be used in conjunction with the standard
pulse output.
• When the active pulse output (Optional code
EM) is selected, do not be short-circuit between
the DO+ and DO– terminals to avoid damaging
the instrument.
• When the active pulse output (Optional code
EM) is selected, the range of pulse rate must
be set to 2 pps maximum.
4-6 IM 01E20D01-01E
4. WIRING
Status Input
AXF integral flowmeter DO+ (or DIO+) Protective diode
IMPORTANT Load
DO- (or DIO-)
Status inputs are designed for use with no- This connection is not possible.
voltage (dry) contacts. Be careful not to connect
the status to any signal source carrying voltage.
AXF integral flowmeter DO+ (or DIO+)
Applying voltage may damage the input circuit. Relay
Electromagnetic
valve
DO- (or DIO-)
DIO- NOTE
No-voltage status input
For status and alarm outputs from the DO or
F0414.EPS
DIO terminals, parameters must be set. Refer to
Figure 4.1.13 Status Input Connection “Parameter Description” in the user’s manual of
the AXF Integral Flowmeter [Software Edition]
(IM 01E20C02-01E).
NOTE
IMPORTANT
4-7 IM 01E20D01-01E
4. WIRING
4.2 Wiring the Remote • Ground the remote flowtube and the converter
separately.
Flowtube • Cover each shield of the signal cable with vinyl
tube or vinyl tape to avoid contact between two
This section describes the wiring of the remote
shields or between a shield and a case.
flowtube only. For information relating to the wiring of
• When waterproof glands or union equipped
the converter, refer to the user’s manual of the
waterproof glands are used, avoid tightening
AXFA11 Magnetic Flowmeter Remote Converter (IM
the glands with an excessive torque.
01E20C01-01E) or the AXFA14 Magnetic Flowmeter
• Be sure to turn the power off before opening
Remote Converter (IM 01E20C02-01E).
the terminal box cover.
• Before turning the power on, tighten the
WARNING terminal box cover securely.
• The terminal box cover of size 2.5 mm to 1000
The wiring of the magnetic flowmeter must be
mm is locked by the special screw. In case of
performed by expert engineer or skilled person-
opening the terminal box cover, use the hex-
nel. No operator shall be permitted to perform
agonal wrench attached. For handling the
procedures relating to wiring.
locking screw, refer to Figure 4.2.8.
• Be sure to lock the cover of size 2.5 mm to
1000 mm by the special screw using the
hexagonal wrench attached after installing the
CAUTION
cover. For handling the locking screw, refer to
Once all wiring is complete, check the connec- Figure 4.2.20.
tions before applying power to the instrument. • Explosion protected types must be wired in
Improper arrangements or wiring may cause a accordance with specific requirement (and, in
unit malfunction or damage. certain countries, legal regulations) in order to
preserve the effectiveness of their explosion
protected features.
4.2.1 Wiring Precautions • When submersible type or optional code DHC
is selected, waterproof glands, signal and
Be sure to observe the following precautions when
excitation cables are attached.
wiring:
In order to preserve the effectiveness of
waterproof features, the terminal box cover and
CAUTION waterproof glands must not be detached from
flowmeter.
• In cases where the ambient temperature
exceeds 50°C (122°F), use external heat-
resistant wiring with a maximum allowable
temperature of 70°C (158°F) or above.
• Do not connect cables outdoors in wet weather
in order to prevent damage from condensation
and to protect the insulation, e.g. inside the
terminal box of the flowtube.
• Do not splice the cable between the flowtube
terminal and the converter if it is too short.
Replace the short cable with a cable that is the
appropriate length.
• All the cable ends must be provided with round
crimp-on terminals and be securely wired.
• The signal cables must be routed in separate
steel conduit tubes 16 (JIS C 8305) or flexible
conduit tubes 15 (JIS C 8309).
• Keep conduits or flexible tubes watertight using
sealing tape.
4-8 IM 01E20D01-01E
4. WIRING
4.2.2 Cables
CAUTION
(1) Dedicated Signal Cable (AXFC)
• As crimp terminals A, B, SA, SB and C have
Conductors (A and B) their own electrical potentials, securely insulate
Shields (SA and SB) them so as not to come in contact with one
Tape
another.
• To prevent a shield from coming in contact with
Outer jacket 10.5 (0.413") another shield or the case, cover each shield
with a vinyl tube or wrap it in vinyl tape.
Shield (C)
Insulation
Insulation
NOTE
F0417.EPS
B
A
50 (1.97)
SA
SB
C
60 (2.36)
90 (3.54)
90 (3.54)
C
70 (2.76)
1505
(5.9)
(5.9)
On the On the
converter flowtube
side side
10.5 (0.4)
L (Specified Dimensions)
20 (0.8)
AXFC
F0418.EPS
4-9 IM 01E20D01-01E
4. WIRING
EX1
EX2
EX2
EX1
IMPORTANT
On the converter side On the flowtube side
The wiring port is sealed with a cap (not water-
proof). Do not remove the cap from the unused
wiring port. If waterproof property is necessary,
please use waterproof glands.
F0419.EPS
(1) When waterproof property is unnecessary
Figure 4.2.3 End Treatment of Excitation Cable (When there are no particular optional
specifications)
NOTE
The wiring port is sealed with a cap (not water-proof)
that must be removed before wiring. At this time,
• For excitation cables, always use a crimp handle the wiring port in accordance with the JIS
terminal with an insulation cover. C0920 mentioned above. Do not remove the cap from
• Use crimp tools from the manufacturer of the the unused wiring port.
crimp terminal you want to use to connect the
crimp terminal and cable.
(2) When waterproof property is necessary
• Use crimp tools that are appropriate for the
(Wiring using waterproof glands)
diameter of the cable to be connected.
IMPORTANT
Washer
Gasket
Waterproof gland
Cable F0420.EPS
4-10 IM 01E20D01-01E
4. WIRING
WARNING
Washer
Tightening gland
(1) Removing Cover (Size 2.5 to 1000 mm (0.1
to 40 in.))
Cable Loosen the cover locking screw clockwise using a
hexagonal wrench (nominal size 3) to unlock the cover.
When working on conduit pipes or flexible pipes (PF1/2 only)
(Upon shipment from the manufacturing plant, the
F0421.EPS
cover is unlocked.) Hold the flowtube with your hand
Figure 4.2.5 Waterproof Gland with Union Joint and remove the cover by turning it in the direction of
(Optional code EU)
the arrow as shown below.
Gasket
Plastic gland
F0422.EPS
Terminal
Symbols Description
A
B Flow signal output
C
EX1
Excitation current input
EX2
Drain valve Protective grounding
(Outside of the terminal)
F0425.EPS
F0423.EPS
Figure 4.2.9 Terminal Configuration (General-Purpose
Figure 4.2.7 Conduit Wiring Use, Submersible Type, Sanitary Type, Size
2.5 to 400 mm (0.1 to 16 in.))
4-11 IM 01E20D01-01E
4. WIRING
F042401.EPS
FUSE
AXFA11 converter
2.5A 250V
N/– L/+ EX1 EX2 P+ P– SI1+ SI2+ COM
Figure 4.2.10 Terminal Configuration (Explosion proof POWER SUPPLY EXCIT ATION PULSE OUT STATUS IN
A EX2
EX2
Excitation cable
B
C
EX1 AXFC dedicated
EX2
Excitation current input signal cable SA
Protective grounding A
(Outside of the terminal)
F040212.EPS B
Figure 4.2.12 Terminal Configuration (General-purpose SB
Use, submersible Type, Size 1100 to 2600 Remote
Converter
mm (44 to 104 in.)) flowtube
SA Taping*
A A
A EX2 B B
SB Taping*
B C C
EX1 EX1
EX2 EX2
EX1
C * Individually tape and insulate the
Remote flowtube shields corresponding to SA and
SB on the remote flowtube side.
F0427.EPS
4-12 IM 01E20D01-01E
4. WIRING
(4) Wiring the Remote Flowtube (Explosion 2) Connection with the AXFA14 converter
Proof Type) with Converters In case of explosion proof type for ATEX, FM, CSA,
IECEx and TIIS certification, connect wiring as shown
IMPORTANT in the figure below.
In case of the explosion proof type, the protective
In case of TIIS Flameproof type, a remote
grounding of remote flowtube must be connected to
flowtube is available for combined use with
a suitable IS grounding system. In that case,
AXFA14 converter only.
(functional grounding terminal) need not be connected.
4-13 IM 01E20D01-01E
4. WIRING
(7) Grounding
I+ I– SO1+ SO2+ COM AL+ AL– C SA A B SB
CURRENT OUT STATUS OUT ALARM OUT SIGNAL
4-20mA out
FUSE
AXFA11 converter
2.5A 250V
N/– L/+
POWER SUPPLY
EX1 EX2
EXCIT ATION
P+ P–
PULSE OUT
SI1+ SI2+
STATUS IN
COM CAUTION
Shortbar
IMPORTANT
(6) Wiring the Remote Flowtube (General-
Purpose Use, Submersible Type, Size 1100 Improper grounding can have an adverse affect
to 2600 mm (44 to 104 in.)) with Convert- on the flow measurement. Ensure that the
ers instrument is properly grounded.
1) Connection with the AXFA11 converter
Connect wiring as shown in the figure below.
FUSE
AXFA11 converter
2.5A 250V
N/– L/+ EX1 EX2 P+ P– SI1+ SI2+ COM
POWER SUPPLY EXCIT ATION PULSE OUT STATUS IN
Shortbar
AXFC dedicated
signal cable
Excitation
cable Remote
Converter
flowtube
SA Taping*
A A
B B
SB Taping*
A EX2 C C
EX1 EX1
B EX2 EX2
* Individually tape and insulate
the shields corresponding to SA
EX1 and SB on the remote flowtube
C side.
Remote flowtube F040218.EPS
4-14 IM 01E20D01-01E
4. WIRING
The electromotive force of the magnetic flowmeter is (8) Installing the Cover (Size 2.5 to 1000 mm
minute and it is easy to be affected by noise. And also (0.1 to 40 in.))
that reference electric potential is the same as the Install the cover to the flowtube by turning it in the
measuring fluid potential. Therefore, the reference direction of the arrow as shown below. Tighten the
electric potential (terminal potential) of the flowtube cover locking screw counterclockwise using a hexago-
and the converter also need to be the same as the nal wrench (nominal size 3) to lock the cover.
measuring fluid. Moreover, that the potential must be
the same with ground. The magnetic flowmeter is
equipped with an grounding ring that makes a connec-
tion with the charge of the measured fluid for ground-
ing and protects the lining.
Be sure to ground according to Figure 4.2.19.
• Class D requirements (ground resistance. Figure 4.2.20 Installing the Terminal Box Cover (Remote
100 or less). Flowtube)
• TIIS Flameproof type: Class C requirements
(ground resistance, 10 or less) or Class A
requirements (ground resistance, 10 or less)
• Explosion proof type except TIIS: Domestic
electrical requirements as regulated in each
country.
4-15 IM 01E20D01-01E
5. MAINTENANCE
5. MAINTENANCE
WARNING
CAUTION
nance and be disassembled and reassembled (4) Using the hexagonal wrench, loosen the screw in
to the original state. the neck.
• The cover is locked by the special screw. In
(5) Turn the converter or the terminal box in the
case of opening the cover, use the hexagonal
desired direction.
wrench attached.
• Be sure to lock the cover by the special screw
using the hexagonal wrench attached after NOTE
installing the cover.
The converter and the terminal box can be
turned –140 degree to +180 degree from the
arrow mark indicating the flow direction. Do not
5.1 Changing Direction of exceed these angle.
Electrical Connection (6) Using the hexagonal wrench, retighten the neck
screw.
IMPORTANT
5-1 IM 01E20D01-01E
5. MAINTENANCE
5.2 Removing, Cleaning, (4) Using end A of the tool, open the electrode cover.
F9807SK
F0503.EPS
WARNING
5-2 IM 01E20D01-01E
5. MAINTENANCE
(6) Move aside the electrode lead wire and avoiding 5.2.2 Cleaning Replaceable Elec-
the screw, insert end A of the tool to the electrode trodes
and loosen the electrode holder using the torque
(1) Clean the electrode surface (wetted part) with
driver.
alcohol or other cleaning agents.
O-ring
Electrode surface
Electrode holde
End A
Torque driver
F0506.EPS
(7) Screw end B of the tool into the screw block of the
electrode, and pull the tool straight out to remove
the electrode holder.
F0508.EPS
NOTE
5-3 IM 01E20D01-01E
5. MAINTENANCE
5.2.3 Installing Replaceable Elec- (2) Using end A of the tool, tighten the electrode
trodes holder.
NOTE
NOTE
(1) Push the electrode holder with end B of the tool Using the torque wrench or torque driver, tighten
screwed into it, straight into the flowtube. the electrode holder to the following torque
values:
PFA lining:
8.0 N·m ±1 N·m (80 kgf·cm ±10 kgf·cm)
Polyurethane rubber lining:
3.0 N·m ±0.5 N·m (30 kgf·cm ±5 kgf·cm)
Electrode lead wire:
Red: Electrode A
White: Electrode B
WARNING
F0510.EPS
(3) Fill the interior of the flowtube with the fluid and
check that there is no leakage from the electrode.
NOTE
• Check that the O-ring has not come off. (4) Hold down the lead wire and terminal lug and
• Do not interchange electrodes A and B when screw the mounting screw into the electrode using
installing them. Install electrode A to the the Phillips screwdriver.
electrode boss with the red lead wire and
electrode B to the one with the white lead wire.
5-4 IM 01E20D01-01E
5. MAINTENANCE
(5) Put the lead wire in the electrode boss, and install
the electrode cover to the boss and tighten it using
end A of the tool.
Mini-flange
Gasket
Adapter
mounting screw
Electrode boss
F0316.EPS
CAUTION
(6) Turn on the power to the flowmeter and start
• In case of standard gasket (EPDM rubber),
normal operation.
tighten the adapter mounting screw according
to the torque values in Table 5.3.1.
• In case that optional code GH (Silicon rubber)
is selected, tighten the adapter mounting screw
5.3 Removing and In- according to the torque values in Table 5.3.2.
stalling Adapters for • Tighten the adapter mounting screw in diagonal
order step by step.
Sanitary Types • After tightening of screw, confirm that gaskets
protrude inside adapter. Protruding of gasket is
necessary to keep the sanitary requirements.
CAUTION
• This section describes how to remove and Table 5.3.1 Tightening Torque Values of adapter for
install a butt weld adapter as an example. EPDM rubber gasket
Apply the same procedure to clamp and union Tightening torque values (N-m / {kgf-cm} / [in-lbf])
adapters. Gasket type EPDM rubber
Size mm (inch) (standard)
• Turn off the power of the magnetic flowmeter,
15 (0.5), 25 (1.0), 32 (1.25), 40 (1.5) 3.0 to 3.5 / {30.59 to 35.69} / [26.55 to 30.98]
and confirm that there is no fluid inside the pipe
50 (2.0), 65 (2.5) 4.5 to 5.0 / {45.89 to 50.99} / [39.83 to 44.25]
before removing the instrument from the pipe 80 (3.0) 8.0 to 9.0 / {81.58 to 91.78} / [70.81 to 79.65]
line. 100 (4.0), 125 (5.0) 10 to 11 / {102.0 to 112.2} / [88.50 to 97.35]
• When installation, refer to section 3.3.6. T0502.EPS
5-5 IM 01E20D01-01E
5. MAINTENANCE
CAUTION
5-6 IM 01E20D01-01E
5. MAINTENANCE
1
2
IMPORTANT
Excercise care so as not to directly touch the Figure 5.4.1 Removing the Display Cover
electronic parts or circuit patterns on the board,
for example, by preventing static electrification 5.4.2.2 Changing the Display Unit Direction
90 Degrees
by using grounded wrist straps when handing
the assembly. (1) Hold the display unit with your hand and remove
the two mounting screws.
(2) Turn the display unit 90 degrees clockwise and
5.4.1 Fuse Replacement confirm the assembling position, taking care of the
connector and wire of the display unit.
CAUTION At this time, do not remove the connector.
(3) Secure the display unit using its two mounting
Please contact Yokogawa’s service office for fuse screws.
replacement. Also be sure to use the fuse that
was supplied by Yokogawa’s sales or service of-
fices.
5-7 IM 01E20D01-01E
5. MAINTENANCE
IMPORTANT
5.4.2.3 Installing the Cover Modification of the burnout direction must be carried
(1) Taking care not to entangle the cables, install the out using the setting switch from the amplifier’s CPU
cover to the flowmeter by turning it in the direc- board (i.e., Switch 1) (See Figure 5.5.1).
tion of the arrow as shown below.
Table 5.5.1 Output Setting Pins for Burnout
(2) Tighten cover locking screw 1 counterclockwise Position of Burnout Burnout
using a hexagonal wrench (nominal size 3) to lock Remarks
Pin Direction Output
the cover. Low High Set to High before
High 25 mA
shipment
Low High Set to Low for optional
Low 0 mA code C1
T0501.EPS
NOTE
1
2
5-8 IM 01E20D01-01E
5. MAINTENANCE
Low
Switch 1
High
2 ← Burnout setting switch
5.6 Regular Inspection
Switch 2
Enable
1 ← Write protect setting switch
Protect
Items
(1) Inspection of moisture-proofing inside the
terminal box: Once/year
(2) Retightening of piping joint screws: About
twice/year
(3) Inspection of electrodes and lining (in
case of adhesive and/or abrasive fluids,
etc.)
Determine the period of regular inspection as
necessary.
5-9 IM 01E20D01-01E
5. MAINTENANCE
Coil Circuit
Checking is possible even if the pipe is filled with
fluid.
Signal Circuit
Before testing, be sure to empty and dry the interior of
the pipe, checking that there is no adhesive material.
Also undo the wiring connection on the converter side
before testing.
5-10 IM 01E20D01-01E
5. MAINTENANCE
5.8 Troubleshooting
Although magnetic flowmeters rarely require maintenance, failures may occur when the
instrument is not operated correctly. This section describes troubleshooting procedures
where the cause of the breakdown is identified through receiver indication.
5.8.1 No Indication
START
Is anything else NO
being displayed?
YES
Does the
NO
arrow on the tube match the
flow direction?
Is the flowtube NO
full of fluid?
YES
Is foreign NO
matter adhering to
the walls?
YES
5-11 IM 01E20D01-01E
5. MAINTENANCE
START
YES
Are valves leaking? Replace the valve(s).
NO
5-12 IM 01E20D01-01E
5. MAINTENANCE
START
Are parameters NO
Set the parameters correctly.
set correctly?
YES
Is the NO
Completely ground the converter and
flowmeter completely flowtube (Refer to the section on
grounded? grounding in the relevant user's manual.)
YES
Does the NO
conductivity stay within the Use within the specified conductivity range.
specified range?
Signal terminals A-C and B-C: NO Dry the tube fully to restore insulation
100 M or more? resistance.
Excitation terminals EX1-C:
1 M or more?*1
5-13 IM 01E20D01-01E
6. OUTLINE
6. OUTLINE
■ STANDARD SPECIFICATIONS HART:
Refer to IM 01E20F02-01E for FOUNDATION Fieldbus Load Resistance:
communication type and IM 01E20F12-01E for 250 to 600Ω (including cable resistance)
PROFIBUS PA communication type marked with “”
Note: HART is a registered trademark of the HART
Converter (Integral flowmeter) Communication Foundation.
The contents of (*1) and (*2) described in the converter Data Security During Power Failure:
specifications are follows. Data (parameters, totalizer value, etc.) storage by
*1: Select two points from: one pulse output, one alarm EEPROM. No back-up battery required.
output, one status input, or two status outputs. Indicator (*2):
*2: For models without an indicator, the configuration tool Full dot-matrix LCD (32132 pixels)
(Such as HTT (handheld terminal) or FieldMateTM, etc.) Lightning Protector:
is necessary to set parameters. The lightning protector is built into the current output and
Excitation Method: pulse/alarm/status input and output terminals. When
• Standard dual frequency excitation: optional code A is selected, the lightning protector is
Size 2.5 to 400 mm (0.1 to 16 in.) built into the power terminals.
• Enhanced dual frequency excitation: Protection:
Size 25 to 200 mm (1.0 to 8.0 in.) General-purpose Use/Sanitary Type/TIIS Flameproof
(Optional code HF1 or HF2) type:
Input Signal (*1) “”: IP66, IP67
One Status Input: Dry contact Explosion proof type except TIIS:
Load Resistance: 200 Ω or less (ON), 100 kΩ or more In case of explosion proof type except TIIS, refer to
(OFF) description of "Enclosure" in "HAZARDOUS AREA
Output Signals “”: CLASSIFICATION".
• One Current Output: 4 to 20 mA DC (load resistance: Coating:
750Ω maximum, including cable resistance) Case and Cover: Corrosion-resistant coating
• One Pulse Output (*1): Coating Color; Mint green coating (Munsell 5.6 BG
Transistor contact output (open collector) 3.3/2.9 or its equivalent)
Contact capacity: 30 V DC (OFF), 200 mA (ON) Converter Material:
Output rate: 0.0001 to 10,000 pps (pulse/second) Case and Cover : Aluminum alloy
• One Alarm Output (*1):
Mounting/Shapes (Integral Flowmeter):
Transistor contact output (open collector)
• Electrical Connection:ANSI 1/2 NPT female
Contact capacity: 30 V DC (OFF), 200 mA (ON)
ISO M20 1.5 female
• Two Status Outputs (*1):
JIS G1/2 (PF1/2) female
Transistor contact output (open collector)
• Direction of Electrical Connection: The direction can be
Contact capacity: 30 V DC (OFF), 200 mA (ON)
changed even after delivery.
Communication Signals “”: • Terminal Connection: M4 size screw terminal
BRAIN or HART communication signal Grounding:
(Superimposed on the 4 to 20 mA DC signal) Grounding resistance 100 Ω or less
Distance from Power Line: 15 cm (6 in.) or more When optional code A is selected, grounding resis-
(Parallel wiring should be avoided.) tance 10 Ω or less shall be applied.
BRAIN: * In case of explosion proof type except TIIS, follow the
domestic electrical requirements as regulated in each
Communication Distance: country.
Up to 1.5 km (0.93 miles), when polyethylene insulated * In case of TIIS Flameproof type, refer to description of
PVC-sheathed cables (CEV cables) are used. “HAZARDOUS AREA CLASSIFICATION”.
Communication distance varies depending on the type
of cable and wiring used.
Load Resistance:
250 to 450Ω (including cable resistance)
Load Capacitance: 0.22 µF or less
Load Inductance: 3.3 mH or less
Input Impedance of Communicating Device:
10 kΩ or more (at 2.4 kHz)
6-1 IM 01E20D01-01E
6. OUTLINE
6-2 IM 01E20D01-01E
6. OUTLINE
6-3 IM 01E20D01-01E
6. OUTLINE
*1: AXF standard lay length dimensions for wafer type ceramics linings are the same as those for ADMAG ceramics linings. T21.EPS
*2: AXF standard lay length dimensions for union joint type ceramics linings are the same as those for ADMAG ceramics linings.
*3: Enhanced dual frequency excitation is not available for models with High grade accuracy.
*4: Not available with 32 mm (1.25 in.), 125 mm (5.0 in.)
*5: Not available with 125 mm (5.0 in.)
*6: Not available with 15 mm (0.5 in.), 125 mm (5.0 in.)
6-4 IM 01E20D01-01E
6. OUTLINE
Size of AXF Flowtubes: Replacement model for earlier ADMAG or ADMAG AE (Lay length code 2)
Unit: mm (in.)
High Grade Enhanced Dual Frequency
Use Process Connection Lining Remote Flowtube Integral Flowmeter Replaceable Electrode
Accuracy 0.2% of Excitation
(Electrode structure code 2)
Rate (Optional code HF1,HF2)
2.5 (0.1), 5 (0.2), 10 (0.4), 15 (0.5), 25 (1.0), 40 (1.5), 50 (2.0),
PFA 25 (1.0), 40 (1.5), 50 (2.0), 80 (3.0), — 80 (3.0), 100 (4.0), —
100 (4.0), 150 (6.0), 200 (8.0) 150 (6.0), 200 (8.0)
Wafer (*6)
25 (1.0), 40 (1.5), 50 (2.0), 80 (3.0), 25 (1.0), 40(1.5), 50 (2.0),
General- Polyurethane —
100(4.0), 150 (6.0), 200(8.0) 80(3.0), 100 (4.0), —
purpose use rubber
150 (6.0), 200 (8.0)
PFA 150 (6.0), 200 (8.0), 250 (10) — 150 (6.0), 200 (8.0) 150 (6.0), 200 (8.0), 250 (10)
Flange (*7) Polyurethane
150 (6.0), 200 (8.0), 250 (10) — 150 (6.0), 200 (8.0) 150 (6.0), 200 (8.0), 250 (10)
rubber
15 (0.5), 25 (1.0), 25 (1.0), 40 (1.5),
40 (1.5), 50 (2.0), 50 (2.0), 80 (3.0), 100 (4.0),
PFA — — —
80 (3.0), 100 (4.0), 150 (6.0), 200 (8.0)
150 (6.0), 200 (8.0)
Wafer (*6)
25 (1.0), 40 (1.5), 25 (1.0), 40 (1.5), 50 (2.0),
Submersible Polyurethane 50 (2.0), 80 (3.0), — — 80 (3.0), 100 (4.0), —
Type rubber 100 (4.0), 150 (6.0), 150 (6.0), 200 (8.0)
200 (8.0)
150 (6.0), 200 (8.0), — — 150 (6.0), 200 (8.0)
PFA —
250(10)
Flange (*7)
Polyurethane 150 (6.0), 200 (8.0), — — 150 (6.0), 200 (8.0) —
rubber 250 (10)
2.5 (0.1), 5 (0.2), 10 (0.4), 15 (0.5), 25 (1.0), 40 (1.5), 50 (2.0),
Explosion Wafer (*6) PFA 25 (1.0), 40 (1.5), 50 (2.0), 80 (3.0), — 80 (3.0), 100 (4.0), —
proof Type 100 (4.0), 150 (6.0), 200 (8.0) 150 (6.0), 200 (8.0)
Flange (*7) PFA 150 (6.0), 200 (8.0), 250 (10) — 150 (6.0), 200 (8.0) —
T22.EPS
*6: ADMAG lay length dimensions for wafer type of 250 mm (10 in.), and 300 mm (12 in.) are the same as those for AXF Standard.
And, in case of “platinum-iridium (grounding ring code P) or tantalum (grounding ring code T) or None (grounding ring code N)” in wafer type of 2.5 mm (0.1 in.)
to 15 mm (0.5 in.), the lay lengths of Replacement model are the same as those for AXF Standard.
*7: ADMAG lay length dimensions for flange type of 15 mm (0.5 in.) to 100 mm (4.0 in.), or 300 mm (12 in.) to 2600 mm (104 in.) are the same as those for AXF
Standard. However, in case of platinum-iridium (grounding ring code P) or tantalum (grounding ring code T) or None (grounding ring code N) in flange type of 15
mm (0.5 in.) to 100 mm (4.0 in.), the lay length of AXF Standard are longer by approx. 4mm (0.16 in) than those of earlier ADMAG or ADMAG AE.
6-5 IM 01E20D01-01E
6. OUTLINE
Submersible Type: Non-tar epoxy coating (black) Size 25 mm (1.0 in.) to 125 mm (5.0 in.)
6-6 IM 01E20D01-01E
6. OUTLINE
Size 150 mm (6.0 in.) to 400 mm (16 in.) Grounding Ring/Grounding Electrode:
• Grounding Ring(plate type)
Part Name Material Stainless steel-JIS SUS316 (AISI 316 SS/EN 1.4401
Housing Carbon steel-JIS SPCC equivalent),
equivalent
Stainless steel-JIS SUS304
Stainless steel-JIS SUS316L (AISI 316L SS/EN
Process Connection code: B** or SUSF304 (AISI 304 1.4404 equivalent),
Flange SS/EN 1.4301 equivalent) Hastelloy*1 C276 equivalent, Titanium,
Process Connection code: C** Carbon steel-JIS SS400 or
SFVC 2A Stainless steel-JIS SUS304 (AISI 304 SS/EN 1.4301
Wafer Type equivalent) for size 500 to 1000mm (20 to 40 in.) only,
Carbon steel-JIS SS400 or
PFA/Polyurethane Rubber/Natural Soft Rubber/ SS400 carbon steel lined with stainless steel-JIS
SFVC 2A
EPDM Rubber lining
Mini- SUS316 (AISI 316 SS/EN 1.4401 equivalent) for size
Flange Wafer Type Stainless steel-JIS SUS304
Ceramics lining [available with (AISI 304 SS/EN 1.4301 1100 to 2600mm (44 to 104 in.) only.
150 mm (6.0 in.), 200 mm (8.0 in.)] equivalent) • Grounding Electrode(electrode type)
Flange Type/Wafer Type Stainless steel-JIS SUS304 Fluorocarbon PFA lining + grounding electrode
PFA/Polyurethane Rubber/Natural Soft Rubber/ (AISI 304 SS/EN 1.4301 (Tantalum, Platinum-Iridium)
EPDM Rubber lining equivalent)
Pipe *1: Hastelloy is a registered trademark of Haynes
Wafer Type International Inc.
Ceramics lining [available with Alumina ceramics (99.9%) *2: Available with sizes 2.5 to 200mm (0.1 to 8.0 in.), PFA
150 mm (6.0 in.), 200 mm (8.0 in.)]
and ceramics linings only. However, the permeable
Terminal Box (Remote Flowtube) Aluminum alloy fluids (such as nitric acid, hydrofluoric acid, or sodium
T05.EPS hydroxide at high temperature) are unusable.
Electrode:
Stainless steel-JIS SUS316L (AISI 316L SS/EN 1.4404
equivalent), Hastelloy*1 C276 equivalent, Titanium,
Tantalum, Platinum-Iridium, Tungsten Carbide, Plati-
num-Alumina cermet(only for ceramics lining)
Note : For size over 500 mm and sanitary type, SUS316L
only.
6-7 IM 01E20D01-01E
6. OUTLINE
Gasket: Joints:
General-Purpose Use / Submersible Type /
Use Ceramics
Explosion proof Type Lining
Union Joints (size 10 mm or less)
PFA/Polyurethane Rubber/
Lining Natural Soft Rubber/ Ceramics weld joint
EPDM Rubber (GUW)
nut
grounding grounding
screw joint
ring ring (GUN, GUR)
Standard
Gasket within Flowtube
Standard Materials for Union Joint
Process Connection Code
No gasket within Flowtube Gasket within Flowtube Stainless steel
GUW: Union Joint (weld joint)
(JIS SUS316L
Gasket Material (within Flowtube)
Process Connection Code (ANSI 316L SS/EN
Fluororesin with 1.4404
— ceramic fillers GUN, GUR: union joint
(screw joint) equivalent))
(Valqua #7020)
grounding grounding Note: Contact Yokogawa office if PVC union joint is required.
ring ring T23-3.EPS
6-8 IM 01E20D01-01E
6. OUTLINE
Replacement model for earlier ADMAG or ADMAG AE: Standard High Grade
Size mm Flow Velocity Accuracy Flow Velocity Accuracy
(in.) V m/s (ft/s) (Calibration V m/s (ft/s) (Calibration
Process Available Electrode code B) code C)
Use Lining
Connection Size Material
2.5 (0.1) V 0.3 (1) 1.0 mm/s
JIS SUS316L
to 0.3 ! V ! 10 0.35% of —
General- 150 to 250 mm PFA/ (AISI 316L
15 (0.5)
Flange (1) (33) Rate
Purpose Use (6.0 to 10 in.) Polyurethane SS/EN
Rubber 1.4404 V 0.15 (0.5) 0.5 mm/s V 0.15 (0.5) 0.5 mm/s
equivalent)(*1) 25 (1.0)
T07.EPS 0.15 ! V 1 0.18% of Rate
to 0.15 ! V ! 10 0.35% of (0.5) (3.3) 0.2mm/s
200 (8.0) (0.5) (33) Rate
*1: If any other electrode materials are required, please 1 ! V !10 0.2% of Rate
contact Yokogawa office. (3.3) (33)
Mounting/Shapes (Remote Flowtube): V 0.15 (0.5) 0.5 mm/s
250 (10)
• Electrical Connection: ANSI 1/2 NPT female to 0.15 ! V ! 10 0.35% of —
ISO M20 1.5 female 400 (16) (0.5) (33) Rate
JIS G1/2 (PF1/2) female T08.EPS
Enhanced dual frequency excitation(Option code HF2):
• Direction of Electrical Connection: The direction can be Standard accuracy 1 mm/ s
changed even after delivery.
Note: In case of submersible types, an optional code DHC, Size 2.5 mm (0.1 in.) to 15 mm (0.5 in.)
and sizes of 1100 mm or larger, the direction can not % of Rate
1.2
be changed after delivery.
• Terminal Connection at Terminal Box: M4 size screw 1.0
0.8
Grounding:
Grounding resistance 100 Ω or less Pulse 0.6
High
Standard
gradeAccuracy
Accuracy
Output
* In case of explosion proof type except TIIS, follow the Accuracy 0.4
domestic electrical requirements as regulated in each 0.35
0.2
country.
* In case of TIIS Flameproof type, refer to description of 0
“HAZARDOUS AREA CLASSIFCATION”. 0 1 2 6 8 10
V[m/s]
F14.EPS
Combined Converter:
• A remote flowtube for sizes of up to 400 mm can be Size 25 mm (1.0 in.) to 400 mm (16 in.)
combined with the AXFA11 Converter or the AXFA14 % of Rate
Converter. If a combined converter is changed from 1.2
6-9 IM 01E20D01-01E
6. OUTLINE
Polyurethane Rubber /Natural Soft Rubber / Size 2200 mm (88 in.) to 2600 mm (104 in.)
EPDM Rubber Lining: % of Rate
2.4
6-10 IM 01E20D01-01E
6. OUTLINE
Fluid Conductivity:
Size 2.5 to 10 mm (0.1 to 0.4 in.): 5 µS/cm or larger
Size 15 to 125 mm (0.5 to 5 in.): 1 µS/cm or larger
Size 150 to 400 mm (6 to 16 in.): 3 µS/cm or larger
Note: For a fluid with large flow noise (pure water, pure
alcohol or others), or a fluid wth low conductivity and
low viscosity, the output fluctuates and is impossible
to measure accurately.
6-11 IM 01E20D01-01E
6. OUTLINE
Cable Length and Liquid Conductivity (Remote Measurable Flow Rate Range:
Flowtube): SI Units (Size: mm, Flow rate: m3/h)
Size 2.5 to 10 mm (0.1 to 0.4 in.) Size 0 to Min. Span Flow Rate 0 to Max. Span Flow Rate
(mm) (0.1 m/s) (10 m/s)
m (ft) Add 1% of Rate to the As show in accuracy
accuracy shown in description 2.5 0 to 0.0018 m3/h 0 to 0.1767 m3/h
accuracy description
Cable length
5 0 to 0.0071 0 to 0.7068
100 (330) Combined
with AXFA11 10 0 to 0.0283 0 to 2.8274
50 (170) or AXFA14
converter 15 0 to 0.0637 0 to 6.361
3 (10)
0 (0) 25 0 to 0.1768 0 to 17.671
1 5 10 20 50 (µS/cm)
Liquid conductivity 32 0 to 0.2896 0 to 28.952
converter
100 0 to 2.8275 0 to 282.74
100 (330) Combined with
AXFA14 125 0 to 4.418 0 to 441.7
converter
25 (85) 50 150 0 to 6.362 0 to 636.1
0 (0)
1 3 5 10 20 (µS/cm) 200 0 to 11.310 0 to 1,130.9
Liquid conductivity
250 0 to 17.672 0 to 1,767.1
Size 150 to 400 mm (6.0 to 16 in.) 300 0 to 25.447 0 to 2,544.6
Add 1% of Rate to As show in accuracy
the accuracy shown in 350 0 to 34.64 0 to 3,463
m (ft) description
accuracy description
Combined with 400 0 to 45.24 0 to 4,523
200 (660)
AXFA11
Cable length
converter
500 0 to 70.69 0 to 7,068
100 (330) Combined with 600 0 to 101.79 0 to 10,178
AXFA14
50 (170) converter 700 0 to 138.55 0 to 13,854
25 (85) 50
0 (0) 800 0 to 180.96 0 to 18,095
1 3 5 10 20 (µS/cm)
900 0 to 229.03 0 to 22,902
Liquid conductivity
F03.EPS 1000 0 to 282.75 0 to 28,274
Note: In case that size 250 or 300 mm (10 or 12 in.) is used T11.EPS
6-12 IM 01E20D01-01E
6. OUTLINE
English Units (Size: in., Flow rate: GPM) Fluid Temperature and Pressure:
0 to Max. Span Flow Rate Note *1 The following figures show maximum allowable
Size 0 to Min. Span Flow Rate
(33ft/s) fluid pressure for the flowtube. Further fluid
(in.) (0.33ft/s)
pressure should also be limitted according to
0.1 0 to 0.0078 GPM 0 to 0.7780 GPM flange rating.
*2 For fluid temperature of the explosion proof type,
0.2 0 to 0.0312 0 to 3.112 refer to descriptions of “HAZARDOUS AREA
0.4 0 to 0.1245 0 to 12.44 CLASSIFICATION”.
3.0 0 to 7.003 0 to 700.2 Pressure 250, 300 mm (10, 12 in.) (wafer type)
MPa (psi) 350, 400 mm (14, 16 in.) (flange type)
4.0 0 to 12.45 0 to 1,244
4 (580)
5.0 0 to 19.46 0 to 1,945 *1
64 0 to 9,561 0 to 318,689 *1: For wafer types of 32 mm to 300 mm(1.25 to 12 in.), and
for carbon steel flange types (process connection code:
72 0 to 12,101 0 to 403,341
C**) of 50 to 400 mm (2.0 to 16 in.), the minimum tem-
80 0 to 14,939 0 to 497,952 perature is –10°C (14°F).
*2: For fluid temperature of the explosion proof type, refer
88 0 to 18,076 0 to 602,522 to descriptions of “HAZARDOUS AREA CLASSIFI-
96 0 to 21,512 0 to 717,051 CATION”.
6-13 IM 01E20D01-01E
6. OUTLINE
General-Purpose Use, Remote Flowtube (electrode General-Purpose Use and Explosion proof Type,
structure code 2: replaceable electrode) Integral flowmeter (electrode structure code 1: Non-
replaceable electrode)
25 to 200 mm (1.0 to 8.0 in.) (flange type, wafer type)
250, 300 mm (10, 12 in.) (flange type) Pressure 2.5 to 50 mm (0.1 to 2.0 in.)
350, 400 mm (14 to 16 in.) (flange type) MPa (psi) 80 to 200 mm (3.0 to 8.0 in.)
Pressure 250, 300 mm (10, 12 in.) (wafer type)
4 (580)
MPa (psi)
3.25 (471)
2 (290)
1 (145) 2 (290)
– 0.1 (–14.5)
–10 0 40 100 130 160
(14) (32) (104) (212) (266) (320) –0.1(–14.5)
–10 0 100 130 180
Temperature °C (°F) F18-3.EPS (14) (32) (212)(266) (356)
Temperature °C (°F)
General-Purpose Use, Integral Flowmeter (electrode F19-2.EPS
structure code 2: replaceable electrode) *1: For fluid temperature of the explosion proof type, refer
to descriptions of “HAZARDOUS AREA CLASSIFI-
25 to 200 mm (1.0 to 8.0 in.) (flange type, wafer type) CATION”.
250, 300 mm (10, 12 in.) (flange type)
Pressure 350, 400 mm (14 to 16 in.) (flange type) Polyurethane Rubber Lining
MPa (psi)
250, 300 mm (10, 12 in.) (wafer type) General-Purpose Use and Submersible Type, Remote
2 (290) Flowtube (electrode structure code 1: Non-replace-
1 (145) able electrode)
25 to 50 mm (1.0 to 2.0 in.) (flange type, wafer type)
– 0.1 (–14.5)
–10 0 40 100 130
65 to 200 mm (2.5 to 8.0 in.) (flange type, wafer type)
(14) (32) (104) (212) (266)
250, 300 mm (10, 12 in.) (flange type)
Temperature °C (°F)
F18-4.EPS
250, 300 mm (10, 12 in.) (wafer type)
Note: For replaceable electrodes for fluid temperatures of 350 to 1000 mm (14 to 40 in.) (flange type)
Pressure
-10°C (14°F)or less, please contact Yokogawa office. MPa (psi) 1100 to 2600 mm (44 to 104 in.) (flange type)
4 (580)
Sanitary Type (electrode structure code 1: Non-
replaceable electrode)
2 (290)
15 to 125 mm (0.5 to 5.0 in.) (remote flowtube)
Pressure 1 (145)
15 to 125 mm (0.5 to 5.0 in.) (integral flowmeter) 0.75(109)
MPa (psi)
1(145) –0.1(–14.5)
–10 0 40
(14) (32) (104)
Temperature °C (°F) F20-1.EPS
–0.1(–14.5)
–10 0 130 160
(14) (32) (266) (320) General-Purpose Use, Integral Flowmeter (electrode
Temperature °C (°F) F18-5.EPS structure code 2: replaceable electrode)
Note: In case of 120 to 160°C (248 to 320°F) of fluid
25 to 200 mm (1.0 to 8.0 in.) (flange type, wafer type)
temperature, please select optional code GH. 250, 300 mm (10, 12 in.) (flange type)
2 (290)
–0.1 (–14.5)
–10 0 100 180
(14) (32) (212) (356)
Temperature °C (°F)
F19-1.EPS
6-14 IM 01E20D01-01E
6. OUTLINE
Natural Soft Rubber Lining Reasonable Figure for Thermal Shock of Creamics Lining:
General-Purpose Use and Submersible Type, Remote Size 2.5 to 25 mm (0.1 to 1.0 in.)
Flowtube (electrode structure code 1: Non-replace- ∆T°C(°F)
able electrode) 120 (248)
Increase
50 mm (2.0 in.) (flange type, wafer type) 100 (212)
65 to 200 mm (2.5 to 8.0 in.) (flange type, wafer type)
250, 300 mm (10, 12 in.) (flange type)
Pressure
250, 300 mm (10, 12 in.) (wafer type)
MPa (psi) Decrease
350, 400 mm (14 , 16 in.) (flange type)
4(580) 60 (140)
6-15 IM 01E20D01-01E
6. OUTLINE
EX1
C
C
EX1
Centering device (wafer type only): 1 pc.
EX2
B
B
EX2
Hexagonal wrench: 2 pcs.
A
A
■ TERMINAL CONFIGURATION AND size 2.5 to 400 mm (0.1 to 16 in.) (Only for Explosion proof type)
TERMINAL WIRING
Integral Flowmeter “”
EX1
C
C
EX1
EX2
Terminal configuration
B
B
EX2
A
A
+ L/+ N/ -
SUPPLY
POWER
size 500 to 1000 mm (20 to 40 in.) size 1100 to 2600 mm (44 to 104 in.)
I
Terminal wiring
+
DO
Terminal
-
Symbols Description
+
DIO
A
-
6-16 IM 01E20D01-01E
6. OUTLINE
6-17 IM 01E20D01-01E
6. OUTLINE
*1: For a wafer type of 2.5 to 10 mm (0.1 to 0.4 in.), prepare 15 mm (0.5 in.) diameter nominal flanges on the process pipe side.
(Process connection codes: AA1, AA2, AD4, AJ1, and AJ2)
*2: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection code: AD4) because there is no
difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection code: AD2) because there is no difference
in the dimensions of the mating faces. 1 2 N
*3: Mating dimensions are based on standards as follow:
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*4: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*5: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is available
only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11. Available
only for JIS G1/2 female electrical connections.
*6: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*7: In case of integral flowmeters of the TIIS flameproof type, select “with indicator”(code 1 or 2).
*8: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*9: For FOUNDATION Fieldbus protocol, refer to IM 01E20F02-01E
*10: For PROFIBUS PA protocol, refer to IM 01E20F12-01E.
*11: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*12: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
6-18 IM 01E20D01-01E
6. OUTLINE
6-19 IM 01E20D01-01E
6. OUTLINE
*1: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection code : AD4)
because there is no difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection code : AD2) because there is no difference
in the dimensions of the mating faces.
*2: Mating dimensions are based on standards as follow: 1 2 N
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*3: AXF standard lay length dimension for ceramics linings are the same as
those for ADMAG ceramics linings.
*4: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*5: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is
available only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11.
Available only for JIS G1/2 female electrical connections.
*6: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*7: In case of integral flowmeters of the TIIS flameproof type, select “with indicator”(code 1 or 2).
*8: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*9: For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*10: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*11: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*12: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
6-20 IM 01E20D01-01E
6. OUTLINE
AXF STANDARD (Flange Type) Size 2.5 mm (0.1 in.) to 400 mm (16 in.)
General-purpose Use/Submersible Type/Explosion proof Type, PFA/Polyurethane Rubber /Natural Soft Rubber/EPDM
Rubber Lining
Model Suffix Code Description Applicable Model
AXF002 ···················· Size 2.5 mm (0.1 in.) Integral Flowmeter/Remote Flowtube
AXF005 ···················· Size 5 mm (0.2 in.) Integral Flowmeter/Remote Flowtube
AXF010 ···················· Size 10 mm (0.4 in.) Integral Flowmeter/Remote Flowtube
AXF015 ···················· Size 15 mm (0.5 in.) Integral Flowmeter/Remote Flowtube
AXF025 ···················· Size 25 mm (1.0 in.) Integral Flowmeter/Remote Flowtube
AXF032 ···················· Size 32 mm (1.25 in.) Integral Flowmeter/Remote Flowtube
AXF040 ···················· Size 40 mm (1.5 in.) Integral Flowmeter/Remote Flowtube
AXF050 ···················· Size 50 mm (2.0 in.) Integral Flowmeter/Remote Flowtube
AXF065 ···················· Size 65 mm (2.5 in.) Integral Flowmeter/Remote Flowtube
AXF080 ···················· Size 80 mm (3.0 in.) Integral Flowmeter/Remote Flowtube
AXF100 ···················· Size 100 mm (4.0 in.) Integral Flowmeter/Remote Flowtube
AXF125 ···················· Size 125 mm (5.0 in.) Integral Flowmeter/Remote Flowtube
AXF150 ···················· Size 150 mm (6.0 in.) Integral Flowmeter/Remote Flowtube
AXF200 ···················· Size 200 mm (8.0 in.) Integral Flowmeter/Remote Flowtube
AXF250 ···················· Size 250 mm (10 in.) Integral Flowmeter/Remote Flowtube
AXF300 ···················· Size 300 mm (12 in.) Integral Flowmeter/Remote Flowtube
AXF350 ···················· Size 350 mm (14 in.) Integral Flowmeter/Remote Flowtube
AXF400 ···················· Size 400 mm (16 in.) Integral Flowmeter/Remote Flowtube
Use G · · · · · · · · · · · · · · · · · · · · · General-Purpose Use
W · · · · · · · · · · · · · · · · · · · · · Submersible Type Size 15 mm (0.5 in.) to 400 mm (16 in.), Remote Flowtube only
C · · · · · · · · · · · · · · · · · · · · · Explosion proof Type (*6) PFA lining only
Converter -D · · · · · · · · · · · · · · · · · Integral Flowmeter with 4 to 20 mA DC Output and BRAIN Communication
Output Signal -E · · · · · · · · · · · · · · · · · Integral Flowmeter with 4 to 20 mA DC Output and HART Communication
and -F · · · · · · · · · · · · · · · · · Integral Flowmeter with Digital communication
Communication (FOUNDATION Fieldbus protocol) (*10)
-G · · · · · · · · · · · · · · · · · Integral Flowmeter with Digital communication
(PROFIBUS PA protocol) (*11)
-N · · · · · · · · · · · · · · · · · Remote Flowtube for Combined Use with AXFA11
-P · · · · · · · · · · · · · · · · · Remote Flowtube for Combined Use with AXFA14
Power Supply 1 ················ Integral Flowmeter, 100 V to 240 V AC or 100 to 120 V DC
2 ················ Integral Flowmeter, 24 V AC/DC
N ················ Remote Flowtube
Lining (*9) A ·············· Fluorocarbon PFA
U ·············· Polyurethane Rubber Size 25 mm (1.0 in.) to 400 mm (16 in.)
D ·············· Natural Soft Rubber Size 50 mm (2.0 in.) to 400 mm (16 in.)
G ·············· EPDM Rubber Size 50 mm (2.0 in.) to 400 mm (16 in.)
Electrode Material (*9) L · · · · · · · · · · · · · JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
P ············· Platinum-iridium PFA lining only
H ············· Hastelloy C276 Equivalent
T ············ Tantalum PFA lining only
V ············· Titanium
W ············· Tungsten Carbide Size 15 mm (0.5 in.) to 400 mm (16 in.),
PFA/Polyurethane Rubber lining only
Electrode Structure 1 · · · · · · · · · · · · · Non-replaceable
2 · · · · · · · · · · · · · Replaceable General-Purpose use, Size 25 mm (1.0 in.) to 400 mm (16 in.)
PFA/Polyurethane Rubber lining only
Electrode Material: JIS SUS316L only
Grounding Ring and N ··········· None
Grounding Electrode S ··········· JIS SUS316 (AISI 316 SS/EN 1.4401 Equivalent)
Material (*9) L ··········· JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
P ··········· Platinum-iridium Size 2.5 mm (0.1 in.) to 200 mm (8.0 in.), PFA lining only
H ··········· Hastelloy C276 Equivalent
T ··········· Tantalum Size 2.5mm (0.1 in.) to 200 mm (8.0 in.), PFA lining only
V ··········· Titanium
Process Connection -BA1 · · · · · · · ANSI Class 150 Flange (Stainless Steel)(*1) Size 2.5 mm (0.1 in.) to 400 mm (16 in.)
(*4)(*12) -BA2 · · · · · · · ANSI Class 300 Flange (Stainless Steel)(*1) Size 2.5 mm (0.1 in.) to 300 mm (12 in.)
-BD1 · · · · · · · DIN PN 10 Flange (Stainless Steel)(*2) Size 200 mm (8.0 in.) to 400 mm (16 in.)
-BD2 · · · · · · · DIN PN 16 Flange (Stainless Steel)(*2) Size 65 mm (2.5 in.) to 300 mm (12 in.)
-BD4 · · · · · · · DIN PN 40 Flange (Stainless Steel)(*1)(*2) Size 2.5 mm (0.1 in.) to 50 mm (2.0 in.)
-BJ1 · · · · · · · JIS 10K Flange (Stainless Steel)(*1) Size 2.5 mm (0.1 in.) to 400 mm (16 in.)
-BJ2 · · · · · · · JIS 20K Flange (Stainless Steel)(*1) Size 2.5 mm (0.1 in.) to 300 mm (12 in.)
-BG1 · · · · · · · JIS F12 (JIS75M) Flange (Stainless Steel) Size 80 mm (3.0 in.) to 400 mm (16 in.)
-CA1 · · · · · · · ANSI Class 150 Flange (Carbon Steel) Size 50 mm (2 in.) to 400 mm (16 in.), Not available for TIIS flameproof type
-CA2 · · · · · · · ANSI Class 300 Flange (Carbon Steel) Size 50 mm (2 in.) to 300 mm (12 in.), Not available for TIIS flameproof type
-CD1 · · · · · · · DIN PN 10 Flange (Carbon Steel)(*2) Size 200 mm (8.0 in.) to 400 mm (16 in.), Not available for TIIS flameproof type
-CD2 · · · · · · · DIN PN 16 Flange (Carbon Steel)(*2) Size 65 mm (2.5 in.) to 300 mm (12 in.), Not available for TIIS flameproof type
-CD4 · · · · · · · DIN PN 40 Flange (Carbon Steel)(*2) Size 50 mm (2.0 in.) only, Not available for TIIS flameproof type
-CJ1 · · · · · · · JIS 10K Flange (Carbon Steel) Size 50 mm (2.0 in.) to 400 mm (16 in.), Not available for TIIS flameproof type
-CJ2 · · · · · · · JIS 20K Flange (Carbon Steel) Size 50 mm (2.0 in.) to 300 mm (12 in.), Not available for TIIS flameproof type
-CG1 · · · · · · · JIS F12 (JIS75M) Flange (Carbon Steel) Size 80 mm (3.0 in.) to 400 mm (16 in.), Not available for TIIS flameproof type
-DD4 · · · · · · · DIN PN 40 Flange (Stainless Steel), DN10(*2)(*3)" Size 2.5 mm (0.1 in.) to 10 mm (0.4 in.)
-DJ1 · · · · · · · JIS 10K Flange (Stainless Steel ), 10 mm Diameter Nominal (*3) Size 2.5 mm (0.1 in.) to 10 mm (0.4 in.)
-DJ2 · · · · · · · JIS 20K Flange (Stainless Steel ), 10 mm Diameter Nominal (*3) Size 2.5 mm (0.1 in.) to 10 mm (0.4 in.)
Lay Length 1······· Standard
Electrical Connection (*7) -0 · · · · JIS G1/2 female
-2 · · · · ANSI 1/2 NPT female Not available for Submersible Type
-4 · · · · ISO M201.5 female Not available for Submersible Type
Indicator (*5)(*8) 1 · · · Integral Flowmeter with indicator (Horizontal)
2 · · · Integral Flowmeter with indicator (Vertical)
N · · · Integral Flowmeter without indicator /Remote Flowtube
Calibration B · · · Standard
C · · · High Grade (*13) Size 25 mm (1.0 in.) to 200 mm (8.0 in.), PFA lining only
/ Optional code (See the Table of Optional Specifications)
T17.EPS
6-21 IM 01E20D01-01E
6. OUTLINE
*1: For a flange type of 2.5 to 10 mm (0.1 to 0.4 in.), prepare 15 mm (0.5 in.) diameter nominal flanges on the process pipe side.
(Process connection codes: BA1, BA2, BD4, BJ1, and BJ2)
*2: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection codes: BD4,CD4 and DD4) because
there is no difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection codes: BD2 , CD2) because there is no
difference in the dimensions of the mating faces.
*3: For a flange type of 2.5 to 10 mm (0.1 to 0.4 in)(Process connection codes: DJ1, DJ2, and DD4), prepare 10 mm (0.4 in.) diameter nominal flanges on the
process pipe side.
*4: Mating dimensions are based on standards as follow: 1 2 N
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*5: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*6: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is
available only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or
G11. Available only for JIS G1/2 female electrical connections.
*7: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*8: In case of integral flowmeters of the TIIS flameproof type, select “with indicator”(code 1 or 2).
*9: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150°C [302°F] or above).
Contact Yokogawa for detailed information of the wetted parts material.
*10:For FOUNDATION Fieldbus protocol, refer to IM 01E20F02-01E
*11: For PROFIBUS PA protocol, refer to IM 01E20F12-01E.
*12: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*13: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
6-22 IM 01E20D01-01E
6. OUTLINE
AXF STANDARD (Flange Type) Size 500 mm (20 in.) to 2600 mm (104 in.)
General-purpose Use/Submersible Type, Polyurethane Rubber Lining
6-23 IM 01E20D01-01E
6. OUTLINE
*4: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*5: For FOUNDATION Fieldbus protocol, refer to IM 01E20F02-01E
*6: For PROFIBUS PA protocol, refer to IM 01E20F12-01E.
*7: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*8: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
6-24 IM 01E20D01-01E
6. OUTLINE
6-25 IM 01E20D01-01E
6. OUTLINE
*1: For a wafer type of 2.5 to 10 mm (0.1 to 0.4 in.), prepare 15 mm (0.5 in.) diameter nominal flanges on the process pipe side.
(Process connection codes: AA1, AA2, AD4, AJ1, and AJ2)
*2: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection code: AD4) because there is no
difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection code: AD2) because there is no difference
in the dimensions of the mating faces. 1 2 N
*3: Mating dimensions are based on standards as follow:
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*4: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*5: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is available
only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11. Available
only for JIS G1/2 female electrical connections.
*6: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*7: In case of integral flowmeters of the TIIS flameproof type, select “with indicator”(code 1 or 2).
*8: In case of platinum-iridium (grounding ring code P) or tantalum (grounding ring code T) or None (grounding ring code N) in wafer type of 2.5 mm (0.1 in.) to 15
mm (0.5 in.), the lay lengths of Replacement model are the same as those for AXF Standard.
In this case, AXF Standard shall be selected.
*9: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*10:For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*11: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*12: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
6-26 IM 01E20D01-01E
6. OUTLINE
6-27 IM 01E20D01-01E
6. OUTLINE
*1: Even when DIN PN10 is required for a 150 (6.0 in.)-mm model, select PN16 (Process connection code: CD2) because there is no difference in the dimensions of the
mating faces.
*2: Mating dimensions are based on standards as follow:
ANSI: ASME B 16.5, DIN:DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*3: N shall be always selected for remote flowtubes.
In the case of an integral, select from among the following figures:
1 2 N
*4: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is available only for
combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11. Available only for JIS G1/2
female electrical connections.
*5: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*6: In case of integral flowmeters of the TIIS flameproof type, select “with indicator”(code 1 or 2).
*7: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is also
possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature steam
(150°C [302°F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*8: For FOUNDATION Fieldbus protocol, refer to IM 01E20F02-01E
*9: For PROFIBUS PA protocol, refer to IM 01E20F12-01E.
*10: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
6-28 IM 01E20D01-01E
6. OUTLINE
G
D
N
E
AXF***G- P
AXF***C- P
AXF***W- P
AXF***H- P
AXF***G- F
AXF***C- F
AXF***H- F
For District Urethane resin potting is applied in the terminal box of a remote flowtube.
Heating and Select JIS G1/2 for the electrical connections. 30-meter signal and excitation
– – – – – DHC
Cooling or cables are pre-wired and waterproof grands with union joints are attached at
Condensation-proof factory.
User-specified Available for the submersible type and a model with optional code DHC. The
Signal and Excitation cable length is limited up to 200 meters when combined with an AXFA11
Cable Length converter, or 100 meters when combined with an AXFA14 converter.
Following “L,” specify the cable length in three digits as a multiple of 1 meter – – – – L***
(e.g., 001, 002, or 005) for a length up to 5 meters or as a multiple of 5
meters (i.e., 005, 010, 015, or the like) for a length of 5 meters or more. If this
optional code is not selected, a 30-meter length cable is attached.
100 ms
Mass Unit Setting The flow rate span, transmission pulse weight, and totalizer display pulse
weight can be set in terms of mass unit. Specify the density of the process
fluid when ordering in addition to the mass flow rate span, transmission pulse
wight (for mass unit), and totalizer display pulse weight (for mass unit).
When ordering a remote flowtube, parameters for 'Mass Unit Setting' will be
set in the corresponding converter before shipment.
1. Density
a. Available density Numerics:
Specify the numeric within the value of 500 to 2000 kg/m3, 4.2 to
16.7 lb/gal, or 31.2 to 124.8 lb/cf. And it can be up to five digits, to a
maximum of 32000 ignoring the decimal point. A fraction is limited to
the fourth decimal place.
b. Available density units: kg/m3, lb/gal, lb/cf
Example: A water density is about 1000kg/m3. In this case specify
“1000kg/m3”.
MU
However a density is changed by temperature.
Specify the actual density.
(The 1000kg/m3 is equivalent to 8.345lb/gal and 62.43lb/cf.)
2. The mass flow rate span, transmission pulse weight, and totalizer display
pulse weight
a. Available density Numerics:
Specify the numeric within the value of 0.0001 to 32000. And it can
be up to five digits, to a maximum of 32000 ignoring the decimal
point. A fraction is limited to the fourth decimal place.
b. Mass Units Available mass units: t, kg, g, klb, lb
Available time units: /d, /h, /min, /s
Note1: In case of specifying the mass flow span, calculate the volumetric flow
span by the setting density, and specify the available value in the mass
flow span.
Note2: In case of transmission pulse weight and totalizer display pulse weight,
specify the mass unit which was specified as the flow unit.
T26-1.EPS
6-29 IM 01E20D01-01E
6. OUTLINE
Table of Optional Specifications (Size 2.5 mm (0.1 in.) to 400 mm (16 in.)) (continued)
: Available –: Not available
Applicable Model
General Explosion proof Submersible Sanitary
Integral Remote Integral Remote Remote Integral Remote
Item Specifications Flowmeter Flowtube Flowmeter Flowtube Flowtube Flowmeter Flowtube Code
G
D
D
N
N
E
E
AXF***G- P
AXF***C- P
AXF***W- P
AXF***H- P
AXF***G- F
AXF***C- F
AXF***H- F
G3/4 Female Waterproof glands for G3/4 conduits or flexible tubes are attached to the
– – – – – EW
Waterproof Glands electrical connections. Available only for JIS G1/2 female electric connections.
Waterproof Glands Waterproof glands are attached to the electrical connections. Available
– – – EG
only for JIS G1/2 female electric connections.
Waterproof Glands Waterproof glands with union joints are attached to the electrical
– – – EU
with Union Joints connections. Available only for JIS G1/2 female electric connections.
Plastic Glands Plastic glands are attached to the electrical connections. Available only for
– – – EP
JIS G1/2 female electric connections.
Mirror Finished PFA Mirror finishing on the PFA lining inside of the tube to the smoothness
Lining lining. Available for 15 mm (0.5 in.) and larger sizes.
The Ra is average of measured values on several point. PM
Size 15 to 200 mm (0.5 to 8.0 in.) : Ra 0.05 to 0.15 µm
Size 250 to 400 mm (10 to 16 in.) : Ra 0.05 to 0.25 µm
Mirror Finished Mirror finishing on the inside of the ceramics tube to Ra # 0.1 µm.
Ceramics Available for 5 mm (0.2 in.) and larger sizes. – – – CM
The Ra is average of measured values on several point.
Stainless Steel Tag A pendant tag plate of JIS SUS304 (AISI 304 SS/EN 1.4301 equivalent) is
Plate provided. Choose this option when a pendant tag plate is required in
SCT
addition to the standard nameplate with the tag number inscribed on it.
Dimension (Height Width): Appr. 12.5 (4.92) 40 (15.7) mm (inch)
Direction change of $90 degrees rotated converter (or terminal box) to change the direction of
RA
the electrical the electrical connection.
connection (*1) $180 degrees rotated converter (or terminal box) to change the direction of
RB
the electrical connection.
%90 degrees rotated converter (or terminal box) to change the direction of
RC
the electrical connection.
Bolts, Nuts, and Bolts, nuts, and Bolts and nuts: Carbon steel;
– – BCC
Gaskets (*2) gaskets are Gaskets: Chloroprene rubber (*3)
provided for wafer Bolts and nuts: Carbon steel;
– – BCF
connections. Gaskets: PTFE-sheathed non-asbestos (*4)
Available only for Bolts: JIS SUS304 (AISI 304 SS stainless steel
ANSI 150, JIS10K, equivalent);
or, JIS20K wafer Nuts: JIS SUS403 (AISI 403SS stainless steel – – BSC
connections. equivalent) ;
Gaskets: Chloroprene rubber (*3)
Bolts: JIS SUS304 (AISI 304 SS stainless steel
equivalent);
Nuts: JIS SUS403 (AISI 403SS stainless steel – – BSF
equivalent) ;
Gaskets: PTFE-sheathed non-asbestos (*4)
Special Gaskets (*5) Viton® gaskets for use with a PFA or ceramics lining with PVC piping.
Allowable temperature and pressure are equivalent to Valqua #4010,
– – GA
special fluororubber not mixed. Available for 2.5 mm (0.1 in.) to 200 mm
(8.0 in.) of PFA lining or 15 to 200 mm (0.5 to 8 in.) sizes of ceramics lining.
Acid-resistant Viton® gaskets for use with a PFA or ceramics lining with PVC
piping. Allowable temperature and pressure are equivalent to Valqua #4010,
– – GC
special fluororubber mixed (mixing #D2470). Available for 2.5 mm (0.1 in.) to 200
mm (8.0 in.) of PFA lining or 15 to 200 mm (0.5 to 8 in.) sizes of ceramics lining.
Alkali-resistant Viton® gaskets for use with a PFA or ceramics lining with PVC
piping. Allowable temperature and pressure are equivalent to Valqua #4010,
– – GD
special fluororubber mixed (mixing #RCD970). Available for 2.5 mm (0.1 in.) to 200
mm (8.0 in.) of PFA lining or 15 to 200 mm (0.5 to 8 in.) sizes of ceramics lining.
Alkali-resistant carbonized fluororesin gaskets for use with a ceramics lining with
– – – GF
metal piping. Allowable temperature and pressure are equivalent to Valqua #7026.
Silicon rubber gaskets for Sanitary Type, provided between the lining and
– – – – – GH
the adapter. For the condition of fluid temp. 120 to 160°C (248 to 320°F).
Oil-prohibited Use Electrodes, linings, and grounding rings are assembled and packed with
polyethylene after being cleaned with water and acetone and dried with air. – – – K1
The label ‘OIL FREE’ is affixed.
T26-2.EPS
6-30 IM 01E20D01-01E
6. OUTLINE
Table of Optional Specifications (Size 2.5 mm (0.1 in.) to 400 mm (16 in.)) (continued)
: Available –: Not available
Applicable Model
General Explosion proof Submersible Sanitary
Integral Remote Integral Remote Remote Integral Remote
Item Specifications Flowmeter Flowtube Flowwmeter Flowtube Flowtube Flowmeter Flowtube Code
G
D
N
E
AXF***G- P
AXF***C- P
AXF***W- P
AXF***H- P
AXF***G- F
AXF***C- F
AXF***H- F
Oil-prohibited Use with Electrodes, linings, and grounding rings are assembled and packed with
Dehydrating Treatment polyethylene including desiccants after being cleaned with water and – – – K5
acetone and dried with air. The label ‘OIL & WATER FREE’ is affixed.
Painting Color Change Coated in black (Munsell N1.5 or its equivalent.) – P1
Coated in jade green (Munsell 7.5 BG 4/1.5 or its equivalent.) – P2
Coated in metallic silver. – P7
Epoxy Resin Coating Epoxy resin coating which has alkali-resistance instead of standard
– – – X1
polyurethane resin coating. The color is same as standard type.
High Anti-corrosion Three-layer coating (polyurethane coating on two-layer epoxy resin
Coating coating) in the same range as that for the standard coating. The color is – – – X2
same as standard type. Salt/alkali/acid/weather-resistance.
Material Certificate Reproduced material certificates for :
PFA/polyurethane: Pipe, electrodes, grounding rings/grounding
M01
electrodes, flanges or mini flanges, adapters (for sanitary type)
Ceramics: only grounding rings or grounding electrodes or union joint
Hydrostatic Test The test verifies the absence of leaks by applying the following water
pressures (which are determined under process connection conditions) to
linings for ten minutes. Test results are described in the Note column of a
test certificate (QIC).
Process Connection: Water Pressure: – – T01
ANSI Class 150, DIN PN10, JIS 10K 1.5 MPa
ANSI Class 300, DIN PN16, JIS 20K 3.0 MPa
DIN PN40, Union joint (Ceramics lining) 6.0 MPa
JIS F12 1.25 MPa
Calibration Certificate Level 2: The Declaration and the Calibration Equipment List are issued. L2
Level 3: The Declaration and the Primary Standard List are issued. L3
Level 4: The Declaration and the Yokogawa Measuring Instruments Control
L4
System are issued.
Vent Hole With a vent hole provided for permeable fluids (such as nitric acid,
hydrofluoric acid, or sodium hydroxide at high temperature). – – – H
Available only for a PFA lining flange type tube.
Enhanced Dual Available for 25 to 200 mm (1.0 to 8.0 in.) sizes.
Frequency Excitation Products are delivered with the Standard dual frequency excitation mode
(*6) and the Enhanced dual frequency excitation mode enabled.
Although the meter factors for the Standard dual frequency excitation mode
HF1
and the Enhanced dual frequency excitation mode are inscribed on the
nameplate, the flow calibration of optional code HF1 is not performed in
Enhanced dual frequency excitation.Excitation mode select optional code
HF2 when an accurate flow measurement is required.
Available for 25 to 200 mm (1.0 to 8.0 in.) sizes.
Products are delivered with the Standard dual frequency excitation mode
and the Enhanced dual frequency excitation mode enabled.
The meter factor for the Enhanced dual frequency excitation obtained by HF2
flow calibration is inscribed on the nameplate and set into the combined
converter in addition to the meter factor for the Standard dual frequency
excitation.
T26-3.EPS
6-31 IM 01E20D01-01E
6. OUTLINE
Table of Optional Specifications (Size 2.5 mm (0.1 in.) to 400 mm (16 in.)) (continued)
: Available –: Not available
Applicable Model
General Explosion proof Submersible Sanitary
Integral Remote Integral Remote Remote Integral Remote
Item Specifications Flowmeter Flowtube Flowmeter Flowtube Flowtube Flowmeter Flowtube Code
G
D
N
E
AXF***G- P
AXF***C- P
AXF***W- P
AXF***H- P
AXF***G- F
AXF***C- F
AXF***H- F
Five-point Calibration A flow test near 0, 25, 50, 75, and 100% of the user-specified span is
in User-specified performed instead of the flow test of the standard 2m/s span and a test
Span certificate (QIC) is submitted. Specify the span (100% flow span) whose
corresponding flow velocity lies between 0.5 to 10 m/s (0.8 to 10m/s for
sizes 32, 65, and 125mm) and that is less than the maximum line capacity.
Selectable range of flow rate span is showing below.
*1: $90-degree rotation $180-degree rotation %90-degree rotation *2: When specifying the optional code BCC or BSC for a PFA
Standard
Optional Code RA Optional Code RB Optional Code RC or ceramics lining, it is advisable to specify the optional
code GA, GC, or GD at the same time to prevent potential
Electrical Indicator Electrical Indicator
Integral Flowmeter Connection Connection leakage caused by the difference in elasticity between the
flowtube and chloroprene gaskets. Refer to description of
“Gasket” in the “Wetted Part Material”.
Electrical Electrical
Remote Flowtube Connection Connection
*3: Allowable temperature and pressure with the optional code BCC or BSC (only for Gaskets:Chloroprene rubber) are equivalent to Valqua #2010.
*4: Allowable temperature and pressure with the optional code BCF or BSF (only for Gaskets:PTFE-sheathed non-asbestos) are equivalent to Valqua #7030 (S).
*5: Special gaskets are inserted between the flowtube and the grounding ring or grounding electrode.
*6: Enhanced dual frequency excitation is not available for models with calibration code C (High Grade Accuracy).
*7: Applicable only for Australia, New Zealand, Singapore and India.
*8: Select optional code G12 or G11 when TIIS Flameproof type with wiring using a flameproof packing adapter. Available only for JIS G1/2 female electric connection.
*9: The TIIS flameproof type is only available for AXF***C-P (remote flowtube for combined use with AXFA14).
T26-4.EPS
*10: In the case of Fieldbus communication type, optional codes C1, C2, C3, EM and G11 are unable to select.
6-32 IM 01E20D01-01E
6. OUTLINE
Table of Optional Specifications (Size 500 mm (20 in.) to 2600 mm (104 in.))
: Available –: Not available
Applicable Model
General Submersible
Remote Remote
Item Specifications Flowtube Flowtube Code
AXF***W-N
AXF***G-N
For District Heating Urethane resin potting is applied in the terminal box of a remote flowtube. Select JIS G1/2 for the electrical
and Cooling or connections. 30-meter dedicated and excitation cables are pre-wired and waterproof glands with union joints are
– DHC
Condensation-proof attached at factory.
User-specified Signal Available for the submersible type and a model with optional code DHC. The cable length is limited up to 200 meters
and Excitation Cable when combined with an AXFA11 converter. Following “L,” specify the cable length in three digits as a multiple of 1
Length meter (e. g., 001, 002, or 005) for a length up to 5 meters, or as a multiple of 5 meters (i.e., 005, 010, 015, or the like) L***
for a length of 5 meters or more. If this optional code is not selected, a 30-meter length cable is attached.
Mass Unit Setting The flow rate span, transmission pulse weight, and totalizer display pulse weight can be set in terms of mass unit.
Specify the density of the process fluid when ordering in addition to the mass flow rate span, transmission pulse
weight (for mass unit), and totalizer display pulse weight (for mass unit).
When ordering a remote flowtube, parameters for 'Mass Unit Setting' will be set in the corresponding converter before
shipment.
1. Density
a. Available density Numerics:
Specify the numeric within the value of 500 to 2000 kg/m3, 4.2 to 16.7 lb/gal, or 31.2 to 124.8 lb/cf. And it can be
up to five digits, to a maximum of 32000 ignoring the decimal point. A fraction is limited to the fourth decimal place.
b. Available density units: kg/m3, lb/gal, lb/cf
Example: A water density is about 1000kg/m3. In this case specify “1000kg/m3”.
However a density is changed by temperature. Specify the actual density. MU
(The 1000kg/m3 is equivalent to 8.345lb/gal and 62.43lb/cf.)
2. The mass flow rate span, transmission pulse weight, and totalizer display pulse weight
a. Available density Numerics:
Specify the numeric within the value of 0.0001 to 32000. And it can be up to five digits, to a maximum of
32000 ignoring the decimal point. A fraction is limited to the fourth decimal place.
b. Mass Units Available mass units: t, kg, g, klb, lb
Available time units: /d, /h, /min, /s
Note1: In case of specifying the mass flow span, calculate the volumetric flow span by the setting density, and specify
the available value in the mass flow span.
Note2: In case of transmission pulse weight and totalizer display pulse weight, specify the mass unit which was
specified as the flow unit.
G3/4 Female Waterproof glands for G3/4 conduits or flexible tubes are attached to the electrical connections. Available only for JIS
Waterproof Glands G1/2 female electric connections. – EW
Waterproof Glands Waterproof glands are attached to the electrical connections. Available only for JIS G1/2 female electric connections. – EG
Waterproof Glands Waterproof glands with union joints are attached to the electrical connections. Available only for JIS G1/2 female
with Union Joints electric connections. – EU
Stainless Steel Tag Screwd JIS SUS304 (AISI 304 SS/EN 1.4301 stainless steel equivalent) stainless steel tag plate for size 1100 to 2600
Plate mm, or a pendant tag plate of JIS SUS304 is provided for size 500 to 1000 mm. Choose this option when a SS tag
plate is required in addition to the standard nameplate with the tag number inscribed on it. SCT
Dimension (Height Width): Appr. 12.5 (4.92) 40 (15.7) mm (inch)
Direction Change of +90 degrees rotated terminal box to change the direction of the electrical connection. Available for 1000 mm
Electrical Connection (*1) (40 in.) and smaller sizes. RA
+180 degrees rotated terminal box to change the direction of the electrical connection. Available for 1000 mm (40 in.)
and smaller sizes. RB
–90 degrees rotated terminal box to change the direction of the electrical connection. Available for 1000 mm (40 in.)
and smaller sizes. RC
Material Certificate Material certificates are provided for pipe, electrodes, grounding rings, and flanges. M01
Electrical Electrical
Remote Flowtube Connection Connection
T26-5.EPS
6-33 IM 01E20D01-01E
6. OUTLINE
Table of Optional Specifications (Size 500 mm (20 in.) to 2600 mm (104 in.)) (continued)
: Available –: Not available
Applicable Model
General Submersible
Remote Remote
Item Specifications Flowtube Flowtube Code
AXF***W-N
AXF***G-N
Hydrostatic Test The test verifies the absence of leaks by applying the following water pressures (which are determined
under process connection conditions) to lining for ten minutes. Test results are described in the Note column
of a test certificate(QIC).
T01
Process Connection: Water Pressure:
JIS 10K, ANSI Class 150, DIN PN10 1.5 MPa
JIS F12 1.25 MPa
Calibration Certificate Level 2: The Declaration and the Calibration Equipment List are issued. L2
Level 3: The Declaration and the Primary Standard List are issued. L3
Level 4: The Declaration and the Yokogawa Measuring Instruments Control System are issued. L4
Five-point Calibration A flow test near 0, 25, 50, 75, and 100% of the user-specified span is performed instead of the flow test of
in User-specified the standard 2m/s span and a test certificate (QIC) is submitted. Specify the span (100% flow span) whose
Span corresponding flow velocity lies between 0.5 to 10 m/s and that is less than the maximum line capacity.
Selectable range of flow rate span is showing below.
Size : mm Selectable range of flow rate span : m3/h
(in.) (Flow rate span velocity : m/s)
500 (20) 354 (0.5) to 7068 (10.00)
600 (24) 509 (0.5) to 8200 (8.06)
700 (28) 693 (0.5) to 8200 (5.92)
800 (32) 905 (0.5) to 8200 (4.53)
900 (36) 1146 (0.5) to 8200 (3.58) SC
1000 (40) 1414 (0.5) to 8200 (2.90)
1100 (44) 1711 (0.5) to 8200 (2.40)
1200 (48) 2036 (0.5) to 8200 (2.01)
1350 (54) 2577 (0.5) to 30000 (5.82)
1500 (60) 3181 (0.5) to 30000 (4.71)
1600 (64) 3620 (0.5) to 30000 (4.14)
1800 (72) 4581 (0.5) to 30000 (3.27)
2000 (80) 5655 (0.5) to 30000 (2.65)
2200 (88) 6843 (0.5) to 30000 (2.19)
2400 (96) 8143 (0.5) to 30000 (1.84)
2600 (104) 9557 (0.5) to 30000 (1.57)
T26-6.EPS
6-34 IM 01E20D01-01E
6. OUTLINE
■ EXTERNAL DIMENSIONS
AXF Standard, AXF002-AXF015, Wafer Type, PFA Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
197(7.76)*1 Ground Terminal
Ground Terminal (M4)
154(6.06) (M4) 66*1 51.5 51.5 28(1.1)
*4 ø86(3.38) (2.6) (2.03) (2.03)
D 48 (1.89) 111(4.37)
AXF002
73(2.87) 49(1.93)
ø128(5.04)
G E 1
AXF005
W F 2 A 1 A 1
AXF010
C G N
70(2.76)
AXF015
N
P
Hi
Hr
*4: D, E, F, G; Integral Flowmeter,
)
(ød
63.5(2.50)
N, P; Remote Flowtube
H1
3(0.12)
D
* No infra-red switches are furnished 72 (2.83)
L*2 4- ø6.2(0.24) 58 (2.28)
for Fieldbus communication type.
Size code 002 005 010 015 *1: When indicator code N is selected, subtract 12 mm (0.47 inch)
from the value in the figure.
Model Size 2.5(0.1) 5(0.2) 10(0.4) 15(0.5) In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional
Lining code A A A A code, add the following value to L (face-to-face length).
Face-to-face *2
length L 81(3.19) Grounding Ring Code S, L, H, V P, T N
None +0 +26(1.02) -2(0.08)
Remote Outside dia. D 44(1.73) Option
flowtube Code GA, GC, GD +6(0.24) +28(1.10) –
Inner diameter of (Special Gaskets)
Grounding ring ød 15(0.59)
Integral
flowmeter H1
*3: When submersible type or option code DHC is selected,
Height 144(5.67)
waterproof glands and a 30m long cable are attached.
Hr 268(10.55) Add 9.5kg(20.9lb) to the weight in the table.
Remote Max. Height
flowtube 3
Weight kg (lb) * 2.4(5.3)
AXF Standard, AXF025-AXF125, Wafer Type, PFA /Polyurethane Rubber /Natural Soft Rubber /EPDM
Rubber Lining
Unit : mm (approx. inch)
ø128(5.04)
AXF050 1
W F 2 D A 1
AXF065 2
C G N G
AXF080
(2.76)
N
Hi
70
AXF100
P
Hr
AXF125
d)
øD
*5: D, E, F, G; Integral Flowmeter,
(ø
N, P; Remote Flowtube
Hr
6-35 IM 01E20D01-01E
6. OUTLINE
AXF Standard, AXF150-AXF300, Wafer Type, PFA /Polyurethane Rubber /Natural Soft Rubber /EPDM
Rubber Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
197(7.76)*1 Ground Terminal
Ground Terminal
(M4) (M4)
154(6.06) 66*1 51.5 51.5 28(1.1)
*5 ø86(3.38) (2.6) (2.03) (2.03)
111(4.37)
AXF150 D A
48 (1.89)
73(2.87) 49(1.93)
G E 1 U
ø128(5.04)
AXF200 1
W F 2 D A 1
AXF250 2
C G N G
70(2.76)
AXF300
N
P
*5: D, E, F, G; Integral Flowmeter,
d)
(ø
Hi
N, P; Remote Flowtube * In case of size
Hr
250mm or 300mm,
øD
position of grounding
H1
ring is different
from the figure to
45-degree.
Size code 150 200 250 300 *1: When indicator code N is selected, subtract 12 mm (0.47 inch)
from the value in the figure.
Model Size 150(6) 200(8) 250(10) 300(12) In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
A,U A,U A,U A,U *2: Depending on the selection of grounding ring code and optional
Lining code
D,G D,G D,G D,G code, add the following value to L (face-to-face length).
Face-to-face *2
length L 200(7.87) 250(9.84) 300(11.81) 350(13.78) Nominal Size: 150 to 200mm
Grounding Ring Code S, L, H, V P, T N
Remote Outside dia. øD 202(7.95) 252(9.92) 310(12.20) 358(14.09) None +0 +32(1.26) -2(0.08)
Flowtube Inner diameter of Option
Grounding ring ød 146.1(5.75) 193.6(7.62) 243.7(9.59) 294.7(11.60) Code GA, GC, GD +10(0.39) +38(1.5) –
(Special Gaskets)
Integral
*3
Flowmeter Width W 202(7.95) 252(9.92) 310(12.20) 358(14.09) Nominal Size: 250 to 300mm
Height H1 243(9.57) 293(11.54) 354(13.94) 402(15.83) Grounding Ring Code S, L, H, V P, T N
Option Code is “None” +0 – -2(0.08)
Hr
Remote Max. Height 367(14.45) 417(16.42) 478(18.82) 526(20.71)
Flowtube *3: When electrode structure 2 is selected, add the following value to
Weight kg (lb)*4 14.5(32.0) 22.1(48.7) 39.0(86.0) 48.3(106.5)
W(width).
Integral Max. Height Hi 405(15.93) 455(17.89) 516(20.31) 564(22.20) Nominal size 150 200 250 300
Flowmeter W +49(1.93) +50(1.97) +49(1.93) +53(2.09)
Weight kg (lb) 16.2(35.7) 23.8(52.4) 40.7(89.7) 50.0(110.2)
*4: When submersible type or option code DHC is selected,
waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table. F24.EPS
ø128(5.04)
C
AXF010 G N R
N
70(2.76)
P
Hi
N, P; Remote Flowtube
63.5(2.50)
H1
3(0.12)
6-36 IM 01E20D01-01E
6. OUTLINE
73(2.87) 49(1.93)
ø128(5.04)
E 1
G
AXF015 F 2 C E 1 A 1
C
G N
70(2.76)
N
Hi
P
Hr
*3: D, E, F, G; Integral Flowmeter,
)
N, P; Remote Flowtube (ød
63.5(2.50)
H1
3(0.12)
D
72 (2.83)
L*2 4- ø6.2(0.24) 58 (2.28)
* No infra-red switches are furnished
for Fieldbus communication type.
E 1
73(2.87) 49(1.93)
AXF040
G
AXF050 F 2 C E 1 A 1
C
AXF080 G N
70(2.76)
AXF100 N
Hi
Hr
P
øD d)
*3: D, E, F, G; Integral Flowmeter, (ø
H1
N, P; Remote Flowtube
6-37 IM 01E20D01-01E
6. OUTLINE
ø128(5.04)
(2.87) (1.93)
73 49
70(2.76)
Hi
Hr ) øD
(ød
H1
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the
Size code 150 200 value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
Model Size 150(6) 200(8) *2: Depending on the selection of grounding ring code and optional code,
Lining code add the following value to L (face-to-face length).
C C
Grounding Ring Code S, L, H, V P, T N
Face-to-face *2
length L 232(9.13) 302(11.89) L(Face-to-face length) +0 +30(1.18) -6(0.24)
Remote
Flowtube Outside dia. D 214(8.43) 264(10.39)
Integral Inner diameter of
ød 144(5.67) 192(7.56)
Flowmeter Grounding ring
Height H1 254(10.00) 304(11.97)
Hr
Remote Max. Height 378(14.88) 428(16.85)
Flowtube
Weight kg (lb) 20.2(44.5) 33.5(73.9)
6-38 IM 01E20D01-01E
6. OUTLINE
Ground Terminal
(M4)
Ground Terminal 197(7.76)*1
B J 1 154(6.06) (M4) 66*1 51.5 51.5 28(1.1)
*4
B J 2 ø86(3.38) (2.6) (2.03) (2.03)
D
AXF002 B A 1 48 (1.89) 111(4.37)
G E 1
73(2.87) 49(1.93)
AXF005 B A 2
ø128(5.04)
W F 2 A 1 1
AXF010 B D 4
C G N
AXF015 D J 1
70(2.76)
N
D J 2
P
Hi
D D 4 t *2
*4: D, E, F, G; Integral Flowmeter, N-øh
Hr
(H2)
N, P; Remote Flowtube øD
øC
H1
(ød)
(2.38)
60.5
θ°
* No infra-red switches are furnished L*2
for Fieldbus communication type.
Process Connection BJ1(JIS10K) BJ2(JIS20K) BA1(ANSI Class 150) BA2(ANSI Class 300) BD4(DIN PN40) DJ1(JIS10K) DJ2(JIS20K) DD4(DIN PN40)
Size code 002 005 010 015 002 005 010 015 002 005 010 015 002 005 010 015 002 005 010 015 002 005 010 002 005 010 002 005 010
Model
2.5 5 10 15 2.5 5 10 15 2.5 5 10 15 2.5 5 10 15 2.5 5 10 15 2.5 5 10 2.5 5 10 2.5 5 10
Size (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.1) (0.2) (0.4) (0.1) (0.2) (0.4)
Lining code A A A A A A A A A A A A A A A A A A A A A A A A A A A A A
2 200
Face-to-face length L -30 * 150(5.91) 200
(7.87) 150(5.91) 200
(7.87) 150(5.91) 200
(7.87) 150(5.91) 200
(7.87) 150(5.91) (7.87) 150(5.91) 150(5.91) 150(5.91)
Outside dia. øD 95(3.74) 95(3.74) 88.9(3.50) 95.3(3.75) 95(3.74) 90(3.54) 90(3.54) 90(3.54)
18 to 22 16 20 to 24 18 17 to 22 15.2 20 to 25 18.2 21 to 25 20 18 to 22 20 to 24 21 to 25
Thickness t *2 (0.71 to 0.87) (0.63) (0.79 to 0.94) (0.71) (0.67 to 0.87) (0.60) (0.79 to 0.98) (0.72) (0.83 to 0.98) (0.79) (0.71 to 0.87) (0.79 to 0.94) (0.83 to 0.98)
Inner diameter of ød
Remote Grounding ring 15(0.59) 15(0.59) 15(0.59) 15(0.59) 15(0.59) 12(0.47) 12(0.47) 12(0.47)
Flowtube Pitch circle dia. øC 70(2.76) 70(2.76) 60.5(2.38) 66.5(2.62) 65(2.56) 65(2.56) 65(2.56) 60(2.36)
θ˚
Bolt hole interval 45 45 45 45 45 45 45 45
Integral
Flowmeter
Hole dia. øh 15(0.59) 15(0.59) 15.7(0.62) 15.7(0.62) 14(0.55) 15(0.59) 15(0.59) 14(0.55)
Number of holes N 4 4 4 4 4 4 4 4
Height H1 141(5.54) 141(5.54) 141(5.54) 141(5.54) 141(5.54) 141(5.54) 141(5.54) 141(5.54)
Height H2 80(3.15) 80(3.15) 80(3.15) 80(3.15) 80(3.15) 80(3.15) 80(3.15) 80(3.15)
Remote Max. Height Hr 265(10.43) 265(10.43) 265(10.43) 265(10.43) 265(10.43) 265(10.43) 265(10.43) 265(10.43)
Flowtube Weight kg (lb)*3 3.4(7.5) 3.5 3.6(7.9) 3.7 3.2(7.1) 3.3 3.6(7.9) 3.7 3.8(8.4) 3.9 3.3(7.3) 3.4(7.5) 3.6(7.9)
(7.7) (8.2) (7.3) (8.2) (8.6)
Integral Max. Height Hi 303(11.91) 303(11.91) 303(11.91) 303(11.91) 303(11.91) 303(11.91) 303(11.91) 303(11.91)
Flowmeter Weight kg (lb) 5.1(11.2) 5.2 5.3(11.7) 5.4 4.9(10.8) 5.0 5.3(11.7) 5.4 5.5(12.1) 5.6 5.1(11.2)
(11.5) (11.9) (11.0) (11.9) (12.4) 5.0(11.0) 5.3(11.7)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to “L” (face-to-face length) and “t” (thickness
of flange).
L t L t L t
Grounding Ring Code S, L, H, V P, T N
None +0 +0 +26(1.02) +13(0.51) -2(0.08) -1(0.04)
Option
Code GA, GC, GD +8(0.31) +4(0.16) +30(1.18) +15(0.59) – –
(Special Gaskets)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table. F29.EPS
6-39 IM 01E20D01-01E
6. OUTLINE
AXF Standard, AXF025-AXF050, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
Ground Terminal
(M4)
Ground Terminal 197(7.76)*1
(M4)
154(6.06) 66*1 51.5 51.5 28(1.1)
*4 B A
ø86(3.38)
(2.6) (2.03) (2.03)
D 48 (1.89) 111(4.37)
AXF025 A B J
G E 1
73(2.87) 49(1.93)
ø128(5.04)
AXF032 U 1 B D 4
W F 2 C A
AXF040 D 2
C G N C J
70(2.76)
AXF050 G
N C D 4 t *2
N-øh
Hi
P
)
Hr
(ød
(H2)
*4: D, E, F, G; Integral Flowmeter,
N, P; Remote Flowtube
H1
øD
* No infra-red switches are furnished øC
for Fieldbus communication type.
L*2 θ°
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to “L” (face-to-face length) and “t” (thickness
of flange).
L t L t L t
Grounding Ring Code S, L, H, V P, T N
None +0 +0 +26(1.02) +13(0.51) -2(0.08) -1(0.04)
Option
Code GA, GC, GD +8(0.31) +4(0.16) +30(1.18) +15(0.59) – –
(Special Gaskets)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F30.EPS
6-40 IM 01E20D01-01E
6. OUTLINE
AXF Standard, AXF065-AXF125, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
ø128(5.04)
73(2.87) 49(1.93)
B A
*4
B D 2
D
70(2.76)
AXF065 A B J N-øh
G E 1 t *2
AXF080 U 1 B G 1
W F 2 1
AXF100 D 2 C A
øD
C G N
Hi
AXF125 G C D 2 øC
Hr
(H2)
N (ød)
C J
P
C G 1
H1
*4: D, E, F, G; Integral Flowmeter,
N, P; Remote Flowtube
L*2 θ°
* No infra-red switches are furnished
for Fieldbus communication type.
Process Connection BJ1/CJ1(JIS10K) BJ2/CJ2(JIS20K) BG1/CG1(JIS F12) BA1/CA1(ANSI Class 150) BA2/CA2(ANSI Class 300) BD2/CD2(DIN PN16)
Size code 065 080 100 125 065 080 100 125 080 100 125 065 080 100 125 065 080 100 125 065 080 100 125
Model 65 80 100 125 65 80 100 125 80 100 125 65 80 100 125 65 80 100 125 65 80 100 125
Size (2.5) (3) (4) (5) (2.5) (3) (4) (5) (3) (4) (5) (2.5) (3) (4) (5) (2.5) (3) (4) (5) (2.5) (3) (4) (5)
A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U
Lining code D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G
2 200 200 250 250 200 200 250 250 200 250 250 200 200 250 250 200 200 250 250 200 200 250 250
Face-to-face length L -30 * (7.87) (7.87) (9.84) (9.84) (7.87) (7.87) (9.84) (9.84) (7.87) (9.84) (9.84) (7.87) (7.87) (9.84) (9.84) (7.87) (7.87) (9.84) (9.84) (7.87) (7.87) (9.84) (9.84)
175 185 210 250 175 200 225 270 211 238 263 177.8 190.5 228.6 254.0 190.5 209.6 254.0 279.4 185 200 220 250
Outside dia. øD (6.89) (7.28) (8.27) (9.84) (6.89) (7.87) (8.86) (10.63) (8.31) (9.37) (10.35) (7.00) (7.50) (9.00) (10.00) (7.50) (8.25) (10.00) (11.00) (7.28) (7.87) (8.66) (9.84)
2 22 22 22 24 24 26 28 30 22 22 24 26.4 27.9 27.9 27.9 29.4 32.4 35.8 39.1 22 24 24 26
Thickness t* (0.87) (0.87) (0.87) (0.94) (0.94) (1.02) (1.10) (1.18) (0.87) (0.87) (0.94) (1.04) (1.10) (1.10) (1.10) (1.16) (1.28) (1.41) (1.54) (0.87) (0.94) (0.94) (1.02)
Inner diameter of 66 77 102 128 66 77 102 128 77 102 128 66 77 102 128 66 77 102 128 66 77 102 128
Remote Grounding ring ød (2.60) (3.03) (4.02) (5.04) (2.60) (3.03) (4.02) (5.04) (3.03) (4.02) (5.04) (2.60) (3.03) (4.02) (5.04) (2.60) (3.03) (4.02) (5.04) (2.60) (3.03) (4.02) (5.04)
140 150 175 210 140 160 185 225 168 195 220 139.7 152.4 190.5 215.9 149.4 168.1 200.2 235.0 145 160 180 210
flowtube Pitch circle dia. øC (5.51) (5.91) (6.89) (8.27) (5.51) (6.30) (7.28) (8.86) (6.61) (7.68) (8.66) (5.50) (6.00) (7.50) (8.50) (5.88) (6.62) (7.88) (9.25) (5.71) (6.30) (7.09) (8.27)
Bolt hole interval ˚ 45 22.5 22.5 22.5 22.5 22.5 22.5 22.5 45 45 30 45 45 22.5 22.5 22.5 22.5 22.5 22.5
22.5 45 22.5 22.5
Integral 19 19 19 23 19 23 23 25 19 19 19 19.1 19.1 19.1 22.4 22.4 22.4 22.4 18
22.4 18 18 18
Hole dia. øh (0.75) (0.75) (0.75) (0.91) (0.75) (0.91) (0.91) (0.98) (0.75) (0.75) (0.75) (0.75) (0.75) (0.75) (0.88) (0.88) (0.88) (0.88) (0.88)
(0.71) (0.71) (0.71) (0.71)
flowmeter
Number of holes N 4 8 8 8 8 8 8 8 4 4 6 4 4 8 8 8 8 8 8 4 8 8 8
Height H1 176 187 211 245 176 195 218 255 200 225 252 177 190 220 247 184 200 233 260 181 195 216 245
(6.93) (7.36) (8.30) (9.65) (6.93) (7.68) (8.59) (10.04) (7.87) (8.85) (9.90) (6.97) (7.48) (8.66) (9.72) (7.24) (7.87) (9.16) (10.22) (7.13) (7.68) (8.49) (9.65)
Height 89 95 106 120 89 95 106 120 95 106 120 89 95 106 120 89 95 106 120 89 95 106 120
H2 (3.50) (3.74) (4.16) (4.73) (3.50) (3.74) (4.16) (4.73) (3.74) (4.16) (4.73) (3.50) (3.74) (4.16) (4.72) (3.50) (3.74) (4.16) (4.72) (3.50) (3.74) (4.16) (4.72)
300 311 335 369 300 319 342 379 324 349 376 301 314 344 371 308 324 357 384 305 319 340 369
Remote Max. Height Hr (11.81) (12.24) (13.18) (14.53) (11.81) (12.56) (13.47) (14.92) (12.76) (13.73) (14.79) (11.85) (12.36) (13.54) (14.61) (12.13) (12.76) (14.04) (15.11) (12.01) (12.56) (13.37) (14.53)
flowtube Weight kg (lb)*3 9.0 9.6 12.4 17.4 9.3 12.4 16.9 24.7 12.2 15.5 19.5 10.8 12.9 17.7 20.8 12.6 16.6 26.8 34.9 9.8 11.9 14.5 19.3
(19.8) (21.2) (27.3) (38.3) (20.5) (27.3) (37.3) (54.5) (26.9) (34.2) (43.1) (23.7) (28.5) (39.1) (45.9) (27.7) (36.6) (59.1) (76.9) (21.6) (26.2) (32.0) (42.5)
338 349 372 407 338 357 380 417 362 386 413 339 352 382 409 346 362 394 421 343 357 377 407
Integral Max. Height Hi (13.31) (13.74) (14.65) (16.01) (13.31) (14.06) (14.95) (16.40) (14.25) (15.21) (16.26) (13.35) (13.86) (15.02) (16.08) (13.62) (14.25) (15.52) (16.58) (13.50) (14.06) (14.85) (16.00)
flowmeter Weight kg (lb) 10.7 11.3 14.1 19.1 11.0 14.1 18.6 26.4 13.9 17.2 21.2 12.5 14.6 19.4 22.5 14.3 18.3 28.5 36.6 11.5 13.6 16.2 21.0
(23.5) (25.0) (31.0) (42.1) (24.3) (31.0) (41.0) (58.3) (30.7) (37.8) (46.8) (27.5) (32.2) (42.8) (49.6) (31.4) (40.4) (62.8) (80.7) (25.4) (29.9) (35.7) (46.2)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to “L” (face-to-face length) and “t” (thickness of
flange).
L t L t L t
Grounding Ring Code S, L, H, V P, T N
None +0 +0 +26(1.02) +13(0.51) -2(0.08) -1(0.04)
Option
Code GA, GC, GD +8(0.31) +4(0.16) +30(1.18) +15(0.59) – –
(Special Gaskets)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F31.EPS
6-41 IM 01E20D01-01E
6. OUTLINE
AXF Standard, AXF150, AXF200, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
ø128(5.04)
N-øh
(2.87)
(2.76)
t *2
73
70
øD
Hi
(H2)
Hr
øC
H1
(ø
d)
* No infra-red switches are furnished L*2
θ°
for Fieldbus communication type.
Process Connection BJ1/CJ1(JIS10K) BJ2/CJ2(JIS20K) BG1/CG1(JIS F12) BA1/CA1(ANSI Class 150) BA2/CA2(ANSI Class 300) BD1/CD1(DIN PN10) BD2/CD2(DIN PN16)
Size code 150 200 150 200 150 200 150 200 150 200 200 150 200
Model 150 200 150 200 150 200 150 200 150 200 200 150 200
Size (6) (8) (6) (8) (6) (8) (6) (8) (6) (8) (8) (6) (8)
A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U
Lining code D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G
Face-to-face 0 2 300 350 300 350 300 350 300 350 300 350 350 300 350
length L -3 * (11.81) (13.78) (11.81) (13.78) (11.81) (13.78) (11.81) (13.78) (11.81) (13.78) (13.78) (11.81) (13.78)
280 330 305 350 290 342 279.4 342.9 317.5 381.0 340 285 340
Outside dia. øD (11.02) (12.99) (12.01) (13.78) (11.42) (13.46) (11.00) (13.50) (12.50) (15.00) (13.39) (11.22) (13.39)
2 27 27 33 35 27 29 30.4 33.4 43.5 46.1 29 27 29
Thickness t* (1.06) (1.06) (1.30) (1.38) (1.06) (1.14) (1.20) (1.31) (1.71) (1.81) (1.14) (1.06) (1.14)
Inner diameter of 146.1 193.6 146.1 193.6 146.1 193.6 146.1 193.6 146.1 193.6 193.6 146.1 193.6
Remote Grounding ring ød (5.75) (7.62) (5.75) (7.62) (5.75) (7.62) (5.75) (7.62) (5.75) (7.62) (7.62) (5.75) (7.62)
flowtube Pitch circle dia. 240 290 260 305 247 299 241.3 298.5 269.7 330.2 295 240 295
øC (9.45) (11.42) (10.24) (12.01) (9.72) (11.77) (9.50) (11.75) (10.62) (13.00) (11.61) (9.45) (11.61)
Integral Bolt hole interval ˚ 22.5 15 15 15 30 22.5 22.5 22.5 15 15 22.5 22.5 15
flowmeter Hole dia. 23 23 25 25 19 19 22.4 22.4 22.4 25.4 22 22 22
øh (0.91) (0.91) (0.98) (0.98) (0.75) (0.75) (0.88) (0.88) (0.88) (1.00) (0.87) (0.87) (0.87)
Number of holes N 8 12 12 12 6 8 8 8 12 12 8 8 12
281 331 294 341 286 337 281 337 300 357 336 284 336
Height H1 (11.06) (13.03) (11.56) (13.43) (11.26) (13.27) (11.05) (13.29) (11.80) (14.04) (13.23) (11.16) (13.23)
141 166 141 166 141 166 141 166 141 166 166 141 166
Height H2 (5.55) (6.54) (5.55) (6.54) (5.55) (6.54) (5.55) (6.54) (5.55) (6.54) (6.54) (5.55) (6.54)
405 455 418 465 410 461 405 461 424 481 460 408 460
Remote Max. Height Hr (15.94) (17.91) (16.44) (18.31) (16.14) (18.15) (15.93) (18.17) (16.68) (18.92) (18.11) (16.04) (18.11)
flowtube 27.8 37.3 37.1 51.9 29.9 43.2 30.9 49.2 52.5 78.8 42.5 28.7 41.9
Weight kg (lb) *3 (61.3) (82.2) (81.8) (114.4) (65.9) (95.3) (68.0) (108.4) (115.7) (173.7) (93.7) (63.2) (92.5)
443 493 456 503 448 499 443 499 462 519 498 446 498
Integral Max. Height Hi (17.42) (19.39) (17.95) (19.80) (17.64) (19.65) (17.43) (19.66) (18.18) (20.41) (19.61) (17.54) (19.61)
flowmeter 29.5 39.0 38.8 53.6 31.6 44.9 32.6 50.9 54.2 80.5 44.2 30.4 43.6
Weight kg (lb) (65.0) (86.0) (85.5) (118.2) (69.7) (99.0) (71.8) (112.2) (119.5) (177.5) (97.5) (66.9) (96.2)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to “L” (face-to-face length)
and “t” (thickness of flange).
L t L t L t
Grounding Ring Code S, L, H, V P, T N
None +0 +0 +32(1.26) +16(0.63) -2(0.08) -1(0.04)
Option
Code GA, GC, GD +10(0.39) +5(0.20) +38(1.5) +19(0.75) – –
(Special Gaskets)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table. F32.EPS
6-42 IM 01E20D01-01E
6. OUTLINE
AXF Standard, AXF250-AXF400, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
B A Unit : mm (approx. inch)
*4 B D
D
AXF250 A B J
G E 1
AXF300 U 1 B G 1
W F 2 1
AXF350 D 2 C A
C G N
AXF400 G C D
N
C J
P
C G 1
*4: D, E, F, G; Integral Flowmeter,
N, P; Remote Flowtube Integral Flowmeter Remote Flowtube
Integral Flowmeter Remote Flowtube Ground Termonal
(M4)
Ground Terminal 197(7.76)*1
154(6.06) ø86 (3.38) (M4)
48(1.89) 66*1 51.5 51.5 28(1.1)
49 (1.93)
ø128(5.04)
(2.87)
73
70
N- øh
øD
(H2)
øC
Hi
H3
Hr
H1
(ød
)
t*2
L*2
θ°
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to “L” (face-to-face length)
and “t” (thickness of flange).
Nominal Size: 250 mm to 300 mm Nominal Size: 350 mm to 400 mm
L t L t L t L t
Grounding Ring Code S, L, H, V N Grounding Ring Code S, L, H, V N
Option Code is “None” +0 +0 -6(0.24) -3(0.12) Option Code is “None” +0 +0 -10(0.39) -5(0.20)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table. F33.EPS
6-43 IM 01E20D01-01E
6. OUTLINE
Ground Terminal
Ground Terminal 197(7.76)*1
(M4)
AXF015 (M4)
154(6.06) ø86(3.38) 66*1 51.5 51.5 28(1.1)
AXF025 *3 48 (2.6) (2.03) (2.03)
AXF032 D (1.89) 111(4.37)
ø128(5.04)
AXF040 E 1 H A B
(2.87) (1.93)
49
AXF050 H F 2 A L 1 N H D B 1
AXF065 G N H K B
70(2.76)
AXF080 N
73
Hi
AXF100 P
Hr
AXF125
øD
*3: D, E, F, G; Integral Flowmeter, (ød)
H1
N, P; Remote Flowtube Adapter for clamp connection
Process Connection HAB / HDB / HKB *1: When indicator code N is selected, subtract 12 mm
(0.47 inch) from the value in the figure.
Size code 015 025 032 040 050 065 080 100 125 *2: When option code DHC is selected, waterproof
Model 15 25 32 40 50 65 80 100 125 glands and a 30m long cable are attached.
Size (0.5) (1) (1.3) (1.5) (2) (2.6) (3) (4) (5) Add 9.5kg(20.9lb) to the weight in the table.
Lining code A A A A A A A A A
Face-to-face 166 166 166 166 176 196 216 246 316
length L (6.55) (6.55) (6.55) (6.55) (6.94) (7.73) (8.52) (9.70) (12.46)
73 73 73 86 99 117 129 155 183
Outside dia. øD (2.87) (2.87) (2.87) (3.39) (3.90) (4.61) (5.08) (6.10) (7.20)
Remote 15.7 22.1 34.8 47.5 60.2 72.9 97.4
Flowtube HAB – –
(0.62) (0.87) (1.37) (1.87) (2.37) (2.87) (3.83)
16 26 32 38 50 66 81 100 125
Integral Inner dia. ød HDB
øC
(0.63) (1.02) (1.26) (1.50) (1.97) (2.60) (3.19) (3.94) (4.92)
Flowmeter
15.2 22.6 31.3 35.6 48.6 60.3 72.9 97.6 135.7
HKB (0.60) (0.89) (1.23) (1.40) (1.91) (2.37) (2.87) (3.84) (5.34)
98 98 98 111 129 147 157 183 212
Height H1
øA
øE
øF
øB
ød
(3.86) (3.86) (3.86) (4.37) (5.08) (5.79) (6.18) (7.20) (8.35)
222 222 222 235 253 271 281 307 336
Max. Height Hr (8.74) (8.74) (8.74) (9.25) (9.96) (10.67) (11.06) (12.09) (13.23)
Remote
Flowtube 2 2.7 2.5 2.6 2.9 3.6 4.8 5.7 8.1 12.1
Weight kg (lb)* (6.0) (5.5) (5.7) (6.4) (7.9) (10.6) (12.6) (17.9) (26.7)
260 260 260 273 291 309 319 345 374
Max. Height Hi (14.70)
G
Integral (10.22) (10.22) (10.22) (10.73) (11.44) (12.15) (12.54) (13.56)
Flowmeter 4.4 4.2 4.3 4.6 5.3 6.5 7.4 9.8 13.8
Weight kg (lb) (9.7) (9.3) (9.5) (10.1) (11.7) (14.3) (16.3) (21.6) (30.4) Adapter for clamp connection
6-44 IM 01E20D01-01E
6. OUTLINE
(2.87) (1.93)
AXF050 H F 2 A L 1 N J D B 1 111(4.37)
49
ø128(5.04)
AXF065 G N J S B
(2.76)
AXF080 N
70
73
Hi
AXF100 P
AXF125
Hr
*3: D, E, F, G; Integral Flowmeter,
N, P; Remote Flowtube
H1
øD
(ød)
Adapter for union
L connection
øC
øA
øB
øE
JKB:
ød
Process Connection JKB / JDB / JSB F G
Size code 015 025 032 040 050 065 080 100 125 H
Model 15 25 32 40 50 65 80 100 125
Size (0.5) (1) (1.3) (1.5) (2) (2.6) (3) (4) (5) G
øC
Lining code A A A A A A A A A
166 166 166 166 176 196 216 246 –
JKB (6.55) (6.55) (6.55) (6.55) (6.94) (7.73) (8.52) (9.70)
Face-to-face 166 166 166 166 176 196 236 266 326
øA
øB
øE
øF
ød
length L JDB (6.55) (6.55) (6.55) (6.55) (6.94) (7.73) (9.31) (10.49) (12.85) JDB:
166 166 166 176 196 216 276 –
JSB –
(6.55) (6.55) (6.55) (6.94) (7.73) (8.52) (10.88)
Remote 155 183
73 73 73 86 99 117 129
Flowtube Outside dia. øD (2.87) (2.87) (2.87) (3.39) (3.90) (4.61) (5.08) (6.10) (7.20)
15.2 22.6 31.3 35.6 48.6 60.3 72.9 97.6 H
Integral JKB –
Flowmeter (0.60) (0.89) (1.23) (1.40) (1.91) (2.37) (2.87) (3.84)
16 26 32 38 50 66 81 100 125
Inside dia. ød JDB (0.63) (1.02) (1.26) (1.50) (1.97) (2.60) (3.19) (3.94) (4.92)
22.5 29.6 35.5 48.5 60.5 72.9 97.6 G
JSB – –
(2.38) (2.87) (3.84)
øC
ød
øA
Max. Height Hr (8.74) (8.74) (8.74) (9.25) (9.96) (10.67) (11.06) (12.09) (13.23) JSB:
Remote
Flowtube *2 2.6 2.6 2.7 3 3.8 4.9 5.9 8.2 13
Weight kg (lb) (5.7) (5.7) (6.0) (6.6) (8.4) (10.8) (13.0) (18.1) (28.7)
260 260 260 273 291 309 319 345 374 3
Max. Height Hi (10.24) (10.24) (10.24) (10.73) (11.44) (12.17) (12.54) (13.56) (14.70)
Integral
Flowmeter 4.3 4.3 4.4 4.7 5.5 6.6 7.6 9.9 14.7 F
Weight kg (lb) (9.5) (9.5) (9.7) (10.4) (12.1) (14.6) (16.8) (21.8) (32.4) H
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the
figure. Adapters for union connection
*2: When option code DHC is selected, waterproof glands and a 30m long cable are
attached. Add 9.5kg (20.9lb) to the weight in the table.
Process JKB (ISO2853 Union)
Connection JDB (DIN 11851 Union) JSB (SMS1145 Union)
Nominal 15 25 32 40 50 65 80 100 15 25 32 40 50 65 80 100 125
Size 25 32 40 50 65 80 100
70 70 70 83 96 114 126 152 70 70 70 83 96 114 126 152 180 70 70 83 96 114 126 152
øA (2.76) (2.76) (2.76) (3.27) (3.78) (4.49) (4.96) (5.98) (2.76) (2.76) (2.76) (3.27) (3.78) (4.49) (4.96) (5.98) (7.09) (2.76) (2.76) (3.27) (3.78) (4.49) (4.96) (5.98)
16 22.2 29.4 34.6 47.6 59.5 72.3 97 16 22.2 29.4 34.6 47.6 59.5 72.3 97 123 22.2 29.4 34.6 47.6 59.5 72.3 97
øB (0.63) (0.87) (1.16) (1.36) (1.87) (2.34) (2.85) (3.82) (0.63) (0.87) (1.16) (1.36) (1.87) (2.34) (2.85) (3.82) (4.84) (0.87) (1.16) (1.36) (1.87) (2.34) (2.85) (3.82)
18 25.6 34.3 38.6 51.6 64.1 76.7 102.5 20 30 36 42 54 70 85 104 129 25.4 32 38.1 51 63.5 76.2 102.5
øC (0.71) (1.01) (1.35) (1.52) (2.03) (2.52) (3.02) (4.04) (0.79) (1.18) (1.42) (1.65) (2.13) (2.76) (3.35) (4.09) (5.08) (1.00) (1.26) (1.50) (2.01) (2.50) (3.00) (4.04)
15.2 22.6 31.3 35.6 48.6 60.3 72.9 97.6 16 26 32 38 50 66 81 100 125 22.5 29.6 35.5 48.5 60.5 72.9 97.6
ød (0.60) (0.89) (1.23) (1.40) (1.91) (2.37) (2.87) (3.84) (0.63) (1.02) (1.26) (1.50) (1.97) (2.60) (3.19) (3.94) (4.92) (0.89) (1.17) (1.40) (1.91) (2.38) (2.87) (3.84)
21.2 29.2 38.2 42.7 56.2 69.9 82.6 108.7 18 30 36 42 54 71 85 104 130 32 40 48 61 73.5 86 120
øE (0.83) (1.15) (1.50) (1.68) (2.21) (2.75) (3.25) (4.28) (0.71) (1.18) (1.42) (1.65) (2.13) (2.80) (3.35) (4.09) (5.12) (1.26) (1.57) (1.89) (2.40) (2.89) (3.39) (4.72)
13.5 13.5 13.5 13.5 13.5 13.5 13.5 13.5 25.8 39.8 45.8 51.8 63.8 80.8 94.8 113.8 141.8 11 13 15 15 19 19 30
øF (0.53) (0.53) (0.53) (0.53) (0.53) (0.53) (0.53) (0.53) (1.02) (1.57) (1.80) (2.04) (2.51) (3.18) (3.73) (4.48) (5.58) (0.43) (0.51) (0.59) (0.59) (0.75) (0.75) (1.18)
3 3 3 3 3 3 3 3 RD34 RD52 RD58 RD65 RD78 RD95 RD110 RD130 RD160 RD40 RD48 RD60 RD70 RD85 RD98 RD132
G (0.12) (0.12) (0.12) (0.12) (0.12) (0.12) (0.12) (0.12) ×1/8″ ×1/6″ ×1/6″ ×1/6″ ×1/6″ ×1/6″ ×1/4″ ×1/4″ ×1/4″ ×1/6″ ×1/6″ ×1/6″ ×1/6″ ×1/6″ ×1/6″ ×1/6″
50 50 50 50 50 50 50 50 50 50 50 50 50 50 60 60 65 50 50 50 50 50 50 65
H (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (2.36) (2.36) (2.56) (1.97) (1.97) (1.97) (1.97) (1.97) (1.97) (2.56)
F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811
Parts No. LA LB LC LD LE LF LG LH KR KS KT KU KV KW KX KY KZ LK LL LM LN LP LQ LR
F35.EPS
6-45 IM 01E20D01-01E
6. OUTLINE
73(2.87) 49(1.93)
AXF040 E 1
ø128(5.04)
K K B
AXF050 H F 2 A L 1 N 1
K D B
AXF065 G N
70(2.76)
AXF080 N
Hi
AXF100 P
Hr
AXF125
øD
*3: D, E, F, G; Integral Flowmeter, (ød)
H1
N, P; Remote Flowtube Adapter for butt weld connection
Process Connection KKB / KDB *1: When indicator code N is selected, subtract 12 mm
(0.47 inch) from the value in the figure.
Size code 015 025 032 040 050 065 080 100 125 *2: When option code DHC is selected, waterproof
Model 15 25 32 40 50 65 80 100 125 glands and a 30m long cable are attached.
Size (0.5) (1) (1.3) (1.5) (2) (2.6) (3) (4) (5) Add 9.5kg (20.9lb) to the weight in the table.
Lining code A A A A A A A A A
Face-to-face 126 126 126 126 136 156 176 206 276
length L (4.98) (4.98) (4.98) (4.98) (5.37) (6.16) (6.94) (8.13) (10.88)
73 73 73 86 99 117 129 155 183
Outside dia. øD (2.87) (2.87) (2.87) (3.39) (3.90) (4.61) (5.08) (6.10) (7.20)
Remote 15.2 22.6 31.3 35.6 48.6 60.3 72.9 97.6 135.7
Flowtube KKB (0.60) (0.89) (1.23) (1.40) (1.91) (2.37) (2.87) (3.84) (5.34)
Inner dia. ød
16 26 32 38 50 66 81 100 125
Integral KDB (0.63) (1.02) (1.26) (1.50) (1.97) (2.60) (3.19) (3.94) (4.92)
Flowmeter
98 98 98 111 129 147 157 183 212
øA
øB
øC
ød
Height H1 (3.86) (3.86) (3.86) (4.37) (5.08) (5.79) (6.18) (7.20) (8.35)
222 222 222 235 253 271 281 307 336
Max. Height Hr (8.74) (8.74) (8.74) (9.25) (9.96) (10.67) (11.06) (12.09) (13.23)
Remote
Flowtube 2 2.6 2.3 2.5 2.8 3.4 4.5 5.3 7.1 11
Weight kg (lb)* (5.7) (5.1) (5.5) (6.2) (7.5) (9.9) (11.7) (15.7) (24.3)
260 260 260 273 291 309 319 345 374
Max. Height Hi (10.24) (10.24) (10.24) (10.73) (11.44) (12.17) (12.54) (13.56) (14.70)
Integral E
Flowmeter 4.3 4 4.2 4.5 5.1 6.2 7 8.8 12.7
Weight kg (lb) (9.5) (8.8) (9.3) (9.9) (11.2) (13.7) (15.4) (19.4) (28.0)
Adapter for butt weld connection
Process
Connection KKB (ISO2037 Butt Weld) KDB (DIN 1185 Butt Weld)
Nominal 15 25 32 40 50 65 80 100 125
Size 15 25 32 40 50 65 80 100 125
70 70 70 83 96 114 126 152 180 70 70 70 83 96 114 126 152 180
øA (2.76) (2.76) (2.76) (3.27) (3.78) (4.49) (4.96) (5.98) (7.09) (2.76) (2.76) (2.76) (3.27) (3.78) (4.49) (4.96) (5.98) (7.09)
16 22.2 29.4 34.6 47.6 59.5 72.3 97 123 16 22.2 29.4 34.6 47.6 59.5 72.3 97 123
øB (0.63) (0.87) (1.16) (1.36) (1.87) (2.34) (2.85) (3.82) (4.84) (0.63) (0.87) (1.16) (1.36) (1.87) (2.34) (2.85) (3.82) (4.84)
18 25.6 34.3 38.6 51.6 64.1 76.7 102.5 141.2 20 30 36 42 54 70 85 104 129
øC (0.71) (1.01) (1.35) (1.52) (2.03) (2.52) (3.02) (4.04) (5.56) (0.79) (1.18) (1.42) (1.65) (2.13) (2.76) (3.35) (4.09) (5.08)
15.2 22.6 31.3 35.6 48.6 60.3 72.9 97.6 135.7 16 26 32 38 50 66 81 100 125
ød (0.60) (0.89) (1.23) (1.40) (1.91) (2.37) (2.87) (3.84) (5.34) (0.63) (1.02) (1.26) (1.50) (1.97) (2.60) (3.19) (3.94) (4.92)
30 30 30 30 30 30 30 30 40 30 30 30 30 30 30 30 30 40
E (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.57) (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.18) (1.57)
F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811 F9811
Parts No. NN NP NQ NR NS NT NU NV NW ND NE NF NG NH NJ NK NL NM
F36.EPS
6-46 IM 01E20D01-01E
6. OUTLINE
Replacement model for Earlier ADMAG or ADMAG AE, AXF002-AXF015, Wafer Type, PFA Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
197(7.76)*1 Ground Terminal
Ground Terminal (M4)
154(6.06) (M4) 66*1 51.5 51.5 28(1.1)
*4 ø86(3.38) (2.6) (2.03) (2.03)
D 48 (1.89) 111(4.37)
AXF002
G E 1
73(2.87) 49(1.93)
ø128(5.04)
AXF005
W F 2 A 1 A 2
AXF010
C G N
AXF015
70(2.76)
N
P
Hi
Hr
*4: D, E, F, G; Integral Flowmeter,
N, P; Remote Flowtube )
(ød
H1
D
3(0.12)
(2.50)
63.5
72 (2.83)
L*2 4- ø6.2(0.24) 58 (2.28)
* No infra-red switches are furnished
for Fieldbus communication type.
Size code 002 005 010 015 *1: When indicator code N is selected, subtract 12 mm (0.47 inch)
from the value in the figure.
Model Size 2.5(0.1) 5(0.2) 10(0.4) 15(0.5) In case of explosion proof type with indicator, add 5 mm (0.2
inch) to it.
Lining code A A A A *2: Depending on the selection of grounding ring code and optional
code, add the following value to L (face-to-face length).
Face-to-face *2 85(3.35)
length L Grounding Ring Code S, L, H, V P, T N
Remote
Flowtube Outside dia. None +0 +22(0.87) -6(0.24)
D 44(1.73) Option
Code GA, GC, GD +2(0.08) +24(0.94) –
Integral Inner diameter of (Special Gaskets)
Flowmeter Grounding ring
ød 15(0.59)
*3: When submersible type or option code DHC is selected,
Height H1 144(5.67) waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
Remote Max. Height Hr 265(10.43)
Flowtube 3
Weight kg (lb)* 2.4(5.3)
Replacement model for Earlier ADMAG or ADMAG AE, AXF025-AXF100, Wafer Type,
PFA /Polyurethane Rubber Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube Integral Flowmeter Remote Flowtube
Ground Terminal
(M4)
154(6.06) ø86(3.38) 197(7.76)*1 Ground Terminal
*4 (M4)
48(1.89) 66*1 51.5 51.5 28(1.1)
AXF025 D (2.6) (2.03) (2.03)
73(2.87) 49(1.93)
AXF040 G E 1 111(4.37)
A
AXF050 W F 2 1 A 2
ø128(5.04)
U
AXF080 C G N
(2.76)
AXF100 N
70
Hi
Hr
P
*4: D, E, F, G; Integral Flowmeter,
H1
d)
øD
(ø
N, P; Remote Flowtube
L*2
* No infra-red switches are furnished
for Fieldbus communication type.
Size code 025 040 050 080 100 *1: When indicator code N is selected, subtract 12 mm (0.47 inch)
from the value in the figure.
Model Size 25(1) 40(1.5) 50(2) 80(3) 100(4) In case of explosion proof type with indicator, add 5 mm (0.2
inch) to it.
Lining code A,U A,U A,U A,U A,U *2: Depending on the selection of grounding ring code and optional
Face-to-face Grounding Ring Code S, L, H, V P, T N
*2 93(3.66) 106(4.17) 120(4.72) 160(6.30) 180(7.09)
length L None +0 +22(0.87) -6(0.24)
Remote Option
Flowtube Outside dia. øD 67.5(2.66) 86(3.39) 99(3.90) 129(5.08) 155(6.10) Code GA, GC, GD +2(0.08) +24(0.94) –
(Special Gaskets)
Inner diameter of
Integral Grounding ring ød 27(1.06) 40(1.57) 52(2.05) 81(3.19) 98(3.86) *3: When submersible type or option code DHC is selected,
Flowmeter waterproof glands and a 30m long cable are attached.
Height H1 92(3.62) 111(4.37) 129(5.08) 157(6.18) 183(7.20) Add 9.5kg(20.9lb) to the weight in the table.
216(8.50) 235(9.25) 253(9.96) 281(11.06) 307(12.09)
Remote Max. Height Hr
Flowtube 3
Weight kg (lb)* 3.1(6.7) 3.5(7.7) 4.2(9.3) 6.4(14.1) 8.0(17.6)
6-47 IM 01E20D01-01E
6. OUTLINE
Replacement model for Earlier ADMAG or ADMAG AE, AXF150, AXF200, Wafer Type,
PFA /Polyurethane Rubber Lining
Unit : mm (approx. inch)
Ground Terminal
(M4)
Ground Terminal 197(7.76)*1
(M4)
66*1 51.5 51.5 28(1.1)
154(6.06) ø86(3.38)
*4 (2.6) (2.03) (2.03)
D 48(1.89) 111(4.37)
73(2.87) 49(1.93)
ø128(5.04)
G E 1
AXF150 A
W F 2 1 A 2
AXF200 U
C G N
70(2.76)
N
P
d)
*4: D, E, F, G; Integral Flowmeter,
(ø
Hi
N, P; Remote Flowtube
Hr
øD
H1
* No infra-red switches are furnished L*2
for Fieldbus communication type.
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the
Size code 150 200
value in the figure.
Size In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
Model 150(6) 200(8)
*2: Depending on the selection of grounding ring code and optional code, add
Lining code A,U A,U the following value to L (face-to-face length).
6-48 IM 01E20D01-01E
6. OUTLINE
Replacement model for Earlier ADMAG or ADMAG AE, AXF150-AXF250, JIS/ANSI/DIN Flange Type,
PFA /Polyurethane Rubber Lining
*4
*4: D, E, F, G; Integral Flowmeter,
D N, P; Remote Flowtube Unit : mm (approx. inch)
C A
AXF150 G E 1
A 1 C D
AXF200 W F 2 2
U 2 C J
AXF250 C G N
C G 1
N
P Integral Flowmeter Remote Flowtube
Integral Flowmeter Remote Flowtube Ground Terminal
(M4)
197(7.76)*1
Ground Terminal
154(6.06) ø86 (3.38) (M4) 66*1 51.5 51.5 28(1.1)
48(1.89) (2.6) (2.03) (2.03)
111(4.37)
ø128(5.04)
73 49 (1.93)
Eye Bolt
(2.76)
(2.87)
70
N- øh
øD
(H2)
Hi
Hr
øC
H3
H1
(ød
)
* No infra-red switches are furnished t *2 θ°
for Fieldbus communication type. L*2
Process Connection CJ1(JIS10K) CJ2(JIS20K) CG1(JIS F12) CA1(ANSI Class 150) CA2(ANSI Class 300) CD1(DIN PN10) CD2(DIN PN16)
Size code 150 200 250 150 200 250 150 200 250 150 200 250 150 200 250 200 250 150 200 250
Model
150 200 250 150 200 250 150 200 250 150 200 250 150 200 250 200 250 150 200 250
Size (6) (8) (10) (6) (8) (10) (6) (8) (10) (6) (8) (10) (6) (8) (10) (8) (10) (6) (8) (10)
Lining code A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U
Face-to-face 270 340 430 270 340 430 270 340 430 270 340 430 270 340 430 340 430 270 340 430
length L*2 (10.63) (13.39) (16.93) (10.63) (13.39) (16.93) (10.63) (13.39) (16.93) (10.63) (13.39) (16.93) (10.63) (13.39) (16.93) (13.39) (16.93) (10.63) (13.39) (16.93)
280 330 400 305 350 430 290 342 410 279.4 342.9 406.4 317.5 381.0 444.5 340 395 285 340 405
Outside dia. øD (11.02) (12.99) (15.75) (12.01) (13.78) (16.93) (11.42) (13.46) (16.14) (11.00) (13.50) (16.00) (12.50) (15.00) (17.50) (13.39) (15.55) (11.22) (13.39) (15.94)
31 31 33 37 39 43 31 33 33 34.4 37.4 39.2 47.5 50.1 56.7 33 35 31 33 35
Thickness t *2 (1.22) (1.22) (1.30) (1.46) (1.54) (1.69) (1.22) (1.30) (1.30) (1.30) (1.38) (1.22) (1.30) (1.38)
(1.35) (1.47) (1.54) (1.87) (1.97) (2.23)
Inner diameter of 140.7 188.9 243 140.7 188.9 243 140.7 188.9 243 140.7 188.9 243 140.7 188.9 243 188.9 243 140.7 188.9 243
Remote Grounding ring ød (5.54) (7.44) (9.57) (5.54) (7.44) (9.57) (5.54) (7.44) (9.57) (5.54) (7.44) (9.57) (5.54) (7.44) (9.57) (7.44) (9.57) (5.54) (7.44) (9.57)
flowtube 240 290 355 260 305 380 247 299 360 241.3 298.5 362.0 269.7 330.2 387.4 295 350 240 295 355
Pitch circle dia. øC (9.45) (11.42) (13.98) (10.24) (12.01) (14.96) (9.72) (11.77) (14.17) (9.50) (11.75) (14.25) (10.62) (13.00) (15.25) (11.61) (13.78) (9.45) (11.61) (13.98)
Integral
flowmeter Bolt hole interval ˚
22.5 15 15 15 15 15 30 22.5 22.5 22.5 22.5 15 15 15 11.25 22.5 15 22.5 15 15
23 23 25 25 25 27 19 19 23 22.4 22.4 25.4 22.4 25.4 28.4 22 22 22 22 26
Hole dia. øh
(0.91) (0.91) (0.98) (0.98) (0.98) (1.06) (0.75) (0.75) (0.91) (0.88) (0.88) (1.00) (0.88) (1.00) (1.12) (0.87) (0.87) (0.87) (0.87) (1.02)
Number of holes N 8 12 12 12 12 12 6 8 8 8 8 12 12 12 16 8 12 8 12 12
284 334 400 297 344 415 289 340 405 284 340 403 303 360 422 339 397 287 339 402
Height H1
(11.18) (13.15) (15.75) (11.69) (13.54) (16.34) (11.38) (13.39) (15.94) (11.18) (13.39) (15.87) (11.93) (14.17) (16.61) (13.35) (15.63) (11.30) (13.35) (15.83)
141 166 197 141 166 197 141 166 197 141 166 197 141 166 197 166 197 141 166 197
Height H2
(5.55) (6.54) (7.76) (5.55) (6.54) (7.76) (5.55) (6.54) (7.76) (5.55) (6.54) (7.76) (5.55) (6.54) (7.76) (6.54) (7.76) (5.55) (6.54) (7.76)
325 375 454 350 395 484 335 387 464 324 387 460 362 426 499 385 449 330 385 459
Height H3
(12.80) (14.76) (17.87) (13.78) (15.55) (19.06) (13.19) (15.24) (18.27) (12.76) (15.24) (18.11) (14.25) (16.77) (19.65) (15.16) (17.68) (12.99) (15.16) (18.07)
408 458 524 421 468 539 413 464 529 408 464 527 427 484 546 463 521 411 463 526
Max. Height Hr
Remote (16.06) (18.03) (20.63) (16.57) (18.43) (21.22) (16.26) (18.27) (20.83) (16.06) (18.27) (20.75) (16.81) (19.06) (21.50) (18.23) (20.51) (16.18) (18.23) (20.71)
flowtube 29 39 64 38.3 53.6 92.5 31.1 44.6 67.4 32.1 50.9 77.4 53.7 80.5 127.0 44.2 67.0 29.9 43.6 68.8
Weight kg (lb)*3 (63.9) (86.0) (141.1) (84.4) (118.2) (203.9) (68.6) (98.3) (148.6) (70.7) (112.2) (170.6) (118.4) (177.5) (279.9) (97.5) (147.7) (65.8) (96.2) (151.7)
Hi 446 496 562 459 506 577 451 502 567 446 502 565 465 522 584 501 559 449 501 564
Integral Max. Height (17.56) (19.53) (22.13) (18.07) (19.92) (22.72) (17.76) (19.76) (22.32) (17.56) (19.76) (22.24) (18.31) (20.55) (22.99) (19.72) (22.01) (17.68) (19.72) (22.20)
flowmeter 30.7 40.7 65.7 40.0 55.3 94.2 32.8 46.3 69.1 33.8 52.6 79.1 55.4 82.2 128.7 45.9 68.7 31.6 45.3 70.5
Weight kg (lb) (67.7) (89.7) (144.8) (88.2) (121.9) (207.7) (72.3) (102.1) (152.3) (74.4) (115.9) (174.4) (122.1) (181.2) (283.6) (101.2) (151.5) (69.6) (100.0) (155.5)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to “L” (face-to-face length) and “t” (thickness of
flange).
Nominal Size: 150, 200 mm Nominal Size: 250 mm
L t L t L t L t L t
Grounding Ring Code S, L, H, V P, T N Grounding Ring Code S, L, H, V N
None +0 +0 +28(1.1) +14(0.55) -6(0.24) -3(0.12) Option Code is “None” +0 +0 -6(0.24) -3(0.12)
Option
Code GA, GC, GD +2(0.08) +1(0.04) +30(1.81) +15(0.59) – –
(Special Gaskets)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F40.EPS
6-49 IM 01E20D01-01E
6. OUTLINE
N- øh
(H2)
øD
øC
Hr
Flange Detail (ø
d)
H1
4- ø24(0.94)
(Anchor bolt M20)
125
(4.92) 35 L2 35
L1 (1.38) (1.38)
L
Process Connection CJ1(JIS10K) CG1(JIS F12) CA1(ANSI Class 150) CD1(DIN PN10)
Size code 500 600 700 800 900 10L 500 600 700 800 900 10L 500 600 500 600 700 800 900 10L
Model 500 600 700 800 900 1000 500 600 700 800 900 1000 500 600 500 600 700 800 900 1000
Size (20) (24) (28) (32) (36) (40) (20) (24) (28) (32) (36) (40) (20) (24) (20) (24) (28) (32) (36) (40)
Lining code U U U U U U U U U U U U U U U U U U U U
Face-to-face 750 800 900 1050 1200 1300 750 800 900 1050 1200 1300 750 800 750 800 900 1050 1200 1300
length L (29.53) (31.50) (35.43) (41.34) (47.24) (51.18) (29.53) (31.50) (35.43) (41.34) (47.24) (51.18) (29.53) (31.50) (29.53) (31.50) (35.43) (41.34) (47.24) (51.18)
350 400 450 550 700 800 350 400 450 550 700 800 350 400 350 400 450 550 700 800
Support interval L1 (13.78) (15.75) (17.72) (21.65) (27.56) (31.50) (13.78) (15.75) (17.72) (21.65) (27.56) (31.50) (13.78) (15.75) (13.78) (15.75) (17.72) (21.65) (27.56) (31.50)
350 400 500 550 650 700 350 400 500 550 650 700 350 400 350 400 500 550 650 700
Mounting bolt interval L2 (13.78) (15.75) (19.69) (21.65) (25.59) (27.56) (13.78) (15.75) (19.69) (21.65) (25.59) (27.56) (13.78) (15.75) (13.78) (15.75) (19.69) (21.65) (25.59) (27.56)
675 795 905 1020 1120 1235 706 810 928 1034 1156 1262 698.5 812.8 670 780 895 1015 1115 1230
Outside dia. øD (26.57) (31.30) (35.63) (40.16) (44.09) (48.62) (27.80) (31.89) (36.54) (40.71) (45.51) (49.69) (27.50) (32.00) (26.38) (30.71) (35.24) (39.96) (43.90) (48.43)
30 32 34 36 38 40 30 33 35 37 39 41 42.9 47.7 28 28 30 32 34 34
Thickness t (1.18) (1.26) (1.34) (1.42) (1.50) (1.57) (1.18) (1.30) (1.38) (1.46) (1.54) (1.61) (1.69) (1.88) (1.10) (1.10) (1.18) (1.26) (1.34) (1.34)
Inner diameter of 468 563 665 765 855 942 485 589 689 788 888 990 468 563 468 563 665 765 855 942
Grounding ring ød (18.43) (22.17) (26.18) (30.12) (33.66) (37.09) (19.09) (23.19) (27.13) (31.02) (34.96) (38.98) (18.43) (22.17) (18.43) (22.17) (26.18) (30.12) (33.66) (37.09)
620 730 840 950 1050 1160 639 743 854 960 1073 1179 635 749.3 620 725 840 950 1050 1160
Remote Pitch circle dia. øC (24.41) (28.74) (33.07) (37.40) (41.34) (45.67) (25.16) (29.25) (33.62) (37.80) (42.24) (46.42) (25.00) (29.50) (24.41) (28.54) (33.07) (37.40) (41.34) (45.67)
Flowtube Bolt hole interval
θ˚ 9 7.5 7.5 6.4 6.4 6.4 15 11.25 11.25 9 9 7.5 9 9 9 9 7.5 7.5 6.4 6.4
27 33 33 33 33 39 27 27 33 33 33 33 31.7 35 26 30 30 33 33 36
Hole dia. øh (1.06) (1.30) (1.30) (1.30) (1.30) (1.54) (1.06) (1.06) (1.30) (1.30) (1.30) (1.30) (1.25) (1.38) (1.02) (1.18) (1.18) (1.30) (1.30) (1.42)
Number of holes N 20 24 24 28 28 28 12 16 16 20 20 24 20 20 20 20 24 24 28 28
450 500 550 600 650 700 450 500 550 600 650 700 450 500 450 500 550 600 650 700
Height H1 (17.72) (19.69) (21.65) (23.62) (25.59) (27.56) (17.72) (19.69) (21.65) (23.62) (25.59) (27.56) (17.72) (19.69) (17.72) (19.69) (21.65) (23.62) (25.59) (27.56)
426 474 529 584 633 682 435 486 536 589 645 697 426 474 426 474 529 584 633 682
Height H2 (16.77) (18.66) (20.83) (22.99) (24.92) (26.85) (17.13) (19.13) (21.10) (23.19) (25.39) (27.44) (16.77) (18.66) (16.77) (18.66) (20.83) (22.99) (24.92) (26.85)
930 1028 1133 1238 1337 1436 939 1040 1140 1243 1349 1451 930 1028 930 1028 1133 1238 1337 1436
Max. Height Hr (36.61) (40.47) (44.61) (48.74) (52.64) (56.54) (36.97) (40.94) (44.88) (48.94) (53.11) (57.13) (36.61) (40.47) (36.61) (40.47) (44.61) (48.74) (52.64) (56.54)
260 400 510 680 870 1200 245 300 450 620 770 980 360 450 253 382 490 659 844 1160
Weight kg (lb)*1 (573.2) (881.8) (1124.4) (1499.1) (1918.0) (2645.5) (540.1) (661.4) (992.1) (1366.9) (1697.6) (2160.5) (793.7) (992.1) (557.8) (842.2) (1080.3) (1452.8) (1860.7) (2557.4)
*1: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached. F48.EPS
6-50 IM 01E20D01-01E
6. OUTLINE
H
Width S (7.87) (7.87) (7.87)
1280 1380 1540
Support Length Y (50.39) (54.33) (60.63)
H1 1211 1261 1366
Interval L1 (47.68) (49.65) (53.78)
720 2 780 2 880 2
16-øR P (28.35) (30.71) (34.65)
(t)
1100 3 1200 3 1350 3
Remote Mounting Interval Q (43.31) (47.24) (53.15)
T T P Flowtube Bolt T 130 1 130 1 130 1
(5.12) (5.12) (5.12)
S S Q 19 19 19
L1 Y Hole dia. øR (0.75) (0.75) (0.75)
L 1366 1470 1642
Outside dia. øD (53.78) (57.87) (64.65)
78 80 82
Thickness t*1 (3.07) (3.15) (3.23)
Flange Detail Pitch circle dia. øC
1283 1387 1552
(50.51) (54.61) (61.10)
(t) Bolt hole interval θ˚ 7.5 6.4 6.4
33 33 39
Hole dia. øh (1.30) (1.30) (1.54)
24 28 28
Gasket Material Number of holes N (0.94) (1.10) (1.10)
+10 +10 +10
Butadiene-styrene-rubber 950 0 1000 0 1100 0
Height H1
(37.40) (39.37) (43.31)
+Natural Rubber 1698 1803 1974
Max. Height H (66.85) (70.98) (77.72)
Material SS400 Weight kg (lb) *2
1650 1910 2420
Material SUS316 (3637.6) (4210.8) (5335.2)
*1: The value before the gasket is tightened.
*2: When submersible type or option code DHC is selected,
waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F45.EPS
(73.23)
Shackle
Size B SB30
200
H1
Width S (7.87)
1700
Support Length Y (66.93)
16-øR
(t) 1490
Interval L1 (58.66)
980 2
P (38.58)
T T P 1500 3
S S Q Remote Mounting Interval Q (59.06)
L1
Flowtube Bolt T 130 1
Y (5.12)
L 19
Hole dia. øR (0.75)
1800
Outside dia. øD (70.87)
Flange Detail 85
Thickness t*1 (3.35)
(t) 1710
Pitch circle dia. øC (67.32)
Bolt hole interval θ˚ 5.6
39
Gasket Material Hole dia. øh (1.54)
Butadiene-styrene-rubber 32
Number of holes N (1.26)
+Natural Rubber 1200 0
+16
Height H1
(47.24)
Material SS400 2155
Material SUS316 Max. Height H (84.84)
3150
Weight kg (lb) *2 (6944.6)
*1: The value before the gasket is tightened.
*2: When submersible type or option code DHC is selected,
waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F46.EPS
6-51 IM 01E20D01-01E
6. OUTLINE
AXF16L
AXF18L Unit : mm (approx. inch)
AXF20L G
NNUL 1S CG11
AXF22L W
AXF24L Process Connection CG1(JIS F12)
AXF26L Size code 16L 18L 20L 22L 24L 26L
Model 1600 1800 2000 2200 2400 2600
Size (64) (72) (80) (88) (96) (104)
A1 A2
Lining code U U U U U U
0 0 0 0 0 0
Face-to-face 2400 -10 2610 -10 2800 -10 2970 -10 3120 -10 3300 -10
length L
Eye-nut B θ˚ (94.49) (102.76) (110.24) (116.93) (122.83) (129.92)
N-øh 1834 2022 2191 2325 2421 2550
Interval 1 A1 (72.20) (79.61) (86.26) (91.54) (95.31) (100.39)
1872 2078 2300 2520 2724 2946
Eye-nut Interval 2 A2 (73.70) (81.81) (90.55) (99.21) (107.24) (115.98)
Size B M48 M48 M48 M48 M64 M64
øD
250 250 250 300 300 300
øC Width
H
Flameproof Packing Adapter for TIIS Flameproof Type (Optional code G12 or G11)
T1
Unit : mm
Adapter body(M. Screw) (Approx. inch)
O-Ring
Packing case
16.5(0.65)
Hexagon socket set screw
Packing *
18(0.71) Hexagon socket set screw
O-Ring
C
F L O-Ring
G Washer
Union nut
Packing gland
*Packing D
Clamp ring
(Choose from the table below
depend on cable outside diameter) Clamp nut
T2 O-Ring
B.coupling
Cable(user's scope) F43.EPS
6-52 IM 01E20D01-01E
6. OUTLINE
Unless otherwise specified, difference in the dimensions are refer to the following table.
1, 01 350
2, 2 0
2, 02 00
0 00
0 0
20 ,0
80 ,6
60 ,4
50 1,
1, 000
1, 00
1,100
60 0
0
0
90 0
0
5
400
0
1, 0
0
0
5
700
65
32
2
25
80
50
20
40
50
80
25
2.
35
30
15
10
12
5
15
10
10.0
5.0
Flow velocity
(m/s)
1.0
0.5
0.3
0.1
0.001 0.01 0.1 1 10 100 1,000 10,000 100,000
Flowrate (m3 /h)
64 60
5
8072
44 0
1.2
0.1
4
1.5
0.4
0.5
0.2
2.5
24
54
28
20
32
16
88
48
14
36
12
96
10
10
5
2
3
1
6
4
33
Flow velocity
(ft/s)
10
5.0
2.0
1.0
0.33
0.01 0.1 1.0 10 100 1,000 10,000 100,000 1,000,000
F44.EPS
Flowrate (GPM)
6-53 IM 01E20D01-01E
7. PED (PRESSURE EQUIPMENT DIRECTIVE)
Module: H
Type of Equipment: Piping (2) Installation
Type of Fluid: Liquids and gas
Group of Fluid: 1 and 2 WARNING
General-purpose Use / Submersible Type / Explosion • Tighten the bolts of the piping joints according
Proof Type to the prescribed torque values.
MODEL
DN PS PS DN
CATEGORY(*2)
• Take measures to protect the flowmeters from
(mm) (*1) (MPa) (*1) (MPa · mm)
forces caused by vibration channeled through
Article 3, (*3)
AXF002G/C 2.5 4 10
paragraph 3
the piping.
Article 3, (*3)
AXF005G/C 5 4 20
paragraph 3
Article 3, (*3) (3) Operation
AXF010G/C 10 4 40
paragraph 3
Article 3, (*3)
AXF015G/W/C 15 4 60
paragraph 3 WARNING
Article 3, (*3)
AXF025G/W/C 25 4 100
paragraph 3 • The instrument should be operated with the
AXF032G/W/C 32 4 128 II temperature and pressure of the fluid under
AXF040G/W/C 40 4 160 II normal operating conditions.
AXF050G/W/C 50 4 200 II • The ambient temperature should be that of
AXF065G/W/C 65 2 130 II normal operating conditions.
AXF080G/W/C 80 2 160 II • Take measures to prevent excessive pressure
AXF100G/W/C 100 2 200 II such as water hammer, etc. To avoid water
AXF125G/W/C 125 2 250 II hammer prevent the pressure from exceeding
AXF150G/W/C 150 2 300 II the PS (maximum allowable pressure) by
AXF200G/W/C 200 2 400 III setting the system’s safety valves, etc. appro-
AXF250G/W/C 250 2 500 III priately.
AXF300G/W/C 300 2 600 III • Should external fire occur, take safety mea-
AXF350G/W/C 350 1 350 II sures at the device itself or system-wide
AXF400G/W/C 400 1 400 III prevent it having an effect on the flowmeters.
T0701.EPS
• Avoid using fluids exceeding the corrosion
Sanitary Type proof limitations of the lining and electrodes.
• Take measures not to abrade the metal pipe,
DN PS PS D
MODEL CATEGORY (*2) and avoid abrading the lining by using fluids
(mm) (*1) (MPa) (*1) (MPa · mm)
Article 3, (*3) such as slurry and sand are contained.
AXF015H 15 1 15
paragraph 3
Article 3, (*3)
AXF025H 25 1 25
paragraph 3
AXF032H 32 1 32 I
AXF040H 40 1 40 I
AXF050H 50 1 50 I
AXF065H 65 1 65 I
AXF080H 80 1 80 I
AXF100H 100 1 100 I
AXF125H 125 1 125 II
T0702.EPS
7-1 IM 01E20D01-01E
8. EXPLOSION PROTECTED TYPE INSTRUMENT
The covers of explosion proof type products are T3 +130°C (+266°F) –40°C (–40°F)
T0801.EPS
locked. Use the attached hexagonal wrench to
Ambient Temp.: –40°C to +60°C (–40°F to +140°F)
open and close the cover. Before opening the
cover, be sure to check that the power of ATEX Type of Protection “Dust”
flowmeter has been turned off. Once the cover is Group: II
closed, be sure to re-lock the product. Category: 1D
Be sure to lock the cover with the special screw Electrode Circuit Um: 250 Vac/dc
using the hexagonal wrench attached after Maximum power supply voltage: 250 Vac/130 Vdc
tightening the cover. Excitation Circuit: 140V max
Enclosure: IP66, IP67
8-1 IM 01E20D01-01E
8. EXPLOSION PROTECTED TYPE INSTRUMENT
8-2 IM 01E20D01-01E
8. EXPLOSION PROTECTED TYPE INSTRUMENT
(Remote Flowtube)
L
H
MPa MAX.
˚C
˚C
COMB.NO.
mm
Made in *3)
8-3 IM 01E20D01-01E
8. EXPLOSION PROTECTED TYPE INSTRUMENT
8-4 IM 01E20D01-01E
8. EXPLOSION PROTECTED TYPE INSTRUMENT
8.3 CSA T6
T5
+70°C (+158°F)
+85°C (+185°F)
–40°C (–40°F)
–40°C (–40°F)
(1) Technical Data
T4 +120°C (+248°F) –40°C (–40°F)
*AXF002C – AXF400C
T3 +130°C (+266°F) –40°C (–40°F)
Applicable Standard: T0807.EPS
For CSA C22.2 Series; Ambient Temp.: –40°C to +60°C (–40°F to +140°F)
C22.2 No 0, C22.2 No 0.4, C22.2 No 0.5,
C22.2 No 25, C22.2 No 30, C22.2 No 94,
C22.2 No 157, C22.2 No 1010.1 (Remote Flowtube)
For CSA E79 Series; For CSA C22.2 Series
CAN/CSA-E79-0, CAN/CSA-E79-1, Explosion proof for Class I, Division 1, Groups A,
CAN/CSA-E79-7, CAN/CSA-E79-11, B, C & D.
CAN/CSA-E79-18 Dust-ignition proof for Class II/III, Division 1,
Certificate: 1481213
Groups E, F & G.
Process Sealing Certification:
Intrinsically safe (electrodes) for Class I, Division 1,
Dual Seal certified by CSA to the requirements of
ANSI/ISA 12.27.01. Groups A, B, C & D.
No additional sealing required. “SEAL ALL CONDUITS WITHIN 50 cm OF THE
Primary seal failure annunciation; ENCLOSURE”
Deterioration of the flowrate output at nonzero “WHEN INSTALLED IN DIV. 2, SEALS NOT
flow point. REQUIRED”
Unstable flowrate output at zero flow point.
Electrode Circuit Um: 250 Vac/dc
(Integral Flowmeter) Excitation Circuit: 170V max
For CSA C22. 2 Series
Enclosure: Type 4X
Explosion proof for Class I, Division 1, Groups A, Temperature Code:
B, C & D.
Dust-ignition proof for Class II/III, Division 1, Temperature Maximum Process Minimum Process
Code Temperature Temperature
Groups E, F & G.
Intrinsically safe (electrodes) for Class I, Division 1, T6 +70°C (+158°F) –40°C (–40°F)
WARNING WARNING
• All wiring shall comply with Canadian Electrical WARNING : DE-ENERGIZE BEFORE OPEN-
Code Part I and Local Electrical Codes. ING.
• In hazardous location, wiring shall be in conduit OUVRIR LE CIRCUIT AVANT
as shown in Figure 8.3.1. D’ENLEVER LE COUVERCLE.
WARNING : SEAL ALL CONDUITS WITHIN • Take care not to generate mechanical spark
50cm OF THE ENCLOSURE’. when access to the instrument and peripheral
UN SCELLEMENT DOIT ÉTRE devices in hazardous locations.
INSTALLÉ À MOINS DE 50cm DU
BOÎTIER.
• When installed in Division 2, “SEALS NOT
REQUIRED” (4) Maintenance and Repair
WARNING
For CSA E79 Series
The instrument modification or parts replacement
WARNING by other than authorized representative of
YOKOGAWA Electric Corporation or
• All wiring shall comply with local installation
YOKOGAWA Corporation of AMERICA is
requirements and local electrical code.
prohibited and will void Canadian Standards
• In hazadous locations, the cable entry devices
Explosionproof Certification.
shall be of a certified flameproof type, suitable
for the conditions of use and correctly installed.
HAZARDOUS LOCATIONS
• Unused apertures shall be closed with suitable Sealing Fitting
Conduit
flameproof certified blanking elements. (The Magnetic Flowmeter
plug attached is flameproof certified.)
(3) Operation
For CSA C22.2 Series 50 cm Max.
F0803.EPS
8-6 IM 01E20D01-01E
8. EXPLOSION PROTECTED TYPE INSTRUMENT
(Remote Flowtube)
8.4 IECEx IECEx Flameproof Type
Ex demb[ia] IIC T6...T3
WARNING Electrode Circuit Um: 250 Vac/dc
Excitation Circuit: 170V max
Only trained persons use this instrument in Enclosure: IP66, IP67
industrial locations. Temperature Class:
Temperature
Process Temperature
Class
(1) Technical Data
T6 –40°C to +70°C (–40°F to +158°F)
*AXF002C – AXF400C
Applicable Standard: T5 –40°C to +85°C (–40°F to +185°F)
(Integral Flowmeter)
IECEx Flameproof Type IECEx Type of Protection "Dust"
Ex demb[ia] IIC T6...T3 Ex tD A21 IP6x T95˚C, T105˚C, T120˚C, T135˚C
Electrode Circuit Um: 250 Vac/dc Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc Excitation Circuit: 170V max
Excitation Circuit: 140V max Enclosure: IP66, IP67
Enclosure: IP66, IP67 Maximum surface temperature:
Temperature Class: Maximum Surface
Process Temperature
Temperature
Temperature
Process Temperature T95°C (+203°F) –40°C to +70°C (–40°F to +158°F)
Class
T6 –40°C to +70°C (–40°F to +158°F) T105°C (+221°F) –40°C to +85°C (–40°F to +185°F)
T5 –40°C to +85°C (–40°F to +185°F) T120°C (+248°F) –40°C to +120°C (–40°F to +248°F)
T4 –40°C to +120°C (–40°F to +248°F) T135°C (+275°F) –40°C to +150°C (–40°F to +302°F)
T0812.EPS
T3 –40°C to +130°C (–40°F to +266°F)
T0809.EPS
Ambient Temp.:
Ambient Temp.: PFA Lining; –40˚C to +60˚C (–40˚F to +140˚F)
PFA Lining; –40˚C to +60˚C (–40˚F to +140˚F) Ceramics Lining; –15˚C to +60˚C (+5˚F to +140˚F)
Ceramics Lining; –15˚C to +60˚C (+5˚F to +140˚F)
(2) Installation
IECEx Type of Protection "Dust"
Ex tD A21 IP6x T95˚C, T105˚C, T120˚C, T130˚C WARNING
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc • All wiring shall comply with local installation
Excitation Circuit: 140V max requirements and local electrical code.
Enclosure: IP66, IP67 • In hazardous locations, the cable entry devices
Maximum surface temperature: shall be of a certified IECEx flameproof type,
suitable for the conditions of use and correctly
Maximum Surface
Temperature
Process Temperature installed.
T95°C (+203°F) –40°C to +70°C (–40°F to +158°F)
• Unused apertures shall be closed with suitable
flameproof certified blanking elements. (The
T105°C (+221°F) –40°C to +85°C (–40°F to +185°F)
plug attached is certified as the flameproof and
T120°C (+248°F) –40°C to +120°C (–40°F to +248°F)
IP66 or IP67 as a part of this apparatus.)
T130°C (+266°F) –40°C to +130°C (–40°F to +266°F)
• In case of ANSI 1/2 NPT plug, ANSI hexagonal
T0810.EPS
WARNING WARNING
In case that ambient temperature exceeds 50°C, For the TIIS flameproof type with wiring using a
use heat-resistant cables with maximum allow- flameproof packing adapter, wire cables through
able temperature of 70°C or above. the packing adapters approved by Yokogawa
(optional code G12 or G11).
(Remote Flowtube)
• Construction: Ex de[ia] IIC T4
: Terminal box ; Explosion proof Unit : mm
T1 (Approx. inch)
Flowtube; Increased Safety and
Adapter body(M. Screw)
Intrinsically Safety(ia) O-Ring
Packing case
Electrode; Intrinsically Safety(ia) 16.5(0.65) Hexagon socket set screw
Packing *
Um=250VAC 50/60Hz, 250VDC, 18(0.71) Hexagon socket set screw
Uo=250V*, Io=3.37mA*, Po=0.211W L
O-Ring
C
F O-Ring
*Uo and Io are rms value. G Washer
Union nut
• Ignition and Explosion class of gas or vapour: IIC T4 Packing gland
*Packing Clamp ring D
• Ambient Temperature: –20 to 60°C (Choose from the table below Clamp nut
• Fluid Temperature: 120°C max depend on cable outside diameter) T2 O-Ring
B.coupling
• Electrode Circuit: 250 V AC/DC Cable(user's scope)
• Grounding: JIS Class C(grouding resistance 10Ω or
Dimension Packing diameter Identification Weight
less) or JIS Class A(grounding resistance 10Ω Cable outer diameter mark kg
T1 T2 C D L F G (lb)
or less) 10.0(0.39)
35 39 94.5 8.0 to 10.0 (0.31 to 0.39) 20.0 16 8-10 0.26
G 1/2 G 1/2
(1.38) (1.54) (3.72) 10.0 to 12.0 (0.39 to 0.47) 12.0(0.47) (0.79) 16 10-12 (0.57)
F0809.EPS
WARNING
Figure 8.4.1 Flameproof Packing Adapter
* In case of TIIS Flameproof type, a remote
flowtube is available for combined use with the
• Apply a nonhardening sealant to the terminal box
AXFA14 only.
connection port and to the threads on the flameproof
packing adapter for waterproofing.
• The same wiring as described below is required for
WARNING both of the terminal box connection ports except
when a four-wire cable is used for power input and
In case that ambient temperature exceeds 50°C, signal output with DC power supply in the Integral
use heat-resistant cables with maximum flowmeter.
allowable temperature of 70°C or above.
8-9 IM 01E20D01-01E
8. EXPLOSION PROTECTED TYPE INSTRUMENT
Adapter body
Drain plug O-ring
F0810.EPS Apply a non-
hardnening sealant to
Figure 8.4.2 Typical Wiring Using Flexible Metal Conduit the threads for
waterproofing.
F0811.EPS
Follow the procedure for flameproof packing adapter Figure 8.4.3 Installing Flameproof Packing Adapter
setting. (refer to Figure 8.4.3)
8-10 IM 01E20D01-01E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS FLAMEPROOF EQUIPMENT
2. Electrical Apparatus of Flame- Note: The permissible sizes of gaps between joint surfaces, the path
proof Type of Explosion-Pro- length of a joint surface and the number of joint threads are
determined by such factors as the enclosure’s internal
tected Construction volume, joint and mating surface construction, and the
explosion classification of the specified gases and vapours.
Electrical apparatus which is of flameproof construction is
subjected to a type test and certified by the Japanese Ministry
of Labour aiming at preventing explosion caused by electrical
apparatus in a factory or any location where inflammable
1 EX-B03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS FLAMEPROOF EQUIPMENT
2 EX-B03E
INSTALLATION AND OPERATING PRECAUTIONS FOR TIIS FLAMEPROOF EQUIPMENT
CAUTION
3 EX-B03E
REVISION RECORD
Title: AXF Magnetic Flowmeter Integral Flowmeter/Remote Flowtube
[Hardware Edition]
Manual No.: IM 01E20D01-01E
Edition Date Page Revised Item
6th Oct. 2005 3-22 to 3-24 Added the "Confirmation of adapter mounting screw"
5-1 (7) Added the sentence.
5-5 Added the note of installation
6-7 Added the note of permeable fluids to the grounding ring/grounding
electrode.
6-12 Added the "IECEx" to the hazardous area classification.
6-14 Added the "SF2 (IECEx)" to the withstand voltage.
6-21 to 6-27 Changed the note for wetted parts material.
6-28 MU Changed description of the "mass unit setting (MU)".
6-29 PM Changed the value of the mirror finished PFA lining (PM).
Changed from "Ra 0.05 to 0.15 µm" to "Size 15 to 200 mm :
Ra 0.05 to 0.15 µm, Size 250 to 400 mm : Ra 0.05 to 0.25 µm"
6-31 SF2 Added the "IECEx Certification (SF2)" to the optional specifications.
6-39 Corrected the "Weight kg (lb)" of BD4.
6-40 Corrected the "Weight kg (lb)" of BJ1/CJ1, BJ2/CJ2 and BG1/CG1.
6-48 Corrected the "Inner diameter of grounding ring (ød)" of size 250mm.
8-7 to 8-8 8.5 Added the "IECEx".
7th June 2006 1-1 Added the postscript about FOUNDATION Fieldbus protocol type.
3-11 Table 3.3.7 Changed the Torque Value.
3-12 Table 3.3.8 Changed the Torque Value.
3-13 Table 3.3.9 Changed the Torque Value.
3-16 Table 3.3.14 Changed the Torque Value.
3-17 Table 3.3.15 Changed the Torque Value.
3-18 Table 3.3.16 Changed the Torque Value.
3-22 5) Added the sentence.
3-23 5) Added the sentence.
3-24 5) Added the sentence.
Changed the Figure 3.3.10.
4-3 (2) Added the postscript about FOUNDATION Fieldbus protocol type.
4-6 (6) Added the postscript about FOUNDATION Fieldbus protocol type.
4-13 1), 2) Added the “IECEx”.
5-5 5.3 Added the CAUTION.
Changed the Figure 5.3.1.
5-8 5.4.3 Added the postscript about FOUNDATION Fieldbus protocol type.
5-9 5.5.1 Added the postscript about FOUNDATION Fieldbus protocol type.
5-10 5.5.2 Added the postscript about FOUNDATION Fieldbus protocol type.
6-1, 14, 20, 28 Added the postscript about FOUNDATION Fieldbus protocol type.
6-7 Corrected from “SPCC” to “SS400” in Housing of size 500 mm to
2600 mm
6-17 to 20 Corrected the value of Pressure on figure.
6-21 to 6-23 Added the “-F” (FOUNDATION Fieldbus protocol).
6-25 to 6-27 Added the “-F” (FOUNDATION Fieldbus protocol).
6-34 to 6-48 Added the postscript about FOUNDATION Fieldbus protocol type.
IM 01E20D01-01E
Edition Date Page Revised Item
8th Apr. 2007 1-2 (4) Changed the warning note of "Maintenance".
1-5 1.4 Added the ATEX documentation.
4-3, 4-10 4.3 Added the important note for the wiring ports.
4-4 (4) Added the important note for a 24 V power supply version (power
supply code 2).
2) Corrected the graph of "Supply Voltage and Cable Length".
5-1 5 Changed the warning note of "Maintenance".
5-7 5.4 Changed the warning note and important note of "Maintenance".
5.4.1 Changed the caution note of "Fuse Replacement".
Deleted the Figure 5.4.1.
5-7 to 5-8 5.4.2 Changed the content of section "Display Unit Replacement" to
"Changing the Direction of the Display Unit".
5-8 5.4.3 Deleted the section "Amplifier Replacement".
5-9 5.5 Added the important note for "Setting of Switches".
6-1 Deleted the "Communication Distance" and "Cable length
for Specific Applications" from item of "HART".
6-13 Changed the certificate number of TIIS.
6-15 Corrected the graph of "Supply Power and Cable Length for Power
Supply Code 2".
6-16 Corrected the value of span flow rate of size 32 mm.
6-17 Corrected the value of span flow rate of size 1.25 inch and size 36
inch.
6-28 Added the note to optional codes C1, C2, C3, EM, and G13.
6-29 Changed the gasket part number of manufacturer for optional code GC.
6-30 Changed the descriptive text of HF1.
6-31 Corrected the value of span flow rate of size 32 mm.
Added the note to optional codes C1, C2, C3, EM, and G13.
8-8 8.5 (1) Changed the certificate number of TIIS.
9th Jan. 2008 1-1 Added the postscript about PROFIBUS PA protocol type.
1-2 1.1 Added the warning note on "write protect".
4-3 4.1.4 (2) Added the postscript about PROFIBUS PA protocol type.
4-6 Added the postscript about PROFIBUS PA protocol type.
5-11 5.8.1 Changed the flow chart.
6-1 Added the postscript about PROFIBUS PA protocol type.
6-7 Corrected the frequency of "Input Impedance of Communicating
Device".
Changed the material of housing from "Carbon steel-JIS SPCC" to
"Carbon steel-JIS SPCC equivalent".
Corrected from "EN1.4404 equivalent" to "EN1.4401 equivalent" in
grounding ring of size 1100mm to 2600mm.
6-15 Added the CAUTION to "EMC Conformity Standard".
Changed the note of "Fluid Conductivity".
6-19 Changed the unit of "Vibration Conditions" from "G" to "m/s2".
6-21 to 6-32 Added the "-G" (PROFIBUS PA protocol).
Added the note of process connection code.
Added the note of calibration code C.
Added the applicable size for the electrode material code W.
6-33 to 6-36 Added the code F and G to item name.
Changed the value of mirror finished ceramics for optional code CM.
6-39 to 6-53 Added the code G.
IM 01E20D01-01E
Edition Date Page Revised Item
IM 01E20D01-01E
6-33 Corrected the sentence of /MU.
6-34 Changed the selectable range of flow rate span for optional code
"SC".
6-43 Correced the figure.
6-49 Corrected the value of ød of CG1.
6-53 Moved the item of Recommended Gaskets Between Flowtubes And
User's Flanges to P6-8.
8-1 Added the year to applicable standard numbers of CENELEC ATEX.
8-5 Added the "Process Sealing Certification".
8-8 Changed the certificate numbers of TIIS Flameproof.
IM 01E20D01-01E