You are on page 1of 13

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/324733001

Measurement uncertainty and metrological traceability of whole blood


cyclosporin A mass concentration results obtained by UHPLC-MS/MS

Article  in  Clinical Chemistry and Laboratory Medicine · April 2018


DOI: 10.1515/cclm-2018-0120

CITATIONS READS

0 231

3 authors, including:

Pedro Alía Francesca Canalias


Hospital Universitari de Bellvitge Autonomous University of Barcelona
35 PUBLICATIONS   387 CITATIONS    69 PUBLICATIONS   426 CITATIONS   

SEE PROFILE SEE PROFILE

Some of the authors of this publication are also working on these related projects:

metrological traceability of enzyme measurements View project

All content following this page was uploaded by Francesca Canalias on 22 June 2018.

The user has requested enhancement of the downloaded file.


Clin Chem Lab Med 2018; aop

Raül Rigo-Bonnin*, Pedro Alía and Francesca Canalias

Measurement uncertainty and metrological


traceability of whole blood cyclosporin A mass
concentration results obtained by UHPLC-MS/MS
https://doi.org/10.1515/cclm-2018-0120 quite similar. This fact would confirm that the top-down
Received December 21, 2017; accepted February 21, 2018 approach could be sufficient for estimating uncertainty of
Abstract CsA mass concentrations in whole blood results in clini-
cal laboratories. Finally, we hope that this study can help
Background: Traceable and accurate results of cyclo- and motivate clinical laboratories to describe metrologi-
sporine A (CsA) mass concentrations in whole blood are cal traceability and to perform measurement uncertainty
required to ensure the monitoring of immunosuppressive studies based on the simpler top-down approach.
therapy in transplant recipients. Metrological traceability
Keywords: bottom-up; cyclosporine A; top-down; trace-
and measurement uncertainty can allow ensuring reliabil-
ability; ultra-high-performance liquid chromatography-
ity and comparability of these results over time and space.
tandem mass spectrometry (UHPLC-MS/MS); uncertainty.
In this study, we provide a practical and detailed example
of how the traceability and uncertainty of mass con-
centration of CsA results, obtained using an ultra-high-­
performance liquid chromatography-tandem mass spec- Introduction
trometry (UHPLC-MS/MS) procedure, can be described
and estimated. Cyclosporine A (CsA) is an immunosuppressive drug that is
Methods: Traceability was described mainly according to widely administered to recipients of solid organ transplants
ISO 17511 and information obtained from certificates facil- [1–3]. Therapeutic drug monitoring of CsA in these patients
itated with the manufacturer’s calibrators. Uncertainty is required because of the narrow therapeutic intervals
estimation was performed using the bottom-up and top- and significant interindividual variability in whole blood
down approaches. For the bottom-up approach, the most concentrations. Trough mass concentration of CsA in
relevant sources of uncertainty were identified and later whole blood (cCsA) is commonly monitored and generally
used to estimate the standard, combined and expanded regarded as a good surrogate for CsA exposure [1–3]. Thus,
uncertainties. For the top-down approach, expanded dose adjustments, critical to regulate the appropriate level
uncertainty was estimated directly using intralab quality of immunosuppression, are made in part based on labora-
control data mainly. tory results. Thereby, traceable and accurate results of cCsA
Results: Mass concentration of CsA results was trace- should be provided by laboratories to ensure the monitor-
able to the manufacturer’s product calibrators used to ing of immunosuppressive therapy. Application of concepts
calibrate the UHPLC-MS/MS procedure. The expanded like metrological traceability and measurement uncertainty
uncertainties estimated by the bottom-up and top-down can help ensure reliability and comparability of these
approaches were 7.4% and 7.2%, respectively. results over time and space. Additionally, the knowledge of
Conclusions: After performing the bottom-up and top- metrological traceability of the results and the estimation
down approaches, we observed that their results were of measurement uncertainty of measured values are man-
datory for clinical laboratories aiming to achieve ISO 15189
compliance and accreditation [4].
*Corresponding author: Raül Rigo-Bonnin, Laboratori Clínic, Metrological traceability is defined as “property of a
IDIBELL, Hospital Universitari de Bellvitge, L’Hospitalet de Llobregat, measurement result whereby the result can be related to a
Barcelona, Spain, Phone: +34932607543, Fax: +34932607546, reference through a documented unbroken chain of calibra-
E-mail: raulr@bellvitgehospital.cat tions, each contributing to the measurement uncertainty”
Pedro Alía: Laboratori Clínic, IDIBELL, Hospital Universitari de
[5]. Ideally, traceability should be materialized by relat-
Bellvitge, L’Hospitalet de Llobregat, Barcelona, Spain
Francesca Canalias: Laboratori de Referència d’Enzimologia Clínica,
ing a measurement result to a stated reference through
Departament de Bioquímica i Biologia Molecular, Universitat an unbroken chain of calibrations. Stated references may
Autònoma de Barcelona, Bellaterra, Spain range from a standard to a certified primary reference

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
2      Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS

material that embodies a unit of the International System was purchased from Recipe (Darmstadt, Munich, Germany). To carry
of Units (SI). These reference materials must have certain out the experimental part of this study, the Liquicheck™ Whole Blood
Immunosuppressant Control Level 2 (lot: 26242) from Bio-Rad Labora-
well-defined characteristics, such as homogeneity and
tories (Hercules, CA, USA) was used, as well as blood samples from
stability, as described in the ISO 17511 [6]. Reference meas- patients treated with cCsA having values near to those of the quality
urement procedures must be characterized, validated and control. Calibrator and quality control materials were reconstituted
documented according to specifications outlined in ISO with LC-water and stored as recommended by manufacturers.
15193 [7]. Different sets of ClinMass® internal standard (IS), lyophilized, for
immunosuppressants (lot: 1312) were supplied by Recipe, containing
The measurement uncertainty can provide a quanti-
CsA-D12 (purity and isotopically purity ≥ 98%). The working IS solu-
tative indication of the quality and accuracy of measure- tion (CsA-D12 25 μg/L) was prepared in acetonitrile and stored as indi-
ment values. Measurement uncertainty is a non-negative cated by Rigo-Bonnin et al. [12].
parameter characterizing the dispersion of the quantity
values being attributed to the measurand based on the
information used [5]. There are two different approaches Measurement procedure and equipment
for the estimation of uncertainty: the bottom-up and the
cCsA was measured using a previously described procedure [12]
top-down [8–10]. In the bottom-up approach, all conceiva-
based on an UHPLC-MS/MS. The measurement system used was
ble sources of uncertainty are fundamentally considered, Acquity® UPLC®-TQD® (Waters, Milford, MA, USA). Furthermore, the
and standard uncertainty is estimated either by a direct following equipment was used: ADA-120/L analytical balance (Adam
experiment (type A) or from other sources of informa- Equipment, Bletchley, UK), Biofuge 13 centrifuge (Heraeus Holding
tion (type B), or a combination of both to obtain then a GmbH, Hanau, Germany), Acura® 825 adjustable 100–1000 μL vol-
combined uncertainty for, finally, estimate an expanded ume micropipette (Socorex Isba, Ecublens, Switzerland), Nichipet®
EX II adjustable 20–100 μL volume micropipette, Nichimate® ­Stepper
uncertainty of a measured value. The top-down approach
repetitive dispenser (Nichiryo Co. Ltd., Koshigaya-shi, Saitama,
considers the uncertainty as a whole. Uncertainties are Japan) and 1000-mL and 250-mL BLAUBRAND® volumetric flasks
evaluated directly using validation data or intralab quality (BRAND GmbH + Co. KG, Wertheim, Germany).
control data produced by a measurement system.
Unfortunately, no high-order reference material or
reference measurement procedure exists for cCsA in the
Joint Committee for Traceability in Laboratory Medicine
Methodology
database [11]. This fact compromises the reliability and
comparability of cCsA results and, consequently, hinders Metrological traceability
their interpretation. The aim of this study was to describe
the measurement traceability of cCsA results obtained by Description of metrological traceability was based on the
a measurement system based on ultra-high-performance ISO 17511 [6] and IUPAC [13]. Information related to the
liquid chromatography (UHPLC)-tandem mass spectro- selection of metrological references, calibration hierarchy
metry (MS/MS). We also aimed to provide to clinical labo- and suitably validated measurement procedures, acqui-
ratory specialists a detailed estimation of the uncertainty sition and verification of end user’s calibrators and end
of cCsA values using the bottom-up and the top-down user’s measurement on system or sample to obtain meas-
approaches. urement results was considered. The information was
obtained from certificates and statements facilitated by
the calibrator’s manufacturer (Recipe).
Materials and methods
Measurement uncertainty
Reagents and materials
Bottom-up approach
LC-MS-grade ammonium acetate (purity ≥ 98%), zinc sulfate
(purity ≥ 99.5%) and formic acid (purity ≥ 98%) were purchased
The steps followed to estimate the uncertainty were as
from Sigma-Aldrich (St. Louis, MO, USA). Liquid chromatography
coupled to mass spectrometry (LC-MS)-grade acetonitrile (quality of
follows [8, 9]:
solvent ≥ 99.9%), LC-MS methanol (quality of solvent ≥ 99.9%) and
LC-MS water (quality of solvent ≥ 99.99%) were supplied by Merck
KGaA (Darmstadt, Germany).
Specification of the measurand
A commercially available lyophilized calibrator set ClinCal® The measurand is the quantity intended to be measured
Whole Blood Calibrators, for Immunosuppressants Levels 0–6 (lot: 405) [5]. According to this definition, the measurand should

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS      3

be well defined and sufficient information of the meas- where As is the peak area obtained for a patient sample;
urement should be detailed. Thus, we have described the AIS, the peak area obtained for the IS in the sample; cIS, the
quantity according to the IUPAC and IFCC recommenda- mass concentration of IS in the sample preparation solu-
tions [14]. tion; aw, the intercept of the weighted regression line of
the calibration curve; bw, the slope of the weighted regres-
sion line of the calibration curve; fsol, the factor for the
Identification of uncertainty sources
solutions preparation; fsp, the factor for the sample prepa-
The bottom-up approach requires the development of
ration; fMS, the factor for mass spectrometer parameters;
a measurement function in which the relation between
fUHPLC, the factor for chromatograph parameters; fdata, the
measurand values (dependent variable [y]) and rel-
factor for data processing; and fδ, correction factors for
evant influencing factors (independent variables [x]) is
biases.
described. We used the cause and effect diagram [9–11]
to identify the different sources of uncertainty, as shown
in Figure 1. Considering that the influencing factors may Estimation of standard uncertainties
themselves be viewed as measurands and depend on Standard uncertainty is expressed as a standard deviation
other quantities, we proposed only an approximation of or as a relative standard deviation. Therefore, the standard
measurement function: uncertainties for each uncertainty source were estimated
 AS  by choosing an appropriate distribution of their values, by
 A ⋅ cIS − aw  using either a type A or a type B evaluation. A type A eval-
cCsA (µg / L) =  IS 
 ⋅ fsol ⋅ fsp ⋅ fMS ⋅ fUHPLC ⋅ fdp ⋅ fδ uation is based on statistical analysis of values obtained
 bw
from direct experiments – e.g. values from measurements

Figure 1: Cause and effect diagram of the most relevant measurement uncertainty sources of cyclosporin A mass concentration in whole
blood using the bottom-up approach.

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
4      Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS

repeated under defined conditions. However, it was not sw, the estimated standard deviation for x0; x̅w, the weight-
always possible to obtain experimental data for individ- ing mean values of xi; and y̅w, the weighting mean values
ual uncertainty sources, so a type B evaluation was used. of yi.
Therefore, helpful information needed for a type B evalua- To estimate ufitt, a blood sample from a patient having
tion was obtained from calibration certificates and manu- cCsA value near to the internal quality control was pro-
facturers’ data. cessed using three different calibration curves.
According to the cause and effect diagram (Figure 1),
standard uncertainties were estimated for calibration pro- Solutions preparation: Solutions needed to assess the
cedure, solutions preparation, samples preparation, mass measurement procedure were mobile phases A and B,
spectrometer parameters, UHPLC parameters, data pro- working IS solution and precipitation solution.
cessing and bias. Uncertainties associated with solutions preparation
were mainly due to the purity of reagents, the quality of
Calibration procedure: The standard uncertainty associ- solvents, the mass weighted into the balance, the volu-
ated with the calibration procedure is generally close to metric flasks and the pipetted volumes and the molar
the individual uncertainties of the commercial calibra- mass of reagents. Standard uncertainties were obtained
tors, which are mainly due to their assigned values, the from manufacturers’ data (purity of reagents and quality
pipetted volumes used for their reconstitution and the sta- of solvents) and calibration certificates (balance, pipettes
bility of reconstituted calibrators, and with the calibration and volumetric flasks). Furthermore, keeping in mind
curve fitting. that the evaporation of organic solvents used to prepare
According to the manufacturer’s data, the cCsA the working IS solution may lead to changing concentra-
assigned values and associated expanded uncertainties tion of IS over the time and affect to the cCsA values, an
(k = 2) were 25.8 ± 0.64, 49.0 ± 1.1, 95.7 ± 2.5, 181 ± 4, 439 ± 11 experiment was performed. Different aliquots from the
and 1243 ± 23 μg/L. same patient’s blood sample with a cCsA value near to the
The measuring volume from pipetting was i­ndicated internal quality control – preserved at −80 °C to minimize
by the certificate of external calibration as 1000 ± 2.6  μL the possible loss of stability, were processed in quintupli-
(k = 2). cate using fresh IS working solution and the same solution
The relative standard uncertainty related to the stabil- after 5 days (maximum time of conservation) at (2–8) °C.
ity of reconstituted calibrators was estimated from experi- Their uncertainty was estimated from the cCsA percentage
mental data, ≤5.9% for at least 6 months [12]. deviation obtained.
Furthermore, the standard uncertainty of the indi-
vidual measurements from weighted linear calibration Samples preparation: Standard uncertainties associ-
curves (ufitt) was calculated according to the following ated to samples preparation were related to the sample,
equation [15, 16]: the working IS and the precipitant solution pipetted
volumes, and the centrifugation was performed during the
sw 1 1 ( y0 − yw )2 protein precipitation process. Uncertainties were obtained
ufitt = ⋅ + +
bw w0 n b2 ⋅ n
from calibration certificates of dispensers, micropipettes
w ∑
( wi ⋅ xi2 − n ⋅ xw2 )
i =1
and centrifuges.
n

sw =
∑ i =1
wi ⋅ ( yi − (aw − bw ⋅ xi ))2
Mass spectrometer and UHPLC parameters: Several
n−2
mass spectrometer and UHPLC parameters were consid-
n n
ered as possible sources of uncertainty (Figure 1). Uncer-
xw =
∑ i =1
( w i ⋅ xi )
                yw =
∑ i =1
( w i ⋅ yi )
tainties associated to these parameters were obtained
n n
from manufacturer’s data.
where n is the number of experimental points used to
perform the calibration curve; xi, the calibrator values; Data processing: The CsA and CsA-D12 peak areas inte-
yi, the response values for calibrators; wi, the weighting gration, peak fitting and peak areas variability, as well as
factor (equal to 1/xi); aw, the y-weighting intercept of the the retention times peaks variability, were identified as the
linear regression equation; bw, the weighting slope of main sources of standard uncertainty related to data pro-
the linear regression equation; x0, the determined cCsA cessing. To estimate them, a blood sample from a patient
of unknown sample calculated through the calibration with cCsA value near to the internal quality control was
curve; y0, the response of x0; w0, the weighting factor for x0; processed in quintuplicate. The standard uncertainty

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS      5

related to peak integration (uinteg(y)) was calculated as the described in the measurement procedure should be
relative standard error of the mean values using pooled included.
relative standard deviations (sinteg(x)), of inter-replicate –– Perform replicates of the cCsA for each condition
data, given by the equation facilitated by the manufac- defined above.
turer of mass spectrometer [17]: –– Represent values graphically. The x-axis describes
the values of the uncertainty source, and the y-axis
m νi ⋅ sinteg
2
(x)
stands for cCsA in percentages.
∑ i =1 νi
uinteg ( y ) = –– Calculate the linear regression function of values. The
m slope of the function represents the sensitivity coef-
  n
 
2 ficient value, which is expressed as the percentage
n  ∑ i i 
( I ⋅ t ) of measurand value variation per uncertainty source
∑ i=1  Ii ⋅  ti − i=1 n   unit (e.g. X% of cCsA variation per °C on column
 
 ∑ i=1Ii  
sinteg ( x ) = chamber).
n
∑ i =1 i
I
A critical step in the sample preparation is associated
to the shaking mix time variability of the precipitated
where m is the total number of replicates; ti, the ­retention
solution. For this reason, the uncertainty source related
time value for the observation i; and Ii, the intensity
to this process was included in the uncertainty estima-
(signal: number of ions per second) associated to ti.
tion, and an experimental study to obtain the sensitivity
The standard uncertainty related to peak fitting
coefficient was performed. A blood sample from a patient
(ufitt (y)) was calculated as follows [17]:
with cCsA value near to the internal quality control was
uinteg ( y ) processed in triplicate at shaking mix times of 10, 20, 30,
ufitt ( y ) =
((2 ⋅WinHalfsize ) + 1)0.5 ⋅ Iterations0.25 40 and 50 s.
Furthermore, given the high degree of variability that
where Iterations is the number of smooths applied to the different mass spectrometer parameters (cone voltage
raw data, and WinHalfsize is the half-size of the smooth- and collision energy) and UHPLC parameters (column
ing window. chamber and autosampler temperatures) could contribute
The standard uncertainties related to peak areas and in the uncertainty estimation, sensitivity coefficients for
retention times variabilities were estimated as the relative these parameters were obtained experimentally. To esti-
standard error of the mean values using their respective mate them, different blood samples from patients with
standard deviations obtained in the replication study. cCsA values near to the internal quality control were pro-
cessed in triplicate at cone voltages of 15, 18, 22, 25 and 28
V; at collision energies of 11, 15, 19, 23 and 27; at column
Estimation of combined uncertainty
chamber temperatures of 40, 45, 50, 55 and 60 °C; and at
Once the contribution of uncertainty sources to the
autosampler temperatures of 10, 15, 20 and 25 °C.
overall uncertainty was quantified as relative standard
deviations, we combined them to give relative combined
standard uncertainty (uc (y)) according to the following Estimation of expanded uncertainty
equation: Relative expanded uncertainty (U) was obtained by mul-
n
tiplying the relative combined standard uncertainty by an
uc ( y ) = ∑(cs ⋅ u ( x ))
i s i
2
appropriate coverage factor (k):
i =1
U = k ⋅ uc ( y )
where us(xi) is the relative standard uncertainty and csi,
the sensitivity coefficient, which denotes uncertainty in According to the GUM [8], the coverage factor was
(y) arising from uncertainty in (x) [8, 9]. Sensitivity coef- obtained from t distribution table with a level of confi-
ficient describes how the measurand value varies with dence of 95%, according to the effective degree of freedom
changes in the value of each source of uncertainty [18]. (veff ) based on the Welch-Satterthwaite formula:
Sensitivity coefficients were experimentally obtained in
uc4 ( y )
the laboratory following the steps below [19]: νeff =
n us4 ( xi )
–– Define different values to each uncertainty source ∑ i =1 νi
(or measurement condition), in which the value

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
6      Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS

where us (xi) is each individual relative standard uncer- relative bias (δr) and its uncertainty (uδ (y)) were calcu-
tainty, and vi is the degree of freedom of us (xi). lated as follows:
x−µ
δr = ⋅ 100
µ
Top-down approach
uδ ( y ) = δ2r + ucal
2
( y ) + up2 ( y ) + uµ2 ( y )

The steps followed to estimate the measurement uncer-


where x̅ is the mean obtained in our laboratory after col-
tainty were as follows [9, 10]: (1) specification of the
lecting 425 internal quality control values; μ, the con-
measurand, (2)  estimation of the uncertainty associated
ventional value calculated as the mean of the means of
to the intermediate precision of the measurement system,
all laboratories participating in the UNITY™ program that
(3) identification of any sources of uncertainty that are
use HPLC-MS/MS measurement procedures; ucal (y), the
not adequately covered by the imprecision data, (4) esti-
relative uncertainty associated with the calibration pro-
mation of combined uncertainty and (5) estimation of
cedure; up (y), the relative uncertainty associated to the
expanded uncertainty. Steps 1, 3, 4 and 5 for the top-down
­measurement system intermediate imprecision; and uμ (y),
approach were performed in the same manner as the bot-
the r­ elative uncertainty associated with the assigned value
tom-up approach described above.
of the reference material calculated as [20]:
Estimation of the precision of the measurement
system was performed using Liquicheck™ Whole Blood s/ µ
uµ ( y ) = 1.25 ⋅
⋅ 100
Immunosuppressant Control Level 2 quality control data. p
Specifically, 425 internal quality control values were col- where s is the robust standard deviation obtained from
lected over 11  months from August 1, 2016, to June 31, all laboratories participating in the UNITY™ program, and
2017. The relative uncertainty associated to the measure- p is the number of laboratories.
ment system imprecision (up (y)) was calculated as the A correction factor for bias was applied and included
relative standard of values using pooled relative standard as source of uncertainty if
deviations, of intermonth data, given by the following
δr > 2 ⋅ uδ ( y )
equation:
n
νi ⋅ si2 ( x ) To estimate the bias related to the REC, the ME, the
up ( y ) = ∑
i =1
νi C-O and the SEL, data from a previously validated meas-
urement procedure [12] were used. Biases were estimated
where n is the number of values for the month i; si (x), the using the following equations:
relative standard deviation obtained in the month i; and δREC = 1 − REC
vi, the degree of freedom of si.
δME = 1 − ME
δC-O = C-O − 0

Estimation of uncertainty associated to the bias δSEL = SEL − 0

The uncertainty associated to the REC (uREC(y)), the


The different possible sources of bias related to the cali-
ME (uME(y)), C-O (uC-O(y)) and the SEL (uSEL(y)) were calcu-
bration procedure, the recovery of extracted samples
lated as follows:
(REC), the matrix effect (ME), the carryover (C-O) and the
selectivity (SEL) were evaluated using data of a previously uREC ( y ) = δREC
2
+ us2−REC ( y )
validated measurement procedure [12]. Studies of compat-
uME ( y ) = δ2ME + us2−ME ( y )
ibility of measurement results were performed to known if
the biases were significant. uC-O ( y ) = δC-O
2
+ us2−C-O ( y )
Because there is no high-order reference material or
uSEL ( y ) = δSEL
2
+ us2−SEL ( y )
material with a value assigned by a reference measure-
ment procedure to be used, we utilized the Liquicheck™ where us−REC(y) and us−ME(y) were the standard error of the
Whole Blood Immunosuppressant Control Level 2 and the mean values obtained in the recovery and ME of the vali-
UNITY™ Interlaboratory Program (BioRad) data to evalu- dated measurement procedure [12]; us-C-O(y) and u ­ s-SEL(y)
ate the bias associated to the calibration procedure. The were estimated using a right-angled triangle distribution

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS      7

(type B approach) from data of the measurement proce- Manufacturer’s standing measurement system.
dure [12]. LC-MS/MS measurement system was used to assign the
Correction factors for biases were applied and final mass concentration of each CsA working calibrator.
included as source of uncertainty if each absolute bias
value was higher than k-times its respective uncertainty. Manufacturer’s product calibrators. ClinCal® Whole
The k values were obtained using the Welch-Satterthwaite Blood Calibrators for Immunosuppressants were the
formula in the same way described above. working calibrators whose CsA values were assigned
using the standing measurement procedure.

End-user’s routine measurement procedure. Routine


Results human samples were processed using the end-user’s
measuring system (Acquity® UPLC®-TQD®) accord-
Metrological traceability ing to the procedure described in Rigo-Bonnin et  al.
[12], and using the product calibrators with associated
The metrological traceability that applies to cCsA in
calibration.
patients’ results was as follows:
Bearing in mind the calibration hierarchy described,
cCsA in patients’ results were metrologically traceable to
Manufacturer’s selected measurement proce-
Recipe’s ClinCal® Whole Blood Calibrators for Immunosup-
dure. Recipe’s selected measurement procedure was
pressants. The metrological traceability chain and calibra-
based on gravimetry. This procedure is neither an inter-
tion hierarchy are shown in Figure 2.
national conventional reference procedure nor an interna-
tional conventional calibrator, and there is no metrological
traceability to the SI. Measurement uncertainty
Manufacturer’s working calibrators. Six standard solu- Bottom-up approach
tions of CsA in a hemolyzed human whole blood. Each
standard solution was prepared from the material C-093 The measurand was defined as the mass concentration
(1.000 ± 0.005 g/L of CsA in acetonitrile) from Cerilliant (μg/L) of the cyclosporine A in human whole blood meas-
Corp. (Round Rock, TX, USA) and a hemolyzed human ured according to an in-house measurement procedure
whole blood solution and, then, subjected to a freeze- using an Acquity® UPLC®-TQD® measurement system. Fur-
drying process. thermore, the quantity can be described as follows:

Manufacturer’s selected Gravimetry


measurement procedure
n
sig
As
Materials prepared from the
material C-093 (Cerilliant Manufacturer’s working
Corp.) and a hemolyzed calibrators Ca
human whole blood solution lib
ra
te

Manufacturer’s standing
Metrological traceability chain

measurement procedure LC-MS/MS


n
sig
As

ClinCal Calibrator for


Relative uncertainty

Manufacturer’s product
immunosuppressants calibrators Ca
(Recipe) lib
ra
te

Clinical laboratory routine


Acquity UPLC-TQD
measurement procedure
n
sig
As

Routine patient
samples

Measurement result

Figure 2: Scheme of metrological traceability chain and calibration hierarchy for the cyclosporin A mass concentration in whole blood results.

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
8      Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS

Blood–Cyclosporine A; mass concentration (ClinCal® chamber and autosampler temperatures for the UHPLC
Whole Blood Calibrators for Immunosuppressants Ref. were 0.315% · s−1, 0.152% · V−1, 0.162% · eV−1, 0.177% · °C−1
9933; UHPLC-MS/MS). and 0.196% · °C−1, respectively. Figure 3 shows, as an
The cause and effect diagram in Figure 1 indicates example, the estimation of shaking mix time vari-
the most relevant sources of uncertainty associated to the ability of the precipitated solution of sample sensitivity
cCsA values. In this study, the sources of uncertainty were coefficient.
identified and classified into seven main contributions The combined uncertainty obtained considering only
as follows: calibration procedure, solutions preparation, the sources with the greatest contribution was of 3.5%
samples preparation, mass spectrometer parameters, (Table 1). Figure 4 shows a block diagram of the relative
UHPLC parameters, data processing and bias. standard uncertainties contributing to the combined
Relative standard uncertainty values obtained and uncertainty of cCsA.
considered to calculate the relative combined uncertainty Given that the calculated effective degree of freedom
are summarized in Table 1. was 16.7, the coverage factor used to calculate the expanded
Sensitivity coefficients of shaking mix time vari- uncertainty, from relative combined uncertainty, was
ability of the precipitated solution, cone voltage and 2.110. The relative expanded uncertainty obtained using
collision energy for the mass spectrometer, and column the bottom-up approach was 7.4% (Table 1).

Table 1: Uncertainty budget for the measurement of cyclosporin A mass concentration in whole blood using bottom-up approach.

Uncertainty   Category   Variable/   Distribution   Origin of data   Cs  us, %  (Cs · u)2, %


source factor type

Calibration   Weighted linear regression   cCsAcal   A   Experimental   1  1.40  1.96


procedure   aw, bw          
  Calibrators assigned values   cCsAcal   B   Manufacturer   1  1.11  1.23
Calibrators reconstitution   cCsAcal   B   Calibration certificate   1  0.09  0.01
  Calibrators stability   cCsAcal   A   Experimental [12]   1  1.39  1.93
Solutions   Mobile phases   fsol   B   Manufacturer   1  1.19  1.42
preparation         Calibration certificates      
  IS working solution   fsol   B   Manufacturer   1  0.83  0.68
        Calibration certificates      
  Precipitant solution   fsol   B   Manufacturer   1  0.19  0.04
        Calibration certificates      
  Organic solvent evaporation in   cIS   A   Experimental   1  0.28  0.08
IS working solution
Samples   Solutions pipetting and   fsp   B   Manufacturer   1  0.50  0.25
preparation centrifuge calibration       Calibration certificates      
  Shaking mix time variability of   fsp   A   Experimental   0.315  1.05  0.11
the precipitated solution
Mass   Cone voltage   fMS   B/A   Manufacturer/experimental  0.152  1.86  0.08
spectrometer   Collision energy   fMS   B/A   Manufacturer/experimental  0.162  2.15  0.12
parameters   Others   fMS   B   Manufacturer’s data   1  1.62  2.63
UHPLC   Column chamber temperature   fUHPLC   B/A   Manufacturer/experimental  0.177  0.74  0.02
parameters   Autosampler temperature   fUHPLC   B/A   Manufacturer/Experimental  0.196  2.72  0.28
  Others   fUHPLC   B   Manufacturer   1  1.15  1.31
Data   CsA peaks   fdp   A   Experimental   1  0.33  0.11
processing   CsA-D12 peaks   fdp   A   Experimental   1  0.31  0.10

Combined uncertainty, %   3.5

Expanded uncertainty, %, k = 2.110   7.4

Cs, sensitivity coefficient; us (%), relative standard uncertainty; cCsAcal, mass concentration of cyclosporin A calibrators values; aw, inter-
cept of the weighted regression line of the calibration curve; bw, slope of the weighted regression line of the calibration curve; CsA,
­cyclosporin A; fsol, factor for the solutions preparation; IS, internal standard (cyclosporin A-D12); cIS, mass concentration of internal standard
in working IS; fsp, factor for the sample preparation; fMS, factor for mass spectrometer parameters; fUHPLC, factor for chromatograph para-
meters; fdata, factor for data processing.

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS      9

115.0 other conditions involving changes, such as different oper-


Cyclosporine A mass concentration, %

ators, solutions, samples and calibrators preparation, lot-


110.0 y = 0.315x + 94.34
R2 = 0.966
to-lot reagents (except for calibrators and internal quality
control), different calibrations, possible variabilities on
105.0 the measurement system parameters and data processing.
The relative uncertainty associated to the measurement
100.0
system intermediate imprecision was 3.50%.
Of all possible sources of uncertainty not adequately
95.0
covered by the imprecision data, only the uncertainty
associated with the assigned values of calibrators was
90.0
0 10 20 30 40 50 60 considered (1.11%).
Time, s As indicated below, biases were negligible, and no
Figure 3: Sensitivity coefficient of shaking mix time variability of source of uncertainty related to bias was considered.
the precipitated solution. Relative uncertainties were combined to obtain a rela-
The linear regression and the determination coefficient (R2) are shown. tive combined uncertainty of 3.7%. Furthermore, given
Symbols are the mean values of replicated cCsA measurements for that the calculated effective degree of freedom was 631, the
each uncertainty source condition. For a shaking mix of 20 s, the 100%
coverage factor used was 1.972 and the relative expanded
cCsA mean value corresponds to a cCsA mean value of 165 μg/L.
uncertainty obtained using the top-down approach was
7.2% (Table 2).
Top-down approach

The sources of uncertainty considered in the top-down Uncertainty associated to the bias
approach were associated with the assigned values of
commercial calibrators, the intermediate precision and Absolute relative bias associated with the calibration
biases of the measurement system. procedure and its expanded uncertainty were 3.1% and
Internal quality control replicates were performed 13.7%, respectively. Furthermore, biases related to the
using the same measurement procedure, which includes REC, the ME, the C-O and the SEL were 0.037, 0.010, 2.20

Figure 4: A block diagram about the relative standard uncertainties contributing to the combined uncertainty of cyclosporin A mass concen-
tration in whole blood for the bottom-up approach.

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
10      Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS

Table 2: Uncertainty budget for the measurement of cyclosporin A mass concentration in whole blood using top-down approach.

Uncertainty source Category Variable Distribution type Origin of data Cs us, % (Cs · u)2, %

Calibration procedure Calibrators assigned values cCsAcal B Manufacturer 1 1.11 1.23


Intermediate precision Internal quality control data CV A Experimental 1 3.50 12.2

Combined uncertainty, % 3.7

Expanded uncertainty, %, k = 1.972 7.2

Cs, sensitivity coefficient; us (%), relative standard uncertainty; cCsAcal, mass concentration of cyclosporin A calibrators values; aw,
intercept of the weighted regression line of the calibration curve; bw, slope of the weighted regression line of the calibration curve;
CsA, ­cyclosporin A; CV, coefficient of variation.

and 3.20, respectively; and their expanded uncertainties, necessary to become familiar with the rationale, we have
0.093, 0.039, 5.50 and 8.00, respectively. tried to show it as simple as possible. For the bottom-up
All biases were lower than their respective expanded approach, we found that the main contribution to the
uncertainty values. Thus, biases were negligible, no cor- uncertainty measurement was the UHPLC autosampler
rection factor was necessary to be applied and, in con- temperature parameter, followed by all mass spectrometer
sequence, no source of uncertainty related to bias was parameters, the weighted linear regression and calibrators
considered to estimate the uncertainty associated to cCsA stability included into the calibration procedure and the
values. mobile phases preparation (Table 1 and Figure 4). For the
top-down approach, measurement uncertainty was evalu-
ated directly using intralaboratory quality control data
Discussion produced by the measurement system and data related
with the assigned values of calibrators, showing that
Values of cCsA are commonly used by the clinicians for the major contribution was from intermediate precision
monitoring the status of a transplant patient and for check- (Table 2). Our study showed that both approaches gave
ing whether the administered dose of CsA is effective. Thus, similar uncertainty results (7.4% and 7.2%), as observed
clinical laboratories must provide traceable cCsA results, by different experts [21, 22, 26, 27]. However, several con-
and as accurate as possible. To achieve it, several clinical siderations must be taken into account. To simplify, the
laboratories use measurement procedures mainly based described example of the estimation of uncertainty used
on HPLC-MS/MS or UHPLC-MS/MS. Unfortunately, few only one cCsA value within therapeutic interval. In a real
of them use metrological traceability and measurement situation, considering the heterocedasticity of the meas-
uncertainty data in order to express the comparability and urement systems based on UHPLC-MS/MS, an uncertainty
accuracy of their reported results, although the importance profile covering the measuring interval should be studied.
of these concepts is increasing [21–25]. Therefore, in this In addition, because no high-order reference material or
study, we described the measurement traceability of cCsA reference procedures exist for cCsA, a commercially availa-
results obtained by a measurement system based on an ble internal quality control was used as reference material
UHPLC-MS/MS procedure. Information from certificates to estimate the uncertainty associated to the intermedi-
and statements, facilitated by manufacturers of calibra- ate precision and calibration bias. This material could be
tors and reference materials, was sufficient to describe non-commutable with human samples, thus affecting to
the metrological traceability. Thus, patients’ cCsA results the results. Furthermore, to our knowledge, there is only
were traceable to the Recipe’s product calibrators. We one published study [28] related to the estimation of meas-
also aimed to provide clinical laboratory specialists with a urement uncertainty for cCsA results using a system based
detailed estimation of the uncertainty of cCsA values using on HPLC-MS/MS. Our expanded uncertainty values differ
the bottom-up and the top-down approaches, accord- substantially from those obtained in the study, perhaps
ing to different guidelines [8–10]. It should be noted that because of the use of a different measurement system, or
although the bottom-up is considered the best approach the way the performance characteristics were studied and
to estimate the uncertainty, in the actual situation of clini- the sources of uncertainty considered.
cal laboratories this approach is labor intensive, time con- In conclusion, we showed a detailed example to
suming and typically too complex and cumbersome to be describe metrological traceability and to estimate the
implemented. Although a certain level of statistical skill is measurement uncertainty for the cCsA results. The results

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS      11

were traceable to the manufacturer’s product calibra- 8. Joint Committee for Guides in Metrology. Evaluation of measure-
tors used to calibrate the UHPLC-MS/MS measurement ment data. Guide to the expression of uncertainty in meas-
urement (GUM). JCGM 100:2008. http://www.bipm.org/en/
system. After performing the bottom-up and top-down
publications/guides/#gum. Accessed: Jan 11, 2018.
approaches, we observed that their results were quite 9. EURACHEM/CITAC. Guide CG4: Quantifying uncertainty in
similar. This fact would confirm that the simpler top- analytical measurement, 3rd ed. Eurachem, 2012. https://
down approach should be sufficient for estimating uncer- www.eurachem.org/index.php/publications/guides/quam.
tainty of CsA mass concentration in whole blood results Accessed: Jan 11, 2018.
10. Clinical and Laboratory Standards Institute. Expression of
by HPLC-MS/MS in clinical laboratories. Finally, we hope
measurement uncertainty in laboratory medicine; approved
that this study can help and motivate clinical laboratories guideline. CLSI EP29-A. Wayne, PA: CLSI, 2012.
to describe metrological traceability and to perform meas- 11. Joint Committee for Guides in Metrology. https://www.bipm.org/
urement uncertainty studies based on the simpler top- en/committees/jc/jcgm/. Accessed: Jan 11, 2018.
down approach for this pharmacological quantity as well 12. Rigo-Bonnin R, Arbiol Roca A, González de Aledo-Castillo
as for other quantities that they may measure. JM, Alía P. Simultaneous measurement of cyclosporine
A, ­everolimus, sirolimus and tacrolimus concentrations
in human blood by UPLC–MS/MS. Chromatographia
Author contributions: All the authors have accepted 2015;78:1459–74.
responsibility for the entire content of this submitted 13. De Bièvre P, Dybkaer R, Fajgelj A, Hibbert DB. Metrological
manuscript and approved submission. traceability of measurement results in chemistry: concepts
Research funding: None declared. and implementation (IUPAC Technical Report). Pure Appl Chem
2011;83:1873–935.
Employment or leadership: None declared.
14. International Union of Pure and Applied Chemistry, International
Honorarium: None declared. Federation of Clinical Chemistry. Properties and units in the
Competing interests: The funding organization(s) played laboratory sciences. Part X. Properties and units in general clini-
no role in the study design; in the collection, analysis, and cal chemistry. Pure Appl Chem 2000;72:747–972.
interpretation of data; in the writing of the report; or in the 15. Kim JY, Kwon W, Kim HS, Suh S, In MK. Estimation of measure-
decision to submit the report for publication. ment uncertainty for the quantification of 11-Nor-delta-9-tetrahy-
drocannabinol-9-carboxylic acid and its glucuronide in urine
using liquid chromatography–tandem mass spectrometry. J Anal
Toxicol 2014;38:164–70.
16. Mansilha C, Melo A, Rebelo H, Ferreira IM, Pinho O, Domingues
References V, et al. Quantification of endocrine disruptors and pesticides
in water by gas chromatography-tandem mass spectrometry.
1. Zhang Y, Zhang R. Recent advances in analytical methods for the Method validation using weighted linear regression schemes. J
therapeutic drug monitoring of immunosuppressive drugs. Drug Chromatogr A 2010;1217:6681–91.
Test Anal 2018;10:81–94. 17. Waters Corporation. MassLynx 4.1: Peak Integration and quanti-
2. McShane AJ, Bunch DR, Wang S. Therapeutic drug monitoring of tation algorithm guide–71500122009/Revision A. Milford, MA,
immunosuppressants by liquid chromatography-mass spectrom- 2005.
etry. Clin Chim Acta 2016;454:1–5. 18. Farrance I, Frenkel R. Uncertainty of measurement: a review of
3. Freudenberger K, Hilbig U, Gauglitz G. Recent advances in rules for calculating uncertainty components through functional
therapeutic drug monitoring of immunosuppressive drugs. TrAC relationships. Clin Biochem Rev 2012;33:49–75.
2016;79:257–68. 19. International Organization for Standardization. Guidance for
4. International Organization for Standardization. Medical labo- the use of repeatability, reproducibility and trueness estimates
ratories. Requirements for quality and competence. ISO 15189. in measurement uncertainty evaluation. ISO 21748. Geneva,
Geneva, Switzerland: ISO, 2012. Switzerland: ISO, 2017.
5. Joint Committee for Guides in Metrology. International vocabulary 20. International Organization for Standardization. Statistical meth-
of metrology. Basic and general concepts and associated terms ods for use in proficiency testing by interlaboratory compari-
(VIM), 3rd ed. JCGM 200:2012. http://www.bipm.org/en/publica- sons. ISO 13528. Geneva, Switzerland: ISO, 2005.
tions/guides/vim.html. Accessed: Jan 11, 2018. 21. Oosterhuis WP, Bayat H, Armbruster D, Coskun A, Freeman KP,
6. International Organization for Standardization. In vitro diag- Kallner A, et al. The use of error and uncertainty methods
nostic medical devices. Measurement of quantities in biological in the medical laboratory. Clin Chem Lab Med 2018;56:
samples. Metrological traceability of values assigned to calibra- 209–19.
tors and control materials. ISO 17511. Geneva, Switzerland: ISO, 22. Kallner A. Estimation of uncertainty in measurements in the
2003. clinical laboratory. Clin Chem Lab Med 2013;51:2249–51.
7. International Organization for Standardization. In vitro diagnos- 23. Comite Europeen de Normalisation. Health Informatics. Expres-
tic medical devices. Measurement of quantities in samples of sion of results of measurements in health sciences. ENV 12435.
biological origin. Requirements for content and presentation of Brussels: CEN, 2006.
reference measurement procedures. ISO 15193. Geneva, Switzer- 24. Kessler A. Mass spectrometry. A key technique for traceability in
land: ISO, 2009. clinical chemistry. Trends Analyt Chem 2016;84:74–9.

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
12      Rigo-Bonnin et al.: Traceability and uncertainty of cyclosporine A results obtained by UHPLC-MS/MS

25. Vesper HW, Thienpont LM. Traceability in laboratory medicine. estimation of measurement uncertainty. Clin Chem Lab Med
Clin Chem 2009;55:1067–75. 2015;53:1025–32.
26. Rami L, Canalias F. An approach to establish the uncertainty 28. Grote-Koska D, Czajkowski S, Klauke R, Panten E, Brand K,
budget of catalytic activity concentration measurements in a ­Schumann G. A candidate reference measurement procedure
reference laboratory. Clin Chem Lab Med 2015;53:743–51. for cyclosporine A in whole blood. Accred Qual Assur
27. Lee JH, Choi JH, Youn JS, Cha YJ, Song W, Park AJ. Comparison 2014;19:147–57.
between bottom-up and top-down approaches in the

Brought to you by | Universitat Autònoma de Barcelona


Authenticated
Download Date | 4/25/18 1:06 PM
View publication stats

You might also like