The world leader in serving science
• Warning Messages
1. Indicates improper operating condition
2. May be cleared by opening the lid and correcting the problem
3. May be cleared by pressing stop
4. Refer to individual system manuals
• Error Codes
1. Indicates a malfunction
2. Message cleared by disconnecting and reconnecting power
3. Usually requires service
2
Warning Messages
• LId
• The cover was not properly closed when the start button was pressed
or the cover was opened during a run.
1. Possible lid switch/lid switch logic failure
• PFL / PFAIL
• Power to the Centrifuge was lost during a run.
1. Check power to the centrifuge.
• bAL
• Indicates an unbalanced rotor
1. Check rotor balance, bucket pivot lubrication, rotor locking
mechanism, adapters for cracks and voids, imbalance sensor.
• HEAd
• No rotor installed when START key pressed
• Recognition system fails to read a valid rotor
1. Verify rotor is installed
2. Check rotor for magnets, verify magnet pattern in manual,
check rotor recognition ring per instructions.
3
Warning Messages
• ch hd
• Change head. The rotor recognized is not in the list of supported
rotors, or, in a program mode, the current program is set up to
run a different rotor. (Rotor Rec System)
1. Press stop, open the lid and install the correct rotor
• rAD
• Radius error ( rAD flashes with max value). The rotor radius value
input exceeds the maximum. (Rotor Rec System)
1. Press stop to clear
2. In manual mode, enter correct value
3. In Program mode, save key will default (max value) for rotor
• rCF
• rCF, (Relative Centrifugal Force), (flashes with max value), the
input RCF is greater than the max for the rotor or the speed is
greater than the max allowable for the rotor. (Rotor Rec System)
1. Press stop to clear
2. Enter a new RCF value within max allowable or a new radius
4
Warning Messages
• SPD
• Speed (SPD appears and flashes with max value) if the input
speed is greater than the rotor max. (Rotor Rec System)
1. Press stop to clear
2. Enter a new speed value within the max allowed
• Cool
• Temperature of MosFet case (motor drive) exceeds 65C.
1. Check cooling / ventilation
2. Check motor / binding / replace PCB
• bruSH
• Brush appears at a timed interval ( varies by system) at the
end of each run if it is time to inspect the brushes.
Only appears on systems with brush counters
1. Power off to reset until brush counter reset
2. To reset, see individual system manual
5
Error Codes
6
Error Codes
7
Error Codes
8
Error Codes
• Er 10
• = MOTOR OVERVOLTAGE. CL2 error. Motor voltage detected over
maximum limit (94V). Rotor coasts to a stop
1. Power off and on to clear.
2. Check brushes, load, bearings, motor voltage.
• Err 14
• =INCORRECT ROTOR SIGNATURE. The rotor detected does not
correspond with the database of rotors. Rotor coasts to a stop
(Rotor Rec System)
1. Power off and on to clear
2. Check if correct rotor for system
3. Check rotor magnet pattern
4. Inspect magnets for cleanliness/strength (paper clip)
5. Use diagnostic beeper to verify (see individual manual)
9
Error Codes
• DIR
• = WRONG DIRECTION OF ROTATION. Processor detects wrong
direction of rotation. (3 phase brushless DC motor). Rotor coasts to
stop
1. Power off and on to clear
2. Check hall sensors
3. Check brake condition
4. Check logic signal (REV)
5. Replace PCB
• norEF
• =NO REFRIGERATION TEMP SENSOR. Refrigerated only.
Processor does not detect a temperature measurement from the
temp sensor.
1. Power off and on to clear.
2. Check temp display
3. Check temp sensor
4. Replace PCB
10
Error Codes
• norr
• =ROTOR RECOGNITION HARDWARE NOT DETECTED DURING
POWER ON. Blocks operation until Rotor Rec hardware is installed
(Rotor Rec System)
1. Check rotor recognition installed
2. Check rotor rec ring connection at logic pcb
3. Verify function w/ magnet
4. Replace logic PCB.
11