Automatic Analysers: Oillab Range

OilLab 580
Noack
CH-6864 Arzo,Mendrisio, Switzerland
tel. +41 91 6300703, fax +41 91 6300719
www.lin-tech.ch – info@lin-tech.ch
Linetronic Technologies SA
Via Onorio Longhi 2

ASTM D5800 Crucible, Crucible Cover • A udible alarm and displayed messages
IP 421 and Heating Block at the end of the analysis and in case
• S ame dimensions and volume of errors and/or malfunctions
Subject
as ASTM D5800 Diagnostic Menu
Determination of the evaporation loss
• E lectrically heated new designed aluminium • S electable value displaying: °C / Volt
of lubricating oils (particularly engine oils).
block, no Woods metal needing C alibration Menu
Procedure A uses the Noack evaporative
• A utomatic calibration of each temperature
tester equipment. Heating unit
probe
Procedure B uses the automated • E lectrical
resistance, 500 W
• L ast calibration date referred to each single
non-Woods metal Noack evaporative
Vacuum Pump probe displayed and relative data printable
apparatus.
• E quipped with high resistant Kalrez valve, • D isplay of calibration diagram
Measuring Noack Principle inlet filter to remove product residuals • I nsertion of offset values
A quantity of 65 grams of sample is heated • A utomatic electronic control system able D ata Utilities
to a specific temperature and maintained to maintain the -2mB during the analysis • F ields for introduction
for 1 hour while it is enclosed in a crucible, • L ow voltage power supply of operator and product name
the crucible’s cover is shaped to allow • A rchive viewer for files recall
a constant vacuum of -2 mbar to remove Integrated Touch Screen Panel PC • A ll analysis stored in Excel ®
from the crucible the evaporating portion • TFT/LCD 8” compatible format
of the sample. At the end of the test, • R esolution 1024 × 768, 16.2 M colours • S torage capacity for more
the sample is cooled and then reweighted: • 2 USB ports for connection to an external than 60’000 analysis
the difference, reported in percentage, printer and/or external PC • LIMS compatible
represent the sample’s Evaporation Loss • S torage capacity for more than 60’000
by the Noack Method. analysis Dimensions
Method A: bath is controlled at 250°C; • L in-Tech operating software Lab-Link • L ength 400 mm
Method B: the sample is controlled running in Windows ambient • W idth 450 mm
at 245.2°C. • A utomatic reading of the weight suggested • M ax. height: 450 mm
balance LT/B-2470/BCA500 INT- CAL
Measuring Temperature Probe Weight
• P latinum resistance PT100 class A Software Features • 22
Kg
• A ll analytical parameters recorded
Measuring Parameters
• C ustomizable analysis parameters Electrical Supply
• Temperatures: in °C
and methods • 2 20V ± 15% / 50 to 60 Hz
• Testing range: +225°C to +275°C
• C ustomizable results report • 1 15V ± 15% / 60 Hz
• M easuring range: 0°C … +320°C
• P rintable graphs and results
• R esolution: 0.01 °C
Cord cable
• A ccuracy: ± 0.1 °C The software includes:
• 3
conductors flexible cable 2 m (7 feet)
• R epeatability / Reproducibility:
Analysis Menu length with PVC sheath oil and heat resistant
according ASTM D5800 or better
• S tandard method as per ASTM / IP / ISO / as per CENELEC directives
Electronic regulator for automatic control EN / DIN… norms of reference
of differential pressure • Unknow sample Ambient Temperature
• D ifferential pressure 20 mm H20, • D irect access to all analog, digital, inputs • Max 35°C
accuracy 1% and outputs • H.R. 80%
Automatic Analysers: Oillab Range

OilLab 580
Noack

Particular attention has been payd

CH-6864 Arzo,Mendrisio, Switzerland


tel. +41 91 6300703, fax +41 91 6300719
www.lin-tech.ch – info@lin-tech.ch
Linetronic Technologies SA
Via Onorio Longhi 2
to the integrated vacuum pump
that is also protected by an inlet
filter for residual recovery.

Crucible holder made in Noack Tester ASTM D5800 Stand Alone includes:
stainless steel keeps the cup • integrated vacuum pump with inlet filter
assembly ready to use. • e vaporation crucible for procedure B
• 1 0 test balls
• nozzle cleaner
• crucible holder
• protection gloves
• hook wrench
• pliers

Accessories
• L AB-580-1001: glassware acc. CEC L40-A-93, 1 complete set
comprising 2 glass bottles 2 litres capacity, with the necessary rubber bungs,
glass delivery tubes (internal Ø 4 mm) and silicone tubings
• L AB-580-1002: stand for glass bottles, including inclined manometer
0 to 50 mm water and Fresenius column
• L AB-580-1003: evaporation crucible
• L AB-580-1009: Noack reference oil, 1 Ltr
Protection gloves ans hook
• L AB-580-0010: Noack software evaluation tools
wrench, high quality materials
for safety operations. Spare parts
• L AB-580/008-12: PT100 sample
• L AB-580/11-01: silicon tubing 2 m
• L AB-580/013-02: air filter
• L AB-580/007-01: main electronic board Noack
• L AB-580-1003: evaporation crucible
• L AB-580-0011: hook wrench
• LAB-580-0012: pliers
• L AB-580-0013: crucible holder
• L AB-580-0014: test balls (pack of 10 pcs.)
• L AB-580-0015: nozzle cleaner
• LAB-580-0016: gloves
• L AB-580/006-03: main electronic board
• LAB-580/05-23: heater
• L AB-580/08-14: PT100 bath
With the pliers the nut of the
Calibration Tools
crucible cover is accurate and
• O illab 80: calibration decade box – PT100 simulator
easily fixed.
• O illab 81: set of connectors and cables for cold range