You are on page 1of 2

Automatic Analysers: Oillab Range

OilLab 600
Pensky Martens
CH-6864 Arzo,Mendrisio, Switzerland
Linetronic Technologies SA

tel. +41 91 6300703, fax +41 91 6300719


www.lin-tech.ch – info@lin-tech.ch
Via Onorio Longhi 2

ASTM D93 procedures A, B, C Measuring Pensky Martens Principle Test Cup


ASTM D3941 - ASTM E502 The sample is heated and stirred at • M ade of brass and provided with
DIN EN 22719 specified rates, using one of three defined an high temperature resistant handle
EN 22719 procedures (A, B, or C). An ignition source • W ith sample level mark
IP 34 is directed into the test cup at regular
Stirrer
ISO 2719 intervals with simultaneous interruption
• A nelectric motor drives a flexible
of the stirring, until a flash is detected.
Subject transmission coil allowing the stirring of
Flash Point on petroleum products, gas oils, Measuring Parameters the product. Stirring speed as per selected
fuel oils, lubrificants, biodiesel. Suitable • Temperatures: in °C procedures A, B, C
for flash point detection on different • M easuring range: +35°C … +370°C
Shutter
substances, waste materials, solvents… • R esolution: 0.01 °C
• A utomatic mechanism opening the shutter
• A ccuracy: ± 0.1 °C
for the dip-in of the ignition device
• R epeatability / Reproducibility:
as per standards methods or better Safety features
• G as valve for closing the gas supply
Ignition system
(max 30 mBar), at the end of the test
G as with flame exposure device and / or
• O verheating protection with auto shut off
electric lighter with electrical ignitor
during the test
Measuring Temperature Devices • A uto fire detection by means of thermal fuse
• S ample temperature is measured with with acoustic alarm
a platinum resistance PT100 Class A
with SS sheath and high temperature Software Features
resistant silicone cable • All analytical parameters recorded
• B ath temperature is measured with • C ustomizable analysis parameters
a PT100 sensor and methods
• C ustomizable results report
Dual flash point detection system
• P rintable graphs and results
• B y ionisation ring
• B y thermal sensor The software includes:

Ionisation ring Analysis Menu


• Barometric sensor • A utomatic execution of the analysis
• B uilt-in sensor with automatic correction of in accordance to the selected procedure
results to a barometric pressure of 101.3 (Standard method as per ASTM / IP / ISO
kPa automatically performed by the software / EN / DIN… norms of reference as well
at the end of analysis as costumized procedures)
• A utomatic handling of samples with
Heating
unknown flash point
• E lectrical
heater with heating rates as per
• D isplay in real time of all the analysis
procedures A, B, C
parameters and status
Cooling • F ields for introduction of operator
• B uilt-in
forced air fan and product name
Automatic Analysers: Oillab Range

OilLab 600
Pensky Martens
CH-6864 Arzo,Mendrisio, Switzerland
Linetronic Technologies SA

tel. +41 91 6300703, fax +41 91 6300719


www.lin-tech.ch – info@lin-tech.ch
Via Onorio Longhi 2

Electric lighter
with electrical ignitor.

• E xpected flash point temperature Electrical Supply


programmable • 2 20V ± 15% / 50 to 60 Hz
• A udible alarm and displayed messages • 1 15V ± 15% / 60 Hz
at the end of the analysis and in case
Cord Cable:
of errors and/or malfunctions
• 3conductors flexible cable
• C onfiguration menu with up to 20 preset
with schuko plug
samples and expected flash point
• A utomatic barometric correction of results Ambient Temperature
Diagnostic Menu • Max 35°C
Gas with flame exposure device.
• D irect access to all analog, digital, • H.R. 80%
inputs and outputs
Dimensions
• S electable value displaying:
• w idth 48 cm
°C / Volt
• d epth 30 cm
Calibration Menu
• h eight 52 cm
• A utomatic calibration of each temperature
probe Weight
• U p to 100 calibration points (standard • 27 Kg
with 5 and dynamic with up to 100 points)
• P rogrammable calibration frequency with Accessories
selectable validity period and notice/lock- • L AB-600/05-23: heater collar
down at expiration date • L AB-600/06-21: gas valve
• L ast calibration date referred to each single • L AB-600/07-01: electrical ignitor
probe displayed and relevant data printable • L AB-600/07-03: micro switch
• D isplay of calibration diagram • LAB-600/07-04: handle
Data Utilities • L AB-600/07-05: gas ignitor
• A rchive viewer for files recall • L AB-600/08-12: PT100 product
Dual flash point detection
• A ll analysis stored in Excel ® compatible • L AB-600/08-13: detection / ionisation cable
system: by ionisation ring,
format • L AB-600/08-14: PT100 Bath
by thermal sensor.
• LIMS compatible • L AB-600/09-04: gas reducer
• L AB-600/09-05: calibrated brass crucible
Integrated Touch Screen Panel PC
• L AB-600/09-06: calibrated brass crucible complete with movement
• TFT/LCD 8”
• L AB-600/09-07: cover cup movement only
• R esolution 1024 × 768, 16.2 M colours
• L AB-600/10-04: PCB fuses, box of 10
• 2 USB ports for connection to an external
• L AB-600/10-05: main electronic board
printer and/or external PC
• L AB-600/11-01: silicon tubing, 1 meter
• S torage capacity for more than 60’000
• L AB-600/11-02: stirrer / flexible
analysis
• L AB-600/12-01: voltage transformer for ignitor
• L AB-600/20-01: support PT100 Teflon

Calibration Tools
• O ilLab 80: calibration decade box – PT100 simulator
• O ilLab 81: set of connectors and cables for cold range

You might also like