Automatic Analysers: NewLab Range

NewLab 300
Pour Point
CH-6864 Arzo,Mendrisio, Switzerland
Linetronic Technologies SA

tel. +41 91 6300703, fax +41 91 6300719


www.lin-tech.ch – info@lin-tech.ch
Via Onorio Longhi 2

ASTM D97 (correlated) Measuring Pour Point Principle The software includes:
ASTM D5853 According to the methods, the sample
Analysis Menu
ASTM D5950 is cooled down at a specified rate and,
• S tandard method as per ASTM / IP / ISO /
ASTM D6922 at the prescribed temperature intervals,
EN / DIN… norms of reference:
IP 15 the mechanical arm of the analyser lifts
· (internal) with sample pre-heating
IP 441 the test jar from the cooling jacket and tilts
· (external) without sample pre-heating
ISO 3016 it in order to bring it in horizontal position
• Optional methods:
to test the flow of the product.
· T-sample – T-bath (Delta T constant)
Subject The sample movement is detected by
· cooling rate °C / h
Pour Point of petroleum products, the thermal probes (PT100 detection) placed
· selectable bath steps
crude oils, motor and engine oils, additives, above the sample surface which react if
· fast bath
lubricating oils, … touched by the cooled sample.
· selectable tilt out test temperature
Determination (D6922) if an automotive
Measuring Pour Point Devices • A udible alarm and displayed messages
engine oil is homogeneous and will remain
• Two PT100 detection probes placed (at the end of the analysis and in case
so, and if it is miscible with certain standard
on the surface of the product of errors and/or malfunctions)
reference oils after being submitted to a
• M echanical moving arm bringing the test jar Diagnostic Menu
prescribed cycle of temperature changes.
in horizontal position • D irect access to all analog, digital, inputs
and outputs
Measuring Temperature Probe
• S electable value displaying: °C / Volt
• P latinum
resistance PT100 class A
Calibration Menu
Measuring Parameters • Automatic calibration
• Temperatures: in °C of each temperature probe
• Measuring range: -110°C … +100°C • L ast calibration date referred
• Range of analysis: -90°C … +60°C (300/2-SA) to each single probe displayed
• R esolution: 0.06 °C and relative data printable
• A ccuracy: ± 0.1 °C • D isplay of calibration diagram
• R epeatability / Reproducibility: • I nsertion of offset values
as per standards methods or better • S tandard and advanced calibration modes

Software Features Data Utilities


• N ew LabLink software able to manage • F ields for introduction of operator
up to 6 analytical heads simultaneously and product name
(stand alone) • A rchive viewer for files recall
• U ser friendly interface • A ll analysis stored in Excel ®
• A ll analytical parameters recorded compatible format
• C ustomizable analysis parameters • S torage capacity for more than
and methods 60’000 analysis
• C ustomizable results report • LIMS compatible
• P rintable graphs and results
• S elf-identification of the typology
of the analysers connected

Automatic Analysers: NewLab Range

NewLab 300
Pour Point

CH-6864 Arzo,Mendrisio, Switzerland

www.lin-tech.ch – info@lin-tech.ch
Linetronic Technologies SA

tel. +41 91 6300703, fax +41 91 6300719


Via Onorio Longhi 2
NewLab 300 ST

Integrated Touch Screen Panel PC Cord Cable NewLab 300 ST


• TFT/LCD 12” • 3
conductors flexible cable 2 m (7 feet) • M easuring range: -110°C … +100°C
• R esolution 1024 × 768, 16.2 M colours length with PVC sheath oil and heat resistant • R ange of analysis: -110°C … +55°C
• 2 USB ports for connection to an external as per CENELEC directives • R esolution: 0.01 °C
printer and/or external PC • W idth: 34 cm
• S torage capacity for more than 60’000 Ambient Temperature • D epth: 60 cm
analysis • M ax 32 °C • H eight: 80 cm
• H.R. 80% • Weight: 34 kg
Test Jar
• S ame dimensions and volume as described Dimensions and weight
by the standard test methods • 1 test pos.: w 66 × d 60 × h 80 cm, 60 kg
• P roduct level mark • 2 test pos.: w 66 × d 60 × h 80 cm, 90 kg / 100 kg
• S mall edge on the top in order to fix the • 3 test pos.: w 100 × d 60 × h 80 cm, 130 kg
glass cell to the analytical head • 4 test pos.: w 134 × d 60 × h 80 cm, 160 kg
• 6 test pos.: w 130 × d 75 × h 170 cm, 280 kg
Cooling System
• I ntegrated gas CFC free motor compressors: Spare Parts
· Single stage • L AB-xxx/005-03: heather + auto adhesive + insulation
(for temperatures up to -40°C / 1) • L AB-xxx/005-04: thermo switch
· Double stage • L AB-xxx/005-06: PT100 bath
(for temperatures up to -80°C / 2) • L AB-xxx/007-02: static relay
• E quipped with an automatic energy power • L AB-xxx/007-04: PCB fuse 1.6 A, box of 10 pcs.
save system. After 15 minutes from • L AB-xxx/006-01: cooling fluid valve + fitting
the end of the analysis the cooling system (only for motor-compressor units)
goes in stand-by mode. • L AB-300/007-01: main electronic board Pour Point
• L AB-300/002-16: precision potentiometer
Safety Devices • L AB-300/008-12: PT100 product w/connector
• P ressure controller for 1st stage motor • L AB-300/008-13: PT100 detection
compressor • L AB-300/008-04: calibrated test jar
• P ressure controller for 2nd stage motor • L AB-300/008-041: o-ring for test jar
compressor
• T hermostat for 2nd stage activation Calibration Tools
• T hermo-switch for each cooling / heating • O ilLab 80: calibration decade box – PT100 simulator
jacket • O ilLab 81: set of connectors and cables for cold range
• M otor compressors equipped with internal
overload devices

Electrical Supply
• 2 20V ± 15% / 50 to 60 Hz
• 1 15V ± 15% / 60 Hz