SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Technical News
Subject: New software of BC-6600/6800 is launched

RECORD IDENTIFIER

Ref No.: TXQ-13012-BC-6800&BC-6600 Relate To:


Version: 01.08.00.14079 Software/Manual Upgrade
Author: Mars Zhang Parts Change
Date: May 10th,2013 Appearance
Model: BC-6800&BC-6600 Stop Production
Comments: N/A Others

CONTENT

To Whom It May Concern,


Below content is only for your information.

New software is launched. Please refer to figure 1 for version information.

SPU DMU

Figure-1

Some improvements are made in this new software which are listed as follows:

1: The range of -40Kpa is changed to (-50Kpa~-33Kpa). It is suggested to adjust it to be


38+/-0.5Kpa

2: When there are too many sample results and many samples use the same sample ID
DMU may work slowly. New software solved this problem.

3: Improved DMU performance in Greek operation system


Ref No.: TXQ-13012-BC-6800&BC-6600 MRSZ/R04N03.04.02(4.0) Page 1 of 2
Technical News- For reference only
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

4: Sheath filter, reagent detection tube, fluorescence cap assembly and diluent preheat
bath are changed to new version which are more reliable and durable. New assembly is
compatible to old one. Part numbers of these assemblies keep the same as before.

If you need download information or have any questions, please feel free to contact
Mindray regional service engineer or Mindray Service Department (email:
service@mindray.com).
Thank you for your kind attention and cooperation.

Sincerely yours,
Customer Service Departments Mindray Co., Ltd.

Ref No.: TXQ-13012-BC-6800&BC-6600 MRSZ/R04N03.04.02(4.0) Page 2 of 2


Technical News- For reference only