You are on page 1of 3

ENGINEERING CHANGE RECORD

No: 312G004
TITLE: Program Update to Version 00-07
Included ECR No:
PRIORITY: Important
DATE RELEASED: 2012.08.13
MODEL: XP Series

Revision Information:
N/A

Applicable To:
XP-100 S/N: A1002 and thereafter

XP-300 S/N: A1030 and thereafter

Summary:
The XP Series program has been updated to version 00-07.

Detailed Information:
The XP Series program has been updated to version 00-07. Changes are
as follows:

<Changed Specifications>

1. "@" key has been added to the alphabet and numeric keypads to allow "@"
to be used for POP user name, etc.
For upper case, "QZ" key has been changed to "QZ@".
For lower case, "qz" key has been changed to "qz@".

[Image of new alphabet and numeric keypads]

312G004 1/3
<Resolved issues>

1. Previously, on instruments for China, some characters were garbled in


"Settings Error" screen. This issue has now been resolved.
2. Previously, when there was no paper in a graphic printer (GP) and
multiple sample data were sent to the printer from stored data, "GP paper
empty" error occurred. There were also cases where only data for the first
sample was printed and data for the remaining samples were not printed
even though paper was subsequently refilled in the printer and [Retry]
button was pressed in the help screen.
Also, when the XP instrument was not connected to a graphic printer
(GP) and multiple sample data were sent to the printer from stored data,
"GP printout error" occurred. There were also cases where only data for
the first sample was printed and data for the remaining samples were not
printed even though the printer was subsequently connected to the
instrument and [Retry] button was pressed in the help screen.

These issues have now been resolved.

Please refer to the program version histories in "Attachments" for detailed


information.

<XP Series Program Upgrade Procedure>


Refer to the "XP Series Program Upgrade Procedure" in "Attachments". It
has been confirmed that upgrade can be done from either version 00-04,
00-05 or 00-06 to version 00-07.

Action To Be Taken :
This program will not be provided free of charge for instruments in the field. If you
need to update the program version, an order should be placed using the part
number listed below.

Parts Quantity Changes:


Old New Lead
No. Description Status Category
Qty Qty Time
BD347455 CARD_ASSY 1 1 Changed 90 days E
NO.25(1XP1X)
NOTE: Lead time is only an approximate delivery time for your reference.

312G004 2/3
Reference Document :
None

Attachments :
CARD1XP1X XP Series Program Version History
XP Series Program Upgrade Procedure

Person in
Chieko Takeda
Charge:
Reviewed by: Junko Horitaka, Hikaru Nakai, Mark Dorfner
Authorized by: Atsumasa Sone

312G004 3/3