0   PROCEDURE FOR HANDLING OF DYNAMIC PASS BOX:


 Precautions during handling of dynamic pass box:

o Ensure that the UV light shall not glow while opening the door.

o Ensure that the pass box doors closed properly.

o Do not open the doors on both sides at the same time to avoid contamination.
 Cleaning of Dynamic Pass Box:

o Open the pass box door by pushing the Press Release

o Clean the inner sides as well as the outer side of Pass Box with clean & lint-free cloth.

o Sanitize its inner surface using 70 % IPA.

o Sanitize the outer surface of Pass Box using 70 % IPA and close the door.

o Fill the cleaning status as per Annexure-1 and a respective logbooks.

o Cleaning frequency: Daily morning and whenever required.
 Operation of Dynamic Pass Box:

o Switch on the main Power Supply.

o After Power Supply the motor blower shall start automatically.

o After 30 Minutes record the Pressure differential of Magnehelic gauge in mm of the water column in the pressure
differential record of dynamic pass box as per Annexure-2.

o Switch “ON” the UV light and let it in operation for 30 minutes.

o After 30 minutes switch OFF the UV light and enter the burning hour details of UV as per Annexure-2

o Check the burning hours of UV light by its auto recorder.

o Total burning hours should not be greater than 1000 Hours.

o If the working hours are more than 1000 hrs. replace the UV light by the new one.

o Open of the pass box door by pushing the Press Release Switch.

o Place the required materials.

o Motor blower is continuously switched on until the material is there in the pass box.

o After replacing the material from the pass box close the door of the pass box.

o If any drop in pressure differential of the Magnehelic gauge of dynamic pass box is observed during operation,

o Then inform maintenance personnel for pre-filter cleaning and enter the details in Annexure-2.
 Monitoring Performance of Dynamic Pass Box:

o Prepare the Plates of Soyabean casein digest Agar with glycerol (if required) as per respective SOP.

o Pre incubate the plates before exposure at 30°C to 35°C for 24 Hours.

o Ensure that the dynamic pass box is ON for a minimum of 30 minutes before exposing the plates.

o Expose these plates for 4 hours on two locations for Passive air sampling & one location for Active air sampling
(1000 L Air) of the dynamic pass box.

o After exposure, plates are incubated at 30-35°C for 48-72 hours to get the no. of bacterial colonies & further the
same plate incubate at 20-25°C for 48-72 hours to get the fungal count.

o Enter the result of total viable counts in Annexure-3.

o Keep one negative control plate without exposure.

o Limit: Active Air sampling: NMT 200 CFU/m3        Passive Air sampling: NMT 100 CFU/Plate

o Frequency: Weekly
 Qualification of Dynamic Pass box
o
o Frequency: Once in six months ±15 days.
o
o The qualification of DPB required:

o If any failure observed during routine monitoring.

o In case of breakdown, maintenance, preventive maintenance.

o In case of a change in the installation location.

o As per schedule
 Operational tests of Dynamic Passbox Qualification:

o The following tests are performed during qualification and results are within acceptance criteria.

o
 PAO/DOP testing (integrity testing) of the HEPA filter.

o
 Particle count test

o
 Air velocity test.

o PAO/DOP Test ( filter integrity testing):

o Feed the PAO (Poly Alfa Olefin)/DOP Aerosol at the suction of blower.

o Measure the upstream concentration of PAO/DOP as a 100% concentration.

o Do the scanning after HEPA filter.

o Acceptance criteria: The penetration level should be less than 0.01%.

o Particle count test:

o Ensure that particle count of the work station is done by using a calibrated airborne particulate matter counter.

o Ensure that the sampling station is cleaned.

o Run the blower at least 1 hour before taking the particle count.

o Take the samples at 5 different locations.

o Acceptance criteria: Particle count of size 0.5 micron and large should not be more than 100 particles per cubic
feet & 5 micron and large should not be more than 0 particles per cubic feet

o Air Velocity Test:

o Start the pass box.

o Take the air velocity by calibrated anemometer 6” away from the grill at five different locations in ft./min.

o Acceptance criteria: Average air velocity should 90±20 ft./min.

o Qualification shall be done by the external agency if the facility not available at the location.

o After Completion of Qualification writes the details and pest as per Annexure-3.
7.0   ANNEXURES:
Annexure-1: Cleaning Status Label
CLEANING STATUS LABEL

Name of Instrument

Instrument ID

Cleaning Done on

Cleaning Due on

Cleaned By (Sign & date)

Checked By (Sign & date)


 

Annexure-2: Operation Record of Dynamic Pass box 


Date Cleaning UV Light Pressure Operated Checked Remark
(8 to 18 mm by by
of WC)

Start Time End Start Time End Time Total


Time Burning
Hours