You are on page 1of 26

The Babcock & Wilcox Company

Chapter 40
Pressure, Temperature, Quality
and Flow Measurement

Instruments for measuring pressure, temperature,


fluid flow, and the quality and purity of steam are
essential in the operation of a power plant. The mea-
surements obtained permit safe, economical and reli-
able operation of the steam generating equipment.
Instrumentation may range from a simple indicating
device to a complex automatic measuring device. The
fundamental purpose of the instrument is to convert
some physical property of the material being measured
into useful information. The methods used to obtain
the measurement depend on available technology,
economics, and the purpose for which the information
is being obtained.
Test instrumentation, often portable, may be used
to determine flow, pressures, and temperatures re-
quired to satisfy the user and the boiler equipment Fig. 1 Bourdon gauge.
supplier that the design operating conditions have
been achieved. Requirements for these instruments
are summarized in the American Society of Mechani- Manometers are considered an accurate means of
cal Engineers (ASME) Performance Test Codes. Many pressure or pressure differential measurement. These
test instruments require skilled technical operators, instruments, containing a wide variety of fluids de-
careful handling, and frequent calibration. Some in- pending on the pressure, are capable of high accuracy
struments and measurement methods may not be with careful use. The fluids used vary from those
suitable for long-term continuous operation because lighter than water for low pressures to mercury for
of a harsh environment, material limitations, or ex- relatively high pressures. Fig. 2 illustrates an inclined
cessive cost. Instruments used for continuous mea- manometer used to measure low pressure differentials
surement may require compromises in accuracy for
long-term, dependable, reliable operation.

Pressure measurement
The pressure gauge is probably the earliest instru-
ment used in boiler operation. Today, with complex
control systems in operation and more than one hun-
dred thirty years after the first water tube safety
boiler, a pressure gauge is still used to determine steam
drum pressure. The Bourdon tube pressure gauge is
shown in Fig. 1. Although improvements have been
made in construction and accuracy, its basic principle
of operation remains unchanged. A closed end oval
tube in a semicircular shape straightens with inter-
nal pressure. The movement of the closed end is con-
verted to an indication (needle position).
Pressure measuring instruments take various forms,
depending on the magnitude of the pressure, the ac- Fig. 2 Inclined differential manometer (courtesy of Dwyer Instru-
curacy desired and the application. ments, Inc.).

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-1


The Babcock & Wilcox Company

at low static pressure. Differential diaphragm gauges


with a magnetic linkage are also used for low pres-
sure measurement. Fig. 3 shows a high pressure mer-
cury manometer. For greater precision in measuring
small pressure differentials, such as the measurement
of flow orifice differentials, hook gauges or
micromanometers may be used.
Bourdon tube pressure gauges are available for
measuring a wide range of static pressures in vary-
ing degrees of precision and accuracy. Pressure gauges
used as operating guides need not be of high preci-
sion and normally have scale subdivisions of about 1%
of the full scale range. For certain tests, such as hy-
drostatic testing of pressure parts and boiler efficiency,
more precision is required. Gauges with 0.1% full scale
subdivisions are used. For performance calculations,
high precision temperature and pressure measure-
ments are required to accurately determine steam and
water enthalpies. Dead weight gauges or calibrated
pressure transmitters are preferable to Bourdon
gauges for high precision pressure measurement.
Diaphragm type gauges are used for measuring
differential pressures. A slack diaphragm pressure
gauge can be used to measure small differentials
where total pressure does not exceed about 1.0 psig
(6.9 kPa). For high static pressures, opposed bellows
gauges (Fig. 4) read a wide range of differential pres-
sures. They are suitable for measuring differential
pressure across boiler circuits and can be used to mea-
sure differentials from 2 to 100 psi (0.01 to 0.69 MPa)
at static pressures up to 6000 psi (41.37 MPa).
Incorporated into many pressure measurement
devices is the capability of producing an output sig-
nal proportional to the measurement. The output sig-
nal may be transmitted to a central measurement or Fig. 3 High pressure mercury manometer (courtesy of Meriam
control system. Pneumatic transmitted signals are Process Technologies).

Fig. 4 Opposed bellows gauge (courtesy of ITT Barton, a unit of ITT Industries, Inc.).

40-2 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

used in control systems, but the more common designs der certain circumstances a through the wall pressure
use electrical circuitry to sense the changing position connection may not be representative. It may be nec-
of basic mechanical motion and to produce an electri- essary to use a probe incorporating a tip configura-
cal output signal to be transmitted. These are readily tion that minimizes the effects of the flowing fluid to
adaptable to computer based systems. Some transduc- obtain a representative static pressure measurement.
ers use a piezoelectric crystal which changes electri- Connecting lines to instruments should be as short
cal resistance as the element is deformed under pres- and direct as possible and must be leak-free. For differ-
sure. Other pressure transmitter designs use a dia- ential pressure readings, it is preferable to use a differ-
phragm with strain gauges attached; minor changes ential pressure measuring device rather than to calcu-
in diaphragm strain are transferred to the strain late the difference between the readings on two instru-
gauges, which change resistance. Electrical circuits ments. Particular attention should be paid to the place-
recognize the resistance change in the crystal or strain ment of the instrument, the potential for condensation
gauges and produce an indication of pressure. Ad- in the impulse lines, and the accumulation of debris or
vanced applications have used an optic glass fiber non-condensable gases in the completed installation.
embedded in a metal diaphragm, with pressure mea-
surement determined as a change in the light beam
traversing the fiber. Pressure transducers incorporat- Temperature measurement
ing electrical components have the advantage of fewer Today, the Fahrenheit and Celsius (formerly centi-
mechanical parts for improved durability. However, grade) scales are the most common and firmly estab-
this may restrict the installation location due to tem- lished temperature scales.
perature limitations of the electrical components. In- The Fahrenheit scale is fixed at the freezing point
struments configured to generate electric signals have (32F) and the boiling point (212F) of pure water at
been adapted to incorporate computing capability to atmospheric pressure, with 180 equal degrees between
assist in calibration. Pressure transmitters and trans- the two points. The centigrade scale was originally based
ducers have been readily adapted for power plant on the same freezing (0C) and boiling (100C) points of
computer data acquisition and control systems. water at standard atmospheric pressure, using the con-
venient 100 degree interval of the decimal system.
In 1960, the General Conference of Weights and
Instrument corrections Measures changed the defining fixed point from the
In recording and reporting pressure readings, cor- freezing point to the triple point (0.0C) of water. The
rections may be necessary for water leg and for con- triple point is the condition under which the three
version to absolute pressure by adding atmospheric phases of matter (solid, liquid and vapor) coexist in
pressure. Effective water leg is the added pressure equilibrium. This point is more easily and accurately
imposed on the gauge by the leg of water standing reproduced than the freezing point. At standard at-
above the gauge. Fig. 5 illustrates a water leg correc- mospheric pressure the interval between the triple
tion to a pressure gauge reading. The density of the point for water and steam on the centigrade scale
water in the leg may change with ambient tempera- became 99.99C instead of 100C.
ture and, for precise measurement, the water leg tem- In 1990, temperature standards were further re-
perature should be monitored. On some gauges it may fined by the International Temperature Scale 1990
be possible to zero the gauge at zero operating pres-
sure with the water leg completely filled to compen-
sate for the water leg static head. Care must be taken
in performing water leg adjustments as this may af-
fect the instrument calibration over the operating
range. Differential pressure measurements require a
water leg correction for both sides of the measuring
instrument. For some liquid level devices, the differ-
ential pressure measurements from two water legs of
different elevations are used to provide the indication
of liquid level. Differential pressure measurements
across orifices, nozzles or pitot tubes to measure flow
are described below.

Location of pressure measurement


connections
The criteria for selecting a connection location to the
pressure source for the measuring device are the same
regardless of the magnitude of the pressure, the type
of measuring device or the fluid being measured. Pres-
sure connections, or taps, in piping, flues or ducts
should be located to avoid impacts or eddies; this as- Fig. 5 Application of water leg correction to pressure gauge reading
sures an accurate static pressure measurement. Un- (Correction ∆P = gρL/gc).

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-3


The Babcock & Wilcox Company

(ITS-90). These changes, however small, reflect the Expansion properties


continued review of temperature measurement. As Most substances expand when heated and, for
measured by the current standard, water at standard many materials, the amount of expansion is almost
conditions boils at 211.953F (99.974C). directly proportional to the change in temperature.
In thermometer calibrations, the fusion and vapor- This effect is the basis of various thermometer types
ization of various pure substances have been carefully using gases, liquids or solids.
determined and are known as fixed points. Two addi- Gases The expansion of gases follows the relation-
tional scales in scientific work use the absolute zero of ship, in English units:
temperature, i.e., –273.16C or –459.69R. The absolute
scale using Celsius degrees as the temperature interval PvM = RT (1)
is called the Kelvin (K) scale; the absolute scale using
Fahrenheit degrees is called the Rankine (R) scale. where
P = absolute pressure, lb/ft2
Methods of measuring temperature vM = volume, ft3/mole of gas
R = universal gas constant, 1545 ft lb/mole R
Heat affected properties of substances, such as ther-
T = absolute temperature, R = F + 460
mal expansion, radiation and electrical effects, are
incorporated into commercial temperature measuring At high pressures or near the condensation point, gases
instruments. These instruments vary in their preci- deviate considerably from this relationship. Under
sion depending on the property measured, the sub- other conditions, however, the deviation is small.
stance used and the design of the instrument. The care Two types of gas thermometers are based on this re-
taken in selecting the correct instrument largely de- lationship. In one, a constant gas volume is maintained
termines the accuracy of the results. It is the and pressure changes are used to measure changes in
engineer’s responsibility to make certain that the ap- temperature. Very accurate instruments of this type
plication is correct and that the results are not affected have been developed for laboratory work. The constant
by extraneous factors. volume type thermometer is widely used commercially.
Nitrogen is commonly used for the gas-filled ther-
Changes of state mometer in industrial applications. It is suitable for a
Fusion For a pure chemical element or compound temperature range of –200 to 1000F (–129 to 538C).
such as mercury or water, fusion, or change of state The construction is similar to the vapor pressure ther-
from solid to liquid, occurs at a fixed temperature. The mometer with nitrogen gas replacing the liquid and
melting points of these materials are suitable fixed vapor. Expansion of the heated nitrogen in the bulb
reference points for temperature scales. increases the pressure in the system and actuates a
The fusion of pyrometric cones is widely used in the temperature indicator.
ceramic industry as a method of measuring high tem- Liquids Liquid expansion properties may be incor-
peratures in refractory heating furnaces. These cones, porated into a thermometer by using a bulb and cap-
small pyramids about 2 in. (50.8 mm) high, are made illary to confine the liquid. In this thermometer, the
of oxide and glass mixtures that soften and melt at bulb and capillary tube are completely filled with liq-
established temperatures. The pyrometric cone is suit- uid and a calibrated pressure gauge is used to mea-
able for temperatures ranging from 1100 to 3600F sure temperature. Mercury has been used for a tem-
(593 to 1982C). Its use in the power industry is gen- perature range of –40 to 1000F (–40 to 538C). Instru-
erally limited to the laboratory. An adaptation of this ment readings can be in error if the capillary tubing
method of temperature determination is used when is subjected to temperature changes.
observing ash and slag accumulations soften and flow The liquid-in-glass thermometer is a simple, direct
in a coal-fired furnace or superheater.
Fusion pyrometers are also made in the form of
crayons, paint and pellets. The crayons and paint,
applied to a metal surface, produce a flat-finish mark.
A change in the finish from flat to glossy indicates that
the surface temperature has reached or exceeded a
selected value. The pellets melt at specified tempera-
tures when in contact with a hot surface and may be
seen more easily than crayon marks.
Vaporization The vapor pressure of a liquid depends
on its temperature. When the liquid is heated to its
boiling temperature, the vapor pressure is equal to the
total pressure above the liquid surface. The boiling
points of various pure chemical elements or compounds
at standard atmospheric pressure [29.92 in. Hg (101.33
kPa)] can be used as thermometric fixed points. If the
liquid and vapor are confined, the increase in vapor
pressure may be used to measure temperature using a
calibrated pressure gauge. (See Fig. 6.) Fig. 6 Schematic assembly of vapor pressure thermometer.

40-4 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

reading, portable instrument. Low precision thermom- Within the visible range, two types of temperature
eters are inexpensive and instruments of moderate measuring instruments, the optical pyrometer and the
precision are available for laboratory use. This ther- radiation pyrometer, sense the emitted radiation to
mometer features a reservoir of liquid in a glass bulb indicate temperature.
that is connected directly to a glass capillary tube with Optical pyrometers The optical pyrometer visually
graduated markings. Mercury, the most common liq- compares the brightness of an object to a reference
uid, is satisfactory from –40F (–40C), just above its freez- source of radiation. The typical internal calibration
ing point, to about 600F (316C) if the capillary space source is an electrically heated tungsten filament. A
above the mercury is evacuated. The upper temperature red filter may also be used to restrict this visual com-
limit may be 900F (482C) or higher if this space is filled parison to a particular wavelength. This instrument
with pressurized nitrogen or carbon dioxide (CO2). measures the temperature of surfaces with an emis-
The use of unprotected glass thermometers is re- sivity of 1.0, which is equivalent to that of a black-
stricted to laboratory or research applications. For body. By definition, a blackbody absorbs all radiation
more severe service there are various designs of in- incident upon it, reflecting and transmitting none.
dustrial thermometers with the bulb and stem pro- When calibrated and used properly, the pyrometer
tected by a metal casing and usually arranged in a yields excellent results above 1500F (816C). Tempera-
thermometer well. Response to rapid temperature ture measurement of the interior surface of a uni-
changes is slower than with the unprotected labora- formly heated enclosure, such as a muffle furnace, is
tory type instrument. a typical application. When used to measure the tem-
Solids The expansion of solids when heated is rec- perature of an object outside a furnace, the optical py-
ognized in thermometers using a bimetallic strip. Flat rometer always reads low. The error is small [20F
ribbons of two metals with different coefficients of (11C)] for high emissivity bodies, such as steel ingots,
thermal expansion are joined face to face by riveting and large [200 to 300F (111 to 167C)] for unoxidized
or welding to form a bimetallic strip. When the strip liquid steel or iron surfaces.
is heated, the expansion is greater for one metal than The optical pyrometer is widely used for measuring
for the other. A flat strip bends, or changes curvature objects in furnaces at steel mills and iron foundries. It
if it was initially in a spiral form. Bimetallic strips are is not applicable for measuring gas temperature be-
seldom used in power plant thermometers but are cause clean gases do not radiate in the visible range.
widely used in inexpensive thermometers, household
Radiation pyrometers In one type of radiation py-
thermostats and other temperature control and regu-
rometer, all radiation from the hot body, regardless of
lating equipment. They are particularly useful for
wavelength, is absorbed by the instrument. The heat
automatic temperature compensation in other instru-
absorption is measured by the temperature rise of a
ment mechanisms.
delicate thermocouple (TC) within the instrument. The
Radiation properties thermocouple is calibrated to indicate the temperature
All solid bodies emit radiation. The amount is very of the hot surface at which the pyrometer is sighted,
small at low temperatures and larger at high tempera- on the assumption that the surface emissivity equals
tures. The quantity of radiation may be calculated by 1.0. The hot surface must fill the instrument’s entire
the Stefan-Boltzman formula: field of view.
This type of radiation pyrometer has high sensitiv-
q/S = σ ε T4 (2) ity and precision over a wide range of temperatures.
The instrument gives good results above 1000F (538C)
where when used to measure temperatures of high emissiv-
q =
radiant energy per unit time, Btu/h ity bodies, such as the interiors of uniformly heated
S =
surface area, ft2 enclosures. Because operation of the instrument does
σ =
Stefan-Boltzman constant, 1.71× 10–9 Btu/h ft2 R4 not require visual comparison, radiation pyrometers
ε =
emissivity of the surface, dimensionless (usu- may be used as remotely operated temperature indi-
ally between 0.80 and 0.95 for boiler materials) cators. Errors in temperature measurements of hot
T = absolute temperature, R = F + 460 bodies with emissivities of less than 1.0, especially if
they are in the open, are extremely large.
At low temperatures the radiation is primarily in Radiation pyrometers sensitive to selective wave-
the invisible infrared range. As the temperature rises, lengths in the infrared band give good results when
an increasing proportion of the radiation is in shorter measuring temperatures of bodies or flames. One de-
wavelengths, becoming visible as a dull red glow at sign uses a system of lenses, a lead sulfide cell and
about 1000F (538C) and passing through yellow to- electronic circuitry to produce a measurement.
ward white at higher temperatures. The temperature An advancement of the single band pyrometer is
of hot metals [above 1000F (538C)] can be estimated by the two color pyrometer, which measures the intensi-
color. For iron or steel, the approximate color scale is: ties of two selected wave bands of the visible spectrum
dark red 1000F (538C) emitted by a heated object. This unit computes the total
medium cherry red 1250F (677C) of the emitted energies and converts it into a tempera-
orange 1650F (899C) ture indication. As with the optical pyrometer, indications
yellow 1850F (1010C) depend on sighting visible rays and therefore its mini-
white 2200F (1204C) mum temperature measurement is 1000F (538C).

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-5


The Babcock & Wilcox Company

Infrared radiation is produced by all matter at tem-


peratures above absolute zero and detectors sensitive
to the infrared band may be used to measure objects
at low temperatures where the radiation is not in the
visible band. The development of metal-silicide detec-
tors permits sensitivities in the 1 to 5 µm band. Metal-
silicides are formed by the reaction of metals, such as
platinum or palladium, with p-type silicon. When
coupled with electronic circuitry for producing an im-
age, the detectors are sophisticated devices for low
temperature measurement.1
Infrared thermal imaging equipment is available
for quantitative measurement or qualitative imaging.
The instruments for making quantitative measure-
ments and providing target temperature indications
are called thermographic imagers or imaging radiom-
eters. Qualitative imaging instruments are called ther-
mal viewers. These viewers are mechanically or elec-
tronically scanned; the electronically scanned units are
called pyroelectric vidicons or pyrovidicons.2
Infrared temperature measurement is applicable to
surveys of surfaces, such as boiler casings and insu-
lated steam pipes, for locating temperature variations.
These instruments may also assist in determining elec-
trical component temperatures in operating circuits.
Radiation pyrometers are not capable of determining
Fig. 7 Diagrams of electrical circuits for resistance thermometers
gas temperatures. (courtesy of Leeds and Northrup Company).
Optical properties
Optical fibers have been used in fixed (end) and coil. From room temperature to 250F (121C), commer-
distributed (average temperature along length) tem- cial instruments usually have resistance coils of nickel
perature measurement. Development is continuing in or copper. Platinum is used for higher temperatures
the application of this method to temperatures above and in many high precision laboratory instruments
450F (232C).3 to cover a wide range of temperatures.
The electrical resistance thermometer can be used
Electrical properties for the remote operations of indicating, recording or
Two classes of widely used temperature measuring automatic controlling. If proper precautions are taken,
instruments, the thermocouple and the electrical re- it is stable and accurate, but it is less rugged and less
sistance thermometer, are based on the relation of versatile than a thermocouple.
temperature to the electrical properties of metals.
Because of its versatility, convenience and durabil- Acoustic properties
ity, the thermocouple is of particular importance in Acoustic properties of gases vary with temperature
power plant temperature measurements. A separate and are used as the basis for temperature measuring
section later in this chapter is devoted to the thermo- instruments. The velocity of sound in a gas varies di-
couple and its applications. rectly with the square root of its absolute temperature,
Resistance thermometer The electrical resistance ther- as indicated in the following formula:
mometer, used over a temperature range of –400 to
1800F (–240 to 982C), depends on the increase in elec- c = gc kRT / M (3)
trical resistance of metals with increasing temperature;
this relationship is nearly directly proportional. If the where
electrical resistance of a calibrated wire material is mea-
c = speed of sound, ft/s
sured (by a Wheatstone bridge or other device), the tem-
gc = proportionality constant, 32.17 lbm ft/lbf s2
perature of the wire can be determined.
k = specific heat ratio, dimensionless
In the simplest form of the Wheatstone bridge sys-
R = universal gas constant, 1545 ft lb/mole R
tem, shown in Fig. 7a, the reading would be the sum
T = absolute temperature, R = F + 460
of the resistances of the calibrated wire and the leads
M = molecular weight, lb/mole
connecting this wire to the Wheatstone bridge. How-
ever, this value would be subject to error from tem- By measuring the time required for sound to traverse
perature changes in the leads. By using a more re- a specific distance, the average gas temperature can
fined circuit shown in Fig. 7b, the resistance of the be calculated. This phenomenon has been commer-
leads can be eliminated from the instrument reading. cially applied to measuring furnace gas temperatures.
To localize the point of temperature measurement, the By developing a number of independent temperature
resistance wire may be made in the form of a small paths a thermal contour may be developed.

40-6 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

Other properties
Other material properties that vary with tempera-
ture are also used as the basis for temperature mea-
suring instruments. By flowing gas through two ori-
fices in series, a method has been developed for mea-
suring gas temperatures up to 2000F (1093C) com-
mercially and up to 4500F (2482C) in the laboratory.
Temperatures ranging from absolute zero to 5000F
(2760C) have been accurately determined by using a
laboratory instrument that measures electron motions.
High temperature research has been stimulated by
gas turbine and jet propulsion applications, and fur-
ther development of new and improved temperature
measuring instruments is expected.

Thermocouples Fig. 8 Principle of the thermocouple illustrated.


A thermocouple consists of two electrical conductors
of dissimilar materials joined at the ends to form a cir- parallel to obtain a single reading of average tempera-
cuit. If one of the junctions is maintained at a tem- ture. In this case, the resistance of each thermocouple
perature higher than the other, an electromotive force must be the same. The emf across the terminals of
(emf) is generated, producing current flow through such a circuit is the average of all the individual emf
the circuit, as shown in Fig. 8. values and may be read on a potentiometer normally
The magnitude of the net emf depends on the tem- used for a single thermocouple.
perature difference between the two junctions and the
materials used for the conductors. No unbalance, or Practical application and multiple circuits
net emf, is generated if the two junctions are at the The limitation of the potentiometer circuitry to
same temperature or if the conductors are of the same measure small emf values has led to various circuits
material at different temperatures. If one junction of of multiple thermocouples connected in series or par-
a calibrated thermocouple is maintained at a known allel. These circuits make it possible to measure ex-
temperature, the temperature of the other junction tremely small temperature variations or average tem-
can be determined by measuring the net emf produced; peratures. Thermocouple circuit considerations of wire
this emf is almost directly proportional to the tempera- diameter and length, while less important for steady-
ture difference of the two junctions. The relationship
between emf and the corresponding temperature dif-
ference has been established by laboratory tests
throughout the temperature ranges for common ther-
mocouple materials. These values are plotted in Fig.
9. The thermocouple is a low cost, versatile, durable,
simple device providing fast response for accurate tem-
perature measurement. It also allows convenient cen-
tralized reading.
In the potentiometer circuit, copper leads may con-
nect the reference junction terminals to the measur-
ing instrument without affecting the net emf of the
thermocouple. If the instrument is at a uniform tem-
perature, no emf is set up between the copper conduc-
tors and slide wire materials within the potentiometer.
However, if temperature differences exist within the
instrument, there will be disturbing emf values in the
circuit and the readings will be affected.
Multiple circuits
If two or more thermocouples are connected in se-
ries, the net emf at the outside terminals is equal to
the sum of the emf values developed by the individual
thermocouples. When all of the individual hot and cold
junctions are maintained at the same respective tem-
peratures, as in the device known as the thermopile,
this multiplied emf value makes it possible to detect
and measure extremely small temperature variations. Fig. 9 Relation of temperature to electromotive force produced by
Two or more thermocouples may be connected in several commonly used thermocouples.

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-7


The Babcock & Wilcox Company

state measurement, will affect thermocouple readings


during dynamic temperature conditions. These appli-
cations require detailed circuit design investigation.
The application of thermocouples requires attention
to the insulation of the lead wire to prevent erroneous
circuits. Parasitic couples may be introduced by mechani-
cal, soldered, or brazed connections if temperature varia-
tions exist within different metals of the connections.
Selection of materials The combinations of metals
and alloys most frequently used for thermocouples are
listed in Table 1 with their general characteristics and
useful temperature ranges. Selection depends largely
on the ability to withstand oxidation attack at the
maximum expected service temperature. Durability
depends on wire size, use or omission of protection
tubes, and nature of the surrounding atmosphere.
All thermocouple materials deteriorate when ex-
posed at the upper portion of their temperature range Fig. 10 Sheathed-type thermocouple.
to air or flue gases and when in contact with other
materials. Platinum in particular is affected by me- and resistance to oxidizing, reducing, or otherwise
tallic oxides and by carbon and hydrocarbon gases corrosive atmospheres. Sheathed thermocouples are
when used at temperatures above 1000F (538C) and available as grounded or nongrounded types, as illus-
is subject to calibration drift. The upper temperature trated in Fig. 10. The grounded type has a more rapid
limits shown in Table 1 are well below the tempera- response to temperature change but can not be used
tures at which rapid deterioration occurs. for connection in series or parallel, because it is
For high temperature duty in a permanent instal- grounded to the sheath. For these applications, the
lation or where destructive contact is likely, service life nongrounded type should be used. The grounded type
may be extended, at some sacrifice in response speed, is susceptible to separation or parting of the thermo-
by using closed-end protection tubes of alloy or ceramic couple wire where long leads at high temperatures are
material. The arrangement should permit removal of used, whereas the nongrounded type appears satis-
the thermocouple element from the protection tube for factory for this service.
calibration or replacement. Omission of the protection Sheathed thermocouples may be equipped with
tube may require frequent calibration and corrections pads for welding to tubes, as illustrated in Fig. 11. Pad-
for calibration drift. For short periods of service within type thermocouples should not be used when the pad
the useful range of the thermocouple selected, the is exposed to temperatures that are markedly differ-
correction for calibration change is usually negligible. ent from that of the body being measured, because this
Sheathed-type thermocouple Sheathed-type mag- added metal surface can radiate or absorb heat. Typi-
nesium oxide insulated thermocouples have been in cal applications of pad-type thermocouples are drum,
use for some time. The thermocouple wires are insu- superheater header, or tube surface temperature
lated with inert magnesium oxide insulation that pro- measurements. (See Fig. 12.)
tects the wires from the deteriorating effects of the en- Thermocouple and lead wire There are two classes
vironment. Sheaths can be made of stainless steel and of wire for thermocouples, the closely standardized
other resistant materials, ensuring relatively long life and matched thermocouple wire and the less accurate

Table 1
Types of Thermocouples in General Use
Type of Useful Temperature Maximum Millivolts at 500F Magnetic
Thermocouple (Note 1) Range, F (C) Temperature, F (C) (260C) (Note 2) Wire

(+) Copper to constantan (−) −300 to 650 (−184 to 343) 1100 (593) 13.24 −
(+) Iron to constantan (−) 0 to 1400 (−18 to 760) 1800 (982) 15.01 Iron (+)
(+) Chromel to constantan (−) −300 to 1600 (−184 to 871) 1800 (982) 17.94 −
(+) Chromel to alumel (−) 0 to 2300 (−18 to 1260) 2500 (1371) 11.24 Alumel (−)
(+) 90% Pt  10% Rh to
platinum (−) 900 to 2600 (482 to 1427) 3190 (1754) (Note 3) 2.048 −

Notes:
1. Nominal composition: constantan, 55% Cu, 45% Ni; chromel, 90% Ni, 10% Cr; alumel, 95% Ni, 5% Al, Si and Mn
combined.
2. Reference junction 0F (−18.0C).
3. Melting point.

40-8 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

Fig. 11 Sheathed thermocouple with pad.

compensating lead wire. For thermocouples of noble


metal, extension leads of copper and copper-nickel
alloy, which have an emf characteristic approximat-
ing the noble metal pair, are used to reduce cost. For Fig. 13 Thermocouple and lead wire arrangement to measuring
instrument through selector switch.
thermocouples of base metal, the extension lead wire
is the same composition as the thermocouple wires, but
it is less expensive because the control in manufac- by joining wires of the same composition. Polarity is
ture and calibration is less rigorous. usually identified by color code or tracers in the wire
For accuracy, the matched thermocouple wire should covering.
be used at the hot junction and continued through the Hot junction Wire and pipe-type elements for hot
zone of greatest temperature gradient to a point close junctions may be purchased or may be made from stock
to room temperature, where a compensating lead wire thermocouple wire, as shown in Fig. 14. These types
may be spliced in for connection to the reference junc- are useful for direct immersion in a gas stream, for
tion. The reference junction is usually located at the insertion into thermometer wells, or for thermal con-
recorder or central observation point. Where a num- tact with solid surfaces.
ber of thermocouples are used, it may be economical Where the temperature of a metallic surface is to
to establish a zone box (see Fig. 13) from which cop- be measured, a versatile hot junction may be formed
per conductors are extended to the measuring instru- by peening the wires separately into holes drilled in
ment. the surface of the metal. Steps for making this type of
Splicing of extension lead wires may be done by junction are illustrated in Fig. 15.
twisting the wire ends together, by using screw clamp The peened junction has the advantage of being
connectors, or by fusion welding, soldering or brazing. completely mechanical, with a tight and remarkably
Care should be taken to maintain correct polarity strong attachment and minimum interference with
the temperature of the object. The temperature indi-
cated is essentially that of the metallic surface, which
is the first point of contact with the conductors. The
depth of the drilled hole has no temperature effect; it
only provides mechanical strength. However, the junc-

Fig. 12 Application of pad-type thermocouple. Fig. 14 Thermocouple elements and hot junctions.

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-9


The Babcock & Wilcox Company

mocouple, the reference junction temperature is usu-


ally automatically compensated for by an internal
resistor circuit. This circuit compensates for deviations
from standard room temperature. The resistor circuit
characteristics are specific to the type of thermocouple
material for which the potentiometer is calibrated.
Some potentiometers provide a selector switch to di-
rectly read different types of thermocouples.
Temperature recorders
In temperature recorders, a power operated mecha-
nism automatically adjusts the slide wire of a potenti-
ometer circuit to balance the emf received from the
thermocouple. The action of the slide wire is coordi-
nated with the movement of a recording pen or print
wheel that is drawn over, or impressed on, a moving
temperature graduated chart.
As a result, an essentially continuous record may
be traced for one thermocouple, or a rapid scanning
of many thermocouples may be recorded.
Checking recorder calibration Compared with the
manually operated potentiometer, all recorders are
inherently less accurate because of wear and lost
Fig. 15 Procedure for installing a peened-junction thermocouple to
motion in the actuating mechanism, shrinkage, or
measure surface temperature of metal. printing inaccuracies of paper and chart scales.

tion contact points are subject to thermal conduction Temperature of fluids inside pipes
in the wires. Where this type of error is significant,
precautions should be taken to prevent temperature The temperature of a fluid (liquid, gas or vapor)
gradients in the wires. Drilled holes for the wire ends flowing under pressure through a pipe is usually
should allow a snug fit before peening. For installa- measured by a glass thermometer, an electrical resis-
tions requiring a significant number of thermocouples, tance thermometer, or a thermocouple. Each thermom-
there are portable spot welding machines available to eter is inserted into a well, or thermowell, projecting
join the ends of the wire directly to a steel surface. into the fluid. The thermowell is preferred but a ther-
mocouple properly attached to the outside of a pipe
Millivolt potentiometer wall can provide good results. The thermowell provides
The temperature-emf relationships of standardized
thermocouple wires, as established and published by
the manufacturer, should be used to convert the po-
tentiometer readings to equivalent temperature val-
ues. These tabulated values represent the net emf
impressed in the potentiometer terminals when the
reference junction is at 0F (–18C) and the measuring
junction is at the temperature listed. The millivolt
potentiometer may be used with any type of thermo-
couple. It is frequently used when several different
thermocouple types are installed.
It is usually impractical to maintain the reference
junction at 0F (–18C) while taking readings. Correc-
tion for any reference junction temperature can be
made by adding to each observed emf reading the emf
value corresponding to the reference junction tem-
perature. The temperature of the measuring junction
can then be determined. Most millivolt potentiometers
are equipped with a compensator, which should be
used to correct for the reference junction temperature.
Direct millivolt readings then correspond to the actual
thermocouple temperature.
Direct reading potentiometer
A potentiometer may be graduated to read the hot
junction temperature directly instead of in millivolts.
When calibrated for use with a specific type of ther- Fig. 16 Thermocouple well installation.

40-10 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

an average temperature if temperature stratification


exists in the fluid, because thermal conduction along
the well equalizes the temperature.
A metal tube closed at one end, screwed into or
welded to a pipe wall, and projecting into the fluid,
serves as a thermowell (Fig. 16). It must have strength
and rigidity to withstand hydrostatic pressure, resis-
tance to bending and vibration caused by its resistance
to fluid flow, and material compatibility for a welded
attachment. To minimize heat conduction to the sur-
roundings and to give rapid response to temperature
changes, the well should be as small as possible. The
portion outside the pipe should also be small. The
material of the well must resist the erosive or corro-
sive action of the fluid. Because the temperature may
be locally depressed by acceleration of a compressible
fluid through a constriction in a pipe, the well should
be carefully positioned, taking these factors into ac-
count. Care should also be observed when locating the
well for measuring steam temperature downstream of
spray water attemperator injection points, as inad-
equate mixing of the steam and spray water may re-
sult in fluid temperature measurement errors. Project-
ing parts of the thermowell and the pipe wall should
be thoroughly insulated to prevent heat loss. Substan-
tial error can result from air circulation within the
thermowell and should be minimized by packing with
insulating material.
Thermowell designs vary. If the fluid is a liquid or
a saturated vapor, good heat transfer is assured and
a plain well is satisfactory. If the fluid is a gas or su-
perheated vapor, a finned well is sometimes used. The
method of attachment, a mechanical thread or welded
attachment to the pipe wall, is optional provided safety Fig. 17 Thermocouple wires extending from hot junction disposed
code requirements are met (see Fig. 16). on pipe wall before leading outside.

Tube temperature measurement Furnace wall tubes


It is frequently desirable to know the metal tem- In measuring water-cooled furnace wall tube tem-
perature of tubes in different classes of service. These peratures, special protection for the thermocouple and
classes include furnace wall or boiler bank tubes that its lead wires must be provided because of the destruc-
are cooled by water and steam at saturation tempera- tive high temperature furnace atmosphere and, in
ture, economizer tubes cooled by water below the tem- some cases, the accumulation and shedding of ash and
perature of saturation, and superheater and reheater slag deposits. If a simple peened thermocouple is used,
tubes cooled by steam above saturation temperature. the results may lack accuracy because of errors from
These temperature measurements may be used to conduction; in most cases, the service life would be
determine the safety of pressure parts, the uniformity short because of physical damage or wire deteriora-
among tubes in parallel flow circuits, or the fluid tem- tion from overheating. Cover plates welded to the
perature increase between inlet and outlet conditions. tubes have been used to protect the wires, but these
The peened-type hot junction, illustrated in Fig. 17, plates interfere with the normal heat transfer and
is satisfactory in many cases and is probably the sim- cause local ash deposits to create abnormal conditions
plest form of thermocouple application. Sheathed-type at the temperature measurement point.
thermocouples with pads are generally used to per- The use of chord-drilled holes through which the
manently monitor temperatures of tubes not exposed thermocouple wires are laced (Fig. 18) was developed
to external heat. (See Fig. 11.) Properly applied insu- by The Babcock & Wilcox Company (B&W); this is a
lation of sufficient thickness in the vicinity of the ther- satisfactory temperature measurement method for fur-
mocouple attachment will reduce the effects of ther- nace wall tubes. The tube surface is free of projections,
mal conduction along the tube. The thermocouple wire the wires are protected, and the thermal conduction
wrapped around the tube will minimize the conduc- effect at the hot junction is minimized because the
tion along the thermocouple wire. These techniques wires pass through an essentially isothermal zone
are illustrated in Fig. 17. When properly installed the before emerging into cooler surroundings at the rear
surface thermocouple may be used to measure both of the tube. To minimize the effect on the heat flow pat-
metal and fluid temperatures. tern within the tube metal, the chord-drilled holes should

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-11


The Babcock & Wilcox Company

time lag between surface and depth thermocouples.


By using the surface and depth dimensions, the
equivalent lengths can be determined and the tem-
perature gradients can be plotted. In practice, the heat
flows through a tube wall in a path of decreasing sec-
tional area and the temperature gradient through the
metal is not linear with respect to thickness. The gra-
dient, however, may be drawn as a straight line if tem-
peratures are plotted against an equivalent length of
flow path (le). This represents an equivalent flat plate
having a uniform flow path with an equivalent but
greater thickness. This is illustrated in Fig. 19.
The equivalent plate thickness is calculated in the
following manner:
le = R n ( R / r ) (4)
where
R = outside radius of tube, in.
r = inside radius of tube, in.
l e = equivalent flat plate thickness, in.
In many cases, thermocouple error may represent
a significant portion of the gradient being measured.
Ideally, lines drawn through the surface and depth
temperature (Fig. 20) should intersect the inner sur-
face equivalent thickness line slightly above the fluid
Fig. 18 Chordal thermocouple.
temperature as measured by a thermocouple peened
on the outside surface of the tube opposite the furnace.
be as small as possible. The effect of these holes on the The amount above the fluid temperature represents
strength of the tube is small. This effect is minimal in the the temperature drop across the fluid film and this
critical direction of hoop stress and is readily tolerated in drop increases with increased heat input. It is diffi-
the direction of longitudinal stress. The use of chord-drilled cult to obtain consistently drilled thermocouple holes
holes in the thicker, high pressure wall tubes is a practi- that reflect the pattern shown in Fig. 20. If, at low
cal method of obtaining surface metal temperatures. inputs, the line drawn through the surface and depth
temperature intersects the inner surface equivalent
Gradient thermocouple thickness line at fluid temperature or slightly above,
Measuring the temperature gradient through a it may be considered as satisfactory. If it falls below or
tube wall is a means of determining the heat flow rate well above this point, a correction or calibration may
through the wall and of detecting the accumulation be made by ignoring the measured equivalent depth
of certain types of internal deposits. dimension, plotting a line through the outer and in-
To obtain the best results, a large temperature gra- ner surface temperatures, then adjusting the equiva-
dient through the wall is desirable. However, ideal lent depth to a value represented by the intersection
conditions may be compromised to permit an arrange- of the adjusted gradient and depth thermocouple tem-
ment that provides measurable metal temperatures perature (Fig. 21).
and is practical to fabricate. Fig. 18 shows a section Heat flux measurement Once a series of temperature
through a tube which illustrates the drilling and in- plots shows consistent results, heat inputs may be cal-
stallation of typical surface and depth thermocouples. culated:
Calibration Sheathed thermocouples are suitable
for this temperature gradient application. For greater k (T1 − T2 )
accuracy the thermocouples may be laboratory cali- q′′ = (5)
brated. No procedure is available to accurately or eco- le
nomically calibrate the thermocouple tube assembly where
in the field. The most satisfactory method is as follows. q″ = heat flux, Btu/h ft2
Following installation, obtain a series of tempera- k = thermal conductivity, Btu/h ft F
ture readings on all thermocouples at two or three T1 = outside surface thermocouple temperature, F
different ratings on the boiler, from a low rate to maxi- T2 = depth thermocouple temperature, F
mum rate, while the tube is known to be internally le = equivalent plate thickness (surface to depth), ft
clean; read all temperatures as simultaneously as (Fig. 19)
possible. Several series of readings should be taken
on the thermocouples at each rating. An average Scale detection If the temperature gradients are to
should be calculated for each thermocouple to correct be used as a scale detector, then thermocouples should
for minor differences in temperatures resulting from be read periodically, plotted, and checked for a change
heat input variations. This also compensates for the in the temperature above saturation.

40-12 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

Fig. 19 Thermocouple installation.

To assure a close comparison, the thermocouple tem- Limitations of gradient thermocouples The gradient
peratures should always be read with the calibrated thermocouple method of measuring heat flow
equipment and under operating conditions close to through tube walls is a guide rather than an abso-
those existing during the original setup. An accumu- lute measure. Relatively small dimensions and small
lation of internal deposits in the thermocouple zone is differential temperatures exist between surface ther-
marked by an increased temperature difference from mocouple and depth thermocouple locations. Also, any
inner surface to saturation. small error in determining temperatures or in metal
When the average of the surface and depth tem- thickness between thermocouples usually represents
peratures approaches the limit for the tube metal, the a large percentage of this small difference.
boiler should be shut down and the inside tube sur- The effectiveness of the gradient temperature plot
faces inspected. Sufficient allowance should be made as an internal scale detector is also dependent on the
for tube to tube temperature variations. nature of the scale. Certain types of scale, such as
carbonate or silica deposits, accumulate uniformly,
while iron oxide deposits can accumulate irregularly.
The location of nonuniform deposits is uncertain and
the gradient thermocouple method is not reliable for
detecting such accumulations.
For a quick determination of internal tube deposit
changes, the difference between surface and depth
temperature readings can be plotted against the dif-
ference between surface and fluid temperature read-
ings. With a clean tube, points should fall along a straight
line over a range of inputs. As deposits form, the differ-
ence between surface and fluid temperature increases
for an unchanging surface and depth temperature dif-
ference. Fig. 22 shows a typical plot of such conditions.

Fig. 20 Sample plot of temperature gradients for low, medium and


high heat rates. Fig. 21 Adjustment of temperature gradient.

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-13


The Babcock & Wilcox Company

vection transfer from the hot gases in which it is im-


mersed, it is also subject to heat exchange by radiation
to and from the surrounding surfaces and by conduc-
tion through the instrument itself. If the temperature
of the surrounding surfaces does not differ from that
of the gas (gas flowing through an insulated duct), the
temperature indicated by the instrument should accu-
rately represent the gas temperature. If the tempera-
ture of the surrounding surfaces is higher or lower than
that of the gas, the indicated temperature will be cor-
respondingly higher or lower than the gas temperature.
The variation from the true gas temperature de-
pends on the temperature and velocity of the gas, the
temperature of the surroundings, the size of the tem-
perature measuring element, and the construction of
the element and its supports. To correct for the errors
in temperature measurement caused by the surround-
ings, it is best to calibrate the instrument to a known
and reliable source.
As an example, consider a 22 gauge (0.7112 mm)
bare thermocouple. When it is used to measure the gas
temperature in boiler, economizer, or air heater cavi-
ties with surrounding walls cooler than the gas, the
error in the observed readings may be found from Fig.
23 (line for bare TC).
High velocity thermocouple
The design and operation of steam generating units
depend on the evaluation of gas temperatures in the
furnace and superheater sections of the equipment.
Boiler design to achieve successful thermal perfor-
mance must take into account the limitations imposed
Fig. 22 Plot of chordal thermocouple temperatures showing effect by the allowable metal temperatures of superheater
of internal deposits.

Superheater and reheater applications


Chordal surface thermocouples can be used to mea-
sure metal temperatures of superheater or reheater
tubes in the gas stream, using the principles previ-
ously described. However, special provision must be
made to protect the wires between the point of mea-
surement and the exit from the boiler setting. This can
be done by containing the sheathed thermocouple in
stainless steel tubing that is welded to the superheater
or reheater tube. This maintains the sheath and pro-
tection tubing at a temperature which is approxi-
mately that of the superheater or reheater tube. Ad-
ditional cooling of the stainless steel protection tube
is frequently required.

Gas temperature measurement


Temperature measurement of combustion gas at dif-
ferent locations within the boiler is important to the
boiler design engineer and the operating plant engi-
neer. Accurate temperature measurements of gas en-
tering and leaving the heat absorbing components can
confirm design predictions and operating performance.
In all cases of gas temperature measurement, the
temperature sensitive element approaches a tempera-
ture that is in equilibrium with the conditions of its Fig. 23 General magnitude of error in observed readings when
environment. While it receives heat primarily by con- measuring temperature in boiler cavities with thermocouples.

40-14 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

tubes and by the fusing characteristics of ash and slag fer to the thermocouple by convection is proportional
from the current or expected fuel. The overall complex- to the gas mass velocity and the temperature differ-
ity of combustion and heat transfer relationships pre- ence between gas and thermocouple, the junction tem-
vents exact calculation and accurate measurement of perature may be brought closer to the true tempera-
the gas temperatures provides information to confirm ture of the gas by increasing the rate of mass velocity
calculation methods. and convection heat transfer at the thermocouple while
The optical pyrometer and radiation pyrometer are shielding the junction from radiation.
not designed to measure gas temperature. Excessive A special high velocity thermocouple probe for mea-
error is also encountered when gas temperatures in suring high gas temperatures in boilers is illustrated
the furnace and superheater areas are measured with in Fig. 25. This portable assembly is primarily used
a bare thermocouple. for making test traverses in high duty zones by in-
The high velocity and multiple-shield high veloc- sertion through inspection doors or other test open-
ity thermocouples (HVT and MHVT), developed to ings in the setting.
correct for radiation effects, are the best instruments This thermocouple is supported by a water-cooled
available for measuring high gas temperatures in probe. The measuring junction is surrounded by a
cooler surroundings or low gas temperatures in hot- tubular porcelain radiation shield through which gas
ter surroundings. Cross-sections through single and flow is induced at high velocity by an attached aspi-
multiple-shield high velocity thermocouples developed rator. The gas aspiration rate over the thermocouple
for use in boiler testing are shown in Fig. 24. can be checked by an orifice incorporated with the
The surfaces (water-cooled walls, or superheater or aspirator and connected to the probe by a flexible hose.
boiler tube banks) surrounding the usual location of The gas mass velocity (or mass flux) over the thermo-
a gas traverse are cooler than the gases. Consequently, couple junction should be at least 15,000 lb/h ft2 (20.34
the readings from a bare unshielded thermocouple kg/m2 s). Convection heat transfer to the junction and
indicate lower temperatures than those obtained with shield is simultaneous and both approach the tempera-
an HVT. For the same reason, an HVT generally in- ture of the gas stream. Radiation transfer at the junc-
dicates lower values than an MHVT. A comparison be- tion is diminished by the shield. Because the shield is
tween bare thermocouple, HVT and MHVT results in exposed to the radiation effect of the surroundings, it
typical boiler furnaces and cavities is given in Fig. 23. may gain or lose heat and its temperature may be
MHVT measurements closely approach true gas slightly different from the junction temperature.
temperatures. In this design, the thermocouple junc- With the increasing size of steam generators, han-
tion is surrounded by multiple shields, all of which dling long HVT probes has become more difficult but,
receive heat by convection induced by the high gas
flow rate. In this manner, the heat transfer by radia-
tion is so reduced that there is virtually no heat ex-
change between the junction and the innermost shield.
Because of the small flow areas that rapidly become
clogged by ash, use of the MHVT is limited to clean gas
conditions. Where traverses are taken in dust- or slag-
laden gases, it is usually necessary to use the HVT. The
readings are corrected by comparison with results ob-
tained under clean gas conditions from an MHVT.
For temperatures exceeding 2200F (1204C), noble
metal thermocouples are required and it is important
to protect the thermocouple from contamination by the
gases or entrained ash. Various coverings, shown in
Fig. 24, provide some protection for the wires, espe-
cially when fouling occurs from molten slag at tem-
peratures above 2400F (1316C). When platinum ther-
mocouples are used in gas above 2600F (1427C), ap-
preciable calibration drift may occur even while tak-
ing measurements requiring only several minutes of
exposure time. The thermocouple elements should be
checked before and after use with corrections applied
to the observed readings. When the error (∆T) reaches
40 to 60F (22 to 33C), the contaminated end of the ther-
mocouple should be removed and a new hot junction
should be made using the sound portion of the wire.
Because heat transfer by radiation is proportional
to the surface area, emissivity, and difference of the
source and receiver absolute temperatures, the effects
of radiation increase as the temperature difference
between the thermocouple hot junction and the sur- Fig. 24 Shield assemblies for high velocity thermocouple (HVT) and
rounding surfaces increases. Also, because heat trans- multiple-shield, high velocity thermocouple (MHVT).

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-15


The Babcock & Wilcox Company

Fig. 25 Rugged, water-cooled, high velocity thermocouple (HVT) for determining high gas temperatures.

to date, no acceptable point by point measurement of and insert these thermocouples using a sootblower
high temperature gases has been fully developed. carriage. The remotely insertable thermocouple is
called a thermoprobe.
Measurement evaluations
Large boilers may have significant variation in ac- Insulation and casing temperature
tual measurements at a particular measurement
plane. Where significant data variation exists, data Outer surfaces
reduction may require mathematical methods to The measurement method should be carefully se-
weight-average individual measured values during lected to avoid significant errors in measuring
the data reduction process. uncased insulation surface temperature. Portable con-
Flue gas temperature measurements at the econo- tact thermocouple instruments designed to be pressed
mizer outlet during performance tests are an example against a surface are unsatisfactory on insulated sur-
of a large number of data points which require reduc- faces because the instrument cools the surface at the
tion to a single representative temperature. point of contact, and the low rate of heat transfer
A weighted average temperature becomes time con- through the insulating material prevents adequate
suming and the need for such accuracy must be justi- heat flow from surrounding areas to the contact point.
fied. An average of the individual temperature mea- Thermocouple attachment to the insulation surface
surements is usually acceptable if a sufficient num- must not appreciably alter the normal rate of heat
ber of points are obtained. The average gas tempera- transmission through the insulation and from the
ture may be approached by increasing the number of surface to the surroundings. Fine wires can be at-
points or by instrument design to satisfy a particular tached and maintained at the surface temperature
requirement. For example, in a heat loss efficiency test, more easily than heavy wires. If the insulation is plas-
multiple measuring locations are used to indicate flue tic at the time of application, press the thermocouple
gas exit temperatures for determining dry gas loss. The junction and several feet of lead wire into the surface
number of points required by the ASME Performance of the insulation; the thermocouple will adhere when
Test Code permits efficiency accuracies within 0.05%. the insulation hardens. If the insulation is hard and
Under certain circumstances, the maximum, rather dry, the junction and lead wires may be cemented to
than average, gas temperature may be required for the surface using a minimum of cement. Fastening the
equipment protection. A moveable probe with one or wires to the surface with staples introduces conduc-
two thermocouples may be used to determine the lo- tion errors. Covering the wires with tape changes the
cation of this high temperature. For instance, during heat transfer characteristics of the surface and imposes
boiler pressurization and before steam is flowing an undesirable insulation layer between the wire and
through the superheater or reheater tubes, a bare the ambient air.
thermocouple temporarily installed in the gas stream
immediately before the tubes may be used to indicate Steel casings
the highest temperature to prevent tube overheating. The temperature of steel boiler casings may be ac-
These thermocouples are normally removed after curately measured with portable contact thermocouple
steam flow is adequate for cooling. On units designed instruments, because lateral heat flow from adjoining
for remote operation, it is possible to remotely retract metal areas quickly compensates for the small quan-

40-16 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

tity of heat drawn by the instrument at the contact The four steam quality measurement methods may
point. Thermocouple wires may be peened into or be summarized as follows:
fused onto the metal surface to form the hot junction.
1. The sodium tracer method, with continuous re-
Thermal contact between the lead wires and the sur-
cording of total dissolved solids in steam, is used
face should be maintained for several feet. The wire
for the highest accuracy.
installation should minimize any disturbance of heat
2. The conductivity method is useful within certain
transfer from the surface to the surrounding air. Ap-
limitations but is less accurate.
proximate surface temperatures may be conveniently
3. The calorimeter method is not suitable for measur-
measured with fusion paints or crayons.
ing extremely small quantities of carryover and
Thermometers are sometimes fastened to metal
at pressures exceeding 600 psi (4.14 MPa).
surfaces with putty. This method gives an approximate
4. Gravimetric analyses require large samples and
surface temperature indication if the metal is massive
do not detect carryover peaks.
and near ambient temperature. It is not recommended
for boiler casing temperature measurement and is
completely unsatisfactory for measuring insulation Obtaining the steam sample
surface temperature.
Through-steel components, such as metal ribs im- If the steam quality results are to be accurate, the
bedded in insulation and studs or door frames extend- instruments must be supplied with a representative
ing through insulation, cause considerable local up- steam sample. The method of obtaining a steam
sets in surface temperatures and their influence may sample and the operation of the boiler during testing
spread laterally along a metal casing. These effects are fundamentally the same for each testing method.
must be considered in planning and interpreting sur- The sampling nozzle design should be as recom-
face temperature measurement. mended by ASME Performance Test Code 19.11. It
Infrared cameras may be used to measure surface should be located after a run of straight pipe equal to
temperatures. These instruments can minimize the need at least ten diameters. Locations in the order of pref-
for complex thermocouple installations, especially if large erence are:
surface areas are to be measured. Improved equipment 1. vertical pipe, downward flow,
permits recording for further review. 2. vertical pipe, upward flow,
3. horizontal pipe, vertical insertion, and
Measurement of steam quality 4. horizontal pipe, horizontal insertion.
and purity The nozzle should be installed in the saturated steam
pipe in the plane of a preceding bend and in a position
The most common methods of field testing for steam that ensures the sampling ports directly face the steam
quality or purity are: flow. On boilers with multiple superheater supply tubes,
1. sodium tracer (flame photometry), sampling nozzles should be located in tubes spaced across
2. electrical conductivity (for dissolved solids), the width of the drum. These sampling points should be
3. throttling calorimeter (for direct determination of at no greater than 5 ft (1.52 m) intervals.
steam quality), and When a calorimeter is used the connection from the
4. gravimetric (for total solids). sampling nozzle to the calorimeter must be short and
well insulated to minimize radiation losses. Connections
Each of these methods is described in ASME Per- must be steam-tight so the insulation remains dry.
formance Test Code 19.11, Water and Steam in the When steam purity is tested by conductivity or so-
Power Cycle. The throttling calorimeter determines dium tracer methods, the tubing from the sampling
steam quality directly, whereas the other methods nozzles to the condenser should be steel (preferably
determine the steam solids content. stainless), with an inside diameter not exceeding 0.25
Most of the solids content of steam comes from the boiler in. (6.4 mm). It should be of minimal length to reduce
water, largely in the carryover of water droplets. (See also the storage capacity of the line. Multiple connections
Chapter 5.) It is customary to relate steam quality and can be run to a common line and then to the condenser.
solids content of steam by the following equation: However, these connections should be valved so that
each one can be sampled individually to isolate pos-
 solids in steam × 100  sible selective carryover. Cooling coils, or condensers,
x = 100 −   (6)
 solids in boiler water  should be located close to the sampling nozzles to mini-
mize settling of solids in the sample line.
where x equals % steam quality in percent by weight
and solids are expressed as equivalent parts per mil- Sodium tracer method
lion (ppm) (by weight) of steam or water. The sodium tracer technique permits measuring
By the use of Equation 6, steam quality can be de- dissolved solids impurities (carryover) in steam con-
termined if the solids content is known. This relation- densate to as low as 0.001 ppm.
ship is subject to error resulting from carryover of sol- Sodium is present in all boiler water where chemi-
ids in solution in the steam or in vaporized form. This cal treatment is in the form of solids. The ratio of the
effect occurs principally with silica at pressures above total dissolved solids in the steam condensate to the
2000 psi (13.79 MPa). total dissolved solids in the boiler water is proportional

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-17


The Babcock & Wilcox Company

to the ratio of the sodium in the steam condensate to adjusts the position of the prism. The selected wave-
the sodium in the boiler water. By determining the length is directed back to the plane mirror where it is
sodium content of the steam condensate and boiler reflected through the adjustable exit slit and lens. The
water and the total dissolved solids in the boiler wa- light impinges on the photomultiplier tube, causing a
ter, the solids content of the steam and the percent current gain which registers on the meter. The amount
moisture carryover may be calculated. The sodium of sodium in the sample is obtained by comparing the
content of steam condensate and water is usually de- emission from the water sample to emissions obtained
termined by using a flame photometer. The effect of from solutions of known sodium concentration.
any change in boiler operating conditions on carryover
is promptly indicated, facilitating problem analysis. Conductivity method
The operation of a flame photometer is illustrated
in Fig. 26. The condensed steam sample is aspirated Electric conductivity can be used to determine steam
through a small tube in the burner into the oxygen- purity in certain boilers. This method is applicable to
hydrogen flame. The flame, at 3000 to 3500F (1649 units operating with significant solids concentration
to 1927C), vaporizes the water and excites the sodium in the boiler water and with a total steam solids con-
atoms, which emit a characteristic yellow light hav- tent in excess of 0.5 ppm.
ing a definite wavelength. The intensity of the emitted The conductivity method is based on the fact that
yellow light is a measure of the sodium in the sample. dissolved solids, whether acids, bases or salts, are com-
The intensity of the light is measured with a spectropho- pletely ionized in dilute solution and, therefore, con-
tometer equipped with a photomultiplier attachment. duct electricity in proportion to the total solids dis-
The light from the flame is focused by the condens- solved. On the basis of solids normally present in boiler
ing mirror and is directed to the diagonal entrance water, the solids content in parts per million equals
mirror. The entrance mirror deflects the light through the electrical conductivity of the sample in micromho
the entrance slit and into the monochromator to the per mm times 0.055.
plane mirror. Light striking the plane mirror is reflected The condensed sample should ideally be free from
to the fiery prism where it is dispersed into its compo- dissolved gases, especially ammonia (NH3) and CO2,
nent wavelengths. The desired light wavelength is which contribute nothing to the solids content of the
obtained by rotating a wavelength selector which steam but have a significant effect on conductivity.

Fig. 26 Schematic arrangement of flame photometer.

40-18 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

Throttling calorimeter In obtaining calorimeter readings, the temperature


When steam expands adiabatically without doing of the expanded and superheated steam is measured
work, as through an orifice, the enthalpies of the high by a thermometer inserted in the thermometer well.
and low pressure steam are equal, provided there is no Due to radiation from the calorimeter installation,
net change in steam velocity. Such an expansion is thermometer corrections and orifice irregularities, the
termed throttling. As can be seen from a Mollier chart observed temperature is generally lower than the ac-
(see Chapter 2), wet steam with an enthalpy exceeding tual temperature. To determine a suitable correction,
1150 Btu/lb (2675 kJ/kg) becomes superheated when the as-installed temperature of the calorimeter should
throttled to atmospheric pressure. The temperature of be determined. This can be done by taking readings
the expanded steam determines the enthalpy, which may when the boiler is known to be delivering dry satu-
be used with the pressure of the wet steam to determine rated steam. This is the case when the boiler output
the percent moisture in the wet steam sample. is steady at about 20% of rated capacity with low
The foregoing principle is used in various forms of water concentration and steady water level. The nor-
throttling calorimeters. Each incorporates a small ori- mal correction is obtained by subtracting the as-in-
fice for expansion of the steam sample into an exhaust stalled temperature from the theoretical temperature,
chamber, where the temperature of the expanded read from the zero moisture curve of Fig. 28. The nor-
steam is measured at atmospheric pressure. Velocity mal correction should not exceed 5F (2.8C). If it does,
changes in properly designed calorimeters are negli- the calorimeter orifice is clogged, the insulation is
gible. The variations in the different designs are chiefly faulty, or some other test feature is incorrect. When
in the means of shielding the unit from external tem- the unit is used to determine the quality of a wet steam
perature influences. At pressures below 600 psi (4.14 sample, the percent moisture can be found from the
MPa), the calorimeter shown in Fig. 27 is used. This is curves of Fig. 28 using measured drum pressure and
a small, low cost, simple, accurate instrument which corrected calorimeter temperature. When properly
gives good results for steam having appreciable quanti- installed, insulated and operated, the calorimeter can
ties of moisture: up to 4.3% at 100 psi (0.69 MPa), 5.6% give accurate results for steam moisture contents as
at 200 psi (1.38 MPa) and 7.0% at 400 psi (2.76 MPa). low as 0.25% for low pressure boilers. For pressures
Fig. 27 shows the calorimeter installed and ready above 600 psi (4.14 MPa) and for greater accuracy of
for use. The connection should be short; the connec- steam purity measurement in the low ppm range,
tion and the calorimeter should be well insulated. It other methods of measurement should be used.
is essential that the calorimeter discharge be com- Gravimetric analysis
pletely unobstructed, so no backpressure can build in
the exhaust chamber. The orifices must be clear, full Gravimetric analysis is used to determine solids lev-
opening and of the correct diameter: 0.125 in. (3.175 els, particularly total solids, in a condensed steam
mm) for pressures from atmospheric to 450 psi (3.10 sample. It consists of evaporating a known quantity
MPa) and 0.0625 in. (1.588 mm) from 451 to 600 psi
(3.11 to 4.14 MPa). The thermometer should be im-
mersed in oil of suitably high flashpoint. The calorim-
eter is operated by fully opening the shutoff valve and
letting the steam discharge through the unit to the
atmosphere. The calorimeter must be thoroughly
warmed up and in temperature equilibrium before
normal temperature or test readings are taken.

Fig. 27 Throttling calorimeter showing sampling tube in steam pipe. Fig. 28 Percent moisture in steam versus calorimeter temperature.

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-19


The Babcock & Wilcox Company

of condensed steam to complete dryness and weigh- Y = compressibility factor, dimensionless, equals
ing the residual. This can be done on a batch basis or 1.0 for most liquids and for gases where the
by a continuous sample system. pressure drop across the device is less than
While a gravimetric analysis provides an accurate 20% of the initial pressure
measure of total solids in water, its main disadvantage A = cross-sectional area of throat, ft2
is that it requires a relatively large quantity of water g c = proportionality constant, 32.17 lbm ft/lbf s2
obtained over an extended period of time. It therefore P1 = upstream static pressure, lb/ft2
does not localize carryover peaks when they occur. P2 = downstream static pressure, lb/ft2
Time lag in other methods of steam purity measure- β = ratio of throat diameter to pipe diameter, di-
ment is usually small enough to permit attributing mensionless
peaks of carryover to periods of high water level, high ρ 1 = density at upstream temperature and pres-
boiler water solids, or other operating upsets. sure, lb/ft3
Details of primary element sizing, fabrication and
Flow measurement flow calculations can be obtained from the ASME pub-
lication Fluid Meters, Their Theory and Application.
Fluid flow includes water, steam, air and gas flow.
Table 2 is extracted from the ASME Performance Test
While there are many means of measuring flow, the
Code 19.5, Section 4, Flow Measurement, to show the
basic methods for the accuracy required by the ASME
advantages and disadvantages of the three primary
Performance Test Code use the orifice, flow nozzle (flow
element types. The throat tap nozzle has the highest
tube), or venturi tubes as primary elements. The pres-
accuracy of the primary flow measuring devices.
sure drop or differential pressure created by these
The primary elements listed should be fabricated
restrictions can be converted into a flow rate.
of erosion and corrosion resistant materials. The ori-
The flow of any fluid through an orifice, nozzle or
fice and flow nozzles are shown in Fig. 29; the ven-
venturi tube may be determined by the equation:
turi tube is shown in Fig. 30.
Even though these elements can be sized with con-
2 gc ρ1 ( P1 − P2 ) siderable accuracy by calculation, they should be labo-
 = Cq YA
m (7)
1 − β4 ratory calibrated prior to precision testing. Calibration
where is usually by weighed water tests using scales. For
commercial use, calculated flow rates or rates based
m = flow rate, lb/s on prototype testing or calibration are adequate.
Cq = coefficient of discharge, dimensionless, depen- Certain important factors of the primary element
dent on the device used, its dimensions and installation should be considered for accurate mea-
installation surement. These include:

Table 2
Advantages and Disadvantages of Various Types of Primary Elements
Advantages Disadvantages

Orifice
1. Lowest cost 1. High nonrecoverable head loss
2. Easily installed and/or replaced 2. Suspended matter may build up at the inlet side of
3. Well established coefficient of discharge horizontally installed pipe unless eccentric or
4. Will not wiredraw or wear in service during test period segmental types of orifices are used with the hole
5. Sharp edge will not foul with scale or other suspended flush with the bottom of the pipe
matter 3. Low capacity
4. Requires pipeline flanges, unless of special
construction
Flow Nozzle
1. Can be used where no pipeline flanges exist 1. Higher cost than orifice
2. Costs less than venturi tubes and capable of handling 2. Same head loss as orifice for same capacity
same capacities 3. Inlet pressure connections and throat taps when used
must be made very carefully
Venturi Tube
1. Lowest head loss 1. Highest cost
2. Has integral pressure connections 2. Greatest weight and largest size for a given size line
3. Requires shortest length of straight pipe on inlet side
4. Will not obstruct flow of suspended matter
5. Can be used where no pipeline flanges exist
6. Coefficient of discharge well established

40-20 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

Pipe Wall Tap Flow Nozzle

Fig. 29 Orifice and flow nozzles.

1. location in the piping in relation to bends or For example, the draft loss across the gas side of an
changes in cross-section, air heater may be used as an index of the flue gas flow.
2. possible need for approach straightening vanes, Because pressure differentials vary with a change in
3. location and type of pressure taps, surface cleanliness, these are not highly reliable meth-
4. dimensions and condition of surface or piping be- ods. The use of orifices, flow nozzles, venturi air foils,
fore and after the element, or impact suction tubes in areas free of entrained dust
5. position of the element relative to direction of fluid and at relatively stable temperatures prove more sat-
flow, and isfactory. Fig. 31 illustrates two types of venturi sec-
6. type and arrangement of piping from primary ele- tions used in air ducts for measuring combustion air
ment to differential pressure measuring instrument. flow. Fig. 32 illustrates an impact suction pitot tube
Details of these requirements for precision testing are arrangement for metering primary air flow to a pul-
also covered in the ASME publication Fluid Meters, Their verizer. For a dependable determination of air flow,
Theory and Application. These devices produce a differ- the primary element should be calibrated at normal
ential pressure across the element and a differential pres- operating pressure and temperature. Test connections,
sure gauge should be used for the flow measurement. or sampling points, for this purpose are located at a
Flow measurement of combustion air or flue gas zone in the duct or pipe where good flow characteris-
generally does not require a high degree of precision. tics are obtainable, and the calibrating flow measure-
Orifices, flow nozzles, or venturis are also used, but ments are made using a hand held pitot tube or
the rigid requirements of construction and location are equivalent probe. The pitot tube, when inserted fac-
usually not possible because of space limitations. As a ing the air or gas flow stream, measures velocity pres-
substitute, a component of the steam generator, located sure, that is, the difference between total and static pres-
in the flow path, can create a suitable pressure dif- sure. The velocity pressure measurement can be con-
ferential, which is used to obtain a flow measurement. verted into a velocity reading for English units only by:

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-21


The Babcock & Wilcox Company

Fig. 30 Dimensional proportions of classical (Herschel) venturi tubes with a rough-cast, convergent inlet cone. (Source: ASME, Fluid Meters,
Their Theory and Application, Sixth Edition, 1971.)

hV probe. It also has a coefficient of 1.0 and incorporates


V = 1097 (8) a null balance feature that permits determining when
ρ the probe faces the direction of gas flow. This probe is
where illustrated in Fig. 35. In this schematic, the outer holes
in the probe are each at exactly 39.25 deg from the
V = velocity, ft/min centerline hole, placing them at a point of zero impact
h V = velocity head as indicated by pitot tube differ- pressure and providing a reading of true static pres-
ential, in. wg
ρ = density at the temperature of the sampling lo-
cation, lb/ft3
A typical pitot tube and measuring manometer are
illustrated in Fig. 33. The pitot tube has a coefficient
of 1.0, which is applied to velocity head readings,
eliminating the need for corrections. Fig. 34 shows a
method of traversing a round duct using a pitot tube.
An additional portable flow measuring device that
can be used for calibration purposes is the Fechheimer

Fig. 31 Venturi sections for air ducts. Fig. 32 Averaging pitot tubes.

40-22 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

Fig. 33 Arrangement of pitot tube and manometer.

sure. When the probe faces the gas flow, a manom-


eter connected across the two outer holes shows a zero,
or null, balance. Because the centerline hole receives
full impact or total pressure when facing into the gas
stream, a manometer connected across one outer hole
and the centerline hole gives true velocity pressure,
that is, impact pressure less static pressure.
The impact suction or reversed type pitot tube, il-
lustrated in Fig. 36, produces a differential pressure
greater than one velocity head, which gives a magni-
fied reading. This type of pitot tube measures flow in
either direction. To use the reversed type pitot tube
for flow measurement, calibration is required for the Fig. 34 Pitot tube traverse.
Reynolds number range as well as for the geometry
of the flow channel and its orientation in the stream.
Permanently installed pitot tubes are used where
relatively high pressure losses due to the primary
measuring element are undesirable. Flow tubes are
also used in these cases. These tubes differ from most
other primary elements in that the flow channel cross-
section varies symmetrically in the inlet and exit di-
rections from the contoured throat section.
The throat section is equipped with two sets of pres-
sure taps, one set pointing upstream and one down-
stream, similar to impact suction pitot tubes. Each set
of taps is interconnected and the two sets in turn are
connected to a differential pressure gauge. The pres-
sure taps are located at cross-sections of equal area so
that the differential developed is a result of impact and
suction pressure differences only and is, therefore, a
function of the velocity head alone.
Many other types of flow measuring devices exist.
The pitot tube grid incorporates an array of pitot tubes
in a flow path. The primary signal is a pressure dif-
ferential. The array of pitot tubes measures a num-
ber of flows and minimizes measurement errors caused
by flow unbalance and upstream conditions. The hot
wire anemometer measures flow using a probe with a
small wire at the tip. Flow over the wire removes heat,
and flow may be measured when the hot wire is in
equilibrium with electrical power input and heat lost.
Doppler phenomena may be applied to flow measure- Fig. 35 Fechheimer probe.

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-23


The Babcock & Wilcox Company

rapid updates to operating information and provide a


check on measurement errors. Measurements of tem-
perature, pressure and flow that might be in error be-
cause of instrument calibration are identified rapidly.
Process control
A computer based control system receives its basic
information from a data acquisition system which accu-
mulates information from the plant instrumentation.
Process control is achieved according to specific math-
ematical algorithms. The control system, updated by
current data or operator interface, produces output sig-
nals to modulate components based on these algorithms.
Data acquisition system
A typical data acquisition system consists of com-
puter hardware and related software. The basic hard-
ware components are the computer, a signal multi-
plexer and the measuring instrument. The computer
program (software) controls the hardware operation
and data recording. In operation, the program causes
the signal multiplexer to read a specific instrument
analog output signal. The signal is converted from an
analog to a digital signal, which is conveyed to the
computer and stored in memory. The first processing
of the data is the storage of the digitized signal in a
Fig. 36 Combined-type pitot tube with reversed static tube. (Source: data array. Data acquisition, incorporating many in-
ASME, Fluid Meters, Their Theory and Application, Sixth Edition, 1971.) struments, other computer outputs and manual in-
puts, requires that each of the signals be assigned a
ment with laser or acoustic transmissions. The trans- specific location in the array. The stored signal may
mission signal and type of fluid make these methods be converted to engineering units and stored in an-
very application specific. other array. Furthermore, the data may be averaged
over time and then stored. Performance calculations
are generally performed using data converted to en-
Computer application to measurement gineering units and averaged over time. Portable data
The application of computer technology to instru- acquisition systems are generally used for specific test
mentation provides the capability to further under- purposes and may use high precision transmitters and
stand the operating characteristics of many interre- thermocouples.
lated power plant components and processes at a
much faster rate than was previously possible. Mea- Signal multiplexer
surements have progressed from readings recorded at The signal multiplexer in a data acquisition system
local gauges, strip charts, or manually recorded test provides an interface between the transmitted instru-
gauges to significant quantities of information ac- ment signal and the computer. The typical multiplexer
quired in seconds. provides signal conditioning, analog to digital signal
The instrument readings are not made directly at conversion and switching capability for multiple inputs.
the instrument, but rather transmitted signals from It produces a digital output signal for computer process-
the instrument are entered into a data acquisition ing and assignment to a position in the data array. Sig-
system and the reading is indicated at a computer. The nificant engineering design is required to accommodate
computer gathers, stores, interprets and displays the the variety of signals entering the multiplexer.
information. The display appears as an extension of
the instrument. The information gathered may be Data acquisition computer program
further manipulated for process control or equipment The data acquisition software may be program ori-
performance computations. ented or menu oriented but should readily permit
changes to the quantity of signals, the rate at which
Performance computation the computer reads each signal and the computations
Computer computation permits calculations in near performed with each signal. Data storage and re-
real time for boiler efficiency and other plant perfor- trieval are but two of many design considerations for
mance indicators. These performance indicators provide computer program configurations.

40-24 Steam 41 / Pressure, Temperature, Quality and Flow Measurement


The Babcock & Wilcox Company

References
1. Tower, J.R., “Staring PtSi IR cameras: More diversity, 3. Weiss, J., and Esselman, W., Electric Power Research
more applications,” Photonics Spectra, (USPS 448870), Institute and R. Lee Foster-Miller Inc., “Assess fiberoptic
Laurin Publishing Company, Pittsfield, Massachusetts, sensors for key power plant measurements,” Power, p. 55,
February, 1991. October, 1990.
2. Kaplan, H., “What’s new in IR thermal imagers,”
Photonics Spectra, (USPS 448870), Laurin Publishing
Company, Pittsfield, Massachusetts, February, 1991.

Bibliography
1986 Digital Systems Techniques, Instruments and Ap- The OMEGA Temperature Handbook and Encyclopedia,
paratus, Supplement to ASME Performance Test Codes, Vol. MMV, Fifth Ed., Omega Engineering, Inc., 2004.
ANSI/ASME PTC 19.22, American National Standards Temperature Measurement Designer’s Guide, Thermo
Institute/American Society of Mechanical Engineers, New Electric Company Inc., Saddle Brook, New Jersey, 1986.
York, New York, 1986. Also available online at www.thermo-electric-direct.com.
Manual on the Use of Thermocouples in Temperature
Measurement, ASTM Special Technical Publication 470B,
American Society for Testing and Materials, Philadelphia,
Pennsylvania, 1981.

Steam 41 / Pressure, Temperature, Quality and Flow Measurement 40-25


The Babcock & Wilcox Company

Virtually all areas of a power plant are continually monitored to ensure efficient and reliable daily operation.

40-26 Steam 41 / Pressure, Temperature, Quality and Flow Measurement

You might also like