Professional Documents
Culture Documents
MAINTENANCE MANUAL
50000-12131
of 1 .
.
99011
`. / Copyright© 1993
© Bear Medical Systems, Inc. Serial #
12/97
WARNING
means there is the possibility of personal injury or death to yourself and others.
CAUTION
NOTE
indicates points of particular interest for more efficient and convenient operation.
It is recommended that the reader take particular notice of the warnings, cautions,
and notes provided throughout this manual.
PREFACE
PROPRIETARY INFORMATION interrelate to the base maintenance manual.
25
ss
the test equipment and from the test equipment to extended to the original buyer purchasing the
the device being tested must be capable of equipment directly from Bear Medical Systems, Inc. or
withstanding a minimum supply pressure of 100 through its authorized dealers. All warranty periods,
psi (7.03 kg/cm’), The use of tubing and fittings where applicable, commence on the date of first
not capable of withstanding this pressure could purchase but not more than six months after shipment
cause the tubing to rupture, resulting in personal from Bear Medical Systems, Inc. Bear Medical Systems’
injury or property damage. sole obligation and liability under this warranty is
When verifying the operation of this medical limited to (at Bear Medical Systems option) the repair or
device, do not breathe directly from the machine. replacement by Bear Medical Systems authorized
Always use a fresh bacterial filter and test circuit. personnel of any parts or assemblies which, upon test
A hazard to the health of the service person may and examination by Bear Medical Systems prove to be
result. defective. This equipment may be returned prepaid to
Bear Medical Systems after prior notification has been
If any of the following procedures cannot be given and approval obtained for the return.
verified as outlined in this document, do not use
this device on a patient and refer it to Bear This warranty does not cover normal maintenance such
Medical Systems or a Bear Medical Systems as cleaning, adjustment, or lubrication and updating of
Authorized Service Facility or a Bear Medical equipment or parts thereof. This warranty shall be void
Systems Trained Hospital Service Technician. and not apply if the equipment, including any of its
parts, is modified without Bear Medical Systems
WARNING authorization; is attempted to be repaired by personnel
not authorized by Bear Medical Systems,; is not
Do not release the ventilator for use if it does not maintained in accordance with the prescribed schedule;
pass all of the verification procedures specified in is used with accessories or parts not authorized; or is
the checklist. To do so could result in personal damaged due to misuse, mishandling, abuse,
injury including death or property damage. Refer negligence, accident, fire, or inadequate packaging by
the ventilator to a Bear Medical Service Technician owner for shipment. Bear Medical Systems makes no
or a Bear Medical Systems trained service guarantee of clinical results.
technician for appropriate repair and/or
calibration. THE WARRANTY STATED ABOVE (INCLUDING ITS
LIMITATIONS) IS THE ONLY WARRANTY MADE BY
WARNING BEAR MEDICAL SYSTEMS AND IS IN LIEU OF ALL
When replacing air 02 inlet filters, mark and
OTHER WARRANTIES, WHETHER EXPRESSED OR
IMPLIED, INCLUDING ANY WARRANTY OF
remove one and only one fitting at a time.
MERCHANTABILITY OR FITNESS FOR A
Replace the filter and return the fitting to its
PARTICULAR PURPOSE. BEAR MEDICAL SYSTEMS
original location before removing the second SHALL NOT BE LIABLE FOR CONSEQUENTIAL OR
fitting. Failure to follow this procedure may INCIDENTAL DAMAGES OF ANY KIND.
result in injury, including death, to the
patient. Warranty period is one year with the following
exceptions:
1. Preventive maintenance (90 day) warranty on all
parts and labor included in all procedures.
Specifications ri rnrees 1
Theory of Operations <<< 2
Operational Verification Procedures .............. 3
Updates Rs 4
Calibration Rs 5
Schematics ra 6
Troubleshoofind ea 7
Removal and Replacement …… せ ee 8
Preventative Maintenance,… scenerne 9
Parts List eee eee een nenene 10
Repacking /Shipping Instructions.................. 1]
Section]
BEAR® 1000 Ventilator
Specifications
When the breath interval elapses (as deter- Tolerance: 0-60 bpm-+1bpm
mined by the preset breath rate), a mandato- 61-100 bpm - + 2% of setting
breath window will be opened. A volume 101-120 bpm - + 3% of setting
breath will be delivered when the patient
activates the assist trigger while the window Sigh Breath Rate (N/A FOR BASIC UNIT)
is open. As soon as the volume breath is trig-
gered, the window will be closed and the Range: Every 100 breaths
patient may breathe spontaneous (or pres-
sure supported) breaths until the next breath Sigh Interval Assist CMV: 2 x Normal Breath
interval elapses when the cycle starts over. Interval
Note that the breath interval is reset as soon SIMV /CPAP: Normal Breath
as it elapses, even though a volume breath Interval
has not necessarily been delivered.
Limitation: Not active in pressure control
If a breath interval elapses and the mandato- mode.
breath window is still open, then a volume
breath will be delivered. The window will Assist Sensitivity (Assist Trigger)
remain open until the patient triggers the
assist trigger. Range: 0.2 to 5.0 cmH20
Manual Breath
End Expiratory Hold (N/A FOR BASIC
The manual breath control delivers a single UNIT)
volume breath (in Assist CMV ος Shall not allow the patient to inspire, or
SIMV/CPAP) or a pressure controlled breath exhale any gas. End expiratory hold shail be
(in Press Control mode) if the ventilator is initiated upon the detection of the next
not in the inspiratory cycle of a breath. breath.
3
50000-12131 » NOVEMBER 1993
SPECIFICATIONS
г ‘ Pressure Augmentation (N/A FOR BASIC Deliver 100% oxygen for 3.0 + 0.1 minutes
UNIT) regardless percent oxygen setting.
Mean Airway Pressure (N/A FOR BASIC Range: 0 to 100 psig, or 0 to 160 psig
UNIT)
Tolerance: +5 psig
Monitor the average pressure in the patient
circuit for the last 30 second interval. Resolution: 5 psig or better
Range: 0 to 140 cmH20 Oxygen Inlet Pressure
Continuously monitor the pressure in the A non-resettable meter used to measure and
patient circuit. display the total amount of time that the ven-
tilator has been on. This shall be independent
Range: -10 to 120 cmH20 of whether or not gas is being delivered.
Tolerance: +3 cmH2O or 3%, whichever is Range: 0 to 99,999 hours
greater.
Tolerance: + 2% of reading
Resolution: 1 cmH2O
Resolution: 0.1 hour
Air Inlet Pressure
b) LE Override active (N/A FOR Limitation: Not active for Sigh Breaths
BASIC UNIT
B. Sigh High Peak Pressure Alarm
LE Ratio Interval: (1/f) * .80 (4:1)
Alarm if the pressure in the patient circuit
f = Cycles/Second exceeds the sigh high peak pressure alarm
during a sigh breath cycle.
Limitation: Not active for demand breaths
(spontaneous and pressure sup- Range: 1.5 * (Normal High Peak
ported). Pressure) to a maximum of 120
cmH20.
Pressure Alarms
Tolerance: + 1 cmH?20 of the monitored
Low Inspiratory Pressure peak pressure.
Alarm if the peak pressure during an inspira- Limitation : Only active for Sigh Breaths.
tion is less than the low inspiratory pressure
limit. Low Baseline Pressure (N/A FOR BASIC
UNIT)
Range: 3 to 99 cmH20
Alarm if the baseline pressure (PEEP) is less
Tolerance: + 1cmH20O of monitored peak than the Low Baseline Pressure Limit.
pressure.
Range : 0 to 50 cmH20
Resolution: 1 cmH2O
Tolerance: + 1cmH20 or 5% of the moni-
Limitation: _ Not active for spontaneous tored PEEP, whichever is
breaths and pressure support- greater
ed/ pressure control brea’
when PEEP + Insp. Press <3 Resolution: 1cmH20
cmH20.
Limitation: The alarm is off if it is set to
High Peak Pressure Limit Zero.
If the high peak pressure limit is violated, the High Baseline Pressure (N/A FOR BASIC
inspiration shall be terminated and circuit UNIT)
pressure shall return to PEEP + 5 + 1.5
cmH20 before the next breath may be deliv- Alarm if the baseline pressure (PEEP) is
ered. greater than the High Baseline Pressure
imit.
A. Normal High Peak Pressure Alarm
Range : 0 to 55 cmH2O
Alarm if the pressure in the patient circuit
exceeds the normal high peak pressure alarm Tolerance: + 1 cmH20 or 5% of the moni-
except during sigh breath cycles. tored PEEP, whichever is
greater.
Range: 0 to 120 cmH20
If power is removed from the unit, an audible Range: -300 to 200 lpm
indicator shall occur for a minimum of 5 min-
utes. A “Failed to Cycle” condition shall also Scale: 1 jpm/10 mV
occur . The audible indicator shall continue
to occur when power is restored until the Tolerance: + 10% of reading or + 30 mV,
alarm silence is activated. whichever is greater.
Run Diagnostics Zero Offset: 3 Vdc at 0lpm
When the ventilator detects a fault that will С) Inspiratory /Expiratory Volume
allow a certain degree of ventilation to be
maintained, the ventilator shall continue to Range: 0.00 to 3.00 1
ventilate the patient and the “Run
Diagnostics” alarm will be activated to notify Scale: 0.01 1/25 mV
the clinician so that he may rectify the condi-
tion. This will occur only if it is more effec- Tolerance: + 10% of reading or + 75 mV,
tive to continue ventilation than to assert the whichever is greater.
“Failed to Cycle” condition.
Zero Offset: 0 Vdc at OL.
Cross Contamination of Supply
Gases RS-232 Port
The ventilator shall not allow the external The ventilator shall contain an RS-232 port
10
50000-12131 « NOVEMBER 1993
SPECIFICATIONS
b) Ambient Operatin:
10 to 40° C (50 to 104° F)
12
50000-12131 + NOVEMBER 1993
SPECIFICATIONS
Sterilization
Methods of Sterilization
Liquid Sterilization
a) Cidex
b) Sonacide
13
50000-12131 « NOVEMBER 1993
SPECIFICATIONS
ALARMS /INDICATORS
ELEC. INTERFACES
15
50000-12131 + NOVEMBER 1993
一
Section 2
BEAR® 1000 Ventilator
Theory of Operation
ÖVETMVİSW eren
Ventilator PREUMATICS...... 이 니 디 디 디 이 디 이 이 미 미 리 이 이니
Са МЕТ... r εκ κανει
Blender Rs
ACCUMUICTON erer
Flow Control Valve ns
Pneumatic Diagram << ex
SOPR VAIVE. rr
Low Pressure Control System.........................
Exhalation Valve 4 eine ios
Nebulizer System. ee
Ventilator Electronics <<< ssiwzee
EİECİTONİCS......... er
Flow Delivery Logic... eee
Volilume LOgiC eee
Pressure LOgİcC............e res
о T
O_o TS ONE
A © ©
OA 4
© ἕνα
1 ELECTRO/PNEUMATIC
1 “INTERFACE (EPİ)
nomen
Boo
Е FLOW
x (DER
BOARD
|
FAN
1
sole FLOW
sour TRANSDUCER
am
sx cross
sanser
EXHALATION
VALVE
PROXIMAL
PRESSURE
PORT
AR
SUPPLY μα...
PATIENT
SUSaBIENT
INLET
IPNEUMATGIS|
OXYGEN
BLENDER ACCUMULATOR:
FLOW
CONTROL el OUTLET
VALVE Rei E F
OO
VA e
OXYGEN
SUPPLY πι | NEBULIZER
NEBULIZER
PORT
} SYSTE M
上 EJ
OVERBOARD
EXHAUST
Figure 2-1
1993
I BAWA 100 HADNOSNYEL HIDNASNYUL
imino | F vozu
MOLVINDAH IUNSSIHA
N3OAXO
anss3dd
uly
ANALY MUZE | KHA 932zmne3N
内 зо
50000-12131 #“NOVEMBER
ARNO ZAMA) BAVA
[T
yoaHg| 9HS
м ano
AAIVA 4103105
Figure 2-2
| Agos) 3318 B ヨ N318
A Hana JAVA 4338
VIBa12V8 ABNSSAHEZAO | нзопавмуни эзамала
' HiNalaWYBnS | BUNSSStd INIHOVIN a Эва в,
ETNOHO) |
oh Ч
NADAXO
ldoly1neaB
Hiv
ορ]
ati
ао ||
dV1 3HTiSS
ヨ Hd XOHdがAA uva y
HALTS NOLLO3NNOO
rhe LIGDHIO LNALLY
<
vidgiove sunssaug |
1N3HYd ART
TE TXQHd =
T 900 [ouiNos
| on HaonGSNYHL
| dava 0553ba
BAIA MOTI
dv GION310S
LVSNa0N0S NOIIVTYHX ヨ ное
ON y BUNSSİYd MOT
ΛΗΛ. ino
-©
NOILYTVHX ヨ ) Ê E Pp is! ON
ZATA
QiON ヨ1QS
| TS dJO-LNHS
get Maz oun TRE)
авопозмеы | | по Ni dnd ar JAWA
Е 31033N
1041N00 dů
| | i 933d
золу AUNSSSIHA FE —® Her -©
ΑΝΜΗΙΝ Ἰνβιχοῦίά
| H3ONGSNYHL HAONGSNVE
38085349 XOHd aunssaud
| 74393349
ΒΊΝΝΘΙ6ΊΟΜΙΝΟΟ. * STVNDIS HZONASNVYL
|
THEORY OF OPERATION
Whatever the output flow established by the the 140 cmH20 normally supplied to the SOPR |
flow delivery logic, the temperature sensor valve diaphragm falls to zero, and the SOPR
built into the FCV enables the ventilator to relief port opens. When the relief port opens,
deliver accurate volumes based upon a stan- the patient can breathe room air.
dard temperature correction, 77°F, zero satura~
tion and the set barometric pressure. Beside providing the patient with access to
room air, the SOPR has a second function. It
The FCV delivers peak flows in the range of: acts as an independent relief valve to mechani-
cally limit the maximum pressure the venti-
Volume-controlled Breaths:10 to 150 LPM
lator can deliver. This mechanical limit func-
Demand or Pressure Breaths:0 to over 200LPM tions only when ventilator outlet pressure
exceeds 175 cmH20. Such a high pressure on
SOPR VALVE the front side of the SOPR relief port causes the
port to open and thereby relieve excess gas
From the FCV, the gas flows through the
Subambient Overpressure Relief Valve (SOPR)
through the outlet labeled “Overboard
Exhaust.”
and out of the ventilator through the port
labeled “Main Flow to Patient.” LOW PRESSURE CONTROL SYSTEM
During normal operation, the Low Pressure The Low Pressure Control (LPC) System serves
Regulator supplies 140 cmH2O to the SOPR three functions:
valve diaphragm, thereby closing its relief
port. Because this port is closed, gas from the + It provides 140 cmH20 to the SOPR valve
FCV passes directly through the SOPR and out diaphragm, and i
to the patient. ¢ It controls the Exhalation Valve Diaphragm
to determine breath phase (inspiration vs.
When the Fail State System activates (due to a exhalation) and PEEP level,
FAILED TO CYCLE alarm or a power failure),
¢ It supplies gas to the Proximal Purge.
PEEP
CONTROL
NEEDLE
VALVE JET PUMP
0-55 CM HO
SHUT-OFF
CEI
SOLENOID Cr EXH
VI [o o
INC) ps
< 5 3 EXHALATION
TE mi Im VALVE
一 一= ra LOW PRESSURE IN No
REGULATOR
2PSIG EXHALATION
SOLENOID CONDENSATE
VALVE
} The LPC System consists of the following matic drive to the Low Pressure Regulator
components: which in turn, cuts off gas pressure to the
© Shut-off Solenoid Valve back side of the SOPR Valve, thereby causing
* Low Pressure Regulator its relief port to open, and allowing the
* PEEP Needle Valve patient to inhale room air.
* PEEP Jet Pump 2.Exhalation Valve and PEEP Control: The
¢ Exhalation Solenoid Valve closing of the Shut-off Solenoid Valve
deprives the PEEP Needle Valve of drive
After entering the ventilator and passing pressure. The PEEP Jet Pump consequently
through the regulator, air is tapped off before exhausts all pressure applied to the
entering the Blender. This sidestream of 18 psig Exhalation Valve. Because the Exhalation
gas feeds directly into the Shut-off Solenoid Valve is also de-energized by the Fail State
Valve. System, it references this zero PEEP pressure
During normal operation, the Shut-off being supplied by the PEEP Needle Valve
Solenoid Valve is energized, passing 18 psig and Jet Pump. As a consequence the
gas from the Gas Inlet System to both the Low Exhalation Diaphragm opens completely,
Pressure Regulator and the PEEP Needle allowing the patient to exhale, unimpeded.
Valve. Under this circumstance, the LPC
System functions as follows: EXHALATION VALVE
1.SOPR Valve: The 140 cmH20 from the Low The Exhalation Valve regulates all flow out of
Pressure Regulator enters the SOPR valve the patient circuit. Exhaled patient gases flow
diaphragm, thereby holding its relief port through the expiratory leg of the patient cir-
closed. cuit, past the Exhalation Valve Diaphragm and
2.Exhalation Valve and PEEP Control: The Seat, and through the External Flow Sensor.
Exhalation Solenoid Valve switches the refer- The Exhalation Valve is designed to minimize
ence pressure for the Exhalation Valve pressure fluctuations as well as audible noise.
Diaphragm between:
The pressure required to:
e The 140 cmH20 supplied by the Low * Close the Exhalation Valve during a positive
Pressure Regulator, and pressure inspiration and
+ The output pressure from the combina- ¢ Control the baseline (PEEP) pressure
tion of the PEEP Needle Valve and the
PEEP Jet Pump. The Exhalation Valve also acts as a check valve
so the patient cannot rebreathe gas through the
This allows the Exhalation Valve to close dur- expiratory leg of the circuit.
ing inspiration and maintain PEEP during
exhalation. A condensate jar attached to the Exhalation
Valve Manifold collects condensed water from
By contrast, during Fail State Operation (a the circuit and manifold.
FAILED TO CYCLE alarm or a power failure),
the LPC System functions as follows: EXTERNAL FLOW SENSOR
} 1.SOPR Valve: The Shut-off Solenoid Valve de- The External Flow Sensor, connected to the
energizes and closes. This cuts off the pneu- Exhalation Valve Manifold, measures exhaled
5
§0000-12131 * NOVEMBER 1993
THEORY OF OPERATION
tidal volumes. The flow sensor operates on the Flow output port) and volume are maintained
principle of hot wire anemometry. at the clinician-set levels.
The hot wire flow sensing system is a constant The oxygen concentration of the nebulizing
temperature device. The unidirectional opera- gas is the same as that on the 02% control.
tion is achieved with the two platinum wires
as follows:
+ À reference wire monitors the ambient gas VENTILATOR ELECTRONICS
temperature and communicates this temper- The five major components the Monitor PCB,
ature to the sensor electronics.
the Control PCB, the Electro-Pneumatic
A hot wire is held at a constant high temper-
Interface (EPI) PCB, the Flow Sensor PCB, and
ature level based upon the temperature feed-
back from the reference wire. the Power Supply
ELECTRONICS
The rate of heat loss from the hot wire varies Monitor PCB & Control PCB
directly with the velocity of the exhaled gases
as they pass through the flow sensor. As the The user operates the ventilator through the
gas flow rate increases, the electronic circuitry front panel controls and displays.
applies additional electrical current to the wire
to maintain a preset temperature. The resulting The Monitor and Control PCBs constantly scan
voltage change can then be translated into a the front panel for current settings or for clini-
gas flow rate. This measured flow rate is then cian-generated changes. If an appropriate mem;
integrated over time by the ventilator’s micro- brane key is selected, the Monitor and Control
processor to yield the measures of exhaled vol- PCBs couple this parameter to the control knob.
umes which are displayed on the front panel An optical encoder senses knob rotation and
monitors. increases or decreases the currently displayed
value accordingly.
NEBULIZER SYSTEM
The Monitor and Control PCBs receive patient
When the NEBULIZER key on the ventilator
and machine status information from the
control panel is activated, the Nebulizer
Electro-Pneumatic Interface (EPI) PCB. Based
Regulator and Nebulizer Solenoid Valve togeth-
er allow nebulization during the inspiration on clinician-selected control settings and EPI
phase of any breath type so long as flow feedback, algorithms within the Control PCB
exceeds approximately 20 LPM. Thus nebuliz- calculate the required delivered gas flow, O2
ing is synchronized with patient breathing. concentration, and Exhalation Valve state. The
Control PCB transmits this information to the
The Nebulizer Regulator reduces the 18 psig EPI PCB, which executes changes to the appro-
pressure from the Accumulator to 10 psig. priate electro-mechanical devices.
While the solenoid valve is open, 6 LPM is Another major function of the Monitor and
assumed to flow through the nebulizer (the Control PCBs is to monitor and display the
actual amount depends upon the specific ventilator’s clinical data. When monitored
nebulizer being used.) This 6 LPM is subtract-
parameters deviate beyond clinician-set alarm ,
ed from the flow delivered by the FCV. In this
limits, the Monitor and Control PCBs create,
manner the overall flow (nebulizer plus Main
audible and visual alarms. |
6
50000-12131 + NOVEMBER 1993
THEORY OF OPERATION
Input
Input TIDAL VOLUME
INSP PRES/PRES SUP PEAK FLOW
PRESSURE AUGMENT COMPLIANCE COMP
PRESSURE SLOPE WAVEFORM
PEEP Feedback: Actual Volume
Feedback: Proximal Pressure delivered.
PRESSURE VOLUME
LOGIC LOGIC
FLOW
signal
FCV
When a patient passively receives a volume- The flow delivery logic for a pressure-con-
controlled breath, the Volume Logic uses the trolled breath is similar to that of a pressure-
preset TIDAL VOLUME, PEAK FLOW, and supported breath except that:
WAVEFORM to calculate a flow signal that is
sent to the “Comparison” Logic. e The breath trigger can be machine-con-
trolled, as well as patient controlled, and
Because this is a passively-received breath, the * The end of inspiration is determined by the
Pressure Logic sends a zero flow signal to the clinician-set INSPIRATORY TIME control
“Comparison” Logic. The “Comparison” Logic rather than by a drop in flow.
then compares these two signals and sends the
larger one (which in this case is the Volume Pressure Slope
flow signal) to the FCV.
The Volume Logic uses feedback information The PRESSURE SLOPE control affects the rate
from the FCV regarding actual delivered vol- at which the flow signal increases to its peak
ume to determine when to switch to sending a value. In essence it operates through the
“zero flow” signal to the “Comparison” Logic. Pressure Logic to determine how fast the FCV
This occurs when the TIDAL VOLUME and opens. Thus it affects how slowly or rapidly
__) the optional COMPLIANCE COMP volume the preset inspiratory pressure level is
have been delivered. achieved at the airway.
Flow and Volume Augmentation This increase in flow is termed “flow augmen:
tation.”
The goals of flow and volume augmentation
are to: If the patient continues to demand flow even
after the clinician-set TIDAL VOLUME has
1. Ensure that the patient is never starved of
been delivered, the Volume Logic will start
flow, and
sending a. “zero flow” signal to the
2. Reduce stacking of volume breaths (auto- “Comparison” Logic. However, the Pressure
cycling). Logic will continue to request flow to meet
patient demand and to maintain pressure at
To achieve these goals, delivered flows must baseline. The “Comparison” Logic, as usual,
meet or exceed patient demand. To begin with, chooses the larger of these two flow signals
the Volume Logic begins to deliver a volume- (that from the Pressure Logic) and sends the
controlled breath according to the clinician- flow signal to the FCV. When patient demand
selected PEAK FLOW, WAVEFORM, and for flow drops to approximately 30% of its
TIDAL VOLUME settings. Then, if the patient peak value, the Pressure Logic switches to a
demands more flow than that delivered by the “zero flow” signal, and the FCV stops deliver-
Volume Logic, proximal airway pressure ing flow. The patient receives a larger tidal vol-
drops.When pressure drops below baseline, ume than set on the controls because the
the Pressure Logic sends a signal to increase patient has demanded a larger volume. The
flow beyond the flow signal from the Volume increase in tidal volume is termed “volume
Logic to maintain airway pressure at baseline. augmentation.” i
Clinician
Ipereases Intermittent
Autocycie ssist Trigger Leak
No Leak у To Avoid Disappears
Makeup Autocycle
Makeup
BaseineL Z mm) Work of Breathing ¢
ASSIST Asynchrony +
SENSITIVITY 4
Pt. Trigger
10
50000-12131 « NOVEMBER 1993
THEORY OF OPERATION
11
50000-12131 + NOVEMBER 1993
Section 3
BEAR® 1000 Ventilator
Operational Verification Procedure
Operational Verification Procedure................ 1
Test Equipment ss 1
Exterior Inspection and Cleaning... 2
Cleaning ................. ee 2
Filters iene aeree 2
Plugs and Cables <<< 2
Automated Built-in Diagnostics..................... 3
Test SETUP nr E LEE À
Entering Operator Diagnostics... 4
Display Panel Verification..................... 4
Manual Testing een 5
Air/O2 Inlet Check Valve Verification.............. 7
Air/O2 Inlet Pressure Gauge Verification ........ 7
Alarm Loüdnesslesting.......................... 7
Breath Rate VErificOtion .... iii 8
Oxygen Blending Verification... 8
… Assist Sensitivity Verification........................». 9
Delivered/Exhaled Volume Verification........ 10
Power Loss Test... o 10
APPENCIX A... rernre 1]
Checklist.............0. ee 13
Active Functions and Displays for
Operator DIagnosfics „ce 14
OPERATIONAL VERIFICATION
PROCEDURE Tube, Circle 72” x 3/4 (2 ea.) .....51000-02759*
Patient Wye -
(Operator Diagnostics)
Chamber, VH-820
This procedure is to assist a Qualified Operator, Filter, Proximal Line.
Hospital Service Technician, or Bear Medical Systems Filter, Mainflow
Service Technician ensure that the BEAR® 1000 Tubing, PVC, 1/4” ID, 72”
Ventilator is in proper operating condition. A checklist Plug for Patient Wye
is included and should be completed during each opera- Plug for Humidifier Water Port
Oxygen Analyzer, 1% Accuracy
tional verification. A copy of the checklist should be
Pressure Gauge, 0-100 psi +1% F/S or better—Ametek
maintained in the hospital file. The attached checklist
Model 1903, Ashcroft type 1279, or equivalent
may be copied, providing forms for future use.
Pressure Gauge, 0-10 cmH20 +1% F/S—
VERIFICATION PROGRAM Marshalltown Model 86F, with scale in cmH20,
Wallace & Tierman Model FA 14170 or equivalent
The Operational Verification Procedure (OVP) is divid- Timeter RT-200 or equivalent
ed into two paris. Test Lung, R5, C10
Regulated Air/Oz Source
Part 1, consists of an automated series of built-in diag- Shut Off Valves
nostics tests and should be performed at least monthly. Stopwatch
Part 2 consists of additional manual verifications and *(or equivalent)
should be performed at least twice a year.
CAUTION
CAUTION
Do not operate the ventilator without the filters in
The following cautions must be read and understood
before performing the procedures described in this sec-
tion. PLUGS AND CABLES
Do not use MEK or Trichloroethylene, as damage to Plugs and cables may be wiped down with an appropri-
surface may result. Do not allow any liquid to spill or ate bactericidal or germicidal agent.
drip into the ventilator.
Do not gas sterilize the ventilator. The internal materi-
als are not compatible with gas sterilization techniques.
Before using any test equipment [electronic or pneu-
matic] for calibration procedures [other than opera-
tional verification], the accuracy of the instruments
must be verified by a testing laboratory. The laboratory
master test instruments must be traceable to the NIST
(National Institute of Standards Technology) or equiv-
alent. When variances exist between the indicated and
actual values, the calibration curves [provided for each
instrument by the testing laboratory] must be used to
establish the actual correct values. This certification
procedure should be performed at least once every six
months. More frequent certification may be required
based on usage and environment.
} PART 1
Automated Built-In Diagnostics
Plug or Cap
Patient Wye Temp Probe
and Humidifier
Water
Port s Feed | Plug or Cap
Humidifier
Figure 3-1
Patient Circuit Connection
0-100 PSIG
TEST GAUGE
2-100 PSIG
TEST REGULATOR
Figure 3-2
External Gas Source Connections
TEST SETUP ENTERING THE OPERATOR DIAGNOSTIC MODE
1. Connect the test circuits to the ventilator as shown NOTE
in Figures 3-1 and 3-2. PEEP must be turned off completely prior to entering
the Diagnostic Mode to avoid test failure. Be sure the
2. Plug the open end of the patient wye and the water
inlet port of the humidifier. PEEP control is set completely counter clockwise.
3. Record the reading from the Hourmeter on the 1. With the ventilator power off, depress and hold the
checklist. TEST Key.
15. Select one of the following tests by pressing the Test 9. Turn off ventilator to exit Diagnostic mode.
Number Key and rotating the SET knob. The selec-
tion will appear in the Test Number display win- PART 2
dow.
MANUAL TESTING
a. All All Diagnostic Tests (d 1 through d 7)
Figure 3-3
0-100 PSIG
| SHUT-OFF VALVE TEST GAUGE
#1
Я 2-100 PSIG
し Ч TEST REGULATOR
ュー BEAKER
He AIR SUPPLY
で ーー
Figure 3-4
) AIR/O2 INLET CHECK VALVE VERIFICATION IR/O2 INLET PRESSURE GAUGE VERIFICATION
1. Connect the ventilator as shown in Figure 3-3. 1. Connect the ventilator as shown in Figure 3-5.
2. Adjust the controls as shown in Table 1 except 2. Sequentially apply 10, 30, and 50 psig.
change the Oxygen % to 60%.
3. Verify that at each pressure the inlet gauges indicate
Connect the ventilator as shown in Figure 3-4. +5 psig of the applied pressure.
Adjust the 100 psi regulator for an indication of 5. Close valve #2.
approximately 50 psi on the test gauge.
6. Verify that the Gas Supply failure alarm doesn't
Tum ON the ventilator power, ignore any alarms. come on.
Close Valve No.1 and connect the test gauge, regula- 2. Create any condition that will cause a ventilator
tor and shut-off valve to the air inlet. Connect the alarm (i.e. Pressure limit).
hose and beaker to 02 inlet. Repeat Steps 7 and 8.
3. While the alarm is sounding, adjust the alarm loud-
10. Turn the regulator down to zero and close Valve ness control on the rear of the ventilator.
No.1.
4. Verify that the alarm loudness can be varied but not
a. Turn the ventilator power OFF, and repeat the pro- turned completely off.
cedure with the power turned OFF.
5. Reset alarm to original setting or new setting if
12. Check the appropriate block on the OVP checklist. desired.
LOUDNESS
CONTROL
0 - 106 PSIG
TEST GAUGE
고 2 - 100 PSIG
TEST REGULATOR
Figure 3-5
Figure 3-6
BREATH RATE VERIFICATION 6. Disconnect the oxygen source and reconnect the air
source.
1. Set the ventilator as shown in Table 1.
7. Return the ventilator to the settings shown in Table
2. Allow the ventilator to cycle and using a stopwatch, 1.
verify the following rates +2 BPM:
NOTE
Due to a 3.5 liter accumulator chamber within the venti-
lator, oxygen percentage changes will not be immediate.
Allow time for stabilization to occur before verifying
specifications.
8
50-12129-00 + NOVEMBER 1993
OPERATION VERIFICATION PROCEDURE
Figure 3-7
ASSIST SENSITIVITY VERIFICATION 8. Repeat the preceding test to assure that the ventila-
tor is operating properly.
1. Set the ventilator as shown in Table 1, except set the
Rate to 0 BPM
DISCONNECT
Figure 3-8
DELIVERED/EXHALED VOLUME VERIFICATION 5. Reconnect expiratory limb and verify that exhaled
volume readings at 0.10 and 0.50 liters are within a
NOTE +0 % to -15% or +0.00 to -0.03 Liter (whichever is
Volume tolerances reflect values without heating or greater) tolerance range of the delivered volumes
humidification measured in step 3.
1. Connect the ventilator and analyzer as shown in
Figure 3-8. Table 3—Tidal Volume Settings
4. Verify that the Delivered Volumes are within the 2. Disconnect AC power at the rear of the ventilator.
specified tolerances.
3. Verify that the Power Loss alarm sounds.
10
50-12129-00 « NOVEMBER 1993
OPERATION VERIFICATION PROCEDURE
Monitor Fields Character one is not used at this time and the
position will be blank.
DELIV FLOW
Character two indicates that the Air Inlet Pressure
Shows the Delivered Flow in the range of 0 to 278 Transducer is sensing sufficient inlet pressure.
LPM.
Character three indicates that the Oxygen Inlet
EXT FLOW TRANSDUCER Pressure Transducer is sensing sufficient inlet
pressure.
Shows the Exhaled Flow in the range of 0 to 300
LPM. BLENDER/FLOW TRANSDUCER
0 0 14.7
500 152 14.4
1000 305 14.2
1500 457 13.9
2000 610 13.7
2500 762 13.4
3000 914 13.2
3500 1067 12.9
4000 1219 12.7
5000 1524 12.2
6000 1829 118
7000 2133 114
8000 2438 11.0
9000 2743 105
available as follows:
12 = 1200 Baud
24 = 2400 Baud
9.6 = 9600 Baud
19.2 = 19200 Baud
When this field is selected, the user may select one of the
other three rates using the Set knob.
12
50-12129-00 + NOVEMBER 1993
OPERATION VERIFICATION PROCEDURE
VERIFICATION STEPS
WARNING
Do not release the ventilator for use if it does not pass all of the verification procedures specified in the
checklist. To do so could result in personal injury including death or property damage. Refer the ventilator
to a Bear Medical Service Technician or a Bear Medical Systems trained service technician for appropriate
repair and/or calibration.
Signature.
PROCEDURE COMPLETE
13
50-12129-00. « NOVEMBER 1993
OPERATION VERIFICATION PROCEDURE
IR) CRI) (E
CO CSJCICSCS
o BUN DIAGNOSTICS
CS CS 回 Lİ
し DA ...
pun test |/ CYCLE CANCEL,
ONCE TEST TEST
ur codes
FIGURE 3-9
CautionsandWamings.........................
Test EQUIPMENT... e αλλ κεε
Proximal Pressure Transducer.......................
Differential Pressure Transducer......................
Machine Pressure Transducer..........................
FCV Pressure Transducer に に に に eee
Air & O2 Transducers i
D/A Converter ss
Exhalation Valve Reguldtor...................
Inlet Check Valves..................
GG renere rrrns
Air / O2 Balance <<ze<<<<<<<zrré
Nebulizer Calibration Gn
INTRODUCTION
The information in this section is intended for use by a WARNING
“Bear Medical Systems factory trained service technician.
Before opening the console or removing assemblies to per-
form calibration a reference should be made to the Oxygen Vigorously accelerates combustion. DO NOT
use any gauges, valves or other equipment that has been
Instruction Manual and Sections 1, 2, 3, and 7 of this
exposed to oi] or grease contamination to avoid violent
Manual to assure that the ventilator is being properly uti- ignition.
lized.
WARNING
NOTE
The Operational Verification Procedure (OVP) is always WARNING
performed prior to use of this section.
When high pressure gas sources are connected to
PROCESS the ventilators, always use extreme caution when
Complete the test or series of calibration verification tests
attempting to measure internal pressures.
necessary. Make an analysis and judgment as to what
Abnormal conditions may exist where measured
/assemblies to calibrate and/or replace. If calibration is
pressures exceed the burst pressure of the gauge,
required, perform the calibration procedures following the
resulting in possible injury.
calibration verification procedure in this document. If
replacement of an assembly is required, refer to the
Replacement Section (Section 8) for instructions.
CAUTION
TEST EQUIPMENT
Part Number Description г
Service Diagnostics
JE É
All specification values are nominal. Variance in test
equipment must be taken into consideration. Test equip- i È E
ment must be periodically checked to verify conformance - О -
Remove the single screw to the right of the Oxygen Inlet. EPI PCB
Tilt the top of the unit up until it remains in an upright
position.
Press and hold the diagnostic push switch while turning 0 - 200 cmH20
0 - 2 psi
on the units power switch. The diagnostic push switch is Gauge
Regulator d
located on the left side of Monitor board as viewed from
the rear (See Figure 5-1).
Figure 5-3
Active Controls for Calibration Mode
Select Calibration Point P.
P
Select and set 00.00 cmH2O in the Pressure 3 MSD Machine Pressure Transducer Calibration:
and Pressure 2 LSD windows
Adjust R1, if required, to obtain a reading of 716 Remove the tubing from the Machine Pressure
F
+10 in the Current Count window. Transducer.
Press the Set Point key to program the new calibra- Select Channel Number 3.
tion point. Select Calibration Point 0.
Connect the Differential Transducer as shown in Adjust R3, if required, to obtain a reading of 204
Figure 5-4 and apply 40 +.5 cmH20 pressure. +10 in the Current Count window.
Select Calibration Point P. Select and set 00.00 cmH20 in the Pressure 3 MSD
Select and set 40.00 cmH2O in the Pressure 3 MSD and Pressure 2 LSD windows.
and Pressure 2 LSD windows. Press the Set Point key to program the new calibra-
10. Press the Set Point to program the new calibration tion point.
point.
1. Connect the Differential Transducer as shown in
Figure 5-4 except to the opposite port (negative)
and apply 95 +.5 cmH20 pressure.
12. Select Calibration Point n. P
0-30 psi
Regulator_
+
LEE
FCV ノア
Ipazia
Transducer
0 - 200 cmH20
Gauge
Machine
Transducer
0-2 psi
Regulator
ra] Ra “ ee
x Nİ - +
18 È È
~ - + -
EPI PCB
삐 || RS s 삐 || RI
Figure 5-8
Interface Pin Locations
Figure 5-7
Air & 02 Transducer Calibration
Select and set 00.00 cmFRO in the Pressure 3 MSD
and Pressure 2 LSD windows. Figure 5-9
D/A Calibration Potentiometers
Press the Set Point key to program the new calibra-
tion point. Air € 02 Pressure Transducers Calibration:
Connect the FCV Transducer as shown in Figure 5-
6 and apply 12 +.5 psig pressure.
Connect an external gas source and gauge to the
Select Calibration Point P. ventilator as shown in Figure 5-7.
Select and set 12.00 psig in the Pressure 3 MSD and Select Channel Number 5 (Air).
Pressure 2 LSD windows.
DN
11. Select and set 28.50 psi in the Pressure 3 MSD and
Connect the tubing back to the FCV Pressure
Pressure 2 LSD windows.
Transducer.
Press the Set Point to program the new calibration
point.
Select Channel Number 6 (O2).
O (Chun pasnosnes
2,
804 POSITION ) (し gwuanowsOENOD ) ( SuiOREaBEEDSrENOGS )( NEmOUZERSOLEMOD )( ouuwAmowsotB*OD )
ER EN EM EM E
= OEE MARY,
Figure 5-10
Active Controls in Manual Mode
9 - 150 cmH20
OUTPUT
FITTING EXHALATION VALVE
REGULATOR ADJUSTMENT
X SCREW
COVEA
To Exhalation
Solenoid <
T
ID
3310
ADI
ANIHOWN
や Ha 573. ( )
XOUd
Е
|
cal
Nod
Figure 5-11
Exhalation Valve Regulator Calibration
8. Select Calibration Point P. Remove the pressure gauge and turn off (display =
0) both solenoids.
29. Press the Set Point to program the new calibration
point. Press the FCV POSITION key and turn the SET
D/A Converter Calibration knob until the Flow Control valve opens to maxi-
mum position.
NOTE Disconnect the Proximal Airway line at the BEAR®
1000 and connect it to the pressure gauge.
There have been two designs of the Control PCB. These
are interchangeable but the location of the potentiometers 5-0-5cmH20
0-100 PSIG GAUGE DIFFERENTIAL GAUGE
and their designations. Refer to Figure 5-9.
3. Press the Exhalation Solenoid key to turn on (dis- 13. Turn off the Shut-Off solenoid, Exhalation solenoid
play = 1) the Exhalation solenoid See Figure 5-10). and close the Flow Control valve.
4. Press the Shut Off key to turn on (display = 1) the 14. Connect a patient circuit to the Manley test lung
Shut Off solenoid. from the ventilator and set the Manley to C50 and
zero resistance.
5. Verify that the pressure gauge reads 140 + 10
cmH20. Readjust the low pressure regulator if nec- 15. Verify that the gas can flow from the dump port to
essary. the patient outlet port by pulling the Manley's bel-
lows.
6. If required, remove the Adjustment Screw Cover
and adjust the Exhalation Vaive regulator to
achieve 140 +10 cmH20
©
| =
i ῇ E Nebulizer = ANE
legulator
a© (©
BO djTD 日
Ho om tm
= 2 Figure 5-14
- Needle 25 LPM Nebulizer Regulator
PSI Gauge Valve Flowmeter
5. Set the Crossover Solenoid to 0 (closed). 5. The gauge should read 10 + 0.5 psi.
6. If required, adjust the Nebulizer Regulator by
pulling up the locking ring and turning the center
Air/O2 Pressure Balance knob (Ref Figure 5-14).
8
50000-12131 ・ NOVEMBER 1993
Section 6
BEAR® 1000 Ventilator
Schematics
+ dO 上 Ja3HS a Au
19980-0001S
WVHSVI DNIHIA NILSAS
5140000015 “ON DMMVIO 335 IS SHM SSIMIVA MIMOd DY 809 Ἡ
SORMAYA JANOd 00 30 4400 10087
831139
| Bah
Lett
po an arsine
| [se 7
130100005
1
70009 |
1993
AINCON HIMOd
ANGIOHS
Arad
« NOVEMBER
dem ASSY
πο
W
e
50000-12131
©
© A
5500 0015
4990
joyeg901
ABSY HO
polvos KE)“=:
i ouinoo
HOLMES Sia
100 BOWNY
TINA 30 aaisino
1590000015 1994]
4200043 0140
OHIO,
+ 30 1 133HS
85810-50828
80d
FATA TOHINO9 MOTI
Αναθνια OILVW
ヨHOS
‘91580-00015 AIBWISSV 494 κ
‘MPI "9 SUV ONY SPHO MM QISSIUAX SINIVA HOISISIU a
“NOS ABV SANA HOLOVAVO es
ou oi Y 个
a BRIT
tae
1993
uondo pus siseyo
24?
tom
Cru
nm
+ NOVEMBER
OC fer TO tuo)
1
tuer
dolay 12445
o
PD
50000-12131
T dA
soi
W Or E
す
A) A + r 3
ol T SOS 下
a
テー y
tua E $ SSH
oo
ILE)
was
δή
V
Ast ŘP HOT
sa
ng
+
q TK
ssa 1531
и Ey esmjetaduis | seg
pa 5] Y08, 198 a mos
Twas | 다
SCHEMAIICS
| dO + 1336
SCHEMATICS
57810-30879
ヨ1HQOW
AIdANS YAMOd
AVH9VIO OLVN3HOS
ano (54
x oso Ly
fine ae | за ONAS | 7 1906
nohy 45 masas Tat[ar |
Ges
, age | LANL 回
00 anco” Hiz: + E 39660"
Vi я È ANODDIBA $ vid Ange 120 ©
55 eH ange
Ete
的 κ ©
99
る3
거라
EIS 1151405
LA 5 =
gi)
omuz sueno Ly) 2
a xovadaas | e
us
È va |: | a
Oldun
vão
а E =
90
HOF
>
Нм ele > Agr
수 Cer tea A + s
HNIdSIQ
ΝΑ ©
CZ) = ET
u
之
ㆍ
ano
το
oa
AS FEN NPE
bsorisa
50000-12131
i
20
Узда
| 4+ 町
8 o
genes 01090
eS AG = KOU VIA, 9
HONAS ANAYUNO VS NO NVIS POSTO Y
NA
т
| 4
OLSUNN
MA
BHO
+ JO + 133HS | 9 Λ88
65810-90879 “VEBBO-OGIS ANBWISSV 43H
999 1091 №09
AVHOVIQ OLLVNAHOS
Jeog 05071 545 38 004 50010 NI UISSIUARI SANIVA BOLSISTH
'AOS 3UY SanTYA dOLIOVdyo
‘A93NNO9 ON Θ3ΙΝΟΙΟΝΙ ΟΝ.
“ABSV NO G2 71V1SN1 LON (gen) NOWdA "ASSY NO OBTIVISNÍ (981) 13008
1993
* NOVEMBER
o o VLELOHYL | 22
ol 02 BLELOHYL | bz
099g 576 88108 EL
o 00 9A91TVO | 22
o 16 $1628 0
と νε 19164 1869
50000-12131
+ À νε ΨΕΙΟΗΝΕ | 29
4 νε ФЕЗОНУ | 99
oz oy 2692998 |s9
é νε νοιοηνὲ |6?
4 tl. bLLOHbe | er
4 4 bh POLSHbL =| Lv
上 SL 2V695 oF
1 si 245698 |5
i s Bys689 | ph
E 와 209689 [eb
y El 219689 Br
+ 와 evsses | ip
ot oz ОНР | 88 上 s 209689 | 05
o oz VLLOHPL | 28 上 si 209609 | 66
bi e 215042 98 8 oh VE6SZLOHPL | ez
vi 一 一 ャSg9 se + EL Ένοθθα Sg
8 9ι BELLOHPZ | #8 8 st 9rfezNS | 0%
0 00 ELELOHbL | ES ol oz SPELOHPZ | B
SCHEMATICS
0 06 PLELOHPZ | 8 와 oz 3HVdS +
UNO DIA ‘0s3q “η ONS SOA “0630 “η
SCHEMATICS
*
=
ÉEm |B 6
Se
so 19[Lo
>
22
FE
[91
juju
EE lolz
£2 1B] 4
шо |
Z“ (©
5
の
o
=ù
x
>”
EST
UBA
TAT
U65
2x22222%
dd
23233398
ETES
174
RN13
108)
Ta HE E
z $
с
ESF