You are on page 1of 82

Model QW-400

Vertical Wall Frame

Models QW-420, QW-420-S,


and QW-420-D
Vertical Wall Stand/Frame

Service Manual

Manual Part No. DC30-040 Revision F


This manual is copyrighted and all rights are reserved. No portion of this document
may be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine readable form without prior consent in writing from Quantum
Medical Imaging, LLC. (QMI)

Copyright© 2013 QMI

Quantum Medical Imaging, LLC Made in U.S.A.


Division of Carestream
2002-B Orville Drive North
Ronkonkoma, New York 11779 USA
Phone: (631) 567-5800
Fax: (631) 567-5074
E-mail: info@qmiteam.com
Revision History

REVISION DATE TYPE OF MODIFICATION


A 2004-05-24 Initial Release

B 2004-10-05 Incorporated ECO# 1664, updated to new address

C 2007-06-22 Incorporated ECO# 1761, 1788, 1824

D 2010-09-21 Incorporated ECO# 2015, 2123, 2167, 2173, 2231, 2266,


2271, 2285, 2286, 2371

E 2010-12-27 Incorporated ECO 2399, 2432

F 2013-6-18 Changed EU Representative, transfer to ETL marking,


Incorporated ECO 2465, 2482, 2486, 2573, 2721, 2722,
2845, 2898

Page Page Page


Number Rev Number Rev Number Rev
i-x F

1-1 thru 1-12 F

2-1 thru 2-16 F

3-1 thru 3-4 F

4-1 thru 4-26 F

5-1 thru 5-4 F

A-1 thru A-8 F

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC i
Revision History

THIS PAGE INTENTIONALLY LEFT BLANK

Revision F Wall Stand/Frame - Service Manual


ii Quantum Medical Imaging, LLC
Table of Contents

SAFETY NOTICES
GENERAL SAFETY INFORMATION .............................................................v
WARNINGS, CAUTIONS, NOTES ...............................................................v
REGULATORY COMPLIANCE ................................................................... vii
CLASSIFICATION ............................................................................. vii
ACCOMPANYING DOCUMENTATION ................................................ viii
Applicable Standards ...................................................................... viii
ENVIRONMENTAL PROTECTION ........................................................ ix
ABBREVIATION DEFINITION ................................................................... ix

CHAPTER 1 - SPECIFICATIONS
OVERVIEW ......................................................................................... 1-3
PHYSICAL SPECIFICATIONS ................................................................. 1-3
ELECTRICAL SPECIFICATIONS ............................................................. 1-5
SYSTEM OPTIONS ............................................................................... 1-5
SYSTEM OPERATING ENVIRONMENT .................................................... 1-6
NON-OPERATING ENVIRONMENT ......................................................... 1-6

CHAPTER 2 - ASSEMBLY & INSTALLATION


OVERVIEW ......................................................................................... 2-3
UNPACKING ........................................................................................ 2-3
REQUIRED TOOLS ............................................................................... 2-3
ASSEMBLY/INSTALLATION INSTRUCTIONS ........................................... 2-3
MODELS QW-420, QW-420-D, QW-420-S
ASSEMBLY/INSTALLATION ............................................................. 2-4
MODEL QW-400 ASSEMBLY/INSTALLATION ...................................2-10
POST INSTALLATION CHECK .........................................................2-15

CHAPTER 3 - THEORY OF OPERATION


OVERVIEW ......................................................................................... 3-3
MECHANICAL OPERATION - Models QW-420, QW-420-D,
AND QW-420-S ............................................................................. 3-3
MECHANICAL OPERATION - Model QW-400 .................................... 3-3
ELECTRICAL OPERATION - Models QW-420, QW-420-D,
AND QW-420-S ............................................................................. 3-3
ELECTRICAL OPERATION - Model QW-400 ..................................... 3-3

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC iii
Table of Contents

CHAPTER 4 - SERVICE INSTRUCTIONS


OVERVIEW ..........................................................................................4-3
SERVICE MAINTENANCE .................................................................4-3
Cleaning ..................................................................................4-3
Visual Inspection ......................................................................4-4
REMOVAL/REPLACEMENT PROCEDURES ..........................................4-5
Replacing the Vertical Pulley Assembly
(Models QW-420, QW-420-D, QW-420-S Only) ...........................4-5
Cassette Tray Removal Procedure ..............................................4-8
FILM CABINET load side reverse procedure ............................................4-9
GENERAL INFORMATION ................................................................4-9
Disassembly of the Film Cabinet ..............................................4-11
Re-assembly of Reversed Load Side Film Cabinet ......................4-13
REPLACEMENT PARTS AND ORDERING INFORMATION .........................4-18
ORDERING INFORMATION ............................................................4-18

CHAPTER 5 - DIAGRAMS
Figure 5-1 Models QW-420, QW-420-D, QW-420-S Vertical Wall
Stand/Frame Wiring Diagram .................................................................5-3
Figure 5-2 Model QW-400 Vertical Wall Frame Wiring Diagram .................5-4

APPENDIX A
OVERVIEW ......................................................................................... A-3
VIEWORKS QXR 9C DIGITAL RECEPTOR INSTALLATION
PROCEDURE ................................................................................. A-3

Revision F Wall Stand/Frame - Service Manual


iv Quantum Medical Imaging, LLC
Safety Notices

GENERAL SAFETY INFORMATION


Quantum products are designed to meet stringent safety standards. All medical elec-
trical equipment requires proper installation, operation, and maintenance (particularly
with regard to safety).
It is vital that the user read, understand, note, and where applicable, strictly observe
all Warnings, Cautions, Notes and Safety markings within this document and on the
equipment, and that the user strictly follow all safety directions in this manual to help
ensure the safety of users and patients.
Every reasonable precaution has been taken during manufacture to safeguard the
health and safety of persons who will operate this equipment. The following precau-
tions must be observed at all times.
WARNINGS, CAUTIONS, NOTES
The following samples show how warnings, cautions, and notes appear in this docu-
ment. The text explains their intended use.

WARNING Indicates injury or death is possible if the instruc-


tions are not obeyed. Instructs users to refer to
documentation if displayed without warning text.

CAUTION Indicates that damage to equipment is possible if


the instructions are not obeyed.

NOTE Notes provide advice and highlight unusual


points. A note is not intended as an instruction.

The purpose of safety icons, such as those shown below, is to indicate at a glance the
type of caution, warning or danger.

WARNING Ionizing radiation: indicates the possibility of


increased levels of radiation.

WARNING Dangerous voltage: indicates the presence of high


voltage.

WARNING Warning, hot surface.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC v
Safety Notices

WARNING
Quantum Medical Imaging, LLC disclaims all responsibility from any injury
resulting from improper application of this equipment.
This equipment is sold to be used exclusively under the prescribed direc-
tion of a person who is licensed by law to operate equipment of this nature.
This equipment must be used in accordance with all safety procedures
described in this manual and must not be used for purposes other than
those described herein.
Quantum Medical Imaging, LLC cannot assume responsibility for any mal-
functioning of this equipment resulting from improper operation, mainte-
nance, or repair, or from damage or modification of its components.
Failure to observe these warnings may cause serious injuries.

WARNING
X-rays are hazardous to both patient and operator unless established safe
exposure factors and operating instructions are observed.
Only qualified and authorized personnel shall operate this system. In this
context, qualified means those legally permitted to operate this equipment
in the jurisdiction in which the equipment is being used, and authorized
means those authorized by the authority controlling the use of the equip-
ment. Full use must be made of all radiation protection features, devices,
systems, procedures and accessories.
It is important that everyone having anything to do with x-radiation be
properly trained and fully acquainted with the recommendations of the
National Council on Radiation Protection and Measurements as published
in NCRP Reports available from NCRP Publications, 7910 Woodmont Ave-
nue, Suite 800, Bethesda, Maryland 20814-3095 (www.ncrp.com), and of
the International Commission on Radiological Protection (www.icrp.org),
and take adequate steps to protect against injury

WARNING
X-ray equipment may cause injury if used improperly. The instructions con-
tained in this manual must be read and followed when operating this unit.
Personal radiation monitoring and protective devices are available. You are
urged to use them to protect against unnecessary X-ray exposure.

Revision F Wall Stand/Frame - Service Manual


vi Quantum Medical Imaging, LLC
Safety Notices

REGULATORY COMPLIANCE
This certified Quantum Medical Imaging, LLC medical device has been designed,
manufactured, and calibrated to comply with governing Federal Regulations 21 CFR
Subchapter J and the performance standards attendant thereto. Upon installation, all
certified products require the filing of Form FD-2579 "Report of Assembly of a Diag-
nostic X-ray System" by the assembler (i.e., the installer) with the appropriate agen-
cies; the "Installation Quality Assurance Checklist" must also be completed and
properly distributed upon installation. A copy of each form (pink copy) is provided to
the user. The installation report is completed by the installer and returned to Quan-
tum Medical Imaging, LLC.
Those responsible for the planning of X-ray equipment installations must be thor-
oughly familiar and comply completely with NCRP Report No. 49, "Structural Shielding
Design and Evaluation for Medical Use of X-rays and Gamma Rays of Energies up to
10 MeV", as revised or replaced in the future. Those authorized to operate, test, par-
ticipate in or supervise the operation of the equipment must be thoroughly familiar
and comply completely with the currently established safe exposure factors and pro-
cedures described in publications such as Subchapter J of Title 21 of the Code of Fed-
eral Regulations, "Diagnostic X-ray Systems and Their Major Components," and NCRP
Report No. 102, “Medical X-ray, Electron Beam and Gamma Ray Protection for Ener-
gies Up to 50 MeV—Equipment Design and Use” as revised or replaced in the future.
This equipment must only be used in rooms that comply with all applicable laws or
regulations that have the force of law, concerning electrical safety for this type of
equipment.
Scheduled maintenance is essential to the assurance of continued integrity of this
equipment with respect to regulatory compliance. The continuance of certified perfor-
mance to the regulatory standard is incumbent upon the user's diligent conformance
to recommended maintenance instructions. Do not use this equipment until you are
sure that the planned maintenance program is up to date.
Complies with FDA Center for Devices and Radiological Health (CDRH) - Title 21 CFR
Subchapter J, and EC Directive 93/42/EEC for Medical Devices

CLASSIFICATION
This product has been classified as Class I, Type B, not suitable for use in the
presence of a flammable anesthetic mixture of air with oxygen or with nitrous
oxide. Protection against Harmful Ingress of Water (Ordinary), enclosed
equipment without protection against ingress of liquids.

MEDICAL ELECTRICAL EQUIPMENT


WITH RESPECT TO ELECTRIC SHOCK, FIRE,
MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH
UL STD 60601-1 AND CAN/CSA C22.2 NO. 601.1
3195606

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC vii
Safety Notices

ACCOMPANYING DOCUMENTATION
The documentation consists of a Service Manual (this document) and related doc-
umentation:
• Operator’s Manual P/N DC30-039: Contains instructions for proper operation
and care of this equipment. The documentation shall be kept with the
system for easy reference.
APPLICABLE STANDARDS
This equipment complies with the following regulatory standards:
• EN 60601-1: 1990 + A1:1993 + A2:1995 + A13:1996
• IEC 60601-2-32: 1994
• CAN/CSA-C22.2 No. 601.1-M90, 2005 (Medical Electrical Equipment, part 1:
General Requirements for Safety)
• UL 60601-1, 1st Edition, 2006-04-26 (Medical Electrical Equipment, part 1:
General Requirements for Safety)
• IEC 60601-1 Medical electrical equipment, Part 1: General requirements for
safety
• IEC 60601-1-2: 2007

EU Authorized Representative:

Carestream Health France


1, rue Galilée
93192 NOISY-LE-GRAND CEDEX
France

The following symbols may be used for marking on this equipment or equipment doc-
umentation:

Non-ionizing Radiation Type B Equipment

Attention, consult accompany-


Earth (ground) ing documents

Protective Earth (ground)

Revision F Wall Stand/Frame - Service Manual


viii Quantum Medical Imaging, LLC
Safety Notices

ENVIRONMENTAL PROTECTION
This equipment contains certain materials and chemical compounds incidental to
the manufacture of electrical and electronic equipment, and improper "end-of-
life" disposal of such equipment can result in environmental contamination.
Therefore, this equipment should not be disposed of as ordinary household
waste, but should instead be delivered to a designated electrical and electronic
waste disposal or recycling center. For further information on disposing of electri-
cal and electronic waste, contact the cognizant authority within the jurisdiction.

ABBREVIATION DEFINITION
The following abbreviations and acronyms may be found in this document. The defi-
nition associated with each entry is intended to support its meaning within the con-
text of medical equipment manufacture, operation and service:

°C Degrees Celsius
A Ampere
CE Conformité Européenne
dc direct current
DR Digital Receptor
EC European Commission
EEC European Economic Community
hPa Hectopascal
kg kilograms
lb pounds
mmAl Millimeters of Aluminum
OCP Operator Control Panel

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC ix
Safety Notices

THIS PAGE INTENTIONALLY LEFT BLANK

Revision F Wall Stand/Frame - Service Manual


x Quantum Medical Imaging, LLC
Chapter
1

SPECIFICATIONS

1-1
1-2
Chapter 1 Specifications

OVERVIEW
This manual provides needed information for installing and maintaining the Models
QW-420, QW-420-D, and QW-420-S referred to herein as the Vertical Wall Stand/
Frame.

WARNING! EQUIPMENT INSTALLATION AND SER-


VICING PROCEDURES SHOULD BE PERFORMED
BY PROPERLY TRAINED AND QUALIFIED SERVICE
PERSONNEL ONLY.

PHYSICAL SPECIFICATIONS
The following are physical specifications of the Vertical Wall Stand/Frame (see Figures
1-1 through 1-6):

Model Model Model Model


QW-400 QW-420 QW-420-D QW-420-S

Height: 2159 mm 2134 mm 2134 mm 2134 mm


(85 in.) (84 in.) (84 in.) (84 in.)

Width: 673 mm 699 mm 640 mm 947 mm


(26.5 in.) (27.5 in.) (25.19 in.) (37.25 in.)

Distance from Wall 187 mm 324 mm 337 mm 203 mm


to Front of (7.36 in.) (12.75 in.) (13.25 in.) (8 in.)
Receptor:

Distance from N/A 114 mm 114 mm 203 mm


Front of Column to (8 in.)
(4.5 in.)Note 1 (4.5 in.)Note 1
Front of Receptor:

Receptor Front Fiber Resin Fiber Resin N/A Fiber Resin


Cover Material: (phenolic) (phenolic) (phenolic)

Front Cover <1 mmAl <1 mmAl N/A <1 mmAl


Density:

Receptor Types: 356 mm x 432 mm 356 mm x 432 mm Quantum Q-DRT 356 mm x 432
(14 in. x 17 in.) (14 in. x 17 in.) (used with mm
grid or bucky, grid or bucky, portable DR (14 in. x 17 in.)
356 mm x 914 mm 356 mm x 914 mm detectors) grid or bucky
(14 in. x 36 in.) (14 in. x 36 in.) Quantum Q-FDR
grid or bucky grid or bucky (used with fixed-
mounted DR
detectors)

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 1-3
Chapter 1 Specifications

Model Model Model Model


QW-400 QW-420 QW-420-D QW-420-S

Focal Spot-to-Floor 635 ±12.7 mm 381 ±12.7 mm 356 ±12.7 mm 381 ±12.7 mm
Distance (with (25 ±0.5 in.) with (15.0 ±0.5 in.) (14.0 ±0.5 in.) (15.0 ±0.5 in.)
356 mm x 432 mm fixed grid
[14 in. x 17 in.] 679 ±12.7 mm
receptor):Note 2 (27 ±0.5 in.) with
bucky

Focal Spot-to-Floor 886 ±12.7 mm 406 ±12.7 mm N/A N/A


Distance (with (34.88 ±0.5 in.) (16.0 ±0.5 in.)
356 mm x 914 mm with fixed grid
[14 in. x 36 in.] 962 ±12.7 mm
receptor): (37.88 ±0.5 in.)
with bucky

Collimator Manual, Manual, Manual, Manual,


Compatibility: Selectable, Selectable, Selectable, Selectable,
Automatic Automatic Automatic Automatic

Weight: Approx. 45.4 kg Approx. 98.0 kg Approx. 90.7 kg Approx. 98.0 kg


(100 lb) (225 lb) (200 lb) Note 3 (225 lb)

Image Receptor 1080 ±25.4 mm 1524 ±25.4 mm 1524 ±25.4 mm 1524 ±25.4 mm
Travel (with (42.5 ±1.0 in.) (60.0 ±1.0 in.) (60.0 ±1.0 in.) (60.0 ±1.0 in.)
356 mm x 432 mm with fixed grid
[14 in. x 17 in.] 1035 ±25.4 mm
cabinet)Note 2 (40.75 ±1.0 in.)
with bucky

Image Receptor 578 ±25.4 mm 1181 ±25.4 mm N/A N/A


Travel (with (22.75 ±1.0 in.) (46.5 ±1.0 in.)
356 mm x 914 mm with fixed grid with fixed grid
(14 in. x 36 in. 502 ±25.4 mm 1067 ±25.4 mm
cabinet) (19.75 ±1.0 in.) (42.0 ±1.0 in.)
with bucky with bucky

Notes:
1. Up to 8.0 inches (196 mm) from front of column to
front of receptor with optional extension QW-EX8
2. Measured with 25.4 mm (1.0 inch) space between
receptor cabinet bottom edge and floor.
3. All digital receptors for integration with QW-420-D
Wall Stands must not exceed 160 lb., including mounting
plate if required, and must comply with the following

Revision F Wall Stand/Frame - Service Manual


1-4 Quantum Medical Imaging, LLC
Chapter 1 Specifications

standards:
• UL/IEC 60601-1 Part 1, CAN/CSA C22.2 No. 601.1 (Safety
Requirements).
• UL/IEC 60601-1 Part 1, CAN/CSA C22.2 No. 601.1 (Leakage
Requirements).
• IEC 60601-1-2, Part 2 (Electromagnetic Compatibility Require-
ments and Tests).

ELECTRICAL SPECIFICATIONS
Mode of Operation: Continuous
Locking System Type: Electromagnetic, Fail-Safe
Locking System Power: 24 Vdc/1.0A
Power Supply Type: Double Insulated (or reinforced); must meet con-
struction/performance criteria set forth in UL/IEC
60601-1 Part 1, CAN/CSA C22.2 No. 601.1

Note: Systems equipped with a bucky require separate


120/220-240 Vac power supply.

SYSTEM OPTIONS
QW-HG20: Patient “Side-Mounted” Handgrips (Used on Model
QW-420 only)
QW-HG30: Patient “Overhead” Handgrip (Used on Models
QW-420 and QW-420-S only)
QW-EX8: 8-inch Bucky Extension Bracket (Used on Models
QW-420 and QW-420-D only)
QW-AT: Auto-Tracking Kit for synchronization of Wall
Stand with Siemens Motorized Tubestand
QW-DC: Mounting Configuration for Canon CXDI-40G
Imager. (Available on Model QW-420-D only.)

Table 1-1: Compatible Receptor Options

Used On Reference Reference


Option Type
(Wall Stand Model) Number Document

Varian 4343R non-portable QW-420-D Q-FDR-V-R, DC30-110


DR panel Q-FDR-V-L

Canon CXDI-40G COMPACT, QW-420-D Q-FDR-C-R, DC30-110


non-portable DR panel Q-FDR-C-L

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 1-5
Chapter 1 Specifications

Table 1-1: Compatible Receptor Options

Used On Reference Reference


Option Type
(Wall Stand Model) Number Document

Agfa DX-D 4336R Portable QW-420-D Q-DRT-VR, DC30-092


DR panel, non-rotating tray Q-DRT-VL

Agfa DX-D 4336R Portable QW-420-D Q-DRT-VR-R, DC30-092


DR panel, rotating tray Q-DRT-VL-R

Carestream DRX-1 Portable QW-420-D Q-DRT-DRXR-R, DC30-092


DR panel, rotating tray Q-DRT-DRXL-R

Carestream DRX-1 QW-420-D Q-DRT-WOD-R DC30-092


Portable DR Panel (Tethered) Q-DRT-WOD-L

Vatech (Samsung) QW-420-D Q-DRT-SAMR-R DC30-092


Xmaru 1417 Q-DRT-SAML-R

Vieworks QXR-9C non-porta- QW-420-D QXR9-C RA21-093-003


ble DR panel

Vieworks FXRD Portable DR QW-420-D Q-DRT-VWR-R DC30-092


Panel Q-DRT-VWL-R

Vieworks FXRD Fixed DR QW-420-D Q-FDR-VW-R DC30-110


Panel Q-FDR-VW-L

SYSTEM OPERATING ENVIRONMENT


Ambient Temperature: 10° C to 40° C (50° F to 104 ° F)
Relative Humidity: 30 to 75 %, non-condensing
Atmospheric Pressure: 700 hPa to 1060 hPa

NON-OPERATING ENVIRONMENT
Ambient Temperature: -18° C to +70° C (-0.4° F to 158° F)
Relative Humidity: 20 to 95 %, non-condensing
Atmospheric Pressure: 500 hPa to 1060 hPa

Revision F Wall Stand/Frame - Service Manual


1-6 Quantum Medical Imaging, LLC
Chapter 1 Specifications

Figure 1-1. Model QW-420 (with 14 in. x 17 in. Receptor) Dimensional Diagram

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 1-7
Chapter 1 Specifications

Figure 1-2. Model QW-420 (with 14 in. x 36 in. Receptor) Dimensional Diagram

Revision F Wall Stand/Frame - Service Manual


1-8 Quantum Medical Imaging, LLC
Chapter 1 Specifications

Figure 1-3. Model QW-420-S Dimensional Diagram


(Parallel Wall Mount Configuration)

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 1-9
Chapter 1 Specifications

Figure 1-4. Model QW-420-S Dimensional Diagram


(Perpendicular Wall Mount Configuration)

Revision F Wall Stand/Frame - Service Manual


1-10 Quantum Medical Imaging, LLC
Chapter 1 Specifications

DETAIL A Ø.38 X .75


SLOT HOLE
2X
18.25
[464]

MOUNTING
STRAP

34.94
[887]
SEE NOTE 1
26.63
[676]
SEE NOTE 1 C

WALL FRAME 19.63


QW-400 [498]

FOCAL 8'-9" MAX


SPOT [2667]
7'-9" MIN
[2362]
76
[1929]
Chiro - X

Chiro - X

A A

39.88
[1013]

Ø.31 MOUNTING
HOLE
2X SECTION A-A LEG 2X

Figure 1-5. Model QW-400 Dimensional Diagram

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 1-11
Chapter 1 Specifications

Figure 1-6. Model QW-420-D Model Dimensional Diagram

Revision F Wall Stand/Frame - Service Manual


1-12 Quantum Medical Imaging, LLC
Chapter
2

ASSEMBLY &
INSTALLATION

2-1
2-2
Chapter 2 Assembly & Installation

OVERVIEW
This chapter describes the steps required to assemble, install and align the Vertical
Wall Stand/Frame and how to install the completed assembly in your facility, including
making all required electrical connections.

UNPACKING
Note: Examine all cartons and crates carefully at time of
delivery. If damage is apparent, have delivery driver write a
"Demand Shipment Note" on copies of the freight bill, sign,
and file appropriate carrier claim. Should you discover
concealed damage, immediately notify the transporting
agent and ask for an "Inspection of Damage." Carrier will
not accept concealed damage claim if filed after 15 days
from date of receipt of merchandise.
The Vertical Wall Stand/Frame is shipped in separate pieces as follows:
• Vertical Wall Stand/Frame Assembly
• Associated system components, depending on the model and options - a
Base Cover, Top Cap, trim weights, cable covering (with bucky and/or Ion
Chamber equipped systems only), cassette tray, power input cable,
hardware bag (including coil cord tie-down clamp and ground stud), patient
handgrips (optional), and Operator and Service Manuals.
Open the crate or carton marked "packing list enclosed" first. Locate and remove the
packing list. Do not dispose of packing material until packing list is matched with
actual parts received. If any damaged parts are found, notify the shipping or freight
company immediately (the manufacturer is relieved of any responsibility for damage
during shipment after unit is picked up by the carrier). Should there be a shortage of
parts, notify Quantum Medical Imaging’s Service Department.

REQUIRED TOOLS
The following tools and materials are required to complete the assembly and installa-
tion procedures:
• Socket Wrench Set
• Assorted Phillips and slot blade screwdrivers
• Drill (and masonry bit if concrete construction)
• Bubble level

ASSEMBLY/INSTALLATION INSTRUCTIONS
The following paragraphs describe how to assemble the Vertical Wall stand/Frame.
Assembly instructions for Model QW-400 are described separately because of the dif-
ferences in this device’s design and construction. Assembly instructions for Models
QW-420, QW-420-D and QW-420-S are very similar in design and construction and

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 2-3
Chapter 2 Assembly & Installation

are therefore combined into one set of instructions with model differences noted
where necessary.
MODELS QW-420, QW-420-D, QW-420-S ASSEMBLY/INSTALLATION
To assemble the Model QW-420, QW-420-D or QW-420-S Vertical Wall Stand,
refer to Figures 2-1 through 2-7 and proceed as follows:
1. Rest the bottom end of the Vertical Column on soft support material.
2. Remove shipping bolt located in Flat Base Mounting Plate (see Figure 2-1).

Figure 2-1. Removing the Shipping Bolt


3. Model QW-420-D with IDC XPLORER Digital Receptor: Refer to
Appendix A for installation instructions. Model QW-420-D with QXR-9C
Digital Receptor: Refer to Appendix B for installation instructions.

WARNING! The Vertical Column weighs


approximately 102.1 kg (225 lb) and requires two
persons to lift.

4. Carefully raise the Vertical Column to an upright position. The Receptor Cab-
inet will rise up the column because of the pull of the counterweight.
5. Connect red and black wires from Quantum generator (Interface Board) or
from a double (or reinforced) insulated 24 Vdc power source (must meet con-
struction/performance criteria per UL 2601-1) to Terminal Board TB1 termi-
nals 1 and 2, respectively (see Models QW-420, QW-420-D, QW-420-S
Electrical Wiring Diagram in Chapter 5, DIAGRAMS). Use cable ties, as
required, to manage cable.

Revision F Wall Stand/Frame - Service Manual


2-4 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

6. Install ground stud (supplied with system) into threaded hole on back side of
Vertical Column. Connect #10AWG yellow/green ground wire between Verti-
cal Wall Stand/Frame’s ground stud (located on back side of Vertical Column
near base) and x-ray generator (see Figure 2-2).

Star washer
Ground Lug Nut
AY51-279-84

Nut
PROTECTIVE
EARTH

Locking Nut

Rear Wall of Column Threaded stud

T:\Storage\Manuals\QW-420\Service Manual \DRAWINGS\Ground Wire Detail.cdr

Figure 2-2. Ground Wire - Detail View


7. On systems equipped with a moving grid (bucky), connect bucky power cable
to terminal block on bucky (refer to bucky manufacturer’s documentation,
provided with system, for specific wiring instructions) and tie-wrap cable to
inside of Receptor Cabinet along the bottom surface. Route cable through
opening in rear side of Receptor Cabinet.
8. If Patient “Overhead” Handgrip option (QW-HG30) is supplied, install as fol-
lows:
a. Loosen the small black thumbscrew located on the Patient Handle
Mount Bracket until threads are no longer visible in mounting hole for
Patient Handgrip.
b. While pulling the Index Plunger out, insert the Patient Handgrip into
mounting hole in Patient Handle Mount Bracket and then release
plunger (see Figure 2-3).

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 2-5
Chapter 2 Assembly & Installation

c. Tighten the small black thumbscrew located on the Patient Handle


Mount Bracket.

INDEX
PLUNGER

PATIENT
HANDGRIP

THUMBSCREW

Figure 2-3. QW-HG30 Patient "Overhead" Handgrip Mounting


9. If Patient “Side-Mounted” Handgrips option (QW-HG20) is supplied, mount
each Side Handgrip to Receptor Cabinet using six (6) 10-32 x 1/2” pan head
phillips screws and star washers.

Note: If Patient “Overhead” Handgrip option (QW-HG30)


is also installed, add a Spacer (part number ME11-122,
provided with handgrips) between the lower end of Side
Handgrip that attaches to the Overhead Patient Handle
Mount Bracket. Secure handgrip on this side using six
(6) 10-32 x 5/16” pan head phillips screws (see Figure
2-4).

Revision F Wall Stand/Frame - Service Manual


2-6 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

SIDE
MOUNTED
HANDGRIP

SPACER
REQUIRED
HERE IF
OVERHEAD
HANDGRIP
ALSO
INSTALLED

Figure 2-4. QW-HG20 Patient "Side-Mounted" Handgrips Installation


10. Turn on the collimator light.
11. Align the collimator cross hair lines with the centerline on grid, cassette tray
or digital receptor.
12. On Models QW-420 and QW-420-S only: Insert a (25.4 cm x 30.5 cm
(10 in. x 12 in.) film cassette into tray in Receptor Cabinet.
13. Verify the counterweight is balanced properly so it provides a smooth up and
down motion of the Receptor Cabinet. If additional counterweight is neces-
sary, add trim weights as follows:
a. Bring the Receptor Cabinet down to minimum vertical travel position.
b. Remove two screws securing the Column Center Cover to top of the
column (see Figure 2-5) and carefully tilt the center cover outward
enough to gain access to Access Opening (see Figure 2-6).

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 2-7
Chapter 2 Assembly & Installation

COLUMN
CENTER
COVER
COLUMN
CENTER
COVER
SCREWS
(2X)

Figure 2-5. Column Center Cover Top Mounting Screws

UP TO 16 TRIM WEIGHTS
CAN FIT IN CONTAINER,
WHEN ORIENTED AS
S H O W N I N D E TA I L

TRIM
WEIGHTS

AS VIEWED LOOKING DOWN


THROUGH ACCESS OPENING
ACCESS OPENING

Figure 2-6. Trim Weight Access Opening Location

Revision F Wall Stand/Frame - Service Manual


2-8 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

c. Adjust the vertical position of the Receptor Cabinet so the top of the
counterweight is visible through the Access Opening.
d. Place trim weights (each weighs approximately 0.7 kg [1.5 lb]) inside
compartment on top of counterweight (see Figure 2-6) until proper
balance (motion) is achieved.
e. Re-install Column Center Cover mounting screws.
f. Re-check counterbalancing and adjust if necessary.
14. Secure the Vertical Column to the floor through the Flat Base Mounting Plate
using appropriate hardware (must be capable of supporting 100 lbs. pull-
load).
15. Secure the Vertical Column to the wall using the Wall-Mounting Bracket (and
optional Wall-Mounting Bar) with appropriate hardware (must be capable of
supporting 27.2 kg [60 lb] pull load), see Figure 2-7.

PULLEY/ WALL
BRAKE MOUNTING
ASSEMBLY BRACKET

Figure 2-7. Vertical Wall Mounting Bracket

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 2-9
Chapter 2 Assembly & Installation

16. On top of the Vertical Column, remove the two (2) phillips head screws from
standoffs. Install the Column Top Cap over top of column and secure by re-
installing the two (2) phillips head screws.
17. Move Receptor Cabinet through its entire vertical movement, making sure the
light beam remains on the center line on grid, cassette tray or digital receptor.
18. On Models QW-420 and QW-420-S only: Insert a film cassette into cas-
sette tray in Receptor Cabinet.
19. Take a projection into the cassette/receptor for verification of compliance.
(Refer to the calibration section of the Collimator manual.)
20. If equipped, dress bucky and AEC cables using corrugated tubing and cable
tie-wraps as appropriate.
21. Attach Base Cover to Flat Base Mounting Plate using two (2) 6-32 x 1/4 in.
pan head screws.
22. On Models QW-420 and QW-420-S only: Attach Front Cover to Receptor
Cabinet and align centerline using collimator light.
MODEL QW-400 ASSEMBLY/INSTALLATION
To assemble and install the Model QW-400 Vertical Wall Frame, refer to Figures
2-8 through 2-12 and proceed as follows:
1. Lift off the fiber-resin (phenolic) Front Cover from the Vertical Wall Frame.
2. Typically, the Receptor Cabinet is pre-installed into the Vertical Wall Frame at
the factory. However, instructions for assembly and installation of the Recep-
tor Cabinet are provided in steps 2a through 2k if needed for service or
replacement purposes:
a. Place the Receptor Cabinet on a flat surface with front side facing up,
top end away, and bottom end toward you.
b. Loosely install the two (2) Spring-Loaded Bearing Brackets on the top
and bottom non-film loading side of the Receptor Cabinet using two
(2) 1/4-20 x 3/4 in. screws on each bracket.
c. Loosely install Bearing Bracket (without spring), part number ME10-
013, on the bottom film-loading side of the Receptor Cabinet using
two (2) 1/4-20 x 3/4 in. screws.
d. Loosely install Bearing Bracket (without spring), part number ME20-
014, and Control Handle Mounting Bracket (part number ME20-012)
on the top film-loading side of Receptor Cabinet using two (2) 1/4-20
x 3/4 in. screws.
e. Install Receptor Cabinet by sliding it onto Bearing Rails in the Vertical
Wall Frame.
f. The slotted mounting holes in the Bucky Bearing Brackets are pro-
vided to adjust tension on bearings (wheels). When adjusted prop-
erly, the wheels on all four brackets should rotate throughout the

Revision F Wall Stand/Frame - Service Manual


2-10 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

entire travel distance of the Receptor Cabinet.


g. Slide Receptor Cabinet all the way down to bottom of Vertical Wall
Frame.
h. Lift the counterweight so that the Pulley Cables slip over the Pulleys.
i. Secure the Pulley Cables to two top Bucky Bearing Brackets on
Receptor Cabinet using 1/4-20 x 1 in. hex head bolts and flat wash-
ers.
j. Slowly allow the counterweight to pull Receptor Cabinet to top of Ver-
tical Wall Frame.
k. Check bumper stops making sure that the Receptor Cabinet does not
contact Vertical Wall Frame.

Figure 2-8. Receptor Cabinet Installed in QW-400 Wall Frame


3. Install the Control Handle with bracket to right side Bucky Bearing Bracket
(for right-hand film loading) or left side Bucky Bearing Bracket (for left-hand
film loading).
4. Connect the black coil cable wires to Terminal Board TB1 terminals 2 and 3 as
shown in Model QW-400 Electrical Wiring Diagram in Chapter 5, DIAGRAMS.
5. On systems configured for right-hand film loading, tie-wrap coil cord cable as
shown in Figure 2-9.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 2-11
Chapter 2 Assembly & Installation

6. On systems configured for left-hand film loading, tie-wrap coil cord cable as
shown in Figure 2-10.
7. Re-attach the Front Cover on the Vertical Wall Frame.
8. Stand the Vertical Wall Frame against wall in proper position. Level Vertical
Wall Frame front-to-back and left-to-right using bubble level (use shims to
level left-to-right, as necessary).
9. Turn on collimator light.
10. Align the collimator cross-hair lines to the center line on the Front Cover of
the Vertical Wall Frame. Adjust position of Vertical Wall Frame as required.
11. Drill holes in the floor using the holes in the Vertical Wall Frame’s floor mount-
ing bracket as a guide.

COIL CORD
TIE-WRAPPED
TO FRAME

COIL CORD TIE-


WRAPPED TO
COUNTERWEIGHT
CABLE

Figure 2-9. Model QW-400 Coil Cord Cable Routing (Right Hand Load Systems)

Revision F Wall Stand/Frame - Service Manual


2-12 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

COIL CORD
TIE-WRAPPED
TO FRAME

Figure 2-10. Model QW-400 Coil Cord Cable Routing (Left Hand Load Systems)
12. Secure the Vertical Wall Frame to the floor through the Floor Mounting Brack-
ets using the appropriate hardware for floor construction (must be capable of
supporting 45.4 kg [100 lb] pull-load) as shown in Figure 2-11.

MOUNTING
BRACKET
HOLES (2X)

Figure 2-11. Vertical Wall Frame Floor Mounting Brackets

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 2-13
Chapter 2 Assembly & Installation

13. Remove Front Cover from the Vertical Wall Frame.


14. Secure the Vertical Wall Frame to the wall through mounting plate in rear of
Vertical Wall Frame using appropriate hardware (must be capable of support-
ing 27.2 kg [60 lb] pull-load).

MOUNTING
BRACKET
HOLES (2X)

Figure 2-12. Vertical Wall Frame Wall Mounting Bracket


15. Connect red and black wires from Quantum generator (Interface Board) or
from a double (or reinforced) insulated 24 VDC power source (must meet
construction/performance criteria per UL 60601-1) to Terminal Board TB1 ter-
minals 1 and 2, respectively (see Model QW-400 Electrical Wiring Diagram in
Chapter 5, DIAGRAMS). Use cable ties, as required, to manage and secure
cable.
16. Ground the Vertical Wall Frame by connecting ground wire from system
ground to the ground stud located on the left leg (rear side) of the Vertical
Wall Frame Assembly.
17. If using a moving grid (bucky), connect bucky power cable to terminal block
on bucky (refer to bucky manufacturer’s documentation, provided with sys-
tem, for specific wiring instructions). Use cable ties to secure bucky cable
along top of Receptor Cabinet and to Control Handle Coil Cable (leave a ser-
vice loop to enable full travel of Receptor Cabinet).
18. Remove grid. Insert cassette into tray in Receptor Cabinet.
19. Move Receptor Cabinet to top of Vertical Wall Frame. Make sure that the Fail-
Safe Brake is de-energized by pressing the LOCK RELEASE button on the con-
trol handle attached to the Receptor Cabinet.
20. Line up the image receptor with collimator light field.
21. Tighten Bucky Bearing Brackets slightly.
22. Move Receptor Cabinet to the bottom of the Vertical Wall Frame.

Revision F Wall Stand/Frame - Service Manual


2-14 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

23. Re-check alignment of the image receptor with the collimator light field.
24. Ensure that control handle coil cord remains in Vertical Wall Frame groove.
Check for smoothness of Receptor Cabinet motion throughout range of travel.
Adjust Bucky Bearing Brackets if necessary.
25. Fully tighten the Bucky Bearing Brackets.
26. Install grid.
27. Take projection into cassette for verification of compliance. (Refer to Collima-
tor Manual for alignment procedures).
28. Re-attach Front Cover to Vertical Wall Frame Assembly.
POST INSTALLATION CHECK
All mounting procedures must be approved before, and checked after, installation
by the hospital or facility engineering department and a registered, professional,
structural engineer. It is recommended that a leakage current measurement be
taken following installation to verify system meets allowable levels.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 2-15
Chapter 2 Assembly & Installation

THIS PAGE INTENTIONALLY LEFT BLANK

Revision F Wall Stand/Frame - Service Manual


2-16 Quantum Medical Imaging, LLC
Chapter
3

THEORY
OF OPERATION

3-1
3-2
Chapter 3 Theory of Operation

OVERVIEW
Models QW-400, QW-420, , QW-420-D, and QW-420-S feature sufficient vertical travel to
accommodate all upright exams, and fail-safe electromagnetic locks for secure image
receptor positioning.
MECHANICAL OPERATION - MODELS QW-420, QW-420-D, AND QW-420-S
Smooth vertical (upward and downward) image receptor movement is achieved by
attachment of the receptor to a Vertical Trolley, which rides on eight spring-loaded
radial ball bearings along with four metal shielded guide bearings bolted to it.
MECHANICAL OPERATION - MODEL QW-400
Smooth vertical (upward and downward) image receptor movement is achieved by
attachment of the receptor to a set of four spring-loaded Receptor Bearing Guides.
ELECTRICAL OPERATION - MODELS QW-420, QW-420-D, AND QW-420-S
See Figure 5-1 in Chapter 5, DIAGRAMS.
Two fail-safe brakes (24 Vdc electromagnets) are located at the top of the vertical col-
umn (part of the Vertical Pulley/Brake Assembly). When the electromagnets are de-
activated (i.e., the Receptor Cabinet Lock Release Button is not pressed), the fail-safe
brakes "grab" the counterweight cable pulleys securely, locking the vertical motion of
the receptor cabinet. Pressing the Receptor Cabinet Lock Release Button energizes
the 24 Vdc electromagnets, releasing the fail-safe brakes and allowing the counter-
weight cable pulleys to spin freely thereby enabling vertical motion of the receptor
cabinet.
ELECTRICAL OPERATION - MODEL QW-400
See Figure 5-2 in Chapter 5, DIAGRAMS.
A fail-safe brake (24 Vdc electromagnets) is located at the top right corner of the ver-
tical frame. When the electromagnet is de-activated (i.e., the Receptor Cabinet Lock
Release Button is not pressed), the fail-safe brake "grabs" the counterweight cable
pulley securely, locking the vertical motion of the receptor cabinet. Pressing the
Receptor Cabinet Lock Release Button energizes the 24 Vdc electromagnet, releasing
the fail-safe brake and allowing the counterweight cable pulley to spin freely thereby
enabling vertical motion of the receptor cabinet.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 3-3
Chapter 3 Theory of Operation

THIS PAGE INTENTIONALLY LEFT BLANK

Revision F Wall Stand/Frame - Service Manual


3-4 Quantum Medical Imaging, LLC
Chapter
4

SERVICE
INSTRUCTIONS

4-1
4-2
Chapter 4 Service Instructions

OVERVIEW
This chapter provides instructions to assist the service technician in maintaining the
smooth operation of the Vertical Wall Stand/Frame. The Vertical Wall Stand/Frame
will function reliably when maintained according to the instructions provided in this
chapter. Only properly trained service personnel should service or maintain this
equipment. Safe equipment performance requires the attention of service personnel
who are specifically trained and experienced with medical x-ray apparatus. Applicable
preventive maintenance or any repair service should be performed by these skilled
individuals. Failure to follow manufacturer’s or service personnel’s recommendations
may result in serious injury.

WARNING! Personnel engaged in maintenance


activities should exercise normal caution and care
while working with electromechanical equipment.
Only properly trained service personnel should
service or maintain this equipment. Failure to fol-
low manufacturer's or service personnel's recom-
mendations may result in serious injury. Before
removing or opening any electrical power panels
or covers, verify that the incoming power supply is
turned OFF. In the event maintenance procedures
require power to be supplied to the unit, extreme
care MUST be exercised to insure the safety of
service or other personnel in the area.

WARNING! Only qualified and authorized persons


shall work on this equipment. In this context,
qualified means those legally permitted to work
on the equipment, and authorized means those
specifically authorized by local management.

SERVICE MAINTENANCE

WARNING! Always disconnect the equipment


from the main power supply prior to any cleaning.
Verify that the equipment is properly grounded
before attempting any electrical operation. Never
use abrasive polish on this equipment.
Cleaning
The Vertical Wall Stand/Frame’s Front Cover surface should be wiped
clean. Painted metal surfaces should be cleaned using a clean cloth
slightly moistened in warm soapy water (use mild soap). Wipe dry with a
clean wet cloth.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-3
Chapter 4 Service Instructions

Visual Inspection
A complete series of inspections and functional checks was conducted at
installation to insure proper operation of the system. However, periodic
routine inspection and maintenance of the Vertical Wall Stand/Frame
should be performed by a qualified service technician. The following
inspection and adjustment procedures are recommended to maintain the
system in its original operating condition:
Perform Every Six Months
• Visually inspect for wear and cleanliness
• Check for evidence of loose hardware or loose wires
• Check all bearings and bearing surfaces for cleanliness and corrosion
• Clean Bearing Tracks
• Conduct a general inspection for worn or damaged parts
• Check for smoothness of Receptor Cabinet motion
• Verify that all ground conductors are properly and securely installed
and free of corrosion or damage
• Check all electrical cabling and wiring for wear and fraying
• Check external covers for proper fit; inspect for scratches that may
indicate cover misalignment or damage caused by collisions with
footstools, trolleys, or other equipment
• Verify vertical electro-magnetic lock holding force exceeds 15.9 kg
(35 lb) for Models QW-420, TW-420-D and QW-420-S, QW-420-D or
8.2 kg (18 lb) for Model QW-400.

Revision F Wall Stand/Frame - Service Manual


4-4 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

REMOVAL/REPLACEMENT PROCEDURES
The following procedures provide step-by-step instructions for removal and
replacement of the repairable components in the Vertical Wall Stand/Frame.
Replacing the Vertical Pulley Assembly (Models QW-420, QW-
420-D, QW-420-S Only)
1. Remove the Top Cap and Base Cover from the Vertical Column
(see Figure 4-1).
COLUMN
CENTER COVER

TOP CAP

NOTE: MODEL
QW-420 IS SHOWN IN
THIS DIAGRAM, MODELSS COLUMN BASE
QW-420-D AND Q W-420- COVER
S HAVE IDENTICAL PART
LOCATIONS

Figure 4-1. Top Cap, Base Cover and Center Column Cover Locations
2. Remove the Column Center Cover from the Vertical Column (see
Figure 4-1).
3. Remove all trim weights from counterweight tray.
4. Move the receptor cabinet to upper most vertical position.
5. Turn off power to Vertical Wall Stand. Disconnect +24 Vdc power
cable from terminal block TB1 located on top of Vertical Column.
6. Disconnect ground wire from ground lug on back side of Vertical
Column.
7. If wall stand is equipped with a bucky and/or ion chamber, free
up bucky/ion chamber cable drape to allow enough slack for wall
stand to be laid down on floor. If sufficient drape cannot be
attained, disconnect bucky power and/or ion chamber cables
from wall stand.
8. Remove all remaining wires from terminal block TB1.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-5
Chapter 4 Service Instructions

TOP BRACKET
TERMINAL
MOUNTING
BLOCK TB1
SCREWS (4X)

Figure 4-2. Removing Vertical Pulley Assembly Mounting Screws


9. Remove four (4) ¼-20 x 3/4 in. hex head screws securing the
Vertical Column Wall Mount Bracket to Vertical Column (see Fig-
ure 4-2). Retain hardware for re-assembly.
10. Remove hardware securing the Vertical Column Base to floor.

WARNING! The Vertical Column requires two per-


sons to lift. Before lowering the column, prepare a
resting location for laying the column on; lay
down two pieces of 2 in. x 4 in. lumber about
three feet apart (cover lumber with soft cloth or
cushioning material to prevent damage to column
or scratching of painted surfaces).

11. Carefully, slide the Vertical Column forward until clear of Wall
Mount Bracket, then slowly lower column until horizontal on floor.
12. Remove two (2) ¼-20 x 3/4 in. screws securing Vertical Pulley
Assembly to Vertical Column (see Figure 4-3).

Note: Before disconnecting counterweight cables, tag the


cable on the left side “LEFT” and the cable on the right
side “RIGHT” (as viewed from top). This will ensure
correct position of cables on replacement Vertical Pulley
Assembly during re-installation.

Revision F Wall Stand/Frame - Service Manual


4-6 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

Figure 4-3. Removing Vertical Pulley Assembly Mounting Screws

Figure 4-4. Tilting the Vertical Pulley Assembly to Access Counterweight Cables
13. Tilt the Vertical Pulley out from Vertical Column enough to gain
access to counterweight cable mounting hardware (see Figure
4-4).
14. Remove two (2) ¼-20 x 7/8 in. screws securing each counter-
weight cable to Trolley Assembly.
15. Remove the Vertical Pulley Assembly from column.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-7
Chapter 4 Service Instructions

16. Position the replacement Vertical Pulley Assembly on the Vertical


Column and tilt forward as shown in Figure 4-4.
17. Route the LEFT and RIGHT counterweight cables over the appropri-
ate ends of the Vertical Pulley Assembly. (Do not cross left and right
cables!). Secure each cable to Trolley Assembly using two 1/4-20 x
7/8 in. screws. Ensure screws are tightened securely.
18. Insert the Vertical Pulley Assembly fully into Vertical Column and
secure to column using two (2) ¼-20 x 3/4 in. screws.

WARNING! The Vertical Column requires two per-


sons to lift.

19. Carefully lift Vertical Column to vertical position and slide into the
location on floor where it was originally installed.
20. Install four (4) ¼-20 x 3/4 in. hex head screws securing the Vertical
Column Wall Mount Bracket to Vertical Column (see Figure 4-2).
21. Install hardware to secure the Vertical Column Base to floor.
22. Return all previously removed trim weights to counterweight tray.
23. Re-connect wiring to terminal block TB1.
24. Return-bucky and/or ion chamber cables to original draping posi-
tions. If disconnected, re-connect the bucky and/or ion chamber
cables to wall stand.
25. Connect ground wire to ground lug on back side of Vertical Column.
26. Re-connect +24 VDC power cable to terminal block TB1.
27. Install the Column Center Cover, Top Cap and Base Cover on the
Vertical Column (see Figure 4-1).
28. Turn on power to Vertical Wall Stand. Verify proper receptor vertical
motion and locking functionality.
Cassette Tray Removal Procedure
To remove the cassette tray from the receptor cabinet housing for cleaning
or repair, proceed as follows:
1. Pull the cassette tray out by the handle until it stops.
2. Grasp both sides of the tray with your right hand placed near the
vertical wall stand/frame and your left hand towards the tray front.
3. Locate the finger hole on the right side of the tray bottom. A
spring-loaded plate covers this hole. Press the plate with your right
index finger, lift the cassette tray up slightly and pull the tray away
from the housing.

Revision F Wall Stand/Frame - Service Manual


4-8 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

FILM CABINET LOAD SIDE REVERSE PROCEDURE


It is possible to reconfigure the film cabinet on the Model QW-420 Wall Stand, to
achieve opposite side loading, if such a change should become necessary. Typically
the Wall Stand is ordered for either left- or right-loading characteristics, however with
the following procedure, and one replacement Cabinet Side Cover, referenced in the
Replacement Parts List (Table 4-1), a standard film cabinet, or a cabinet equipped
with a bucky can be easily converted. Inasmuch as changing the load side on the
standard film cabinet, and the cabinet equipped with a bucky are parallel processes
(with the bucky unit requiring a few extra steps) the diagram in Figure 4-5 provides
reference information supported by step-by-step instructions.
GENERAL INFORMATION
Prior to beginning the procedures described in this section, the Service Technician
must obtain the correct Cabinet Side Cover for the loading configuration desired.
Table 4-1 contains the specific information required to properly order the neces-
sary part. Care should be taken when ordering the Side Cover to specify whether,
or not, the cabinet is equipped with a Bucky, and from which side the cabinet will
load after the conversion has been completed.

Note: The procedures for the removal of covers and


handles are performed while the Film Cabinet is still
mounted to the Vertical Trolley of the Wall Stand. The
description below is based on a Cabinet with Overhead
Patient Hand Grip, and Bucky options. If the Film
Cabinet is not equipped with the Overhead Patient Hand
Grip option, the long mounting bracket will not be
present.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-9
Chapter 4 Service Instructions

Figure 4-5. Load Side Reverse Procedure Diagram

Revision F Wall Stand/Frame - Service Manual


4-10 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

Disassembly of the Film Cabinet


The film cabinet is equipped with four outer covers, the EZ-Glide Handle
(brake release), and possibly optional patient hand grips or handles,
which must be removed in order to change the film cabinet’s load side.
Proceed with disassembly as follows:
1. Working from the back of the Film Cabinet, remove all of the
Exterior Covers, except the Load Side Cover, by removing the
10-32 x 1/4 in. Phillips Head screw from the end of each Cover.
2. Working from the side of the Film Cabinet, remove the Load Side
Cover, by removing four (4) 10-32 x 1/4 in. Phillips Head screws
from the face of the Cover (see Figure 4-6).

Figure 4-6. Exterior Covers Removed from Film Cabinet


Figure 4-7 shows the location and mounting scheme for optional Patient
Handle Grips, and the EZ-Glide Handle (break release):
• Bucky Cabinet Back Plate (Item 1).
• EZ-Glide Handle (Item 2) - shown mounted to a Bucky Cabinet
equipped with a bracket for the Overhead Patient Hand Grip
option.
• Patient Side Hand Grips (right side shown), including spacer
installed at lower attachment point (Item 3).

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-11
Chapter 4 Service Instructions

• Optional Patient Overhead Hand Grip Pivot Mounting Bracket (Item


4). Refer to Chapter 2, ASSEMBLY & INSTALLATION for additional
information.

Figure 4-7. Back Plate - Bucky Cabinet


Proceed with removal of the handles shown in Figure 4-7, as follows:
1. Remove the EZ-Glide Handle assembly by removing four (4)
screws from the bracket to which rotating handle is attached.
2. The optional Overhead Patient Hand Grip is supported by the
Bushing Block shown above (Item 4). Prior to removing the Bush-
ing Block, the Overhead Hand Grip should be removed:
a. Remove the thumbscrew installed in the Support Block.
b. Pull the Index Plunger out as far as it will go.
c. Lift the Overhead Patient Handle out of the Support Block,
and release the Index Plunger and reinstall the Thumbscrew.
3. Remove twelve (12) mounting screws (3 at each attachment
point) from the Patient Hand Grip Pivot Handle Mounting Bracket,
mounted to the Film Cabinet. Remove the Handle Grip, the
spacer and long mounting bracket.

Revision F Wall Stand/Frame - Service Manual


4-12 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

4. Carefully lift Phenolic Panel off of the Bucky by releasing the vel-
cro strips around the perimeter of the Phenolic Panel.
5. Remove the Grid Panel from its mounting position to gain access
to the bolts that mount the Back Plate to the Trolley.
6. Remove eight (8) 1/4-20 bolts, nuts and washers, securing the
Film Cabinet Back Plate to the Vertical Trolley.
7. Remove the eight (8) 1/4-20 x 3/4 in. button head screws holding
the Back Plate to the Bucky, and lift the Bucky out of its mounted
position.
8. Refer to Figure 4-8, and remove the screws holding the End Plate
on the "Non Load Side" of the film Cabinet. Remove the End
Plate and reinstall it on the opposite side of the Bucky assembly
using the same hardware.

Figure 4-8. Non-Load Side Bucky Plate Removal


Re-assembly of Reversed Load Side Film Cabinet
1. Re-mount the Bucky assembly to the Back Plate in exactly the
same orientation it was in when it was removed.
2. Re-mount the Patient Hand Grip Pivot Handle Mounting Bracket.
3. Re-mount the Back Plate (including Bucky assembly) to the Verti-
cal Trolley, using eight (8) 1/4 in. x 20 bolts and washers and
Locking Nuts.
4. Re-install the Grid Panel into the Bucky assembly.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-13
Chapter 4 Service Instructions

5. Re-install the Phenolic Panel on the face of the Bucky Cabinet.


6. Locate and install the replacement Load Side Exterior Cover, using
the 10-32 x 1/4 in. Phillips Head screws (4X).
• Bucky Cabinet Exterior Load Side Cover:
Left Load: P/N: ME10-051
Right Load: P/N: ME10-032
• Grid Cabinet Exterior Load Side Cover:
Left Load: P/N: ME10-194
Right Load: P/N: ME10-195
7. Re-install the remaining Exterior Covers in their respective posi-
tions on the Bucky Cabinet.

Note: When performing a Load Side Reversal, the EZ-


Glide Handle Assembly must be reversed and installed on
the opposite side of the Film Cabinet.

Figure 4-9 shows an EZ-Glide Handle configured for a right load cabinet.
The handle reversing process is depicted in Figure 4-9.
8. Disassemble the EZ-Glide Handle to reverse its mounting orienta-
tion for installation on the other side of the Film Cabinet as fol-
lows:
a. Remove Finishing Plug (Figure 4-10, Item 8) of the EZ-Glide
Handle, and remove the 5/16 in. - 20 Lock Nut (Item 7).
b. Carefully remove the Flattened Shaft Washer (Item 6), two
Spring Washers (Item 5), Steel Shim Washer (Item 4), and
the Nylon Flat Washer (Item 9) from the flattened shaft of
the Pivot Handle Bracket (Item 1).
c. Separate the EZ-Glide Handle from the mounting bracket,
making sure to extract the Pivot Handle Spacer along with
the Pivot Shaft on the Bracket.

Revision F Wall Stand/Frame - Service Manual


4-14 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

The EZ-Glide Handle shown here is for a Once the Finishing Cap has been removed,
Right Load Film Cabinet. the Lock Nut holding the Mounting Bracket
in place can be removed.

The flattened pivot shaft and the rotation The EZ-Glide Handle disassembled with
limit pin are part of the Mounting internal components displayed.

Figure 4-9. EZ-Glide Handle - Steps to Disassembly


d. Turn the bracket over (longitudinally), so the threaded stud
passes through the aluminum handle from the opposite side,
and the pivot pin aligns properly with the guide.
e. Rebuild the Handle assembly as shown in Figure 4-10.
f. Pivot the EZ-Glide Handle about the bracket to ensure that it
pivots freely, without binding, and then re-install the Finish
Cap.
9. Mount the EZ-Glide Handle assembly to the back plate of the Film
Cabinet on the same side as the Film Cassette is loaded.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-15
Chapter 4 Service Instructions

Figure 4-10. EZ-Glide Handle Assembly Diagram

Revision F Wall Stand/Frame - Service Manual


4-16 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

10. If the Film Cabinet is equipped with Patient Side Hand Grips,
install them on the Back Plate. Make sure to install the Spacer
(P/N: ME11-122) to the lower Side Handle of the right side, if the
Film Cabinet is equipped with optional Overhead Patient Hand
Grip.
11. Install the Overhead Patient Hand Grip, by inserting the shaft into
the Bushing Block while the Index Plunger is pulled back, and the
Thumbscrew is unscrewed far enough to permit the shaft of the
Hand Grip to seat.
12. Release the Index Plunger and tighten the Thumbscrew suffi-
ciently to keep the Hand Grip from accidental withdrawal from the
Bushing Block.

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-17
Chapter 4 Service Instructions

REPLACEMENT PARTS AND ORDERING INFORMATION


Table 4-1 below provides a list of replaceable parts for the Vertical Wall Stand/Frame.
See Figures 4-11 through 4-19 for locations of replaceable parts. Use only original
replacement parts, as supplied or recommended by Quantum Medical Imaging, LLC.
ORDERING INFORMATION
To order replacement parts for the Vertical Wall Stand/Frame, contact the Service
Department at;
Quantum Medical Imaging, LLC.
2002-B Orville Drive North
Ronkonkoma, New York 11779 USA
Phone: (631) 567-5800 (x2)
Fax: (631) 567-5074
e-mail: service@qmiteam.com
When ordering replacement parts, supply the following information:
• Model and serial number of equipment
• Part number
• Part description
• Quantity required
• P.O. Number
• Shipping Instructions
When ordering components or parts not listed in Table 4-1, a complete description
of the part, including its function and location should be provided with the model
number and serial number of the unit.

Revision F Wall Stand/Frame - Service Manual


4-18 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

Table 1: Replaceable Parts

Part Model(s) Used Qty Used


Item Description
Number On in System

1 Switch, Gray Push button EL50-002 QW-400 1


QW-420 1
QW-420-D 1
QW-420-S 1
2 Handle Assembly AY20-022 QW-400 1
QW-420 1
QW-420-D* 1
QW-420-S 1
3 R6ZZ Bearing ME40-018 QW-420 4
QW-420-D 4
QW-420-S 4
4 1628-ZZ Bearing ME40-017 QW-420 8
QW-420-D 8
QW-420-S 8
5 Vertical Pulley Assembly AY20-081 QW-420 1
QW-420-D 1
QW-420-S 1
6 Brake, Fail-Safe EL80-002 QW-400 1
QW-420 2
QW-420-D 2
QW-420-S 2
7 Trim Weight (not shown) ME10-105 QW-420
QW-420-D As Req’d
QW-420-S
8 24-Volt Power Cable (not EL75-005 QW-400
shown) QW-420 35 feet
QW-420-D
QW-420-S
9 Chin Rest ME30-243 QW-420 1
QW-420-S 1
10 Front Panel (phenolic), 14 in. ME31-002 QW-420 1
x 17 in. QW-420-S 1
11 Cover, Cabinet, Top ME10-048 QW-420 1
QW-420-S 1
12 Cover, Cabinet, Bottom ME10-047 QW-420 1
QW-420-S 1

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-19
Chapter 4 Service Instructions

Part Model(s) Used Qty Used


Item Description
Number On in System

13 Cover, Cabinet, Non-Load ME10-049 QW-420 1


Side QW-420-S 1
14 Cover, Cabinet, Load Side, ME10-051 QW-420 1
Left Load Bucky QW-420-S
OR
Cover, Cabinet, Load Side, ME10-194
Left Load Grid Cabinet
15 Cover, Cabinet, Load Side, ME10-032 QW-420 1
Right Load Bucky QW-420-S
OR
Cover, Cabinet, Load Side, ME10-195
Right Load Grid Cabinet
16 Switch, Push button, Auto- EL50-030 QW-420 with 1
Track Enable QW-AT Option
17 Potentiometer, 5K, 10 Turn, EL20-024 QW-420 with 1
Precision QW-AT Option
18 Front Panel (phenolic) ME31-004 QW-400 1
19 12 in. Control Handle Retract- EL75-001 QW-400 1
able Coil Cord
20 Receptor Bearing Guides ME30-008 QW-400 4

Revision F Wall Stand/Frame - Service Manual


4-20 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

Figure 4-11. Replaceable Parts Location Diagram - All Models

4 4

Figure 4-12. Replaceable Parts Location Diagram -


Models QW-420, QW-420-D, QW-420-S

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-21
Chapter 4 Service Instructions

6
Figure 4-13. Replaceable Parts Location Diagram -
Models QW-420, QW-420-D, QW-420-S

Figure 4-14. Replaceable Parts Location Diagram -


Models QW-420, QW-420-S

Revision F Wall Stand/Frame - Service Manual


4-22 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

11

14 15
13

10 10

12

LEFT SIDE LOAD RIGHT SIDE LOAD

Figure 4-15. Replaceable Parts Location Diagram -


Models QW-420, QW-420-S

16

Figure 4-16. Replaceable Parts Location Diagram -


Model QW-420 (with QW-AT Option)

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-23
Chapter 4 Service Instructions

17

Figure 4-17. Replaceable Parts Location Diagram -


Model QW-420 (with QW-AT Option)

18 2

Figure 4-18. Replaceable Parts Location Diagram - Model QW-400

Revision F Wall Stand/Frame - Service Manual


4-24 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

19

20

Figure 4-19. Replaceable Parts Location Diagram - Model QW-400

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC 4-25
Chapter 4 Service Instructions

THIS PAGE INTENTIONALLY LEFT BLANK

Revision F Wall Stand/Frame - Service Manual


4-26 Quantum Medical Imaging, LLC
Chapter
5

DIAGRAMS

5-1
5-2
Chapter 5 Diagrams

Notes:

1. INSERT BLACK SHRINK 4. REFER TO GENERATOR'S SERVICE MANUAL


TUBING QMI P/N EL73-016 FOR DETAILED CONNECTIONS FOR
(2.2") OVER DIODE AND BLACK WIRES INSTALLED RECEPTOR/BUCKY TYPE.
CONNECTED TO BRAKES.

2. INSERT BLACK SHRINK TUBING


QMI P/N EL73-016 (2.0")
OVER RED WIRE CONNECTED TO TB1-3.

3. NOT USED ON Q-DRT


BUCKY WALL STAND

WALL STAND COLUMN 24 VDC FAIL SAFE VERTICAL BRAKES RECEPTOR HOUSING
G

X-RAY GENERATOR

BLACK

BLACK

BLACK

BLACK
D1
1N4005
A9 INTERFACE
RED D2
BOARD 1N4005
RED
TB1

A9TB2 1 24 VDC RED 1


2 24 VDCRTN BLACK 2
3
RED SEE NOTE 1
BLACK

BLACK SEE NOTE 2 BLACK


TO GENERATOR
BLACK A9 INTERFACE BOARD
HAND SWITCH RECEPTOR/BUCKY
(OPTION)
SW1 6 (SEE NOTE 4)
5
4 STUD OR

N/O PROTECTIVE TERMINAL


COM
3 EARTH (SEE NOTE 3)
BLACK 2 STUD
1 GREEN/YELLOW GREEN/YELLOW
RED
AY51-279-084 AY51-279-016

GREEN/YELLOW TO X-RAY GENERATOR MAIN GROUND STUD (OR PROTECTIVE EARTH)


GND STUD

SC50-006T Rev. G

Figure 5-1. Models QW-420, QW-420-D, QW-


420-S Vertical Wall Stand/Frame Wiring
Diagram

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging 5-3
Chapter 5 Diagrams

Figure 5-2. QW-400 Vertical Wall Frame


Wiring Diagram

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging 5-4
Appendix
A

VIEWORKS MODEL QXR


9C DIGITAL RECEPTOR
INSTALLATION

A-1
A-2
Appendix A Digital Receptor Installation

OVERVIEW
This section provides instructions for installing the Vieworks Model QXR 9C Digital
Receptor (hereinafter referred to as the Digital Receptor) onto the QW-420-D Vertical
Wall Stand. For additional information concerning the digital receptor, refer to the
Digital Receptor installation and operation documentation.
VIEWORKS QXR 9C DIGITAL RECEPTOR INSTALLATION PROCEDURE
The QW-420-D Vertical Wall Stand is designed to permit easy installation of the
Digital Receptor. The following steps describe how to mount the Digital Receptor
onto the QW-420-D Vertical Wall Stand. Proceed as follows:

CAUTION! To minimize the risk of damage, the


digital receptor requires two persons to install.
One person should be supporting the digital
receptor on each side when lowering it in its
mounting location within the housing.

1. Remove the Digital Receptor from its packaging, and position it for convenient
handling during mounting process, which requires a 2-man lift of the assem-
bly and attachment to the Wall Stand Trolley Adaptor Bracket, shown in Fig-
ure A-2.

MOUNTING
BRACKET

Figure A-1. Vieworks QXR 9C Digital Receptor Ready for Mounting

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC A-3
Appendix A Digital Receptor Installation

Note: If the steel Mounting Bracket (shown in Figure


A-1) is not already mounted to the Digital Receptor
assembly, mount it with six (6) M8 socket head screws.

2. Position a person on either sides of the Digital Receptor assembly, so it can be


lifted and rotated into position for mounting to the Wall Stand Trolley Adaptor
Bracket.
3. Carefully, lift the Digital Receptor (in a coordinated fashion), and position it so
the tapped holes in the Mounting Bracket (three on each side) align with
those provided in the Wall Stand Trolley Adaptor Bracket.
4. Install six (6) 5/16 in. x 3/4 in. socket head screws to attach the Mounting
Bracket to the Wall Stand Trolley Adaptor Bracket.

MODEL
QW-420-D
WALL STAND

WALL STAND
TROLLEY ADAPTOR
BRACKET

TROLLEY

Figure A-2. Mounting the Digital Receptor to the Wall Stand


5. Attach the Hand Control Assembly (AY20-201) to the Adaptor Bracket, using
two (2) 10-32 x ½ in. Phillips Head Screws as shown in Figure A-3.
6. Connect the wires (plug) from the short pigtail (Figure A-3) to the wire (plug)
coming from the Vieworks Camera All Lock Release (located at the bottom of
the cover at the front of the Digital Receptor). This will electrically connect
the Hand Control Handle and the Vieworks Camera All Lock Release in series,
allowing Fail Safe Brake release from either location.

Revision F Wall Stand/Frame - Service Manual


A-4 Quantum Medical Imaging, LLC
Appendix A Digital Receptor Installation

HAND CONTROL 10-32X1/2 in.


MOUNTING ASSEMBLY
(AY20-201) MOUNTS PHILLIPS HEAD
TO THE TROLLEY SCREWS (2X)
ADAPTOR BRACKET (FEX1008-1)

SHORT
PIGTAIL

Figure A-3. Mounting The Hand Control Handle

Figure A-4. Electrical Connections for Fail Safe Brake System

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC A-5
Appendix A Digital Receptor Installation

7. Connect the wires (plug) from the long pigtail to the remaining plug, as
shown in Figure A-4. This will connect the circuit to 24Vdc when the opposite
end of the long pigtail is connected to TB1.
8. Connect lugs on long pigtail to TB1 atop the wall stand (under the cap).

Figure A-5. Power Connection Is Made On TB1 (On Top Of Wall Stand)

Note: Refer to Figure 5-1 and Figure A-7 for wiring dia-
grams.

9. Install Digital Receptor covers.

Figure A-6. Model QXR 9C Digital Receptor Mounted to QW-420-D Wall Stand

Revision F Wall Stand/Frame - Service Manual


A-6 Quantum Medical Imaging, LLC
Appendix A Digital Receptor Installation

Figure A-7. Vieworks All Lock Interface Diagram

Wall Stand/Frame - Service Manual Revision F


Quantum Medical Imaging, LLC A-7
Appendix A Digital Receptor Installation

THIS PAGE INTENTIONALLY LEFT BLANK

Revision F Wall Stand/Frame - Service Manual


A-8 Quantum Medical Imaging, LLC