Replacement of Parts Cios Select
Replacement of Parts Cios Select
Replacement of Parts
System
Replacement of Parts
2 Document Version / Disclaimer / Copyright
Document Version
Siemens Healthcare GmbH reserves the right to change its products and services at any
time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens Healthineers office to order a current version or refer to
our website http://www.siemens-healthineers.com.
Disclaimer
Siemens Healthcare GmbH provides this documentation “as is” without the assumption of
any liability under any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.
Copyright
“© Siemens Healthineers, 2014” refers to the copyright of a Siemens Healthineers entity
such as:
Siemens Healthcare GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 2 of 155 © Siemens Healthineers, 2014
07.18
Table of Contents 3
1 General Information 5
1.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.1 Required Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.2 Required Tools, Test Equipment and Aids . . . . . . . . . . . . . . . . . . . 5
1.1.3 Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.1.4 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 3 of 155 Cios Select /Cios Select FD
07.18
4 Table of Contents
4.6.3 Grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
4.7 X-ray Tube, Collimator, DAP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
4.7.1 DAP (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
4.7.2 Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
4.7.3 Single-tank tube assembly (monoblock). . . . . . . . . . . . . . . . . . . . 87
4.7.4 Installing the monoblock cover . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
4.8 Mechanics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
4.8.1 Steering system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
4.8.2 Wheel chain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
4.8.3 Foot brake and wheels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
4.8.4 Neck covering installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
4.8.5 Control Parts for C-arm Orbital Movement . . . . . . . . . . . . . . . . . . 108
4.8.6 Angular movement force adjustment . . . . . . . . . . . . . . . . . . . . . . 116
4.8.7 Swivel movement force adjustment . . . . . . . . . . . . . . . . . . . . . . . 117
4.9 Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
4.9.1 C-arm cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
4.10 CCP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
4.11 TUI CCP, AIC Board, Speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
5 Trolley Covers 126
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 4 of 155 © Siemens Healthineers, 2014
07.18
General Information 1 5
1.1 General
n
USB keyboard (English) see SPC
n
Screw jack size 70/100 (only for replacing the wheel and the vertical 10759620
toothed belt)
n
Sealing compound (for monoblock covers) 7905490
n
Ratchet tension belt (for monoblock replacement) 10400644
n
Torx key set, Tx6 to Tx25 9900663
n
Tool case 9702457
n
TONNA S68 OIL (200 ML) 8126430
n
GREASE Tribol GR 100-2 PD - 20g 3491271
n
Paint sticks / spray cans See SPC
n
Spring balance up to 200 N 4415113
n
Loctite 221 2048874
n
Loctite 243 5507038
n
Digital multimeter See SPC
n
Dig. water level 800 mm 7414209
n
Torque wrench, 5 - 25 Nm 7059975
n
Pipe wrench 11145191
n
Dynamic test case 3790156
n
2.1 mm precision radiation filter 9900598
n
Precision radiation filter set (Two, 0.6 mm) 10860929
n
Resolution test kit (Type 41 is needed) 2871820
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 5 of 155 Cios Select /Cios Select FD
07.18
6 1 General Information
n
Resolution test, Type 52 kit 9900051
n
Safety tester for leakage current and protective conductor resistance test see SPC
n
SMFit spot meter 7752848
n
Dosimeter e.g., Unfors see SPC
1.1.3 Abbreviations
WARNING
Risk of death, injury or material damage.
Non-compliance can lead to death, injury, or material damage.
Please note:
The product-specific safety information in these instructions,
the general safety information in TD00-000.860.01...
the safety information provided by your service organization and
the safety information in XPR2.140.860.01...
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 6 of 155 © Siemens Healthineers, 2014
07.18
General Information 1 7
WARNING
Electrical voltage after power off
Even after the system switches off the power supply, electrical voltage is present
in the system.
Live lines can cause electric shock or mechanical damage through short-circuiting.
Risk of death, serious physical injury, or damage to property.
V
De-energize the system
If checks or measurements require working with a live line, cover all ener‐
gized parts to prevent unintentional contact.
Use extreme caution when working on live lines.
Use only measurement equipment and devices that are suited to the task
Read and observe the warning labels attached next to conducting parts
HZID_06H384798
WARNING
Electrical voltage on power filter Z1
There is no protective cover on the connector of power filter Z1. Live voltage is
present at the Z1.
Even after the system switches off the power supply, electrical voltage is present.
Live lines can cause electric shock.
Risk of death, serious physical injury, or damage to property.
V Deenergize the system.
If checks or measurements require working with a live line, cover energized
parts to prevent unintentional contact.
Use extreme caution when working on live lines.
Use only measurement equipment and devices that are suited to the task.
Read and observe the warning labels attached next to conducting parts.
HZID_06H459443
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 7 of 155 Cios Select /Cios Select FD
07.18
8 1 General Information
WARNING
Capacitors on D620, dangerous voltage
Non-compliance can lead to death, injury, or material damage.
Before starting work on relevant parts of the main unit, wait at least 1 minute
after the system switches off the power supply.
Check the LED indicator of the capacitor’s residual voltage on D620, the LED
goes off 1 minute after the system is switched off.
HZID_06H384800
WARNING
Electrical safety!
Non-compliance can lead to severe bodily injury or even death, as well as proper‐
ty damage.
After the cover panels are opened, parts under voltage are accessible. To
avoid danger, disconnect the system from the power supply before opening
the covers.
V
An uninterruptible power supply (UPS) is installed in the system. Unplug the
power input connector of the imaging system or switch off the UPS prior to
beginning service work.
If steps have to be performed using electrical power, the general safety infor‐
mation in “Medical Products; Safety Information; General Safety
Notes“ (TD00-000.860.01.xx) must be observed.
HZID_06H384658
CAUTION
Front cover touches the screw stem of the transformer and the metal parts of the
main unit
Risk of injury to service personnel and damage to the property
Do not touch the screw stem of the transformer to any part during service
HZID_06H466435
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 8 of 155 © Siemens Healthineers, 2014
07.18
General Information 1 9
CAUTION
Laser emissions!
This product contains lasers.
Disregarding safety precautions can lead to bodily injury, especially to the retina
L of the eye, resulting in irreversible damage to vision.
Do not look into the laser diodes (I.I. side or tube side)
Observe local safety guidelines regarding laser emissions.
HZID_06H383909
CAUTION
Hot rear surface of TCP
Risk of burns
I
When replacing the trolley control panel (TCP), please note that its rear sur‐
face may be hot. Handle with care!
HZID_06H423193
CAUTION
Hot rear surface of CCP
Risk of burns
I
When replacing the chassis control panel (CCP), please note that its rear sur‐
face may be hot. Handle with care!
HZID_06H537290
CAUTION
Radiation released from a footswitch in another room
Non-clinical exposure to ionizing radiation
Upon service activities with the wireless footswitch, it shall be checked
whether the location labeling (i.e.which system the footswitch belongs) is in
place and is properly visible
HZID_06H488922
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 9 of 155 Cios Select /Cios Select FD
07.18
10 1 General Information
NOTICE
Dose adjustment
After each dose adjustment an IQAP measurement is necessary.
Perform IQAP measurement after each dose adjustment.
HZID_06H384277
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 10 of 155 © Siemens Healthineers, 2014
07.18
Adjustment and IQ tests after replacement 2 11
These IQAP tests should be performed after all other service adjustments have been
made. Create a partial inspection (Quality Assurance > New). After performing all re‐
quired IQ tests, lock the active inspection to save the results by clicking "Lock" in the
IQAP Control Functions section.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 11 of 155 Cios Select /Cios Select FD
07.18
12 2 Adjustment and IQ tests after replacement
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 12 of 155 © Siemens Healthineers, 2014
07.18
Adjustment and IQ tests after replacement 2 13
If several parts are replaced, the combined tests required for each part must be per‐
formed.
* Only available if the system has a DAP meter.
** see chapter( Optical Sharpness, Image Rotation, Centering, Dose TV Iris and IQAP
test / Page 72)
If the “Dose Sensitivity” IQ test returns a result that is out of tolerance, perform a “Dose
TV-iris”/ “Dose DEXI” adjustment in Tune-up.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 13 of 155 Cios Select /Cios Select FD
07.18
14 3 Main Unit Covers
With some covers, the scale is not part of the cover and must be ordered separately if a
replacement scale is needed.
6
4
2
5
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 14 of 155 © Siemens Healthineers, 2014
07.18
Main Unit Covers 3 15
NOTICE
Cable damage without main unit covers
After replacing the main covers, during C-arm movement the C-arm cable can be
damaged on sharp edges.
Always move the C-arm carefully without main covers.
Prerequisites
n Switch off system
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 15 of 155 Cios Select /Cios Select FD
07.18
16 3 Main Unit Covers
Work steps
n Loosen 2 screws on each side
n Open the left/right cover
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 16 of 155 © Siemens Healthineers, 2014
07.18
Main Unit Covers 3 17
n Release the latch on the upper and lower side of each cover
Fig. 5: Side cover
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 17 of 155 Cios Select /Cios Select FD
07.18
18 3 Main Unit Covers
CAUTION
Front cover touches the screw stem of the transformer and the metal parts of the
main unit
Risk of injury to service personnel and damage to the property
Do not touch the screw stem of the transformer to any part during service
HZID_06H466435
Work steps
n Unplug the power supply cable and foot switch cable from the front cover
n Remove 4 screws from the front cover
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 18 of 155 © Siemens Healthineers, 2014
07.18
Main Unit Covers 3 19
(1) Disconnect the power supply cables from the filter, and the grounding cable from the base of the main unit.
(2) Remove the footswitch cable from the front cover.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 19 of 155 Cios Select /Cios Select FD
07.18
20 3 Main Unit Covers
When installing the front cover, the footswitch cable should go through under the hex
rod. Also make sure that none of the cables touch to prevent abrasion.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 20 of 155 © Siemens Healthineers, 2014
07.18
Main Unit Covers 3 21
Fig. 11:
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 21 of 155 Cios Select /Cios Select FD
07.18
22 3 Main Unit Covers
n Loosen the 3 securing screws of the I.I. top cover (with handle), and remove handle
and top cover.
Disconnect the ground cable from the cover
Fig. 13: I.I. top cover
Hold the cover carefully to avoid dropping when loosening the attachment screws if
turning the C-arm vertically and the FD is on top
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 22 of 155 © Siemens Healthineers, 2014
07.18
Main Unit Covers 3 23
n Unwind the grounding cable, move FD out and then disconnect them to remove FD
n Untie and disconnect FD laser cable (if present)
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 23 of 155 Cios Select /Cios Select FD
07.18
24 3 Main Unit Covers
n Loosen the two attachment screws to release FD signal cable holding plate
Fig. 15: FD supporting frame
Lead shielding is not part of the FD top cover. Tear it off and attach it to the new one
when replacing FD top cover
Most of the lead shielding will be held by the FD supporting frame. Secure it with silicon
glue for the part not in the frame.
The monoblock cover has been sealed all around against contamination and dirt with a
sealing compound.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 24 of 155 © Siemens Healthineers, 2014
07.18
Main Unit Covers 3 25
Work steps
n Loosen the three cover screws of the monoblock cover
Fig. 16: Remove monoblock cover Fig. 17: Remove monoblock cover
n Use a sharp knife to cut open the sealing compound all around the cover
n Lift off the monoblock cover
n Remove the sealing compound residue from the cover and from the tube
n After all service procedures are finished, place the cover on the tube and seal it with
sealing compound all around.
n Use a paper towel to wipe off any excess sealing compound
n Refasten the monoblock cover with the cover screws. If necessary, apply lateral pres‐
sure to the monoblock cover until the sealing compound hardens.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 25 of 155 Cios Select /Cios Select FD
07.18
26 4 Main unit
4.1.1 Orientation
Fig. 18:
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 26 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 27
6 4
5
(1) CCP
(2) D610
(3) D620
(4) Main wheel
(5) D621
(6) Wheel chain
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 27 of 155 Cios Select /Cios Select FD
07.18
28 4 Main unit
1
3
5
2
(1) MCU
(2) D142
(3) U2
(4) Vertical motor
(5) Foot brake
(6) SG cable connector
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 28 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 29
(1) T1
(2) T2
(3) Filter
(4) Lower end switch assembly
(5) C-arm cable holder
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 29 of 155 Cios Select /Cios Select FD
07.18
30 4 Main unit
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 30 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 31
** For VA10, the speaker is positioned beside steering rod while for VA20 the speaker is
positioned under TUI CCP and AIC bard.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 31 of 155 Cios Select /Cios Select FD
07.18
32 4 Main unit
1 2
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 32 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 33
4.3.1 MCU
It will be necessary to apply for new licenses and service keys when the MCU is replaced.
Prerequisites
n Switch off the system.
n Open the left side cover as described in “Main Unit Covers” ( Main Covers / Page 15).
Work steps
n Unplug all connectors from the MCU board
n Remove the shielding plates from MCU, install the shielding plates on the new MCU
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 33 of 155 Cios Select /Cios Select FD
07.18
34 4 Main unit
Do not interrupt the download procedure to the MCU because this will damage the MCU.
Wait until the message “Download completed” appears in the action bar.
After the download is completed, “Version Check” and “Config Check” requests can be ig‐
nored until a MCU restore has been performed.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 34 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 35
For VA10, D611 (STU board) is integrated into the D610 board, STU board cannot be re‐
placed separately
For VA20, D211 (STU board) is separate from D610 board, can be replace separately.
After the D610 and D611 are replaced, FC2_Min will need to be adjusted. Perform the
adjustment with an oscilloscope.
After the D610 and D611 are replaced, the filament current heating (IH) will need to be
adjusted. Perform the adjustment with an oscilloscope.
After the D610 and D611 are replaced, the NTC temperature will need to be adjusted. It
is recommended to use an infrared thermometer to perform the adjustment.
After the D610 and D611 are replaced, “Set Wake On LAN” and “Set On Time” will need
to be adjusted.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 35 of 155 Cios Select /Cios Select FD
07.18
36 4 Main unit
WARNING
Capacitors on D620, dangerous voltage
Non-compliance can lead to death, injury, or property damage.
Before starting work on relevant pants of the main unit, wait at least 1 minute
after the system switches off the power.
Check the LED indicator of the capacitor’s residual voltage on D620, the LED
goes off 1 minute after the system is switched off.
HZID_06H384800
Prerequisites
n Switch off the system.
n Open the right side cover as described in “Main Unit Covers” ( Main Cov‐
ers / Page 15).
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 36 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 37
Work steps
n Disconnect all cables from D610
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 37 of 155 Cios Select /Cios Select FD
07.18
38 4 Main unit
PTC (Positive Temperature Conductor) can be damaged easily. Be careful not to touch it
unintentionally.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 38 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 39
n Perform the firmware upgrade following the instructions in the online help.
If the upgrade procedure is interrupted, the CPU may be damaged such that it is no lon‐
ger usable.
Do not touch anything until the upgrade is completed.
n Exit the service software and PMS, restart the entire system
n Navigate to the version check, check whether the installation was successful
Adjustments: FC2_Min / Heating / Set Wake On LAN /Set On Time /NTC Temp
n Move the single tank into a safe exposure position and close the collimator.
n Set the SERV button on D610 to ON ( position 2 )
While the SERV button is OFF (position 1), no values are displayed.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 39 of 155 Cios Select /Cios Select FD
07.18
40 4 Main unit
n Select Adjustment > Heating (IH), and then perform the adjustment following the in‐
structions in the online help.
n Select Adjustment->Set WakeOn LAN, perform the adjustment following the instruc‐
tions in the online help
n Select Adjustment >Set On Time, and then perform the adjustment following the in‐
structions in the online help.
n Select Adjustment > NTC Temp, and then perform the adjustment following the in‐
structions in the online help.
For VA20, D211 (STU board) is separate from D610 board, can be replace separately.
After the D211 are replaced, “Set Wake On LAN” and “Set On Time” will need to be adjus‐
ted.
Prerequisites
n Switch off the system.
n Open the right side cover as described in “Main Unit Covers” ( Main Cov‐
ers / Page 15).
Work steps
n Disconnect all cables from D211
n Remove the screws to remove D211
n Install the new D211
n Check the firmware and hardware version, upgrade the firmware if necessary.
Upgrading the firmware
n Switch on the system
n Log in to the SSW, perform the version check
n If a firmware upgrade is required, exit the version check, select Tune Up->Local Gener‐
ator Tool -> Firmware Upgrade
n Perform the firmware upgrade following the instructions in the online help.
If the upgrade procedure is interrupted, the CPU may be damaged such that it is no lon‐
ger usable.
Do not touch anything until the upgrade has been completed.
n Exit the service software and PMS, restart the entire system
n Navigate to the version check, check whether the installation was successful
Adjustments: Set Wake On LAN/Set On Time
n Move the single tank into a safe exposure position and close the collimator.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 40 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 41
n Set the SERV button on D610 to ON (position 2)( Fig. 28 Page 39)
While the SERV button is OFF (position 1), no values are displayed.
n Log in to the SSW, and then select Tune Up > Local Generator Tools.
n Select “Technical Service” and then enter the Service password.
n Select Adjustment >Set WakeOn LAN, perform the adjustment following the instruc‐
tions in the online help
n Select Adjustment >Set On Time, and then perform the adjustment following the in‐
structions in the online help.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 41 of 155 Cios Select /Cios Select FD
07.18
42 4 Main unit
After the D620 is replaced, the FC2_Min will need to be adjusted. Perform the adjust‐
ment with an oscilloscope.
After the D620 is replaced, the filament current heating (IH) will need to be adjusted.
Perform the adjustment with an oscilloscope.
WARNING
Capacitors on D620, dangerous voltage
Non-compliance can lead to death, injury, or property damage.
Before starting work on relevant parts of the main unit, wait at least 1 minute
after the system switches off the power supply.
Check the LED indicator of the capacitor’s residual voltage on D620, the LED
goes off 1 minute after the system is switched off.
HZID_06H384800
Prerequisites
n Switch off the system.
n Open the right side cover as described in “Main Unit Covers” ( Main Cov‐
ers / Page 15).
Work steps
n Disconnect the cables from D610
n Remove the D610
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 42 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 43
6 4 2
5 3 1
n Unscrew the D620 and replace it with a new one. Apply heat sink grease if necessary.
n Connect all cables.
Adjustments: FC2_Min / Heating
n Move the single tank into a safe exposure position and then close the collimator.
n Set the SERV button on the D610 to ON ( position 2 )( Fig. 28 Page 39)
While the SERV button is OFF (position 1), no values are displayed.
n Log in to the SSW, and then select Tune Up > Local Generator Tools.
n Select “Technical Service” and then enter the Service password.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 43 of 155 Cios Select /Cios Select FD
07.18
44 4 Main unit
n Select Adjustment > F2_Min, and then perform the adjustment following the instruc‐
tions in the online help
n Select Adjustment > Heating (IH), and then perform the adjustment following the in‐
structions in the online help.
After the D621 is replaced, the FC2_min will need to be adjusted. Perform the adjust‐
ment with an oscilloscope.
After the D621 is replaced, the filament current heating (IH) will need to be adjusted.
Perform the adjustment with an oscilloscope.
Instructions in this document apply only for system software versions VA10C and higher.
Instructions for system software version VA10B with system serial numbers from 10000
to 10099 can be found in the documentation for UI XP038/16/P.
WARNING
Capacitors on D620, dangerous voltage
Non-compliance can lead to death, injury, or property damage.
Before starting work on relevant parts of the main unit, wait at least 1 minute
after the system switches off the power supply.
Check the LED indicator of the capacitor’s residual voltage on D620, the LED
goes off 1 minute after the system is switched off.
HZID_06H384800
Prerequisites
n Switch off the system.
n Open the back and right side covers as described in “Main Unit Covers” ( Main Cov‐
ers / Page 15).
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 44 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 45
n Loosen and remove the top 4 long fixing nuts ( 2/Fig. 31 Page 45)
( 4/Fig. 31 Page 45)and 2 screws on the left ( 3/Fig. 31 Page 45)
n Take out the D621 and replace the new board
Fig. 31: D621
4
3
1
2
n Install the 2 long fixing nuts on the right ( 4/Fig. 31 Page 45) to fasten D621
n Attach the plasticized glass cover plate with the 4 screws( 1/Fig. 31 Page 45).
Adjustments: FC2_Min / Heating
n Move the single tank into a safe exposure position and then close the collimator.
n Set the SERV button on the D610 to ON ( position 2 )( Fig. 28 Page 39)
While the SERV button is OFF (position 1), no values are displayed.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 45 of 155 Cios Select /Cios Select FD
07.18
46 4 Main unit
n Log in to the SSW, and then select Tune Up > Local Generator Tools.
n Select “Technical Service” and then enter the Service password.
n Select Adjustment > F2_Min, and then perform the adjustment following the instruc‐
tions in the online help
n Select Adjustment > Heating (IH), and then perform the adjustment following the in‐
structions in the online help.
4.3.6 D142
D142 is located near the MCU, on the left side of main unit chassis.
Fig. 32: D142
(1) MCU
(2) D142
Prerequisites
n Switch off the system.
n Remove the left and right side covers as described in “Main Unit Covers” ( Main Cov‐
ers / Page 15).
Work steps
n Remove the four nuts from the back cover as described in “Main Unit Covers” ( Main
Covers / Page 15).
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 46 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 47
n Pull out the back cover slowly to show the 4 screws of D142, be careful of the cable
connections on back cover.
Fig. 33: Pull out the back cover
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 47 of 155 Cios Select /Cios Select FD
07.18
48 4 Main unit
4.3.7 U1
Power supply U1 is located in the middle of main unit chassis.
Prerequisites
n Switch off the system.
n Remove the left and right covers as described in “Main Unit Covers” ( Main Cov‐
ers / Page 15).
Work steps
n Unscrew the back cover as described in “Main Unit Covers” ( Main Covers / Page 15).
Pull out the back cover to access the U1.
n Disconnect cables from U1
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 48 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 49
Fig. 36:
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 49 of 155 Cios Select /Cios Select FD
07.18
50 4 Main unit
4.3.8 U2
Power supply U2 is located on the left side of main unit chassis.
Prerequisites
n Switch off the system.
n Open the left cover as described in “Main Unit Covers” ( Main Covers / Page 15).
Work steps
n Disconnect cables from U2
n Unscrew U2 and replace it with a new one
Fig. 37: Remove U2
(1) U2
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 50 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 51
4.3.9 T1
Transformer T1, T2, and power filter Z1 are located at the front of main unit chassis.
Fig. 38: Front view of main unit chassis
2
4
(1) T1
(2) T2
(3) Filter Z1
(4) Plastic caps
Prerequisites
n Switch off the system.
n Remove the left and right covers as described in “Main Unit Covers” ( Main Cov‐
ers / Page 15).
n Remove the front cover as described in “Main Unit Covers” ( Main Covers / Page 15).
Work steps
n Disconnect cables from T1
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 51 of 155 Cios Select /Cios Select FD
07.18
52 4 Main unit
T1 weights 24 kg, please be careful when removing T1 from the metal rod.
Do not discard the plastic caps and ensure they are installed after replacement.
Do not forget to apply LOCTITE 243 (5507038) to secure the nut, and tightening torque
25Nm for to secure T1.
4.3.10 T2
Prerequisites
n Switch off the system.
n Remove the left and right covers as described in “Main Unit Covers” ( Main Cov‐
ers / Page 15).
n Remove the front cover as described in “Main Unit Covers” ( Main Covers / Page 15).
Work steps
n Disconnect cables from T2
n Unscrew T2 and replace it with a new one
Do not discard the plastic caps and ensure they are installed after replacement.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 52 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 53
4.3.11 Filter Z1
WARNING
Electrical voltage on power filter Z1
There is no protective cover on the connector of power filter Z1. Live voltage is
present at the Z1.
Even after the system switches off the power supply, electrical voltage is present.
Live lines can cause electric shock.
Risk of death, serious physical injury, or damage to property.
V
Deenergize the system.
If checks or measurements require working with a live line, cover all ener‐
gized parts to prevent unintentional contact.
Use extreme caution when working on live lines.
Use only measurement equipment and devices that are suited to the task.
Read and observe the warning labels attached next to conducting parts.
HZID_06H459443
Prerequisites
n Switch off the system.
n Remove the left and right covers as described in “Main Unit Covers” ( Main Cov‐
ers / Page 15).
n Disconnect power cable from system
n Remove the front cover as described in “Main Unit Covers” ( Main Covers / Page 15).
Work steps
n Disconnect cables from filter Z1
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 53 of 155 Cios Select /Cios Select FD
07.18
54 4 Main unit
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 54 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 55
If the limit switch is defective, turn off the system and manually drive the lifting motor
by applying an Allen wrench to the back of the motor.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 55 of 155 Cios Select /Cios Select FD
07.18
56 4 Main unit
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 56 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 57
n The position of the switch can be adjusted by loosening the 2 set screws and moving
the metal plate; tighten them after adjustment
Fig. 42: Loosen the two set screws.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 57 of 155 Cios Select /Cios Select FD
07.18
58 4 Main unit
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 58 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 59
n The position of the switch can be adjusted by loosening the 2 set screws; tighten
them after adjustment
n Switch the system on and move the lifting column to the highest and lowest position
to check the switch function.
n Reinstall the covers
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 59 of 155 Cios Select /Cios Select FD
07.18
60 4 Main unit
n Insert the maintenance rod (delivered together with the vertical drive) straight into
the two holes from left side to right side
Fig. 45: Insert the maintenance rod
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 60 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 61
n Then slowly lower the C- arm lifting column downwards until its bottom securely sits
on the rod; apply an Allen wrench to the back of the motor to drive it.
Fig. 46: Manually drive the motor
For older delivery systems with serial nos. 10000 to 10002, the motors cannot be driven
manually by applying an Allen wrench.
In this case, please refer to( Lifting C-arm manually by pipe wrench (P/
N:11145191) / Page 63) to manually lift the C-arm.
n Disconnect the energy chain by removing the 2 Allen screws and its cable from MCU
x152.
n Disconnect the motor power cable from the terminal bar.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 61 of 155 Cios Select /Cios Select FD
07.18
62 4 Main unit
Fig. 47: Remove energy chain Fig. 48: Disconnect the motor power cable.
n Unscrew the 2 screws connecting the vertical drive assembly to the lifting column and
remove the joint key.
n Remove the pin and pull the shaft back, then the vertical drive assembly can be re‐
moved.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 62 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 63
Fig. 49: Remove the 2 screws and joint key Fig. 50: Remove the vertical drive assembly.
2 1
2
1
(1) Remove the the pin.
(2) Draw back the shaft.
n Disassemble the rubber chain and install it on the new vertical drive assembly
n Reinstall the vertical drive assembly in the reverse order of removal
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 63 of 155 Cios Select /Cios Select FD
07.18
64 4 Main unit
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 64 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 65
Camera (spare part) with material no. 08888492 is applicable with system SW versions
VA10A and higher, while 10793970 is only applicable with system SW versions VA10E
and higher
Prerequisites
n Remove the I. I. cover as described in “Main Unit Covers” ( I.I. top covers (with han‐
dle) / Page 21).
When installing the optics, make sure that no dust or dirt contaminates the I.I. output
window or the optics.
4 3
5
(1) Camera
(2) Camera set screws (3)
(3) Clamping ring
(4) Clamping ring set screws (3)
(5) Optics
(6) Optics set screws (3)
Work steps
n Disconnect the plugs from the camera.
n Remove the three optics set screws ( 6/Fig. 52 Page 65).
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 65 of 155 Cios Select /Cios Select FD
07.18
66 4 Main unit
Camera (spare part) with material no. 08888492 is applicable with system SW versions
VA10A and higher, while 10793970 is only applicable with system SW versions VA10E
and higher.
Prerequisites
n Remove the I. I. cover as described in “Main Unit Covers” ( I.I. top covers (with han‐
dle) / Page 21).
Fig. 53: II camera
Work steps
n Disconnect all plugs from the camera.
n Loosen the three camera set screws ( 2/Fig. 52 Page 65) from the camera flange and
remove the camera.
After camera/optics replacement, make sure to avoid dust and immediately in‐
stall new camera/optics.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 66 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 67
4.5.3 Optics
Prerequisites
n Remove the I.I. top cover
Work steps
n Remove the camera from the optics as described under “Camera Replacement”
( Camera Replacement / Page 66).
n Loosen the three screws ( 6/Fig. 52 Page 65) for the optics and remove the optics
from the I.I.
n Install the new optics and camera in the reverse order of their removal.
n Continue with ( Optical Sharpness, Image Rotation, Centering, Dose TV Iris and IQAP
test / Page 72).
When installing the optics, make sure that no dust or dirt contaminates the I.I. output
window or the optics
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 67 of 155 Cios Select /Cios Select FD
07.18
68 4 Main unit
3 2
Work steps
n Remove the Allen screw ( 1/Fig. 54 Page 68) and remove the bracket.
n Remove the three Allen screws ( 2/Fig. 54 Page 68) (behind the bracket).
n Disconnect all the cables from the I.I. mini power supply.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 68 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 69
n Install the new I.I. mini power supply in the reverse order of its removal.
Adjustment
n Pull out the I.I. mini power supply.
If some cables are too short, disconnect them temporarily.
n Place the I.I. mini power supply ( Fig. 55 Page 68) off to the side.
DIP switches 1 2 3 4
Normal status ON OFF OFF OFF
Adjustment E1 (U1) ON OFF ON OFF
Adjustment E2 (U2) ON ON OFF OFF
Adjustment E3 (U3) ON ON ON OFF
n Adjust the voltages of the I.I. mini power supply with the potentiometer
( Fig. 57 Page 69) if necessary.
Do not make readjustments in cases where there are only slight deviations
from the values on the test protocol (measurement device tolerances).
n Check Optical sharpness, Dose TV Iris, and perform the IQAP test. Refer to ( Optical
Sharpness, Image Rotation, Centering, Dose TV Iris and IQAP test / Page 72).
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 69 of 155 Cios Select /Cios Select FD
07.18
70 4 Main unit
When installing the optics, make sure that no dust or dirt contaminates the I.I. output
window or the optics.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 70 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 71
Work steps
n Remove the I.I. power supply as described under “I.I. power supply” ( I.I. Power Sup‐
ply / Page 67).
n Remove the cable clamps and disconnect the C-arm cables at the I.I.
n Remove the rubber stop in front of the I.I. mount and the two Allen screws
( Fig. 58 Page 70).
The I.I. is mechanically detached from the C-arm.
n Remove the camera and optics, as described under “Camera including Optics” ( Cam‐
era including optics / Page 65).
n On the new I.I., install the camera and optics, as described under “Camera including
Optics”( Camera including optics / Page 65).
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 71 of 155 Cios Select /Cios Select FD
07.18
72 4 Main unit
Be sure not to pinch the C-arm cables when tightening the I.I. screws.
T Apply the Loctite 221 to the 2 I.I. attachment screws and tighten them with 10
Nm of torque.
n Reconnect all plugs.
n Continue with ( Optical Sharpness, Image Rotation, Centering, Dose TV Iris and IQAP
test / Page 72).
n Reinstall and seal the cover
n Reinstall the rubber stop
4.5.6 Optical Sharpness, Image Rotation, Centering, Dose TV Iris and IQAP
test
n Centering adjustment: Loosen the three optics set screws( 6/Fig. 52 Page 65) and
move the camera with optics into the I.I. center by using the ( Diagnostic Tool for
Centering / Page 74).
n Optical sharpness adjustment:
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 72 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 73
After adjusting the optical sharpness, it is necessary to check / adjust the image rotation.
Loosen the three clamping ring set screws ( 4/Fig. 52 Page 65) to release the
clamping ring ( 3/Fig. 52 Page 65)
Slightly loosen the camera set screws( 2/Fig. 52 Page 65).
Adjust for maximum optical sharpness by slightly moving the camera sleeve
upwards or downwards to change the camera focus.
Tighten the camera set screws when the adjustment is finished.
Attach the clamping ring up to the camera and tighten its 3 set screws
n Image rotation: Loosen the three optics set screws ( 6/Fig. 52 Page 65), rotate the
camera until the image orientation is correct, and then tighten the Allen screws again.
Use a soldering wire or similar on the I.I. surface for image rotation orientation.
n Dose TV Iris: Perform “Dose TV Iris” as specified in the chapter “Dose Adjustment” in
the Troubleshooting Guide.
Service -> Tune up -> System Adjustment -> Dose TV-iris
n IQAP tests
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 73 of 155 Cios Select /Cios Select FD
07.18
74 4 Main unit
1
2
2
Central beam
Service -> FLC -> Diagnostic -> Image Diagnostics
n Select “Fluoro” ( 1/Fig. 61 Page 74) and click “Start” ( 2/Fig. 61 Page 74).
The “Blanking circle” window appears.
n Mark the checkbox “Check II centering” ( 1/Fig. 62 Page 74) and uncheck “Blanking
circle off” ( 2/Fig. 62 Page 74).
The centering lines are visible in the image area
The blanking circle disappears.
Fig. 63: Centering example
1
2
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 74 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 75
n Check that if the I.I. edge is centered with the Blanking circle.
If necessary, loosen the three optics set screws and move the camera with op‐
tics to adjust the centering.
Retighten the 3 optics set screws
n Check that the collimator blades are equal at the I.I. edge.
If necessary, slightly loosen the two tube attachment screws and tilt the tube
to adjust it
Retighten the tube screws with 10 Nm of torque.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 75 of 155 Cios Select /Cios Select FD
07.18
76 4 Main unit
1 2
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 76 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 77
3
1 2
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 77 of 155 Cios Select /Cios Select FD
07.18
78 4 Main unit
The laser lights cannot be repaired; therefore the complete light localizer including laser
light must be replaced.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 78 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 79
4.6.1 FD
Work Steps
n Remove FD bottom cover
n Turn C-arm to position 180°deg Angulation (vertically and FD downside)
n Loose the FD attachment screws, disconnect the Ethernet cable and signal cable
n Unwind the grounding cable, move FD out and then disconnect them to remove FD
n Install the FD in the reverse order of removal
n Perform FD installation (including detector drive disc and VCP disc), FD calibration,
Dose DEXI, Centering and IQAP tests
4.6.2 FD laser
Work Steps
n Remove FD bottom cover
n Turn C-arm to position 180°deg Angulation (vertically and FD downside)
n Untie and disconnect FD laser cable
n Loosen the attachment screw to remove the FD laser
n Install the FD laser in the reverse order of removal
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 79 of 155 Cios Select /Cios Select FD
07.18
80 4 Main unit
n Adjust the position in the adjustment hole so that two laser beams cross over at the
red point of the monoblock cover
2 1
4.6.3 Grid
Work steps
n Loosen the screw and slide the grid out
n Slide grid in and tighten the screw
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 80 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 81
4.7.2 Collimator
Work steps
n Position the C-arm at 0 °deg vertically, block all movement brakes
When removing C-arm parts, place an object under the rear chassis to keep the main
unit from tilting unexpectedly ( Fig. 97 Page 101).
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 81 of 155 Cios Select /Cios Select FD
07.18
82 4 Main unit
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 82 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 83
For systems with serial numbers 10382 and lower, the collimator must be configured af‐
ter it has been replaced.
Work steps
1. Download the Zip file from the CB-DOC section on the intranet.
https://productinformation.healthcare.siemens.com/ProductInformation.aspx?
uuid=d4e86517-ca1b-4c10-a1d9-a4df6a4be7b4
or
CS / Product Information / AT / Cios Select / Software
2. Extract the Zip files, and copy them to a USB stick or burn them to a CD.
3. Insert the USB stick or CD into the FLC.
4. Log in to the Service Software.
5. Press <Ctrl> and then <Esc> and keep them pressed simultaneously to open the Win‐
dows Start menu.
6. Click Windows Explorer, and then navigate to the USB stick or CD.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 83 of 155 Cios Select /Cios Select FD
07.18
84 4 Main unit
When you click the Replace button, the old file is backed up and replaced with
a new one
When you click the Restore button, the old file that was backed up in the “re‐
place” process is restored.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 84 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 85
8. Click the Replace button, and then confirm each pop-up window (“Successfully...”)
one by one.
Click OK to confirm the pop-up window for the successful backup of the file
“ccu.cdf”.
Fig. 73: CCU backup
Click OK to confirm the pop-up window for the successful replacement of the
file “ccu.cdf”.
Fig. 74: CCU.cdf replaced
Click OK to confirm the pop-up window for the successful backup of the file
“system.cdf”.
Fig. 75: system cdf backup
Click OK to confirm the pop-up window for the successful replacement of the
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 85 of 155 Cios Select /Cios Select FD
07.18
86 4 Main unit
file “system.cdf”.
Fig. 76: system cdf replaced
If a backup already exists for any file, a “...won’t backup...” pop-up window will appear.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 86 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 87
11. Enter Configuration. Click Config Check. The checking process starts automatically.
If any errors are found, select Auto Conf to correct the configuration.
Fig. 78: Config check result
12. When the prompt "No errors detected" is displayed, click Exit.
13. Exit the Service software.
14. Reboot the system.
After the single-tank tube is replaced, the filament current heating (IH) will need to be
adjusted. Perform the adjustment with an oscilloscope.
After the single-tank tube is replaced, the NTC temperature will need to be adjusted. It is
recommended to use a thermometer to perform the adjustment.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 87 of 155 Cios Select /Cios Select FD
07.18
88 4 Main unit
Work steps
n Position the C-arm at 90 °deg vertically, block all movement brakes
When removing C-arm parts, place an object under the rear chassis to keep the main
unit from tilting unexpectedly ( Fig. 97 Page 101).
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 88 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 89
n Remove the monoblock cover as described in “Main Unit Covers” ( Monoblock cov‐
er / Page 24).
Fig. 80: Single tank
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 89 of 155 Cios Select /Cios Select FD
07.18
90 4 Main unit
n Place the new X-ray tube unit on the guide bolts and secure it with both Allen screws.
When doing this, use one hand or ask a second person to hold the tube unit.
Be sure not to pinch the C-arm cables when tightening the tube screws.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 90 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 91
2. Apply uniform horizontal beads of silicone glue to the left and right sides of the
monoblock, and uniform vertical beads of glue to the back. The beads of glue should
just cover the edges. Apply no glue to the front.
Fig. 82: Apply glue to left and right sides Fig. 83: Apply glue to back
3. Install the cover on the monoblock, and tighten the two screws on the back.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 91 of 155 Cios Select /Cios Select FD
07.18
92 4 Main unit
4. Apply uniform horizontal beads of silicone glue to the back of the cover.
Fig. 84: Apply glue to cover
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 92 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 93
5. Carefully smooth the silicone glue and remove the excess with your finger.
Wipe the excess silicone glue off your finger with a paper towel at regular intervals.
Smoothing the bead is easier when your finger is kept clean.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 93 of 155 Cios Select /Cios Select FD
07.18
94 4 Main unit
7. Check the appearance of the silicone glue on the monoblock and the cover.
Fig. 86: Silicon glue appearance
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 94 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 95
4.8 Mechanics
Fig. 87: Loosen the chain Fig. 88: Remove the handle.
2
(1) Lock nuts
(2) Adjuster
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 95 of 155 Cios Select /Cios Select FD
07.18
96 4 Main unit
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 96 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 97
1
2
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 97 of 155 Cios Select /Cios Select FD
07.18
98 4 Main unit
3
2
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 98 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 99
The Adjustment Nut should be nearly in the middle of the two pulleys on the left and
right sides of the chassis.
The tension force of three segments of chain is 20N +/- 2 N; the chain displacement is 3
mm.
Adjustment nut 1 is under the base frame. Measure the chain tension force at adjust‐
ment nut 2 and 3. It is sufficient to check the chain tension force at adjustment nut 1 by
hand.
Carefully and slowly remove the front chain, in order to prevent the left and right chains
from dropping off the bearings, which may then require much more effort to reinstall
and adjust them to the correct positions.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 99 of 155 Cios Select /Cios Select FD
07.18
100 4 Main unit
n Lift the main unit up on one side with the rear handles.
Fig. 94: Put steering handle in middle position Fig. 95: Pull up the main unit
Do not remove the back cover to pull up the main unit, otherwise the top plate of the
main unit may get bent.
n Put a wooden block or screw jack under the chassis near the main wheel and keep the
wheels in the 0°deg orientation.
n Lift up the other side and put the wooden block or screw jack under the chassis near
the other main wheel to keep the wheels in the 0°deg orientation.
n Now you can access the chain from below
Put an object under the rear chassis to keep the main unit from tilting unexpectedly
( Fig. 97 Page 101).
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 100 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 101
Fig. 96: Put on the wooden blocks. Fig. 97: Chassis support
n Loosen the lock nuts and open the chain by turning the adjustment nut.
n Draw out the left / right chain.
n Replace the chain with a new one, reinstall the chain in the reverse order of removal
and according to the chain layout
n Turning the adjustment nut to adjust the chain tension to 3 mm with 20+/-2 N.
The normal procedure to adjust is to tighten adjustment nut 1 and lock it via its lock
nuts, and then to adjust adjustment nuts 2 and 3, in order to ensure that both main
wheels maintain the 0°deg position. If the main wheels deflect a little, then tighten ad‐
justment nuts 2 and 3 with a little more force, then loosen adjustment nut 1 a little bit,
then tighten the lock nuts.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 101 of 155 Cios Select /Cios Select FD
07.18
102 4 Main unit
Fig. 98: Tap out the pedal axle. Fig. 99: Remove the pedal and its base.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 102 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 103
2
1
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 103 of 155 Cios Select /Cios Select FD
07.18
104 4 Main unit
If the ends of the hex head rod (pedal axle) are heavily oxidized and therefore really diffi‐
cult to remove from the foot brake pedal, completely remove the hex head rod (pedal
axle) and replace it with the new one provided in the delivery.
If the hex head rod is still in good condition and easy to remove, then it does not need to
be replaced and can be reused.
n If needed (see note), completely remove / tap out the hex head rod and replace it with
the one provided in the delivery.
WARNING
Risk of injury!
The basic unit is extremely heavy, and if tilted too far to the side, may tip over
completely and cause serious bodily injury or damage to property.
Tilt the basic unit very carefully and only as far on its side as you need to in
order to move the wooden block under the side beam.
With two people, tilt the basic unit.
A third person is needed to move the wooden block under the side beam of
the transport frame. Take hold of the wooden block on its sides only and place
it under the beam; do not put your hands, arm or fingers under the side beam.
Place the basic unit down on the wooden block.
WARNING
Risk of injury!
When placing the wooden block underneath the beam, take great care to keep
your fingers, hand or arm from getting crushed.
When placing the wooden block under the side beam of the transport frame,
take hold of the wooden block only on its sides and move it under the beam.
Do not put your fingers, hands or arms under the side beam.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 104 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 105
11
10
8
1
6
2
3 4 5
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 105 of 155 Cios Select /Cios Select FD
07.18
106 4 Main unit
n Remove the caps and the spring rings, then pull out the main wheel axle and remove
the main wheel
n Remove the foot brake rod
n Loosen the wheel chain and pull the chain off the toothed gear
n Then remove the spring ring to remove the wheel holder welding
4.8.3.4 Cleaning and lubricating the sleeve bearing and removing paint damage
n Clean rust and deposits off the sleeve bearing. To do this, you can use fine sandpaper
and a nail file. Make sure you do not widen the diameter of the sleeve bearing, just
clean it.
n Remove dirt residue.
n In case of paint damage and surface rust on the transport frame itself, use water to
remove as much rust as possible, let it dry and then spray paint it. For paint work
around the sleeve bearing, cover it up first to avoid paint getting into it.
Do NOT paint the existing grounding bolts (potential balance) on the transport frame. If
necessary, put tape on the grounding bolts first and then remove it again afterwards.
For those systems on which aggressive surface disinfectants have been used, surface
paint damage here and there and heavy instances of rust may appear on the transport
frame.
Any extensive rust removal measures and paintwork have to be approved beforehand by
the customer.
Complete elimination of oxidation damage is, in such cases, no longer possible.
Refurbishing such areas to perfect condition is possible, but requires great time and ef‐
fort.
Further intensive use of aggressive surface disinfectants can cause renewed occurrences
of oxidation damage.
4.8.3.5 Install the new main wheels and foot brake assembly
n Reinstall the wheel and foot brake assembly in the reverse order of removal
Make sure the 2 shim rings are installed between the wheel and wheel holder welding
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 106 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 107
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 107 of 155 Cios Select /Cios Select FD
07.18
108 4 Main unit
WARNING
Electrical voltage after power-off
Even after the system switches off the power supply, electrical voltage is present
in the system.
Short-circuiting of live lines can cause electric shock or mechanical damage.
Risk of death, serious physical injury, or damage to property.
WARNING
Electrical safety!
Non-compliance can lead to severe bodily injury or even death, as well as proper‐
ty damage.
After the cover panels are opened, parts under voltage are accessible. To
avoid danger, disconnect the system from the power supply before opening
the covers.
V
An uninterruptible power supply (UPS) is installed in the system. Unplug the
power input connector of the imaging system or switch off the UPS prior to
beginning service work.
If steps have to be performed using electrical power, the general safety infor‐
mation in “Medical Products; Safety Information; General Safety Notes”
(TD00-000.860.01.xx) must be observed.
HZID_06H384658
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 108 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 109
CAUTION
Laser emissions!
This product contains lasers.
L
Disregarding safety precautions can lead to bodily injury, especially to the retina
of the eye, resulting in irreversible damage to vision.
Do not look into the laser diodes (I.I. side or tube side)
Observe local safety guidelines regarding laser emissions.
HZID_06H383909
4.8.5.1 C-arm
The C-arm is not a spare part, but has to be removed and installed when replacing orbital
movement control parts.
When removing C-arm parts, place an object under the rear chassis to keep the main
unit from tilting unexpectedly ( Fig. 97 Page 101).
a) Remove one of the C-arm handles, which may collide with the supporting trolley
or frame in the next step.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 109 of 155 Cios Select /Cios Select FD
07.18
110 4 Main unit
b) Position the C-arm carefully at about 0°deg horizontally, use supporting trolley or
frame to secure it against dropping, then block all brakes to secure the C-arm
against unintended movement.
c) Remove the I.I. handle and I.I. cover, disconnect all cables from the chassis, then
re-install the I.I. cover to keep out dust and prevent unintended collision
d) Remove the rubber stop and loosen the two fixing Allen screws
( Fig. 58 Page 70), detach the I.O.C. from the C-arm carefully.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 110 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 111
e) Open I.I. cover and carefully remove C-arm cables from I.O.C.
Be sure not to pinch the C-arm cables when tightening the I.I. and tube screws.
T Apply Loctite 221 to the 2 I.I. screws and tighten them with 10 Nm of torque.
T Tighten the 2 tube screws with 10 Nm of torque
4. Function check:
- Ensure C-arm orbital movement works properly.
- Smooth movement of C-arm without any blocking point throughout the entire
orbital range
- Max. torque ≤ 30 Nm for C-arm manual orbital movement using the C-arm
handles with the brake released
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 111 of 155 Cios Select /Cios Select FD
07.18
112 4 Main unit
4.8.5.2 Rollers
Work steps
1. Remove C-arm
2. Loosen the fixing nuts to remove the rollers
3. Replace with the new rollers and tighten the nuts to secure the rollers
1
1
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 112 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 113
- After all C-arm parts are installed, adjust the eccentric roller to ensure max. torque
≤ 30Nm for C-arm manual orbital movement using the C-arm handles with the
brake released
Fig. 108: Brake and pre-friction screws Fig. 109: Brake screw
1
2
4. Replace with the new brake and tighten the screw to secure it.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 113 of 155 Cios Select /Cios Select FD
07.18
114 4 Main unit
5. After all the C-arm parts are installed, adjust the brake by loosening and tightening
the screw to ensure
- Max. torque ≤ 30 Nm for C-arm manual orbital movement using the C-arm han‐
dles with the brake released
- Min. torque ≥ 60 Nm for C-arm manual orbital movement using the C-arm handles
with the brake triggered.
- Smooth movement of C-arm without any blocking point throughout the entire or‐
bital range
- No abnormal noise during orbital movement
Fig. 110: Orbital force
4.8.5.4 Pre-friction-asm
Work steps
1. Remove C-arm
2. Remove the C-arm cable holder plate( Fig. 107 Page 113)
3. Loosen the fixing screw to remove the pre-friction-asm( 2/Fig. 108 Page 113)
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 114 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 115
4. Replace with the new pre-friction-asm, insert the spring, washer, and screw in se‐
quence, then tighten the screw.
5. Without twisting it, position the C-arm cable outside of C-arm, and use a finger to
push against the pre-friction-asm when sliding in the C-arm
6. After all the C-arm parts are installed, adjust pre-friction-asm by loosening and tight‐
ening the screw to ensure
- Max. torque ≤ 30 Nm for C-arm manual orbital movement using the C-arm han‐
dles with the brake released
- No obvious unintended movement by C-arm itself, < 5°deg at normal position, <15
°deg at limited position.
- Smooth movement of C-arm without any blocking point throughout the entire or‐
bital range
- No abnormal noise during orbital movement
7. Reinstall the C-arm cable holder plate
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 115 of 155 Cios Select /Cios Select FD
07.18
116 4 Main unit
Fig. 113: Angular force fixing nut and adjustment screw Fig. 114: Angular force adjustment screw
Work steps
1. Remove the rubber cap.
2. Loosen the fixing nut.
3. Turn the adjustment screw clockwise to tighten it for greater force, and counterclock‐
wise to loosen it for less force.
4. Test the regular movement force.
Acceptance criteria:
- Max. torque ≤ 25 Nm for C-arm manual angular movement using the C-arm han‐
dles with the brake released and C-arm in 0 position
- Min. torque ≥ 55 Nm for C-arm manual angular movement with the brake trig‐
gered and C-arm in 180 position
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 116 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 117
1 2
Work steps
1. Remove the rubber cap.
2. Loosen the fixing nut.
3. Turn the adjustment screw clockwise to tighten it for greater force, and counterclock‐
wise to loosen it for less force.
4. Test the regular movement force with C-arm in 90 position.
Acceptance criteria:
- Max. torque ≤ 25 +/-5 Nm for manually swiveling the C-arm (left-right turn) using
the C-arm handles with the brake released
- Min. torque ≥ 60 Nm for manually swiveling the C-arm (left-right turn) using the
C-arm handles with the brake triggered
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 117 of 155 Cios Select /Cios Select FD
07.18
118 4 Main unit
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 118 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 119
4.9 Cables
After the C-arm cable is replaced, the filament current heating (IH) will need to be adjus‐
ted. Perform the adjustment with an oscilloscope.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 119 of 155 Cios Select /Cios Select FD
07.18
120 4 Main unit
Fig. 118:
Work steps
n Switch off the system.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 120 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 121
a) Remove one of the C-arm handles, which may collide with the supporting trolley
or frame in the next step.
b) Position the C-arm carefully at about 0°deg horizontally, use supporting trolley or
frame to secure it against dropping, then block all brakes to secure the C-arm
against unintended movement..
c) Remove the I.I. handle and I.I. cover, disconnect all cables from the chassis, then
re-install the I.I. cover to keep out dust and prevent unintended collision.
d) Remove the rubber stop, loosen the two Allen screws ( Fig. 58 Page 70), and de‐
tach the I.O.C. from the C-arm carefully.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 121 of 155 Cios Select /Cios Select FD
07.18
122 4 Main unit
e) Open the I.I. cover and carefully remove the C-arm cables from the I.O.C.
f) Re-install the I.I. cover to protect it, then remove the I.O.C. component.
n Remove the cable holders from the C-arm and the main unit.
n Pull the cables from the I.I./FD and single tank through the C-arm
For VA10, the cable cannot be pulled out of C-arm directly with connector housing. Re‐
move the connector housing before pulling the cable out. Take a photo in advance be‐
cause each connector has a front- and back-facing direction.
Be sure not to pinch the C-arm cables when tightening the I.I. and tube screws.
Apply Loctite 221 to the two I.I. screws and tighten them with 10 Nm of torque. Tight‐
en the two tube screws with 10 Nm of torque.
n Perform a test of system functions
n Perform the filament current heating (IH) adjustment following the instructions in the
online help.
n Perform IQ test
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 122 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 123
4.10 CCP
Fig. 121:
Fig. 122:
Work steps
n Switch off the system.
n Open the left and right covers of the main unit
n Disconnect the cables from the CCP
n Unscrew 8 nuts from the CCP
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 123 of 155 Cios Select /Cios Select FD
07.18
124 4 Main unit
n Reconnect the cable and reassemble the system in the reverse order of removal
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 124 of 155 © Siemens Healthineers, 2014
07.18
Main unit 4 125
Fig. 123: TUI CCP attachment screws Fig. 124: TUI CCP replacement
n Push and lift TUI CCP with AIC board and speaker to remove them from Main Unit
- Disconnect the cable and unscrew the attachment screws to remove the speaker.
- Disconnect the cables and unscrew the attachment screws to remove the AIC
board.
- Remove the old gasket if needed
n Attach the new gasket
n Install the TUI CCP, AIC board and speaker in the reverse order of removal
n Perform a TUI CCP adjustment
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 125 of 155 Cios Select /Cios Select FD
07.18
126 5 Trolley Covers
Work steps
n Remove 4 screws from the front cover
n Pull the cover out and unscrew the grounding cable.
n Remove the front cover.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 126 of 155 © Siemens Healthineers, 2014
07.18
Trolley Covers 5 127
Work steps
n Remove 6 screws of rear cover.
n Pull the cover out and unscrew the grounding cable.
n Remove the rear cover.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 127 of 155 Cios Select /Cios Select FD
07.18
128 6 Trolley
6.1 Trolley
6.1.1 SG cable
WARNING
Electrical voltage
Non-compliance can lead to severe bodily injury or even death, as well as proper‐
ty damage.
After the cover panels are opened, parts under voltage are accessible. To
avoid danger, disconnect the system from the power supply before opening
the covers.
An uninterruptible power supply (UPS) is installed in the system. Unplug the
power input connector of the imaging system or switch off the UPS prior to
beginning service work.
If steps have to be performed using electrical power, the general safety infor‐
mation in “Medical Products; Safety Information; General Safety Notes”
(TD00-000.860.01.xx) must be observed.
HZID_06H384658
Prerequisites
n Switch off the system.
n Unplug the main power connection and the SG cable from the main unit
Work steps
n Remove the trolley rear cover
n Remove the SG cable entrance cover
Fig. 127: Trolley
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 128 of 155 © Siemens Healthineers, 2014
07.18
Trolley 6 129
6.1.2 PC
See CB DOC.
The label on the PC housing only contains information about the FLC that pertains to fac‐
tory delivery. The label does not contain any information about the PC itself. Please refer
to the SPC for details about PC spare part material numbers. For example, in the SPC,
Cios Select > Trolley Unit > FLC S3 I.I. 10862072, 10860674 is the actual spare part ma‐
terial number for the W530 PC.
It will be necessary to apply for new licenses and service keys when the PC is replaced.
After the PC is replaced, “Set Wake On LAN” and “Set On Time” will need to be adjusted.
6.1.3 UPS
WARNING
Electrical voltage
Non-compliance can lead to severe bodily injury or even death, as well as proper‐
ty damage.
After the cover panels are opened, parts under voltage are accessible. To
avoid danger, disconnect the system from the power supply before opening
the covers.
An uninterruptible power supply (UPS) is installed in the system. Unplug the
power input connector of the imaging system or switch off the UPS prior to
beginning service work.
If steps have to be performed using electrical power, the general safety infor‐
mation in “Medical Products; Safety Information; General Safety Notes”
(TD00-000.860.01.xx) must be observed.
HZID_06H384658
Work steps
n Switch off the system.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 129 of 155 Cios Select /Cios Select FD
07.18
130 6 Trolley
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 130 of 155 © Siemens Healthineers, 2014
07.18
Trolley 6 131
n Install the new UPS in the reverse order of removal, but do not connect the power
plugs to the UPS.
Use the “enter” button get into the service menu and confirm the selected item.
Use the “up/down” buttons to move up and down in the menus.
Use the “ESC” button to go back to the start menu.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 131 of 155 Cios Select /Cios Select FD
07.18
132 6 Trolley
If the UPS menu language and voltage have already been initially confirmed, the menu
can be found under the following paths:
Menu language: Setting > Local Settings > Language
Output voltage: Settings > In/Out settings > Output voltage
UPS Configuration
n Initial programming of the UPS language and output voltage
- Connect the UPS to 220 V AC and switch on the UPS
- The symbol “~” lights green
- The language selection appears on the menu
- Confirm language English with the “Enter” button
- The voltage selection appears on the menu
- Confirm the preselected voltage “220V” with the “Enter” button
Confirm the message “Please register” with the “Enter” button
n Programming UPS sensitivity
- With the up/down buttons select the “Settings” menu and confirm with “Enter”
- With the up/down buttons select the “In/Out settings” submenu and confirm with
“Enter”
- With the up/down buttons select the “Sensitivity” description and confirm with “En‐
ter”
- With the up/down buttons set “Low” and confirm with “Enter”
- Return with “ESC” to the “Settings” menu. (press twice)
n Programming the “Energy saving” mode
- Select the “ON/OFF settings” submenu and confirm with “Enter”
- Select “Energy saving” with the up/down buttons and confirm with “Enter”
- Select “Load” with the up/down buttons and confirm with “Enter”
- With the up/down buttons and confirm with “Enter”
- Select “Time” with the up/down buttons and confirm with “Enter”
- With the up/down buttons, set the value to 5 min and confirm with “Enter”
- Return to the start menu by pressing the ESC button several times.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 132 of 155 © Siemens Healthineers, 2014
07.18
Trolley 6 133
WARNING
Electrical voltage
Non-compliance can lead to severe bodily injury or even death, as well as proper‐
ty damage.
After the cover panels are opened, parts under voltage are accessible. To
avoid danger, disconnect the system from the power supply before opening
the covers.
An uninterruptible power supply (UPS) is installed in the system. Unplug the
power input connector of the imaging system or switch off the UPS prior to
beginning service work.
If steps have to be performed using electrical power, the general safety infor‐
mation in “Medical Products; Safety Information; General Safety Notes”
(TD00-000.860.01.xx) must be observed.
HZID_06H384658
CAUTION
Risk of burns and of destroying the battery component due to possible high short-
circuit current.
Wearing metal objects such as rings, watches, and bracelets, and using tools with
uninsulated handles during procedures can lead to a short-circuit of the battery
contacts. This can destroy the battery component and heat up the metal objects,
causing a short-circuit at very high temperatures.
Before servicing the battery components, remove watches, rings, bracelets
and all other metal objects from the hands and arms. Use tools with insulated
handles.
Prerequisites
n Switch off the system.
n Remove the front trolley cover
n Switch off the UPS
n Remove the UPS from the trolley
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 133 of 155 Cios Select /Cios Select FD
07.18
134 6 Trolley
Work steps
n Remove the front panel of the UPS
Fig. 130: Remove battery - 1
n Disconnect the battery module by separating the two connector parts (never pull on
the wires)
Fig. 131: Remove the battery
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 134 of 155 © Siemens Healthineers, 2014
07.18
Trolley 6 135
n Pull the plastic tab to remove the battery block and replace it.
Fig. 133: Remove the battery
Take care to firmly press together the two parts of the connector during remounting
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 135 of 155 Cios Select /Cios Select FD
07.18
136 6 Trolley
CAUTION
Hot surface
Risk of burns
When replacing the trolley control panel (TCP), the back surface of the TCP
may be hot. Handle with care!
Prerequisites
n Switch off the system.
n Remove the trolley rear cover
Work steps
n Disconnect the TCP cable
n Remove 6 bolts of TCP
- 3 bolts can be reached from the trolley front
Fig. 134: Trolley front view
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 136 of 155 © Siemens Healthineers, 2014
07.18
Trolley 6 137
n Reconnect the cable and reassemble the system in the reverse order of removal
n Perform a TUI TCP adjustment if applicable
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 137 of 155 Cios Select /Cios Select FD
07.18
138 6 Trolley
6.1.7 U3
Power supply U3 is located near the AIC board
Prerequisites
n Switch off the system.
n Remove the trolley rear cover
Work steps
n Disconnect the U3 cable
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 138 of 155 © Siemens Healthineers, 2014
07.18
Trolley 6 139
6.1.8 U4
Power supply U4 is located beside the UPS
Prerequisites
n Switch off the system.
n Remove the trolley rear cover
Work steps
n Disconnect the U4 cable
n Remove two attachment screws of U4
n Remove U4 from the trolley
n Install the new one in the reverse order of removal
6.1.9 Monitor
Prerequisites
n Switch off the system.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 139 of 155 Cios Select /Cios Select FD
07.18
140 6 Trolley
Work steps
n Disconnect the cables from monitor
n Remove the 2 Allen screws (lower position).
Loosen the 2 Allen screws (upper position)
Fig. 138: Right monitor - rear view Fig. 139: Left monitor - rear view
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 140 of 155 © Siemens Healthineers, 2014
07.18
Trolley 6 141
Work steps
n Remove 2 Allen screws to remove video splitter holder
Disconnect all cables from the video splitter
Unscrew the video splitter from the holder
Fig. 140: DVI video splitter Fig. 141: DVI video splitter
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 141 of 155 Cios Select /Cios Select FD
07.18
142 6 Trolley
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 142 of 155 © Siemens Healthineers, 2014
07.18
Trolley 6 143
Fig. 143: DVI adapter Fig. 144: DVI adapter (trolley left view)
6.1.11 Wheels
Lifting the monitor trolley requires two persons for a short time
Prerequisites
n Switch off the system.
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 143 of 155 Cios Select /Cios Select FD
07.18
144 6 Trolley
(1) Axle
(2) Lateral cover
(3) Wheel mount
(4) Cable deflectors
Work steps
n Activate the brakes of the monitor trolley
n Jack the trolley up on a service stamp, or use any other suitable object to lift trolley up
and set it down on the wooden block.
n Remove the axle
n Remove the lateral covering from the castor
n Remove the castor
n Unscrew the castor mount
n Install the new castor mount
When replacing a castor with direction lock, lock the direction lock before installation
and straighten the castor mount in travel direction.
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 144 of 155 © Siemens Healthineers, 2014
07.18
Trolley 6 145
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 145 of 155 Cios Select /Cios Select FD
07.18
146 6 Trolley
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 146 of 155 © Siemens Healthineers, 2014
07.18
Trolley 6 147
(1) Terminal XG
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 147 of 155 Cios Select /Cios Select FD
07.18
148 7 Fuses
WARNING
Electrical voltage after power-off
Even after the system switches off the power supply, electrical voltage is present
in the system.
Short-circuiting of live lines can cause electric shock or mechanical damage.
Risk of death, serious physical injury, or damage to property.
WARNING
Electrical voltage on power filter Z1
There is no protective cover on the connector of power filter Z1. Live voltage is
present at the Z1.
Even after the system switches off the power supply, electrical voltage is present.
Live lines can cause electric shock.
Risk of death, serious physical injury, or damage to property.
V Deenergize the system.
If checks or measurements require working with a live line, cover all ener‐
gized parts to prevent unintentional contact.
Use extreme caution when working on live lines.
Use only measurement equipment and devices that are suited to the task.
Read and observe the warning labels attached next to conducting parts.
HZID_06H459443
Main unit
n Switch off the system.
n Remove the stand frame cover as described under “Main Unit Covers”( Main Cov‐
ers / Page 15).
The following fuses can be replaced:
- F1: 12 AT @230 V AC
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 148 of 155 © Siemens Healthineers, 2014
07.18
Fuses 7 149
- F2: 2 AT @230 V AC
- F3: 2 AT @ 72 V AC
- F4: 10 AT @100 V AC
- F5: 3 AT @230 V AC
- F6: 2 AT @24 V AC
- F7: 4 AT @230 V AC
- F8: 15 AT @250 V AC (15 AT Fuse for voltage 240 V/230 V/220 V/200 V)
or 20 AT @250 V AC (20 AT Fuse for voltage 127 V/120 V/110 V/100 V)
- F9: 15 AT @250 V AC (15 AT Fuse for voltage 240 V/230 V/220 V/200 V)
or 20 AT @250 V AC (20 AT Fuse for voltage 127 V/120 V/110 V/100 V)
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 149 of 155 Cios Select /Cios Select FD
07.18
150 7 Fuses
1
2
3
4 5 6
(1) F1
(2) F2
(3) F3
(4) F9
(5) F8
(6) F4
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 150 of 155 © Siemens Healthineers, 2014
07.18
Fuses 7 151
1
2
(1) F6
(2) F5
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 151 of 155 Cios Select /Cios Select FD
07.18
152 7 Fuses
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 152 of 155 © Siemens Healthineers, 2014
07.18
Changes to Previous Version 8 153
01 Initial version.
02 Special requirement added for vertical drive, D621, I.I. replacement, I.O.C, adjustments for
optical sharpness, image rotation, centering, etc.
03 Update to new version D621, add C-arm orbital movement control parts, modify X-ray
tube, modify I.I. replacement
04 1). Note added regarding FLC label and PC spare part material number.
2). Section added regarding collimator tool.
3). Modifications made to test/adjustment cases and tools after parts replacement for
D610, D620, D621, single-tank tube, C-arm cable, FLC, etc.
4). Note added regarding license and service key re-application after FLC and MCU replace‐
ment.
5). Instructions added for C-arm movement force adjustment.
6). Instructions added for applying silicone glue to monoblock and cover.
etc.
05 1). HZID 06H537290 added.
2). Wheel chain instructions optimized.
3). C-arm cable instructions optimized.
4). Valid scope for two different (spare part) cameras added.
5). Table IQ tests (post-replacement) optimized.
06 To involve Cios, select FD (VA20)
© Siemens Healthineers, 2014 XPR2-140.841.01.06.02 Page 153 of 155 Cios Select /Cios Select FD
07.18
154 9 List of Hazard IDs
Cios Select /Cios Select FD XPR2-140.841.01.06.02 Page 154 of 155 © Siemens Healthineers, 2014
07.18