You are on page 1of 8

Design Considerations in Optimizing

a Breast omosynthesis System


Andrew Smith, Ph.D., Vice President - Imaging Science, Hologic

Introduction The Physics of Breast Tomosynthesis Imaging


Breast tomosynthesis, also re erred to as three-dimen- Conventional x-ray mammography is a 2D imaging
sional (3D) mammography, has been available or clinical modality, and as such, pathologies o interest are sometimes
use in numerous countries outside o the United States di cult to visualize. As shown in Figure 1, overlapping or
(U.S.) since 2009. In February 2011, the U.S. Food and superimposed tissues create a clutter o signals above and
Drug Administration (FDA) approved the Hologic Selenia® below the objects o interest. Tis occurs because the signal
Dimensions® 3D (breast tomosynthesis) system, making detected at a location on the digital detector is dependent
it the frst, and currently, the only commercially available upon the total attenuation o all the tissue between the
breast tomosynthesis system in the U.S. As such, interest in detector and the x-ray beam.
the area o breast tomosynthesis has grown tremendously. In a 2D image, overlapping tissue can hide objects o
Te FDA based its approval on Hologic’s clinical stud- interest, potentially resulting in missed cancers. Conversely,
ies, which showed that tomosynthesis imaging combined normal structures in the breast that overlap may give the
with conventional digital mammography results in superior appearance o lesions, leading to unnecessary recalls.
system per ormance compared to 2D mammography alone.1 omosynthesis is a 3D imaging modality that can
Tis paper looks at the physics o a breast tomosynthesis reduce or eliminate the tissue superimposition e ect.
system and discusses the parameters that Hologic considered
in designing the Selenia Dimensions 2D/3D system
(Dimensions).

Selenia Dimensions 2D/3D Technical Parameters


Parameter Value
Scan time < 4 seconds
Total scan angle 15 degrees
Number of projections 15
Acquisition modes 2D, 3D, Combo
(2D and 3D in one compression)
Figure 1. Overlapping structures can hide lesions in 2D that are clearly seen
kVp range 25-49 kVp in a 3D cross-sectional slice.
X-ray tube current 200 mA maximum
Reconstructed slice separation 1 mm Figure 2 shows the basic design o a breast tomosyn-
Reconstruction time 2-5 seconds thesis system. Te breast is compressed in the standard way
Reconstructed pixel size Approximately 100 microns between a compression paddle and the detector housing.
2D pixel size 70 microns While keeping the breast stationary, the x-ray tube is moved
in an arcuate motion, and a series o low-dose images,
Table 1. Technical parameters used in the Hologic Dimensions® known as projections, are taken at di erent angular
tomosynthesis system

1
The Hologic clinical studies presented to the FDA as part of Hologic’s PMA submission showed that 2D plus 3D imaging resulted in superior system performance compared to 2D mammography
alone. See http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfTopic/pma/pma.cfm?num=P080003
locations o the tube. Te total angular range covered by the • Provides ast scan time to reduce patient motion
x-ray tube is known as the scan angle. • Can acquire tomosynthesis images in all desired
Following the scan, the projections undergo a orientations (i.e. CC, MLO, LM, ...)
reconstruction process, which consists o computing high- • Provides rapid reconstruction to support
resolution images with planes parallel to the breast support screening patient throughput
plate. Te number o reconstructed slices will depend upon
• Facilitates comparison to priors
the thickness o the compressed breast and the desired
separation between slices, which is typically around 1 mm. • Supports tomosynthesis-guided interventions
Te reconstructed slices are then transmitted to a diagnostic and biopsies
workstation or review by a radiologist. • Keeps radiologist reading times short enough
to be suitable or screening

In addition to overcoming the e ects o superimposed


tissue, the optimal tomosynthesis system must also provide
the image quality needed to properly visualize calcifcations,
masses, architectural distortions and asymmetries.
Although the highest value or tomosynthesis may be in
screening applications, the optimum tomosynthesis system
must also work well in diagnostic imaging. Te system
must allow all standard conventional diagnostic – as well as
tomosynthesis – imaging. Conventional diagnostic imaging
is needed because some views, such as magnifcation views,
cannot be done with tomosynthesis and may still be needed
or patient work-ups. In addition, an ideal breast tomosyn-
thesis system must be able to support tomosynthesis-guided
interventional procedures, as some abnormalities that require
Figure 2. Schematic showing principle of operation of tomosynthesis system. biopsy may only be visible in the tomosynthesis mode.
Te ollowing sections describe di erent design
Selection of System Design Parameters parameters and how they a ect the unctionality o a
Te design o a tomosynthesis system involves the tomosynthesis system.
optimization o many di erent parameters. First and
oremost, a breast tomosynthesis system must success ully Scan Time
reduce the e ect o overlapping tissue. O equal impor- Scan time is probably the single most important
tance, the design criteria should be balanced to ensure an per ormance metric or a breast tomosynthesis system.
end product that unctions properly in both screening and Scan times must be short or two key reasons: to sup-
diagnostic applications. Design parameters or an optimally port a reasonable patient throughput and more importantly,
e ective system should satis y the ollowing requirements: to reduce blurring caused by patient motion. Currently,
no tomosynthesis system is capable o matching the short
• Works well or screening imaging scan times associated with conventional mammography.
(asymptomatic women) Even mammography has occasional cases o patient motion;
there ore, it is expected that patient motion will be present
• Works well or diagnostic imaging
in some tomosynthesis cases. Tis problem is exacerbated
(symptomatic patients)
or tomosynthesis scans longer than a ew seconds.
• Supports both conventional 2D imaging Trough the evolution o its tomosynthesis technology,
and tomosynthesis Hologic has seen a signifcant reduction in the incidence
• Improves clinical per ormance over digital o patient motion, rom 18 seconds in its early prototype
mammography alone to 10 seconds in the system used or the clinical trial to
3.7 seconds in its commercially released Hologic Selenia Another issue with wide scans angles is the potential
Dimensions system. 2 Reduction o scan time to an absolute or the edge o the compression paddle to inter ere with the
minimum is there ore critical, as patient motion as small beam path, as seen in Figure 4. While this can be solved by
as 0.1 mm can degrade the sharpness o microcalcifcations making the compression paddle wider than the detector, it
and tumor spiculations. Figure 3 demonstrates the e ect o can lead to breast positioning issues. However, by ar the
patient motion on microcalcifcations. most signifcant disadvantage o a wide-angle scan is the
di culty in keeping the scan times short enough to reduce
the likelihood o patient motion and to support screening
throughput.

Figure 3. Effect of patient motion on sharpness of microcalcifications

Scan Angle
Tere are a number o actors that must be considered
in the selection o the scan angle to be used in a breast
tomosynthesis system. Determining the optimum scan angle Figure 4. The edge of the compression paddle can interfere with projections at
should be based on which actors most e ectively support wide scan angles.
system design goals. Te scan angle or a system designed
or both screening and diagnostic applications must support Field of View Versus Scan Angle
rapid scan times, while preserving the ability to e ectively Te feld o view (area that is visible) is a ected by scan
visualize microcalcifcation clusters and masses. angle. With a stationary detector, shadows o some parts o
Narrow scan angles allow the most rapid scans. How- the breast will miss the detector when large scan angles are
ever, because the depth o feld decreases as the scan angle used, reducing feld o view. It will always be possible or
increases3, they may not reduce overlapping structures as tissue to be outside o the feld o view, even in 2D imaging;
much as wider scans. In general, as the scan angle increases, however, the raction o additional missed tissue is
the out-o -plane resolution is superior; however, the trade- insignifcant or shallow projection angles and increases
o is a degradation o the in-plane resolution. to signifcant levels or wide scan angles.
As scan angles increase, con ounding e ects o super- Figure 5 illustrates the feld o view or a 50° total scan
imposed tissue might decrease, but microcalcifcations will angle compared to the 0° stationary conventional mammo-
appear less sharp due to the degradation o the in-plane graphic acquisition. As the scan angle increases, the e ective
resolution. size o the breast that can be ully imaged decreases. For an
Some other issues with a wide scan angle include: 8-cm thick breast and a 50° total scan angle (±25°), the feld
• A reduced feld o view due to shadows o the breast o view is reduced by approximately 4 cm on each breast
missing the detector at the widest angular projections side. Tis reduces the size o the breast that can be ully
• Increased noise and scatter due to the thicker e ective imaged by almost 8 cm.
breast thickness
• Degraded detector resolution due to oblique
incidence o the x-rays.

2
Ren B, Zhang Y, Ruth C, et al. Automatic patient motion detection in digital breast tomosynthesis. Proc SPIE 2011 (in press).
3
Ren B, Ruth C, Stein J et al. Design and performance of the prototype full-field breast tomosynthesis system with selenium-based flat-panel detector. Proc SPIE 5745 (2005): 550-561.
Effect of Scan Angle on Image Resolution
Objects appear sharpest or the narrowest scan angles,
with the limit o a non-scan such as 2D mammography
having the highest theoretical resolution. Tis somewhat
non-intuitive result is shown in Figure 7. In this fgure,
the appearance in a given slice o an object that extends
axially or a certain distance is simulated, shown here to be
about 3 in arbitrary units. With a single overhead exposure
(0º scan), the object’s width is seen to be its original size o
1.00. For a 15º scan, the projection into the slice is broad-
ened slightly to about 1.05, and or a 50º tomosynthesis
scan, the object’s size is blurred to 2.11. A larger scan angle
can broaden the object, decreasing the resolution and
sharpness o objects in the slice.
Figure 5. The field of view is impacted by the tomosynthesis scan angle. As the
scan angle increases, the effective size of the breast that can be fully imaged
decreases.

Figure 6 shows the feld o view reduction that occurs


as a unction o tomosynthesis total scan angle or an 8-cm
thick breast. Wider scan angles lead to a progressively
larger loss o feld o view. Projections rom the wider scan
angles do not contribute to the reconstructions at the upper
slices. I not properly normalized and accounted or, they
can increase noise due to ewer projections available in the
reconstruction; in addition arti acts can be introduced. Arti-
acts can also occur at wider scan angles unless care is taken
Figure 7. In-plane sharpness of objects degrades at larger scan angles.
to ensure such objects as the ace shield and/or the patient’s
head do not appear in the feld o view.
Oblique Incidence Versus Scan Angle
Te absorption o x-rays in a radiation detection layer
will occur at various depths. For x-ray trajectories at larger
scan angles, this will degrade the resolution o the image.4
As illustrated in Figure 8, the x-ray could be absorbed at
points “a,” “b” or anywhere along the line. I the absorption
occurred at point “a,” it would be detected in pixel 1, whereas
i it were absorbed at point “b,” it would be detected in
pixel 2. Tis e ect o resolution degradation becomes worse
as the scan angle increases. Tis e ect occurs in both cesium
iodide (CsI) and amorphous selenium (a-Se) detectors.

Figure 6. The total field-of-view reduction at the top of an 8 cm breast as a func-


tion of scan angle. Larger scan angles can reduce the field of view significantly.
The field-of-view loss is smaller for smaller breasts, larger for larger breasts.

4
Badano A, Kyprianou I, Jennings R. Anisotropic imaging performance in breast tomosynthesis. Med Phys. 34 (11): 4076-4091 (2007).
Continuous Scan Versus Step-and-Shoot Scans
T ere are two common methods o moving the x-ray
tube during the acquisition: continuous and step-and-shoot
. In step -and -shoot systems , the x-ray tube moves to each
angular position , then stops and takes an exposure , then
moves to the next location. In a continuous scan system, the
x-ray tube moves in a generally uni orm velocity throughout
the scan.
Te rationale o the step-and-shoot motion is that the x
-ray tube is stationary during the exposure , avoiding
Figure 8. Resolution degrades due to oblique incidence at the largest scan
angles. An incoming x-ray can be detected in more than one pixel. problems with ocal spot blur due to tube motion. However
, it is very di cult to make a step -and -shoot system that
As seen in Figure 9, or shallower scan angles, there mechanically works in a very rapid scan time. In addition ,
is no resolution degradation. Regardless o whether the x-ray any residual mechanical shaking or instability associated
is absorbed at point “a” or “b,” the signal will be detected with stopping the tube in such a system can also introduce
in the same pixel 1. ocal spot blur.
With a continuous scan, the x-ray tube is being pulsed
during the tube motion . Te challenge is to have a short
pulse width to avoid blurring o the image. Pulse widths can
be made quite small through the use o high mA x-ray
generators and x-ray flters designed specifcally or 3D
imaging . Te existing 2D flters are usually thicker and
would reduce x-ray ux , resulting in longer exposures . A
smaller scan angle will reduce the angular speed or a given
scan time, contributing to a reduction in ocal spot blurring.
On the contrary , a larger scan angle will contribute to
Figure 9. The resolution does not degrade when the incidence angle of the
increased ocal spot blurring with a potential negative e ect
x-ray is narrow. on microcalcifcation sharpness . Longer scan times reduce
the ocal spot blur through slower tube velocity , but a
Number of Projections more appropriate solution or a screening mammography
Another tomosynthesis system design consideration system is to narrow the scan angle.
intertwined with scan angle re ers to projection sampling , aking all actors into consideration , the continuous
or the number o projections per degree -scan-angle taken scan motion is probably pre erable because o the ast
during the scan . oo ew projections result in missed scan capability , and because , despite the continuous
sampling and can generate reconstruction arti acts . oo motion , the level o pulse width blurring is clinically
many projections result in a fxed total dose being divided insignif cant.
among many projections, which results in a very small dose
per projection. With too many projections, electronic noise Spatial Resolution
could begin to dominate . Wider scan angles need more Pixel size or f nal image spatial resolution must meet
projections to avoid the under -sampling problem and the the clinical requirements or mammography . A commonly
introduction o image arti acts. 5 Even i electronic noise used eature to decrease detector readout time and reduce
is not a signif cant design concern , it is di cult to keep noise is binning . T is re ers to the summation o multiple
scan times short as the number o projections increase . smaller pixels in a detector to generate a larger signal .
Detector readout times can then become a limiting actor, State-o -the-art detectors can seamlessly switch rom binned
un avorably increasing the scan time. to non -binned mode as a mammography unit switches
between 3D and 2D imaging modes.

5
Ren B, Wu T, Smith A. The dependence of tomosynthesis imaging performance on the number of scan projections. Astley et al (Eds.): IWDM 2006, LNCS 4046: 517-524 (2006).
Binning detector pixels does result in a reduction o comparison to 2D priors, or evaluating le t/right asymmetry
the intrinsic spatial resolution, but with a-Se detectors, and or the rapid appreciation o microcalcifcation clusters.
their intrinsically high resolution results in a high-resolution Microcalcifcations are not as a ected by superimposed
image, even with 2x2 binning. Tis is illustrated in tissues as masses are, and they are generally well visualized
Figure 10. As shown, the modulation trans er unction in a 2D image, particularly in systems with intrinsically high
(M F) resolution per ormance o a 2x2 binned 70-micron resolution. Te 3D image is valuable or identi ying masses
a-Se detector is, in act, superior up to about 4 lp/mm, and distortions, as well as or overcoming the e ects o
compared to a 100-micron mammography detector using superimposed tissue.
CsI as the radiation detection element.6 Studies have
also shown how appropriate reconstruction methods can Combo-mode
generate extremely high-resolution tomosynthesis images Combo-mode imaging with the Selenia Dimensions
using binned detectors.7 system provides the ability to acquire 2D images and 3D
scans in the same compression. Not only is this convenient
or the technologist, but it also benefts patients because it
minimizes the number o compressions and the time under
compression. In combo-mode, the Hologic system frst
per orms a 3D scan, and then the x-ray tube returns to the
perpendicular and acquires a 2D image. Because the 2D and
3D scans are acquired in the same compression, the image
sets are co-registered. Tere ore, any object in a given x, y
location in the 3D image will have the same object in the
same x, y location in the 2D image.8 Reviewing co-registered
images allows readers to toggle back and orth between the
two datasets. Combo-mode is also use ul or Computer-
aided Detection (CAD), which will be discussed in urther
detail later in this paper.
Researchers are currently looking into methods or cre-
ating a reconstructed 2D image rom the 3D dataset, which
would eliminate the need to separately acquire a 2D image
Figure 10. MTF of binned selenium detector compared to 100 micron during the procedure. When this technology is optimized,
CsI detector
clinical trials will be needed to validate the per ormance o
the reconstructed 2D image.
Te advantages o detector binning in allowing aster
scan times can outweigh the disadvantage o lower resolu- Reconstructed Slice Thickness
tion – as long as the fnal resolution meets the clinical needs.
I reconstruction is done with too ew slices, a sharp
Ultimately, clinical trials prove the per ormance and e cacy
object may never be properly visualized. I there are too many
o the system.
slices, reading time may be increased without improving
diagnostic outcome. For example, with a scan angle o 15°,
2D and 3D Imaging
a small microcalcifcation object will remain sharply in
A breast tomosynthesis system that can per orm both ocus in approximately two 1-mm-thick slices. In this case,
2D and 3D imaging is convenient and has a number o reconstructing the object into 0.5 mm slices will double the
advantages. In normal clinical use, alternating between 2D number o slices that need to be reviewed and would likely
and 3D imaging o di erent patients requires a system with not improve diagnostic per ormance because now, the same
the capability o switching modes quickly. Tere ore, being calcifcation will be seen sharply in our slices. Tere will be
able to rapidly switch between both modes is key. too many adjacent slices that appear similar. I , on the other
In screening applications, acquiring images or both the hand, reconstruction is done in slices with 2 mm separation,
2D and 3D examinations under the same compression is potentially in no slice would an object appear maximally
also desired. Te 2D image is use ul or many reasons: or

6
Ren B, Ruth C, Stein J et al. Design and performance of the prototype full-field breast tomosynthesis system with selenium-based flat-panel detector. Proc SPIE 5745 (2005): 550-561.
7
Acciavatti RJ, Maidment ADA. Investigating the potential for super resolution in digital breast images. Proc SPIE 2011 (in press).
8
Ren B, Ruth C, Wu T et al. A new generation FFDM/tomosynthesis fusion system with selenium detector. . Proc SPIE 7622 (2010): doi: 10.1117/12.844555
sharp. Tus, there is a relationship between scan angle and through the slices, as well as measuring x, y, z locations.
slice thickness. For wider scan angles, there is a need to Some researchers suggest tools like “slabbing,” where the
reconstruct objects into thinner slices to avoid the problem e ective slice thickness is increased, to better acilitate
o the objects never appearing sharply in ocus. Tinner the appreciation o microcalcifcation clusters.10 Finally,
slices mean more slices and an increase in the radiologist’s tools to toggle between the 2D and 3D image may enable
reading time. At the moment, no breast tomosynthesis radiologists to more quickly review images.
clinical studies exist showing any value or slices thinner Te development o such workstations, particularly by
than 1 mm. PACS companies, is in its in ancy, and as a result, signifcant
Tinner slices can also degrade the user’s ability to advances in reviewing e ciencies are expected in the uture.
appreciate a microcalcifcation cluster, since in a given slice,
only a subset o the calcifcations will appear sharply in ocus.
One way to improve calcifcation cluster visibility is to do
some type o summation to create thicker slices; however,
these summation algorithms o ten degrade resolution, and
add additional steps or the reviewing radiologist.

Reconstruction Algorithms
Reconstruction time is another important design pa-
rameter in tomosynthesis systems. O course, image quality
is critical, but reconstruction times must obviously be ast
enough to support screening patient throughputs, and they
must be per ormed in seconds i tomosynthesis is going to
be used in interventional procedures.
Investigations have been made using such iterative
reconstructions methods including variants o Maximum
Likelihood (ML), Algebraic Reconstruction echniques
Figure 11. Example of tomosynthesis diagnostic workstation
(AR ), as well as the simpler, but more computationally
e cient Filtered Back Projection (FBP) method. Filtered
Tomosynthesis in the Future
back-projection algorithms are able to reconstruct a volume
much more rapidly, and through the use o appropriate Tomosynthesis Interventional Procedures
fltering, they appear virtually indistinguishable rom the Because tomosynthesis exams may uncover lesions
slower iterative methods.9 Hologic uses FBP because that would not be visible using other imaging modalities, a
customer pre erence studies ound no signifcant image tomosynthesis system should support the ability to per orm
quality di erences between ML, AR and FBP. Moreover, tomosynthesis-guided interventional procedures. Consider-
the rapid reconstruction speed is enormously advantageous, ation should be given to the system requirements needed to
particularly or screening throughput. support tomosynthesis-guided biopsy. For example, a long
source-to-image distance (SID) is use ul to allow room or
Reading Time, File Sizes, PACS Issues, installation and deployment o biopsy device hardware. In
Workstation Tools and More addition, image reconstruction algorithms that minimize
Tere are many issues that arise in a tomosynthesis the arti acts rom metal objects such as needles are needed.
system based on the large datasets inherent in a 3D image. Reconstruction time a ects how long the patient is con-
Te Picture Archiving and Communication System (PACS) tinuously under compression, as well as enables an e cient
must have adequate bandwidth to transmit the large images, work ow, which is particularly important in a busy practice.
as well as su cient storage. Perhaps the biggest challenges As a result, reconstruction time in the order o seconds is
are in the design o diagnostic workstations. Tese critically important in tomosynthesis-guided interventional
workstations must support virtually instantaneous loading procedures. A biopsy procedure can entail several acquisi-
and display o the tomosynthesis images. In addition, they tions, including scout, pre-fre and post-fre scans, so rapid
must provide tools to support viewing and navigating scans are also highly desirable in such a system.

9
Ren B, Ruth C, Zhang Y, et. al. The CNR method in scan angle optimization of tomosynthesis and its limitations. Proc. SPIE 7258 (2009); doi:10.1117/12.813918.
10
Kopans DB. Breast Imaging. Lippincott Williams & Wilkins. 3rd edition, e.g. 1075-1077.
Computer-aided Detection (CAD) Glossary
3D CAD could potentially work in a superior way
compared to current 2D CAD per ormance on a tomosyn- 2D Conventional digital mammography
thesis dataset. However, the large number o cases needed 3D Breast tomosynthesis
or developing a CAD algorithm will probably limit the ART Algebraic Reconstruction Techniques, a method of
availability o 3D CAD in the near uture. reconstructing tomosynthesis images
One advantage o per orming a combo-mode exam is a-Se Amorphous selenium, a radiation detector material
that 2D CAD is available on the 2D image. With a combo- used in direct conversion detectors
mode scan acquiring 2D and 3D in the same compression, Binning The summation of multiple smaller pixels in a detector
the 2D CAD mark can be placed in the appropriate and to generate a larger signal and faster image readout
correct x, y location on the 3D image. Tis allows the use CAD Computer aided detection. Software algorithm that
marks areas with characteristics suggestive of
o 2D CAD, while gaining the advantages o tomosynthesis breast malignancy.
imaging.
Combo-mode A protocol where both the 3D and 2D images are
acquired in the same compression, with the second
Conclusion acquisition following immediately after the first
Te design o a tomosynthesis system requires care ul Continuous scan A method of moving the x-ray tube during a
consideration o a large number o parameters. From many tomosynthesis scan, where the x-ray tube rotates
years o research and clinical experience, it is obvious that continuously throughout the scan
rapid scan times are essential in order or a commercial CsI Cesium Iodide, a radiation detector material used in
indirect conversion detectors
tomosynthesis system to be success ul. Hologic chose to
create a completely new mammography unit – the Selenia FBP Filtered Back Projection, a method of reconstructing
tomosynthesis images
Dimensions 2D/3D system – to deliver the rapid scan time
Field of view Area that is visible
needed, along with other optimized parameters, including
scan angle, pixel size and slice thickness. Tis design ML Maximum Likelihood, a method of reconstructing
tomosynthesis images
ensures that the needs o both screening and diagnostic
MTF Modulation Transfer Function. The ability of an imaging
mammography may be met e ectively.
system to faithfully translate the modulation of the
transmitted x-ray beam into a visible image is
measured by the Modulation Transfer Function.
Projections The low dose images taken at differing angles around
the breast as the x-ray tube rotates. In the Hologic
Dimensions system, there are 15 projections acquired
over a total 15° arc
Reconstruction The process of computing high-resolution images
(slices) from projections. The planes of the recon-
structed slices are parallel to the breast support plates
Scan angle The total angular range covered by the x-ray tube
SID Source-to-image distance
Slabbing Increasing the effective slice thickness
Step-and-shoot A method of moving the x-ray tube during a tomo-
scan synthesis scan, where the x-ray tube comes to a stop
for each projection’s exposure, and then moves to the
next angular location.

United States / Latin America


35 Crosby Drive
Bedford, MA 01730-1401 USA
Tel: +1.781.999.7300
Sales: +1.781.999.7453
Fax: +1.781.280.0668
www.hologic.com

WP-00046 (4/11) U.S./International

You might also like