Benefusion VP5– Syringe pump
NO.:MSK-17006-Benefusion VP5 FP
Benefusion
By VP5–
Service Engineer Mindray Infusion
Medical Indonesia pump
Benefusion VP5 Introduction
Appearance
Infusion Pump
Benefusion
VP5
Benefusion VP5 Appearance Side-looking
Alarm light
Keypad,easy to use
RS232:
AC IN Data transfer to PC/software upgrade/update the
Dock system
Nurse Call
DC IN
Benefusion VP5 Operation-Key function introduction
A total of 12 buttons
CLEAR/BAC
K Edit state, press to clear values.
For high/middle level alarm, press to eliminate the alarm sound for
SILENCE
2 minutes, for low level alarm, press to clean the alarm.
Operation
Infusion Pump
Benefusion
VP5
Infusion Flow Chart
Turn Pump
Install IV Set Purge
on
Finish
Benefusion VP5 BOOT Self testing
⚫ Self testing progarm: Memory, communication, alarm lamp, power supply, the pressure
coefficient, etc.;
⚫ Maximum guarantee the machine running normally before set parameters, to ensure
safety in advance.
Install IV Set according to the following method:
Select brand
Select infuse mode Default: Flow Rate mode, Press “Menu” to select others
Connect to patient
Bolus
• Press “Bolus”button
Automatic • Set bolus parameters
Bolus Rate:0.10-2000ml/h
Benefusion VP5 DPS & Anti-Bolus
Anti-Bolus:
KVO (Keep Vein Open) means to keep the vein open, during which the
infusion pump continues infusion at a very low rate after finishing the
infusion in order to prevent blood backflow or vascular blockage.
Principle
Infusion Pump
Benefusion
VP5
20
Benefusion VP5 Principle- Half extrusion peristalsis
Microcomputer system control pulse stepper motor rotation, and stepping motor driven
extrusion/check liquid pump peristalsis, the infusion of liquid into the body.
Benefusion VP5 Principle-Hardware(Microcomputer)
Benefusion VP5 Principle-Hardware(Microcomputer)
Master control Host CPU
Keyboard PCBA:
◆Key management section is responsible for (a total of 12 key, the power
switch directly to the CPU power supply module, other key by management
CPU) scanning processing, alarm lamp driver, LCD backlight control (8
gears adjustable), and other functions.
Calibration
Infusion Pump
Benefusion
VP5
Benefusion VP5 Maintenance-Accuracy calibration
The accuracy of the infusion pump don't need to do a routine check. But for the first
time use, change new iv brands or when you suspect infusion flow deflection need
to be calibrated.
Detection
Infusion Pump
Benefusion
VP5
Benefusion VP5 Testing-AD value
Steps to enter AD value:System---Input passport---AD value
Upgrade
Infusion Pump
Benefusion
VP5
Benefusion VP5 Maintenance-Upgrate
Login PW:25931501
Multifunction
serial port line
Symptom
Infusion Pump
Benefusion
VP5
Benefusion VP5 Symptom-Possible Cause and Action
Symtom Possible Cause Action
Pressure
Down pressure detect board
value not Replace down pressure detect board
trouble
change
Pressure
Down pressure detect board
value over Replace down pressure detect board
trouble
range
Alarm up Up pressure detect board
Replace up pressure detect board
occlusion trouble
Alarm Loose or damage to the door
Repair or replace the switch
door open to detect line switch
Special 1.Special tube detection
circuit fault 1.Replacement air bubble detection board
tube
failure 2.The wire from Bubble 2.Repair bubble board to the main control
detect board to the main board wire
control panel is loose
Benefusion VP5 Symtom-Possible Cause and Action
Symtom Possible Cause Action
1, ultrasonic launch
ultrasonic transmission 1.Repair ultrasonic launch
circuit wafer wire loose or wafer wire or replace keypad;
False damaged; 2, ultrasonic 2.Repair ultrasonic receiving
positives receiving chip loose wires or chip wire or replace bubble
bubbles ultrasonic testing circuit detection plate;
failure; 3, bubble test board 3.Repair bubble test board to
to the main control board the main control board wire.
wire is loose.
1.Motor testing board
failure; 1.Replace Motor testing board;
Motor fault 2.Motor wire poor contact; 2.Replace or repair motor wire
3.Stepper motor drive circuit 3.Replace mainboard or motor
fault fault;
1.Hydraulic clamp board
1.Replace bubble detect board
Hydraulic fault
clamp 2.Repair the wire from bubble
2.The wire from Bubble
failure detect board to main control
detect board to the main
panel
control panel is loose
Benefusion VP5 Maintenance and Repair
Disassembly
Infusion Pump
Benefusion
VP5
Benefusion VP5 Disassemble -Whole machine struction
Whole
machine
Pump body
Front shell After shell transmission
components
Hydraulic
Pump body pump CAM&Motor
clamp
installed base mounting base components
component
Benefusion VP5 Disassemble -Whole machine
The supplied pan-head screws, M3x8 double gasket
Upper cover
The supplied pan-head screws, M3x8 double gasket The supplied pan-head screws, M3x8 double gasket
Benefusion VP5 Disassemble -Front shell
Ultrasonic wire
Benefusion VP5 Disassemble -Pump body
M3X8 double gasket screw
Door buffering
door axle spring
Pressure sensor
Ultrasonic sensor