Professional Documents
Culture Documents
Sensor module
The sensor module, available in
stainless steel or copper, is easily
removed, which facilitates cleaning
and decontamination. The fan and
HEPA CO2 filter are also easy to
change.
02
UV light
UV irradiation at a wavelength of OPTIMUM
254 nm is effective against bacteria,
fungi, yeasts, viruses, mycoplasmas
and spores, and disinfects the
PROTECTION
surface of the inner casing.
Glass door safety device Support frame 6-door gas tight screen
The glass door safety device CO2 incubators should never be The gas tight screen reduces the
makes it impossible to close the installed directly onto the floor area exposed to the external atmos-
outer door unless the inner glass because of the risk of contami- phere by a factor of 6 when opening
door has been properly locked. nation from the ground. Support the incubator, reducing the risk of
This prevents unwanted airborne frames with a clearance of 300 mm contamination considerably whilst at
bacteria from entering the work or 780 mm substantially reduce the same time maintaining incubation
space when the incubator is in this risk. conditions. Recovery times are
operational mode. greatly accelerated, gas consumption
is reduced and work space atmos-
phere interruptions are kept to an
absolute minimum.
03
STABLE INCUBATION CONDITIONS
for good growth
120
ration from culture media. Rapid recovery times work space and wall temperatures, and so
100
are particularly important when doors have to prevents condensation.
80
be opened frequently or when culture media
volumes are low.
60
Attention to safety
Alarm system
Any deviation from set parameters activates a
reliable acoustic and visual alarm.
04
USEFUL EQUIPMENT
specifically ... for your applications
Lockable door
The lockable door prevents unauthorised or
unwanted access to samples. This is a major
advantage, and is of particular significance
when important or sensitive cultures are being
handled.
05
ONE INCUBATOR FOR ALL APPLICATIONS
BBK 6220 CO2 Incubator with refrigeration system
06
TECHNICAL DATA
Type Unit Value/Description
BB 6060 BB 6220 BBK 6220
Dimensions
External dimensions (w/h/d) mm 744/595/567 895/860/747 895/860/747
Internal dimensions (w/h/d) mm 564/405/405 607/669/585 607/669/585
Total volume l 60 220 220
One-piece shelves
Dimensions (w/d) mm 409/340 560/500 560/500
Standard/maximum number no. 2/6 3/12 3/12
Divided shelves
Dimensions (w/d) mm – 260/500 –
Standard/maximum number no. – 6/24 (181)) –
Net weight kg 60 100 110
Microprocessor control PID CO2 and O2/N2 temperature control system
Temperature control
Temperature control range °C tA 2) … 50 °C tA 3) … 50 °C 104)…50 °C
Temperature deviation (spatial) (DIN 12 880) °C ≤ ± 0.4 ≤ ± 0.5 ≤ ± 0.4
Temperature deviation (time) (DIN 12 880) °C ≤ ± 0.1 ≤ ± 0.1 ≤ ± 0.1
CO2 control auto-start and auto-zero routine
CO2 control range % 0…20 0…20 0…20
O2 /N2 control (option) auto-cal routine
O2 control range % 3…80 3…80 –
N2 supply for range % 3…21 3…21 –
O2 supply for range % 21…80 21…80 –
Humidity control
Constant humidity % > 95 98 > 954)
High/low humidity (option), reversible % – 98/92 –
Electrical data
Power kW 0,60 1,20 1,20
Voltage/frequency V~; Hz 230 5); 50/60 230 5); 50/60 230 5); 50/60
1) 2) 3)
In combination with gas screen tA at least 7 °C above ambient temperature tA at least 5 °C above ambient temperature
4) 5)
Depending on thermostat flow temperature 120 V version also available
ORDER NUMBERS
Model Order no.
BB 6060 with stainless steel inner casing 51008947
BB 6060 with solid copper inner casing 51008948
BB 6220 with stainless steel inner casing and divided shelves 51007413
BB 6220 with solid copper inner casing and divided shelves 51007412
BB 6220 with stainless steel inner casing plus O2 control, 6-door gas tight screen and divided shelves 51007494
BB 6220 with solid copper inner casing plus O2 control, 6-door gas tight screen and divided shelves 51007918
BBK 6220 with stainless steel inner casing and one-piece shelves (refrigeration thermostat available on request) 50035416
07
FAST AND ORDER NUMBERS
PRECISE Accessory
Stacking frame for 2 units
BB 6060
50027660
BB 6220/BBK 6220
50027661
Dimensions
860
120
Internet http://www.kendro.com
Registered to ISO 9001. Kendro Laboratory Products meet or exceed stringent quality and product safety standards. ©2002 Kendro Laboratory Products. All Rights Reserved.
Printed in Germany 1C 01/03 4t Druckerei – Artikelnummer
Accessories for CO2 - incubators
List of contents
Information:
When fitted the castors will raise the
support frame by 90 mm.
Information:
When fitted the castors will increase
the height of the support frame by 90
mm.
1. Standard shelf
Full width inc. support rails, SS 50 065 793
Full width inc. support rails, Cu 50 065 794
Width: 560 mm (Overall)
Depth: 500 mm (Overall)
Thickness: 1 mm
Information:
The cytoperm 2 or BBD 6220 should
stand on the adaptor plate
Information:
Alternatively another unit can be
stacked in the upper position, e.g.
BB6220 or HERAcell 240
Information:
When fitted the castors will raise the
support frame by 90 mm.
Anschrift:
Kendro Laboratory Products GmbH
Robert – Bosch – Straße 1
D - 63505 Langenselbold
Deutschland
Telefon:
Vertrieb/
Service + 49 (0) 6184 / 906940
Fax:
Vertrieb/
Service + 49 (0) 6184 / 907474
PURPOSE:
EQUIPMENT AFFECTED:
DESCRIPTION / INFORMATION:
Spare part 50029957 will be replaced by 50082995 Fan comp. for TCD
Important Note Although the connector is different (5 pin DIN) it will fit the
existing socket and the functionality is the same.
Author Reviewed By
S. Czwak R. Bloomfield
Kendro Service & Support
Service Bulletin
Product Line Equipment Bulletin Issue Date
ADDITIONAL INFORMATION:
Author Reviewed By
S. Czwak R. Bloomfield
Kendro Service & Support
Service Bulletin
Product Line Equipment Bulletin Issue Date
BBK 6220, BB16, BK 6160,
BB 5060, B 5061, BB 6060/6220 Fan 0503-LSB-INC-009 12. April 2005
Reason:
EQUIPMENT AFFECTED:
Identnumber 50082995 Fan complete, TCD. As a spare part for the listed products
above.
DESCRIPTION / INFORMATION:
To perform this you need a multimeter. The pin description is shown in the picture on
the next page.
If the measured value is infinite, the diode inside is in the reverse direction mounted.
Caution! If this fan is installed, it will cause a short circuit and destroy the capacitor.
F. Born R Bloomfield
Kendro Service & Support
Service Bulletin
Product Line Equipment Bulletin Issue Date
BBK 6220, BB16, BK 6160,
BB 5060, B 5061, BB 6060/6220 Fan 0503-LSB-INC-009 12. April 2005
PIN 4
PIN 2
PIN 1
PIN 5
PIN 3
ADDITIONAL INFORMATION:
F. Born R Bloomfield