You are on page 1of 240

TRIED AND TESTED

BB 6060/BB 6220 CO2 Incubators for cell and tissue culture


BBK 6220 CO2 Incubator with refrigeration system
YOU CAN RELY Solid copper inner casing
BB 6060 and BB 6220 CO2
Smooth inner casing
The smooth inner casing, which
Incubators can be supplied with has rounded corners, is made
ON US optional solid copper inner
casings and shelving systems.
from corrosion-resistant electro-
polished stainless steel (1.4301).
Copper has a fungicidal effect The sensor module and fan are
and prevents the spread of both plug-in units, so they can be
Over the past 30 years, we have contamination, providing perma- changed without difficulty. This
nent protection for samples. enables problem-free cleaning
worked closely with our custo- and safe decontamination of the
mers to develop and improve entire interior.

our CO2 incubator range. This


has resulted in better reliability,
consistent improvements in
quality and innovative product
solutions.

The Heraeus® BB 6060 and


BB 6220 CO2 Incubators offer
many advantages:
■ Available with a wide range
of equipment to provide
maximum protection against
contamination
■ Safe and stable incubation
conditions ensure optimum
growth
■ Protection against drying out
through extremely rapid
humidity recovery times

Sensor module
The sensor module, available in
stainless steel or copper, is easily
removed, which facilitates cleaning
and decontamination. The fan and
HEPA CO2 filter are also easy to
change.

02
UV light
UV irradiation at a wavelength of OPTIMUM
254 nm is effective against bacteria,
fungi, yeasts, viruses, mycoplasmas
and spores, and disinfects the
PROTECTION
surface of the inner casing.

Glass door safety device Support frame 6-door gas tight screen
The glass door safety device CO2 incubators should never be The gas tight screen reduces the
makes it impossible to close the installed directly onto the floor area exposed to the external atmos-
outer door unless the inner glass because of the risk of contami- phere by a factor of 6 when opening
door has been properly locked. nation from the ground. Support the incubator, reducing the risk of
This prevents unwanted airborne frames with a clearance of 300 mm contamination considerably whilst at
bacteria from entering the work or 780 mm substantially reduce the same time maintaining incubation
space when the incubator is in this risk. conditions. Recovery times are
operational mode. greatly accelerated, gas consumption
is reduced and work space atmos-
phere interruptions are kept to an
absolute minimum.
03
STABLE INCUBATION CONDITIONS
for good growth

Rapid recovery time Air jacket heating system


High humidity in the incubator reduces evapo- The patented heating system ensures uniform
Humidity recovery time (min)

120

ration from culture media. Rapid recovery times work space and wall temperatures, and so
100
are particularly important when doors have to prevents condensation.
80
be opened frequently or when culture media
volumes are low.
60

In all Heraeus incubators equipped with a direct Gas tight screen


40
humidifying system, the tray at the bottom of The innovative gas screen allows culture
20
the work space serves as a reservoir for the conditions to be maintained virtually unchanged
humidifying water. This design allows heat from during loading, and also ensures extremely
0 the heating system in the base to be transferred rapid recovery times. The gas screen's six
BB 6220 CO2 incubators directly to the water, enabling a high level of individual doors seal off the work space and
with separate water tray
humidity to be built up very rapidly in the work render it gas tight, allowing separate access to
space. individual compartments. The advantage of the
Humidity recovery time – BB 6220 compared By comparison, a separate water tray – and gas screen lies in the reduction of the work
with traditional CO2 incubators with separate more importantly, the layer of air between the space's opening profile. In other words, the
water tray tray and the bottom of the incubator – block screen leads to lower gas and energy con-
the transfer of heat to the humidifying water, sumption, reduced risk of contamination and
resulting in extended humidity recovery times protection against cultures drying out.
and high rates of evaporation from cultures.
Compared with ordinary incubators with a
water tray, the BB 6220's humidity recovery
times are up to five times faster.

Automatic CO2 control with Microprocessor controlled metering


auto-zero system and control system
The patented, self-balancing CO2 control All key incubation parameters, such as the
system ensures constant, long-term stability of functioning of components essential for opera-
CO2 levels and so of pH levels in the culture tions, are monitored and controlled by a micro-
medium as well. This means that there is no processor. Software routines detect special
longer any need for time consuming recalibration operating conditions – opening the door, for
of the measuring cell. The encapsulated mea- instance – and ensure that incubation condi-
suring cell, which is protected against bacterial tions revert to normal as rapidly as possible.
infiltration, is easy to remove from the work
space. Microprocessor controlled metering and
control technology ensures that, once set, the
CO2 concentration remains unaffected by other
gases.

Attention to safety

Easy to read display and Independent overtemperature


control panel protection
The display and control panel is fitted with an Two independent platinum precision resistors
easy to clean, soft-touch keypad. Separate fitted in the work space serve as sensors,
displays make it easy and quick to monitor ensuring a uniform climate and providing
incubation parameters. secure protection against overheating.

Alarm system
Any deviation from set parameters activates a
reliable acoustic and visual alarm.

04
USEFUL EQUIPMENT
specifically ... for your applications

Control of the O2 content and Bottle turning device


drift-free metering via an auto-cal The bottle turning device enables bottles of
system different sizes to be accommodated. The
Controlling the O2 content in the work space rotational speed of each level can be set inde-
atmosphere makes it possible to reproduce the pendently. The drives themselves are designed
partial oxygen pressure of various organs in for minimal heat emission and smooth, silent
vitro. running. Bottle turning devices can be com-
bined with metal shelves.
Oxygen levels of >21% can be realised by fee-
ding in O2, and <21% by feeding in N2. Oxygen
levels are determined by means of a Clark-type
O2 electrode.

The electrode is calibrated automatically by a


built-in auto-cal system. A monitor indicates
when the electrolyte solution needs to be
changed. This guarantees the accuaracy of the Magnetic stirrer control module
set value and so ensures the highest level of Magnetic stirrers are suitable for mass cell
safety for your cultures. cultivation in spinner bottles. The control
module is designed for the connection of up to
two (BB 6060) or four (BB 6220) magnetic
High/low humidity system stirrers. The connector socket is built into
Humidity within the work space can be reduced the work space. Each magnetic stirrer can
from 98% to 92% of the rH value to prevent accommodate two spinner bottles.
condensation forming on large cell culture
bottles after they have been out of the incubator
for a long time.

Additional safety features Cytolink data capture


Cytolink monitors all relevant parameters via a
PC. Up to ten BB 6060 and BB 6220 Incuba-
tors can be connected to a single bus system if
desired (Cytolink with Cytolink modem).

Gas cylinder monitor


Automatic switching of gas cylinders for safety.

Lockable door
The lockable door prevents unauthorised or
unwanted access to samples. This is a major
advantage, and is of particular significance
when important or sensitive cultures are being
handled.

05
ONE INCUBATOR FOR ALL APPLICATIONS
BBK 6220 CO2 Incubator with refrigeration system

The BBK 6220 with built-in refri-


geration is a general purpose
CO2 incubator, suitable for all
applications, which completes
the 6000 range. Its additional
refrigeration function offers the
possibility of incubation at low
temperatures. This capability is
particularly important in marine
biology and fish farming.

Equipment Refrigeration system


■ Refrigeration unit with connection to a Connection to a simple laboratory refrigeration
compact table top refrigeration thermostat thermostat enables work space temperature
■ Microprocessor control settings from 10 to 50 °C. The combination of
■ auto-start automatic start-up refrigeration system and air jacket heating
■ Fully automatic auto-zero adjustment of CO2 system guarantees optimum temperature and
measuring cell humidity. Uniform heat distribution prevents
■ Independent electronic overtemperature condensation build-up, even during cooling.
protection with separate sensor
■ Heated external door To enable all working temperatures from 10
■ Glass internal door to 50 °C can be covered, we recommend a
■ One-piece metal shelves refrigeration thermostat with a lowest working
■ All other equipment options, such as oxygen temperature of –15 °C and a refrigeration
control, high/low humidity, gas screen, capacity of 0.2 kW.
copper construction, bottle turning device,
Cytolink data capture, access ports and To maintain an incubator temperature of 37 °C
lockable door, are available for the BBK 6220 in a laboratory with a high ambient temperature,
on request. a thermostat with a much lower refrigeration
capacity can also be used, or the incubator can
be connected to a water cooling system.

06
TECHNICAL DATA
Type Unit Value/Description
BB 6060 BB 6220 BBK 6220
Dimensions
External dimensions (w/h/d) mm 744/595/567 895/860/747 895/860/747
Internal dimensions (w/h/d) mm 564/405/405 607/669/585 607/669/585
Total volume l 60 220 220
One-piece shelves
Dimensions (w/d) mm 409/340 560/500 560/500
Standard/maximum number no. 2/6 3/12 3/12
Divided shelves
Dimensions (w/d) mm – 260/500 –
Standard/maximum number no. – 6/24 (181)) –
Net weight kg 60 100 110
Microprocessor control PID CO2 and O2/N2 temperature control system
Temperature control
Temperature control range °C tA 2) … 50 °C tA 3) … 50 °C 104)…50 °C
Temperature deviation (spatial) (DIN 12 880) °C ≤ ± 0.4 ≤ ± 0.5 ≤ ± 0.4
Temperature deviation (time) (DIN 12 880) °C ≤ ± 0.1 ≤ ± 0.1 ≤ ± 0.1
CO2 control auto-start and auto-zero routine
CO2 control range % 0…20 0…20 0…20
O2 /N2 control (option) auto-cal routine
O2 control range % 3…80 3…80 –
N2 supply for range % 3…21 3…21 –
O2 supply for range % 21…80 21…80 –
Humidity control
Constant humidity % > 95 98 > 954)
High/low humidity (option), reversible % – 98/92 –
Electrical data
Power kW 0,60 1,20 1,20
Voltage/frequency V~; Hz 230 5); 50/60 230 5); 50/60 230 5); 50/60
1) 2) 3)
In combination with gas screen tA at least 7 °C above ambient temperature tA at least 5 °C above ambient temperature
4) 5)
Depending on thermostat flow temperature 120 V version also available

ORDER NUMBERS
Model Order no.
BB 6060 with stainless steel inner casing 51008947
BB 6060 with solid copper inner casing 51008948
BB 6220 with stainless steel inner casing and divided shelves 51007413
BB 6220 with solid copper inner casing and divided shelves 51007412
BB 6220 with stainless steel inner casing plus O2 control, 6-door gas tight screen and divided shelves 51007494
BB 6220 with solid copper inner casing plus O2 control, 6-door gas tight screen and divided shelves 51007918
BBK 6220 with stainless steel inner casing and one-piece shelves (refrigeration thermostat available on request) 50035416

Options Suitable for Order no.


O2 control with auto-cal function for stainless steel inner casing BB 6060 51900089
O2 control with auto-cal function for solid copper inner casing BB 6060 51900145
Connection for central monitoring BB 6060/BB 6220 51900087
Gas screen with 6 gas tight glass doors for separate access BB 6220 51900085
High/low humidity system BB 6220 51900086
UV light for stainless steel version BB 6220 51900088
UV light for solid copper version BB 6220 51900144
Access port BB 6060/BB 6220 51900238
Lockable external door BB 6060 51900280
BB 6220 51900281
Cytolink module – PC monitoring and documentation BB 6060/BB 6220 51900177
Magnetic stirrer control module for stainless steel inner casing BB 6060/BB 6220 51900178
for solid copper inner casing BB 6060/BB 6220 51900179
for stainless steel inner casing and fitted with UV light BB 6220 51900180
for solid copper inner casing and fitted with UV light BB 6220 51900181
Bottle turning device, including drive module in module frame for stainless steel inner casing, 1 level BB 6220 51900171
2 levels BB 6220 51900172
3 levels BB 6220 51900173
Bottle turning device, including drive module in module frame for solid copper inner casing, 1 level BB 6220 51900174
2 levels BB 6220 51900175
3 levels BB 6220 51900176

07
FAST AND ORDER NUMBERS
PRECISE Accessory
Stacking frame for 2 units
BB 6060
50027660
BB 6220/BBK 6220
50027661

SERVICE Support frame 300 mm high


Support frame 780 mm high
Additional one-piece stainless steel shelf
50031349
50029890
50011400
50031348
50029597
50029945
Additional one-piece solid copper shelf 50025629 50029944
Additional divided stainless steel shelf (BB 6220 only) – 50029943
Additional divided solid copper shelf (BB 6220 only) – 50029942
Gas cylinder monitor with automatic gas cylinder switching,
reversible, 230 V-, 50/60 Hz and 120 V-, 50/60 Hz 50046033 50046033
CO2 cylinder pressure reducer 03429937 03429937
N2 cylinder pressure reducer 03429942 03429942
O2 cylinder pressure reduce 03429943 03429943
Cytolink modem, 230 V-, 50/60 Hz 50034330 50034330
Magnetic stirrer 50034465 50034465

Dimensions

Front view 50 895 50 747 Side view


BB 6220/BBK 6220 BB 6220/BBK 6220

860
120

Fast response times, com-


Front view 50 744 50 567 Side view
petent solutions: that is
BB 6060 BB 6060
our performance promise
to our customers all over
the world. Our comprehen-
sive service package does
not only include repairs 597
and maintenance.

For us, service means


technological competence, 120
advice on applications
and excellent logistics.
The basis for all this is our Your Sales Contact
equipment's outstanding
quality.

For that reason, your


decision to work together
with Kendro is also the first
step towards a reliable
and reassuring service
partnership.

For Ordering or Technical Information


Asia Pacific Kendro Laboratory Products (H.K.) Limited · Hong Kong · Tel. +852 2711 3910 · Fax +852 2711 3858 · info@kendro.com
Europe, Middle East, Africa
Kendro Laboratory Products International Sales · Hanau · Germany · Tel. +49 (0) 1805 - 536 376 · Fax +49 (0) 1805 -112 114 · info@kendro.de
Latin America Kendro Laboratory Products International Sales · Newtown, CT · USA · Tel. +1 203 - 270 2080 · Fax +1 203 - 270 2210 · info@kendro.com
North America Kendro Laboratory Products · Newtown, CT · USA · Tel. +1 800 - 522 7746 · Fax +1 203 - 270 2166 · info@kendro.com

Internet http://www.kendro.com

Quality Products – Lifetime Care

Registered to ISO 9001. Kendro Laboratory Products meet or exceed stringent quality and product safety standards. ©2002 Kendro Laboratory Products. All Rights Reserved.
Printed in Germany 1C 01/03 4t Druckerei – Artikelnummer
Accessories for CO2 - incubators

Illustrations and descriptions

Version: 07.2003 / 50078754


Accessories for CO2 - incubators

List of contents

List of contents ................................................................................................................................... 2

1 Accessories for HERAcell / HERAcell 150 .............................................................................. 3

1.1 Support frames..................................................................................................................... 3


1.2 Shelves / HERAtrays ............................................................................................................ 4
1.3 Stacking adapters ................................................................................................................ 4
2 Accessories for HERAcell 240................................................................................................... 5

2.1 Support frames..................................................................................................................... 5


2.2 Shelves / HERAtrays ............................................................................................................ 6
2.3 Stacking adapters ................................................................................................................ 7
3 Accessories for cytoperm 2 / BBD 6220 ................................................................................... 8

3.1 Support frames..................................................................................................................... 8


3.2 Shelves / HERAtrays .......................................................................................................... 10
3.3 Stacking adapters .............................................................................................................. 11

Stand 07.2003 / 50078754 2


Accessories for CO2 - incubators

1 Accessories for HERAcell / HERAcell 150

1.1 Support frames

1. Support frame 200 mm 50 051 376


Width: 637 mm
Height: 200 mm
Depth: 600 mm

2. Support frame 780 mm 50 051 436


Width: 637 mm
Height: 780 mm
Depth: 600 mm

3. Support frame 185 mm (with castors) 50 057 161


Width: 637 mm
Height: 185 mm
Depth: 600 mm

4. Support cart with drawers and 50 056 459


castors
Width: 637 mm
Height: 890 mm
Depth: 600 mm

5. Castor set for 1 & 2 above. (Set of 4) 50 052 528

Information:
When fitted the castors will raise the
support frame by 90 mm.

Version: 07.2003 / 50078754 3


Accessories for CO2 - incubators

1.2 Shelves / HERAtrays

1. Shelf inc. support rails, SS 50 051 909


Shelf inc. support rails, Cu 50 051 910
Width: 423 mm (Outside
Depth: 465 mm (Outside)
Thickness: 1 mm

2 HERAtray 1/3 Width SS (3 pc.) 50 051 913


HERAtray 1/3 Width Cu (3 pc.) 50 051 914
Width: 135 mm (Overall)
Depth: 440 mm (Overall)
Thickness: 1 mm

3. HERAtray 1/2 Width SS (2 pc.) 50 058 672


HERAtray 1/2 Width Cu (2 pc.) 50 061 050
Width: 205 mm (Overall)
Depth: 440 mm (Overall)
Thickness: 1 mm

1.3 Stacking adapters

1. Plate for stacking a HERAcell or 50 068 677


HERAcell 150 on a HERAcell 240
Width: 780 mm
Depth: 772 mm

2. Plate for stacking a HERAcell or 50 051 938


HERAcell 150 with a BB 16
Width: 647mm
Depth: 660 mm

3. Plate for stacking a HERAcell or 50 060 612


HERAcell 150 on a BB 6220
Width: 920 mm
Depth: 670 mm

4. Plate for stacking a HERAcell or 50 060 736


HERAcell 150 on a 5060 or B 5061
Width: 890 mm
Depth: 665 mm

Version: 07.2003 / 50078754 4


Accessories for CO2 - incubators

2 Accessories for HERAcell 240

2.1 Support frames

1. Support frame 200 mm 50 065 754


Width: 780 mm
Height: 200 mm
Depth: 652 mm

2. Support frame 780 mm 50 065 753


Width: 780 mm
Height: 780 mm
Depth: 652 mm

3. Support frame 185 mm (with castors) 50 067 224


Width: 780 mm
Height: 185 mm
Depth: 652 mm

4. Castor set for 1 & 2 above (set of 4) 50 052 528

Information:
When fitted the castors will increase
the height of the support frame by 90
mm.

Version: 07.2003 / 50078754 5


Accessories for CO2 - incubators

2.2 Shelves / HERAtrays

1. Standard shelf
Full width inc. support rails, SS 50 065 793
Full width inc. support rails, Cu 50 065 794
Width: 560 mm (Overall)
Depth: 500 mm (Overall)
Thickness: 1 mm

2. Reinforced shelf (2 mm Thick)


Full width inc. 2 support rails, SS
Full width inc. 2 support rails, Cu 50 077 367
Width: 560 mm (Overall) 50 077 365
Depth: 500 mm (Overall)
Thickness: 2 mm

3. HERAtray 1/4 Width, SS (3 pc.) 50 065 807


HERAtray 1/4 Width, Cu (3 pc.) 50 065 808
Width: 135 mm (Overall)
Depth: 485 mm (Overall)
Thickness: 1 mm

4. HERAtray 1/3 Width, SS (3 pc.) 50 065 805


HERAtray 1/3 Width, Cu (3 pc.) 50 065 806
Width: 180 mm (Overall)
Depth: 485 mm (Overall)
Thickness: 1 mm

5. Shelf for units with gas screen

Half width inc. support rails, SS 50 065 795


Half width inc. support rails, Cu 50 065 796
Width: 260 mm (Overall)
Depth: 500 mm (Overall)
Thickness: 1 mm

6. For units with gas screen

HERAtray 1/2 Width, SS (2 pc.) 50 065 809


HERAtray 1/2 Width, Cu (2 pc.) 50 065 810
Width: 125 mm
Depth: 485 mm
Thickness: 1 mm

Version: 07.2003 / 50078754 6


Accessories for CO2 - incubators

2.3 Stacking adapters

1. Plate for stacking 2 x HERAcell 240 50 068 677


Width: 780 mm
Depth: 772 mm

2. Plate for stacking a HERAcell 240 50 066 094


with a BB 6220 or BBD 6220 or
cytoperm 2 or B 5060 or B 5061
Width: 895 mm
Depth: 721 mm

Information:
The cytoperm 2 or BBD 6220 should
stand on the adaptor plate

Version: 07.2003 / 50078754 7


Accessories for CO2 - incubators

3 Accessories for cytoperm 2 / BBD 6220

3.1 Support frames

1. Support frame 300 mm 50 031 349


Width: 885 mm
Heidht: 300 mm
Depth: 645 mm

2. Support frame 780 mm 50 029 597


Width: 885 mm
Height: 780 mm
Depth: 645 mm

3. Stacking frame with castors for 50 053 628


stacking 2 x BBD 6220 or cytoperm 2

Support frame for lower unit:


Width: 920 mm
Height: 200 mm
Depth: 670 mm

Support frame for upper unit:


Width: 1080 mm
Height: 1215 mm
Depth: 770 mm

Information:
Alternatively another unit can be
stacked in the upper position, e.g.
BB6220 or HERAcell 240

Version: 07.2003 / 50078754 8


Accessories for CO2 - incubators

5. Castor set for 1 & 2 above. (Set of 4) 50 052 528

Information:
When fitted the castors will raise the
support frame by 90 mm.

Version: 07.2003 / 50078754 9


Accessories for CO2 - incubators

3.2 Shelves / HERAtrays

1. Shelf full width, SS, inc. support rails, 50 029945


Width: 560 mm (Overall)
Depth: 500 mm (Overall)
Thickness: 1 mm

3. HERAtray 1/4 Width, SS (3 pc.) 50 065 807


Width: 135 mm (Overall)
Depth: 485 mm (Overall)
Thickness: 1 mm

4. HERAtray 1/3 Width, SS (3 pc.) 50 065 805


Width: 180 mm (Overall)
Depth: 485 mm (Overall)
Thickness: 1 mm

5. Shelf for units with gas screen

Half width, SS, inc. support rails 50 029 943


Width: 260 mm (Overall)
Depth: 500 mm (Overall)
Thickness: 1 mm

6. For units with gas screen

HERAtray 1/2 Width, SS (2 pc.) 50 065 809


Width: 125 mm
Depth: 485 mm
Thickness: 1 mm

Version: 07.2003 / 50078754 10


Accessories for CO2 - incubators

3.3 Stacking adapters

1. Plate to stack BBD 6220, cytoperm 2 50 066 094


with a BB 6220 or HERAcell 240 or
B 5060 or B 5061
Width: 895 mm
Depth: 721 mm

Version: 07.2003 / 50078754 11


Accessories for CO2 - incubators

Anschrift:
Kendro Laboratory Products GmbH
Robert – Bosch – Straße 1
D - 63505 Langenselbold
Deutschland

Telefon:
Vertrieb/
Service + 49 (0) 6184 / 906940

Fax:
Vertrieb/
Service + 49 (0) 6184 / 907474

Version: 07.2003 / 50078754 12


Kendro Service & Support
Service Bulletin
Product Line Equipment Bulletin Issue Date

Incubator - Heraeus Spare Part 1204-LSB-INC-004 10. December.04

New TCD Fan

PURPOSE:

To inform field service of a spare part change.

EQUIPMENT AFFECTED:

BB 5060/BB 6060/BB 6220/Cytomat 6000 series

The spare part 50029957 Fan for TCD is no longer available.

DESCRIPTION / INFORMATION:

Spare part 50029957 will be replaced by 50082995 Fan comp. for TCD

The new TCD fan is fully compatible with existing incubators.

Important Note Although the connector is different (5 pin DIN) it will fit the
existing socket and the functionality is the same.

Author Reviewed By
S. Czwak R. Bloomfield
Kendro Service & Support
Service Bulletin
Product Line Equipment Bulletin Issue Date

Incubator - Heraeus Spare Part 1204-LSB-INC-004 10. December.04

ADDITIONAL INFORMATION:

Fan new (#50082995 )

Fan old (#50029957)

Author Reviewed By
S. Czwak R. Bloomfield
Kendro Service & Support
Service Bulletin
Product Line Equipment Bulletin Issue Date
BBK 6220, BB16, BK 6160,
BB 5060, B 5061, BB 6060/6220 Fan 0503-LSB-INC-009 12. April 2005

Fan for measure cell

Reason:

To inform the field of a wrongly manufactured connector.

EQUIPMENT AFFECTED:

Identnumber 50082995 Fan complete, TCD. As a spare part for the listed products
above.

DESCRIPTION / INFORMATION:

Checking the connections!

To perform this you need a multimeter. The pin description is shown in the picture on
the next page.

- Choose highest Ohm – measuring range


- Connect the „COM“(-) connection of the multimeter at PIN 2
- Connect the „V“ (+)connection of the multimeter at PIN 1

The measured value should be between 2 - 3 Mega Ohms.

This value rises during measuring because there is a capacitor in circuit.

If the measured value is infinite, the diode inside is in the reverse direction mounted.
Caution! If this fan is installed, it will cause a short circuit and destroy the capacitor.

Author Contact Information Reviewed By

F. Born R Bloomfield
Kendro Service & Support
Service Bulletin
Product Line Equipment Bulletin Issue Date
BBK 6220, BB16, BK 6160,
BB 5060, B 5061, BB 6060/6220 Fan 0503-LSB-INC-009 12. April 2005

PIN 4

PIN 2

PIN 1
PIN 5

PIN 3

ADDITIONAL INFORMATION:

Send any incorrectly wired fan back to Kendro Langenselbold.

Author Contact Information Reviewed By

F. Born R Bloomfield

You might also like