The new Heraeus IR-CO2 gas tester from Kendro Laboratory Products uses a
miniaturised infra-red cell to monitor CO2-concentrations in percent
concentration ranges.
Features
■ Maintenance-free infra-red cell
■ Measuring range: 0 – 60 % CO2
■ Automatic compensation for
changes in elevation, temperature,
and barometric pressure
■ Pumped model
■ Data storage and logging facilities
■ Compact and lightweight –
weighs only 250 g
■ 9 hours use on charge
■ Can be recharged
in approx. 2 hours
■ Backlight
▼
Australia Kendro Laboratory Products · Lane Cove, Sydney · NSW 2066 · Tel. +61 (2) 9936 1540 · Fax +61 (2) 9427 9765 · info@kendro.com.au
Austria Kendro Laboratory Products GmbH · Vienna · Tel. +43 (1) 801 40-0 · Fax +43 (1) 801 40 40 · office@kendro.at
Canada Kendro Laboratory Products International Sales · Newtown, CT · USA · Tel. 1 (203) 270-2080 · Fax 1 (203) 270-2166 · info@kendro.com
China Kendro Laboratory Products Beijing Rep. Office · Beijing · Tel. +86 (10) 6501-1573 · Fax +86 (10) 6501-4229 · kendrobj@163bj.com
Kendro Laboratory Products (H.K.) Limited · Hong Kong · Tel. +852 2711-3910 · Fax +852 2711-3858 · info@kendro.com
Kendro Laboratory Products Shanghai Rep. Office · Shanghai · Tel. +86 (21) 5490-0216 · Fax +86 (21) 5490-0230 · kendrosh@public4.sta.net.cn
Denmark Kendro Laboratory Products AB · Albertslund · Tel. +45-43 62 46 47 · Fax +45-43 62 46 41 · info@kendro.dk
France Kendro Laboratory Products · Courtaboeuf cedex · Tel. +33 (1) 69 18 77 77 · Fax +33 (1) 60 92 00 34 · info@kendro-lab.fr
Germany Kendro Laboratory Products GmbH · Hanau · Tel +49 (1805) 536 376 · Fax +49 (1805) 112 114 · info@kendro.de
India Kendro Laboratory Products (India) Pvt. Ltd. · New Delhi · Tel. +91 (11) 618 48 40 · Fax +91 (11) 618 53 97 · kendro.india@vsnl.com
Italy AHSI S.p.A. · Cavenago Brianza · Tel. +39 (02) 95 08 11 · Fax +39 (02) 95 08 12 77 · info@ahsi.it
Japan Nippon Kendro Co. Ltd. · Tokyo · Tel. +81 (0) 3 3517-1661 · Fax +81 (0) 3 3517-1664
New Zealand Kendro Laboratory Products Pty Ltd · Auckland · Tel. +64 (9) 525 03 33 · Fax +64 (9) 525 03 37
Poland Kendro Spólka z.o.o. · Warsaw · Tel. +48 (22) 663 43 23 · Fax +48 (22) 663 43 25 · kendro.warszawa@kendro.com.pl
Portugal Heraeus S.A. · Cacem · Tel. +351 (1) 912 08 65 · Fax +351 (1) 912 08 60 · heraeus@mail.telepac.pt
Spain Heraeus S.A. · Madrid · Tel. +34 (91) 358 19 96 · Fax +34 (91) 358 20 67 · laboratorio@heraeus.es
Sweden Kendro Laboratory Products AB · Sollentuna · Tel. +46 (8) 585 777 50 · Fax +46 (8) 623 15 45 · info@kendro.se
Switzerland Kendro Laboratory Products AG · Zurich · Tel. +41 (1) 454 12 12 · Fax +41 (1) 454 12 99 · kendro-ag@swissonline.ch
Kendro Laboratory Products SA · Carouge-Genève · Tel. +41 (22) 343 21 67 · Fax +41 (22) 342 38 31 · kendro-sa@swissonline.ch
U.K./Ireland Kendro Laboratory Products Ltd. · Bishop’s Stortford · Herts · Tel. +44 (1279) 827700 · Fax +44 (1279) 827750 · kendro@kendro.co.uk
USA Kendro Laboratory Products · Newtown, CT · Tel. +1 (800) 522-7746 · Fax +1 (203) 270-2166 · info@kendro.com
Internet http://www.kendro.com
H BB-GT 2e
Registered to ISO 9001. Kendro Laboratory Products meet or exceed stringent quality and product safety standards: CE for the European Union, and UL, cUL or CSA standards for
North America. ©2001 Kendro Laboratory Products. All Rights Reserved. Printed in Germany XC X/00 4t Druckerei
Operating Instructions
f1
f2
f3
Contents
3. Operation .............................................................................................................................................................. 6
3.1 Switching the Unit ON/OFF........................................................................................................................... 6
3.2 Measurement Procedure .............................................................................................................................. 6
3.3 Saving Measured Values .............................................................................................................................. 7
3.4 Battery Charge State .................................................................................................................................... 8
3.5 Display Backlight........................................................................................................................................... 8
3.6 Acoustic CO2 Level Output ........................................................................................................................... 8
3.7 Fault Message .............................................................................................................................................. 8
3.8 Software and Serial Number......................................................................................................................... 9
4. Calibration .......................................................................................................................................................... 10
4.1 Calibrating the Gas Tester .......................................................................................................................... 10
4.2 Calibrating the Zero-point ........................................................................................................................... 10
4.3 Calibrating the Metering Range .................................................................................................................. 10
6. Technical Specifications................................................................................................................................... 14
6.1 CO2 Gas Tester .......................................................................................................................................... 14
6.2 Mains Power Supply ................................................................................................................................... 14
Telephone
Sales ++ (0 61 81) 35 – 3 00
Spare Parts ++ (0 61 81) 35 – 4 38
Fax
Sales ++ (0 61 81) 35 – 59 73
1.3 Standard Delivery
Spare Parts ++ (0 61 81) 35 – 6 67
• Anagas CD 98 CO2 Gas Tester 1
• Mains power supply 1
• Silicon tubing, 1 meter 1
Copyright©
• Filter with Luer lock connection 3
Kendro Laboratory Products GmbH
• Operating instructions 1
• Carrying case 1
When using the gas tester, be sure the supplied filter The CO2 gas tester continuously monitors the charge
has been installed over the gas intake opening on the level of the battery. If “BATTERY LOW” appears, the
top of the unit and that the gas outlet opening on the unit should be turned off as soon as possible and
front of the unit, above the display, is not covered. The connected to the supplied charger. The unit must be
filter’s PTFE membrane prevents particles and water turned off during charging. The charger connection is
vapor from entering the unit’s metering cell. Operating located on the right side of the gas tester. During
the CO2 gas tester without a filter can lead to opera- charging (with the unit turned off), the message,
tional problems up to and including destruction of the “CHARGING”, is displayed. Charging is complete
unit. when an acoustic signal is not generated for a period
of 10 seconds, and the message, “CHARGED”, ap-
Replace the filter when it becomes discolored. pears.
Refer to the section “Spare Parts and Accessories” for
replacement filter information. Optimal charging takes place at ambient temperatures
between approx. 20 and 25 °C. Charging should not
Because of the highly humid air sucked into the unit be carried out at temperatures below 12 °C as this
during continuous metering of CO2 gas absorption in- may result in damage to the battery because the sys-
cubators, condensation will form on the filter’s PTFE tem is unable to recognize the actual charge level. We
membrane. For these types of applications, metering recommend charging the unit once every month, even
times should be restricted to a maximum of 10 min- if the gas tester was not used. If the battery has not
utes or other measures should be instituted to dehu- been used for several months, it should be charged
midify the intake air. During pauses in metering, the prior to use, then drained by operating the unit, and fi-
condensation on the filter will eventually dry and nally, recharged again. This charging cycle ensures
should not affect filter function. that the battery is fully charged.
2. Unit Description
The Anagas CD 98 CO2 Gas Tester is a portable unit • Storage of up to 100 current measured values
to measure carbon dioxide which employs a low- • Storage of up to 800 records (measured value,
volume infrared metering cell operating at two differ- date, and time)
ent wavelengths to measure CO2 concentration. The • Automatic compensation for changes in altitude,
gas tester is equipped with an integrated metering gas air pressure, and temperature
pump with a throughput of approx. 120 ml/min to sim- • Calibration function
plify the extraction of a gas sample. The metering • Battery charge indicator
range is from 0 to 60 % CO2. • Display backlighting
• IrDA interface for data transfer to a PC via PC
The gas tester can be used for either short-term COM software (accessory item, not part of the
measurements, whereby the results are stored, or can standard delivery).
be operated in the continuous mode if connected to
the mains power. In this case, the results can be
saved at specified intervals.
6 IrDA interface 1 6
% CO2
7 LED, metering gas pump ON 00.04
7
8 Dual-function button,
- Gas tester ON/OFF
- Metering gas pump ON/OFF
12 8
9 Charger connection socket f1
11 9
10 Dual-function button, f3 f2
- Display backlight ON
- Acoustic CO2 level output 10
f3
11 Dual-function button, f2
- Battery charge status
- Calibrate
3. Operation
3. Operation
NOTE - Software!
3. Operation
3. Operation
After you release the f1 button, the gas tester’s serial ID No.
number will be displayed for approx. 5 seconds. xxxx
4. Calibration
f1 -> N2
f3 -> Gas
4. Calibration
4. Calibration
• Press and hold down the f2 button for at least 3 User Cal
seconds, until the message, “User Cal”, appears
on the display.
• Release the f2 button. A selection menu appears
f1 -> N2
on the display.
f3 -> Gas
• Press the f3 button. The required CO2 concentra-
tion of the calibration gas appears on the top line,
while the current measured value is shown on the f1 -> 05.0
bottom line. 00.0
• As soon as the current measured value stabilizes,
press the f1 button. The message, „Gas Cal oK“,
appears on the display.
Gas Cal
• If the metering range calibration was not per-
oK
formed properly, a corresponding error message
appears on the display. In this case, repeat the
calibration. If the error message still appears, the
Gas Cal
gas tester must be returned for inspection.
failed
The manufacturer only warrants the safety and proper Only original spare parts and accessories tested and
operation of the CO2 gas tester under the conditions approved by Kendro Laboratory Products may be
that: used. The use of other parts carried unknown risks
and will void the unit’s warranty.
• The unit is used properly and in accordance with
these instructions; Order No. Spare part / Accessory
• No modifications are made; 50060286 Silicon tubing, 1 meter
• Only original manufacturer spare parts are used; 50060287 Filters, 5-pack
• The back of the unit has not been opened. 50060288 Mains power supply with adapters
50060289 PM-COM software with IrDA inter-
face adapter and cable
50060290 Operating Instructions
5.2 Maintenance
The unit is maintenance-free. Only the filter needs to
be replaced if the membrane becomes discolored.
Refer to the section “Spare Parts and Accessories” for
replacement filter information.
5.3 Cleaning
Occasionally wipe the plastic gas tester housing down
with a damp towel. Never using any cleaning or abra-
sive materials as these will damage the housing and,
in particular, the display window.
6. Technical Specifications
Telephone
Sales ++ (0 61 81) 35 – 3 00
Spare Parts ++ (0 61 81) 35 – 4 38
Fax
Sales ++ (0 61 81) 35 – 59 73
Spare Parts ++ (0 61 81) 35 – 6 67
1. Do not charge the unit whilst in use – the charging circuitry and battery will produce additional heat which
may effect the reading.
2. Do not use the unit immediately after charging – allow a few minutes to cool.
3. Before using the CO2 tester, switch on (with pump off) and allow to stabilise for 10 minutes before taking a
measurement.
4. The calibration should be checked regularly (every 2 months) using 5% calibration gas with a certificate of
analysis.
a) Carry out air zero – this part of the calibration must be carried out using air from outside the building.
i) Turn on CO2 tester (pump off) and allow to stabilise for 10 minutes.
ii) Turn on cylinder valve and set flow rate to just below the first graduation.
iii) Connect tube outlet to CO2 tester. – ensure pump is off.
iv) Wait for stable reading.
v) If reading is outside 5% +/- 5% of reading, re-calibration is required.
Calibrate as follows
Note – to achieve a factory reset switch on whilst pressing key F2 and keep held until display shows
“Reset OK”