You are on page 1of 136

Manual ver1.

EPX-Series
Portable X-ray Systems
Manual

D/N:ECM-P-SERISE,Rev1.3

1
Manual ver1.3

(This page intentionally left blank)

2
Manual ver1.3

CONTENTS

1. INTRODUCTION----------------------------------------------------------------------------------------6

2. NOTICE OF THE SAFE OPERATION ------------------------------------------------------------8

3. RADIATION SAFETY ----------------------------------------------------------------------------------9

4. SAFETY AND SPECIFICATIONS -----------------------------------------------------------------10

5. SYSTEM CONFIGURATION------------------------------------------------------------------------25

6. OPERATION MODE ----------------------------------------------------------------------------------27

7. MAINTENANCE PROCEDURE -------------------------------------------------------------------29

8. QUALITY WARRANTY ------------------------------------------------------------------------------30

SERVICE CHAPTER

1. CALIBRATION-------------------------------------------------------------------------------------------31

2. DIP SWITCH SETTING--------------------------------------------------------------------------------37

3 .HARD-WIRED hand switch control---------------------------------------------------------------38

4. COLLIMATOR---------------------------------------------------------------------------------------------39

5. TROUBLE SHOOTING---------------------------------------------------------------------------------41

6. SERVICE CHECK LIST---------------------------------------------------------------------------------42

3
Manual ver1.3

REVISON HISTORY
Revision Number Date Description

A SEP 11,2013 First Edition

ADVISORY SYMBLOS
The following advisory symbols are used throughout this manual.
Their application and meaning are described below.

WARNING

Warning symbol used to indicate a potential hazard lead operators, service personnel to serious injury, death or
radiation exposure.

CAUTION

Caution symbol used to indicate a potential hazard lead operators, service personnel to injury or damage of
equipment.

NOTE

Note symbol used to indicate important information need to proper use and right operation of equipment.

Keep this service manual with the equipment at all times, and review the important information whenever
required.

4
Manual ver1.3

(This page intentionally left blank)

5
Manual ver1.3

1.INTRODUCTION

The EPX Series Innovation portable x-ray system features excellent x-ray Image, quality, accuracy,
reproducibility and long-term stability. Fast kV rising time minimize unnecessary patient dose and supply
superior Image quality, fast Image response.

The operator control console is designed to be simple and user-friendly and one can select or change x-
ray parameters easily using a large soft membrane switch.

Intended use
The EPX Series is intended for use to produce X-ray images of patient during medical operations. The
EPX Series consists of a Main power board and OP control board, SMPS, FET drive board, X-ray
generator.
This system can be applied in emergency room, operation room, cast room, outside room or tec. of
hospital.

6
Manual ver1.3

1.1 COMPOSITION
This product is composed of the X-ray component part, accessories part. Detail information of each component
follows below.
(1) X-ray component part
 Main Power board

 SMPS

 Mono block X-ray tank

 OP control board

 FET drive board

 Collimator

(2) Accessories
 Hand switch

 Foot switch(Option)

 Operator’s Manual

This manual contains important safety information.


An understanding of this information is critical to the safe operation of your
equipment.
Please ensure that you read the warning notices before using the equipment

7
Manual ver1.3

2. NOTICE OF THE SAFE OPERATION

This user’s guide is designed to ensure correct use and operation of EPX series. Please read all the
lines thoroughly before you use this equipment.

Incorrect use and operation exceeding described conditions in this manual may occur damage of the
machine and shorten its life. Particular attention must be paid to all the warnings, cautions and notes
incorporated herein.

This equipment should be used only by the legally qualified persons and practitioners.

EPX series is designed with the due consideration for users’ safety and product reliability. It, however, is
advisable to follow under mentioned rules to keep your additional safety and health.

1. This product should be operated only by or under the supervision of legally qualified persons.

2. EPX series is designed for the radiographic uses and not for fluoroscopy or other associated
applications.

3. EPX series should be used for the diagnosis, not for the therapy.

4. Do not modify the equipment at your discretion and in case any modification is required unavoidably,
ask the help of Ecotron or its authorized dealer for the service.

5. This system has been calibrated for optimal operations.

No protection against the ingress of the liquids

8
Manual ver1.3

3. RADIATION SAFETY

1. Users and operators should wear appropriate protecting devices and clothes.

2. Stay distance from the radiant sources and all the possible secondary radiation zones.

3. Eliminate all unnecessary objects near the exposure zones.

4. The distance from the focus to skin should be kept at least 8 inch (20cm).

5. For the experimental uses, apply the lowest possible values of kV, mAs and exposure time.

6. Be careful not to exceed the limited radiograms in the exposure area.

The ionizing radiation could be dangerous for patients and operators


unless following safety regulations are strictly observed.

9
Manual ver1.3

4. SAFETY AND SPECIFICATIONS


4.1 SAFETY AND WARNING SYMBOLS

Never allow unqualified personnel to operate the X-ray


system.

High voltage symbol used to indicate the presence of


high voltage.

Warning symbol used to indicate a potential hazard to


operators, service personnel or to the equipment.. It
indicates a requirement to refer to the accompanying
documentation for details.

HV 3 MINUTES WARNING LABEL

This label is attached on the main power board.


The DC link capacitor(approximately 325 VDC with line voltage
at 230VAC) will remain charged for up to 3 minutes after the AC
mains is disconnected or the AC switch is switched off.

DANGER HIGH VOLTAGE LABEL


This label attached to the main fuse cover, main noise filter
cover and FET driver board. Main voltage is present inside the x-
ray system whenever the main disconnect is switched on.
Additionally, the DC link capacitors will remain charged for up to
3 minutes after the AC mains is disconnected or the AC switch is
switched off.

10
Manual ver1.3

Protection earth symbol

L Live line among the single phase line powers.

N Neutral line among the single phase line powers.

L1 First phase line power among the three phase line


powers.

L2 Second phase line power among the three phase line


powers.

L3 Third phase line power among the three phase line


powers.

V~ Single phase AC line voltage.

X-ray radiation exposure may be damaging to health, with some effects being cumulative and extending
over periods of many months or even years. X-ray operators should avoid any exposure to the
primary beam and take protective measures to safeguard against scatter radiation. Scatter radiation is
caused by any object in the path of the primary beam and may be of equal or less intensity than the
primary beam that exposes the film.
No practical design can incorporate complete protection for operators or service personnel who do not
take adequate safety precautions. Only authorized and properly trained service and operating
personnel should be allowed to work with this X-ray system equipment. The appropriate personnel
must be made aware of the inherent dangers associated with the servicing of high voltage equipment
and the danger of excessive exposure to X-ray radiation during system operation.

● Wear protective clothing. Protective aprons and gloves with an equivalent of a minimum of
1/64”(0.35mm) of lead are recommended.
● To protect the patient against radiation, always use radiation protection accessories in addition to
devices which are fitted to the X-ray system
● Keep as large a distance as possible away from the object being exposed and the X-ray tube assembly.
● Never operate this X-ray system in areas where there is a risk of explosion. Detergents and

11
Manual ver1.3

disinfectants, including those used on patients, may create explosive mixtures of gases. Please
observe the relevant regulations.
● The operator console or anything electrically connected to it, must never be used within 6ft(1.8m) of
the patient environment.
● Do not place liquids(coffee, beverages, flowers, etc) on the X-ray system.
● Do not operate the x-ray system in direct sunlight or near any heat sources.
● Do not operate the x-ray system near strong magnetic fields(microwave ovens, speakers, etc), and
avoid routing the x-ray system near these devices.
● The x-ray system must be operated in locations that are clean(free of excess dust, dirt, debris, etc),
stable(free of vibration).
● Only trained maintenance staff may remove the covers of the x-ray system.

12
Manual ver1.3

4.2 COMPOSITION LABEL

5.1 EPX-F1600

IDENTIFICATION LABEL-HIGH TENSION TANK

HV MONO BLOCK TANK


This is label is attached on side of the HV mono block
MODEL : EMB-F1600
SN
: tank inside of the Portable x-ray system.(Example)
Input Power : AC200~240V/Single phase 50/60Hz
Max. output Power : 1.6kW
kV Range: 40kV~90kV
mAs Range:0.4~32mAs/ mA Range: 13~30mA

Inserted X-ray Tube


Model : D-0814
SN :
Focal spot : 0.8mm / Target Angle : 16degrees.
Anode Heat Storage Capacity: 7kJ
Inherent Filtration : 0.5mmAl

Manufacture: ECOTRON Co.,Ltd


701Ho, Hanshin IT Tower II, 60-18, Gasan-Dong, Geumcheon-Gu,
Seoul Korea. 153-023 / TEL:+82-2025-3760

PET LINE A.E


FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004
SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924

REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM

PORTABLE X-RAY UNIT


This label is attached to rear cover of the Portable x-ray
MODEL : EPX-F1600
SN
: system
Input Power : AC200~240V/Single phase 50/60Hz
Max. output Power : 1.6kW (Example)
kV Range: 40kV~90kV
mAs Range:0.4~32mAs/ mA Range: 13~30mA

Inserted X-ray Tube


Model : D-0814
SN :
Focal spot : 0.8mm / Target Angle : 16degrees.
Anode Heat Storage Capacity: 7kJ
Inherent Filtration : 0.5mmAl

Manufacture: ECOTRON Co.,Ltd


701Ho, Hanshin IT Tower II, 60-18, Gasan-Dong, Geumcheon-Gu,
Seoul Korea. 153-023 / TEL:+82-2025-3760

PET LINE A.E


FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004
SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924

13
Manual ver1.3

5.2 EPX-F2400

IDENTIFICATION LABEL-HIGH TENSION TANK

This is label is attached on side of the HV mono block


tank inside of the Portable x-ray system.(Example)

REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM

This label is attached to rear cover of the Portable x-ray


system
(Example)

14
Manual ver1.3

5.3 EPX-F2800

IDENTIFICATION LABEL-HIGH TENSION TANK

This is label is attached on side of the HV mono block


tank inside of the Portable x-ray system.(Example)

REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM

This label is attached to rear cover of the Portable x-ray


system
(Example)

15
Manual ver1.3

5.4 EPX-F3200

IDENTIFICATION LABEL-HIGH TENSION TANK

HV MONO BLOCK TANK


This is label is attached on side of the HV mono block
MODEL : EMB-F3200
SN
:
Input Power : AC200~240V/Single phase 50/60Hz
tank inside of the Portable x-ray system.(Example)
Max. output Power : 3.2kW
kV Range: 40kV~100kV
mAs Range:0.4~100mAs/ mA Range: 25~60mA

Inserted X-ray Tube


Model : D-205B
SN :
Focal spot : 2.0mm / Target Angle : 20degrees.
Anode Heat Storage Capacity: 28kJ
Inherent Filtration : 0.5mmAl

Manufacture: ECOTRON Co.,Ltd


701Ho, Hanshin IT Tower II, 60-18, Gasan-Dong, Geumcheon-Gu,
Seoul Korea. 153-023 / TEL:+82-2025-3760

PET LINE A.E


FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004
SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924

REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM

PORTABLE X-RAY UNIT


This label is attached to rear cover of the Portable x-ray
MODEL : EPX-F3200
SN
:
Input Power : AC200~240V/Single phase 50/60Hz
system
Max. output Power : 3.2kW
kV Range: 40kV~100kV (Example)
mAs Range:0.4~100mAs/ mA Range: 25~60mA

Inserted X-ray Tube


Model : D-205B
SN :
Focal spot : 2.0mm / Target Angle : 20degrees.
Anode Heat Storage Capacity: 28kJ
Inherent Filtration : 0.5mmAl

Manufacture: ECOTRON Co.,Ltd


701Ho, Hanshin IT Tower II, 60-18, Gasan-Dong, Geumcheon-Gu,
Seoul Korea. 153-023 / TEL:+82-2025-3760

PET LINE A.E


FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004
SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924

16
Manual ver1.3

5.5 EPX-F4000

IDENTIFICATION LABEL-HIGH TENSION TANK

This is label is attached on side of the HV mono block


tank inside of the Portable x-ray system.(Example)

REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM

This label is attached to rear cover of the Portable x-ray


system
(Example)

17
Manual ver1.3

5.6 EPX-F5000
IDENTIFICATION LABEL-HIGH TENSION TANK

This is label is attached on side of the HV mono block


tank inside of the Portable x-ray system.(Example)

REPRESENTATIVE IDENTIFICATION LABEL X-RAY SYSTEM

This label is attached to rear cover of the Portable x-ray


PORTABLE X-RAY UNIT
MODEL : EPX-F5000
system
:
(Example)
SN

Input Power : AC200~240V/Single phase 50/60Hz


Max. output Power : 5kW
kV Range: 40kV~120kV
mAs Range:0.4~200mAs/ mA Range: 20~100mA

Inserted X-ray Tube


Model : CEI OX/110-15
SN :
Focal spot : 1.8mm / Target Angle : 15degrees.
Anode Heat Storage Capacity: 30kJ
Inherent Filtration : 0.5mmAl

Manufacture: ECOTRON Co.,Ltd


701Ho, Hanshin IT Tower II, 60-18, Gasan-Dong, Geumcheon-Gu,
Seoul Korea. 153-023 / TEL:+82-2025-3760

PET LINE A.E


FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004
SPATA – GREECE TEL:+30-210-6069837,FAX:+30-210-6634924

18
Manual ver1.3

4.3 Symbol on the Packing of X-ray system

SYMBOL Description

KEEP DRY: This symbol means keep away from rain

Fragile: This symbol means do not drop of product

UP: This symbol means “This way up”

INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY


RESULT IN INJURY OR EQUIPMENT DAMAGE.

THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS


SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE
OBSERVED.

19
Manual ver1.3

PROPER USE AND SAFE OPERATING PRACTICES WITH RESPECT TO X-RAY


SYSTEMS ARE THE RESPONSIBILITY OF THE USERS OF SUCH X-RAY
SYSTEMS.
ECOTRON.,Ltd PROVIEDS INFORMATION ON ITS PRODUCTS AND
ASSOCIATED HAZARDS, BUT ASSUMES NO RESPONSIBILITIES FOR AFTER-
SALE POERATING AND SAFETY PRACTICES.
ECOTRON.,Ltd ACCEPTS NO RESPONSIBILITY FOR ANY GENERATOR NOT
MAINTAINED OR SERVICED ACCORDING TO THE SERVICE MANUAL OR ANY
X-RAY SYSTEM THAT HAS BEEN MODIFIED IN ANY WAY.
ECOTRON.,Ltd ALSO ASSUMES NO RESPONSIBILITY FOR X-RAY RADIATION
OVEREXPOSURE OF PATIENTS OR PERSONNEL RESULTING FROM POOR
OPERATING TECHNIQUES OR PROCEDURS.

DO NOT EXCEED THE TUBE MAXIMUM POERATING LIMITS SHOWN IN THE X-


RAY TUBE DATA SECTION AT THE END OF THE POERATOR’S MANUAL.
INTENDED LIFE AND RELIABILITY WILL NOT BE OBTAINED UNLESS X-RAY
SYSTEMS ARE POERATED WITHIN PUBLISHED SPECIFICATION.

20
Manual ver1.3

4.4. EPX-Series DUTY CYCLE LIMIT

THE FOLLOWING SECTION CONTAINS IMPORTANT INFORMATION. PLEASE


READ AND UNDERSTAND THIS MATERIAL BEFORE CONTINUING.

Internal X-ray system components will heat up during normal use of the x-ray system. This is similar to X-ray tube
heating during normal x-ray system operation. The amount of heat produced is proportional to the product of kV,mA
and time.

THIS X-RAY SYSTEM HAS TEMPERATURE MONITORING OF FET-DRIVE


BOARD TO PROTECT THE EXCESSIVE HEAT BUILD-UP.
IF THE X-RAY SYSTEM DETECTS THE OVER-HEAT OF FET-DRIVE, AN ERROR
CODE “E04” OR “E05” WILL BE DISPLAYED. EXPOSURE WILL BE INHIBITED
WHEN THIS MESSAGE IS DISPLAYED, AND IT SHOULD BE UNDERSTOOD
THAT CONTINUING TO MAKE EXPOSURES MIGHT CAUSE X-RAY SYSTEM
DAMAGE DUE TO OVERHEATING. THE GENERATOR SHOULD BE ALLOWED
TO COOL SUFFICIENTLY SUCH THAT THIS MESSAGE IS NO LONGER
DISPLAYED.

21
Manual ver1.3

4.5 DAILY X-RAY TUBE WARM-UP PROCEDURE

THE FOLLOWING PROCEDURE PRODUCES X-RAYS. OBSERVE ALL SAFETY


PRECAUTIONS TO PROTECT PERSONNEL.

Use the procedure when the generator has not been used for several days. This procedure provides for exposures
at medium power before the tube is used at maximum mA or kV values. This will reduce the possibility of damaging
the X-ray tube and high voltage components. No test setup is required.
For maximum stability and reliability, use the following techniques at start up:
Select the following:

● 60kV/5mA.

● Make three to five exposures at 30 second intervals.

22
Manual ver1.3

4.6 APPLICABLE STANDARDS

The EPX series comply with the regulatory requirements and design standards in this section as follows:
This product complies with DHHS requirements of 21 CFR Subchapter J.

1) SAFETY
● IEC 60601-1:1988+A1:1991+A2:1995
● IEC 60601-2-7:1998(2nd Edition)

2) EMC
● EN60601-2:2007
● EN55011:2007+A2:2007GROUP 1 CLASS A
● EN61000-3-2:2000+A2:2005
● EN61000-3-3:1995+A1:2001
The EPX series X-ray system with CE marking comply with the European Council Directive concerning Medical
Devices. One of the harmonized standards of this Directive defines the permitted levels of electromagnetic
emission from this equipment and its required immunity from the electromagnetic emission of other devise.
It is not possible, however to exclude with absolute certainty the possibility that other high frequency electronic
equipment, which is fully compliant to the EMC regulations, will not adverse effect the operation of this x-ray system.
If the other equipment has a comparatively high level of transmission power and is in close proximity to the
generator, these EMC concerns (the risk of interference) may as mobile telephones, cordless microphones and
other similar mobile radio equipment be restricted from the vicinity of this X-ray system.

3) OTHERS

● EN980:2008 Graphical symbols for use in the labeling of medical devices


● IECTR60878:2003 Graphical symbols for electrical equipment n medical practice
● IEC60417-1:2007 Graphical symbols for use on equipment-part1:overview and application
● ENISO14971:2007 Medical devices
-Application of risk management to medical devices(ISO 14971:2007)
● ENISO13485:2003 Medical devices –
Quality management systems-Requirements for regulatory purpposes
● ISO9001:2008 Quality management systems-Requirements
● 93/42/EEC:1993 Council Directive concerning medical devices
● EN1041:2008 Information supplied by the manufacturer with medical devices
● EN62304:2006 Medical device software-Software lifecycle processes

23
Manual ver1.3

4) CLINICAL EVALUATION

● EN ISO 14155-1:2003 Clinical investigation of medical devices for human subjects - Part1 : General
requirements(ISO14155-1:2003)
● MEDDEV.2.7.1 Evaluation of Clinical Data: A Guide for Manufacturers and Notified Bodies

24
Manual ver1.3

4.7 SPECIFICATION
4.7.1 SPECIFICATION TABLE
5.1 EPX-F1600

Max Output Power 1.6 kW


Input Power Voltage 110 V-120V / 220V-240V
Phase & Frequency Single / 50/60 Hz
Tube voltage current mAs
40KV ~ 50KV 30mA 0.4 ~ 20
40KV ~ 50KV 25mA 25~32
51KV ~ 60KV 25mA 0.4 ~ 20
51KV ~ 60KV 20mA 25~32
Radiography kV Range in 1kV steps 61KV ~ 70KV 20mA 0.4 ~ 20
61KV ~ 70KV 15mA 25~32
71KV ~ 80KV 18 mA 0.4 ~ 20
71KV ~ 80KV 14 mA 25~32
81KV ~ 90KV 16 mA 0.4 ~ 20
81KV ~ 90KV 13 mA 25~32
MAs Range 0.4mAs – 32mAs, 20steps
Max. kV Deviation ±3 %
Max. mAs Deviation ±5 %
Display KV /mAs: 7-segment LED
Model Name D-0814

Focal Spot 0.8mmx0.8mm


X-ray Tube
Target Angle 16 degree
Anode Heat Storage 7 KHU
Total Filtration Tube:0.5mmAl/collimator:0.5mmAl/filter:1.5m
mAl
2.5 mm Al eq. @ 80kV
Min. X-ray Field Size 5cm x 5cm @1m SID
Collimator
Max X-ray Field Size 40cm x 40cm @ 75cm SID
with Laser
Timer Push button illuminator with 30 sec timer
Pointer
Lamp 12V 50W Halogen
Weight 6.8 Kg

25
Manual ver1.3

5.2 EPX-F2400

Max Output Power 2.4 KW


Input Power Voltage 110 V-120V / 220V-240V
Phase & Frequency Single, 50/60 Hz
Tube voltage current mAs
40KV ~ 60KV 40mA 0.4 ~ 100
61KV ~ 70KV 30mA 0.4 ~ 64
61KV ~ 70KV 25mA 80 ~ 100
71KV ~ 80KV 30mA 0.4 ~ 32
Radiography KV Range in 1KV steps
71KV ~ 80KV 25mA 40 ~80
81KV ~ 90KV 25mA 0.4 ~ 40
81KV ~ 90KV 20mA 50 ~ 80
91KV ~ 100KV 20mA 0.4 ~ 50
91KV ~ 100KV 16mA 64 ~ 80
mAs Range 0.4mAs – 100mAs, 25steps
Max. KV accuracy ±3 %
Max. mAs Deviation ±5 %
Display KV / mAs: 7-segment LED
Model TOSHIBA D-125S

Focal Spot Size 1.2mm x 1.2mm


X-ray Tube Target Angle 16 degree
Anode Heat Storage 35 kHU
Heat Dissipation 250 HU/sec
Total Filtration 2.5 mm Al eq. @ 100kV
Type Double slit type, manual operation
Min. X-ray Field Size 5cm x 5cm @1m SID
Collimator
Max X-ray Field Size 40cm x 40cm @ 75cm SID
with Laser
Laser Pointer Class ⅢA, 5mW
Pointer
Timer Push button illuminator with 30 sec timer
Lamp 12V, 50W Halogen
Weight 11 Kg

26
Manual ver1.3

5.3 EPX-F2800
Max Output Power 2.8 kW
Input Power Voltage 110 V-120V / 220V-240V
Phase & Frequency Single / 50/60 Hz
Tube voltage current mAs
40KV ~ 60KV 40mA 0.4 ~ 100
61KV ~ 70KV 30mA 0.4 ~ 64
61KV ~ 70KV 25mA 80 ~ 100
71KV ~ 80KV 35mA 0.4 ~ 10
71KV ~ 80KV 30mA 13 ~ 32
71KV ~ 80KV 25mA 40 ~ 80
81KV ~ 90KV 25mA 0.4 ~ 10
Radiography kV Range in 1kV steps 81KV ~ 90KV 25mA 13 ~ 40
81KV ~ 90KV 20mA 50 ~ 80
91KV ~ 100KV 20mA 0.4 ~ 13
91KV ~ 100KV 20mA 16 ~ 50
91KV ~ 100KV 16mA 64 ~ 80
101KV ~ 110KV 18mA 0.4 ~ 32
101KV ~ 110KV 16mA 40 ~ 64
111KV ~ 120KV 16mA 0.4 ~ 20
111KV ~ 120KV 14mA 25 ~ 50
MAs Range 0.4mAs – 100mAs, 25steps
Max. kV Deviation ±3 %
Max. mAs Deviation ±5 %
Display KV /mAs: 7-segment LED
Model Name D-125S TOSHIBA

Focal Spot 1.2mmx1.2mm


X-ray Tube Target Angle 16 degree
Anode Heat Storage 35 kHU
Heat Dissipation 250 HU/sec
Total Filtration 2.5 mm Al eq. @ 120kV
Min. X-ray Field Size ≦5cm x 5cm @1m SID
Collimator Max X-ray Field Size 40cm x 40cm @ 75cm SID
with Laser Laser Pointer Class : ⅢA 5mW
Pointer Timer Push button illuminator with 30 sec timer
Lamp 12V 50W Halogen
Weight 11.5 Kg

27
Manual ver1.3

5.4 EPX-F3200

Max Output Power 3.2 KW


Input Power Voltage 220V-240V
Phase & Frequency Single / 50/60 Hz
Tube voltage current mAs
40KV ~ 50KV 60mA 0.4 ~ 100
51KV ~ 60KV 50mA 0.4 ~ 80
61KV ~ 70KV 40mA 0.4 ~ 100
71KV ~ 80KV 40mA 0.4 ~ 32
Radiography kV Range in 1KV steps
71KV ~ 80KV 35mA 40 ~80
81KV ~ 90KV 30mA 0.4 ~ 40
81KV ~ 90KV 25mA 50 ~ 80
91KV ~ 100KV 30mA 0.4 ~ 50
91KV ~ 100KV 25mA 64 ~ 80
MAs Range 0.4mAs – 100mAs, 25steps
Max. kV Deviation ±3 %
Max. mAs Deviation ±5 %
Display KV /mAs: 7-segment LED
Model Name D-205S TOSHIBA

Focal Spot 2.0mmx2.0 mm


X-ray Tube Target Angle 16 degree
Anode Heat Storage 28 kHU
Heat Dissipation 265 HU/sec
Total Filtration 2.5 mm Al eq. @ 100KV
Type Double slit type, Manually operation
Min. X-ray Field Size ≦5cm x 5cm @1m SID
Collimator
Max X-ray Field Size 40cm x 40cm @ 75cm SID
with Laser
Laser Pointer Class : ⅢA 5mW
Pointer
Timer Push button illuminator with 30 sec timer
Lamp 12V 50W Halogen
Weight 12 Kg (26.4 lb.)

28
Manual ver1.3

5.5 EPX-F4000

Max Output Power 4 KW


Input Power Voltage 200V-250V
Phase & Frequency Single / 50/60 Hz
Tube voltage current mAs Sec
50KV 80mA 0.4~16mAs 0.2
50KV 70mA 18~32mAs 0.45
50KV 60mA 40~100mAs 1.666
51KV ~ 60KV 60mA 0.4~32mAs 0.533
51KV ~ 60KV 50mA 40~100mAs 2
61KV ~ 70KV 50mA 0.4~32mAs 0.64
61KV ~ 70KV 40mA 40~100mAs 2.5
Radiography kV Range in 1KV steps
71KV ~ 80KV 50mA 0.4~32mAs 0.64
71KV ~ 80KV 40mA 40~80mAs 2
81KV ~ 90KV 40mA 0.4~32mAs 0.8
81KV ~ 90KV 30mA 40~80mAs 2.666
91KV ~ 100KV 35mA 0.4~32mAs 0.914
91KV ~ 100KV 25mA 40~80mAs 3.2
101kV ~ 110KV 30mA 0.4~32mAs 1.066
101kV ~ 110KV 20mA 40~80mAs 4
MAs Range 0.4mAs – 100mAs
Max. kV Deviation ±3 %
Max. mAs Deviation ±5 %
Display KV /mAs: 7-segment LED
Model Name CEI OX/105-2

Focal Spot 1.8mmx1.8mm


X-ray Tube
Target Angle 19 degree
Anode Heat Storage 30000J
Total Filtration 2.5 mm Al eq. @ 100KV
Type Double slit type, Manually operation
Min. X-ray Field Size ≦5cm x 5cm @1m SID
Collimator
Max X-ray Field Size 40cm x 40cm @ 75cm SID
with Laser
Laser Pointer Class : ⅢA 5mW
Pointer
Timer Push button illuminator with 30 sec timer
Lamp 12V 50W Halogen

29
Manual ver1.3

5.6 EPX-F5000

Max Output Power 5 KW


Input Power Voltage 200V-240V
Phase & Frequency Single / 50/60 Hz
Tube voltage current mAs Sec
50KV 100mA 0.4~10mAs 0.1
50KV 80mA 13~32mAs 0.4
50KV 70mA 40~100mAs 1.42
51KV ~ 60KV 80mA 0.4~10mAs 0.2
51KV ~ 60KV 70mA 13~32mAs 0.457
51KV ~ 60KV 60mA 40~100mAs 1.666
61KV ~ 70KV 70mA 0.4~10mAs 0.142
61KV ~ 70KV 60mA 13~32mAs 0.533
Radiography kV Range in 1KV steps 61KV ~ 70KV 50mA 40~100mAs 2
71KV ~ 80KV 60mA 0.4~10mAs 0.166
71KV ~ 80KV 50mA 13~32mAs 0.64
71KV ~ 80KV 40mA 40~100mAs 2.5
81KV ~ 90KV 40mA 0.4~32mAs 0.8
81KV ~ 90KV 30mA 40~100mAs 3.33
91KV ~ 100KV 35mA 0.4~32mAs 1.066
91KV ~ 100KV 30mA 40~100mAs 3.33
101kV ~ 110KV 30mA 0.4~32mAs 1.066
101kV ~ 110KV 20mA 40~100mAs 5
MAs Range 0.4mAs – 100mAs
Max. kV Deviation ±3 %
Max. mAs Deviation ±5 %
Display KV /mAs: 7-segment LED
Model Name CEI OX/110-15

Focal Spot 1.8mmx1.8mm


X-ray Tube
Target Angle 15 degree
Anode Heat Storage 30000J
Total Filtration 2.5 mm Al eq. @ 100KV
Type Double slit type, Manually operation
Collimator
Min. X-ray Field Size ≦5cm x 5cm @1m SID
with Laser
Max X-ray Field Size 40cm x 40cm @ 75cm SID
Pointer
Laser Pointer Class : ⅢA 5mW

30
Manual ver1.3

Timer Push button illuminator with 30 sec timer


Lamp 24V 50W Halogen

31
Manual ver1.3

4.7.2 kV, mAs, mA TABLE

1) EPX-F1600

2) EPX-F2400

mAs TABLE
40KV ~ 60KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
Tube current : 40mA 35mA
61KV ~ 70KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
30mA
71KV ~ 80KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
30mA 25mA
81KV ~ 90KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
25mA 20mA
91KV ~ 100KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
20mA

32
Manual ver1.3

2) EPX-F2800

mAs TABLE
40KV ~ 60KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
Tube current : 40mA(0)
61KV ~ 70KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
30mA(0) 25mA(0)
71KV ~ 80KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
35mA(1) 30mA(1) 25mA(1)
81KV ~ 90KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
30mA(2) 25mA(2) 20mA(0)
91KV ~ 100KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
25mA(3) 20mA(1) 16mA(0)

101KV ~ 110KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64
18mA(0) 16mA(1)
111KV ~ 120KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50
16mA(2) 14mA(0)

3) EPX-F3200

mAs TABLE
40KV ~ 50KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
Tube current : 60mA(0)
51KV ~ 60KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
50mA(0)
61KV ~ 70KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
40mA(0)
71KV ~ 80KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
40mA(1) 35mA(0)
81KV ~ 90KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
30mA(0)
91KV ~ 100KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
30mA(1) 25mA(0)

33
Manual ver1.3

4) EPX-F4000

5) EPX-F5000

40~50KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100 125 150 175 200
100mA(0) 80mA(0) 70mA(0) 50mA(0)

51~60KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100 125 150
80mA(1) 70mA(1) 60mA(0) 45mA(0)

61~70KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
70mA(2) 60mA(1) 40mA(0)

71~80KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
60mA(2) 50mA(1) 35mA(0)

81~90KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
40mA(1) 30mA(0)

91~100KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
35mA(0) 25mA(0)

101~110KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
30mA(1) 20mA(0)

111~120KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
25mA(0) 20mA(1)

34
Manual ver1.3

4.7.3 MECHANICAL SPECIFICATION

EPX-F1600 Dimension (unit : mm)

Side view

Front view

35
Manual ver1.3

EPX-F2400/F2800/F3200/F4000/F5000 Dimension (unit : mm)

Side view

Front view

36
Manual ver1.3

4.8 WORKING ENVIRONMENT

Avoid the following places for the normal operation and safe storage:

a. where the equipment is exposed to water vapor.


b. where the equipment is exposed to direct sunlight.
c. where the equipment is exposed to dust.
d. where the equipment is exposed to high humidity.
e. where there is a ventilation problem.
f. where the equipment is exposed to salty atmosphere.
g. where the equipment is exposed to chemicals or gas.

For safe operation of the equipment, it should be kept away from the place with strong vibration and
maintain proper environment and conditions.

Operation Environment

Temperature range 50°F ~ 104°F


Relative Humidity Range 30% ~ 75% RH

Optimal Temperature and Humidity

Temperature range 62°F ~ 73°F


Relative Humidity Range 40% ~ 60% RH

For safe storage and transportation, the following range of temperature and humidity must be kept.

Environment for storage and transportation

Temperature range -13°F ~ +140°F


Relative Humidity Range 10% ~ 95% RH

37
Manual ver1.3

4.9 CUSTOMER SUPPORT


Address any questions regarding X-ray system operation to:

ECOTRON Co.,Ltd
7th Fl, Hanshin IT Tower II, 60-18, Gasang-Dong,
Geumcheon-Gu, Seoul 153-023, Korea
TEL:+82-2-2025-3760, FAX:+82-2-2025-3764
E-mail:Ecotron@ecotron.co.kr
Web-site:http://www.ecotron.co.kr

In USA
Contact DRGEM USA Inc.
7018 NW 50TH Terrace, Gainsville, Florida, 32653, USA
TEL:201-370-6672, FAX:352-337-1271
E-mail:drgemusa@yahoo.com

PET LINE A.E


FINOS FILM ROAD - THESI CHOUSMIZA -P.O. BOX 77 -19004 SPATA – GREECE
TEL:+30-210-6069837,FAX:+30-210-6634924

38
Manual ver1.3

5. SYSTEM CONFIGURATION

EPX Series is a radiological device for veterinary uses. It should be applied for the radiographic
diagnosis and operated by qualified practitioners. Users have to comply with safety and health
regulations concerning the ionizing radiation protection and the electrical and mechanical safety of the
veterinary medical devices.

PREPARATION FOR OPERATION

1. Operation Practices

a. Wear a lead apron while making exposures.

b. Stay at least 6.5feet (2m) away from the unit or extend the cord of hand switch as long as possible
in order to secure enough space for your safety.

c. Use proper field size and technical values for each procedure to minimize x-ray exposure dose and
obtain the best radiographic result.

d. When you make exposure to the patient on the bed, ask visitors to leave the room first and keep
appropriate distance from the patient.

e. Pay attention to the maintenance schedule of the device and keep up with it.

f. Cumulative radiation dose usually does not exceed recommended maximum levels. However, if you
make radiographic exposures using high kV and mAs quite often, the evaluation of the specialist is
needed to check whether extra protective devices are necessary for the user or not.

2. Positioning of the patient against the collimator

a. Place a loaded cassette on the patient’s backside.

b. Arrange the SID (Source to Image Receptor Distance) using a tape measure located on the side of
the collimator.

c. Turn on the collimator lamp and laser pointer by the pushing collimator/laser switch is on.

d. Adjust the size of the x-ray beam in accordance with the film size using the x-ray field

39
Manual ver1.3

adjustment knobs. The collimator and laser pointer will be turned off after 30 seconds by internal
timer.

Please check whether the voltage and frequency of the input power
are in accordance with the figures written in the system labels,
which attached on the body of the machine. The level of input
WARNING voltage fluctuation should be within ±10% of the nominal values.

The operation of the machine should not be started unless all


verifications and connections are fully ceked.
CAUTION

For the first time installation of the (machine) or operation after long
period of non-use, it needs to be preheated to secure long service
life of the tube. Following method of pilot test will relieve the
damage on the x-ray tube. Operate the (machine) with low KV/mAs
values (50KV/5mAs) three times in row and execute with higher
CAUTION
values (70KV/5mAs) three times in row again. Then high KV/mAs
values (90kV/5mAs) will be applied without trouble.

40
Manual ver1.3

6. OPERATION MODE
5.1 EPX-F1600

① ④

② ⑤


⑥,⑦,⑧,⑨ ⑩



No. Description Function

1 KV indicator Display KV value

2 mAs indicator Display mAs value

3 mAs & sec indicator Display mAs & sec indicator

4 KV up/down control switch Select KV value by up or down buttons

5 mAs up/down control switch Select mAs value by up or down buttons

6 Ready indicator Lighted push the hand switch one time for preparation

7 Exposure indicator Alert lamp during x-ray exposure

8 Error lamp Lighted in case of system problem

9 Wait lamp Lighted until the second exposure started

10 Collimator lamp on Switch for operation of the lamp inside the collimator

11 Laser pointer To adjust exposure focus

12 Display reversal switch Help to read LED value of kV, mAs reversed

13 mAs & sec select switch The switch select between mAs and sec

14 APR Switch (1-6) Memory setting is available for 25 APR data

15 Save APR Data Save the selected APR Data

41
Manual ver1.3

1. Normal Mode
a. Connect power plug into the device.

b. Turn on the line switch located on the back of the equipment.

c. Select the KV values using the KV up/down switch.

d. Select the mAs values using the mAs up/down switch.

e. Positioning the machine at the ready condition by pushing the hand one time
switch.

f. When the “Ready LED” lighted, press the hand switch again and make exposure. During the x-
ray exposure, “X-ray LED” will be lighted on.

g. After the exposure is ended, “wait LED” sign will be lighted on for the waiting time.

2. APR Mode
a. Connect power plug into the device.

b. Turn on the line switch.

c. Main APR memory is flashing.

Main APR 1 Main APR 2 Main APR 3 Main APR 4 Main APR 5 Main APR 6 Total memory

2,3,4,5,6 1,3,4,5,6 1,2,4,5,6 1,2,3,5,6 1,2,3,4,6 1,2,3,4,5,6 30


If you want to change Main APR, push Main APR you want to choose for 2 seconds.

d. Select the kV values using the KV up/down switch.

e. Select the mAs values using the mAs up/down switch.

f. Save the selected KV and mAs figures with pushing the “save” button.

g. When the “Ready LED” sign appears, press the hand switch again and make
exposure. During the x-ray exposure, the “X-ray LED” will be lighted on.

h. After the exposure is ended, [5] “wait LED” will be lighted on for the waiting time.
i.

42
Manual ver1.3

5.2 EPX-F2400/F2800/F3200/F4000/F5000

No. Description Function

1 KV indicator Display KV value

2 mAs indicator Display mAs value

3 mAs & sec indicator Display mAs & sec indicator

4 KV up/down control switch Select KV value by up or down buttons

5 mAs up/down control switch Select mAs value by up or down buttons

6 Ready indicator Lighted push the hand switch one time for preparation

7 PC interface connecting indicator Lighted when EPX series is connected to PC

8 Exposure indicator Alert lamp during x-ray exposure

9 Error lamp Lighted in case of system problem

10 Collimator lamp on Switch for operation of the lamp inside the collimator

11 Laser pointer To adjust exposure focus

12 Display reversal switch Help to read LED value of kV, mAs reversed

13 mAs & sec select switch The switch select between mAs and sec

14 Wait lamp Lighted until the second exposure started

15 APR Switch (1-6) Memory setting is available for 6 APR data

16 Save APR Data Save the selected APR Data

43
Manual ver1.3

3. Normal Mode
a. Connect power plug into the device.

b. Turn on the line switch located on the back of the equipment.

c. Select the KV values using the KV up/down switch.

d. Select the mAs values using the mAs up/down switch.

e. Positioning the machine at the ready condition by pushing the hand one time
switch.

f. When the “Ready LED” lighted, press the hand switch again and make exposure. During the x-
ray exposure, “X-ray LED” will be lighted on.

g. After the exposure is ended, “wait LED” sign will be lighted on for the waiting time.

4. APR Mode
a. Connect power plug into the device.

b. Turn on the line switch.

c. Select the APR memory out of 1~6.

j. Select the kV values using the KV up/down switch.

k. Select the mAs values using the mAs up/down switch.

l. Save the selected KV and mAs figures with pushing the “save” button.

m. When the “Ready LED” sign appears, press the hand switch again and make
exposure. During the x-ray exposure, the “X-ray LED” will be lighted on.

i. After the exposure is ended, [5] “wait LED” will be lighted on for the waiting time.

44
Manual ver1.3

7. MAINTENANCE PROCEDURE

(1) General caution

It is recommended to follow the maintenance procedure described below, for the reliable operation. The

routine inspection should be committed by a trained expert

(2) MAINTENANCE SCHEDULE

Maintenance/control item Period

mA :

1 Should be within accuracy boundary of ±3% , with the voltage 1 Year

set by 40 kV and 120 kV.

mAs :

2 Should be within accuracy boundary of ±5%, when the mAs 1 Year

value is between 0.4mAs and 200 mAs.

KV :

4 Should be within accuracy boundary of ±5%, when the kV value 1 Year

is between 40kV and 120kV.

Optical Maintenance :

5 When dirt appears on the image, remove dirt on Collimator 1 Year

window with soft brush.

(3) CLEANING

Keep the machine clean when it gets dirt.

To clean the main-body, turn off the power and pull the plug out from the outlet.

Scrub surfaces smoothly with soft fabric or gauze.

45
Manual ver1.3

8. QUALITY WARRANTY
 Scope & Duration of Warranty
EPX seriese manufactured by Ecotron Co.,Ltd are warranted to be free from defects
for a period of two years after purchase date. If during the warranty period the product
you purchased is found to be defective, it will be repaired free of charge.
In the case of one of the following, however, a certain amount of service fees will be charged.
• Defect or damage found after the warranty period.
• Defect or damage in appearance which is not related to main function of the system.
• Damage caused by a natural disaster; such as fire, earthquake, or lightning strikes.
• Damage resulting from either improper movement or inattention to the precautions.
• Damage resulting from repair or modification by someone other than Ecotron or the authenticated by Ecotron
Co.,Ltd.
• Damage caused by a natural disaster; such as fire, earthquake, or lightning strikes.
• Incidental or indirect loss caused during system manipulation.

Any or all defect or damages in appearance, which do not affect the main functions of the
product are not covered by this free of charge warranty.

 Prerequisites for Repair Request


- When a defect is found, stop the using immediately. It is strongly recommended to refer to related material
on the Service Manual.
- Before a service request, must power off the entire system and check the model number, serial number,
and the purchase date. Then contact an authorized service office
 Any product with a defect in appearance only shall not be returned to nor replaced by Ecotron Co.,Ltd. Ecotron
Co.,Ltd. shall not be liable any incidental or consequential damages arising out of or relating to the use of the
product
 Ecotron Co.,Ltd. shall not be liable for any damages or losses occurring after the warranty period
 This Quality Warranty prevails over the detailed Warranty for fitness or all other warranties in relation to the
product

46
Manual ver1.3

Service chapter
This manual is designed to ensure correct use and operation of EPX model portable x-ray units. Please
read all the lines thoroughly before you use the equipment.

Incorrect use and operation exceeding described conditions in this manual may cause damage of the
equipment and shorten its life. Particular attention must be paid to all the warnings, cautions and notes
incorporated herein.

This equipment should be used only by the legally qualified persons and practitioners.

Regarding the AC cable and software, only the ones developed and supplied by Ecotron Inc. should be
used.

No responsibility is taken by Ecotron for any infringement of patients or other rights of the third parties
which many results from the use of this manual.

Always keep the manual at hand for your reference.

NOTE: If the output calibration is needed after periodical inspections or repairs, please observe following
rules and methods described in this manual.

CAUTION
The ionizing radiation is dangerous for the operator if the following safety measures
are not strictly observed.

47
Manual ver1.3

1. EPX-Series Calibration
5.1 EPX-F1600
1.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.2 Calibration Procedure

1.2.1 Tube voltage confirmation and calibration.


a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(30mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 90KV and 5mAs(16mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 4.5V, adjust KV by turning VR3 on the OP board.

48
Manual ver1.3

1.2.2 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(30mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 5.1 V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

where, C00~C8 are calibration code of tube current at AC110V model


and C9~C18 are calibration code of tube current at AC220V model.
C00 is calibration code of tube current (30mA at 40kV~50kV) at AC110V
and 30.0 is calibration data of tube current (30mA at 40kV~50kV).

49
Manual ver1.3

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 5.1V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C9 is calibration code of tube current (25mA at 51kV~60kV).


and 25.0 is calibration data of tube current (25mA at 51kV~60kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 4.25V, set the calibration data again.

50
Manual ver1.3

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration
data is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.
a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

where, P00~P6 are calibration code of filament preheat current at AC110V,


P07~P14 are calibration code of filament preheat current at AC220V,
P00 or P07 is Filament preheat calibration code (Preheat calibration code of 30mA),
and 0.90 is Filament preheat calibration data (Preheat calibration data of 30mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

51
Manual ver1.3

where, P01 or P8 is Filament preheat calibration code (Preheat calibration code of 25mA)
and 1.00 is Filament preheat calibration data (Preheat calibration data of 25mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

52
Manual ver1.3

5.2 EPX-F2400
1.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.2 Calibration Procedure

1.2.1 Tube voltage confirmation and calibration.


a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(40mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 100KV and 5mAs(20mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 5V, adjust KV by turning VR3 on the OP board.

53
Manual ver1.3

1.2.3 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 (GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(40mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 6V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

where, C00~C7 are calibration code of tube current at AC110V model


and C7~C14 are calibration code of tube current at AC220V model.
C00 is calibration code of tube current (40mA at 40kV~50kV) at AC110V
and A00.0 is calibration data of tube current (100mA at 40kV~50kV).

54
Manual ver1.3

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 5.1V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C23 is calibration code of tube current (80mA at 51kV~60kV).


and 80.0 is calibration data of tube current (80mA at 51kV~60kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 4.08V, set the calibration data again.

55
Manual ver1.3

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration
data is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.
a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

where, P00~P10 are calibration code of filament preheat current at AC110V,


P11~P21 are calibration code of filament preheat current at AC220V,
P00 or P11 is Filament preheat calibration code (Preheat calibration code of 100mA),
and 2.23 or 2.30 is Filament preheat calibration data (Preheat calibration data of 100mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

56
Manual ver1.3

where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA)
and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

57
Manual ver1.3

5.3 EPX-F2800
1.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.2 Calibration Procedure

1.2.1 Tube voltage confirmation and calibration.


a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP17 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(100mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 120KV and 5mAs(20mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 6V, adjust KV by turning VR3 on the OP board.

58
Manual ver1.3

1.2.4 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP17 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(100mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 5.1 V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

where, C00~C21 are calibration code of tube current at AC110V model


and C22~C43 are calibration code of tube current at AC220V model.
C00 is calibration code of tube current (100mA at 40kV~50kV) at AC110V
and A00.0 is calibration data of tube current (100mA at 40kV~50kV).

59
Manual ver1.3

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 5.1V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C23 is calibration code of tube current (80mA at 51kV~60kV).


and 80.0 is calibration data of tube current (80mA at 51kV~60kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 4.08V, set the calibration data again.

60
Manual ver1.3

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration
data is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.
a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

where, P00~P10 are calibration code of filament preheat current at AC110V,


P11~P21 are calibration code of filament preheat current at AC220V,
P00 or P11 is Filament preheat calibration code (Preheat calibration code of 100mA),
and 2.23 or 2.30 is Filament preheat calibration data (Preheat calibration data of 100mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

61
Manual ver1.3

where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA)
and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

62
Manual ver1.3

5.4 EPX-F3200
1.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.3 Calibration Procedure

1.2.1 Tube voltage confirmation and calibration.


a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP17 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(100mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 120KV and 5mAs(20mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 6V, adjust KV by turning VR3 on the OP board.

63
Manual ver1.3

1.2.5 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP17 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(100mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 5.1 V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

where, C00~C21 are calibration code of tube current at AC110V model


and C22~C43 are calibration code of tube current at AC220V model.
C00 is calibration code of tube current (100mA at 40kV~50kV) at AC110V
and A00.0 is calibration data of tube current (100mA at 40kV~50kV).

64
Manual ver1.3

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 5.1V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C23 is calibration code of tube current (80mA at 51kV~60kV).


and 80.0 is calibration data of tube current (80mA at 51kV~60kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 4.08V, set the calibration data again.

65
Manual ver1.3

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration
data is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.
a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

where, P00~P10 are calibration code of filament preheat current at AC110V,


P11~P21 are calibration code of filament preheat current at AC220V,
P00 or P11 is Filament preheat calibration code (Preheat calibration code of 100mA),
and 2.23 or 2.30 is Filament preheat calibration data (Preheat calibration data of 100mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

66
Manual ver1.3

where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA)
and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

67
Manual ver1.3

5.5 EPX-F4000
1.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.4 Calibration Procedure

1.2.1 Tube voltage confirmation and calibration.


a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP17 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(100mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 120KV and 5mAs(20mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 6V, adjust KV by turning VR3 on the OP board.

68
Manual ver1.3

1.2.6 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP17 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(100mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 5.1 V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

where, C00~C21 are calibration code of tube current at AC110V model


and C22~C43 are calibration code of tube current at AC220V model.
C00 is calibration code of tube current (100mA at 40kV~50kV) at AC110V
and A00.0 is calibration data of tube current (100mA at 40kV~50kV).

69
Manual ver1.3

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 5.1V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C23 is calibration code of tube current (80mA at 51kV~60kV).


and 80.0 is calibration data of tube current (80mA at 51kV~60kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 4.08V, set the calibration data again.

70
Manual ver1.3

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration
data is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.
a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

where, P00~P10 are calibration code of filament preheat current at AC110V,


P11~P21 are calibration code of filament preheat current at AC220V,
P00 or P11 is Filament preheat calibration code (Preheat calibration code of 100mA),
and 2.23 or 2.30 is Filament preheat calibration data (Preheat calibration data of 100mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

71
Manual ver1.3

where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA)
and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

72
Manual ver1.3

5.6 EPX-F5000
1.1 Line Power Switch

Line switch should be at off position.


Connect power line cord to the power outlet.
Turn on the line switch.
The FND on the control panel should illuminate.

1.5 Calibration Procedure

1.2.1 Tube voltage confirmation and calibration.


a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one probe of the oscilloscope to TP17 (IP F/B).
d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40KV and 5mAs(100mA), respectively.
h. Then observe the KV waveform of the oscilloscope channel 2.
i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 120KV and 5mAs(20mA), respectively.


k. Then observe the KV waveform of the oscilloscope channel 2.
l. If it is not 6V, adjust KV by turning VR3 on the OP board.

73
Manual ver1.3

1.2.7 Tube current confirmation and calibration.

a. Turn off the line switch.


b. Detach the control panel from the unit to access the PCB OP BOARD.
c. Connect the channel one (1) probe of the oscilloscope to TP17 (IP F/B).
d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B).
e. Connect the ground to TP13 or TP14 or TP23(GND) on the OP board.
f. Turn on the line switch.
g. Set the kV and the mAs at 40kV and 5mAs(100mA), respectively.
h. Then observe the mA waveform of the oscilloscope channel 1.
i. If it is not 5.1 V, adjust mA using the calibration data on the OP BOARD.
j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

where, C00~C21 are calibration code of tube current at AC110V model


and C22~C43 are calibration code of tube current at AC220V model.
C00 is calibration code of tube current (100mA at 40kV~50kV) at AC110V
and A00.0 is calibration data of tube current (100mA at 40kV~50kV).

74
Manual ver1.3

a. Adjust mA by using the mAs Up-Down switch (see operation sheet).


b. After it is adjusted, push the STORE button during 3sec(see operation sheet).
c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray.
e. If it is not 5.1V, set the calibration data again.
f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
g. Then push the kV UP switch (see operation sheet).
h. The display should indicate as follows,

Where, C01 or C23 is calibration code of tube current (80mA at 51kV~60kV).


and 80.0 is calibration data of tube current (80mA at 51kV~60kV).

i. Adjust mA by using the mAs Up-Down switch (see operation sheet).


j. After it is adjusted, push the STORE switch during 3sec(see operation sheet).
k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
l. If it is not 4.08V, set the calibration data again.

75
Manual ver1.3

Filament preheat calibration


This is preheating filament for making tube current.
The overshooting or the undershooting may occur from the tube current if the Preheat calibration
data is set by mistake.
Changing the calibration data is possible in accordance with the tube condition.
a. Turn off the line switch.
b. Detach the control panel from the unit to access the PCB OP BOARD.
c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position.
d. Turn on the line switch.
f. Push APR S/W ① and then the display should indicate as follows,

where, P00~P10 are calibration code of filament preheat current at AC110V,


P11~P21 are calibration code of filament preheat current at AC220V,
P00 or P11 is Filament preheat calibration code (Preheat calibration code of 100mA),
and 2.23 or 2.30 is Filament preheat calibration data (Preheat calibration data of 100mA).

g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,

76
Manual ver1.3

where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA)
and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA).

m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

77
Manual ver1.3

2. DIP Switch setting

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.

78
Manual ver1.3

3. Hard-wired hand switch control

3.1 Introduction
The Exposure Hand Switch consists of a two step hand switch.
The first (1) stage of the hand switch is called the “Preparation (Prep or READY)” stage.
The second (2) stage of the hand switch is called the “Exposure (Exp or X-Ray Out)“ stage.

SWITCH 1
SWITCH 2

3.2 Exposure process

There are two ways to use the hand switch to get the exposure.
a. Press and hold the Switch #1 for longer than 1 second, which will get the “Prep or
Ready” status on, then press Switch #2 to get the “Exposure”.

b. Press and hold the Switch #1 and Switch #2 simultaneously for longer than 1
second. This action will result in “Prep or Ready” status and “Exposure” at the same
time.

79
Manual ver1.3

4. Colimator

a. Halogen Lamp * 1
b. Laser pointer * 1
c. Horizontal and vertical lighting range control

i. Block diagram

ii. Layout

a. External Layout

80
Manual ver1.3

b. Internal Layout

81
Manual ver1.3

c. Knob Layout

5. EPX Series Trouble shooting

If you have any error message when you are using portable x-ray.

Please check below service check list

5.1 EPX-F1600 Error Table.

82
Manual ver1.3

ERROR # SPEC 표시내용

0001 reseved 001

0002 THERMAL ERROR (60℃ over) 002

0004 stanby FIL FB ERROR( Over [0 mA + MAX value x 10%] ) 010

0008 stanby EP FB ERROR( Over [0 KV + MAX value x 10%] ) 011

0010 stanby IP FB ERROR( Over [0 mA + MAX value x 10%] ) 012

0020 run FIL FB ERROR(out of range -5% ~ +5% from reference Filament current ) 014

0040 run IP FB LOW ERROR(out of range -5% from reference Filament current ) 018

0080 run IP FB HIGH ERROR(out of range +5% from reference Filament current ) 016

0100 run EP FB LOW ERROR(out of range -5% from reference Filament current ) 015

0200 run EP FB HIGH ERROR(out of range +5% from reference Filament current ) 017

0400 READY SW ERR (Initial state isn’t OFF) 007

0800 EXPOSURE SW ERR(Initial state isn’t OFF) 008

5.2 EPX-F2400/F2800/F3200/F4000 Error Table.

ERROR
No. Description Indication
Code

1 002 THERMAL ERR(INTERLOCK) Excess from 60℃

2 010 standby FIL FB ERROR Excess from 200mA (10% of max. current, 2A)

3 011 standby EP FB ERROR Excess from 12KV (10% of max. voltage, 120KV)

83
Manual ver1.3

4 012 standby IP FB ERROR Excess from 3.5mA (10% of MAX current, 35mA)

5 014 run FIL FB ERR Out of range from reference±10%

6 018 run IP FB LOW ERR Out of range from reference-10%

7 016 run IP FB HIGH ERR Out of range from reference+10%


run EP FB or AC POWER LOW Out of range from reference-10% or AC power
8 015 ERR line poor

9 017 run EP FB HIGH ERR Out of range from reference+10%

10 007 READY SW ERR READY switch malfunction

11 008 EXPOSURE SW ERR EXPOSURE switch malfunction

5.6 EPX-F5000 Error Table.

ERROR
No. Description Indication
Code

1 001 LOW VOLTAGE kV low voltage

2 002 THERMAL ERR(INTERLOCK) Excess from 60℃

3 007 READY SW ERR READY switch malfunction

84
Manual ver1.3

4 008 EXPOSURE SW ERR EXPOSURE switch malfunction

5 010 standby FIL FB ERROR Excess from 200mA (10% of max. current, 2A)

6 011 standby EP FB ERROR Excess from 12KV (10% of max. voltage, 120KV)

7 012 standby IP FB ERROR Excess from 3.5mA (10% of MAX current, 35mA)

8 014 run FIL FB ERR Out of range from reference±10%


run EP FB or AC POWER LOW Out of range from reference-10% or AC power line
9 015 ERR poor

10 016 run IP FB HIGH ERR Out of range from reference+10%

11 017 run EP FB HIGH ERR Out of range from reference+10%

12 018 run IP FB LOW ERR Out of range from reference-10%

13 028 Pulse by Pulse ERR FET board fault

14 029 INC_OC ERR  Feedback error

15 030 KV_OVP ERR  kV over voltage(Hardware)

16 031 FIL_OC ERR  FIL over current(Hardware)

17 032 IP_OCP ERR mA over current(Hardware)

85
Manual ver1.3

6. EPX Series Service Check List


6.1 EPX-F2400

Please fill in the blank.

kV/mAs condition(When error


Product S/N Error code
is occurred)

Open upper cover of portable x-ray below picture.

86
Manual ver1.3

1. Voltage check
1) Turn on AC switch

TP3 TP4 TP5 TP6


VDC Range 4.8V~5.V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V
Pass Fail Pass Fail Pass Fail Pass Fail
Pass/Fail
Measured Voltage
If it is fail case

87
Manual ver1.3

2. kV,mA waveform check


1)Turn off AC switch
2)Please connect probe to EP_FB(kV feedback_TP11) ,IP_FB(mA feedback_TP8) test point on operation
board below picture.
3)Turn on AC switch.

88
Manual ver1.3

4) 40kV/5mAs waveform

Please set kV, mAs below table and take exposure

Exposure condition : 40kV/5mAs


CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage
If kV,mA waveforms are fail, please attach your kV,mA waveform

5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

89
Manual ver1.3

If kV,mA waveforms are fail, please attach your kV,mA waveform


6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 60kV 3V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 70kV 3.5V 4.5V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

90
Manual ver1.3

8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 4.5V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 3.75V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

91
Manual ver1.3

10) 100kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 100kV 5V 3V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

92
Manual ver1.3

3. Waveform of poor AC power condition


If AC power is under 100V or 200VAC, You have below abnormal waveform.
Error 015,016,017,018 is occurred by poor AC power.

93
Manual ver1.3

4. kV , mA PWM frequency check


1) Turn off AC swtich
2) Please connect probe to 200kHz(TP15) , 140kHz(TP8) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mAPWM frequency
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency

94
Manual ver1.3

5. mA calibration data check.

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C015(AC110V), C16~C31(AC220V)


Please check Input data.

95
Manual ver1.3

Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
C00 C09 40~60 0.4 100 40.0 40.0
C01 C10 61~70(1) 0.4 64 30.0 30.0
C02 C11 61~70(2) 80 100 25.0 25.0
C03 C12 71~80(1) 0.4 32 30.0 30.0
C04 C13 71~80(2) 40 80 25.0 25.0
C05 C14 81~90(1) 0.4 40 25.0 25.0
C06 C15 81~90(2) 50 80 20.0 20.0
C07 C16 91~100(1) 0.4 50 20.0 20.0
C08 C17 91~100(2) 64 80 16.0 16.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

96
Manual ver1.3

6. Filament Preheat Calibration Code and Data check

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P04(AC110V), P05~P09(AC220V)


Please check Input data.
Input data and Tube current should be the same.
Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA(A)


40 P00 P05
30 P01 P06
2.4kW 25 P02 P07
20 P03 P08
16 P04 P09

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

97
Manual ver1.3

6.3 EPX-F2800

Please fill in the blank.

kV/mAs condition(When error


Product S/N Error code
is occurred)

Open upper cover of portable x-ray below picture.

98
Manual ver1.3

1. Voltage check
1) Turn on AC switch

TP3 TP4 TP5 TP6


VDC Range 4.8V~5.V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V
Pass Fail Pass Fail Pass Fail Pass Fail
Pass/Fail
Measured Voltage
If it is fail case

99
Manual ver1.3

2. kV,mA waveform check


1) Turn off AC switch
2) Please connect probe to EP_FB(kV feedback_TP11) ,IP_FB(mA feedback_TP8) test point on operation
board below picture.
3) Turn on AC switch.

100
Manual ver1.3

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

101
Manual ver1.3

6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 60kV 3V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 70kV 3.5V 5.15V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

102
Manual ver1.3

8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 5.15V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

103
Manual ver1.3

10) 100kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 100kV 5V 4.27V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

104
Manual ver1.3

3. Waveform of poor AC power condition


If AC power is under 200VAC, You have below abnormal waveform.
Error 015,016,017,018 is occurred by poor AC power.

105
Manual ver1.3

4. kV , mA PWM frequency check


1) Turn off AC swtich
2) Please connect probe to 200kHz(TP15) , 140kHz(TP8) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mAPWM frequency
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency

106
Manual ver1.3

5. mA Calibration data check

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C015(AC110V), C16~C31(AC220V)

107
Manual ver1.3

Please check Input data.


Input data and Tube current should be the same.
Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
C00 C16 40~60 0.4 100 35.0 35.0
C01 C17 61~70(1) 0.4 64 30.0 30.0
C02 C18 61~70(2) 80 100 25.0 25.0
C03 C19 71~80(1) 0.4 10 35.0 35.0
C04 C20 71~80(2) 13 32 30.0 30.0
C05 C21 71~80(3) 40 80 25.0 25.0
C06 C22 81~90(1) 0.4 10 30.0 30.0
C07 C23 81~90(2) 13 40 25.0 25.0
C08 C24 81~90(3) 50 80 20.0 20.0
C09 C25 91~100(1) 0.4 13 25.0 25.0
C10 C26 91~100(2) 16 50 20.0 20.0
C11 C27 91~100(3) 64 80 16.0 16.0
C12 C28 101~110(1) 0.4 32 18.0 18.0
C13 C29 101~110(2) 40 64 16.0 16.0
C14 C30 111~120(1) 0.4 20 16.0 16.0
C15 C31 111~120(2) 25 50 14.0 14.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

108
Manual ver1.3

6. Filament Preheat Calibration Code and Data check


1) Turn off AC switch
2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P06(AC110V), P07~P13(AC220V)

109
Manual ver1.3

Please check Input data.


Input data and Tube current should be the same.
Fill in the blank below table because P_DATA is different for each product.
Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA(A)
35 P00 P07
30 P01 P08
25 P02 P09
2.8kW 20 P03 P10
18 P04 P11
16 P05 P12
14 P06 P13

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

110
Manual ver1.3

6.4 EPX-F3200

Please fill in the blank.

kV/mAs condition(When error


Product S/N Error code
is occurred)

Open upper cover of portable x-ray below picture.

111
Manual ver1.3

1. Voltage check
1) Turn on AC switch

TP3 TP4 TP5 TP6


VDC Range 4.8V~5.V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V
Pass Fail Pass Fail Pass Fail Pass Fail
Pass/Fail
Measured Voltage
If it is fail case

112
Manual ver1.3

2. kV,mA waveform check


1)Turn off AC swtich
2)Please connect probe to EP_FB(kV feedback_TP11) ,IP_FB(mA feedback_TP8) test point on operation
board below picture.
3)Turn on AC switch.

113
Manual ver1.3

4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 40kV 2V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

114
Manual ver1.3

6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 60kV 3V 5V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 70kV 3.5V 4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

115
Manual ver1.3

8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 4V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage
If kV,mA waveforms are fail, please attach your kV,mA waveform

9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 3V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

116
Manual ver1.3

10) 100kV/5mAs waveform


Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 100kV 5V 3V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage
If kV,mA waveforms are fail, please attach your kV,mA waveform

117
Manual ver1.3

3. Waveform of poor AC power condition


If AC power is under 200VAC, You have below abnormal waveform.
Error 015,016,017,018 is occurred by poor AC power.

118
Manual ver1.3

4. kV , mA PWM frequency check


1) Turn off AC swtich
2) Please connect probe to 200kHz(TP15) , 140kHz(TP8) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mAPWM frequency
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency

119
Manual ver1.3

5. mA Calibration data check

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C07(AC110V), C08~C15(AC220V)


Please check Input data.

120
Manual ver1.3

Input data and Tube current should be the same.


Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
C00 C08 40~50 0.4 100 60.0 60.0
C01 C09 51~60 0.4 80 50.0 50.0
C02 C10 61~70 0.4 100 40.0 40.0
C03 C11 71~80(1) 0.4 32 40.0 40.0
C04 C12 71~80(2) 40 80 35.0 35.0
C05 C13 81~90(1) 0.4 80 30.0 30.0
C06 C14 91~100(1) 0.4 50 30.0 30.0
C07 C15 91~100(2) 64 80 25.0 25.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

121
Manual ver1.3

6. Filament Preheat Calibration Code and Data check


1) Turn off AC switch
2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P05(AC110V), P07~P11(AC220V)


Please check Input data.

122
Manual ver1.3

Input data and Tube current should be the same.


Fill in the blank below table because P_DATA is different for each product.
Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA(A)
60 P00 P06
50 P01 P07
40 P02 P08
3.2kW
35 P03 P09
30 P04 P10
25 P05 P11

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

123
Manual ver1.3

6.6 EPX-F5000
Open upper cover of portable x-ray below picture.

Please fill in the blank.


kV/mAs condition(When error
Product S/N Error code
is occurred)

124
Manual ver1.3

1.1 Voltage check

1) Turn on AC switch

TP22 TP26 TP24 TP25 TP6


VDC Range 23V~25V 4.8V~5.2V 11.5V~12.5V -11.5V~-12.5V +3V~3.6V
Pass Fail Pass Fail Pass Fail Pass Fail Pass Fail
Pass/Fail
Measured
Voltage
If it is fail
case

125
Manual ver1.3

1.2 kV,mA waveform check

1)Turn off AC swtich


2)Please connect probe to EP_FB(kV feedback_TP11) ,IP_FB(mA feedback_TP17) test point on operation
board below picture.
3)Turn on AC switch.

126
Manual ver1.3

1) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure

Exposure condition : 50kV/5mAs


CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 50kV 2.5V 5.12V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage
If kV,mA waveforms are fail, please attach your kV,mA waveform

2) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure

Exposure condition : 60kV/5mAs


CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 60kV 3V 4.12V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

127
Manual ver1.3

3) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure

Exposure condition : 70kV/5mAs


CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div

Setting value EP_FB IP_FB


Ref voltage 70kV 3.5V 3.72V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage
If kV,mA waveforms are fail, please attach your kV,mA waveform

4) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 80kV 4V 3.12V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

128
Manual ver1.3

5) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure

Exposure condition : 90kV/5mAs


CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 90kV 4.5V 2.6V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

6) 100kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 100kV 5V 2.36V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

129
Manual ver1.3

7) 110kV/5mAs waveform
Please set kV, mAs below table and take exposure

Exposure condition : 100kV/5mAs


CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div

Setting value EP_FB IP_FB


Ref voltage 100kV 5.5V 2.0V
Pass/Fail Pass Fail Pass Fail
Measured kV feedback voltage

If kV,mA waveforms are fail, please attach your kV,mA waveform

130
Manual ver1.3

1.3 Waveform of poor AC power condition


If AC power is under 200VAC, You have below abnormal waveform.
Error 015,016,017,018 is occurred by poor AC power.

131
Manual ver1.3

1.4 kV , mA PWM frequency check

1) Turn off AC swtich


2) Please connect probe to 60kHz(TP15) , 120kHz(TP16) test point on operation board below picture.
3) Turn on AC switch.

4) kV / mAPWM frequency

CH 1:
CH 2 :
25msec/div

CH CH 1 mA PWM CH 1 kV PWM
Frequency 30kHz 120kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency

132
Manual ver1.3

1.5 mA Calibration data check

1) Turn off AC switch


2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.

Calibration code

Calibration Input data

Up/Down switches increase and decrease C00~C015(AC110V), C16~C31(AC220V)

Please check Input data.

Input data and Tube current should be the same.

133
Manual ver1.3

Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
C00 C22 40~50 0.4 10 100 A00
C01 C23 40~50 13 32 80 80
C02 C24 40~50 40 100 70 70
C03 C25 40~50 125 200 50 50
C04 C26 51~60 0.4 10 80 80
C05 C27 51~60 13 32 70 70
C06 C28 51~60 40 100 60 60
C07 C29 51~60 125 150 45 45
C08 C30 61~70 0.4 10 70 70
C09 C31 61~70 13 32 60 60
C10 C32 61~70 40 100 40 40
C11 C33 71~80 0.4 10 60 60
C12 C34 71~80 13 32 50 50
C13 C35 71~80 40 100 35 35
C14 C36 81~90 0.4 32 40 40
C15 C37 81~90 40 100 30 30
C16 C38 91~100 0.4 32 35 35
C17 C39 91~100 40 100 25 25
C18 C40 101~110 0.4 32 30 30
C19 C41 101~110 40 100 20 20
C20 C42 111~120 0.4 32 25 25
C21 C43 111~120 40 100 20 20

4) If mA calibration is done, Please dip switch 4 set to OFF position.

134
Manual ver1.3

1.6 Filament Preheat Calibration Code and Data check


1) Turn off AC switch
2) Dip switch on op board

DIP Switches are used for following setting.


SW #1: ERROR Enable (ON)/Disable (OFF)
SW #2: Reserved
SW #3: Initialize memories (reserved)
SW #4: Calibration mode (ON)/Normal mode (Off)
For normal operation, only the DIP SW#1 must be set to ON position.
DIP SWITCH #4 on the OP BOARD PCB is set to ON position.
Then the display should indicate as follows,

3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P10(AC110V), P11~P21(AC220V)


Please check Input data.
Input data and Tube current should be the same.

135
Manual ver1.3

Fill in the blank below table because P_DATA is different for each product.

Model Current(mA) P_Mode(110V) P_Mode(220V) P_DATA(A)


100 P00 P11
80 P01 P12
70 P02 P13
60 P03 P14
50 P04 P15
5kW 45 P05 P16
40 P06 P17
35 P07 P18
30 P08 P19
25 P09 P20
20 P010 P21

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

136

You might also like