Professional Documents
Culture Documents
EPX-series Manual Ver 0.9 PDF
EPX-series Manual Ver 0.9 PDF
EPX-Series
Portable X-ray Systems
Manual
D/N:ECM-P-SERISE,Rev1.3
1
Manual ver1.3
2
Manual ver1.3
CONTENTS
1. INTRODUCTION----------------------------------------------------------------------------------------6
5. SYSTEM CONFIGURATION------------------------------------------------------------------------25
SERVICE CHAPTER
1. CALIBRATION-------------------------------------------------------------------------------------------31
4. COLLIMATOR---------------------------------------------------------------------------------------------39
5. TROUBLE SHOOTING---------------------------------------------------------------------------------41
3
Manual ver1.3
REVISON HISTORY
Revision Number Date Description
ADVISORY SYMBLOS
The following advisory symbols are used throughout this manual.
Their application and meaning are described below.
WARNING
Warning symbol used to indicate a potential hazard lead operators, service personnel to serious injury, death or
radiation exposure.
CAUTION
Caution symbol used to indicate a potential hazard lead operators, service personnel to injury or damage of
equipment.
NOTE
Note symbol used to indicate important information need to proper use and right operation of equipment.
Keep this service manual with the equipment at all times, and review the important information whenever
required.
4
Manual ver1.3
5
Manual ver1.3
1.INTRODUCTION
The EPX Series Innovation portable x-ray system features excellent x-ray Image, quality, accuracy,
reproducibility and long-term stability. Fast kV rising time minimize unnecessary patient dose and supply
superior Image quality, fast Image response.
The operator control console is designed to be simple and user-friendly and one can select or change x-
ray parameters easily using a large soft membrane switch.
Intended use
The EPX Series is intended for use to produce X-ray images of patient during medical operations. The
EPX Series consists of a Main power board and OP control board, SMPS, FET drive board, X-ray
generator.
This system can be applied in emergency room, operation room, cast room, outside room or tec. of
hospital.
6
Manual ver1.3
1.1 COMPOSITION
This product is composed of the X-ray component part, accessories part. Detail information of each component
follows below.
(1) X-ray component part
Main Power board
SMPS
OP control board
Collimator
(2) Accessories
Hand switch
Foot switch(Option)
Operator’s Manual
7
Manual ver1.3
This user’s guide is designed to ensure correct use and operation of EPX series. Please read all the
lines thoroughly before you use this equipment.
Incorrect use and operation exceeding described conditions in this manual may occur damage of the
machine and shorten its life. Particular attention must be paid to all the warnings, cautions and notes
incorporated herein.
This equipment should be used only by the legally qualified persons and practitioners.
EPX series is designed with the due consideration for users’ safety and product reliability. It, however, is
advisable to follow under mentioned rules to keep your additional safety and health.
1. This product should be operated only by or under the supervision of legally qualified persons.
2. EPX series is designed for the radiographic uses and not for fluoroscopy or other associated
applications.
3. EPX series should be used for the diagnosis, not for the therapy.
4. Do not modify the equipment at your discretion and in case any modification is required unavoidably,
ask the help of Ecotron or its authorized dealer for the service.
8
Manual ver1.3
3. RADIATION SAFETY
1. Users and operators should wear appropriate protecting devices and clothes.
2. Stay distance from the radiant sources and all the possible secondary radiation zones.
4. The distance from the focus to skin should be kept at least 8 inch (20cm).
5. For the experimental uses, apply the lowest possible values of kV, mAs and exposure time.
9
Manual ver1.3
10
Manual ver1.3
X-ray radiation exposure may be damaging to health, with some effects being cumulative and extending
over periods of many months or even years. X-ray operators should avoid any exposure to the
primary beam and take protective measures to safeguard against scatter radiation. Scatter radiation is
caused by any object in the path of the primary beam and may be of equal or less intensity than the
primary beam that exposes the film.
No practical design can incorporate complete protection for operators or service personnel who do not
take adequate safety precautions. Only authorized and properly trained service and operating
personnel should be allowed to work with this X-ray system equipment. The appropriate personnel
must be made aware of the inherent dangers associated with the servicing of high voltage equipment
and the danger of excessive exposure to X-ray radiation during system operation.
● Wear protective clothing. Protective aprons and gloves with an equivalent of a minimum of
1/64”(0.35mm) of lead are recommended.
● To protect the patient against radiation, always use radiation protection accessories in addition to
devices which are fitted to the X-ray system
● Keep as large a distance as possible away from the object being exposed and the X-ray tube assembly.
● Never operate this X-ray system in areas where there is a risk of explosion. Detergents and
11
Manual ver1.3
disinfectants, including those used on patients, may create explosive mixtures of gases. Please
observe the relevant regulations.
● The operator console or anything electrically connected to it, must never be used within 6ft(1.8m) of
the patient environment.
● Do not place liquids(coffee, beverages, flowers, etc) on the X-ray system.
● Do not operate the x-ray system in direct sunlight or near any heat sources.
● Do not operate the x-ray system near strong magnetic fields(microwave ovens, speakers, etc), and
avoid routing the x-ray system near these devices.
● The x-ray system must be operated in locations that are clean(free of excess dust, dirt, debris, etc),
stable(free of vibration).
● Only trained maintenance staff may remove the covers of the x-ray system.
12
Manual ver1.3
5.1 EPX-F1600
13
Manual ver1.3
5.2 EPX-F2400
14
Manual ver1.3
5.3 EPX-F2800
15
Manual ver1.3
5.4 EPX-F3200
16
Manual ver1.3
5.5 EPX-F4000
17
Manual ver1.3
5.6 EPX-F5000
IDENTIFICATION LABEL-HIGH TENSION TANK
18
Manual ver1.3
SYMBOL Description
19
Manual ver1.3
20
Manual ver1.3
Internal X-ray system components will heat up during normal use of the x-ray system. This is similar to X-ray tube
heating during normal x-ray system operation. The amount of heat produced is proportional to the product of kV,mA
and time.
21
Manual ver1.3
Use the procedure when the generator has not been used for several days. This procedure provides for exposures
at medium power before the tube is used at maximum mA or kV values. This will reduce the possibility of damaging
the X-ray tube and high voltage components. No test setup is required.
For maximum stability and reliability, use the following techniques at start up:
Select the following:
● 60kV/5mA.
22
Manual ver1.3
The EPX series comply with the regulatory requirements and design standards in this section as follows:
This product complies with DHHS requirements of 21 CFR Subchapter J.
1) SAFETY
● IEC 60601-1:1988+A1:1991+A2:1995
● IEC 60601-2-7:1998(2nd Edition)
2) EMC
● EN60601-2:2007
● EN55011:2007+A2:2007GROUP 1 CLASS A
● EN61000-3-2:2000+A2:2005
● EN61000-3-3:1995+A1:2001
The EPX series X-ray system with CE marking comply with the European Council Directive concerning Medical
Devices. One of the harmonized standards of this Directive defines the permitted levels of electromagnetic
emission from this equipment and its required immunity from the electromagnetic emission of other devise.
It is not possible, however to exclude with absolute certainty the possibility that other high frequency electronic
equipment, which is fully compliant to the EMC regulations, will not adverse effect the operation of this x-ray system.
If the other equipment has a comparatively high level of transmission power and is in close proximity to the
generator, these EMC concerns (the risk of interference) may as mobile telephones, cordless microphones and
other similar mobile radio equipment be restricted from the vicinity of this X-ray system.
3) OTHERS
23
Manual ver1.3
4) CLINICAL EVALUATION
● EN ISO 14155-1:2003 Clinical investigation of medical devices for human subjects - Part1 : General
requirements(ISO14155-1:2003)
● MEDDEV.2.7.1 Evaluation of Clinical Data: A Guide for Manufacturers and Notified Bodies
24
Manual ver1.3
4.7 SPECIFICATION
4.7.1 SPECIFICATION TABLE
5.1 EPX-F1600
25
Manual ver1.3
5.2 EPX-F2400
26
Manual ver1.3
5.3 EPX-F2800
Max Output Power 2.8 kW
Input Power Voltage 110 V-120V / 220V-240V
Phase & Frequency Single / 50/60 Hz
Tube voltage current mAs
40KV ~ 60KV 40mA 0.4 ~ 100
61KV ~ 70KV 30mA 0.4 ~ 64
61KV ~ 70KV 25mA 80 ~ 100
71KV ~ 80KV 35mA 0.4 ~ 10
71KV ~ 80KV 30mA 13 ~ 32
71KV ~ 80KV 25mA 40 ~ 80
81KV ~ 90KV 25mA 0.4 ~ 10
Radiography kV Range in 1kV steps 81KV ~ 90KV 25mA 13 ~ 40
81KV ~ 90KV 20mA 50 ~ 80
91KV ~ 100KV 20mA 0.4 ~ 13
91KV ~ 100KV 20mA 16 ~ 50
91KV ~ 100KV 16mA 64 ~ 80
101KV ~ 110KV 18mA 0.4 ~ 32
101KV ~ 110KV 16mA 40 ~ 64
111KV ~ 120KV 16mA 0.4 ~ 20
111KV ~ 120KV 14mA 25 ~ 50
MAs Range 0.4mAs – 100mAs, 25steps
Max. kV Deviation ±3 %
Max. mAs Deviation ±5 %
Display KV /mAs: 7-segment LED
Model Name D-125S TOSHIBA
27
Manual ver1.3
5.4 EPX-F3200
28
Manual ver1.3
5.5 EPX-F4000
29
Manual ver1.3
5.6 EPX-F5000
30
Manual ver1.3
31
Manual ver1.3
1) EPX-F1600
2) EPX-F2400
mAs TABLE
40KV ~ 60KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
Tube current : 40mA 35mA
61KV ~ 70KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
30mA
71KV ~ 80KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
30mA 25mA
81KV ~ 90KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
25mA 20mA
91KV ~ 100KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
20mA
32
Manual ver1.3
2) EPX-F2800
mAs TABLE
40KV ~ 60KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
Tube current : 40mA(0)
61KV ~ 70KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
30mA(0) 25mA(0)
71KV ~ 80KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
35mA(1) 30mA(1) 25mA(1)
81KV ~ 90KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
30mA(2) 25mA(2) 20mA(0)
91KV ~ 100KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
25mA(3) 20mA(1) 16mA(0)
101KV ~ 110KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64
18mA(0) 16mA(1)
111KV ~ 120KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50
16mA(2) 14mA(0)
3) EPX-F3200
mAs TABLE
40KV ~ 50KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
Tube current : 60mA(0)
51KV ~ 60KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
50mA(0)
61KV ~ 70KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
40mA(0)
71KV ~ 80KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
40mA(1) 35mA(0)
81KV ~ 90KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
30mA(0)
91KV ~ 100KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
30mA(1) 25mA(0)
33
Manual ver1.3
4) EPX-F4000
5) EPX-F5000
40~50KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100 125 150 175 200
100mA(0) 80mA(0) 70mA(0) 50mA(0)
51~60KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100 125 150
80mA(1) 70mA(1) 60mA(0) 45mA(0)
61~70KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
70mA(2) 60mA(1) 40mA(0)
71~80KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
60mA(2) 50mA(1) 35mA(0)
81~90KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
40mA(1) 30mA(0)
91~100KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
35mA(0) 25mA(0)
101~110KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80 100
30mA(1) 20mA(0)
111~120KV
0.4 0.5 0.6 0.8 1.0 1.3 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8.0 10 13 16 20 25 32 40 50 64 80
25mA(0) 20mA(1)
34
Manual ver1.3
Side view
Front view
35
Manual ver1.3
Side view
Front view
36
Manual ver1.3
Avoid the following places for the normal operation and safe storage:
For safe operation of the equipment, it should be kept away from the place with strong vibration and
maintain proper environment and conditions.
Operation Environment
For safe storage and transportation, the following range of temperature and humidity must be kept.
37
Manual ver1.3
ECOTRON Co.,Ltd
7th Fl, Hanshin IT Tower II, 60-18, Gasang-Dong,
Geumcheon-Gu, Seoul 153-023, Korea
TEL:+82-2-2025-3760, FAX:+82-2-2025-3764
E-mail:Ecotron@ecotron.co.kr
Web-site:http://www.ecotron.co.kr
In USA
Contact DRGEM USA Inc.
7018 NW 50TH Terrace, Gainsville, Florida, 32653, USA
TEL:201-370-6672, FAX:352-337-1271
E-mail:drgemusa@yahoo.com
38
Manual ver1.3
5. SYSTEM CONFIGURATION
EPX Series is a radiological device for veterinary uses. It should be applied for the radiographic
diagnosis and operated by qualified practitioners. Users have to comply with safety and health
regulations concerning the ionizing radiation protection and the electrical and mechanical safety of the
veterinary medical devices.
1. Operation Practices
b. Stay at least 6.5feet (2m) away from the unit or extend the cord of hand switch as long as possible
in order to secure enough space for your safety.
c. Use proper field size and technical values for each procedure to minimize x-ray exposure dose and
obtain the best radiographic result.
d. When you make exposure to the patient on the bed, ask visitors to leave the room first and keep
appropriate distance from the patient.
e. Pay attention to the maintenance schedule of the device and keep up with it.
f. Cumulative radiation dose usually does not exceed recommended maximum levels. However, if you
make radiographic exposures using high kV and mAs quite often, the evaluation of the specialist is
needed to check whether extra protective devices are necessary for the user or not.
b. Arrange the SID (Source to Image Receptor Distance) using a tape measure located on the side of
the collimator.
c. Turn on the collimator lamp and laser pointer by the pushing collimator/laser switch is on.
d. Adjust the size of the x-ray beam in accordance with the film size using the x-ray field
39
Manual ver1.3
adjustment knobs. The collimator and laser pointer will be turned off after 30 seconds by internal
timer.
Please check whether the voltage and frequency of the input power
are in accordance with the figures written in the system labels,
which attached on the body of the machine. The level of input
WARNING voltage fluctuation should be within ±10% of the nominal values.
For the first time installation of the (machine) or operation after long
period of non-use, it needs to be preheated to secure long service
life of the tube. Following method of pilot test will relieve the
damage on the x-ray tube. Operate the (machine) with low KV/mAs
values (50KV/5mAs) three times in row and execute with higher
CAUTION
values (70KV/5mAs) three times in row again. Then high KV/mAs
values (90kV/5mAs) will be applied without trouble.
40
Manual ver1.3
6. OPERATION MODE
5.1 EPX-F1600
① ④
② ⑤
③
⑬
⑥,⑦,⑧,⑨ ⑩
⑪
⑭
⑫
6 Ready indicator Lighted push the hand switch one time for preparation
10 Collimator lamp on Switch for operation of the lamp inside the collimator
12 Display reversal switch Help to read LED value of kV, mAs reversed
13 mAs & sec select switch The switch select between mAs and sec
41
Manual ver1.3
1. Normal Mode
a. Connect power plug into the device.
e. Positioning the machine at the ready condition by pushing the hand one time
switch.
f. When the “Ready LED” lighted, press the hand switch again and make exposure. During the x-
ray exposure, “X-ray LED” will be lighted on.
g. After the exposure is ended, “wait LED” sign will be lighted on for the waiting time.
2. APR Mode
a. Connect power plug into the device.
Main APR 1 Main APR 2 Main APR 3 Main APR 4 Main APR 5 Main APR 6 Total memory
f. Save the selected KV and mAs figures with pushing the “save” button.
g. When the “Ready LED” sign appears, press the hand switch again and make
exposure. During the x-ray exposure, the “X-ray LED” will be lighted on.
h. After the exposure is ended, [5] “wait LED” will be lighted on for the waiting time.
i.
42
Manual ver1.3
5.2 EPX-F2400/F2800/F3200/F4000/F5000
6 Ready indicator Lighted push the hand switch one time for preparation
10 Collimator lamp on Switch for operation of the lamp inside the collimator
12 Display reversal switch Help to read LED value of kV, mAs reversed
13 mAs & sec select switch The switch select between mAs and sec
43
Manual ver1.3
3. Normal Mode
a. Connect power plug into the device.
e. Positioning the machine at the ready condition by pushing the hand one time
switch.
f. When the “Ready LED” lighted, press the hand switch again and make exposure. During the x-
ray exposure, “X-ray LED” will be lighted on.
g. After the exposure is ended, “wait LED” sign will be lighted on for the waiting time.
4. APR Mode
a. Connect power plug into the device.
l. Save the selected KV and mAs figures with pushing the “save” button.
m. When the “Ready LED” sign appears, press the hand switch again and make
exposure. During the x-ray exposure, the “X-ray LED” will be lighted on.
i. After the exposure is ended, [5] “wait LED” will be lighted on for the waiting time.
44
Manual ver1.3
7. MAINTENANCE PROCEDURE
It is recommended to follow the maintenance procedure described below, for the reliable operation. The
mA :
mAs :
KV :
Optical Maintenance :
(3) CLEANING
To clean the main-body, turn off the power and pull the plug out from the outlet.
45
Manual ver1.3
8. QUALITY WARRANTY
Scope & Duration of Warranty
EPX seriese manufactured by Ecotron Co.,Ltd are warranted to be free from defects
for a period of two years after purchase date. If during the warranty period the product
you purchased is found to be defective, it will be repaired free of charge.
In the case of one of the following, however, a certain amount of service fees will be charged.
• Defect or damage found after the warranty period.
• Defect or damage in appearance which is not related to main function of the system.
• Damage caused by a natural disaster; such as fire, earthquake, or lightning strikes.
• Damage resulting from either improper movement or inattention to the precautions.
• Damage resulting from repair or modification by someone other than Ecotron or the authenticated by Ecotron
Co.,Ltd.
• Damage caused by a natural disaster; such as fire, earthquake, or lightning strikes.
• Incidental or indirect loss caused during system manipulation.
Any or all defect or damages in appearance, which do not affect the main functions of the
product are not covered by this free of charge warranty.
46
Manual ver1.3
Service chapter
This manual is designed to ensure correct use and operation of EPX model portable x-ray units. Please
read all the lines thoroughly before you use the equipment.
Incorrect use and operation exceeding described conditions in this manual may cause damage of the
equipment and shorten its life. Particular attention must be paid to all the warnings, cautions and notes
incorporated herein.
This equipment should be used only by the legally qualified persons and practitioners.
Regarding the AC cable and software, only the ones developed and supplied by Ecotron Inc. should be
used.
No responsibility is taken by Ecotron for any infringement of patients or other rights of the third parties
which many results from the use of this manual.
NOTE: If the output calibration is needed after periodical inspections or repairs, please observe following
rules and methods described in this manual.
CAUTION
The ionizing radiation is dangerous for the operator if the following safety measures
are not strictly observed.
47
Manual ver1.3
1. EPX-Series Calibration
5.1 EPX-F1600
1.1 Line Power Switch
48
Manual ver1.3
49
Manual ver1.3
50
Manual ver1.3
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
51
Manual ver1.3
where, P01 or P8 is Filament preheat calibration code (Preheat calibration code of 25mA)
and 1.00 is Filament preheat calibration data (Preheat calibration data of 25mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
52
Manual ver1.3
5.2 EPX-F2400
1.1 Line Power Switch
53
Manual ver1.3
54
Manual ver1.3
55
Manual ver1.3
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
56
Manual ver1.3
where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA)
and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
57
Manual ver1.3
5.3 EPX-F2800
1.1 Line Power Switch
58
Manual ver1.3
59
Manual ver1.3
60
Manual ver1.3
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
61
Manual ver1.3
where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA)
and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
62
Manual ver1.3
5.4 EPX-F3200
1.1 Line Power Switch
63
Manual ver1.3
64
Manual ver1.3
65
Manual ver1.3
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
66
Manual ver1.3
where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA)
and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
67
Manual ver1.3
5.5 EPX-F4000
1.1 Line Power Switch
68
Manual ver1.3
69
Manual ver1.3
70
Manual ver1.3
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
71
Manual ver1.3
where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA)
and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
72
Manual ver1.3
5.6 EPX-F5000
1.1 Line Power Switch
73
Manual ver1.3
74
Manual ver1.3
75
Manual ver1.3
g. Adjust data by using the mAs Up-Down switch (see operation sheet).
h. After it is adjusted, push the STORE button during 3sec(see operation sheet).
i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.
j. If overshooting or undershooting occurs in the waveform, set the calibration data again.
k. Then push the KV UP switch (see operation sheet).
l. The display should indicate as follows,
76
Manual ver1.3
where, P01 or P12 is Filament preheat calibration code (Preheat calibration code of 80mA)
and 2.08 or 2.10 is Filament preheat calibration data (Preheat calibration data of 80mA).
m. Adjust data by using the mAs Up-Down switch on the Control Panel.
n. After it is adjusted, push the Store switch during 3sec(see operation sheet).
o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
p. If overshooting or undershooting occurs in the waveform, set the calibration data again.
77
Manual ver1.3
78
Manual ver1.3
3.1 Introduction
The Exposure Hand Switch consists of a two step hand switch.
The first (1) stage of the hand switch is called the “Preparation (Prep or READY)” stage.
The second (2) stage of the hand switch is called the “Exposure (Exp or X-Ray Out)“ stage.
SWITCH 1
SWITCH 2
There are two ways to use the hand switch to get the exposure.
a. Press and hold the Switch #1 for longer than 1 second, which will get the “Prep or
Ready” status on, then press Switch #2 to get the “Exposure”.
b. Press and hold the Switch #1 and Switch #2 simultaneously for longer than 1
second. This action will result in “Prep or Ready” status and “Exposure” at the same
time.
79
Manual ver1.3
4. Colimator
a. Halogen Lamp * 1
b. Laser pointer * 1
c. Horizontal and vertical lighting range control
i. Block diagram
ii. Layout
a. External Layout
80
Manual ver1.3
b. Internal Layout
81
Manual ver1.3
c. Knob Layout
If you have any error message when you are using portable x-ray.
82
Manual ver1.3
0020 run FIL FB ERROR(out of range -5% ~ +5% from reference Filament current ) 014
0040 run IP FB LOW ERROR(out of range -5% from reference Filament current ) 018
0080 run IP FB HIGH ERROR(out of range +5% from reference Filament current ) 016
0100 run EP FB LOW ERROR(out of range -5% from reference Filament current ) 015
0200 run EP FB HIGH ERROR(out of range +5% from reference Filament current ) 017
ERROR
No. Description Indication
Code
2 010 standby FIL FB ERROR Excess from 200mA (10% of max. current, 2A)
3 011 standby EP FB ERROR Excess from 12KV (10% of max. voltage, 120KV)
83
Manual ver1.3
4 012 standby IP FB ERROR Excess from 3.5mA (10% of MAX current, 35mA)
ERROR
No. Description Indication
Code
84
Manual ver1.3
5 010 standby FIL FB ERROR Excess from 200mA (10% of max. current, 2A)
6 011 standby EP FB ERROR Excess from 12KV (10% of max. voltage, 120KV)
7 012 standby IP FB ERROR Excess from 3.5mA (10% of MAX current, 35mA)
85
Manual ver1.3
86
Manual ver1.3
1. Voltage check
1) Turn on AC switch
87
Manual ver1.3
88
Manual ver1.3
4) 40kV/5mAs waveform
5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div
89
Manual ver1.3
7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div
90
Manual ver1.3
8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
91
Manual ver1.3
92
Manual ver1.3
93
Manual ver1.3
4) kV / mAPWM frequency
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency
94
Manual ver1.3
3) Turn on AC switch.
Calibration code
Input data
95
Manual ver1.3
Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
C00 C09 40~60 0.4 100 40.0 40.0
C01 C10 61~70(1) 0.4 64 30.0 30.0
C02 C11 61~70(2) 80 100 25.0 25.0
C03 C12 71~80(1) 0.4 32 30.0 30.0
C04 C13 71~80(2) 40 80 25.0 25.0
C05 C14 81~90(1) 0.4 40 25.0 25.0
C06 C15 81~90(2) 50 80 20.0 20.0
C07 C16 91~100(1) 0.4 50 20.0 20.0
C08 C17 91~100(2) 64 80 16.0 16.0
96
Manual ver1.3
3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.
P_Calibration code
Input data
97
Manual ver1.3
6.3 EPX-F2800
98
Manual ver1.3
1. Voltage check
1) Turn on AC switch
99
Manual ver1.3
100
Manual ver1.3
4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div
5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div
101
Manual ver1.3
6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div
7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div
102
Manual ver1.3
8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
103
Manual ver1.3
104
Manual ver1.3
105
Manual ver1.3
4) kV / mAPWM frequency
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency
106
Manual ver1.3
3) Turn on AC switch.
Calibration code
Input data
107
Manual ver1.3
108
Manual ver1.3
3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.
P_Calibration code
Input data
109
Manual ver1.3
110
Manual ver1.3
6.4 EPX-F3200
111
Manual ver1.3
1. Voltage check
1) Turn on AC switch
112
Manual ver1.3
113
Manual ver1.3
4) 40kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 40kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div
5) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 50kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div
114
Manual ver1.3
6) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 60kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div
7) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 70kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
10msec/div
115
Manual ver1.3
8) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
9) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 90kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
116
Manual ver1.3
117
Manual ver1.3
118
Manual ver1.3
4) kV / mAPWM frequency
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
CH CH 1 mA PWM CH 2 kV PWM
Frequency 200kHz 140kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency
119
Manual ver1.3
3) Turn on AC switch.
Calibration code
Input data
120
Manual ver1.3
121
Manual ver1.3
3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.
P_Calibration code
Input data
122
Manual ver1.3
123
Manual ver1.3
6.6 EPX-F5000
Open upper cover of portable x-ray below picture.
124
Manual ver1.3
1) Turn on AC switch
125
Manual ver1.3
126
Manual ver1.3
1) 50kV/5mAs waveform
Please set kV, mAs below table and take exposure
2) 60kV/5mAs waveform
Please set kV, mAs below table and take exposure
127
Manual ver1.3
3) 70kV/5mAs waveform
Please set kV, mAs below table and take exposure
4) 80kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 80kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
128
Manual ver1.3
5) 90kV/5mAs waveform
Please set kV, mAs below table and take exposure
6) 100kV/5mAs waveform
Please set kV, mAs below table and take exposure
Exposure condition : 100kV/5mAs
CH 1: kV 2V/div
CH 2 : mA 2V/div
25msec/div
129
Manual ver1.3
7) 110kV/5mAs waveform
Please set kV, mAs below table and take exposure
130
Manual ver1.3
131
Manual ver1.3
4) kV / mAPWM frequency
CH 1:
CH 2 :
25msec/div
CH CH 1 mA PWM CH 1 kV PWM
Frequency 30kHz 120kHz
Pass/Fail Pass Fail Pass Fail
Measured frequency
132
Manual ver1.3
3) Turn on AC switch.
Calibration code
133
Manual ver1.3
Tube Current
Tube Voltage mAs Tube Current INPUT DATA
Parameter
AC110V AC220V KV MIN MAX mA mA
C00 C22 40~50 0.4 10 100 A00
C01 C23 40~50 13 32 80 80
C02 C24 40~50 40 100 70 70
C03 C25 40~50 125 200 50 50
C04 C26 51~60 0.4 10 80 80
C05 C27 51~60 13 32 70 70
C06 C28 51~60 40 100 60 60
C07 C29 51~60 125 150 45 45
C08 C30 61~70 0.4 10 70 70
C09 C31 61~70 13 32 60 60
C10 C32 61~70 40 100 40 40
C11 C33 71~80 0.4 10 60 60
C12 C34 71~80 13 32 50 50
C13 C35 71~80 40 100 35 35
C14 C36 81~90 0.4 32 40 40
C15 C37 81~90 40 100 30 30
C16 C38 91~100 0.4 32 35 35
C17 C39 91~100 40 100 25 25
C18 C40 101~110 0.4 32 30 30
C19 C41 101~110 40 100 20 20
C20 C42 111~120 0.4 32 25 25
C21 C43 111~120 40 100 20 20
134
Manual ver1.3
3) Turn on AC switch.
4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.
P_Calibration code
Input data
135
Manual ver1.3
Fill in the blank below table because P_DATA is different for each product.
136