SI Pressure Instruments Ltd. Garretts Green Lane, Birmingham, B33 OYA. UK. Tel: +44 (0)121 784 6855 Fax: +44 (0)121 784 4795. sales@si-pressure.com
Page i
SAFETY WARNING!
HIGH PRESSURE:
Uncontrolled release of high pressure is hazardous to personnel and may cause
damage to equipment. Before connection of any pressure component to the PC6-
PRO ensure that the component(s) is/are isolated from the pressure supply and any
internal pressure is released slowly.
Page ii
QUICK REFERENCE
Keys Menu Selection
ON/OFF Press to turn PC6-PRO on and off Pressure
Display
MENU Press to access the menus
1st Press for Menu 1
MENU
Press to accept functions or settings 2nd Press for Menu 2
MENU
Press to move cursor up/left or increase values 3rd Press back to Pressure Display
MENU
Select menu position cursor press the Enter key Tare Hold down the key for 4 seconds while in
under function the pressure display mode. The unit will bleep after two
seconds when the Max. and Min. values reset. After
four seconds another bleep will sound as the displayed
Exit pressure now resets to zero. This is a temporary
To exit out of any menu select 'Exit' option, function and will be cancelled when the unit is
or press MENU key. switched off.
Page iii
MANUFACTURER'S SPECIFICATIONS
Model Pressure Range Accuracy (FS)
* Temperature coefficient ± 0.005% of reading / °C w.r.t. 20°C
bar psi up to Pressure Units bar, mbar, Mpa, kPa, p.s.i, kg/cm², atm, inH20,
PC6-PRO-0.1-C G -0.1 to 0.1 1.5 0.025% mH20, mmH20, inHg, mmHg, unit1, unit2 (user
PC6-PRO-1-C G or A -1 to 1 15 0.025% defined, factory set only)
PC6-PRO-2-C G or A -1 to 2 30 0.025% Overload Warning Flashing display and audible tone at 110% of FS
PC6-PRO-5-C G or A -1 to 5 75 0.025% Temperature Display Temperature of pressure sensor module
PC6-PRO-10-C G or A -1 to 10 150 0.025% (Accuracy: ±0.5°C)
PC6-PRO-20-C G or A -1 to 20 300 0.025% Pressure Switch Test Status OPEN/CLOSED
PC6-PRO-25-C G or A -1 to 25 360 0.025% Operating Temperature 0 to 50°C
PC6-PRO-40-C G or A -1 to 40 600 0.025% Calibration Temperature 20°C ± 2°C
PC6-PRO-70-C SG 0 to 70 1000 0.025% Calibrated Temp. range 0 to 50°C
PC6-PRO-100-C SG 0 to 100 1500 0.025% Storage Temperature -20°C to 70°C
PC6-PRO-200-C SG 0 to 200 3,000 0.025% Temperature Effects 0.001% of reading/°C between 10-40°C
PC6-PRO-350-C SG 0 to 350 5,000 0.025% RS232 Parameters baud rate 1200, 2400, 4800, 9600, stop bits 1 or
PC6-PRO-700-C SG 0 to 700 10,000 0.025% 2, status ON/OFF
PC6-PRO-1000-C SG 0 to 1000 15,000 0.025% Zero Reset Manual keyboard operation
Data Capacity 437 records into a maximum of 20 files, 32k
memory additional 585 records, ( 32k memory
G = Vented Gauge, A = Absolute, SG = Sealed Gauge option available )
( For Indicators, substitute ‘C’ with an ‘I’ ). Display Backlit LCD, 16 character x 2 line alpha numeric
Alternative pressure ranges are available to special order
Humidity 5 to 95% Relative Humidity non condensing
Overload Pressure 10% FS Resolution Select Increase or decrease by a factor of 10
Pressure Display Resolution 6 Digits Power Supply 6V NI-MH rechargeable battery pack (see warning
EPM External Pressure Modules are available in page ii)/ mains operation via charger (supplied).
all of the above ranges. Differential units Voltage Output Regulated D.C supply 10V ± 0.1% max current
are available as EPM’s. ( See Appendix for 30mA
EPM details ).
Unregulated D.C supply 24V ± 10% max current
Measurement Range Resolution Accuracy Input resistance 50mA
Voltage* 0 to 100mV 1uV 0.025%FS 1 MW Battery Life Fully charged: up to 10 hours
(Auto Range ) 0 to 5V 0.1mV 0.025%FS 1 MW Recharge Time 14-16 hours for full charge
0 to 50V 1mV 0.025%FS 1 MW Low Battery Continuous check, audio and visual warning
Current* 0 to 50mA 1uA 0.025%FS 3W Dimensions 92 x 110 x 59mm
Weight 850 grams (Calibrator head only)
Sourcing Range Resolution Accuracy Max. Load
Microprocessor H8/2138
Voltage* 0 to 10V 0.2mV 0.025%FS 20mA Recommended Recalibration Period 1 Year (Max)
Current* 0 to 50mA 1uA 0.025%FS 330 Ohm
Page iv
MANUFACTURER'S SPECIFICATIONS (continued)
Electrical Connections
Ground Standard 2mm socket
Voltage/Current Input/output Standard 2mm sockets
Pressure Switch Input Standard 2mm sockets
Battery 2 pole miniature round connector
RS232 8 pole miniature round connector
EPM (P2) 14 pole miniature round connector
Page v
DESCRIPTION
Page 1
1.0 Introduction
The PC6-PRO Pressure Calibrator is a microprocessor controlled, precision instrument powered by an internal 6V
rechargeable battery pack. It is portable and capable of accurate pressure measurement, voltage or current
measurements and sourcing when calibrating transducers, pressure switches, etc. This section describes the equipment
from an external viewpoint, allowing the user to become familiar with the various controls and connections provided.
1.1 General
The PC6-PRO is housed in a three part anodised aluminium case with a liquid crystal display unit at the front and a six
button membrane keyboard. The main body contains the circuitry, connections, internal transducer and battery pack.
1.2 Keypad
The Keypad is a non-tactile membrane keyboard with six buttons. The keyboard layout is shown below.
MENU ENTER
ZERO
PRINT ON/OFF
LOG
Page 2
1.3 Liquid Crystal Display (LCD) 1.4.2 Pressure connections are made to
either a 3/8"BSP (1/4”BSP / 1/4”NPT
adaptors) male for low pressure systems
The LCD is a 16 character x 2 line alphanumeric LED (below 40 bar) or 3/8”BSP female (1/4”BSP COUPLING
backlit display which is capable of displaying special / 1/4”NPT adaptors) for systems greater
characters. than 40 bar.
The connection panel is located at the top of the case and 1.4.5 The 14 way circular miniature EPM
provides input / output connections as detailed in Fig.2 connector is used to connect to the EPM
(External Pressure Module). These optional
modules cover pressure ranges from 0 to
1.4.1 Six standard 2mm electrical sockets mounted along 1000 bar.
the top of the case provide input or output current and
voltage and switch status.
Voltage outputs of 10 and 24V are available for use with
sensors, transducers, transmitters, etc.
Page 3
ARRANGEMENT OF PC6-PRO
Front View: LCD & keypad Top View: connection panel
Fig 1 Fig 2
BATTERY
CHARGER Battery Charger Input Socket
Display. + - See Page 11
RS232 Communication
Socket See Page 11
MENU ENTER
ZERO
Keypad
See Page 2 Quick Connect Coupling. +24 V Volts output.
+10 V Volts output.
GND Common Ground Socket
mA Milliamp Input (controls input current in
(See Page 10 source mode).
54-0158 ¼”BSP 54-0101 3/8”BSP for V/mA V Volt Input (output in source mode).
54-0363 ¼”NPT 54-0364 ¼”NPT connections) SWITCH Pressure Switch Input.
PC6-PRO- 40 bar PC6-PRO- Above 40 bar
or below
General Purpose Adaptors
Page 4
FUNCTIONS (MENU 1)
PRESSURE DISPLAY - Allows the user to operate the PC6 for pressure measurements. (Display Functions)
Menu 1
UNITS Selection of pressure unit.
bar bar mbar millibar
MPa MegaPascals kPa kiloPascals
p.s.i pounds force per square inch kg/cm² kilograms force per square centimetre
atm atmospheres inH2O inches of Water (at 4°C)
mH2O metres of water (at 4°C) mmH2O millimetres of water (at 4°C)
inHg inches of Mercury (at 0°C) mmHg mm of Mercury (at 0°C)
Unit1 user defined unit (Factory Set Only) Unit2 second user defined unit (Factory Set Only)
Exit Returns to menu 1
MAX Displays the maximum pressure measured since the last reset (50 samples per second).
MIN Displays the minimum pressure measured since the last reset (50 samples per second).
% Displays the pressure as a percentage between user pre-set high and low alarm values.
TARE Displays the pressure at the time of tare.
V / mA Volts / milliamps functions
Measure Displays the following input measurements.
Actual Displays actual input measurements.
Converted Displays converted input measurements.
Exit Returns to Pressure Display.
Source Displays the following outputs.
Actual Displays actual output measurements.
Converted Displays converted output measurements.
P to V / P to mA Displays Pressure to Volts / milliamps conversion.
Exit Returns to Pressure Display.
Exit Returns to pressure display
SWITCH Indicates the status of the pressure switch under test (OPEN/CLOSED).
Leak Allows the PC6 to measure pressure changes over time.
Start Starts the leak test.
Set-Time Sets the leak test time.
Exit Returns to Pressure Display.
Temp C Allows the user to monitor the temperature around the selected pressure transducer in pressure display mode.
Date & Time Allows the user to monitor the date & time of the on board real time clock in pressure display mode.
File Allows the user to view or manipulate a file previously stored in the filing system.
View Allows the user to see logged file.
Export Outputs a logged file
Import Imports a file ( refer to ‘SiCal’ Instructions for exporting files to PC6 PRO)
Delete Deletes a logged file.
Exit Returns to Pressure Display.
Exit Returns to Pressure Display mode. (SHORTCUT: To abort any menu level, simply press the MENU key).
Page 5
SETTING MENU - To set user specific functions. (Set Functions)
Menu 2
High Allows the user to set a High Pressure Alarm Limit.
(by using the increase/decrease facility above which an alarm sounds.)
Low Allows the user to set a Low Pressure Alarm Limit.
(by using the increase/decrease facility below which an alarm sounds.)
V / mA
Measure Voltage / current input settings for Converted Measurement.
Zero Set Zero.
FS Set Full scale value.
Text User defined measurement text.
Root Switch measurement root function ‘on’ or ‘off’.
Exit Reverts back to Pressure Display.
Source Voltage / current output settings.
Range Sets the following values:-
Zero Sets Zero value
FS Sets Full Scale value
Exit Returns to Source sub-menu.
Convert Sets the following converted values:-
Zero Sets converted Zero values.
FS Sets converted Full Scale values.
Text User defined converted text.
Exit Returns to Source sub-menu.
P to V / mA Sets Pressure to V / mA conversion factors.
Zero Set zero values.
FS Set Full Scale values.
Exit Returns to Source sub-menu.
Setup Sets the parameters for voltage/current sourcing.
Status Switches output on or off.
Root Switch sourcing root function ‘on’ or ‘off’.
Step Set step value (%FS)
Ramp Settings for Ramp mode
Status Switches ramp mode on or off.
Rate Sets the ramp rate (V or mA/Sec).
Dwell_time Sets the time for which the output remains at the same level.
Exit Returns to Source sub-menu.
Exit Returns to Source sub-menu.
Exit Returns to Pressure Display.
Vout Switches the Power Supply outputs +10V & +24V ON or OFF
Continued on Display Light Switches ON or OFF the LED back light.
Page 7
Page 6
FUNCTIONS (MENU 2) Continued.
Resolution Selects low, normal or high resolution.
Event Sets up the event status for RS232 output and for logging functions.
Key Sets key driven output or logging.
Timed Sets timed output defined by the user.
Exit Reverts back to Pressure Display.
Log Allows the user to setup, amend and view files for logging.
Log To select a log file and start logging.
View To look at a file.
Delete Deletes a file.
Status To enable or disable logging.
Export To export logged files.
Import To import files from a PC. (see ‘SiCal’ software instructions for exporting files to the PC6)
Exit Reverts back to Pressure Display.
RS2332 Settings for RS232 communications.
Status To enable RS232 communication.
Baud-rate To select the baud rate.
Stop-bits To select the number of stop bits.
Exit Reverts back to Pressure Display.
Date/Time Sets up Date & Time parameters
Set Sets the date & time.
View To look at date & time.
Exit Reverts back to Pressure Display.
Language Allows user to change language.
Zero Will reset pressure reading to Zero. ( This will remove Zero offset permanently / NOT available for absolute
units ).
Port Selects current pressure module (P1, P2, DPM)
Power_Off Allows the user to set Power_Off time or to switch the function on or off.
Key Switches the auto-power function off.
Timed Allows the user to set the auto-power off time in minutes.
Exit Reverts back to Pressure Display.
Exit Returns to Pressure Display
Page 7
FUNCTIONS (MENU 3)
CAL MENU
Menu 3
ZERO Allows user to set zero. ( for absolute units, see page 32 )
CONTINUE Starts zero calibration procedure.
EXIT Returns to Cal Menu.
SPAN Allows user to set span.
CONTINUE Starts span calibration procedure.
EXIT Returns to Cal Menu.
HISTORY Allows user to see overload and zero-span history.
OVERLOAD Displays period of time over which overload occurred.
ZERO-SPAN Displays date last calibrated, last span set, range and overload value.
PC6 S/No Displays the unit’s serial number.
PM S/No Displays the serial number of the currently selected pressure module.
EXIT Returns to Cal Menu.
Language Allows user to download different languages.
EXIT Returns to Pressure Display.
Page 8
OPERATING
INSTRUCTIONS
Page 9
1.5 Introduction
The PC6-RPO enables the user to carry out checks on pressure EPM
transducers, transmitters, pressure switches, pressure gauges
and process indicators quickly and accurately. It is intended for
use as a portable , on-site calibrator, used in close proximity
with the equipment under test and without the need to remove
the unit from its parent system. This section describes the
connections to be made when checking switches, transmitters,
etc. It also lists menu options, selection procedures and
purpose of options available. Fig.3
1.6 Connections
1.6.1 Pressure Input The PC6-PRO measures pressure
directly via the connector found underneath the case (i.e. the
connection between the PC6-PRO and a hand held pressure
EPM
test system). Alternatively, connect the item under test to the CHARGER RS232
Page 10
1.6.4 Switch Test The switching pressure of a pressure Single Shot - This output is always available in Display mode
switch can be monitored by connecting the contacts of the when RS232 status & Log status are OFF. The output is obtained
pressure switch between the INPUT 'GND' and 'SWITCH' each time the PRINT
LOG key is pressed (provided any previous output
sockets on the connection panel and pressurizing the is complete).
pressure switch. Warning: The PC6-PRO must be switched off prior to
connecting or disconnecting the RS232 cable.
1.6.5 RS232 Timed - This output is available in Display mode when RS232
If the RS232 facility is to be used then connections to the 8- status is ON & Log status is OFF. The output cycle is initiated by
way socket must be made to the specifications given in the pressing the PRINT
LOG key. Pressing the MENU key cancels the output.
table below. Alternatively an RS232 communication cable Once initiated, the output is repeated at the selected time interval
can be purchased from SI suitable for most RS232 (see ‘Event' function, page 26 ) until the MENU key is pressed.
applications as shown below.
1.6.6 DPM (DIgital Pressure Module) An alternative to the EPM,
To RS232 Standard 9-way the DPM communicates via the RS232 socket and is available in
socket on PC6 'D'type female -1 to 1000bar/15,000psi pressure ranges with accuracies up to
Cable Part No: 01-0625 connector 0.025%
RS232 Connections Table
PIN No. INPUT/OUTPUT SIGNAL STATE 1.6.7 Battery Charger The internal battery pack may be
1 Output Data Terminal Ready (DTR) Held High recharged (when indicated by the battery low message) by
2 Output Transmit Data (TxD) plugging the charger (supplied) into the 'CHARGER' socket.
3 Input Receive Data (RxD)
4 Output Request To Send (RTS) CAUTION: When recharging internal batteries, only the charging
5 Input Clear To Send (CTS) Active High unit supplied with the PC6-PRO should be used.
6 Output Do not use
1.6.8 Voltage Output A voltage of +10V or +24V may be
7 Common Ground
sourced from the PC6-PRO by connecting to the 2mm sockets.
8 Input Do not use
A typical application for this voltage output would be to power a
Note: Pin 6 and pin 8 should not be used.
pressure transmitter. ( see Fig. 3. page 10 )
RS232 Output To switch on voltage output function:-
Connect the PC6-RPO to a printer or computer using a suitable a) Press the MENU key twice to activate 'Set' menu.
cable (not supplied) as detailed above and set the parameters as b) Select 'V-out' and press the ENTER ZERO key.
detailed in Para. 2.6 & 2.8. The output can now be obtained as c) Select 'ON' and press the ENTER ZERO key.
‘single shot' or ‘timed' as follows:
Page 11
1.7 Messages & Parameters 1.7.7 Auto Switch off The PC6-PRO will automatically turn itself of
after a set period (See Page 31 for instructions)
The PC6-PRO registers pressure input at switch on after the initial start-
up message. The pressure reading is displayed on the top line of the LCD 1.7.8 Pressure Overload (Bleep rate 4 Hz) ‘OVERLOAD' occurs when
(see Page 3). Pressure values are monitored, alarm values are input pressure exceeds overload pressure level.
compared, alarm warnings and other information messages displayed
until the unit is switched off. 1.7.9 Low Battery Detect (Bleep rate 1 Hz). ‘Low Battery' occurs when
Memory Errors: internal battery pack drops below 5.5V. Message flashes for 4 seconds,
Any read / write errors to memory detected at switch on will be reported after a further 30 seconds, low battery detection is re-enabled.
as: Memory Error ‘Error No.’
Error Numbers:- 1.7.10 Pressure Switch State Change (Bleep rate 2 Hz)
0 - Onboard eeprom; 8 - IPM; 9 - 2nd Onboard eeprom. Occurs in pressure mode with the pressure switch state display being
Errors with Pressure Modules will be indicated by the message "No P1 selected. When the connected pressure switch changes state, the
(P2 or DPM) proceeded by "PM EEPROM ERROR" at power on. pressure value (upper line) is frozen and the pressure switch state (lower
line) display is flashed at 1 Hz unless a log file is open. Normal display
1.7.1 When switched on, the unit registers the same state as when last update resumes when the ENTER ZERO key is pressed.
switched off. The display mode (Operating/Set), units of measurement,
function and parameter settings, alarm limits and RS232 output control, 1.7.11 Printer Busy (Bleep rate 1 Hz)
EPM selection, logged parameters are all retained at switch off, with the If the receiving device is not ready for input (e.g. device not connected or
exception of the following conditions: handshake LOW), 10 seconds after starting an output cycle (single shot
or timed output) the message 'Printer Busy' is flashed on the display,
1.7.2 A battery low state occurring prior to the switch off. When switched together with an audible warning. If the device does not accept input
on, the unit registers the latest retained set of parameters. within the next 5 seconds then the output cycle is cancelled and output is
stopped until the PRINT
LOG key is pressed again.
1.7.3 Any TARE value set during use is not retained and therefore
should be set, if required, each time the unit is switched on. 1.7.12 Outside User Set Alarm Values (Bleep rate 2 Hz)
Occurs when measured pressure is lower than the user set Low alarm
1.7.4 Maximum and Minimum values are not retained. Current values value or greater than the user set High alarm value and the ’outside
reflect pressure monitored from the start of the pressure input. limits’ message is flashed. Indication ceases when pressure is equal to
either alarm value or between the values.
1.7.5 If the unit is switched off during a LEAK test then the test will be
void. When switched on again the unit will start up in LEAK test mode. 1.7.13 Display Hold (Bleep rate 2 Hz)
Occurs in pressure mode. If the ENTER
ZERO key is pressed once (less than 1
1.7.6 Switch off When the ON/OFF key is pressed the 'switch off' second) the pressure value (upper line) is frozen and the lower line is
message is displayed for two seconds accompanied by an audible bleep flashed with the ‘Display Hold’ message at 1 Hz. Normal display update
(Bleep rate 1 Hz). Normal operation may be resumed by pressing any key resumes when any top row key is pressed.
during switch off. Note: The display contents can be printed out during display hold by
pressing the PRINT
LOG key.
Page 12
1.7.14 Connection Fault: 1.7.18 Alarm Flags A single character is output to indicate the various
In voltage / current source mode, if the PC6-PRO is unable to control the pressure alarm conditions as given below:
voltage/current output (e.g. because there is no loop back in the current * overload pressure level exceeded
mode), the message “Connection Fault” will flash on the lower line (pressure warnings, 1.7.8 & 1.7.12)
accompanied by a bleep at 1 Hz. The message will continue until the fault > pressure exceeds user set high limit
is rectified or the user exits from source mode. < pressure below user set low limit
“ ” a space if no alarms apply.
1.7.15 Short Detected: c connection fault
If the 24V output is ON and it is shorted to 0V, the message “*Short s short detected
detected*” will be displayed on the lower line accompanied by a bleep at Note: If two (or more) conditions occur which require a message display, the
1 Hz. This message will be cleared when any key is pressed. In this highest priority condition in the above list will be displayed. The sequence of
condition the 24V supply will be switched off for safety. messages given above set the priority except 'Switch Off' which comes last.
Page 14
1.8 Units
To change the units of pressure, proceed as follows:
• Press the MENU key once, then press ENTERZERO key to choose
the units sub-menu To measure input voltage or current select ‘Measure’ and
• Select kPa by pressing the key and then press ENTER
ZERO key. press the ENTER
ZERO key.
The display will indicate the pressure in kPa and the pre-selected Two options are now available; ‘Actual’ and ‘Convert’.
2nd line function. • ‘Actual’ is selected by pressing the ENTER
ZERO key.
An example display below shows the Measure Mode symbol
1.9 Max., Min., Switch, Temp °C, Date & Time, % on the second line.
To display any one of these functions, proceed as follows:
• Press the MENU key once and then scroll to the required
function using the or key and then press ENTER
ZERO
‘Tare’ from the menu using the key and pressing the ENTER
ZERO key. key.
To activate Tare function: hold down the ENTER
ZERO key for 4 seconds Three options become available; ‘Actual’, ‘Convert’ and
while in the pressure display mode. The unit will bleep after two ‘P V’ / mA.
seconds when the Max. and Min. values reset. At four seconds • ‘Actual’ is selected by pressing the ENTER
ZERO key.
another bleep will sound as the displayed pressure now resets to An example display below shows the Source Mode symbol
zero. This is a temporary function and will be cancelled when the on the second line.
unit is switched off.
• By pressing and holding the MENU key for more than two
seconds whilst in ‘Source Mode’ you will activate a ‘Nudge’
function allowing the source value to be increased or
decreased, the longer you hold down the or key,
the larger the nudge increments will be (see Page 25 for
more information).
By pressing and holding the MENU key again for more than
two seconds, the nudge function will be cancelled.
Page 16
2.0 Leak Test Repeat procedure for third number and press ENTER
ZERO key.
To carry out leak tests, proceed as follows:
• Press the MENU key once and scroll to the 'Leak'
function using the key. Press ENTER
ZERO key to select.
Presented with the leak test sub menu, the options 'Start' and Now select the units of time measurement ( e.g. seconds,
'Set Time' are given as shown below. minutes, hours, days, etc. ) Scroll through the units with the
keys and press ENTER
ZERO
Leak Sub-menu Once the leak test time has been set press the MENU key to
return to the leak test sub-menu.
'Set Time' To set the duration of the leak test.
Scroll using the key to select 'Set Time' and press the
ENTER
ZERO key. • Select 'Start' using the keys and press ENTER
ZERO
key. ( e.g. time set to 999 min.) The top line displays the difference in pressure while the
bottom line shows the start pressure and the time counting
down.
• When the leak test has finished, to start the leak test again
press the ENTER
ZERO key once to reset the time and again to
With the second digit flashing, press ENTER
ZERO and again select start the test.
the number required and press ENTER
ZERO key. • If no other leak test is to be carried out, press the MENU key
to return to the leak test sub-menu.
Page 17
2.1 File
This function allows the user to view, export, Import and Date & Time
delete previously stored files.
To use the filing system proceed as follows:
One
• Switch the unit on. Pressure Data Complete
Reading
• Press the MENU key once and scroll to 'File' using the
key and press ENTER
ZERO .
2nd line function
file is required then scroll through the filenames using through the records. Once the last record has been viewed
the keys and then press ENTER key. an **END** message is displayed.
ZERO
The records then 'wrap around' so the user can move to the
Number of records
first or last file record using the keys.
‘N’ = Normal Log
‘C’ = Calibration procedure
Note: At the time of logging, if the pressure is above or
• To move through the record now selected, press the below the set limits then a 'High' or 'Low' message will
key or the ENTER
ZERO key. This will first show the appear after the Pressure Data screen.
‘IPM / EPM' serial number from which data was read.
Note: Files can be exported If you are sure that the file is to be deleted then the ‘Yes'
to ‘SiCal’ software option should be selected using the key and then
2.13 To Import files, proceed as follows: entered using the ENTERZERO key.
• Scroll to ‘Import’ using the key. Once you finished Once the operation is completed the display will return to the
a calibration procedure in ‘SiCal’, click on ‘Export’, 'File' sub-menu.
press the ENTER
ZERO key on the PC6-PRO to download. The If another file is to be deleted then repeat the procedure.
message “downloading Cal” will be displayed. Once If no more files are to be deleted then press the MENU key to
SiCal gives “All done” message, press the MENU key return to the pressure display.
• Choose ‘Yes’ when prompted to ‘Stop?’
Note: Pressing the MENU key once at any time during the
2.14 To Delete files, proceed as follows:
procedure will take you back to the 'File' sub-menu.
• Scroll the cursor using the key and select ‘Delete'
and press the ENTER
ZERO key.
If all the files in the PC6-PRO are to be deleted then select
The following options are available:
‘ALL’ in the Delete sub-menu and press the ENTERZERO key.
The following message is displayed:
Page 19
SET- MENU
(MENU 2)
Page 20
2.2 Alarm settings ‘High’, ‘Low’
To set High or Low pressure warning limits proceed as
follows:
• Press the ON/OFF key to switch on the PC6-PRO.
• Select the 'Set' menu by pressing the MENU key twice.
• Select 'High' or Low' by pressing the key and then
press the ENTER
ZERO key.
• Select the digits to be set by moving the cursor with the
keys and pressing the ENTER
ZERO key.
• Set the number to the required value using the
keys and then press the ENTER
ZERO key.
• Repeat the previous two steps to change other digits for
the alarm value.
• Once finished setting the alarm limit value press the MENU
Page 21
2.3 V, mA (Measure mode) FS (Full Scale):
The PC6-PRO can measure voltage or current and simply • From the setup menu select 'FS' using the key and
display it or convert it using the ‘Converted, Setup’ options then press the ENTER
ZERO key.
and then display the measured value via a conversion factor. • As with the 'zero' above, set the conversion factor for a full
• Select menu 2 by pressing the MENU key twice. scale input using the keys and the ENTER
ZERO key. Press
the MENU key to finish editing.
• Select V or mA by pressing the key and then press
ENTER
ZERO (e.g. mA is selected).
e.g. Full Scale 20mA will
• The following options will appear, as shown below: be converted to 10 bar
Text:
The ‘Text’ function is used for changing the displayed units
when the ‘converted’ (see 1.9.2 on page 15) option is selected.
• Select ‘Measure’ using the key and press the ENTER
ZERO key If 20 mA is being measured from a device then this could be
converted and displayed as pressure in bar for example.
(6 characters max.)
Zero:
• Select 'Zero' using the key and then press the ENTER
ZERO key. Root:
• Select ‘Root’ using the key and press the ENTERZERO key.
• Using the keys and the ENTER
ZERO key, change the • Select ‘On’ or ‘Off’ using the key and press the ENTER
ZERO
value of the converted zero. Press the MENU key to finish key.
editing.
With ‘On’ is selected, any measurements will be displayed
having the root function applied until ‘Off’ is next selected.
e.g. 4.00 mA will be
converted to 0.00 bar
Note: Due to display constraints, the six character ‘Text’ readout
will be shortened to display the first three characters when the
‘Root’ function is switched on. ‘Root’ function is available in both
Actual and Converted modes.
Page 22
2.3.1 V, mA (Source mode) To set up conversion factors for Zero, Full Scale or a user
defined (Text) value.
• Select ‘Zero’ by pressing the ENTER
ZERO key.
• Change the zero conversion values as required using the
The second option for V and mA is ‘Source’ (output) mode. keys and the ENTER
ZERO key.
• Select ‘FS’ (full scale) using the key and press the ENTER
ZERO
• Select ‘FS’ using the key and press the ENTER key.
to each other as follows: ZERO
keys..
Status:
• Select ‘Status’ using the key and press the ENTER
ZERO key.
• Select ‘Off’ or ‘On’ and press the ENTER
ZERO key.
Root:
When switched on, this provides a voltage or current output
with the application of square root function.
• Select ‘Root’ using the key and press the ENTER
ZERO key.
• Switch the function ‘Off’ or ‘On’ using the and ENTER
ZERO key.
Status
• Select ‘Status’ by pressing the ENTER
ZERO key.
• Select ‘Off’ or ‘On’ using the key and press the ENTER
ZERO key
Rate
• Select ‘Rate’ with the key and press the ENTER
ZERO key.
• Change the V/sec or mA/sec value using the keys
and press the ENTER
ZERO key.
Dwell_Time
• Select ‘Dwell_time with the key and press the ENTER
ZERO key.
Page 24
Ramp Mode: (Not available in P to V / mA mode) Nudge Mode press and hold the MENU key for 2 seconds
V / mA
Ramp Mode to its starting point. In Step Mode, pressing the
Keys together returns to the first Step level.
Page 25
2.4 V-out, Display Light 2.6 Event
These functions may be selected as follows: • Press the MENU key twice to select the Set menu.
• Switch on PC6-PRO. • Select 'Event' using the key and then press the
• Press the MENU key twice to select the Set menu. ENTER
ZERO key. The Event menu is shown below:
• Select the required function using the keys
and then press the ENTER
ZERO key. Each of these functions has
an On or Off setting.
• Select the setting using the key and then press the Key:
ENTER key.
ZERO This results in RS232 (status ON) output / data storage
triggered by key press.
Note: The display light can also be set On or Off by holding • Select 'Key' using the key and then press the ENTER
ZERO key.
down the ON/OFF key for more than 2 seconds. The back light Timed:
is turned off when the PC6-PRO is switched off. This results in RS232 (status ON) output / data storage at fixed
time intervals set by the user.
Port: • Select 'Timed' using the key and then press the ENTER
ZERO
This function allows the user to select the current (displayed) key.
pressure module.
• Press the MENU key twice to select the Set Menu.
• Scroll to 'Port' using the key and press the ENTER
ZERO key
• Select a module using the keys and press the • Select the digit to be modified using the key and
ENTER
ZERO key to activate. press the ENTER
ZERO key. Use the keys to change the
value and press the ENTER
ZERO key.
2.5 Resolution • Repeat for next digit and press the ENTER
ZERO key.
To change the resolution proceed as follows: • Change the time measurement units in the same way and
• Press the MENU key twice to select the Set menu. press the ENTER
ZERO key.
• Select 'Resolution' using the keys and • Press the MENU key to return to the pressure display.
press the ENTER
ZERO key.
• Select the required resolution ( Low, Normal or High )
using the keys and then press the ENTER
ZERO key.
Note: Resolution will change by a factor of 10.
Page 26
2.7 Log • As with the filename, define the record count and then
press the MENU key when completed.
• Switch on PC6-PRO.
• If a file has already been set up with the same name
• Press the MENU key twice, scroll to and select 'Log' and
press ENTER
ZERO . then the ‘Overwrite' option screen will be displayed.
The display gives the option to log a file, view a file or
change the log status as shown below:
Page 27
2.7 Log continued Delete:
• With the file now selected, use the or ENTER
ZERO key to To delete a logged file, proceed as follows:
show the file contents in the sequence shown below. • Select ‘Delete' from the log menu using the key and
then press the ENTER
ZERO key.
• Select ‘File', ‘All' or ‘Exit' using the key and then
File Name press the ENTER
ZERO
key.
Number of Records
File:
One Record
• Using the keys, select a file and then press
the ENTER
ZERO key.
• A confirmation to delete file prompt is given. Select ‘Yes'
or ‘No' and then press the ENTER
ZERO key.
All:
• Select the ‘All' option using the key and then press
the ENTER
ZERO key.
Last Record
• A confirmation to delete all prompt is given. Select ‘Yes'
or 'No' and then press the ENTER
ZERO key.
All Files stored on the PC6-PRO will be deleted.
Exit:
Note: Use the or ENTER
ZERO key to scroll the other way. The • Select the ‘Exit' option using the key and then
direction in which the scrolling moves when using the press the ENTER
ZERO key.
ENTER
ZERO key depends on the last key pressed.
Press MENU twice to return to the 'Log' menu.
Page 28
2.7 Log (continued) 2.8 RS232
Status: RS232 standard communication protocol is available on the
Use this option to enable or disable logging function. PC6-PRO. Select the RS232 functions as follows:
• Select 'Status' from the log menu using the key and • Connect the PC6-PRO to a printer or computer as
then press the ENTER
ZERO key. described in paragraph 1.6.5 on page 11.
• Switch on PC6-PRO.
• Select the Set menu by pressing the MENU key twice.
• Select 'RS232' option using the key, then press the
• Select 'On' or 'Off' and then press the ENTER
ZERO key. ENTER key. The RS232 menu is displayed as below:
ZERO
• The display returns to the log menu.
Once a file has been set using the instructions given, the
logging process is shown as active with a disc shaped icon in
the top left hand corner of the display.
Status:
If you should press the MENU key whilst logging is in progress
then the following screen appears: • Select 'Status' using the ENTER
ZERO key. Choose 'On' or 'Off'
option with the key and then press the ENTERZERO key.
The display returns to the RS232 menu.
Baud Rate:
• Select 'Baud-rate' using the key and then press
Select 'Yes' or 'No' and then press the ENTER key.
ZERO
the ENTER
ZERO key. Select the required baud rate (1200, 2400,
Note: Pressing the PRINT
LOG key allows direct access to the log 4800 or 9600) using the keys and then press
menu. This is only possible if the log status icon is flashing in the ENTER key.
ZERO
the top left hand corner of the display.
Stop Bits:
• Select the 'Stop bits' using the key and then press
Export:
the ENTER
ZERO key. Select either 1 or 2 bits with the
See 2.12 on page 19 For Export procedure. keys and then press the ENTER key.
ZERO
Import:
See 2.13 on page 19 for Import procedure.
Page 29
2.8.1 DPM 2.9 Date & Time
To use the DPM (Digital Pressure Module) with the PC6-PRO:- To select Date & Time proceed as follows:
• Switch off the PC6-PRO and connect the RS232 cable from • Switch on PC6-PRO.
the DPM (supplied) to the 'RS232' socket of the PC6-PRO. • Press the MENU key twice to select Set menu.
• Switch the PC6-PRO back on. • Select 'Date & Time' option using the key and
• Press the MENU key twice. Scroll to 'Port' option using the then press the ENTER
ZERO key.
key and press the ENTER
ZERO key to select.
• Select 'DPM' using the key and press the ENTER
ZERO key.
File Export, File Import, downloading of languages, Cal Zero • Select the digit you want to change using the key and
and Cal Span are also not available. press ENTER
ZERO key. Use the keys to change the
value and then press ENTER
ZERO key.
While communicating with the DPM, the PC6-PRO will set (and • Repeat the above step to change the other digits to set
permanently change) the communications settings of the PC6- time and proceed to the next set of digits to change the
PRO and the DPM to the following: date.
baud-rate 9600 • Press the MENU key to return to the 'Date & Time' menu.
stop bits 1
To View date & time:
To alter the DPM RS232 settings (i.e. for use with a line printer) • Select 'View' using the key and then press the ENTER
ZERO
it will be necessary to use the software supplied with the DPM. key.
To return the RS232 settings on the PC6-PRO refer to section • Press the MENU key to return to the Date & Time menu.
2.8 on page 29.
Page 30
3.0 Language 3.2 Power_Off
• Press the MENU key twice.
• Scroll to ‘Language’ using either arrow key and press The PC6-PRO has an auto-power off function. The time after
ENTER
ZERO key to select. which the unit automatically switches off is adjustable from 1 to
999 minutes. However, the auto-power off will not operate under
the following conditions:-
If a key has been pressed during the auto-power off
• Select the desired language from the options provided time period.
and press the ENTER
ZERO key. If a leak test is running.
If the RS232 Status is On.
Note: To download different languages, see 3.5 on page 33. If the Log Status is On.
If the unit is online with Sical Pro
3.1 Zero ( not available on absolute units )
Note: Unlike the 'Tare' function (page 11), 'Zero' reset is a • Press the MENU key twice.
permanent setting. • Scroll to 'Power_Off' function using the key and press
To reset pressure reading to zero: the ENTER key.
ZERO
• Press the MENU key twice to select the Set menu. • Select 'Key' to turn the auto-power function off.
• Select ‘Zero’ option using the key and then press the • Select 'Timed' to set the auto-power off time (in minutes)
ENTER
ZERO key. and press the ENTER
ZERO key. This automatically turns the auto-
power function on.
• Set to desired time using the keys and press the
ENTER
ZERO key.
• Press the ENTERZERO key as instructed.
The Pressure Display will now show the pressure value as
zero. If however the pressure reading is not within 5% Full
Scale the following message will be displayed and the zero
reset will not be possible.
Page 31
CAL-MENU
(MENU 3)
Important note:
Should you wish to change zero or span settings at any time, the original calibration made at time of
manufacture by SI Pressure Instruments will no longer apply. We therefore take no responsibility for any false
or inaccurate readings that occur after any changes to zero and/or span have been made.
Calibration Procedure:
If necessary then the zero and span calibrations should be made using a deadweight tester with an accuracy
of 0.01% of reading or better, and carried out in a temperature controlled environment at 20°C ±2°.
To maintain the accuracy of the calibrator, do not increase the value of ‘tol-ppm’ above 50 ( see flow charts ).
For full calibration which includes Temperature Compensation between 0-50°C the PC6-PRO should be sent
back to SI Pressure Instruments.
Page 32
To access the Cal Menu, press all four top row buttons at the • The following message is then displayed:
same time. A beep will accompany this action. Press the MENU
key three times. The functions available are as follows: Example
Value
Note: Password
To gain access to the ‘Zero’ and ‘Span’ functions, you will be
• Press the ENTER
ZERO key twice and wait.
asked to enter a password. Enter the four-digit serial number
• The unit will now take fifty readings. If the readings taken are
located on the label on the rear panel of the unit.
not within the tolerance set then you will be prompted to
repeat the step. The tolerance may need adjusting in order to
3.2 Zero: Not available with the DPM (carry out zero/span continue.
operations using the software supplied with the DPM). • If the readings are within tolerance then a ‘ ** completed **
To calibrate zero : press ENTER’ message will be displayed. Press the ENTER key
ZERO
• Select ‘Zero’ using the key and press ENTER
ZERO key. to finish.
• Select ‘Continue’ option pressing the ENTER
ZERO key.
• A ‘Warming up...’ message will be displayed and after several Notes: If at any time during the procedure you wish to exit and
seconds you will be prompted to release pressure and then return to the ‘Cal menu’ then press the MENU key.
press ‘ENTER’ to continue. The message: 'Sure ? No Yes' appears. Select 'Yes' to exit or
• The following screen will then be displayed asking you to set 'No' to continue with the zero calibration.
a tolerance.
Absolute: For absolute units you will be presented with the
following screen. It displays the current pressure reading on the
top line and on the second line you set the corrected value.
To set span:
• Select ‘Span’ using the key and press the ENTER
ZERO key
• The procedure for setting span is exactly the same as the or ENTER
ZERO
zero calibration up to the ‘reg_val’ section. (Not applicable to
absolute units). MENU
Exit mA
File Date Switch
& Time Temp C Leak
Exit Resolution
Power_Off Event
Date Log
Port Zero Language & Time RS232
Page 36
Menu 1 sub-menu unit & Leak
Unit
2
ENTER
ZERO bar mbar MPa kPa p.s.i kg/cm
atm
Exit
inH2O
Leak ENTER
ZERO Start Set
time Exit
ENTER
ENTER ZERO
ZERO
Pressure ENTER
Display ZERO
ENTER
ZERO
000 sec/min/
Yes
ENTER
ZERO
ENTER
ZERO
010 ENTER
ZERO
ENTER
ZERO
010 ENTER
ZERO
ENTER
ZERO
hr/day
ENTER
ZERO MENU
re-start
?
No
MENU
Page 37
Menu 1 sub-menu V, mA
Status must be switched
on to acess source functions
(Menu 2)
V ENTER
ZERO Measure Source Exit
ENTER ENTER
ZERO ZERO ENTER
ZERO
mA Actual ENTER
ZERO Actual ENTER
ZERO
Pressure
Display
Actual reading Actual reading
shown in display shown in display
Converted ENTER
ZERO Converted ENTER
ZERO Converted reading
shown in display
Converted reading
shown in display
Exit P to V / mA ENTER
ZERO P to V / mA conversion
shown in display
Pressure
Display
Exit
Pressure
Display
Page 38
Menu 1 sub-menu File
File ENTER
ZERO View Export Import Delete Exit
Latest stop No
Latest
File File import ? File All Exit
Yes
ENTER ENTER ENTER
ZERO ZERO ZERO
MENU
First
ENTER
ZERO * ENTER
ZERO
stop ? File Sure ?
stop
Browse export ? Yes No
File Yes No
ENTER ENTER ENTER
ZERO ZERO
Yes No ZERO
or ENTER
ZERO
Completes
MENU Import Sure ?
No
Exit ? ENTER
ZERO
Yes stop ?
Yes No
MENU
Yes No
MENU
ENTER
ZERO
ENTER
ZERO
ENTER
* ZERO not necessary for file export to ‘SiCal PRO’.
Page 39
Menu 2 High & Low
For setting negative values
-000.0000
*
High ENTER
ZERO
ENTER
ZERO or ENTER
ZERO
Low -000.0000 or
*
ENTER ENTER ENTER
ZERO ZERO ZERO
Page 40
Menu 2 sub-menu V , mA (Measure)
V ENTER
ZERO Measure Source Exit
ENTER
ZERO
mA
Zero FS Text Root Exit
ENTER ENTER ENTER
ZERO ZERO ENTER ENTER
ZERO ZERO ZERO
ENTER ENTER
ZERO ZERO
ENTER
ZERO
Pressure
Measure Z. Conv
000.000 000.000
Measure FS Conv.
050.000 050.000
Off ENTER
ZERO
Display
Measure Text:
XXXXXX
ENTER ENTER
ZERO ZERO
ENTER
On ENTER
ZERO
ZERO
No Finished No Finished
Editing ? Editing ?
No Finished
MENU
Yes Yes
Editing ?
Menu Menu
Yes
Menu
Measure Z. Conv Measure FS Conv.
025.010 000.000 050.000 100.000
ENTER ENTER
ZERO ZERO
No Finished No Finished
Editing ? Editing ?
Yes Yes
Menu Menu
Page 41
Menu 2 sub-menu V , mA (Source - Range)
V ENTER
ZERO Measure Source Exit
Pressure
ENTER Display
ZERO
mA ENTER
ZERO
ENTER
ZERO
S.Range Z. Yes
Zero ENTER
ZERO 000.000
Change
Value?
ENTER
ZERO
ENTER
ZERO
No
Change value as required
MENU
Yes
S.Range FS. Change
FS ENTER ENTER ENTER
ZERO ZERO
ZERO 010.000 Value?
No
Change value as required
MENU
Exit ENTER
ZERO
Page 42
Menu 2 sub-menu V , mA (Source - Setup)
V ENTER
ZERO Measure Source Exit
Pressure
ENTER Display
ZERO
mA ENTER
ZERO
ENTER
ZERO
No
Change values as required
MENU
Yes
S.Conv. FS Conv. Change
FS ENTER ENTER ENTER
Value? ZERO ZERO
ZERO 010.000 010.000
No
Change values as required
MENU
Yes
S.Convert Text:
Text ENTER
ZERO XXX
Enter
text?
ENTER
ZERO
ENTER
ZERO
No
Change text as required
MENU
Exit ENTER
ZERO
Page 43
Menu 2 sub-menu V , mA (Source - P to V / mA)
V ENTER
ZERO Measure Source Exit
Pressure
ENTER Display
ZERO
mA ENTER
ZERO
ENTER
ZERO
No
Change values as required
MENU
Yes
S.P V FS Conv. Change
FS ENTER ENTER ENTER
ZERO Value? ZERO ZERO
010.000 010.000
No
Change values as required
MENU
Exit ENTER
ZERO
Page 44
Menu 2 sub-menu V , mA (Source - Setup)
V ENTER
ZERO Measure Source Exit Pressure
Display
ENTER
ZERO
ENTER
mA
ZERO
Status ENTER
ZERO On Off
ENTER ENTER
ZERO ZERO
Root ENTER
ZERO On Off
ENTER ENTER
ZERO ZERO
Step: Yes
Step ENTER
ZERO 025 (%FS)
Change
Value?
ENTER
ZERO
ENTER
ZERO
No
Change value as required
MENU
Ramp ENTER
ZERO Page 45
ENTER
ZERO Exit
Page 45
Menu 2 sub-menu V , mA (Source - Setup - Ramp)
Status
Root
Step
Ramp ENTER
ZERO Status Rate Dwell_time Exit
ENTER ENTER ENTER ENTER
ZERO ZERO ZERO ZERO
ENTER ENTER
ZERO ZERO
ENTER ENTER
On ENTER
ZERO
ZERO ZERO
No Finished No Finished
MENU
Editing ? Editing ?
Yes Yes
Menu Menu
Page 46
Menu 2 sub-menu Display light, V-out, Resolution, Event & RS232.
ENTER
ZERO
ENTER
ZERO
ENTER
ZERO
Pressure Display V-out
Resolution ENTER
ZERO Low Normal High
Display light
ENTER
ZERO
ENTER
On
Event
ZERO
ENTER
ZERO Key Timed Exit
Pressure
ENTER ENTER
ZERO
ENTER
ZERO
Display
ZERO
Pressure ENTER
Display Off ZERO
Pressure ENTER
ZERO
Display
sec/min/
100 ENTER
ZERO
ENTER
ZERO 1 10 ENTER
ZERO
ENTER
ZERO 111 ENTER
ZERO
ENTER
ZERO
hr/day
ENTER
ZERO MENU
Pressure
Display
RS232 ENTER
ZERO Status Baud-rate Stop bits Exit
ENTER ENTER ENTER ENTER
ZERO ZERO ZERO ZERO
On 1200 1 Pressure
ENTER ENTER Display
ZERO ZERO
2400
Off 4800 2
ENTER ENTER
ZERO ZERO
MENU
9600
MENU
ENTER or MENU
ZERO
Page 47
Pressure ENTER
display ZERO
Log ENTER
ZERO Log View Delete Status Export Import
ENTER ENTER
ZERO ZERO ENTER ENTER ENTER ENTER
ZERO ZERO ZERO ZERO
Yes Select File
File Exists ?
No
Last File
ENTER
* ZERO not necessary for file export to ‘SiCal PRO’.
Page 48
Menu 2 sub-menu Date&time , Language, Zero , Port, Power_Off
Language
Date&time ENTER
ZERO Set View Exit ENTER
ZERO
ENTER ENTER
ENTER
ZERO ZERO ZERO
Set: English
e.g. German Exit
Menu Pressure
ENTER
ZERO
Display
Change No
Language? ‘Exit’ or Menu
Finished Yes
000000 000000 ENTER
ZERO editing Menu Yes ENTER
Time ? ZERO
No
Edit next Pressure
digit Display
Select
language
ENTER
Port
ZERO
ENTER
ZERO P1 P2 DPM
PC6-PRO resets
ENTER ENTER ENTER
ZERO ZERO ZERO
Pressure
Display
Zero
Power_Off ENTER
ZERO Key Timed Exit ENTER
ZERO
Page 49
Menu 3 cal-menu zero
ENTER
ZERO
ENTER
ZERO
1.0100 bar Sure? Yes No
ENTER Yes No
Set: 1.1100 ZERO
Set: 1.11
Yes
Absolute unit ? Warming up . . . ENTER
ZERO
ENTER
ZERO
No
** COMPLETED **
Warming up . . . MENU Stop? Yes ENTER
ZERO Press ENTER
ENTER
ZERO No Pressure Display ENTER
ZERO
No
Stop? MENU Release Pressure
Press ENTER to cont..
Yes
ENTER
ZERO
Yes No
ENTER
ZERO
ENTER
ZERO
Change tol_ppm.
050
ENTER
ZERO 050 ENTER
ZERO
ENTER
ZERO
050 ENTER
ZERO
ENTER
ZERO
050 ENTER
ZERO
ENTER
ZERO
MENU
Pressure Display
No
Example
50 Within
ENTER ** COMPLETED ** counts 50 ppm ?
ENTER
ZERO
reg-val 00001f75
ZERO
Press ENTER Yes done Yes Press ENTER to cont..
x2
No MENU
Stop?
ENTER
ZERO Yes No ENTER
ZERO
Page 50
Menu 3 cal-menu Span
ENTER
ZERO
ENTER
ZERO Exit Continue
ENTER
ZERO
ENTER
ZERO
Yes No MENU
ENTER
ZERO
No No No
Example
Within 50 50
ENTER ** COMPLETED ** Tolerance ? counts
Within ENTER Set 20.00 bar counts
Within ENTER reg-val 00001f75
ZERO 50 ppm ? ZERO 50 ppm ? ZERO
Press ENTER Yes Yes done Yes Press ENTER to cont.. Yes done Yes Press ENTER to cont..
x2
No No MENU
Maximum
pressure Stop?
of transducer
ENTER
ZERO Yes No ENTER
ZERO
Page 51
Menu 3 cal-menu History
ENTER
ZERO Exit
Range 0000
Overload 0000
ENTER
ZERO or MENU
Page 52
Menu 3 cal-menu Language
Completes import
No Stop
import ?
Yes
MENU
Stop? Yes No
Yes No
ENTER ENTER
ZERO ZERO
Completes import
Page 53
PRESSURE
GENERATION
The PC6-PRO Pressure Calibrator is supplied with one of two pressure generation systems:-
These systems are not supplied with the PC6-PRO Indicator but can be purchased
separately from SI Pressure Instruments.
Instructions for use of these systems are available on the following pages.
Page 54
INSTRUCTIONS FOR USE OF HYDRAULIC HAND HELD
PRESSURE TEST SYSTEM - HTP1
WARNING : HIGH PRESSURE:- Uncontrolled release of high pressure is hazardous to personnel and may
cause damage to equipment. Before connection of any pressure component to the HTP1 ensure that the
component(s) is/are isolated from the pressure supply and any internal pressure is released slowly. Do not connect
HTP1 to external pressure source. Pressure built up internally during use can be extremely
high, ensure that all connections are made correctly. 1 4
9 *
KEY:
H
H IG
H
HIG
7. Priming / High pressure selector.
8. Fluid inlet tube. 7
9. Rear port: Reserved for PRV12 Pressure Relief Valve ONLY. PRIME / HIGH PRESSURE
( * Do not use for any other purpose ) SELECTOR
10. Pump Handles
Pump shown with
10 selector and handles
in 'Prime' position.
1. Remove filling plug (6) and fill reservoir (5) to the recommended level with the appropriate fluid and replace plug.
2. Connect master instrument to pump via connection (1) using the appropriate seals and adaptors.
3. Connect instrument under test to flexible hose / gauge adaptor and attach to pump via the quick-fit connection (4).
4. Adjust the Fine control (3) to 'mid-travel'.
5. Ensure the pressure release valve (2) is open ( turn fully clockwise then one turn anticlockwise ).
Fully squeeze handles 'in' and turn the selector (7) to the 'prime' position.
Page 55
6. Operate handles several times to expel air from the pump. ( Ensure that the fluid inlet tube (8) remains immersed in fluid at all
times).
8. Prime system by squeezing handles together and then releasing, allowing the oil to enter the pump cylinder. Repeat as
necessary until system is fully primed and low pressure is indicated on either the master or test instrument.
9. With handles fully squeezed 'in' select the 'high' pressure position on selector (7) and operate handles to generate
approximate pressure.
NOTE:- Smaller handle strokes enable easier pressure generation at high pressures.
10. Adjust pressure to required value using the fine control (3).
NOTE:- Pressure will fall slightly, immediately after pressure generation due to Thermodynamic effects, but will stabilise after
a short time.
11. To totally release pressure from the system turn release valve (2) one turn anticlockwise and select the 'prime' position on
selector (7) after first squeezing handles fully in.
NOTE:- Careful use of the release valve (2) and fine control (3) enables a controlled release of pressure, essential for
calibration purposes.
12. RESERVOIR FLUID LEVEL: If the fluid level in the reservoir falls considerably during use, a partial vacuum can be created
in the reservoir which may affect the pump performance. To avoid this, simply allow air to enter the reservoir by partly
unscrewing the filling plug (6).
13. SEAL REPLACEMENT: Dependant on the frequency of use, the Main Piston Seal
(and others) will need replacing. Replacement seals and instructions for fitting are
contained in the seal set ( HTPK1 ).
Page 56
OPERATING INSTRUCTIONS FOR THE HAND HELD PRESSURE TEST SYSTEM - TP1
8 1
Key: 2
7
8
The system is a portable dual source of vacuum and pressure. Each system incorporates a vacuum / pressure selector, a
volume control for fine adjustment and adjustable stroke to provide over pressure protection.
RELEASE VALVE (3)
This can be used to reduce or release the pressure in the system. The rate of pressure reduction is dependent upon the
degree of rotation when opening the valve. Minimal force is required to seal the system.
Page 57
IMPORTANT
Under no circumstances should the fine adjustment valve (2) be wound back beyond the red line
indicator on the body. Should this occur, then the pressure must be released from the system
before attempting to re-engage the fine adjustment valve.
1. Connect item under test to pressure port (6) at the end of the flexible hose (7) choosing correct adaptors and seals from
the set provided.
Note: Ensure seals are fitted and adaptors tightened to a maximum torque of 15 Nm.
2. Screw fine adjustment valve (2) fully clockwise.
3. Screw fine adjustment valve (2) 4 - 6 full turns anticlockwise.
4. Screw pressure release valve (3) fully clockwise, tightening to ensure good seal.
5. Operate handles until the pressure is close to that finally required. Ensure handles are fully squeezed together on each
stroke to achieve maximum pressure output.
6. Wind the fine adjustment valve (2) clockwise to increase pressure or anticlockwise to decrease pressure until required
pressure is reached.
Note: The pressure may take up to 1 minute to settle after increasing pressure due to thermodynamic effects, settling of seals
Page 58
and expansion of the flexible hose.
Caution: NEVER screw the fine adjustment valve (2) beyond the red line indicator.
7. Reductions in pressure can also be achieved by careful use of the pressure release valve (3).
8. Vacuum is achieved using the above procedure and having the changeover valve (4) pushed completely towards the
vacuum position.
Fault Investigation.
In the event that the system appears to lose pressure then the procedure above should be repeated ensuring new seals
are used, adaptors are tightened sufficiently and the pressure release valve (3) is tightened firmly.
Note: The connections to the hand held test system are sealed with 'o' ring or bonded seals and should not leak. The pipe
to body connection can be checked but tightened no more than 2 Nm.
DO NOT attempt to tighten the other fittings to the test system as this could lead to damage of sealed joints.
When testing for leaks it may be noticed that air is drawn in or expelled from around the changeover valve.
This is normal and should cause no concern.
Page 59
OPERATING INSTRUCTIONS FOR THE LOW PRESSURE HAND PUMP - LTP1
Key: 1
8. Nylon seals for BSP adaptors ( see seal kit provided ) DO NOT
use 'PTFE' tape for sealing with parallel threads.
9. Pressure Relief Valve.
4
The pump is a portable dual source of vacuum and pressure. Each pump
incorporates a vacuum / pressure selector and fine adjustment control.
6
RELEASE VALVE (1)
7
This can be used to reduce or release the pressure in the system. The rate of pressure
reduction is dependent upon the degree of rotation when opening the valve. Minimal force is 9 8
required to seal the system.
NOTE
The pump should only be used for pressurising small volumes due to its small displacement.
If the pump has not been used for a period of time, it could be difficult to operate on the first stroke.
For seal replacement, refer to service kit LTPK1 instructions.
1.3 Screw fine adjustment control (2) in fully and then out 4 - 6 full turns.
1.4 Screw pressure release valve (1) in fully, tightening to ensure good seal.
1.5 Using a small screwdriver, adjust the pressure relief valve (8) to set the desired maximum output pressure. Turn the grub
screw located in the main piston clockwise to increase or anti-clockwise to decrease the pressure setting.
1.6 Operate handle (6) until the pressure is close to that finally required.
1.7 Wind the fine adjustment control (2) in to increase pressure or out to decrease pressure until required pressure is reached.
Page 61
Note: The pressure may settle for up to 30 seconds after increasing pressure due to thermodynamic effects, settling of
seals and expansion of the flexible hose.
Caution: STOP unscrewing the fine adjustment control (2) when the top of the pump body becomes visible.
1.8 Reductions in pressure can also be achieved by careful use of the pressure release valve (1).
1.9 Vacuum is achieved using the above procedure and having the pressure / vacuum selector (3) pushed completely towards
the vacuum position.
When used with instruments such as the PC6-PRO Calibrator the connections and use are as for gauges above, however
the higher resolution available will amplify the visibility of the thermodynamic effects as mentioned in paragraph 1.7. These
will settle to useable values within one minute of pressurisation.
Note: On very high resolutions such as 1 mbar or 0.1 inches of water, small movements of the pipe may result in
noticeable pressure changes.
3. Fault Investigation.
In the event that the pump appears to lose pressure then the procedure above should be repeated ensuring new seals are
used, adaptors are tightened sufficiently and the pressure release valve (1) is tightened firmly.
Note: The connections to the hand held test system are sealed with 'o' ring or bonded seals and should not leak.
DO NOT attempt to tighten the other fittings to the pump as this could lead to damage of sealed joints.
When testing for leaks it may be noticed that air is drawn in or expelled from around the pressure / vacuum selector. This is
normal and should cause no concern.
Page 62
INDEX E
enter............................. iii
memory errors ........... 12
menu selection............ iii
safety warning............. ii
serial number ............. 6, 23, 34, 52
A EPM............................. iv,3,10,26 messages & parameters..12 set menu..................... 20, 24
absolute ...................... iv, 33 event............................ 26, 47 microprocessor............ iv short detected ........... 13
accuracy....................... iv export .......................... 18, 48 minimum (Min) ........... 15 software....................... v
activate a function........ iii mA ............................... 15, 21 span ............................ 32, 34, 51
alarm. ........................ 21, 21 F specifications............... Iv, v
arrangement................ 4 file ............................... 18, 39 N status .......................... 23, 24, 29
auto switch off.............. 12 filename ...................... 27 nudge ......................... 23 step ............................. 22, 24
flowcharts.................... 35 stop bits....................... 29
B functions...................... 5 O storage......................... 19
back-light.................... 26 FS (full scale) ............. iv, 22 overload....................... iv, 12 source ......................... 15, 23
battery......................... iv, 8 over range.................... 10 switch test ................... 11
baud-rate.................... iv, 29 H
hand pumps................. 54 P T
C high.............................. 20, 40 P to V / mA ................. 23 tare............................... ii, 15
cal menu..................... 32 history ......................... 34, 52 password .................... 33 temperature.................. 15
calibration procedure .. 32 PM_S/No ................... 8, 34 text .............................. 22
charger....................... 11 I Port ........................ 7, 30, 49 timed ........................... 11, 24
connection panel........ 3, 4 import .......................... 18, 48 Power_Off .................. 7, 31, 49
connections................ 3, 10 Indicator....................... i pressure connections .. v, 10 U
connection fault ......... 13 input............................. 2 pressure generation..... 54 unit .............................. 15, 36
convert ...................... 23 pressure media ........... v
coupling...................... 3, 4 K pressure range............ iv V
current input................ 10 keypad......................... 2 print.............................. 12 v, mA ........................... 15, 22, 37, 40
current overscale ....... 13 keys ............................ iii voltage input................ 10
Q voltage output.............. 11
D L quick reference............ iii voltage overscale
data entry.................... iii language .. ................. 31, 34, 49, 53 quick connect coupling.. v, 3, 10 v-out ........................... 26, 46
date & time.................. 30, 49 LCD............................ 3 (Quick-Fit)
display (LCD).............. 3 leak. (test) ................. 17, 37 R W
display light................. 26, 47 log............................... 26, 30, 48 range .......................... 23 warning........................ ii
display hold................. 13 low............................... 20, 40 ramp ........................... 22, 24, 25, 46
downloading ............... 18 rate ............................. 24, 46 Z
display symbols ......... 13 M resolution..................... iv, 26, 47 zero......................... 22, 30, 32, 49, 50
DPM ........................... 3, 7, 11, 13, 30 root ............................. 22, 29 zero calibration .......... 32, 33, 50
dwell-time ................... 24, 46 maximum (Max)........... 15 RS232 ......................... 11, 27, 47
measure ..................... 15, 22 S
Page 63
APPENDIX:
EPM
Datasheet
Page 64
EXTERNAL PRESSURE MODULE (EPM)
The EPM has been developed to work in conjunction with the PC6-PRO Pressure Calibrator to provide additional pressure ranges
without incurring the full cost of extra calibrators ( see illustration on Page 65 for typical application ).
The EPM is an intelligent module which contains a pressure transducer, temperature sensor, analogue to digital converter, and the
necessary calibration data stored during EPM calibration.
During operation the EPM sends pressure and temperature related information digitally to the PC6-PRO which acts as an intelligent
terminal.
The EPM or the PC6-PRO internal pressure sensor can be selected during calibration using the calibrator menu system.
SPECIFICATIONS
Model Pressure Range Acuracy (FS) Model Pressure Range Acuracy (FS)
bar psi up to bar psi up to
EPM-0.1 -0.1 to 0.1 1.5 0.025% EPM-40 -1 to 40 600 0.025%
EPM-1 -1 to 1 15 0.025% EPM-70 0 to 70 1000 0.025%
EPM-2 -1 to 2 30 0.025% EPM-100 0 to 100 1500 0.025%
EPM-5 -1 to 5 75 0.025% EPM-200 0 to 200 3,000 0.025%
EPM-10 -1 to 10 150 0.025% EPM-350 0 to 350 5,000 0.025%
Alternative pressure ranges and
EPM-20 -1 to 20 300 0.025% EPM-700 0 to 700 10,000 0.025% differential input versions
EPM-25 -1 to 25 360 0.025% EPM-1000 0 to 1000 15,000 0.025% are available to special order.
Overload Pressure 10% FS
Pressure Display Resolution 6 Digits
Operating Temperature: 0 to 40°C
Calibration Temperature: 20°C ±2°C
Storage Temperature: -20 to 70°C
Calibrated Temp. Range: 10 to 40°C
Temperature Effects: 0.001% of reading/°C between 10-40°C
Pressure Connection: ¼“BSP Male
Dimensions: 140mm x 31mm DIA, 1 meter cable
Weight: 450 grams approx.
Material: All Stainless Steel
Page 65
ITEM
UNDER
TEST
350 bar
EXTERNAL PRESSURE SOURCE
PC6
CALIBRATION
SYSTEM
10 bar
ITEM
UNDER
TEST
Page 66