Operator’s Manual
[Acoustic Power Data and Surface Temperature Data]
Table of Contents
i
© 2007-2013 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.
For this Operator’s Manual, the issue date is 2013-09.
Introduction
This manual gives all the transducers Acoustic Output Power data and Surface
Temperature Data for this Diagnostic Ultrasound System. Please refer to correlative
tables in use.
The acoustic output data of transducers are shown respectively in the following imaging
modes. The mode in the parenthesis refers to the imaging mode mentioned in Operator’s
Manual (Basic Volume).
z M Mode (M)
z B Mode (B)
z PW Mode (D)
z CW (D)
IMPORTANT!
I
How to Read the Acoustic Power Tables
NOTE: About the detail of acoustic power, please refer to the regulation of ALARA in
operator’s manual (Basic Volume).
1
Transducer Maximum Surface Temperature
* The measurement data were obtained under the test conditions employed at
Mindray.
Transducer model Maximum surface temperature Maximum surface temperature
*1
(°C) (°C)*1
*1: Values following the "±" mark indicate the expanded uncertainty with a confidence
level of 95%, t.975 = 4.30.
2
Acoustic Output Reporting Tables
10L4
Transducer Model: 10L4
Imaging Mode: M
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1cm 2
2 scan
1cm
Maximum Index Value 8.61E-01 1.22E-01 3.11E-01 1.86E-01
pr.a 2.35E+00
P 3.43E+00 3.43E+00 3.43E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.33E+00
Parameter z at max.Ipi,α 1.42E+00
deq(zb) 1.17E-01
fawf 7.60E+00 7.60E+00 7.60E+00 7.60E+00
Dim of X 2.06E-01 2.06E-01 2.06E-01
Aaprt Y 3.49E-01 3.49E-01 3.49E-01
td 2.82E-07
prr 2.00E+03
Other
pr at max. Ipi 3.09E+00
Information
deq at max. Ipi 1.13E-01
Ipa,α at max. MI 2.46E+02
3
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 9.19E-01 5.25E-01 5.17E-01
pr.a 2.53E+00
P 3.63E+01 3.63E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter
z at max.Ipi,α 1.40E+00
deq(zb)
fawf 7.61E+00 7.61E+00 7.61E+00
Dim of Aaprt X 2.09E-01 2.09E-01
Y 3.36E-01 3.36E-01
td 2.80E-07
prr 1.56E+04
Other
pr at max. Ipi 3.09E+00
Information
deq at max. Ipi
Ipa,α at max. MI 2.73E+02
Control 1 Focus Number: one focus; Focus Position: 2.0cm for B-mode
4
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1 cm 2 scan
1 cm2
Maximum Index Value 8.61E-01 4.58E-01 1.22E-01 3.11E-01 6.28E-01
pr.a 2.35E+00
P 3.17E+01 3.43E+00 3.43E+00 3.50E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.33E+00
Parameter
z at max.Ipi,α 1.42E+00
deq(zb) 1.17E-01
fawf 7.60E+00 7.60E+00 7.60E+00 7.60E+00
Dim of X 2.06E-01 2.06E-01 2.06E-01
Aaprt Y 3.49E-01 3.49E-01 3.49E-01
td 2.82E-07
prr 2.00E+03
Other
pr at max. Ipi 3.09E+00
Information
deq at max. Ipi 1.13E-01
Ipa,α at max. MI 2.46E+02
Control 5
5
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>1
Aaprt≤1 cm 2 scan
cm2
Maximum Index Value 6.39E-01 4.62E-01 1.20E+00 6.98E-01
pr.a 1.77E+00
P 1.29E+01 1.29E+01 1.29E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.37E+00
Parameter z at max.Ipi,α 1.47E+00
deq(zb) 1.09E-01
fawf 7.58E+00 7.58E+00 7.58E+00 7.58E+00
Dim of X 1.74E-01 1.74E-01 1.74E-01
Aaprt Y 2.84E-01 2.84E-01 2.84E-01
td 5.05E-07
prr 6.30E+03
Other
pr at max. Ipi 2.47E+00
Information
deq at max. Ipi 1.04E-01
Ipa,α at max. MI 2.00E+02
Control 1 Sampling Volume Position: at the 2.0cm focus of B-mode
6
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI
Scan Aaprt> Non-scan
2
Aaprt≤1 cm
1cm2
Maximum Index Value 6.43E-01 2.23E-01 1.29E-01 3.33E-01 4.11E-01
pr.a 1.78E+00
P 1.52E+01 3.62E+00 3.62E+00 1.88E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.40E+00
Parameter
z at max.Ipi,α 1.50E+00
deq(zb) 1.10E-01
fawf 7.59E+00 7.79E+00 7.59E+00 7.59E+00
Dim of X 1.89E-01 1.77E-01 1.77E-01
Aaprt Y 2.88E-01 2.86E-01 2.86E-01
td 5.05E-07
prr 1.80E+03
Other
pr at max. Ipi 2.42E+00
Information
deq at max. Ipi 1.05E-01
Ipa,α at max. MI 1.93E+02
Control 1 Focus Number: one focus, Focus Position: 2.0cm for B-mode
7
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt>1
Aaprt≤1cm 2
2 scan
cm
Maximum Index Value 8.41E-01 7.83E-01 8.48E-01
pr.a 2.29E+00
P 2.82E+01 2.82E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb
Parameter
z at max.Ipi,α 1.43E+00
deq(zb)
fawf 7.65E+00
Dim of X
Aaprt Y
td 2.79E-07
prr 1.56E+04
Other
pr at max. Ipi 2.82E+00
Information
deq at max. Ipi
Ipa,α at max. MI 2.24E+02
Control 1 B-Mode Focus Number: one focus; Focus Position: 2.0cm for B-mode
Color Sampling Gate Position: at the 2.0cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
Control 3 Depth: 2.50 cm
Operating
Control Control 4 Working Frequency: 8.0MHz for B-mode, and 6.6MHz for Color-mode
Conditions Scan Line Mode: minimum scan angle for B-mode, and lower-density
Control 5
line for B-mode and Color-mode, and color sensitivity : 3 .
Control 6 PRF:15600 Hz for B-mode, and 7800Hz for Color-mode
8
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>1
Aaprt≤1 cm 2 scan
cm2
Maximum Index Value 6.67E-01 4.92E-01 1.35E-01 3.55E-01 7.28E-01
pr.a 1.83E+00
P 1.71E+01 3.78E+00 3.78E+00 2.08E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.38E+00
Parameter
z at max.Ipi,α 1.47E+00
deq(zb) 1.10E-01
fawf 7.59E+00 7.59E+00 7.59E+00
Dim of X 1.82E-01 1.82E-01
Aaprt Y 2.85E-01 2.85E-01
td 4.96E-07
prr 1.80E+03
Other pr at max. Ipi 2.44E+00
Information deq at max. Ipi 1.05E-01
Ipa,α at max.
MI 1.91E+02
Control 1 B mode Focus Number : one focus; Focus Position: 2.0cm for B-mode
Sampling Volume Position: at the 2.0cm focus of B-mode;
Control 2 Color Sampling Gate Position: at the 2.0cm focus of B-mode.
9
Acoustic Output Reporting Tables
3C5A
Transducer Model: 3C5A
Imaging Mode: M
TIS TIB
Non-scan
Index Label MI TIC
Non-
Scan Aaprt≤ Aaprt> scan
2 2
1cm 1cm
Maximum Index Value
1.24E+00 4.71E-01 2.18E+00 9.07E-01
pr.a
1.87E+00
P
4.90E+01 4.90E+01
Min of [Pα(zs), 3.67E+0
Ita,α(zs)] 1
zs 1.00E+0
0
zbp 2.02E+0
Associate 0
Acoustic zb
4.25E+00
Parameter
z at max.Ipi,α
4.47E+00
deq(zb)
3.12E-01
fawf 2.36E+0
2.36E+00 0 2.36E+00 2.36E+00
Dim of X
Aaprt 5.04E-01 5.04E-01 5.04E-01
Y 7.31E-01 7.31E-01 7.31E-01
td
7.69E-07
prr
2.00E+03
Other pr at max. Ipi
Information 2.11E+00
deq at max. Ipi
3.00E-01
Ipa,α at max. MI
3.05E+02
Control 2 Depth:7.40cm
Operating
Control Control 3 Working Frequency: H6.0MHz
Conditions
Control 4 PRF: 2000Hz
10
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value
1.24E+00 2.99E-01 5.16E-01
pr.a
1.87E+00
P
1.17E+02 1.17E+02
Min of [Pα(zs),
Ita,α(zs)]
zs
zbp
Associate
Acoustic zb
Parameter
z at max.Ipi,α
4.47E+00
deq(zb)
fawf
2.36E+00 2.36E+00 2.36E+00
Dim of X 5.04E-01 5.04E-01
Aaprt
Y 7.31E-01 7.31E-01
td
7.69E-07
prr
Other 3.90E+03
Informatio pr at max. Ipi
2.11E+00
n deq at max. Ipi
Ipa,α at max. MI
3.05E+02
Control 1 Focal Zone Number: one; Focal Position: 6.0cm for B-mode
11
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI TIC
Non-
Scan Aaprt≤ Aaprt> 1 scan
1 cm
2 cm2
Maximum Index Value
1.24E+00 1.52E-01 4.71E-01 2.18E+00 1.14E+00
pr.a
1.87E+00
P
5.94E+01 4.90E+01 1.03E+02
Min of
[Pα(zs),
Ita,α(zs)] 3.67E+01
zs
1.00E+00
Associate zbp
2.02E+00
Acoustic zb
Parameter 4.25E+00
z at max.Ipi,α
4.47E+00
deq(zb)
3.12E-01
fawf
2.36E+00 2.36E+00 2.36E+00 2.36E+00 -
Dim of X
Aaprt 5.04E-01 5.04E-01 5.04E-01
Y 7.31E-01 7.31E-01 7.31E-01
td
7.69E-07
prr
2.00E+03
Other
pr at max. Ipi
Informatio 2.11E+00
n deq at max.
Ipi 3.00E-01
Ipa,α at max.
MI 3.05E+02
Control 1 Focal Position: at the 6.0cm of B-mode
12
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 2
Aaprt>1 cm scan
cm2
Maximum Index Value 1.23E+00 6.58E-01 3.07E+00 1.16E+00
pr.a 1.95E+00
P 6.78E+01 6.78E+01
Min of [Pα(zs),
Ita,α(zs)] 5.12E+01
zs 1.02E+00
Associate zbp 2.19E+00
Acoustic
zb 4.50E+00
Parameter
z at max.Ipi,α 4.62E+00
deq(zb) 2.63E-01
fawf 2.51E+00 2.51E+00 2.51E+00 2.51E+00
Dim of X 3.40E-01 3.40E-01 3.40E-01
Aaprt Y 7.30E-01 7.30E-01
td 1.51E-06
prr 1.80E+03
Other
pr at max. Ipi 2.49E+00
Information
deq at max. Ipi 2.57E-01
Ipa,α at max. MI 2.65E+02
Control 1 Sampling Volume Position: at the 6.0cm focus of B-mode
13
Acoustic Output Reporting Tables
14
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan
Aaprt≤1cm 2
Aaprt>1cm 2 scan
15
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt≤1
Aaprt>1 cm
2 scan
cm2
Maximum Index Value 1.04E+00 6.75E-01 5.47E-01 2.51E+00 2.08E+00
pr.a 1.65E+00
P 1.61E+02 5.66E+01 2.17E+02
Min of [Pα(zs),
Ita,α(zs)] 4.39E+01
zs 1.05E+00
Associate zbp 2.19E+00
Acoustic
zb 4.63E+00
Parameter
z at max.Ipi,α 4.85E+00
deq(zb) 2.62E-01
fawf 2.52E+00 2.52E+00 2.52E+00
Dim of X 3.43E-01 3.43E-01
Aaprt Y 7.20E-01 7.20E-01
td 1.51E-06
prr 2.40E+03
Other pr at max. Ipi 2.08E+00
Information deq at max. Ipi 2.54E-01
Ipa,α at max.
MI 1.70E+02
Control 1 B mode Focus Number : one focus; Focus Position: 6.0cm for B-mode
Sampling Volume Position: at the 6.0cm focus of B-mode;
Control 2 Color Sampling Gate Position: at the 6.0cm focus of B-mode.
16
Acoustic Output Reporting Tables
6CV1
Transducer Model: 6CV1
Imaging Mode: M
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1cm 2
2 scan
1cm
Maximum Index Value 1.19E+00 2.02E-01 6.74E-01 3.72E-01
pr.a 2.74E+00
P 7.97E+00 7.97E+00 7.97E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.55E+00
Parameter z at max.Ipi,α 1.60E+00
deq(zb) 1.52E-01
fawf 5.32E+00 5.32E+00 5.32E+00 5.32E+00
Dim of X 3.17E-01 3.17E-01 3.17E-01
Aaprt Y 3.52E-01 3.52E-01 3.52E-01
td 3.11E-07
prr 2.00E+03
Other
pr at max. Ipi 3.45E+00
Information
deq at max. Ipi 1.49E-01
Ipa,α at max. MI 3.99E+02
17
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 1.19E+00 1.02E-01 1.29E-01
pr.a 2.74E+00
P 7.66E+00 7.66E+00
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb
Parameter
z at max.Ipi,α 1.60E+00
deq(zb)
fawf 5.32E+00 5.32E+00 5.32E+00
Dim of X 3.17E-01 3.17E-01
Aaprt Y 3.52E-01 3.52E-01
td 3.11E-07
prr 2.40E+03
Other
pr at max. Ipi 3.45E+00
Information
deq at max. Ipi
Ipa,α at max. MI 3.99E+02
Control 1 Focus Number: one focus; Focus Position: 2.0cm for B-mode
18
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.19E+00 4.37E-02 2.02E-01 6.74E-01 4.26E-01
pr.a 2.74E+00
P 3.28E+00 7.97E+00 7.97E+00 1.12E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.55E+00
Parameter
z at max.Ipi,α 1.60E+00
deq(zb) 1.52E-01
fawf 5.32E+00 5.32E+00 5.32E+00 5.32E+00
Dim of Aaprt X 3.17E-01 3.17E-01 3.17E-01
Y 3.52E-01 3.52E-01 3.52E-01
td 3.11E-07
prr 2.00E+03
Other
pr at max. Ipi 3.45E+00
Information
deq at max. Ipi 1.49E-01
Ipa,α at max. MI 3.99E+02
Control 1 Focus Position: at the 2.0cm focus of B-mode
Operating Depth: 29.60cm
Control 2
Control
Control 3 Working Frequency: 5.0MHz
Conditions
Control 4 Highest transmission power (100%)
19
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 7.91E-01 3.24E-01 8.91E-01 6.70E-01
pr.a 1.72E+00
P 1.44E+01 1.44E+01 1.44E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.70E+00
Parameter
z at max.Ipi,α 1.85E+00
deq(zb) 1.82E-01
fawf 4.69E+00 4.69E+00 4.69E+00 4.69E+00
Dim of X 4.82E-01 4.82E-01 4.82E-01
Aaprt Y 3.04E-01 3.04E-01 3.04E-01
td 8.17E-07
prr 2.40E+03
Other
pr at max. Ipi 2.04E+00
Information
deq at max. Ipi 1.73E-01
Ipa,α at max. MI 1.58E+02
Control 1 Sampling Volume Position: at the 3.7cm focus of B-mode SV=0.5mm
20
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI
Scan Aaprt≤1 Aaprt> Non-scan
cm2 1 cm2
Maximum Index Value 5.34E-01 1.13E-01 7.44E-02 2.41E-01 2.61E-01
pr.a 1.15E+00
P 7.53E+00 3.38E+00 3.38E+00 1.04E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.67E+00
Parameter
z at max.Ipi,α 1.87E+00
deq(zb) 1.97E-01
fawf 4.71E+00 5.51E+00 4.71E+00 4.71E+00
Dim of Aaprt X 5.15E-01 5.02E-01 5.02E-01
Y 3.03E-01 2.92E-01 2.92E-01
td 8.20E-07
prr 1.70E+03
Other
pr at max. Ipi 1.31E+00
Information
deq at max. Ipi 1.85E-01
Ipa,α at max. MI 6.08E+01
Control 1 Focus Number: one focus, Focus Position: 3.7cm for B-mode
21
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤1 Aaprt>1
2 2 scan
cm cm
Maximum Index Value 4.40E-01 2.10E-01 2.81E-01
pr.a 1.04E+00
P 1.04E+01 1.04E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter z at max.Ipi,α 1.93E+00
deq(zb)
fawf 5.48E+00
Dim of X
Aaprt Y
td 3.24E-07
Other prr 1.04E+04
Informatio pr at max. Ipi 1.20E+00
n deq at max. Ipi
Ipa,α at max. MI 3.43E+01
Control 1 B-Mode Focus Number: one focus; Focus Position: 3.7cm for B-mode
Color Sampling Gate Position: at the 3.7cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
Control 3 Depth: 4.90 cm
Operating
Control Control 4 Working Frequency: 5.0MHz for B-mode, and 4.6MHz for Color-mode
Conditions Scan Line Mode: minimum scan angle for B-mode, and lower-density
Control 5
line for B-mode and Color-mode, and color sensitivity : 3 .
Control 6 PRF:10400 Hz for B-mode, and 5300Hz for Color-mode
22
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>
Aaprt≤1 cm 2 scan
1 cm2
Maximum Index Value 4.40E-01 1.37E-01 1.81E-01 5.56E-01 5.48E-01
pr.a 9.46E-01
P 7.58E+00 8.16E+00 8.16E+00 1.54E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.67E+00
Parameter
z at max.Ipi,α 1.83E+00
deq(zb) 1.96E-01
fawf 4.71E+00 4.71E+00 4.71E+00
Dim of Aaprt X 4.83E-01 4.83E-01
Y 3.34E-01 3.34E-01
td 8.21E-07
prr 5.70E+03
Other
pr at max. Ipi 1.10E+00
Information
deq at max. Ipi 1.86E-01
Ipa,α at max. MI 3.64E+01
Control 1 B mode Focus Number : one focus; Focus Position: 3.7cm for B-mode
Sampling Volume Position: at the 3.7cm focus of B-mode; SV=0.5mm
Control 2 Color Sampling Gate Position: at the 3.7cm focus of B-mode.
23
Acoustic Output Reporting Tables
7L4A
Transducer Model: 7L4A
Imaging Mode: M
TIS TIB
Non-scan
Index Label MI TIC
Non-
Scan Aaprt>
Aaprt≤1cm2 scan
2
1cm
Maximum Index Value
9.34E-01 2.09E-01 2.53E-01 2.42E-01
pr.a
2.54E+00
P
6.56E+00 6.56E+00 6.56E+00
Min of [Pα(zs),
Ita,α(zs)]
zs
zbp
Associate
Acoustic zb
1.27E+00
Parameter
z at max.Ipi,α
1.40E+00
deq(zb)
1.86E-01
fawf
6.99E+00 6.99E+00 6.99E+00 6.99E+00
Dim of X
Aaprt 2.68E-01 2.68E-01 2.68E-01
Y 2.44E-01 2.44E-01 2.44E-01
td
2.36E-07
prr
2.00E+03
Other
pr at max. Ipi
Informatio 2.96E+00
n
deq at max. Ipi
1.45E-01
Ipa,α at max. MI
2.73E+02
Control 1 Focal Position: at the 2.0cm of B-mode
24
Acoustic Output Reporting Tables
TIS TIB
zbp
Associate
Acoustic zb
Parameter
z at max.Ipi,α
1.40E+00
deq(zb)
fawf
6.99E+00 6.99E+00 6.99E+00
Dim of X
Aaprt 2.68E-01 2.68E-01
Y 2.44E-01 2.44E-01
td
2.36E-07
prr
5.70E+03
Other
pr at max. Ipi
Informatio 2.96E+00
n
deq at max. Ipi
Ipa,α at max. MI
2.73E+02
Control 1 Focal Zone Number: one; Focal Position: 2.0cm for B-mode
25
Acoustic Output Reporting Tables
TIS TIB
zbp
Associate
Acoustic zb
1.27E+00
Parameter
z at max.Ipi,α
1.40E+00
deq(zb)
1.86E-01
fawf
6.99E+00 6.99E+00 6.99E+00 6.99E+00 -
Dim of X
Aaprt 2.68E-01 2.68E-01 2.68E-01
Y 2.44E-01 2.44E-01 2.44E-01
td
2.36E-07
prr
2.00E+03
Other pr at max. Ipi
2.96E+00
Information
deq at max. Ipi
1.45E-01
Ipa,α at max.
MI 2.73E+02
Control 1 Focal Position: at the 2.0cm of B-mode
Control 2 Depth:11.10cm
Operating
Control Control 3 Working Frequency: 7.5MHz
Conditions
Control 4 PRF: 2000Hz
26
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>1
Aaprt≤1 cm 2 scan
cm2
Maximum Index Value 1.10E+00 2.64E-01 1.27E+00 4.53E-01
pr.a 2.39E+00
P 1.16E+01 1.16E+01 1.16E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.65E+00
Parameter
z at max.Ipi,α 1.68E+00
deq(zb) 1.30E-01
fawf 4.75E+00 4.75E+00 4.75E+00 4.75E+00
Dim of X 2.60E-01 2.60E-01 2.60E-01
Aaprt Y 2.22E-01 2.22E-01 2.22E-01
td 8.17E-07
prr 2.40E+03
Other
pr at max. Ipi 2.82E+00
Information
deq at max. Ipi 1.27E-01
Ipa,α at max. MI 3.06E+02
Control 1 Sampling Volume Position: at the 2.8cm focus of B-mode
27
Acoustic Output Reporting Tables
28
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt>1
Aaprt≤1cm 2
2 scan
cm
Maximum Index Value 6.32E-01 7.53E-01 8.12E-01
pr.a 1.48E+00
P 3.63E+01 3.63E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb
Parameter
z at max.Ipi,α 2.33E+00
deq(zb)
fawf 5.50E+00
Dim of X
Aaprt Y
td 6.93E-07
prr 1.14E+04
Other
pr at max. Ipi 1.71E+00
Information
deq at max. Ipi
Ipa,α at max. MI 1.53E+02
Control 1 B-Mode Focus Number: one focus; Focus Position: 2.8cm for B-mode
Color Sampling Gate Position: at the 2.8cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
Control 3 Depth: 3.70 cm
Operating
Control Control 4 Working Frequency: 5.0MHz for B-mode, and 5.5MHz for Color-mode
Conditions Scan Line Mode: minimum scan angle for B-mode, and lower-density line
Control 5
for B-mode and Color-mode, and color sensitivity : 3 .
Control 6 PRF:12500 Hz for B-mode, and 11400Hz for Color-mode
29
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>1
Aaprt≤1 cm 2 scan
cm2
Maximum Index Value 5.75E-01 4.79E-01 6.91E-02 2.93E-01 6.29E-01
pr.a 1.35E+00
P 2.86E+01 4.83E+00 4.83E+00 3.26E+01
Min of
[Pα(zs),
Ita,α(zs)]
zs
Associate
Acoustic zbp
Parameter zb 1.60E+00
z at max.Ipi,α 2.42E+00
deq(zb) 1.45E-01
fawf 5.52E+00 4.80E+00 4.80E+00
Dim of X 3.27E-01 3.27E-01
Aaprt Y 2.04E-01 2.04E-01
td 6.92E-07
prr 2.40E+03
Other pr at max. Ipi 1.48E+00
Information deq at max. Ipi 1.39E-01
Ipa,α at max.
MI 1.19E+02
Control 1 B mode Focus Number : one focus; Focus Position: 2.8cm for B-mode
Sampling Volume Position: at the 2.8cm focus of B-mode;
Control 2 Color Sampling Gate Position: at the 2.8cm focus of B-mode.
30
Acoustic Output Reporting Tables
7L6
Transducer Model: 7L6
Imaging Mode: M
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1cm 2
2 scan
1cm
Maximum Index Value 7.75E-01 3.12E-02 9.72E-02 3.63E-02
pr.a 1.95E+00
P 1.10E+00 1.10E+00 1.10E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.62E+00
Parameter z at max.Ipi,α 1.72E+00
deq(zb) 1.20E-01
fawf 6.17E+00 6.17E+00 6.17E+00 6.17E+00
Dim of X 2.32E-01 2.32E-01 2.32E-01
Aaprt Y 1.98E-01 1.98E-01 1.98E-01
td 1.22E-07
prr 2.00E+03
Other
pr at max. Ipi 2.32E+00
Information
deq at max. Ipi 1.15E-01
Ipa,α at max. MI 2.22E+02
Control 1 Focus Position: at the 2.8cm focus of B-mode
31
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 7.75E-01 7.36E-02 8.54E-02
pr.a 1.95E+00
P 1.06E+01 1.06E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter
z at max.Ipi,α 1.72E+00
deq(zb)
fawf 6.17E+00 6.17E+00 6.17E+00
Dim of Aaprt X 2.32E-01 2.32E-01
Y 1.98E-01 1.98E-01
td 1.22E-07
prr 1.25E+04
Other
pr at max. Ipi 2.32E+00
Information
deq at max. Ipi
Ipa,α at max. MI 2.22E+02
Control 1 Focus Number: one focus; Focus Position: 2.8cm for B-mode
32
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1cm2
Maximum Index Value 7.75E-01 5.72E-02 3.12E-02 9.72E-02 9.13E-02
pr.a 1.95E+00
P 8.23E+00 1.10E+00 1.10E+00 8.98E+00
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.62E+00
Parameter
z at max.Ipi,α 1.72E+00
deq(zb) 1.20E-01
fawf 6.17E+00 6.17E+00 6.17E+00 6.17E+00
Dim of Aaprt X 2.32E-01 2.32E-01 2.32E-01
Y 1.98E-01 1.98E-01 1.98E-01
td 1.22E-07
prr 2.00E+03
Other
pr at max. Ipi 2.32E+00
Information
deq at max. Ipi 1.15E-01
Ipa,α at max. MI 2.22E+02
Control 1 Focus Position: at the 2.8cm focus of B-mode
Control 5
33
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>1
Aaprt≤1 cm 2 scan
cm2
Maximum Index Value 8.38E-01 2.58E-01 1.10E+00 3.51E-01
pr.a 1.82E+00
P 1.15E+01 1.15E+01 1.15E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.92E+00
Parameter
z at max.Ipi,α 2.00E+00
deq(zb) 1.23E-01
fawf 4.74E+00 4.74E+00 4.74E+00 4.74E+00
Dim of X 1.93E-01 1.93E-01 1.93E-01
Aaprt Y 2.17E-01 2.17E-01 2.17E-01
td 8.18E-07
prr 2.70E+03
Other
pr at max. Ipi 2.28E+00
Information
deq at max. Ipi 1.20E-01
Ipa,α at max. MI 2.28E+02
Control 1 Sampling Volume Position: at the 2.8cm focus of B-mode
34
Acoustic Output Reporting Tables
35
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt>1
Aaprt≤1cm 2
2 scan
cm
Maximum Index Value 6.53E-01 3.39E-01 5.22E-01
pr.a 1.42E+00
P 3.36E+01 3.36E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb
Parameter
z at max.Ipi,α 2.32E+00
deq(zb)
fawf 4.73E+00
Dim of X
Aaprt Y
td 8.17E-07
prr 7.80E+03
Other
pr at max. Ipi 1.81E+00
Information
deq at max. Ipi
Ipa,α at max. MI 1.45E+02
Control 1 B-Mode Focus Number: one focus; Focus Position: 2.0cm for B-mode
Color Sampling Gate Position: at the 2.0cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
Control 3 Depth: 2.50 cm
Operating
Control Control 4 Working Frequency: 5.0MHz for B-mode, and 4.6MHz for Color-mode
Conditions Scan Line Mode: minimum scan angle for B-mode, and lower-density
Control 5
line for B-mode and Color-mode, and color sensitivity : 3 .
Control 6 PRF:15600 Hz for B-mode, and 7800Hz for Color-mode
36
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>1
Aaprt≤1 cm 2 scan
cm2
Maximum Index Value 7.49E-01 2.58E-01 1.55E-01 6.45E-01 5.37E-01
pr.a 1.63E+00
P 2.62E+01 6.87E+00 6.87E+00 2.82E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 1.95E+00
Parameter
z at max.Ipi,α 2.10E+00
deq(zb) 1.25E-01
fawf 4.75E+00 4.75E+00 4.75E+00
Dim of X 1.87E-01 1.87E-01
Aaprt Y 2.18E-01 2.18E-01
td 8.19E-07
prr 2.70E+03
Other
pr at max. Ipi 1.85E+00
Information
deq at max. Ipi 1.21E-01
Ipa,α at max. MI 1.44E+02
Control 1 B mode Focus Number : one focus; Focus Position: 2.8cm for B-mode
Sampling Volume Position: at the 2.8cm focus of B-mode;
Control 2 Color Sampling Gate Position: at the 2.8cm focus of B-mode.
37
Acoustic Output Reporting Tables
2P2
Transducer Model: 2P2
Imaging Mode: M
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt Aaprt> scan
≤1cm2 2
1cm
Maximum Index Value 1.27E+00 3.67E-01 1.46E+00 7.31E-01
pr.a 2.12E+00
P 4.06E+01 4.06E+01
Min of [Pα(zs), Ita,α(zs)]
2.51E+01
zs 2.00E+00
Associate zbp 2.08E+00
Acoustic zb 3.62E+00
Parameter z at max.Ipi,α 3.73E+00
deq(zb) 2.89E-01
fawf 2.82E+00 2.82E+00 2.82E+00 2.82E+00
Dim of X 3.23E-01 3.23E-01 3.23E-01
Aaprt Y 8.09E-01 8.09E-01 8.09E-01
td 5.39E-07
prr 2.00E+03
Other
pr at max. Ipi 2.78E+00
Information
deq at max. Ipi 2.76E-01
Ipa,α at max. MI 3.15E+02
38
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 1.27E+00 5.38E-01 7.53E-01
pr.a 2.12E+00
P 8.69E+01 8.69E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter
z at max.Ipi,α 3.73E+00
deq(zb)
fawf 2.82E+00 2.82E+00 2.82E+00
Dim of Aaprt X 3.23E-01 3.23E-01
Y 8.09E-01 8.09E-01
td 5.39E-07
Other prr 2.30E+03
Informatio pr at max. Ipi 2.78E+00
n deq at max. Ipi
Ipa,α at max. MI 3.15E+02
Control 1 Focus Number: one focus; Focus Position: 4.0cm for B-mode
39
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤ Aaprt> 1 scan
1 cm2 cm2
Maximum Index Value 1.27E+00 2.35E-01 3.67E-01 1.46E+00 1.05E+00
pr.a 2.12E+00
P 3.80E+01 4.06E+01 7.75E+01
Min of [Pα(zs), Ita,α(zs)] 2.51E+01
zs 2.00E+00
Associate zbp 2.08E+00
Acoustic zb 3.62E+00
Parameter z at max.Ipi,α 3.73E+00
deq(zb) 2.89E-01
fawf 2.82E+00 2.82E+00 2.82E+00 2.82E+00
Dim of Aaprt X 3.23E-01 3.23E-01 3.23E-01
Y 8.09E-01 8.09E-01 8.09E-01
td 5.39E-07
Other prr 2.00E+03
Informatio pr at max. Ipi 2.78E+00
n deq at max. Ipi 2.76E-01
Ipa,α at max. MI 3.15E+02
40
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤ Aaprt>1 scan
1 cm2 cm2
Maximum Index Value 1.36E+00 7.25E-01 2.93E+00 1.55E+00
pr.a 2.16E+00
P 8.59E+01 8.59E+01
Min of [Pα(zs), Ita,α(zs)] 4.51E+01
zs 2.00E+00
Associate zbp 2.08E+00
Acoustic zb 3.70E+00
Parameter z at max.Ipi,α 3.95E+00
deq(zb) 2.99E-01
fawf 2.52E+00 2.52E+00 2.52E+00 2.52E+00
Dim of X 3.46E-01 3.46E-01 3.46E-01
Aaprt Y 8.00E-01 8.00E-01
td 1.51E-06
Other prr 1.70E+03
Informatio pr at max. Ipi 2.50E+00
n deq at max. Ipi 2.88E-01
Ipa,α at max. MI 2.32E+02
41
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI
Scan Aaprt≤ Aaprt> 1 Non-scan
2 2
1 cm cm
Maximum Index Value 8.50E-01 6.58E-01 2.38E-01 1.10E+00 1.31E+00
pr.a 1.46E+00
P 1.00E+02 2.80E+01 1.24E+02
Min of [Pα(zs), Ita,α(zs)]
1.63E+01
zs 2.00E+00
Associate zbp 2.08E+00
Acoustic zb 3.70E+00
Parameter z at max.Ipi,α 4.03E+00
deq(zb) 3.19E-01
fawf 2.51E+00 2.51E+00 2.53E+00 2.53E+00
Dim of Aaprt X 4.61E-01 3.83E-01 3.83E-01
Y 7.94E-01 8.24E-01 8.24E-01
td 7.01E-07
Other prr 1.04E+04
Informatio pr at max. Ipi 1.25E+00
n deq at max. Ipi 3.04E-01
Ipa,α at max. MI 1.21E+02
Control 1 Focus Number: one focus, Focus Position: 4.0cm for B-mode
42
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤1 Aaprt>1
2 2 scan
cm cm
Maximum Index Value 6.15E-01 1.32E+00 2.12E+00
pr.a 1.06E+00
P 1.71E+02 1.71E+02
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter z at max.Ipi,α 4.47E+00
deq(zb)
fawf 2.52E+00
Dim of X
Aaprt Y
td 7.15E-07
Other prr 8.90E+03
Informatio pr at max. Ipi 9.69E-01
n deq at max. Ipi
Ipa,α at max. MI 5.04E+01
B-Mode Focus Number: one focus; Focus Position: 6.0cm for
Control 1
B-mode
Color Sampling Gate Position: at the 6.0cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
43
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt≤ Aaprt>1 scan
1 cm2 cm2
Maximum Index Value 7.07E-01 5.23E-01 4.05E-01 1.99E+00 1.72E+00
pr.a 1.21E+00
P 7.22E+01 4.82E+01 1.20E+02
Min of [Pα(zs), Ita,α(zs)] 2.54E+01
zs 2.00E+00
Associate zbp 2.08E+00
Acoustic zb 3.63E+00
Parameter z at max.Ipi,α 3.95E+00
deq(zb) 3.22E-01
fawf 2.51E+00 2.52E+00 2.52E+00
Dim of Aaprt X 3.73E-01 3.73E-01
Y 8.35E-01 8.35E-01
td 7.28E-07
Other prr 1.04E+04
Informatio pr at max. Ipi 9.38E-01
n deq at max. Ipi 3.08E-01
Ipa,α at max. MI 4.89E+01
B mode Focus Number : one focus; Focus Position: 4.0cm for
Control 1 B-mode
Sampling Volume Position: at the 4.0cm focus of B-mode;
Control 2 SV=0.5mm
Color Sampling Gate Position: at the 4.0cm focus of B-mode.
Control 3 Color Sampling Gate Width : Minimum Gate.
Operating
Control Control 4 Depth: 4.90cm
Conditions Working Frequency : 2.5MHz for B-mode, 2.5MHz for PW-mode, and
Control 5 2.5MHz for Color-mode, and color sensitivity: 3 .
Scan Line Mode: Minimum scan angle for B-mode, and lower-density
Control 6 line for B-mode and Color-mode.
PRF: 10400Hz for B-mode, 5300Hz for PW-mode, 5300 for
Control 7 Color-mode.
Control 8 Acoustic Power: Highest transmission power (100%).
44
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤ Aaprt>1 scan
1 cm2 cm2
Maximum Index Value 1.05E-01 1.16E+00 5.20E+00 2.59E+00
pr.a 1.49E-01
P 1.34E+02 1.34E+02
Min of [Pα(zs), Ita,α(zs)] 7.64E+01
zs 1.17E+00
Associate zbp 1.93E+00
Acoustic zb 1.97E+00
Parameter z at max.Ipi,α 2.55E+00
deq(zb) 4.36E-01
fawf 2.02E+00 2.02E+00 2.02E+00 2.02E+00
Dim of X 4.78E-01 4.78E-01 4.78E-01
Aaprt Y 1.04E+00 1.04E+00 1.04E+00
td
Other prr
Informatio pr at max. Ipi 1.42E-01
n deq at max. Ipi 4.04E-01
Ipa,α at max. MI 7.34E-01
Control 5
45
Acoustic Output Reporting Tables
3C1
Transducer Model: 3C1
Imaging Mode: M-mode
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1cm 2
2 scan
1cm
Maximum Index Value 1.35E+00 7.67E-01 2.38E+00 1.54E+00
pr.a 2.27E+00
P 5.81E+01 5.81E+01 5.81E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 3.02E+00
Parameter z at max.Ipi,α 3.15E+00
deq(zb) 3.25E-01
fawf 2.74E+00 2.74E+00 2.74E+00 2.74E+00
Dim of X 8.23E-01 8.23E-01 8.23E-01
Aaprt Y 6.16E-01 6.16E-01 6.16E-01
td 5.90E-07
prr 2.00E+03
Other
pr at max. Ipi 2.60E+00
Information
deq at max. Ipi 3.18E-01
Ipa,α at max. MI 3.82E+02
46
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 1.35E+00 4.22E-01 7.02E-01
pr.a 2.27E+00
P 1.02E+02 1.02E+02
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter
z at max.Ipi,α 3.15E+00
deq(zb)
fawf 2.74E+00 2.74E+00 2.74E+00
Dim of Aaprt X 8.23E-01 8.23E-01
Y 6.16E-01 6.16E-01
td 5.90E-07
Other prr 5.70E+03
Informatio pr at max. Ipi 2.60E+00
n deq at max. Ipi
Ipa,α at max. MI 3.82E+02
Control 1 Focus Number: one focus; Focus Position: 8.0cm for B-mode
47
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.35E+00 1.55E-01 7.67E-01 2.38E+00 1.79E+00
pr.a 2.27E+00
P 3.73E+01 5.81E+01 5.81E+01 9.51E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 3.02E+00
Parameter z at max.Ipi,α 3.15E+00
deq(zb) 3.25E-01
fawf 2.74E+00 2.74E+00 2.74E+00 2.74E+00
Dim of Aaprt X 8.23E-01 8.23E-01 8.23E-01
Y 6.16E-01 6.16E-01 6.16E-01
td 5.90E-07
Other prr 2.00E+03
Informatio pr at max. Ipi 2.60E+00
n deq at max. Ipi 3.18E-01
Ipa,α at max. MI 3.82E+02
48
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.16E+00 7.39E-01 2.45E+00 1.42E+00
pr.a 2.01E+00
P 5.17E+01 5.17E+01 5.17E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 3.02E+00
Parameter z at max.Ipi,α 3.15E+00
deq(zb) 2.73E-01
fawf 3.00E+00 3.00E+00 3.00E+00 3.00E+00
Dim of X 4.63E-01 4.63E-01 4.63E-01
Aaprt Y 5.78E-01 5.78E-01 5.78E-01
td 1.26E-06
Other prr 1.70E+03
Informatio pr at max. Ipi 2.26E+00
n deq at max. Ipi 2.65E-01
Ipa,α at max. MI 2.51E+02
49
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI
Scan Aaprt≤1 Aaprt> Non-scan
cm2 1 cm2
Maximum Index Value 8.64E-01 2.54E-01 4.80E-01 1.60E+00 1.31E+00
pr.a 1.50E+00
P 5.93E+01 3.35E+01 3.35E+01 9.18E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 3.12E+00
Parameter z at max.Ipi,α 3.30E+00
deq(zb) 2.73E-01
fawf 3.01E+00 2.86E+00 3.01E+00 3.01E+00
Dim of Aaprt X 7.19E-01 4.62E-01 4.62E-01
Y 5.89E-01 5.69E-01 5.69E-01
td 1.26E-06
Other prr 1.70E+03
Informatio pr at max. Ipi 1.85E+00
n deq at max. Ipi 2.62E-01
Ipa,α at max. MI 1.68E+02
Control 1 Focus Number: one focus, Focus Position: 4.0cm for B-mode
50
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤1 Aaprt>1
2 2 scan
cm cm
Maximum Index Value 7.64E-01 1.44E+00 2.26E+00
pr.a 1.32E+00
P 1.34E+02 1.34E+02
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter z at max.Ipi,α 3.68E+00
deq(zb)
fawf 2.98E+00
Dim of X
Aaprt Y
td 1.26E-06
Other prr 5.30E+03
Informatio pr at max. Ipi 1.57E+00
n deq at max. Ipi
Ipa,α at max. MI 9.56E+01
Control 1 B-Mode Focus Number: one focus; Focus Position: 4.0cm for B-mode
Color Sampling Gate Position: at the 4.0cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
Control 3 Depth: 6.20 cm
Operating
Control Control 4 Working Frequency: H5.0MHz for B-mode, and 3.0MHz for Color-mode
Conditions Scan Line Mode: minimum scan angle for B-mode, and lower-density
Control 5
line for B-mode and Color-mode, and color sensitivity : 3 .
51
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>
Aaprt≤1 cm 2 scan
1 cm2
Maximum Index Value 1.08E+0
9.08E-01 0 4.80E-01 1.60E+00 2.58E+00
pr.a 1.58E+00
P 1.18E+0
2 3.35E+01 3.35E+01 1.50E+02
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic
zb 3.10E+00
Parameter
z at max.Ipi,α 3.32E+00
deq(zb) 2.73E-01
fawf 3.01E+00 3.01E+00 3.01E+00
Dim of Aaprt X 4.63E-01 4.63E-01
Y 5.70E-01 5.70E-01
td 1.26E-06
Other prr 1.70E+03
Informatio pr at max. Ipi 1.85E+00
n deq at max. Ipi 2.61E-01
Ipa,α at max. MI 1.68E+02
B mode Focus Number : one focus; Focus Position: 4.0cm for
Control 1 B-mode
Sampling Volume Position: at the 4.0cm focus of B-mode;
Control 2 SV=0.5mm
Color Sampling Gate Position: at the 4.0cm focus of B-mode.
Control 3 Color Sampling Gate Width : Minimum Gate.
Operating
Control Control 4 Depth: 6.20cm
Conditions Working Frequency : H5.0MHz for B-mode, 3.0MHz for PW-mode,
Control 5 and 3.0MHz for Color-mode, and color sensitivity: 3 .
Scan Line Mode: Minimum scan angle for B-mode, and lower-density
Control 6 line for B-mode and Color-mode.
PRF: 8900Hz for B-mode, 1700Hz for PW-mode, 1700 for
Control 7 Color-mode.
Control 8 Acoustic Power: Highest transmission power (100%).
52
Acoustic Output Reporting Tables
7L5
Transducer Model: 7L5
Imaging Mode: M
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1cm 2
2 scan
1cm
Maximum Index Value 7.53E-01 1.03E-01 1.59E-01 1.25E-01
pr.a 2.19E+00
P 2.57E+00 2.57E+00 2.57E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.58E+00
Parameter z at max.Ipi,α 1.67E+00
deq(zb) 1.21E-01
fawf 8.50E+00 8.50E+00 8.50E+00 8.50E+00
Dim of X 2.82E-01 2.82E-01 2.82E-01
Aaprt Y 2.07E-01 2.07E-01 2.07E-01
td 2.12E-07
prr 2.00E+03
Other
pr at max. Ipi 3.10E+00
Information
deq at max. Ipi 1.14E-01
Ipa,α at max. MI 1.75E+02
53
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 7.53E-01 1.72E-01 1.35E-01
pr.a 2.19E+00
P 1.40E+01 1.40E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter
z at max.Ipi,α 1.67E+00
deq(zb)
fawf 8.50E+00 8.50E+00 8.50E+00
Dim of Aaprt X 2.82E-01 2.82E-01
Y 2.07E-01 2.07E-01
td 2.12E-07
Other prr 8.90E+03
Informatio pr at max. Ipi 3.10E+00
n deq at max. Ipi
Ipa,α at max. MI 1.75E+02
Control 1 Focus Number: one focus; Focus Position: 2.8cm for B-mode
54
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 7.53E-01 1.31E-01 1.03E-01 1.59E-01 2.10E-01
pr.a 2.19E+00
P 1.07E+01 2.57E+00 2.57E+00 1.26E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.58E+00
Parameter z at max.Ipi,α 1.67E+00
deq(zb) 1.21E-01
fawf 8.50E+00 8.50E+00 8.50E+00 8.50E+00
Dim of Aaprt X 2.82E-01 2.82E-01 2.82E-01
Y 2.07E-01 2.07E-01 2.07E-01
td 2.12E-07
Other prr 2.00E+03
Informatio pr at max. Ipi 3.10E+00
n deq at max. Ipi 1.14E-01
Ipa,α at max. MI 1.75E+02
55
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 8.88E-01 2.43E-01 8.65E-01 4.77E-01
pr.a 1.93E+00
P 1.08E+01 1.08E+01 1.08E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.55E+00
Parameter z at max.Ipi,α 1.65E+00
deq(zb) 1.25E-01
fawf 4.74E+00 4.74E+00 4.74E+00 4.74E+00
Dim of X 1.94E-01 1.94E-01 1.94E-01
Aaprt Y 3.09E-01 3.09E-01 3.09E-01
td 8.14E-07
Other prr 1.70E+03
Informatio pr at max. Ipi 2.35E+00
n deq at max. Ipi 1.21E-01
Ipa,α at max. MI 1.84E+02
56
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI
Scan Aaprt≤1 Aaprt> Non-scan
cm2 1 cm2
Maximum Index Value 7.99E-01 8.82E-02 1.96E-01 7.00E-01 4.07E-01
pr.a 1.74E+00
P 6.66E+00 8.66E+00 8.66E+00 1.16E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.57E+00
Parameter z at max.Ipi,α 1.63E+00
deq(zb) 1.25E-01
fawf 4.74E+00 9.07E+00 4.74E+00 4.74E+00
Dim of Aaprt X 2.15E-01 1.95E-01 1.95E-01
Y 1.92E-01 2.99E-01 2.99E-01
td 8.16E-07
Other prr 1.70E+03
Informatio pr at max. Ipi 2.19E+00
n deq at max. Ipi 1.22E-01
Ipa,α at max. MI 1.48E+02
Control 1 Focus Number: one focus, Focus Position: 2.0cm for B-mode
57
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤1 Aaprt>1
2 2 scan
cm cm
Maximum Index Value 6.50E-01 5.78E-01 7.79E-01
pr.a 1.41E+00
P 3.10E+01 3.10E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter z at max.Ipi,α 1.88E+00
deq(zb)
fawf 4.73E+00
Dim of X
Aaprt Y
td 8.15E-07
Other prr 1.14E+04
Informatio pr at max. Ipi 1.73E+00
n deq at max. Ipi
Ipa,α at max. MI 8.98E+01
Control 1 B-Mode Focus Number: one focus; Focus Position: 2.0cm for B-mode
Color Sampling Gate Position: at the 2.0cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
Control 3 Depth: 2.50 cm
Operating
Control Control 4 Working Frequency: 10.0MHz for B-mode, and 4.6MHz for Color-mode
Conditions Scan Line Mode: minimum scan angle for B-mode, and lower-density
Control 5
line for B-mode and Color-mode, and color sensitivity : 3 .
58
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>
Aaprt≤1 cm 2 scan
1 cm2
Maximum Index Value 6.99E-01 2.84E-01 1.51E-01 5.30E-01 6.26E-01
pr.a 1.52E+00
P 1.91E+01 6.67E+00 6.67E+00 2.29E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.53E+00
Parameter z at max.Ipi,α 1.68E+00
deq(zb) 1.28E-01
fawf 4.74E+00 4.74E+00 4.74E+00
Dim of Aaprt X 1.98E-01 1.98E-01
Y 3.02E-01 3.02E-01
td 8.16E-07
Other prr 1.70E+03
Informatio pr at max. Ipi 1.94E+00
n deq at max. Ipi 1.22E-01
Ipa,α at max. MI 1.14E+02
B mode Focus Number : one focus; Focus Position: 2.0cm for
Control 1 B-mode
Sampling Volume Position: at the 2.0cm focus of B-mode;
Control 2 SV=0.5mm
Color Sampling Gate Position: at the 2.0cm focus of B-mode.
Control 3 Color Sampling Gate Width : Minimum Gate.
Operating
Control Control 4 Depth: 2.50cm
Conditions Working Frequency : 10.0MHz for B-mode, 4.6MHz for PW-mode,
Control 5 and 4.6MHz for Color-mode, and color sensitivity: 3 .
Scan Line Mode: Minimum scan angle for B-mode, and lower-density
Control 6 line for B-mode and Color-mode.
PRF: 15600Hz for B-mode, 1700Hz for PW-mode, 1700 for
Control 7 Color-mode.
Control 8 Acoustic Power: Highest transmission power (100%).
59
Acoustic Output Reporting Tables
7LT4
Transducer Model: 7LT4
Imaging Mode: M
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1cm 2
2 scan
1cm
Maximum Index Value 1.77E+00 7.83E-02 3.65E-01 1.45E-01
pr.a 4.09E+00
P 3.08E+00 3.08E+00 3.08E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.48E+00
Parameter z at max.Ipi,α 1.52E+00
deq(zb) 1.21E-01
fawf 5.45E+00 5.45E+00 5.45E+00 5.45E+00
Dim of X 2.16E-01 2.16E-01 2.16E-01
Aaprt Y 2.78E-01 2.78E-01 2.78E-01
td 1.47E-07
prr 2.00E+03
Other
pr at max. Ipi 4.01E+00
Information
deq at max. Ipi 1.18E-01
Ipa,α at max. MI 6.17E+02
60
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 1.77E+00 1.42E-01 1.73E-01
pr.a 4.09E+00
P 1.43E+01 1.43E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter
z at max.Ipi,α 1.52E+00
deq(zb)
fawf 5.45E+00 5.45E+00 5.45E+00
Dim of Aaprt X 2.16E-01 2.16E-01
Y 2.78E-01 2.78E-01
td 1.47E-07
Other prr 7.80E+03
Informatio pr at max. Ipi 4.01E+00
n deq at max. Ipi
Ipa,α at max. MI 6.17E+02
Control 1 Focus Number: one focus; Focus Position: 2.0cm for B-mode
61
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.77E+00 1.03E-01 7.83E-02 3.65E-01 2.56E-01
pr.a 4.09E+00
P 1.03E+01 3.08E+00 3.08E+00 1.29E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.48E+00
Parameter z at max.Ipi,α 1.52E+00
deq(zb) 1.21E-01
fawf 5.45E+00 5.45E+00 5.45E+00 5.45E+00
Dim of Aaprt X 2.16E-01 2.16E-01 2.16E-01
Y 2.78E-01 2.78E-01 2.78E-01
td 1.47E-07
Other prr 2.00E+03
Informatio pr at max. Ipi 4.01E+00
n deq at max. Ipi 1.18E-01
Ipa,α at max. MI 6.17E+02
62
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.35E+00 2.76E-01 1.26E+00 5.15E-01
pr.a 2.92E+00
P 1.24E+01 1.24E+01 1.24E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.67E+00
Parameter z at max.Ipi,α 1.75E+00
deq(zb) 1.41E-01
fawf 4.70E+00 4.70E+00 4.70E+00 4.70E+00
Dim of X 3.31E-01 3.31E-01 3.31E-01
Aaprt Y 2.21E-01 2.21E-01 2.21E-01
td 8.15E-07
Other prr 1.80E+03
Informatio pr at max. Ipi 3.09E+00
n deq at max. Ipi 1.37E-01
Ipa,α at max. MI 3.71E+02
63
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI
Scan Aaprt≤1 Aaprt> Non-scan
cm2 1 cm2
Maximum Index Value 9.49E-01 8.06E-02 1.45E-01 6.21E-01 3.31E-01
pr.a 2.06E+00
P 6.78E+00 6.45E+00 6.45E+00 1.20E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.77E+00
Parameter z at max.Ipi,α 1.80E+00
deq(zb) 1.38E-01
fawf 4.72E+00 6.66E+00 4.72E+00 4.72E+00
Dim of Aaprt X 2.92E-01 3.21E-01 3.21E-01
Y 1.87E-01 2.09E-01 2.09E-01
td 8.15E-07
Other prr 1.70E+03
Informatio pr at max. Ipi 2.27E+00
n deq at max. Ipi 1.35E-01
Ipa,α at max. MI 2.12E+02
Control 1 Focus Number: one focus, Focus Position: 2.8cm for B-mode
64
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤1 Aaprt>1
2 2 scan
cm cm
Maximum Index Value 6.10E-01 3.97E-01 5.34E-01
pr.a 1.32E+00
P 2.02E+01 2.02E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter z at max.Ipi,α 1.87E+00
deq(zb)
fawf 4.72E+00
Dim of X
Aaprt Y
td 8.17E-07
Other prr 1.14E+04
Informatio pr at max. Ipi 1.54E+00
n deq at max. Ipi
Ipa,α at max. MI 1.30E+02
Control 1 B-Mode Focus Number: one focus; Focus Position: 2.8cm for B-mode
Color Sampling Gate Position: at the 2.8cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
Control 3 Depth: 3.70 cm
Operating
Control Control 4 Working Frequency: 5.0MHz for B-mode, and 4.6MHz for Color-mode
Conditions Scan Line Mode: minimum scan angle for B-mode, and lower-density
Control 5
line for B-mode and Color-mode, and color sensitivity : 3 .
65
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>
Aaprt≤1 cm 2 scan
1 cm2
Maximum Index Value 7.95E-01 3.30E-01 8.10E-02 3.66E-01 6.00E-01
pr.a 1.73E+00
P 1.99E+01 3.60E+00 3.60E+00 2.35E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.75E+00
Parameter z at max.Ipi,α 1.88E+00
deq(zb) 1.45E-01
fawf 4.73E+00 4.73E+00 4.73E+00
Dim of Aaprt X 3.07E-01 3.07E-01
Y 2.12E-01 2.12E-01
td 8.17E-07
Other prr 1.80E+03
Informatio pr at max. Ipi 1.81E+00
n deq at max. Ipi 1.39E-01
Ipa,α at max. MI 1.41E+02
B mode Focus Number : one focus; Focus Position: 2.8cm for
Control 1 B-mode
Sampling Volume Position: at the 2.8cm focus of B-mode;
Control 2 SV=0.5mm
Color Sampling Gate Position: at the 2.8cm focus of B-mode.
Control 3 Color Sampling Gate Width : Minimum Gate.
Operating
Control Control 4 Depth: 3.70cm
Conditions Working Frequency : 5.0MHz for B-mode, 4.6MHz for PW-mode, and
Control 5 4.6MHz for Color-mode, and color sensitivity: 3 .
Scan Line Mode: Minimum scan angle for B-mode, and lower-density
Control 6 line for B-mode and Color-mode.
PRF: 12500Hz for B-mode, 1800Hz for PW-mode, 1800 for
Control 7 Color-mode.
Control 8 Acoustic Power: Highest transmission power (100%).
66
Acoustic Output Reporting Tables
6C2
Transducer Model: 6C2
Imaging Mode: M
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1cm 2
2 scan
1cm
Maximum Index Value 1.23E+00 1.97E-01 6.45E-01 3.60E-01
pr.a 2.83E+00
P 7.71E+00 7.71E+00 7.71E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.47E+00
Parameter z at max.Ipi,α 1.53E+00
deq(zb) 1.59E-01
fawf 5.26E+00 5.26E+00 5.26E+00 5.26E+00
Dim of X 2.77E-01 2.77E-01 2.77E-01
Aaprt Y 3.96E-01 3.96E-01 3.96E-01
td 3.19E-07
prr 2.00E+03
Other
pr at max. Ipi 3.29E+00
Information
deq at max. Ipi 1.55E-01
Ipa,α at max. MI 3.76E+02
67
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 1.23E+00 2.61E-01 3.23E-01
pr.a 2.83E+00
P 1.90E+01 1.90E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter
z at max.Ipi,α 1.53E+00
deq(zb)
fawf 5.26E+00 5.26E+00 5.26E+00
Dim of Aaprt X 2.77E-01 2.77E-01
Y 3.96E-01 3.96E-01
td 3.19E-07
Other prr 5.20E+03
Informatio pr at max. Ipi 3.29E+00
n deq at max. Ipi
Ipa,α at max. MI 3.76E+02
Control 1 Focus Number: one focus; Focus Position: 2.0cm for B-mode
68
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.23E+00 2.02E-01 1.97E-01 6.45E-01 6.05E-01
pr.a 2.83E+00
P 1.48E+01 7.71E+00 7.71E+00 2.23E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.47E+00
Parameter z at max.Ipi,α 1.53E+00
deq(zb) 1.59E-01
fawf 5.26E+00 5.26E+00 5.26E+00 5.26E+00
Dim of Aaprt X 2.77E-01 2.77E-01 2.77E-01
Y 3.96E-01 3.96E-01 3.96E-01
td 3.19E-07
Other prr 2.00E+03
Informatio pr at max. Ipi 3.29E+00
n deq at max. Ipi 1.55E-01
Ipa,α at max. MI 3.76E+02
69
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 7.83E-01 4.31E-01 1.50E+00 9.97E-01
pr.a 1.69E+00
P 1.93E+01 1.93E+01 1.93E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.45E+00
Parameter z at max.Ipi,α 1.58E+00
deq(zb) 1.76E-01
fawf 4.72E+00 4.72E+00 4.72E+00 4.72E+00
Dim of X 3.57E-01 3.57E-01 3.57E-01
Aaprt Y 3.63E-01 3.63E-01 3.63E-01
td 8.19E-07
Other prr 4.40E+03
Informatio pr at max. Ipi 2.00E+00
n deq at max. Ipi 1.69E-01
Ipa,α at max. MI 1.38E+02
70
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI
Scan Aaprt≤1 Aaprt> Non-scan
cm2 1 cm2
Maximum Index Value 6.16E-01 2.24E-01 9.97E-02 3.35E-01 4.98E-01
pr.a 1.33E+00
P 1.61E+01 4.48E+00 4.48E+00 2.04E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.43E+00
Parameter z at max.Ipi,α 1.60E+00
deq(zb) 1.87E-01
fawf 4.74E+00 5.38E+00 4.74E+00 4.74E+00
Dim of Aaprt X 4.00E-01 3.55E-01 3.55E-01
Y 3.01E-01 3.75E-01 3.75E-01
td 8.21E-07
Other prr 1.65E+03
Informatio pr at max. Ipi 1.51E+00
n deq at max. Ipi 1.76E-01
Ipa,α at max. MI 1.14E+02
Control 1 Focus Number: one focus, Focus Position: 2.8cm for B-mode
71
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤1 Aaprt>1
2 2 scan
cm cm
Maximum Index Value 7.74E-01 5.54E-01 7.51E-01
pr.a 1.78E+00
P 2.21E+01 2.21E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter z at max.Ipi,α 1.98E+00
deq(zb)
fawf 5.31E+00
Dim of X
Aaprt Y
td 3.16E-07
Other prr 1.25E+04
Informatio pr at max. Ipi 2.31E+00
n deq at max. Ipi
Ipa,α at max. MI 1.66E+02
Control 1 B-Mode Focus Number: one focus; Focus Position: 2.8cm for B-mode
Color Sampling Gate Position: at the 2.8cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
Control 3 Depth: 3.70 cm
Operating
Control Control 4 Working Frequency: 5.0MHz for B-mode, and 4.6MHz for Color-mode
Conditions Scan Line Mode: minimum scan angle for B-mode, and lower-density
Control 5
line for B-mode and Color-mode, and color sensitivity : 3 .
72
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>
Aaprt≤1 cm 2 scan
1 cm2
Maximum Index Value 5.86E-01 2.78E-01 3.19E-01 1.09E+00 1.07E+00
pr.a 1.26E+00
P 1.32E+01 1.44E+01 1.44E+01 2.64E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.43E+00
Parameter z at max.Ipi,α 1.60E+00
deq(zb) 1.81E-01
fawf 4.73E+00 4.73E+00 4.73E+00
Dim of Aaprt X 3.45E-01 3.45E-01
Y 3.66E-01 3.66E-01
td 8.21E-07
Other prr 6.30E+03
Informatio pr at max. Ipi 1.44E+00
n deq at max. Ipi 1.71E-01
Ipa,α at max. MI 9.63E+01
B mode Focus Number : one focus; Focus Position: 2.8cm for
Control 1 B-mode
Sampling Volume Position: at the 2.8cm focus of B-mode;
Control 2 SV=0.5mm
Color Sampling Gate Position: at the 2.8cm focus of B-mode.
Control 3 Color Sampling Gate Width : Minimum Gate.
Operating
Control Control 4 Depth: 3.70cm
Conditions Working Frequency : 5.0MHz for B-mode, 4.6MHz for PW-mode, and
Control 5 4.6MHz for Color-mode, and color sensitivity: 3 .
Scan Line Mode: Minimum scan angle for B-mode, and lower-density
Control 6 line for B-mode and Color-mode.
PRF: 12500Hz for B-mode, 6300Hz for PW-mode, 6300 for
Control 7 Color-mode.
Control 8 Acoustic Power: Highest transmission power (100%).
73
Acoustic Output Reporting Tables
6LB7(Convex)
Transducer Model: 6LB7 (Convex)
Imaging Mode: M
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1cm 2
2 scan
1cm
Maximum Index Value 1.39E+00 2.26E-01 7.64E-01 3.56E-01
pr.a 3.33E+00
P 8.26E+00 8.26E+00 8.26E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.38E+00
Parameter z at max.Ipi,α 1.47E+00
deq(zb) 1.40E-01
fawf 5.67E+00 5.67E+00 5.67E+00 5.67E+00
Dim of X 2.85E-01 2.85E-01 2.85E-01
Aaprt Y 3.00E-01 3.00E-01 3.00E-01
td 2.93E-07
prr 2.00E+03
Other
pr at max. Ipi 3.89E+00
Information
deq at max. Ipi 1.36E-01
Ipa,α at max. MI 5.38E+02
74
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 1.39E+00 1.07E-01 1.23E-01
pr.a 3.33E+00
P 8.37E+00 8.37E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter
z at max.Ipi,α 1.47E+00
deq(zb)
fawf 5.67E+00 5.67E+00 5.67E+00
Dim of Aaprt X 2.85E-01 2.85E-01
Y 3.00E-01 3.00E-01
td 2.93E-07
Other prr 2.90E+03
Informatio pr at max. Ipi 3.89E+00
n deq at max. Ipi
Ipa,α at max. MI 5.38E+02
Control 1 Focus Number: one focus; Focus Position: 2.0cm for B-mode
75
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.39E+00 3.74E-02 2.26E-01 7.64E-01 3.98E-01
pr.a 3.33E+00
P 2.92E+00 8.26E+00 8.26E+00 1.11E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.38E+00
Parameter z at max.Ipi,α 1.47E+00
deq(zb) 1.40E-01
fawf 5.67E+00 5.67E+00 5.67E+00 5.67E+00
Dim of Aaprt X 2.85E-01 2.85E-01 2.85E-01
Y 3.00E-01 3.00E-01 3.00E-01
td 2.93E-07
Other prr 2.00E+03
Informatio pr at max. Ipi 3.89E+00
n deq at max. Ipi 1.36E-01
Ipa,α at max. MI 5.38E+02
76
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 9.19E-01 3.47E-01 1.44E+00 7.33E-01
pr.a 1.99E+00
P 1.54E+01 1.54E+01 1.54E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.40E+00
Parameter z at max.Ipi,α 1.47E+00
deq(zb) 1.50E-01
fawf 4.73E+00 4.73E+00 4.73E+00 4.73E+00
Dim of X 3.50E-01 3.50E-01 3.50E-01
Aaprt Y 2.93E-01 2.93E-01 2.93E-01
td 8.20E-07
Other prr 3.00E+03
Informatio pr at max. Ipi 2.16E+00
n deq at max. Ipi 1.46E-01
Ipa,α at max. MI 2.16E+02
77
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI
Scan Aaprt≤1 Aaprt> Non-scan
cm2 1 cm2
Maximum Index Value 6.08E-01 1.08E-01 5.80E-02 2.48E-01 2.38E-01
pr.a 1.48E+00
P 8.17E+00 2.55E+00 2.55E+00 1.06E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.43E+00
Parameter z at max.Ipi,α 1.95E+00
deq(zb) 1.55E-01
fawf 5.90E+00 5.90E+00 4.76E+00 4.76E+00
Dim of Aaprt X 4.17E-01 3.77E-01 3.77E-01
Y 1.90E-01 3.11E-01 3.11E-01
td 2.87E-07
Other prr 1.25E+04
Informatio pr at max. Ipi 1.72E+00
n deq at max. Ipi 1.50E-01
Ipa,α at max. MI 1.06E+02
Control 1 Focus Number: one focus, Focus Position: 2.8cm for B-mode
78
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤1 Aaprt>1
2 2 scan
cm cm
Maximum Index Value 8.58E-01 2.08E-01 2.85E-01
pr.a 2.07E+00
P 8.87E+00 8.87E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter z at max.Ipi,α 1.87E+00
deq(zb)
fawf 5.78E+00
Dim of X
Aaprt Y
td 2.88E-07
Other prr 1.25E+04
Informatio pr at max. Ipi 2.29E+00
n deq at max. Ipi
Ipa,α at max. MI 1.84E+02
Control 1 B-Mode Focus Number: one focus; Focus Position: 2.8cm for B-mode
Color Sampling Gate Position: at the 2.8cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
Control 3 Depth: 3.70 cm
Operating
Control Control 4 Working Frequency: 5.0MHz for B-mode, and 4.6MHz for Color-mode
Conditions Scan Line Mode: minimum scan angle for B-mode, and lower-density
Control 5
line for B-mode and Color-mode, and color sensitivity : 3 .
79
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>
Aaprt≤1 cm 2 scan
1 cm2
Maximum Index Value 5.30E-01 1.47E-01 2.00E-01 8.27E-01 6.13E-01
pr.a 1.30E+00
P 7.44E+00 8.87E+00 8.87E+00 1.60E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.40E+00
Parameter z at max.Ipi,α 1.95E+00
deq(zb) 1.60E-01
fawf 5.93E+00 4.75E+00 4.75E+00
Dim of Aaprt X 3.73E-01 3.73E-01
Y 2.98E-01 2.98E-01
td 2.89E-07
Other prr 1.25E+04
Informatio pr at max. Ipi 1.56E+00
n deq at max. Ipi 1.54E-01
Ipa,α at max. MI 8.40E+01
B mode Focus Number : one focus; Focus Position: 2.8cm for
Control 1 B-mode
Sampling Volume Position: at the 2.8cm focus of B-mode;
Control 2 SV=0.5mm
Color Sampling Gate Position: at the 2.8cm focus of B-mode.
Control 3 Color Sampling Gate Width : Minimum Gate.
Operating
Control Control 4 Depth: 3.70cm
Conditions Working Frequency : 5.0MHz for B-mode, 4.6MHz for PW-mode, and
Control 5 4.6MHz for Color-mode, and color sensitivity: 3 .
Scan Line Mode: Minimum scan angle for B-mode, and lower-density
Control 6 line for B-mode and Color-mode.
PRF: 12500Hz for B-mode, 6300Hz for PW-mode, 6300 for
Control 7 Color-mode.
Control 8 Acoustic Power: Highest transmission power (100%).
80
Acoustic Output Reporting Tables
6LB7(Linear)
Transducer Model: 6LB7 (Linear)
Imaging Mode: M
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1cm 2
2 scan
1cm
Maximum Index Value 1.77E+00 2.14E-01 8.96E-01 2.36E-01
pr.a 4.19E+00
P 8.04E+00 8.04E+00 8.04E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.83E+00
Parameter z at max.Ipi,α 1.85E+00
deq(zb) 9.69E-02
fawf 5.46E+00 5.46E+00 5.46E+00 5.46E+00
Dim of X 1.25E-01 1.25E-01 1.25E-01
Aaprt Y 2.60E-01 2.60E-01 2.60E-01
td 3.11E-07
prr 2.00E+03
Other
pr at max. Ipi 5.47E+00
Information
deq at max. Ipi 9.61E-02
Ipa,α at max. MI 8.36E+02
81
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 1.77E+00 1.30E-01 1.57E-01
pr.a 4.19E+00
P 2.33E+01 2.33E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter
z at max.Ipi,α 1.85E+00
deq(zb)
fawf 5.46E+00 5.46E+00 5.46E+00
Dim of Aaprt X 1.25E-01 1.25E-01
Y 2.60E-01 2.60E-01
td 3.11E-07
Other prr 2.90E+03
Informatio pr at max. Ipi 5.47E+00
n deq at max. Ipi
Ipa,α at max. MI 8.36E+02
Control 1 Focus Number: one focus; Focus Position: 2.0cm for B-mode
82
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.77E+00 4.52E-02 2.14E-01 8.96E-01 2.80E-01
pr.a 4.19E+00
P 8.11E+00 8.04E+00 8.04E+00 1.47E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.83E+00
Parameter z at max.Ipi,α 1.85E+00
deq(zb) 9.69E-02
fawf 5.46E+00 5.46E+00 5.46E+00 5.46E+00
Dim of Aaprt X 1.25E-01 1.25E-01 1.25E-01
Y 2.60E-01 2.60E-01 2.60E-01
td 3.11E-07
Other prr 2.00E+03
Informatio pr at max. Ipi 5.47E+00
n deq at max. Ipi 9.61E-02
Ipa,α at max. MI 8.36E+02
83
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.30E+00 2.42E-01 1.23E+00 4.01E-01
pr.a 2.83E+00
P 1.09E+01 1.09E+01 1.09E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.65E+00
Parameter z at max.Ipi,α 1.72E+00
deq(zb) 1.14E-01
fawf 4.68E+00 4.68E+00 4.68E+00 4.68E+00
Dim of X 1.57E-01 1.57E-01 1.57E-01
Aaprt Y 2.42E-01 2.42E-01 2.42E-01
td 8.16E-07
Other prr 1.80E+03
Informatio pr at max. Ipi 3.34E+00
n deq at max. Ipi 1.12E-01
Ipa,α at max. MI 4.14E+02
84
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI
Scan Aaprt≤1 Aaprt> Non-scan
cm2 1 cm2
Maximum Index Value 9.77E-01 1.65E-01 9.29E-02 5.05E-01 3.43E-01
pr.a 2.37E+00
P 2.78E+01 4.21E+00 4.21E+00 3.21E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.72E+00
Parameter z at max.Ipi,α 2.07E+00
deq(zb) 1.14E-01
fawf 5.77E+00 5.77E+00 4.70E+00 4.70E+00
Dim of Aaprt X 1.08E-01 1.52E-01 1.52E-01
Y 1.98E-01 2.29E-01 2.29E-01
td 2.87E-07
Other prr 1.56E+04
Informatio pr at max. Ipi 3.20E+00
n deq at max. Ipi 1.12E-01
Ipa,α at max. MI 2.98E+02
Control 1 Focus Number: one focus, Focus Position: 2.0cm for B-mode
85
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤1 Aaprt>1
2 2 scan
cm cm
Maximum Index Value 1.05E+00 4.90E-01 6.71E-01
pr.a 2.49E+00
P 4.52E+01 4.52E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter z at max.Ipi,α 1.38E+00
deq(zb)
fawf 5.67E+00
Dim of X
Aaprt Y
td 3.01E-07
Other prr 1.56E+04
Informatio pr at max. Ipi 2.95E+00
n deq at max. Ipi
Ipa,α at max. MI 2.91E+02
B-Mode Focus Number: one focus; Focus Position: 1.2cm for
Control 1
B-mode
Color Sampling Gate Position: at the 1.2cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
86
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>
Aaprt≤1 cm 2 scan
1 cm2
Maximum Index Value 7.60E-01 1.58E-01 2.23E-01 9.67E-01 5.63E-01
pr.a 1.83E+00
P 1.80E+01 1.00E+01 1.00E+01 2.68E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.73E+00
Parameter z at max.Ipi,α 2.13E+00
deq(zb) 1.19E-01
fawf 5.87E+00 4.70E+00 4.70E+00
Dim of Aaprt X 1.62E-01 1.62E-01
Y 2.31E-01 2.31E-01
td 2.86E-07
Other prr 1.56E+04
Informatio pr at max. Ipi 2.45E+00
n deq at max. Ipi 1.15E-01
Ipa,α at max. MI 1.49E+02
B mode Focus Number : one focus; Focus Position: 2.0cm for
Control 1 B-mode
Sampling Volume Position: at the 2.0cm focus of B-mode;
Control 2 SV=0.5mm
Color Sampling Gate Position: at the 2.0cm focus of B-mode.
Control 3 Color Sampling Gate Width : Minimum Gate.
Operating
Control Control 4 Depth: 2.50cm
Conditions Working Frequency : 5.0MHz for B-mode, 4.6MHz for PW-mode, and
Control 5 4.6MHz for Color-mode, and color sensitivity: 3 .
Scan Line Mode: Minimum scan angle for B-mode, and lower-density
Control 6 line for B-mode and Color-mode.
PRF: 15600Hz for B-mode, 7100Hz for PW-mode, 7100 for
Control 7 Color-mode.
Control 8 Acoustic Power: Highest transmission power (100%).
87
Acoustic Output Reporting Tables
6LE7
Transducer Model: 6LE7
Imaging Mode: M
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt>
Aaprt≤1cm 2
2 scan
1cm
Maximum Index Value 1.77E+00 2.14E-01 8.96E-01 2.36E-01
pr.a 4.19E+00
P 8.04E+00 8.04E+00 8.04E+00
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.83E+00
Parameter z at max.Ipi,α 1.85E+00
deq(zb) 9.69E-02
fawf 5.46E+00 5.46E+00 5.46E+00 5.46E+00
Dim of X 1.25E-01 1.25E-01 1.25E-01
Aaprt Y 2.60E-01 2.60E-01 2.60E-01
td 3.11E-07
prr 2.00E+03
Other
pr at max. Ipi 5.47E+00
Information
deq at max. Ipi 9.61E-02
Ipa,α at max. MI 8.36E+02
88
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> 1 scan
cm2 cm2
Maximum Index Value 1.77E+00 1.30E-01 1.57E-01
pr.a 4.19E+00
P 2.33E+01 2.33E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter
z at max.Ipi,α 1.85E+00
deq(zb)
fawf 5.46E+00 5.46E+00 5.46E+00
Dim of Aaprt X 1.25E-01 1.25E-01
Y 2.60E-01 2.60E-01
td 3.11E-07
Other prr 2.90E+03
Informatio pr at max. Ipi 5.47E+00
n deq at max. Ipi
Ipa,α at max. MI 8.36E+02
Control 1 Focus Number: one focus; Focus Position: 2.0cm for B-mode
89
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.77E+00 4.52E-02 2.14E-01 8.96E-01 2.80E-01
pr.a 4.19E+00
P 8.11E+00 8.04E+00 8.04E+00 1.47E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.83E+00
Parameter z at max.Ipi,α 1.85E+00
deq(zb) 9.69E-02
fawf 5.46E+00 5.46E+00 5.46E+00 5.46E+00
Dim of Aaprt X 1.25E-01 1.25E-01 1.25E-01
Y 2.60E-01 2.60E-01 2.60E-01
td 3.11E-07
Other prr 2.00E+03
Informatio pr at max. Ipi 5.47E+00
n deq at max. Ipi 9.61E-02
Ipa,α at max. MI 8.36E+02
90
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI Non-
Scan Aaprt≤1 Aaprt> scan
cm2 1 cm2
Maximum Index Value 1.30E+00 2.42E-01 1.23E+00 4.01E-01
pr.a 2.83E+00
P 1.09E+01 1.09E+01 1.09E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.65E+00
Parameter z at max.Ipi,α 1.72E+00
deq(zb) 1.14E-01
fawf 4.68E+00 4.68E+00 4.68E+00 4.68E+00
Dim of X 1.57E-01 1.57E-01 1.57E-01
Aaprt Y 2.42E-01 2.42E-01 2.42E-01
td 8.16E-07
Other prr 1.80E+03
Informatio pr at max. Ipi 3.34E+00
n deq at max. Ipi 1.12E-01
Ipa,α at max. MI 4.14E+02
91
Acoustic Output Reporting Tables
TIS TIB
Non-scan TIC
Index Label MI
Scan Aaprt≤1 Aaprt> Non-scan
cm2 1 cm2
Maximum Index Value 9.77E-01 1.65E-01 9.29E-02 5.05E-01 3.43E-01
pr.a 2.37E+00
P 2.78E+01 4.21E+00 4.21E+00 3.21E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.72E+00
Parameter z at max.Ipi,α 2.07E+00
deq(zb) 1.14E-01
fawf 5.77E+00 5.77E+00 4.70E+00 4.70E+00
Dim of Aaprt X 1.08E-01 1.52E-01 1.52E-01
Y 1.98E-01 2.29E-01 2.29E-01
td 2.87E-07
Other prr 1.56E+04
Informatio pr at max. Ipi 3.20E+00
n deq at max. Ipi 1.12E-01
Ipa,α at max. MI 2.98E+02
Control 1 Focus Number: one focus, Focus Position: 2.0cm for B-mode
92
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤1 Aaprt>1
2 2 scan
cm cm
Maximum Index Value 1.05E+00 4.90E-01 6.71E-01
pr.a 2.49E+00
P 4.52E+01 4.52E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb
Parameter z at max.Ipi,α 1.38E+00
deq(zb)
fawf 5.67E+00
Dim of X
Aaprt Y
td 3.01E-07
Other prr 1.56E+04
Informatio pr at max. Ipi 2.95E+00
n deq at max. Ipi
Ipa,α at max. MI 2.91E+02
Control 1 B-Mode Focus Number: one focus; Focus Position: 1.2cm for B-mode
Color Sampling Gate Position: at the 1.2cm focus of B-mode;
Control 2
Color Sampling Gate Width: minimum.
Control 3 Depth: 2.50 cm
Operating
Control Control 4 Working Frequency: 5.0MHz for B-mode, and 4.6MHz for Color-mode
Conditions Scan Line Mode: minimum scan angle for B-mode, and lower-density
Control 5
line for B-mode and Color-mode, and color sensitivity : 3 .
93
Acoustic Output Reporting Tables
TIS TIB
Non-scan Non- TIC
Index Label MI
Scan Aaprt>
Aaprt≤1 cm 2 scan
1 cm2
Maximum Index Value 7.60E-01 1.58E-01 2.23E-01 9.67E-01 5.63E-01
pr.a 1.83E+00
P 1.80E+01 1.00E+01 1.00E+01 2.68E+01
Min of [Pα(zs), Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.73E+00
Parameter z at max.Ipi,α 2.13E+00
deq(zb) 1.19E-01
fawf 5.87E+00 4.70E+00 4.70E+00
Dim of Aaprt X 1.62E-01 1.62E-01
Y 2.31E-01 2.31E-01
td 2.86E-07
Other prr 1.56E+04
Informatio pr at max. Ipi 2.45E+00
n deq at max. Ipi 1.15E-01
Ipa,α at max. MI 1.49E+02
B mode Focus Number : one focus; Focus Position: 2.0cm for
Control 1 B-mode
Sampling Volume Position: at the 2.0cm focus of B-mode;
Control 2 SV=0.5mm
Color Sampling Gate Position: at the 2.0cm focus of B-mode.
Control 3 Color Sampling Gate Width : Minimum Gate.
Operating
Control Control 4 Depth: 2.50cm
Conditions Working Frequency : 5.0MHz for B-mode, 4.6MHz for PW-mode, and
Control 5 4.6MHz for Color-mode, and color sensitivity: 3 .
Scan Line Mode: Minimum scan angle for B-mode, and lower-density
Control 6 line for B-mode and Color-mode.
PRF: 15600Hz for B-mode, 7100Hz for PW-mode, 7100 for
Control 7 Color-mode.
Control 8 Acoustic Power: Highest transmission power (100%).
94
Acoustic Output Reporting Tables
D6-2
Transducer Model: D6-2
Imaging Mode: M-mode
TIS TIB
Non-scan
Index Label MI Aaprt Non- TIC
Scan Aaprt>
≤ scan
1cm2
1cm2
Maximum Index Value 1.29E+00 / / 9.76E-01 2.36E+00 2.11E+00
pr.a 2.28E+00 / / / / /
P / / / / 8.35E+01 8.35E+01
Min of [Pα(zs), Ita,α(zs)]
/ / / 7.00E+01 / /
zs / / / 1.34E+00 / /
Associate zbp / / / 2.38E+00 / /
Acoustic zb / / / / 3.73E+00 /
Parameter z at max.Ipi,α 4.00E+00 / / / / /
deq(zb) / / / / 3.68E-01 /
fawf 3.09E+00 / / 3.03E+00 3.09E+00 2.85E+00
Dim of X / / / 1.65E+00 1.05E+00 1.80E+00
Aaprt Y / / / 1.20E+00 1.20E+00 1.20E+00
td 5.69E-01 / / / / /
prr 2.00E+03 / / / / /
pr at max. Ipi 3.44E+00 / / / / /
deq at max. Ipi / / / / 3.59E-01 /
Other
Ipa,α at max. MI 3.04E+02 / / / / /
Information
Focal FLx
Length (cm) / / / 1.10E+01 / /
FLy
(cm) / / / 6.00E+00 / /
Focus position 8cm 8cm 10cm 15cm
95
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤ Aaprt> 1
scan
1 cm2 cm2
Maximum Index Value 1.29E+00 1.78E+00 / / / 2.55E+00
pr.a 2.28E+00 / / / / /
P / 1.15E+02 / / / 1.15E+02
Min of [Pα(zs), Ita,α(zs)]
/ / / / / /
zs / / / / / /
Associate zbp / / / / / /
Acoustic zb / / / / / /
Parameter
z at max.Ipi,α 4.00E+00 / / / / /
deq(zb) / / / / / /
fawf 3.10E+00 3.16E+00 / / / 2.96E+00
Dim of Aaprt X / 1.80E+00 / / / 1.80E+00
Y / 1.20E+00 / / / 1.20E+00
td 5.68E-01 / / / / /
prr 4.09E+03 / / / / /
pr at max. Ipi 3.44E+00 / / / / /
Other deq at max. Ipi / / / / / /
Informatio Ipa,α at max. MI 3.04E+02 / / / / /
n Focal FLx
Length (cm) / 1.40E+01 / / / 1.40E+01
FLy
(cm) / 6.00E+00 / / / 6.00E+00
Focus position 8cm 15cm / 15cm
96
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤ Aaprt> 1
scan
1 cm2 cm2
Maximum Index Value 1.27E+00 2.19E+00 2.02E+00 3.69E+00
pr.a 2.25E+00 / / / / /
P / / / / 6.95E+01 6.95E+01
Min of [Pα(zs), Ita,α(zs)] / / / 5.89E+01 / /
zs / / / 1.00E+00 / /
zbp / / / 2.38E+00 / /
Associate
Acoustic zb / / / / 3.66E+00 /
Parameter z at max.Ipi,α 3.36E+00 / / / / /
deq(zb) / / / / 3.77E-01 /
fawf 3.13E+00 / / 3.05E+00 2.94E+00 /
Dim of Aaprt X / / / 1.65E+00 1.05E+00 /
Y / / / 1.20E+00 1.20E+00 /
td 5.71E-01 / / / / /
prr 2.00E+03 / / / / /
pr at max. Ipi 3.19E+00 / / / / /
Other deq at max. Ipi / / / / 3.62E-01 /
Informatio Ipa,α at max. MI 2.55E+02 / / / / /
n Focal FLx
Length (cm) / 1.10E+01 / 1.10E+01 / /
FLy
(cm) / 6.00E+00 / 6.00E+00 / /
Focus position 6cm 8cm 10cm 15cm
97
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Aaprt Non- TIC
Scan Aaprt>1
≤ scan
cm2
1 cm2
Maximum Index Value 1.39E+00 / / 7.62E-01 2.02E+00 1.96E+00
pr.a 2.43E+00 / / / / /
P / / / / 6.31E+01 6.31E+01
Min of [Pα(zs), Ita,α(zs)]
/ / / 5.41E+01 / /
zs / / / 1.00E+00 / /
Associate zbp / / / 2.27E+00 / /
Acoustic zb / / / / 3.45E+00 /
Parameter z at max.Ipi,α 1.00E+00 / / / /
deq(zb) / / / / 3.44E-01 /
fawf 3.11E+00 / / 3.06E+00 3.09E+00 3.11E+00
Dim of X / / / 1.50E+00 1.13E+00 3.00E-01
Aaprt Y / / / 1.20E+00 1.20E+00 1.20E+00
td 1.47E+00 / / / / /
prr 6.99E+02 / / / / /
pr at max. Ipi 2.70E+00 / / / / /
Other deq at max. Ipi / / / / 3.31E-01 /
Informatio Ipa,α at max. MI 1.77E+02 / / / / /
n Focal FLx
Length (cm) / / / 1.70E+01 / /
FLy
(cm) / / / 6.00E+00 / /
Focus position 4cm 6cm 18cm 2.5cm
98
Acoustic Output Reporting Tables
99
Acoustic Output Reporting Tables
100
Acoustic Output Reporting Tables
101
Acoustic Output Reporting Tables
D6-2A
Transducer Model: D6-2A
Imaging Mode: M-mode
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤ Aaprt>
2 2
scan
1cm 1cm
Maximum Index Value 1.21E+00 5.75E-01 1.49E+00 9.09E-01
pr.a 2.34E+00
P 3.23E+01 3.23E+01 3.23E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate
zbp
Acoustic
zb 1.50E+00
Parameter
z at max.Ipi,α 1.50E+00
deq(zb) 3.42E-01
fawf 3.76E+00 3.76E+00 3.76E+00 3.76E+00
Dim of X 3.26E-01 3.26E-01 3.26E-01
Aaprt Y 9.49E-01 9.49E-01 9.49E-01
td 4.71E-07
prr 2.00E+03
Other
pr at max. Ipi 2.50E+00
Information
deq at max. Ipi 3.41E-01
Ipa,α at max. MI 3.05E+02
102
Acoustic Output Reporting Tables
103
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤ Aaprt>
2 2
scan
1 cm 1 cm
Maximum Index Value 1.21E+00 6.90E-01 1.49E+00 9.80E-01
pr.a 2.34E+00
P 1.92E+01 3.23E+01 3.23E+01 4.80E+01
Min of [Pα(zs),
Ita,α(zs)]
zs
Associate zbp
Acoustic zb 1.50E+00
Parameter z at max.Ipi,α 1.50E+00
deq(zb) 3.42E-01
fawf 3.76E+0
3.76E+00 0 3.76E+00 3.76E+00 -
Dim of X 3.26E-01 3.26E-01 3.26E-01 -
Aaprt Y 9.49E-01 9.49E-01 9.49E-01 -
td 4.71E-07
prr 2.00E+03
Other
pr at max. Ipi 2.50E+00
Information
deq at max. Ipi 3.41E-01
Ipa,α at max. MI 3.05E+02
104
Acoustic Output Reporting Tables
TIS TIB
Non-scan
Index Label MI Non- TIC
Scan Aaprt≤ 2
2 Aaprt>1 cm scan
1 cm
Maximum Index Value 1.01E+00 6.70E-01 3.31E+00 1.50E+00
pr.a 1.61E+00
P 7.29E+01 7.29E+01
Min of [Pα(zs),
Ita,α(zs)] 3.87E+01
zs 1.50E+00
Associate
zbp 1.82E+00
Acoustic
zb 3.15E+00
Parameter
z at max.Ipi,α 3.35E+00
deq(zb) 2.77E-01
fawf 2.55E+00 2.55E+00 2.55E+00 2.55E+00
Dim of X 4.00E-01 4.00E-01 4.00E-01
Aaprt Y 6.73E-01 6.73E-01 6.73E-01
td 1.59E-06
prr 1.70E+03
Other
pr at max. Ipi 2.02E+00
Information
deq at max. Ipi 2.67E-01
Ipa,α at max. MI 2.79E+02
SV 0.5mm
105
Acoustic Output Reporting Tables
106
P/N: 046-000866-00 (V2.0)