Professional Documents
Culture Documents
SERIES
Digital Thermometer
T1 T2 K
T1-T2
°F
°C
mV
TYPE KJTERSN
INTERVAL hour:min
USB MAX MIN AVG MEN
K
T1 T2
: °F
°C
CAL min:sec
TC TYPE C°/F°K
POWER
MAX HOLD
AVG ENTER
MIN SETUP
MS6511
THERMOMETER
CONTENTS CONTENTS
1. Safety Information................................1 5.4 Viewing MAX,
MIN and AVG Readings....................13
2. Product Overview..................................3
5.5 Use Offset value To Adjust
3. Recognize Meter...................................4 Temperature probe error...................14
3.1 Component...................................................4 5.6 User Self-Calibration..........................14
3.2 Display..........................................................5
6. Data Storage...................................16
3.3 Keys Description...........................................6
6.1 Data Storage ......................................16
4. Setting Meter.........................................9 6.2 Data Reading.....................................16
01 02
2. Product Overview 3. Meter Description
This digital thermometer uses a thermocouple with 3.1 Components
microprocessor as the temperature sensor. It has
the following features:
• Suitable for various thermocouples, such as K, J,
T, E, R, S, N(MS6511 and MS6512 are only
5
suitable for J, K, T, E) T1 T2
OFFSET SETUP READ REC HOLD
K
T1-T2
°F
• It can show results with °C, °F and K (Kelvin) °C
mV
TYPE KJTERSN USB MAX MIN AVG MEN
K
• Maximum, minimum and average value INTERVAL hour:min
CAL
:
min:sec
T1 T2
°F
°C 1 2 3 4
measurement
• Data hold
USB - T1 + - T2 +
• Thermocouple deviation compensation TC TYPE C°/F°K
POWER
03 04
3.2 Display 13 Auxiliary display
1 2 3 4 5 6 7 14 USB port
15 Time display
OFFSET SETUP READ REC HOLD
16 Indicate time display min: sec
T1 T2 K 8 17 Indicate calibrating state
T1-T2
°F
°C 18 Indicate auto-save time setting
20
9
mV
10
19 Indicate time display hour: min
TYPE KJTERSN
19 INTERVAL hour:min
USB MAX MIN AVG MEN
K 20 Indicate thermocouple type
T1 T2 11
18 : °F
°C 12
CAL min:sec
2 replaced.
Thermocouple measurements include a
3 HOLD HOLD
deviation value MAX MAX
ENTER ENTER
AVG AVG
4 Status setup indicator with flashing display
MIN MIN
SETUP SETUP
5 Indicator to display saved data
Indicator to save data automatically with
6 flashing display
MS6511 Key MS6512 Key
7 Data hold state
8 Main display unit
9 Main display
10 MAX, MIN, AVG
11 Indicator to save data
12 Auxiliary display unit
05 06
(MS6513 and MS6514 only)
To change setting options or add
POWER
TC TYPE C°/F°K
POWER
T1/T2/T1-T2 C°/F°K
functions, see specific operation
for details
HOLD HOLD
To change setting options or reduce
MAX MAX functions, see specific operation for
AVG
ENTER MEM
AVG
ENTER MEM
details
MIN MIN
SETUP SETUP Read saved data (only for MS6513
READ and MS6514)
PClink READ PClink READ
07 08
4. Setting Meter 4.2 SETUP Option Setting
4.1 SETUP Option 1. Data storage interval setting (INTERVAL):
Press the SETUP to enter the data storage
Press the SETUP key to enter setup mode. “ SETUP ” interval setting mode. “ SETUP ” will flash on the
symbol will flash on the screen, Press it again to screen. Use ▼ to change automatic data
switch setting states sequentally. Save all previous storage time interval. Hold ▼ to quickly
settings before exiting. If the meter is turned off in increase or decrease setting time. Maximum
the process of setting, setting changes will not be interval time is 59:59, minimum interval time
saved. Changed settings will take effect is 00:00 (00:00 enables automatic data storage
immediately after exiting setup mode. function; it only can be stored manually).
Loop sequence: data storage interval setting 2. Thermocouple type setting (TYPE)
(INTERVAL) - thermocouple type setting (TYPE) Enter thermocouple type setting mode. Use
- OFFSET (T1) setting - OFFSET (T2) setting - to set thermocouple types: K, J, T, E (for MS6513
automatic power-off time setting - system time and MS6514, R, S, N can be set, too. For MS6511
setting (S-T) - power frequency setting (LinE) - and MS6513, thermocouple types can be
normal temperature compensation (NTC) switch switched directly with “TC TYPE” key)
setting - auto-calibration switch setting (CAL) - 3. OFFSET (T1) setting
data cleaning (CLr) - save setting and return to Users can adjust the thermometer displayed
normal measurement state (if self-calibration is value to compensate for a certain kind of
set to ON, return to the calibration state, then turn thermocouple error. See “Adjust the temperature
the meter off and restart to return to normal sensor error with deviation value” section.
measurement state). Allowable adjustment range is ±6°C.
Note: only MS6512 and MS6514 have OFFSET After entering OFFSET (T1) setting state, the
(T2) option. offset value setting can be changed with .
Only MS6513 and MS6514 have data storage Note: When this offset value is no longer
interval setting (INTERVAL) and data cleaning required, please restore this offset value to 0.0.
(CLr) options. Changing the thermocouple type will
automatically restore offset value to 0.0.
09 10
4. OFFSET(T2) setting Press ENTER key to select time format, including
Users can adjust the thermometer show value to “hour:min” or “min:sec”. Hold ▼ to quickly
compensate for a certain kind of thermocouple increase or decrease. If this setting is not changed,
error. See “Adjust the temperature sensor error the system time is the current thermometer
with deviation value” section. Allowable running time.
adjustment range is ±6°C. 7. Power frequency setting (LinE)
After entering OFFSET (T2) setting state, the To get the best measurement results, please set
offset value setting can be changed with ▼ . the thermometer's power frequency to local usage
Note: When this offset value is no longer Enter the power frequency setting. “LinE” will
required any more, please restore this offset show on the screen. Select and set interference
value to 0.0. Changing the thermocouple type frequency to 50Hz or 60Hz with ▼ .
will automatically restore offset value to 0.0. 8. Normal temperature compensation (NTC)
5. Automatic power-off time setting (P) switch setting Enter the normal temperature
Enter the automatic power-off time setting mode. compensation (NTC) switch setting mode. “NTC”
“P-” will show on the screen. Set automatic will show on the screen. By default, it is in ON
power-off time (5 to 60 minutes) with ▼ . Hold state upon booting. ON/OFF can be set for
▼ to quickly increase or decrease. Set the normal temperature compensation with ▼ .
sleep time less than 5 minutes to show “OFF”, The thermometer will automatically restore to
turns off the automatic power-off function. ON state after reboot.
When automatic power-off function is active, 9. Auto-calibration switch setting (CAL)
“ ” will show on the screen; otherwise, it Enter the ON/OFF setting of auto-calibration
won't display. (CAL) mode. “CAL” will show on the screen. By
Automatic power-off time is counted from last default, it is in OFF state upon booting. ON/OFF
can be set for calibration function with ▼ . By
key operation.
setting this item to ON, users can enter the
When the thermometer is in automatica data
calibration mode. See “User Self-Calibration”
recording and PC communication state, it won't
for details.
enter automatic power-off state
Note: Use this function with caution!
6. System time setting (S-T) 10. Data clearing (CLr)
System time is the time from powering on. It will Enter the data clearing (CLr) mode. “CLr” will
automatically be cleared after power failure. show on the screen, then press “ENTER” to
Enter the system time setting. “S-T” will show clear all recorded data. At the same time, “CLr”
on the screen. The system time can be set with will flash on the screen. It will stop flashing
▼ . immediately after data is cleared.
11 12
5. Using The Meter 5.5 Use offset Value To Adjust
Temperature Probe Error
5.1 Connecting Thermocouple
1. Insert the thermocouple to input jack Use OFFSET (T1) and OFFSET (T2) in the SETUP
2. Press the power key to turn on the option to adjust the thermometer readings to
thermometer power compensate for a certain kind of thermocouple error.
3. Set the thermocouple type to be consistent 1. Put the thermocouple in a known and stable
with the inserted thermocouple type temperature environment (such as in ice bath or
dry well calibrator)
Note: If the thermocouple is not connected to 2. Stabilize the temperature reading
the selected input end or the thermocouple is 3. Under the SETUP option, adjust OFFSET value
“open”, or when it exceeds measuring range, until the temperature reading on the auxiliary
“OL” will show on the thermometer. display is consistent with the calibrated
5.2 Displaying Temperature temperature (see “SETUP Option Setting”).
1. Press °C/°F/K key to select appropriate 5.6 User Self-Calibration
temperature unit
2. Put the thermocouple to the testing position After users enter calibration mode, the meter can
3. Measurement results will display on the screen be calibrated. Calibration points and calibration
methods are shown here:
5.3 Data Hold 1. Calibration point:
1. Press HOLD to keep reading on the screen. a) Normal Temperature 25°C
“ HOLD ” will show on the screen b) T1 input channel, 0μV and 40.000mV
2. Press HOLD key again to close “HOLD” function c) T2 input channel, 0μV and 40.000mV (only for
and restore measurement state. MS6512 and MS6514)
2. Calibration methods:
5.4 Viewing MAX, MIN and AVG Readings a) Put the thermometer into a thermotank with
1. Press MAX/MIN/AVG key to view MAX, MIN the temperature of 25°C for 3 to 5 minutes.
and AVG readings. b) Power on the thermometer to stabilize the
2. Hold MAX/MIN/AVG key to exit MAX/MIN/AVG thermometer.
view mode c) Set self-calibration (CAL) to ON state in the
“SETUP” setting option, and exit setting state.
It will return to the calibration state.
d) Adjust the screen temperature with ▼ keys
until the temperature is consistent with the
thermotank.
13 14
e) Press “ENTER” key to save the current 6. Data Storage
calibration value.
f) Press “T1/T2/T1-T2” key (or “TC TYPE” key) to
6.1Data Storage
switch to T1 measurement channel. Data is stored with two methods: manual storage
and automatic storage
g) Input 0μV in T1 measurement channel 1. Manual storage is in the normal measurement
h) After stabilizing, press “ENTER” key to save the mode, and data storage interval (INTERVAL) is
current calibration value. set to “00:00”. Press the “MEM” key to store an
i) Input 40.000mV in T1 measurement channel item of data. The storage location is the smallest
j) After stabilizing, press “ENTER” key to save the number in the unused storage space. When
current calibration value. storage space is full of data, “FULL” character
k) Press “T1/T2/T1-T2” key to switch to T2 will display for two seconds.
measurement channel. 2. Automatic storage is in the normal measurement
l) Input 0μV in T2 measurement channel mode, and data storage interval (INTERVAL) is
not set to “00:00”. Press the “MEM” key to start
m) After stabilizing, press “ENTER” key to save storing data. The storage location is the smallest
the current calibration value. number in the unused storage space. At the
n) Input 40.000mV in T2 measurement channel same time, “ REC ” symbol will flash on the screen.
o) After stabilizing, press “ENTER” key to save the Press the “MEM” key again to stop storing data.
current calibration value. Each data storage time interval is the time set in
p) Power off and restart to complete calibration. the data storage interval setting (INTERVAL).
The minimum time interval which can be set is 1
Note: Make sure your standard source is second, the maximum time interval is 59 minutes
accurate, then make calibration. The user is and 59 seconds. When storage space is full of
responsible for using an accurate calibration data, “FULL” character will display for two
source. If the meter is inaccurate due using an seconds, and stop storing data.
inaccurate calibration source, the user bears 6.2 Data Reading
the liability.
In the normal measurement mode, press the “READ”
key to enter data reading mode. At the same time,
“ READ ” symbol will flash on the screen. Press ▼ to
read the next data forward or backward. Hold ▼ ▲
to quickly move forward or backward in the stored
data. At the same time, “ MEM ” and number will shown
on the auxiliary display. If there is no available
storage space, “NULL” character will display for
two seconds.
15 16
6.3 Clearing Stored Data 9.Technical Data
Enter the “Data Clearing (CLr)” item in the SETUP The indicator is accurate within the temperature
option. “CLr” will show on the screen. Phen press range from 18°C to 28°C, when relative humidity is
“ENTER” to clear all recorded data. At the same not more than 80%. Warranty period is one year
time, “CLr” will flash on the screen. It will stop (not including thermocouple error).
flashing immediately after clearing data.
Function MS6511 MS6512 MS6513 MS6514
7. Data Transmission
Connect the thermometer to PC with data line. Thermocouple
K, J, T, E K, J, T, E, R, S, N
Hold “PCLink” key. “ USB ” symbol will show on the type
screen. At this time, the USB port of the Measurement
thermometer is opened. See data transmission T1 T1/T2 T1 T1/T2
channel
protocol for details.
Note: After connecting the USB, the thermometer Measurement K: -200.0°C to +1372°C
will be powered by USB power supply and the range -328.0°F to +2501°F
thermometer's power key is invalid. After J: -210.0°C to +1200°C
disconnecting, the USB key will become valid. -346.0°F to +2192°F
T: -250.0°C to +400°C
8. Meter Maintenance -418.0°F to +752°F
8.1 Replace Battery E: -150.0°C to +1000°C
-238.0°F to +1832°F
When the battery indicator “ ” appears on the R: 0°C to +1767°C
thermometer, the battery should be replaced 32°F to +3212°F
immediately. The procedure is: S: 0°C to +1767°C
1. Turn off the thermometer power 32°F to +3212°F
2. Release the battery cover screw to remove the N: -200.0°C to +1300°C
battery cover -328.0°F to +2372°F
3. Replace with 9V battery
4. Put the battery cover back and tighten screw. Display 0.1°C/ °F / K<1000° (1°C/ °F / K for
resolution R-type and S-type)
8.2 Clean 1°C/ °F / K >1000°
When the thermometer surface is dirty and cleaning K,J,T,E:±(0.2%+0.5°C)
is required, use a soft cloth or sponge to wipe gently Precision T1/ ±[0.5% +0.5°C] R,S:±(0.2%+1°C)
with a little clear water, soap water or commercial T2/T1-T2 N:±(0.2%+0.5°C)
detergent. To avoid damage, don't immerse the
thermometer in water.
17 18
<-10°C: within +0.5°C; <-200°C:
within +1°C
T-type < -200°C for reference only
Time Relative time
Data record 0~999, 1000 groups in total
Temperature
ITS-90
scale
Applicable
NIST-175
standard
Recording interval setting*,
thermocouple type setting,
thermocouple offset setting, automatic
Setting power-off time setting, system time
functions setting, power frequency setting,
temperature compensation switch
setting, self-calibration switch setting,
data clearing*
Power supply 9V Battery
*MS6513, MS6514
19 HYS007044