You are on page 1of 3

Page 1 of 3

Online Parameterize
Online Parameterize / Group Select
Online Parameterize / Group Select / System Units
Parameter Value Lower Value Upper Value State
Flow Unit Cum/h ok
Vol Unit Cum ok
Size Unit mm ok

Online Parameterize / Group Select / Current Output


Parameter Value Lower Value Upper Value State
Range 1 196 Cum/h ok
Dual Range Off ok
PV Damp 1.00 s 0.01 s 100.00 s ok
Current Span 4-20 mA ok
Alarm Type Run ok
Analog Sim Off ok
AO1 4.00 mA ok

Online Parameterize / Group Select / Puls/Freq. Output


Parameter Value Lower Value Upper Value State
Freq/Pulse Pulse ok
Puls scaling 1 Cum ok
Puls width 2.00 s 0.05 s 2.00 s ok
Output passiv/pos ok
FO Alarm Type Low ok
Freq Sim Off ok
FO1 0 Hz ok

Online Parameterize / Group Select / Relays


Parameter Value Lower Value Upper Value State
Rel1 fct Limit 1 ok
Switch on R1 353.43 Cum/h ok
Switch off R1 353.43 Cum/h ok
Rel2 fct Limit 2 ok
Switch on R2 353.43 Cum/h ok
Switch off R2 353.43 Cum/h ok

Online Parameterize / Group Select / Batching


Parameter Value Lower Value Upper Value State
Batch Variable Off ok

Print to PDF without this message by purchasing novaPDF (http://www.novapdf.com/)


Administrator 2016-11-21
Page 2 of 3

Online Parameterize / Group Select / Display


Parameter Value Lower Value Upper Value State
Tot ovfl 0 ok
Display Damping 0s 0s 99 s ok
Display Format x.xxxx ok
LCD Contrast 100 % 0% 100 % ok
Language English ok

Online Parameterize / Group Select / Processing Para.


Parameter Value Lower Value Upper Value State
Lo flo cutoff 1.5 Cum/h ok
Noise sup Medium ok
EPD Off ok
Meas Mode unidirect ok
Flow dir forward ok
Ampl. Mode Mode 4 ok

Online Parameterize / Group Select / System Parameter


Parameter Value Lower Value Upper Value State
PZR Off ok
Customer Code 53 0 9999 ok
Code Entry 0 -1 9999 ok
Self check On ok
SW AMP V3.01.01 ok
SW COM V2.04.02 ok

Online Parameterize / Group Select / System Parameter / Status


Parameter Value Lower Value Upper Value State
Status group 1 ok
Status group 2 Sensor gain failure ok
Status group 3 ok
Status group 4 ok

Online Parameterize / Group Select / Sensor Data


Parameter Value Lower Value Upper Value State
Calf pos 1.1721 0.5000 2.0000 ok
Calf neg 1.1721 0.5000 2.0000 ok
Zero point 7 -1000 1000 ok
Line size 100 mm 2 mm 2000 mm ok
max SAPS 16.7 / s 1.0 / s 60.0 / s ok
act SAPS 16.7 / s 1.0 / s 16.7 / s ok

Print to PDF without this message by purchasing novaPDF (http://www.novapdf.com/)


Administrator 2016-11-21
Page 3 of 3

PV Snsr s/n 570643 ok


EPD electrode Yes ok

Online Parameterize / Device Data


Lower Upper
Parameter Value State
Value Value
Distributor Endress & Hauser ok
Tag @@@@@@@@ ok
Descriptor @@@@@@@@@@@@@@@@ ok
Message @@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@ ok
Date 1960-01-01 ok
Dev id 570643 ok
Write No ok
protect