1009 No data from Oxigraf sensor. Oxigraf Read out error log, evaluate with Product Support
restarted. representative. In case of recurrence, replace processor board
and Oxigraf adapter.
100A MOPS PCB test task started. Occurs No action.
only in production.
100B Software auto-restarted. Evaluate with Product Support representative