*HQHUDOLQWURGXFWLRQWRVHUYLFHPRGH 1. Press the three keys on the front panel all at once (see
Fig. 1/1).
languages.
Basic preconditions for calling service mode:
− Primus is on.
− Primus is in self-test or standby mode
The following sections, and the legends listed in them,
make reference to item numbers (numbers inside
circles). An item number may also follow a figure
(diagram) number. Example: Softkeys (Fig. 1/1).
Caution:
6WDUWLQJVHUYLFHPRGH
To access service mode:
'UlJHU0HGLFDO$* &R.*D$3ULPXV
5HSDLU,QVWUXFWLRQV *HQHUDOLQWURGXFWLRQWRVHUYLFHPRGH
2SHUDWLRQ
− Select the various screens, windows and functions
using the softkeys (Fig. 2/1).
− Use the turn knob (Fig. 2/2) to select and edit the
All rights reserved. Copyright reserved • R5132300SM1_Servicemode.fm 05.04.02 • Repair Instructions • Version 1.0 • Released
'UlJHU0HGLFDO$* &R.*D$3ULPXV
5HSDLU,QVWUXFWLRQV
/RJSDJH³LQIRORJ
/RJSDJH³LQIRORJ
The info log "ORJ SDJH" is the service mode start
screen. All other screens can be accessed from it (see
Fig. 4/2/3/4/5/6/7).
All rights reserved. Copyright reserved • R5132300SM1_Servicemode.fm 05.04.02 • Repair Instructions • Version 1.0 • Released
'UlJHU0HGLFDO$* &R.*D$3ULPXV
5HSDLU,QVWUXFWLRQV
/RJSDJH³LQIRORJ
'UlJHU0HGLFDO$* &R.*D$3ULPXV
5HSDLU,QVWUXFWLRQV V\VWHPLQIRV\VWHPLQIRUPDWLRQ
KRXUV".
4 Softkey activates window (6) "FRQVXPSWLRQ".
5 "RSHUDWLRQKRXUV" = Operating hours. Activated
with softkey (3). Here the operating hours are
totalled and programmed.
'UlJHU0HGLFDO$* &R.*D$3ULPXV
5HSDLU,QVWUXFWLRQV V\VWHPLQIRV\VWHPLQIRUPDWLRQ
Setting the gas consumption: First in the "set" "EDWWHU\YROWDJH" = Voltage of UPS batteries.
box set the gas consumption (0 to 99999) and
then select and confirm the relevant gas. "EDWWHULHFDSDFLW\" = Capacitance of UPS
batteries.
"O2": Totals the O2 gas consumption in mixer
mode in litres times 1000. "SRZHURIIUHPDLQWLPH" = Cut-off delay time.
The time is programmable.
"Air": Totals the AIR gas consumption in mixer
mode in litres times 1000. For more information on the power pack refer to
section 8 "power supply" (power pack).
"N2O": Totals the N2O gas consumption in
mixer mode in litres times 1000.
FRQILJXUDWLRQFRQILJXUDWLRQ
7 "VRIWZDUHYHUVLRQV" = Software versions of the
listed components. From the "system info" screen the "config." (Fig. 5/2)
softkey activates the following screen (see Fig. 6).
8 "FRPSOLDQFHSDWFLUFOH" = Compliance of the
compact breathing system. The value is
determined in the VGC self-test (see Fig.
10/2).
'UlJHU0HGLFDO$* &R.*D$3ULPXV
5HSDLU,QVWUXFWLRQV V\VWHPLQIRV\VWHPLQIRUPDWLRQ
,WHP .H\
5 Softkey activates window (11) "FRORUV".
6 Softkey activates window (12) "SDVVZ
VHTXHQFH".
All rights reserved. Copyright reserved • R5132300SM1_Servicemode.fm 05.04.02 • Repair Instructions • Version 1.0 • Released
'UlJHU0HGLFDO$* &R.*D$3ULPXV
5HSDLU,QVWUXFWLRQV V\VWHPLQIRV\VWHPLQIRUPDWLRQ
Launch the service software for the software "on": Password protection set (for setting see
download. window (12)).
Primus starts in download mode. "yes": The alarm tone can be reduced to a
minimum of 45 dB(A).
Following the software download see (8!!!).
"no": The alarm tone can be reduced to below
If "request" is displayed next to a component 45 dB(A).
(e.g. when a component has been replaced), a
download to update to a compatible software The setting chosen here affects the volume
version is required. setting on the MCP screen under "KDUGZDUH
WHVWVVHWWLQJ" (see Fig. 12/11) and the
configuration menu (customer) in standby.
'UlJHU0HGLFDO$* &R.*D$3ULPXV
5HSDLU,QVWUXFWLRQV V\VWHPLQIRV\VWHPLQIRUPDWLRQ
disabled with the Silence key. The password is changed using the turn knob.
Default setting "8056" (e.g. following a reset of
"warning only": In operation, only warnings can all configuration data (Fig. 4/8).
be disabled with the Silence key. 13 "LQIRORJUHVHW" = Reset of info log. Activated
with softkey (7).
"DFRQH" = A-cone
"ok": The info log is cleared.
"yes": The A-cone valve is fitted.
'UlJHU0HGLFDO$* &R.*D$3ULPXV
5HSDLU,QVWUXFWLRQV 3*03DWLHQW*DV0RGXOH,5,$
From the "log page" screen the "PGM" (Fig. 4/3) ,WHP .H\
softkey activates the following screen (see Fig. 7). 1 "" = IRIA serial number.
All rights reserved. Copyright reserved • R5132300SM1_Servicemode.fm 05.04.02 • Repair Instructions • Version 1.0 • Released
"ON": Breathing phase detection is always "occluded" : When this display appears the
active in service mode. sampling gas line is blocked.
"YES": The IRIA has detected no CO2 for "YES": The IRIA requests zeroing. In window
> 15 seconds. (7) calibrate "general calibration" the IRIA and
the O2 sensor.
"NO": The IRIA has detected CO2.
"NO": The IRIA does not request zeroing.
"YDOYH" = Operates PGM valve 1. If the status "invalid" is displayed, launch the
VISIA program from the service software. Read
"room air" : Room air and analyze the IRIA log (see (2)).
13 Remaining capacity of O2 sensor. The value
"patient" : Sample gas should be > 35%. Lower values indicate a used
O2 sensor.
"UGFXUYH" = Switches the parameters of the
third real-time curve (11). Choice between
primary anaesthetic and N2O.
5 V = 4.8 V to 5.2 V
)LJ Mixer screen
24 V = 22.5 to 25.5 V
"TFFO" = Tank flow frequency offset. The determined in the self-test (test step 4). Manual
frequency shown is proportional to the offset test step start in window (11). For tolerances
voltage of the sensor "PDMIX". The offset is see Test Certificate.
determined in the self-test (test step 55).
Manual test step start in window (11). For "PDMGSLO" = Differential pressure sensor for
tolerances see Test Certificate. fresh gas flow range from 0 L/min to 9 L/min.
Display of decimal actual. The offset is
"MFA" = Mixer flow calibration value. determined in the self-test (test step 5). Manual
Calibration factor of sintered-metal screen test step start in window (11). For tolerances
"RMGS". The calibration factor is determined in see Test Certificate.
the self-test (test step 3). Manual test step start
in window (11). For tolerances see Test
Certificate.
Test step 22 causes the fan to be shut off "PRP" = Measuring range 0 mbar to 3985 mbar
briefly. Manual test step start in window (11). (absolute). Separate "PTANK" pressure
11 "VHUYLFHWHVWV" = Test steps for DrägerService. measuring channel for DrägerService. This is
Here all the steps from the mixer self-test and used to calibrate the pressure regulators
additional service test steps can be activated "PRPxxx".
(more details see document: Mixer self-test
steps). "PSYS" = Measuring range 512 mbar to
1535 mbar (absolute).
The current test step number is shown next to
"results of test:". "TMGS" = Temperature sensors on the mixer
block.
To carry out a manual mixer calibration in
service mode, follow the test step sequence: 4,
5, 10, 2, 3 and 6.
Bit 1: Table for gas mixture O2/N20 upper limit "3X" = Pressure sensor (vacuum pump "VP")
infringed.
Calibration: Remove the compact breathing
Bit 2: Measured flow below zero. system. Select and confirm the pressure sensor
using the turn knob.
Bit 3: Measured flow reaches upper limit.
Only the pressure sensors "3H" and "3X" are
Bit 4: Flowmeter not functioning. calibrated in the self-test.
Bit 2: Measured pressure surpasses upper limit. "PRWRUFXUUHQW" = Motor current. Bias current
approx. 40 mA. The motor current rises when
Bit 3: Pressure gaugee not functioning. the piston is actuated with the softkeys (4/5).
11-2 "VWDWXV" = Pressure status. Not currently used.
Bit 4: Difference Pz Pe high -> Warning.
12 "VXSSO\YROWDJH" = Supply voltages generated
Bit 5: Difference Pz Pe too high -> Alarm. on the POWER PCB. Tolerances:
5 V = 4.8 V to 5.2 V
24 V = 22.5 to 29 V
15 V = 13.85 V to 16.15 V
indication is given of how much volume is errors are binary. A conversion is required.
required to build up a specific pressure. The Status:
pressure is measured with "PAWe".
Bit 0: Internal RAM error.
"APL limit" = Opening pressure of the "APL"
valve. A constant flow of approx. 3 L/min. is Bit 1: Internal register error.
generated with the mixer. The pressure in the
system rises until the "APL" valve opens. The Bit 2: Internal XRAM error.
pressure is measured with "PAWe".
Bit 3: ROM checksum error.
"auto PEEP" = Pressure build-up in system.
Test procedure: Valves "V1" and "V2" are open. Bit 4: External RAM error.
A constant flow is generated with the piston.
The pressure drop at "V1", "V2" and "RV2" is Bit 5: Internal watchdog error.
measured by "PAWe".
Bit 6: Dual port RAM error.
,WHP .H\
12 "7HVWLQJRI/('" = Test of LEDs in keys. If a
key with an integral LED is pressed, the LED
must light up.
All rights reserved. Copyright reserved • R5132300SM1_Servicemode.fm 05.04.02 • Repair Instructions • Version 1.0 • Released
capacitor (Goldcap).
0 = Power switch on 2 = Power switch off
4 Softkey "self test" activates a self-test of the
power pack (no visible response). 0 = Status: OK 4 = Status: Error
,WHP .H\
11-3 "detailed status 2"
0 = Do not replace 2 = Replace battery
battery
All rights reserved. Copyright reserved • R5132300SM1_Servicemode.fm 05.04.02 • Repair Instructions • Version 1.0 • Released