 0L[HU
This section describes the mixer for the AIR, N2O and
O2 gases. The newly generated fresh gas is fed
through the vaporizer to the VGC.
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

)LJ Mixer without hood, for legend see Table 1

7DEOH Legend to Fig. 2


,WHP &RPSRQHQWV
1 PRIMUS CPU PCB
)LJ Mixer position, with rear panel open 2 Gas inlet block
3 Mixer block

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
)XQFWLRQ'HVFULSWLRQ 0L[HU

,WHP &RPSRQHQWV ,WHP &RPSRQHQW


4 A-cone valve 2 CAN bus
5 Fresh gas tank 3 Mixer electronics (PRIMUS CPU PCB/MIXER
6 MIXER PCB PCB)
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

4 Fresh gas outlet


5 Fresh gas flow valve
6 Fresh gas tank
7 Gas inlet valves
8 Pressure regulator

The item numbers and abbreviations occurring in this


section relate to Fig. 3 .
On the MCP (1) the user selects the carrier gas AIR or
N2O as well as the fresh gas flow and the O2
concentration.
A CAN bus (2) transfers the setup parameters to the
mixer electronics (3). The mixer electronics generate
the actuation signals for the gas inlet valves (7).
)LJ Block diagram of fresh gas metering, legend At the inlet of the fresh gas tank (6) the flow (V) is
Table 2 measured. With the measured flow value the switching
times for the gas inlet valves (7) are calculated. The
7DEOH Legend to Fig. 3 selected gas concentration is set in the fresh gas tank.
The pressure in the tank (6) and at the fresh gas outlet
,WHP &RPSRQHQW
(4) is measured and monitored.
1 Monitor Control Panel (MCP)

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
)XQFWLRQ'HVFULSWLRQ 0L[HU

The fresh gas flow valve (5) delivers the selected fresh "24 V PLUG-IN". Only the DC/DC converters (-15 V,
gas flow. The flowmeter at the fresh gas outlet (4) +15 V) for the pressure sensor amplifiers are
regulates the fresh gas flow valve. supplied).
− The pressure sensors for the gases from the central
 2SHUDWLQJPRGHV supply (CS) system and the cylinder supply are read
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

and the compressed gas supply status is indicated


The operating mode is regardless of whether the by LEDs on the front panel (see 1.3.2 Pressure
Primus is powered from the mains via its power plug or status LEDs).
is powered up in battery mode.
When the Primus is switched on at the power switch
 9$PRGH (mains or battery powered) the following safety rule
applies:
Internal leakage may mean that the O2 concentration
in the Primus is above 21% when operation is begun. − Before other modules receive operating voltage, the
"10 VA" mode prevents dangerous operating states fan in the mixer is switched on for at least 10
from occurring. seconds. Only then is 'Normal“ mode activated.

When the power plug is connected to the mains power 


1RUPDO
PRGH
but the power switch is not yet switched on, the
In this mode the mixer CPU PCB controls normal
following safety rule applies:
mixing. All DC/DC converters (+5 V, +24 V, -15 V,
− A supply voltage of only 10 VA is fed into the interior +15 V) are supplied with "24 V SWITCH".
of the mixer (10 VA is the product of the no-load
voltage and short-circuit current of the supply voltage
 /D\RXW
 0,;(53&%
In the "Block diagrams" section you will find a block
diagram of the MIXER PCB.

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
)XQFWLRQ'HVFULSWLRQ 0L[HU

The MIXER PCB holds the following components: this is the fresh gas flow valve 'VMGS“. The
− DC/DC converters (+5 V, + 15 V, - 15 V, +24 V) with proportional valve cannot be monitored by this
voltage monitoring procedure.
− Amplifiers of the pressure and temperature sensors For safety shut-off of all valves, the supply voltage can
be cut to all valves.
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

− Valve power switch and monitoring circuits


− Fan actuation and monitoring  35,086&38 3&% PL[HU
− Logic device for the safety O2 adjuster In the "Block diagrams" section you will find a block
A cable harness connects the mixer with the NEUTRAL diagram of the PRIMUS CPU PCB.
POINT PCB and with the other unit components. By The PRIMUS CPU PCB is based on a 2-processor
way of the cable harness the mixer receives the system with Master and Supervisor. Both 16-bit
unstabilized 18 V to 30 V supply voltage and the 12 V processor systems are identical in structure. The
fan power from the power pack as well as the CAN bus. peripherals are controlled by way of an interface.
The mixer fan is switched on 10 seconds before the Each processor has the following components:
other components and is powered directly by the power
− Quartz
pack (see 1.1.1 10 VA mode).
− undervoltage detector
When the fan wheel rotates a proportional square
signal is generated (approx. 90 Hz). This square signal − External watchdog
is monitored by the PRIMUS CPU PCB. − Flash-PROM
The valves on the mixer block are actuated by the − RAM
Master CPU (PRIMUS CPU PCB). Power drivers on − EEPROM (system configuration)
the MIXER PCB operate the valves. When one of the
A Dual-Port RAM is placed between the two processor
valves is actuated, the current flows through a shunt.
systems for data exchange. A logic circuit prevents
The voltage drop at the shunt is evaluated by a
both processors writing to a memory cell
comparator circuit and monitored by the supervisor
simultaneously.
processor on the PRIMUS CPU PCB. An exception to

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
)XQFWLRQ'HVFULSWLRQ 0L[HU

Each processor system has an isolated serial port and


a CAN bus.
Configuration options for the serial port by jumper:
− Independent serial ports, RS232 level, isolation on
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

jack plug.
− Independent serial ports, TTL level, no isolation,
output to connector strip.
7-segment displays are provided for visual indication of
operating states of the Master and the Supervisor.

 *DVLQOHWEORFN $,52DQG12


)LJ CS/compressed gas cylinder ports
In the "Block diagrams" section you will find a detailed
layout of the pneumatic components of the Primus. Optional outlets for a bronchial suction device with AIR
On the underside of the gas inlet block are the ISO (Fig. 5/1) and for a O2 flowmeter (Fig. 5/2) are
ports for the CS gases (Fig. 4/1) e.g. Nist or DISS and prepared.
the compressed gas cylinder ports for N2O and O2
(Fig. 4/2).

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
)XQFWLRQ'HVFULSWLRQ 0L[HU
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

)LJ Optional outlets AIR (1) and O2 (2)

The following specifications apply to the CS and the


compressed gas cylinders:
− The pressure values of the CS gases must be
between 270 kPa and 690 kPa (19 psi to 80 psi).
− If the CS pressure falls below 270 kPa, the user
must open the O2 or N2O compressed gas cylinder,
as appropriate. If the cylinder pressure is too low, the )LJ Detail, gas inlet block, for legend see Table 3
user is warned on the MCP display. The maximum
cylinder pressure is 19000 kPa.

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
)XQFWLRQ'HVFULSWLRQ 0L[HU

7DEOH Legend to Fig. 6  3QHXPDWLFFRPSRQHQWVJDVLQOHWEORFN


,WHP &RPSRQHQW
1 Pressure regulators (N2O, AIR, O2) regulate the
input pressure to the mixer to between 2.3 and
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

0.1 bar
2 Sintered-metal filter
3 ISO port (e.g. NIST for O2 cylinder supply)
4 Optional gas outlet (plug-in connection for AIR)
5 Non-return valve
6 Non-return valve for N2O gas inlet
7 CS sensor N2O
8 CS sensor AIR
9 CS sensor O2

)LJ Pneumatic components diagram, gas inlet


block, for legend see Table 4

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
)XQFWLRQ'HVFULSWLRQ 0L[HU

7DEOH Legend to Fig. 7 $EE,WHP &RPSRQHQW


$EE,WHP &RPSRQHQW 8 To flush button and safety O2 adjuster
PPO2 Relative pressure sensors with integral 9 To the CS flow metering inlet valves
PPAIR amplifier for CS pressure measurement
The item numbers and abbreviations occurring in this
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

PPN2O section relate to Fig. 7 .


PCO2 High-pressure sensor cylinder pressure
For the following description the O2 gas path was
O2
chosen.
PCN2O High-pressure sensor cylinder pressure
N2O The CS gas (3) passes through the ISO port to the gas
inlet block.
PRPO2 Pressure regulator, O2 line
PRPAIR Pressure regulator, AIR line By way of the sintered-metal filter (4) the gas passes to
the pressure sensor "PPO2". The relative pressure
PRPN2O Pressure regulator, N2O line sensor delivers a signal proportional to the pressure.
PRCO2 Pressure regulator, O2 compressed gas
The sintered-metal filter (4) prevents the non-return
cylinder
valve (5) from being contaminated by particles.
PRCN2O Pressure regulator, N2O compressed
gas cylinder If the CS fails, no CS pressure is indicated. The non-
return valve (5) prevents O2 from escaping out of the
1 Gas inlet block (complete) compressed gas cylinder (2) into the CS system.
2 HP ISO port O2 (NIST or DISS)
The pressure regulator PRPO2 generates a constant
3 CS ISO port O2 (NIST or DISS) input pressure for the mixer (2.3 bar 0.1 bar). The O2
4 Sintered-metal filter in CS ISO port pressure regulator additionally supplies (8) the flush
5 Non-return valve, CS inlet O2 button "VO2+" and the safety O2 adjuster "VSFC".

6 ISO outlet O2 (optional) An optional compressed gas supply for an external


bronchial suction device (7) or an external O2
7 ISO outlet AIR (optional)
flowmeter (6) is provided for.

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
)XQFWLRQ'HVFULSWLRQ 0L[HU

 3UHVVXUHVWDWXV/('V The LEDs are lit either green or red or are unlit,
depending on operating status. The LEDs are made to
The mixer measures and monitors the pressure of the
change colour by reversing the polarity of the supply
CS and the compressed gas cylinders. Depending on
voltage.
the measured values, LEDs on the MCP indicate the
pressure status. If the mixer is in "10 VA" mode, the LEDs are either
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

green or unlit.
The initial condition for Table 5 is:
− Primus is off, but power is connected.

7DEOH LED status, Primus off


*DVVXSSO\ 3UHVVXUHYDOXH /('VWDWXV
O2-CS p < 2.7 bar Off
p ≥ 2.7 bar Green
AIR-CS p < 2.7 bar Off
p ≥ 2.7 bar Green
N2O-CS p < 2.7 bar Off
p ≥ 2.7 bar Green

)LJ Position of LEDs (front panel) O2 cylinder p < 20 bar Off


p ≥ 20 bar Green
The CS pressure and the cylinder pressure is indicated Pressure sensor not Off
by five LEDs on the front of the MCP. The sixth LED is connected
intended for an optional AIR compressed gas cylinder.

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 
)XQFWLRQ'HVFULSWLRQ 0L[HU

*DVVXSSO\ 3UHVVXUHYDOXH /('VWDWXV *DVVXSSO\ 3UHVVXUHYDOXH /('VWDWXV


AIR cylinder p < 20 bar Off N2O-CS p ≥ 2.7 bar Off
p ≥ 20 bar Green p ≥ 2.7 bar Green
Pressure sensor not Off Pressure sensor error Flashing
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

connected green
N2O p < 10 bar Off O2 cylinder p < 20 bar and CS O2 Off
cylinder p ≥ 10 bar Green pressure is ≥ 2.7 bar
Pressure sensor not Off p ≥ 20 bar Green
connected p < 20 bar and the CS Flashing red
O2 pressure is at 1.4 to
The initial condition for Table 6 is:
< 2.7 bar 2.8 Hz
− Primus is on. The mixer is in normal mode.
Pressure sensor not Off
7DEOH LED status, Primus in normal mode connected
AIR cylinder p < 10 bar and the CS Flashing red
*DVVXSSO\ 3UHVVXUHYDOXH /('VWDWXV
N2O pressure is at 1.4 to
O2-CS p < 2.7 bar Off < 2.7 bar 2.8 Hz
p ≥ 2.7 bar Green p ≥ 10 bar Green
Pressure sensor error Flashing Pressure sensor not Off
green connected
AIR-CS p < 2.7 bar Off p < 10 bar and CS Off
p ≥ 2.7 bar Green N2O pressure is
Error Flashing ≥ 2.7 bar

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 


)XQFWLRQ'HVFULSWLRQ 0L[HU

*DVVXSSO\ 3UHVVXUHYDOXH /('VWDWXV  0L[HUEORFN


N2O p < 10 bar and CS Off
cylinder N2O pressure is
≥ 2.7 bar
p ≥10 bar
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

Green
p < 10 bar and CS Flashing red
N2O pressure is at 1.4 to
< 2.7 bar 2.8 Hz
Pressure sensor not Off
connected

)LJ Mixer block, for legend see Table 7

7DEOH Legend to Fig. 9


,WHP &RPSRQHQW 6KRUWQDPH
1 Differential pressure sensors with PDMGSHI
2 piezo measuring bridge PDMGSLO

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 


)XQFWLRQ'HVFULSWLRQ 0L[HU

,WHP &RPSRQHQW 6KRUWQDPH  3QHXPDWLFV\VWHP


3 Absolute pressure sensors with PSYS
4 integral amplifier PTANK
5 Tank vent valve VTANK
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

6 Temperature sensors TANKTEMP,


MGSTEMP
7 Connection for tank volume TANK
8 Gas inlet valves VMIXO2
9 VMIXAIR
10 VMIXN2O
11 Pressure inlets (N2O, AIR, O2) to
the gas inlet valves
12 O2 inlet to safety O2 adjuster and
flush button
13 Differential pressure sensor with PDMIX
piezo measuring bridge
14 Fresh gas flow valve (proportional VMGS
valve)
)LJ Pneumatic components diagram, mixer, for
15 A-cone valve VSWAK legend see Table 8

7DEOH Legend to Fig. 10


,WHP &RPSRQHQW0HDQLQJ
1 Gas inlet block

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 


)XQFWLRQ'HVFULSWLRQ 0L[HU

,WHP &RPSRQHQW0HDQLQJ At a fresh gas flow setting of 12 L/min, the "VMIX"


valves are opened 48 times a minute. The following
2 A-cone valve (optional)
equation illustrates the TANK filling process (48 /min
3 Mixer, flow metering x 0.25 L = 12 L/min).
4 Vaporizer
 3'0,;DQG50
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

I To compact breathing system


With the differential pressure sensor "PDMIX" and the
II To A-cone valve/compact breathing system
flowmeter "RM" the gas flow and volume of the gas
switching valve (MV3) for airway pressure
flowing into the TANK are measured
measurement
(flow x time = volume). If a constant usable 'TANK'
III To A-cone outlet volume of 0.25 L is used then any given gas
concentration can be mixed.
In the "Block diagrams" section you will find a detailed
layout of the pneumatic components of the Primus. The differential pressure sensor "PDMIX" and the
absolute pressure sensor "PTANK" form a single
The abbreviations occurring in the following sections functional unit. The output signal of the pressure
relate to Fig. 10 and Fig. 9. sensor "PTANK" is used for the bridge supply voltage
 90,;YDOYHV for "PDMIX". The output voltage of "PDMIX" is
evaluated on the PRIMUS CPU PCB (mixer).
The "VMIX" valves mix the desired gas concentration.
By way of the "VMIX" valves the CS gas (N2O, AIR,  37$1. SUHVVXUHVHQVRU
O2) passes to the differential pressure sensor The pressure sensor "PTANK" has the following tasks:
"PDMIX", to the pressure sensor "PTANK" and to the
flowmeter section "RM". − Tank pressure monitoring
− Compensation for effect of ambient pressure
The "VMIX" valves fill "TANK". The pressure range in
fluctuations on flow metering with "PDMIX".
'TANK“ is between 1 bar and 1.5 bar. The pressure
difference of 0.5 bar and the "TANK" volume of 0.5 L − Monitoring of flowmeter "RM" by means of the
produce a usable "TANK" volume of 0.25 L (0.5 bar x proportional pressure rise in "TANK".
0.5 L = 0.25 L).

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 


)XQFWLRQ'HVFULSWLRQ 0L[HU

The pressure sensor "PTANK" has a resolution of  36<6 SUHVVXUHVHQVRU


2 mbar and a maximum measuring range of 2.558 bar
The pressure sensor "PSYS" compensates for the
absolute. The pressure sensor "PTANK" also has a
effect of ambient pressure fluctuations on the flow
second amplification branch to the AD converter of the
metering of "PDMGSHI" and "PDMGSLO".
PRIMUS CPU PCB (mixer). To be able to measure
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

higher pressures for service purposes, the resolution is The pressure sensor "PSYS" has a resolution of
4 mbar per digit. The maximum measured value is 1 mbar.
4.092 bar absolute.
 96:$. $FRQHYDOYH
 97$1.YDOYH
The A-cone valve actuated by the MIXER PCB has a
The tank vent valve "VTANK" opens when: bistable switching response.
− O2 and AIR are missing. The valve is switched with a time-limited voltage pulse.
The polarity of the voltage pulse determines the
 90*6 IUHVKJDVIORZYDOYH
switching direction.
The fresh gas flow valve is a proportional valve. It
The valve position is detected by a proximity sensor in
delivers a fresh gas flow of 0.2 L/min to 18 L/min. The
the valve and monitored by the PRIMUS CPU (mixer)
flowmeter at the fresh gas outlet regulates the valve.
PCB.
 3'0*6+,3'0*6/2 GLIIHUHQWLDOSUHVVXUH
 9%$. VDIHW\YDOYH
VHQVRUV
A mechanical safety valve for the A-cone valve.
The differential pressure sensors each have a press-
Opening pressure 80 mbar at 18 L/min +15%.
dependent resistance measuring bridge.
The differential pressure sensors measure the flow  7(037$1.7(03%/2&. WHPSHUDWXUHVHQVRUV
through the flowmeter section "RMGS". With the The volume of the gases "N2O, AIR, O2" is dependent
measured value the PRIMUS CPU (mixer) PCB on the temperature. Consequently, the measured
regulates the fresh gas flow valve "VMGS". values of the temperature sensors "NTC" are
incorporated into the mixing process. The temperature

'UlJHU0HGLFDO$* &R.*D$‡3ULPXV 


)XQFWLRQ'HVFULSWLRQ 0L[HU

sensors ("TEMPTANK", "TEMPBLOCK") used for


temperature compensation generate voltages
proportional to the temperature. The PRIMUS
CPU (mixer) PCB compares and monitors the sensor
voltages.
All rights reserved. Copyright reserved • F5132300T04_Mischer.fm 04.04.02 • Function Description • Version 1.0 • Released

 96)& VDIHW\2DGMXVWHU


The safety O2 adjuster is located outside the mixer
unit. With the manual safety O2 adjuster the user can
set an additional O2 flow of 3 to 12 L/min 30%.
When the overall system power-up test is running the
flow adjuster must be closed. The operating status is
registered by a microswitch in the safety O2 adjuster.
The microswitch is plugged into the NEUTRAL POINT
PCB. The signal is transmitted via the cable harness to
the mixer.

 92 IOXVKEXWWRQ


Mechanical flush button. Output min. 35 L/min.

 1H[WSDUWRIWKHIXQFWLRQGHVFULSWLRQ
The function description relating to the VGC follows.