Technical Information EN 01/2018
Software 9.1A, Rev 1.00.00/February 2018
11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
Copyright
This document is the property of B. Braun Avitum AG with all rights reserved.
Commissioning and Service
Only trained personnel must service the Dialog+, i.e. repair, maintenance, software
installation, firmware update, retrofitting and commissioning of the Dialog+ SW 9.xx.
Servicing must only be performed with proper tools, calibration equipment and be in
accordance with the most recent revision of this service manual/technical information,
which must be clearly and thoroughly understood.
Prevent Electrical Shock Hazard
Switch off the Dialog+ SW 9.xx and disconnect unit from mains if you have to open the
machine for servicing. Do not touch any exposed wiring or conductive surfaces while the
Dialog+ is opened. The voltages present when electrical power is connected to the Dialog+
SW 9.xx can cause serious injury or death.
ESD Information
Pay attention to ESD information, because electronic components are sensitive to
electrostatic discharges.
High Voltage in TFT Monitor
If a battery option is present in the machine:
! High voltage can be present at the backlight inverter board BIB in the TFT monitor, even if
the machine has been disconnected from mains. Pull out the battery compartment in the
base platform and switch off the battery voltage before opening the machine.
Protective Conductor in TFT Housing
If the TFT housing had to be opened during a service job, the tight seat of the protective
conductors in the TFT housing must be checked.
TSM Service Program
Only activate the TSM service program for service activities. It is prohibited to connect a
patient to the Dialog+ SW 9.xx and to run a therapy if the TSM service program is
activated in the Dialog+ SW 9.xx. If the TSM service program is activated the complete
alarm system is disabled. The TSM service program is started in the service mode: digital
board, service switch S1, position 2.
Software
The software is installed in the software mode: digital board, service switch S1, position 3.
Therapy Mode
After completion of all procedures switch back to the therapy mode: digital board, service
switch S1, position 0.
Calibration
Only perform a calibration after the Dialog+ SW 9.xx has reached working temperature
and the machine was disinfected and decalcified. Save the calibration data (CFC) before
you exit the TSM service program: TSM Main Menu, File Operations, Save Calibration Data.
Tubing
Tubing must be replaced only by the same tubing type/length and identical installation
manner. Make sure that the tubings in the machine are not kinked or twisted after
servicing (e.g. if sub-racks are pulled out and inserted again). The tubing must not touch
moving/rotating components (e.g. motors of gear pumps).
Wiring
Wiring must be replaced only by the same cable type/length and identical installation
manner. The cables must not touch moving/rotating components (e.g. motors of gear
pumps).
Prevent Chemical Burns and Scalding
During servicing on running machines: prevent chemical burns and scalding of the skin due
to the penetration of disinfectant or hot liquid.
Contaminated Machines
Protective gear should be worn in case of servicing of assumed contaminated machines.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 2/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
Cover in Rear Door
Servicing of mechanical assembly groups (components in contact with fluid): the cover for
the switch mode power supply microcontroller SMPS-MC in the rear door must be
assembled during servicing because it serves as a spray protection.
O-Rings
Always check o-rings from disassembled groups/components and replace if necessary.
Figures
The displayed figures can differ slightly from the machines on site, due to different
hardware statuses.
Fuses
If fuses are replaced they must exactly match the type and rating specified by the
manufacturer in the spare parts list/technical information. Where applicable: fuses must be
approved by UL/CSA.
Spare Parts
Only use original spare parts manufactured and sold by B. Braun Avitum AG.
System Configuration
The system configuration saved on a FSU USB stick must be downloaded to another
Dialog+ SW 9.xx machine only if:
• the hardware matches and
• the identical software version number is present.
Function Check
Check the respective function of the assembly group/component after servicing. A
complete function check must be performed after every service, according to the repair
matrix.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 3/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018

Table of Contents Page

General Information 6
Software 9.1A for Dialog+ SW 9.xx 6
Ordering Numbers 6
1. Hardware and Software Status 7
2. Software Modifications 8
2.1 Software Modifications from SW 9.12 to SW 9.13 8
2.2 Software Modifications from SW 9.13 to SW 9.16 8
2.3 Software Modifications from SW 9.16 to SW 9.17 9
2.4 Software Modifications from SW 9.17 to SW 9.18 9
2.5 Software Modifications from SW 9.18 to SW 9.19 10
2.6 Software Modifications from SW 9.19 to SW 9.1A 10
3. Hardware Modifications 11
3.1 Modification Matrix 11
3.2 Flow Diagrams (from SW 9.19) 12
3.2.1 Dialog+ with BIC Option and TSDE-S 12
3.2.2 Dialog+ with BIC Option, DF Filter Option and TSDE-S 13
3.2.3 Dialog+ HDF Online with TSDE-S (Molded Air Separator LA) 14
3.2.4 Dialog+ HDF Online with TSDE-S (Injection Molded Air Separator 15
LA)
4. Software Installation 16
4.1 After Installation of Software 16
4.2 Boot Machine/Check Version Number 16
4.3 Select Options 16
4.4 Enter Customer Specific Default Values 16
4.5 Enter Specific Parameters 17
5. Modified Calibration Procedures 18
5.1. Calibration of UF Pump 18
5.2. Calibration of Balance Chamber 19
5.2.1 Repair Matrix Balance Chamber 20
5.3 Calibration Procedure for TSBIC, TSD, TSD-S, TSDE-S and TSDE 21
5.3.1 Calibration Sensor Position for TSBIC, TSD, TSD-S, TSDE-S and 22
TSDE
5.3.1.1 Position of Measurement Sensor if TSDE-S/TSDE are Located at 22
Front
5.3.1.2 Position of Measurement Sensor if TSDE-S/TSDE are Located at 22
Rear
5.4 Calibration of Conductivity Sensors 23
6. Test Run with UF and Temperature Measurement 24
7. Structure TSM Service Program (from SW 9.16) 26
7.1 New/Modified Menus in TSM Service Program 27
TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 4/50
11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018

7.1.1 System Configurations 27


7.1.1.1 Dialysate Side Parameters (from SW 9.16) 27
7.1.1.2 ABPM Parameters (from SW 9.16) 27
7.1.1.3 Miscellaneous Parameters (from SW 9.16) 28
7.1.1.4 DBI Parameters (from SW 9.16) 28
7.1.2 TL Manual Test and Calibration 29
7.1.2.1 Alarm Signal (for Development only) (from SW 9.16) 29
7.1.3 Low Level Manual Test 29
7.1.3.1 Degassing and Heating (from SW 9.16) 29
7.1.3.2 Dialysis Temperature (from SW 9.16) 30
7.1.4 Low Level Manual Calibration 30
7.1.4.1 Degassing Temperature (from SW 9.16) 30
7.1.4.2 Dialysis Temperature (from SW 9.16) 31
7.1.4.3 Pressure Sensors (Water Side) (from SW 9.16) 31
7.1.4.4 Position Setting of BICP, KP and UFP (from SW 9.16) 33
7.1.4.5 Pressure Follower Valve (from SW 9.16) (for Development only) 33
7.1.5 Production Report 34
7.1.5.1 Low Level Options (from SW 9.16) 34
8. Service Overview in Therapy (from SW 9.16) 34
8.1 Degassing 35
8.2 Temperature 35
8.3 Conductivity 35
8.4 Ultrafiltration 35
8.4.1 PFV (n/a - do not select in Low Level Options) 35
9. bioLogic RR Comfort 36
9.1 Pop-up Window if bioLogic RR Comfort Guideline Curve could 36
not be Saved
10. Modification in Self Test 37
10.1 Temperature TSD and TSDE (from SW 9.16) 37
10.2 Test of Level Regulation and Equality Test for Blood Side 38
Pressure Sensors (from SW 9.19)
10.3 Pressure Test Substitution Line S-Online HDF Online (from 9.19) 41
10.4 Substitution Line Connection Test (from SW 9.16) 42
11. Default Table SW 9.01 to SW 9.1A 43

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 5/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
General Information
The technical information TI describes the software and hardware changes from machines
with software 9.12/9.13/9.17/9.18/9.19 to software 9.1A.
Upgrade from SW 9.12/9.13/9.17/9.18/9.19 to SW 9.1A is recommended for all Dialog SW
9.xx machines to achieve a uniform software status in the market.
A modification matrix describes the dependences between software and hardware.
A hardware modification is required before installing the software version 9.1A. It is
described in the assembly instruction AI 01/2018, attached to this technical information.
A particle filter FVE (filter degassing valve, 4.6 mm) was introduced at the position of the
previous temperature sensor TSHE.
Software 9.1A for Dialog+ SW 9.xx
The following component software versions are included in the software version 9.1A:
• Dialog+ TLC SW version 9.80 • Dialog+ LLC FW version 9.14
• Dialog+ LLC SW version 9.80 • Dialog+ LLS FW version 9.14
• Dialog+ LLS SW version 9.80
Language Versions
* 044 9.19
* 049 9.19
Ordering Numbers
Art. No. Description
77091A Software 9.1A for Dialog+ SW 9.xx (including TI 01/2018 and AI 01/2018)
34562274 Retrofit kit relocation temperature sensors (one retrofit kit per machine)

Each machine is marked with a special sticker (included in the retrofitting kit) to indicate
the upgrade. The sticker has to be adhered to the Dialog+ logo on the monitor, covering
the original white plus (+) character.
Dialog+ machines with the software version 9.1A and the hardware upgrade ex factory are
delivered with a modified front panel membrane.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 6/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
1. Hardware and Software Status
See table for hardware and software modifications depending on the installed software
version.
Dialog+ Service Hardware/Software Modifications
Software Documentation for Dialog+ SW 9.xx Machines
SW 9.03 TI 04/2010 Dialog+ SW 9.xx Machines
• We recommend updating to software 9.03 during the next service job.
Note
• All SW 9.02 features are included in the SW 9.03.
• The software does not require any application training.
• A hardware upgrade is not required.
• The firmware version of the LLS (LLS FW) is wrongly displayed as version 9.03. However,
the FW version 9.02 is actually present in the LLS.
SW 9.04 TI 01/2011 Dialog+ SW 9.xx HDF Online Machines
• We recommend updating Dialog+ SW 9.xx HDF online machines to software 9.04, if
repeated unwarranted permanent rinsing of the HDF filter occurs.
Note
• All SW 9.03 features are included in the SW 9.04.
• The software does not require any application training.
• A hardware upgrade is not required.
• The firmware version of the LLS (LLS FW) is wrongly displayed as version 9.04. However,
the FW version 9.02 is actually present in the LLS.
SW 9.12 TI 3.1/2011 Dialog+ SW 9.12
• We recommend updating Dialog+ SW 9.xx machines to software 9.12 during a service
job.
Note
• All SW 9.04 features are included in the SW 9.12.
• The software requires application training.
• A hardware upgrade is not required.
• An instruction for use 9.1x is required only if a Dialog+ SW 9.xx machine is updated
from a SW 9.0x to SW 9.12.
SW 9.13 TI 02/2014 Dialog+ SW 9.13
• We recommend updating Dialog+ SW 9.xx machines to software 9.13 during a service
job if the option bioLogic RR Comfort option is installed and used.
Note
• All features and functions of previous software versions are included in the SW 9.13.
• The software does not require an application training.
• An instruction for use 9.1x is required only if a Dialog+ SW 9.xx machine is updated
from a SW 9.03/9.04 to SW 9.13.
SW 9.16 TI 1.1/2016 Dialog+ SW 9.16
TI not published Note
The software version 9.16 was replaced by software version 9.17 before it was released
(see TI 04/2016).
SW 9.17 TI 04/2016 Dialog+ SW 9.17
TI not published Note
globally The software version 9.17 was released for limited markets only.
SW 9.18 TI 02/2017 Dialog+ SW 9.18
• Updating of Dialog+ SW 9.xx machines to software 9.18 is not permitted.
• The Digital Board Version 5 can only be used in conjunction with SW 9.18.
• The temperature sensor TSHE is changed into a new position as TSDE-S for an additional
monitoring by the LLS. The previous function of TSHE is covered by software in
conjunction with remaining TSE.
Note
• All features and functions of previous software versions are included in the SW 9.18.
SW 9.19 TI 07.1/2017 Dialog+ SW 9.19
TI not published The software version 9.19 was released for limited markets only.
globally Note
• A hardware upgrade is required as per assembly instruction.
• The software does not require an application training.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 7/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
SW 9.1A TI 01/2018 Dialog+ SW 9.1A
• We recommend updating all Dialog+ SW 9.xx machines in the market to software 9.1A
during a service job to achieve a uniform software status in the market.
Note
• A hardware upgrade is required as per assembly instruction.
• The software does not require an application training.

2. Software Modifications
2.1 Software Modifications from SW 9.12 to SW 9.13
Modifications Comments
1 Correction of the backup functionality on patient cards for increase of availability
bioLogic RR Comfort to be able to save more than 100 (relevant only for
therapies on the card. bioLogic RR Comfort)
2 Upline tank overflow due to unwanted open VZ in some TSM increase of availability
menus fixed.

2.2 Software Modifications from SW 9.13 to SW 9.16


Modifications Comments
1 Relocation of TSHE as TSDE-S to be used as additional optimization
supervisor sensor for dialysis fluid temperature monitoring
- Supervision of the TSDE by the LLS via TSDE-S
- Alarm text Temperature too high in case of a TSDE-S
alarm when TSDE-S reaches 41 °C in mainflow
- TSM modification for temperature sensors
2 Self Test Modification optimization
- Temperature TSD and TSDE (Preparation)
- Substitution Line Connection Test (Preparation)
3 Display of the supervisor value for the ENDLF ratio in TSM optimization
menu 2.10 and monitoring the ENDLF ratio
4 Display of the END ratio limit and BIC ratio limit in therapy optimization
technical overview
5 Optimized the flow control for the dialyzing fluid in optimization
Preparation and Treatment mode, using the balance
chamber calibration
6 Supporting the new version of Digital Board optimization
- SW 9.16 supports both, versions 4 and 5 of digital boards
- SW < 9.16 supports only version 4 of digital boards
7 Intermittent rinsing of the piston pumps during the optimization
standby phase to prevent some blockage
8 Warning message for Maximum out of action time optimization
exceeded shall be readable, also when the machine is set
to Automatic preparation start after disinfection.
9 The concentrate is displayed during descaling, instead of optimization
the default disinfectant
10 Nexadia: optimization
- Text messages can be longer than 40 characters
- Message text is cleared after sending
11 Clearing up of some Help texts optimization
12 The user shall be able to override the closed online port optimization
after red detection in Preparation to continue preparation
in case of a defect in the red detector.
13 Message for the user to check the connection and flow optimization
direction of the substitution line
14 Pressure equalization on blood lines of HDF online optimization
machines before the patient is connected
15 Reset of the ABPM in case of an error without turning the optimization
machine off/on
16 Default three point calibration of PDA in TSM. optimization
New calibration points -450 mmHg; 0 mmHg; +450 mmHg
17 Setting value of UFP pos. Pressure parking position is optimization
removed (no longer used).

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 8/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
18 Exit of disinfection is possible only after saving patient optimization
parameters to patient card
19 Compulsory to enter a patient name when using a new optimization
patient card
20 The self test cannot be skipped anymore optimization
21 bioLogic RR Comfort pop-up window to ensure saving optimization
guideline curves
22 bioLogic RR Comfort only possible if patient name or ID is *extended function
entered on patient card
23 The possibility to disable/enable sodium profiles is *extended function
selectable in TSM System Configuration.
24 ABPM module is able to measure during inflation *extended function
25 Flow direction check of level regulation pump. *extended function
During the first level regulation command the flow
direction of the level regulation pump is checked (PV/PA).
26 Adimea improvements: Prediction curve increase of availability
27 Changing of disinfectant is possible, when the machine is increase of availability
set to Confirm with Enter button.
28 Removed Arterial Bolus in disinfection increase of availability
29 Corrected, display of conductivity when exceeding 32 increase of availability
mS/cm
30 The selected concentrate type in mmol/l setting remains increase of availability
after power-fail
31 Corrected delivery rate for the heparin pump when increase of availability
selecting a B-D 20 ml syringe
32 Infinite rinsing in Preparation of HDF online machines due increase of availability
to dialysate side alarms fixed.
33 bioLogic RR Comfort patient data storage on patient card increase of availability
34 Corrected, never ending disinfection suck-in. increase of availability
35 Corrected, unwanted “incorrect heparin rate” warning. increase of availability
* Does not affect existing control mechanisms, operating principles or performance specifications.

2.3 Software Modifications from SW 9.16 to SW 9.17

Modifications Comments
1 The TSDE alarm monitoring was optimized. increase of availability
2 The temperature self test is repeated with the heater ON if increase of availability
the self test failed.
3 The alarm handling (LLC/LLS) for the temperature alarms increase of availability
was optimized.
4 Menu and selection for the pressure follower valve PFV was *extended function
introduced (for Development only).
* Does not affect existing control mechanisms, operating principles or performance specifications.

2.4 Software Modifications from SW 9.17 to SW 9.18

Modifications Comments
1 The temperature control was optimized. increase of availability

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 9/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
2.5 Software Modifications from SW 9.18 to SW 9.19

Modifications Comments
1 Purging of air from the DF filter every 150 seconds for 1 optimization
second, alarm if more than 2 VLA openings in 1 hour
2 Heparin pump: optimization
- control is optimized and earlier "Incorrect heparin rate"
warning for low heparin rates
- pump can only run in TSM with syringe correctly inserted
and syringe not empty
3 Disinfection valve VD is supervised during the disinfection increase of availability
cycle to prevent overflow of disinfectant due to open VD.
4 Supplement of the self test of substitution line optimized optimization
to detect also smaller leakage
5 Setting of date for daylight saving time (summer time) is optimization
corrected.
6 Clearing up of some Help texts optimization

2.6 Software Modifications from SW 9.19 to SW 9.1A

Modifications Comments
1 Alarm delay for TSDE-S modified increase of availability
2 Monitoring of degassing pressure PE during disinfection increase of availability
modified to reduce unwarranted alarms “Emptying of
upline tank not possible”
3 bioLogic RR possible also in combination with Nexadia increase of availability

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 10/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
3. Hardware Modifications
3.1 Modification Matrix

Art. No. Compatibility Hardware Modification


with ex Factory Retrofitting

Digital board DB version 4 3456074B any SW version 9.xx - -

Digital board DB version 5 34562141 only with SW version ≥9.17 - -

Power board valves PBV version 3 34562150 any SW version 9.xx from serial numbers -
218832
304349

Basic board BB version 2 34560726 any SW version 9.xx - -

Basic board BB version 3 34562168 any SW version 9.xx from serial numbers -
218832
304349

Injection molded components at 34562216 (injection injection molded parts not n/a
molded) compatible with milled Standard Machines
the UF sub-rack, i.e. air separator
34562215 (milled) version ≥ 230333
chamber and DDE holder
Standard Machines
≥ 503077
Rear door with mounting for VBE no spare part modified rear door n/a
(HDF online only) compatible with milled HDF Online
version, i.e. backwards ≥ 307941
compatible

Heater inlet temperature sensor - only with SW 9.17 and 9.18 Standard Machines Required
TSHE as dialyzer inlet temperature ≥ 232560*
sensor supervisor TSDE-S Standard Machines
≥ 504034*
HDF Online
≥ 308838*

TSDE and TSDE-S relocated (only - only with ≥9.1A Required


standard machines without DF SW 9.19 and ≥9.1A
filter and HDF Online machines)

Particle filter FVE in degassing 34561501 ≥9.1A ≥9.1A Required


circuit

*Some of these serial numbers of the transitional period might have software 9.12 with TSHE.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 11/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
3.2 Flow Diagrams (from SW 9.19)
3.2.1 Dialog+ with BIC Option and TSDE-S
TSHE relocated as TSDE-S next to TSDE, both directly behind DDE. New particle filter FVE in
the degassing circuit.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 12/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
3.2.2 Dialog+ with BIC Option, DF Filter Option and TSDE-S
TSHE relocated as TSDE-S next to TSDE, both between DF filter outlet and VDE. New
particle filter FVE in the degassing circuit.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 13/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
3.2.3 Dialog+ HDF Online (Milled Air Separator LA) with TSDE-S
TSHE relocated as TSDE-S next to TSDE, both directly behind DDE. VBE and HFB assembled
on the UF sub-rack. New particle filter FVE in the degassing circuit.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 14/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
3.2.4 Dialog+ HDF Online (Injection Molded Air Separator LA) with TSDE-S
TSHE relocated as TSDE-S next to TSDE, both directly behind DDE. VBE and HFB assembled
on the rear door. New particle filter FVE in the degassing circuit.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 15/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
4. Software Installation
The installation of the software is described in detail in the service manual, chapter 3,
Dialog+ software 9.xx.

4.1 After Installation of Software


Additional language installation for the graphical user interface GUI is required if any other
language than English or German is used as local language. The installation of additional
languages with the FSU USB stick is described in the service manual, chapter 3, Dialog+
software 9.xx.
Specific parameters must be selected/entered again in the TSM service program (System
Configuration) and saved with the OK icon after the software installation (e.g. for HDF
online machines: disinfectant volume and other defaults).
4.2 Boot Machine/Check Version Number
1. Switch on machine in TSM mode and watch if it boots correctly and shows the TSM
main menu.
2. Open menu Production Report, Version Number and check the version numbers.
• TLC SW 9.80
• LLC SW 9.80
• LLS SW 9.80
• LLC FW 9.14
• LLS FW 9.14
• Language Versions
* 044 9.19
* 049 9.19

4.3 Select Options


All available options in TSM service program (Production Report and Low Level Options)
must be selected again and saved with the OK icon after the software installation.

Low Level Options


1. Select and activate all present options in the Low Level Options with the OK icon.
Thus, the data is saved when quitting the menu.
Production Report
2. If necessary enter the serial number (see type plate SN) in the Production Report
menu.
3. Select and activate all present options in the Production Report with the OK icon.
Thus, the data is saved when quitting the menu.
4.4 Enter Customer Specific Default Values
The customer specific default values of the machine must be entered after the software
installation, unless the factory default settings are used.
4. Set system configuration (customer specific values) manually or use the FSU USB stick
with the system configuration, if available, and save the settings.
FSU USB-Stick for Treatment Support Settings
The system configuration saved on a USB stick (FSU utility) must only be downloaded to
another Dialog+ SW 9.xx machine after a software update if:

• the hardware matches and


• the machine has the identical software version number.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 16/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018

4.5 Enter Specific Parameters

Confirm that the ratio values are correct and equal for the respective controller and
supervisor, especially for the new value of ENDLF Ratio, Actual Status Supervisor.
For Machines with Pump Head for 7x10 Blood Pump Segment only

The type of pump head has to be selected as 7x10 and saved after a software update.
Check and confirm that the values for the 7x10 tube constant are correctly set, i.e. 9293
µl/head rotation for 7x10 tubing segment.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 17/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
5. Modified Calibration Procedures
5.1 Calibration of UF Pump

The UF pump has to be calibrated prior to the balance chamber calibration. The calibration
procedure for the UF pump has changed from SW 9.18.
1. Clamp and remove outlet tubing from UFP. Connect another tubing (e.g. suction
tubing for disinfectant) to UFP outlet and immerse tubing into an empty container.
2. Press UF Pump START Calibration button to prime the tubing.
Wait a few seconds to fully prime the tubing.
3. Press again UF Pump START Calibration button to stop UFP.
The UFP will rotate back into its initial position.
4. Remove filled tubing carefully from container, measure weight and note value.
5. Immerse tubing carefully into container.
6. Press UF Pump START Calibration button to start calibration procedure.
The UFP runs approx. two minutes and stops. It will rotate back into its initial
position.
7. After UF pump has stopped, remove tubing carefully from container, measure weight
and note value.
8. Press Ref. Value button.
9. Enter calculated delivery volume as reference value:
second measurement – first measurement = delivery volume.
10. Press Calibration button to confirm.
11. Confirm and store completed calibration procedure with OK icon.
Note: The calibration data is saved only after the bit sequence 0-1-0.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 18/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018

5.2 Calibration of Balance Chamber


The balance chamber has to be calibrated because the optimized flow controller requires a
calibrated balance chamber to ensure an accurate flow and smooth transition in chamber
cycles.

12. To avoid air in the dialyzing fluid system, rinse the machine before starting the
balance chamber calibration (use e.g. Low Level Manual Test, 1.11 Dialysis
Temperature).
13. Activate Calibration button and wait until automatic calibration procedure is
completed. This will take a few minutes whereby the system automatically moves
both membranes into their end positions to determine the respective frequencies
(displayed in µs) and chamber volumes at those points.
The values will remain at 0 µs and the LLC Status will appear as 00000800 0000000
(Calibration error balance chambers) if the calibration process fails at one point, e.g. if min.
or max. values are outside the permitted range.
MSBK1 Minimum/Maximum µs Values
MSBK2 Minimum/Maximum µs Values

Check that the min./max. values for MSBK1 and MSBK2 are between 11,000 µs and
35,000 µs respectively.
• The minimum values of MSBK1/2 must be >11,000 µs
• The maximum values of MSBK1/2 must be <35,000 µs
Additionally, the difference between the min. value and max. value of each chamber must
be >5,500 µs.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 19/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
5.2.1 Repair Matrix Balance Chamber

Tests and Calibration to be Executed Exchanged/Repaired/


(see Chapter 3, 4 and 5) Assembled Components

Balance Chamber Membrane Position Sensor MSBK1/2 Boards


Balance Chamber Membrane Position Sensor MSBK1/2

Conductivity Sensors BICLF, ENDLF and ENDLF-S


Temperature Sensors TSDE and TSDE-S
Balance Chamber Membrane
1.15 Flow Pump FPA, Balance Chamber Sensors and Air Separator u u u
1.27 DFS Pressure Test of Therapy u u u
2.5 Calibration: Dialysis Temperature TSDE/TSDE-S u u
2.6 Calibration: Conductivity BICLF, ENDLF and ENDLF-S u u
2.8 Calibration: Flow of UFP u u u
2.17 Calibration: Balance Chambers u u u
Self Tests u u u u u
Conductivity Test Run, Temperature Comparison Measurement u u
Test Run UF Comparison Measurement u u u u u
Perform Disinfection after Repair u u u u u

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 20/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
5.3 Calibration Procedure for TSBIC, TSD, TSD-S, TSDE-S and TSDE

A common calibration of all sensors for the dialysis temperature is recommended. The
supervisor sensors TSD-S and TSDE-S are always automatically calibrated together with the
respective controller sensors.
Refer to the procedure from the service manual, chapter 4 for the requirements and
settings, e.g. setting of pump speed and heater power.
The reference points remain the same, i.e. at 30 °C, 38 °C and 50 °C.
The loop-in positions of the temperature reference meter are changed.

Clamp

Clamp

The measurement sensor for the temperature calibration is connected between FPE and
TSDE-S, bypassing the balance chamber and DDE.
The filters on DF filter machines and HDF online machines are bypassed as well.
Clamp open tubings to avoid spillage.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 21/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
5.3.1 Calibration Sensor Position for TSBIC, TSD, TSD-S, TSDE and TSDE-S
5.3.1.1 Position of Measurement Sensor if TSDE/TSDE-S are Located at Front

Red

Measurement
Sensor

to
FBK1 Blue TSDE-S

TSDE
Clamp 1

to
TSDE-S
FBK1 Clamp 2
Fig.: Front view (the actual hardware might look differently, depending on the serial number)
5.3.1.2 Position of Measurement Sensor if TSDE/TSDE-S are Located at Rear
to
VDE FBK1

Clamp 1

Red
TSDE
Measurement
Sensor

TSDE-S FBK1

to Clamp 2 Blue
TSDE-S
Fig.: Rear view (the actual hardware might look differently, depending on the serial number)
Cover the open tubing ends with an absorbent cloth to prevent water spraying due to
pressure built-up in the balance chamber.
Secure all 4 tubing connections with tubing clamps to prevent accidental disconnection of
tubings due to high pressure and to prevent major damage on the electronics due to fluid
spillage.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 22/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
5.4 Calibration of Conductivity Sensors
The conductivity sensors always have to be calibrated after the temperature sensors were
calibrated.

A common calibration of all sensors for the dialysis fluid conductivity is recommended. The
supervisor sensor ENDLF-S is always automatically calibrated together with the respective
controller sensor.
Refer to the procedure from the service manual, chapter 4 for the requirements and
settings, e.g. setting of pump speed and heater power.
The reference points remain the same, i.e. at 0.0 mS/cm, 3.0 mS/cm, 5.8 mS/cm and 14
mS/cm.
The position of the reference meter is the same as for the temperature calibration, i.e.
between FPE and TSDE-S, bypassing the balance chamber and DDE.
The filters on DF filter machines and HDF online machines are bypassed as well.
Cover the open tubing ends with an absorbent cloth to prevent water spraying due to
pressure built-up in the balance chamber when opening the clamps.

After calibration:
Remove the reference meter and reconnect the tubing.
• connection filter FBK1 to balance chamber
• connection throttle DDE outlet (on standard and HDF online machines) or DF filter
outlet (on DF filter machines) to temperature sensor TSDE-S inlet.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 23/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
6. Test Run with UF and Temperature Measurement

WARNING
Wrongly installed temperature sensors on standard machine (without DF filter) and HDF
Online machines can lead to undetected high temperature fluid going to the dialyzer. Extra
test steps have to be performed during the test run to prevent this.
• Ensure that the temperature sensors TSDE and TSDE-S are correctly installed between
DDE and T-piece in machines without DF filter, i.e. in standard machines and HDF
Online machines.
• The temperature sensors shall not be installed between the T-piece and the valve VBP.
________________________________________________________________________

The test run should be performed under normal environmental condition, i.e. at a room
temperature of 25 °C ±5 °C.
1. Turn service switch S1 to position 0 - Therapy and close both doors.
2. Start machine (fully equipped) in dialysis mode.
3. Insert line system with dialyzer and prime and rinse with fluid – see instructions for
use, i.e. perform preparation (insert heparin syringe and select 2 ml/h and use BIC
cartridge if present).
4. Perform self test – according to instructions for use.
5. Connect a reference meter for temperature and conductivity in series with the
dialyzer (at blue coupling before dialyzer).
6. Set desired temperature to 37 °C and dialyzing fluid flow to 500 ml/min.
7. Set the UF volume to 125 ml and a therapy time to 00:20 h.
The machine must run without errors up to the message Connect Patient.
8. Press the Connect Patient icon to switch to therapy mode. This icon is enabled after
all self tests were performed successfully.
9. Insert the arterial and venous lines in a min. 250 ml container with water and make
sure that no air is drawn in.
10. Set the blood pump so that no blood side alarms are activated and circulate the blood
side.
11. Start therapy by switching from bypass to mainflow.
12. Stop blood pump after about one minute (the machine switches to bypass and UF
stops).
13. Remove blood lines from container.
14. Measure the weight of container filled with water and document as first value.
15. Document UF GROSS-S value from service overview as start value.
16. Insert arterial and venous lines into container with water again.
17. Start blood pump again and switch out of bypass.
18. After about 15 minutes in treatment check temperature sensor readings in Service
Overview screen and compare set temperature with reference meter. The set
temperature shall deviate no more than ± 0.5 °C from the temperature reading of the
reference meter and from TSDE/TSDE-S.
19. Check the conductivity for ENDLF by comparing with reference meter. The values shall
be within the tolerance of ±0.2 mS/cm.
20. Additional check for standard machines (without DF-filter) and HDF-Online machines
! only:
Check the values of the temperature sensors TSDE and TSDE-S in the Service
Overview in mainflow. Switch to bypass and compare the values of the temperature
sensors TSDE and TSDE-S again. The values shall not change more than 1.0 °C within
10 seconds, compared to the values in mainflow mode. A value change of more than
1.0 °C indicates that the temperature sensors might be assembled in the wrong
position and need to be reconfirmed.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 24/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
21. After a UF time of >00:15 h, stop blood pump. The machine switches to bypass and
UF stops.
22. Remove blood lines from the container and measure new weight of container filled
with water and document as second value.
23. Document UF GROSS-S value from Service Overview screen as end value.
24. Calculate weight from container first value minus weight from container second
value.
25. Calculate UF GROSS-S end value minus UF GROSS-S start value.
26. Compare calculation from container, with calculation from UF GROSS-S. The max.
deviation of measured UF volume (container) to displayed UF volume (UF GROSS-S) is
±15 ml.
27. Terminate therapy.
28. Remove disposables.
29. Perform a disinfection/decalcification of machine.
30. Switch off Dialog+.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 25/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
7. Structure TSM Service Program (from SW 9.16)

* from Software 9.16


{}: not applicable

Fig. : TSM service program

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 26/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
7.1 New/Modified Menus in TSM Service Program

7.1.1 System Configurations

7.1.1.1 Dialysate Side Parameters (from SW 9.16)

System Configuration, Dialysate Side Parameters


Sodium Profiles
The use of Sodium Profiles can be enabled/disabled in TSM.
7.1.1.2 ABPM Parameters (from SW 9.16)

System Configuration, ABPM Parameters, ABPM reading during cuff inflation


This feature is only possible for ABPM modules M2015 in combination with the new cuff
(with female connector at the machine).
The YES button is not highlighted and cannot be activated if the ABPM module M2010 is
installed.
Yes activated
ABPM measurement during inflation.
No activated
ABPM measurement during deflation.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 27/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
7.1.1.3 Miscellaneous Parameters (from SW 9.16)

System Configuration, Miscellaneous Parameters


Skip Self Test
The Skip Self Test function was removed.
7.1.1.4 DBI Parameters (from SW 9.16)

Data Set DBI 2.21 (from SW 9.16)


The data set DBI 2.21 is required if the BSL-WAN firmware is 2.21. The BSL firmware is
displayed in the service overview, configuration, BSL FW.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 28/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
7.1.2 TL Manual Test and Calibration

7.1.2.1 Alarm Signal (for Development only) (from SW 9.16)

Alarm signal
This new menu can be used by development to test the alarm and warning reactions in the
different therapy phases.
Phase
Selection of the Phase (e.g. Preparation, Therapy, Disinfection) which has to be tested.
Subsystem
Selection of Subsystem (e.g. TLC, LLC, LLS) which has to be tested.
Bit Number
Selection of Bit Number (alarm identification) which alarm shall be activated.
START
Test start.
7.1.3 Low Level Manual Test

7.1.3.1 Degassing and Heating (from SW 9.16)

Heater Inlet, Temperature Sensor TSHE


TSHE was removed from menu.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 29/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
7.1.3.2 Dialysis Temperature (from SW 9.16)

Dialyzer Inlet Temperature Sensor Supervisor TSDE-S


TSHE was changed to a new position as TSDE-S. The new temperature sensor TSDE-S is
shown in the LL Manual Test, 1.11 Dialysis Temperature menu.
7.1.4 Low Level Manual Calibration
7.1.4.1 Degassing Temperature (from SW 9.16)

Heater Inlet Temperature Sensor TSHE


TSHE was removed from the menu.
The sensor TSHE is no longer present and therefore the reference meter is only connected in
series to TSE sensor (behind TSE) for calibration.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 30/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
7.1.4.2 Dialysis Temperature (from SW 9.16)

Dialyzer Inlet Temperature Sensor Supervisor TSDE-S


TSHE was changed to a new position as TSDE-S. The new temperature sensor TSDE-S is
shown in the LL Manual Calibration, 2.5 Dialysis Temperature menu.
7.1.4.3 Pressure Sensors (Water Side) (from SW 9.16)

Calibration Dialysate Outlet Pressure PDA


A three point calibration is now the default for PDA. New calibration points:
• -450 mmHg/-600 mbar
• 0 mmHg/0 mbar
• +450 mmHg/+600 mbar
1. Activate the LL Manual Calibration menu with the LL Manual Calibration button in the
Manual Test and Calibration menu.
2. Activate the Pressure Sensors Water Side menu by pressing on the line Calibration 2.3
Pressure Sensors Water Side.
3. Disconnect tubing (to filter FVD) from rinsing bridge.
4. Connect manometer and a syringe to the left tubing connector outlet (red dialyzer
coupling) on rinsing bridge.
5. Clamp tubing between blood leak detector BLD and dialysate outlet pressure sensor
PDA.
6. Set measurement unit to mbar with the Pressure in mbar button if you want to
perform a calibration in mbar.
TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 31/50
11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
PDA Calibration Point -450 mmHg/–600 mbar

7. Press Dialysis Pressure PDA Calibration button.


8. Use a syringe to create a negative pressure of (reading on manometer display):
• -450 mmHg ±10 mmHg/-600 mbar ±13 mbar
9. Press Ref. Value button.
10. Enter the stable value from the manometer as reference value.
Wait until digits are stable.
Confirm with the Calibration button.
PDA Calibration Point 0 mmHg/0 mbar
11. Open syringe to atmosphere for 0 mmHg/0 mbar calibration.
12. Press Ref. Value button.
13. Keep the default reference value as 0 mmHg/0 mbar
Wait until digits are stable.
14. Confirm with the Calibration button.
PDA Calibration Point +450 mmHg/+600 mbar
15. Use the syringe to create a positive pressure of (reading on manometer display):
• +450 mmHg ±10 mmHg/+600 mbar ±13 mbar
16. Press Ref. Value button.
17. Enter the stable value from the manometer as reference value.
Wait until digits are stable.
18. Confirm with Calibration button.
19. Confirm and store complete calibration procedure with OK icon.
Note: The calibration data is saved only after the bit sequence 0-1-0 (for detailed
information see service manual 4.7.3.1).
20. Remove calibration devices.
21. Reconnect tubings.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 32/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
7.1.4.4 Position Setting of BICP, KP and UFP (from SW 9.16)

UFP Positive Pressure Test


The parameter UFP Positive Pressure Test was removed from the menu because the software
is no longer using this function.
7.1.4.5 Pressure Follower Valve (from SW 9.16) (for Development only)

PFV not available.


Do not change any value.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 33/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
7.1.5 Production Report

7.1.5.1 Low Level Options (from SW 9.16)

PFV not available, do not select.

8. Service Overview in Therapy (from SW 9.16)

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 34/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
8.1 Degassing
Temperature Sensor

The temperature sensor TSHE was removed.


8.2 Temperature
Temperature Sensor

The temperature sensor TSDE-S was added.


8.3 Conductivity
Displaying Ratio Limit

The limits of BIC and END ratio are displayed.


8.4 Ultrafiltration

8.4.1 PFV (n/a - do not select in Low Level Options)


PFVVBP (Pressure Follower Valve, Valve Bypass)

PFVVBP (PFVVBP-S) is displayed.


Note: For internal development only. PFV must not be selected in Low Level Options, the
PFVVBP valve is displayed as N/A.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 35/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
9. bioLogic RR Comfort

9.1 Pop-up Window if bioLogic RR Comfort Guideline Curve could not be Saved

To ensure that the bioLogic RR Comfort guideline curve was recorded correctly (without any
issues), a new pop-up window was created:
bioLogic RR guideline curve of the current therapy is not saved
The possible reasons of a saving issue are display as well.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 36/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
10. Modifications in Self Test
Modifications in self test from SW 9.03/9.04 to SW 9.12 are described in the technical
information TI 3.1/11.
10.1 Temperature TSD and TSDE (from SW 9.16)
Objective Up to SW 9.12/9.13: The function of the TSD/TSD-S sensors is tested.
From SW 9.16: The function of the TSBIC, as well as TSD/TSD-S and TSDE-S/TSDE sensors is
tested.
Test Code 22
Repeats Until Error Message 1
ID Error Message 1160: Temperature test not OK
Actions after Error Rinse, repeat all tests.
LLC Test LLC Test Steps LLS Test Steps LLS Test
Code Code
From SW 9.16 From SW 9.16
1. Prerequisite for Test: Prerequisite for Test:
LLC heats up dialysis fluid to >41.5 °C at both sensors Temperature readings at TSD-S and TSDE-S have
TSD and TSDE and continuously compares TSBIC sensor exceeded 41 °C at least once during Preparation.
with TSD sensor.
2. 2202x Prerequisites for Test:
From SW 9.16
Test temperature (TSD and TSDE > 41 °C) has to be
reached and TSBIC sensor has to be equal with TSD
sensor (tolerance ±1 °C) at 2 different check points
(more than 2 °C apart).
If test conditions are not reached:
- Set warning 1102
- Wait for max. 3 minutes to reach test conditions
Test conditions reached or timeout elapsed:
- Delete warning 1102
- Perform test
3. Test code = 22 to LLS Test code = 22 back to LLC
4. 2210x From SW 9.16 From SW 9.16 22000 –
If test temperature (TSD >41 °C and TSDE > 41 °C) was Condition for test initialization on HDF Online machines: 22004
not reached: Remaining pre-rinsing volume shall be 0 ml
Test not passed, otherwise: timeout - set timer
5. From SW 9.16 22005
Both TSD-S and TSDE-S shall exceeded 41 °C after
start of Preparation and must be below 50 °C at this
time.
Note: Heater power is reduced by LLC if both sensors
TSD-S and TSDE-S are higher than 41.5°C.
6. Acknowledge with switch-off of VGD 22006 –
22011
7. 2220x Wait for D24OFF (timeout 10 s)
8. VGD ON 22012
9. 2230x Wait for D24ON (timeout 10 s)
10. 2299x Completed

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 37/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
10.2 Test of Level Regulation and Equality Test for Blood Side Pressure Sensors
(from SW 9.19)
up to SW 9.04 The pressure sensors are connected together via the filled tubing system with container
and thus have the same pressure level. Additionally the feedback of the level regulation
valves, the level regulation pump LRP (PPR) and the SAKA are checked.
from SW 9.12 The pressure sensors are connected together via the filled tubing system with bag and thus
have the same pressure level. PV and PA are always checked. PBE is only tested if the
optional level regulation is installed (optional possible from SW 9.12). PBS is tested if the
machine is equipped with double pumps.
Additionally the feedback of the level regulation valves, the level regulation pump LRP
(PPR) and the SAKA are checked. VBT is always tested. VPV, VPE, VPD and LRP (PPR) are
only tested if the optional level regulation is installed. VPA and SAKA are tested if the
machine is equipped with double pumps (i.e. the level regulation is present).
from SW 9.19 The former Equality Test for Blood Side Pressure Sensors was renamed from software
version 9.19 and is now called: Test of Level Regulation and Equality Test for Blood Side
Pressure Sensors. The flow direction of the level regulation pump was integrated (LLC test
codes: 3275x, 3276x and 3277x; and LLS test codes: 32011 and 32012).
Objective Test the equality of pressure sensors with opened tubing system.
Test code 32
Repeats until error message 3
ID error message 1163: Blood side pressure sensor test not OK
Actions after error
Return to window: Repeat blood line self test with EQ key.

LLC Test LLC Test Steps LLS Test Steps LLS Test
Code Code
1. 3202x Prerequisites for Test:
a) Blood side must not be stopped by TLC or LLC.
If stopped: Message 1140 is displayed.
No timeout.
b) If BS is not stopped HDFonline task must enable
the start of the tests. A delay occurs only with
autopriming (set pressure).
Timeout 2 min
2. Test code = 32 to LLS Test code = 32 back to LLC
3. 3210x LLC opens SAKV, VBT, VPA, VPV, VPU and closes SAKA Test initialization 32000
VPU must be opened to prevent a pressure
equalization to the environment and that the levels
in the chambers do not change.
VPE remains closed to prevent a pressure
equalization if PBE is not connected.
4. 3220x Check of current detection for VBT, VPA, VPV, VPE, 32001
VPD, SAKA, SAKV and LRP
Test timeout 60 s and wait for inversion of
LLS_SELFTESTBIT from LLS Acknowledge with inversion of LLS_SELFTESTBIT
5. 3230x LLC closes VBT, VPA, VPV, SAKA
and opens VPE, SAKV, SAKA
6. 3240x Check of current detection for VBT, VPA, VPV, VPE, 32002
VPD, SAKA, SAKV and LRP
Test timeout 60 s and wait for inversion of
LLS_SELFTESTBIT from LLS Acknowledge with inversion of LLS_SELFTESTBIT
7. 3250x LLC closes VPE and VPU
VPD is opened
LRP is activated with the PWM PPRLOW (PWM for
line with small chamber, setting in TSM)
From SW 9.12
If AV-Line without PA Connector is selected, LLC
opens valve VBT to check PA and PV without PA
connection.
8. Initialize mean value storage for PA, PV, PBS and PBE Check of current detection for VBT, VPA, VPV, VPE, 32003
VPD, SAKA, SAKV and LRP

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 38/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
9. 3260x Valve status: Save values from PA, PV, PBS, PBE and compare the 32004
SAKA, SAKV --------------------------------opened pressure sensor PA, PV, PBE and PBS (if present):
VPE, VPV, VPA, VPD, VPU -------------------- closed

VBT is opened if PA connection is not necessary,


otherwise closed

Wait for inversion of LLS_SELFTESTBIT from LLS


(timeout 15 s)
Evaluate every 250 ms continuous mean value from
the 4 values of PA, PV, PBS and PBE.
10. Up to SW 9.04 32005
PBE is always tested
|PV – PBE| must be ≤ 20 mmHg
or |PBE| must be < 10 mmHg.
From SW 9.12
PBE is only tested, if present, i.e. if the optional Level
Regulation is installed.
|PV – PBE| must be ≤ 20 mmHg
or |PBE| must be < 10 mmHg
11. Up to SW 9.04 32006
If PBS is present:
|PV – PBS| must be ≤ 20 mmHg
or |PBS| < 10 mmHg.
From SW 9.12
If PBS is present:
PV – PBS must be ≤ 20 mmHg and PBS – PV must be
≤70 mmHg
or |PBS| < 10 mmHg.
12. Acknowledge with inversion of LLS_SELFTESTBIT 32010
Save PV for later use.
13. 3270x Detected inversion of LLS_SELFTESTBIT
Calculate mean value storage of PA, PV, PBS and PBE
SAKV is closed in advance for pressure test with LRP.
Save PV for later use.
14. 3275x From SW 9.19 Check current detection for VBT, VPA, VPV, VPE, VPD, 32011
Valve status: SAKA, SAKV and LRP
VPE, VPA, SAKV, VBT ------------------------ closed
Check that pressure is increased on PV: PV is 30 mmHg
VPU, VPD, VPV, SAKA -----------------------opened
higher than in LLS 32010
Level regulation pump runs with PWM PPRLOW 32012
Acknowledge with inversion of LLS_SELFTESTBIT
(PWM for small chamber TSM)
Pumping air into system to test flow direction of LRP
(PV will increase as air goes through VPU, VPD, VPV)
VPU is closed if PV is higher than saved PV + 100
mmHg (LRP recirculates). VPU is reopened if PV is
lower than saved PV + 50 mmHg.
Note: if test fails after pressure setup (3275x) it will
jump to 3276x to remove pressure but later the fail
code will be restored.
Test timeout 15 s, wait for inversion of
LLS_SELFTESTBIT from LLS
15. 3276x From SW 9.19
Remove pressure at PV by opening PV to air through
VPV, VPU and VPD
Valve status:
SAKV, VPE, VPA, VPD, VPU, VBT -------------- closed
SAKA, VPV ---------------------------------opened
Stop LRP
Condition:
If PV in SavedPVbeforeLRPtest (3270x) and
-15 mmHg then:
- close VPV
- if 1 s elapsed in this pressure interval, then go to
next state
otherwise
- open VPV
- reset timer
Timeout 15 s since 3275x elapsed
TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 39/50
11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
16. 3277x From SW 9.19
Valve status:
VPE, VPV, VPA, VPD, VPU, VBT---------------- closed
SAKA, SAKV --------------------------------opened
17. 3280x (Error) received B24ON - mean value of PA and PV
deviate more than ±40 mmHg.
18. 3285x (Error) the test prerequisites from HDF online-Task
are not given during one of the test steps (e.g.
autopriming, HDF filter rinsing required after LF
error).
19. 3299x Completed
Valve status:
VPE, VPV, VPA, VPD, VPU, VBT---------------- closed
SAKA, SAKV --------------------------------opened

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 40/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
10.3 Pressure Test Substitution Line S-Online HDF Online (from 9.19)
Objective If the substitution line s-online is connected at the beginning of the blood side self test
and the filling of the tubing system is performed by the machine (autopriming): LLC
performs a pressure test of the port connectors and the lines between the port and OSP.
This tubing is connected with the pressure sensor PA via the arterial tubing adapter. The
test is performed with positive pressure. The build-up is performed in the DFS system with
FPE/FPA via a pressure drop at the throttle DDE and an opened VSAE. The pressure test is
performed for s-online after the equality test of the pressure sensors in the blood side test
block.
If autopriming is not performed s-online can not be tested. As an information LLS
generates an alarm Leakage Check S-Online after the initial start of OSP in therapy. This
information is generated even if the autopriming of the s-online test was not performed by
LLC.
If the s-online is not connected during the pressure test this is not detected and a message
to check the line is not generated after connection and the start of online.
From SW 9.19
The function was optimized to recognize small leakages in the substitution line. Note: The
test will be performed in preparation only, if HDF or HF therapy is selected, the online port
is open and port-to-port priming is used.
Test code 40
Repeats until error message 1
ID error message 1166: Self test substitution line failed
Actions after error
Return to window: Repeat blood line self test with EQ key.
LLC Test LLC Test Steps LLS Test Steps LLS Test
Code Code
1. 4002x Prerequisites for Test:
Blood side must not be stopped by TLC or LLC.
If stopped: Message 1140 is displayed.
No timeout.
2. Test code = 40 to LLS Test code = 40 back to LLC
3. 4010x Opened VSAE Test initialization 40000
4. 4020x Pressure build-up with FPE/FPA in chamber flow
through mode via DDE until PA > 270 mmHg
(no pressure control for pressure sensor PA!)
Valve Status:
VDFF, VDE, VDA, VSB, VSAA ------------------ closed
VBP, VSAE --------------------------------- opened
Timeout 20 s
5. S-online Pressure Test Start condition for the test (for 5 s): 40001 –
Valve Status: 1. Correct valve position (VSAE closed) 40010
VSAE closed, BPA and OSP stopped 2. PA > 250 mmHg
3- PA < 500 mmHg
6. Test 10 s, 40011
Averaging MeanPA,
Determine MinPA and MaxPA
Check –500 mmHg < PA < 500 mmHg,
max. deviation ∆PA < 50 mmHg,
-450 mmHg < PA < 450 mmHg,
7. Check: 40012
- MaxPA - MeanPA < 50 mmHg,
- MeanPA - MinPA < 50 mmHg
From SW 9.19
- MaxPA - MeanPA < 15 mmHg,
- MeanPA - MinPA < 15 mmHg
40013 –
Acknowledge with switch-off VGB 40018
8. 4020x Wait for B24OFF (timeout 30 s)
9. VGB ON 40019
10. 4030x Wait for B24ON (timeout 15 s)
11. 4070x (Error) the test prerequisites from HDFonline task are
not given during one of the test steps (e.g.
autopriming, HDF filter rinsing required after LF error).
12. 4099x Completed

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 41/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
10.4 Substitution Line Connection Test (from SW 9.16)
Objective This function has been integrated to recognize a closed tubing between substitution pump
and extracorporeal circuit.
Note: The test will be performed in preparation only, if HDF or HF therapy is selected, the
online port is open and port-to-port priming is used.
Test Code 41
Repeats Until Error Message 3
ID Error Message 1430: Blood side test failed
Actions after Error Repeat test.
LLC Test LLC Test Steps LLS Test Steps LLS Test
Code Code
1. 4110x Save initial PA. - -
BPA = 0 rpm
BPV = 0 rpm
SAKV, VDE, VDA --------------------------- -- closed
Final Condition:
Wait for stabilized DF flow
Not passed after 15 s timeout
2. 4120x Change to initial PA Start condition 41001
If condition is met: Correct Valve position:
- VSAE, VSB ---------------------------------- open - VSAE, VSB ---------------------------------- open
- All other valves --------------------------- closed - VSAA, VDE, VDA, SAKV ---------------------- closed
Not passed after 5 s timeout -> Substitution line (for PV) ------------------ closed
Not passed after 12 s timeout
3. 4130x Wait for PV to stabilize: Start condition 41002
Less than 5 mmHg/s change Wait for PV to stabilize:
- Less than 5 mmHg/s change
Correct Valve position:
- VSAE, VSB ---------------------------------- open
- VSAA, VDE, VDA, SAKV ---------------------- closed
Substitution line (for PV) --------------------- closed
Correct Valve position:
- Balance chamber in stabilized DF flow
Wait for LLS feedback Send feedback to LLC
Not passed after 5 s timeout Not passed after 12 s timeout
4. 4140x Start BPV. Start condition 41003
Save PV sample Wait for PV to increase by >20 mmHg
BPV speed > 0
BPA speed = 0
Correct Valve position:
- VSAE, VSB ---------------------------------- open
- VSAA, VDE, VDA, SAKV ---------------------- closed
Not passed after 4 s timeout
 Substitution line (at PV) ------------------- closed
5. 4150x Wait for PV to increase by >20 mmHg. Finished. 41005
Wait for LLS feedback. Send feedback to LLC
If final condition is met:
Stop BPV.
Not passed after 3 s timeout
 No PV increase or 5 s timeout for LLS feedback
6a 4160x Test failed Test failed 41004
or
6b 4170x Test passed and finished - -
Close VSAE.

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 42/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
11. Default Table SW 9.01 to 9.1A
System Configuration
Unit HD HDF Online US Default Modified
Dialysate Side Parameters
Na Concentration/Conductivity Measurement Unit - mS/cm mS/cm mS/cm
Acetate Mode Disabled - NO YES YES
Default Conductivity Mode - BIC BIC BIC
Stand-by Function Available in Preparation - YES YES NO
Stand-by after Self Test/Rinsing - YES YES NO
Maximum Stand-by Time h:min 1:00 1:00 5:00
Selected Acetate - B.BRAUN B.BRAUN B.BRAUN
CONCENTRATE 1 CONCENTRATE 1 CONCENTRATE 1
Selected Bicarbonate - BIC without NaCl BIC without NaCl BIC without NaCl
Selected Acid - B.BRAUN ACID 1 B.BRAUN ACID 1 B.BRAUN ACID 1
BIC Cartridge Change with Draining (from SW 9.12) - NO NO NO
Temperature Measurement Unit - °C °C °C
Dialysate Flow ml/min 500 500 600
Dialysate Temperature °C 37.0 37.0 37.0
Sodium Profile (from SW 9.16) - YES YES YES

Settings for Acetate Concentrate Unit


1st to 10th Concentrate Name - B.Braun Concentrate 1 – 10
Conversion Factor mS/cm per mmol/l 0.10500
Default Value mS/cm 14.0

Settings for BIC


1st Bicarbonate without NaCl - BIC without NaCl
Conversion Factor mS/cm per mmol/l 0.0950
Default Value mS/cm 3.0
Acid Selection - B.Braun Acid 1

2nd Bicarbonate with NaCl - BIC with NaCl


Conversion Factor mS/cm per mmol/l 0.0950
Default Value mS/cm 5.8
Acid Selection - B.Braun Acid 2

Setting for Acid -


1st to 20th Concentrate Name - B.Braun Acid 1 – 20
Conversion Factor mS/cm per mmol/l 0.10500
Default Value mS/cm 14.3

Blood Side Parameters -


BS Pressure Test with Equalization of Pressure - NO NO YES
Default Blood Flow at End of Therapy ml/min 100 100 200
AV Line without Chamber is Possible - NO NO NO
AV Line with PBE Connector - YES YES NO
PA Chamber from Start of Preparation (from SW 9.12) - NO NO NO
Min-Max Parameters
Delta PA min. mmHg 70 70 70
Delta PA max. mmHg 70 70 70
Limit Maximum PBE mmHg 700 700 700
Limit Delta PBE mmHg 150 150 600
Limit Maximum TMP mmHg 350 350 500
Limits Low/High TMP - active active active
Low Limit TMP % 40 40 40
High Limit TMP % 40 40 40
Minimum TMP Alarm Window mmHg 20 20 20
Extended TMP Limit Range Button is Displayed in - NO NO NO
Treatment
Select Extended TMP Limit Range - inactive inactive inactive
Arterial Bolus Parameters
Arterial Bolus Volume ml 100 100 100
Arterial Bolus with SAKA Support NO NO NO

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 43/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
Unit HD HDF Online US Default Modified
Single Needle Parameters
Single Needle Valve
Limit Min. PA mmHg -200 -200 -200
Max. Control Min. PV mmHg 150 150 150
Control Min. PV mmHg 100 100 100
Control Max. PV mmHg 350 350 390
Single-Needle Cross-Over
Control PA mmHg -180 -180 -180
Control PV mmHg 360 360 360
Heparin Parameters
Treatment without Heparin - NO NO NO
Heparin Measurement Unit - ml/h ml/h ml/h
Conversion Factor Heparin (IE/ml) - 1000 1000 1000
Heparin Rate ml/h 3.0 3.0 3.0
Heparin Bolus Volume ml 0.0 0.0 0.0
Therapy Beginning Bolus - NO NO NO
Selected Syringe Type - Omnifix 30 ml Omnifix 30 ml B-D 10 ml
Syringe Table
Heparin Pump Compact Syringe Sizes: 10 ml to 30 ml
Syringe Type and volume Inside Diameter [mm]
Braun OPS 20 ml 19.00
Terumo 20/25 ml 20.10
Terumo 30/35 ml 22.57
B-D (for US) 10 ml 14.43
B-D 20 ml 18.90
B-D 30 ml 21.53
Monoject 20 ml 19.83
Dispomed 30 ml 22.17
Fresenius HS 30 ml 22.20
Omnifix 10 ml 16.05
Omnifix 20 ml 19.80
Omnifix 30 ml 22.00
Unit HD HDF Online US Default Modified
Rinsing Parameters
Priming without Recirculation - NO NO NO
Filling BP Rate ml/min 100 100 100
Filling BP Volume ml 700 700 500
Automatic Rinse Program after Self Tests - YES YES YES
User Message for BS Filled? - YES YES YES
Rinsing Blood Flow ml/min 200 200 200
Rinsing Dialysate Flow ml/min 500 500 600
Rinsing Time by UFP h:min 00:59 00:59 00:10
Rinsing Rate by UFP ml/h 203 203 600
Rinsing Volume by UFP ml 200 200 100
Rinsing Time by BPA or BPV h:min - 00:30 -
Rinsing Rate by BPA or BPV ml/h - 6000 -
Rinsing Volume by BPA or BPV ml - 3000 -
Blood Flow for Connecting Patient ml/min 100 100 100
UF Parameters
Minimal UF Rate ml/h 50 50 50
UF Rate Compensation? - YES YES YES
Max. Value of Upper Limit UF Rate (up to SW 9.04) ml/h 3000 3000 3000
Max. Value of Upper Limit UF Rate can be set to 4000 ml/h 3000 3000 3000
ml/h (from SW 9.12)
Default Value of Upper Limit UF Rate (up to SW 9.04) ml/h 2000 2000 2000
Default Value of Upper Limit UF Rate can be set to 4000 ml/h 2000 2000 2000
ml/h if Max. Value of Upper Limit UF Rate was set to
4000 ml/h (from SW 9.12)
UF Volume ml 2000 2000 2000
Therapy Time h:min 04:00 04:00 03:30
Absolute End of Therapy Time Setting Possibility?(from - NO NO NO
SW 9.12)

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 44/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
Unit HD HDF Online US Default Modified
Gross UF Rate vs. Blood Flow - inactive inactive inactive
% 40 40 40
Gross UF Rate vs. Blood Flow - active active active
% 30 30 30
UF Profile Editor - - - -
HDF/HF Online Parameters
HDF Substitution Flow ml/min - 60 -
HF Substitution Flow ml/min - 100 -
HDF Dialysate Flow for Postdilution ml/min - 600 -
HDF Dialysate Flow for Predilution ml/min - 700 -
HDF/HF Infusion Bolus Volume ml - 100 -
Post-/Predilution Selection - - inactive -
Disinfection Parameters
Disinfection after each Therapy - NO YES NO
Automatic Preparation Start after Disinfection - NO NO NO
Thermal Disinfection - YES NO YES
Chemical Disinfection - YES YES YES
Chemical Disinfection, Short - YES NO NO
Central Thermal Disinfection - NO NO NO
Central Manual Chemical Disinfection - NO NO NO
Central Automatic Chemical Disinfection - NO NO NO
Rinsing - NO NO YES
Automatic Switch-On with Rinsing - No Weekly Schedule - inactive inactive inactive
(up to SW 9.04)
Switch-On Time (up to SW 9.04) h:min 06:30 06:30 06:30
Automatic Weekly Disinfection Program (from SW 9.12) - active active active
Automatic Switch-On by Weekly Schedule (up to SW 9.04) - active active active
Weekly Disinfection Program (from SW 9.12) - inactive inactive inactive
Weekly Schedule (see Table)
Start on Sunday - inactive inactive active
Start on Monday - active active inactive
Auto Switch-On without Daily Confirmation - NO NO NO
Weekly Schedule (from SW 9.12) - inactive inactive inactive
Maximum Out of Action Time - inactive inactive active
Days Days 0 0 2
Hours/Minutes h:min 08:00 08:00 00:00
Disinfection Configuration Data Table - Citric Acid 50% Citric Acid 50% Citric Thermal
Central Thermal: Inlet Flow ml/min 250 250 250
Central Thermal: Time h:min 00:30 00:30 00:30
Central Manual Chemical: Inlet Flow ml/min 250 250 250
Central Manual Chemical: Inlet Time h:min 00:05 00:05 00:05
Central Manual Chemical: Rinsing Flow ml/min 800 800 800
Central Manual Chemical: Rinsing Time h:min 00:30 00:30 00:30
Central Automatic Chemical: Inlet Volume ml 260 260 260
Central Automatic Chemical: Switch off for Retention - NO NO NO
without Automatic Switch-On
Central Automatic Chemical: Retention Time h:min 00:20 00:20 00:20
Central Automatic Chemical: Rinsing Flow ml/min 800 800 800
Central Automatic Chemical: Rinsing Time h:min 00:30 00:30 00:30
Rinsing: Inlet Flow ml/min 800 800 800
Rinsing: Time h:min 00:02 00:02 00:02
Automatic Descaling (from SW 9.12):
- Enabled - NO NO NO
- Descaling Time h:min 0:05 0:05 0:05
o
- Descaling Temperature C 40 40 40
Filter Parameters
Name - Diacap-Ultra Diacap-Ultra Diacap-Ultra
Therapy Number - 150 150 150
Filter Operation Time h 900 900 900

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 45/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
Unit HD HDF Online US Default Modified
Dialyzer Parameters (Kt/V)
Default Target Kt/V - 1.20 1.20 1.20
Urea Distribution Volume Formula - Watson Watson Watson
Height Measurement Unit of Watson Formula - cm cm cm
Default Setting for Treatment Mode Warning of Kt/V - Warning active Warning active Warning inactive
Target Projected Deviation (up to SW 9.04)
Default Setting for Treatment Mode Warning of Kt/V - NO NO NO
Target Projected Deviation (from SW 9.12)
Default Setting for Treatment Mode Warning of Kt/V YES YES YES
Target Projected Deviation (from SW 9.16)
Kt/V Table - Manual Input: Setting Clean/Total Blood [%] - YES YES YES
Warning for Saving Kt/V Volume enabled (up to SW 9.04) - NO NO NO
Kt/V-UV
Default Target Kt/V - 1.20 1.20 1.20
Kt/V Correction Mode - Single Pool Single Pool Single Pool
Default Setting for Treatment Mode: Warning of Kt/V - Warning active Warning active Warning inactive
Target-Projected Deviation (up to SW 9.04)
Default Setting for Treatment Mode: Warning of Kt/V - NO NO NO
Target-Projected Deviation (from SW 9.12)
Warning for Saving Kt/V Table - NO NO NO
Timings
Parameter Setting Window Disappearing Time s 500 500 600
Screen Saver Appearance - NO NO NO
Appearing Time s 600 600 600
Language Selection
English - active active active
German - inactive inactive inactive
Summer Time Setting (enter values accordingly)
Begin of Summer Time D/M/Y dd.mm.yyyyy dd.mm.yyyyy dd.mm.yyyyy
h:min xx:xx xx:xx xx:xx
End of Summer Time D/M/Y dd.mm.yyyyy dd.mm.yyyyy dd.mm.yyyyy
h:min xx:xx xx:xx xx:xx
Time Difference h:min xx:xx xx:xx xx:xx
ABPM
Delete Button for Data - YES YES YES
Auto Start of Cyclic Reading (from Patient Identification - NO NO NO
Media)
ABPM Cyclic Reading Stop at Entering Disinfection - YES YES YES
Taking Over Limits from Patient’s Parameters - YES YES YES
ABPM measurement during inflation (from SW 9.16) - NO NO NO
bioLogic RR Comfort
Guide Line Mode - NO NO -
Suggested SYS Lower Limit - NO NO -
bioLogic RR Algorithm ID - 100 100 -
Miscellaneous Parameters
Click Sound - NO NO YES
Warning and Writing (Saving) Patient Parameters - NO NO YES
Skip Self Tests (up to SW 9.13) - NO NO NO
Skip Self Tests deleted (from SW 9.16) - - - - -
Automatic Preparation Start - YES YES NO
Data Validation with - Touch Touch Enter Key
UF Profile Editing in Treatment - YES YES NO
Manual Bypass Confirmation Window Enabled - NO NO YES
End of Therapy Sound Duration - 100 % 100 % 100 %
Chopped Alarm Sound - NO NO NO
Suppression of Warning Sounds in Preparation - NO NO NO
Automatic Reinfusion Start at Entering End of Therapy - YES YES NO
DCI
Dianet Address - 1 1 1
Data Set - MPI 0.12 MPI 0.12 MPI 0.12
Data Set (from SW 9.12) - MPI 0.13 MPI 0.13 MPI 0.13

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 46/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
Unit HD HDF Online US Default Modified
DBI
Baud Rate kBaud 38.4 38.4 38.4
User Logout Time min 15 - ON 15 - ON 15 – ON
Message Warning Time min 15 - OFF 15 - OFF 15 - OFF
Data Set DBI 1.61 - inactive inactive inactive
DBI 1.72 - inactive inactive inactive
DBI 2.00 - inactive inactive inactive
DBI 2.20 - inactive inactive inactive
DBI 2.21 (from SW 9.16) - active active active
DBI 2.22 inactive inactive inactive
DBI 2.30 - inactive inactive inactive
DBI-WAN Interface – Network Name Configuration
Pre-SN String is needed - - - d / YES
SN String is needed - - - - / YES
Post-SN String is needed - - - / NO
Network Name (max. 15 chars.) - - - d-
Preventive Maintenance
Warning for Preventive Maintenance Enabled - NO NO NO
Load Interval - inactive inactive inactive
Elapsed Month - inactive inactive inactive
Setting Elapsed Month months 12 12 12
Elapsed Working Time Counter WTC - inactive inactive inactive
Setting Elapsed WTC hours 4000 4000 4000
Production Report
HD HDF Online US Default Modified
Options
ABPM inactive inactive active
DCI inactive inactive inactive
bioLogic RR inactive inactive inactive
Kt/V inactive inactive inactive
Kt/V UV inactive inactive inactive
Card Reader inactive inactive active
HCT Crit-Line inactive inactive inactive
Low Level Options:
Double Pump 1) 1) 1)
Holder for BIC Cartridge 1) 1) 1)
DF Filter 1) 1) 1)
HDF Online 1) 1) 1)
Level Regulation 1) 1) 1)
Battery 1) 1) 1)
Nexadia-BSL (DBI) 1) 1) 1)
WAN-BSL (DBI) 1) 1) 1)
PFV (from SW 9.16) (not applicable/do not select) 1) 1) 1)
1) Setting depends on the configuration of the machine.
Blood Leak Detector Limit ( TSM Menu 2.7)
HD HDF Online US Default Modified
Standard Limit (0.5 ml/min at Hematocrite HK 0.45) (up to SW 9.13) active active inactive
Standard Limit (0.5 ml/min at Hematocrite HK 0.45) (from SW 9.16) inactive inactive inactive
AAMI Limit (0.35 ml/min at Hematocrite HK 0.25) (up to SW 9.13) inactive inactive active
AAMI Limit (0.35 ml/min at Hematocrite HK 0.25) (from SW 9.16) active active active
Position Setting for BICP, K P and UFP ( TSM Menu 2.8.2)
Unit Default US Default Modified
BICP Parking - 180 180
KP Parking - 180 180
UFP Parking - 180 180
UFP Positive Pressure Test (up to SW 9.13) - 270 270
UFP Positive Pressure Test “deleted” (from SW 9.16) - - - -
UFP Negative Pressure Test - 90 90
BICLF and ENDLF Ratio ( TSM Menu 2.10)
Unit Default US Default Modified
BICLF Ratio - 25 25
ENDLF Ratio - 31 36
BICLF Cart Ratio - 35 35

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 47/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
PV Alarm Window, PA Low Limit ( TSM M enu 2.12)
Unit Default US Default Modified
PV Alarm Window mmHg 100 100
PV Low Limit Position mmHg 35 35
PV Absolute Low Limit mmHg 20 20
PA Absolute Low Limit mmHg -200 -300
PA Connection Selection ( TSM Menu 2.12.1) ( from SW 9.12)
Unit Default US Default Modified
AV Line with PA Connector - active (0) active (0)
AV Line without PA Connector - inactive (1) inactive (1)
Pump Head Selection ( TSM Menu 2.13)
Unit Default US Default Modified
Pump Head 8 x 12 - active active
Pump Head 7 x 10 - inactive inactive
BPA and BPV ( TSM Menu 2.13.1)
Unit Default US Default Modified
BPA µl/Head Revolution 12306 12306
BPV µl/Head Revolution 12306 12306
HDF Online Substitution Pump OSP ( TSM Menu 2.15)
Unit Default US Default Modified
OSP µl/Head Revolution 8300 8300
Level Regulation ( TSM Menu 2.16)
Unit Default US Default Modified
LRP (PPR) Slow % 60 60
LRP (PPR) Fast % 100 100
Weekly Disinfection Program
Time Method Disinfectant Time Method Disinfectant
Monday
Night - - -
Morning 06:30 Chemical Long Citric Acid 50%
Tuesday
Night - - -
Morning 06:30 Chemical Long Citric Acid 50%
Wednesday
Night - - -
Morning 06:30 Chemical Long Citric Acid 50%
Thursday
Night - - -
Morning 06:30 Chemical Long Citric Acid 50%
Friday
Night - - -
Morning 06:30 Chemical Long Citric Acid 50%
Saturday
Night - - -
Morning - - -
Sunday
Night - - -
Morning - - -
Weekly Schedule ( Auto Switch-On) ( from SW 9.12)
Begin of First Shift End of Last Shift
Monday 8:00 16:00
Tuesday 8:00 16:00
Wednesday 8:00 16:00
Thursday 8:00 16:00
Friday 8:00 16:00
Saturday 8:00 16:00
Sunday 8:00 16:00

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 48/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
US Weekly Disinfection Program
Time Method Disinfectant Time Method Disinfectant
Monday
Night - - -
Morning 06:30 Rinsing -
Tuesday
Night - - -
Morning 06:30 Rinsing -
Wednesday
Night - - -
Morning 06:30 Rinsing -
Thursday
Night - - -
Morning 06:30 Rinsing -
Friday
Night - - -
Morning 06:30 Rinsing -
Saturday
Night - - -
Morning - - -
Sunday
Night - - -
Morning - - -
Disinfection Configuration Data
( Standard Machine)
Chemical Disinfection (long or short)
Disinfectant Reaction Time Rinsing Temperature Conductivity
Name Volume Long Short Time min.
(max. 20 Characters) ml h:min h:min h:min o
C mS/cm
Citric Acid 50 % 120 0:15 0:05 0:05 83 2.2
Decalcification with Citric Acid 50 % 120 0:05 0:05 0:05 40 2
* Peracetic acid 1 to < 5 90 0:20 0:20 0:25 40 0.1
* Hydrogen peroxide 1 to < 35
* Acetic acid 1 to < 10
(composition of concentrate [%])
# Tiutol KF 90 0:15 0:05 0:25 60 10.5
* Puristeril 340 90 0:20 0:05 0:25 40 0.1
Doxan 160 0:20 0:05 0:25 40 0.2
Dialox 160 0:20 0:05 0:25 40 0.2
Peresal 110 0:20 0:05 0:25 40 0.1
Citrosteril 120 0:15 0:05 0:20 83 2.0
Maranon H 90 0:20 0:05 0:25 40 10
Diasteril 90 0:15 0:05 0:15 80 0.5
* DF filter: a max. of 150 disinfections can be performed with Citric Acid 50 % or peracetic acid/hydrogen peroxide/acetic acid/Puristeril 340
# DF filter: max. one disinfection can be performed with Tiutol KF at the end of its service life time (service life, 150 treatments/ 900 h). An immediate filter
exchange is required after the use of Tiutol KF, i.e.:
- Default: citro-thermal disinfection (thermal disinfection) with citric acid 50% after every treatment
- Option: after 150 treatments (150 treatments and prior to exchange of filter) with Tiutol KF and then citro-thermal disinfection (after exchange of filter)

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 49/50


11Dialog+
Dialog+ SWSW 9.xx
9.xx Software
Software 9.19
9.1A Technical
TechnicalInformation
Information07/2017
01/2018
Disinfection Configuration Data
( HDF Online Machine with Diacap Ultra)
Chemical Disinfection (long or short)
Disinfectant Reaction Time Rinsing Temperature Conductivity
Name Volume Long Short Time min.
o
(max. 20 Characters) ml h:min h:min h:min C mS/cm
Citric Acid 50 % 130 0:15 0:15 0:05 83 2.2
* Peracetic acid 1 to < 5 120 0:20 0:20 0:25 40 0.1
* Hydrogen peroxide 1 to < 35
* Acetic acid 1 to < 10
(composition of concentrate [%])
# Tiutol KF 120 0:30 0:30 0:30 60 10.5
* Puristeril 340 120 0:20 0:20 0:25 40 0.1
* DF filter: a max. of 150 disinfections can be performed with Citric Acid 50 % or peracetic acid/hydrogen peroxide/acetic acid/Puristeril 340
# DF filter: max. one disinfection can be performed with Tiutol KF at the end of its service life time (service life, 150 treatments/ 900 h). An immediate filter
exchange is required after the use of Tiutol KF, i.e.:
- Default: citro-thermal disinfection (thermal disinfection) with citric acid 50% after every treatment
- Option: after 150 treatments (150 treatments and prior to exchange of filter) with Tiutol KF and then citro-thermal disinfection (after exchange of filter)
US Disinfection Configuration Data
Chemical Disinfection (long or short)
Disinfectant Reaction Time Rinsing Temperature Conductivity
Name Volume Long Short Time min.
o
(max. 20 Characters) ml h:min h:min h:min C mS/cm
Citric Thermal 120 00:15 00:05 00:05 83 2.2
Decalcification Short 120 00:05 00:05 00:05 40 2.2
Bleach 90 00:15 00:05 00:25 30 5.0
* DF filter: a max. of 150 disinfections can be performed with Citric Acid 50 % or peracetic acid/hydrogen peroxide/acetic acid/Puristeril 340

TI_2018_01_DialogP_SW_9.xx_SW 9.1A_EN.doc/pdf Art. No. 38910436 Rev. 1.00.00/February 2018 50/50