Hemodialysis Machine
Technical Manual
Section Page
Section Page
Clear.: XXXml/min Shows the last measured clearance value. Displayed for
1 minute after a successful measurement.
Plasma-Na: XXXmmol/l Shows the plasma sodium concentration of the last single
measurement for 1 minute.
Kt/V: X.XX Shows the current Kt/V. Displayed for 1 minute when the UF
goal has been reached or if the OD senses clear fluid.
Kt/V: Not available There are no Kt/V data available as the calculation of Kt/V was
aborted. Displayed for 1 minute when the UF goal has been
achieved or if the OD senses clear fluid.
HDF/HF < > OCM An OCM and an HDF or HF treatment procedure cannot be
performed simultaneously. Displayed once if OCM is selected
while HDF/HF is active. If HDF/HF is selected while OCM is al-
ready active, the OCM will be aborted and the message “OCM
off” will be displayed. The OCM can, however, be reselected
after HDF/HF has been switched off.
SN < > OCM The OCM cannot be started as a Single Needle treatment (SN
or SN click-clack) has been selected.
Profile< > OCM An OCM cannot be started in conjunction with a Na-/UF profile
1, 5 or 6 and a treatment time of less than 180 minutes. Dis-
played when the OCM is switched on and/or when selecting a
profile.
OCM off/X The OCM was aborted due to a malfunction and will be deacti-
vated until the end of the treatment.
Inform the service technician of the error code (X).
OCM adjust cancelled The OCM adjustment was not successfully completed and will
be repeated after the next treatment.
OCM adjust intended Displays when the optical detector senses dark fluid and if an
adjustment measurement is required on completion of the
treatment in progress.
Input value too low The operator has attempted to enter a value for an OCM para-
meter that is below the minimum limit.
Input value too high The operator has attempted to enter a value for an OCM para-
meter that is above the maximum limit.
1.1.2 Warnings
Kt/V too low Displays if the prescribed Kt/V cannot be reached at the end of
the treatment with the current Kt/V.
OCM OFF The OCM option will be deactivated until the machine is
switched off.
– Power failure with battery backup
– CPU II error
OCM OFF/2 The OCM option will be deactivated until the machine is
switched off.
One of the CD cells has measured a value below 11.5 mS/cm
during more than 3 consecutive OCM measurements.
– Check CD cells for proper connection
– P.C.B. LP 633-5 defective
OCM OFF/3 The OCM option will be deactivated until the machine is
switched off.
– System error
OCM OFF/5 The OCM option will be deactivated until the machine is
switched off.
A downward pulse occurs although an upward pulse was re-
quested.
– Check the bibag/concentrate delivery
– Check the concentrate pumps / mixing system
– Air in the dialysate
OCM OFF/6 The OCM option will be deactivated until the machine is
switched off.
– System error
OCM OFF/7 The OCM option will be deactivated until the machine is
switched off.
Implausibility of the ratio of blood flow to clearance or no appa-
rent change on the CD cell during more than three consecutive
measurements.
– Poor terminal connection of the CD cell
OCM tech. deact./1 The OCM option is decalibrated and can only be reactivated
after a successful pulse calibration.
Deviation between the 8-bit and the 12-bit CD value of the pre-
cell > 0.5 ms/cm for > 20 s.
– Disinfection
– Check OCM calibration and calibrate, if necessary
OCM tech. deact./4 The OCM option is decalibrated and can only be reactivated
after a successful pulse calibration.
Three cyclic pressure holding tests (CPHT) showed a deviation
of > 0.5 ms/cm between CD cell 7 and CD cell 110 for > 4.5 s.
– Disinfection
– Check OCM calibration and calibrate, if necessary
OCM tech. deact./7 The OCM option is decalibrated and can only be reactivated
after a successful pulse calibration.
More than 5 consecutive measurements showed a clearance of
more than 1.3 times the blood flow or below zero.
– Check the CD cells for poor terminal connection.
– Disinfect, if necessary
– Check the calibration
OCM tech. deact./8 The OCM option is decalibrated and can only be reactivated
after a successful pulse calibration.
Zero pulse measurement failure or time of maximum pulse be-
tween pre- and post-CD cell > 13 s.
– Check the hydraulic for proper flow
– Check OCM calibration and calibrate, if necessary
OCM tech. deact./9 The OCM option is decalibrated and can only be reactivated
after a successful pulse calibration.
The correction factor of the zero pulse measurement for upward
or downward pulses is outside the acceptable range of 0.8 to
1.2.
– Perform a disinfection
– Check OCM calibration and calibrate, if necessary
Section Page
Caution
The accuracy of the measuring equipment used during the calibration procedure
is decisive for the accuracy of the OCM measurement.
The measuring equipment used for the calibration of the conductivity must have
an accuracy of 0.05 mS/cm within a temperature range of 35 °C – 39 °C.
We recommend to use the UMED universal measuring device which is available
from Fresenius (part no. M32 403 1).
CALIBRATION
7 8 9 0 Esc
4 5 6
1 2 3 C Best
Confirm
CALIBRATION key
▲▼
DIAGNOSTICS
▲▼
MISCELLANEOUS
▲▼
SETUP MENU
▲▼
CALIBRATION
Confirm
key
▲▼
▲▼
▲▼
▲▼
CALIB. (B)-PUMP-RATE
▲▼
▲▼
▲▼
▲▼
▲▼
▲▼
Confirm
CALIB. TEMPERATURE key see Part 11
▲▼
▲▼
Confirm
CALIB. CONDUCTIVITY key see Part 13
▲▼
▲▼
CALIBRATE BLD
▲▼
▲▼
▲▼
NOVRAM
▲▼
Confirm
back to main menu ? key
▲▼
CALIB. TEMPERATURE
Confirm
key
▲▼
Confirm see Part 11.2
Check TEMPERATURE key
▲▼
Confirm see Part 11.3
OCM TEMP.-Check key
▲▼
Confirm
back to menu? key
▲▼
☞ Note
Accuracy of the external meter to be connected: ±0.2 °C.
Temp. ADJUSTMENT
Confirm
key
adj. temp to 37°C
DATA STORED
➀ NTC3; 8 bit
➁ Control 8 bit
approx. 3 sec
➂ NTC3; 12 bit
Esc
OCM TEMP. Adjust key
➃ NTC109; 8 bit,
Machine in bypass if NTC 109 is activated in SETUP
1 XXXX ADC-digits
DATA STORED
approx. 3 sec
2 XXXX °C
3 XXXX ADC-digits
4 XXXX DAC-digits
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
➀ NTC3; 12 bit
➁ NTC3 temperature
➂ NTC4; 12 bit
➃ Dynamic measurement range switching
Check TEMPERATURE
Confirm
key
use
to select the
Temperature setting field
Specify a temperature by
pressing the +/– keys.
Tone mute
key
ACKNOWLEDGED
approx. 3 sec
➀ NTC3; 8 bit
3 XXX.X °C
4 XXXX ADC-digits
Alpha Temperature
display adjustment
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
OCM TEMP.-Check
Confirm
key
use
to select the
Temperature setting field
Specify a temperature by
pressing the +/– keys.
Tone Mute
key
ACKNOWLEDGED
approx. 3 sec
➀ NTC3; 12 bit
3 XXXX ADC-digits
4 XX.XX °C
Alpha Temperature
display adjustment
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
CALIB. CONDUCTIVITY
Confirm
key
▲▼
Confirm see Part 13.2
COND. with Concentr. key
▲▼
Confirm see Part 13.3
TEMP/COND COMP Calib key
▲▼
Confirm see Part 13.4
OCM PULSE calib. key
▲▼
Confirm see Part 13.5
CONDUCTIVITY Check key
▲▼
Confirm see Part 13.6
COND Check with Conc key
▲▼
Confirm see Part 13.7
OCM COND.-Check key
▲▼
Confirm see Part 13.8
TEMP/COND COMP Check key
▲▼
Confirm
back to menu? key
▲▼
CONDUCTIVITY Set
Confirm
key
Esc
CD set: 12.80 mS/cm key
CD set: 15.70 mS/cm
Use to
move to the Use to
Concentrate setting field. move to the
If no key is Concentrate setting field.
Conc. set = ±0 % pressed for If no key is
approx. 4 sec conc. set = –X % pressed for
approx. 4 sec
Use +/– to reduce
the concentrate pump Use +/– to raise
volume to get to a CD in the concentrate pump
the lower display range. volume to get to a CD in
the lower display range.
e.g.:
Conc. set. = –X %
Conc. set. = +10 %
Tone Mute
key Tone Mute
key
ACKNOWLEDGED
ACKNOWLEDGED
approx. 3 sec
approx. 3 sec
ACKNOWLEDGED
DATA STORED
approx. 3 sec
approx. 3 sec
ADC-digits ➂
1 XXXX
[ADC-digits (OCM)]
➃ CD cell 110; 12 bit/ 8 bit alternating
2 XXXX steps
(concentrate pump)
3 XXXX steps
(bicarbonate pump)
4 XXXX ADC-digits
[ADC-digits (OCM)]
Alpha Concentrate Bicarbonate
display setting setting
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
☞ Note
An adjustment of the conductivity using a calibration concentrate is currently
not possible, since a suitable calibration concentrate is not available.
Confirm
key
act temp = XX.XX °C
1 XX.XX mS/cm
conc. set = +X %
Tone Mute
2 XX.XX mS/cm
key
3 XX.XX mS/cm
ACKNOWLEDGED
4 XX.XX mS/cm
approx. 3 sec
Alpha Temperature
act temp = XX.XX °C display adjustment
Tone Mute
key
DATA STORED
approx. 3 sec
Confirm
key
Tone Mute
key
approx. 3 sec
Tone mute
LED is flashing.
Info sound
☞
alternating
approx. 3 sec
4 XXXX mS/cm
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
CONDUCTIVITY Check
Confirm
key
use to select
the Concentrate
setting field
If no key is
conc. set = ± 0% pressed for
approx. 4 sec
Use the +/– keys
to reduce the
concentrate pump volume
e.g.:
conc. set = –6 %
Tone Mute
key
ACKNOWLEDGED
approx. 3 sec
ADC-digits ➂
1 XXXX [ADC-digits (OCM)]
➃ CD cell 110; 12 bit / 8 bit alternating
2 XXXX steps
(concentrate pump)
3 XXXX steps
(bicarbonate pump)
4 XXXX ADC-digits
[ADC-digits (OCM)]
Alpha Concentrate Bicarbonate Temperature
display setting setting adjustment
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
☞ Note
A calibration of the conductivity using a calibration concentrate is currently not
possible, since a suitable calibration concentrate is not available.
OCM COND.-Check
Confirm
key
OCM Cond.-Check
Use to
select the
concentrate setting field
If no key is
conc. set = ± 0% pressed for
approx. 4 sec
Use the +/– keys
to reduce the
concentrate pump volume.
e.g.:
conc. set = –6 %
Tone Mute
key
ACKNOWLEDGED
approx. 3 sec
OCM Cond.-Check
Verify the conductivity.
Check:
Reference meter, display,
Repeat the LED bargraph display
procedure with
different values
Esc
key
➀ CD cell 7; 12 bit
4 XX.XX mS/cm
Alpha Concentrate Bicarbonate Temperature
display setting setting adjustment
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
Confirm
key
The individual
conductivities and their
compensation factors are
displayed in the fields
(see 4008 H/S)
Esc
key
➀ CD cell 7, compensated
act temp = XX.XX °C ➁ CD cell 7, compensation factor
➂ CD cell 110, compensated
1 XX.XX mS/cm
➃ CD cell 110, compensation factor
2 XXXX mS/(cm•°C)
3 XX.XX mS/cm
4 XXXX mS/(cm•°C)
Alpha Concentrate Bicarbonate Temperature
display setting setting adjustment
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
6.1 Overview
CALIBRATION
▲▼
DIAGNOSTICS
▲▼
MISCELLANEOUS
▲▼
Confirm
SETUP MENU key
▲▼
SETUP MENU
Confirm
key
▲▼ ▲▼
▲▼ ▲▼
▲▼ ▲▼
▲▼ ▲▼
▲▼ ▲▼
▲▼ ▲▼
▲▼ ▲▼
▲▼ ▲▼
Confirm see Part 9
OCM SETTINGS key SET CENTRAL-DELIVERY
▲▼ ▲▼
▲▼ ▲▼
CONFIRM
ACTIV. MONIT_NTC109 BACK TO MAIN MENU ? KEY
▲▼ ▲▼
OCM SETTINGS
Confirm
key
▲▼
Confirm see Part 9.2
OCM ZERO MEASUREMENT key
▲▼
Confirm see Part 9.3
OCM DELAY TIME key
▲▼
Confirm see Part 9.4
OCM CLR CORR FACTOR key
▲▼
Confirm see Part 9.5
OCM BASELINE DIFF key
▲▼
Confirm see Part 9.6
OCM INTEGRAL VALUE key
▲▼
Confirm see Part 9.7
OCM KT/V WARNLEVEL key
▲▼
Confirm
back to menu? key
▲▼
OCM MEASUREMENT
Confirm
key
Confirm
key
☞ Note
If the OCM zero measurement is set to “OFF”, the OCM option will be deacti-
vated.
When the OCM zero measurement is reactivated (ON) it is imperative to per-
form an OCM pulse calibration.
DATA STORED
Confirm
key
Esc
ClrCorr Factor: 1.000 key
Tone Mute
key
DATA STORED
☞ Note
Do not change the default value.
☞ Note
In 4008 H/S machines it is generally possible to define a correction factor for
all clearance measurements. This feature has been provided to be able to
adapt a machine at its place of installation to latest knowledge gained from
scientific discussion that has not been completed yet. Based on present know-
ledge the correction factor for Fresenius Polysulfone membranes must be ad-
justed to “1” (factory setting). The accuracies specified are only applicable for
Fresenius Polysulfone membranes in combination with a correction factor that
was set to “1”.
Confirm
key
Esc
Base Diff.: 0.015 ms/cm key
Tone Mute
key
DATA STORED
☞ Note
Do not change the default value.
Confirm
key
Esc
Integral value: 55 key
Tone Mute
key
DATA STORED
☞ Note
Do not change the default value.
Confirm
key
Esc
Kt/V Warnlevel: 85 % key
Tone Mute
key
DATA STORED
Section Page
8.1 Circuit diagrams and component layout LP 632, CPU 2 .......................................... 8-3
8.2 Circuit diagrams and component layout LP 633-5, Input board ............................. 8-19