Professional Documents
Culture Documents
Hemodialysis Device
Technical Manual
Software version: ≥ 4.5 / ≥ 5.3
Edition: 12A-2016
Part no.: F40003413
0123
Table of contents
4 Adjustment instructions
4.1 Overview of the DIP switches in the 4008 ............................................................................. 4-10
4.1.1 PCB LP631 (CPU1) DIP switch array 1 .................................................................................... 4-10
4.1.2 PCB LP631 (CPU1) DIP switch array 2 .................................................................................... 4-11
4.1.3 PCB LP632 (CPU2) DIP switch array 1 .................................................................................... 4-12
4.1.4 PCB LP632 (CPU2) DIP switch array 2 .................................................................................... 4-13
4.2 Calibration mode ..................................................................................................................... 4-14
4.2.1 Basic conditions......................................................................................................................... 4-14
4.2.2 Messages on the displays on the UF monitor (4008 B) or on the screen (4008 H/S) ............... 4-15
4.3 Hydraulics ................................................................................................................................ 4-16
4.3.1 Reduced water inlet pressure .................................................................................................... 4-16
4.3.2 Degassing pressure................................................................................................................... 4-18
4.3.3 Balancing chamber loading pressure ........................................................................................ 4-20
4.3.4 Flow pump pressure .................................................................................................................. 4-22
4.3.5 UF pump volume ....................................................................................................................... 4-24
4.3.6 CDS (Central Delivery System) pressure switch ....................................................................... 4-27
4.3.7 Checking the bibag® pressure transducer (Envec) ................................................................... 4-29
4.4 Level detector .......................................................................................................................... 4-30
5 Calibration program
6 Diagnostics program
6.1 General information................................................................................................................... 6-1
6.2 Menu structure........................................................................................................................... 6-3
6.3 Reading the analog inputs of CPU1......................................................................................... 6-5
6.4 Reading the analog inputs of CPU2......................................................................................... 6-7
7 Setup menu
7.1 Overview of Setup menu settings............................................................................................ 7-1
7.2 Overview..................................................................................................................................... 7-4
7.3 Main menu for 4008 B ............................................................................................................... 7-5
7.4 Main menu for 4008 H/S .......................................................................................................... 7-30
8 Miscellaneous
Device name The device 4008 B / 4008 H / 4008 S will be referred to as the device in
the following.
Purpose This document is intended for service technicians and is to be used for
first studies (to acquire a basic knowledge) and for reference purposes
(for Technical Safety Checks, maintenance procedures, and repairs).
The document, however, does not replace the training courses offered
by the manufacturer.
Identification The document can be identified by the following information on the title
page and on the labels, if any:
– Edition of the document
– Part number of the document
Styles used in the The following text styles may be used in the document:
document
Style Description
Illustrations The illustrations used in the documents may differ from the original if
this does not have any influence on the function.
Warning
Advises the operator that failure to observe this information can result
in personal injury.
Note
Advises the operator that failure to observe this information can result
in the following:
– Damage to the device.
– Specific functions not being executed at all or not being executed
correctly.
Warning
Risk of infection
Contamination of the device with blood results in a risk of infection.
The following must be observed to avoid an infection:
– Wear protective gloves.
– Disinfect your hands after removing the gloves.
– Wear a surgical mask and goggles if there is a risk of splashes and
aerosols.
– Clean and disinfect the device before handling.
Warning
Risk of infection
Contamination of the device with blood results in a risk of infection.
If there is a possibility that fluid might have passed through the
hydrophobic filter, the device must be checked for contamination after
the treatment.
If the device is contaminated, it must be removed from service. Before
starting the device again, all affected parts must be replaced in
accordance with the manufacturer’s specifications.
Requirements Knowledge of the current Instructions for Use for the respective device.
Background experience in mechanics, electrical, and medical
engineering.
Authorized persons
Warning
Start-up, extensions, adjustments, calibrations, maintenance
procedures, modifications, or repairs may only be carried out by the
manufacturer or persons authorized by the manufacturer.
Test equipment and The activities described in this document require the availability of the
accessories necessary technical test equipment and accessories.
Specifications Technical data is provided in the current version of the Instructions for
Use and must be observed.
Precautions Before turning the power on, repair any visible damage.
Prior to opening the device and when working on the open device, the
following precautions have to be observed:
– Protect the components against ingress of fluids.
– Do not touch live parts.
– All plugs, connections, and components must only be disconnected
or connected when the device is turned off.
ESD precautions When repairing the device and replacing spare parts, observe the
relevant ESD precautions.
Hygienic measures The device and the consumables are generally considered to be
contaminated and must therefore be sufficiently disinfected by the
operator as specified by the manufacturer.
Warning
When the device is returned to use, check that the pressure of the water
supply meets the prescribed minimum pressure.
To be observed after A disinfection, a T1 test, and a check of the electrical safety must be
working on the device performed after working on the device.
The current Instructions for Use must be enclosed with the device.
Additional operator training may be required.
1.5 Addresses
Local service
MODULE TEST
T1 BATTERY
START T1 TEST
TEST OK No SAVE ERROR NUMBER
?
Yes
TEST TEST
BYPASS BLOOD LEAK DETECTOR
Yes Yes
TEST TEST
OPT. DETECTOR TEMPERATURE
Yes Yes
TEST TEST
BLOOD SYSTEM NEG. PRESSURE HOLDING
Yes Yes
TEST TEST
VENOUS PRESSURE SYSTEM POS. PRESSURE HOLDING
Yes Yes
TEST TEST
LEVEL DETECTOR UF FUNCTION
Yes Yes
TEST TEST
DISPLAY CONDUCTIVITY
Yes Yes
Yes
T1 TEST Yes
FAILED DIALYSIS START KEY
No
INCORRECT
TEST STEP
RETURN
Yes
FURTHER
Yes NEXT INCORRECT TEST
INCORRECT
TEST STEPS STEP
No
RETURN
MODULE
T1
START T1 TEST
TEST OK Yes TEST OK Yes TEST OK Yes TEST OK Yes TEST OK Yes TEST OK Yes
? ? ? ? ? ?
No 1 No No No 1 No 1 No 1
SAVE ERROR NUMBER SAVE ERROR NUMBER SAVE ERROR NUMBER SAVE ERROR NUMBER SAVE ERROR NUMBER SAVE ERROR NUMBER
Yes Yes
TEST OK TEST OK
? ?
No No
No 1 No No 1
No No
No No No
TEST OK Yes
? TEST OK Yes TEST OK Yes
? ?
No 1
No 1 No 1
SAVE ERROR NUMBER
SAVE ERROR NUMBER SAVE ERROR NUMBER
No
INCORRECT TEST STEP
RETURN
TEST OK No
? DISPLAY ERROR
Yes
Yes
RETURN
Dialines not conn? The dialyzer tubes are not in the shunt interlock.
Flow alarm Tube to or from the dialyzer kinked, malfunctions in the hydraulics.
Bypass test
Illustration
X632/A26 HOT_RINSE
LP 639
(4008H)
X639/
Power A20
Logic
X632/A10 H_REL_W X639/A12
X632/A9 EM_H_OFF X639/A17
LP 647
(4008S/B)
X632/C25 V24_EN
X632/C26 V24B_EN
X632/B22 CI
LP 632
X632/A6 V26
CPU 2
X632/A6 V26
X632/A5 V24B X634R/ X634R/
X632/A4 V24 A18 C22
X634L/C25
X634L/A25
X634L/C12
X631/A20 X631/A21
LP 631
LP 630 DATA BUS
CPU 1
Mother
board
V24B
V24
V26
Display
board
LP 649
(4008 B) V24 V24B V26
LP 924 LP 922
(4008 H) (4008 S)
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
Error description
F05 Bypass The extended bypass cannot be correctly switched off by CPU2
(V24 = on, V26 = off, V24B = on).
– Acknowledgment (V24, X637/C1) → X632/A4, 0 V missing.
– Acknowledgment (V26, X637/C2) → X632/A6, 24 V missing.
– Acknowledgment (V24B, X637/C23) → X632/A5, 0 V missing.
F06 Bypass CPU1 cannot set the temperature control to hot rinse.
– Control line (HOTRINSE, X634R/C24) → X639/A20, 12 V missing.
– Acknowledgment (HOTRINSE, X634R/C24) → X632/A26,
12 V missing.
F09 Bypass The extended bypass cannot be correctly switched off by CPU1
(V24 = on, V26 = off, V24B = on).
– Acknowledgment (V24, X637/C1) → X632/A4, 0 V missing.
– Acknowledgment (V26, X637/C2) → X632/A6, 24 V missing.
– Acknowledgment (V24B, X637/C23) → X632/A5, 0 V missing.
Illustration
X632/A30 OD_OUT
X633L/C7
X633L/C8
OD_IN
X631/A20 X631/A21
LP 631
LP 630 DATA BUS
CPU 1
Mother
board
X351/5 X351/7
Display Pven
board X351/10
LP 649
(4008 B)
LP 924 LP 922
(4008 H) (4008 S)
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
Error description
F01 opt. Detector CPU1 and CPU2 interpret the optical detector differently.
– Acknowledgment (OD_OUT, X633L/C7) → X632/A30 and the
digital input of PCB LP633 measure different levels.
F04 opt. Detector CPU2 recognizes that the optical detector senses blood (required
because of the test in the cleaning program).
– Acknowledgment X632/A30, 12 V missing.
– AD28 defective.
Illustration
X632/C10 CLP_CTL
X632/C14 LDA2
X632/C21 BL_AL
LP 632
X632/B11 BPSB_VEN
CPU 2
X632/A11 BPSB_ART
BPST_ART
LP 647
(4008S/B)
X633L/ A13 X633L/ X634L/ B15 X634L/
C13 B14 C14
X631/A20 X631/A21
LP 631
BPSST_A
LP 630 DATA BUS
CPU 1
Mother
BPSST_V
board
X348/V1
X348/V3
X348/V6
X348a/1
X348a/3
X348a/6
X348a/2
X351/8
X351/6
Pven
Display
board
LP 649
(4008 B)
LP 924 LP 922
(4008 H) (4008 S)
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
Error description
F09 Blood system Acknowledgment that CPU2 detects that the arterial blood pump is
inactive (blood pump not running).
– Acknowledgment (BPSB_ART, X348a/6) → X632/A11,
12 V missing.
– Control line (BPSTT_ART, X634L/B14) → X348a/1, 12 V missing
or (BPST_ART, X634L/A14) → X348a/3, 12 V missing.
F10 Blood system Acknowledgment that CPU1 detects that the arterial blood pump is
inactive (blood pump not running).
– Acknowledgment (BPSB_ART, X348a/6) → X633L/A11,
12 V missing.
– Control line (BPSTT_ART, X634L/B14) → X348a/1, 12 V missing
or (BPST_ART, X634L/A14) → X348a/3, 12 V missing.
– Level is raised during the T1 test.
F11 Blood system The arterial blood pump cannot be stopped by CPU1.
CPU2 recognizes that the arterial blood pump remains active.
– Control line (BPSTT_ART, X634L/B14) → X348a/1, 0 V missing
and (BPST_ART, X634L/A14) → X348a/3, 0 V missing.
– Acknowledgment (BPSB_ART, X348a/6) → X632/A11,
0 V missing.
– The level is raised during the T1 test, or the up / down key on the
level detector is blocked and constantly raised.
The level is raised.
F12 Blood system The arterial blood pump cannot be stopped by CPU1.
CPU1 recognizes that the arterial blood pump remains active.
– Control line (BPSTT_A, X634L/B14) → X348a/1, 0 V missing and
(BPST_ART, X634L/A14) → X348a/3, 0 V missing.
– Acknowledgment (BPSB_ART, X348a/6) → X633L/A11,
0 V missing.
F13 Blood system Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF, or
4008HDF pump is connected (= ADKS active)
Acknowledgment that CPU2 detects that the pump is inactive (pump
is not running).
– Acknowledgment (BPSB_VEN, X348V/6) → X632/B11,
12 V missing.
– Control line (BPSST_VEN, X634L/B15) → X348V/1, 12 V missing
or (BPST_VEN, X634L/A15) → X348V/3, 12 V missing.
– Transistor T9 on PCB LP754 defective.
– IC5 on PCB LP632 defective.
– In 4008 HDF, an HDF treatment was performed, followed by a
cleaning program with the substituate pump running, then the
T1 test has been re-started.
Substituate pump must be switched off otherwise the test step will
not be passed. (Problem was corrected with SW 3.20 in 4008 H/S:
The substituate pump will be switched off automatically on starting
a cleaning program).
F14 Blood system Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF, or
4008HDF pump is connected (= ADKS active)
Acknowledgment that CPU1 detects that the pump is inactive (pump
is not running).
– Acknowledgment (BPSB_VEN, X348V/6) → X633L/A13,
12 V missing.
– Control line (BPSST_VEN, X634L/B15) → X348V/1, 12 V missing
or (BPST_VEN, X634L/A15) → X348V/3, 12 V missing.
– IC16 on PCB LP633 defective.
– PCB LP 633 detects Single-Needle pump although it is not
connected.
F15 Blood system Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF, or
4008HDF pump is connected (= ADKS active)
CPU1 fails to stop the corresponding blood pump.
CPU2 detects that the pump remains active.
– Control line (BPSST_VEN, X634L/B15) → X348V/1, 0 V missing
and (BPST_VEN, X634L/A15) → X348V/3, 0 V missing.
– Acknowledgment (BPSB_VEN, X348V/6) → X632/B11,
0 V missing.
– Transistor T9 on PCB LP754 defective.
– IC5 on PCB LP632 defective.
– During the test, the tubing system is inserted on the corresponding
pump using the Start / Stop key.
– PCB LP 633 detects Single-Needle pump although it is not
connected.
F16 Blood system Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF, or
4008HDF pump is connected (= ADKS active)
CPU1 fails to stop the corresponding blood pump.
CPU1 detects that the pump remains active.
– Control line (BPSST_VEN, X634L/B15) → X348V/1, 0 V missing
and (BPST_VEN, X634L/A15) → X348V/3, 0 V missing.
– Acknowledgment (BPSB_VEN, X348V/6) → X633L/A13,
0 V missing.
– IC16 on PCB LP633 defective.
– PCB LP 633 detects Single-Needle pump although it is not
connected.
F17 Blood system Applicable for SW 4.91/2.91 and higher, check of ADKS signal
(= ADKS not active)
Although the recognition of the venous blood pump (ADKS) is not
acknowledged, the 24 V supply voltage of the pump can be switched
off.
– Acknowledgment line (ADKS, X348V/7) → X633L/A10,
12 V missing.
– Acknowledgment (BPSB_VEN, X348V/6) → X633L/A13,
12 V missing.
– Acknowledgment (BPSB_VEN, X348V/6) → X632/B11,
12 V missing.
– Online-HDF has already been switched on during the T1 test.
– IC16 on PCB LP633 defective.
F18 Blood system Applicable for SW 5.00/4.10 and higher, check of the BPUS signal
(CPU2, PCB LP632)
At the beginning of the test step a maximum of 40 seconds may pass
until rotation has stopped. If the blood pump is being activated, the
rotation stop alarm must have been cleared.
– Acknowledgment line (BPUS, X348A/8) → X632/A13, 0 V missing.
– Acknowledgment line (BPUS, X348A/8) → X632/A13,
12 V missing.
– Blood pump speed is set to "0": Preset speed during the T1 Test.
F19 Blood system Applicable for SW 5.00/4.10 and higher, check of the BPUS signal
(CPU1, PCB LP631 via PCB LP633)
At the beginning of the test step a maximum of 40 seconds may pass
until rotation has stopped. If the blood pump is being activated, the
rotation stop alarm must have been cleared.
– Acknowledgment line (BPUS, X348A/8) → X633L/A12,
0 V missing.
– Acknowledgment line (BPUS, X348A/8) → X633L/A12,
12 V missing.
Test description Verification of the lower limit by checking the venous zero point. The
upper limit is tested by detuning the venous pressure unit in positive
direction. (The venous occlusion clamp is closed during the test.)
Illustration
LP 632
X632/C16 LDSA
CPU 2
X632/C18 PV_DET
X632/C17 P_VEN
X632/B27 X632/B28
LP 633 LP 634
Input Output
board board
X633L/B5 X634R/C18
VENT_V
X631/A20 X631/A21
LP 631
CPU 1 LP 630 DATA BUS
Mother
board
X351/1 X351/2
P VEN
Display
board
LP 649
(4008 B) X351/4 X351/10
LP 924 LP 922
(4008 H) (4008 S)
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
Error description
F01 Venous CPU1 (input board) shows a venous zero point deviation of more than
±12 mmHg (60 seconds).
– Control (VENT_VALVE, X634R/C18) → X351/1 of the vent valve
in the LD is defective.
– Acknowledgment (P_VEN, X351/4) → X633L/B5, voltage value is
outside zero point tolerance.
– P-venous not calibrated.
F02 Venous CPU2 has a venous zero point deviation of more than ±12 mmHg
(60 seconds).
– Control (VENT_VALVE, X634R/C18) → X351/1 of the vent valve
in the LD is defective.
– Acknowledgment (P_VEN, X351/4) → X632/C17, voltage value is
outside zero point tolerance.
– P-venous not calibrated.
F03 Venous With detuning in positive direction, the achieved change in the venous
display is less than 100 mmHg (7 seconds).
– Test detuning defective (PV_DET, X632/C18) → X351/2.
– Acknowledgment (P_VEN, X351/4) → X633L/B5, the change in
voltage is too low.
– P-venous not calibrated.
F04 Venous The deviation in the measured value between CPU1 and CPU2 is
more than ±12 mmHg (if Pven > 100 mmHg).
– Acknowledgment (P_VEN, X351/4) → X633L/B5 and X632/C17
measure different voltage values.
– P-venous not calibrated.
Test description – Test of the level detector by checking the alarm state.
– Switch-off of the venous occlusion clamp in the level detector
module.
Illustration
Display
board
LP 649
(4008 B)
LP 924 LP 922
(4008 H) (4008 S)
Error description
F01 Level detector CPU1 and CPU2 interpret level detector signal differently.
– Acknowledgments (LDA1, X351/14) → X632/C13 and X633L/C10
recognize different signal levels.
F02 Level detector The level detector alarm is not detected by CPU2.
– Acknowledgment (LDA1, X351/14) → X632/C13, 0 V missing.
– Transmission weakening (LDSA, X632/C16) → X351/10,
12 V missing.
F03 Level detector Level detector clamps acknowledgment (CPU2) activated (clamp
closed)
– Acknowledgment (LDA2, X351/6) → X632/C14, 24 V missing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V missing.
F04 Level detector Level detector clamps acknowledgment (CPU1) activated (clamp
closed)
– Acknowledgment (LDA2, X351/6) → X633L/C13, 24 V missing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V missing.
F05 Level detector The blood alarm signal has not been cleared (indicates an alarm).
– Acknowledgment (BL_AL, X634L/C15) → X632/C21,
12 V missing.
If the HDF option is used, this signal is not tested (special function).
F06 Level detector The level detector clamp could not be closed via the CPU2 control line.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 0 V missing.
– Acknowledgment (LDA2, X351/6) → X632/C14, 0 V missing.
F07 Level detector The level detector clamp could not be opened via the CPU2 control
line.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V missing.
– Acknowledgment (LDA2, X351/6) → X632/C14, 24 V missing.
F08 Level detector The level detector clamp could not be closed via the CPU1 control line,
or the CPU2 acknowledgment is incorrect.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 0 V missing.
– Acknowledgment (LDA2, X351/6) → X632/C14, 0 V missing.
F09 Level detector The level detector clamp could not be closed via the CPU1 control line,
or the CPU1 acknowledgment is incorrect.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 0 V missing.
– Acknowledgment (LDA2, X351/6) → X633L/C13, 0 V missing.
F11 Level detector Level detector clamps acknowledgment (CPU2) activated (clamp
closed)
– Acknowledgment (LDA2, X351/6) → X632/C14, 24 V missing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V missing.
F12 Level detector Level detector clamps acknowledgment (CPU1) activated (clamp
closed)
– Acknowledgment (LDA2, X351/6) → X633L/C13, 24 V missing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V missing.
F13 Level detector The blood alarm signal has not been cleared (indicates an alarm).
– Acknowledgment (BL_AL, X634L/C15) → X632/C21,
12 V missing.
If the HDF option is used, this signal is not tested (special function).
F14 Level detector Raise level key on the level detector is constantly active.
– Acknowledgment (LEVEL_UP, X351/3) → X632/C11, 0 V missing.
F15 Level detector Acknowledgment of mains power for ultrasonic output stage after
3 seconds not between 6.5 and 13.5 V
– Adapter board AD28 not connected.
– Acknowledgment (X351/11 → X633L/25A jumper to X633L/B7)
12 V missing.
– Relay on AD28 failed to drop.
F16 Level detector Acknowledgment of mains power for ultrasonic output stage after
3 seconds not > 14.5 V.
– Adapter board AD28 not connected.
– Acknowledgment (X351/11 → X633L/25A jumper to X633L/B7), no
voltage 16 V/24 V.
– Relay on AD28 is not tripped.
– No 10 Hz signal at ALARM_REST (X351/12)
F17 Level detector Acknowledgment of mains power for ultrasonic output stage after
3 seconds not between 6.5 and 13.5 V
– Adapter board AD28 not connected.
– Acknowledgment (X351/11 → X633L/25A jumper to X633L/B7)
12 V missing.
– Relay on AD28 failed to drop
Display test
Test description Check of all displays and indicators on the monitor front
– Display test
– Status LED
– Alarm LED
– Seven-segment display, all dark
– Seven-segment display, all 8888
– Bar graph
– CPU1 / CPU2 audible alarm
This display test must be monitored by the user!
Illustration
CPU 2
X634R/A16
LP 633 LP 634
X632/ B27 X632/ B28 Input Output
board board
+LS X634L/A13
-LS X634L/B13
X631/ A20 X631/ A21
LP 631
LP 630 DATA BUS
CPU 1
Mother
board
Display
board Test Display
LP 649
(4008 B) 8888
8888
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
Error description
F01 Display CPU1 failed to start the display test within 5 seconds.
– The "test started" information transmitted via the serial interface is
missing.
F02 Display CPU1 failed to complete the display test within 120 seconds.
– The "test completed" information transmitted via the serial interface
is missing.
Test description Test of the arterial pressure unit by electronic detuning in positive or
negative direction.
Illustration
Display
board
LP 649
(4008 B)
Error description
F01 Arterial With detuning in negative direction, the change achieved on the
arterial display is less than 100 mmHg (2 seconds).
– Acknowledgment (P_ART, X348A/7) → X633L/B12, insufficient
voltage change.
– Test detuning defective (PA_DET, X632/A17) → X348A/9.
F02 Arterial With detuning in positive direction, the change achieved on the arterial
display is less than 100 mmHg (2 seconds).
– Acknowledgment (P_ART, X348A/7) → X633L/B12, insufficient
voltage change.
– Test detuning defective (PA_DET, X632/A17) → X348A/9.
Battery test
Illustration
TESTBATT
LP 632
CPU 2
X634R/C23 X639/A10
X633L/B21 X639/A2
X639/A3
X639/A4
X631/A20 X631/A21
LP 631
LP 630 DATA BUS
CPU 1
Mother U_ACCU
board
Display
board
LP 649
(4008 B)
LP 924 LP 922
(4008 H) (4008 S)
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
Error description
F01 Battery CPU1 failed to complete the battery test within 5 seconds.
– The "test completed" information transmitted via the serial interface
is missing.
F02 Battery The battery charge is insufficient for 1 minute of audible alarm (maybe
no battery connected).
– The battery voltage (U_ACCU, ...) → X633L/B21 dropped below
17.6 V.
– Acknowledgment (U_ACCU, ...) → X633L/B21 of the battery
voltage defective.
Test description Test of the blood leak detector by lowering the capacity of the
transmitting diode.
Illustration
Display
board
LP 649
(4008 B)
LP 924 LP 922
(4008 H) (4008 S)
Error description
F01 Blood leak Blood leak channel and dimness not in alarm-free condition during the
T1 test.
– Dimness channel contaminated (calcium precipitate, etc.)
– Acknowledgment (BLL, X637/A18) → X633L/B10, voltage value
within the alarm tolerances (< 3 V).
– Acknowledgment (BLL_DIM, X637A/21) → X633L/B11, voltage
value within the alarm tolerances (< 1.5 V / > 8 V).
– DAC_BLL or DAC_DIM not within the tolerances (check
calibration).
F02 Blood leak Blood leak alarm / dimness alarm not detected in case of test
detuning.
– Acknowledgment (BLL, X637/A18) → X633L/B10, voltage value
not within the alarm tolerances.
– Acknowledgment (BLL_DIM, X637A/21) → X633L/B11, voltage
value not within the alarm tolerances (< 1.5 V).
– Test detuning (BLL_DET, X632/A25) → X633L/B27, 5 V missing.
– Calibration of DAC_BLL or DAC_DIM is too high.
– Detuning (DAC_DIM, X634R/A11) → X633L/C3 not possible.
– Dimness calibration is set to potentiometer calibration (BR6 from
pos. 1/2 to 2/3).
F03 Blood leak After test detuning, the blood leak channel and dimness fail to enter
the alarm-free state.
– Dimness channel contaminated (calcium precipitate, etc.)
– Acknowledgment (BLL, X637/A18) → X633L/B10, voltage value
within the alarm tolerances.
– Test detuning (BLL_DET, X632/A25) → X633L/B27, 0 V missing.
– Acknowledgment (BLL_DIM, X637A/21) → X633L/B11, voltage
value within the alarm tolerances (< 1.5 V / > 8 V).
– DAC_BLL or DAC_DIM not within the tolerances (check
calibration).
Temperature test
Test description Test of the upper alarm limit by electronically detuning the temperature
display in positive direction.
Illustration
LP 639
(4008H)
LP 632 X632/A26 HOTRINSE
X633R/C21 X633L/B16
X633R/ X634R/
X632/B27 X632/B28 LP 633 A20 A13
LP 634
Input BIBAG_TE Output
board board
X633R/C25
X633R/C15
X634R/C24
X631/A20 X631/A21
LP 631
LP 630 DATA BUS
CPU 1
Mother
board
MON_NTC
NTC_BIB
X639/A20
LP 639
(4008H)
Display
board Power
LP 649 Logic
(4008 B) LP 647
LP 924 LP 922 (4008S/B)
(4008 H) (4008 S)
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
Error description
F02 Temperature The actual temperature is less than 35.0 °C (test running time
> 15 minutes).
– Calibrate the temperature.
– The heater rod has failed.
– Acknowledgment (T_DIAL1, X633L/B16) → X632/A24, voltage
stuck.
F03 Temperature The actual temperature is higher than 39.0 °C (test running time
> 15 minutes).
– Calibrate the temperature.
– The regulating sensor (NTC-2) is defective.
– Acknowledgment (T_DIAL1, X633L/B16) → X632/A24, voltage
stuck.
F05 Temperature Detuning in positive direction not more than 3 °C (10 seconds).
– Acknowledgment (T_DIAL1, X633L/B16) → X632/A24, insufficient
change in voltage.
– Detuning (T_DETADJ, X632/A23) → X633R/C21 insufficient.
F08 Temperature CPU1 failed to transmit a bibag® status message within 3 seconds.
– Runtime problem (software).
F10 Temperature bibag® Temperature display out of measuring range (15–45 °C).
– Acknowledgment (NTC_BIB, X633R/C15) → ADW on PCB LP
633.
Test description: Within a specific time period, the actual value of the dialysate pressure
transducer should change within certain limits only.
Illustration
X632/A19 ACKN_ASP
LP 632
X632/B22 CI
CPU 2
X632/C27 UF_P_CTL
X632/A29 P_DIAL
LP 633 LP 634
Input Output X634L/B10
board board
X633R/C28 X634R/A24
X631/A20 X631/A21
LP 631
CPU 1 LP 630 DATA BUS
Mother
board
UF_P_CTL
+P_DIAL
Display
board
LP 649
(4008 B)
LP 924 LP 922
(4008 H) (4008 S)
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
Error description
F01 neg. Pressure During the start phase, a negative pressure of more than 450 mmHg
was set (max. test running time 120 seconds).
– Hydraulic system contaminated.
– Air separation pump started.
F02 neg. Pressure The dialysate pressure could not be set to the test pressure
(–300 mmHg to –450 mmHg) (max. test running time 120 seconds).
On repeating the measurement, the range was extended from
–260 mmHg to –490 mmHg.
– Leakage in the hydraulic system.
– UF pump defective.
– If the HDF filter test was skipped: Clamp the HDF Filter
F03 neg. Pressure The working point (116 digits) of the differential amplifier cannot be set
correctly (max. test running time 120 seconds).
– Pressure fluctuations are too large.
– The D-A converter (IC11) on PCB LP632 is defective.
– The operational amplifier (IC1/IC3) on PCB LP 632 is defective.
– Acknowledgment (P_DIAL, X633L/B6) → X632/A29 is defective.
– No CI signal (PCB LP 632 → X632/B22).
F04 neg. Pressure Pressure measurement could not be completed (max. test running
time 120 seconds).
– The D-A converter (IC11) on PCB LP632 is defective.
– The operational amplifier (IC1/IC3) on PCB LP 632 is defective.
– Acknowledgment (P_DIAL, X633L/B6) → X632/A29 is defective.
F05 neg. Pressure The air separation pump started running during the measurement
phase.
– Acknowledgment (ACKN_ASP, X634L/B10) → X632/A19,
0 V missing.
– ASP has been interrupted electrically.
F06 neg. Pressure Negative pressure holding test failed. The dialysate pressure drop
exceeds ±40 mmHg (related to ten balancing chamber switchings).
– Leakage in the hydraulic system.
In devices with HDF option, the negative pressure holding test is performed internally only; i.e., V24, V24B
are closed and V26 is open.
Test description Valves V24, V24B, and V26 are checked for proper function
(mechanical).
Test of the TMP unit by detuning it electronically in positive direction.
With the dialysate flow turned off, positive pressure is applied to the
balancing system. The actual value of the dialysate pressure transducer
is now monitored for a defined period of time.
Test the pump segment of P97.
Illustration
Display
board
LP 649
(4008 B)
LP 924 LP 922
(4008 H) (4008 S)
Error description
F01 pos. Pressure The mandatory filling program of CPU1 has not been completed
(10 seconds).
– The solenoid valve V43 is not closed.
F25 pos. Pressure No pressure increase above 150 mmHg (change in pressure) after
valve switching.
– Control signals of V24 and V24B mistaken for each other.
– Leakage in the external system (shunt interlock, dialysate lines,
etc.)
– If the HDF filter test was skipped: Clamp the HDF Filter
F26 pos. Pressure No pressure compensation after opening of V43 (–125 mmHg to
55 mmHg).
– V24 stuck (mechanically open).
– V43 not open.
– V26 leaking.
F27 pos. Pressure No pressure compensation after opening of V43 (–125 mmHg to
55 mmHg).
– V24 stuck (mechanically open).
– V43 not open.
– V189 (retentate valve) leaking.
F02 pos. Pressure The loading pressure cannot be measured via solenoid valve V26 in
the hydraulic system (P-Dial. < 600 mmHg, 15 seconds).
– Solenoid valve V26 mechanically not open.
– Solenoid valve V43 mechanically not closed.
The balancing chamber is switched to passage during this test
sequence. V24, V24B, and V43 are closed; V26 is open.
F03 pos. Pressure The hydraulic system cannot be deaerated via the solenoid valve V43;
the zero point of –125 to 55 mmHg has not been reached
(15 seconds).
– Solenoid valve V26 mechanically not closed.
– Solenoid valve V43 mechanically not open.
– Zero point outside the –125 to 55 mmHg range.
The balancing chamber is switched to passage during this test
sequence. V24, V24B, and V26 are closed; V43 is open.
F04 pos. Pressure The first working point (220 digits) of the differential amplifier cannot
be set.
– Pressure fluctuations are too large.
– The D-A converter (IC11) on PCB LP632 is defective.
– The operational amplifier (IC1/IC3) on PCB LP 632 is defective.
– Acknowledgment (P_DIAL, X633L/B6) → X632/A29 is defective.
F05 pos. Pressure Test detuning results in a change in the measuring range of more than
95 mmHg (60 seconds).
– The operational amplifier (IC2) on PCB LP632 is defective.
– Acknowledgment (P_DIAL, X633L/B6) → X632/A29, change in
voltage too large.
– Detuning defective (P_DETADJ, X632/C20) → X633R/C22.
– The balancing chamber valve V36 or V38 (waste water valve) is
leaky.
F06 pos. Pressure Test detuning results in a change in the measuring range of less than
85 mmHg (60 seconds).
– The D-A converter (IC11) on PCB LP632 is defective.
– Acknowledgment (P_DIAL, X633L/B6) → X632/A29, change in
voltage insufficient.
– Detuning defective (P_DETADJ, X632/C20) → X633R/C22.
– V26 leaking.
F07 pos. Pressure After test detuning, there is a difference (P Diff. > ±9 mmHg) between
the display and the differential amplifier.
– The voltage divider R23/R9 or the operational amplifier IC2 is
defective.
– The operational amplifier IC1/IC3 is defective.
– The balancing chamber valve V36 or V38 (waste water valve) is
leaky.
F08 pos. Pressure Test detuning results in a change in the measuring range of more than
400 mmHg (20 seconds).
– The operational amplifier (IC2) on PCB LP632 is defective.
– Acknowledgment (P_DIAL, X633L/B6) → X632/A29, change in
voltage too large.
– Detuning defective (P_DETADJ, X632/C20) → X633R/C22.
F09 pos. Pressure Test detuning results in a change in the measuring range of less than
350 mmHg (20 seconds).
– The D-A converter (IC11) on PCB LP632 is defective.
– Acknowledgment (P_DIAL, X633L/B6) → X632/A29, change in
voltage insufficient.
– Detuning defective (DIAL_DET_ADJ, X632/C20) → X633R/C22.
F10 pos. Pressure The second working point (116 digits) of the differential amplifier
cannot be set correctly.
– The D-A converter (IC11) on PCB LP632 is defective.
– The operational amplifier (IC1/IC3) on PCB LP 632 is defective.
F11 pos. Pressure Change in the dialysate pressure after closing of the solenoid valve
V43 (zero point change from –20 mmHg to +80 mmHg within
15 seconds).
– The solenoid valve V24B is not closed.
– The balancing chamber valve V36 or V38 (waste water valve) is
leaky.
The balancing chamber is switched to passage during this test
sequence. V43, V24B, and V26 are closed; V24 is open.
F12 pos. Pressure The loading pressure cannot be measured via the solenoid valves V24
and V24B in the hydraulic system (P Dial. < 600 mmHg, 15 seconds).
– Solenoid valve V24 or V24B mechanically not open.
The balancing chamber is switched to passage during this test
sequence. V43 and V26 are closed; V24 and V24B are open.
F13 pos. Pressure The hydraulic system cannot be deaerated via the solenoid valve V43;
(P-Dial. not equal to –125 to 55 mmHg, 20 seconds).
– The solenoid valve V24 is not closed.
– V43 not opening electrically or mechanically.
The balancing chamber is switched to passage during this test
sequence. V24 and V26 are closed; V24B and V43 are open.
F14 pos. Pressure Zero point change after closing of solenoid valve V43 (20 seconds).
Standard: P-Dial. not equal to –125 to 55 mmHg.
HDF option: P-Dial. not equal to –125 to 60 mmHg.
– The solenoid valve V24 is not closed.
The balancing chamber is switched to passage during this test
sequence. V24, V26 and V43 are closed; V24B is open.
F15 pos. Pressure The loading pressure is below 780 mmHg ±30 mmHg (10 seconds).
– The loading pressure is too low.
F16 pos. Pressure During the start phase, the pressure dropped below 620 mmHg
(measurement tolerance: ±30 mmHg, max. test running time
120 seconds).
– Major leakage in the hydraulic system.
– The UF pump spring is defective.
– The loading pressure is too low.
– ASP fails to occlude.
– Relief valve (78) or V43 is leaky.
F17 pos. Pressure In the start phase the dialysate pressure could not be reduced below
760 mmHg (measurement tolerance: ±30 mmHg, test running time
120 seconds).
– The loading pressure is too high.
– UF pump defective.
F18 pos. Pressure The working point (116 digits) of the differential amplifier cannot be set
correctly (test running time 120 seconds).
– The pressure variations in the system are too large.
F19 pos. Pressure Pressure measurement could not be completed (max. test running
time 120 seconds).
– The D-A converter (IC11) on PCB LP632 is defective.
– The operational amplifier (IC1/IC3) on PCB LP 632 is defective.
– Acknowledgment (P_DIAL, X633L/B6) → X632/A29 is defective.
F20 pos. Pressure Positive pressure holding test failed. While the flow was off, a pressure
drop of more than ±80 mmHg/min was detected in the hydraulic
system.
– Leakage in the hydraulic system.
– The UF pump spring is defective.
– ASP fails to occlude.
– Relief valve leaking.
– V84 leaking.
F21 pos. Pressure The dialysate pressure cannot be set to a value between 460 and
760 mmHg ±30 mmHg (10 seconds).
– The heat exchanger is defective.
– Hydraulic problem.
F22 pos. Pressure The air separation pump is not running during the test phase
(2 seconds).
– Control line (AIR_SEP+/A22) → ASP/… 24 V missing.
– Control line (AIR_SEP–/C22) → ASP/… 0 V missing.
– Acknowledgment (ACKN_ASP, X634L/B10) → X632/A19,
12 V missing.
F23 pos. Pressure Pressure drop in the hydraulic system during the measuring phase
(8 seconds).
Change more than +4 digits or more than –8 digits.
– Leakage in the pump segment of the air separation pump.
– Leakage in the heat exchanger.
– Acknowledgment (P_DIAL, X633L/B6) → X632/A29, change in
voltage too large.
F28 pos. Pressure ASP functional test (running and delivery test)
– ASP tube segment is occluded.
– ASP tube segment has been incorrectly inserted (check direction
of delivery).
– ASP is not running (electrically or mechanically).
– V87 electrically or mechanically closed.
UF function test
Illustration
X632/C7 UF_P2
X632/A7 UF_P1
LP 632
X632/C28 UF_P_EN
CPU 2
X632/C27 UF_P_CTL X634R/A24
X632/B24 UF_P2CTL
X632/A29 P_DIAL
X633L/ X634R/ X634R/
X632/B27 X632/ B6 C11 A22
B28
LP 633 LP 634
Input Output
board board
LP 631
CPU 1 LP 630 DATA BUS
.
Mother
board
.
UF_P2 (nur bei 4008 HDF)
UF_P1
+P_DIAL
Display
board
LP 649
(4008 B)
LP 924 LP 922
(4008 H) (4008 S)
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
Error description
F01 UF Function The pause between the strokes of the UF pump 1 was shorter than
220 milliseconds. Correct volume delivery is not ensured due to too
short a return.
– CPU1 issued too high a pump rate.
F02 UF Function The pulse time for the UF pump 1 is shorter than 180 milliseconds.
Correct volume delivery is not ensured due to too short an emission
time.
– The monoflop on PCB LP634 is defective (IC42/R82/C47).
F03 UF Function The pulse time for the UF pump 1 is longer than 500 milliseconds. A
maximum rate of 5000 ml/h is not possible.
– The monoflop on PCB LP634 is defective (IC42/R82/C47).
F07 UF Function The change in pressure after a stroke is less than 20 mmHg.
– The UF pump 1 is mechanically defective.
– Control line (UF_P1_CTL, X632/C27) → X634R/A24, no LOW
pulse.
F09 UF Function Dialysate pressure is outside the measuring range (15 seconds).
– UF pressure transducer defective.
– The D-A converter (IC11) on PCB LP632 is defective.
– The operational amplifier (IC1/IC3) on PCB LP 632 is defective.
F11 UF Function The pause between the strokes of the UF pump 2 was shorter than
220 milliseconds. Correct volume delivery is not ensured due to too
short a return.
– CPU1 issued too high a pump rate.
F12 UF Function The pulse time for the UF pump 2 is shorter than 180 milliseconds.
Correct volume delivery is not ensured due to too short an emission
time.
– The monoflop on PCB LP634 is defective (IC42/R65/C45).
F13 UF Function The pulse time for the UF pump 2 is longer than 500 milliseconds. A
maximum rate of 5000 ml/h is not possible.
– The monoflop on PCB LP634 is defective (IC42/R65/C45).
F09 UF Function Dialysate pressure is outside the measuring range (15 seconds).
– UF pressure transducer defective.
– The D-A converter (IC11) on PCB LP632 is defective.
– The operational amplifier (IC1/IC3) on PCB LP 632 is defective.
F17 UF Function The change in pressure after a stroke of the UF pump 2 is less than
20 mmHg.
– The UF pump 2 is mechanically defective.
– Control line (UF_P2_CTL, X632/B24) → X634R/C11, no HIGH
pulse.
F20 UF Function The difference in volume between UF pump 1 and UF pump 2 is higher
than 25% (range of tolerance 15%–35%)
– The stroke volume of UF pump 1 or UF pump 2 has been
misadjusted.
Conductivity test
Test description Test of the alarm limits by electronically detuning the conductivity by
+5% or by –5%.
Illustration
X632/A26 HOTRINSE
LP 632
CPU 2 X632/A21 COND_DET
X632/A22 COND_SIG
7b
LP 633 LP 634
X633R/A16
Input Output
board baord
X633R/
C17
X633R/C27
X631/A20 X631/A21
LP 631
CPU 1 LP 630 DATA BUS
Mother
board
COND_C108
COND_BIB
COND_C1
Display
board (Bibag-LF-Zelle)
LP 649
(4008 B)
LP 924 LP 922
(4008 H) (4008 S)
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
Error description
F01 Conductivity The conductivity failed to reach the scale limits or to stabilize within
10 minutes (±0.1 mS/10 sec).
– No concentrate connected.
– Acknowledgment (COND_SIG, X633L/B8) → X632/A22, voltage
outside the measuring range or unstable.
F02 Conductivity Detuning in positive direction not more than 0.5 mS (10 seconds).
– Acknowledgment (COND_SIG, X633L/B8) → X632/A22
insufficient.
– Detuning (COND_DET, X632/A21) → X633L/B31 insufficient.
F03 Conductivity Detuning in negative direction not more than 0.5 mS (10 seconds).
– Acknowledgment (COND_SIG, X633L/B8) → X632/A22
insufficient.
– Detuning (COND_DET, X632/A21) → X633L/B31 insufficient.
F05 Conductivity CPU1 failed to transmit a bibag® status message within 3 seconds.
– Runtime problem (software).
F08 Conductivity CPU1 cannot increase the working point (with conductivity
< 40 mS/cm uncompensated) for the bibag® conductivity by > 5 digits
– Detuning (HOT_RINSE, X634R/C24) → X633R/A16,
12 V missing.
– PCB LP633 T2 or IC26 defective.
Test description Test of the filters by testing the volume of the internal capillary and
pressure holding test.
Illustration
X632/B5 V_DSAFE
X632/C25 V24_EN
X632/C26 V24B_EN
LP 632
CPU 2 X632/A6 V26
X634R/A18
X 6 3 4 R /C 2 2
X632/A29 P_DIAL
X633L/
X632/B27 X632/B28 B6
X634L/A-C28
LP 633 LP 634
X634L/C5
Input Output
.
X634L/C7
board board
.
X634L/A7
X634L/C25
X633R/
C28
X634L/
X634L/
A-C23
A-C30
X631/A20 X631/A21
LP 631
CPU 1 LP 630 DATA BUS
Mother
board
DEGAS+P1
V112
FLOW+P1
+P_DIAL
V36
UF_P1
V35
V26
Display V32
board
LP 649
(4008 B)
LP 924 LP 922
(4008 H) (4008 S)
V26
V35
V35 V36
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
Error description
F02 Diasafe CPU1 failed to stop the balancing chamber (24 seconds).
– The message via the serial interface from CPU1 to CPU2 is
missing.
– The current rise pulse is missing (CI, X634R/A23) → X633L/C31,
5 V pulse missing.
F04 Diasafe CPU1 failed to complete one balancing chamber switching within
20 sec (30 ml fluid not removed?).
– The message via the serial interface from CPU1 to CPU2 is
missing.
– The current rise pulse is missing (CI, 634R/A23) → 633L/C31,
5 V pulse missing.
F06 Diasafe During the pressure build-up phase, a negative pressure of less than
–450 mmHg has developed (24 seconds).
– Diasafe valve is not open, control line (V_DSAFE, X632/B5) →
X637/C16, 0 V missing.
F07 Diasafe After the maximum fluid volume of 145 ml + 30 ml has been removed,
the expected negative pressure of –300 mmHg to –450 mmHg failed
to build up.
– Major leakage in the Diasafe filter membrane and/or filter housing.
– Major leakage in the O-rings on filter holder / couplings.
– V26 electrically or mechanically not closed.
F08 Diasafe The negative test pressure of more than –300 mmHg has developed
before the minimum fluid removal of 145 ml –30 ml has been
achieved.
– Diasafe filter contaminated.
– The Diasafe filter was not correctly deaerated at the start of the
test.
– V112 electrically or mechanically not open.
F09 Diasafe The zero point for pressure measurement cannot be set. The
maximum test time has been exceeded (max. test time 5 min).
– Leakage in the Diasafe filter membrane and/or filter housing.
– Leakage in the O-rings on filter holder / couplings.
– PCB LP632, IC3/pin12 not in socket or IC defective (differential
amplifier).
F10 Diasafe The negative pressure to be achieved in the test failed to stabilize
within the maximum test time of 5 minutes
(change > ±16.7 mmHg/min).
– Leakage in the Diasafe filter membrane and/or filter housing.
– Leakage in the O-rings on filter holder / couplings.
– Leakage in the hydraulic system.
– V 26 electrically or mechanically not closed.
F20 Diasafe It was not possible to prime (deaerate) the dialysate filter within
2 minutes.
– Flow problems.
– The fill program is permanently active (level sensor, osmosis
water, or PCB LP633, IC36 defective).
F01 HDF Filter Diasafe option has not been set although the ONLINE HDF has been
selected.
– On CPU2, the DIP switch array 2, switch 1 is not set to ON.
F02 HDF Filter CPU1 failed to stop the balancing chamber (24 seconds).
– The message via the serial interface from CPU1 to CPU2 is
missing.
– The current rise pulse is missing (CI, X634/A23) → X633L/C31,
5 V pulse missing.
F04 HDF Filter CPU1 failed to complete one balancing chamber switching within
20 sec (30 ml fluid not removed?).
– The message via the serial interface from CPU1 to CPU2 is
missing.
– Verify the current rise pulse.
F06 HDF Filter During the pressure build-up phase, a negative pressure of less than
–370 mmHg has developed (24 seconds).
– The HDF filter is clamped/clogged.
– Diasafe valve is not open, control line (V_DSAFE, X632/B5) →
X637/C16, 0 V missing.
F07 HDF Filter After the maximum fluid volume of 255 ml + 60 ml has been removed,
the expected negative pressure of –220 mmHg–370 mmHg failed to
develop.
– Major leakage in the Diasafe / HDF filter membrane and/or filter
housing.
– Major leakage in the O-rings on filter holder / couplings.
– V26 electrically or mechanically not closed.
F08 HDF Filter The negative test pressure of less than –220 mmHg has developed
before the minimum fluid removal of 255 ml–60 ml has been achieved.
– Diasafe / HDF filters are contaminated.
– The Diasafe/HDF filters were not correctly deaerated at the start of
the test.
– V112 electrically or mechanically not open.
F09 HDF Filter The zero point for pressure measurement cannot be set. The
maximum test time has been exceeded (10 minutes).
– Leakage in the Diasafe / HDF filter membrane and/or filter housing.
– Leakage in the O-rings on filter holder / couplings.
F10 HDF Filter The negative pressure to be achieved in the test failed to stabilize
within the maximum test time of 10 minutes
(change > ±13.3 mmHg/min).
– Leakage in the Diasafe / HDF filter membrane and/or filter housing.
– Leakage in the O-rings on filter holder / couplings.
– Leakage in the hydraulic system.
– V26 electrically or mechanically not closed.
F20 HDF Filter It was not possible to prime (deaerate) the Diasafe filter within
2 minutes.
– Flow problems.
– The fill program is permanently active (level sensor, osmosis
water, or PCB LP633, IC36 defective).
F21 HDF Filter It was not possible to correctly rinse / prime the HDF filter within
5 minutes (before the test).
– Flow problems.
– No conductivity
– Conductivity at the upper or lower end of the scale range.
– The HDF pump has stopped (e.g. cover is open).
– The delivery rate of the HDF pump is less than 380 ml/min.
– Tube diameter not set to 8 mm.
– NTC6 permanently fails to detect fluid.
– Sieve on V43 clogged.
F22 HDF Filter It was not possible to correctly rinse / prime the HDF filter within
5 minutes (after the test).
– Flow problems.
– No conductivity
– Conductivity at the upper or lower end of the scale range.
– Particularly for bibag® devices: Check the filter on V43.
– The HDF pump has stopped (e.g. cover is open).
– The delivery rate of the HDF pump is less than 380 ml/min.
– Tube diameter not set to 8 mm.
– NTC6 permanently fails to detect fluid.
– Sieve on V43 clogged.
Error description
F01 ONLINE plus Available options and DIP switch settings do not match.
F01 DIASAFE plus CPU1 system status (MST), HPU status, and DIP switch / Array2
F01 HPU changed during the test running time.
– ONLINEplus™:
CPU2: Array2, DipSw2 not set to OFF
CPU2: Array2, DipSw3 not set to ON
– DIASAFE® plus:
CPU2: Array2, DipSw2 not set to OFF
CPU2: Array2, DipSw3 not set to OFF
– MST transmitted by CPU1 not matching with the set DIP switch of
array2
– Dip switch/Array2 changed while the test was in progress
– HPU logged off
F34 ONLINE plus Pressure holding test failed. Max. number of treatments exceeded?
F34 DIASAFE plus – Diasafe and HDF filter membranes leaking / worn.
F02 ONLINE plus Dialysate outlet pressure (DA1) outside the permissible ranges
F02 DIASAFE plus (10 seconds). Test range for DA1 –125 mmHg ≤ P_dial ≤ 55 mmHg
F02 HPU – Acknowledgment DA 1 (P_DIAL, X633L/B6) → X632/A29
– Acknowledgment line DA 2 (see Chapter 9.4 on page 9-4)
F41 ONLINE plus Test valve V183 leaking. Pressure increase in the system ΔP(DA2)
F41 DIASAFE plus > 30 mmHg within 4 seconds
F41 HPU – V183 open, contaminated, or mechanically defective
– HPU, output stage, etc., defective
F42 ONLINE plus No pressure increase of ΔP(DA2) > 200 mmHg within 4 seconds after
F42 DIASAFE plus opening test valve V183 in the system.
F42 HPU – V183 fails to open or mechanically defective
– Air pump defective, is not running
– HPU, V183, and/or air pump output stage, etc., defective
F43 ONLINE plus Lower pressure test range of ΔP(DA2) > 300 mmHg not achieved
F43 DIASAFE plus within 1 second after closing test valve V183.
F43 HPU – HPU, output stage, etc., defective
– Online filter leaking
F44 ONLINE plus The upper pressure test range of ΔP(DA2) 750 mmHg was exceeded
F44 DIASAFE plus within 4 seconds after closing of test valve V183.
F44 HPU – HPU, output stage, etc., defective
F04 ONLINE plus The air pump is running although valve V43 is closed.
F04 DIASAFE plus – HPU, output stage, etc., defective
F05 ONLINE plus Door on the Online Sys module open during the rate test.
– Close the module door.
F06 ONLINE plus Port 1 open during the first pressure build-up phase.
– Close port 1 (substituate port).
F07 ONLINE plus Port 2 open during the first pressure build-up phase.
– Close port 2 (rinse port).
F08 ONLINE plus Failure to reach the test pressure PDIAL2 > 795 mmHg within
12 seconds.
– Calibrate dialysate pressure.
– Replace DA 2 (re-calibration required)
– Air pump (185) or test valve (V183) defective
– Hydraulic system or valve ONL3 (191) leaking
– Air pump control (185) based on V43 status defective
(HPU defective)
F09 ONLINE plus ONLINE system pump failed to comply with the first test rate of
100 ml/min ±9 ml/min.
– ONLINE system pump control defective
F10 ONLINE plus Monitoring unit (Hall sensor) of the pump rotor detects incorrect
rotation of the rotor (desired rate 300 ml/min ±25%).
– ONLINE system pump control defective (outside the tolerance of
±25%)
– Hall sensor / electronics defective
F11 ONLINE plus ONLINE system pump failed to comply with the second test rate of
300 ml/min ±9 ml/min.
– ONLINE system pump control defective
F12 ONLINE plus After the ONLINE system pump was switched off in the test, the
monitoring unit (Hall sensor) detects that the rotor failed to stop
correctly.
– Pump stop (output stage) defective
– Hall sensor / electronics defective
F13 ONLINE plus After the ONLINE system pump was switched off in the test, the
module detects an actual rate > 0 ml/min.
– Pump stop (output stage) defective
– Rotary transducer electronics defective
F14 ONLINE plus After activation of the substituate pump, the monitoring unit (Hall
sensor) of the pump rotor detects incorrect rotation of the rotor
(desired rate 300 ml/min).
– Pump control defective (outside the tolerance of ±25%)
– Hall sensor / electronics defective
F15 ONLINE plus After activation of the substituate pump, the system pump failed to
comply with the test rate of 300 ml/min ±9 ml/min.
– Pump control defective
F16 ONLINE plus Port 1 open during ONL valve test sequence.
– Close port 1 (substituate port).
F17 ONLINE plus Port 2 open during ONL valve test sequence.
– Close port 2 (rinse port).
F18 ONLINE plus Acknowledgment of ONL1 (V193) differs from the desired state of the
valve.
– Valve control in the ONLINE Sys module defective
– Valve acknowledgment in the ONLINE Sys module defective
F19 ONLINE plus Acknowledgment of ONL2 (V192) differs from the desired state of the
valve.
– Valve control in the ONLINE Sys module defective
– Valve acknowledgment in the ONLINE Sys module defective
F20 ONLINE plus Acknowledgment of ONL3 (V191) differs from the desired state of the
valve.
– Valve control in the ONLINE Sys module defective
– Valve acknowledgment in the ONLINE Sys module defective
F24 ONLINE plus After the valves ONL1 to 3 were opened, the pressure drop in the
system was insufficient (DP < –100 mmHg).
– Valve ONL1 / ONL2 / ONL3 electrically or mechanically not open
– Kinked tubing
– HDF filter strongly contaminated
F26 ONLINE plus Insufficient test pressure (P < 750 mmHg) in the system.
F26 DIASAFE plus – HDF filter membrane leaking (major leakage)
– No HDF filter installed
– Hydraulic system leaking
F27 ONLINE plus After the valve V189 was opened, the pressure drop in the system was
F27 DIASAFE plus insufficient (ΔP < –70 mmHg).
– Valve V189 electrically or mechanically not open
– Diasafe filter strongly contaminated
– Filter before/after V43 strongly contaminated
F28 ONLINE plus No pressure increase in the system above P > 760 mmHg.
F28 DIASAFE plus – Diasafe filter membrane leaking (major leakage)
– No Diasafe filter installed
F29 ONLINE plus Pressure holding test failed. Pressure drop within a measurement time
F29 DIASAFE plus of 30 seconds (ΔP > –10 mmHg) excessive.
– Diasafe and/or HDF filter membrane leaking
F30 ONLINE plus During the pressure holding test valve(s) ONL1, 2 or 3 and/or V189
F30 DIASAFE plus was (were) closed (according to electronic acknowledgment).
– Valve control failed
F32 ONLINE plus Valve(s) ONL1, 2, or 3 closed and/or V24 open, or port 1 or 2 open
during the rinse phase (according to electronic acknowledgment).
– Valve control failed
– Operator opened ports too early.
F34 ONLINE plus See between F01 and F02 ONLINE plus / DIASAFE plus
F34 DIASAFE plus
F41 ONLINE plus See between F01 and F02 ONLINE plus / DIASAFE plus
F41 DIASAFE plus
F42 ONLINE plus See between F03 and F04 ONLINE plus / DIASAFE plus
F42 DIASAFE plus
F43 ONLINE plus See between F03 and F04 ONLINE plus / DIASAFE plus
F43 DIASAFE plus
F44 ONLINE plus See between F03 and F04 ONLINE plus / DIASAFE plus
F44 DIASAFE plus
V84 monitoring
Error description
F01, F02 and F03 cause the V84 monitoring flag to be set. This means after one of these error messages
has occurred Bergström or sequ. dialysis is no longer possible, since it is not possible to switch the flow
off. The V84 malfunction can be eliminated by correctly performing Dis I–IV. Another possible method of
eliminating the malfunction is to use the calibration program (to be performed by a service technician only)
under the NOVRAM menu item (Delete V84 faultiness).
PSW (pressure switch) monitoring during free rinse program (only with devices with CDS)
The following – DIP switch 8 DIP array 2 on PCB LP631 must be switched to ON.
requirements must be – Free rinse followed by disinfection or hot disinfection (Dis I–V)
fulfilled to run the PSW test or
Mandatory rinse as individual program
START
No V104 opens
PSW 104 open
Pressure reduction by
membrane pumps Yes
Yes
No V102 opens
PSW 102 open Pressure reduction by
membrane pumps Yes
Yes
Error message
Rinse Failure F05 to Rinse
PSW 104 No Failure F06 or Rinse
and
Failure F12
PSW 102 open Yes
(PSW 104 and PSW102
closed) System stopped
PSW
Yes Start monitoring
Pressure reduction in the
Yes rinse chambers (3 strokes
Mandatory rinse each by conc. and bic.
Pressure build-up required pump)
V91/104 open
V100 closed
No
No
No
Specifications
Error description
Rinse Failure F05 Free rinse program with following Dis, HDIS, or mandatory rinse as
individual program in Dis I–V.
Pressure switch for PSW_104 (S124) (bicarbonate) could not be
opened.
– Pressure on distribution piping > 500 mbar (according to
specification max. 500 mbar permissible). Pressure peaks on
distribution piping: Frequently occurs in distribution pipings with
user points if, e.g., several patients are disconnected
simultaneously and disinfection is started.
– Switching point of pressure switch too low:
Desired value = 700 mbar ± 20 mbar
– Check acknowledgment of pressure switch on PCB LP633:
Bicarbonate: X633L/A19
Rinse Failure F06 Free rinse program with following Dis, HDIS, or mandatory rinse as
individual program in Dis I–V.
It was impossible to open the pressure switch for PSW_102 (S123)
(concentrate).
– Pressure on distribution piping > 500 mbar (according to
specification max. 500 mbar permissible). Pressure peaks on
distribution piping: Frequently occurs in distribution pipings with
user points if, e.g., several patients are disconnected
simultaneously and disinfection is started.
– Switching point of pressure switch too low:
Desired value = 700 mbar ± 20 mbar
– Check acknowledgment of pressure switch on PCB LP633:
Concentrate: X633L/A20
Rinse Failure F07 Free rinse program, Dis, HDIS, or mandatory rinse in Dis. I–V.
Pressure drop during the monitoring phase on PSW_104 (S124)
(bicarbonate) or pressure build-up impossible.
– Check switching point of pressure switch.
– Check loading pressure.
(Possibly splinter or contamination in orifice 151; remove tube and
blow through tube from both ends)
– Check negative pressure and orifice (89).
(To do so, remove the tube and purge from both ends)
– Check check valve (118) and filter (120).
– Check CDS valve (104).
– Check CDS path for leaks.
– Check acknowledgment of pressure switch on PCB LP633:
Bicarbonate: X633L/A19
– Cartridge filter upstream of degassing pump clogged or wrong filter
(filter for disinfectant container) installed. Filters can be
distinguished by different adapters.
Rinse Failure F08 Free rinse program, Dis, HDIS, or mandatory rinse in Dis. I–V.
Pressure drop during the monitoring phase on PSW_102 (S123)
(concentrate) or pressure build-up impossible.
– Check switching point of pressure switch.
– Check loading pressure.
(Possibly splinter or contamination in orifice 151; remove tube and
blow through tube from both ends)
– Check negative pressure and orifice (89).
(To do so, remove the tube and purge from both ends)
– Check check valve (117) and filter (119).
– Check CDS valve (102).
– Check CDS path for leaks.
– Check acknowledgment of pressure switch on PCB LP633:
Concentrate: X633L/A20
– Cartridge filter upstream of degassing pump clogged or wrong filter
(filter for disinfectant container) installed. Filters can be
distinguished by different adapters.
Rinse Failure F09 5 minutes before the end of the mandatory rinse in Dis I–V.
Pressure switch PSW_104 (S124) (bicarbonate) or PSW_102 (S123)
(concentrate) did not open after pressure reduction.
See Rinse Failure F12.
Rinse Failure F12 Free rinse program with following Dis, HDIS, or mandatory rinse as
individual program in Dis I–V.
The pressure switches for PSW_104 (S124) (bicarbonate) and for
PSW_102 (S123) (concentrate) could not be opened.
– Membrane pumps fail to run.
– V 102 or 104 fails to open.
– Pressure on distribution piping > 500 mbar (according to
specification max. 500 mbar permissible). Pressure peaks on
distribution piping: Frequently occurs in distribution pipings with
user points if, e.g., several patients are disconnected
simultaneously and disinfection is started.
– Switching point of pressure switch too low:
Desired value = 700 mbar ± 20 mbar
– Check acknowledgment of pressure switch on PCB LP633:
Bicarbonate: X633L/A19
Rinse Failure F13 Free rinse program with following Dis, HDIS, or mandatory rinse as
individual program in Dis I–V.
Pressure drop during the monitoring phase on PSW_104 (S124)
(bicarbonate) and PSW_102 (S123) (concentrate) or pressure
build-up impossible.
– Check switching point of pressure switch.
– Check loading pressure.
(Possibly splinter or contamination in orifice 151; remove tube and
blow through tube from both ends)
– Check negative pressure and orifice (89).
(To do so, remove the tube and purge from both ends)
– Check check valve (117/118) and filter (119/120).
– Check CDS valve (102/104).
– Check CDS path for leaks.
– Check acknowledgment of pressure switch on PCB LP633:
Bicarbonate: X633L/A19
Concentrate: X633L/A20
– Cartridge filter upstream of degassing pump clogged or wrong filter
(filter for disinfectant container) installed. Filters can be
distinguished by different adapters.
In case of F07, F08 and F13, the DO NOT SWITCH OFF !! message can, in addition, be alternately
displayed.
However, this message is displayed only if a mandatory rinse program is requested, since the concentrate
and bicarbonate lines still have to be emptied before the device is switched off.
Yes No
Yes
V91/99/100/102
open for 900 ms, V104 closed
Yes
Error description
Rinse Failure F11 3 minutes before the end of the mandatory rinse in Dis I–V.
The pressure switch PSW_102 (S123) (concentrate) did not open
after pressure reduction.
– Pressure on distribution piping > 500 mbar (according to
specification max. 500 mbar permissible). Pressure peaks on
distribution piping: Frequently occurs in distribution pipings with
user points if, e.g., several patients are disconnected
simultaneously and disinfection is started.
– Switching point of pressure switch too low:
Desired value = 700 mbar ± 20 mbar
– Membrane pumps fail to run.
– V102 fails to open electrically or mechanically
– Check acknowledgment of pressure switch on PCB LP633:
X633L/A20
V91/V100 Failure 3 minutes before the end of the mandatory rinse in Dis I–V.
V91 or V100 cannot be opened.
– V91 or V100 fail to open electrically: PCB LP634:
V91 = X634L/A12; V100 = X634L/C13
– V91 or V100 mechanically not open: Check sieve (148) upstream
of V100, or valves clogged.
– V99 constantly open (electrically PCB LP 634: X634L/B12 or
mechanically)
– V102 not open
– Pressure switch for PSW_102 (S123) fails to switch.
V99 Failure 3 minutes before the end of the mandatory rinse in Dis I–V.
V99 cannot be opened.
– V99 fails to open electrically:
PCB LP634: X634L/B12
– V99 fails to open mechanically:
Check sieve (149) upstream of V99, or V99 clogged.
– Pressure switch for PSW_102 (S123) fails to open.
V130 Failure 3 minutes before the end of the mandatory rinse in Dis I–V.
V130 cannot be opened (only in devices with bibag®).
– V130 electrically defective:
PCB LP634: X634L/A4
– V130 mechanically defective or clogged
– Pressure switch for PSW_102 (S123) fails to open.
– Check tubing for bicarbonate suction line and bibag block.
F14 Shortly before the end of the mandatory rinse in Dis I–V.
Rinse section test not completed correctly. Possibly caused by flow
problems.
Error description
F14 3 minutes before the end of the mandatory rinse in Dis I–V.
Flow could not be adjusted to 750 ml/min ±50 ml/min. V91 defective.
V91 Failure 3 minutes before the end of the mandatory rinse in Dis I–V.
After V91 has opened, a flow > 950 ml/min failed to develop. V91 or
valve V98 defective.
F14 Shortly before the end of the mandatory rinse in Dis I–V.
Rinse section test not completed correctly. Possibly caused by flow
problems.
The following conditions must be fulfilled for carrying out the rinse
section test:
1. The test is run during the last 3 minutes of the mandatory rinse
program only.
2. DIP switch 7 Dip array 2 on PCB LP631 must be set to ON.
No
Mandatory rinse V100 Failure,
time No
Pressure switch program stop
≤3 min open
Yes
Yes
Pressure reduction
Pressure build-up,
using BIC pump
V91/V100 open
(completed 5 times)
No
Rinse Failure F15, Rinse Failure F16,
Pressure switch No
program stop Pressure switch program stop
open
closed
Yes
Yes
Yes
Yes
Pressure reduction
via V100 through Pressure build-up
concentrate pump Pressure build-up
(completed 8 times) V91 open V91/100 open
V100 closed
Pressure switch No Rinse Failure F17, Pressure switch No Rinse Failure F20,
closed program stop closed program stop
Yes
Yes
No
Test completed
Error description
Rinse Failure F15 3 minutes before the end of the mandatory rinse in Dis I–V.
Pressure switch (bibag® pressure switch 134) could not be opened at
the beginning of the test.
– Check pressure switch: Switching point: Desired: 130 mbar + 30
– Suction error on bicarbonate pump.
– V91 constantly electrically or mechanically open.
– V99/100 constantly electrically or mechanically closed.
V91 Failure 3 minutes before the end of the mandatory rinse in Dis I–V.
It is impossible to build up pressure on pressure switch
(bibag® pressure switch 134) via V91.
– Pressure switch fails to close mechanically: Check switching point.
– V91 fails to open electrically: PCB LP634: X634L/A12
– V91 fails to open mechanically (possibly clogged)
– V130 electrically not closed: PCB LP634: X634L/A4
– V130 fails to close mechanically (possibly clogged).
– bibag® connector leaking (check O-rings)
– Sealing on the bicarbonate suction tube leaking.
– Check acknowledgment of pressure switch on PCB LP633:
X633L/A8
– V99 constantly electrically or mechanically open.
V100 Failure 3 minutes before the end of the mandatory rinse in Dis I–V.
It is not possible to reduce pressure on pressure switch
(bibag® pressure switch 134) via V100.
– V100 fails to open electrically: PCB LP634: X634L/C13
– V100 fails to open mechanically (possibly clogged)
– V91 constantly electrically or mechanically open.
– Concentrate pump fails to pump.
– Filter (148) clogged.
– Pressure switch fails to open.
Rinse Failure F16 3 minutes before the end of the mandatory rinse in Dis I–V.
Pressure switch (bibag® pressure switch 134) cannot be closed.
V99 or V130 is leaking.
– V91 not opening electrically or mechanically.
– V99 constantly electrically or mechanically open.
– V130 constantly electrically or mechanically open.
– Sealing on the concentrate suction tube leaking.
– Pressure switch fails to close.
V99 Failure 3 minutes before the end of the mandatory rinse in Dis I–V.
Pressure switch (bibag® pressure switch 134) cannot be opened.
V99 does not open.
– V99 not opening electrically or mechanically.
– V100 fails to open electrically or mechanically
– Pressure switch fails to open.
– V91 electrically or mechanically open
– Filter (149) upstream of V99 clogged
Rinse Failure F17 3 minutes before the end of the mandatory rinse in Dis I–V.
Pressure switch (bibag® pressure switch 134) cannot be closed.
– V91 not opening electrically or mechanically.
– V130 electrically or mechanically open.
– V100 electrically or mechanically open
– Pressure switch fails to close.
V130 Failure 3 minutes before the end of the mandatory rinse in Dis I–V.
Pressure switch (bibag® pressure switch 134) cannot be opened.
– V130 not opening electrically or mechanically.
– Pressure switch fails to open.
– Check tubing for bicarbonate suction line and bibag block.
– Bicarbonate line squeezed at strain relief.
– Narrowing in the reducer on the bibag® connector
Rinse Failure F20 Pressure switch (134) cannot be closed via V91/100
– V91 not opening electrically or mechanically.
– V130/V188 electrically or mechanically open.
– Pressure switch fails to close.
V188 Failure The pressure on pressure switch (134) cannot be reduced via V188.
– V188 not opening electrically or mechanically.
– Pressure switch fails to open.
– Check tubing for carbonate suction line and air separator block.
– Concentrate line squeezed at strain relief.
Rinse Failure F14 Shortly before the end of the mandatory rinse in Dis I–V.
Rinse section test not completed correctly. Possibly caused by flow
problems.
2-62
Mandatory rinse ≤ 1 min V39 test
Pressure increase?
(dav1+50 mmHg < dav2) Yes
Wait 10 seconds
only.
Chapter 2: Description of device functions and errors
No
V39 open;
Wait 5 seconds
Yes
12A-2016
Chapter 2: Description of device functions and errors
Error description
Error description
Shunt Cover open Start of a cleaning program or during the course of RI I–II, HR I–III,
Dis I–V.
The shunt interlock is not closed.
Dialines not conn? Start of a cleaning program in RI I–II, HR I–III, Dis I–V.
The dialysate couplings are not connected to the shunt interlock.
No LD alarm Priming of the blood line system in RI I–II, HR I–III, Dis I–V.
The drip chamber in the level detector module does not recognize any
alarm.
Conc line not conn Start of a cleaning program in RI I–II, HR I–III, Dis I–V.
Or end of the disinfectant suction phase in Dis V.
The concentrate plug is not connected to the rinse chamber.
Reconnect the concentrate plug to the rinse chamber.
Bic line not conn Start of a cleaning program in RI I–II, HR I–III, Dis I–V, or end of the
disinfectant suction phase in Dis V.
The bicarbonate plug is not connected to the rinse chamber.
Reconnect the bicarbonate plug to the rinse chamber.
Upper Flow Alarm During a cleaning program in RI I–II, HR I–III, Dis I–V.
The cleaning flow increases to >1000 ml/min. The program has
stopped.
The error can be acknowledged by pressing the respective cleaning
program key.
Dial. Valve failed During a cleaning program in RI I–II, HR I–III, Dis I–V.
V24 or V24b is closed although it should be open.
The program has stopped.
The error message can be acknowledged by pressing the respective
program key.
Bypass Valve failed During a cleaning program in RI I–II, HR I–III, Dis I–V.
V26 is closed although it should be open.
The program has stopped.
The error message can be acknowledged by pressing the respective
program key.
Float-Switch Failure During a disinfection program in the suction phase in Dis I–V.
The lower switching point of the float switch is not reached within
20 seconds.
The program has stopped.
Rinse after Disinf. Selection of a cleaning program when a mandatory rinse is required in
HR.
A disinfection program has been stopped and subsequently a rinsing
or hot rinsing program started.
BIBAG cover open At the start of a cleaning program in RI I–II, HR I–III, Dis I–V.
The bibag® connector is not closed (cap not attached).
Heater Failure During the CDS rinsing phase at the end of a hot rinsing program or a
hot disinfection program in CDS: HR I–III, Dis II–IV.
Heating signal (PCB LP 633: X633R/A26) does not change for
> 40 seconds.
Error description
Watchdog Error This error message can only be displayed shortly after switch-on.
Switch the device off and on again.
Check the WD relay and components. Check CPU2/CPU1.
Check the plug connectors on the monitor.
NTC 109 switched off No valid value was stored on start-up in the NOVRAM.
The difference in temperature between NTC 109 and NTC 3 is too
large.
Switch off NTC 109 in the setup menu, or recalibrate the temperature.
Error description
Cyclic PHT F03 IC1 or IC3 on PCB LP 632 is defective, or system is leaky.
Cyclic PHT F04 Test could not be completed within a specific time interval.
Shunt Cover open – PCB LP 633 C24 (100n) short circuits sporadically.
(temporarily) – Shunt interlock defective (check switches).
UF1 volume - Error UF pump test failed. The capacity of the secondary air separator is
UF2 volume - Error outside the tolerance of 100 ml ±4 ml.
Possible causes:
– The affected UF pump fails to deliver correctly (not calibrated or
mechanical defect).
– If the test result is > 104 ml, the cause may stem from air
transported from a poorly deaerated dialyzer.
F327 UF-failure Pause between two UF1 pump strokes less than 220 milliseconds.
Possible causes:
– CPU1 defective.
F328 UF-failure Pulse time of one UF1 pump stroke is less than 180 milliseconds.
Possible causes:
– Controlling monoflop on PCB LP634 defective.
F329 UF-failure Pulse time of one UF1 pump stroke exceeds 500 milliseconds.
Possible causes:
– Controlling monoflop on PCB LP634 defective.
F331 UF-failure Desired / actual rate of the UF1 pump deviates by more than ±10%.
Possible causes:
– System error
F332 UF-failure UF1 pump stopped for more than the maximum time period.
Possible causes:
– Controlling output stage on PCB LP634 defective.
– UF pump interruption
– System error
F333 UF-failure Volume changes by more than 10 ml during desired idle time (only
monitored if OD senses blood).
Possible causes:
– System error
F334 UF-failure Pause between two UF2 pump strokes less than 220 milliseconds.
Possible causes:
– CPU1 defective.
F335 UF-failure Pulse time of one UF2 pump stroke is less than 180 milliseconds.
Possible causes:
– Controlling monoflop on PCB LP634 defective.
F336 UF-failure Pulse time of one UF2 pump stroke exceeds 500 milliseconds.
Possible causes:
– Controlling monoflop on PCB LP634 defective.
F338 UF-failure Desired / actual rate of the UF2 pump deviates by more than 10%.
Possible causes:
– System error
F339 UF-failure UF2 pump stopped for more than the maximum time period.
Possible causes:
– Controlling output stage on PCB LP634 defective.
– UF pump interruption
– System error
F340 UF-failure Change in UF2 volume is more than 10 ml despite the UF being
switched off.
Possible causes:
– System error
F350 UF-failure A difference of more than 100 ml between the CPU1 and the CPU2
volume is detected during a UF data transfer after turning the UF unit
on.
Possible causes:
– System error
F351 UF-failure CPU2 could not detect plausibility of the CPU1 UF parameters.
Possible causes:
– System error
F361 UF-failure CPU1 sent the UF parameter set to CPU2 and has not received an
enable from CPU2 after a timeout of 30 seconds.
Possible causes:
– System error
F363 UF-failure CPU2 repeatedly did not receive a complete UF parameter set.
Possible causes:
– System error
F364 UF-failure UF1 volume change although the UF goal has already been reached.
Possible causes:
– System error
F370 UF-failure UF2 volume change although the UF goal has already been reached.
Possible causes:
– System error
HPU Error
Error description
HPU Error F00 The HPU logs off with index STATUS_ER; no bit is set in the error bit
field.
– Problem on PCB LP 941
– Problem on CAN distributor board
– Problem on PCB LP 763
– Problem on PCB LP 630
HPU Error F01 Cyclic communication has failed for more than 2 seconds.
– System error
HPU Error F04 Voltage drop (24 V_SW) during HPU operation.
– 24 V voltage supply on PCB LP 941 failed (watchdog dropped)
HPU Error F07 The HPU was logged off by the monitor. Will not be displayed since
CPU1 has already stopped the communication.
– System error
HPU Error F08 General valve error. Can occur in HPU SW 2.01 or 3.00. (Software
versions before evaluation of the HPU errors).
– System error
HPU Error F15 Error in the HPU software. Valves are activated incorrectly.
– System error
HPU Error F98 Transition to the T1 test is not allowed after restart.
– System error
Error description
ONL Error F05 +24 V_WD dropped to less than 17 V or was switched off
ONL Error F07 A transmission from the module was not confirmed by the monitor.
The blood pump ensures a sufficient blood flow in the extracorporeal blood circuit. It is absolutely
necessary that sterility is maintained and that the blood is prevented from becoming contaminated.
The blood pump integrated in the device is a roller pump designed as a field-replaceable plug-in module.
The blood line is inserted between the circular roller track of the pump casing (stator) and the rollers, which
are attached to the rotor. The pressure of the rollers causes the development of a narrowing and seal. If
the rollers are moving in the direction of delivery, the blood is pushed in this direction.
A microprocessor controls the stepper motor with quartz accuracy, depending on the selected delivery
rate, the set tube diameter, and the monitor signals.
The pressure measuring equipment comprises a piezo-resistive pressure transducer. The pressure-
proportional voltage is indicated on the monitor on a quasi-analog LED scale.
Functions of the blood – RAM and CRC test after turning power on
pump – Control and monitoring of the function by a dual-processor system
– Emergency stop in case of alarm: Stop detection (15 or 30 seconds)
– Setting of the speed to 180 ml/min during priming
– Measurement of the arterial pressure or the Single-Needle pressure
(depending on model)
– Semi-automatic loading and unloading of the tube segment
Error messages
Essentially, the blood pump (Single-Needle) is identical with the arterial blood pump. The difference lies
in the Single-Needle control. During SN operation, the pressure outlet of the compliance vessel is
connected to the pressure connector of the SN pump. The pressure transducer is protected by a
hydrophobic filter both in the external and the internal tubing system.
The SN stroke volume can be set within a range from 10 ml to 50 ml in increments of 5 ml.
Upper switching point 110 130 150 172 195 219 244 270 299
(mmHg) ±7 mmHg
Since the blood flows through an extracorporeal circuit during hemodialysis, coagulation would occur
within a short time. The heparin pump allows continuous heparinization of the blood causing the
coagulation time to be prolonged. Since the heparin volume required during dialysis depends on the
respective patient concerned, it must be determined by the attending physician.
A syringe plunger is moved by a drive rod. The drive rod is connected to a threaded spindle via a slide. A
microprocessor-controlled stepper motor causes the spindle to rotate. Depending on the activation, the
piston will move up or down. A Hall sensor indicates the upper end position of the plunger. The protective
system of the pump comprises a speed monitoring unit (slotted disc with optical sensor) as well as a motor
current monitoring unit.
Warning
The syringes classified as approved among the syringes indicated
below were successfully tested with the device.
If syringes that are classified as not approved in the table below are
used, it is the responsibility of the responsible operator to ensure correct
function of the device.
The manufacturer does not assume any responsibility or liability for
personal injury or other damage and excludes any warranty for damage
to the device resulting from the use of non-approved or unsuitable
consumables or accessories.
0 20 ml BD / Fresenius Approved
3 10 ml BD Not approved
4 30 ml BD Not approved
5 50 ml BD Not approved
C Not assigned –
D Not assigned –
E Not assigned –
F Not assigned –
Warning
Do not change the coding switch position during operation.
Function of the heparin – RAM and CRC test after turning power on
pump – Delivery rate adjustable between 0.1 ml and 10 ml in increments of
0.1 ml
– Delivery time selection (stop watch) ranges between 1 min and
9 h 59 min
– Bolus administration
Error codes
E56 Error in step counting (more than 8 pulses during transition of the
slotted disc; the slotted disc is oscillating)
The penetration of air into the patient's extracorporeal blood circuit may cause an air embolism. In order
to catch limited amounts of air and to separate accompanying air bubbles, the venous blood line is
expanded (venous drip chamber). A major task of the level detector is to monitor the filling level in the
venous drip chamber.
The protection system against air infusion uses the method of ultrasonic transmission. Ultrasonic
converters are attached on either side of the venous bubble catcher. At periodic intervals of approx.
90 ms, a transmitting resonator generates attenuated ultrasonic vibrations at a natural resonance of
approx. 90 kHz, which are absorbed by a receiving resonator. The amplitude of the signal received is
dependent upon the medium between the converters. Its value is at its minimum with the bubble catcher
empty (air) and at its maximum with bubble-free fluids. The amplitude decreases with increasing air
content (foam). The signal path is fail-safe up to and including the receiving resonator, i.e., the failure of
any component always leads to a smaller amplitude and, thus, to an alarm. Starting at the receiving
resonator, the signal voltage is always sent onto two independent receiver paths. As soon as the signal
is too weak, one of these receiver paths causes the blood pump to stop and the other the venous
occlusion clamp to close.
The and keys are used to raise and lower the blood level in the venous bubble catcher. As long as
the key is pressed, the venous occlusion clamp closes. The vent valve in the level detector module
opens, and the blood level rises. The blood pump runs at reduced speed (180 ml/min). As long as the
key is pressed, the venous occlusion clamp remains open. The vent valve in the level detector module
opens, the ventilation pump is running, and the blood level sinks. The blood pump runs at the preselected
speed.
Optical detector
The optical detector detects if there is blood, NaCl solution, or air in the venous return tube downstream
of the bubble catcher. In the device, the dialysis phase is defined by presence of a dark medium and the
preparation phase by presence of a clear medium.
The venous pressure measuring equipment comprises a piezo-resistive pressure sensor provided on the
PCB with following operational amplifier. The pressure-proportional output voltage is supplied onto the
logic PCB in the monitor. There, the pressure is indicated on a quasi-analog LED scale, and the
transmembrane pressure is computed by determining the difference between the dialysate pressure and
the venous pressure.
Legend
As soon as the inlet valve (41) opens, the water flows through the pressure reducing valve (61) into the
chamber (66a) of the heater block and across the heat exchanger (77) into the heater rod chamber (66b).
The concentrate pump (23) admixes concentrate to the inflowing water per balancing chamber phase.
The overflow prevents pressure from building up in chambers b and c. In Hot rinse mode, the developing
vapor can escape through the overflow.
While it is rising, the fluid is warmed up to the preset temperature by the heater (54). The heater is
controlled by the temperature sensor (2).
From chamber b, the dialysate flows into the chamber (66c). Incorporated into this chamber is a float
switch (5), which controls the solenoid valve (41), thus ensuring the correct fluid level.
The degassing pump (29) draws in the dialysate via the degassing orifice (89). This generates a negative
pressure of approximately 0.8 bar.
In the tubes and the following chamber (88a), the dialysate is degassed to a level which is sufficient for
dialysis.
Via the degassing pump (29), dialysate and released air are directed tangentially into the primary air
separator (88c), where air bubbles and the airless dialysate are separated. The air accumulates at the top
of the chamber (88c). Then, together with the recirculation flow and via the loading pressure valve (65) as
well as the chamber (66c), the air escapes into the atmosphere.
Chamber 88c is provided with a separating disc (standard hydraulics only), which serves to prevent
bicarbonate, if added, from being recirculated via the heater rod chamber (66b).
At the bottom of chamber (88c), the degassed dialysate is pressed out and into the balancing
chamber (68) by means of the loading pressure.
Together with the eight solenoid valves (31 to 38), the balancing chamber (68) constitutes the balancing
system. Each of the two sections of the balancing chamber comprises two compartments separated by
an elastic membrane each. Hence, there are two chambers with four spaces:
– F1 and F2: Fresh solution
– A1 and A2: Used solution (waste)
As soon as one of the chambers (A1 or A2) is filled with dialysate, the solenoid valves are reversed in
groups of four. The valves are reversed by the electronic evaluation of the current rise pulse of the drive
motor of the pump (21), which receives this pulse upon membrane abutment. Within the filling phase, F1
or F2 is filled with fresh dialysate by means of the loading pressure. In order to obtain a continuous flow,
a second chamber is switched parallel to the first chamber. The second chamber is operated in an inverse
sequence.
Each time the chamber is changed over (membrane end position), all valves are closed for approximately
100 milliseconds (dead time).
From the balancing chamber, the dialysate flows through the conductivity cell (7) with integrated
temperature sensor (3). The measured conductivity values are indicated on the monitor in ms/cm, related
to 25 °C.
If the actual values (temperature or conductivity) of the dialysate exceed or fall below the alarm limits, the
bypass valve (26) opens, and the dialyzer inlet valve (24) is closed. The device is now in the bypass mode.
The dialysate is discharged into the drain not via the dialyzer, but via the secondary air separator (88b)
and the balancing chamber (68).
If the actual conductivity and temperature values of the dialysate are within the set limits, the dialyzer
valve (24) opens. The valve (26) is closed. The dialysate flows to the dialyzer.
After the dialyzer, the dialysate which is now loaded with the substances usually eliminated with the urine
flows into the secondary air separator (88b) via a filter (73), the valve (24b) and the blood leak
detector (8). The secondary air separator (88b) comprises the pressure transducer (9) and the level
sensor (6).
With a hematocrit of 0.25, blood losses of 0.5 ml per minute are recognized in the dialysate by the blood
leak detector.
Together with the venous back pressure, the signal of the pressure transducer (9) is evaluated and
indicated on the monitor as TMP. The fluid level in the secondary air separator (88b) is monitored by the
level sensor (6). Due to the secondary air separator (88b), only airless dialysate is always delivered into
the balancing chamber (68). Any presence of air bubbles in the balancing chamber (68) would cause
balancing errors.
The dialysate is pressed into the balancing chamber (68) by the flow pump (21). As mentioned above, the
balancing chamber valves are reversed by the current rise pulses of the drive motor of the flow pump.
Using the speed of this pump, the dialysate flow can be adjusted in the dialysis program: 300, 500, and
800 ml/min. In the cleaning programs, the flow of the dialysate is fixed.
The relief valve (78) is used to limit the pressure of the flow pump before the balancing chamber to
approximately 2 bar.
After the balancing chamber, the dialysate flows through the valve (30), the heat exchanger (77) and the
valve (87) into the drain.
The valves (86) and (87) serve to recirculate fluid during the hot rinsing and disinfection programs.
(Standard program)
1st cycle:
30
31 32 33 34
F1 A1 F2 A2
68
35 36 37 38
21
2nd cycle:
30
31 32 33 34
F1 A1 F2 A2
68
35 36 37 38
21
This system ensures that equal amounts of fluid enter and exit the dialyzer. This leads to an exact
balancing of the dialysate and, in conjunction with the ultrafiltration pump (22), a controlled volumetric
ultrafiltration.
As soon as the fluid level in the secondary air separator (88b) has dropped below the level sensor (6), this
sensor activates the air separation pump (97). If the fluid level has reached the level sensor (6) within a
given time period, the FILL PROGRAM is started.
Note
In order to recognize the fluid level, the level sensor (6) requires fluid
with a certain minimum conductivity, which is definitely achieved in all
dialysis programs. Separation of air is only required in the dialysis
programs.
In all other programs, the air separation pump (97) and the valve (43)
are force-actuated.
1st cycle:
30
31 32 33 34
76 97
43
F1 A1 F2 A2
68
6
35 36 37 38
6
88b
21
9
84
2nd cycle:
30
31 32 33 34
76 97
43
F1 A1 F2 A2
68
6
35 36 37 38
6
88b
21
9
84
If not enough air was separated and the fluid level is still below the level sensor (6), the FILL PROGRAM
is activated.
The pump (21) fills either chamber A1 or chamber A2. Valves (36), (38), and (30) close. Valve (43) opens.
The air can escape into the drain.
1st cycle Chamber F1 is filled. This forces the fluid from chamber A1 into
chamber A2. The fluid is then forced into the secondary air
separator (88b) by chamber A2 via the dialyzer.
2nd cycle Chamber F2 is filled. This forces the fluid from chamber A2 into
chamber A1. The fluid is then forced into the secondary air
separator (88b) by chamber A1 via the dialyzer.
Filling is performed in this way to prevent a change in conductivity. As in the standard program, one stroke
of the concentrate pump is still executed per balancing chamber cycle (30 ml) here as well.
A fill program is always activated at the beginning of dialysis (to fill the dialyzer). If it is activated during
the treatment (OD senses blood), this is shown on the display.
Note
Repeated activation of the fill program during treatment indicates a
defect (leakages).
The Central Delivery System is connected to the connectors 121/122. The concentrate flows into the rinse
chambers via the inlet filters and the valves 102/104. Through the connected concentrate suction tubes,
the concentrate pumps deliver the concentrate to the mixing point.
During dialysis, the valves 91/99 and 100 are closed. Depending on the Central Delivery System, the
V102 and / the V104 are opened.
During the cleaning programs, the valves 102 and 104 are closed. During the suction phase of the
concentrate pump and bicarbonate pump, the valves 91 and 99 open for 500 ms upon each balancing
chamber switching. Valve 100 is open.
In order to check the tightness of the valves 102 and 104, the pressure switch is tested during the free
rinse phase with following disinfection or hot disinfection or a mandatory rinse. To perform this test,
pressure is applied to the two lines between the check valves 117/118 and the valves 102/104. The
pressure switches P123 and P124 are used to monitor the pressure. Three minutes before the mandatory
rinse program is completed, a functional check of the valves 91/99 and 100 is performed.
End of dialysis
Dialysis
PGM Program
R Rinse
–R– Rinse endless
endless
F Free rinse
HR Hot rinse
C Cooling rinse
D Disinfection
D(F) Cleaning
Disinfectant drawn in from the front (concentrate suction tube).
HDIS Hot disinfection
M Mandatory rinse
IHR Integrated hot rinse
At the end of the set program, the rinse chamber is evacuated for approx. 1 minute.
Any statements on time refer to the factory setting. Shorter or longer program times can be set at any time
using the SETUP menu (see Chapter 7 on page 7-1).
Rinse
PGM 1: –R–
T/°C
37
33
t/min
Evacuation of
Rinse rinse chambers
per 10 strokes
15–30 min (Setup) approx. 1 min
Start End
(Rinse key)
T/°C
37
33
t/min
Rinse endl.
Start End
(Rinse key) (Program abort)
Hot rinse
PGM 1: –F–HR–C–
T/°C
80
37
34,5
Evacuation of
t/min
Heat-up Hot rinse time-counting above Cooling rinse to
rinse chambers
Free rinse 80 °C 34.5 °C
per 10 strokes
approx. 6 min 15–30 min (Setup) approx. 8 min approx. 1 min
4–10 min approx.
(Setup) 4 min End
Start
(Hot rinse key)
PGM 2: –F–HR–
T/°C
80
37
Evacuation of
t/min
Heat-up Hot rinse time-counting above
rinse chambers
Free rinse 80 °C per 10 strokes
approx. 6 min 15–30 min (Setup) approx. 1 min
4–10 min approx.
(Setup) 4 min End
Start
(Hot rinse key)
PGM 3: –IHR–
T/°C
80
Evacuation of t/min
rinse chambers
Integrated hot rinse
per 10 strokes
approx. 1 min
15–40 min (Setup)
Start End
(Hot rinse key)
PGM 3: –IHR–C–
T/°C
approx. 80
approx. 35
t/min
Evacuation of
Heat-up Integrated hot rinse Cooling rinse rinse chambers
per 10 strokes
Disinfection
PGM 1: –F–D–M–
T/°C
37
33
Evacuation of
t/min
Free rinse Pre.* Disinfection Mandatory rinse rinse chambers
per 10 strokes
4–10 min approx. 10–20 min 15–30 min approx. 1 min
(Setup) 1 min (Setup) (Setup)
Start Mandatory rinse required End
(Disinfection key)
PGM 2: –F–HDIS–M–
T/°C
80
37
34,5
Evacuation of
t/min
Free rinse Pre.*
Hot disinfection Mandatory rinse rinse chambers
Heat-up per 10 strokes
4–10 min approx. approx. 10–20 min (Setup) 15–30 min (Setup) approx. 1 min
(Setup) 1 min 4 min
approx. 6 min
Start Mandatory rinse required End
(Disinfection key)
PGM 3: –F–D–M–HR
T/°C
80
37
33
Evacuation of t/min
Disinfec- Mandatory rinse Hot rinse rinse chambers
Free rinse Pre.* tion Heat-up per 10 strokes
4–10 min approx. 10–20 min 15–30 min approx. 15–30 min
(Setup) 1 min 4 min (Setup) approx. 1 min
(Setup) (Setup)
approx. 6 min
Start End
(Disinfection key)
PGM 4: –F–HDIS–M–HR–
T/°C
80
37
Evacuation of t/min
Free rinse Pre.* Hot Mandatory rinse Hot rinse
rinse chambers
disinfection Heat-up
Heat-up per 10 strokes
4–10 min approx. approx. 10–20 min 15–30 min approx. 15–30 min
4 min (Setup) approx. 1 min
(Setup) 1 min 4 min (Setup) (Setup)
approx. 6 min approx. 6 min
Start Mandatory rinse End
(Disinfection key) required
PGM 5: –F–D(F)–M–
T/°C
37
33
t/min
Evacuation of
Free rinse Pre.* Disinfection Mandatory rinse rinse chambers
per 10 strokes
4–10 min approx. 10–20 min (Setup) 15–30 min (Setup)
1 min approx. 1 min
(Setup)
Start Mandatory rinse required
End
(Disinfection key)
The following messages are possible depending on the setting of the DIP switches (see Chapter 4.1.2 on
page 4-11):
Message Description
RO-Mandatory rinse Message occurs after cleaning the single-station reverse osmosis system.
– Run the device's mandatory rinse program.
For initial start-up only The following information is only intended for the initial start-up. This
information does not apply to recommissioning devices that have been
taken out of service even temporarily.
Tester's qualification The tests may only be performed by persons qualified to perform the
specified checks correctly owing to their educational background,
training, knowledge, and experience. Furthermore, the persons
performing the checks must be permitted to do so independently and
without outside interference.
Warning
Start-up, extensions, adjustments, calibrations, maintenance
procedures, modifications, or repairs may only be carried out by the
manufacturer or persons authorized by the manufacturer.
Test equipment and The activities described in this document require the availability of the
accessories necessary technical test equipment and accessories.
Correcting settings Settings must only be corrected if the measured values fall outside the
specified tolerances.
Precautions Before turning the power on, repair any visible damage.
Prior to opening the device and when working on the open device, the
following precautions have to be observed:
– Protect the components against ingress of fluids.
– Do not touch live parts.
– All plugs, connections, and components must only be disconnected
or connected when the device is turned off.
ESD precautions When repairing the device and replacing spare parts, observe the
relevant ESD precautions.
4008 B / 4008 H /
Initial start-up report
4008 S
Remarks:
#FMESIG_1##
Remarks:
#FMESIG_2##
Warning
On completion of these procedures it is imperative to run a disinfection.
Re 1 Visual inspections
4008 H:
Connect the cable to the battery.
1
Type : IPX1
SN : EC :
: 2
Power :
Current :
Index_H
Fresenius Medical Care AG & Co.KGaA
D-61346 Bad Homburg
2011 +49 (0) 6172 / 609-0
Re 1.6 Color coding of the dialyzer couplings and of the shunt interlock
adapted to customer specifications (counter-current principle)
Operating state: Off
Adapt coding of the dialyzer couplings and of the shunt interlock as
desired by the customer. Ensure counter-current principle by correct
color coding.
Document the change in the medical device register.
State upon delivery:
Dialysate inlet (right): Dialyzer coupling and shunt interlock = red
Dialysate outlet (left): Dialyzer coupling and shunt interlock = blue
Note
Risk of tubing constriction
Risk of damage to the connectors
Tie the tubes up so that they are not subject to any kinking, tension,
torsion, and squeezing.
Note
DIP switch 6 (field 1) is provided for servicing / troubleshooting only and
must be set to OFF for treatment.
Warning
DIP switches 3 and 8 (field 1) allow the operator to skip test steps that
are requested by the device.
If the switches are set to the "can be skipped" position, it is important to
know that the operator can then bypass the automatic test of the safety
systems.
The person requesting this setting is responsible for this procedure.
4008 H:
Establish a connection.
Re 2.5 V91, V99, V100 are functional and tight (test not applicable with
activated CDS or with activated rinse section test)
Operating state: Off
Check the external valves to make sure they are functional and tight.
Activate the valves with the diagnostics program.
This test is not required
– when CDS is activated, PCB LP 631 SW2 DIP switch 8 is ON,
– when rinse section test is activated, PCB LP 631 SW2 DIP switch 7
is ON.
Re 3.1 Water inlet pressure (reduced): 1.2 bar (+0.2 bar / –0.3 bar)
Operating state: CALIBRATION
Connect a pressure gage to measuring point A in the hydraulic unit
before MV41.
Check the water inlet pressure when MV41 is closed.
Desired: 1.2 bar (+0.2 bar / –0.3 bar).
Re 3.2 Loading pressure: 1.45 bar (±0.05 bar) [3mix: 1.25 bar (±0.05 bar)]
Operating state: CALIBRATION
Connect a pressure gage to the pressure side of the degassing
pump on measurement point B in the HU.
Check loading pressure.
Desired: 1.45 bar (±0.05 bar) [3mix: 1.25 bar (±0.05 bar)].
Re 6 Check of temperature
Re 7 Check of conductivity
(If the bibag® option is used, connect a bibag®)
Re 10.1 Blood pumps: Check of blood pump rate (BP rate TEST)
Operating state: CALIBRATION
Check the blood pump rate (BP rate TEST).
Re 11.1 Making settings in the setup menu and transferring data to the
dialysis device
Operating state: CALIBRATION
Adjust the setup settings in the PC service program.
Transfer data to the dialysis device.
1
9
2
3
8
6
Legend:
1 Upper rear panel (screw)
2 Heat sink (power supply unit)
3 Monitor rear panel (plate)
4 Power supply plate
5 Heater rod housing (hydraulic unit open)
6 Dialyzer tube ports
(hydraulic unit open / earthing screw)
7 Bonding studs
8 Hydraulic unit rear panel (plate on the push-on blade inside)
Caution: The protective conductor must be connected.
9 Monitor rear panel (ports housing)
4008 B, 4008 S:
Test the protective earth resistance at the following measurement
points.
Desired: Protective earth resistance < 0.3 Ω.
Legend:
1 Upper rear panel (screw)
2 Heat sink (power supply unit)
3 Heater rod housing (hydraulic unit open)
4 Dialyzer tube ports (adapters)
5 Bonding studs
or
Device leakage current – direct measurement according to fig. 4
Additional conditions:
If the value of the leakage current normalized to the nominal voltage is
higher than 90% of the alarm limit (450 µA), the last measured value or
the first measured value must additionally be considered for the rating.
If the device leakage current has considerably increased since the last
measurement or has continuously increased since the first
measurement (slow deterioration of the insulation), or if the sum
composed of the current value plus the difference since the last
measurement is > 500 µA, the measurement is considered "not
passed".
Example 2 – Additional conditions passed:
Device leakage current: 470 µA
Last measured value: 450 µA
470 µA + (470 µA – 450 µA) = 470 µA + 20 µA = 490 µA: passed.
Example 3 – Additional conditions not passed:
Device leakage current: 470 µA
Last measured value: 390 µA
470 µA + (470 µA – 390 µA) = 470 µA + 80 µA = 550 µA: not passed.
Re 13 Functional check
Re 14 Final checks
Re 14.2 Air separation: Activation of the air separation pump. If more air
must be separated and OD senses blood, the display shows:
Fill program
Operating state: Dialysis
Simulate OD sensing blood.
Activate the air separation pump.
Check the text display by separating more air.
Desired: Text display Fillprogram.
Re 14.6 Entries recorded in the medical device register and machine card
Operating state: Off
Record entries in the medical device register.
Record entries in the machine card.
Re 14.7 Instructions for Use and accessories package are complete and
match the device
Operating state: Off
Check whether the Instructions for Use and accessories package
are complete and that they match the device.
Assessment of the test The device is released for its intended use
(attach inspection label)
There must be no risk to patients, staff, or third parties during the
intended use of the device.
Within the scope of the overall assessment, the tester must make a
definite decision as to whether or not the device may be used. The
responsible organization must be informed without delay of any defects
detected.
Decide whether or not the system can be released for its intended
purpose.
Inform the responsible organization of any defects in the device
without delay.
Date of next check:
Document the date of the next test in the report.
Respect the intervals specified by the manufacturer.
Remarks:
Record any irregularities which influenced your assessment.
Date, signature, stamp:
Confirm the initial start-up assessment with the date, tester’s
signature, and stamp.
This chapter lists all necessary Technical Safety Checks (TSC) and Maintenance procedures (MA). The
Technical Measurement Checks (TMC) additionally stipulated for specific countries form part of the
Technical Safety Checks (TSC).
These checks must be performed every 24 months if all of the following requirements have been met:
– Unique assignment of the rotors to the appropriate tube roller pumps
– Software:
4008 B system: from 4.951 to < 5.00 or with 5.201 or higher
4008 H/S systems: from 2.951 to < 3.00 or from 4.311 to < 10.00
– DIP switch PCB LP632 array 2 SW 5 set to OFF.
– DIP switch PCB LP631 array 2 SW 7 set to ON.
– DIP switch PCB LP631 array 2 SW 8 set to ON. (With CDS option only)
– Heater rod made of Titanium
If one of the requirements mentioned above has not been met, the checks have to be performed
every 12 months.
Performance of the Technical Safety Checks must be recorded in the Medical Device Register.
Warning
Start-up, extensions, adjustments, calibrations, maintenance
procedures, modifications, or repairs may only be carried out by the
manufacturer or persons authorized by the manufacturer.
The activities described in the Technical Manual require the availability of the necessary technical test
equipment and accessories.
When working on the open device, the following precautions must be respected:
– Protect the components against ingress of fluids.
– Do not touch live parts (e.g., connectors on the power cable or heater).
When repairing and replacing spare parts, observe the applicable ESD precautions (e.g., EN 100 015-1).
TSC / MA report The TSC / MA report (see TSC / MA report on page 3-25) and the
TSC report TSC report (see TSC report on page 3-42) differ in the following points:
– Maintenance procedures (MA)
– Category TSC or MA in first column
Explanatory notes for The explanatory notes on the report are contained in a separate chapter
completing the report (see Chapter 3.2.3 on page 3-29).
4008 B / 4008 H /
TSC / MA report
4008 S
These checks must be performed every 24 months if all of the following requirements have been met: Interval:
Unique assignment of the rotors to the appropriate tube roller pumps 24 months
Software: 12 months
4008 B system: from 4.951 to < 5.00 or with 5.201 or higher (Please check)
4008 H/S systems: from 2.951 to < 3.00 or from 4.311 to < 10.00
DIP switch PCB LP632 array 2 SW 5 set to OFF.
DIP switch PCB LP631 array 2 SW 7 set to ON.
DIP switch PCB LP631 array 2 SW 8 set to ON. (With CDS option only)
Heater rod made of Titanium
If one of the requirements mentioned above has not been met, the checks have to be performed
every 12 months.
Remarks:
#FMESIG_3##
Remarks:
#FMESIG_4##
Warning
A disinfection must be performed on completion of the checks.
These checks must be performed every 24 months if all of the following requirements Interval:
have been met: 24 months
Unique assignment of the rotors to the appropriate tube roller pumps 12 months
Software:
(Please check)
4008 B system: from 4.951 to < 5.00 or with 5.201 or higher
4008 H/S systems: from 2.951 to < 3.00 or from 4.311 to < 10.00
DIP switch PCB LP632 array 2 SW 5 set to OFF
DIP switch PCB LP631 array 2 SW 7 set to ON
DIP switch PCB LP631 array 2 SW 8 set to ON (With CDS option only)
Heater rod made of Titanium
If one of the requirements mentioned above has not been met, the checks have to be
performed every 12 months.
These inspections must be carried out by persons who are qualified to properly perform the specified
Technical Safety Checks owing to their educational background and training, their knowledge and
experience gained in practice, and who are not subject to any directions with regard to this inspection
activity.
1 Visual inspections
TSC 1.1 Fuses accessible from the Must comply with the maximum permissible
outside values.
TSC 1.4 Damage and contaminations There must not be any detectable damage or
contamination.
TSC 1.4.1 Rotors of the tube roller pumps The rotors show no signs of damage. Check the
rotors of the tube roller pumps.
– Rotors installed in the proper location and
1 identification present.
– Handle disks (3) tight and undamaged.
– Check all rollers (2) for smooth running.
2
– No mechanical damage on the guide
pulleys (1) and other parts of the rotors
(see drawing)
3 Damaged rotors must be replaced by new ones.
TSC 1.4.2 PCB LP 928 (CAN bus distributor Check PCB LP 928 with covering to verify the
board) with covering (only 4008 S following:
with E-Code 905 and lower) – There is no moisture.
– There is no precipitation.
– PCB LP 928 is not mechanically damaged in
any other way.
MA 1.6.1 Replace the sealing in the concentrate / bicarbonate suction tubes and lubricate with
Unisilkon. Replace the rivet in the suction tubes, if necessary.
MA 1.6.2 Check the rubber in the rinse chambers for proper function.
MA 1.6.5 Replace the filter sieves; upstream of the UF pump (filter 74), downstream of MV43
(filter 76), between MV99 and rinse chamber (filter 149), between MV100 and rinse
chamber (filter 148).
MA 1.6.6 Replace the filter sieve in the dialyzer tube; replace the complete filter (73),
if necessary.
MA 1.6.8 Check the tube in the sampling valve (116) dialysate circuit for proper function,
replace the complete valve, if necessary.
MA 1.6.10 Check the air separation pump (97): replace the belt ribbon and the tube segment.
Observe direction of delivery.
2 General checks
TSC 2.2 Check level sensor Draw in air via the dialysate couplings.
The air separation pump is activated.
If more air is detected, the device will switch to
the fill program, depending on the dialysate flow.
Text displayed when the OD senses blood:
Fillprogram
TSC 2.3 Check DIP switches PCB LP 631 (CPU1) DIP switch array 2 SW7 is
set to ON.
With Central Delivery System:
PCB LP 631 (CPU1) DIP switch array 2 SW8 is
set to ON.
MA 3.1 Check water inlet pressure Connect a pressure gage before MV41 to
(reduced) and correct if measuring point A in the hydraulic unit.
necessary When the valve MV41 is closed, the pressure
should range between 0.9 and 1.4 bar.
MA 3.2 Check the balancing chamber Connect a pressure gage to the pressure side of
loading pressure and correct, the degassing pump (measuring point B in the
if necessary hydraulic unit).
The pressure should be between 1.2 and 1.3 bar.
In case of the equipment codes
4008 B - EC495
4008 H - EC295
4008 S - EC275
and higher, the pressure must be 1.45 bar
±0.05 bar.
MA 3.3 Check the degassing pressure. Connect a pressure gage to the suction side of
the degassing pump (measuring point D in the
hydraulic unit).
Degassing pressure: –0.81 to –0.85 bar.
MA 3.4 Check the balancing chamber Connect a pressure gage to the pressure side of
relief pressure at a flow of the flow pump (measuring point C in the hydraulic
800 ml/min (relief valve 78). unit).
The relief pressure depends on the loading
pressure set (see MA 3.2):
Loading pressure: Relief pressure:
1.2 to 1.3 bar 1.9 to 2.1 bar
1.45 ±0.05 bar 2.2 ±0.05 bar
TSC 4.1 Check the delivery volume of the Collect 60 ml of dialysate in an appropriate
UF pump. measuring cylinder (see Chapter 4.3.5 on
page 4-24).
60 strokes = 60 ml ±0.5 ml
(59.6 g ±0.5 g dialysis water)
Correct the setting of the UF pump, if necessary.
5 Treatment
TSC 5.4 Verify the conductivity display If the bibag® option is used, connect a bibag®!
Measure the conductivity with a reference meter
connected between the dialyzer couplings.
The conductivity measured must agree with the
value on the device.
Correct any deviations with the calibration
program.
6 Extracorporeal components
MA 6.1 Arterial pressure transducer Check the slope of the pressure transducer.
After applying a pressure of approx. 200 mmHg
to the pressure transducer, the value displayed
must agree with the reading shown on the
external reference meter (tolerance ±10mmHg).
Correct any deviations with the calibration
program.
MA 6.2 Venous pressure transducer Check the slope of the pressure transducer.
After applying a pressure of approx. 300 mmHg
to the pressure transducer, the value displayed
on the device must agree with the reading shown
on the external reference meter (tolerance
±10 mmHg).
Correct any deviations with the calibration
program.
TSC 6.3 Arterial and Single Needle blood Check the blood pump rate (calibration program:
pump rate BP-Rate CHECK).
TSC 6.4 SN switching points Check the switching points according to the table
in the Technical Manual.
TSC 6.5 Check the blood pump stop Opening the blood pump door will trigger the
alarm. blood pump stop alarm after 30 s (factory
setting).
TSC 6.6 Level detector In the event of a blood alarm, the venous tube
clamp must close.
TSC 6.7 Level detector Generate pressure of about 2 bar in the venous
bubble catcher.
Ensure that the pressure does not drop by more
than 0.1 bar within 3 minutes (see Chapter 4.4
on page 4-30).
7 Options
7.1 bibag®
7.2 DIASAFE
MA 7.3.1 Filter life of the DIASAFE® plus Check the filter life.
Filter life: 12 weeks
TSC 7.4.1 Check the delivery rate of the 2nd Collect 60 ml of dialysate in an appropriate
UF pump. measuring cylinder (see Chapter 4.3.5 on
page 4-24).
60 strokes = 60 ml ±0.5 ml
(59.6 g ±0.5 g dialysis water)
If necessary, correct the value.
MA 7.5.4 HDF pump rotor Check the rotor for smooth running and wear.
or
MA 7.6.1 Filter life of DIASAFE® plus and Check the filter life
ONLINEplus™ filters Filter life of DIASAFE® plus and ONLINEplus™
filters: 12 weeks or 100 treatments
MA 7.6.3 Substituate port (195) and rinse Replace the O-rings. Grease the O-rings with
port (194) Unisilkon grease before installation.
TSC 7.6.4 Line pinch valve 193 (ONL1) Replace the tube.
TSC 7.6.5 Check of DIP switch With HPU (hydraulic processing unit)
PCB LP632 (CPU2) DIP switch array 2 switch 5
is set to OFF.
MA 7.8.1 Labels and inscriptions Must be present and clearly legible. Check of the
actual condition.
MA 7.8.2.1 Check whether the tube connector is correctly attached to the device.
MA 7.8.2.2 Check whether the internal blood pressure module, the printed circuit boards, and all
cable connections are correctly fixed.
TSC 7.8.3 Leakage test Perform the leakage test with the cuff and
TMC pressure tube connected. The pressure leakage
rate must be less than 6 mmHg/min.
(See Technical Manual BPM 4008, chapter 3.1)
TSC 7.8.6 Blood pressure measurement Measure the blood pressure in manual mode.
TMC Check the results for plausibility.
TSC 8.2 Protective earth resistance Maximum 0.3 Ω (with power cable)
or
Device leakage current – direct measurement according to fig. 4
Additional conditions:
If the value of the leakage current normalized to the nominal voltage is higher than
90% of the alarm limit (450 µA), the last measured value or the first measured value
must additionally be considered for the rating.
If the device leakage current has considerably increased since the last measurement
or has continuously increased since the first measurement (slow deterioration of the
insulation), or if the sum composed of the current value plus the difference since the
last measurement is > 500 µA, the measurement is considered "not passed".
Example 2 – Additional conditions passed:
Device leakage current: 470 µA
Last measured value: 450 µA
470 µA + (470 µA – 450 µA) = 470 µA + 20 µA = 490 µA: passed.
Example 3 – Additional conditions not passed:
Device leakage current: 470 µA
Last measured value: 390 µA
470 µA + (470 µA – 390 µA) = 470 µA + 80 µA = 550 µA: not passed.
9 Functional check
Warning
A disinfection must be performed on completion of the checks.
Test 4008 H
1
2
3
4
6
7
Legend
1 Upper rear panel (screw)
2 Monitor rear panel (ports housing)
3 Heat sink (power supply unit)
4 Monitor rear panel (plate)
5 Power supply plate
6 Heater rod housing (hydraulic unit open)
7 Hydraulic unit rear panel (plate on the push-on blade inside)
Caution: The grounding cable must be connected.
8 Dialyzer tube ports
(hydraulic unit open / earthing screw)
9 Bonding stud
Legend
1 Upper rear panel (screw)
2 Heat sink power supply unit
3 Heater rod housing (hydraulic unit open)
If in place: outward measurement point (screw head) at bottom right on the rear of the device
4 Dialyzer tube ports (adapters)
5 Bonding stud
3. Measurement conditions
The measurements must be taken in the dialysis mode in the "ON phase" of the heater control system.
The scales must be moved out to such an extent that neither of the two end switches are actuated (middle
position).
TSC / MA report The TSC / MA report (see TSC / MA report on page 3-25) and the
TSC report TSC report (see TSC report on page 3-42) differ in the following points:
– Maintenance procedures (MA)
– Category TSC or MA in first column
Explanatory notes for The explanatory notes on the report are contained in a separate chapter
completing the report (see Chapter 3.2.3 on page 3-29).
4008 B / 4008 H /
TSC report
4008 S
These checks must be performed every 24 months if all of the following requirements have been met: Interval:
Unique assignment of the rotors to the appropriate tube roller pumps 24 months
Software: 12 months
4008 B system: from 4.951 to < 5.00 or with 5.201 or higher (Please check)
4008 H/S systems: from 2.951 to < 3.00 or from 4.311 to < 10.00
DIP switch PCB LP632 array 2 SW 5 set to OFF.
DIP switch PCB LP631 array 2 SW 7 set to ON.
DIP switch PCB LP631 array 2 SW 8 set to ON. (With CDS option only)
Heater rod made of Titanium
If one of the requirements mentioned above has not been met, the checks have to be performed
every 12 months.
Remarks:
#FMESIG_5##
Remarks:
#FMESIG_6##
Warning
A disinfection must be performed on completion of the checks.
3.3 Decommissioning
Precautions for working on Before working on the device, first read the information in the following
the device chapter (see Chapter 1.4 on page 1-2).
Warning
Start-up, extensions, adjustments, calibrations, maintenance
procedures, modifications, or repairs may only be carried out by the
manufacturer or persons authorized by the manufacturer.
Article Description
Explanatory notes for The explanatory notes on the report are contained in a separate chapter
completing the report (see Chapter 3.3.4 on page 3-48).
4008 B / 4008 H /
Decommissioning report
4008 S
Antifreeze used:
(Durability assured throughout decommissioning period):
#FMESIG_7##
Identification sheet
Decommissioned device
The decommissioning of this device may only be revoked by the service support organization.
Antifreeze used:
Date of decommissioning:
Validity of decommissioning:
Note
Battery discharge
The rechargeable battery must also be maintained while a system is
decommissioned.
Follow Instructions for Use.
Re 1 Cleaning / Disinfection
Re 4 Disinfectant connector
Re 5 Hydraulics
Re 5.1 Service switch set to ON (heater off) after starting up the device
Operating state: On
Turn the device on.
Set the service switch to ON (Service mode) after starting up the
device.
The heater is deactivated.
Connect the dialysis water inlet tube (3) to the tubing connector (2).
Secure the tubing set (1) and dialysis water inlet tube (3) with tube
clips.
Insert the tubing set (1) into the arterial blood pump as shown.
Connect the suction side of the tubing set (4) to the suction tube of
the 4008 decommissioning kit.
Close off the tube (5).
Place the suction tube in the antifreeze canister (6).
Re 6 Final procedures
Precautions for working on Before working on the device, first read the information in the following
the device chapter (see Chapter 1.4 on page 1-2).
Warning
Start-up, extensions, adjustments, calibrations, maintenance
procedures, modifications, or repairs may only be carried out by the
manufacturer or persons authorized by the manufacturer.
Article Description
Explanatory notes for The explanatory notes on the report are contained in a separate chapter
completing the report (see Chapter 3.4.4 on page 3-56).
4008 B / 4008 H /
Decommissioning extension report
4008 S
#FMESIG_8##
Re 1 Recommissioning
Re 2 Decommissioning
Precautions for working on Before working on the device, first read the information in the following
the device chapter (see Chapter 1.4 on page 1-2).
Warning
When the device is recommissioned into operation, check that the
pressure of the dialysis water (permeate) supply meets the prescribed
minimum pressure.
Warning
Start-up, extensions, adjustments, calibrations, maintenance
procedures, modifications, or repairs may only be carried out by the
manufacturer or persons authorized by the manufacturer.
Article Description
Explanatory notes for The explanatory notes on the report are contained in a separate chapter
completing the report (see Chapter 3.5.4 on page 3-60).
Remarks:
#FMESIG_9##
Warning
On completion of these procedures it is imperative to run a disinfection.
Re 1 Preparation
Re 1.3 Dialysis water inlet tube and water drain tube connected
Operating state: Off
Connect the dialysis water inlet tube.
Connect the water drain tubing.
Re 2 Disinfectant connector
Re 4 Hydraulics
Re 5 Cleaning / Disinfection
Re 6 Functional check
Confirmation of The device is released for its intended use (attach inspection label)
recommissioning
There must be no risk to patients, staff, or third parties during the
intended use of the device.
Within the scope of the overall assessment, the tester must make a
definite decision as to whether or not the device may be used. The
responsible organization must be informed without delay of any defects
detected.
Decide whether or not the system can be released for its intended
purpose.
Inform the responsible organization of any defects in the device
without delay.
Date of next check:
Document the date of the next test in the report.
Respect the intervals specified by the manufacturer.
Remarks:
Record any irregularities which influenced your assessment.
Date, signature, stamp:
Confirm recommissioning with date, tester's signature and stamp.
Precautions for working on Before working on the device, first read the information in the following
the device chapter (see Chapter 1.4 on page 1-2).
Removal from service Removal from service is the definitive withdrawal of a device from
routine operation. Instructions for removal from service and disposal are
given in the latest applicable Instructions for Use, and must be followed.
Removal from service must be entered in the medical device register.
4 Adjustment instructions
Test equipment and accessories
IC extractor tool
The current list of the test equipment and accessories with part numbers can be found in the electronic
Spare Pars Catalog.
Fig.: Hydraulics flow diagram with DIASAFE® plus (optional) without HPU (basic hydraulics)
CDS - Option
ZKV - Option
Legend
Legend
LP 636
LP 630
SW1
DIP-SW1
P1 Current Increasing
1 Stromerhöhung
SW2
DIP-SW2
P3 TMP-Gain
TMP-Steilheit
(LP 763)
SH1
DIP-SW2
SH2
DIP-SW1
LP 631
Alpha-Display (4008B)
Legend
Note
Dip switch 6 is provided for service purposes/troubleshooting only and
must be set to the OFF position for treatment mode.
*
4008 H/S only
**
4008 H Prometheus only
Warning
DIP switches 3 and 8 allow the operator to skip test steps which are
requested by the device.
If the switches are set to the "can be skipped" position, it is important to
know that the operator can then bypass the automatic test of the safety
systems.
The person demanding this switch position shall be solely responsible
for such a procedure.
Warning
If the Service switch is set to ON during the treatment or the cleaning
program, the heater relay will be turned off.
4.2.2 Messages on the displays on the UF monitor (4008 B) or on the screen (4008 H/S)
4008 B
UF Goal ml
I/O Prog.
4008 H/S
CALIBRATION
004.7 Volt
Watchdog supply voltage (4.5 V–5.5 V)
011.9 Volt
12 V supply (11.5 V–12.5 V)
023.6 Volt
24 V supply (23 V–25 V)
020.4 Volt
Battery charging voltage (> 20 V)
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
4.3 Hydraulics
Note
Measuring equipment for measurement points in the hydraulic unit:
UMED, HMED, or pressure gage with measuring range from –1 to
+2.2 bar, minimum quality rating: 1.6
Condition: Flow on
Check / adjustment:
– Checking the reduced water inlet pressure
Connect the measuring equipment to measurement port A.
Measure the water pressure with MV 41 closed.
Rated value of water inlet pressure: 0.90–1.40 bar
If it deviates from the rated value, the water inlet pressure must be adjusted.
– Adjusting the reduced water inlet pressure
Pull out the knurled nut on the pressure reducing valve (61).
Turn the knurled nut to set the water pressure to the desired value
(clockwise: +, counter-clockwise: –).
Push the knurled nut back in.
MV41
Measurement port A
MV41
Pressure reducing
valve
Measurement
port A
Check / adjustment:
– Checking the degassing pressure
Connect the measuring equipment to the measurement port D.
Measure the degassing pressure.
Desired degassing pressure: –0.81 to –0.85 bar.
If it deviates from the desired value, the degassing pressure must be adjusted.
– Adjusting the degassing pressure
Enter the CALIBRATION menu, select and start the option CAL. DEGAS. PRESSURE
(see Part 7: Calibrating the degassing pressure on page 5-12).
Note
If the degassing pressure is changed, make sure to check the loading
pressure and adjust if necessary.
If 4008 devices are operated at high-altitude sites (observe sea level), the specified degassing pressure
cannot be reached. The calibration process has to be performed as follows:
The setting for the degassing pump has to be increased in increments from a low speed until no significant
increase of the negative pressure can be detected anymore.
This setting can also be stored.
91
117
104
Measurement
port D
Measurement
port D
118
Measurement port D
Check / adjustment:
– Checking the loading pressure of the balancing chamber
Connect the measuring equipment to the measurement port B.
Measure the loading pressure of the balancing chamber.
Rated value of the balancing chamber loading pressure: 1.2–1.3 bar
From 4008 B EC495, 4008 H EC295, 4008 S EC275 (change to the pressure spring in the loading
pressure valve):
The loading pressure of the balancing chamber has been increased to 1.45 bar ±0.05 bar.
If it deviates from the rated value, the loading pressure of the balancing chamber must be adjusted.
– Adjusting the loading pressure of the balancing chamber
Use the loading pressure valve (65) to adjust the loading pressure to the rated value.
Turning the adjusting screw clockwise will increase the loading pressure.
Note
During the balance chamber fill phase, the loading pressure will drop to
approximately 1.0 bar.
Note
If the loading pressure was changed, make sure to check the degassing
pump pressure and readjust, if necessary.
Loading pressure
valve 65 Measurement
port B
MV41
Measurement
port B
Check / adjustment:
– Checking the pressure of the flow pump
Connect the measuring equipment to the measurement port C.
Shut off the water supply. Water alarm, the balancing chamber stops.
Measure the pressure of the flow pump.
The desired pressure value of the flow pump depends on the loading pressure set:
Loading pressure: Flow pump pressure:
1.2 to 1.3 bar 1.9 to 2.1 bar
1.45 ±0.05 bar 2.2 ±0.05 bar
If it deviates from the rated value, the pressure of the flow pump must be adjusted.
– Adjusting the pressure of the flow pump
Use the relief valve (78) to adjust the rated value.
Measurement port C
Relief valve 78
4008 B/S
91
Relief valve 78
117
104
Measurement port C
118
4008 H
Measurement port C
Relief valve 78
Note
Incorrect measurement result when using a scale
The volume of fluid can be determined with a scale only if the specific
gravity of the fluid is known.
– Remove all concentrates.
– Measure weight with dialysis water (permeate).
Note
Incorrect measurement result when using an incompatible
graduated cylinder
– Use a graduated cylinder with a maximum tolerance of ±0.5%.
Check / adjustment:
– Checking the UF pump volume:
Remove the outlet tube / UF (2) of the UF pump (3) from the T-piece (1) and close the T-piece.
Place the outlet tube / UF (2) in the graduated cylinder.
Go to the CALIBRATION service menu and select and run the ADJ. UF-PUMP VOLUME 1 option
(see Part 6: Adjusting the UF pump volume on page 5-11).
Desired value: 60 strokes = 60 ml ±0.5 ml (59.6 g ±0.5 g dialysis water)
– Adjusting the UF pump:
Remove the protective cap.
Loosen the lock nut.
Change the delivery volume using the adjusting screw (turning the adjusting screw clockwise reduces,
turning it counter-clockwise increases the stroke volume).
Retighten the lock nut.
Verify the delivery volume.
4008 B/S
4008 H
Measuring equipment: UMED, HMED, or pressure gage (e.g., 0–1 bar, accuracy ±1%) and
syringe. For measurement setup see the following illustrations.
Check / adjustment:
– Connect the test equipment as illustrated below.
– Select the DIAGNOSTICS service menu and then go to:
READ INPUTS
READ DIGITAL INPUTS
CPU1: RD DIGITAL INP
I: CPU1_PSW_V102 or I: CPU1_PSW_V104
– Activate the audible alarm by pressing the ALARMTONE MUTE key
(depressurized: audible alarm on).
– Under this menu option, the solenoid valves 102 and 104 are closed.
– Use the syringe to create a pressure of 0.7 bar.
– Use artery forceps to clamp the tube (1) so that a pressure of 0.7 bar continues to act on the pressure
switch.
– Check the switching point by means of the audible alarm.
Rated values:
Envec pressure switch: 0.68-0.72 bar
4008 B/S
General information on the The check can be performed by means of a multimeter or the
bibag® pressure DIAGNOSTICS service menu.
transducer
To read in the corresponding signal, select the menu item
I: CPU1_BIBAG_PSW menu item must be selected in the diagnostics
program.
Calibration of the pressure switch is not possible. The pressure switch
is set to a fixed position and can be checked in the DIAGNOSTICS
service menu.
The audible signal can be activated by pressing the ALARMTONE
MUTE key.
Options in the 1. The DIAGNOSTICS service menu can only be accessed in Service
DIAGNOSTICS service mode.
menu
– The device must be switched off.
– Set the Service switch to ON (up).
– Turn the device on.
Wait briefly until the Service mode screen is displayed.
– Use the and keys to select DIAGNOSTICS.
Confirm with Conf key.
2. In the DIAGNOSTICS service menu, continue as follows:
– Use the and keys to select DIAGNOSTICS.
Confirm with Conf key.
– Use the and keys to select READ INPUTS.
Confirm with Conf key.
– Use the and keys to select READ DIGITAL INPUTS.
Confirm with Conf key.
– CPU1: Use the and keys to select RD DIGITAL INP.
Confirm with Conf key.
– Use the and keys to select I:CPU1_BIBAG_PSW.
Confirm with Conf key.
– The active levels 0000 or 1111 are still shown on the display.
Note
In the menu item I:CPU1_BIBAG_PSW, the active level is identified by
1111.
The active level of the circuit output is LOW (< 1 V).
Warning
For adjusting the level detector, the device must be in "Calibration"
mode. The ambient temperature should be between 15 °C and 35 °C.
Warning
The calibration using the level detector calibration set is only applicable
to devices running with Fresenius tubing systems.
Observe the "use by" date!
Adjusting the level detector LD22, using the set for the level detector calibration
(see also adjustment instructions)
The device must be in Service mode and the bridge J1 / PCB LP450 in the calibration position.
Fill the hemispherical recesses of the adjusting block with grease and use a spatula to remove any excess
grease so that only the hemispherical recesses are completely filled with grease.
Introduce the greased reference block into the drip chamber holder (angled edges first). Ensure that the
ultrasonic sensors click correctly into place in the hemispherical recesses of the block. The reference
block must not touch the walls of the holder but hang freely between the sensors.
Turn the potentiometer 1 and then the potentiometer 2 on PCB LP450 clockwise until LED D5 and
LED D10 on PCB LP450 have turned dark.
Slowly (caution: allow for response time) turn the potentiometer 1 on PCB LP450 counterclockwise, until
the LED D5 on PCB LP450 lights up.
Slowly (caution: allow for response time) turn the potentiometer 2 on PCB LP450 counterclockwise, until
the LED D10 on PCB LP450 lights up.
Take the reference block out of the drip chamber and completely remove the grease from the drip
chamber holder, using only lint-free cloth and permissible disinfectants.
Check:
Fill the hemispherical recesses of the test block with grease and use a spatula to remove any excess
grease so that only the hemispherical recesses are completely filled with grease.
Insert the greased reference block into the drip chamber holder. Ensure that the ultrasonic sensors click
correctly into place in the hemispherical recesses of the block. The reference block must not touch the
walls of the holder but hang freely between the sensors.
Both LED D5 and LED D10 must be illuminated. If one or two of the LEDs is not illuminated, repeat the
adjustment process.
Take the reference block out of the drip chamber and completely remove the grease from the drip
chamber holder, using only lint-free cloth and permissible disinfectants.
Potentiometer P1
LED D5
Potentiometer P2
LED D10
Jumper J1
Potentiometer P5
Luer-lock adapter
Fluid level
Alternative: Adjusting the level detector without the set for the level detector calibration
Check / adjustment:
– Adjusting the ultrasonic detector
Install the measurement set-up before checking / adjusting the level detector. Do not yet place the line
in the occlusion clamp.
Jumper J1 / PCB LP450 set to calibration.
Fill the bubble catcher. The fluid level must be set to approx. 10 mm above the top edge of the sensor
holder.
Turn potentiometer 1 and potentiometer 2 on PCB LP450 clockwise, until the LED D5 and LED D10
on PCB LP450 are dark.
Slowly (caution: allow for response time) turn the potentiometer 1 on PCB LP450 counterclockwise,
until the LED D5 on PCB LP450 lights up.
Slowly (caution: allow for response time) turn the potentiometer 2 on PCB LP450 counterclockwise,
until the LED D10 on PCB LP450 lights up.
After completion of the calibration procedure, set the jumper J1 / PCB LP450 back to the operation
position.
– Testing
Lowering the level in the bubble catcher: An alarm must be generated.
Raising the level in the bubble catcher again: It must be possible to clear the alarm; both LEDs are off.
– Checking the venous occlusion clamp (see fig.)
Place the tube into the venous line clamp.
Open the clamp and generate pressure of around 2 bar with syringe.
Close the clamp.
Ensure that the pressure does not drop by more than 0.1 bar within 3 minutes.
– Adjusting the optical detector
Use the 1-layer gray filter.
Diagnostics menu, read digital inputs CPU1, menu option I: CPU1_OD_IN.
Install the 1-layer gray filter; close the flap.
Slowly turn potentiometer P5 on PCB LP450 clockwise, until the UF display indicates 1111.
Slowly turn potentiometer P5 counterclockwise, until the display jumps to 0000. Continue to turn the
potentiometer counterclockwise for half a turn.
Avoid incident light from external sources.
5 Calibration program
Key
CALIBRATION Conf
LM
DIAGNOSTICS
LM
MISCELLANEOUS
LM
SETUP MENU
LM
Adjusting the blood pump stop alarm (see Without menu display: Adjusting the
(blood pump or HDF pump) blood pump stop alarm (blood pump or HDF
pump) on page 5-5)
Calibrating the Single Needle blood pump (see Without menu display: Calibrating the
Single Needle blood pump (SN pressure)
(option) on page 5-10)
Adjusting the current rise pulse (see Adjusting the current rise pulse: on
page 5-13)
Adjusting the Hall sensor in the heparin pump (see Without menu display: Adjusting the Hall
sensor in the heparin pump on page 5-40)
In the Calibration, Diagnostics, Setup, and Miscellaneous programs, the function of the keys differs
between 4008 B and 4008 H/S systems.
In the description of the steps, the keys to be used on 4008 H/S systems are shown in brackets.
Main menu
CALIBRATION
Key
Conf
LM
Key
CAL. ART. P_MODULE see part 2
Conf
LM
Key
CAL. VENOUS PRESSURE see part 3
Conf
LM
Key
CAL. VEN. P_MODULE Conf see part 4
LM
Key
CALIB. (B)-PUMP-RATE Conf see part 5
LM
LM
Key
CAL. DEGAS. PRESSURE Conf
see part 7
LM
Key
CAL. FLOW 300 ml/min Conf
see part 8
LM
Key
CAL. FLOW 500ml/min see part 9
Conf
LM
Key
CAL. FLOW 800ml/min Conf
see part 10
LM
Key
CALIB. TEMPERATURE Conf
see part 11
LM
LM
Key
CALIB. CONDUCTIVITY Conf see part 13
LM
Key
CAL. DIAL. PRESSURE Conf
see part 14
LM
Key
CALIBRATE BLD Conf see part 15
LM
Key
CALIB. BIBAG VALUES Conf
see part 16
(option)
LM
Key
RESET FAILURE RECORD Conf see part 17
LM
Key
NOVRAM Conf see part 18
LM
Key
BACK TO MAIN MENU ? Conf
LM
Note
Before calibrating the hydraulics, remove any existing precipitate by
running an appropriate disinfection program.
Key
Conf
Note
Pressure gage accuracy: ±1% of the measured value.
Set the hex switch in the module (PCB LP 624, pos. 1) to position F.
Should the error message E02 appear on the blood pump display, clear the message by pressing the
Start/Stop key.
Key
Conf
Select key
art.press: XXX mmHg (Esc)
Syringe connected to
external reference meter,
connect to arterial pressure
transducer
Without menu display: Adjusting the blood pump stop alarm (blood pump or HDF pump)
Set the hex switch in the module (PCB LP 624, pos. 1) to position B.
Should the error message E02 appear on the blood pump display, clear the message by pressing the
Start/Stop key.
DATA STORED
Note
Pressure gage accuracy: ±1% of the measured value.
Key
Conf
Select key
ven.press: XXX mmHg (Esc)
ven.press: ±0 mmHg
Select key
(Esc)
Note:
Check zero point and slope; repeat
the procedure if necessary.
Note:
If adjustments are made to the level
detector, the CAL. VENOUS
PRESSURE menu item must be
run.
Potentiometer P3
Potentiometer P4
CALIB. (B)-PUMP-RATE
Key
Conf
LM
Key
CALIB. SN. BP-RATE see part 5.2
Conf
LM
LM
Key
Back to menu ? Conf
LM
Key
Conf
LM
Set a rate of ≥ 550 ml/min* on
Key the arterial blood pump
art. BP-Rate CHECK Conf (B)P Rate=550 ml/min
Note: Set the tube diameter to 8 mm before starting the calibration procedure and press Start/Stop on the blood pump.
* The BP rate of 550 ml/min is a default value! It can be changed with (+/-).
Part 5.2: Calibrating the Single Needle blood pump rate (option)
Key
Conf
Using a syringe, set the pressure at the P-SN
connector according to the set stroke volume Key Select key
CALIB. SN. BP-RATE Conf (B)P Rate=550 ml/min (Esc)
(see table).
LM
Set a rate of ≥ 550 ml/min* on
Key the SN blood pump
SN. BP-Rate CHECK Conf (B)P Rate=550 ml/min
Note: Set the tube diameter to 8 mm before starting the calibration procedure and press Start/Stop on the blood pump.
* The BP rate of 550 ml/min is a default value! It can be changed with (+/-).
Upper switching point 110 130 150 172 195 219 244 270 299
(mmHg) ±7 mmHg
CALIB. HDF-PUMP-RATE
Key
Conf
Note: This function is only possible if ONLINE HDF is activated with the DIP switch.
Without menu display: Calibrating the Single Needle blood pump (SN pressure) (option)
Set the hex switch in the module (PCB LP 624, pos. 1) to position F.
Should the error message E02 appear on the blood pump display, clear the message by pressing the
Start/Stop key.
Upper switching point 110 130 150 172 195 219 244 270 299
(mmHg) ±7 mmHg
Note
It may be necessary to change the Single Needle stroke volume: Press
the and Start/Stop keys at the same time.
Press the and keys to set the stroke volume and confirm with the
Start/Stop key.
Key
Optional Conf
Optional
Enter the number of strokes with
(+/-)
Key
UF-Pump 2 Conf
Override key (ALARM-
LM TONE MUTE)
Key
Back to menu ? Conf ACKNOWLEDGED
Select key
Press uf key (Esc)
Key
UF I/O uf pulses left = 60
Note: The remaining UF strokes are
Check the volume. Adjust the UF pump if necessary and indicated on the display.
repeat the procedure (see Chapter 4.3.5 on page 4-24). The number of preselected
strokes is shown on the UF
Goal display.
Note
Incorrect measurement result when using a scale
The volume of fluid can be determined with a scale only if the specific
gravity of the fluid is known.
– Remove all concentrates.
– Measure weight with dialysis water (permeate).
Note
Incorrect measurement result when using an incompatible
graduated cylinder
– Use a graduated cylinder with a maximum tolerance of ±0.5%.
Key
Conf
Select key
Adjust degas.-press. (Esc)
DATA STORED
The messages
– Fill program active
– Set flow on
Key
Conf
Select key
Flow (300) = XXX (Esc)
DATA STORED
Important:
If it is impossible to adjust the 300 – 500 – 800 flow volumes, or if problems caused by flow alarms occur
after the "calibrate flow" message has appeared, this can be caused by the adjusted current rise pulse.
Note:
(only for devices with advanced hydraulics)
The flow selected first is accompanied by the DIASAFE-filling act. message which is displayed for the
duration of 17 balancing chamber switchings.
MP1 1 V/cm
GND
MP8 2 V/cm
1. The system runs with regular balancing 2. The system is in the "Intrinsic clock pulse" mode.
chamber switching. Display: Flow (300) = 147
Display: Flow (300) = XXX – Turn the potentiometer P1 clockwise, until
– If necessary, correct the flow with the the system switches from "intrinsic clock
and (+/-) keys until the actual flow is pulse mode" to regular balancing chamber
approx. 300. switching (wait for approx. 10 seconds after
– Turn the potentiometer P1 counterclockwise each rotation!).
(wait for at least 10 seconds after each – Display: Flow (300) = XXX
rotation), until the system switches to – If necessary, correct the flow with the
"intrinsic clock pulse mode". and (+/-) keys until the actual flow is
– Display: Flow (300) = 147 approx. 300.
– Now turn the potentiometer P1 clockwise – Turn the potentiometer P1 counterclockwise
(wait for at least 10 seconds after each (wait for at least 10 seconds after each
half-rotation!), until the actual flow again rotation), until the system switches to
indicates approx. 300. "intrinsic clock pulse mode".
– Turn the potentiometer P1 clockwise for – Display: Flow (300) = 147
another 2 rotations. – Now turn the potentiometer P1 clockwise
(wait for at least 10 seconds after each
half-rotation!), until the actual flow again
indicates approx. 300.
– Turn the potentiometer P1 clockwise for
another 2 rotations.
After adjusting the current rise pulse, check and, if necessary, readjust the 300 / 500 / 800 flow settings.
Key
Conf
Select key
Flow (500) = XXX (Esc)
DATA STORED
Key
Conf
Select key
Flow (800) = XXX (Esc)
DATA STORED
CALIB. TEMPERATURE
Key
Conf
Key
Temp. ADJUSTMENT Conf
see part 11.1
LM
Key
Check TEMPERATURE see part 11.2
Conf
LM
Key
OCM TEMP.-Check Conf
see part 11.3
LM
Key
Back to menu ? Conf
LM
Note
Accuracy of the measuring instrument to be connected externally:
±0.2 °C.
Temp. ADJUSTMENT
Key
Conf adj. temp to 37 °C
Key
adj. temp to 37 °C Esc
XXXX ADC-digits
Use the +/- keys to adjust the
digital value in the UF Rate () XXXX ADC-digits
window
Key
ALARMTONE MUTE 7 8 9 0 Esc
4 5 6
DATA STORED
1 2 3 C Conf
After approx. 3 seconds
NTC3; 8 bit
Key
OCM TEMP.-Adjust Esc Controlled variable, 8 bit
Device in bypass
NTC3; 12 bit (with OCM option only)
NTC109; 8 bit,
Wait for the temperature to
stabilize (37 °C).
When NTC 109 is activated in the SETUP
Display in the UF Rate
() window
MENU
Key
ALARMTONE MUTE
DATA STORED
OCM TEMP.-Adjust
After approx. 3 seconds
XXXX ADC-digits
XXXX °C
XXXX ADC-digits
XXXX DAC-digits
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
Note
The hydraulics unit must be installed, the rear panel closed, and the
DIASAFE® plus covered.
Check TEMPERATURE
Key
Conf
Specify a temperature by
pressing the +/- keys
Key
ALARMTONE MUTE
ACKNOWLEDGED
NTC3; 8 bit
act. temp = 37.0 °C
NTC109; temperature
NTC109; 8 bit
XXXX ADC-digits
XXXX °C
XXXX ADC-digits
Alpha- Temperature
display setting
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
Part 11.3: Checking the dialysate temperature for the OCM option
OCM TEMP.-Check
Key
Conf
Specify a temperature by
pressing the +/- keys
Key
ALARMTONE MUTE
ACKNOWLEDGED
NTC3; 12 bit
OCM TEMP.-Check NTC3; temperature
NTC4; 12 bit
NTC4; temperature
XXXX ADC-digits
XX.XX °C
XXXX ADC-digits
XX.XX °C
Alpha- Temperature
display setting
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
CAL. MIXING-SYSTEM
Key
Conf
Key
Run-In PUMPS see part 12.1
Conf
LM
Key
DET. BAL.CHAMBER Vol Conf see part 12.2
LM
Key
cal. CONC.-PUMP-VOL Conf see part 12.3
LM
Key
CONC. PUMP VOL det. Conf see part 12.4
LM
Key
calib. BIC.-PUMP-VOL Conf see part 12.5
LM
Key
BIC. PUMP VOL det. see part 12.6
Conf
LM
Key
check CONC/BIC VOL. see part 12.7
Conf
LM
Key
Back to menu ? Conf
LM
The membrane pumps have to run-in so that the concentrate/bicarbonate pumps reach their operating
temperatures before the pump volume is verified.
Run-In PUMPS
Key
Conf
* Select key
MAND. FILLING active (Esc)
After 9 seconds
Select key
( Press ALARMTONE MUTE ) Press OVERRIDE (Esc)
Override key
(ALARMTONE MUTE)
Select key
mem puls left = 1000 (Esc)
Key
Conf
Select key
BC-Volume = 30.00 ml (Esc)
Override key
(ALARMTONE MUTE)
DATA STORED
Note
Incorrect measurement result when using a scale
The volume of fluid can be determined with a scale only if the specific
gravity of the fluid is known.
– Remove all concentrates.
– Measure weight with dialysis water (permeate).
Note
Incorrect measurement result when using an incompatible
graduated cylinder
– Use a graduated cylinder with a maximum tolerance of ±0.5%.
cal. CONC.-PUMP-VOL
Key
Conf
Select key
CONP-Vol = 0.828 ml (Esc)
Override key
(ALARMTONE MUTE)
DATA STORED
Key
Conf
* Select key
MAND. FILLING active (Esc)
After 9 seconds
Select key
conc pulses = 100 (Esc)
Override key
(ALARMTONE MUTE)
ACKNOWLEDGED
( )
Select key
Press ALARMTONE MUTE Press OVERRIDE (Esc)
Override key
(ALARMTONE MUTE)
Select key
conc puls left = 100 (Esc)
Notes:
The factory setting is 100 strokes. This setting can be changed by pressing
the (+/-) keys (depending on the graduated cylinder used). However, on
returning to CAL. MIXING-SYSTEM, the display will show the factory
setting again.
Note
Incorrect measurement result when using a scale
The volume of fluid can be determined with a scale only if the specific
gravity of the fluid is known.
– Remove all concentrates.
– Measure weight with dialysis water (permeate).
Note
Incorrect measurement result when using an incompatible
graduated cylinder
– Use a graduated cylinder with a maximum tolerance of ±0.5%.
calib. BIC.-PUMP-VOL
Key
Conf
Select key
BICP-Vol = 1.050 ml (Esc)
Override key
(ALARMTONE MUTE)
DATA STORED
Key
Conf
Select key
START MAND. FILLING? (Esc)
Override key
(ALARMTONE MUTE)
Select key
MAND. FILLING active (Esc)
After 9 seconds
Select key
bic pulses = 50 (Esc)
Override key
(ALARMTONE MUTE)
Select key
ACKNOWLEDGED (Esc)
Override key
(ALARMTONE MUTE)
Select key
bic pulse left = 50 (Esc)
Notes:
Note
Incorrect measurement result when using a scale
The volume of fluid can be determined with a scale only if the specific
gravity of the fluid is known.
– Remove all concentrates.
– Measure weight with dialysis water (permeate).
Note
Incorrect measurement result when using an incompatible
graduated cylinder
– Use a graduated cylinder with a maximum tolerance of ±0.5%.
Key
Conf
Select key
START MAND. FILLING? (Esc)
Override key
(ALARMTONE MUTE)
Select key
MAND. FILLING active (Esc)
After 9 seconds
Select key
conc pulses = 100 (Esc)
Override key
(ALARMTONE MUTE)
Select key
ACKNOWLEDGED (Esc)
( )
Select key
Press ALARMTONE MUTE Press OVERRIDE (Esc)
Override key
(ALARMTONE MUTE)
Select key
mem puls left = 50 (Esc)
Notes:
This test step verifies the concentrate or bicarbonate pump
volumes in accordance with the parameters entered for the mixing
system (dilution, BC volume, CONC, and BIC pump volume).
Note
Incorrect measurement result when using a scale
The volume of fluid can be determined with a scale only if the specific
gravity of the fluid is known.
– Remove all concentrates.
– Measure weight with dialysis water (permeate).
Note
Incorrect measurement result when using an incompatible
graduated cylinder
– Use a graduated cylinder with a maximum tolerance of ±0.5%.
CALIB. CONDUCTIVITY
Key
Conf
Key
CONDUCTIVITY Set see part 13.1
Conf
LM
Key
TEMP/COND COMP Calib see part 13.2
Conf
(with OCM option only)
LM
LM
Key
CONDUCTIVITY Check Conf see part 13.4
LM
Key
OCM COND.-Check Conf see part 13.5
(with OCM option only)
LM
LM
Key
Back to menu ? Conf
LM
CONDUCTIVITY Set
Key
Conf
Use to select the Concentrate setting field
If no key is
If no key is conc. set = –X% pressed for
conc. set = ±0% pressed for approx.
approx. 4 seconds
4 seconds
Use (+/-) to reduce the Use (+/-) to raise the concentrate
concentrate pump volume to get to a pump volume to get to a CD in the
CD in the lower display range. upper display range.
e.g.,:
ACKNOWLEDGED ACKNOWLEDGED
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
Part 13.2: Setting the temperature/conductivity compensation (with OCM option only)
Key
Conf act temp = XX.XX °C
Device in bypass
If no key is
conc. set = ±0% pressed for 7 8 9 0 Esc
approx.
4 seconds 4 5 6
Use +/- to reduce/increase the
concentrate pump volume until the
conductivity is within the window 13.9 to 1 2 3 C Conf
14.5 mS/cm.
ACKNOWLEDGED
XX.XX mS/cm
After approx. 3 seconds
XX.XX mS/cm
act temp = XX.XX °C
XX.XX mS/cm
Verify that the preset temperature of 35 °C
is achieved and that it has stabilized. XX.XX mS/cm
Check:
Display, LED display tolerance: ±0.5 °C
Alpha- Temperature
display setting
7 8 9 0 Esc
ACKNOWLEDGED
4 5 6
DATA STORED
Part 13.3: Calibrating the OCM pulse (with OCM option only)
Key
Conf
Key
Start PULSE calib. ? Esc
Key
OCM PULSE UP activ Esc
Key
OCM PULSE DOWN activ Esc
LED
ALARMTONE MUTE is flashing.
Info sound.
DATA STORED
Note
If the value is outside ±25, check the device’s conductivity measuring
system.
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
CONDUCTIVITY Check
Key
Conf
If no key is
conc. set = ±0% pressed for
approx.
4 seconds
Use the +/- keys to reduce the concentrate
pump volume.
e.g.,:
Key
ALARMTONE MUTE
ACKNOWLEDGED
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
Part 13.5: Checking the OCM conductivity (with OCM option only)
OCM COND.-Check
Key
Conf
OCM COND.-Check
If no key is
conc. set = ±0% pressed for
approx.
4 seconds
Use the +/- keys to reduce the concentrate
pump volume.
e.g.,:
Key
ALARMTONE MUTE
ACKNOWLEDGED
OCM COND.-Check
Check conductivity.
Check:
Reference meter, display, LED display
Key
Esc
CD cell 7; 12 bit
OCM COND.-Check CD cell 7
CD cell 110; 12 bit
CD cell 110
XXXX ADC-digits (OCM)
XX.XX mS/cm
XXXX ADC-digits (OCM)
XX.XX mS/cm
Alpha- Concentrate Bicarbonate Temperature
display setting setting setting
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
Key
Conf
Key
Esc
CD cell 7, compensated
act temp = XX.XX °C CD cell 7, compensation factor
CD cell 110; compensated
CD cell 110, compensation factor
XX.XX mS/cm
XXXX mS/(cm*°C)
XX.XX mS/cm
XXXX mS/(cm*°C)
Alpha- Concentrate Bicarbonate Temperature
display setting setting setting
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
Part 14: Calibrating the dialysate pressure (stainless steel pressure transducer)
Key
Conf
Key
DIALYSATE Pressure Conf
see part 14.1
LM
Key
TMP-Check Conf see part 14.2
LM
Key
PDIAL2 press-check Conf
see part 14.3
LM
Key
Back to menu ? Conf
LM
Note
Measuring instrument accuracy: ±1% of the measured value
DIALYSATE Pressure
Key
Conf
Select key
adj.zero.: +XXX mmHg (Esc)
Set flow on
Set flow on
Repeat the
procedure until 0 and Enter the value indicated by the external
–500 mmHg match Not reference meter by pressing the
the external OK (+/-) keys
reference meter
Override key
(ALARMTONE MUTE)
Override key
(ALARMTONE MUTE)
DATA STORED
TMP-Check
Key
Conf
Select key
(Esc)
Note:
Observe the venous
pressure!
Note:
Switch the flow on and off
again once in a while, to
maintain the operating
temperature of the pressure
transducer.
Select key
(Esc)
Note:
Observe the venous
pressure!
PDial2 Flow
display setting
7 8 9 0 Esc
4 5 6
1 2 3 C Conf
CALIBRATE BLD
Key
Conf
DATA STORED
Key
Conf
Override key
LM
(ALARMTONE MUTE)
Key Select key
BIBAG Cond.-Adjust Conf connect 104Ω ±0.1% (Esc)
DATA STORED
Connect the calibration resistor (104 kΩ) to
connector X108 (with advanced hydraulics: X13)
After approx. 3 seconds
Override key
(ALARMTONE MUTE)
ACKNOWLEDGED
Select key
connect 56.2Ω ±0.1% (Esc)
Override key
(ALARMTONE MUTE)
DATA STORED
LM
** Select key
Key
BIBAG Cond.-Check Conf act.cond: XX.X mS/cm (Esc)
LM
Key * With test plug: 25 °C
Back to menu ? Conf
** With test plug: 45.7 mS/cm or 84.5 mS/cm
LM
see also BIBAG Technical Manual
Key
Conf
Audible alarm
Select key
Are you sure ? (Esc)
Override key
(ALARMTONE MUTE)
ACKNOWLEDGED
Part 18: Initializing the NOVRAM, resetting the mandatory rinse, resetting the V84 malfunction
Audible alarm
Select key
Are you sure ? (Esc)
Override key
(ALARMTONE MUTE)
NOVRAM ACKNOWLEDGED
Key
Conf After approx. 3 seconds
Key
Init NOVRAM BACK TO MAIN MENU ?
Conf
LM
Audible alarm
Key Select key
Reset MAND. RINSE Conf Are you sure ? (Esc)
LM Override key
Audible alarm (ALARMTONE MUTE)
Select key
Key Are you sure ?
Reset CDS MAND. RINSE Conf
(Esc)
ACKNOWLEDGED
LM Override key
(ALARMTONE MUTE)
After approx. 3 seconds
ACKNOWLEDGED
BACK TO MAIN MENU ?
After approx. 3 seconds
Audible alarm
Key Select key
Reset V84 Are you sure ? (Esc)
Conf
LM Override key
(ALARMTONE MUTE)
Key
Back to menu ? Conf
ACKNOWLEDGED Note:
LM The NOVRAM menu item can be entered only if DIP switch 1
(PCB LP 631, field 2) is operated in Calibration mode.
After approx. 3 seconds For this, the switch must be operated once and then returned to its
initial position.
Without menu display: Adjusting the Hall sensor in the heparin pump
Setting the hall sensors 1 Briefly press the "" key (syringe plunger holder moves to its upper
end position)
Disconnect Hall sensor 2 (remove plug)
Insert syringe which is selected on the hex switch (figure 2)
Press and hold the "" key (until the syringe plunger holder is in the
lower end position).
Repeat the following steps until the end of the syringe's scale (0 ml)
and the bottom of the plunger are on the same level:
– Unscrew the screw of Hall sensor 1 and slightly shift the sensor
(undefined), fix the screw.
– To verify, first press the ““ key to move the syringe plunger holder
up, then press the ““ key to move the plunger holder down.
– Check whether the syringe pump automatically stops (lowest point)
and whether the bottom of the plunger and the end of the syringe's
scale (0 ml) are on the same level.
If the bottom of the plunger stops before having reached 0 ml, the
sensor must be unscrewed and adjusted downwards, if the syringe
pump moves beyond 0 ml, the sensor must be unscrewed and
adjusted upwards. If the error message "E55" occurs during the Hall
sensor adjustment, the message can be acknowledged by pressing
the "Start/Stop" key. If Hall sensor 1 was adjusted, the error
message "E55" must not occur anymore.
Each time Hall sensor 1 was adjusted,. repeat the check whether the
syringe pump automatically stops at the described position.
Briefly press the "" key (syringe plunger holder moves to its upper
end
After adjusting Hall senor 1, Hall sensor 2 has to be re-connecte
Setting the hall sensors 2 Repeat the following steps until the end of the syringe's scale (0 ml) and
the bottom of the plunger are on the same level and the bottom of the
plunger is pressed against the bottom of the syringe:
Press and hold the "" key until the syringe plunger holder is
approx. 2 cm before its lower end position
Manually turn the threaded spindle in delivery direction (one scale
line on the syringe)
Press and hold the "" key pressed (until the syringe plunger holder
has reached its lower end position)
Unscrew the screw of Hall sensor 2 and slightly shift the sensor
(undefined), fix the screw.
To verify, first press the ““ key to move the syringe plunger holder
up, then press the ““ key to move the plunger holder down.
Check whether the syringe pump automatically stops (lowest point),
whether the bottom of the plunger and the end of the syringe's scale
(0 ml) are on the same level and whether the bottom of the plunger
is pressed against the bottom of the syringe.
If the bottom of the plunger stops before having reached 0 ml, the
sensor must be unscrewed and adjusted downwards, if the syringe
pump moves beyond 0 ml, the sensor must be unscrewed and
adjusted upwards.
If the error message "E55" occurs during the Hall sensor adjustment,
the message can be acknowledged by pressing the "Start/Stop" key. If
Hall sensor 2 was adjusted, the error message "E55" must not occur
anymore.
Each time Hall sensor 2 was adjusted, repeat the check whether the
syringe pump automatically stops at the described position.
Briefly press the "" key (syringe plunger holder moves to its upper
end position)
After having adjusted both Hall sensors, apply screwlock to the
screws.
The hex switch position is not changed while the syringe type remains
the same.
Hall sensor 1
Hall sensor 2
6 Diagnostics program
The diagnostics program can be used to activate all of the device’s inputs and outputs.
Activation refers to CPU1 (PCB LP 631) and CPU2 (PCB LP 632), as well as to the output board
(PCB LP 634) and the input board (PCB LP 633).
This program allows technicians to make their own settings for testing error symptoms.
– READ INPUTS
– READ ANALOG INPUTS
– CPU1: RD ANALOG INP.
– CPU2: RD ANALOG INP.
– READ DIGITAL INPUTS
– CPU1: RD DIGITAL INP
– CPU2: RD DIGITAL INP
– WRITE OUTPUTS
– WRITE ANALOG OUTPUTS
– CPU1: WR ANALOG OUTP
– CPU2: WR ANALOG OUTP
– WRITE DIGIT. OUTPUTS
– CPU1: WR DIGIT. OUTP
– CPU2: WR DIGIT. OUTP
– INP/OUTP COMBINATION
– CPU1: COMBINATION
– CAN-COMPONENTS
– HPU
– ONLINE PLUS MODULE
In order to indicate the corresponding levels, all UF-seven-segment displays as well as the status indicator
(monitor), the external traffic light and the loudspeaker are used in the "READ DIGITAL INPUTS" menu.
The active signal state (which may correspond to both present and absent voltage) is indicated by 1111
on the UF displays, activated traffic light (status indicator), and audible signal.
The audible signal can be deactivated by pressing the alarm ALARMTONE MUTE key. If the audible
signal is deactivated, the ALARMTONE MUTE LED alarm flashes to alert the operator.
The audible signal allows the operator to evaluate the signal state without having to have a view of the
monitor. This is advantageous if measurements must be taken behind the device (e.g., hydraulics unit).
Note
In the diagnostics program, the signals are listed in the order of their
electric connection, i.e., in latch groups of 8 signals each, according to
the 8-bit data bus and according to the latch numbering on the circuit
diagram
(e.g., PCB 633: CS_LATCH0 – CS_LATCH6).
The signals are not grouped by relationship (e.g., all bibag® signals in a
row). The exceptions here are the signals for activating the solenoid
valves. If possible, these are listed in the menu by their number to
facilitate finding an individual valve because, usually, several valves
must be activated at the same time for troubleshooting purposes.
Since the signals are assigned to their respective connections (latch
groups), it is always possible to use the circuit diagram to locate the
respective signal in the menu, even if the signal name has changed.
Within one latch group, only one known signal is needed to find the
renamed signal by counting through the menu.
All voltage values specified can differ due to the tolerances in the
particular devices.
The item I:CPU1_KEY_TESTING is located under the menu item CPU1: RD DIGITAL INP.
This menu item can be used to test the keys.
The key pressed is shown on the alpha-display.
The UP, DOWN, ACKNOWLEDGED, SELECT and I/O keys have not been implemented, since their
function can be tested by selecting the corresponding menu.
CALIBRATION
LM
Key
DIAGNOSTICS Conf
LM
MISCELLANEOUS
LM
SETUP MENU
LM
WRITE OUTPUTS
INP/OUTP COMBINATION
Key
Conf
CAN-COMPONENTS
READ INPUTS
Key
Conf
Key
Back to menu ?
Conf
WRITE OUTPUTS
Key
Conf
Key
Back to menu ?
Conf
Key
Back to menu ?
Conf
Key
Back to menu ? Conf
Key
Back to menu ? Conf
Explanation:
UF Volume display: ADC value
UF Time Left display: Analog voltage (in 0.1 V), converted to the value at the input of PCB LP 633
Key
Prescribed venous blood pump rate
Confirm
0–8 V, IC 27/3, ADC 0–171
(Tube diameter 8 mm)
I: CPU1_PWR_WD
I: CPU1_ART_BPR_SET
I: CPU1_BPR_VEN I: CPU1_REF1
I: CPU1_PR_HDF I: CPU1_U_ACCU
Battery voltage
HDF blood pump rate e.g., 22 V, IC 28/26, ADC, e.g., 184
0–3.3 V, IC 27/1, ADC 0–72
(HDF switched on,
tube diameter: 8 mm)
I: CPU1_P_BIC
I: CPU1_P_ART I: CPU1_TEMP_DIAL3
Arterial pressure NTC temperature
0–10.5 V, IC 28/1, ADC 0–225 (Slot X107 / PCB LP 747)
0–12 V, IC 29/1, ADC 0–255
I: CPU1_BLL_DIM
I: CPU1_COND_SIGNAL3
Blood leak dimness voltage
5.0 V, IC 28/2, ADC 108 CD cell (slot X108 / PCB LP 747)
0–10.8 V, IC 29/2, ADC 0–231
I: CPU1_BLL
I: CPU1_FREE2
Blood leak voltage
5.0 V, IC 28/3, ADC 108 Not used, input open
IC 29/3
I: CPU1_COND_SIGNAL1
I: CPU1_U_BATT_SW
CD display
0–10.8 V, IC 28/4, ADC 0–231
Voltage for audible alarm if battery relay off
10.6 V, IC 29/4, ADC 110
I: CPU1_REF2
I: CPU1_FREE1
Key
Confirm
I: CPU1_TEMP_DIAL2
I: CPU1_COND_SIGNAL2
Not used
0 V, IC 29/28, ADC 0
Explanation:
UF Volume display: ADC value
UF Time Left display: Analog voltage (in 0.1 V), converted to the value at the input of PCB LP 632
I: CPU2_BPR_ART
I: CPU2_BLL
Arterial blood pump
0–9.6 V, IC 12/20, ADC 0–223 Blood leak voltage
5.0 V, IC 12/13, ADC 116
I: CPU2_P_ART
I: CPU2_+10V
Arterial pressure
0–10.6 V, IC 12/19, ADC 0–245 Reference voltage D-A converter / CPU2
10 V, IC 12/13, ADC 234
I: CPU2_P_VEN
I: CPU2_NC6
Venous pressure
0–11 V, IC 12/18, ADC 0–255 Not used
0 V, IC 12/13, ADC 0
I: CPU2_P_DIAL
Back to menu ?
Dialysate pressure
0–10 V, IC 12/17, ADC 0–231
Key
Confirm
I: CPU2_COND_SIGNAL
CD display CPU2: RD ANALOG INP.
0–10.8 V, IC 12/16, ADC 0–251
I: CPU2_TEMP_DIAL1
Temperature display
0–10.8 V, IC 12/15, ADC 0–251
I: CPU2_P_DIAL2
Explanation:
All UF displays on 0000, traffic lights red, yellow, green, off: Low level applied to latch on PCB LP 633
All UF displays on 1111, traffic lights red, yellow, green, on: High level applied to latch on PCB LP 633
If high level is applied, an audible alarm is active at the same time. This can be suppressed by pressing
the alarm ALARMTONE MUTE key. In this case, the alarm ALARMTONE MUTE LED is lit.
I: CPU1_COND_V84
I: CPU1_ABG_ON
V 84 CD detection
IC 19/2 Not used
Test by pulling off / short-circuiting the sensors IC 19/8
I: CPU1_ABG_ALARM
I: CPU1_LDA1
Not used
IC 19/9
Level detector alarm channel 1
IC 19/3
Test level detector alarm / alarm-free
I: CPU1_V43
Valve 43
IC 13/2
I: CPU1_OD_IN Test by opening / closing the valve
Valve 26
IC 13/3
I: CPU1_FL_SWITCH1 Test by opening / closing the valve
Float switch
IC 19/5
Trigger water deficiency in calibration mode: 0 I: CPU1_V24b
Open V 41 until water leaks out of the overflow tube: 1
Valve 24b
IC 13/4
Test by opening / closing the
valve
I: CPU1_CI
Balance chamber switching pulse
IC 19/6
I: CPU1_V24 I: CPU1_REED_RINSE
I: CPU1_UF_P1
I: CPU1_BIBAG_C
Acknowledgment from UF pump 1
IC 13/6 Microswitch 138 / connector
Test only possible in combination menu IC 14/5
Test by connecting / removing the cap
I: CPU1_LDA2
I: CPU1_PSW_V102
Level detector alarm channel 2
Concentrate pressure switch
IC 13/7
IC 14/6
Preparation:
Test by building up / relieving pressure at the
Level detector alarm-free, and set CLAMP_CTRL
pressure switch
(CPU1: WR DIGIT. OUTP) to 1.
Test: Trigger level detector alarm
I:CPU1_PSW_V104
I: CPU1_SUB_W_P
Bicarbonate pressure switch
Acknowledgment from UF pump 2 IC 14/7
IC 13/8 Test by building up / relieving pressure at the
pressure switch
I: CPU1_LC11
I: CPU1_PWR_OFF
IC 13/9
Power off
IC 14/8
I: CPU1_REED_BIC
I: CPU1_BIBAG
I: CPU1_LA32
Microswitch 137 / connector
IC 14/3 IC 15/2
Test by connecting / removing the BIBAG
I: CPU1_SW_ON_OFF I: CPU1_SN
I: CPU1_PWR_FAIL
I: CPU1_ADKS
Power failure detection
IC 15/4 Single Needle blood pump recognition
connected
IC 16/3
Connect / disconnect blood pump (only when device
I: CPU1_SHUNT_OUTP is switched off)
I: CPU1_LEV_SIGNAL I:CPU1_BPUS_VEN
I: CPU1_HEP_ALARM I: CPU1_ACKN_CONC
I: CPU1_ACKN_BIC
Not used
I: CPU1_BIB_LEVEL IC 20/8
Not used
IC 21/2
I: CPU1_SERVICE_MODE
HDF on
IC 21/3
Press the HDF On / Off key
I: CPU1_LEVEL_UP
Raise LD level
IC 20/4
Press the Raise level key I: CPU1_V102
Acknowledgment from valve 102
IC 21/4
Open / close valve
I: CPU1_LEVEL_DOWN
Lower LD level
IC 29/5
Preparation: I: CPU1_V104
Level detector alarm-free, set CLAMP_CTRL
(CPU1: WR DIGIT. OUTP) to 1 Acknowledgment from valve 104
Test: Press the Lower level key IC 21/5
Open / close valve
I: CPU1_ADS_SN I: CPU1_CSS_REED
Not used
IC 21/6
IC 20/6
I: CPU1_HEAT_CLK I: CPU1_LATCH7_FREE7
IC 21/7 Not used
IC 7/8
I: CPU1_BYP_REQ
I: CPU1_LATCH7_FREE8
IC 21/8
Not used
IC 7/9
I: CPU1_CLP_REQ
I: CPU1_DIP2: 00000000
I: CPU1_LATCH7_FREE2
Dip switch CPU1 / array 2
Not used PCB LP 631/IC 13/2 – 9
IC 7/3 The position of the dip switch is shown on the
alpha-display
(1: Dip switch "ON")
I: CPU1_LATCH7_FREE3
Not used
IC 7/4 I: CPU1_KEY_TESTING
I: CPU1_RCU_KEY_TEST
I: CPU1_LATCH7_FREE5
Touch panel test for
Not used RCU 4008 remote control
IC 7/6 The key pressed is shown on the alpha-display, the
associated LED keys light up
I: CPU1_LATCH7_FREE6
Key
Confirm
Explanation:
All UF displays on 0000, traffic lights red, yellow, green, off: Low level applied to latch on PCB LP 632
All UF displays on 1111, traffic lights red, yellow, green, on: High level applied to latch on PCB LP 632
If high level is applied, an audible alarm is active at the same time. This can be suppressed by pressing
the alarm ALARMTONE MUTE key. In this case, the alarm ALARMTONE MUTE LED is lit.
I: CPU2_NC3
I: CPU2_V43
Not used
IC 4/2 Acknowledgment from valve 43
IC 5/2, open / close valve
I: CPU2_UF_P1
I: CPU2_BL_ALARM
Acknowledgment from UF pump 1
Blood pump rate switching SN / HDF
IC 4/3
IC 5/3
I: CPU2_ACKN_AIRSEP
I: CPU2_PWR_OFF
Acknowledgment from air separation pump
IC 4/4 Power off
IC 5/4
I: CPU2_UF_P2 I: CPU2_FL_SWITCH+5V
Float switch
Acknowledgment from UF pump 2
IC 5/5
IC 4/5
Trigger water deficiency in calibration mode = 0
Open V 41 until water leaks out
of the overflow tube = 1
I: CPU2_CI
Balance chamber switching pulse
IC 4/6
I: CPU2_LDA1
I: CPU2_V24b
I: CPU2_LDA2 I: CPU2_BPST_VEN
I: CPU2_BPUS_VEN
I: CPU2_BPSB_VEN
Venous blood pump rotation stop
Venous blood pump stop confirmation IC 6/6
IC 5/8 Preparation:
Press the start / stop key on the venous blood pump Set SNST (CPU1: WR DIGIT. OUTP) to 1 and wait
for venous blood pump alarm field on.
Test: Clear the alarm field by pressing
start / stop key on the venous blood pump
I: CPU2_BPSB_ART
I: CPU2_LEVEL_UP
I: CPU2_BPST_ART Raise LD level
IC 6/8, press Raise level
Acknowledgment from arterial blood pump special
control
IC 6/3
Test by setting BPST_ART
I: CPU2_LEVEL_DOWN
(CPU1: WR DIGIT. OUTP) to 1/0
Lower LD level
IC 6/9
Preparation:
I: CPU2_BPUS_ART Level detector alarm-free, set CLAMP_CTRL
(CPU1: WR DIGIT. OUTP) to 1
Arterial blood pump rotation stop Test: Press the Lower level key.
IC 6/4
Arterial blood pump alarm field on = 0
(Delete with start / stop key on blood pump)
I: CPU2_RINSE
Concentrate reed contact
IC 7/2
Actuate rinse chamber / concentrate reed contact
I: CPU2_V147 I: CPU2_SERVICE_MODE
I: CPU2_REED_BIC
I: CPU2_HOT_RINSE
Bicarbonate reed contact
IC 7/4 Hot rinse switchover
Press rinse chamber / bicarbonate reed contact IC 8/4
Test by setting HOT_RINSE
(CPU1: WR DIGIT. OUTP) to 0/1
I: CPU2_PSW_104
I: CPU2_OD_OUT
Bicarbonate pressure switch
IC 7/5 Level detector optical system
Test by building up / relieving pressure at the IC 8/5, optical detector dark / light
pressure switch
I: CPU2_SNST
I: CPU2_V145
Single Needle control
IC 7/6 IC 8/6
I: CPU2_SHUNT_OUTP I: CPU2_24V_SW
I: CPU2_SN
I: CPU2_DIP2: 11000001
Dip switch CPU2, array 2
IC 10/2 – 9
Position of the dip switch is shown on the
alpha-display
(1: Dip switch "ON")
Back to menu ?
Key
Confirm
Explanation:
UF Rate display: DAC value (adjustable with: 4008 B: UF rate, 4008 H/S: +/-)
UF Time Left display: Analog voltage on PCB LP 634 in 0.1 V
O: CPU1_TEMP_SET
Prescribed temperature O: CPU1_STEUER_EP
IC 53/9
Degassing pump speed setting
0–5 V (X634R/C20, 0.2–10 V)
IC 53/3
0–4.4 V (X634L / between A, B, C 27
and A, B, C 28, 0–21 V)
O: CPU1_DAC_DIM
Dimness calibration voltage O: CPU1_STEUER_FP
IC 53/8
0–5 V (C634R/A11, 0–5 V) Flow pump speed setting
IC 53/2
0–4.4 V (X634L / between A, B, C 29
and A, B, C 30, 0–21 V)
O: CPU1_TEMP_ADJ
Temperature control balance voltage
IC 53/7 Back to menu ?
0–5 V (X634R/C21, 0–5 V)
Key
Confirm
O: CPU1_DAC_BLL
Blood leak calibration voltage CPU1: WR ANALOG OUTP
IC 53/6
0–5 V
(X634R/A12, 0–5 V)
O: CPU1_BIBAG_TEMP_AJ
Calibration voltage for temperature NTC
(Slot X107 / PCB LP 747)
IC 53/5
0–5 V (X634R/A13, 0–5 V)
Explanation:
UF Rate display: DAC value (adjustable with: 4008 B: UF rate, 4008 H/S: +/-)
UF Time Left display: Analog voltage on PCB LP 632 in 0.1 V
O: CPU2_TEMP_DET_ADJ
Key
Confirm
O: CPU2_DIAL_DET_ADJ
Dialysate pressure display detuning
IC 11/1, 0–10 V (X632/C 20, 0–10 V)
CPU2: WR ANALOG OUTP
O: CPU2_P_ADS_DET
Not used
IC 11/20, 0–10 V (X632/A 20, 0–10 V)
O: CPU2_PV_DET
O: CPU2_PA_DET
O: CPU2_COND_DET
CD display detuning
IC 11/19, 0–10 V (X632/A 21, 0–10 V)
Explanation:
UF Rate display:
0000 = not active
1111 = active (level PCB LP 634)
(adjustable with: 4008 B: UF rate, 4008 H/S: +/-)
O: CPU1_V24 O: CPU1_V34
Valve 24 Valve 34
IC 10/19 IC 12/16
O: CPU1_V24b
O: CPU1_V35
Valve 24b
Valve 35
IC 10/13
IC 12/15
O: CPU1_V26
O: CPU1_V36
Valve 26
Valve 36
IC 10/18
IC 12/14
O: CPU1_V130
O: CPU1_V37
Valve 130
Valve 37
IC 10/17
IC 12/13
O: CPU1_V30
O: CPU1_V38
Valve 30
Valve 38
IC 7/16
IC 12/12
O: CPU1_V31 O: CPU1_V41
Valve 31
IC 12/19 Valve 41
IC 7/13
(Valve automatically closes after a short time
to prevent water from overflowing)
O: CPU1_V32
Valve 32
IC 12/18
O: CPU1_V43 O: CPU1_V102
Valve 43 Valve 102
IC 7/15 IC 10/12
(The valve can only be activated
if a mandatory rinse is not required)
O: CPU1_V84
Valve 84 O: CPU1_V104
IC 7/18
Note: Rinsing is required Valve 104
following activation of valve V84 IC 7/14
(The valve can only be activated
if a mandatory rinse is not required)
O: CPU1_V86
Valve 86
O: CPU1_V126
IC 7/17
Not used
O: CPU1_V87
O: CPU1_V145
Valve 87
IC 10/15 Not used
O: CPU1_V91 O: CPU1_V147
O: CPU1_V99 O: CPU1_V173
Valve 99 Not used
IC 7/19
Comment: The valve will be closed on exiting
the menu item (back to "CPU1: WR DIGIT.
OUTP"). O: CPU1_AIR_SEP_PUMP
Air separation pump
IC 4/18, 19
(1111: clockwise)
Comment: The ASP stops running when this
O: CPU1_V100 menu item is exited
Valve 100
IC 4/17
Comment: The valve will be closed on exiting
the menu item (back to "CPU1: WR DIGIT.
OUTP").
O: CPU1_STOP_EP O: CPU1_FILL_ONE_CHAM
O: CPU1_STOP_FP
UF pump 1 activation
IC 4/14
(0/1 jump = 1 stroke) O: CPU1_CO_L:XXXXXXXX
O: CPU1_BI_L:XXXXXXXX
O: CPU1_SET_EN_UF2
Bicarbonate pump step number
Not used
Condition: Bicarbonate plug reed contact open
IC 4/12
O: CPU1_SET_FLOW_ON O: CPU1_ALARM_SOUND
Flow on
Audible alarm
Data word to Gal 23:
IC 5/18, 19 to 1: active
0000 0010 (active, V 32 open)
0000 0011 (inactive, V 31, 32 open)
O: CPU1_WARN_SOUND
Info sound
O: CPU1_EMPTIING_PRG IC 5/18 to 0, 19 to 1: active
Emptying program
Data word to Gal 23: O: CPU1_CLK_OVERLAP
0001 0010 (V 32, 35 open)
Balancing chamber dead time switchover
IC 5/17 to 0: 1 kHz, to 1: 2 Hz
O: CPU1_EN_IN_PULSE O: CPU1_CPU_OFF
O: CPU1_BC_PULSE O: CPU1_EN_PF_AT
O: CPU1_PIC_RA3
O: CPU1_BC_FUNCTION
Not used
Balancing chamber Gal switchover IC 13/16
IC 5/13
O: CPU1_PROG_LOG1
O: CPU1_STEPPER_PULS
Program logic 1, HDF pump
Intrinsic clock pulse IC 13/14
IC 5/12 (0: speed 200, 1: speed 400)
O: CPU1_FL_SWITCH_EN O: CPU1_PROG_LOG2
O: CPU1_CLR_ALARM O: CPU1_BPSST_ART
Battery test
IC 11/16
O: CPU1_BPSST_VEN
Heater blockage
IC 11/14
O: CPU1_BL_ALARM
O: CPU1_TL_YELLOW
O: CPU1_TL_RED
External traffic light yellow
External traffic light red IC 6/13
IC 11/12 The associated LED traffic light display for the
(The associated LED traffic light display for the display board is also activated
display board is also activated)
O: CPU1_TL_GREEN
O: CPU1_BPST_ART
External traffic light green
Arterial blood pump special control IC 6/12
IC 6/19 The associated LED traffic light display for the
display board is also activated
O: PUMP SLOWER
4008 H/S:
O: CPU1_BP_SLOWER
Back to menu ?
Blood pump rate adjustment –
(RCU 4008)
PCB LP 649/X4.5; PCB LP 922/X5.5; Key
PCB LP 924/X6.5 Confirm
O: CPU1_V_ADS
Not used
O: CPU1_NC_I
Not used
O: CPU1_NC_II
Not used
O: CPU1_NC_III
Not used
O: CPU1_ACKN_FLOW
Not used
Explanation:
UF Rate display:
0000 = not active
1111 = active (level PCB LP 632)
(adjustable with: 4008 B: UF rate, 4008 H/S: +/-)
O: CPU2_WD_RES
Watchdog reset
IC 24/18
Preparation: O: CPU2_UF_P_EN
Set WD_SET (CPU2: WR DIGIT. OUTP)
to 0, then back to 1 Enable UF pumps
Test: IC 24/13
Briefly set WD_RES to 0; when reset to 1,
WD relay is tripped
O: CPU2_CPU2_ALARM
Watchdog set
IC 24/17
Watchdog relay drops, 24 V is switched off, O: CPU2_UF_P2_CTRL
audible signal is sounded.
To reset the device, switch it off and UF pump 2 activation
back on again. IC 24/11, 0/1 jump = 1 stroke
Otherwise calibration is not possible. Preparation: Set CPU2_UF_P_EN to 1
O: CPU2_V24_EN
O: CPU2_4066_ENABLE_1
Enable V 24
IC 24/16 Analog switch for P_ADS_DET
Preparation: Switch on V 24 IC 20/13 (X632/A20)
(CPU1: WR DIGIT. OUTP)
Test: Switch valve on / off with V 24_EN
O: CPU2_4066_ENABLE_2
Enable V 24b
IC 24/15
Preparation: Switch on V 24B
O: CPU2_4066_ENABLE_3
(CPU1: WR DIGIT. OUTP)
Test: Switch valve on / off with V 24B_EN Analog switch for PV_DET
IC 20/6 (X632/C18)
O: CPU2_4066_ENABLE_4 O: CPU2_CLAMP_CTRL
O: CPU2_4066_ENABLE_5
O: CPU2_NC5
Analog switch, not used
IC 27/13 Not used
X632/B25
O: CPU2_4066_ENABLE_6 O: CPU2_NC7
Analog switch for COND_DET Not used
IC 20/5 (X632/A21) X632/B10
O: CPU2_4066_ENABLE_7 O: CPU2_BLL_DET
Analog switch for BLL_DIM
Blood leak detector detuning
IC 27/6
X632/A25
O: CPU2_4066_ENABLE_8 O: CPU2_SN_EN
Analog switch for BLL
Enable Single Needle
IC 27/12
X632/C19
O: CPU2_SN_ART O: CPU2_NC10
Arterial Single Needle control Not used
X632/A15 X632/B4
O: CPU2_LDSA O: CPU2_V26
Audible level detector alarm weakened Valve 26
X632/C16 X632/A6
Preparation:
Level detector alarm-free, and set
CLAMP_CTRL
(CPU1: WR DIGIT. OUTP) to 1
Test: Set LDSA to 1 to close the clamp on the O: CPU2_V42
level detector
Not used
X632/C4
O: CPU2_ODSA
O: CPU2_VENT_DSAFE O: CPU2_LED6
O: CPU2_LED7
O: CPU2_EM_HEAT_OFF
PCB LP 632 LED 7
Heater relay IC 21/13
X632/A9
After having been activated, the relay switches
off again for safety reasons
O: CPU2_LED8
Not used
X632/B9
Back to menu ?
O: CPU2_NC9
Key
X632/C6 Confirm
IC 29/13
O: CPU2_LED1
O: CPU2_LED2
O: CPU2_LED3
O: CPU2_LED4
O: CPU2_LED5
PCB LP 632 LED 5
IC 21/15
Explanation:
UF Volume display: Acknowledgment / input (the 3 traffic light LEDs are also lit up with 1111)
UF Rate display: Activation / output (adjustable with: 4008 B: UF rate, 4008 H/S: +/-)
CPU1: COMBINATION
CPU1_COMBI: V104
Key Valve 104
Confirm Activation PCB LP 634 / IC 7/14
Acknowledgment PCB LP 633 / IC 21/5
(The valve can only be activated if a
CPU1_COMBI: V24 mandatory rinse is not required)
Valve 24
Activation PCB LP 634 / IC 10/19
Acknowledgment PCB LP 633 / IC 13/5
CPU1_COMBI: UF1_PUMP
UF pump 1
Activation PCB LP 634 / IC 4/14
CPU1_COMBI: V24b Acknowledgment PCB LP 633 / IC 13/6
(When setting from 0 to 1 = 1 stroke,
Valve 24b acknowledgment is a brief jump to 1)
Activation PCB LP 634 / IC 10/13
Acknowledgment PCB LP 633 / IC 13/4
CPU1_COMBI: AIR_SEP
CPU1_COMBI: V26
ASP activation / deactivation
Valve 26
Activation PCB LP 634 / IC 10/18
Acknowledgment PCB LP 633 / IC 13/3
Back to menu ?
Key
CPU1_COMBI: V43 Confirm
Valve 43
Activation PCB LP 634 / IC 7/15
CPU1: COMBINATION
Acknowledgment PCB LP 633 / IC 13/2
CPU1_COMBI: V102
Valve 102
Activation PCB LP 634 / IC 10/12
Acknowledgment PCB LP 633 / IC 21/4
(The valve can only be activated if a
mandatory rinse is not required)
ONLINE-PLUS-MODUL
READ ANALOG INPUTS Key
Key Conf
Conf
Key
WRITE OUTPUTS Key
Conf Back to menu ?
Conf
Key
Back to menu ?
Conf
Key
WRITE DIGIT. OUTPUTS Conf
O: ONL+_ONL1 (V193)
PCB LP 785 X1/6a
Valve Online 1 V193
Key
Back to menu ?
Conf
O: ONL+_ONL2 (V192)
O: ONL+_ONL3 (V191)
Key
Back to menu ? Conf
6.13 HPU
HPU
READ ANALOG INPUTS Key
Key Conf
Conf
Key
WRITE OUTPUTS Key
Conf Back to menu ?
Conf
Key
Back to menu ?
Conf
Key
WRITE DIGIT. OUTPUTS Conf
O: WTR_AIR_P
Compressor
185
Key
Back to menu ?
Conf
O: WTR_V_TEST
Test valve
V183
O: WTR_V39
O: WTR_RETENT_V
Retentate valve
V189
O: WTR_V_EVAC_INDI
O: WTR_V_EVAC_SOD
Evacuation valve
V188
Key
Back to menu ? Conf
7 Setup menu
7.2 Overview
CALIBRATION
LM
DIAGNOSTICS
LM
MISCELLANEOUS
LM
Key
SETUP MENU Conf
LM
SETUP MENU
Key
Conf
Key Key
SET ALARM/WARN TIME see part 1 ACTIV. STD UF-DATA Conf see part 12
Conf
Key Key
SETUP CLEANING PGM see part 2 SET STD. PRIME-TIME see part 13
Conf Conf
Key Key
SETUP DILUTION see part 3 SOUND I/O-SWITCH Conf
see part 14
Conf
Key Key
SET CONDUCT. LIMIT see part 5 SET RINSE-VOLUME Conf see part 16
Conf
Key Key
INFO SOUND ( C-PGM ) Conf
see part 6 T1-TEST AUTOSTART Conf
see part 17
Key Key
SET DIAL PARAMETERS Conf
see part 7 ONLINE plus SETTINGS Conf
see part 18
(option)
Key Key
DIALYSIS TIME Conf see part 8 SET CENTRAL-DELIVERY Conf see part 19
Key Key
CALC.CUMUL.BLOOD-VOL Conf see part 9 AutoOFF after AutoON see part 20
Conf
Key Key
AUTOM. SN-START Conf see part 10 Init. CAMUS-baudrate Conf see part 21
Key Key
ACTIV. MONIT_NTC109 see part 11 STORE DEFAULT VALUES see part 22
Conf Conf
Key
BACK TO MAIN MENU ? Conf
Key
Conf
Key
Set ART-AL DELAYTIME Conf
see part 1.1
LM
Key
Set VEN-AL DELAYTIME see part 1.2
Conf
LM
Key
Set FLOW-OFF W-TIME Conf
see part 1.3
LM
Key
Set UF-WARNING-TIME see part 1.4
Conf
LM
Key
Set MUTE-TIME Conf
see part 1.5
LM
Key
Back to menu ? Conf
LM
Key
Conf
Key
Art Al Delay = 5 s Select
Key
Override
DATA STORED
Key
Conf
Key
Ven Al Delay = 5 s Select
Key
Override
DATA STORED
Key
Conf
Key
Flow-Off-T = 30 min Select
Key
Override
DATA STORED
Set UF-WARNING-TIME
Key
Conf
Key
UF-Warn-Time = 10 min Esc
Key
Override
DATA STORED
Set MUTE-TIME
Key
Conf
Key
MUTE-Time = 1 min Select
Key
Override
DATA STORED
Key
Conf
LM
Key
Pgm COMBINATION see part 2.2
Conf
LM
Key
Back to menu ? Conf
LM
Note
The default values and the adjustable range for the cleaning times are
not indicated, as they depend on the particular device options.
CLEANING Times
Key
Conf
LM
Key
Hotrinse TIME see part 2.1.2
Conf
LM
Key
Disinfection TIME see part 2.1.3
Conf
LM
Key
Rinsing Free TIME see part 2.1.4
Conf
LM
Key
Hot-Disinf TIME see part 2.1.5
Conf
LM
Key
Mandatory Rinse TIME see part 2.1.6
Conf
LM
Key
CITRO-Mandat-Ri-Time see part 2.1.7
Conf
LM
LM
Key
Back to menu ? Conf
LM
Rinsing TIME
Key
Conf
Key
Rinsing Time = xxmin Select
Key
Override
DATA STORED
Hotrinse TIME
Key
Conf
Key
H-Rinse Time = xxmin Select
Key
Override
DATA STORED
Disinfection TIME
Key
Conf
Key
Disinf. Time = xxmin Select
Key
Override
DATA STORED
Key
Conf
Key
R.-Free Time = xmin Select
Key
Override
DATA STORED
Hot-Disinf TIME
Key
Conf
Key
H-Disinf Time= xxmin
Select
Key
Override
DATA STORED
Key
Conf
Key
M-Rinse Time = xxmin Select
Key
Override
DATA STORED
CITRO-Mandat-Ri-Time
Key
Conf
Key
CITRO-MRTime = xxmin Select
Key
Override
DATA STORED
INTEGRATED-HR Time
Key
Conf
Key
INT. HR-Time = xxmin Select
Key
Override
DATA STORED
Pgm combination
Key
Conf
Key
RINSE Pgm see part 2.2.1
Conf
LM
Key
HOTRINSE Pgm see part 2.2.2
Conf
LM
Key
DISINFECTION Pgm see part 2.2.3
Conf
LM
Key
Back to menu ? Conf
LM
RINSE Pgm
Key
Conf
HOTRINSE Pgm
Key
Conf
DISINFECTION Pgm
Key
Conf
Key
Select
(see Part 19: Setting the parameters for Central Delivery System on page 7-28)
SETUP DILUTION
Key
Conf
LM
(see Part 19: Setting the parameters for Central Delivery System on page 7-28)
SETUP DILUTION
CDS 1+34
After approx.
Key DATA STORED
Key 3 seconds
Canister 1+44 C ACKNOWLEDGED Override
Override LM
Key
Select
CDS 1+44 C
LM Key
Select
CDS 1+34
Key After approx.
Key DATA STORED 3 seconds
Canister 1+44 ACF ACKNOWLEDGED Override
Override LM
Key
Select
LM CDS 1+44 ACF
Key
Select
LM
(see Part 19: Setting the parameters for Central Delivery System on page 7-28)
SETUP DILUTION
Key
Conf
Warning
The operator or technician is informed about his duty of care to enter the
component parameters and settings correctly or to check them and to
set the CD alarm window to the expected conductivity value.
When using the programmable mixing ratio (dilution), make sure to use
the right concentrate.
Using a bibag® in combination with the freely adjustable dilution is not
allowed.
Only enter authorized, programmable mixing ratios (dilutions).
SETUP DILUTION
Key
Conf
LM
Key Key
VARIABLE SETTING Conf B-Comp[part]: 0.800 Select
Key
Override
ACKNOWLEDGED
After approx.
3 seconds
Key
RO[part]: 30.000
Select
Key
Override
ACKNOWLEDGED
After approx.
3 seconds
Key
HCO3-[mmol]: 25
Select
Key
Override
ACKNOWLEDGED
After approx.
3 seconds
Key
Na(B)[mmol]: 25 Select
Key
Override
DATA STORED
After approx.
3 seconds
Part 4: Setting up the HDF-DILUTION – only for the ONLINE HDF (option)
HDF-DILUTION
Key
Conf
Key
Conf
Key
Cd Limit: 12.8 mS/cm Select
Key
Override
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Conf
Key
SET Flow Parameter see part 7.1
Conf
LM
Key
SET Temp. Parameter Conf see part 7.2
LM
Key
SET Na/Bic Parameter see part 7.3
Conf
LM
Key
Back to menu ? Conf
LM
Key
Conf
Key
Flow[ml/min]: 500 Select
Key
Override
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Temp.[°C]: 37.0 Select
Key
Override
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Base Na+ 135mmol Select
Key
Override
ACKNOWLEDGED
After approx.
3 seconds
Key
Prescr. Na+ 135mmol Select
Key
Override
ACKNOWLEDGED
After approx.
3 seconds
Key
Bicarbonate: ±0mmol Select
Key
Override
DATA STORED
After approx.
3 seconds
DIALYSIS TIME
Key
Conf
Key
Select
CALC.CUMUL.BLOOD-VOL
Key
Conf
AUTOM. SN-START
Key
Conf
ACTIV. MONIT_NTC109
Key
Conf
Key
Conf
Key
Conf
Key
Prime-Time = 2min Select
Key
Override
DATA STORED
After approx.
3 seconds
SOUND I/O-SWITCH
Key
Conf
BPR/UFR-WARNING
Key
Conf
SET RINSE-VOLUME
Key
Conf
Key
RINSE-VOL: 1000 ml Select
Key
Override
DATA STORED
After approx.
3 seconds
T1-TEST AUTOSTART
Key
Conf
Key
Conf
Key
ONLINE plus HD Conf
see part 18.1
LM
Key
ONLINE plus HDF see part 18.2
Conf
LM
Key
ONLINE plus HF Conf
see part 18.3
LM
Key
ONLINE plus MISC. see part 18.4
Conf
LM
Key
Back to menu ? Conf
LM
ONLINE plus HD
Key
Conf
LM
Key
SET Rinsing Volume see part 18.1.2
Conf
LM
Key
Back to menu ? Conf
LM
Key
Conf
Key
Volume[ml]: 0 Select
Key
Override
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Volume[ml]: 1000 Select
Key
Override
DATA STORED
After approx.
3 seconds
Key
Conf
Key
SET UF-Volume F/R see part 18.2.1
Conf
LM
Key
SET Rinsing Volume Conf see part 18.2.2
LM
Key
SET Substit.-Volume see part 18.2.3
Conf
LM
Key
Back to menu ? Conf
LM
Key
Conf
Key
Volume[ml]: 500 Select
Key
Override
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Volume[ml]: 1000 Select
Key
Override
DATA STORED
After approx.
3 seconds
SET Substit.-Volume
Key
Conf
Key
Volume[l]: 12 Select
Key
Override
DATA STORED
After approx.
3 seconds
ONLINE plus HF
Key
Conf
LM
Key
SET Rinsing Volume see part 18.3.2
Conf
LM
Key
SET Substit.-Volume see part 18.3.3
Conf
LM
Key
Back to menu ? Conf
LM
Key
Conf
Key
Volume[ml]: 1000 Select
Key
Override
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Volume[ml]: 1000 Select
Key
Override
DATA STORED
After approx.
3 seconds
SET Substit.-Volume
Key
Conf
Key
Volume[l]: 20 Select
Key
Override
DATA STORED
After approx.
3 seconds
Key
Conf
Key
SET Reinf.-Volume see part 18.4.1
Conf
LM
Key
Back to menu ? Conf
LM
SET Reinf.-Volume
Key
Conf
Key
Volume[ml]: 240 Select
Key
Override
DATA STORED
After approx.
3 seconds
SET CENTRAL-DELIVERY
Key
Conf
LM Key
Select
Key After approx. After approx.
central Bic SETUP DILUTION 3 seconds
DATA STORED 3 seconds
Override
LM Key
Select
After approx. After approx.
central Acid Key SETUP DILUTION 3 seconds DATA STORED 3 seconds
Override
LM Key
Select
Key After approx. After approx.
central Acid + Bic Override
SETUP DILUTION 3 seconds DATA STORED 3 seconds
LM Key
Select
Key After approx. After approx.
centr acetate-supply SETUP DILUTION 3 seconds DATA STORED 3 seconds
Override
LM Key
Select
Key
Conf
Init. CAMUS-baudrate
Key
Conf
Key
Select
Key
Conf
Key
Press OVERRIDE Select
Press
the Override key
to store
the default values
DATA STORED
After approx.
3 seconds
SETUP MENU
Key
Conf
Key Key
SET ALARM/WARN TIME see part 1 ACTIV. STD UF-DATA Conf see part 13
Conf
Key Key
see part 3 SOUND I/O-SWITCH see part 15
SETUP DILUTION Conf Conf
Key Key
SET CONDUCT. LIMIT see part 5 BPR/UFR-WARNING Conf see part 17
Conf
Key Key
INFO SOUND ( C-PGM ) see part 6 SET RINSE-VOLUME Conf see part 18
Conf
Key Key
SET DIAL PARAMETERS see part 7 T1-TEST AUTOSTART Conf
see part 19
Conf
Key Key
CALC.CUMUL.BLOOD-VOL see part 8 ONLINE plus SETTINGS Conf
see part 20
Conf (option)
Key Key
AUTOM. SN-START see part 11 Init. CAMUS-baudrate see part 23
Conf Conf
Key
BACK TO MAIN MENU ? Conf
Key
Conf
Key
Set ART-AL DELAYTIME Conf
see part 1.1
LM
Key
Set VEN-AL DELAYTIME see part 1.2
Conf
LM
Key
Set FLOW-OFF W-TIME Conf
see part 1.3
LM
Key
Set UF-WARNING-TIME see part 1.4
Conf
LM
Key
Set MUTE-TIME Conf
see part 1.5
LM
Key
Back to menu ? Conf
LM
Key
Conf
Key
Art Al Delay = 5 s Esc
Key
ALARMTONE MUTE
DATA STORED
Key
Conf
Key
Ven Al Delay = 5 s Esc
Key
ALARMTONE MUTE
DATA STORED
Key
Conf
Key
Flow-Off-T = 30 min Esc
Key
ALARMTONE MUTE
DATA STORED
Set UF-WARNING-TIME
Key
Conf
Key
UF-Warn-Time = 10 min Esc
Key
ALARMTONE MUTE
DATA STORED
Set MUTE-TIME
Key
Conf
Key
MUTE-Time = 1 min Esc
Key
ALARMTONE MUTE
DATA STORED
Key
Conf
LM
Key
DEFAULT Cleaning Pgm see part 2.2
Conf
LM
Key
Back to menu ? Conf
LM
Note
The default values and the adjustable range for the cleaning times are
not indicated, as they depend on the particular device options.
CLEANING Times
Key
Conf
LM
Key
Hotrinse TIME Conf
see part 2.1.2
LM
Key
Disinfection TIME see part 2.1.3
Conf
LM
Key
Rinsing Free TIME see part 2.1.4
Conf
LM
Key
Hot-Disinf TIME see part 2.1.5
Conf
LM
Key
Mandatory Rinse TIME see part 2.1.6
Conf
LM
Key
CITRO-Mandat-Ri-Time see part 2.1.7
Conf
LM
LM
Key
Back to menu ? Conf
LM
Rinsing TIME
Key
Conf
Key
Rinsing Time = xxmin Esc
Key
ALARMTONE MUTE
DATA STORED
Hotrinse TIME
Key
Conf
Key
H-Rinse Time = xxmin Esc
Key
ALARMTONE MUTE
DATA STORED
Disinfection TIME
Key
Conf
Key
Disinf. Time = xxmin Esc
Key
ALARMTONE MUTE
DATA STORED
Key
Conf
Key
R.-Free Time = xmin Esc
Key
ALARMTONE MUTE
DATA STORED
Hot-Disinf TIME
Key
Conf
Key
H-Disinf Time= xxmin Esc
Key
ALARMTONE MUTE
DATA STORED
Key
Conf
Key
M-Rinse Time = xxmin Esc
Key
ALARMTONE MUTE
DATA STORED
CITRO-Mandat-Ri-Time
Key
Conf
Key
CITRO-MRTime = xxmin Esc
Key
ALARMTONE MUTE
DATA STORED
INTEGRATED-HR Time
Key
Conf
Key
INT. HR-Time = xxmin Esc
Key
ALARMTONE MUTE
DATA STORED
Key
Conf
(see Part 21: Setting the parameters for Central Delivery System on page 7-56)
SETUP DILUTION
Key
Conf
Key
+/– Esc
+/–
(see Part 21: Setting the parameters for Central Delivery System on page 7-56)
SETUP DILUTION
CDS 1+34
Key
Conf
+/– Key ALARM- After approx.
TONE MUTE DATA STORED 3 seconds
Key ALARM-
Canister 1+34 TONE MUTE ACKNOWLEDGED CDS 1+44 C
Key
Esc
+/–
+/–
CDS 1+44 ACF
Key
Esc
CDS 1+34
Key ALARM- After approx.
Key ALARM- TONE MUTE
DATA STORED 3 seconds
Canister 1+44 C ACKNOWLEDGED +/–
TONE MUTE Key
Esc
CDS 1+44 C
+/– Key
Esc
CDS 1+34
Key ALARM- After approx.
Key ALARM- TONE MUTE DATA STORED 3 seconds
Canister 1+44 ACF TONE MUTE ACKNOWLEDGED +/–
Key
Esc
+/– CDS 1+44 ACF
Key
Esc
Key
VARIABLE SETTING Conf see part 3.1
+/–
(see Part 21: Setting the parameters for Central Delivery System on page 7-56)
SETUP DILUTION
Key
Conf
Warning
The operator or technician is informed about his duty of care to enter the
component parameters and settings correctly or to check them and to
set the CD alarm window to the expected conductivity value.
When using the programmable mixing ratio (dilution), make sure to use
the right concentrate.
Using a bibag® in combination with the freely adjustable dilution is not
allowed.
Only enter authorized, programmable mixing ratios (dilutions).
SETUP DILUTION
Key
Conf
+/–
Key Key
VARIABLE SETTING Conf B-Comp[part]: 0.800 Esc
Key
ALARMTONE MUTE
ACKNOWLEDGED
After approx.
3 seconds
Key
RO[part]: 30.000
Esc
Key
ALARMTONE MUTE
ACKNOWLEDGED
After approx.
3 seconds
Key
HCO3-[mmol]: 25
Esc
Key
ALARMTONE MUTE
ACKNOWLEDGED
After approx.
3 seconds
Key
Na(B)[mmol]: 25 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Part 4: Setting up the HDF-DILUTION – only for the ONLINE HDF (option)
HDF-DILUTION
Key
Conf
Key
+/– Esc
Key ALARM- After approx.
HDF–POST–dilution TONE MUTE DATA STORED 3 seconds
Key
+/– Esc
Key
Conf
Key
Cd Limit: 12.8 mS/cm Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Key
Conf
Key
+/– Esc
Key ALARM- After approx.
Info Sound: OFF TONE MUTE DATA STORED 3 seconds
Key
+/– Esc
Key
Conf
Key
SET Flow Parameter see part 7.1
Conf
LM
Key
SET Temp. Parameter Conf
see part 7.2
LM
Key
SET Na/Bic Parameter see part 7.3
Conf
LM
Key
Back to menu ? Conf
LM
Key
Conf
Key
Flow[ml/min]: 500 Esc
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Temp.[°C]: 37.0 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Base Na+ 135mmol Esc
Key
ALARMTONE MUTE
ACKNOWLEDGED
After approx.
3 seconds
Key
Prescr. Na+ 135mmol Esc
Key
ALARMTONE MUTE
ACKNOWLEDGED
After approx.
3 seconds
Key
Bicarbonate: ±0mmol Esc
Key
ALARMTONE MUTE
ACKNOWLEDGED
After approx.
3 seconds
Key
Limit Na/Base 13mmol Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
CALC.CUMUL.BLOOD-VOL
Key
Conf
HAEMOGLOBIN UNIT
Key
Conf
OCM SETTINGS
Key
Conf
Key
OCM MEASUREMENT Conf
see part 10.1
LM
Key
OCM ZERO MEASUREMENT Conf
see part 10.2
LM
Key
OCM MEASURE DEL.TIME see part 10.3
Conf
LM
Key
OCM CLR CORR FACTOR Conf
see part 10.4
LM
Key
OCM BASELINE DIFF. Conf
see part 10.5
LM
Key
OCM INTEGRAL VALUE see part 10.6
Conf
LM
Key
OCM KT/V WARNLEVEL see part 10.7
Conf
LM
Key
Back to menu ? Conf
LM
OCM MEASUREMENT
Key
Conf
Note
If the OCM zero measurement is set to "OFF", the OCM option is
deactivated.
If the OCM option is reactivated ("ON"), an OCM pulse calibration must
be performed.
Key
Conf
Device Time
Key
Conf
Key
Delaytime: XX s Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Note
Do not change the default value.
Note
The 4008 H/S offers the theoretical possibility of defining a correction
value for all clearance measurements. This allows the device to be
adapted on site to incorporate new findings from ongoing scientific
research. According to the current standard of knowledge, this
correction value has to be set to "1" (factory setting) for Fresenius
polysulfone membranes. The indicated accuracy specifications are
valid only for Fresenius polysulfone membrane combined with this
correction value setting of "1".
Key
Conf
Key
ClrCorr factor:1.000 Esc
Display
shows the
default value
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Note
Do not change the default value.
Key
Conf
Key
Baseline Diff.:0.015 Esc
Display
shows the
default value
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Note
Do not change the default value.
Key
Conf
Key
Integral value: 55 Esc
Display
shows the
default value
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Kt/V Warnlevel: 85% Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
AUTOM. SN-START
Key
Conf
ACTIV. MONIT_NTC109
Key
Conf
Key
Conf
Key
Conf
Key
Prime-Time = 2min Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
SOUND I/O-SWITCH
Key
Conf
SET KEY-CLICK
Key
Conf
BPR/UFR-WARNING
Key
Conf
SET RINSE-VOLUME
Key
Conf
Key
RINSE-VOL: 1000 ml Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
T1-TEST AUTOSTART
Key
Conf
Key
Conf
Key
ONLINE plus HD Conf
see part 20.1
LM
Key
ONLINE plus HDF see part 20.2
Conf
LM
Key
ONLINE plus HF Conf
see part 20.3
LM
Key
ONLINE plus MISC. see part 20.4
Conf
LM
Key
Back to menu ? Conf
LM
ONLINE plus HD
Key
Conf
LM
Key
SET Rinsing Volume see part 20.1.2
Conf
LM
Key
Back to menu ? Conf
LM
Key
Conf
Key
Volume[ml]: 0 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Volume[ml]: 1000 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Key
Conf
Key
SET UF-Volume F/R see part 20.2.1
Conf
LM
Key
SET Rinsing Volume Conf see part 20.2.2
LM
Key
SET Substit.-Volume see part 20.2.3
Conf
LM
Key
Back to menu ? Conf
LM
Key
Conf
Key
Volume[ml]: 500 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Volume[ml]: 1000 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
SET Substit.-Volume
Key
Conf
Key
Volume[l]: 12 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
ONLINE plus HF
Key
Conf
Key
SET UF-Volume F/R see part 20.3.1
Conf
LM
Key
SET Rinsing Volume Conf see part 20.3.2
LM
Key
SET Substit.-Volume see part 20.3.3
Conf
LM
Key
Back to menu ? Conf
LM
Key
Conf
Key
Volume[ml]: 1000 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Key
Conf
Key
Volume[ml]: 1000 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
SET Substit.-Volume
Key
Conf
Key
Volume[l]: 20 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Key
Conf
Key
SET Reinf.-Volume see part 20.4.1
Conf
LM
Key
Back to menu ? Conf
LM
SET Reinf.-Volume
Key
Conf
Key
Volume[ml]: 240 Esc
Key
ALARMTONE MUTE
DATA STORED
After approx.
3 seconds
Note
Central delivery is not available for the 3mix option.
SET CENTRAL-DELIVERY
Key
Conf
Key
Conf
Init. CAMUS-baudrate
Key
Conf
Key
Conf
Key
Press ALARMTONE MUTE Esc
DATA STORED
After approx.
3 seconds
8 Miscellaneous
CALIBRATION
LM
DIAGNOSTICS
LM
Key
MISCELLANEOUS Conf
LM
SETUP MENU
LM
Main menu
Service switch to
MISCELLANEOUS Treatment
Key Key
Conf Conf
Key Key
SYSTEM CLOCK Conf see part 1 SW-VERSION-NUMBER Conf see part 2
LM LM
Key Key
BACK TO MAIN MENU ? Conf SYSTEM CLOCK Conf see part 1
LM LM
Key
BACK TO MAIN MENU ? Conf
LM
Part 1
SYSTEM CLOCK
Key
Conf
Time Date
Key
XX:XX:XX XX:XX:XXXX Select
(Esc)
Press the (+/-) keys to set
the flashing hours
Key
Conf
Key
Conf
Key
Override (ALARMTONE MUTE)
DATA STORED
After approx.
3 seconds
Part 2
SW-VERSION-NUMBER
Key
Conf
Key
CPU_1-Ver-No.: X.XX Select
(Esc)
LM (+/–)
Key
CPU_2-Ver-No.: X.XX Select
(Esc)
LM (+/–)
Key
MDC-Ver-No.: X.XX Select
(Esc)
LM (+/–)
Key
HPU-Ver-No.: X.XX Select
(Esc)
LM (+/–)
Key
Online-Vers.: X.XX Select
(Esc)
LM (+/–)
Key
BVM-Ver-No.: X.XX Select
(Esc)
LM (+/–)
9.1
+24V_SW
Netzteil
serielle 4008
+24V_SW +24V_EM
TM-EN
+5V +12V Hydraulik Einschübe
Hydraulic Modules Heater Board
WD-Set 230V 24V_US
WD-Reset +5V
LP 632 Watchdog +12V
CPU 2
12A-2016
+24V_SW
+24V_EM
Circuit diagrams and circuit descriptions
9-1
Chapter 9: Circuit diagrams and circuit descriptions
Chapter 9: Circuit diagrams and circuit descriptions
RS232 CPU
RS232
RAM
RS232 CPU
X639
TM-EN
U-ACCU
24V_US
Adress-/Databus
Block diagram of screen 4008 H/S
MDC-Board TFT-Display
Display board
12A-2016
LP922 (4008S) ROM
LP 924 (4008H)
RS232
Video
RAM DRAM
9-3
Chapter 9: Circuit diagrams and circuit descriptions
9-4
9.4
Rate: ml/min Rate: ml/min Rate: ml/h
(Ø: mm) (Ø: mm) (Bolus: ml)
( :h.min) Pven.
Bolus
Start Start
Stop Stop Rate
Start
Stop
XA1 UF pump 22
LP 928/774
Kurz- Flow pump 21
CAN-BUS Verteiler
schluß- Degassing pump 29
Art SN CAN-BUS Distribution
Diagnose teil Air separation pump 97
Diagnosis Shunt XA2 Valves V24 V24B V26 V30 V41 V43 V84
LP 624 LP 624 LP 643-3 LP 450
interlock X M9 X 1/774 V91 V112
X 348 X 348 LP 950 LP 450-2
X 6/928 XA3 Balancing chamber valves V31–V38
ST 1 X 351
Valves V86 V87
XA4 CDS V99 V100 V102 V104 V126 V130
XA5 V183 V188 V189 V9
Compressor 185
X DIAG X CAN X 637A X 348A X 348V X 350 X 351 X SHUNT CO1 Concentrate pump
Sensoren / Sensors
BI1 Bicarbonate pump
XR1 OCM_NTC 4
LP 636 X ACDA XR2 V84 sensor
Pumpen / Pumps XR3 Bibag pressure switch
Extern X 636 X 637B
I/O X2 NTC2
Ventile / Valves
X HDF X3 NTC3
HPU wiring diagram
X5 Float switch
LP 632 X 637C X6 Level sensor 6
CPU 2
X 632
X CAN
X 633L
X 634L
X7 Conductivity measuring cell
X8 Dialysate pressure transducer 1
X ACCU
Akku
Accu X9 Reed contact 10 concentrate
Chapter 9: Circuit diagrams and circuit descriptions
X PWR
Input Output X12 Bibag NTC 133
board board X13 Bibag conductivity measuring cell
Netzteil X14 NTC109
4008 X15 OCM conductivity measuring cell 110
X RS
Power XS1 Dialysate pressure transducer 2 (182)
Supply LP 763
XS2 NC
LP 631 Multi
X63Z
X 631
X 639A
XS4 NC
X 633R
X 634R
Board
XS5 Level sensor concentrate 202
XS6 NTC INDI 308
XS7 INDI conductivity measuring cell 309
X 635
LP 630
Motherboard
Display
board X 82 X C1 X A7 X1
X S1 P-Dial 2
Pumpen Ventile Sensoren CAN
X S2 NC
N.C.
X R1 OCM_NTC4 X2 NTC2 X 16 X S3 NC
Conc.P. Bic.P.
TM-EN
X 10 ZKV PSW X9 Reedk. 10 X B1 X R2 V89 Sens.
LP 941 Blutleckdetektor
Wasserteilrechner / Hydraulic Processing Unit Blood Leak Detector
12A-2016
Chapter 9: Circuit diagrams and circuit descriptions
US-Sender LP 450-2
US-Transmitter
X156 20 +24V
19 +24V
18 0 (24V)
X155 0 (24V)
17
16 +12V
US-Empfänger 0 (12V)
15
US-Receiver
14 LDA1 = L
1 13 Niveau senken / Level down = H
2 X154 Dialyse Start / Dialysis start = H
3 12
11 n.c.
Niveau heben/senken
X351 LDSA = H
Level up/down 10
9 Potentialausgleich / Ground
1
2 8 Klemme zu, Clamp closed = L
LP 248 X160 ODSA = H
3 7
4 6 LDA2 = H
Potentialausgleich
Ground
31.07.14
02.04.14
Q
&>
&&
$
#
$
&$
&
& &" & &#
&%
$
#
)
)
)
,
,
"
"
) " ( 5 "
%
%
#
@: $ D9
E
K
K
&
& A"5CE<F(5<)F
<
<
&Q
"5E<F#<)F)
J
J
I
I
"%
""
"#
H
H
%>
"
!
!
G
G
$ $
%& # #
-689&$6
%#
%% % %
:
/6;
PCB LP 630 Motherboard
%
"
%
&
&
Q
!
"#
)
)
)
)
,
,
"
%
&
TM-EN
&
%
%
#
# #
K
K
K
K
<
<
<
<
J
J
J
#
I
I
I
"
""
"#
"&
H
H
H
"Q
H
!
!
!
!
&
12A-2016
G
G
G
" "
& %
%
"
"$
&
% %
A5BC A(BC
) " %%&>>"
&
) ) -/ %%&>>"?@J
= & '& &
-/
">
!"# -/
$
% ) -/
&'& ()(*+*, )-/
"> #"%
<
%
9-11
Chapter 9: Circuit diagrams and circuit descriptions
Chapter 9: Circuit diagrams and circuit descriptions
WATCHDOG
RTC
2 CHANNEL
SIO
LED's
512 KByte
EPROM
DIP-SW2
DIP-SW1
BATRAM
8 KByte
PAL
128 KByte
RAM
Address bus
Control bus
Data bus
Addresses Data
controller
Micro-
80188
9- 6-
ANALOG DAC
inputs outputs
ANALOG
C515C
Fig.: Block diagram PCB LP632 CPU2
Micro-
Control bus
9.10 PCB LP 632 CPU2
controller 8-
16-
DIGITAL DIGITAL
inputs outputs
PORT
(DIRECT) Adress bus (BUFFER)
Chapter 9: Circuit diagrams and circuit descriptions
ADRESSES
2x8 8-
DIP Data bus POWER
SWITCH outputs
DATA
(BUFFER) (BUFFER)
8-
DIGITAL
outputs
PORT
(DIRECT)
Data bus
TM-EN
12A-2016
Chapter 9: Circuit diagrams and circuit descriptions
05.12.12
analog digital
BUFFER 6
V104 MF OD_OUT
BUFFER 0
P_V102 CSS_REED FL_SWITCH_+5V
5V/12V
SPARE CI
P_VEN SPARE ABG_BYP
SPARE ABG_ON
ADC 0
BPR_ART
V43
VEN_BPR_SET V26
BUFFER 7
BUFFER 1
V24B
ART_BPR_SET V24
LD_SERIAL
UF_P1
Uref LDA2
SUB_W_P
SPARE
U_ACCU REED_BIC
BIBAG
BUFFER 2
P_V104 REED_RINSE
BIBAG_C
24V_EM PSW_V102
PSW_V104
ADC 1
P_ART PWR_OFF
HEP_ON
BLL_DIM
SPARE
BLL/BLL_OLD SW_ON_OFF
BUFFER 3 PWR_FAIL
COND_SIGNAL SHUNT_OUTP
SHUNT_INP
Uref SHUNT
SERV_EN
LEV_SIGNAL
SPARE SN
ADKS
BUFFER 4
TEMP_DIAL2 BPSB_ART
BPUS_ART
COND_SIGNAL2 BPSB_VEN
BPUS_VEN
ADC 2
BIBAG_TEMP HEP_ALARM
BIB_LEVEL
BIBAG_COND
EXT_ALARM
SPARE SERVICE_MODE
BUFFER 5
LEVEL UP
U_BATT_SW LEVEL DOWN
ADS_SN
Uref ACKN_CONC
ACKN_BIC
BIBAG_PSW
Microprocessor Interface
STEUER–BUS
DATEN–BUS
ADR–BUS
CPU 1
18.10.10
UF_P_CTRL UF_P1
UF_P_EN
UF_P2_CTRL UF_P2
EN3
Latch CONC_P
Micro-
control-
EN4 ler
Latch BIC_P
&
EN_CPU2
EN6 V24
& V24 V91
V24B V24b V102
Latch V26 V126
Control bus
Address V87 V130
decodin
Address bus
g logic
EN7
V30 V86
Latch V41 V99
V43 V104
Data bus Bus V84
driver
EN8
SN_EN HDF_LOG1
Latch SNST PROG_LOG2
CPU_OFF VENT_VALVE
V_ADS
EN5
Latch GAL
Oscilla-
tor
EN11
Latch Counter FLOW_P
D/A DEGAS_P
TEMP_SET DAC_BLL
DAC_DIM DAC_X
TEMP_ADJ DAC_P_ZE
U_ref STEUER_EP BIBAG_TEMP_AJ
STEUER_FP
EN12 V_ZKV1
Latch V_ZKV2
VY5
FREE_OUTP
16.03.12
entf.
A
B
E
C
D
1
1
C22 R33
R32
D1
D2
C23
R29
IC3
C21
IC4
C24
R30
ST2
2
2
C20
R20
R27
C7
C8
1 IC14
C19
C17
R28
C27
C49
R19
R45 C16
R50
C31 C28 C15
D10
DR1
IC12
3
3
T1
R42 R49
R48
1
2
R1
DR6
DR34
T2
C26
C41
DR15
DR5
DR16 R52
ST1
DR7
DR4 R44
C48
D7
DR2
D8
DR3
1
2
C47
D11
BR2
ST4
2
4
4
DR32
DR33
DR27
DR28
DR25
DR14
DR31
DR30
DR17
DR19
1
C18
C4
DR26
DR24
DR21
DR22
DR23
DR10
1
DR20
DR18
DR29
ST3
DR11
IC5
R24
IC2
SW1
R7 R8
C34
T4
C6
T3
5
5
R10 R11
T6
C12
ÄNDERUNG
T5
R13 R21
T8
C13
IC18
T7
a neue Sachnummer
R22 R23
T10
C14
T9
DATUM
R3 R9
25.02.15
C11 Q1
NAME
C2 C1
R4
BE
GE
C33
C3
R12
6
6
ZD2
IC1
R5
DATUM
R59
LAYOUT
08.03.99
06.10.98
R46
R43
R63
Ott
Sw
1
2
IC11
NAME
FRESENIUS
R16
R17
LP950M
R18
MEDICAL CARE
TD7
IC8
R62
R31
R15
R2DR12
C30 C29
ST5 1
R6
R47
R41
IC13
Q2
C25
7
7
C37
DR13 R36
LP950
ST6 DR8 R14 C9
1 DR9 C5 R53 D3
T11
R58 C32 R34
R37
D4 IC17
C10
R35
T12
8
8
M56173
CONTROL BOARD (HEP)
/
A3
BP
BLATT
F
A
B
E
C
D
9.14 PCB LP 950 Control board (HEP)
Chapter 9: Circuit diagrams and circuit descriptions
Chapter 9: Circuit diagrams and circuit descriptions
CS 1 – 12
Bargraph
Duty Art
Ven
Cycle TMP
Conductivity
CS 16, 17
Status LED's
Traffic Light
CS
WR
RD CS 18
A0 Chip-Select Alarm LED
A1 Decoding Key click
A2
A3
A4
CS 20
Key interrupt Keyboard
Encoding
Key matrix (Plug X2)
CS 21
Brightness
Control
CS 22
Alpha
Display
CS 23
UF Goal
UF Time
UF Rate
Data bus D0 – D7 UF Volume
2 2
X12 CAN X1 zu 729/X1
1 1
IC9
D3
C27
F1 2
BVM X8
C26 C25 1
R13
DR10
F2
DR9
2
UT1
DR5
C24
2 zu 751/X1 X14
IC8 1
R1
R5
2
DR11
DR12
DR4
DR3
DR8
DR7
DR6
C11 C9 X9
1
R4
R3
R2
R6
R8
C13
IC3
R10
C8 C10
R7
T1
C22
T2
C7 C5
IC10
R11
R12
C19
IC2
C20
C4 C6
C21
R9
12 R16
R20
32
22
11
C2
33
IC6
R17
C15
RN3
Q1
R18
C32 C35
R21 IC12
C31
IC1
43
IC13
1
6 2 44 40
2
C33
R24
C34
C3
7 39
43
R25
3 1
C30
3
C2
1
C12
IC5 1
48
R22 X15
75
2
53
IC11
54
64
74
17 29
R19
8
C2
C17
IC4
R23
18 28 C29
off
R15 C18
S1
RN1
Q2
RN4
C16 IC7
DR1
C1
RN2
C
32 1
A
XZ1
j neues LT-G
i neues LT-F
h X2-X5 entfällt 20.12.00 Ko
CI29
IC29 IC1 IC30 IC31 IC32
R40
C11
RN10
LD27
R41
IC20
LD22 LD23 LD24
R33
T16 T14
IC21
LD25
LD26
T13 T15
IC22
T2 T17
D5
CI28
IC33
T9
R29
R30
R32
R31
R28
RN5
RN9
T11
RN3
R55 C5 C7
CI23
RN6 IC24 RN7 IC25 IC26 IC17 R51
R25
C8
X1
R52 C6
T12
CI24
T8
R53 F1
D4
R24
T10 LD18 LD19 R54
R26
R45
CI10
R39
LD17
LD21
T1
RN4
D3
LD20
R42
CI21 R44
IC36
IC10 R43
R15
CI11
XA2
X2
1
R16
48
R21
R20
R17
R18
R19
X5
CI14
Q1
T3
RN2
IC11
IC14 IC15 IC16 X6 IC34
RN8
IC12 T6
CI1
C2
XB2
CI13
LD9 LD10 LD11 LD13
R58
R46
R47
R48
R49
R50
CI5 T5 T4 LD12
R61
R22
LD8 R64 R60 R59
C9
48
1
LD14
LD15
LD16
C1
IC2
R7
R8
LS1 CI7
C4
C3
R6
R10 R62 CI8
R4
R63
R11
CI9
CI19
R12
R13
R14
IC9
CI6
D2
LD1 LD2 LD3 LD4 LD5 LD6 LD7
IC7
IC8
RN1
TM-EN
R5
IC3
IC4
IC5
IC6
D7 D9
R34
D10
CI2
T19 T18 D1
R2
C16 D11
CI3 C13 D6 D8
R1 R3
T22
IC13
C15
C14 T21 T20
R57 R56
9.21 PCB LP 922 Display board (4008 S)
12A-2016
g IC19/27 gesockelt
f R44-45 hinzu 05.12.08 Schg
e R40-43 hinzu 09.06.04 SLu
d Option IC34 17.12.03 SLu
i neues LT-K
h neues LT-J
9-29
Chapter 9: Circuit diagrams and circuit descriptions
Chapter 9: Circuit diagrams and circuit descriptions
$~]q?@j\]^qj^'q^`}
}
"
5
(
#
%
)
)
, ,
" #
$~]q?@jj]^qj^'q^`}}jj}
Chapter 9: Circuit diagrams and circuit descriptions
$
$
$
$
$
$
$
$
$$$
$
$
$$
$
$
!$
"#$
, ,
$
$
$
%%>&"jj $=\]^`j^q{^`
%">&" '
$ $=\]q]j`q|\jq}~ $
$
Q%
}]qjq}~^~q~
^~{{q]^~]q~
\^`{?^~
$
)
&66
) "
TM-EN
12A-2016
& % " # $
)( )( )($
<
) (
#
"
#
)# )
""
"#
J
J
"
")
J)
"
)" #
"
)(
)
)#
"
$# $, $ #
)
) )
) "
J))
"
)5
")
) (
(
#
"
)
")
""
"
"#
$" $)
#
"
5
)
""
"
J)"
5 5
"
)
#
"
$
$
$
$
$
$
$
$
$$$
$
$
$$
$
$
"
)
J#
(
(
# 5
5
"
TM-EN
J)
#5
J
( ( #
#
"
"
(#
(
#
#
)
5
(
)
5
#(
))
(
5 J
"
"
5
J)
"
5
#5
#(
)
5)
5"
5
J)
)
# )
) (
(
)
##
"5
"
J)(
(
(
5 (
#
##
#
)
)" J J( "
"
"5
#
J
)
J)
#
( #)
J
J)5
(
#"
#
"
#
()
("
J5
## #
"5
J(
)
J"
#"
"
"
"
)
)
((
(5
5
"
# "
55 5 #
12A-2016
J"
"5
(
"(
)
J
J
5
#
5
5
5
( )
))
#
)
#
(
) #
)
# #(
#)
5
J)
<
(
9.25 PCB LP 941 Hydraulic processing unit
5#
5 J5
) "(
#
#
(
J
J#
J#
J"
J
#5
(
J
) ) "
#
J
!$
"#$
5
)
) "
J)# "
#"
5(
6G&-@//
5 # $
$
$
9
':
( " ,; &Q>&Q
!{^{ ]]^`j^q{^?@~^`
) " -/ %">&Q^\| '
$ j^`\~qj $
:
$
Q#&
-($D9 ('
/-&6/
/ &:@ # ('
G"/
) )'
Q#& %"
b $
$ $"
& % " # $
9-33
Chapter 9: Circuit diagrams and circuit descriptions
Chapter 9: Circuit diagrams and circuit descriptions