URIT-660
Microplate Reader
Service Manual
1
URIT Medical Electronic Group Co., Ltd.
Contents
1. APPEARANCE ............................................................................................................................... 3
APPEARANCE...............................................................................................................................
2. INSTALLATION ................................................................................................................................ 3
INSTALLATION................................................................................................................................
4. MAINTENANCE ................................................................................................................................ 9
MAINTENANCE................................................................................................................................
5.TROUBLESHOOTING
TROUBLESHOOTING ..................................................................................................................
TROUBLESHOOTING.................................................................................................................. 10
..................................................................................................................10
2
URIT Medical Electronic Group Co., Ltd.
1. Appearance
Figure 1-1
Figure 1-2
2. Installation
3
URIT Medical Electronic Group Co., Ltd.
Spring bolts
4
URIT Medical Electronic Group Co., Ltd.
3. Operation Instructions
3.1 Start up
Instrument will initialize after power on. Then the system will connect the PC automatically. Click
“cancel”, and display will show the main menu.
If user doesn’t touch any parts on the touch display, instrument will close the
light of LCD automatically to save power; user can continue to operate by clicking
any part of LCD.
Please operate it with special touch pen attached by URIT-660 microplate reader, do not use
accurate or hard materials (such as metal, glass etc.) to touch screen surface to avoid damage.
5
URIT Medical Electronic Group Co., Ltd.
User can set up the plate direction, move mode, and shake parameters in this
window. Press “OK”, enter into plate setting window.
After selecting the program, click the little box on the screen. The box represents
the well on the plate. There will be a number in the little box. If you modify sample
No., please click that well again, enter the number in pop up window, max number is
999.
If user wants delete marks, press “Del”, then clicks the numbers to delete, and
reset them.
Press”Auto”, the system can mark the plate automatically according to your
setting.
Press “Clear” will clear all the marks.Press “Start”, system will test the plate. If
any faults occur, system will pop up windows to warning. Press “stop” to quit, return
to plate setting window, tray reset automatically.
6
URIT Medical Electronic Group Co., Ltd.
7
URIT Medical Electronic Group Co., Ltd.
3.4 QC Management
Select the QC program in listing, then select QC method, enter QC number, press
“OK” to store.
Press “QC Inquiry” into inquiry window.
Select program want to inquire, and select month from month listing, press “OK”
to get the result.
3.5 Report Management
All tested samples are stored and listed according to test time.
User can edit, delete or view the repot by pressing the bottom on the right side of
the menu.
8
URIT Medical Electronic Group Co., Ltd.
1) Date and time: Please setting according to standard time, illegal date or time
is invalidation.
2) Printer : This instrument equipped internal heat sensor specially, which is
shown as Figure 8-1.
3) Sound switching: Setting about turn on or off the buzz. Sound switching on,
every warning will be sounded, or not.
4) Touch panel calibration : Touch panel and pen is the entering way for
URIT-660., touch screen takes resistance circuit, in some changeable external
environment(temperature), touch screen may have slight excursion(click position is
coincide with arrow), which may cause result error. At this time user should mark
touch screen again. Press “Touch screen calibration” to enter windows.
3.7 Power off
a) Press “Power off” in the main menu, select “OK”, screen will
display “Thank you for using, Please turn off!” At this time, turn off the power
switch on the rear panel of instrument.
b) If you don’t want to turn off the instrument, select “cancel “,
instrument will return to the main menu.
4. Maintenance
9
URIT Medical Electronic Group Co., Ltd.
Caution:
Organic solvent with strong corrosive acid, alkali and volatility such as acetone, ethoxyethane,
chloroform to clean instrument’s surface is forbidden. User can only use neutral scour to clean.
5.Troubleshooting
Troubleshooting
10
URIT Medical Electronic Group Co., Ltd.
6. System Instruction
11
URIT Medical Electronic Group Co., Ltd.
b) Main board
If you find out that the display is not very clear, please adjust the potentiometer
12
URIT Medical Electronic Group Co., Ltd.
Using multimeter to test the charged main board, the voltage value is:
JP606: +5V(±0.15V), +15V(±0.45V), -15V(±0.45V)
JP605: +24V(±1.2V), +12V(±0.36V)
13
URIT Medical Electronic Group Co., Ltd.
14
URIT Medical Electronic Group Co., Ltd.
15