Professional Documents
Culture Documents
Table 1 General characteristics (average values) of the samples from the five different sewage treatment plants in each day (samples were collected
from 16 Feb 2006 to 25 Feb 2006, DO: dissolved oxygen, SS: suspended solid)
thought to be mainly responsible for the COD, has been amounts of glucose to distilled water. The COD of the
described.18 It was demonstrated that the catalytic activity of solutions were measured by the conventional COD test meth-
copper was able to promote the oxidation of organic matter at od.25 The synthetic wastewater was then autoclaved at 121 1C
the electrode surface to water and carbon dioxide. Previous for 15 min to prevent the action of the microorganisms and
research clearly showed that the electrons released during stored at 4 1C before use.4 In the case of real wastewater,
oxidation can be measured as an electrical current, which is samples were obtained from five different sewage treatment
proportional to the COD value of the solution being ana- plants in Korea (Nanji, Anyang, Jungnang, Tancheon, and
lyzed.18,21,22,23 The electrochemical oxidation mechanisms of Konkuk University sewage treatment plants) from 16 Feb to
organic matter and the stoichiometric concept of the oxidation 25 Feb 2006. All samples were collected at a depth of 50 cm at
were also proposed.21,22 Frequently, the term ‘electrochemical the first sedimentation tank at 9:00 am, 11:00 am, 13:00 pm,
oxygen demand’ (EOD), instead of COD, is used for these 15:00 pm and were kept below 4 1C. The general character-
electrochemical methods, which is defined as the oxygen istics of the samples (at the first sedimentation tank) are
equivalent of the charge consumed during electrochemical described in Table 1. Conventional COD determinations on
oxidation of the organic species of a sample.21,22,23 As both the collected samples were pre-established using the official
parameters express the concentration of total organic matter test method.25 All experiments were carried out within 12 h
in solution, the only difference between EOD and COD is the after sample collection.
oxidation method of these organic species.14 However, it
Electrodes and electrochemical cell
should be noted that a major drawback associated with the
electrochemical method using a metal electrode is the contam- A COD sensor with a three-electrode electrochemical cell
ination and/or the accumulation of inactive metal oxidant at configuration (working volume: 3 mL) and the electrode-
the surface of the electrode in the form of residues left after surface grinding unit was prepared (Fig. 1). A copper tube
reaction. In continuous operation at a practical site, the (99.99%, 30 mm outside diameter, 20 mm thickness, 10 mm
inactivation of metal electrodes by contamination presents a length, 463 mm2 surface area, Poongsan Metal, Korea), a
major problem, and may lead to inaccurate data.23,24 stainless steel cylinder (40 mm outside diameter, 3 mm thick-
In the authors’ previous work,24 a COD sensor using a ness, 348 mm2 surface area, stainless steel 316 L, Yonnam
copper working electrode and an electrode-surface grinding Metal, Korea), and an Ag/AgCl electrode (Model No.
unit was examined, and it was found that the method can be 2731401-35-S8, Phoenix, USA) were used, respectively, as
used successfully for COD measurement for a limited opera- working electrode, counter electrode and reference electrode.
tion time (i.e. up to 24 h of measurement). The experiments Using a compressed diamond grinding block (3 3 30 mm,
focused on establishing the basic experimental conditions for DIA grinding stone, Dawoori, Korea) mounted on a servo-
the electrochemical COD measurement using glucose solutions motor unit (HS-85MG, Hitec RCD, Korea), the electrode-
and real wastewater from a plant as laboratory samples.24 surface grinding action was achieved by rotation of the grind-
The present study describes demonstrations of a COD ing block. A small stick of acrylic stick was fitted with the
sensor system capable of on-line, near real-time monitoring working electrode surface for the preparation of a grinding
of COD for real wastewater treatment, and is divided into two block rest. Generally, the grinding operation was performed
parts. Firstly, using wastewaters obtained from sewage treat- for 30 s (10 times, forward and backward, over a 901 arc) every
ment plants, the COD measurement capability of the system 8 h and the pressure of the working electrode upon the
was demonstrated. Secondly, at a wastewater treatment plant diamond grinding block was maintained using a coiled spring.
using the real wastewater, the sensor system was examined for After the grinding action, the grinding block was set at the
field applicability, reproducibility and operational durability. acrylic non-conductive block rest to prevent formation of
electrochemical noise and the measurement was performed.
The procedures ensured that a relatively clean electrode sur-
Materials and methods face was exposed to the solution under test, thus avoiding
significant effects arising from surface adsorption of organic
Wastewater samples
oxidation products.
For the estimation of the detection range of the COD sensor,
Configuration and operation of the system
glucose solution was used as synthetic wastewater. The COD
concentration (generally 10 to 2000 mg L 1) of the synthetic The schematics of the COD monitoring system, including the
wastewater was maintained by additions of the appropriate COD sensor and the grinding unit, are represented in Fig. 2.
This journal is
c The Royal Society of Chemistry 2007 J. Environ. Monit., 2007, 9, 1352–1357 | 1353
View Article Online
Published on 23 October 2007. Downloaded by University of Arizona on 30/05/2013 02:17:21.
Fig. 1 Schematic representation (A) and photo (B) of the electrochemical cell. During the measurement operation, the grinding block was set at
the non-conductive block rest (an acrylic block).
For the system operation, 1 : 1 (v : v) of wastewater sample : ground current, the same volume of distilled water was fed to
electrolyte (0.1 M NaOH) mixture was added to the sensor the sensor at intervals of 300 s, instead of wastewater sample.
(flow rate: 1 mL min 1) using a programmable multi-channel DC amperometry was used for electrochemical COD measure-
peristaltic pump (ISM597A, Ismatec, Switzerland). To distin- ment of the wastewater samples.20,22–24,26 Oxidation of organic
guish the oxidation current of the organic matters from back- matter by the copper as an electro-catalyst was achieved using
a potentiostat (WMPG 1000, IBM PC controlled, Wonatech
Co. Korea), at a constant potential of 600 mV versus Ag/AgCl
(i.e. below 700–800 mV, which is usually regarded as the onset
potential of water oxidation).24 This process had a constant
temperature (30 1C). For the precise control of the temperature
sensor unit, a temperature control system using a thermo-electric
chip module (SKT, Korea) was used. The data was recorded
using a personal computer with the exclusive software of WMPG.
This software was used to calculate the Coulombic yield (current
(I) time (t)) produced from the COD sensor. The measured
data were given as an on-line graphic display, which showed
values at 1 min intervals or as required. The data (i.e. Coulombic
yield) from the sensor were calculated as EOD, estimated from a
Fig. 2 Schematic representation of the COD measurement system. pre-established calibration curve (COD vs. Coulombic yield).
Table 2 General characteristics of the samples collected from the of a known concentration of COD (synthetic wastewater) to
Jungnang sewage treatment plant during 3 months (18 Feb 2006– the sensor showed an increase Coulombic yield with increasing
18 May 2006, DO: dissolved oxygen, SS: suspended solid)
COD concentration (Fig. 4A). When the grinding unit was
Parameter Average Range operated, no significant signal was observed, apart from a
minor perturbation at the initial stage of grinding. When the
Temp/1C 15.7 8.3–22.6
pH 7.4 7.0–7.7 grinding unit was not operated, however, a decrease in current
DO/mg L 1 1.4 0.3–2.8 was observed under the same concentration of COD (Fig. 4B).
COD/mg L 1 45.2 25.4–91.2 Fig. 4C shows the correlation between Coulombic yield and
SS/mg L 1 87 21–245
Total nitrogen/mg L 1 18.4 7.2–25.8 the different concentrations of COD fed over the range of 10
to 2000 mg L 1, with and without the grinding. In the case of
Published on 23 October 2007. Downloaded by University of Arizona on 30/05/2013 02:17:21.
1
Total phosphorous/mg L 2.1 1.1–4.9
grinding, a clear linear relationship was obtained between the
range of 10 to 1000 mg L 1 (line (b): correlation coefficient;
Field application site
r2 = 0.98). However, over 1000 mg L 1, the Coulombic yield
The application site was located in Jungnang sewage treatment reached a plateau with the COD increase (Fig. 4C, line (a))
plant (Seoul, Korea). The COD monitoring system was in- The plateau at the high concentration of COD range was
stalled at the first sedimentation tank in the plant. The inflow probably due to the saturation of organic matter at the limited
to the system was taken from the wastewater from the tank. surface area of the electrode28 or was probably due to the
The general characteristics of the inflow sample (at the first presence of other effects, such as corrosion or deactivation or
sedimentation tank) are described in Table 2. The temperature polymerfilm formation.29 In the case of the COD measure-
was varied from 8.3 to 22.6 1C (average: 15.7 1C). The COD ment without the grinding, a relatively low correlation (r2 = 0.781)
was varied from 25.4 to 91.2 mg L 1 (average: 45.2 mg L 1). between the COD of wastewaters and Coulombic yield was
In the case of field application, a particle remover for the observed (Fig. 4C, line (c)) compared to the measurement with
wastewater was also used to prevent clogging of the system.27 the grinding, probably because of attachment of oxidant residue
and/or accumulation of inactive metal oxidant at the surface of the
electrode. From these results, it can be deduced that the Coulombic
Results and discussion yields (i.e. EOD) from the sensor, when the electrode-surface
grinding procedure was in operation, could be conveniently con-
Estimation of the detection range of the COD sensor verted into COD values for a limited measurement range.
Typical traces of the current change due to the oxidation of Electrochemical COD measurement with various real
organic matter at the surface of the electrode with different wastewater samples
concentrations of COD (synthetic wastewater) are shown in
Fig. 3. The general form of DC amperometric response to the Using various real wastewater samples obtained from five
addition of the synthetic wastewater–NaOH mixture (COD sewage treatment plants, the reproducibility and accuracy of
load), was a rise in current as a function of time, decaying the COD sensor system for real wastewater samples were
when the NaOH–distilled water solution was fed to the cell, so examined. To obtain different COD concentrations of waste-
removing the organic matter. The measured Coulombic yields water, samples were collected at 9:00 am, 11:00 am, 13:00 pm,
varied with the concentrations of COD of wastewater fed to and 15:00 pm in the day from each sewage treatment plant.
the sensor. Using various concentrations of the synthetic The grinding of the electrode surface was set at once every 8 h
wastewater, the detection range of the sensor was established. for a period of 30 s. Fig. 5 shows the results of electrochemical
Measurement of the current change after successive additions measurement of organic matter using different concentrations
of COD. In general, all the wastewaters used in this study, at
different kinds of organic matter with different concentrations,
generated an oxidation current, and the current generation
pattern was closely related to the conventional COD value.
When the correlation between the produced EOD data and the
conventional COD value were compared, reasonable correla-
tions (correlation coefficient, r2, ranged from 0.92 to 0.95) were
obtained. These results indicate that the COD sensor system
can be applied to advantage in practical sewage treatment
plants. The differences of EOD obtained for the various waste-
water samples, however, may be due to differences of waste-
water characteristics of each plant. Therefore, for a precise
measurement of COD using the system, a pre-calibration
between EOD and conventional COD would be required.1–3,24
Long term application of the COD monitoring system at a
sewage treatment plant
Fig. 3 Electrochemical signal generation patterns from the electro- To examine the on-line capability and durability of the system,
chemical cell with different concentrations of synthetic wastewater. the COD sensor system was installed at the wastewater
This journal is
c The Royal Society of Chemistry 2007 J. Environ. Monit., 2007, 9, 1352–1357 | 1355
View Article Online
Published on 23 October 2007. Downloaded by University of Arizona on 30/05/2013 02:17:21.
Fig. 4 Signal generations from the electrochemical cell with different from the system (Fig. 6B) for 90 days at the plant. The
concentrations of synthetic wastewater (A: using grinding unit, arrows conventional COD profile during the 90 days is also shown
indicate grinding points, B: without grinding unit) and correlation in Fig. 6A. As shown in Fig. 5, the EOD pattern in this
curves (C) between the conventional COD and the Coulombic yield experiment (Fig. 6) varied with the variation in COD concen-
with different COD concentrations (a): using grinding unit, (b): ideal tration of the real wastewater. When the Coulombic yield (i.e.
correlation, (c): without grinding unit.
EOD) from the system was compared with the conventional
COD measurement results, a clear linear relationship (correla-
treatment plant. A particle remover was also installed to tion coefficient; r2 = 0.98) was observed (Fig. 6C). During the
prevent clogging of the sensor. Over 90 days, the measurement experimental period, there were several hours of interruption
system was successfully operated without any serious pro- due to the sudden power failure and the exchange of the
blems. Fig. 6 illustrates COD measurement data using the reference electrode. However, when the experimental condi-
sensor system (Fig. 6A) and the current-generation pattern tions were recovered, no significant variation in COD
Acknowledgements
Published on 23 October 2007. Downloaded by University of Arizona on 30/05/2013 02:17:21.
References
1 S. Zhang, H. Zhao, D. Jiang and R. John, Anal. Chim. Acta, 2004,
514, 89.
2 S. Ai, M. Gao, Y. Yang, J. Li and L. Jin, Electroanalysis, 2004, 16, 404.
3 A. Cuesta, J. L. Todoli, J. Mora and A. Canals, Anal. Chim. Acta,
1998, 372, 399.
4 Y. C. Kim, S. Sasaki, K. Yano, K. Ikebukuro, K. Hashimoto and
I. Karube, Anal. Chim. Acta, 2001, 432, 59.
5 B. Vallejo-Pecharroman, A. Izquierdo-Reina and M. D. Luque de
Castro, Analyst, 1999, 124, 1261.
6 T. Korenaga and H. Ikatsu, Anal. Chim. Acta, 1982, 141, 301.
Fig. 6 Application of the COD monitoring system to the Jungnang 7 M. Kataoka, N. Unjyo and T. Kambara, Talanta, 1983, 30, 741.
sewage treatment plant (Seoul, Korea) for 3 months (18 Feb 2006– 8 T. Korenaga, X. Zhou, K. Okada and T. Moriwake, Anal. Chim.
18 May 2006). (A) EOD profile and conventional COD (J: conven- Acta, 1993, 272, 237.
tional COD concentration measured by standard method); (B) current 9 J. M. H. Appleton and J. F. Tyson, Anal. Chim. Acta, 1986, 179, 269.
generation pattern from the system; (C) correlation curve between the 10 W. F. Jardim and J. R. Rohwedder, Water Res., 1989, 23, 1069.
11 A. A. Beliustin, A. M. Pisarevsky, G. P. Lepnev, A. S. Sergeyev
conventional COD concentration and Coulombic yield from the
and M. M. Shultz, Sens. Actuators, B, 1992, 10, 61.
system. 12 T. Korenaga, X. Zhou, T. Moriwake, H. Muraki, T. Naito and S.
Sanuki, Anal. Chem., 1994, 66, 73.
measurement performance was observed. These experimental 13 S. J. Edwards and M. Allen, Analyst, 1984, 109, 671.
14 P. Westbrock and E. Temmerman, Anal. Chim. Acta, 2001, 437, 95.
results show that the real-time, on-line monitoring of COD at
15 E. Tonning, S. Sapelnikova, J. Christensen, C. Carlsson, M.
a practical sewage treatment plant could be possible using a Winter-Nielsen, E. Dock, R. Solna, P. Skladal, L. Norgaard, T.
simple copper electrode and surface grinding unit. It should be Ruzgas and J. Emneus, Biosens. Bioelectron., 2005, 21, 608.
noted that the durability of the sensor over three months was 16 K. H. Lee, T. Ishikawa, S. J. McNiven, Y. Nomura, A. Hiratsuka,
S. Sasaki, Y. Arikawa and I. Karube, Anal. Chim. Acta, 1999, 398,
achieved using the grinding unit. About 1 mm thickness of the 161.
rotating working electrode compartment was worn during the 17 B. Jin, Y. He, J. Shen, Z. Zhung, X. Wang and F. S. C. Lee, J.
operation. According to previous research, major important Environ. Monit., 2004, 6, 673.
drawbacks of the electrochemical sensor for wastewater mon- 18 K. H. Lee, T. Ishikawa, S. McNiven, Y. Nomura, S. Sasaki, Y.
Arikawa and I. Karube, Anal. Chim. Acta, 1999, 386, 211.
itoring, especially for long-term operation, are the contamina- 19 M. Z. Luo and R. P. Baldwin, J. Electroanal. Chem., 1995, 387, 87.
tion of the electrode surface by oxide film formation and 20 P. F. Luo, F. Z. Zhang and R. P. Baldwin, Anal. Chim. Acta, 1991,
residual organic matter.23,24,29,30 Several attempts have been 244, 169.
21 C. Comninellis, Electrochim. Acta, 1994, 39, 1857.
made to regenerate the electrode by electrochemical, physical, 22 C. A. Martinez-Huitle and S. Ferro, Chem. Soc. Rev., 2006, 35,
and chemical methods.31 These approaches, however, have 1324.
solved only some of the problems associated with electroche- 23 R. Roessler-Fromme and F. Scholz, Fresnius’ J. Anal. Chem., 1996,
mical methods, and there remains a continuing need for 356, 197.
24 S. M. Yoon, C. H. Choi, B. S. Park, G. J. Jin, B. G. Jeong, M. S.
convenient, accurate, and economic COD measurement meth- Hyun, J. M. Park, S. S. Lee, D. H. Yi and H. J. Kim, J. Korean
ods that will allow more efficient monitoring and control of Soc. Environ. Eng., 2006, 28, 453.
COD of wastewater. Because previous regeneration methods 25 APHA-AWWA-WPCF, Standard methods for the examination of
water and wastewater, American Public Health Association,
have generally involved sophisticated devices, such as a spe- Washington, DC, 21st edn, 2005, pp. 5.1–5.15.
cific operation program with electrochemical devices, a water 26 A. Lynggaard-Jensen, Talanta, 1999, 50, 707.
jet, or an ultrasonic sonicator, the operation procedures 27 H. J. Kim, C. H. Choi, S. M. Yoon, K. K. Kwon, B. G. Jeong, G.
require careful and periodic maintenance. In the present J. Jin and M. S. Hyun, Kor. Pat. 10-2005-0121834, 2005.
28 M. Panizza, P. A. Michaud, G. Cerisola and C. Comninellis,
studies, however, the attachment of a simple rotating grinder Electrochem. Commun., 2001, 3, 336.
onto the electrode surface facilitates the continuous applica- 29 C. A. Martinez-Huitle, S. Ferro and A. D. Battisti, Electrochim.
tion of the metal electrode. Pre-calibration is an important Acta, 2004, 49, 4027.
parameter for the operation of the system, especially in the 30 M. Kim, S. M. Youn, S. H. Shin, J. G. Jang, S. H. Han, M. S.
Hyun, G. M. Gadd and H. J. Kim, J. Environ. Monit., 2003, 5, 640.
practical field application. Other factors that influence the 31 J. M. Skowronski and P. Krawczyk, J. Solid State Electrochem.,
monitoring performance include the surface area of the 2004, 8, 442.
This journal is
c The Royal Society of Chemistry 2007 J. Environ. Monit., 2007, 9, 1352–1357 | 1357