Professional Documents
Culture Documents
On 19 September 2014, the current version of the “Guideline of the German Medical
Association on Quality Assurance in Medical Laboratory Examinations” was published. It
featured an introduction by the German Medical Association.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 27
Central laboratory means that the medical laboratory Deviation of measurement, root mean square of the
examinations are usually performed for the entire insti-
tution (e.g., hospital) by one single organisational unit The root mean square of the deviation of measurement is
(“medical laboratory”) and by appropriately qualified a scatter of distribution of the measured values around
technical personnel. The central laboratory can also be the (conventional) true value of the measurand (here, the
an external laboratory managed by another legal entity/ target value of the control sample). It is calculated using
operator. the formula
1 n
∆= ∑( x − x )2
n i=1 i 0
Control strain
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
28 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
n− 1 2 2
∆= s +δ Influencing factor
n
This relates to the patient being examined. These are
where
changes in the composition of body fluids as a result of
s represents the empirical standard deviation of a
illness or defects (diagnostically relevant) or other biolog-
random sample
ical phenomena (diagnostically irrelevant). They reflect
δ represents the systematic deviation of measurement
the conditions within the patient.
Document
This includes medical laboratory analyses that are per-
formed directly as single measurements without sample
A document includes information and its carrier. For
preparation.
example, records, instructions (including quality regula-
An important criterion for these laboratory examina-
tions), method descriptions, specifications, calibration
tions is the immediate deduction of therapeutic actions
tables, reference ranges, drawings, reports, results, legal
from the laboratory analysis performed.
provisions or standards.
Equipment Location
Equipment includes, but is not limited to, devices, rea- The geographical location (postal address) of a company
gents, control samples, reference materials, consumables or an institution where medical laboratory examinations
and analysis systems. are performed.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 29
Measurement
Precision of measurement
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
30 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
Reference method volume dosing is not included in this definition. If the col-
lection system contains additives included by the manu-
A thoroughly investigated measurement method whose facturer, this does also not constitute sample preparation.
results have an uncertainty of measurement commensu-
rate with its intended use, such as in assessing the true-
Nominal value
ness of other measurement procedures used for the same
measurand and for the characterisation of reference
Target value determined without using a reference meas-
materials.
urement method.
This includes all changes made to the material to be exam- Unit use reagents
ined prior to loading it into the measuring device or instru-
ment by the person collecting the sample or by the person These are reagents which are portioned for single meas-
carrying out the measurement. Pipetting of the sample and urements and are fully consumed after one analysis.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 31
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
32 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
that can impact on the quality of these medical laboratory 6.1.4 Written instructions outlining the correct method of
examinations. collection and handling of specimens have to be available
These records should contain, at a minimum: to the persons responsible. These instructions have to be
(1) Name of system or device summarised in the document for collecting the specimen.
(2) Name of the manufacturer, model and serial number 6.1.5 The document for collecting the specimen has to
or other form of identification specifically contain the following:
(3) Date of initial operation (1) The list of the medical laboratory examinations
(4) Instructions for use, operating instructions and other offered, or reference to this
information from the manufacturer, or justification, if (2) Instructions on:
they are not available (a) Preparing the patient
(5) Functional tests (b) Completing the examination request form or
(6) Instrument maintenance intervals and results of filling in an electronic entry mask
conducted controls including date, time and type of (c) The required information about the patient
inspection and other maintenance work (d) The type and quantity of the specimen to be
(7) The nature, dates and times of equipment outage, collected
malfunctions, structural modifications and repairs. (e) The specific time constraints for collecting, stor-
These records have to be kept for 2 years beyond the ing and transporting the specimen, if required
working life of the equipment and have to be promptly (f) Collecting the specimen, with a description of
accessible. the containers for the specimen and all required
5.4.4 Authorised and trained members of staff may only additives
operate devices and analysis systems. Instructions for oper- (g) Unmistakably labelling the specimen
ating and servicing the equipment have to be kept up-to-date (h) All of the measures that are to be taken between
and be accessible to the members of staff in the workplace. the time the specimen is collected and its arrival
at the medical laboratory, and
(i) The period of time within which further medical
6 Medical laboratory examinations laboratory examinations can be requested.
(3) Information and instructions to be provided to the
6.1 Pre-analytical phase patients regarding preparation measures for collect-
ing the specimen and, if necessary, patient consent
6.1.1 The sender of the specimen to be examined has to forms for collecting the specimen and for performing
be given a list of the laboratory examinations offered by the medical laboratory examinations and
the medical laboratory relevant to their purposes as well (4) Patient information regarding the self-collection of
as a document stating the details of specimen collection. the specimen and storage and transportation details
6.1.2 The medical laboratory has to provide profession- of the self-collected specimen.
ally competent advice regarding the examinations offered,
including the examination method to be selected, the type 6.1.6 Criteria have to be established for rejecting medical
of specimen to be used and the evaluation of results of laboratory examinations.
examinations. 6.1.7 The submitted specimen and portions thereof have
6.1.3 The examination request form submitted by the to be clearly assigned to one patient. If this is not fulfilled,
sender of the material has to contain the following the medical laboratory may not process them. The sender
information: of the material has to be notified and the incident has to
(1) Identification of the patient – including gender and be documented.
date of birth in the case of age and gender-specific If the specimen cannot be assigned beyond a doubt to
measurands one patient and a specimen of the same quality cannot be
(2) Identification of the sender of the material and the collected or was collected when the patient was in a critical
recipient of the report if not the same condition, the medical laboratory shall decide, after consult-
(3) Type of specimen and, if relevant, the anatomical site ing with the sender of the specimen, if the requested medical
of collection and time of collection laboratory examinations should nevertheless be performed.
(4) Requested examinations and The result of the consultation has to be documented.
(5) Clinically relevant patient information required for 6.1.8 When the specimen arrives, the medical labora-
the requested examination. tory has to check whether there is any indication that a
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 33
timely delivery has not occurred for the requested medical (13) Measures to be taken in the case of abnormal results
laboratory examinations or whether the conditions estab- (14) Safety precautions, and
lished in the document governing the collection, han- (15) References.
dling, storage and transport of the specimen were not 6.2.4 If the medical laboratory modifies one of its examina-
met. If such indications are identified, the medical labo- tion procedures so that the results, and thus the interpreta-
ratory has to decide whether the examinations are to be tion, change in a clinically significant way, the sender of the
performed despite this, or whether a new specimen is to material has to be informed as soon as possible in written
be requested. This incident has to be documented. form.
6.1.9 If necessary, the medical laboratory must have
documented procedures in place covering the acceptance,
labelling and processing of specimens, and the reporting
6.3 Post-analytical phase
of medical laboratory examinations that are considered to
be urgent.
6.3.1 The results have to be technically validated and,
using the available clinical data, medically validated.
6.2
Procedures for conducting medical laboratory The medical laboratory must have procedures in
examinations place for releasing the examination results. These should
include information about who may authorise the release
6.2.1 The medical laboratory may only use examinations of reports and whom they may be issued to. The proce-
procedures that meet medical requirements. dures have also to contain a guideline for the immediate
6.2.2 The medical laboratory may only use validated issuance of reports to patients.
examinations procedures. It has to document the proce- It has to be documented which persons carried out the
dure used for validation and the results obtained. technical and medical validation.
6.2.3 All medical laboratory examination procedures 6.3.2 The reports have to be easy to read and contain at
have to be documented in procedural instructions. These least the following information:
instructions have to be written in such a way that medical (1) Date, and if required, time the report was issued
laboratory staff can understand them. They have to be (2) Identification of the patient
available at the workstations at all times. (3) Name or other means of identifying the sender of the
The manufacturer’s instructions for use and addi- specimen and, if required, his address; the address of
tional remarks as appropriate are deemed to be part of the the recipient of the report if not the same as that of
procedural instructions. the sender
Each procedural instruction has to contain the follow- (4) Name of the medical laboratory
ing, where applicable: (5) Date and time when the specimen arrived at the medi-
(1) Identification of the document cal laboratory
(2) Principle of the procedure used for examination (6) Date and time when the specimen was collected, if
(methods) this information is available and important for inter-
(3) Individual steps of the procedure preting the examination results
(4) Calibration procedure as appropriate or as available (7) Type of specimen
(5) Procedure used to calculate the result as appropriate (8) Name of the laboratory examinations and the meth-
or as available ods used, if the latter is important for interpreting the
(6) Required specimen (including details of the type of examination results
specimen container and the necessary additives) (9) Examination results and corresponding units as
(7) Required instruments, reagents, culture media and necessary
test systems (10) Reference intervals or other remarks for interpreting
(8) Specification of the performance of the examination the examination results, and
procedure (11) Identification of the person responsible for releasing
(9) Information on possible interference and cross-reactivity the report.
(10) Reference ranges for healthy probands 6.3.3 If there is a possibility that the examination result
(11) Objective of the medical laboratory examination was affected by the condition of the specimen, this has
(medical indication) to be stated in the report. As the case may be, the report
(12) Possible causes of deviating results should state that the results are conditional.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
34 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
6.3.4 The medical laboratory must have written policies (b) Examination procedures, handling equipment,
and procedures in place for subsequent amendments to reagents and other relevant consumables, valida-
reports. The changes have to be marked with the date, tion of the examinations procedures
time and name of the person responsible for the changes. (c) Ensuring the analytical quality of the examina-
The original results have to remain accessible. tions procedures through internal and external
6.3.5 The medical laboratory must have procedures in quality assurance and regular discussions about
place for immediately notifying a physician (or other the results of the quality assurance
clinical personnel responsible for patient care) if exami- (d) Post-analytical procedures and generating and
nation results exceed “alarming” or “critical” limits. This transmitting reports
includes reports from referral laboratories. (e) Technical and medical validation of the examina-
6.3.6 Specimens and samples have to be stored in such tion results
a way that enables repeat or additional medical labora- (f) Document control procedure
tory examinations to be performed over a period of time (g) Keeping, storing and archiving records
as established by the medical laboratory. (h) Resolving with complaints
(i) Determining errors and corrective measures
(j) Preventive measures
7 Quality management system (k) Communication and other interaction especially
with patients, medical personnel, referral labora-
7.1 Quality manual tories, and
(l) Internal audits.
7.1.1 The quality management system and the docu- 7.1.2 If the medical laboratory is a part of an organisation
mentation used in the medical laboratory have to be which already has a quality management system in place,
compiled in a quality manual. This quality manual has it is not necessary to have a separate quality manual for
to include or make reference to all procedures. Labo- the medical laboratory, if the corresponding section in
ratory personnel have to be instructed on the use and the organisation’s quality manual contains requirements
practical application of the quality manual and all ref- similar to the ones contained in this guideline. The same
erenced documents. The quality manual shall be kept applies to 7.2 and 7.3 below.
up-to-date at all times.
The quality manual must contain the following infor-
mation where applicable: 7.2 Document control
(1) Introduction: description of the medical laboratory,
its legal status and its main tasks The medical laboratory has to define, document and
(2) Objectives and strategies: description of the quality maintain procedures for the control of all quality assur-
assurance policy ance documents and information (internal and external).
(3) Management: description of their responsibilities and A copy of each version of these documents has to be stored
qualifications for future reference. Management has to establish the
(4) Staff: archive period, taking legal requirements into considera-
(a) Qualifications, briefings, training and continuing tion. A procedure has to be introduced which guarantees
education and that only the current version of the documents is acces-
(b) Health protection and safety procedures sible at the place where they are being used.
(5) Resources and partnerships:
(a) Rooms
(b) Equipment 7.3 Resolving complaints
(c) Environmental conditions
(d) Partnerships (referral laboratories, external ser- The medical laboratory has to establish and implement
vice providers and suppliers) a procedure for the documentation and handling of com-
(e) Environmental issues plaints. Records documenting the complaints and the
(6) Processes: investigation, as well as preventive and corrective meas-
(a) Procedures in accordance with the document for ures taken by the medical laboratory, are to be kept and
collecting the specimen maintained.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 35
7.4 Examinations in referral laboratories quality assessment programmes in accordance with the
procedures described in Chapter B of this guideline.
7.4.1 The medical laboratory has to keep a list of all
referral laboratories it commissions. All medical labora-
tory examinations sent to a referral laboratory must be
documented.
B Special parts
7.4.2 The commissioning medical laboratory is responsi-
ble for ensuring that the original sender receives the exam- B1 Quantitative medical laboratory
niation results and findings from the referral laboratory. examinations
7.4.3 When hiring referral laboratories outside the scope
of this guideline, the commissioning medical laboratory 1 Principles of quality assurance
has to ensure that the referral laboratory possesses the
required competencies and that a similar quality manage- (1) Part B1 specifies minimum requirements to assess the
ment system is in place. quality of quantitative results of medical laboratory
examinations. These minimum requirements include
internal and external quality assurance.
7.5 Non-conforming examination results (2) All of the quantitative examinations performed by
the medical laboratory are subject to internal quality
The medical laboratory has to define and apply a proce- assurance. If several devices or measuring stations
dure for corrective measures for non-conforming exami- are used to perform a medical laboratory examina-
nation results. tion, internal quality assurance is to be performed on
Management must specifically ensure that: each of these device or measuring stations.
(1) Persons responsible for problem resolution are named (3) In addition, all measurands listed in Table B1a to c of
(2) The medical significance of the incorrect result is this Part are subject to external quality assurance.
considered and, if necessary, the sender is informed (4) The measurands in Table B1a to c are listed alpha-
thereof betically based on the type of specimen. The criteria
(3) Examinations are halted and reports are withheld as used for including a measurand in the table are, spe-
necessary cifically, the frequency of the measurement procedure
(4) Corrective action is taken immediately and its medical relevance according to the current
(5) Already-released examination results are recalled or state of science. Deviations of measurements as listed
the recipient is appropriately informed of the error in the table are determined based on medical require-
(6) There is a designated person responsible for the ments and current state of analytical technology.
recalling of the examination results Table B1 is continuously updated.
(7) Causes and corrective measures taken are docu- (5) This Part of the guideline does not apply to hemacy-
mented and tometer counting of corpuscular components in body
(8) The success of any corrective action taken is verified fluids, determination of erythrocyte sedimentation
to ensure that all identified non-conformities have rate and examination of pH test strips.
been eliminated.
The records documenting the identified non-
conformities and the corrective measures have to be 2 Procedure of quality assurance
kept for 2 years.
2.1 Internal quality assurance
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
36 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
(3) On days when a measuring procedure is used to ana- 2.1.3 Calculating and evaluating the root mean square of
lyse patient samples, a single measurement of a con- the deviation of measurement after completing a
trol sample has to be performed at least twice within a control period
24-hour period and, at the latest, after 16 hours. (1) Based on the results from all single measurements
(4) In addition, a single measurement of a control sam- of control samples that have led to the release of the
ple has to be performed after every intervention to the measuring procedure or of the patient results, the rela-
measuring system. tive root mean square of the deviation of measurement
Interventions into the measuring system are: has to be calculated immediately after the completion of
(a) Restarting the device after it has been switched a control period. A control period generally consists of 1
off completely calendar month. If, per control period, there are fewer
(b) Calibration by the user than 15 results from single measurements of control
(c) Repair or maintenance work on devices relevant samples per measurement procedure that have led to the
for the results of the medical laboratory examina- release of a measurement, this period will be extended
tion and by 1 month until at least 15 such results are available.
(d) Changing reagent lots The total period of time may not exceed 3 months.
(5) The control samples must be as similar as possible to (2) If the relative root mean square of the deviation of meas-
the patient samples being examined. Within the same urement for a control sample exceeds the value given in
measuring procedure the control and calibration Table B1a to c, Column 3, the examination procedure
materials must not be identical. has to be blocked from further use in measuring patient
(6) Control samples have to be used with known target specimens. The measurement procedure must not be
values that are within the measurement interval rel- made available for measurements until the functional-
evant for medical decisions. ity of the procedure has been proven through appropri-
(7) Control samples with target values in at least two dif- ate measures. The entire process has to be documented.
ferent concentration intervals have to be used alter- (3) If the value given in Table B1a to c, Column 3 is again
nately, if available. exceeded in the subsequent control period for the
same control sample, and user-related causes can be
excluded, apart from taking the appropriate measures
2.1.2 E valuation of the results of single measurements according to (2) the responsible federal authorities
of control samples have to be informed, if this can be defined as an “inci-
(1) The evaluation of the results of the single meas- dent” according to § 2 of the German Safety Plan for
urements of control samples has to be performed Medical Devices (MPSV).
without delay as after the results are available. Eval- (4) As to measurands that are not listed in Table B1a to
uation is based on the deviations of measurement as c (2) applies accordingly. Instead of the maximum
listed in Table B1a to c Column 3, or either on the permissible deviation as listed in Table B1a to c, Col-
basis of the internal laboratory deviation limits or umn 3, the laboratory’s internal Δmax has to be used
on the intervals of the manufacturers of the control as established by the laboratory in accordance with
samples. 2.1.4. The measurement procedure must not be made
(2) If a single measurement of a control sample exceeds available for measurements until the functionality of
the deviation of measurements, the measuring pro- the procedure has been proven through appropriate
cedure will initially be blocked for further use in measures. The entire process has to be documented.
measuring specimens from patients. The cause of
the failure of performance has to be sought and, if
possible, rectified. Taking medical relevance into 2.1.4 E stablishing internal laboratory deviation limits
consideration, the responsible person must decide for measurands not listed in Table B1
whether the measuring procedure can be re-author- (1) In order to establish internal laboratory deviation
ised or whether further measures must be taken, limits for single measurements of control samples of
e.g., whether all of the examinations preceding measurands not listed in Table B1a to c, one control
and including the control examination have to be sample result per day is chosen for a minimum of 15
repeated, or whether the sender has to be notified days, or for a maximum of one control period, for each
about already communicated results. The entire pro- control sample used. Values are selected based on a
cess has to be documented. pattern, i.e., either the first, the nth or the last value
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 37
is used in the calculation. Randomly selected control a single measurement of a control sample has to be
results may also be used. performed at least once a week if the procedure is used
The limits of deviation are then calculated from the during that calendar week to test patient specimens.
target value x0 plus or minus Δmax. The following In the case of devices that do not use electronic/physi-
formula is used to calculate Δmax: cal standards, or where there is no other form of inte-
grated testing of the device’s functionality to prevent
∆max = k 2 ∗ sep
2 2
+ δep , the output of erroneous measuring results, only the
regulations set forth in (2) and (4 letter a) of Section
where: 2.1.1 shall be waived.
k = 3, coverage factor for calculating the internal labo- (3) The evaluation of the single measurements of the
ratory deviation limits. control samples and any resulting consequences
sep, empirical standard deviation of the control sample have to be carried out in accordance with Section
measurements used in the calculations during the 2.1.2 (2). For measurands not listed in Table B1, Sen-
pre-evaluation period. tence 1 applies accordingly. The permitted deviation
δep, systematic deviation of measurement of the limits are those stated by the manufacturer of the
control sample measurements used in the calcula- control samples.
tions during the evaluation period (ep). (4) Calculation and evaluation of the root mean square
For simplification purposes, variance sep 2
is not of the error of measurement according to Section
corrected with (n-1)/n. 2.1.3 are omitted as well as a graphic illustration, as
To calculate relative internal laboratory deviation required in Section 2.1.7 (3).
limits, Δmax is to be divided by the target value x0.
In justified cases an internal laboratory devia-
tion limit that deviate from this procedure can also be 2.1.6 Measurands with low examination frequencies
defined. The reasons and the chosen procedure have (1) Measurands that are likely to be analysed on fewer
to be documented in a transparent way. than 15 days in 3 months have to be verified by at
(2) The acceptability limits of the manufacturer of the least two control samples with target values in differ-
control samples have to be used while the laboratory ent concentration ranges, if available, on the days on
is establishing its own internal laboratory deviation which the patient samples are examined.
limits. (2) The evaluation of single measurements of control
(3) The internal laboratory deviation limits must be samples and the consequences in accordance with
within the interval provided by the manufacturer of Section 2.1.2 (2) have to be performed for all control
the control sample. samples.
(4) No internal laboratory deviation limits need to be cal- Sentence 1, applies accordingly for measurands not
culated for control samples with a batch life of less listed in Table B1. The permitted deviation limits
than 12 weeks. The interval given by the manufacturer are those stated by the manufacturer of the control
of the control samples have to be applied. samples.
(3) Calculation and evaluation of the root mean square
of the error of measurement according to Section
2.1.5 Patient near immediate laboratory examinations 2.1.3 are omitted as well as a graphic illustration, as
with unit-of-use reagents required in Section 2.1.7 (3).
(1) Are unit-use reagents and the corresponding measur-
ing systems applied in patient near immediate diag-
nostic, they have to be checked in accordance with 2.1.7 Documentation
the manufacturer’s instructions on quality control. (1) All results of the internal quality assurance have to be
The results have to be documented. documented in a structured way by measurand and
(2) The provisions set forth in Section 2.1.1 (2), (3) and (4 by type of sample material taking into account the
letter a) may be waived if electronic/physical stan measuring procedure and measuring station. Upon
dards are used daily, or where there is another form the request of the authority responsible for inspecting
of integrated testing of the device’s functionality that adherence to this guideline, the respective documen-
prevents the output of erroneous results. In such cases tation has to be presented.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
38 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
(2) The documentation must include the following: (2) The participant of the EQA examines the EQA samples
(a) Name of the medical laboratory under routine conditions, and conveys results and
(b) Name of the measuring station information as required by the reference institution.
(c) Date and time of the measurement When communicating results, the participant con-
(d) Measurand, sample material and unit firms that the analysis was performed in accordance
(e) Measuring method with this guideline, in the participant’s laboratory,
(f) Measured value of the control sample and under his supervision.
(g) Target value of the control sample (3) The obligations pursuant to (1) do not apply to exami-
(h) The relative or absolute deviation from the target nations with unit-of-use reagents as part of patient
value and the evaluation according to Table B1a near immediate diagnostic in:
to c, Column 3, or the internal laboratory devia- (a) Doctors’ offices and medical services without a
tion limits or the ranges stated by the manufac- central laboratory
turer of the control samples (b) Hospitals, if the central laboratory is responsible
(i) Release or lock flag for internal quality assurance and determines the
(j) Corrective measures taken measurand itself.
(k) Manufacturer, name and lot number of the con- (4) If the participant does not receive a certificate for a
trol sample, and measurand because one of the participant’s results
(l) Name, cipher or signature of the investigator. exceeds the authorised limits of error as specified in
(3) In addition, the measured values of the control sam- Table B1a to c, Column 5, the participant is obliged
ples should be represented graphically. to determine the causes and rectify them if this lies
(4) All measurement results of the quality assurance within their responsibility. The entire process has to
must be stored for 5 years together with the respective be documented.
calculations after the control period and evaluations, (5) The EQA participation certificate and the acquired
as well as the protocols on actions taken when limits EQA certificates have to be retained for a period of 5
of deviation were exceeded, unless longer archiving years unless longer periods of time are stipulated by
periods are stipulated by other regulations. other regulations.
1 2 3 4 5 6
No. Measurand Permissible relative Rili-BAEK applicable Permissible Type of target
deviation of a single result concentration intervals relative deviation value in EQA
or the relative root mean of columns 3 and 5 in EQA
square, respectively
From To Unit
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 39
(Table B1a: Continued)
1 2 3 4 5 6
No. Measurand Permissible relative Rili-BAEK applicable Permissible Type of target
deviation of a single result concentration intervals relative deviation value in EQA
or the relative root mean of columns 3 and 5 in EQA
square, respectively
From To Unit
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
40 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
(Table B1a: Continued)
1 2 3 4 5 6
No. Measurand Permissible relative Rili-BAEK applicable Permissible Type of target
deviation of a single result concentration intervals relative deviation value in EQA
or the relative root mean of columns 3 and 5 in EQA
square, respectively
From To Unit
RMV, reference method value; NV, nominal value specific to measuring method.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 41
1 2 3 4 5 6
No. Measurand Permissible relative deviation Rili-BAEK applicable concentration Permissible Type of target
of a single result or the relative intervals of columns 3 and 5 relative value in EQA
root mean square, respectively deviation in EQA
From To Unit
1 2 3 4 5 6
No. Measurand Permissible relative deviation Rili-BAEK applicable concentration Permissible Type of
of a single result or the relative intervals of columns 3 and 5 relative target value
root mean square, respectively deviation in EQA in EQA
From To Unit
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
42 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
assurance. If several instruments or measuring (3) The control samples must be as similar as possible to
stations are used to perform a medical laboratory the patient samples being examined. Within the same
examination, internal quality assurance has to be measuring procedure, control and calibration materi-
performed on each of these instruments or measur- als must not be identical.
ing stations. (4) Control samples have to be used with known results
(3) In addition, all measurands and nominal characteris- that are within the measurement interval relevant for
tics listed in Table B2-2 of this Part are subject to exter- making medical decisions.
nal quality assurance. (5) When unit-use reagents and their corresponding
(4) The measurands in Tables B2-1 and B2-2 are listed measuring systems are used in patient near immedi-
alphabetically. The criteria used for including a ate examination, the requirements of (1) Sentence 2
measurand in the tables are, specifically, the fre- and (2) Sentence 2 (a) do not need to be met, as long
quency of the measurement procedure and its as process control measures indicating the display of
medical relevance according to the current state erroneous examination results are integrated into the
of science. Tables B2-1 and B2-2 are continuously test.
updated.
(5) This section of the guideline does not apply to qualita-
tive tissue examinations and to examinations whose 2.1.2 Evaluation of the results
internal and external quality assurance requirements (1) The evaluation of results of control sample measure-
are stated in further special Parts B. ments has to be performed without delay. The evalua-
tion is performed using the target objectives assigned
to the control sample.
2 Procedure of quality assurance (2) If the requirements are not met, the measuring pro-
cedure will initially be blocked from further use in
2.1 Internal quality assurance measuring specimens from patients. The cause of
the failure of performance has to be sought and, if
2.1.1 Procedure possible, rectified. Taking medical relevance into
(1) Regarding the type and frequency of internal quality consideration, the responsible person must decide
assurance the specifications of the manufacturer have whether the measuring procedure can be re-author-
to be followed. ised or whether further measures must be taken,
Irrespective of this, internal quality assurance has to e.g., whether all of the examinations preceding
be performed: and including the control examination have to be
(a) In accordance with Table B2-1 for the examina- repeated, or whether the sender has to be notified
tions listed therein about already communicated results. The entire pro-
(b) Adequately and regularly in accordance with med- cess has to be documented.
ical necessity and with the required examination
frequency of patient specimens, in case the exami-
nations are not listed in Table B2-1. 2.1.3 Documentation
The requirements of (1) Sentence 2 are considered (1) All of the results of the internal quality assurance
to be met, if corresponding controls that ensure the have to be documented in a structured way by type of
accurateness of the results, are integrated within the the sample material, taking into account the measur-
applied measuring system. ing procedure and measuring station or instrument.
(2) In addition, internal quality assurance has to be per- Upon the request of the authority responsible for
formed after every intervention to the examination inspection adherence to this guideline the respective
procedure. Interventions to the examination proce- documentation has to be presented.
dure include: (2) The documentation must include the following:
(a) Restarting the device after it has been switched (a) Name of the medical laboratory
off completely (b) Name of the measuring station or device
(b) Calibration by the user
(c) Repair or maintenance work on devices relevant 1 This also includes changes to the composition of the reagents, such
to the medical laboratory examination and as the production of dilutions or, in the case of in-house production,
(d) Changing reagent lots.1 the reiterated preparation of reagents.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 43
(c) Date and, where relevant, time of the examination Table B2-1: Internal quality assurance.
(d) Examination, sample material, and, if necessary,
the unit of measurement No. Measurand/Examination Frequency
of controls
(e) Measuring method
(f) Result of control sample 1 6-Acetylmorphine Daily
2 ABO characteristics Weekly
(g) Target objectives of the control sample 3 Amphetamines Daily
(h) The evaluation 4 Barbiturates Daily
(i) Release or lock flag 5 Benzodiazepines Daily
6 Borrelia burgdorferi, antibodies against Daily
(j) Corrective measures taken 7 Buprenorphine Daily
(k) Manufacturer, name and lot number of the con- 8 Candida albicans, antibodies against Daily
trol sample and 9 Cannabinoides Daily
10 Chromatographic analysis with identification Daily
(l) Name/ cipher or signature of the investigator. of the active substance (STA)
(3) The documentation on the performed internal qual- 11 Cocaine and metabolites Daily
ity assurance has to be stored for 5 years, along with 12 Direct Coombs test Weekly
13 dsDNA, autoantibodies against Daily
the evaluations as well as the protocols of the meas- 14 Echinococcus, antibodies against Daily
ures taken when the target objectives were not met, 15 Electrophoresis with immunoreaction Monthly
unless longer archiving periods are stipulated by 16 Entamoeba histolytica, antibodies against Daily
17 Erythrocyte antigens, antibodies against Daily
other regulations.
(Coombs test)
18 Extractable nuclear antigens, autoantibodies Daily
against
19 Smooth muscle, autoantibodies against Daily
2.2 External quality assurance (EQA)
20 Glutaminase, antibodies against Daily
21 HBc antigen, antibodies against Daily
(1) The medical laboratory is required to participate in an 22 HBe antigen, antibodies against Daily
23 HBs antigen, antibodies against Daily
external quality assurance for every measurand listed
24 Hepatitis A virus, antibodies against Daily
in Table B2-2 in the frequency of participation speci- 25 Hepatitis C virus, antibodies against Daily
fied there. In a medical laboratory comprising several 26 HIV, antibodies against Daily
sites, each site has to participate for the examinations 27 IgE antibodies, allergen-specific single Weekly
allergen test
it provides. Participation is mandatory for the exami- Method-specific control in a rotating
nations listed in the table regardless of whether the procedure with a leading allergen
examination result is quantitatively or qualitatively 28 Immune complexes, circulating Daily
29 Nuclei (ANA), autoantibodies against Daily
stated in the report or in the findings. 30 Methadone and metabolites Daily
(2) The participant in the external quality assessment 31 Methaqualone Daily
programme examines the EQA samples under routine 32 Mitochondria (AMA), autoantibodies against Daily
33 Opiates Daily
conditions, and conveys results and information as 34 Phencyclidine Daily
required by the reference institution. By communi- 35 Plasmodium, antibodies against Daily
cating these results, the participant confirms that the 36 Rhesus type Weekly
37 Rheumatoid factor (RF) Daily
tests were performed in accordance with this guide- 38 Ribonucleoprotein (RNP), autoantibodies against Daily
line, in the participant’s laboratory, and under his 39 Rubella virus, antibodies against Daily
supervision. 40 Schistosoma, antibodies against Daily
41 Scl-70 antigen, autoantibodies against Daily
(3) If a participant does not receive a certificate for an 42 Sm antigen, autoantibodies against Daily
examination because one or more of the participant’s 43 SS-A antigen, autoantibodies against Daily
results do not meet the target objectives of the respec- 44 SS-B antigen, autoantibodies against Daily
45 Streptococcal desoxyribonuclease, antibodies Daily
tive reference institute, the participant is obligated to against
determine the causes and rectify them insofar as this 46 Streptolysin O, antibodies against Daily
lies within the participant’s responsibility. The entire 47 Toxoplasma gondii, antibodies against Daily
48 Treponema pallidum, antibodies against Daily
procedure is to be documented.
49 Tricyclic anti-depressives Daily
(4) Certificates of both participation and successful 50 Cytoplasmic components of neutrophil Daily
participation in the external quality assurance pro- granulocytes (C-ANCA, P-ANCA),
autoantibodies against
gramme have to be kept for 5 years, unless a longer
archiving period is stipulated by other regulations. Daily, each calendar day on which patient samples are tested;
weekly, each calendar week in which patient samples are tested, etc.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
44 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 45
type of pathogen or testing method(s) applied. The (4) Control samples must be as similar as possible to the
criteria used for the listing of an examination in the examined patient samples. Within the same measur-
tables include, specifically, the frequency of the tests ing procedure, control and calibration materials must
applied and their medical relevance according to the not be identical.
current state of the knowledge. The respective tables (5) Control samples with known results have to be used
are continuously updated. unless otherwise stated.
(6) Statistics have to be kept and evaluated with respect
to the frequency of pathogens detected and their sus-
2 Implementation of quality assurance ceptibility to anti-infective agents.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Table B3-1: Internal quality assurance.
Microscopy
Gram stain Characteristic staining of Gram-negative No deviation Daily
and Gram-positive bacteria on a control
specimen (e.g., tongue swab)
Ziehl-Neelsen stain Characteristic staining of acid-fast bacilli No deviation Daily
on a control specimen
Giemsa stain Characteristic staining of erythrocytes and No deviation Daily
leucocytes in a smear, if applicable, from
the patient being tested examined
pH value of the buffer 6.8–7.2 Weekly
Microscopic detection of pathogens, e.g., Recognition of characteristic pathogen A maximum of a 20% deviation (based on Annually
parasites structures, e.g., using image charts or the number of specimens assessed)
external quality assurance programme /
other preserved specimens (“consensus
training”)
Negative contrasting in transmission Use of samples with defined viruses/ No deviations With every new lot of film-coated copper
electron microscopy of viruses virus quantities (testing the integrity of mesh
the carrier film, its binding properties,
negative contrasting and amplification
factor) clear detection of the viruses / virus
Unangemeldet
groups
Culture procedures
Non cell culture-based procedures
Pathogen identification
Examination of individual methods for Pathogen-typical reactions No deviation Daily
(orientating) pathogen identification with
control strains:
catalase, oxidase, indole, coagulase, germ
tube test, urease
Verification of commercial systems for Pathogen-typical reactions No deviation When changing lots
pathogens identification of control strains
Examination of the inoculum purity in Pure culture No deviation Each isolate
commercial systems used in identifying
pathogens
Susceptibility testing
Beta-lactamase Positive and negative controls using No deviation Daily
control strains
Verification of the susceptibility test On 20 consecutive work days using 1 out of 20 findings per pathogen- Before initial use and when requirements
appropriate control strains, evaluation of antibiotic combination outside the of the current internal quality assurance
the findings of the pathogen-antibiotic permissible tolerance range are not met
combinations
On-going internal quality assurance of the Compliance with tolerance ranges for the If deviation occurs more than twice for Weekly and when changing batches; for
susceptibility test normative control strains a pathogen-antibiotic combination: systems used less than once a week: each
Unangemeldet
troubleshooting, correction and, if time the system is used
required, re-verification of the test system
Inoculum purity Examination of the inoculum purity No deviation Each isolate
Reaction components Conformity testing of the reagents No deviation In the case of new reagent batch or newly
(primers, polymerase, nucleotides dissolved reagent
and probes) by nucleic acid/signal
amplification of the target sequence with
old and new reagent batch
Pathogen-specific nucleic acid detection Positive and negative control according to No deviation Each procedure
2.1.2.3
Sequence-based methods (NAT, FISH and Checking database of the primer and probe No deviation impacting on the test results At least once a year or in accordance with
other hybridisation techniques sequences used in each detection method provision by the producer
with respect to the declared species
specificity
Immunological procedures
Diagnostic antibodies Positive and, if applicable, negative control No deviation When changing batches
Antigen detection (EIA, ELFA,CLIA and other Positive and negative controls No deviation Daily
immunochemical detection methods)
Antigen detection using diagnostic rapid Positive and, if applicable, negative control No deviation Once per test package
tests (e.g., immunochromatographic tests)
with integrated function controls (e.g.,
stool pathogens)
Direct immunofluorescence test Positive and negative controls No deviation Daily
Unangemeldet
(e.g., respiratory viruses, Legionella,
Pneumocystis jirovecii, Giardia lamblia)
Antigen detection using particle / Checking function by using known positive No deviation Daily
Daily, every calendar day on which patient samples are tested; weekly, every calendar week in which patient samples are tested.
a
In the case of commercial culture media, this inspection can be documented by a corresponding batch certificate from the manufacturer.
b
Growth, colony morphology and biochemical reactivity are tested using the same control strains if possible. Compliance with the specifications required for culture media (e.g., growth, colony
morphology, if applicable, biochemical reactivity) can also be tested by regularly sub-cultivating of the internal quality assurance control strains.
c
If there are validation data on the efficient nucleic acid extraction from the relevant target organism for closed, mechanised systems, an extraction control, and possibly an inhibition control do
not have to be performed.
48 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 49
Table B3-1a: Internal quality assurance when examining nucleic acid concentration in blood/plasma/serum.
1 2 3 4 5
No. Analyte Permissible absolute deviation Rili-BAEK applicable concentration intervals of Frequency
of the logarithmic (base column 3 of control
10) single value from the examination
From To Unit
logarithmic nominal valuea
Alternatively a control sample can be used that has a designated target interval with a maximum span of one log10 step.
a
(3) The virus dose for cell culture-based neutralisation deviation of the number of cycles from the nominal
tests must be determined and documented using a value (cycle threshold/Ct, crossing point/Cp, cycle
TCID50 assay or a comparable procedure. quantitative/Cq) has to be established. Likewise, the
(4) Sensitive and non-sensitive control strains must be validity interval for the quantitative positive control
used as positive and negative control samples for phe- must be established and documented.
notypic resistance testing. The extent of inhibition by (5) The base sequence of amplification products must be
the antiviral control substance must be documented. verified.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
50 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
1 Gram stain Every half year NV 27 Toxoplasma gondii, genome Every half year RLV
2 Cultivation, identification and Every half year RLV detection
sensitivity testing of bacteria Yeasts
3 Cultivation, identification and Every half year RLV 28 Cultivation and identification Every half year RLV
sensitivity testing of fast growing of yeasts and hyphomycetes
bacteria and, if applicable, (mycoses of mucosa, organ,
detection of accompanying flora systemic or resulting from trauma)
of the urogenital system 29 Identification of dermatophytes, Every half year RLV
4 Bordetella pertussis, genome Every half year NV yeasts and moulds (pathogens
detection causing dermatomycoses and
5 Borrelia burgdorferi sensu lato, Every half year NV mucosal yeast infections)
genome detection 30 Candida, antigen detection Every half year NV
6 Chlamydia pneumoniae, genome Every half year NV 31 Cryptococcus neoformans, Every half year NV
detection antigen detection
7 Chlamydia trachomatis, antigen Every half year NV
Viruses
detection
32 Adenoviruses, genome detection Every half year NV
8 Chlamydia trachomatis, genome Every half year NV
33 Cytomegalovirus, genome Every half year NV
detection
detection
9 EHEC/STEC (Shiga toxin), genome Every half year NV
34 Enteroviruses, genome detection Every half year NV
detection 35 Epstein-Barr virus, genome Every half year NV
10 Helicobacter pylori, genome Every half year NV
detection
detection 36 Hepatitis A virus, genome detection Every half year NV
11 Legionella pneumophila, genome Every half year NV 37 Hepatitis B virus, genome detection Every half year NV
detection 38 Hepatitis B virus, HBs antigen Every half year NV
12 Listeria monocytogenes, genome Every half year NV detection
detection 39 Hepatitis B virus, HBe antigen Every half year NV
13 Methicillin-resistant Every half year NV detection
Staphylococcus aureus (MRSA), 40 Hepatitis C virus, genome detection Every half year NV
genome detection 41 Hepatitis C virus genotyping, Every NV
14 Mycoplasma pneumoniae, Every half year NV genome detection calendar year
genome detection 42 Hepatitis C virus, HCV antigen Every half year NV
15 Neisseria gonorrhoeae, genome Every half year NV detection
detection 43 Herpes simplex virus type 1 / type Every half year NV
16 Salmonella enterica, genome Every half year NV 2, genome detection
detection 44 HIV-1 (RNA), genome detection Every half year NV
17 Coxiella burnetii, genome Every half year NV 45 HIV-1, p24 antigen detection Every half year NV
detection 46 Human papillomavirus, genome Every half year NV
18 Francisella tularensis, genome Every half year NV detection
detection 47 Influenza A and B viruses, Every half year NV
Mycobacteria Every half year NV genome detection
19 Microscopic detection of Every half year NV 48 Influenza A and B viruses, antigen Every half year NV
mycobacteria detection
20 Cultivation of mycobacteria Every half year NV 49 Parvovirus B19, genome detection Every half year NV
21 Differentiation of tuberculosis Every half year NV 50 Respiratory syncytial virus, Every half year NV
bacteria genome detection
22 Susceptibility test of Every half year NV 51 Respiratory syncytial virus, Every half year NV
Mycobacterium tuberculosis antigen detection
23 Identification of mycobacteria Every half year NV 52 Varicella zoster virus, genome Every half year NV
24 Mycobacterium tuberculosis, Every half year NV detection
genome detection a
RLV, reference laboratory value: the target values of the external
Parasites quality assurance programme are calculated by reference laborato-
25 Parasites in the blood, Every half year RLV ries as the arithmetic average or median (if applicable); NV, nominal
microscopic detection value: the target values are calculated from the results of the
26 Parasites in the stool, Every half year RLV external quality assurance programme as the arithmetic average or
microscopic detection median (as appropriate).
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 51
Table B3-2a: External quality assurance when examining nucleic acid concentration in blood/plasma/serum.
1 2 3 4 5 6
No. Analyte Permissible absolute deviation of the Rili-BAEK applicable concentration Target value Frequency of
logarithmic (base 10) single value from the intervalsof column 3 in EQA EQA
logarithmic nominal value in EQA
From To Unit
1 CMV DNA -0.8 to +0.8 5000 5,000,000 IU/mL NV Every half year
2 HBV DNA -0.6 to +0.6 500 5,000,000 IU/mL NV Every half year
3 HCV RNA -0.6 to +0.6 500 5,000,000 IU/mL NV Every half year
4 HIV-1 RNA -0.6 to +0.6 500 5,000,000 Copies/mL NV Every half year
about results already communicated. The entire pro- frequency. In a medical laboratory comprising several
cess has to be documented. sites, each site has to participate for the examinations
it provides.
(2) The participant in the external quality assurance pro-
2.1.4 Documentation gramme should examine the EQA samples under rou-
(1) All of the results of the internal quality assurance tine conditions, and convey results and information
have to be documented in a structured way accord- as required by the reference institution. By communi-
ing to type of the sample material, taking into account cating these results, the participant confirms that the
the measurement procedure and site or device. Upon examinations were performed in accordance with this
request of the inspection authority charged with con- guideline, in the participant’s laboratory, and under
trol of adherence to this guideline the respective doc- his supervision.
umentation has to be presented. (3) If a participant does not receive a certificate for an
(2) The documentation must include the following: examination because one or more of the results do not
(a) The name of the medical laboratory meet the target objectives of the respective reference
(b) The name of the site or device institute, the participant is obligated to determine the
(c) Date and, if relevant, time of the test causes and rectify them insofar as this lies within the
(d) Examination, samples and, if required, the unit of participant’s responsibility. The entire procedure has
measurement to be documented.
(e) Examination method (4) Certificates of both participation and successful par-
(f) Result of the control sample ticipation in the external quality assessment pro-
(g) Objectives for the control sample gramme must be kept for 5 years, unless a longer
(h) Evaluation archiving period is stipulated by other regulations.
(i) Release or lock flag
(j) Corrective measures taken
(k) Manufacturer, name and batch number of the
control sample as appropriate and B4 Examination of ejaculate
(l) Name/cipher or signature of the investigator.
(3) The documentation on the performed internal quality 1 Principles of quality assurance
assurance must be stored for a minimum of 5 years,
along with the evaluations as well as the protocols of (1) The minimum requirements of quality assurance for
the corrective measures taken when the target objec- the results of ejaculate examinations are specified in
tives were not met, unless longer archiving periods Part B4. These minimum requirements include inter-
are stipulated by other regulations. nal and external quality assurance.
(2) Ejaculate examinations, as defined in this guideline,
are examinations for sperm concentration, motility
2.2 External quality assurance (EQA) and morphology.
(3) All of the ejaculate examinations performed by the
(1) The medical laboratory is required to participate in medical laboratory are subject to internal and external
an external quality assurance for every examination quality assurance. If a test is performed on multiple
listed in Tables B3-2 and B3-2a in the therein specified devices or at multiple sites, internal quality assurance
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
52 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 53
to be blocked from being used for measuring patient spec- (3) If a participant does not receive a certificate for an
imens. The measuring procedure cannot be released until examination because one or more of the participant’s
the functionality of the procedure has been established results do not meet the target objectives of the respec-
through suitable measures. The entire process has to be tive reference institute, the participant is obligated
documented. to determine the causes and rectify them, if this lies
within the participant’s responsibility. The entire pro-
cedure has to be documented.
2.1.3 Documentation (4) Certificates of both participation and successful par-
(1) All of the results of the internal quality assurance ticipation in the external quality assessment pro-
have to be documented in a structured way by type of gramme have to be kept for 5 years, unless a longer
the sample material, taking into account the measur- archiving period is stipulated by other regulations.
ing procedure and measuring station or device. Upon
the request of the authority responsible for inspection
of adherence to this guideline the respective docu-
B5 Molecular-genetic and cytogenetic
mentation has to be presented
(2) The documentation must include the following: medical laboratory examinations
(a) The name of the medical laboratory
1 Principles of quality assurance
(b) The name of the measuring station
(c) The period of evaluation
(1) Part B5 specifies minimum requirements to assess
(d) Examination, specimen, unit of measurement
the quality of molecular-genetic and cytogenetic
(e) Examination methods (counting chamber used,
medical laboratory examinations. These minimum
dyeing technique)
requirements include internal and external quality
(f) Examination results including the single values
assurance.
of the repeat determinations
(2) Molecular-genetic and cytogenetic medical laboratory
(g) Evaluation in accordance with the corresponding
examinations, as defined in this Part of the guide-
formula
line, are all medical laboratory examinations on the
(h) Release or lock flag
human genome and transcriptome aiming to detect
(i) Corrective measures taken, and
known sequence variants, identify unknown vari-
(j) Name/cipher or signature of the investigator
ants, and establish the structure or copy number of
(3) The documentation on the performed internal qual-
genomic segments or to detect epigenetic modifica-
ity assurance has to be stored for 5 years, along with
tions of genomic segments. They include molecular
the evaluations as well as the protocols of the meas-
karyotyping using array analysis (e.g., array CGH,
ures taken when the target objectives were not met,
SNP arrays).
unless longer archiving periods are stipulated by
Cytogenetic medical laboratory examinations, as
other regulations.
defined in this Part of the guideline, include all
medical laboratory examinations of postnatal cytoge-
2.2 External quality assurance (EQA) netic diagnostics, prenatal cytogenetic diagnostics
and cytogenetic diagnosis of tumours.
(1) Every location of the medical laboratory has to par- (a) Postnatal cytogenetic diagnostics, as defined in
ticipate once every half-year in an EQA to examine this guideline, is the cytogenetic investigation of
concentration, morphology and motility if the medi- a blood sample, tissue sample, cytologic smear or
cal laboratory provides this examination. a cell culture from a body tissue after birth.
(2) The participant in the external quality assessment (b) Prenatal cytogenetic diagnostics, as defined
programme examines the EQA samples under rou- in this guideline, is the cytogenetic investiga-
tine conditions and conveys results and information tion of amniotic cells, chorionic villi or foetal
as required by the reference institution. By commu- lymphocytes.
nicating the results, the participant confirms that the (c) Cytogenetic diagnosis of a tumor, as defined in
examinations were performed in accordance with this this guideline, is the analysis of neoplastic cells.
guideline, in the participant’s laboratory, and under This includes the analysis of cells from bone mar-
their supervision. row, blood, lymph nodes and other tissues.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
54 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
Cytogenetic medical laboratory examinations in the (b) Regularly and adequately according to medical
areas of postnatal cytogenetic diagnostics, prenatal necessity and testing frequency of patient sam-
cytogenetic diagnostics and cytogenetic diagnosis ples if the examinations are not listed in Tables
of tumours include the application of conventional B5-1 and B5-2a.
molecular cytogenetics [ISH, usually fluorescence in The requirements of (1) Sentence 1 may be waived
situ hybridisation (FISH)]. if suitable controls are integrated into the applied
(3) All of the molecular-genetic and cytogenetic exami- analysis system that ensures the accuracy of the
nations performed by the medical laboratory (mea results.
surands and nominal characteristics) are subject to (2) Additionally, internal quality assurance has to be
internal quality assurance. If an examination is per- performed after there has been an intervention to the
formed on multiple instruments or at multiple meas- examination procedure.
urement stations, internal quality assurance has to be Interventions to the examination procedure include:
performed at each of these instruments or measuring (a) Restarting the device after it has been switched
stations. off completely
(4) In addition, all of the examinations listed in Column (b) Calibration by the user
7 of Table B5-1 and the quantities listed in Table B5-2b (c) Repair or maintenance work on devices relevant
are subject to external quality assurance. For molecu- for the results of the medical laboratory examina-
lar genetic examinations that are not listed in Table tion, and
B5-1 Column 7, external quality assurance has to be (d) Changing reagent lots.3
performed in the form of participation in an external
quality assessment programme that examines the 2. Molecular-genetic medical laboratory examinations
method used, if such an external quality assessment (1) The control samples must be as similar as possible to
programme is offered. The requirements of Part B5, the patient samples being examined. Within the same
Sentence 2 are deemed to be met when the applied examination procedure control material and calibra-
methodology has been included in an external qual- tion material must not be identical.
ity assessment programme listed in Table B5-1 and (2) Control samples with known results have to be used.
participation has occurred. When examining for known sequence variations, the
(5) The examinations and quantities are divided into control samples shall represent the known alleles or
molecular-genetic and cytogenetic examinations in allelic ranges of the known sequence variants, struc-
Tables B5-1, B5-2a and B5-2b. They are included in ture variants, or copy number of genomic segments,
the tables based on the frequency of their testing and if available.
their medical significance in accordance with the (3) Where nucleic acid amplification procedures are used
current state of science. The tables are continuously for examining, control samples capable of detecting
updated. contaminations must be used.
(4) In the case of array analyses, known control param-
eters must be used to ensure that, at minimum, the
requirements of the manufacturers have been met for
2 Procedure of quality assurance
the analysis.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 55
2.1.2 Evaluation of results If the limits given in Column 4 of Table B5-2a are
2.1.2.1 For molecular-genetic medical laboratory exceeded, the examination procedure has to be initially
examinations based on control sample blocked for testing further patient samples. The examina-
examinations tion procedure cannot be released until the functionality
(1) The control sample examination and/or the control of the procedure has been established through suitable
measurand have to be evaluated as soon as the results measures. If it is likely that less than 50 results of patient
are available. The assessment is done based on the samples will be released within a 3-month period, it is not
target objectives. necessary to calculate the relative percentage of violation
(2) If the requirements are not met, the examination pro- of limits as specified in Sentence 1. If, in this case of low
cedure will initially be blocked from further use in testing numbers, the limits set forth in Column 3 of Table
examining patient specimens. The cause of the fail- B5-2a are exceeded five times in 3 months, action has to be
ure of performance has to be sought and, if possible, taken as specified in Sentences 4 and 5.
rectified. Taking medical relevance into considera- The entire process has to be documented.
tion, the responsible person must decide whether the
examination procedure method can be re-authorised
2.1.4 Documentation
or whether further measures must be taken, e.g.,
(1) All of the results of the internal quality assurance
whether all of the examinations preceding the control
have to be documented in a structured way by type of
sample and including the control examination have
the sample material, taking into account the measur-
to be repeated or whether the sender needs to be noti-
ing procedure and measuring station or device. Upon
fied about already communicated results. The entire
the request of the authority responsible for inspection
process has to be documented.
adherence to this guideline the respective documen-
tation has to be presented.
(2) The documentation must include the following:
2.1.2.2 For cytogenetic medical laboratory examinations
(a) Name of the medical laboratory
based on the quantities
(b) Name of the measuring station or device
The quality of each patient sample has, where appropriate,
(c) Date and, where relevant, time of the examination
to be tested according to the quantities listed in Table B5-2a.
(d) Examination, sample material, and, if necessary,
If one of the quantities exceeds the limits given in Column
the unit of measurement
4 of Table B5-2a, the person responsible has to decide
(e) Examination method
whether the patient sample should be re-examined. If the
(f) Result of control sample or quantity
limit is also exceeded when the examination is repeated,
(g) Specification of the control sample or quantity
the cause must be sought and, if possible, rectified. Based
(h) The evaluation
on medical relevance, the person responsible has to decide
(i) Release or lock flag
whether test results can still be obtained using this sample
(j) Corrective measures taken
and, with corresponding comments in the findings, be
(k) Manufacturer, name and lot number of the con-
used.
trol sample, as far as applicable, and
(l) Name/cipher or signature of the investigator.
(3) The documentation on the performed internal qual-
2.1.3 A
ssessing the results of cytogenetic examination
ity assurance has to be stored for 5 years, along with
based on quantities at the end of a control period
the evaluations as well as the protocols of the meas-
A control period generally comprises 1 calendar month.
ures taken when the target objectives were not met,
If more than 50 approved results of patient samples
unless longer archiving periods are stipulated by
are obtained after the period of a calendar month, the
other regulations.
medical laboratory has to calculate the relative percent-
age, by which the limits given in Column 4 of Table B5-2a
are exceeded. 2.2 External quality assurance (EQA)
If less than 50 approved results of patient samples
are available, the period of time has to be extended by (1) The medical laboratory is required to participate in
1-month intervals until at least 50 of the same results are an external quality assurance for every examination
obtained. The total time period is not allowed to exceed 3 or quantity listed in Tables B5-1 Column 7 and B5-2b
months. in the frequency of participation specified there. In a
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
56 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
Tables B5
Daily, every calendar day on which patient samples are tested; weekly, every calendar week in which patient samples are tested, etc.
a
Due to genetic heterogeneity, “molecular genetic categories” are – by definition of the placeholder concept – listed as classifiers of
genetic alterations:
Point mutation and/or single nucleotide polymorphism (MUT/SNP), insertion/deletion (IN/DEL), changes in the copy number of a genomic
segment or a genomic sub-segment (CNV), repeat expansion (EXP), methylation defect (METH).
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 57
medical laboratory comprising several sites, each site (b) The chairs of the Expert Groups listed in Chapter B
has to participate for the examination it provides. (c) A representative from the German Medical
(2) The participant in the external quality assessment Association
programme examines the EQA samples under rou- (d) A representative from the National Association of
tine conditions and conveys results and information Statutory Health Insurance Physicians
as required by the reference institution. By commu- (e) A representative from the German Hospital
nicating the results, the participant confirms that the Federation
tests were performed in accordance with this guide- (f) A representative from the German Association of
line, in the participant’s laboratory, and under their Medical Technologists and Analysts
supervision. (g) A representative from a competent industrial
(3) If a participant does not receive a certificate for an association
examination because one or more of the participant’s (h) Three state representatives
results do not meet the target objectives of the respec- (i) A representative from the German Federal Minis-
tive reference institute, the participant is obligated try for Health
to determine the causes and rectify them, if this lies (j) A representative from the Federal Institute for
within the participant’s responsibility. The entire pro- Drugs and Medical Devices (BfArM), and
cedure has to be documented. (k) A representative from the Physikalisch-Tech
(4) Certificates of both participation and successful par- nische Bundesanstalt (PTB)
ticipation in the external quality assessment pro- (5) The business of the Advisory Board shall be admin-
gramme have to be kept for 5 years, unless a longer istrated by the German Medical Association. The
archiving period is stipulated by other regulations. German Medical Association shall bear the costs for
conducting the Advisory Board meetings. The partici-
pation costs for the members shall be borne by the
delegating institutions.
C Advisory Board (6) The Advisory Board shall issue for itself and the
Expert Groups by procedural rules in accordance with
(1) An Advisory Board “Quality assurance in Medical Part D of this guideline.
Laboratory Examination” shall be established at the
German Medical Association that shall primarily per-
form the following duties:
(a) Advising the German Medical Association in all
D Expert Groups
aspects of this guideline
(1) There shall be an Expert Group for each Part B.
(b) Dealing with questions pertaining to the applica-
(2) The composition of the Expert Group and their tasks
tion of this guideline
shall be set forth in the special D Parts.
(c) Collecting, assessing and formulating sugges-
(3) The business of the Expert Groups shall be admin-
tions for updating this guideline.
istrated by the German Medical Association. The
(2) The members of the Advisory Board shall be recom-
German Medical Association shall bear the costs for
mended by the institutions listed under (4) below,
conducting the Expert Group meetings. Participation
and appointed by the Executive Board of the German
costs incurred by the members shall be borne by the
Medical Association, for a period of 4 years. Extraor-
delegating institutions.
dinary appointments during the current term shall
remain in effect until the end of the term. Re-appoint-
ments are permitted. The Advisory Board shall elect a
chairman from among its members. The members of D1 Expert Group “Quantitative Medical
the Advisory Board may be represented by proxy, with Laboratory Examinations”
approval of the chairman.
(3) The Advisory Board may commission external experts. (1) The Expert Group “Quantitative Medical Laboratory
(4) The Advisory Board is made up of representatives Examinations” shall be established at the German
from the following institutions: Medical Association and have the following tasks:
(a) Three representatives from the relevant medical (a) Advising the German Medical Association in all
scientific societies questions pertaining to Parts B1 and E1
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
58 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
Postnatal analyses
Lymphocytes Banding resolution At least 400 bphs A max. of 5% of the samples with banding
resolution < 400 bphs
Number of overlapping points At most 12 per A maximum of 5% of the samples > 12
in the case of bphs < 400 metaphase
Number of overlapping points At most 20 per A maximum of 5% of the samples > 20
in the case of bphs ≥ 400 metaphase
Degree of lightness At least 3 A maximum of 5% of the samples < 3
Prenatal analyses
Amniotic cells Banding resolution At least 400 bphs A maximum of 5% of the samples < 400 bphs
Chorionic villus cells Banding resolution At least 300 bphs A maximum of 5% of the samples < 300 bphs
Amniotic and chorionic villus Number of overlapping points At most 12 per A maximum of 5% of the samples > 12
cells at bphs < 400 metaphase
Number of overlapping points At most 20 per A maximum of 5% of the samples > 20
in the case of bphs ≥ 400 metaphase
Degree of lightness At least 3 A maximum of 5% of the samples < 3
FISH (interphase) constitutional Hybridisation efficiency N/A A maximum of 10% without signals of the
and tumour cytogenetics control probe
Postnatal analyses
Lymphocytes Nominal chromosome numbera No deviation Calendar year
Banding resolution None of the samples < 400 bphs Calendar year
Number of overlapping points in None of the samples > 12 Calendar year
the case of bphs < 400
Number of overlapping points in None of the samples > 20 Calendar year
the case of bphs ≥ 400
Degree of lightness None of the samples < 3 Calendar year
Prenatal analyses
Nominal chromosome numbera No deviation Calendar year
Amniotic cells Banding resolution None of the samples < 400 bphs Calendar year
Chorionic villus cells Banding resolution None of samples < 300 bphs Calendar year
Amniotic and chorionic villus cells Number of overlapping points in None of samples > 12 Calendar year
the case of bphs < 400
Number of overlapping points in None of samples > 20 Calendar year
the case of bphs > 400
Degree of lightness None of samples < 3 Calendar year
FISH (interphase) constitutional Hybridisation efficiency None of samples > 10% without Calendar year
and tumour cytogenetics signal of the control probe
Molecular cytogenetics DLRS value None of samples > 0.4 Calendar year
(OligoArray)
Nominal chromosome number, e.g., 45, X (Turner syndrome); 46, XX (normal female); 47, XXY (Klinefelter syndrome).
a
(b) Establishing the pass modalities for the external (2) The members of this Expert Group are recommended
quality assurance programmes by the institutions listed in (3) below, and are
(c) Dealing with questions pertaining to the applica- appointed by the Executive Board of the German Med-
tion of Parts B1 and E1 ical Association for a period of 4 years. Extraordinary
(d) Collecting, assessing and formulating sugges- appointments during the current term shall remain in
tions for updating Parts B1 and E1 effect until the end of the term. Re-appointments are
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 59
permitted. The Expert Group shall elect a chairman (b) A representative from the German Medical
from among its members. The members of the Expert Association
Group may be represented by proxy with approval of (c) A representative from the National Association of
the chairman. Statutory Health Insurance Physicians
The Expert Group may commission external experts. (d) A representative from the German Hospital
(3) Members of the Expert Group include: Federation
(a) Three representatives from relevant medical sci- (e) A representative from the German Association of
entific societies Medical Technologists and Analysts
(b) A representative from the German Medical (f) A representative from a competent industrial
Association association
(c) A representative from the National Association of (g) Two state representatives
Statutory Health Insurance Physicians (h) A representative from the Physikalisch-Technis-
(d) A representative from the German Hospital che Bundesanstalt (PTB)
Federation
(e) A representative from the German Association of
Medical Technologists and Analysts
D3 Expert Group “Direct Detection and
(f) A representative from a competent industrial
association Characterisation of Infectious Agents”
(g) Two state representatives, and
(1) The Expert Group “Quality assurance for Medical Lab-
(h) A representative from the Physikalisch-Technis-
oratory Examinations for Direct Detection and Char-
che Bundesanstalt (PTB)
acterisation of Infectious Agents” shall be established
at the German Medical Association and have the fol-
D2 Expert Group “Qualitative lowing tasks:
Medical Laboratory Examinations” (a) Advising the German Medical Association in all
questions pertaining to Parts B3 and E3
(1) The Expert Group “Qualitative Medical Laboratory (b) Establishing the pass modalities for the external
Examinations” shall be established at the German quality assurance programmes
Medical Association and have the following tasks: (c) Dealing with questions in the application of Parts
(a) Advising the German Medical Association in all B3 and E3,
questions pertaining to Parts B2 and E2 (d) Collecting, assessing and formulating sugges-
(b) Establishing the pass modalities for the external tions for updating Parts B3 and E3.
quality assurance programmes (2) The members of this Expert Group are recommended
(c) Dealing with questions pertaining to the applica- by the institutions listed in (3) below, and are
tion of Parts B2 and E2 appointed by the Executive Board of the German Med-
(d) Collecting, assessing and formulating sugges- ical Association for a period of 4 years. Extraordinary
tions for updating Parts B2 and E2 appointments during the current term shall remain in
(2) The members of this Expert Group are recommended effect until the end of the term. Re-appointments are
by the institutions listed in (3) below, and are permitted. The Expert Group shall elect a chairman
appointed by the Executive Board of the German Med- from among its members. The members of the Expert
ical Association for a period of 4 years. Extraordinary Group may be represented by proxy with approval of
appointments during the current term shall remain in the chairman.
effect until the end of the term. Re-appointments are The Expert Group may commission external
permitted. The Expert Group shall elect a chairman experts.
from among its members. The members of the Expert (3) Members of the Expert Group include:
Group may be represented by proxy with approval of (a) Five representatives from relevant medical scien-
the chairman. tific societies
The Expert Group may commission external experts. (b) A representative from the German Medical
(3) Members of this Expert Group include: Association
(a) Five representatives from relevant medical scien- (c) A representative from the National Association of
tific societies Statutory Health Insurance Physicians
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
60 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
(d) A representative from the German Hospital (f) A representative from a competent industrial
Federation association
(e) A representative from the German Association of (g) A state representative
Technologists and Analysts (h) A representative from the Physikalisch-Tech
(f) A representative from a competent industrial nische Bundesanstalt (PTB)
association
(g) A state representative, and
(h) A representative from the Physikalisch-Tech- D5 Expert Group “Molecular-
nische Bundesanstalt (PTB), the Robert Koch- genetic and Cytogenetic
Institute (RKI), the Paul-Ehrlich-Institute (PEI) Medical Laboratory Examination”
and the Federal Institute for Drugs and Medical
Devices (BfArM), respectively. (1) The Expert Group “Molecular-genetic and Cytogenetic
Medical Laboratory Examination” shall be estab-
lished at the German Medical Association and have
D4 Expert Group “Examination of the following tasks:
ejaculate” (a) Advising the German Medical Association in all
questions pertaining to Parts B5 and E5
(1) The Expert Group “Examination of ejaculate” shall be (b) Establishing the pass modalities for the external
established at the German Medical Association and quality assurance programmes
have the following tasks: (c) Dealing with questions pertaining to the applica-
(a) Advising the German Medical Association in all tion of Parts B5 and E5
questions pertaining to Parts B4 and E4 (d) Collecting, assessing and formulating sugges-
(b) Establishing the pass modalities for the external tions for updating Parts B5 and E5.
quality assurance programmes (2) The members of this Expert Group are recommended
(c) Dealing with questions pertaining to the applica- by the institutions listed in (3) below, and are
tion of Parts B4 and E4 appointed by the Executive Board of the German Med-
(d) Collecting, assessing and formulating sugges- ical Association for a period of 4 years. Extraordinary
tions for updating Parts B4 and E4. appointments during the current term shall remain in
(2) The members of this Expert Group are recommended effect until the end of the term. Re-appointments are
by the institutions listed in (3) below, and are permitted. The Expert Group shall elect a chairman
appointed by the Executive Board of the German Med- from among its members. The members of the Expert
ical Association for a period of 4 years. Extraordinary Group may be represented by proxy with approval of
appointments during the current term shall remain in the chairman.
effect until the end of the term. Re-appointments are The Expert Group may commission external experts.
permitted. The Expert Group shall elect a chairman (3) Members of this Expert Group include:
from among its members. The members of the Expert (a) Three representatives from relevant medical sci-
Group may be represented by proxy with approval of entific societies
the chairman. (b) A representative from the German Medical
The Expert Group may commission external experts. Association
(3) The Expert Group includes: (c) A representative from the National Association of
(a) Five representatives from relevant medical scien- Statutory Health Insurance Physicians
tific societies (d) A representative from the German Hospital
(b) A representative from the German Medical Federation
Association (e) A representative from the German Association of
(c) A representative from the National Association of Medical Technologists and Analysts
Statutory Health Insurance Physicians (f) A representative from a competent industrial
(d) A representative from the German Hospital association
Federation (g) A state representative
(e) A representative from the German Association of (h) A representative from the Physikalisch-Tech
Medical Technologists and Analysts nische Bundesanstalt (PTB)
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 61
(i) A representative from the Robert Koch-Institute conjunction with reference laboratories and labo-
(RKI) ratories for determination of nominal values
(e) Taking further measures if problems with the
external quality assurance programme samples
arise, and involving the affected manufacturer if
E General requirements for necessary.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
62 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
(4) The reference institutions charge suitable reference The certificate is valid for 6 months.
laboratories with the determination of the reference (9) If the reference institution establishes that partici-
method values of samples for external quality assur- pants frequently are not receiving a certificate for a
ance as long as this is required according to Table 1a measurand with reagents or devices from a particu-
to c of Part B1. The reference laboratories are deemed lar manufacturer and if the cause for this cannot be
suitable if they are accredited calibration laboratories traced to the medical laboratories taking part in the
in accordance with DIN EN ISO/IEC 17025 and DIN external quality assurance programme or the refer-
EN ISO 15195. This only applies to measurands where ence institution, the appropriate federal authorities
an accreditation is offered by an accreditation body. have to be notified if this can be defined as an “inci-
Only those accreditation bodies that are included in dent” according to Section 2 of Germany’s Medical
the Multilateral Agreement on the Mutual Acceptance Devices Safety Plan Ordinance.
of Calibration Certificates of the European Co-oper- (10) Further details for conducting external quality assurance
ation for Accreditation (EA) may be used. Moreover, programmes and analysing external quality assurance
the leader of a reference laboratory must have special programme test results are set forth in implementation
expertise and experience in the area of the reference regulations. These are published by the German Medical
measurement methods and be capable of examining Association and by the reference institutions.
new methods.
(5) For every external quality assurance programme, the
reference institutions commission each participant
2 Determining target values
to examine at least two external quality assurance
programme samples with varying concentrations or
(1) After consulting with its competent committees and
activities of the measurands.
after formal consultation of the parties concerned,
(6) The reference institutions send the external quality
the German Medical Association establishes and
assurance programme samples to each external qual-
announces the type of target value has to be used for
ity assurance programme participant with informa-
the measurands. Reference methods have to be used,
tion on handling the samples and conveying their
where possible, when determining the target values in
measurement results.
control samples.
(7) Reference institutions shall only analyse measurement
(2) The reference institutions establish the test plans for
results that were submitted by the external quality assur-
determining the target values of the external quality
ance programme participant before the set deadline.
assurance programme samples, commission the ref-
(8) A certificate showing the submission date of the exter-
erence laboratories, analyse the measurement results
nal quality assurance programme has to be issued to
and merge these into a target value.
every external quality assurance programme partici-
(3) The reference institutions must store the documen-
pant giving the information which of his results from
tation used in determining the target values for
the examinations of these measurands are within the
a period of at least 5 years beginning from when
admissible evaluation limits. In addition, a participa-
they were used in the external quality assurance
tion certificate has to be issued for all measurands
programmes.
where participation in an external quality assurance
programme occurred. The certificate and participa-
tion certificate shall be sent to the participants no
later than four weeks before the next external quality 2.1 Determining reference method values
assurance programme.
External quality assurance programme partici- (1) The reference laboratory commissioned by the refer-
pants are also to be informed of: ence institution uses a reference method to determine
(a) Target values and evaluation limits of the exter- the reference method value for a measurand.
nal quality assurance programme samples (2) The reference method values for external quality
(b) The position and measures of scattering of the assurance programme samples must be available
measurement results of all participants and the before the start of the external quality assurance
measurement procedures used programme. Exceptions are permitted in special
(c) Number of participants, as appropriate, listed cases (e.g., very limited shelf life of the control
according to measurement procedure. sample).
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 63
2.2 Determining the nominal values that they will conduct for the examinations in accord-
ance with (1). This announcement includes:
The nominal values depend on the measurement method (a) The deadline for registrations for participation in
and are calculated from the external quality assurance the external quality assurance programmes
programmes as an arithmetic average or as a median. (b) The respective date of sample shipment and the
deadline for sending back the results
(c) The examinations involved in the external quality
3 A
ssessing the external quality assurance assurance programme, if necessary, with details
programme results on the examination procedure
(d) Type of specimen, the sample volumes of the liq-
(1) Assessment is performed based on column 5 in Table uid or reconstituted external quality assurance
B1a-c. programme samples.
(2) If the entire population or method-dependent sub- (3) The reference institutions select the external quality
populations of the participants’ results show a con- assurance programme samples and check their suit-
siderable deviation to the target value, i.e., a deviation ability. The suitability of the selected external qual-
which influences the pass rate, the reference institu- ity assurance external programme samples for these
tions must research the cause and, if possible, rectify examinations, whose evaluation is carried out on the
this in co-operation with the affected manufacturer basis of reference method values, needs to be checked
of the external quality assurance programme sample prior to their use in the external quality assurance
or with experts. They are to check whether, in such programmes under routine conditions and using a
a case, extending the pass limits or changing the routine measurement procedure.
target value would allow for a proper assessment. (4) For every external quality assurance programme, the
They decide whether the results have to be analysed reference institutions commission each participant to
according to the acceptance limits listed in column 5 examine at least two external quality assurance pro-
or according to the modified pass limits, or whether gramme samples.
the external quality assurance programme has to be (5) The reference institutions send the external quality
repeated for this measurand. assurance programme samples to each external qual-
The process has to be substantiated and documented. ity assurance programme participant with informa-
The participants of the external quality assurance tion on handling the samples and conveying their
programme participants and the Expert Group at the measurement results.
German Medical Association in accordance with Part (6) Reference institutions shall only analyse examina-
D1 have to be informed. tion results that were submitted by the external qual-
ity assurance programme participant before the set
deadline.
(7) A certificate showing the submission date of the
E2 Special requirements for external quality
external quality assurance programme has to be
assurance programme of qualitative issued to every external quality assurance pro-
medical laboratory examination gramme participant giving the information which
of his results from the examinations are within the
1 Obligations of the reference institutions admissible evaluation limits. In addition, a participa-
tion certificate has to be issued for all examinations
(1) Each of the reference institutions ensure that a suf- where participation in an external quality assurance
ficient number of external quality assurance pro- programme occurred. The certificate and participa-
grammes are offered for all of the examinations listed tion certificate shall be sent to the participants no
in Table B2-2 so that every medical laboratory can par- later than four weeks before the next external quality
ticipate at the intervals stipulated in Table B2-2. They assurance programme date.
may only deviate from this if they can prove that there External quality assurance programme participants
is an insufficient amount of suitable external quality are also to be informed of:
assurance programme samples. (a) Target results and, if applicable, evaluation lim-
(2) The reference institutions announce for 1 year in its of the external quality assurance programme
advance, the external quality assurance programmes samples
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
64 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
(b) Position and distribution of the examination samples. The external quality assurance programme
results of all participants, and the examination organisations have to inform the participants of the
procedures used assessment criteria.
(c) Number of participants, as appropriate, listed (2) If the entire population or method-dependent sub-
according to examination procedure. populations of the participants’ results show a consid-
The certificate is valid for double the interval given in erable deviation to the target result, i.e., a deviation
Table B2-2 which influences the pass rate, the reference institu-
(8) If the reference institution establishes that partici- tions must search for the cause and, if possible, to
pants frequently are not receiving a certificate for an rectify this in co-operation with the affected manufac-
examination with reagents or devices from a particu- turer of the external quality assessment programme
lar manufacturer, and if the cause for this cannot be sample, the manufacturers of the respective exami-
traced to the medical laboratories taking part in the nation systems or with experts. They have to check
external quality assurance programme or the refer- whether, in such a case, changing the target result
ence institution, the appropriate federal authorities would allow for a proper assessment. They decide
have to be notified if this can be defined as an “inci- whether the external quality assurance programme
dent” according to Section 2 of Germany’s Medical has to be repeated for this measurand.
Devices Safety Plan Ordinance. This process has to be substantiated and documented.
(9) Further details for conducting external quality assur- The participants of the external quality assurance
ance programmes and analysing external quality programme and the Expert Group at the German
assurance programme results are set forth in the Medical Association are to be informed in accordance
implementation regulations. These are published by with Part D2.
the German Medical Association and by the reference
institutions.
E3 Special requirements for external
quality assurance programme of
2 Determining the target results medical laboratory detection and
characterisation of infectious agents
(1) After consulting with its competent committees and
after formal consultation of the parties concerned, 1 Obligations of the reference institutions
the German Medical Association establishes and
announces the type of target result to be used for the (1) Each of the reference institutions ensure that a suf-
examinations. Reference methods have to be used, ficient number of external quality assurance pro-
where possible, when determining the target values grammes are offered for the examinations listed in
in control samples. Tables B3-2 and B3-2a so that every medical labora-
(2) The reference institutions establish the examination tory can participate at the intervals stipulated in
plans for determining the target values of the external Tables B3-2 and B3-2a. They may only deviate from
quality assurance programme samples, commission this if they can prove that there is an insufficient
the reference laboratories, analyse the results and amount of suitable external quality assessment pro-
merge these into a target result. gramme samples.
(3) The reference institutions must store the documen- (2) The reference institutions announce for 1 year in
tation used in determining the target results for a advance the external quality assurance programmes
period of at least 5 years starting from when they were that they will conduct for the examinations in accord-
used in the external quality assurance programmes. ance with (1). This announcement includes:
(a) The deadline for registrations for participation in
the external quality assurance programmes
3 A
ssessing the external quality assurance (b) The respective date of sample shipment and the
programme results deadline for sending back the results
(c) The examinations involved in the external quality
(1) Assessment is performed based on the target results. assurance programme, if necessary, with details
The assessment criteria must be fulfilled for all on the examination procedure
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 65
(d) Type of specimen, the sample volumes of the liq- particular manufacturer and if the cause for this can-
uid or reconstituted external quality assurance not be traced to the medical laboratories taking part
programme samples in the external quality assurance programme or the
(e) Pass modalities. reference institution, the appropriate federal authori-
(3) The reference institutions select the external quality ties are to be notified if this can be defined as an “inci-
assurance programme samples and check their suit- dent” according to Section 2 of Germany’s Medical
ability. The suitability of the selected external quality Devices Safety Plan Ordinance.
assurance programme samples needs prior to their (9) Further details for conducting external quality assur-
use in the external quality assurance programme ance programmes and analysing external quality
be checked under routine conditions using routine assurance programme results are set forth in imple-
examination procedures. mentation regulations. These are published by the
(4) For every external quality assurance programme, the German Medical Association and by the reference
reference institutions commission each participant to institutions.
examine at least two external quality assurance pro-
gramme samples.
(5) The reference institutions send the external quality 2 Determining target results
assurance programme samples to each external qual-
ity assurance programme participant with informa- (1) After consultation with its competent committees,
tion on handling the samples and conveying their and after formal consultation of the parties con-
measurement results. cerned, the German Medical Association establishes
(6) Reference institutions shall only analyse examina- and announces the type of target result to be used for
tion results that were submitted by the external qual- the examinations and how the target results have to
ity assurance programme participant before the set be determined. Reference methods have to be used
deadline. whenever possible to determine target results in con-
(7) A certificate showing the submission date of the exter- trol samples.
nal quality assurance programme submission date (2) The reference institutions establish the test plans for
has to be issued to every external quality assurance determining the target results of the external qual-
programme participant giving the information which ity assurance programme samples, commission the
of his results from the examinations of these measur- reference and normal value laboratories, analyse the
ands are within the admissible evaluation limits. In results and merge these into a target result.
addition, a participation certificate has to be issued (3) The reference institutions must store the documenta-
for all examinations where participation in an exter- tion used in determining the target results for a period
nal quality assurance programme occurred. The cer- of at least 5 years starting from when they were used
tificate and participation certificate shall be sent to in the external quality assurance programmes.
the participants no later than 4 weeks before the next
external quality assurance programme date.
External quality assurance programme participants 3 A
ssessing the external quality assurance
are also to be informed of: programme results
(a) Target results and, where necessary, evaluation
limits of the external quality assurance pro- (1) Analysis is performed using the target results. The
gramme samples assessment criteria must be fulfilled for all samples.
(b) Position and distribution of the examination The external quality assurance programme organi-
results of all participants and the examination sations are to inform the participants of assessment
procedures used criteria.
(c) Number of participants, as appropriate, listed (2) If the entire population or method-dependent sub-
according to examination procedure. populations of the participants’ results show a consid-
The certificate is valid for double the interval listed in erable deviation to the target results, i.e., a deviation
Tables B3-2 and B3-2a. which influences the pass rate, the reference institu-
(8) If the reference institution establishes that partici- tions must search for the cause and, if possible, rectify
pants frequently are not receiving a certificate for this in co-operation with the affected manufacturer of
an examination with reagents or devices from a the external quality assurance programme sample,
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
66 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
the manufacturers of the respective test systems, or programme under routine conditions and using rou-
with other experts. They have to check whether, in tine examination procedures.
such a case, changing the target result would allow for (5) For every external quality assurance programme, the
a proper assessment. They decide whether the exter- reference institutions commission each participant
nal quality assurance programme is to be repeated for to examine at least two sets of external quality assur-
this examination. ance programme materials.
This process is to be substantiated and documented. (6) The reference institutions send the external qual-
The participants of the external quality assurance ity assurance programme materials to each exter-
programme and the Expert Group at the German nal quality assurance programme participant with
Medical Association are to be informed in accordance instructions on handling the material and conveying
with Part D3. their measurement results.
(7) Reference institutions shall only analyse measure-
ment results that have been submitted by the external
E4 Special requirements for external quality
quality assurance programme participant before the
assurance programme of ejaculate set deadline.
examinations (8) A certificate showing external quality assurance
programme submission date has to be issued to
1 Obligations of the reference institutions every external quality assurance programme giving
the information which of his results are within the
(1) Each of the reference institutions ensures that a admissible evaluation limits. In addition, a participa-
sufficient number of external quality assurance tion certificate has to be issued for all examinations
programmes are offered for all listed examinations, where participation in an external quality assurance
so that every medical laboratory can participate in programme occurred. The certificate and participa-
at least one external quality assurance programme tion certificate shall be sent to the participants no
per half year. They may only deviate from this if later than four weeks before the next external quality
they can prove that there is an unsufficient amount assurance programme date.
of suitable external quality assurance programme External quality assurance programme participants
samples. are also to be informed of:
(2) The external quality assurance programme is com- (a) The target results and evaluation limits of the
prised of an examination for sperm concentration, external quality assurance programme materials
sperm motility and sperm morphology. (b) Position and distribution of the measurement
(3) The reference institutions announce for 1 year in results of all participants and the examination
advance, the external quality assurance programmes procedures used
that they will conduct for the examinations in accord- (c) Number of participants, as appropriate, listed
ance with (2). This announcement includes: according to the measurement procedure.
(a) The deadline for registration for participation in The certificate is valid for 12 months.
the external quality assurance programmes (9) If the reference institution establishes that partici-
(b) The respective date of control material shipment pants frequently are not receiving a certificate for
and the deadline for sending back the results specific measurement procedure and if the cause for
(c) The examinations included in the external qual- this cannot be traced to the medical laboratory tak-
ity assurance programme, if necessary, with ing part in the external quality assurance programme
details on the measuring procedure or the reference institution, the appropriate federal
(d) The type of external quality assurance pro- authorities have to be notified if this can be defined
gramme material, the sample volumes of the liq- as an “incident” according to Section 2 of Germany’s
uid or reconstituted external quality assurance Medical Devices Safety Plan Ordinance.
programme materials. (10) Further details for conducting external quality assur-
(4) The reference institutions select the external quality ance programmes and analysing external quality
assurance programme materials and check their suit- assurance programme results are set forth in the
ability. The suitability of the selected external quality implementation regulations. These are published by
assurance programme samples needs to be checked the German Medical Association and by the reference
prior to their use in the external quality assurance institutions.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 67
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
68 Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations
(6) Reference institutions shall only analyse examination German Medical Association establishes and announces
results that were submitted by the external quality assur- the type of target values to be used for the examinations
ance programme participant before the set deadline. and how the target results have to be determined. Ref-
(7) A certificate showing the submission date of the erence methods have to be used whenever possible to
external quality assurance programme has to be determine target results in control samples.
issued to every external quality assurance pro- (2) The reference institutions establish the examination
gramme participant giving the information which plans for determining the target results of the external
of his results are within the admissible evaluation quality assurance programme samples, commission
limits. In addition, a participation certificate has the reference laboratories, analyse the results and
to be issued for all examinations where participa- merge these into a target result.
tion in an external quality assurance programme (3) The reference institutions must store the docu-
occurred. The certificate and participation certifi- mentation used in determining the target results
cate shall be sent to the participants no later than for a period of at least 5 years starting from when
four weeks before the next external quality assur- they were used in the external quality assurance
ance programme date. programmes.
External quality assurance programme participants
have also to be informed of:
(a) The target results and, if applicable, evaluation 3 A
ssessing the external quality assurance
limits of the external quality assurance pro- programme results
gramme samples
(b) Position and distribution of the examination (1) Assessment is based on the target results. The assess-
results of all participants and the examination ment criteria have to be fulfilled for all samples. The
procedures used external quality assurance programme organisations
(c) Number of participants, as appropriate, listed have to inform the participants of the assessment
according to examination procedure. criteria.
The certificate is valid for double the interval listed in (2) If the entire population or method-dependent sub-
Table B5-1 Column 7 and Table B5-2b. populations of the participants’ results show a consid-
(8) If the reference institution establishes that partici- erable deviation to the target result, i.e., a deviation
pants frequently are not receiving a certificate for a which influences the pass rate, the reference institu-
measurand with reagents or devices from a particu- tions must search for the cause and, if possible, rectify
lar manufacturer, and if the cause for this cannot be this in co-operation with the affected manufacturer
traced to the medical laboratories taking part in the of the external quality assurance programme sample
external quality assurance programme or the refer- or with experts. They have to check whether, in such
ence institution, the appropriate federal authorities a case, changing the target result would allow for a
have to be notified if this can be defined as an “inci- proper assessment. They decide whether the external
dent” according to Section 2 of Germany’s Medical quality assurance programme has to be repeated for
Devices Safety Plan Ordinance. this measurand.
(9) Further details for conducting external quality assur- This process has to be substantiated and documented.
ance programmes and analysing external quality The participants of the external quality assurance
assurance programme results are set forth in the programme and the Expert Group at the German
implementation regulations. These are published by Medical Association are to be informed in accordance
the German Medical Association and by the reference with Part D5.
institutions.
Unangemeldet
Heruntergeladen am | 16.03.15 15:47
Guideline of the German Medical Association on Quality Assurance in Medical Laboratory Examinations 69
Unangemeldet
Heruntergeladen am | 16.03.15 15:47