The new EXAxt 450 series has been designed to meet the
highest market requirements of today’s industry. This series of
analyzers is unique as it will no longer be the uncertainty factor of
your quality demands, but instead will bring you to a higher level.
The EXAxt will help you and guide you as the unique Human
Machine Interface (HMI) will not only present you reliable process
data, but diagnostics, trends, logbooks and step by step
calibration routines to become your friend in analytical
measurement and help you gaining that higher level of quality.
The instrument is easy to set up, very intuitive and can be used
without the need of an instruction manual. The HMI has a clear
menu structure in the language of your preference (English,
French, German, Spanish, or Italian). The touch screen interface
provides access to the display features. The main display gives
three process values at the same time.
Features
• Unique intuitive HMI menu structure in 6 languages.
• Process data trending up to 2 weeks.
• Predefined OIML standard solutions.
• Predefined matrices for precise temperature compensation.
• Two mA-outputs and four SPDT relay contacts with display
indicators.
• %weight on display.
• HART® Communications.
• FM Class 1, Div. 2, Group ABCD, T6 for Ta -20 to 55ºC
System Configuration
Sensors
Sensors Cables
Cables Fittings
Fittings Transmitters
Transmitters Accessories
Accessories
GS 12D8B5-E-E
2nd Edition
2
GS 12D8B5-E-E
3
The main screen displays: Soft alarm detected. Maintenance is recommended for
- The primary variable in large font (user selectable) best accuracy.
- Other process variable(s) in small font
- Unit symbols Hard alarm is detected indicating malfunction that is
- Tagnumber (user programmable) critical for good analysis. When this key is pressed
- Process description (user programmable) details are displayed with regards to detected
- Status of contact output(s) malfunction and troubleshooting guidelines are
- Status indicator during HOLD and WASH situation displayed to resolve the malfunction.
- Main function keys
Trend screen
Maximum
ISC 450
Live reading
120.0
SC
90.0
Minimum
60.0
Maintenance screen
30.0 The maintenance screen gives access to calibration,
109.3 µS/cm commissioning and setup of the instrument. These levels can be
protected by passwords.
T
Example:
Maximum
Average
Minimum
GS 12D8B5-E-E
4
Output and Alarm Functions Four SPDT relays are included as standard, and can be
configured by the user as conventional process alarms,
Output signal or in one of 2 control modes:
The standard ISC450 features two 4-20 mA current outputs
available for registration, and indication or control functions.
% controller output
The user selectable application can represent:
- the measured conductivity value 100 toff > 0.1 sec
- the concentration in wt%
- the measured temperature value Duty cycle
In addition the following output functions are available:
GS 12D8B5-E-E
5
Measurement Principle
V3 V1
Unlike 2- or 4-electrode conductivity systems, the EXA ISC450G Receive coil Drive coil
analyses the conductivity without any contact between
electrodes and process fluid. The measurement is based on
inductive coupling of 2 ring transformers (toroids) by the liquid.
The EXA ISC450 supplies a reference voltage (V1) at a high N3 N1
frequency to the ”drive coil”. The core of this coil is of a high
permeability magnetic material, and a strong magnetic field is
generated in the toroid.
The liquid passes through the hole in the toroid and can be
considered as a ”one turn” secondary winding.
The magnetic field induces a voltage (V2) in this liquid winding. I=V2 x Gliquid
The induced current thus made to flow is proportional to this
voltage and the conductance of the liquid ”one turn winding” is
according to Ohm’s law.
The conductance (G=1/R) is proportional to the specific Inductive conductivity measurement principle
conductivity and a constant factor that is determined by the
geometry of the sensor (length divided by surface area of the
hole in the toroid) and the installation of the sensor.
There are 2 toroids mounted in the doughnut shaped sensor.
The liquid also flows through the second toroid and therefore the
liquid turn can be considered as a primary winding of the second
ring transformer. The current in the liquid will create a magnetic
field in the second toroid. The induced voltage (V3) being the
result of this magnetic field can be measured as an output. The
output voltage of this ”receive coil” is therefore proportional to the
specific conductivity of the process liquid.
Functional Description
The EXA ISC450G is real time micro-controller operated
conductivity-analyzing system. It uses a dedicated micro-
controller to control all functions necessary in such a system.
The input and output functions are concentrated in the analog
section of the instrument. Even these functions are operated Grounding
through special interfaces designed to minimize interference with
the digital functions. All functions are executed separately.
GS 12D8B5-E-E
6
40 (1.57)
The ambient temperature and humidity of the installation
environment must be within the limits of the instrument wrench opening
specifications. YOKOGAWA
20 ( 0.79 )
wrench opening
ca 240 (9.45)
G3/4 32 ( 1.42 )
Mounting methods
The EXA ISC450G transmitter has universal mounting possibilities: Ø 40 ( 1.57)
- Panel mounting using optional brackets.
100 (3.94)
- Surface mounting on a plate (by bolts from the back).
- Wall mounting on a bracket (e.g. thick brick wall).
- Pipe mounting using a bracket on a horizontal or vertical pipe
(maximum diameter 50 mm).
1.25
Up to 50 meters of WF10 extension cable may be used with a
BA10 junction box. The installation factor of the ISC40 is the 1.20
ratio of the measured conductivity at the sensor and the specific 1.15
non conductive
conductivity of the solution. The unit is cm-1 just as the cell piping
1.10
constant of a contact electrode system. This factor is 1.88 cm-1 D
for the ISC40 if the sensor is installed with a minimum of 25 mm 1,05
(1 in) of process fluid surrounding the donut. 1.00
- Installed in an ISC40FF-S stainless steel flow cell,
0.95
the factor is 1.7 cm-1. conductive D in millimeters
piping
- Installed in an ISC40FF-P polypropylene flow cell, 0.90
the factor is 1.88 cm-1. 0 10 20 30 40 50
- The factor may be estimated from diagram (PEEK and PFA) for
actual installations not using the standard flow assemblies. Installation factor as a function of free distance around
- The I.F. or cellcontact of ISC40 sensors made of PFA is 3.0/cm. donut
GS 12D8B5-E-E
7
**If the tagnumber is predefined with the purchase, Yokogawa will inscript the tagplate with the specified tagnumber and program the
tagnumber in the transmitter.
Wiring Diagram
Connection Diagram
31
C
32 21
NC S1
33 Input contact
NO
41 22
C
42 NC
S2
Sensor Inputs
INPUTS
43 11
NO
+
Relay 51
C
Temp
- 12
Contacts 52 Receive 13
NC S3 coil
53 Shield 17
NO
71 14
C
Fail Safe
72 NO
S4
73 Drive coil 15
NC
Shield 16
GS 12D8B5-E-E
Dimension and mounting
wall mounting pipe mounting pipe mounting
(vertical) (horizontal)
80
(3.15")
2x ø6.5
(0.26")
200
(7.87")
4x ø10
(0.4")
141.5
(5.57")
70
(2.75")
min.195(7.75)
144(5.67")
138
138
24.5(1")
138(5.43)
27
(1.06")
M6
114.5(4.51")
138 M5
(5.43)
M6
Housing dimensions and layout of glands Option/UM. Universal mounting kit, panel mounting
diagram
YOKOGAWA HEADQUARTERS YOKOGAWA CORPORATION OF AMERICA Yokogawa has an extensive sales and
9-32, Nakacho 2-chome, 2 Dart Road
Musashinoshi Newnan GA 30265 distribution network.
Tokyo 180 United States
Japan Tel. (1)-770-253-7000 Please refer to the European website
Tel. (81)-422-52-5535 Fax (1)-770-251-2088
Fax (81)-422-55-1202 www.yokogawa.com/us (www.yokogawa.com/eu) to contact your
www.yokogawa.com
nearest representative.
GS 12D8B5-E-E